The present invention relates to a device comprising a piston sliding in a body like, for example, a syringe.
In most of commercial syringes, the piston sliding in the hollow barrel of a syringe body is made of a resilient material, such as rubber or thermoplastic elastomer, to absorb the irregularity in the shape of the syringe body. In order to allow the sliding and ensure that the syringe does not become leaky when pressure is applied, the sliding piston is coated with a silicone lubricant. One disadvantage of the use of silicone coated pistons is that that the silicone oils contaminate the content of the syringe body, e.g. a liquid medicament to be applied with the syringe.
In order to avoid such effect, laminated pistons were developed and disclosed in the prior art. Those pistons are of a silicone-free type in which it is not necessary to coat the sliding portion with a silicone oil layer as a lubricant.
In U.S. Pat. No. 6,090,081, pistons (rubber stoppers) are described which are capable of satisfying both the sealing property and slidable property without using silicone oils and having high sanitary and safety property. Such pistons are coated with a tetrafluoroethylene-ethylene copolymer resin as disclosed in Japanese Patent Laid-Open Publication No. 139668/1987, or with a polytetrafluoroethylene resin film, as disclosed in Japanese Patent Laid-Open Publication No. 97173/1988. The content of the said US patent and the Japanese Patent Publications are incorporated herein by reference.
Efforts were made to develop new materials for the body of the syringes which can be combined with the laminated pistons.
It is the merit of the present invention that it was surprisingly found that it is possible to combine certain pistons with certain coating with conventional hollow barrel bodies made of polypropylene in order to obtain a device fulfilling all functional, sanitary and regulatory requirements for use for medical purposes, like air and water tightness and required sliding forces.
In one embodiment of the invention, the surface of the resilient piston is coated with a laminated layer of polytetrafluoroethylene resin film, in another embodiment the surface of the resilient piston is coated with a laminated layer tetrafluoroethylene-ethylene resin film. The coating can be performed as described in U.S. Pat. No. 6,090,081 and the Japanese Patent Laid-Open Publication No. 139668/1987, or Japanese Patent Laid-Open Publication No. 97173/1988. In a further embodiment of the present invention, a tetrafluoroethylene polymer coated piston Flurotec commercially available from West Pharmaceutical-Daikyo as specified in Example 1 is used.
A hollow barrel polypropylene body according to the present invention may be any body made of polypropylene which is a hollow barrel intended for use in combination with a sliding piston, e.g. conventional syringe bodies or the like. It is within the ordinary skill of a worker in the field to be capable to combine a piston with a certain design with the appropriate hollow barrel in order to achieve the functional requirements such as air and water tightness and requiring a sliding force that fulfills the acceptance criteria of regulatory authorities.
Therefore, the invention concerns a device comprising a combination of a polypropylene body with a laminated piston defined above as possible embodiment of the invention.
The device exemplified in detail in the following example shall be another embodiment of the invention. However, the examples shall illustrate the invention and not be used to limit the scope of the teaching given herein.
In a TRICOS-device (see
The different components of one embodiment of the TRICOS-Fluoro-Device are described in detail in the following table:
The use of a tetrafluoroethylene polymer coated piston from Daikyo with the polypropylene syringe body, presents a lot of advantages for the development of the final product, as it does not require the use of silicone oil to facilitate the sliding of the piston inside of the syringe body. This is a tremendous advantage from a regulatory but also manufacturing point of view: easy to store, does not stick, inexpensive process and equipment, no transfer of the silicone oil to the granules of calcium phosphate.
“5 ml, 10 ml and 20 ml” tetrafluoroethylene polymer coated pistons (i.e pistons foreseen for conventional 5 ml, 10 ml, and 20 ml syringes, respectively) from Daikyo were successfully used with 3.5 ml, 7.0 ml and 17 ml TRICOS syringes (design History file: 001-DHF-NIV). Of course, it is understood that also pistons of other size could be used with the appropriate hollow barrel (TRICOS syringe) to produce a functional TRICOS-Fluoro-Device.
Tests were performed to evaluate if the tetrafluoroethylene polymer resin coated piston can fulfill the acceptance criteria of the standards applying for syringe like container made of polypropylene HD810MO.
These standards are applied for commercial syringe made of polypropylene with a piston that is siliconized.
This test challenges the ability of the syringe like container to resist to leakage and piston detachment from the plunger under negative pressure. This test is an attribute test based on the ISO 7886-1, Annex B of the norm.
A pass or fail determination was made based on a visual observation for replacing bubbles and piston detachment.
Protocol of test is described in EXAMPLE 3.
No leak at piston is accepted and no piston detachment is accepted.
The pressure may not increase during the 60's test under vacuum
75 non-sterile units and 75 sterile units were tested.
All units passed successfully the piston detachment test and no increase in pressure during the 60 seconds of vacuum was observed for any of the units tested. No piston leak was detected.
All tested units passed successfully test “Air leakage past piston during aspiration, and for separation of piston and plunger as per ISO 7886-1 (annex B of the norm)” and by that it can be stated with 95% confidence that there is less than 3.916% defective units.
This test challenges the ability of the piston to remain engaged with the plunger when exposed to a potential pull out force. A pass or fail determination was made.
Protocol of test is described in EXAMPLE 4.
No piston detachment from the plunger is accepted
75 non sterile units and 75 sterile units were tested.
All units passed successfully the test.
All tested units passed successfully “Piston pull-out test” and by that it can be stated with 95% confidence that there are less than 3.916% defective units as per test.
This test challenges the ability of the piston/plunger to remain inserted into the body syringe when exposed to a potential pull out force. The force needed to remove the piston/plunger from the body syringe was measured thanks to a tensile machine and the maximum pull out force has to be higher than 29 N (precision movement sustained male—DEF STAN 00-25—part 3) and it is preferable that the maximum pull force is higher than 59 N (precision movement momentary male—DEF STAN 00-25—part 3).
Protocol of test is described in EXAMPLE 5.
50 non-sterile units and 50 sterile units were tested. For both the sterile and the non-sterile units there were 3 units where the part of the plunger attached to the tensile machine broke before the plunger was removed. This means that the actual force needed to remove the plunger is above the value registered.
All tested units successfully passed the test, and it can be stated with 95% confidence that at least 99% of the units of an equal production, when tested according to test, will result in a peak force above 75.8 N for the non-sterile samples and above 97.8 N for the sterile samples.
The test challenges the ability of the syringe piston to resist leakage under axial pressure. This test is based on the ISO 7886-1.
A pass or fail determination was made.
Protocol of test is described in EXAMPLE 6.
No leak is accepted
75 non sterile units and 75 sterile units were tested.
No leak was detected for any of the units tested.
All test units passed successfully test “Liquid leakage at syringe piston under compression” and by that it can be stated with 95% confidence that there is less than 3.916% defective units.
The test purpose is to measure the force, which is required to initiate the movement of the plunger inside of the syringe body. This test is based on the ISO 7886-1:1993 annex G
In ISO 7886-1:1993 annex G there is no strict requirement on the force required to initiate the movement of the plunger, but a proposed value of <25 N is given.
It is known in the art that a piston cannot slide into the syringe body without coating with silicone oil.
Protocol of test is described in EXAMPLE 7.
50 sterile units were tested.
The force needed to initiate the movement of the plunger is below the proposed limit of 25 N for all units tested and it can be stated with 95% confidence that at least 99% of the units of an equal production when tested according to the above test, will result in an initial force to move the plunger below 25.9 N. These results are acceptable since there is no difficulty to move the piston at the forces obtained in this study
The tests are performed before sterilization and after beta sterilization at a dose of 50 kGy onto the overall dimensions of the TRICOS devices.
This test is important to show that the tetrafluoroethylene polymer resin coated piston from Daikyo keeps its dimensions after sterilization and therefore the its functionality when mounted in the TRICOS device as shown in tests 1 to 4
Traceability:
Description:
25 pistons were tested as received by the West supplier, while 25 other pistons were packed into an HDPE overpouch and sent to Ionisos for beta sterilization at 50 kGy before dimensional test.
Test description: visual inspection with a calibrated caliper
Performance: All 50 pistons were inspected. The 25 pistons for sterilization were inspected both before and after sterilization.
Result: No defective units were observed.
Performance: 25 sterile and 25 non-sterile pistons were measured as per the attached blueprint. A letter as indicated on the blueprint identified each dimension.
All piston measurements performed were within the limits.
No significant differences between sterile and non-sterile units were observed.
The piston keeps its characteristics after irradiation at a dose of 50 kGy.
Design History File: 001-DHF-NIV
| Filing Document | Filing Date | Country | Kind | 371c Date |
|---|---|---|---|---|
| PCT/EP2006/004494 | 5/12/2006 | WO | 00 | 7/2/2008 |
| Number | Date | Country | |
|---|---|---|---|
| 60685830 | May 2005 | US |