This invention relates to methods and systems for the processing of aerosols from solvents of low vapor pressure (such as water or alcohol) containing a solute or suspension, and the delivery of solute or suspension as a solid respirable aerosol into the respiratory system. More particularly, it pertains to the delivery of aqueous-based aerosols containing large molecule therapeutic materials such as drugs, biologics, proteins, surfactants and genetic materials to the pulmonary system as respirable dry particle aerosols at a high dose rate.
The characteristics of aerosol inhalation patterns for deposition in the respiratory tract, as well as their clearance from the respiratory tract, have been well documented (Yeates, D. B. and Mortensen J. Deposition and Clearance, In: Murray J F. Nadel J A: Textbook of Respiratory Medicine, et 3. Philadelphia, WB Saunders Company, Vol. 1. Chapter 15 pp. 349-384, 2000).
Reduction in Aerosol Inhalation Times
Aerosol drug therapies from liquid solutions, liquid suspensions, and dry powder suspensions have been used in hospitals, outpatient clinics, and at home. Of the new drugs being developed, it has been estimated that 16% will be delivered via the respiratory tract. This includes the treatment of both respiratory and non-respiratory diseases. These new agents include antibiotics, anticancer agents, surfactants, hormones, proteins and peptides. Whereas there are excellent nebulizers for the delivery of microgram quantities of many small molecules, efficient and effective aerosol delivery of many of the above agents presents new challenges for aerosol delivery systems. A result of the limitations of current technology is that, in order to provide a therapeutic level of drug to the pulmonary system, an extended time period for inhalation is required. The necessity, in some cases, to deliver milligram masses to the lungs rather than microgram masses has led to inhalation treatment times of up to 2-3 hours per day. This is to the detriment of patient comfort and, possibly, compliance. For instance, the Respimat™ metered-dose inhaler, is a small device in which two pressurized liquid jets collide to produce a fine aerosol with 11-15 μl being dispensed at each actuation (breath). However, it will only nebulize liquids, not suspensions. The dose per breath is very limited and the aerosol will readily change in size due to evaporation. There are numerous jet-type nebulizers available with perhaps the ePARI, PARI LC Star™ and the Aerotech II™ representing the best of these. These nebulizers have fluid flow rates of up to 0.89, 0.62 ml/min and 0.39 ml/min, respectively, with the PARI LC and Aerotech II™ respectively, having 80% and 77%, particles being less than 5 μm in diameter. Typically, jet type nebulizers produce aerosols with geometric standard deviations of about 1.8-2.3. The polydispersity does not make them ideal for the delivery of drugs to the lower respiratory tract; however they are relatively inexpensive. The markedly higher dose rate (4-5 ml/min) provided by the use of the APIS will result in reduced treatment times.
Reduction in Shear Degradation
Some polymeric molecules and biologics are sensitive to shear degradation and consequent loss of desired activity. Shear degradation can be minimized by the generation of droplets which are too large to penetrate into the respiratory tract.
Delivery of Sparsely Soluble Agents
Some agents are sparsely soluble. In these cases, both large volumes of fluid and long inhalation times are required to deliver an effective dose of the active agent. The controlling (critical) parameters are a) the liquid flow rate b) the total input of unsaturated air and c) the output particle size. The initial size of the aqueous/solvent-based aerosol generated is dependent on the mass of the agent in solution or suspension such that on complete evaporation of the solvent, the residual particles have aerodynamic diameters between 1 and 7 μm. For example, a sparsely soluble compound will require the generation of very large droplets whereas very soluble compounds such as NaCl or sugar require the generation of smaller droplets. To optimize the rate of drug delivery the aqueous solvent flow rate should be 1-5 ml/min. Thus, aerosol generators should be chosen to generate the optimal size droplets such that on evaporation a solid phase aerosol of the desired respirable size, typically between 1 and 7 μm aerodynamic diameter results.
Alternate Method for the Generation and Delivery of Dry Powder Aerosols
Suspension of powders to form respiratory aerosols is difficult and sometimes impractical due to the surface forces between the molecules and agglomeration of the composite particles. Excipients are used to help facilitate aerosolization but these reduce the drug load per particle. Dry powder inhalers can require relatively rapid inhalation rates of 1 l/sec to disperse the powder, resulting in bronchial rather than deep lung aerosol deposition. Due to formulation issues, the drug and the inhaler are often designed to work together. There are several hand held devices available, including Rotahaler™, Turbuhaler™ and Diskhaler™. The Nektar™ dry power inhaler uses an independent power source to disperse the powder from a “blister”. The powder contains drug particles (<5 μm MMAD), lactose or glucose particles (>30 μm diameter) or micronized particles. Typical doses delivered range from 4 to 450 μg with the Nektar product, providing 2-5 mg of solids per puff.
Compact Design
Large devices have obvious disadvantages in use. For example, an evaporator/concentrator previously described, (Pillai, R. S., Yeates, D. B., Eljamal, M., Miller, I. F. and Hickey, A. J. Generation of Concentrated Aerosols for Inhalation Studies. J. Aerosol Sci., 25(1):187-197, 1994.) had a volume of 200 l and was 5 ft long and 1 ft in diameter.
Inhalation Regulated Aerosol Delivery—Respiratory Aerosol Control System, RACS
Manually operated and breath-activated metered dose inhalers are the most commonly used devices for aerosol administration of medications. In a manual metered dose inhaler the drug delivery is manually activated by the patient. This requires the patient to have good coordination skill to operate these devices for efficient drug delivery. It is estimated that more than half of patients are unable to use the device properly and efficiently. Major problems with a manual metered dose inhaler include timing coordination between activation of drug delivery and inspiration of aerosol medication, multiple activation of drug delivery during inspiration, improper breath-holding, and operation difficulty with insufficient hand strength (young child, elderly or seriously ill). Another problem with metered dose inhalers and dry powder inhalers is that the inspiratory effort required to activate drug release results in a high inspiratory flow that causes excess aerosol deposition in the oral cavity and larynx. All these limitations make these manual metered dose inhalers sub-optimal for delivering aerosol medication.
Delivery of aerosolized agents has generally been limited to a given mass of fluid that is aerosolized at the beginning of each breath. When the mass aerosolized is independent of the size and depth of breathing, optimal use of the patient's breathing pattern is not utilized to achieve maximal delivery of the drug. Some devices are either operator-activated or activated by the flow caused by the initiation of the breath. In these devices a set dose is delivered independent of the size of the breath. In many situations the drug solution is placed in the reservoir of a nebulizer and the patient is instructed to inhale the medication until the medication has been completely aerosolized. The mass of drug leaving the nebulizer and that deposited within the desired region of the respiratory tract can vary greatly depending on the technique. These methods do not provide a dose which is dependent on breath volume. The inhaled volume is, in part, determined by the size of the patient.
To provide a better inhaler for aerosol delivery of medication, a variety of delivery systems and methods have been attempted and are the subject of U.S. patents. The major focus of these patents has been the provision of a breath-activated apparatus for the timing of the actuation of a metered dose inhaler, MDI and the assessment of inspiratory flow using a) measurements of the pressure drop across a resistive element b) a Venturi flow meter or c) the negative pressure caused by an inspiration. This signal has been used to regulate the valve on an MDI and provide flow and volume information to the user. These patents include:
Nichols, et al (U.S. Pat. No. 6,491,233; U.S. Pat. No. 6,854,461) discloses an aerosol generator and breath-activated methods of delivering an aerosol, in which the aerosol is generated by heating a medicated fluid as it flows through a capillary tube. It utilizes a pressure drop to trigger delivery of a given dose of the aerosolized agent at the beginning of a breath. Once the pressure drop is detected, the aerosol can be delivered to the user. However, this system and method makes no provision for concentrating the generated aerosol particles, as does the present invention. It also makes no provision for an outlet for the patient's exhalation and a constant air supply, so that the patient must disconnect his/her mouth from the mouthpiece for the next inhalation. It is unsuitable for operation as a respiratory control delivery device in conjunction with APIS which has a constant air supply flowing through the concentrator and a positive pressure at the output.
Cox, et al (U.S. Pat. No. 6,516,796; U.S. Pat. No. 6,557,552) discloses an aerosol generator and methods for using it. The generator comprises a heated flow passage, a source of material to be volatilized, a valve to regulate material flow, and a pressurization arrangement to cause material to flow. However, this method makes no provision for concentrating the generated aerosol, as does the present invention.
Poole (U.S. Pat. No. 6,158,431) discloses a portable system and method for delivering therapeutic material to the pulmonary system, comprising a droplet dispersion chamber, a droplet generating assembly, an assembly for heating and evaporating the droplets, and a delivery system. However, this system and method makes no provision for producing monodisperse aerosol particles in the optimum 1-7 μm diameter size range, nor does it make any provision for concentrating the generated aerosol particles, as does the present invention.
Lloyd et al. (U.S. Pat. No. 5,469,750) and Goodman et al. (U.S. Pat. No. 5,813,397) disclose a breath-activated microcontroller-based apparatus for delivery of aerosol medication for inspiration from a metered dose. However, this apparatus makes no provision for providing an outlet for the patients' exhalation and a constant air supply flowing (such that patients have to disconnect their mouth from the mouth piece for the next inhalation). The present invention enables continuous breathing without the necessity to disconnect from the device.
In all of the above-reported systems, the primary focus has been the development of hand-held devices which deliver doses of up to a few micrograms of active drug per breath. None of the systems and methods described in U.S. patents or in the literature for commercially available products deliver aqueous-based respirable aerosols at high dose rates of over a milligram of active agent per breath as a stable solid phase aerosol. The technology and methods that are the subject of this invention are particularly suitable for delivery of aqueous-based aerosols containing large molecules, genetic material and other therapeutic agents to the pulmonary system.
Thus, there is a need for a new aerosol processing and inhalation system that is suitable for delivery of aqueous-based aerosols containing large molecules, genetic materials and other therapeutic agents to the pulmonary system at a high dose rate, and in a manner that is both therapeutically effective and comfortable for the patient. The present invention provides a method and system for generating 10-30 μm aqueous or nonaqueous droplets, evaporating solvent from the droplets, and concentrating the aerosols to produce 1-7 μm aerodynamic diameter dry particles. By removing most of the carrier air and the unwanted vapor, respirable aerosols are delivered at a high dose rate.
APIS is designed to deliver aqueous- or nonaqueous-based aerosols at a high dose rate. In a preferred configuration, a large aqueous aerosol with droplets 10-30 μm in diameter is generated using one or more jet-type aerosol generators, enabling the generation of large molecules and biologics while minimizing the risk of shear-induced degradation. Other aerosol generators can be used. The aerosol jet is arrested by virtual impaction with a warm coaxial counter flow air jet in the opposite direction, and is rapidly evaporated with a warm swirling annular air flow within the evaporation chamber. Aerosol disposition on the chamber is minimized by the use of sheath air. The evaporation of an aqueous solvent can be further enhanced by connective heat transfer, and when necessary, the use of radiative heat transfer such as an infrared source with or without infrared reflectors to enhance the effect. The resulting residual dry particles with mass median aerodynamic diameter (MMAD) of about 3 μm are concentrated using a passive virtual impactor, PVI. Most of the dilution air and the unwanted vapor are removed through the exhaust ports of the PVI. The resulting small fraction of the air passing through the virtual impaction plates carries most of the dry particles and flows toward the output.
APIS results in stable dry particles containing up to 100% active agents.
Aerosol production can be regulated during inhalation using a respiratory aerosol control system and providing flow derived visible indicators.
For a better understanding of the present invention, reference is made to the accompanying drawings, which are incorporated herein by reference.
The present invention relates to a method and system for aerosol processing and the inhalation control for generation and evaporation of agents dissolved or suspended in a fluid with a vapor pressure similar to or less than that of water, and their subsequent concentration and controlled delivery as solid phase respirable aerosols. To perform these tasks, the invention generates relatively large droplets (10-30 μm) with a relatively narrow size distribution at fluid flow rates up to 5 ml per minute. The liquid is evaporated from the droplets, and the resulting aerosol, containing ˜3 μm mass median aerodynamic diameter particles, is concentrated by a passive virtual impactor aerosol concentrator to provide a concentrated respirable aerosol. The mechanism to control the delivery of aerosol to a patient throughout each inspiratory cycle at a self-regulated flow rate is contained within the method and system. The method and system is able to provide at least 1 mg/breath of therapeutic aerosol with a particle mass median aerodynamic diameter (MMAD) of approximately 3 μm. This system, when used as recommended by the practitioner, maximizes the mass of aerosol deposition and provides improved reproducibility of the generated aerosol deposited within the subject's respiratory tract within a much shorter period of drug administration than is possible with other technologies.
The Aerosol Processing and Inhalation System, APIS and its clinical counterpart SUPRAER™, described herein is suitable for the evaporation of aqueous and/or nonaqueous solvents of low vapor pressure, such as water or alcohol from solutions or suspensions, and the concentration of the resulting 1-7 μm aerodynamic diameter residual solid particles and their delivery to the respiratory tract at inspiratory flow rates of 15-30 l/min. Aqueous liquid flow rates are typically between 1 and 5 ml/min with input air flow rates totaling 200 to 300 l/min. It is recognized that aerosol generation and processing systems are often specific to the active agent/solvent combination and to the properties of these, both individually and in concert. While the APIS does not provide a panacea for all aerosol delivery applications, it does enable many unresolved issues related to the delivery of aerosolized drugs and biologics to be overcome. These issues and their resolution by the invention described herein are addressed below.
The APIS reduces the size of these large droplets while minimizing shear degradation to provide solid aerosols within the respiratory range. Aerosol generation of small particles requires more energy and higher shear forces than is required for the generation of larger particles. The use of the APIS in conjunction with a large aerosol generator, such as those demonstrated in the Examples herein, enables the generation of large molecules and biologics while minimizing the risk of their shear-induced degradation. It provides, following the evaporation of the fluid carrier, a respirable aerosol comprised of the residual solids. This configuration enables a high respirable drug load of solid phase aerosol particles with a good deposition probability to be generated for sparsely soluble compounds. Respirable particles of 1-7 μm aerodynamic diameter can be targeted.
The use of APIS enables these agents to be generated as solutions potentially without the use of excipients and enables delivery of 100% active agent as a solid phase aerosol in an effort-independent manner at low inhalation velocities.
The APIS is 12 inches long and 3.5 inches in outer diameter with the aging chamber having an internal diameter of 3 inches and length of 6 inches. To achieve this marked reduction in size, the APIS combines multiple features to “condition” and concentrate the aerosol within a relatively small device. To achieve this end, the APIS contains several novel features. The aerosol generation jet is arrested, dispersed and partially evaporated with a coaxial counterflow air jet. Evaporation of the solvent is augmented by a sheaf of swirling airflow. The counterflow air jet and the sheaf of swirling air are warmed to further augment evaporation. Combined, these augment the evaporation of the aerosol through increased mixing which increases the energy transfer to the droplet to provide the latent heat of evaporation. This design also results in a more uniform transit time of the particles through the aging chamber.
The APIS is the only “hand held” or compact bench aerosol delivery device to include a passive aerosol concentrator. This concentrator is a small nine radial slit virtual impactor which requires no negative air suction to remove the unwanted air and solvent vapor. The output has a slight positive pressure which assists the patient inhalation. This obviates the effort dependent dispersion of dry powers and their subsequent proximal deposition due to their rapid inhalation. When using aqueous based aerosols, concentration factors of 6 are typical.
The use of the APIS in conjunction with a respiratory aerosol control system, RACS, enables the delivery of the medication throughout the breath together with both the optimization of inhalation rate and measurement of the volume of active agent delivered. This will result in both reduced treatment time and greater accuracy of the metered mass of the agent delivered to the respiratory tract. This is predicted to provide improved patient outcomes with less overdosing-related toxic and side effects.
An object of the invention is to provide a method for delivering aqueous-based aerosols at high dose rates, higher than present commercial nebulizers by a factor of 4-10.
Another object is to provide a method for delivering, surfactant, substantial masses of sparsely soluble agents, large molecules such as DNA, plasmids, liposomes and viral vectors.
An object of the invention is to provide a breath-activated respiratory aerosol control system for aerosol flow regulation when the clinical version of the invention (SUPRAER™) is used to deliver medication to patients.
Another object of the invention is to provide a simple, visible and cheap readout to directly measure and monitor the amount of medication remaining in the drug container during medication delivery.
It is another object to provide two different types of readouts for the aerosol control system: a LED bar dosage readout and a translucent window readout for direct visualization of the cartridge reservoir.
Another object of the invention is to measure respiratory flow using a differential pressure transducer and to simultaneously detect if the patient's mouth is connected to the mouthpiece or if the pressure signal is not measured with a pressure transducer within the flow tube. The outputs of these devices are utilized for monitoring breathing pattern and respiratory flow, and for operation of a pinch valve to automatically release active agent during inspiration.
Another object of the invention is to monitor the inhalation flow with a series of yellow, green, and red LEDs, which indicate that an inhaled flow is not yet high enough, in the range of an optimum flow, or too high, respectively. This provides a visual signal for the patient to control his/her breath pattern for optimal effect of aerosol delivery of medication.
Another object of the invention is to automatically deliver medication into the aerosol generator for aerosolization when a patient's inhalation flow detected by a differential pressure transducer is greater than a preset threshold and the patient's mouth is connected to the mouthpiece, which is detected by a pressure transducer. In the event that the patient's respiratory flow is insufficient to activate drug delivery, the preset threshold can be decreased by a threshold adjustment knob.
Another object of the invention is to apply two signals, flow rate and mouth-on, simultaneously measured by a differential pressure transducer and a pressure transducer, for decision making with regard to fluid delivery. The pressure at the output of the passive aerosol concentrator is positive, i.e. above atmospheric pressure, in both mouth-on and mouth-off states, so the mouth-on detection is required to avoid incorrectly delivering medication in the mouth-off state.
Another object of the invention is to apply a normally closed solenoid pinch valve for control of breath-activated drug delivery, in which fluid contacts only a piece of tubing that is squeezed by a solenoid. This provides a conveniently operated drug delivery controller for high-purity fluid applications.
Another object of the invention is to provide a breath-activated inhaler for patients to breathe normally during aerosol inhalation therapy. It is another object to deliver variable amounts of the entire dose of medication proportional to the amount of the patient's inspiratory time in one breath cycle. The longer the inspiratory period, the more medication delivered in one breath cycle.
Another object of the invention is to provide a breath-activated battery-powered electronic apparatus for the respiratory aerosol control system, which is designed with low-power and low-cost electronic elements housed in the SUPRAER's case. This reduces the cost of production, thereby making the device available to a greater number of users.
A feature of the invention is that the size of the initial droplets generated by the atomizer/atomizers can be chosen to obtain dry particles with a mass median aerodynamic diameter of about 3 μm.
Another feature of the invention is that the pressure, flow, temperature and dimensions of the heated coaxial counter flow air jet, and the heated swirling annular dilution air jet can be adjusted with consideration of the desired solution, particle size and dose rate.
Another feature of the invention is that the dimensions of the passive virtual impactor aerosol concentrator can be chosen with consideration to the flow rate and particle size desired to obtain the maximum aerosol delivery efficiency.
An advantage of the invention is that the patient treatment time can be shortened due to higher dose rate drug delivery.
An advantage of the invention is that it can deliver respirable stable particles for respiratory drug delivery (aerodynamic diameter ˜1-7 μm), consistent with a maximum drug load.
An advantage of the invention is that the inhaled particles can contain up to 100% of the active agent.
An advantage of the invention is that the production of initially large particles requires less energy than small particles, resulting in lower shear stresses.
An advantage of the invention is that aerosol generators which produce large particles can be incorporated, minimizing the degradation of large molecules such as DNA, plasmids, liposomes and viral vectors can be avoided. This can be achieved with microfluidic focused flow.
An advantage of the invention is that the particle size and fluid flow output are easily adjusted.
An advantage of the invention is that the passive virtual impactor aerosol concentrator can remove most of the air and unwanted vapor, thus preventing the vapor from condensing and thereby maintaining a stable particle size. This is particularly well achieved using the two stage concentrator described herein.
An advantage of the invention is that a small positive pressure of the PVI can reduce the effort by the patient.
Another advantage of the invention is that the solvent can be typically water, alcohol or other volatile solvent such as hydrofluoroalkane for hydrophilic and lipophilic drugs.
Another advantage of the invention is that the heated coaxial counter flow air jet which arrests the high velocity aerosol jet augments the mixing of aerosol and thus its evaporation. This markedly reduces the length of the device.
Another advantage of the invention is that the heated swirling annular dilution air jet surrounding the aerosol jet augments the mixing of aerosol and thus its evaporation.
Another advantage of the invention is that the heated swirling annular dilution air flow surrounding the aerosol jet reduces impaction of aerosol onto the inner wall of the chamber thereby increasing aerosol delivery efficiency.
Another advantage of the invention is that the respiratory aerosol control system can automatically deliver drug to the aerosol generator during inhalation and enable the patient to monitor inhalation flow rate so as to achieve the maximum aerosol delivery efficiency.
Another advantage of the invention is that the aerosol generator can be comprised of many jet-type aerosol generators which generate large particles.
An aspect of the invention is that the solution containing the active agent is contained in a cylindrical corrugated collapsible bag. This provides a uniform pressure and thus constant fluid flow is ensured by the fixed diameter of the corrugated bag, yielding a constant area for compression. A low compliance delivery tube enables the drug flow to a microfluidic focused flow aerosol generator to be controlled by a pinch valve. This can be sterilized and disposable resulting in greater patient safety and compliance.
Another advantage of the invention is that it enables the aerosol to be delivered to the patient only during optimal inhalation flow and for the entire duration of this optimal inhalation flow while monitoring the dose delivered. This reduces treatment time and increases the consistency of the dose delivered to the respiratory tract. Also, the ability of the patient to optimize inspiratory flow increases the deposition efficiency within the targeted region of the respiratory tract.
The detailed descriptions provide herein are for illustrative purposes only, and should not be construed as limiting the scope of the invention as described in the appended claims.
The assembly drawing of the invention is shown in
Aerosol Generator
The initial diameter of the aqueous/solvent-based aerosol generated is governed by the mass of the agent, in solution or suspension, that on complete evaporation of the solvent, will result in residual particles with aerodynamic diameters between 1 and 7 μm. For instance, a sparsely soluble compound will require the generation of large droplets whereas very soluble compounds such as NaCl or sugar require the generation of smaller droplets. The initial size will also be dependent on the density of the residual particles. To optimize the rate of drug delivery, the aqueous solvent flow rate should be about 1-5 ml/min. Thus, many aerosol generators can be used in conjunction with APIS. The Microjet aerosol generator 1000 in
The single-orifice aerosol generator shown in
Aerosol generation of small particles requires more energy and higher shear forces than that of larger particles. Thus APIS, when used in conjunction with a large particle aerosol generator, enables the generation of large molecules and biologics while minimizing the risk of their shear-induced degradation, and provides a resultant respirable aerosol. This configuration also enables sparsely soluble compounds to be generated with a high respirable drug load of solid aerosol particles with a good deposition probability. Final respirable particles size of 1-7 μm aerodynamic diameter can be targeted.
A multi-orifice aerosol generation system is illustrated in
Aerosol Evaporator
The aerosol processing and inhalation system contains several features to augment the evaporation of these 10-30 μm droplets in a small volume evaporation chamber 1100 as shown in
The inclination of the heat transfer vanes, as shown in
Aerosol Concentrator
APIS is the only “hand held” or compact bench aerosol inhalation delivery device to include a passive aerosol concentrator. The aerosol concentrator system assembly is shown in
Based on a cutoff aerodynamic diameter of 3 μm (density=1000 kg/m3) and particles with input air flow rates of 200 and 300 l/min, respectively, the calculated dimensions of the nozzles are shown in Table 1. The width of the rectangular impactor, W, is considered as a reference dimension. The distance between the nozzle and the plate S is S=1.5W, the thickness of the nozzle T=W, and the width of the slits in the plates aligned with the nozzles is 1.3W. The design incorporates nine slit nozzles radially arranged to have the capacity for processing large input volumetric flow rates (200 and 300 l/min).
The dimensions of the passive virtual impactor concentrator can be adjusted with consideration of the flow rate and particle size desired to obtain the maximum aerosol delivery efficiency.
Two Stage Concentrator/Evaporator
In a one stage concentrator, it is a prerequisite that the exhaust air is at less than 100% humidity for complete evaporation of the aqueous aerosol to dryness. In the case of saline aerosol (aqueous sodium chloride solution) the humidity must be below 70% to ensure complete dryness of the residual salt particle. This requires very high dilution air flows. This airflow can be substantially reduced using a two stage device. The air removed during the first stage can be 100% humidified and thus, in the case of sodium chloride, either less total air is required or more saline solution can be aerosolized. For example, for 100 l/min into the first concentrator with a fluid flow rate of 5 ml/min and temperature of 34° C., 40 mg/l×100 l/min or 4 ml will evaporate. Dilution of the 20 l/min output containing approximately 1 ml of water with 80 l/min of dry air results in a relative humidity at the exit of the second stage of 1000 mg/100 l/min or 10 mg/l, or 25% relative humidity at about 34° C., thus insuring complete evaporation of the water from the sodium chloride solute. As the total airflow is only 200 l/min, the dimensions of the concentrator can be markedly reduced in both diameter and length and a smaller air pump can be used. The assembly drawing of the invention with a two-stage aerosol concentrator is shown in
The two-stage concentrator has been designed for the present invention, based on a cutoff aerodynamic diameter of 3 μm water (density=1000 kg/m3) and particles at input air flow rates of 100 l/min. The dimensions of the nozzles are shown in Table 2. The width of the rectangular impactor, W, is considered as a reference dimension. The distance between the nozzle and the plate S=1.5W, the thickness of the nozzle T=W, and the width of the slits in the plates aligned with the nozzles is 1.3W The design incorporates 9 slit nozzles radially arranged to have the capacity for processing input volumetric flow rates of 100 l/min.
Invention Assembly for Clinical Applications
The device that is the subject of this invention can be configured for clinical applications. This configuration, SUPRAER™, is capable of delivering inhalable aqueous-based aerosols containing large molecules or genetic materials at a high dose rate. A clinical or home use design of SUPRAER™ is shown in
Most of SUPRAER™ is reusable and easy to clean. However, to avoid contamination and ensure proper drug dosage, a fluid reservoir, and focused flow fluidic aerosol generator has been designed into one single or multiuse disposable cartridge. A schematic of this cartridge is shown in
The aerosol generators can be atomizers or nebulizers which generate aqueous aerosols of 10-30 μm with a narrow size distribution. They can be single orifice generators or multi-orifice generators. An air jet 2300 (see
An aerosol evaporation and aging chamber 2100, as shown in
A combination 900 of LEDs and photoelectric cells, (see
When the drug bag 2003 is full of medication fluid, the infrared light from all 5 infrared light emitting diodes 904-908 is blocked by the bag 2003, so no infrared light can reach 5 infrared phototransistors 909-913. All 5 infrared phototransistors 909-913 are completely turned off. Thus, all 5 transistors 919-923, LED drivers of the LED bar dosage readout 605, can get a base current via resistors 914-918, respectively. This puts all 5 transistors 914-918 in the saturation state with a maximum collector current. All 5 green LED diodes 924-928 of the LED bar dosage readout 605 driven by the transistors 919-923 are lit to indicate a full dosage of fluid.
When fluid is released in response to the patient's respiration, the fluid bag 2003 is gradually compressed toward the right. When infrared light of an infrared light emitting diode, for example, infrared diode 904, is not blocked by the bag 2003, the light will reach the relative infrared phototransistor 909. The phototransistor 909 inverts completely from the turn-off state to the saturation state with a collector-emitter voltage of near zero. This collector-emitter voltage of the phototransistor 909 is unable to provide enough base current to retain the transistor 919 in the saturation state. The transistor 919 inverts the saturation state to the turn-off state with a zero collector current. The LED 924 of the LED bar dosage readout 605 is off and all 4 other LEDs 925-928 stay on, which indicates a non-full dosage of fluid.
When the entire dose of fluid is delivered, the fluid bag 2003 will be collapsed. All 5 phototransistors 909-913 receive the infrared light from 5 infrared diodes 904-908. Thus, all 5 phototransistors 909-913 are in the saturation state, which puts all 5 transistors 919-923 in the turn-off state. All LEDs 924-928 of the LED bar dosage readout 605 are thus off to indicate an empty dosage bag.
The aerosol exiting the chamber enters the large ends of nine radially arranged slit nozzles of the aerosol concentrator 2200 as shown in
Finally, the flow is regulated by a Respiratory Aerosol Control System 2400, RACS. This system makes use of a differential pressure sensor (Venturi effect flow meter), a one-way straight-through non-obstructive silicone valve at the exit of the concentrator, a pressure sensor at the output and a resistive exhalation valve. The aerosol inhalation flow, breathing pattern and aerosol generation are governed by logic circuits which in turn activate both flow indicator lights and inspiratory aerosol generation. These are further described below.
A diagram of the flow rate measuring system 500 is shown in
Referring to
Referring to
The pressure in the tube 502 is measured by the pressure sensor 510 at port 511 and used to detect if the patient's mouth is connected to the mouthpiece 503. A white LED 604 is used to indicate that the patient's mouth is connected to the mouthpiece.
The pressure differential measured by the transducer 507 has 4 possible stages corresponding to 4 states: 1) mouth off, 2) mouth on, 3) inhale, and 4) exhale, which is schematically shown in the top row of
The control circuitry 700 of the respiratory aerosol control system is shown in
When the patient's mouth is not connected to the mouthpiece 503, the air flow 505 passes through the tube 501 to the tube 502 and goes out via the mouthpiece 503, while the check valve 513 remains in the closed state. The output voltage VP 701 of the pressure sensor 510 is near zero. The voltage VP is connected to the positive input of an amplifier 704 through a resistor 702, amplified by the amplifier 704, and applied to the positive input of a comparator 708. The output voltage of the amplifier 704 is compared with a reference voltage 705 that is set by a resistor “voltage divider” composed of 2 resistors 706, 707 and led to the negative input of the comparator 708. By setting the reference voltage 705 to be greater than the output voltage of the amplifier 704, the output of the comparator 708 is held to logic zero (ground). This logic zero is applied to one input pin of each of 4 two-input “AND” gates, 709, 721, 730, and 740, so the outputs of all 4 gates, 709, 722, 732, and 740, are logic 0, i.e. (G1,G6,G8,G4)=(0,0,0,0). In this case, the gates 709, 722, 732, and 740 are unable to provide a current to the bases of 4 pairs of transistors (711 & 712, 724 & 725, 734 & 735, 743 & 744). These transistors are completely cut off with no collector current. All LEDs, 715, 728, 738, and 747, driven by the collector current of these transistors are turned off.
When a patient's mouth is connected to the mouthpiece 503, the pressure in the tubes 502 and 504 increases above the pressure threshold of the check valve 513. The valve 513 opens, so the airflow 505 passes through the valve 513 to ambient air. The voltage VP 701 increases to and remains at a positive value as shown in the lower part of
When the pressure differential (ΔP) 716 caused by an inhaled flow is in the range between the ΔPmin and ΔP1, the output voltage of the amplifier 719 is greater than the reference voltage 754 and less than both reference voltages 753 and 752. The output of the comparator 720 is logic 1, whereas the outputs of both comparators 729 and 739 are logic 0. The logic outputs of the 4 “AND” gates, 709, 722, 732, and 740, are (G1, G6,G8,G4)=(0,1,0,0), so only the yellow LED 728 (comprising 4 yellow LEDs) is on, indicating insufficient inhalation flow. At this time, the others, white 715, green 738 and red 747 LEDs are turned off. When an inhaled flow is within the optimum range, the output voltage of the amplifier 719 is greater than both the reference voltages 754 and 753 and less than the reference voltage 752. The outputs of both comparators 720 and 729 are logic 1, whereas the output of the comparator 740 is logic 0. The logic outputs of the 4 “AND” gates, 709, 722, 732, and 740, are (G1,G6,G8,G4)=(0,0,1,0). Only the green LED 738 (comprised of 4 green LEDs) is turned on to indicate an optimum inhalation flow and all others are off. When an inhaled flow causes a pressure differential greater than ΔP2, the output voltage of the amplifier 719 is greater than the reference voltages, 754753, and 752. All 3 comparators, 720, 729, and 739, have an output held at logic 1. The logic outputs of the 4 “AND” gates, 709, 722, 732, and 740, are (G1,G6,G8,G4)=(0,0,0,1) so that only the red LED 747 composed of 4 red LEDs is on, indicating that the inhalation flow rate is too high.
During the patient's exhalation, the one-way valve 512 is closed and the pressure in the tube 504 increases to above the pressure threshold of the check valve 513. The valve 513 opens, so the patient's exhaled flow passes through the valve 513 to the outside. The pressure differential (ΔP) 716 decreases to below ΔPmin. The output voltage of the amplifier 719 is less than all 3 reference voltages, 754753, and 752. All 3 comparators, 720, 729, and 739, output a logical 0. The logical outputs of the 4 “AND” gates, 709, 722, 732, and 740, are (G1,G6,G8,G4)=(1,0,0,0). Only the white LED 715 is lit to indicate the exhalation.
Fluid delivery to the nebulizer is controlled by a normally closed, direct-current (DC) solenoid-operated 2-way pinch valve (Cole-Parmer Canada Inc.) (2008 in
During inhalation, when the pressure differential measured by the pressure differential transducer 507 is greater than the ΔPmin the output voltage of the amplifier 719 is greater than the reference voltage 754. The output of the comparator 720 is held to logic 1. As both inputs of the “AND” gate 721 are logic 1, the output 755 of the gate 721 is logic 1. This control signal 755 provides a current via a resistor 801 to the base of transistor 802 (2N3053). The transistor 802 is completely open with the maximum collector current. While the collector current of the transistor passes through the wire coil wrapped on the solenoid electromagnet tube 805 of the pinch valve 804, the pinch valve 804 inverts from the normally closed state 809 to the opened state 810. Fluid goes from inlet 807 to outlet 808 of the pinch valve 804 and is delivered to the nebulizer for aerosolization. In the event that the patient's respiratory flow is insufficient to trigger the valve 804 for drug delivery, i.e., the output voltage of the amplifier 719 is less than the reference voltage 754, the reference voltage 754 at the preset threshold can be reduced by adjusting the value of the variable resistor 751. As the output voltage of amplifier 719 is greater than the reference voltage 754, the valve 804 can be triggered for drug delivery.
When the patient's mouth is off or during the exhalation, the output 755 of the “AND” gate 721 is held to logic 0 which does not provide a current to the base of the transistor 802 The transistor 802 is closed completely. The wire coil wrapped on the solenoid electromagnet tube 805 of the pinch valve 804 has no power supply and goes back to the normally closed state 809. No fluid is delivered.
To evaluate the APIS, five different orifice hole diameters, D=0.15 mm, 0.20 mm, 0.30 mm, 0.38 mm, and 0.5 mm; two capillary tube radii, R0=0.28 mm, 0.55 mm; and two distances H, defined as the tube mouth to the orifice, H=1.0 mm and 1.5 mm; have been tested in a microfluidic flow aerosol generation unit.
To evaluate the virtual impaction and arrest of the aerosol plume by the counter-flow air jet, photographs of the plume with and without the counter-flow air jet were taken.
The effectiveness of the processes engaged to rapidly evaporate the large particle aerosols within a small volume aging chamber were evaluated. After optimization with the 12 inch aging chamber, we were able to completely evaporate the aqueous droplets in less than 6 inches when the fluid flow was 4 ml/min. Thus, further experiments were carried out using the 6 inch aging chamber. The mean axial velocity of 300 l/min air in the 3.0 inch inner diameter evaporation chamber is 1 m/sec. The heated annular swirling dilution air jet can be adjusted with consideration of the solution particle size and the dose desired.
To increase the droplet evaporation rate, the temperature of the swirling annular air jet and counter-flow air jet were maintained at 60° C. using two mechanisms. First, four cartridge heaters (2 inches long, 0.25 inch wide) were embedded in the atomizer aluminum body (see
After optimization with the 12-inch aging chamber, we were able to completely evaporate the aqueous droplets in less than 6″ when the fluid flow rate was 4 ml/min. Thus, a 3 inch inner diameter 6 inch long Pyrex glass tube was used as an aging chamber.
To evaluate the performance of the APIS, the output aerosol was characterized using an aqueous saline solution (NaCl). An 8% saline solution containing 0.05% fluoroscein was injected into the aerosol generator at 1˜5 ml per minute. The fluid in the aerosol was evaporated, producing an aerosol of crystalline salt particles. The particle size distribution of the residual salt crystals was determined using a five-stage Marple-Miller Cascade Impactor (Model #: 150, MSP Corporation) operated at a flow rate of 30 l/min. The cut-off diameters of 1, 2, 3, 4 and 5 stages are 9.9 μm, 5 μm, 2.5 μm, 1.13 μm, 0.63 μm, respectively. A 47 mm membrane filter paper with 0.45 μm pore size was used to collect particles smaller than 0.63 μm at the end-stage of the cascade impactor. The NaCl containing fluoroescein collected in each stage of the cascade impactor and the filter paper was washed with 3 ml phosphate buffer solution with pH=7.4. The fluorescein concentration was measured at 490 nm using an LKB spectrophotometer (Model ULTROSPEC II). Thus, the relative mass of NaCl collected at each stage of the cascade impactor and the filter paper was determined and the cumulative mass deposited on each stage of the impactor was expressed as percent of the total mass deposited. Experiments were carried out using a 6 inch long Pyrex glass tube as the aging chamber. The mean axial velocity of 300 l/min air in the 3.0 inch inner diameter evaporation chamber was 1 m/sec.
The distribution of particles generated by the single-orifice aerosol generator at fluid flow rate Q=3 ml/min and various compressed air pressures of 8-40 psi is shown in Table 3. An orifice diameter of D=0.5 mm, capillary tube diameter of 2R0=1.1 mm, distance from the tube mouth to the orifice H=1.0 mm, input air flow Qair
Thus, the mass median aerodynamic diameters of the droplets generated and the particles are well within the design parameters of the system. The liquid flow rate does not greatly affect the mass median aerodynamic diameters of the particles generated, as shown in Table 4.
The particle distribution generated by the multi-orifice aerosol generator operated at a compressed air pressure of 35 psi with a fluid flow rate of Q=3 ml/min and 4 ml/min were also determined using a five stage Marple-Miller Cascade Impactor, and is shown in Table 5. An orifice with a diameter D=0.3 mm, the capillary tube with a diameter of 2R0=0.56 mm, the distance from open end of the tube mouth to the orifice H=0.5 mm, and input air flow Qair
The concentration factor, the ratio of the output aerosol concentration to the input aerosol concentration, was determined using saline solution. When coaxial counterflow air jet nozzle with a diameter of 1.8 mm, the counterflow jet flowrate Qcounter
The embodiment of this part of the invention is solely for purposes of illustration and should not be construed as limiting the scope of the present invention.
The concentration factor shown is well within the design parameters of the invention, and can be adjusted according to the desired performance by small design changes.
Evaluation of the Performance of APIS
The objects of this invention can be achieved by many techniques and methods known to persons who are skilled in this field. To those skilled and knowledgeable in the arts to which the present invention pertains, many widely differing embodiments will be suggested by the foregoing without departing from the intent and scope of the present invention. The descriptions and disclosures herein are intended solely for purposes of illustration and should not be construed as limiting the scope of the present invention as described in the appended claims.
The present invention was made with U.S. Government support from the National Institutes of Health, National Heart, Lung, and Blood Institute, under grant No. HL78281. The U.S. Government has certain rights in this invention.