This invention relates generally to the field of intraocular lenses (IOL) and, more particularly, to accommodative IOLs.
The human eye in its simplest terms functions to provide vision by transmitting light through a clear outer portion called the cornea, and focusing the image by way of a crystalline lens onto a retina. The quality of the focused image depends on many factors including the size and shape of the eye, and the transparency of the cornea and the lens.
When age or disease causes the lens to become less transparent, vision deteriorates because of the diminished light which can be transmitted to the retina. This deficiency in the lens of the eye is medically known as a cataract. An accepted treatment for this condition is surgical removal of the lens and replacement of the lens function by an artificial intraocular lens (IOL).
In the United States, the majority of cataractous lenses are removed by a surgical technique called phacoemulsification. During this procedure, an opening is made in the anterior capsule and a thin phacoemulsification cutting tip is inserted into the diseased lens and vibrated ultrasonically. The vibrating cutting tip liquifies or emulsifies the lens so that the lens may be aspirated out of the eye. The diseased lens, once removed, is replaced by an artificial lens.
In the natural lens, bifocality of distance and near vision is provided by a mechanism known as accommodation. The natural lens, early in life, is soft and contained within the capsular bag. The bag is suspended from the ciliary muscle by the zonules. Relaxation of the ciliary muscle tightens the zonules, and stretches the capsular bag. As a result, the natural lens tends to flatten. Tightening of the ciliary muscle relaxes the tension on the zonules, allowing the capsular bag and the natural lens to assume a more rounded shape. In the way, the natural lens can be focus alternatively on near and far objects.
As the lens ages, it becomes harder and is less able to change shape in reaction to the tightening of the ciliary muscle. This makes it harder for the lens to focus on near objects, a medical condition known as presbyopia. Presbyopia affects nearly all adults over the age of 45 or 50.
Prior to the present invention, when a cataract or other disease required the removal of the natural lens and replacement with an artificial IOL, the IOL was a monofocal lens, requiring that the patient use a pair of spectacles or contact lenses for near vision. Advanced Medical Optics has been selling a bifocal IOL, the Array lens, for several years, but due to quality of issues, this lens has not been widely accepted.
Several designs for accommodative IOLs are being studied. For example, several designs manufactured by C&C Vision are currently undergoing clinical trials. See U.S. Pat. Nos. 6,197,059, 5,674,282, 5,496,366 and 5,476,514 (Cumming), the entire contents of which being incorporated herein by reference. The lens described in these patents is a single optic lens having flexible haptics that allows the optic to move forward and backward in reaction to movement of the ciliary muscle. A similar designs are described in U.S. Pat. No. 6,302,911 B1 (Hanna), U.S. Pat. Nos. 6,261,321 B1 and 6,241,777 B1 (both to Kellan), the entire contents of which being incorporated herein by reference. The amount of movement of the optic in these single-lens systems, however, may be insufficient to allow for a useful range of accommodation. In addition, as described in U.S. Pat. Nos. 6,197,059, 5,674,282, 5,496,366 and 5,476,514, the eye must be paralyzed for one to two weeks in order for capsular fibrosis to entrap the lens that thereby provide for a rigid association between the lens and the capsular bag. In addition, the commercial models of these lenses are made from a hydrogel or silicone material. Such materials are not inherently resistive to the formation of posterior capsule opacification (“PCO”). The only treatment for PCO is a capsulotomy using a Nd:YAG laser that vaporizes a portion of the posterior capsule. Such destruction of the posterior capsule may destroy the mechanism of accommodation of these lenses.
There have been some attempts to make a two-optic accommodative lens system. For example, U.S. Pat. No. 5,275,623 (Sarfarazi), WIPO Publication No. 00/66037 (Glick, et al.) and WO 01/34067 A1 (Bandhauer, et al), the entire contents of which being incorporated herein by reference, all disclose a two-optic lens system with one optic having a positive power and the other optic having a negative power. The optics are connected by a hinge mechanism that reacts to movement of the ciliary muscle to move the optics closer together or further apart, thereby providing accommodation. In order to provide this “zoom lens” effect, movement of the ciliary muscle must be adequately transmitted to the lens system through the capsular bag, and none of these references disclose a mechanism for ensuring that there is a tight connection between the capsular bag and the lens system. In addition, none of these lenses designs have addressed the problem with PCO noted above.
Prior art accommodative two lens systems using a movable “zoom” lens have inherently limited movement. The maximum sensitivity or movement magnification a (a unitless ratio) is defined as the axial movement of the lens per unit zonule movement and is derived by the following equation:
α=−B/A
where B is the projected distance of the zonule length which is in the order of 1.0 to 2.0 mm; and
A is the axial distance between the middle plane between the dual lens and the anterior surface of the anterior lens where the zonules terminate.
Practically speaking, because of the lens thickness and dual lens separation requirement, A cannot be less than ˜1 mm. Therefore, α cannot be larger than 2, which defines the limit of the known dual lens accommodative approaches. This limit is too low for the dual optics design to achieve the objective of creating the greater than 2.25 diopters of accommodative amplitude that patients need for normal accommodation, which ideally results in a greater than or equal to 4.
Secondly, existing dual optics accommodative implants do not manage any necessary change in the base power of the dual optics lens systems. Such changes can result from the inaccuracy of biometry, surgical variations, implant variations and inter-patient capsule variations. Consequently, patients can have refractive error after the implantation and need additional spectacles corrections that are not desired. In addition, potential post implantation capsule reaction and other ocular changes over time can result in the gradual development of refractive errors over time.
Therefore, a need continues to exist for a safe and stable accommodative intraocular lens that provides accommodation over a broad and useful range and an adjustable base power.
The present invention improves upon the prior art by providing a two optic accommodative lens system. The present invention also contemplates the use of a cam mechanism to adjust the distance power via adjustment of the dual lens separation when the eye is at distance vision stasis. The cam mechanism allows for distance/base power adjustment as needed.
Accordingly, one objective of the present invention is to provide a safe and biocompatible intraocular lens system.
Another objective of the present invention is to provide a safe and biocompatible intraocular lens system that is easily implanted in the posterior chamber.
Still another objective of the present invention is to provide a safe and biocompatible intraocular lens system that is stable in the posterior chamber.
Still another objective of the present invention is to provide a safe and biocompatible accommodative lens system.
Still another objective of the present invention is to provide a safe and biocompatible accommodative lens system having an adjustable base power.
These and other advantages and objectives of the present invention will become apparent from the detailed description and claims that follow.
As best seen in
As best seen in
In order to provide power adjustability to lens assembly 500, as best seen in
This description is given for purposes of illustration and explanation. It will be apparent to those skilled in the relevant art that changes and modifications may be made to the invention described above without departing from its scope or spirit.
Number | Name | Date | Kind |
---|---|---|---|
5275623 | Sarfarazi | Jan 1994 | A |
5476514 | Cumming | Dec 1995 | A |
5496366 | Cumming | Mar 1996 | A |
5674282 | Cumming | Oct 1997 | A |
6197059 | Cumming | Mar 2001 | B1 |
6241777 | Kellan | Jun 2001 | B1 |
6261321 | Kellan | Jul 2001 | B1 |
6302911 | Hanna | Oct 2001 | B1 |
6616691 | Tran | Sep 2003 | B1 |
6695881 | Peng et al. | Feb 2004 | B2 |
6926736 | Peng et al. | Aug 2005 | B2 |
20020107568 | Zadno-Azizi et al. | Aug 2002 | A1 |
20040039446 | McNicholas | Feb 2004 | A1 |
Number | Date | Country |
---|---|---|
WO 0134067 | Nov 1999 | WO |
WO 0066037 | Nov 2000 | WO |
WO 03059196 | Jul 2003 | WO |
WO 03059208 | Jul 2003 | WO |
Number | Date | Country | |
---|---|---|---|
20070050024 A1 | Mar 2007 | US |