This invention is related to a method and device used for measuring Eustachian dysfunction level of patients.
The Eustachian tube is a hollow tube that originates in the back of the nose and connects the nasal cavity to middle ear space. The middle ear space is the hollowed out portion of the skull bone that contains the hearing mechanism, which is covered by eardrum on one side and cochlea on the other side. The Eustachian Tube provides ventilation, drainage and protection of middle ear space against reflux, microorganisms, excessive sound pressure and pressure changes in pharynx. In adults, the Eustachian tube is approximately 35 mm long and approximately 3 mm in diameter. The first part of the Eustachian tube is supported by cartilage and the last part that is close to the middle ear space is located inside the bone structure. The lining tissue of the Eustachian tube is similar to the tissue that covers the nasal cavity and responds to external stimulants just as nasal tissue does.
The main function of the Eustachian tube is to provide ventilation to the middle ear space, ensuring that its pressure remains at slightly negative but close to ambient air pressure. Another function of the Eustachian tube is to drain secretions and debris from the middle ear space. Several small muscles located in the back of the throat and palate control the opening and closing action of the tube. Typical muscular actions like swallowing and yawning cause contractions of these muscles and activate Eustachian tube function time to time to achieve its pressure regulation function.
Eustachian tube is normally closed most of the time to prevent contaminants contained in the nasal cavity to reach the middle ear cavity. Disorders of Eustachian Tube are either “dilatory” type (failure to open adequately) or “patulous” type (failure to maintain tube closed at rest). If dysfunction of Eustachian tube makes the tube always open, this condition is called “patulous” type dysfunction of the Eustachian tube. Patients with “patulous” dysfunctional Eustachian tube suffer from frequent, chronic ear infections. A more common form of the Eustachian tube dysfunction is caused by partial or complete blockage of the Eustachian tube, which causes sensations of popping, clicking, and ear fullness as well as moderate to severe ear pain associated with the condition. This type of dysfunction is the “dilatory” type dysfunction of the Eustachian Tube.
Eustachian tube dysfunction (ETD) is medically defined as “the inability of the Eustachian tube to perform its required functions”, e.g. opening up regularly to pressure equalize the middle ear cavity. A recently published book titled “Interventions for adult Eustachian tube dysfunction: a systematic review” by Llewellyn A, Norman G, Harden M, et al. Health Technology Assessment, No. 18.46, Southampton (UK): MEM Journals Library, July 2014 can be considered as an authoritative book summarizing current practices and technologies used in Eustachian tube function related matters. Excerpts from this reference book will be used for establishing facts related to Eustachian tube assessment.
Diagnosis of Eustachian tube dysfunction (ETD) is not easy. According to the reference book mentioned above;
As it is mentioned above, there is no consensus on diagnosis or assessment of ETD. Most common practice used in identifying ETD is through tympanometry measurement. According to the reference mentioned above; “The UK national survey defined ETD as the presence of a normal or abnormal but intact tympanic membrane with a middle ear pressure of <−100 mmH2O and an air-bone gap of ≥15 decibels (dB)”. Tympanometry tests can only be performed on patients with healthy tympanic membrane. Since most chronic ETD patients have ruptured tympanic membrane due to their condition, tympanometry cannot be used on such patients.
Recent clinical study done by inventors revealed an interesting finding about ETD. The study revealed that individuals with healthy Eustachian tubes receive periodic neural signals to their Eustachian tube muscles. The signals are highly periodic and come with a specific pattern and phase difference between them. Clinical study revealed that patients with ETD problem have indeed problems with neural signals received by the Eustachian tube muscles. The problem appears either with the periodicity of the signals, or the latency between the signals or the amplitude of the signals. Based on the guidelines of this clinical study, a new type of ETD assessment device is designed which assesses the ETD level of the patients according to the signal pattern and functionality of ET muscles.
The invention presented in this document aims to assess condition of ET dysfunction in patients regardless of the condition of their tympanic membrane. It further aims to identify patients who have ET dysfunction due to neurological disorder of the muscles controlling Eustachian Tube.
The basis of the invention starts from a discovery made by inventors during a clinical investigation of the Eustachian Tube muscle behavior. Medical community so far believed that Eustachian tube opens sporadically only during yawning or swallowing action. During clinical trials inventors noticed that Eustachian tube muscles receive neurological signals periodically almost every 20 seconds and activate periodically. The periodicity of the signals resembles almost a heartbeat signal albeit slower. Due to intricate anatomy of the Eustachian tube valve, the muscles need to work synergistically in order to open the valve. One of the muscles needs to be activated first, followed by the other muscle after a precise delay in order to open the Eustachian tube. Any discoordination of the muscle activity causes muscles to contradict to each other, which in turn results in dysfunctional Eustachian tube. After studying dozens of patients, it appeared that the cause of Eustachian Dysfunction in some patients is due to lack of synchronization of neural signals received by the ET muscles. It also appeared that by applying a certain signal pattern to Eustachian Tube muscles through the palate of the patient by subdermal electrodes, it is possible to open the Eustachian tube. Most researchers and medical practitioners so far tried to access Eustachian tube muscles nasally through nose. Nasal area is very sensitive and application of subdermal electrodes in this area is quite painful for most patients. So most researchers use anesthetics to alleviate the pain while recoding muscle activity. Since anesthetics interfere with nerve activity, real nature of the signals so far has never been revealed. Inventors are the first researchers who accessed ET muscles through the palate without using anesthetics and recorded the signals. Analyses of the signals revealed important synergistic and periodic nature of the neural signals.
Eustachian tube is not a simple tube, but a delicate valve controlled by two muscles. These two tubal muscles are the Levator Veli Palatini Muscle (mLVP) and the Tensor Veli Palatini muscle (mTVP). Among the two, Tensor Veli Palatini (mTVP) is the main tubal dilator which performs the Eustachian tube function. However, coordinated synergistic action of mLVP and mTVP together performs an efficient opening action of Eustachian Tube. Any disturbance of the synergy between these two muscles may make them work antagonist to each other which result in ET dysfunction.
Making use of these discoveries, the newly invented diagnostic device tests Eustachian Tube and assesses ETD condition of the patient by doing the following tests. Test steps may be executed fully or partially to assess the ETDA (Eustachian Tube Dysfunction Assessment) condition of the patient.
mLVP and mTVP muscles of the patient are stimulated electrically or optically by using probes placed to the palate of the patient while monitoring the opening of Eustachian tube through sensors placed into external ear canal. In case of electrical stimulation, the probes are subdermal electrodes, in case of optical stimulation they are fiber optic cables or light emitting diode (led) placed at the tip of the probe. While administering subdermal electrodes the patient is not anesthetized, however does not feel any pain at the particular application location due to insensitive nature of the application area and miniscule size of subdermal electrodes. An alternative way of stimulating ET muscles is through optical stimulation. Optical stimulation is a new technique for stimulating muscles and very suitable for this particular application. Optical stimulation pulses are applied through fiber optic cables or by photodiodes installed on distal end of a probe assembly. In both cases optical energy is directed to the same location transpalatally where subdermal electrodes were applied.
Eustachian tube dysfunction assessment is done as follows:
As a result of these tests, it will be possible to identify if the patient has ETD condition due to neurological disorder or due to some other reason. The invention provides clinical evidence about the condition of the ET muscles for the health professional to decide. Having Eustachian tube open with only mTVP stimulation, or only with mLVP stimulation, or with both mTVP and mLVP stimulation together tell information about the state of Eustachian tube dysfunction. Those patients whose ET opens due to stimulation of mTVP, mLVP are a category of patients who are likely to be treated by correcting this neurological disorder. There is a branch of medical practice called “neuromodulation therapy” which aims to correct this sort of disorders. The purpose of the invention is to assess the ETD condition of patients and identify those who may get benefit from neurological treatment.
The operation of the present invention will now be described with the aid of the figures. The purpose of the invention is to diagnose the Eustachian tube dysfunction cases and identify the ones that are suffering because of neurological disorder of Eustachian tube (ET) muscles. As it is explained in the background section of this document, so far Eustachian tube dysfunction (ETD) ailment has been an ill-defined condition with little knowledge about the underlying cause. The invention is intended to diagnose ETD condition in patients and identify the category of patients who have healthy ET muscles and ET valve mechanism, but suffering from ETD because of disorder of neurological signals received by the muscles. Identification of these patients is important because these patients may benefit from neurological therapy. The invention does this by stimulating Eustachian tube muscles of the patients while monitoring pressure in the external ear canal. Stimulation is done from an easily accessible and identifiable location without causing excessive pain or discomfort to patients.
The amplitude, periodicity and phase difference between the two signals received from locations 10 and 12 tell important information about the health of Eustachian tube activity.
In another embodiment of the system reading signals (210), and displaying signals (220) may be omitted and operation of the system may start from block 230 directly.
In the next phase of the assessment, pressure generator is started (230) and maximum of 50 daPa pressure is applied to nasal cavity of the patient through nasal pressure sensor pathway. Pressurization of nasal cavity continues until pressure reaches 50 daPa (240). 50 daPa pressure level is considered “mild” pressure level which is tolerable by most patients. In another embodiment of the system, the pressure level may be adjustable for those patients who may find 50 daPa uncomfortably high.
As a next step, the stimulation pulses are applied to the patient in block 250. Stimulation is given either electrically or optically to mLVP and mTVP muscles. As a first step, only mTVP muscle is stimulated with electrical signal of 0.2 mV amplitude pulse with 0.020 msec. duration. During the application of the pulse, external ear canal sensor output is monitored. Opening of ET is seen as a distinct change in the output of external ear canal sensor output with at least 10 daPa pressure change (270). If the pressure change in not observed, the stimulation step is repeated after changing parameters (280). During this step, both mLVP and mTVP are stimulated by 0.2 mV pulses while mLVP receives longer duration pulse which lasts 0.040 msec. while mTVP receives 0.020 msec. duration pulse. The mLVP muscle stimulation pulse is applied 0.020 msec. before the mTVP pulse is applied. During the stimulation, the external ear canal sensor is monitored again for at least 10 daPa pressure change. If no pressure change is observed, the signal amplitude is increased to 0.3 mV and stimulation is repeated. Although these are the recommended parameters based on the clinical trials, in another embodiment of the invention, the pulse amplitude and pulse duration is made adjustable for finding the best pulse pattern for the patients. Block 260 shows the step where the test results are displayed.
The ETD assessment of the patient is based on the results of the aforementioned test results.
Medical interpretations of these types are beyond the scope of this document but they broadly define the ETD assessment level.
Number | Date | Country | Kind |
---|---|---|---|
2016/03458 | Mar 2016 | TR | national |
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/IB2017/050680 | 2/8/2017 | WO | 00 |
Number | Date | Country | |
---|---|---|---|
Parent | 15018874 | Feb 2016 | US |
Child | 16076821 | US |