The present invention relates to medical devices for monitoring vital signs, e.g., arterial blood pressure.
False alarms generated by conventional vital sign monitors can represent up to 90% of all alarms in critical and peri-operative care, and are therefore a source of concern. A variety of factors cause false alarms, one of which is motion-related artifacts. Ultimately false alarms can have a severe impact on the safety of hospitalized patients: they can desensitize medical professionals toward ‘true positive’ alarms, lead them to set dangerously wide alarm thresholds, or even drive them to completely disable alarms. This can have a particularly profound impact in lower-acuity areas of the hospital, i.e. areas outside the intensive care unit (ICU), emergency department (ED), or operating room (OR), where the ratio of medical professionals to patients can be relatively low. In these areas a single medical professional (e.g. a nurse) often has to care for a large number of patients, and necessarily relies on automated alarms operating on vital sign monitors to effectively monitor their patients.
Studies in critical care environments indicate that the majority of false positive alarms are simple ‘threshold alarms’, meaning they are generated when a patient's vital sign exceeds a predetermined threshold. Patient motion can result in a vital sign having an erroneous high or low value, which in turn can trigger the false alarm. In most cases, these alarms lack any real clinical meaning, and go away after about 20 seconds when they are not acknowledged. Alarms can also be artificially induced when a patient is moved or manipulated, or if there is an actual problem with the vital sign monitor. False alarms due to motion-related artifacts are particularly very high when measured from ambulatory patients.
Blood pressure is a vital sign that is particularly susceptible to false alarms. In critical care environments like the ICU and OR, blood pressure can be continuously monitored with an arterial catheter inserted in the patient's radial or femoral artery. Alternatively, blood pressure can be measured intermittently using a pressured cuff and a technique called oscillometry. A vital sign monitor performs both the catheter and cuff-based measurements of blood pressure. Alternatively, blood pressure can be monitored continuously with a technique called pulse transit time (PTT), defined as the transit time for a pressure pulse launched by a heartbeat in a patient's arterial system. PTT has been shown in a number of studies to correlate to systolic (SYS), diastolic (DIA), and mean (MAP) blood pressures. In these studies, PTT is typically measured with a conventional vital signs monitor that includes separate modules to determine both an electrocardiogram (ECG) and pulse oximetry (SpO2). During a PTT measurement, multiple electrodes typically attach to a patient's chest to determine a time-dependent ECG component characterized by a sharp spike called the ‘QRS complex’. The QRS complex indicates an initial depolarization of ventricles within the heart and, informally, marks the beginning of the heartbeat and a pressure pulse that follows. SpO2 is typically measured with a bandage or clothespin-shaped sensor that attaches to a patient's finger, and includes optical systems operating in both the red and infrared spectral regions. A photodetector measures radiation emitted from the optical systems that transmits through the patient's finger. Other body sites, e.g., the ear, forehead, and nose, can also be used in place of the finger. During a measurement, a microprocessor analyses both red and infrared radiation detected by the photodetector to determine the patient's blood oxygen saturation level and a time-dependent waveform called a photoplethysmograph (‘PPG’). Time-dependent features of the PPG indicate both pulse rate and a volumetric absorbance change in an underlying artery caused by the propagating pressure pulse.
Typical PTT measurements determine the time separating a maximum point on the QRS complex (indicating the peak of ventricular depolarization) and a foot of the optical waveform (indicating the beginning the pressure pulse). PTT depends primarily on arterial compliance, the propagation distance of the pressure pulse (which is closely approximated by the patient's arm length), and blood pressure. To account for patient-dependent properties, such as arterial compliance, PTT-based measurements of blood pressure are typically ‘calibrated’ using a conventional blood pressure cuff and oscillometry. Typically during the calibration process the blood pressure cuff is applied to the patient, used to make one or more blood pressure measurements, and then left on the patient. Going forward, the calibration blood pressure measurements are used, along with a change in PTT, to continuously measure the patient's blood pressure (defined herein as ‘cNIBP). PTT typically relates inversely to blood pressure, i.e., a decrease in PTT indicates an increase in blood pressure.
A number of issued U.S. Patents describe the relationship between PTT and blood pressure. For example, U.S. Pat. Nos. 5,316,008; 5,857,975; 5,865,755; and 5,649,543 each describe an apparatus that includes conventional sensors that measure an ECG and PPG, which are then processed to determine PTT. U.S. Pat. No. 5,964,701 describes a finger-ring sensor that includes an optical system for detecting a PPG, and an accelerometer for detecting motion.
To improve the safety of hospitalized patients, particularly those in lower-acuity areas, it is desirable to have a vital sign monitor operating algorithms featuring: 1) a low percentage of false positive alarms/alerts; and 2) a high percentage of true positive alarms/alerts. The term ‘alarm/alert’, as used herein, refers to an audio and/or visual alarm generated directly by a monitor worn on the patient's body, or alternatively a remote monitor (e.g., a central nursing station). To accomplish this, the invention provides a body-worn monitor that measures a patient's vital signs (e.g. SYS, DIA, SpO2, heart rate, respiratory rate, and temperature) while simultaneously characterizing their activity state (e.g. resting, walking, convulsing, falling). The body-worn monitor processes this information to minimize corruption of the vital signs by motion-related artifacts. A software framework generates alarms/alerts based on threshold values that are either preset or determined in real time. The framework additionally includes a series of ‘heuristic’ rules that take the patient's activity state and motion into account, and process the vital signs accordingly. These rules, for example, indicate that a walking patient is likely breathing and has a regular heart rate, even if their motion-corrupted vital signs suggest otherwise.
The body-worn monitor features a series of sensors that measure time-dependent PPG, ECG, motion (ACC), and pressure waveforms to continuously monitor a patient's vital signs, degree of motion, posture and activity level. Blood pressure, a vital sign that is particularly useful for characterizing a patient's condition, is typically calculated from a PTT value determined from the PPG and ECG waveforms. Once determined, blood pressure and other vital signs can be further processed, typically with a server within a hospital, to alert a medical professional if the patient begins to decompensate.
In other embodiments, PTT can be calculated from time-dependent waveforms other than the ECG and PPG, and then processed to determine blood pressure. In general, PTT can be calculated by measuring a temporal separation between features in two or more time-dependent waveforms measured from the human body. For example, PTT can be calculated from two separate PPGs measured by different optical sensors disposed on the patient's fingers, wrist, arm, chest, or virtually any other location where an optical signal can be measured using a transmission or reflection-mode optical configuration. In other embodiments, PTT can be calculated using at least one time-dependent waveform measured with an acoustic sensor, typically disposed on the patient's chest. Or it can be calculated using at least one time-dependent waveform measured using a pressure sensor, typically disposed on the patient's bicep, wrist, or finger. The pressure sensor can include, for example, a pressure transducer, piezoelectric sensor, actuator, polymer material, or inflatable cuff.
In one aspect, the invention provides a system for processing at least one vital sign from a patient along with a motion parameter and, in response, generating an alarm. The system features two sensors to measure the vital sign, each with a detector configured to detect a time-dependent physiological waveform indicative of one or more contractile properties of the patient's heart. The contractile property, for example, can be a beat, expansion, contraction, or any time-dependent variation of the heart that launches both electrical signals and a bolus of blood. The physiological waveform, for example, can be an ECG waveform measured from any vector on the patient, a PPG waveform, an acoustic waveform measured with a microphone, or a pressure waveform measured with a transducer. In general, these waveforms can be measured from any location on the patient. The system includes at least two motion-detecting sensors (e.g. analog or digital accelerometers) positioned on locations selected from a forearm, upper arm, and a body location other than the forearm or upper arm of the patient. Here, ‘forearm’ means any portion of the arm below the elbow, e.g. the forearm, wrist, hand, and fingers. ‘Upper arm’ means any portion of the arm above and including the elbow, e.g. the bicep, shoulder, and armpit. Each of the motion-detecting sensors generate at least one motion waveform, and typically a set of three motion waveforms (each corresponding to a different axis), indicative of motion of the location on the patient's body to which it is affixed.
A processing component (e.g., an algorithm or any computation function operating on a microprocessor or similar logic device in the wrist-worn transceiver) receives and processes the time-dependent physiological and motion waveforms. The processing component performs the following steps: (i) calculates at least one vital sign (e.g., SYS, DIA, SpO2, heart rate, and respiratory rate) from the first and second time-dependent physiological waveforms; and (ii) calculates at least one motion parameter (e.g. posture, activity state, arm height, and degree of motion) from the motion waveforms. A second processing component, which can be another algorithm or computational function operating on the microprocessor, receives the vital sign and motion parameter and determines: (i) a first alarm condition, calculated by comparing the vital sign to an alarm threshold; (ii) a second alarm condition, calculated from the motion parameter; and (iii) an alarm rule, determined by collectively processing the first and second alarm conditions with an alarm algorithm. The alarm rule indicates, e.g., whether or not the system generates an alarm.
In embodiments, the motion parameter corresponds to one of the following activities or postures: resting, moving, sitting, standing, walking, running, falling, lying down, and convulsing. Typically the alarm rule automatically generates the alarm if the motion parameter is one of falling or convulsing, as these activities typically require immediate medical attention. If the motion parameter corresponds to walking or most ambulatory motions, then the alarm rule does not necessarily generate an alarm for vital signs such as heart rate, respiratory rate, and SpO2. Here, the patient is assumed to be in a relatively safe state since they are walking. However, even while the patient is in this activity state, the alarm rule can still generate an alarm if the heart rate exceeds an alarm threshold that is increased relative to its initial value. If the motion parameter corresponds to standing, and the vital sign is blood pressure, then the alarm rule can generate the alarm if the blood pressure exceeds an alarm threshold that is decreased relative to its initial value. This is because it is relatively normal for a patient's blood pressure to safely drop as the move from a sitting or lying posture to a standing posture.
In embodiments, the vital sign is blood pressure determined from a time difference (e.g. a PTT value) between features in the ECG and PPG waveforms, or alternatively using features between any combination of time-dependent ECG, PPG, acoustic, or pressure waveforms. This includes, for example, two PPG waveforms measured from different locations on the patient's body. The motion parameter can be calculated by processing either a time or frequency-dependent component from at least one motion waveform. For example, the processing component can determine that the patient is walking, convulsing, or falling by: i) calculating a frequency-dependent motion waveform (e.g. a power spectrum of a time-dependent motion waveform); and ii) analyzing a band of frequency components from the frequency-dependent waveform. A band of frequency components between 0-3 Hz typically indicates that the patient is walking, while a similar band between 0-10 Hz typically indicates that the patient is convulsing. Finally, a higher-frequency band between 0-15 Hz typically indicates that a patient is falling. In this last case, the time-dependent motion waveform typically includes a signature (e.g. a rapid change in value) that can be further processed to indicate falling. Typically this change represents at least a 50% change in the motion waveform's value within a time period of less than 2 seconds. In other embodiments, the first processing component determines the motion parameter by comparing a parameter determined from the motion waveform (e.g., from a time or frequency-dependent parameter of the waveform) to a pre-determined ROC threshold value associated with a pre-determined ROC curve.
In embodiments, both the first and second processing components are algorithms or computational functions operating on one or more microprocessors. Typically the processing components are algorithms operating on a common microprocessor worn on the patient's body. Alternatively, the first processing component is an algorithm operating on a processor worn on the patient's body, and the second processing component is an algorithm operating on a remote computer (located, e.g., at a central nursing station).
In another aspect, the invention provides a method for continuously monitoring a patient featuring the following steps: (i) detecting first and second time-dependent physiological waveforms indicative of one or more contractile properties of the patient's heart with first and second body-worn sensors; (ii) detecting sets of time-dependent motion waveforms with at least two body-worn, motion-detecting sensors; (iii) processing the first and second time-dependent physiological waveforms to determine at least one vital sign from the patient; (iv) analyzing a portion of the sets of time-dependent motion waveforms with a motion-determining algorithm to determine the patient's activity state (e.g. resting, moving, sitting, standing, walking, running, falling, lying down, and convulsing); and (v) generating an alarm by processing the patient's activity state and comparing the vital sign to a predetermined alarm criteria corresponding to this state.
In embodiments, the analyzing step features calculating a mathematical transform (e.g. a Fourier Transform) of at least one time-dependent motion waveform to determine a frequency-dependent motion waveform (e.g. a power spectrum), and then analyzing frequency bands in this waveform to determine if the patient is walking, convulsing, or falling. This step can also include calculating a time-dependent change or variation in the time-dependent waveforms, e.g. a standard deviation, mathematical derivative, or a related statistical parameter. In other embodiments, the analyzing step includes determining the motion parameter by comparing a time-dependent motion waveform to a mathematical function using, e.g., a numerical fitting algorithm such as a linear least squares or Marquardt-Levenberg non-linear fitting algorithm.
The analyzing step can include calculating a ‘logit variable’ from at least one time-dependent motion waveform, or a waveform calculated therefrom, and comparing the logit variable to a predetermined ROC curve to determine the patient's activity state. For example, the logit variable can be calculated from at least one time or frequency-dependent motion waveform, or a waveform calculated therefrom, and then compared to different ROC curves corresponding to various activity and posture states.
In another aspect, the invention provides a system for continuously monitoring a group of patients, wherein each patient in the group wears a body-worn monitor similar to those described herein. Additionally, each body-worn monitor is augmented with a location sensor. The location sensor includes a wireless component and a location processing component that receives a signal from the wireless component and processes it to determine a physical location of the patient. A processing component (similar to that described above) determines from the time-dependent waveforms at least one vital sign, one motion parameter, and an alarm parameter calculated from the combination of this information. A wireless transceiver transmits the vital sign, motion parameter, location of the patient, and alarm parameter through a wireless system. A remote computer system featuring a display and an interface to the wireless system receives the information and displays it on a user interface for each patient in the group.
In embodiments, the user interface is a graphical user interface featuring a field that displays a map corresponding to an area with multiple sections. Each section corresponds to the location of the patient and includes, e.g., the patient's vital signs, motion parameter, and alarm parameter. For example, the field can display a map corresponding to an area of a hospital (e.g. a hospital bay or emergency room), with each section corresponding to a specific bed, chair, or general location in the area. Typically the display renders graphical icons corresponding to the motion and alarm parameters for each patient in the group. In other embodiments, the body-worn monitor includes a graphical display that renders these parameters directly on the patient.
Typically the location sensor and the wireless transceiver operate on a common wireless system, e.g. a wireless system based on 802.11, 802.15.4, or cellular protocols. In this case a location is determined by processing the wireless signal with one or more algorithms known in the art. These include, for example, triangulating signals received from at least three different base stations, or simply estimating a location based on signal strength and proximity to a particular base station. In still other embodiments the location sensor includes a conventional global positioning system (GPS).
The body-worn monitor can include a first voice interface, and the remote computer can include a second voice interface that integrates with the first voice interface. The location sensor, wireless transceiver, and first and second voice interfaces can all operate on a common wireless system, such as one of the above-described systems based on 802.11 or cellular protocols. The remote computer, for example, can be a monitor that is essentially identical to the monitor worn by the patient, and can be carried or worn by a medical professional. In this case the monitor associate with the medical professional features a display wherein the user can select to display information (e.g. vital signs, location, and alarms) corresponding to a particular patient. This monitor can also include a voice interface so the medical professional can communicate with the patient.
In another aspect, the invention provides a body-worn monitor featuring optical sensor that measures two time-dependent optical waveforms (e.g. PPG waveforms) from the patient's body, and an electrical sensor featuring at least two electrodes and an electrical circuit that collectively measure a first time-dependent electrical waveform (e.g., an ECG waveform) indicating the patient's heart rate, and a second time-dependent electrical waveform (e.g. a waveform detected with impedance pneumography) indicating the patient's respiratory rate. The monitor includes at least two motion-detecting sensors positioned on two separate locations on the patient's body. A processing component, similar to that described above, determines: (i) a time difference between features in one of the time-dependent optical and electrical waveforms; (ii) a blood pressure value calculated from the time difference; iii) an SpO2 value calculated from both the first and second optical waveforms; (iii) a heart rate calculated from one of the time-dependent electrical waveforms; (iv) a respiratory rate calculated from the second time-dependent electrical waveform; (v) at least one motion parameter calculated from at least one motion waveform; and (vi) an alarm parameter calculated from at least one of the blood pressure value, SpO2 value, heart rate, respiratory rate, and the motion parameter.
In embodiments, the processing component renders numerical values corresponding to the blood pressure value, SpO2 value, heart rate, and respiratory rate on a graphical display. These parameters, however, are not rendered when the motion parameter corresponds to a moving patient (e.g. a walking patient). Using the motion waveforms, the monitor can detect when the patient is lying down, and from the electrical waveforms if their respiratory rate has ceased for an extended period of time (e.g. at least 20 seconds). In this case, for example, the processing component can generate an alarm parameter corresponding to apnea. The time-dependent electrical waveforms can be further processed to determine heart rate along with an additional parameter, such as VFIB, VTAC, and ASY, defined in detail below. Similarly, the processing component can analyze the time-dependent optical waveforms to determine a pulse rate, and can determine a pulse pressure from a difference between diastolic and systolic blood pressures. It determines a ‘significant pulse rate’ if the pulse rate is greater than 30 beats per minute, and the pulse pressure is greater than 10 mmHg. The monitor then generates an alarm parameter corresponding to one of VFIB, VTAC, and ASY if these parameters are determined from the patient and a significant pulse rate is not present.
In other embodiments, the processing component can process at least one motion waveform to determine a number of times the patient moves from lying in a first position to lying in a different position, and generate an alarm parameter if the number is less than a threshold value (e.g. once per four hours). Such an alarm indicates, for example, a ‘bed sore index’, i.e. an index that indicates when the patient may develop lesions due to inactivity. The monitor can also include a temperature sensor, configured, e.g., to attach to a portion of the patient's chest.
In another aspect, the invention provides a body-worn monitor described above for monitoring a patient's vital signs using time-dependent ECG and PPG waveforms. The processing component determines at least one motion parameter measured by a motion-detecting sensor (e.g. an accelerometer) representing the patient's posture, activity state, and degree of motion. The motion parameter is calculated by comparing a component determined from a time or frequency-dependent waveform or a ROC curve to a predetermined threshold value. An alarm is generated by collectively processing a vital sign and the motion parameter with an alarm algorithm. The monitor can include a graphical display, worn on the patient's body, which renders numerical values indicating the patient's vital signs, and iconic images indicating both the motion parameter and the alarm. The graphical display typically includes a first user interface for a patient, and a second user interface for a medical professional that is rendered after the processing unit processes an identifier (e.g. a barcode or radio frequency identification, or RFID) corresponding to the medical professional. The body-worn monitor can also include a wireless transceiver that transmits the vital sign, motion parameter, and alarm to a remote computer which further includes a graphical display for rendering this information.
In another aspect, the invention provides a method for generating an alarm while monitoring vital signs and posture of a patient. A monitor, similar to that described above, measures vital signs from time-dependent waveforms (e.g. any combination of optical, electrical, acoustic, or pressure waveforms) and a patient's posture with at least one motion-detecting sensor positioned on the patient's torso (e.g., an accelerometer positioned on the patient's chest). The processing component analyzes at least a portion of a set of time-dependent motion waveforms generated by the motion-detecting sensor to determine a vector corresponding to motion of the patient's torso. It then compares the vector to a coordinate space representative of how the motion-detecting sensor is oriented on the patient to determine a posture parameter, which it then processes along with the vital sign to generate an alarm. The alarm, for example, is indicated by a variance of the vital sign relative to a predetermined alarm criterion, and is regulated according to the patient's posture.
In embodiments, the method generates the alarm in response to a change in the patient's posture, e.g. if the patient is standing up, or if their posture changes from lying down to either sitting or standing up, or from standing up to either sitting or lying down.
To determine the vector the method includes an algorithm or computation function that analyzes three time-dependent motion waveforms, each corresponding to a unique axis of the motion-detecting sensor. The motion waveforms can yield three positional vectors that define a coordinate space. In a preferred embodiment, for example, the first positional vector corresponds to a vertical axis, a second positional vector corresponds to a horizontal axis, and the third positional vector corresponds to a normal axis extending normal from the patient's chest. Typically the posture parameter is an angle, e.g. an angle between the vector and at least one of the three positional vectors. For example, the angle can be between the vector and a vector corresponding to a vertical axis. The patient's posture is estimated to be upright if the angle is less than a threshold value that is substantially equivalent to 45 degrees (e.g., 45 degrees+/−10 degrees); otherwise, the patient's posture is estimated to be lying down. If the patient is lying down, the method can analyze the angle between the vector and a vector corresponding to a normal axis extending normal from the patient's chest. In this case, the patient's posture is estimated to be supine if the angle is less than a threshold value substantially equivalent to 35 degrees (e.g., 35 degrees+/−10 degrees), and prone if the angle is greater than a threshold value substantially equivalent to 135 degrees (e.g., 135 degrees+/−10 degrees). Finally, if the patient is lying down, the method can analyze the angle between the vector and a vector corresponding to a horizontal axis. In this case, the patient is estimated to be lying on a first side if the angle is less than a threshold value substantially equivalent to 90 degrees (e.g., 90 degrees+/−10 degrees), and lying on an opposing side if the angle is greater than a threshold value substantially equivalent to 90 degrees (e.g., 90 degrees+/−10 degrees).
Blood pressure is determined continuously and non-invasively using a technique, based on PTT, which does not require any source for external calibration. This technique, referred to herein as the ‘composite technique’, operates on the body-worn monitor and wirelessly transmits information describing blood pressure and other vital signs to the remote monitor. The composite technique is described in detail in the co-pending patent application entitled: VITAL SIGN M FOR MEASURING BLOOD PRESSURE USING OPTICAL, ELECTRICAL, AND PRESSURE WAVEFORMS (U.S. Ser. No. 12/138,194; filed Jun. 12, 2008), the contents of which are fully incorporated herein by reference.
Still other embodiments are found in the following detailed description of the invention, and in the claims.
System Overview
The ECG 16 and pneumatic 20 systems are stand-alone systems that include a separate microprocessor and analog-to-digital converter. During a measurement, they connect to the transceiver 12 through connectors 28, 30 and supply digital inputs using a communication protocol that runs on a controller-area network (CAN) bus. The CAN bus is a serial interface, typically used in the automotive industry, which allows different electronic systems to effectively communicate with each other, even in the presence of electrically noisy environments. A third connector 32 also supports the CAN bus and is used for ancillary medical devices (e.g. a glucometer) that is either worn by the patient or present in their hospital room.
The optical system 18 features an LED and photodetector and, unlike the ECG 16 and pneumatic 20 systems, generates an analog electrical signal that connects through a cable 36 and connector 26 to the transceiver 12. As is described in detail below, the optical 18 and ECG 16 systems generate synchronized time-dependent waveforms that are processed with the composite technique to determine a PTT-based blood pressure along with motion information.
The first accelerometer 14a is surface-mounted on a printed circuited board within the transceiver 12, which is typically worn on the patient's wrist like a watch. The second 14b accelerometer is typically disposed on the upper portion of the patient's arm and attaches to a cable 40 that connects the ECG system 16 to the transceiver 12. The third accelerometer 14c is typically worn on the patient's chest proximal to the ECG system 16. The second 14b and third 14c accelerometers integrate with the ECG system 16 into a single cable 40, as is described in more detail below, which extends from the patient's wrist to their chest and supplies digitized signals over the CAN bus. In total, the cable 40 connects to the ECG system 16, two accelerometers 14b, 14c, and at least three ECG electrodes (shown in
To determine posture, arm height, activity level, and degree of motion, the transceiver's CPU 22 processes signals from each accelerometer 14a-c with a series of algorithms, described in detail below. In total, the CPU can process nine unique, time-dependent signals (ACC1-9) corresponding to the three axes measured by the three separate accelerometers. Specifically, the algorithms determine parameters such as the patient's posture (e.g., sitting, standing, walking, resting, convulsing, falling), the degree of motion, the specific orientation of the patient's arm and how this affects vital signs (particularly blood pressure), and whether or not time-dependent signals measured by the ECG 16, optical 18, or pneumatic 20 systems are corrupted by motion. Once this is complete, the transceiver 12 uses an internal wireless transmitter 24 to send information in a series of packets, as indicated by arrow 34, to a remote monitor within a hospital. The wireless transmitter 24 typically operates on a protocol based on 802.11 and communicates with an existing network within the hospital. This information alerts a medical professional, such as a nurse or doctor, if the patient begins to decompensate. A server connected to the hospital network typically generates this alarm/alert once it receives the patient's vital signs, motion parameters, ECG, PPG, and ACC waveforms, and information describing their posture, and compares these parameters to preprogrammed threshold values. As described in detail below, this information, particularly vital signs and motion parameters, is closely coupled together. Alarm conditions corresponding to mobile and stationary patients are typically different, as motion can corrupt the accuracy of vital signs (e.g., by adding noise), and induce changes in them (e.g., through acceleration of the patient's heart and respiratory rates).
General Methodology for Alarms/Alerts
Algorithms operating on either the body-worn monitor or remote monitor generate alarms/alerts that are typically grouped into three general categories: 1) motion-related alarms/alerts indicating the patient is experiencing a traumatic activity, e.g. falling or convulsing; 2) life-threatening alarms/alerts typically related to severe events associated with a patient's cardiovascular or respiratory systems, e.g. asystole (ASY), ventricular fibrillation (VFIB), ventricular tachycardia (VTAC), and apnea (APNEA); and 3) threshold alarms/alerts generated when one of the patient's vital signs (SYS, DIA, SpO2, heart rate, respiratory rate, or temperature) exceeds a threshold that is either predetermined or calculated directly from the patient's vital signs. The general methodology for generating alarms/alerts in each of these categories is described in more detail below.
Motion-Related Alarms/Alerts
The figures indicate that time-dependent properties of both ECG 50, 55, 60, 65 and PPG 51, 56, 61, 66 waveforms are strongly affected by motion, as indicated by the ACC waveforms 52, 57, 62, 67. Accuracy of the vital signs, such as SYS, DIA, heart rate, respiratory rate, and SpO2, calculated from these waveforms is therefore affected as well. Body temperature, which is measured from a separate body-worn sensor (typically a thermocouple) and does not rely on these waveforms, is relatively unaffected by motion.
The ECG waveform 55 measured from the walking patient is relatively unaffected by motion, other than indicating an increase in heart rate (i.e., a shorter time separation between neighboring QRS complexes) and respiratory rate (i.e. a higher frequency modulation of the waveform's envelope) caused by the patient's exertion. The PPG waveform 56, in contrast, is strongly affected by this motion, and becomes basically immeasurable. Its distortion is likely due to a quasi-periodic change in light levels, caused by the patient's swinging arm, and detected by the optical sensor's photodetector. Movement of the patient's arm additionally affects blood flow in the thumb and can cause the optical sensor to move relative to the patient's skin. The photodetector measures all of these artifacts, along with a conventional PPG signal (like the one shown in
The body-worn monitor deploys multiple strategies to avoid generating false alarms/alerts during a walking activity state. As described in detail below, the monitor can detect this state by processing the ACC waveforms shown in
To further reduce false alarms/alerts, software associated with the body-worn monitor or remote monitor can deploy a series of heuristic rules determined beforehand using practical, empirical studies. These rules, for example, can indicate that a walking patient is likely healthy, breathing, and characterized by a normal SpO2. Accordingly, the rules dictate that respiratory rate and SpO2 values that are measured during a walking state and exceed predetermined alarm/alert thresholds are likely corrupted by artifacts; the system, in turn, does not sound the alarm/alert in this case. Heart rate, as indicated by
Convulsing modulates the ACC waveform 62 due to rapid motion of the patient's arm, as measured by the wrist-worn accelerometer. This modulation is strongly coupled into the PPG waveform 61, likely because of the phenomena described above, i.e.: 1) ambient light coupling into the optical sensor's photodiode; 2) movement of the photodiode relative to the patient's skin; and 3) disrupted blow flow underneath the optical sensor. Note that from about 23-28 seconds the ACC waveform 62 is not modulated, indicating that the patient's arm is at rest. During this period the ambient light is constant and the optical sensor is stationary relative to the patient's skin. But the PPG waveform 61 is still strongly modulated, albeit at a different frequency than the modulation that occurred when the patient's arm was moving. This indicates modulation of the PPG waveform 61 is likely caused by at least the three factors described above, and that disrupted blood flow underneath the optical sensor continues even after the patient's arm stops moving. Using this information, both ECG and PPG waveforms similar to those shown in
The ECG waveform 60 is modulated by the patient's arm movement, but to a lesser degree than the PPG waveform 61. In this case, modulation is caused primarily by electrical ‘muscle noise’ instigated by the convulsion and detected by the ECG electrodes, and well as by convulsion-induced motion in the ECG cables and electrodes relative to the patient's skin. Such motion is expected to have a similar affect on temperature measurements, which are determined by a sensor that also includes a cable.
Table 1B, below, shows the modified threshold values and heuristic rules for alarms/alerts generated by a convulsing patient. In general, when a patient experiences convulsions, such as those simulated during the two 12-second periods in
Table 1B also shows the heuristic rules for convulsing patients. Here, the overriding rule is that a convulsing patient needs assistance, and thus an alarm/alert for this patient is generated regardless of their vital signs (which, as described above, are likely inaccurate due to motion-related artifacts). The system always generates an alarm/alert for a convulsing patient.
After a fall, both the ECG 65 and PPG 66 waveforms are free from artifacts, but both indicate an accelerated heart rate and relatively high heart rate variability for roughly 10 seconds. During this period the PPG waveform 66 also shows a decrease in pulse amplitude. Without being bound to any theory, the increase in heart rate may be due to the patient's baroreflex, which is the body's hemostatic mechanism for regulating and maintaining blood pressure. The baroreflex, for example, is initiated when a patient begins faint. In this case, the patient's fall may cause a rapid drop in blood pressure, thereby depressing the baroreflex. The body responds by accelerating heart rate (indicated by the ECG waveform 65) and increasing blood pressure (indicated by a reduction in PTT, as measured from the ECG 65 and PPG 66 waveforms) in order to deliver more blood to the patient's extremities.
Table 1C shows the heuristic rules and modified alarm thresholds for a falling patient. Falling, similar to convulsing, makes it difficult to measure waveforms and the vital signs calculated from them. Because of this and the short time duration associated with a fall, alarms/alerts based on vital signs thresholds are not generated when a patient falls. However, this activity, optionally coupled with prolonged stationary period or convulsion (both determined from the following ACC waveform), generates an alarm/alert according to the heuristic rules.
As described in detail below, the patient's specific activity relates to both the time-dependent ACC waveforms and the frequency-dependent Fourier Transforms of these waveforms.
The ACC waveform corresponding to a resting patient (52 in
Life-Threatening Alarms/Alerts
ASY and VFIB are typically determined directly from the ECG waveform using algorithms known in the art. To reduce false alarms associated with these events, the body-worn monitor calculates ASY and VFIB from the ECG waveform, and simultaneously determines a ‘significant pulse’ from both the PPG waveform and cNIBP measurement, described below. A significant pulse occurs when the monitor detects a pulse rate from the PPG waveform (see, for example, 51 in
The alarm/alert for ASY and VFIB additionally depends on the patient's activity level. For example, if the monitor determines ASY and VFIB from the ECG, and that the patient is walking from the ACC waveforms, it then checks for a significant pulse and determines pulse rate from the PPG waveform. In this situation the patient is assumed to be in an activity state prone to false alarms. The alarms/alerts related to ASY and VFIB are thus delayed, typically by 20-30 seconds, if the monitor determines the patient's pulse to be significant and their current pulse differs from their pulse rate measured during the previous 60 seconds by less than 40%. The monitor sounds an alarm only if ASY and VFIB remain after the delay period and once the patient stops walking. In another embodiment, an alarm/alert is immediately sounded if the monitor detects either ASY or VFIB, and no significant pulse is detected from the PPG waveform for between 5-10 seconds.
The methodology for alarms/alerts is slightly different for VTAC due to the severity of this condition. VTAC, like ASY and VFIB, is detected directly from the ECG waveform using algorithms know in the art. This condition is typically defined as five or more consecutive premature ventricular contractions (PVCs) detected from the patient's ECG. When VTAC is detected from the ECG waveform, the monitor checks for a significant pulse and compares the patient's current pulse rate to that measured during the entire previous 60 seconds. The alarm/alert related to VTAC is delayed, typically by 20-30 seconds, if the pulse is determined to be significant and the pulse rate measured during this period differs from patient's current pulse rate by less than 25%. The monitor immediately sounds an alarm/alert if VTAC measured from the ECG waveform meets the following criteria: 1) its persists after the delay period; 2) the deficit in the pulse rate increases to more than 25% at any point during the delay period; and 3) no significant pulse is measured for more than 8 consecutive seconds during the delay period. The alarm for VTAC is not generated if any of these criteria are not met.
APNEA refers to a temporary suspension in a patient's breathing and is determined directly from respiratory rate. The monitor measures this vital sign from the ECG waveform using techniques called ‘impedance pneumography’ or ‘impedance rheography’, both of which are known in the art. The monitor sounds an alarm/alert only if APNEA is detected and remains (i.e. the patient does not resume normal breathing) for a delay period of between 20-30 seconds.
The monitor does not sound an alarm/alert if it detects ASY, VFIB, VTAC, or APNEA from the ECG waveform and the patient is walking (or experiencing a similar motion that, unlike falling or convulsing, does not result in an immediate alarm/alert). The monitor immediately sounds an alarm during both the presence and absence of these conditions if it detects that the patient is falling, has fell and remains on the ground for more than 10 seconds, or is having a Grand-mal seizure or similar convulsion. These alarm criteria are similar to those described in the heuristic rules, above.
Threshold Alarms/Alerts
Threshold alarms are generated by comparing vital signs measured by the body-worn monitor to fixed values that are either preprogrammed or calculated in real time. These threshold values are separated, as described below, into both outer limits (OL) and inner limits (IL). The values for OL are separated into an upper outer limit (UOL) and a lower outer limit (LOL). Default values for both UOL and LOL are typically preprogrammed into the body-worn monitor during manufacturing, and can be adjusted by a medical professional once the monitor is deployed. Table 2, below, lists typical default values corresponding to each vital sign for both UOL and LOL.
Values for IL are typically determined directly from the patient's vital signs. These values are separated into an upper inner limit (UIL) and a lower inner limit (LIL), and are calculated from the UOL and LOL, an upper inner value (UIV), and a lower inner value (LIV). The UIV and LIV can either be preprogrammed parameters (similar to the UOL and LOL, described above), or can be calculated directly from the patient's vital signs using a simple statistical process described below:
UIL=UIV+(UOL−UIV)/3
In a preferred embodiment the monitor only sounds an alarm/alert when the vital sign of issue surpasses the UOL/LOL and the UIL/LIL for a predetermined time period. Typically, the time periods for the UOL/LOL are shorter than those for the UIL/LIL, as alarm limits corresponding to these extremities represent a relatively large deviation for normal values of the patient's vital signs, and are therefore considered to be more severe. Typically the delay time periods for alarms/alerts associated with all vital signs (other than temperature, which tends to be significantly less labile) are 10 s for the UOL/LOL, and 120-180 s for the UIL/LIL. For temperature, the delay time period for the UOL/LOL is typically 600 s, and the delay time period for the UIL/LIL is typically 300 s.
Other embodiments are also possible for the threshold alarms/alerts. For example, the body-worn monitor can sound alarms having different tones and durations depending if the vital sign exceeds the UOL/LOL or UIL/LIL. Similarly, the tone can be escalated (in terms of acoustic frequency, alarm ‘beeps’ per second, and/or volume) depending on how long, and by how much, these thresholds are exceeded. Alarms may also sound due to failure of hardware within the body-worn monitor, or if the monitor detects that one of the sensors (e.g. optical sensor, ECG electrodes) becomes detached from the patient.
The first module 94 corresponds to a resting patient. In this state, the patient generates ECG, PPG, and ACC waveforms similar to those shown in
Method for Displaying Alarms/Alerts Using Graphical User Interfaces
Graphical user interfaces (GUI) operating on both the body-worn module and the remote monitor can render graphical icons that clearly identify the above-described patient activity states.
The patient view 106 is designed to give a medical professional, such as a nurse or doctor, a quick, easy-to-understand status of all the patients of all the patients in the specific hospital area. In a single glance the medical professional can determine their patients' vital signs, measured continuously by the body-worn monitor, along with their activity state and alarm status. The view 106 features a separate area 108 corresponding to each patient. Each area 108 includes text fields describing the name of the patient and supervising clinician; numbers associated with the patient's bed, room, and body-worn monitor; and the type of alarm generated from the patient. Graphical icons, similar to those shown in
The medical professional view 126 is designed to have a look and feel similar to each area 108 shown in
Algorithms for Determining Patient Motion, Posture, Arm Height, Activity Level and the Effect of these Properties on Blood Pressure
Described below is an algorithm for using the three accelerometers featured in the above-described body-worn monitor to calculate a patient's motion, posture, arm height, activity level. Each of these parameters affects both blood pressure and PTT, and thus inclusion of them in an algorithm can improve the accuracy of these measurements, and consequently reduce false alarms/alerts associated with them.
Calculating Arm Height
To calculate a patient's arm height it is necessary to build a mathematical model representing the geometric orientation of the patient's arm, as detected with signals from the three accelerometers.
The algorithm for estimating a patient's motion and activity level begins with a calculation to determine their arm height. This is done using signals from accelerometers attached to the patient's bicep (i.e., with reference to
The algorithm determines a gravitational vector RGA at a later time by again sampling DC portions of ACC1-6. Once this is complete, the algorithm determines the angle □GA between the fixed arm vector RA and the gravitational vector RGA by calculating a dot product of the two vectors. As the patient moves their arm, signals measured by the two accelerometers vary, and are analyzed to determine a change in the gravitational vector RGA and, subsequently, a change in the angle □GA. The angle □GA can then be combined with an assumed, approximate length of the patient's arm (typically 0.8 m) to determine its height relative to a proximal joint, e.g. the elbow.
B=rBx{circumflex over (l)}+rByĴ+rBz{circumflex over (k)} (1)
At any given time, the gravitational vector RGB is determined from ACC waveforms (ACC1-3) using signals from the accelerometer 132b located near the patient's bicep, and is represented by equation (2) below:
GB[n]=yBx[n]{circumflex over (l)}+yBy[n]Ĵ+yBz[n]{circumflex over (k)} (2)
Specifically, the CPU in the wrist-worn transceiver receives digitized signals representing the DC portion of the ACC1-3 signals measured with accelerometer 132b, as represented by equation (3) below, where the parameter n is the value (having units of g's) sampled directly from the DC portion of the ACC waveform:
yBx[n]=yDC,Bicep,x[n]; yBy[n]=yDC,Bicep,y[n]; yBz[n]=yDC,Bicep,z[n] (3)
The cosine of the angle □GB separating the vector RB and the gravitational vector RGB is determined using equation (4):
The definition of the dot product of the two vectors RB and RGB is:
GB[n]·B=(yBx[n]×rBx)+(yBy[n]×rBy)+(yBz[n]×rBz) (5)
and the definitions of the norms or magnitude of the vectors RB and RGB are:
∥GB[n]∥=√{square root over ((yBx[n])2+(yBy[n])2+(yBz[n])2)} (6)
and
∥B∥=√{square root over ((rBx)2+(rBy)2+(rBz)2)} (7)
Using the norm values for these vectors and the angle □GB separating them, as defined in equation (4), the height of the patient's elbow relative to their shoulder joint, as characterized by the accelerometer on their chest (hE) is determined using equation (8), where the length of the upper arm is estimated as LB:
hE[n]=−LB×cos(θGB[n]) (8)
As is described in more detail below, equation (8) estimates the height of the patient's arm relative to their heart. And this, in turn, can be used to further improve the accuracy of PTT-based blood pressure measurements.
The height of the patient's wrist joint hW is calculated in a similar manner using DC components from the time-domain waveforms (ACC4-6) collected from the accelerometer 132a mounted within the wrist-worn transceiver. Specifically, the wrist vector RW is given by equation (9):
W=rWx{circumflex over (l)}+rWyĴ+rWz{circumflex over (k)} (9)
and the corresponding gravitational vector RGW is given by equation (10):
GW[n]=yWx[n]{circumflex over (l)}+yWy[n]Ĵ+yWz[n]{circumflex over (k)} (10)
The specific values used in equation (10) are measured directly from the accelerometer 132a; they are represented as n and have units of g's, as defined below:
yWx[n]=yDC,Wrist,x[n];yWy[n]=yDC,Wrist,y[n];yWz[n]=yDC,Wrist,z[n] (11)
The vectors RW and RGW described above are used to determine the cosine of the angle □GW separating them using equation (12):
The definition of the dot product between the vectors RW and RGW is:
GW[n]·W=(yWx[n]×rWx)+(yWy[n]×rWy)+(yWz[n]×rWz) (13)
and the definitions of the norm or magnitude of both the vectors RW and RGW are:
∥GW[n]∥=√{square root over ((yWx[n])2+(yWy[n])2+(yWz[n])2)} (14)
and
|W∥=√{square root over ((rWx)2+(rWy)2+(rWz)2)} (15)
The height of the patient's wrist hW can be calculated using the norm values described above in equations (14) and (15), the cosine value described in equation (12), and the height of the patient's elbow determined in equation (8):
hW[n]=hE[n]−LW×cos(θGW[n]) (16)
In summary, the algorithm can use digitized signals from the accelerometers mounted on the patient's bicep and wrist, along with equations (8) and (16), to accurately determine the patient's arm height and position. As described below, these parameters can then be used to correct the PTT and provide a blood pressure calibration, similar to the cuff-based indexing measurement described above, that can further improve the accuracy of this measurement.
Calculating the Influence of Arm Height on Blood Pressure
A patient's blood pressure, as measured near the brachial artery, will vary with their arm height due to hydrostatic forces and gravity. This relationship between arm height and blood pressure enables two measurements: 1) a blood pressure ‘correction factor’, determined from slight changes in the patient's arm height, can be calculated and used to improve accuracy of the base blood pressure measurement; and 2) the relationship between PTT and blood pressure can be determined (like it is currently done using the indexing measurement) by measuring PTT at different arm heights, and calculating the change in PTT corresponding to the resultant change in height-dependent blood pressure. Specifically, using equations (8) and (16) above, and (21) below, an algorithm can calculate a change in a patient's blood pressure (□BP) simply by using data from two accelerometers disposed on the wrist and bicep. The □BP can be used as the correction factor. Exact blood pressure values can be estimated directly from arm height using an initial blood pressure value (determined, e.g., using the cuff-based module during an initial indexing measurement), the relative change in arm height, and the correction factor. This measurement can be performed, for example, when the patient is first admitted to the hospital. PTT determined at different arm heights provides multiple data points, each corresponding to a unique pair of blood pressure values determined as described above. The change in PTT values (□PTT) corresponds to changes in arm height.
From these data, the algorithm can calculate for each patient how blood pressure changes with PTT, i.e. □BP/□PTT. This relationship relates to features of the patient's cardiovascular system, and will evolve over time due to changes, e.g., in the patient's arterial tone and vascular compliance. Accuracy of the body-worn monitor's blood pressure measurement can therefore be improved by periodically calculating □BP/□PTT. This is best done by: 1) combining a cuff-based initial indexing measurement to set baseline values for SYS, DIA, and MAP, and then determining □BP/□PTT as described above; and 2) continually calculating □BP/□PTT by using the patient's natural motion, or alternatively using well-defined motions (e.g., raising and lower the arm to specific positions) as prompted at specific times by monitor's user interface.
Going forward, the body-worn monitor measures PTT, and can use this value and the relationship determined from the above-described calibration to convert this to blood pressure. All future indexing measurements can be performed on command (e.g., using audio or visual instructions delivered by the wrist-worn transceiver) using changes in arm height, or as the patient naturally raises and lowers their arm as they move about the hospital.
To determine the relationship between PTT, arm height, and blood pressure, the algorithm running on the wrist-worn transceiver is derived from a standard linear model shown in equation (17):
Assuming a constant velocity of the arterial pulse along an arterial pathway (e.g., the pathway extending from the heart, through the arm, to the base of the thumb):
the linear PTT model described in equation (17) becomes:
Equation (19) can be solved using piecewise integration along the upper 137 and lower 136 segments of the arm to yield the following equation for height-dependent PTT:
From equation (20) it is possible to determine a relative pressure change Prel induced in a cNIBP measurement using the height of the patient's wrist (hW) and elbow (hE):
As described above, Prel can be used to both calibrate the cNIBP measurement deployed by the body-worn monitor, or supply a height-dependent correction factor that reduces or eliminates the effect of posture and arm height on a PTT-based blood pressure measurement.
Calculating a Patient's Posture
As described above in Tables 1A-C, a patient's posture can influence how the above-described system generates alarms/alerts. The body-worn monitor can determine a patient's posture using time-dependent ACC waveforms continuously generated from the three patient-worn accelerometers, as shown in
The first step in this procedure is to identify alignment of CV in the chest accelerometer coordinate space. This can be determined in either of two approaches. In the first approach, CV is assumed based on a typical alignment of the body-worn monitor relative to the patient. During manufacturing, these parameters are then preprogrammed into firmware operating on the wrist-worn transceiver. In this procedure it is assumed that accelerometers within the body-worn monitor are applied to each patient with essentially the same configuration. In the second approach, CV is identified on a patient-specific basis. Here, an algorithm operating on the wrist-worn transceiver prompts the patient (using, e.g., video instruction operating on the display, or audio instructions transmitted through the speaker) to assume a known position with respect to gravity (e.g., standing up with arms pointed straight down). The algorithm then calculates CV from DC values corresponding to the x, y, and z axes of the chest accelerometer while the patient is in this position. This case, however, still requires knowledge of which arm (left or right) the monitor is worn on, as the chest accelerometer coordinate space can be rotated by 180 degrees depending on this orientation. A medical professional applying the monitor can enter this information using the GUI, described above. This potential for dual-arm attachment requires a set of two pre-determined vertical and normal vectors which are interchangeable depending on the monitor's location. Instead of manually entering this information, the arm on which the monitor is worn can be easily determined following attachment using measured values from the chest accelerometer values, with the assumption that CV is not orthogonal to the gravity vector.
The second step in the procedure is to identify the alignment of CN in the chest accelerometer coordinate space. The monitor can determine this vector, similar to the way it determines CV, with one of two approaches. In the first approach the monitor assumes a typical alignment of the chest-worn accelerometer on the patient. In the second approach, the alignment is identified by prompting the patient to assume a known position with respect to gravity. The monitor then calculates CN from the DC values of the time-dependent ACC waveform.
The third step in the procedure is to identify the alignment of CH in the chest accelerometer coordinate space. This vector is typically determined from the vector cross product of CV and CN, or it can be assumed based on the typical alignment of the accelerometer on the patient, as described above.
The derivation of this algorithm is as follows. Based on either an assumed orientation or a patient-specific calibration procedure described above, the alignment of CV in the chest accelerometer coordinate space is given by:
CV=rCVx{circumflex over (l)}+rCVyĴ+rCVz{circumflex over (k)} (22)
At any given moment, G is constructed from DC values of the ACC waveform from the chest accelerometer along the x, y, and z axes:
G[n]=yCx[n]{circumflex over (l)}+yCy[n]Ĵ+yCz[n]{circumflex over (k)} (23)
Equation (24) shows specific components of the ACC waveform used for this calculation:
yCx[n]=yDC,chest,x[n];yCy[n]=yDC,chest,y[n];yCz[n]=yDC,chest,z[n] (24)
The angle between CV and G is given by equation (25):
where the dot product of the two vectors is defined as:
G[n]·CV=(yCx[n]×rCVx)+(yCy[n]×rCVy)+(yCz[n]×rCVz) (26)
The definition of the norms of G and CV are given by equations (27) and (28):
∥G[n]∥=√{square root over ((yCx[n])2+(yCy[n])2+(yCz[n])2)} (27)
∥CV∥=√{square root over ((rCVx)2+(rCVy)2+(rCVz)2)} (28)
As shown in equation (29), the monitor compares the vertical angle θVG to a threshold angle to determine whether the patient is vertical (i.e. standing upright) or lying down:
if θVG≤45° then Torso State=0, the patient is upright (29)
If the condition in equation (29) is met the patient is assumed to be upright, and their torso state, which is a numerical value equated to the patient's posture, is equal to 0. The torso state is processed by the body-worn monitor to indicate, e.g., a specific icon corresponding to this state, such as icon 105a in
The angle θNG between CN and G determines if the patient is lying in the supine position (chest up), prone position (chest down), or on their side. Based on either an assumed orientation or a patient-specific calibration procedure, as described above, the alignment of CN is given by equation (30), where i, j, k represent the unit vectors of the x, y, and z axes of the chest accelerometer coordinate space respectively:
CN=rCNx{circumflex over (l)}+rCNyĴ+rCNz{circumflex over (k)} (30)
The angle between CN and G determined from DC values extracted from the chest accelerometer ACC waveform is given by equation (31):
The body-worn monitor determines the normal angle θNG and then compares it to a set of predetermined threshold angles to determine which position the patient is lying in, as shown in equation (32):
if θNG≤35° then Torso State=1, the patient is supine
if θNG≥135° then Torso State=2, the patient is prone (32)
Icons corresponding to these torso states are shown, for example, as icons 105h and 105g in
The alignment of CH is determined using either an assumed orientation, or from the vector cross-product of CV and CN as given by equation (33), where i, j, k represent the unit vectors of the x, y, and z axes of the accelerometer coordinate space respectively. Note that the orientation of the calculated vector is dependent on the order of the vectors in the operation. The order below defines the horizontal axis as positive towards the right side of the patient's body.
CH=rCVx{circumflex over (l)}+rCVyĴ+rCVz{circumflex over (k)}=CV×CN (33)
The angle θHG between CH and G is determined using equation (34):
The monitor compares this angle to a set of predetermined threshold angles to determine if the patient is lying on their right or left side, as given by equation (35):
if θHG≥90° then Torso State=3, the patient is on their right side
if θNG<90° then Torso State=4, the patient is on their left side (35)
Table 4 describes each of the above-described postures, along with a corresponding numerical torso state used to render, e.g., a particular icon:
Calculating a Patient's Activity
An algorithm can process information generated by the accelerometers described above to determine a patient's specific activity (e.g., walking, resting, convulsing), which is then used to reduce the occurrence of false alarms. This classification is done using a ‘logistic regression model classifier’, which is a type of classifier that processes continuous data values and maps them to an output that lies on the interval between 0 and 1. A classification ‘threshold’ is then set as a fractional value within this interval. If the model output is greater than or equal to this threshold, the classification is declared ‘true’, and a specific activity state can be assumed for the patient. If the model output falls below the threshold, then the specific activity is assumed not to take place.
This type of classification model offers several advantages. First, it provides the ability to combine multiple input variables into a single model, and map them to a single probability ranging between 0 and 1. Second, the threshold that allows the maximum true positive outcomes and the minimum false positive outcomes can be easily determined from a ROC curve, which in turn can be determined using empirical experimentation and data. Third, this technique requires minimal computation.
The formula for the logistic regression model is given by equation (36) and is used to determine the outcome, P, for a set of buffered data:
The logit variable z is defined in terms of a series of predictors (xi), each affected by a specific type of activity, and determined by the three accelerometers worn by the patient, as shown in equation (37):
z=b0+b1x1+b2x2+ . . . +bmxm (37)
In this model, the regression coefficients (bi, i=0, 1, . . . , m) and the threshold (Pth) used in the patient motion classifier and signal corruption classifiers are determined empirically from data collected on actual subjects. The classifier results in a positive outcome as given in equation (38) if the logistic model output, P, is greater than the predetermined threshold, Pth:
If P≥Pth then Classifier State=1 (38)
x10
The predictor variables described in Table 5 are typically determined from ACC signals generated by accelerometers deployed in locations that are most affected by patient motion. Such accelerometers are typically mounted directly on the wrist-worn transceiver, and on the bulkhead connector attached to the patient's arm. The normalized signal power (x1) for the AC components (yW,i, i=x,y,z) calculated from the ACC is shown in equation (39), where Fs denotes the signal sampling frequency, N is the size of the data buffer, and xnorm is a predetermined power value:
The average arm angle predictor value (x2) was determined using equation (40):
Note that, for this predictor value, it is unnecessary to explicitly determine the angle □GW using an arccosine function, and the readily available cosine value calculated in equation (12) acts as a surrogate parameter indicating the mean arm angle. The predictor value indicating the standard deviation of the arm angle (x3) was determined using equation (41) using the same assumptions for the angle □GW as described above:
The remaining predictor variables (x4-x10) are determined from the frequency content of the patient's motion, determined from the power spectrum of the time-dependent accelerometer signals, as indicated in
XW[m]=am+ibm (42)
Once the FFT is determined from the entire time-domain ACC waveform, the fractional power in the designated frequency band is given by equation (43), which is based on Parseval's theorem. The term mStart refers to the FFT coefficient index at the start of the frequency band of interest, and the term mEnd refers to the FFT coefficient index at the end of the frequency band of interest:
Finally, the formula for the total signal power, PT, is given in equation (44):
As described above, to accurately estimate a patient's activity level, predictor values x1-x10 defined above are measured from a variety of subjects selected from a range of demographic criteria (e.g., age, gender, height, weight), and then processed using pre-determined regression coefficients (bj) to calculate a logit variable (defined in equation (37)) and the corresponding probability outcome (defined in equation (36)). A threshold value is then determined empirically from an ROC curve. The classification is declared true if the model output is greater than or equal to the threshold value. During an actual measurement, an accelerometer signal is measured and then processed as described above to determine the predictor values. These parameters are used to determine the logit and corresponding probability, which is then compared to a threshold value to estimate the patient's activity level.
ROC curves similar to those shown in
Hardware System for Body-Worn Monitor
The body-worn monitor 10 features a wrist-worn transceiver 272, described in more detail in
To determine ACC waveforms the body-worn monitor 10 features three separate accelerometers located at different portions on the patient's arm. The first accelerometer is surface-mounted on a circuit board in the wrist-worn transceiver 272 and measures signals associated with movement of the patient's wrist. The second accelerometer is included in a small bulkhead portion 296 included along the span of the cable 286. During a measurement, a small piece of disposable tape, similar in size to a conventional bandaid, affixes the bulkhead portion 296 to the patient's arm. In this way the bulkhead portion 296 serves two purposes: 1) it measures a time-dependent ACC waveform from the mid-portion of the patient's arm, thereby allowing their posture and arm height to be determined as described in detail above; and 2) it secures the cable 286 to the patient's arm to increase comfort and performance of the body-worn monitor 10, particularly when the patient is ambulatory.
The cuff-based module 285 features a pneumatic system 276 that includes a pump, valve, pressure fittings, pressure sensor, analog-to-digital converter, microcontroller, and rechargeable battery. During an indexing measurement, it inflates a disposable cuff 284 and performs two measurements according to the composite technique: 1) it performs an inflation-based measurement of oscillometry to determine values for SYS, DIA, and MAP; and 2) it determines a patient-specific relationship between PTT and MAP. These measurements are performed according to the composite technique, and are described in detail in the above-referenced patent application entitled: ‘VITAL SIGN MONITOR FOR MEASURING BLOOD PRESSURE USING OPTICAL, ELECTRICAL, AND PRESSURE WAVEFORMS’ (U.S. Ser. No. 12/138,194; filed Jun. 12, 2008), the contents of which have been previously incorporated herein by reference.
The cuff 284 within the cuff-based pneumatic system 285 is typically disposable and features an internal, airtight bladder that wraps around the patient's bicep to deliver a uniform pressure field. During the indexing measurement, pressure values are digitized by the internal analog-to-digital converter, and sent through a cable 286 according to the CAN protocol, along with SYS, DIA, and MAP blood pressures, to the wrist-worn transceiver 272 for processing as described above. Once the cuff-based measurement is complete, the cuff-based module 285 is removed from the patient's arm and the cable 286 is disconnected from the wrist-worn transceiver 272. cNIBP is then determined using PTT, as described in detail above.
To determine an ECG, the body-worn monitor 10 features a small-scale, three-lead ECG circuit integrated directly into a bulkhead 274 that terminates an ECG cable 282. The ECG circuit features an integrated circuit that collects electrical signals from three chest-worn ECG electrodes 278a-c connected through cables 280a-c. The ECG electrodes 278a-c are typically disposed in a conventional ‘Einthoven's Triangle’ configuration which is a triangle-like orientation of the electrodes 278a-c on the patient's chest that features three unique ECG vectors. From these electrical signals the ECG circuit determines up to three ECG waveforms, which are digitized using an analog-to-digital converter mounted proximal to the ECG circuit, and sent through a five-wire cable 282 to the wrist-worn transceiver 272 according to the CAN protocol. There, the ECG is processed with the PPG to determine the patient's blood pressure. Heart rate and respiratory rate are determined directly from the ECG waveform using known algorithms, such as those described in the following reference, the contents of which are incorporated herein by reference: ‘ECG Beat Detection Using Filter Banks’, Afonso et al., IEEE Trans. Biomed Eng., 46:192-202 (1999). The cable bulkhead 274 also includes an accelerometer that measures motion associated with the patient's chest as described above.
There are several advantages of digitizing ECG and ACC waveforms prior to transmitting them through the cable 282. First, a single transmission line in the cable 282 can transmit multiple digital waveforms, each generated by different sensors. This includes multiple ECG waveforms (corresponding, e.g., to vectors associated with three, five, and twelve-lead ECG systems) from the ECG circuit mounted in the bulkhead 274, along with waveforms associated with the x, y, and z axes of accelerometers mounted in the bulkheads 275, 296. Limiting the transmission line to a single cable reduces the number of wires attached to the patient, thereby decreasing the weight and cable-related clutter of the body-worn monitor. Second, cable motion induced by an ambulatory patient can change the electrical properties (e.g. electrical impendence) of its internal wires. This, in turn, can add noise to an analog signal and ultimately the vital sign calculated from it. A digital signal, in contrast, is relatively immune to such motion-induced artifacts.
More sophisticated ECG circuits can plug into the wrist-worn transceiver to replace the three-lead system shown in
As described above, the transceiver 272 features three CAN connectors 204a-c on the side of its upper portion, each which supports the CAN protocol and wiring schematics, and relays digitized data to the internal CPU. Digital signals that pass through the CAN connectors include a header that indicates the specific signal (e.g. ECG, ACC, or pressure waveform from the cuff-based module) and the sensor from which the signal originated. This allows the CPU to easily interpret signals that arrive through the CAN connectors 204a-c, and means that these connectors are not associated with a specific cable. Any cable connecting to the transceiver can be plugged into any connector 204a-c. As shown in
The second CAN connector 204b shown in
The final CAN connector 204c can be used for an ancillary device, e.g. a glucometer, infusion pump, body-worn insulin pump, ventilator, or end-tidal CO2 delivery system. As described above, digital information generated by these systems will include a header that indicates their origin so that the CPU can process them accordingly.
The transceiver includes a speaker 201 that allows a medical professional to communicate with the patient using a voice over Internet protocol (VOIP). For example, using the speaker 101 the medical professional could query the patient from a central nursing station or mobile phone connected to a wireless, Internet-based network within the hospital. Or the medical professional could wear a separate transceiver similar to the shown in
In addition to those methods described above, a number of additional methods can be used to calculate blood pressure from the optical and electrical waveforms. These are described in the following co-pending patent applications, the contents of which are incorporated herein by reference: 1) CUFFLESS BLOOD-PRESSURE MONITOR AND ACCOMPANYING WIRELESS, INTERNET-BASED SYSTEM (U.S. Ser. No. 10/709,015; filed Apr. 7, 2004); 2) CUFFLESS SYSTEM FOR MEASURING BLOOD PRESSURE (U.S. Ser. No. 10/709,014; filed Apr. 7, 2004); 3) CUFFLESS BLOOD PRESSURE MONITOR AND ACCOMPANYING WEB SERVICES INTERFACE (U.S. Ser. No. 10/810,237; filed Mar. 26, 2004); 4) CUFFLESS BLOOD PRESSURE MONITOR AND ACCOMPANYING WIRELESS MOBILE DEVICE (U.S. Ser. No. 10/967,511; filed Oct. 18, 2004); 5) BLOOD PRESSURE MONITORING DEVICE FEATURING A CALIBRATION-BASED ANALYSIS (U.S. Ser. No. 10/967,610; filed Oct. 18, 2004); 6) PERSONAL COMPUTER-BASED VITAL SIGN MONITOR (U.S. Ser. No. 10/906,342; filed Feb. 15, 2005); 7) PATCH SENSOR FOR MEASURING BLOOD PRESSURE WITHOUT A CUFF (U.S. Ser. No. 10/906,315; filed Feb. 14, 2005); 8) PATCH SENSOR FOR MEASURING VITAL SIGNS (U.S. Ser. No. 11/160,957; filed Jul. 18, 2005); 9) WIRELESS, INTERNET-BASED SYSTEM FOR MEASURING VITAL SIGNS FROM A PLURALITY OF PATIENTS IN A HOSPITAL OR MEDICAL CLINIC (U.S. Ser. No. 11/162,719; filed Sep. 9, 2005); 10) HAND-HELD MONITOR FOR MEASURING VITAL SIGNS (U.S. Ser. No. 11/162,742; filed Sep. 21, 2005); 11) CHEST STRAP FOR MEASURING VITAL SIGNS (U.S. Ser. No. 11/306,243; filed Dec. 20, 2005); 12) SYSTEM FOR MEASURING VITAL SIGNS USING AN OPTICAL MODULE FEATURING A GREEN LIGHT SOURCE (U.S. Ser. No. 11/307,375; filed Feb. 3, 2006); 13) BILATERAL DEVICE, SYSTEM AND METHOD FOR MONITORING VITAL SIGNS (U.S. Ser. No. 11/420,281; filed May 25, 2006); 14) SYSTEM FOR MEASURING VITAL SIGNS USING BILATERAL PULSE TRANSIT TIME (U.S. Ser. No. 11/420,652; filed May 26, 2006); 15) BLOOD PRESSURE MONITOR (U.S. Ser. No. 11/530,076; filed Sep. 8, 2006); 16) TWO-PART PATCH SENSOR FOR MONITORING VITAL SIGNS (U.S. Ser. No. 11/558,538; filed Nov. 10, 2006); and, 17) MONITOR FOR MEASURING VITAL SIGNS AND RENDERING VIDEO IMAGES (U.S. Ser. No. 11/682,177; filed Mar. 5, 2007).
Other embodiments are also within the scope of the invention. For example, other techniques, such as conventional oscillometry measured during deflation, can be used to determine SYS for the above-described algorithms.
Still other embodiments are within the scope of the following claims.
This application is a continuation of U.S. patent application Ser. No. 15/905,744, filed Feb. 26, 2018, now U.S. Pat. No. 10,987,004, which is a continuation of U.S. patent application Ser. No. 15/431,459, filed Feb. 13, 2017, now U.S. Pat. No. 9,901,261, which is a continuation of U.S. patent application Ser. No. 14/738,910, filed Jun. 14, 2015, now U.S. Pat. No. 9,566,007, which is a continuation of U.S. patent application Ser. No. 14/090,433, filed Nov. 26, 2013, now U.S. Pat. No. 9,055,928, which is a continuation of U.S. patent application Ser. No. 13/432,976, filed Mar. 28, 2012, now U.S. Pat. No. 8,594,776, which is a continuation of U.S. patent application Ser. No. 12/469,182, filed May 20, 2009, now U.S. Pat. No. 8,180,440, all of which are hereby incorporated in its entirety including all tables, figures and claims.
Number | Name | Date | Kind |
---|---|---|---|
4086916 | Freeman et al. | May 1978 | A |
4263918 | Swearingen et al. | Apr 1981 | A |
4270547 | Steffen et al. | Jun 1981 | A |
4305400 | Logan | Dec 1981 | A |
4577639 | Simon et al. | Mar 1986 | A |
4582068 | Phillipps et al. | Apr 1986 | A |
4653498 | New, Jr. et al. | Mar 1987 | A |
4710164 | Levin et al. | Dec 1987 | A |
4722351 | Phillipps et al. | Feb 1988 | A |
4802486 | Goodman et al. | Feb 1989 | A |
4807638 | Sramek | Feb 1989 | A |
4905697 | Heggs et al. | Mar 1990 | A |
5025791 | Niwa | Jun 1991 | A |
5140990 | Jones et al. | Aug 1992 | A |
5190038 | Polson et al. | Mar 1993 | A |
5197489 | Conlan | Mar 1993 | A |
5224928 | Sibalis et al. | Jul 1993 | A |
5247931 | Norwood | Sep 1993 | A |
5289824 | Mills et al. | Mar 1994 | A |
5316008 | Suga et al. | May 1994 | A |
5339818 | Baker et al. | Aug 1994 | A |
5448991 | Polson et al. | Sep 1995 | A |
5465082 | Chaco | Nov 1995 | A |
5482036 | Diab et al. | Jan 1996 | A |
5485838 | Ukawa et al. | Jan 1996 | A |
5490505 | Diab et al. | Feb 1996 | A |
5515858 | Myllymaki | May 1996 | A |
5517988 | Gerhard | May 1996 | A |
5524637 | Erickson | Jun 1996 | A |
5549650 | Bornzin et al. | Aug 1996 | A |
5575284 | Athan et al. | Nov 1996 | A |
5588427 | Tien | Dec 1996 | A |
5593431 | Sheldon | Jan 1997 | A |
5632272 | Diab et al. | May 1997 | A |
5645060 | Yorkey | Jul 1997 | A |
5649543 | Hosaka et al. | Jul 1997 | A |
5680870 | Hood, Jr. et al. | Oct 1997 | A |
5685299 | Diab et al. | Nov 1997 | A |
5709205 | Bukta | Jan 1998 | A |
5743856 | Oka et al. | Apr 1998 | A |
5766131 | Kondo et al. | Jun 1998 | A |
5769785 | Diab et al. | Jun 1998 | A |
5800349 | Isaacson et al. | Sep 1998 | A |
5820550 | Polson et al. | Oct 1998 | A |
5848373 | Delorme et al. | Dec 1998 | A |
5853370 | Chance et al. | Dec 1998 | A |
5857975 | Golub | Jan 1999 | A |
5865755 | Golub | Feb 1999 | A |
5865756 | Peel, III | Feb 1999 | A |
5873834 | Yanagi et al. | Feb 1999 | A |
5876353 | Riff | Mar 1999 | A |
5895359 | Peel, III | Apr 1999 | A |
5899855 | Brown | May 1999 | A |
5913827 | Gorman | Jun 1999 | A |
5919141 | Money et al. | Jul 1999 | A |
5941836 | Friedman | Aug 1999 | A |
5964701 | Asada et al. | Oct 1999 | A |
5964720 | Pelz | Oct 1999 | A |
5971930 | Elghazzawi | Oct 1999 | A |
6002952 | Diab et al. | Dec 1999 | A |
6011985 | Athan et al. | Jan 2000 | A |
6018673 | Chin et al. | Jan 2000 | A |
6036642 | Diab et al. | Mar 2000 | A |
6041783 | Gruenke | Mar 2000 | A |
6057758 | Dempsey et al. | May 2000 | A |
6067462 | Diab et al. | May 2000 | A |
6081735 | Diab et al. | Jun 2000 | A |
6081742 | Amano et al. | Jun 2000 | A |
6094592 | Yorkey et al. | Jul 2000 | A |
6117077 | Del Mar et al. | Sep 2000 | A |
6129686 | Friedman | Oct 2000 | A |
6157850 | Diab et al. | Dec 2000 | A |
6159147 | Lichter et al. | Dec 2000 | A |
6160478 | Jacobsen et al. | Dec 2000 | A |
6168569 | McEwen et al. | Jan 2001 | B1 |
6198394 | Jacobsen et al. | Mar 2001 | B1 |
6198951 | Kosuda et al. | Mar 2001 | B1 |
6199550 | Wiesmann et al. | Mar 2001 | B1 |
6206830 | Diab et al. | Mar 2001 | B1 |
6236872 | Diab et al. | May 2001 | B1 |
6251080 | Henkin et al. | Jun 2001 | B1 |
6261247 | Ishikawa et al. | Jul 2001 | B1 |
6262769 | Anderson et al. | Jul 2001 | B1 |
6263222 | Diab et al. | Jul 2001 | B1 |
6287262 | Amano et al. | Sep 2001 | B1 |
6334065 | Al-Ali et al. | Dec 2001 | B1 |
6371921 | Caro et al. | Apr 2002 | B1 |
6388240 | Schulz et al. | May 2002 | B2 |
RE37852 | Aso et al. | Sep 2002 | E |
6443890 | Schulze et al. | Sep 2002 | B1 |
6480729 | Stone | Nov 2002 | B2 |
6491647 | Bridger et al. | Dec 2002 | B1 |
6503206 | Li et al. | Jan 2003 | B1 |
6516289 | David | Feb 2003 | B2 |
6526310 | Carter et al. | Feb 2003 | B1 |
6527729 | Turcott | Mar 2003 | B1 |
6533729 | Khair et al. | Mar 2003 | B1 |
6541756 | Schulz et al. | Apr 2003 | B2 |
6544173 | West et al. | Apr 2003 | B2 |
6544174 | West et al. | Apr 2003 | B2 |
6551252 | Sackner et al. | Apr 2003 | B2 |
6584336 | Ali et al. | Jun 2003 | B1 |
6589170 | Flach et al. | Jul 2003 | B1 |
6595929 | Stivoric et al. | Jul 2003 | B2 |
6605038 | Teller et al. | Aug 2003 | B1 |
6606993 | Wiesmann et al. | Aug 2003 | B1 |
6616606 | Petersen et al. | Sep 2003 | B1 |
6650917 | Diab et al. | Nov 2003 | B2 |
6684090 | Ali et al. | Jan 2004 | B2 |
6694177 | Eggers et al. | Feb 2004 | B2 |
6699194 | Diab et al. | Mar 2004 | B1 |
6712769 | Freund et al. | Mar 2004 | B2 |
6732064 | Kadtke et al. | May 2004 | B1 |
6745060 | Diab et al. | Jun 2004 | B2 |
6770028 | Ali et al. | Aug 2004 | B1 |
6790178 | Mault et al. | Sep 2004 | B1 |
6811538 | Westbrook et al. | Nov 2004 | B2 |
6845256 | Chin et al. | Jan 2005 | B2 |
6850787 | Weber et al. | Feb 2005 | B2 |
6879850 | Kimball | Apr 2005 | B2 |
6893396 | Schulze et al. | May 2005 | B2 |
6934571 | Wiesmann et al. | Aug 2005 | B2 |
6947781 | Asada et al. | Sep 2005 | B2 |
6976958 | Quy | Dec 2005 | B2 |
6985078 | Suzuki et al. | Jan 2006 | B2 |
6997882 | Parker et al. | Feb 2006 | B1 |
7004907 | Banet et al. | Feb 2006 | B2 |
7020508 | Stivoric et al. | Mar 2006 | B2 |
7020578 | Sorensen et al. | Mar 2006 | B2 |
7029447 | Rantala | Apr 2006 | B2 |
7041060 | Flaherty et al. | May 2006 | B2 |
7048687 | Reuss et al. | May 2006 | B1 |
7079888 | Oung et al. | Jul 2006 | B2 |
7115824 | Lo | Oct 2006 | B2 |
7156809 | Quy | Jan 2007 | B2 |
7179228 | Banet | Feb 2007 | B2 |
7184809 | Sterling et al. | Feb 2007 | B1 |
7186966 | Al-Ali | Mar 2007 | B2 |
7194293 | Baker, Jr. | Mar 2007 | B2 |
7215984 | Diab et al. | May 2007 | B2 |
7215987 | Sterling et al. | May 2007 | B1 |
7225007 | Al-Ali et al. | May 2007 | B2 |
7241265 | Cummings et al. | Jul 2007 | B2 |
7257438 | Kinast | Aug 2007 | B2 |
7296312 | Menkedick et al. | Nov 2007 | B2 |
7299159 | Nanikashvili | Nov 2007 | B2 |
7301451 | Hastings | Nov 2007 | B2 |
7314451 | Halperin et al. | Jan 2008 | B2 |
7316653 | Sano et al. | Jan 2008 | B2 |
7351206 | Suzuki et al. | Apr 2008 | B2 |
7355512 | Al-Ali | Apr 2008 | B1 |
7373191 | Delonzer et al. | May 2008 | B2 |
7373912 | Self et al. | May 2008 | B2 |
7377794 | Al-Ali et al. | May 2008 | B2 |
7382247 | Welch et al. | Jun 2008 | B2 |
7383069 | Ruchti et al. | Jun 2008 | B2 |
7383070 | Diab et al. | Jun 2008 | B2 |
7384398 | Gagnadre et al. | Jun 2008 | B2 |
7400919 | Petersen et al. | Jul 2008 | B2 |
7420472 | Tran | Sep 2008 | B2 |
7427926 | Sinclair et al. | Sep 2008 | B2 |
7455643 | Li et al. | Nov 2008 | B1 |
7468036 | Rulkov et al. | Dec 2008 | B1 |
7477143 | Albert | Jan 2009 | B2 |
7479890 | Lehrman et al. | Jan 2009 | B2 |
7485095 | Shusterman | Feb 2009 | B2 |
7502643 | Farringdon et al. | Mar 2009 | B2 |
7508307 | Albert | Mar 2009 | B2 |
7509131 | Krumm et al. | Mar 2009 | B2 |
7509154 | Diab et al. | Mar 2009 | B2 |
7522035 | Albert | Apr 2009 | B2 |
7530949 | Al-Ali et al. | May 2009 | B2 |
7539532 | Tran | May 2009 | B2 |
7541939 | Zadesky et al. | Jun 2009 | B2 |
7542878 | Nanikashvili | Jun 2009 | B2 |
7586418 | Cuddihy et al. | Sep 2009 | B2 |
7598878 | Goldreich | Oct 2009 | B2 |
7602301 | Stirling et al. | Oct 2009 | B1 |
7616110 | Crump et al. | Nov 2009 | B2 |
7625344 | Brady et al. | Dec 2009 | B1 |
7628071 | Sasaki et al. | Dec 2009 | B2 |
7628730 | Watterson et al. | Dec 2009 | B1 |
7641614 | Asada et al. | Jan 2010 | B2 |
7648463 | Elhag et al. | Jan 2010 | B1 |
7656287 | Albert et al. | Feb 2010 | B2 |
7668588 | Kovacs | Feb 2010 | B2 |
7670295 | Sackner et al. | Mar 2010 | B2 |
7674230 | Reisfeld | Mar 2010 | B2 |
7674231 | McCombie et al. | Mar 2010 | B2 |
7678061 | Lee et al. | Mar 2010 | B2 |
7684954 | Shahabdeen et al. | Mar 2010 | B2 |
7689437 | Teller et al. | Mar 2010 | B1 |
7698101 | Alten et al. | Apr 2010 | B2 |
7698830 | Townsend et al. | Apr 2010 | B2 |
7698941 | Sasaki et al. | Apr 2010 | B2 |
7715984 | Ramakrishnan et al. | May 2010 | B2 |
7725147 | Li et al. | May 2010 | B2 |
7782189 | Spoonhower et al. | Aug 2010 | B2 |
7827011 | Devaul et al. | Nov 2010 | B2 |
7974689 | Volpe et al. | Jul 2011 | B2 |
7976480 | Grajales et al. | Jul 2011 | B2 |
7983933 | Karkanias et al. | Jul 2011 | B2 |
7993275 | Banet et al. | Aug 2011 | B2 |
8047998 | Kolluri et al. | Nov 2011 | B2 |
8082160 | Collins, Jr. et al. | Dec 2011 | B2 |
8137270 | Keenan et al. | Mar 2012 | B2 |
8167800 | Ouchi et al. | May 2012 | B2 |
8180440 | McCombie et al. | May 2012 | B2 |
8326420 | Skelton et al. | Dec 2012 | B2 |
8419649 | Banet et al. | Apr 2013 | B2 |
8449469 | Banet et al. | May 2013 | B2 |
8594776 | McCombie et al. | Nov 2013 | B2 |
8668643 | Kinast | Mar 2014 | B2 |
9055928 | McCombie et al. | Jun 2015 | B2 |
9149192 | Banet et al. | Oct 2015 | B2 |
9566007 | McCombie et al. | Feb 2017 | B2 |
9901261 | McCombie et al. | Feb 2018 | B2 |
10987004 | McCombie | Apr 2021 | B2 |
20010004234 | Petelenz et al. | Jun 2001 | A1 |
20010013826 | Ahmed et al. | Aug 2001 | A1 |
20010045395 | Kitaevich et al. | Nov 2001 | A1 |
20020013517 | West et al. | Jan 2002 | A1 |
20020032386 | Sackner et al. | Mar 2002 | A1 |
20020072859 | Kajimoto et al. | Jun 2002 | A1 |
20020151805 | Sugo et al. | Oct 2002 | A1 |
20020156354 | Larson | Oct 2002 | A1 |
20020170193 | Townsend et al. | Nov 2002 | A1 |
20020193671 | Ciurczak et al. | Dec 2002 | A1 |
20020193692 | Inukai et al. | Dec 2002 | A1 |
20020198679 | Victor et al. | Dec 2002 | A1 |
20030004420 | Narimatsu | Jan 2003 | A1 |
20030097046 | Sakamaki et al. | May 2003 | A1 |
20030130590 | Bui et al. | Jul 2003 | A1 |
20030135099 | Al-Ali | Jul 2003 | A1 |
20030153836 | Gagnadre et al. | Aug 2003 | A1 |
20030158699 | Townsend et al. | Aug 2003 | A1 |
20030167012 | Friedman et al. | Sep 2003 | A1 |
20030171662 | O'Conner et al. | Sep 2003 | A1 |
20030181815 | Ebner et al. | Sep 2003 | A1 |
20030208335 | Unuma et al. | Nov 2003 | A1 |
20040019288 | Kinast | Jan 2004 | A1 |
20040030261 | Rantala | Feb 2004 | A1 |
20040034293 | Kimball | Feb 2004 | A1 |
20040034294 | Kimball et al. | Feb 2004 | A1 |
20040054821 | Warren et al. | Mar 2004 | A1 |
20040073128 | Hatlestad et al. | Apr 2004 | A1 |
20040077934 | Massad | Apr 2004 | A1 |
20040077958 | Kato et al. | Apr 2004 | A1 |
20040111033 | Oung et al. | Jun 2004 | A1 |
20040133079 | Mazar et al. | Jul 2004 | A1 |
20040162466 | Quy | Aug 2004 | A1 |
20040162493 | Mills | Aug 2004 | A1 |
20040193063 | Kimura et al. | Sep 2004 | A1 |
20040225207 | Bae et al. | Nov 2004 | A1 |
20040267099 | McMahon et al. | Dec 2004 | A1 |
20050027205 | Tarassenko et al. | Feb 2005 | A1 |
20050043598 | Goode, Jr. et al. | Feb 2005 | A1 |
20050059870 | Aceti | Mar 2005 | A1 |
20050070773 | Chin et al. | Mar 2005 | A1 |
20050113107 | Meunier | May 2005 | A1 |
20050113703 | Farringdon et al. | May 2005 | A1 |
20050119586 | Coyle et al. | Jun 2005 | A1 |
20050124866 | Elaz et al. | Jun 2005 | A1 |
20050124903 | Roteliuk et al. | Jun 2005 | A1 |
20050149350 | Kerr et al. | Jul 2005 | A1 |
20050171444 | Ono et al. | Aug 2005 | A1 |
20050187796 | Rosenfeld et al. | Aug 2005 | A1 |
20050206518 | Welch et al. | Sep 2005 | A1 |
20050209511 | Heruth et al. | Sep 2005 | A1 |
20050216199 | Banet | Sep 2005 | A1 |
20050228296 | Banet | Oct 2005 | A1 |
20050228297 | Banet et al. | Oct 2005 | A1 |
20050228298 | Banet et al. | Oct 2005 | A1 |
20050228300 | Jaime et al. | Oct 2005 | A1 |
20050228301 | Banet et al. | Oct 2005 | A1 |
20050234317 | Kiani | Oct 2005 | A1 |
20050240087 | Keenan et al. | Oct 2005 | A1 |
20050261565 | Lane et al. | Nov 2005 | A1 |
20050261593 | Zhang et al. | Nov 2005 | A1 |
20050261598 | Banet et al. | Nov 2005 | A1 |
20050265267 | Hwang | Dec 2005 | A1 |
20050283088 | Bernstein | Dec 2005 | A1 |
20060009697 | Banet et al. | Jan 2006 | A1 |
20060009698 | Banet et al. | Jan 2006 | A1 |
20060036141 | Kamath et al. | Feb 2006 | A1 |
20060047215 | Newman et al. | Mar 2006 | A1 |
20060074321 | Kouchi et al. | Apr 2006 | A1 |
20060074322 | Nitzan | Apr 2006 | A1 |
20060084878 | Banet et al. | Apr 2006 | A1 |
20060122469 | Martel | Jun 2006 | A1 |
20060128263 | Baird | Jun 2006 | A1 |
20060142648 | Banet et al. | Jun 2006 | A1 |
20060155589 | Lane et al. | Jul 2006 | A1 |
20060178591 | Hempfling | Aug 2006 | A1 |
20060200029 | Evans et al. | Sep 2006 | A1 |
20060252999 | Devaul et al. | Nov 2006 | A1 |
20060265246 | Hoag | Nov 2006 | A1 |
20060270949 | Mathie et al. | Nov 2006 | A1 |
20060271404 | Brown | Nov 2006 | A1 |
20060281979 | Kim et al. | Dec 2006 | A1 |
20070010719 | Huster et al. | Jan 2007 | A1 |
20070015976 | Miesel et al. | Jan 2007 | A1 |
20070055163 | Asada et al. | Mar 2007 | A1 |
20070066910 | Inukai et al. | Mar 2007 | A1 |
20070071643 | Hall et al. | Mar 2007 | A1 |
20070094045 | Cobbs et al. | Apr 2007 | A1 |
20070118054 | Pinhas et al. | May 2007 | A1 |
20070118056 | Wang et al. | May 2007 | A1 |
20070129769 | Bourget et al. | Jun 2007 | A1 |
20070142715 | Banet et al. | Jun 2007 | A1 |
20070142730 | Laermer et al. | Jun 2007 | A1 |
20070156456 | McGillin et al. | Jul 2007 | A1 |
20070161912 | Zhang et al. | Jul 2007 | A1 |
20070167844 | Asada et al. | Jul 2007 | A1 |
20070185393 | Zhou et al. | Aug 2007 | A1 |
20070188323 | Sinclair et al. | Aug 2007 | A1 |
20070193834 | Pai et al. | Aug 2007 | A1 |
20070208233 | Kovacs | Sep 2007 | A1 |
20070232867 | Hansmann | Oct 2007 | A1 |
20070237719 | Jones et al. | Oct 2007 | A1 |
20070244376 | Wang | Oct 2007 | A1 |
20070250261 | Soehren | Oct 2007 | A1 |
20070252853 | Park et al. | Nov 2007 | A1 |
20070255116 | Mehta et al. | Nov 2007 | A1 |
20070260487 | Bartfeld et al. | Nov 2007 | A1 |
20070265533 | Tran | Nov 2007 | A1 |
20070265880 | Bartfeld et al. | Nov 2007 | A1 |
20070270671 | Gal | Nov 2007 | A1 |
20070276261 | Banet et al. | Nov 2007 | A1 |
20070282208 | Jacobs et al. | Dec 2007 | A1 |
20070287386 | Agrawal et al. | Dec 2007 | A1 |
20070293770 | Bour et al. | Dec 2007 | A1 |
20070293781 | Sims et al. | Dec 2007 | A1 |
20080004500 | Cazares et al. | Jan 2008 | A1 |
20080004507 | Williams, Jr. et al. | Jan 2008 | A1 |
20080004904 | Tran | Jan 2008 | A1 |
20080027341 | Sackner et al. | Jan 2008 | A1 |
20080039731 | McCombie et al. | Feb 2008 | A1 |
20080077026 | Banet et al. | Mar 2008 | A1 |
20080077027 | Allgeyer | Mar 2008 | A1 |
20080082001 | Hatlestad et al. | Apr 2008 | A1 |
20080082004 | Banet et al. | Apr 2008 | A1 |
20080101160 | Besson | May 2008 | A1 |
20080103405 | Banet et al. | May 2008 | A1 |
20080114220 | Banet et al. | May 2008 | A1 |
20080129513 | Bielas et al. | Jun 2008 | A1 |
20080132106 | Burnes et al. | Jun 2008 | A1 |
20080139955 | Hansmann et al. | Jun 2008 | A1 |
20080146887 | Rao et al. | Jun 2008 | A1 |
20080146892 | LeBoeuf et al. | Jun 2008 | A1 |
20080162496 | Postrel | Jul 2008 | A1 |
20080167535 | Stivoric et al. | Jul 2008 | A1 |
20080183053 | Borgos et al. | Jul 2008 | A1 |
20080194918 | Kulik et al. | Aug 2008 | A1 |
20080195735 | Hodges et al. | Aug 2008 | A1 |
20080204254 | Kazuno | Aug 2008 | A1 |
20080208013 | Zhang et al. | Aug 2008 | A1 |
20080208273 | Owen et al. | Aug 2008 | A1 |
20080214963 | Guillemaud et al. | Sep 2008 | A1 |
20080221399 | Zhou et al. | Sep 2008 | A1 |
20080221404 | Tso | Sep 2008 | A1 |
20080262362 | Kolluri et al. | Oct 2008 | A1 |
20080275349 | Halperin et al. | Nov 2008 | A1 |
20080281168 | Gibson et al. | Nov 2008 | A1 |
20080281310 | Dunning et al. | Nov 2008 | A1 |
20080287751 | Stivoric et al. | Nov 2008 | A1 |
20080294019 | Tran | Nov 2008 | A1 |
20080319282 | Tran | Dec 2008 | A1 |
20080319327 | Banet et al. | Dec 2008 | A1 |
20090018408 | Ouchi et al. | Jan 2009 | A1 |
20090018409 | Banet et al. | Jan 2009 | A1 |
20090018453 | Banet et al. | Jan 2009 | A1 |
20090040041 | Janetis et al. | Feb 2009 | A1 |
20090054752 | Jonnalagadda et al. | Feb 2009 | A1 |
20090058635 | LaLonde et al. | Mar 2009 | A1 |
20090062667 | Fayram et al. | Mar 2009 | A1 |
20090069642 | Gao et al. | Mar 2009 | A1 |
20090076363 | Bly et al. | Mar 2009 | A1 |
20090076397 | Libbus et al. | Mar 2009 | A1 |
20090076405 | Amurthur et al. | Mar 2009 | A1 |
20090082681 | Yokoyama et al. | Mar 2009 | A1 |
20090112281 | Miyazawa et al. | Apr 2009 | A1 |
20090112630 | Collins, Jr. et al. | Apr 2009 | A1 |
20090118590 | Teller et al. | May 2009 | A1 |
20090118626 | Moon et al. | May 2009 | A1 |
20090131759 | Sims et al. | May 2009 | A1 |
20090187085 | Pav | Jul 2009 | A1 |
20090192366 | Mensinger et al. | Jul 2009 | A1 |
20090198139 | Lewicke et al. | Aug 2009 | A1 |
20090221937 | Smith et al. | Sep 2009 | A1 |
20090222119 | Plahey et al. | Sep 2009 | A1 |
20090227877 | Tran | Sep 2009 | A1 |
20090233770 | Vincent et al. | Sep 2009 | A1 |
20090259113 | Liu et al. | Oct 2009 | A1 |
20090262074 | Nasiri et al. | Oct 2009 | A1 |
20090264712 | Baldus et al. | Oct 2009 | A1 |
20090287067 | Dorogusker et al. | Nov 2009 | A1 |
20090295541 | Roof | Dec 2009 | A1 |
20090306485 | Bell | Dec 2009 | A1 |
20090306487 | Crowe et al. | Dec 2009 | A1 |
20090306524 | Muhlsteff et al. | Dec 2009 | A1 |
20090312973 | Hatlestad et al. | Dec 2009 | A1 |
20090318779 | Tran | Dec 2009 | A1 |
20090322513 | Hwang et al. | Dec 2009 | A1 |
20100010380 | Panken et al. | Jan 2010 | A1 |
20100026510 | Kiani | Feb 2010 | A1 |
20100030034 | Schulhauser et al. | Feb 2010 | A1 |
20100030085 | Rojas Ojeda et al. | Feb 2010 | A1 |
20100056881 | Libbus et al. | Mar 2010 | A1 |
20100056886 | Hurtubise et al. | Mar 2010 | A1 |
20100113948 | Yang et al. | May 2010 | A1 |
20100125188 | Schilling et al. | May 2010 | A1 |
20100130811 | Leuthardt et al. | May 2010 | A1 |
20100160793 | Lee et al. | Jun 2010 | A1 |
20100160794 | Banet et al. | Jun 2010 | A1 |
20100160795 | Banet et al. | Jun 2010 | A1 |
20100160796 | Banet et al. | Jun 2010 | A1 |
20100160797 | Banet et al. | Jun 2010 | A1 |
20100160798 | Banet et al. | Jun 2010 | A1 |
20100168589 | Banet et al. | Jul 2010 | A1 |
20100210930 | Saylor | Aug 2010 | A1 |
20100217099 | LeBoeuf et al. | Aug 2010 | A1 |
20100222649 | Schoenberg | Sep 2010 | A1 |
20100234693 | Srinivasan et al. | Sep 2010 | A1 |
20100234695 | Morris | Sep 2010 | A1 |
20100234786 | Fulkerson et al. | Sep 2010 | A1 |
20100241011 | McCombie et al. | Sep 2010 | A1 |
20100280417 | Skelton et al. | Nov 2010 | A1 |
20100280440 | Skelton et al. | Nov 2010 | A1 |
20100298650 | Moon et al. | Nov 2010 | A1 |
20100298651 | Moon et al. | Nov 2010 | A1 |
20100298652 | McCombie et al. | Nov 2010 | A1 |
20100298653 | McCombie et al. | Nov 2010 | A1 |
20100298654 | McCombie et al. | Nov 2010 | A1 |
20100298655 | McCombie et al. | Nov 2010 | A1 |
20100298656 | McCombie et al. | Nov 2010 | A1 |
20100298657 | McCombie et al. | Nov 2010 | A1 |
20100298658 | McCombie et al. | Nov 2010 | A1 |
20100298659 | McCombie et al. | Nov 2010 | A1 |
20100298660 | McCombie et al. | Nov 2010 | A1 |
20100298661 | McCombie et al. | Nov 2010 | A1 |
20100312115 | Dentinger | Dec 2010 | A1 |
20100324384 | Moon et al. | Dec 2010 | A1 |
20100324385 | Moon et al. | Dec 2010 | A1 |
20100324386 | Moon et al. | Dec 2010 | A1 |
20100324387 | Moon et al. | Dec 2010 | A1 |
20100324388 | Moon et al. | Dec 2010 | A1 |
20100324389 | Moon et al. | Dec 2010 | A1 |
20100331640 | Medina | Dec 2010 | A1 |
20110066006 | Banet et al. | Mar 2011 | A1 |
20110066007 | Banet et al. | Mar 2011 | A1 |
20110066008 | Banet et al. | Mar 2011 | A1 |
20110066009 | Moon et al. | Mar 2011 | A1 |
20110066010 | Moon et al. | Mar 2011 | A1 |
20110066037 | Banet et al. | Mar 2011 | A1 |
20110066038 | Banet et al. | Mar 2011 | A1 |
20110066039 | Banet et al. | Mar 2011 | A1 |
20110066043 | Banet et al. | Mar 2011 | A1 |
20110066044 | Moon et al. | Mar 2011 | A1 |
20110066045 | Moon et al. | Mar 2011 | A1 |
20110066050 | Moon et al. | Mar 2011 | A1 |
20110066051 | Moon et al. | Mar 2011 | A1 |
20110066062 | Banet et al. | Mar 2011 | A1 |
20110070829 | Griffin et al. | Mar 2011 | A1 |
20110076942 | Taveau et al. | Mar 2011 | A1 |
20110093281 | Plummer et al. | Apr 2011 | A1 |
20110105862 | Gies et al. | May 2011 | A1 |
20110112442 | Meger | May 2011 | A1 |
20110144456 | Muhlsteff et al. | Jun 2011 | A1 |
20110152632 | Le Neel et al. | Jun 2011 | A1 |
20110178375 | Forster | Jul 2011 | A1 |
20110224498 | Banet et al. | Sep 2011 | A1 |
20110224499 | Banet et al. | Sep 2011 | A1 |
20110224500 | Banet et al. | Sep 2011 | A1 |
20110224506 | Moon et al. | Sep 2011 | A1 |
20110224507 | Banet et al. | Sep 2011 | A1 |
20110224508 | Moon | Sep 2011 | A1 |
20110224556 | Moon et al. | Sep 2011 | A1 |
20110224557 | Banet et al. | Sep 2011 | A1 |
20110224564 | Moon et al. | Sep 2011 | A1 |
20110257489 | Banet et al. | Oct 2011 | A1 |
20110257551 | Banet et al. | Oct 2011 | A1 |
20110257552 | Banet et al. | Oct 2011 | A1 |
20110257554 | Banet et al. | Oct 2011 | A1 |
20110257555 | Banet et al. | Oct 2011 | A1 |
20110275907 | Inciardi et al. | Nov 2011 | A1 |
20120065525 | Douniama et al. | Mar 2012 | A1 |
20120123232 | Najarian et al. | May 2012 | A1 |
Number | Date | Country |
---|---|---|
0443267 | Aug 1991 | EP |
2329250 | Mar 1999 | GB |
9932030 | Jul 1999 | WO |
2006005169 | Jan 2006 | WO |
2007024777 | Mar 2007 | WO |
2007143535 | Dec 2007 | WO |
2008110788 | Sep 2008 | WO |
2010135516 | Nov 2010 | WO |
2010135518 | Nov 2010 | WO |
2010148205 | Dec 2010 | WO |
2011032132 | Mar 2011 | WO |
2011034881 | Mar 2011 | WO |
2011082341 | Jul 2011 | WO |
2011112782 | Sep 2011 | WO |
2011133582 | Oct 2011 | WO |
Entry |
---|
International Search Report and Written Opinion issued in PCT/US2010/035554 dated Sep. 23, 2010 (25 pages). |
Supplemental European Search Report issued in EP 10778376 dated Jan. 31, 2013 (23 pages). |
First Exam Report issued by the India Patent and Trademark Office in 2712/MUMNP/2011 dated Aug. 27, 2018—Incl Engl lang transl (6 pages total). |
Non Final Office Action issued by the United States Patent and Trademark Office in U.S. Appl. No. 13/432,976 dated Dec. 14, 2012 (14 pages). |
Ahlstrom et al., Noninvasive investigation of blood pressure changes using the pulse wave transit time: a novel approach in the monitoring of hemodialysis patients. J Artif Organs. 2005;8(3):192-197. |
Allen et al., Classification of a known sequence of motions and postures from accelerometry data using adapted Gaussian mixture models. Physiol. Meas. 2006;27:935-951. |
Alves et al., CAN Protocol: A Laboratory Prototype for Fieldbus Applications. XIX IMEKO World Congress Fundamental and Applied Metrology Sep. 6-11, 2009, Lisbon, Portugal. 4 pages :454-457 ISBN 978-963-88410-0-1. |
Asada et al., Active Noise Cancellation Using MEMS Accelerometers for Motion-Tolerant Wearable Bio-Sensors. Proceedings of the 26th Annual International Conference of the IEEE EMBS. San Francisco, CA, USA. Sep. 1-5, 2004:2157-2160. |
Benefits of Digital Sensors. Gems Sensors. Feb. 14, 2008. http:/web.archive.org/web/20080214122230/http://www.sensorland.com/HowPage054.html. |
Bowers et al., Respiratory Rate Derived from Principal Component Analysis of Single Lead Electrocardiogram. Computers in Cardiology Conference Proceedings Sep. 2008;35:437-440. |
Bussmann et al., Measuring daily behavior using ambulatory accelerometry: The Activity Monitor. Behav Res Methods Instrum Comput. Aug. 2001;33(3):349-356. |
Clifford et al., Measuring Tilt with Low-g Accelerometers. Freescale Semiconductor, Inc., May 2005:8 pages. |
Cretikos et al., The Objective Medical Emergency Team Activation Criteria: a case-control study. Resuscitation Apr. 2007;73(1):62-72. |
De Scalzi et al., Relationship Between Systolic Time Intervals and Arterial Blood Pressure. Clin Cardiol. 1986;9:545-549. |
Drinnan et al., Relation between heart rate and pulse transit time during paced respiration. Physiol. Meas. Aug. 2001;22(3):425-432. |
Afonso et al., ECG Beat Detection Using Filter Banks. IEEE Trans Biomed Eng. Feb. 1999;46(2):192-202. |
Espina et al., Wireless Body Sensor Network for Continuous Cuff-less Blood Pressure Monitoring. Proceedings of the 3rd IEEE-EMBS. International Summer School and Symposium on Medical Devices and Biosensors. MIT, Boston, USA, Sep. 4-6, 2006:11-15. |
Fieselmann et al., Respiratory rate predicts cardiopulmonary arrest for internal medicine patients. J Gen Intern Med Jul. 1993; 8(7):354-360. |
Flash et al., The Coordination of Arm Movements: An Experimentally Confirmed Mathematical Model. J Neurosci. Jul. 1985;5(7):1688-1703. |
Gallagher, Comparison of Radial and Femoral Arterial Blood Pressure in Children after Cardiopulmonary Bypass. J Clin Monit. Jul. 1985;1(3):168-171. |
Goldhill et al., A physiologically-based early warning score for ward patients: the association between score and outcome. Anaesthesia Jun. 2005;60(6):547-553. |
Hung et al., Estimation of Respiratory Waveform Using an Accelerometer. 5th IEEE International Symposium on Biomedical Imaging: From Nano to Macro, May 14-17, 2008:1493-1496. |
Jackson, Digital Filter Design and Synthesis Using High-Level Modeling Tools. Virginia Polytechnic Institute and State University Masters of Science Thesis. Dec. 1999 (200 pages). |
Jin, A Respiration Monitoring System Based on a Tri-Axial Accelerometer and an Air-Coupled Microphone. Technische Universiteit Eindhoven, University of Technology. Master's Graduation Paper, Electrical Engineering Aug. 25, 2009:1-13. |
Karantonis et al., Implementation of a Real-Time Human Movement Classifier Using a Triaxial Accelerometer for Ambulatory Monitoring. IEEE Transactions on Information Technology in Biomedicine. Jan. 2006;10(1):156-167. |
Khambete et al., Movement artefact rejection in impedance pneumography using six strategically placed electrodes. Physiol. Meas. 2000;21 :79-88. |
Khan et al., Accelerometer Signal-based Human Activity Recognition Using Augmented Autoregressive Model Coefficients and Artificial w Neural Nets. 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. Aug. 20-24, 2008:5172-5175. |
Kim et al., Two Algorithms for Detecting Respiratory Rate from ECG Signal. IFMBE Proceedings 2007; 14(6) JC27:4069-4071. |
Klabunde, Mean Arterial Pressure. Cardiovascular Physiology Concepts. Mar. 8, 2007.http://web.archive.org/web/20070308182914/http://www.cvphysiology.com/Blood%20Pressure/BP006.htm (2 pages). |
Ma and Zhang, A Correlation Study on the Variabilities in Pulse Transit Time, Blood Pressure, and Heart Rate Recorded Simultaneously from Healthy Subjects. Conf Proc IEEE Eng Med Biol Soc. 2005;1:996-999. |
Mason, Signal Processing Methods for Non-Invasive Respiration Monitoring. Doctor of Philosophy dissertation submitted to Department of Engineering Science, University of Oxford 2002 (175 pages). |
Mathie et al., Classification of basic daily movements using a triaxial accelerometer. Med Biol Eng Comput. Sep. 2004;42(5):679-687. |
Mathie, Monitoring and Interpreting Human Movement Patterns using a Triaxial Accelerometer. Faculty of Engineering, The University of New South Wales, PhD Dissertation. Aug. 2003: part1 pp. 1-256. |
Mathie, Monitoring and Interpreting Human Movement Patterns using a Triaxial Accelerometer. Faculty of Engineering, The University of New South Wales, PhD Dissertation. Aug. 2003: part2 pp. 256-512. |
McKneely et al., Plug-and-Play and Network-Capable Medical Instrumentation and Database with a Complete Healthcare Technology Suite: MediCAN. Joint Workshop on High Confidence Medical Devices, Software, and Systems and Medical Device Plug-and-Play Interoperability. 2007:122-129. |
Montgomery et al., Lifeguard—A Personal Physiological Monitor For Extreme Environments. Conf Proc IEEE Eng Med Biol Soc. 2004;3:2192-2195. |
O'Haver, Peak Finding and Measurement, Version 1.6 Oct. 26, 2006. http://web.archive.org/web/20090205162604/http://terpconnect.umd.edu/-toh/spectrum/PeakFindingandMeasurement.htm. |
Otto et al., System Architecture of a Wireless Body Area Sensor Network for Ubiquitous Health Monitoring. Journal of Mobile Multimedia Jan. 10, 2006;1 (4):307-326. |
Packet Definition. The Linux Information Project Jan. 8, 2006, accessed online at: http://www.linfo.org/packet.html (2 bages). |
Park et al., An improved algorithm for respiration signal extraction from electrocardiogram measured by conductive textile electrodes using instantaneous frequency estimation. Med Bio Eng Comput 2008;46:147-158. |
Park et al., Direct Blood Pressure Measurements in Brachia! and Femoral Arteries in Children. Circulation Feb. 1970;41(2):231-237. |
PDF-Pro for iphone & ipod touch User Manual. ePapyrus Jul. 2009;1 :1-25, accessed online at: http://epapyrus.com/en/files/PDFPro%. |
Poon and Zhang, Cuff-Less and Noninvasive Measurements of Arterial Blood Pressure by Pulse Transit Time. Conf Proc IEEE Eng Med Biol Soc. 2005;6:5877-5880. |
Reddan et al., Intradialytic Blood Volume Monitoring in Ambulatory Hemodialysis Patients: A Randomized Trial. J Am Soc Nephrol. Jul. 2005;16(7):2162-2169. |
Reinvuo et al., Measurement of Respiratory Rate with High-Resolution Accelerometer and EMFit Pressure Sensor. Proceedings of the 2006 IEEE Sensors Applications Symposium Feb. 7-9, 2006:192-195. |
RS-232. Wikipedia Dec. 5, 2008, accessed online at: http: //web.archive.org/web/20081205160754/http:/len.wikipedia.org/wiki/RS-232 (1 page). |
Sendelbach and Funk, Alarm Fatigue: A Patient Safety Concern, AACN Advanced Critical Care, 2013, 24(4):3479-396. |
Seo et al., Performance Improvement of Pulse Oximetry-Based Respiration Detection by Selective Mode Bandpass Filtering. Ergonomics and Health Aspects of Work with Computers Lecture Notes in Computer Science, 2007;4566:300-308. |
Signal Strength. Oct. 6, 2008, accessed online at: http://en.wikipedia.org/wiki/Signal_strength. (3 pages). |
Soh et al., An investigation of respiration while wearing back belts. Applied Ergonomics 1997; 28(3):189-192. |
Subbe et al., Effect of introducing the Modified Early Warning score on clinical outcomes, cardiopulmonary arrests and intensive care utilization in acute medical admissions. Anaesthesia Aug. 2003;58(8):797-802. |
Talkowski, Quantifying Physical Activity in Community Dwelling Older Adults Using Accelerometry. University of Pittsburgh, PhD Dissertation. 2008:1-91. |
Thongpithoonrat et al., Networking and Plug-and-Play of Bedside Medical Instruments. Conf Proc IEEE Eng Med Biol Soc. 2008;2008:1514-1517. |
Vuorela et al., Two portable long-term measurement devices for ECG and bioimpedance. Second International Conference on Pervasive Computing Technologies for Healthcare. Jan. 30-Feb. 1, 2008:169-172. |
Wolf et al., Development of a Fall Detector and Classifier based on a Triaxial Accelerometer Demo Board. 2007:210-213. |
Yang et al., Research on Multi-Parameter Physiological Monitor Based on CAN Bus. IFMBE Proceed. 2008; 19:417-419. |
Zeltwanger, Controller Area Network and CANopen in Medical Equipment. Bus Briefing: Med Dev Manuf Technol. 2002:34-37. |
Zitzmann and Schumann, Interoperable Medical Devices Due to Standardized CANopen Interfaces. Joint Workshop on High Confidence Medical Devices, Software, and Systems and Medical Device Plug-and-Play Interoperability. 2007:97-103. |
Number | Date | Country | |
---|---|---|---|
20210251493 A1 | Aug 2021 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 15905744 | Feb 2018 | US |
Child | 17240941 | US | |
Parent | 15431459 | Feb 2017 | US |
Child | 15905744 | US | |
Parent | 14738910 | Jun 2015 | US |
Child | 15431459 | US | |
Parent | 14090433 | Nov 2013 | US |
Child | 14738910 | US | |
Parent | 13432976 | Mar 2012 | US |
Child | 14090433 | US | |
Parent | 12469182 | May 2009 | US |
Child | 13432976 | US |