Field of Invention
The present invention relates to a piezoelectric sensor for measuring shear and compression.
Brief Description of the Prior Art
A typical soft-material/tissue mechanical property tester requires an external force (displacement) applicator and an external displacement (force) gauge.1,2 The external force (displacement) generator may be hydraulic or piezoelectric and the external displacement gauge (force gauge) may be optical or piezoelectric. Regardless of the mechanism of force/displacement generation and displacement/force detection, typical tissue/soft-material mechanical testing is destructive and it requires specimens cut to a disc shape to fit in the tester. In addition, a compressive elastic modulus tester e.g., an Instron is also different from a shear modulus tester, e.g., a rheometer.3 Currently, no single instrument measures both the compressive Young's modulus and the shear modulus.
Over the past decades, many techniques have been developed to image tissue structures.4-9 Computer Tomography (CT)10 takes 360 degree X-ray pictures and reconstructs 3D tissue structures using computer software. Magnetic Resonance Imaging (MRI)11 uses powerful magnetic fields and radio waves to create tissue images for diagnosis. Ultrasound (US)12 transmits high-frequency waves through tissue and captures the echoes to image tissue structures. T-scan (TS)13 measures low-level bioelectric currents to produce real-time images of electrical impedance properties of tissues. Ultrasound elastography (UE)14 evaluates the echo time through tissues under a constant mechanical stress and compares it to that of the same tissue when unstressed. A tissue strain map is then obtained, from which an image of 2D elastic modulus distribution is created by inversion techniques. Tactile imaging tools using array pressure sensors probe spatial tissue stiffness variations. However none of these techniques have the ability to probe tumor interface properties.
The detection of abnormal tissue as cancerous growth requires improvements in screening technologies. The key to successful treatment lies in early detection. Imaging techniques such as mammography in breast cancer screening, detect abnormal tissue by tissue density contrast. Mammography is the only FDA approved breast cancer screening technique, which has a typical sensitivity of 85% that decreases to 65% in radiodense breasts.10 However, in these screening processes there is a high incidence of false positives. In fact, only about 15-30% of breast biopsies yield a diagnosis of malignancy. Changes in tissue stiffness have increasingly become an important characteristic in disease diagnosis. It is known that breast cancers are calcified tissues that are more than seven times stiffer than normal breast tissues.11-14 Thus, contrasting levels of stiffness within the breast may indicate cancerous tissue. Similarly, plaque-lined blood vessels are also stiffer than normal, healthy blood vessels.
The examining physician may detect abnormal tissue stiffness by palpation by taking advantage of the fact that cancerous tissues are stiffer than surrounding normal tissues under compression. Thus, palpation has been a useful tool for experienced physicians to diagnose breast and prostate cancer. However, palpation is not quantitative and depends solely on the experience of the individual physician. So there remains a critical need to improve cancer-screening technology to reduce the number of unnecessary biopsies.
In a first aspect, the invention relates to a sensor for measuring a compression modulus and a shear modulus that has a first layer made of piezoelectric material, a second layer made of a non-piezoelectric material, a first electrode placed on top of the first layer for sensing a displacement of the first layer; and a second electrode placed on top of the first layer for providing a force to the first layer.
In a second aspect of the invention a sensor for measuring a compression modulus and a shear modulus is provided having a first layer made of piezoelectric material for providing a force, a second layer made of non-piezoelectric material, and a third layer made of piezoelectric material for sensing a displacement.
In a third aspect of the invention a method for measuring a compression modulus and a shear modulus is provided. The method has the steps of providing a plurality of sensors made of a piezoelectric material and a non-piezoelectric material at a target, applying a force with at least one of the plurality of sensors, detecting a displacement with at least one of the plurality of sensors, and providing a measurement of a compression modulus and a shear modulus of said target.
A “PEFS” includes a piezoelectric layer bonded to a non-piezoelectric layer to form a cantilever.
In a first aspect of the invention, the PEFS is capable of simultaneously applying a force and detecting the corresponding displacement. The application of a voltage at the driving electrode generates the force and the corresponding displacement is measured by detecting the induced piezoelectric voltage within the sensing electrode. The PEFS can measure both the compressive Young's modulus and the shear modulus of a soft material through the cantilevered tip. Thus, in an aspect of the current invention, the PEFS measures both the compressive Young's modulus and shear modulus using one single device, while at the same time increasing the sensitivity and accuracy of the measurements, relative to some commercially available devices used for this purpose. Comparisons between the shear and compressive measurements using one single device provide clear and accurate information about the interfaces between hard inclusions (tumors) and the surrounding tissue that otherwise could likely not be obtained.
In another aspect of the invention, several PEFS can form an array to measure lateral and in-depth stiffness variations in soft-materials and tissues both under compression and under shear. This ability to self-excite via the driving voltage and self-detect via the sensing electrode allows a PEFS to measure the elastic and shear properties of specimens having complex shapes.
In another aspect of the invention, the PEFS may apply forces and measure displacements at the same time, allowing the device to function using completely electrical means for tissue-stiffness imaging, cancer and disease detection. Thus, the PEFS may be powered with a DC power source allowing the PEFS to take electrical measurements in the DC mode. In this aspect of the invention the PEFS may be part of a portable hand-held device for measuring tissue. This simple all-electrical measurement makes the PEFS look and work like a finger, which may allow for in vivo measurements in tight spaces.
While the sensitivity of the PEFS has been improved by the increased sensitivity of the compression tests with the addition of the shear test to the PEFS, the reduction of the probe size as compared to the bulkiness of current tactile cancer-imaging devices, provides increased versatility as well. In particular, the finger-like shape of the PEFS is now suitable for, for example, prostate cancer detection.
In another aspect of the invention, the PEFS can analyze measurements of various widths allowing the direct experimental determination of a stiffness variation in the thickness direction. Current tumor imaging techniques are incapable of directly measuring a tumor size or position. Instead, these techniques measure the surface mechanical response. The tumor information is generated numerically by the “inversion” technique and used to reconstruct the tumor size and position.15,16
In another aspect of the invention, the PEFS can assess vertical stiffness variations of soft materials/tissues up to several centimeters in depth with increased resolution by use of an array of PEFS's of varying probe widths ranging from less than 1 mm to several cm. Detection of tissue stiffness both under shear and under compression will allow comparison of the stiffness of a hard inclusion such as a tumor, with the stiffness of surrounding normal tissue, not only under compression, but also under shear. Such comparisons will permit a determination of the interfacial properties between the hard inclusion and surrounding tissue, which has the potential to greatly enhance the ability to assess tumor malignancy.
The PEFS is capable of detecting soft-material/stiffness variations in both the shear and the compression modes while under DC power. This allows a portable hand held device to detect soft material/tissue stiffness. Additional advantages result from the ultra-small strains employed for detection (smaller than 1%), and the minimal discomfort that such strains will cause to the patient.
In another aspect of the invention, the PEFS may be fabricated in a variety of shapes including L-shaped, U-L-shaped, U-shaped, square-shaped, O-shaped, tapered, etc. as well as in various lengths and widths. When the PEFS has an L-shaped tip, the PEFS is capable of accurately measuring the shear modulus of soft tissues and materials at very small strain (<0.1%), a capability most of the current commercial rheological instruments lack.
In addition to detection and mapping of a tumor, the PEFS may also be employed as a tissue/soft-material mechanical tester, for breast cancer detection, for prostate cancer detection, for monitoring skin cancer and skin elasticity testing, or for cellular elasticity/plasticity measurements using a miniaturized PEFS. Of course, the PEFS is capable of use in conventional methods for making compression and shear measurements on pliable materials of any kind and its use need not be limited to tissue measurements.
All-Electrical Measurement.
When a voltage is applied to the top PZT layer of a PEFS as shown in
The ability of the PEFS to electrically apply a force and electronically measure the displacement makes it ideal for “electronic palpation” like an “electronic finger.” The PEFS measures the tissue compressive (shear) stiffness by touching (rubbing) the tissue surface. The force generation and displacement sensing are all within the “finger.” The PEFS may be used for in-vivo tissue imaging particularly for breast cancer and prostate cancer detection.
The PEFS can measure elastic stiffness and shear modulus of soft materials, with or without a sensing electrode. In case a sensing electrode is not used, other means such a laser or piezoelectric displacement meter can be provided for displacement determination.
In
A flat-punch indentation compression test is a test whereby the cantilever tip is pressed on the sample surface and the cantilever contact area is much smaller than the sample surface area. This test simulates an in vivo measurement.
wherein ν is the Poisson ratio and E is the Young's modulus of the model tissue. Denoting the spring constant of the cantilever as K and the “free” displacement generated by an applied voltage V is represented as d0, the force F exerted on the model tissue by the applied voltage V with a displacement, δ, is therefore, F=K(d0−δ). It follows that the Young's modulus, E, is then expressed as,
wherein the subscript i of E denotes indentation.
A regular compression test describes a condition where a cantilever is pressed on the sample surface and the contact area is the same as the sample surface area. A schematic of a regular compression test is shown in
wherein h is the height of the sample, d0 the “free” displacement of the cantilever at voltage V, δ the displacement with the sample, K the spring constant of the cantilever and Ec is the elastic modulus measured by the regular compression test. For comparison, the same model tissue sample was tested using both the indentation compression and regular compression test conditions (the left-hand set of data and setup in
Having demonstrated that piezoelectric cantilevers are capable of determining the elastic stiffness of soft materials using the flat-punch indentation test, and that the test can be done with a driving electrode for force generation, and a sensing electrode for displacement/force detection using all-electrical measurements, here it is demonstrated that tissue elastic stiffness profiling can be done using flat-punch indentation. A simulated tissue sample with a hard inclusion was made from gelatin and candle wax. A 7 mm diameter, 5 mm tall cylinder of wax was embedded in an 8 mm thick gelatin matrix. Indentation tests were conducted at 2 mm increments over the entire surface surrounding the inclusion. The elastic modulus at each location was calculated using Eq. (2). The 2-D plot in
Because the indentation test only needs to press a small part of the sample surface, it is a natural configuration for in vivo application. However, because the probe size is smaller than the sample size in an indentation test, only the volume immediately beneath the indenting device is affected by the indentation test. It is thus conceivable that the detection sensitivity depends on the depth of the hard inclusion. To demonstrate this point, the effect of the probe size on the depth limit for detection using cantilevers of 3 mm width and 5 mm width was examined. Both cantilevers are 2 cm in length. The 3 mm wide cantilever had a contact area of 3 mm (the width)×2 mm and the 5 mm wide cantilever had a contact area of 5 mm (the width)×2 mm. With the 3 mm wide cantilever six wax inclusions of 7 mm diameters at varying depths beneath the top surface were embedded in a gelatin sample of 7 mm in height. For the 5 mm wide cantilever, five wax inclusions 15 mm in diameter were embedded in a gelatin sample of 18 mm in height. Cantilever indentation tests were conducted across the central axis of each sample at 1 mm increments. Elastic modulus was calculated at each location using the indentation formula, Equation (2). The result of elastic profiles of wax inclusions 7 mm in diameter embedded at different depths measured using the 3 mm wide cantilever are plotted in
For comparison, the result of elastic profiles of wax inclusions 15 mm in diameter embedded at different depths were measured using the 5 mm wide cantilever and plotted in
The above results indicate that one can probe the elastic stiffness to different depths by using cantilevers having different probe sizes. Furthermore, by carefully analyzing measurements with different probe sizes, it is possible to obtain not only the stiffness variation in the lateral direction, but also in the thickness direction to thereby allow construction of 3-D tissue stiffness maps.
Shear tests can be accomplished using L-shaped cantilevers. A schematic of a regular shear test is shown in
where h and A are the height and the surface area of the sample, respectively, K is the spring constant of the cantilever, and Δx0 and Δx the displacement of the cantilever without and with the sample, respectively.
A schematic of an indentation shear test is shown in
wherein A is the contact area, ν is the Poisson ratio, K is the spring constant of the cantilever, and Δx0 and Δx are the displacements of the cantilever without and with the sample, respectively.
For comparison of all four measurements, regular compression, indentation compression, regular shear and indentation shear, a cantilever that has a U-shaped stainless steel tip as shown in
Both the driving and sensing PZT layers are 127 μm thick and the stainless steel layer is 50 μm thick. The measurements were done on rubber samples with a Young's modulus of 400-500 kPa.
The results of the regular compression, indentation compression, regular shear, and indentation shear measurements on rubber samples that had a Young's modulus of 400-500 kPa using the U-shaped cantilever as shown in
Two tumor models, one with a smooth face and the other with a rough face were made of play dough and were the same size and embedded at the same depth in the model tissue gelatin as shown in
While both the smooth-surfaced and rough-surfaced model tumors were much stiffer than the surrounding gelatin under compression, only the rough-surfaced tumor displayed a stiffer shear. This indicates that the rough-surfaced tumor model was less mobile than the smooth-surfaced tumor model under shear. Thus, the indentation shear measurement with a piezoelectric finger was effective in probing tumor interfacial properties and tumor mobility. Combining both the compression and shear tests offers the potential of not only measuring the stiffness, but also determining the mobility of a tumor, which has great potential for tumor malignancy detection.
The use of the PEFS in excised breast tumors has been evaluated in the laboratory. The lumpectomy specimen was from a 60-year old woman with breast cancer. The known malignancy was 1.4 cm in the largest dimension. After surgical excision, the specimen was oriented with silk sutures, scanned with ultrasound, and images were stored. The PEFS scan was performed in the same orientation to allow later correlation with the ultrasound image. The specimen was sectioned in the same orientation to allow histological confirmation of the PEFS findings as well. Using the PEFS, preliminary elastic modulus measurements were performed on breast lumpectomy samples using an 8 mm wide PEFS with a rectangular tip.
Of note, the PEFS scan identified the known 15×13×12 mm invasive ductal carcinoma at x=15-25 mm and identified a smaller 6×5×3 mm satellite invasive ductal carcinoma at x=5-10 mm. This smaller lesion was not detected by mammogram, ultrasound or the physician's preoperative palpation.
With the second lumpectomy sample (not shown), the elastic modulus, E, and shear modulus, G, profiles have been performed. The tumor was 12-10 mm in size and 5 mm below the surface. The tissue was examined with an 8 mm wide PEFS.
In
A separate experiment (not shown) has determined that the probe depth of a PEFS is twice the PEFS's width. With Egel obtained from locations far away from the inclusion, one can then use the two profiles measured with the two PEFS' to solve the following two equations for the two unknowns, Einclusion and d:
where dp,1 and dp,2 (Emeasured,1 and Emeasured,2) are the probe depths of PEFS 1 and PEFS 2, respectively, Egel the elastic modulus of gelatin that can be obtained from far away from the inclusion, and d and Einclusion are the inclusion depth and the elastic modulus of the inclusion. Using the measured elastic modulus over the top of the center of the inclusion, the known probe depths, and the Egel obtained from far away from the inclusion and solving Eqs. (6) and (7), we obtained d=2.4 mm and Einclusion=109 kPa from
In this model study, a model smooth tumor (the shaded inclusion in the back of
Preliminary measurement on breast tissues after surgery have indicated that a PEFS can detect cancerous tumors as small as 3 mm in size that were missed by mammography, ultrasound, and physician's palpation, offering great potential for early breast cancer detection. Furthermore, it was demonstrated that by using two or more PEFS's of different widths, one can simultaneously determine both the tumor elastic (shear) modulus and its depth. In addition, it has also been demonstrated that tumor mobility can be assessed by measuring the ratio of the shear modulus to the elastic modulus of a tumor, or by sensitive direct tumor mobility measurement using two PEFS's, one for pushing and one for measuring the movement. The tumor mobility measurement offers the potential for non-invasive breast cancer malignancy screening.
It is to be understood, however, that even though numerous characteristics and advantages of the present invention have been set forth in the foregoing description, together with details of the structure and function of the invention, the disclosure is illustrative only, and changes may be made in detail, especially in matters of shape, size and arrangement of parts within the principles of the invention to the full extent indicated by the broad general meaning of the terms in which the appended claims are expressed.
The below list of references is incorporated herein in their entirety.
This application is a continuation of U.S. patent application Ser. No. 14/017,508, filed Sep. 4, 2013, now U.S. Pat. No. 8,826,749, which, in turn, is a continuation of U.S. patent application Ser. No. 13/268,225, filed on Oct. 7, 2011, now U.S. Pat. No. 8,549,933 issued on Oct. 8, 2013, which, in turn, is a continuation of U.S. patent application Ser. No. 12/837,590, filed on Jul. 16, 2010, now U.S. Pat. No. 8,033,185 issued on Oct. 11, 2011, which, in turn, is a continuation of U.S. patent application Ser. No. 12/328,639, filed on Dec. 4, 2008, now U.S. Pat. No. 7,779,707 issued on Aug. 24, 2010, which, in turn, is a continuation of U.S. patent application Ser. No. 11/136,173, filed on May 24, 2005, now U.S. Pat. No. 7,497,133, issued on Mar. 3, 2009, which, in turn, is a non-provisional of U.S. Provisional Application No. 60/573,869, filed May 24, 2004, the entire disclosures of which are hereby incorporated by reference as if set forth fully herein.
This invention was reduced to practice with Government support under Grant No. R01 EB00720-01 awarded by NIH; the Government is therefore entitled to certain rights to this invention.
Number | Name | Date | Kind |
---|---|---|---|
3205464 | Schwartz | Sep 1965 | A |
4093883 | Yamamoto | Jun 1978 | A |
4302694 | Fujishima et al. | Nov 1981 | A |
4349762 | Kitamura et al. | Sep 1982 | A |
4363993 | Nishigaki et al. | Dec 1982 | A |
4403166 | Tanaka et al. | Sep 1983 | A |
4528502 | Rocha | Jul 1985 | A |
4649312 | Robin et al. | Mar 1987 | A |
4802371 | Calderara et al. | Feb 1989 | A |
4944187 | Frick et al. | Jul 1990 | A |
RE33691 | Harnden, Jr. et al. | Sep 1991 | E |
5054323 | Hubbard et al. | Oct 1991 | A |
5289826 | Kovacevic | Mar 1994 | A |
5313535 | Williams | May 1994 | A |
5334835 | Nakayama et al. | Aug 1994 | A |
5338999 | Ramakrishnan et al. | Aug 1994 | A |
5382864 | Morikawa et al. | Jan 1995 | A |
5445008 | Wachter et al. | Aug 1995 | A |
5475318 | Marcus et al. | Dec 1995 | A |
5503010 | Yamanaka | Apr 1996 | A |
5553486 | Bonin | Sep 1996 | A |
5626728 | Ramakrishnan et al. | May 1997 | A |
5689063 | Fujiu et al. | Nov 1997 | A |
5719324 | Thundat et al. | Feb 1998 | A |
5780727 | Gimzewski et al. | Jul 1998 | A |
5807758 | Lee et al. | Sep 1998 | A |
5833634 | Laird et al. | Nov 1998 | A |
5866807 | Elings et al. | Feb 1999 | A |
5874126 | Kahn et al. | Feb 1999 | A |
5948993 | Ting et al. | Sep 1999 | A |
5966787 | Nakayama et al. | Oct 1999 | A |
5996412 | Hansen | Dec 1999 | A |
6075585 | Minne et al. | Jun 2000 | A |
6278379 | Allen et al. | Aug 2001 | B1 |
6280396 | Clark | Aug 2001 | B1 |
6289717 | Thundat et al. | Sep 2001 | B1 |
6336366 | Thundat et al. | Jan 2002 | B1 |
6422069 | Shimizu et al. | Jul 2002 | B1 |
6458327 | Vossmeyer et al. | Oct 2002 | B1 |
6465368 | Inoue et al. | Oct 2002 | B2 |
6500119 | West et al. | Dec 2002 | B1 |
6589727 | Klenerman et al. | Jul 2003 | B1 |
6621080 | Yamamoto | Sep 2003 | B2 |
6734425 | Hantschel et al. | May 2004 | B2 |
6781285 | Lazarus et al. | Aug 2004 | B1 |
6903491 | Irie et al. | Jun 2005 | B2 |
6992421 | Ikeda et al. | Jan 2006 | B2 |
7022540 | Kim et al. | Apr 2006 | B2 |
7055378 | Su et al. | Jun 2006 | B2 |
7083270 | Torii et al. | Aug 2006 | B2 |
7084554 | Xu et al. | Aug 2006 | B2 |
7104134 | Amano et al. | Sep 2006 | B2 |
7112452 | Cho et al. | Sep 2006 | B2 |
7195909 | Klenerman et al. | Mar 2007 | B2 |
7263874 | Fitch et al. | Sep 2007 | B2 |
7458265 | Shih et al. | Dec 2008 | B2 |
7497133 | Shih et al. | Mar 2009 | B2 |
7521257 | Adams et al. | Apr 2009 | B2 |
7744713 | Blessing et al. | Jun 2010 | B2 |
7744773 | Shih et al. | Jun 2010 | B2 |
7942056 | Mutharasan et al. | May 2011 | B2 |
7992431 | Shih et al. | Aug 2011 | B2 |
8474319 | Mutharasan | Jul 2013 | B2 |
8562546 | Shih | Oct 2013 | B2 |
20020094528 | Salafsky | Jul 2002 | A1 |
20020117659 | Lieber et al. | Aug 2002 | A1 |
20020155303 | Wielstra et al. | Oct 2002 | A1 |
20030032293 | Kim et al. | Feb 2003 | A1 |
20030068655 | Bottomley et al. | Apr 2003 | A1 |
20030194697 | Klenerman et al. | Oct 2003 | A1 |
20030224551 | Kim et al. | Dec 2003 | A1 |
20030235681 | Sebastian et al. | Dec 2003 | A1 |
20040022677 | Wohlstadter et al. | Feb 2004 | A1 |
20040265664 | Badding et al. | Dec 2004 | A1 |
20050112621 | Kim et al. | May 2005 | A1 |
20050114045 | Giurgiutiu et al. | May 2005 | A1 |
20050199047 | Adams et al. | Sep 2005 | A1 |
20050277852 | Shih et al. | Dec 2005 | A1 |
20050287680 | Venkatasubbarao et al. | Dec 2005 | A1 |
20060053870 | Berndt | Mar 2006 | A1 |
20060217893 | Li et al. | Sep 2006 | A1 |
20060223691 | Shih et al. | Oct 2006 | A1 |
20060228657 | Masters et al. | Oct 2006 | A1 |
20060257286 | Adams | Nov 2006 | A1 |
20070089515 | Shih et al. | Apr 2007 | A1 |
20070141721 | Vafai et al. | Jun 2007 | A1 |
20070169553 | Mutharasan | Jul 2007 | A1 |
20070218534 | Klenerman et al. | Sep 2007 | A1 |
20080034840 | Mutharasan | Feb 2008 | A1 |
20080035180 | Mutharasan | Feb 2008 | A1 |
20090007645 | Shih et al. | Jan 2009 | A1 |
20090053709 | Mutharasan | Feb 2009 | A1 |
20090078023 | Mutharasan | Mar 2009 | A1 |
20090203000 | Mutharasan | Aug 2009 | A1 |
20100068697 | Shih et al. | Mar 2010 | A1 |
20100224818 | Shih et al. | Sep 2010 | A1 |
20100239463 | Shih et al. | Sep 2010 | A1 |
Number | Date | Country |
---|---|---|
0631319 | Dec 1994 | EP |
1536227 | Jun 2005 | EP |
3093849 | Apr 2000 | JP |
2003-298131 | Oct 2003 | JP |
2004-265899 | Sep 2004 | JP |
2007-67125 | Mar 2007 | JP |
9850773 | Nov 1998 | WO |
2004061991 | Jul 2004 | WO |
2005043126 | May 2005 | WO |
2006031072 | Mar 2006 | WO |
2007087328 | Aug 2007 | WO |
2007109228 | Sep 2007 | WO |
2007133619 | Nov 2007 | WO |
2008020903 | Feb 2008 | WO |
2008021187 | Feb 2008 | WO |
2008021189 | Feb 2008 | WO |
2008101199 | Aug 2008 | WO |
2009014830 | Jan 2009 | WO |
2009035732 | Mar 2009 | WO |
2009035732 | Mar 2009 | WO |
2009046251 | Apr 2009 | WO |
Entry |
---|
Amanuma, K. et al., “Crystallization behavior of sol-gel derived Pb(Zr,Ti)O3 thin films and the polarization switching effect on film microstructure”, Appl. Phys. Lett., 65(24): 3140-3142 (1994). |
Ammari, H. et al., “T-Scan Electrical Impedance Imaging System for Anomaly Detection”, Siam J. Appl. Math., 65(1): 252-266 (2004). |
Baselt, D. R. et al., “Biosensor based on force microscope technology”, J. Vac. Sci. Technol. B, 14(2): 789-793 (1996). |
Birnie, III, D. P. et al., “Coating uniformity and device applicability of spin coated sol-gel PXT films”, Microelectronic Engineering, 29: 189-192 (1995). |
Bondoux, C. et al., “MgO insulating films prepared by sol-gel route for SiC substrate”, J. Europe. Ceramic Soc., 25: 2795-2798 (2005). |
Brito, R. et al., “Adsorption of 3-mercaptopropyltrimethoxysilane and 3-aminopropyltrimethoxysilane at platinum electrodes”, J. Electroanalytical Chem., 520: 47-52 (2002). |
Campbell, G.A., et al., “Piezoelectric excited millimeter-sized cantilever (PEMC) sensor detects Escherichia coli O157:H7 in two-hour incubated samples at 4 CFU per gram of beef,” J. of Rapid Methods and Automation in Mirobiology, 1-39. |
Campbell, G.A., et al., “Detection and quantification of proteins using self-excited PZT-glass millimeter-sized cantilever,” Biosensors and Bioelectronics, 26-36. |
Campbell, G.A., “Piezoelectric-excited millimeter-sized cantilever (PEMC) sensors detect Bacillus anthracis at 300 spores/mL,” Biosensors and Bioelectronics, 37-45. |
Campbell, G.A., et al., “kinetics of bacillus anthracis spore binding to antibody functionalized PEMC sensors in presence of bacillus thuringiensis and bacillus cereus,” J. Publications, Am. Chem. Soc. 25 pages. |
Campbell, G.A., et al., “Escherichia coli O157:H7 detection limit of millimeter-sized PZT cantilever sensors in 700 cells/mL,” Analytical Sci., 11-13. |
Campbell, G.A., et al., “Detection of pathogen Escherichia coli O157:H7 using self-excited PZT-glass microcantilevers,” Biosensors and Boelectronics, 14-25. |
Campbell, G.A., “Detection of Staphylococcus enterotoxin B at pictogram levels using piezoelectric-excited millimeter-sized cantilever sensors,” Submitted on line to J. of Analytical Chem., 1-24. |
Campbell, G.A., et al., “Detect Escherichia coli O157:H7 in ground beef samples using piezoelectric excited millimeter-sized cantilever (PEMC) sensors,” Submitted on-line to Biosensors and Bioelectronics, 2-34. |
Campbell, G.A., et al., “A method for measuring Escherichia coli O157:H7 at 1 cell/mL in 1 liter sample using antibody functional piezoelectric-excited millimeter sized cantilever sensor,” Paper submitted on-line to J. of Analytical Chemistry. 1-23. |
Capobianco, J. A., et al., “Methyltrimethoxysilane-insulated piezoelectric microcantilevers for direct, all-electrical biodetection in buffered aqueous solutions”, Rev. Sci. Instrum., 77: 125105-1-125105-6 (2006). |
Capobianco, J. A., et al., “3-mercaptopropyltrimethoxysilane as insulating coating and surface for protein immobilization for piezoelectric microcantilever sensors”, Rev. Sci. Instrum., 78: 046106-1-046106-3 (2007). |
Carlier, S. G., et al., “Elastography”, J. Cardiovasc Risk, 9(5): 237-245 (2002). |
Carr, D.W., et al., “Fabrication of nanoelectromechanical systems in single crystal silicon using silicon on insulator substrates and electron beam lithography,” J. Vac. Sci. Technology, B, 5(6), 2760-2763. |
Che, G. et al., “Molecular recognition based on (3-mercaptopropyl) trimethoxysilane modified gold electrodes”, J. Electroanalytical Chem., 417: 155-161 (1996). |
Chen, G. Y. et al., “Adsorption-induced surface stress and its effects on resonance frequency of microcantilevers”, J. Appl. Phys., 77(8): 3618-3622 (1995). |
Chen, X. et al., “Electrochemical and Spectroscopic Characterization of Surface Sol-Gel Processes”, Langmuir, 20 (20): 8762-8767 (2004). |
Cho, S. H. et al., “Micro-scale metallization on flexible polyimide substrate by Cu electroplating using SU-8 photoresist mask”, Thin Solid Films, 475: 68-71 (2005). |
Duval, F.F.C. et al., “Stable TiO2/Pt electrode structure for lead containing ferroelectric thick films on silicon MEMS structures”, Thin Solid Films, 444: 235-240 (2003). |
Feili, D. et al., “Encapsulation of organic field effect transistors for flexible biomedical microimplants”, Sensors and Actuators, A120: 101-109 (2005). |
Ferrini, R. et al., “Screening Mammography for Breast Cancer: American College of Preventive Medicine Practice Policy Statement”, www.acpm.org/breast, pp. 1-4 (2005). |
Fritz, J. et al., “Translating Biomolecular Recognition into Nanomechanics”, Science, 288: 316-318 (2000). |
Fung, Y. S. et al., “Self-Assembled Monolayers as the Coating in a Quartz Piezoelectric Crystal Immunosensor to Detect Salmonella in Aqueous Solution”, Anal. Chem., 73: 5302-5309 (2001). |
Gao, L. et al., “Imaging of the elastic properties of tissue: A review”, Ultrasound in Med. & Biol., 22(8): 959-977 (1996). Abstract Only. |
Greenleaf, J. F. et al., “Selected Methods for Imaging Elastic Properties of Biological Tissues”, Annu. Rev. Biomed. Eng., 5: 57-78 (2003). |
Gu, H. et al., “Single-Calcination Synthesis of Pyrochlore-Free 0.9Pb(Mg1/3Nb2/3)O3—0.1PbTiO3 and Pb(Mg1/3Nb2/3)O3 Ceramics Using a Coating Method”, J. Am. Ceram. Soc., 86(2): 217-221 (2003). |
Haccart, T. et al., “Evaluation of niobium effects on the longitudinal piezoelectric coeffecients of Pb(Zr,Ti)O3 thin films”, Appl. Phys. Lett., 76(22): 3292-3294 (2000). |
Han, W. et al., “A magnetically driven oscillating probe microscope for operations in liquids”, Appl. Phys. Lett., 69(26): 4111-4113 (1996). |
Hiboux, S. et al., “Mixed titania-lead oxide seed layers for PZT growth on Pt(111): a study on nucleation, texture and properties”, J. Europe. Ceram. Soc., 24: 1593-1596 (2004). |
Hwang, I.H. et al., “Self-actuating biosensor using a piezoelectric cantilever and its optimization”, Journal of Physics: Conference Series 34, pp. 362-367, 2006. |
Hwang, K.S. et al., “In-situ quantitative analysis of a prostate-specific antigen (PSA) using a nanomechanical PZT cantilever”, Lab Chip, 4: 547-552 (2004). |
Ilic, B. et al., “Mechanical resonant immunospecific biological detector”, Appl. Phys. Lett., 77(3): 450-452 (2000). |
Itoh, T. et al., “Self-excited force-sensing microcantilevers with piezoelectric thin films for dynamic scanning force microscopy”, Sensor and Actuators, A54:477-481 (1996). |
Jung, S.K. et al., “Polymeric Mercaptosilane-Modified Platinum Electrodes for Elimination of Interferants in Glucose Biosensors”, Anal. Chem., 68: 591-596 (1996). |
Kanda, T. et al., “A flat type touch probe sensor using PZT thin film vibrator”, Sensors and Actuators, 83: 67-75 (2000). |
Katiyar, P. et al. “Electrical properties of amorphous aluminum oxide thin films”, Acta Materialia, 53: 2617-2622 (2005). |
Keller, A. et al., “Reliability of Computed Tomography Measurements of Paraspinal Muscle Cross-Sectional Area and Density in Patients With Chronic Low Back Pain”, Spine, 28(13): 1455-1460 (2003). |
Kelly, J. et al., “Effect of Composition on the Electromechanical Properties of (1-x)Pb(Mg1/3Nb2/3)O3—xPbTiO3 Ceramics” J. Am. Ceram. Soc., 80(4): 957-964 (1997). |
Khabari, A. et al., “Partially ionized beam deposition of parylene” J. Non-Crystalline Solids, 351: 3536-3541 (2005). |
Kim, S.H. et al., “Influence of Al2O3 diffusion barrier and PbTiO3 seed layer on microstructurel and ferroelectric charachteristics of PZT thin films by sol-gel spin coating method,” Thin Solid Films, 305: 321-326 (1997). |
Kim, S.J. et al., “Fabrication and Characterization of Pb(Zr,Ti)O3 Microcantilever for Resonance Sensors,” Jpn. J. Appl. Phys., 42(3): 1475-1478 (2003). |
Klissurska, R.D. et al. “Microstructure of PZT sol-gel films on Pt substrates with different adhesion layers,” Microelectronic Engineering, 29: 297-300 (1995). |
Kruse, S.A. et al., “Tissue characterization using magnetic resonance elastography: preliminary results,” Phys. Med. Biol., 45: 1579-1590 (2000). |
Kumar, V. et al., “A Simple System for the Preparation of Submicrometer Lead Titanate Powders by the Sol-Gel Method,” J. Am. Ceram. Soc., 79(10): 2775-2778 (1996). |
Kwok, Clk. et al., “Low temperature perovskite formation of lead zirconate titanate thin films by a seeding process,” J. Mater. Res., 8(2): 339-344 (1993). |
H. Zhang, et al., “A Sensitive and High-Throughput Assay to Detect Low-Abundance Proteins in Serum,” Nature Medicine 12(4) 473-477 (2006). |
J. W. Park, S. Kurosawa, H. Aizawa Y. Goda, M. Takai and K. Ishihara, “Piezoelectric Immunosensor for Bisphenol A Based on Signal Enhancing Step With 2-methacrolyloxyethyl Phosphorylcholine Polymeric Nanoparticle,” Analyst, 131, 155-162 (2006). |
A. M. Smith, G. Ruan, M. N. Rhyner, and S. Nie, “Engineering Luminescent Quantum Dots for In Vivo Molecular and Cellular Imaging,” Ann. Biomed. Eng., 34 (1),3-14 (2006). |
R. E. Jaeger and L. Egerton, “Hot-Pressing of Potassium-Sodium Niobates,” J. Am. Ceram. Soc. 45, 209 (1962). |
H. Birol, D. Damjanovic and N. Setter, “Preparation and Characterization of (K0.5Na0.5)NbO3 Ceramics”, J. Eur. Ceram. Soc. 26, 861 (2006). |
Y. Guo, K. Kakimoto, and H. Ohsato, “Phase Transitional Behavior and Piezoelectric Properties of (Na0.5K0.5)NbO3—LiNbO3 Ceramics,” Appl. Phys. Lett., 85, 4121 (2004). |
Y. Guo, K. Kakimoto, and H. Ohsato, “(Na0.5K0.5)NbO3—LiTaO3 Lead-free Piezoelectric Ceramics,” Mater. Lett., 59, 241 (2005). |
H. Li, W. Y. Shih, and W.-H. Shih, “Effect of Antimony Concentration on the Crystalline Structure, Dielectric and Piezoelectric Properties of (Na0.5K0.5)0.945Li0.055Nb1—xSbxO3 Solid Solutions”, J. Am. Ceram. Soc., 90, 3070 (2007). |
S. Zhang, R. Xia, T. R. Shrout, J. Zang, and J. Wang, “Piezoelectric Properties in Perovskite 0.948(K0.5Na0.5)NbO3-0.052LiSbO3 lead-free ceramics”, J. App. Phys., 100, 104108 (2006). |
X. Li, W. Y. Shih, J. S. Vartuli, D. L. Milius, I. A. Aksay, and W.-H. Shih, “Effect of Transverse Tensile Stress on Electric-Field-Induced Domain Reorientation in Soft PZT: In Situ XRD Study”, S. Am. Ceram. Soc. 85 (4), 844 (2002). |
Q. Zhu, W. Y. Shih, and W.-H. Shih, “Real-Time, Label-Free, All-Electrical Detection of Salmonella typhimurium Using Lead Titanate Zirconate/Gold-Coated Glass Cantilevers at any Relative Humidity,” Sensors and Actuators B, 125, 379-388 (2007). |
Q. Zhu, W. Y. Shih, and W.-H. Shih, “Length and Thickness Dependence of Longitudinal Flexural Resonance Frequency Shifts of a Piezoelectric Microcantilever Sensor due to Young's Modulus Change,” J. Appl. Phys. 104, 074503 (2008). |
Q. Zhu, W. Y. Shih, and W.-H. Shih, “Enhanced Detection Resonance Frequency Shift of a Piezoelectric Microcantilever Sensor by a DC Bias Electric Field in Humidity Detection,” Sensors and Actuators, B 138, 1 (2009). |
Q. Zhu, W. Y. Shih, and W.-H. Shih, “Mechanism of the Flexural Resonance Frequency Shift of a Piezoelectric Microcantilever Sensor in a DC Bias Electric Field,” Appl. Phys. Lett. 92, 033503 (2008). |
Q. Zhu, Drexel University (2008). |
McGovern, J.P. Drexel University (2008). |
Luo, H.Y. Drexel University (2005). |
Shin, S., Kim, J.P., Sim, S.J. & Lee, J. A multisized piezoelectric microcantilever biosensor array for the quantitative analysis of mass and surface stress. Applied Physics Letters 93,—(2008). |
Pang, W. et al. Femtogram mass sensing platform based on lateral extensional mode piezoelectric resonator. Applied Physics Letters 88,—(2006). |
Cherian, S. & Thundat, T. Determination of adsorption-induced variation in the spring constant of a microcantilever. Applied Physics Letters 80, 2219-2221 (2002). |
Lee, J.H., Kim, T.S. & Yoon, K.H. Effect of mass and stress on resonant frequency shift of functionalized Pb(Zr0.52Ti0.48)O3 thin film microcantilever for the detection of C-reactive protein. Applied Physics Letters 84, 3187-3189 (2004). |
Lee, J.H. et al. Immunoassay of prostate-specific antigen (PSA) using resonant frequency shift of piezoelectric nanomechanical microcantilever. Biosensors and Bioelectronics 20, 2157-2162 (2005). |
Shen, Z., Shih, W.Y. & Shih, W.-H. Self-exciting, self-sensing PbZr0.53Ti0.47O3/SiO2 piezoelectric microcantilevers with femtogram/Hertz sensitivity. Applied Physics Letters 89, 023506-3 (2006). |
Zhu, Q., Shih, W.Y. & Shih, W.-H. In situ, in-liquid, all-electrical detection of Salmonella typhimurium using lead titanate zirconate/gold-coated glass cantilevers at any dipping depth. Biosensors and Bioelectronics 22, 3132-3138 (2007). |
McGovern, J.P. et al. Label-free flow-enhanced specific detection of Bacillus anthracis using a piezoelectric microcantilever sensor. Analyst 133, 649-654 (2008). |
McGovern, J.P., Shih, W.Y. & Shih, W.H. In situ detection of Bacillus anthracis spores using fully submersible, self-exciting, self-sensing PMN-PT/Sn piezoelectric microcantilevers. Analyst 132, 777-783 (2007). |
McGovern, J.-P. et al. Label-free flow-enhanced specific detection of Bacillus anthracis using a piezoelectric microcantilever sensor. The Analyst 133, 649-654 (2008). |
McGovern, J.-P., Shih, W.Y. & Shih, W.-H. In situ detection of Bacillus anthracis spores using fully submersible, self-exciting, self-sensing PMN-PT/Sn piezoelectric microcantilevers. The Analyst 132, 777-783 (2007). |
Zhu, Q., Shih, W.Y. & Shih, W.-H. Mechanism of flexural resonance frequency shift of a piezoelectric microcantilever sensor during humidity detection. Applied Physics Letters 92, 183505-3 (2008). |
Su, W.-S., Chen, Y.-F., Shih, W.Y., Luo, H. & Shih, W.-H. Domain switching in lead magnesium niobate-lead titanate polycrystalline sheets at single grain level. Applied Physics Letters 91, 112903-3 (2007). |
Shang, J.K. & Tan, X. Indentation-induced domain switching in Pb(Mg1/3Nb2/3)O3—PbTiO3 crystal. Acta Materialia 49, 2993-2999 (2001). |
Alguero, M., Jimenez, B. & Pardo, L. Rayleigh type behavior of the Young's modulus of unpoled ferroelectric ceramics and its dependence on temperature. Applied Physics Letters 83, 2641-2643 (2003). |
Masys, A.J., Ren, W., Yang, G. & Mukherjee, B.K. Piezoelectric strain in lead zirconate titante ceramics as a function of electric field, frequency, and dc bias. Journal of Applied Physics 94, 1155-1162 (2003). |
Capobianco, J.A., Shih, W.Y., Yuan, Q.-A., Adams, G.P. & Shih, W.-H. Label-free, all-electrical, in situ human epidermal growth receptor 2 detection. Review of Scientific Instruments 79, 076101 (2008). |
Shih, W.Y., Luo, H., Li, H., Martorano, C. & Shih, W.-H. Sheet geometry enhanced giant piezoelectric coefficients. Applied Physics Letters 89, 242913-3 (2006). |
Capobianco, J.A., Shih, W.Y. & Shih, W.-H. 3-mercaptopropyltrimethoxysilane as insulating coating and surface for protein immobilization for piezoelectric microcantilever sensors. Review of Scientific Instruments 78, 046106 (2007). |
Morton, T.A., Myszka, D.G. & Chaiken, I.M. Interpreting Complex Binding-Kinetics from Optical Biosensors—a Comparison of Analysis by Linearization, the Integrated Rate-Equation, and Numerical-Integration. Analytical Biochemistry 227, 176-185 (1995). |
Shuck, P. & Minton, A.P. Kinetic analysis of biosensor data: elementary test of self-consistency. Trends Biochemical Sciences 21, 458-460 (1996). |
McKendry, R. et al. Multiple label-free biodetection and quantitative DNA-binding assays on a nanomechanical cantilever array. Proceedings of the National Academy of Sciences of the United States of America 99, 9783-9788 (2002). |
Ndieyira, J.W. et al. Nanornechanical detection of antibiotic mucopeptide binding in a model for superbug drug resistance. Nature Nanotechnology 3, 691-696 (2008). |
Sofian M. Kanan and Carl P. Tripp, “An Infrared Study of Adsorbed Organophosphonates on Silica: A Prefiltering Strategy for the Detection of Nerve Agents on Metal Oxide Sensors,” Langmuir 2001, 17, 2213-2218, United States of America. |
“Enhanced detection resonance frequency shift of a piezoelectric microcantilver sensor by a DC bias electric field in humidity detection,” Sensors and Actuators B: Chemical, 2009, vol. 138, United States of America. |
J. K. Shang and X. Tan, “Indentation-Induced Domain Switching in Pb(Mg1/3Nb2/3)O3—PbTiO3 Crystal” Acta Mater., 2001, pp. 2993-2999, vol. 49, Urbana, Illinois. |
IEEE Standard on Piezoelectricity IEEE, New York, 1988, Chap. 6. |
L. Bellaiche and David Vanderbilt, Physical Review Letters, 83(7), Aug. 16, 1999, 1347. |
S-F. Liu, W. Ren, B. K. Mukherjee, S. J. Zhang, T. R. Shrout, P. W. Rehrig, and W. S. Hackenberger, Appl. Phys. Lett., 83, 2886 (2003). |
PZT data, IEEE Micro Electro Mechanical Systems Workshop, Jan.-Feb. 1991, Nara, Japan p. 118. |
Xu, et al., “Longtitudinal piezoelectric coefficient measurement for bulk ceramics and thin films using pneumatic pressure rig,” Journal of Applied Physics, Jul. 1, 1999, pp. 588-594, vol. 86, No. 1, Pennsylvania. |
Li, “Sodium Potassium Niobate-based Lead-free Piezoelectric Ceramics: Bulk and Freestanding Thick Films,” Thesis Submitted to Faculty of Drexel University, Jun. 2008, Philadelphia, Pennsylvania. |
Li, “Synthesis of Na0.5K0.5NbO3 Piezoelectrics by a Solution Coating Approach,” Int. J. Appl. Technol., 2009, pp. 205-215, vol. 6, Issue 2, United States of America. |
Hudson, J.B. Surface Science: An Introduction, (Wiley-IEEE, New York, 1998). |
Q. Zhu. “Enhanced detection resonance frequency shift of a piezoelectric microcantilever sensor by a DC bias electric field in humidity detection,” Sensors and Actuators B: Chemical, 2009, pp. 1-4, vol. 138. |
Q. Zhu, W. Y. Shih & W.H. Shih, “mechanism of flexural resonance frequency shift of a piezoelectric microcantilever sensor during humidity detection,” Applied Physics Letters, 2008, vol. 92, United States of America. |
Hudson, J.B. Surface Science: An Introduction, Wiley-IEEE, 1998, pp. 96-98, New York. |
Leckband, D.E. et al. Force Probe Measurements of Antibody-Antigen Interactions. Methods 20, 329-340 (2000). |
O'Sullivan, C.K. & Guilbault, G.G. Commercial quartz crystal microbalances—theory and applications. Biosensors & Bioelectronics 14, 663-670 (1999). |
Lofgren, J.A. et al. Comparing ELISA and surface plasmon resonance for assessing clinical immunogenicity of panitumumab. J Immunol 178, 7467-72 (2007). |
Borghaei, et al., “induction of Adaptive Anti-HER2/neu Immune Responses . . . ” J. Immunother, Jun. 2007, pp. 455, vol. 30, No. 4. |
H. Yengingil, “Breast Cancer Detection and Differentiation Using Piezoelectric Fingers,” PhD Thesis, Drexel University, Philadelphia, PA, Jan. 2009. |
E. E. Konofagou, T. Harrigan, and J. Ophir, “Shear Strain Estimation and Lesion Mobility Assessment in Elastography,” Ultrasonics, 2000, pp. 400-404, vol. 38. |
H.O. Yegingil, W. Y. Shih, W. Anjum, A. D. Brooks and W.-H. Shih, “Soft tissue elastic modulus measurement and tumor detection using piezoelectric fingers,” Mat. Res. Soc. Symp. Proc., 2006, vol. 898E. |
P. S. Wellman, E. P. Dalton D. Krag,, K. A. Kern, R. D. Howe, “Tactile Imaging of Breast Masses: First Clinical Report,” Archives of Surgery 136(2), 204-08 (2001). |
Z. Shen, W. Y. Shih, and W. -H. Shih, “Mass detection sensitivity of piezoelectric cantilevers with a nonpiezoelectric extension,” Rev. Sci. Instrum. 77, 065101 (2006). |
A. Markidou, W. Y. Shih, and W.-H. Shih, “Soft-materials elastic and sear moduli measurement using piezoelectric cantilevers,” Rev. Sci. Ins. 76, 064302 (2005). |
S . T. Szewczyk, W.Y. Shih, and W.-H. Shih, “Palpationlike soft-material elastic modulus measurement using piezoelectric cantilievers,” Rev. Sci. Ins., 77, 044302 (2006). |
H. O. Yegingil, W. Y. Shih, and W.-H. Shih, “All-electrical indentation shear modulus and elastic modulus measurement using a piezoelectric cantilever with a tip,” J. Appl. Phys., 101, 054510 (2007). |
W. Jiang and W. Cao, “Intrinsic and coupling-induced elastic nonlinearity of lanthanum-doped lead magnesium niobate-lead titanate electrostrictive ceramic,” Appl. Phys. Lett., 77,1387 (2000). |
A. W. McFarland, et al., “Influence of surface stress on the resonance behavior of microcantilevers,” Appl. Phys. Lett. 87, 053505 (2005). |
O. Kwon, “T-scan Electrical Impedance Imaging system for anomaly detection,” Siam J. Appl. Math., 2004, pp. 252-266, vol. 65, No. 1. |
Sure Touch Exam [online] retrieved Nov. 29, 2010 from the internet @ http://www.medicaltactile.com/default.htm. |
Q. Ren and Y. P. Zhao, “Influence of surface stress on frequency of microcantilever-based biosensors,” Microsystem Technologies, 2004, pp. 307-314, vol. 10. |
E. Chen, “Ultrasound Tissue Displacement and Tissue Elasticity Imaging,” Ph.D. dissertation, University of Illinois at Urbana-Champaign, (1995). |
Haun, M.J. “Thermodynamic Theory of the Lead Zirconate-Titanate Solid Solution System,” The Pennsylvania State University (1988). |
Wellman, P. S. et al., “Tactile Imaging: A Method for Documenting Breast Lumps”. |
Weng, L. et al., “Effect of acetylacetone on the preparation of PZT materials in sol/gel processing”, Mater. Sci. Engin., B96: 307-312 (2002). |
Wilson, L S et al., “Elastography—the movement begins”, Phys. Med. Biol., 45: 1409-1421 (2000). |
Wilson, L., et al., “Pezoelectric-excited millimeter-sized cantilever (PEMC) sensor provides viscosity and density measurements,” Submitted to Review of Scientific Instruments, 1-26. |
Yi, J. W. et al., “Effect of length, width, and mode on the mass detection sensitivity of piezoelectric unimorph cantilevers”, J. Appl. Phys., 91(3): 1680-1686 (2002). |
Yi, J. W. et al., “In situ cell detection using piezoelectric lead zirconate titanate-stainless steel cantilevers”, J. Appl. Phys., 93(1): 619-625 (2003). |
Zhao, Q. et al., “Array adsorbent-coated lead zirconate titanate (PZT)/stainless steel cantilevers for dimethyl methylphosphonate (DMMP) detection”, Sensors and Actuators, B117(1): 74-79 (2006). Abstract Only. |
Zhou, J. et al., “Zeolite-modified microcantilever gas sensor for indoor air quality control,” Sensors and Actuators B, Oct. 1, 2003, 94(3), 337-342. |
Zhu, D.M. et al., “Thermal conductivity and electromechanical property of single-crystal lead magnesium niobate fitanate”, Appl. Phys. Lett., 75(24): 3868-3870 (1999). |
Data of Commercially Available Product, EDO Corporation: 1-8 (1999). |
Data of Commercially Available Product, APC International, Ltd.: 1-2 (2005). |
Campbell, G.A., et al., “Use of Piezoelectric-Excited millimeter Sized Cantilever Sensors to Measure Albumin Interaction with Self-Assembled Monolayers of Alkanethiols Having Different Functional Headgroups,” Anal. Chem. 78, 2328-2334 (2006). |
Campbell, G.A., et al., “Method of measuring Bacillus anthracis spores in the Presence of copious amounts of Bacillus thurigiensis and Bacillus cereus,” Anal. Chem. 79, 1145-1152 (2007). |
Campbell, G.A., et al., “PEMC sensor's mass change sensitivity is 20 pg/Hz under liquid immersion,” Biosensors and Bioelectronics, 22, 35-41 (2006). |
Campbell, G.A., et al., “Detection of Bacillus anthracis spores and a model protein usings PEMC sensors in a flow cell at 1 mL/min,” Biosensors and Bioelectronics, 22, 78-85 (2006). |
Campbell, G.A., et al., “Detection of airborne Bacillus anthracis spores by an integrated system of an air sampler and a cantilever immunosensor,” Sensors and Actuators B 127, 376-382 (2007). |
Maraldo, et al., “Method for Label-Free Detection of Femtogram Quantities of Biologics in Flowing Liquid Samples,” Anal. Chem. 79, 2762-2770 (2007). |
Maraldo, et al., “Detection and confirmation of Staphylococcal enterotoxin B in apple juice and milk using piezoelectric-excited millimeter-sized cantilever sensors at 2.5 fg/mL,” Anal Chem. 79, 7636-7643 (2007). |
Maraldo, et al., “Method for Quantification of a Prostate Cancer Biomarker in Urine without Sample Preparation,” Anal. Chem. 79, 7683-7690 (2007). |
Maraldo, et al., “10-Minute assay for detecting Escherichia coli O157:H7 in ground beef samples using piezoelectric-excited millimeter-size cantilever sensors.” Journal of Food Protection, vol. 70, No. 7, 1670-1677 (2007). |
Maraldo, et al., “Preapration-Free Method for Detecting Escherichia coli O157:H7 in the Presence of Spinach, Spring Lettuce Mix, and Ground Beef Particulates,” Journal of Food Protection, vol. 70, No. 11, 2651-2655 (2007). |
Rijal, et al., “PEMC-based method of measuring DNA hybridization at femtomolar concentration directly in human serum and in the presence of copious noncomplementary strands,” Anal. Chem., 79, 7392-7400 (2007). |
Rijal, et al., “Method for measuring the Self-Assembly of Alkanethiols on Gold at Femtomolar Concentrations,” Langmuir, 23, 6856-6863 (2007). |
Wilson, et al., “Viscosity and density values from excitation level response of piezoelectric-excited cantilever sensors,” Sensors and Actuators A 138, 44-51 (2007). |
Gu, et al., “Single-Calcination Synthesis of Pyrochlore-Free 0.9Pb (Mg1/3Nb2/3)O3-0.1PbTiO3 and Pb(Mg1/3Nb2/3) O3 Ceramics using a Coating Method,” J. Am. Ceram. Soc., 86 [2] 217-21 (2003). |
Thaysen, et al., “Cantilever-Based Bio-Chemical Sensor Integrated in a Microliquid Handling System,” 401-404 (2001). |
Li, et al., Micromachined Biomimetic Sensor Using a Modular Artificial Hair Cells, pp. 1-3. |
Thaysen, “Label free Detection, BioMEMs Materials and Fabrication Methods,” Track 2, 3:00pm, pp. 1-3, Sep. 7, 2002. |
Lee, C. et al., “Sol-gel derived PZT force sensor for scanning force microscopy”, Mater. Chem. Phys., 44: 25-29 (1996). |
Lee, C. et al., “Self-excited piezoelectric PZT microcantilevers for dynamic SFM—with inherent sensing and actuating capabilities”, Sensors and Actuators, A72: 179-188 (1999). |
Lee, J. H. et al., “Label free novel electrical detection using micromachined PZT monolithic thin film cantilever for the detection of C-reactive protein”, Biosensors and Bioelectronics, 20: 269-275 (2004). |
Lee, J. H. et al., “Effect of mass and stress on resonant frequency shift of functionalized Pb(Zr0.52Ti0.48)O3 thin film microcantilever for the detection of C-reactive protein”, Appl. Phys. Lett., 84(16): 3187-3189 (2004). |
Lee, J. H. et al., “Immunnoassay of prostate-specific antigen (PSA) using resonant frequency shift of piezoelectric nanomechanical microcantilever”, Biosensors and Bioelectronics, 20: 2157-2162 (2005). |
Lee, S. S. et al., “Self-Excited Piezoelectric Cantilever Oscillators”, The 8th International Conference on Solid-State Sensors and Actuators, and Eurosensors IX, Stockholm, Sweden: 417-420 (1995). |
Lee, Y. et al., “A Piezoelectric Micro-Cantilever Bio-Sensor Using the Mass-Microbalancing Technique With Self-Excitation”, The 13th International Conference on Solid-State Sensors, Actuators, and Microsystems, Seoul, Korea: 644-647 (2005). |
Li, S. et al., “The intrinsic nature of nonlinear behavior observed in lead zirconate titanate ferroelectric ceramic”, J. Appl. Phys., 69(10): 7219-7224 (1991). |
Li, X. et al., “Detection of water-ice transition using a lead zirconate titanate/brass transducer”, J. Appl. Phys., 92(1): 106-111 (2002). |
Lin, Z. et al., “Operation of an Ultrasensitive 30-MHz Quartz Crystal Microbalance in Liquids”, Anal. Chem., 65(11): 1546-1551 (1993). |
Liu, W. et al., “Preparation and orientation control of Pb1.1(Zr0.3Ti0.7)O3 thin films by a modified sol-gel process”, Mat. Lett., 46: 239-243 (2000). |
Luo, H. et al., “Synthesis of PMN and 65PMN-35PT Ceramics and Films by a New Suspension Method”, Perovskite, Piezoelectric, and Dielectric Ceramics: 251-260. |
Luo, H. et al., “Comparison in the Coating of Mg(OH)2 on Micron-Sized and Nanometer-Sized Nb2O5 Particles”, Int. J. Appl. Ceram. Technol., 1(2): 146-154 (2004). |
Luo, H., “Colloidal Processing of PMN-PT Thick Films for Piezoelectric Sensor Applications”, A Thesis Submitted to the Faculty of Drexel University in Jun. of 2005. |
Maki, K. et al., “Evaluation of Pb(Kr,Ti)O3 Films Derived from Propylene-Glycol-Based Sol-Gel Solutions”, Jpn. J. Appl. Phys., 39(9B): 5421-5425 (2000). |
Maraldo, D. et al., “Resonant-mode millimeter sized cantilever biosensor for continuous detection of proteins and pathogens in flowing liquids,” Dept. of Chem. and Biological Eng., 1-21. |
Matsui, Y. et al., “Highly Oxidation-Resistant TiN Barrier Layers for Ferroelectric Capacitors”, Jpn. J. Appl. Phys., 36(3B): 1586-1588 (1997). |
Mazza, E. et al., Biomechanics, http://www.zfm.ethz.ch/e/res/bio/, 1-10. |
McGovern, J.P. et al., “Real-Time Salmonella Detection Using Lead Zirconate Titanate-Titanium Microcantilevers”, Mater. Res. Soc. Symp. Proc., 845: AA3.8.1-AA3.8.6 (2005). |
Mueller, V. et al., “Nonlinearity and scaling behavior in donor-doped lead zirconate titanate piezoceramic”, Appl. Phys. Lett., 72(21): 2692-2694 (1998). |
Mulvihill, M. L. et al., “The Role of Processing Variables in the Flux Growth of Lead Zinc Niobate-Lead Titanate Relaxor Ferroelectric Single Crystals”, Jpn. J. Appl. Phys., 35(7): 3984-3990 (1996). |
Niedziolka, J. et al., “Charaterisation of gold electrodes modified with methyltrimethoxysilane and (3-mercaptopropyl)trimethoxysilane sol-gel processed films”, J. Electroanalytical Chem., 578: 239-245 (2005). |
Nguyen, L. T. T. et al., “Synthesis and characterization of a photosensitive polyimide precursor and its photocuring behavior for lithography applications”, Optical Materials, 29: 610-618 (2007). |
Oden, P. I. et al., “Viscous drag measurements utilizing microfabricated cantilevers”, Appl. Phys. Lett., 68(26): 3814-3816 (1996). |
Ohnmacht, M. et al., “Microcoils and microrelays—an optimized multilayer fabrication process”, Sensors and Actuators, 83: 124-129 (2000). |
Park, G.T. et al., “Measurement of piezoelectric coefficients of lead zirconate titanate thin films by strain-monitoring pneumatic loading method”, Appl. Phys. Lett., 80(24): 4606-4608 (2002). |
Park, S.E. et al., “Ultrahigh strain and piezoelectric behavior in relaxor based ferroelectric single crystals”, J. Appl. Phys., 82(4): 1804-1811 (1997). |
Piezo Systems, Inc., “Piezoceraminc Sheets and Their Properties”, Piezo Systems, Inc. Catalog: 1-3 (2007). |
Pons, T. et al., “Solution-phase single quantum dot fluorescence resonance energy transfer”, J. Amer. Chem. Soc., 128(47): 15324-15331 (2006). Abstract Only. |
Ren, W. et al., “Non linear strain and DC bias induced piezoelectric behaviour of electrostrictive lead magnesium niobate-lead titanate ceramics under high electric fields”, J. Phys. D: Appl. Phys., 35: 1550-1554 (2002). |
Ren, W. et al., “Nonlinear behavior of piezoelectric lead zinc niobate-lead titanate single crystals under ac electric fields and dc bias”, Appl. Phys. Lett., 83(25): 5268-5270 (2003). |
Rosenberg, RD et al., “Effects of age, breast density, ethnicity and estrogen replacement therapy on screening mammographic sensitivity and cancer stage at diagnosis: review of 183,134 screening mammograms in Albuquerque, New Mexico”, Radiology, 209(2): 511-5118 (1998). Abstract Only. |
Saito, Y. et al., “Lead-free piezoceramics”, Nature, 432: 84-87 (2004). |
Schemmel, A. et al., “Single molecule force spectrometer with magnetic force control and inductive detection”, Rev. Sci. Instrum., 70(2): 1313-1317 (1999). |
Shen, Z. et al., “Microfabrication of Miniaturized PZT/SiO2 Piezoelectric Microcantilever for Rapid, Direct, In-situ Biosensing”, MRS Fall Meeting, Boston: 1-23 (2005). |
Shen, Z. et al., “Self-exciting, self-sensing PbZr0.53Ti0.47O3/SiO2 piezoelectric microcantilevers with femtogram/Hertz sensitivity”, Appl. Phys. Lett., 89: 023506-1-023506-3 (2006). |
Shih, W. et al., “Simultaneous liquid viscosity and density determination with piezoelectric unimorph cantilevers”, J. Appl. Phys., 89(2): 1497-1505 (2001). |
Shih, W. et al., “Ultrasensitive Pathogen Quantification in Drinking Water Using Highly Piezoelectric Microcantilevers”, Amer. Chem. Soc., Chapter 23, 179-185 (2005). |
Shih, W. et al., “Nanosensors for Environmental Applications”, Nanotechnologies for the Life Sciences, 5: 271-293 (2005). |
Straub, V. et al., “Contrast Agent-Enhanced Magnetic Resonance Imaging of Skeletal Muscle Damage in Animal Models of Muscular Dystrophy”, Magn. Reson. Med., 44: 655-659 (2000). |
Thompson, W. R. et al., “Hydrolysis and Condensation of Self-Assembled Monolayers of (3-Mercaptopropyl) trimethoxysilane on Ag and Au Surfaces”, Langmuir, 13: 2291-2302 (1997). |
Thundat, T. et al., “Detection of mercury vapor using resonating microcantilevers”, Appl. Phys. Lett., 66(13): 1695-1697 (1995). |
Tslonsky, M. et al., “Sol-Gel-Derived Ceramic-Carbon Composite Electrodes: Introduction and Scope of Applications”, Anal. Chem., 66: 1747-1753 (1994). |
Tu, Y. L. et al., “A study of the effects of process variables on the properties of PZT films produced by a single-layer sol-gel technique”, J. Mater. Sci., 30: 2507-2516 (1995). |
Udayakumar, K. R. et al., “Thickness-dependent electrical characteristics of lead zirconate titanate thin films”, J. Appl. Phys., 77(8): 3981-3986 (1995). |
Wang, Q.M. et al., “Nonlinear piezoelectric behavior of ceramic bending mode actuators under strong electric fields”, J. Appl. Phys., 86(6): 3352-3360 (1999). |
Wang, Y. et al., “Tactile Mapping of Palpable Abnormalities for Breast Cancer Diagnosis”. |
Ward, M. D. et al., “In Situ Interfacial Mass Detection with Piezoelectric Transducers”, Science, 249: 1000-1007 (1990). |
Wellman, P. S. et al., “Breast Tissue Stiffness in Compression is Correlated to Histological Diagnosis”, http://biorobotics.harvard.edu/pubs/mechprops: 1-15. |
Wellman, P. S. et al., “Tactile Imaging of Breast Masses”, Arch. Surg., 136: 204-208 (2001). |
File History for U.S. Appl. No. 12/238,639. |
File History for U.S. Appl. No. 12/837,590. |
File History for U.S. Appl. No. 13/268,225. |
File History for U.S. Appl. No. 14/017,508. |
Number | Date | Country | |
---|---|---|---|
20140352448 A1 | Dec 2014 | US |
Number | Date | Country | |
---|---|---|---|
60573869 | May 2004 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14017508 | Sep 2013 | US |
Child | 14459993 | US | |
Parent | 13268225 | Oct 2011 | US |
Child | 14017508 | US | |
Parent | 12837590 | Jul 2010 | US |
Child | 13268225 | US | |
Parent | 12328639 | Dec 2008 | US |
Child | 12837590 | US | |
Parent | 11136173 | May 2005 | US |
Child | 12328639 | US |