This invention relates to the collection and subsequent preparation low volume samples for optical assay, such as biological samples.
Accurate and rapid counting of cells or cellular moieties in biological fluids is a necessity in the biomedical, pharmaceutical, environmental and other fields. It is becoming increasingly important to be able to perform rapid on site quantitative testing. For example, in Africa the majority of AIDS infected patients are in remote villages which can be 8 to 10 hours on foot from test clinics in major cities. Consequently, patients do not get diagnosed and do not get treatment for their disease. It is important to be able to test these patients rapidly, at low cost at their villages. On site testing where there are limited lab resources is a challenge; everything needed for the test must be provided on site. For AIDS diagnosis an absolute CD4 count is required and treatment begins when the cell count is <200 cells/uL.
If only a few uL are available for assay, accurate enumeration of cells is difficult for many 5 analytical systems.
U.S. Pat. No. 6,852,527 discloses an apparatus and method to accurately measure the number of cells in biological fluids with a low or very low cell concentration or with low volumes.
The general requirement for assays includes the following steps:
1. Withdraw sample from a sample source.
2. Dispense the sample into a reaction container.
3. Dispense some combination of reagents into the assay container, including some type of detection agent.
4. If required, concentrate or dilute the sample.
5. Assay the reaction mixture and count resulting detected targets (e.g., labeled cells.)
The drawbacks of such a system include:
There exists a need for a simplified device and system for the preparation of samples for on-site testing where lab resources are not available.
The disclosed embodiments include a sample preparation device including a cartridge housing having a capillary tube held within the cartridge housing and having an open capillary tube end extending from one end of said cartridge housing. The capillary tube has adherent reagent at a selected concentration for the capillary volume of the capillary tube. The embodiments also utilize a resilient dispenser dockable onto the cartridge housing for dispensing the capillary tube. This resilient dispenser may, for example, be a separate dispense module that a user may bring into contact with the housing. Alternatively, the resilient dispenser may be a dispense bulb mounted on the housing. The housing also includes an integrated dock shaped and positioned to allow the resilient dispenser to dock with the housing and dispense a fluid from said capillary. This dock may be shaped and proportioned like the upper end of a pipettor tip, to allow use of standard pipettors in dispensing liquid from the housing. Alternatively, the integrated dock may be detents on a housing section or plunger mounted to the housing section.
In some embodiments the resilient dispenser includes an internal membrane and a liquid within the resilient dispenser by the internal membrane. Such an internal membrane is able to be ruptured by pressing on the resilient dispenser such that liquid flow from a resilient dispenser opening.
In other embodiments the housing also includes a lance held on the housing such that a lance tip extends from the housing. Such a lance may be retractable. The device may also include a hinged safety flap movable from a first position in which said safety flap covers the lance tip and the capillary tube open end and a second position in which the lance tip and the capillary end are exposed. The housing also may include a means for retaining the safety flap.
In another embodiment the resilient dispenser includes a plunger having an apical bulb. Lowering an upper section of the plunger into a lower section of the housing (such as a retaining barrel of the housing) allows fluid communication through an upper open end of the capillary tube. The bulb may then be deformed, as by squeezing the bulb, to displace the contents in the capillary tube if said plunger has been lowered. If such a bulb contains a liquid, then the liquid will act to displace the contents of the capillary tube.
The disclosed embodiments provide a simple low cost means to collect samples combined with reagents and dispense into an optical analytical system. The disclosed embodiments enable rapid quantitative testing to be done at patient side, in the field and where there are no lab facilities.
To effectively perform patient sample testing as well as other testing, all sample preparation steps, such as patient sample collection and sample preparation with assay reagents, must be done on site. Conventional sample preparation method using pipettes, reagents and other lab supplies to prepare patient sample prior to analysis is cumbersome and wasteful and requires a skilled technician. Present embodiments provide a pre-analysis cartridge that integrates all necessary functions and reagents required for sample preparation in a compact disposable cartridge to make patient side testing simple and effective.
One embodiment is shown in
1. Possible dispensing using a pipettor tool.
2. Manipulation of the cartridges using robotic tools designed to manipulate pipette tips.
3. Use of port 12 as a mixing chamber. A pipettor device with an attached tip would dispense a dilution or wash solution into the port. The pipettor would eject the tip, seal onto the port, and then could be used to mix the sample and wash solution. The pipettor then may dispense the solution from the opposite end of the capillary dimension tube.
In this manner the sample collection cartridge would function as a specialized pipette tip. Like pipette tips, these devices could be distributed in sealed boxes, and fabricated from plastic material allowing the device to be sterilized (as by using an autoclave for steam/pressure sterilization or other means).
This device could also have a wash reagent module 22 having a dispense tip 20 which fits into docking port 12. The capillary dimension tube 14 in this embodiment terminates at the bottom of docking port 12, as is shown in
In some embodiments the capillary dimension tube 14 is coated with a dried reagent for performing a desired assay. For example, the capillary dimension tube 14 could be coated with EDTA, an anti-CD4 antibody/Alexa 610 conjugage (PE, Foster City, Calif.) and an anti CD14 antibody/Cy 5.5 conjugate (PE, Foster City, Calif.). Once the sample is introduced into the tube, the dried reagent goes into solution. Incubation may take place within the tube in about two minutes. The sample may then be dispensed directly into the analysis cartridge, slide, injection post, or other labware or instrument onto which the sample is dispensed.
With reference to
The lance 18 is shown mounted on a lance retraction slide 17. Lance retraction slide 17 may move along lance retraction guide 19. This allows lance 18 to be fully retracted into the interior of housing 10.
The integrity of the lance 18 and capillary dimension tube 14 is protected by safety flap 16. Safety flap 16 is attached to housing 10 by hinge 16A. As shown in
The drawing of
In another alternative the wash reagent module 22 is entirely filled with a diluent. Upon docking of dispense tip 20 with docking port 12 and squeezing the resilient sides of wash reagent module 22 the fluid from wash reagent module 22 would wash the sample and sample reagents into a receiving reservoir of the sample assay device as shown in
With reference to
Again with reference to
With reference to
An alternative embodiment is shown in
With reference to
With respect to
To dispense the sample the bulb 102 is pressed as shown in
There are two sections to the cartridge assembly, the body which holds the capillary for sample collection and the plunger in which the reagent is encapsulated.
For the white cell count preparation cartridge, a 5 uL capillary is used for sample collection. 20 uL of Popidium Iodide (PI) is loaded into the reagent chamber (as seen in
Using the capillary a precise amount of sample is drawn. Then the barrel and plunger are collapsed together. This action inserts the capillary into the reagent chamber. The bulb is then compressed to dispense the reagent and the sample on to the Inovx Test Cartridge. The PI stains the nucleus of the white cells collected on the surface of the test cartridge. The test cartridge is then loaded into the optical analysis unit for analysis.
For the Red Cell count preparation cartridge, a 5 uL capillary is used for sample collection. 20 uL of PBS is loaded into the reagent chamber, as illustrated in
There are two options for Red cell testing:
1. Pipette 10 uL of cerebral sample spinal fluid into a tube. Add 10 uL of Glycophorin/PE-Cy5. Incubate for about 3 minutes. Draw incubated sample into the cartridge capillary. Wait 2 minutes. Using the bulb, dispense PBS wash and sample onto the Test cartridge for testing in the analyzer.
2. In this option, the reagent is initially dried in the cartridge capillary. 20 uL of PBS is loaded into the reagent chamber. Draw CSF sample into the capillary. The sample mixes with the pre-dried reagent as it enters the capillary. Incubate for 2-3 minutes. Collapse the body and plunger together. Dispense the PBS and incubated sample on to the Test cartridge for analysis in the analysis system.
The use of the collection cartridge requires a simple manual step to complete the assay.
Typical pre-analysis steps needed for quantitive analysis are:
1. Draw patient blood sample
2. Pipette desired volume of sample
3. Pipette reagents to sample
4. Incubate for a length of time (10 mins.)
5. Additional reagent steps, (wash, lyse, etc.)
6. Load sample onto a slide or other labware and place in an analyzer for analysis. (Alternately, the cell sample may be injected into an analytical instrument, (e.g., flow cytometer)).
The method using the disclosed embodiment:
1. The disclosed cartridge provides a retractable lancet for finger stick to obtain patient blood sample
2. It includes a micro channel capillary to collect precise amount of sample. The capillary fills by capillary action and stops filling at the end of the length. Thus, the length of the capillary defines the volume of sample captured, without need for any additional mechanism to drive sample flow.
3. The reagents needed for the assay are dried within the capillary and mixes with the blood as it fills the capillary.
4. Incubation time is greatly reduced (2 mins.) from (10 mins.) for conventional method (in a test tube). The need for pipetting or other processes is also reduced.
5. An additional reagent pouch is provided for lyse, wash or other solutions/diluents if needed.
6. As shown in the embodiment of
7. The single use/single assay cartridge can be safely disposed. Minimization of the cartridge material requirements makes risks of contamination lower, and reduces the burden of proper disposal.
The cartridge can be configured for a variety of specific assays. Any reagent that is dryable onto the surface of the tube is adaptable, as is any detectable label. These could include optical detection chemestives including fluorescent, colorimetric, luminescent assays, or radiological and other labels. Capillary volume can be tailored for each test. It could also be possible to use multiple capillary tubes in a single cartridge if additional volume of sample is needed. The cartridge could also include multiple capillaries for multiple assays on the patient sample. Thus a single lance would provide enough blood for several tests.
Each test uses a minimum amount of sample compared to a typical blood draw. Minimized sample requires minimum amount of reagent. This lowers cost per test. The cartridge is made of inexpensive materials (the housing is plastic) and be made at low cost. Self-contained single use dry reagent cartridge enables convenient distribution and storage, and could be ideal for use in remote locations where storage of reagents in a controlled environment is not available. No refrigeration is required. This design reduces the risks of bio-hazards.
The disclosed system allows development of a variety of tests using the same modular components. Wash modules and capillary tube unit could be sold separately, wash modules may be used for multiple assays. Other known cartridges in current use commonly use the analyzer to actuate valves and piercing mechanisms to obtain a precise amount of sample, dispense reagents and transport sample to analysis location. This approach makes the cartridge complicated, expensive and unreliable and not easily adaptable for other tests. It also makes the analyzer more complicated and expensive. In contrast the disclosed embodiments are inexpensive components and do not require the instrument to provide any sample prep functions.
Obtaining blood sample from a finger stick eliminates the need for a trained nurse to be present for a puncture blood draw.
Using a minimum volume of blood sample, i.e., 1 uL has added advantages since it reduces the amount of reagents needed for the assay consequently reduces the cost per test. Small sample size also presents a detection challenge particularly for samples with low concentration of cells.
The cartridge may be dispensed into an assay device, such as the once disclosed in U.S. Pat. No. 6,852,527, hereby incorporated for all purposes herein. The invention could be characterized in a number of ways, including:
1. A device for obtaining a sample. A capillary is used to capture a precise amount of sample, and within the capillary a dried reagent is included in an amount sufficient to identify targets within the sample.
2. The device above, also including a reagent module configured to dock with the collection cartridge, and also configured to allow manual dispensing capillary by the reagent module. The reagent module could be the illustrated squeeze bulb, a syringe, or other device.
3. A method to capture a desired volume of sample and incubate this sample with reagents that label targets within the sample.
4. The method above, further including dispensing the sample into an analytical device that measures all targets within a sample. This device could be the assay device of U.S. Pat. No. 6,852,527. This device allows all the targets to be visualized at one optically level surface, providing all targets in a field of view. Further, the device has, a well over the sample collection membrane. Incubation could take place within this well. The suction requires to move the targets onto the membrane could be simple centrifugal rotation or vacuum force. A simple suction bulb should be able to draw sample onto this membrane. The assay measures the number of targets in a precise volume. The membrane could be system.
This application claims priority from provisional application Ser. No. 61/405,596, filed Oct. 21, 2010.
Number | Date | Country | |
---|---|---|---|
61405596 | Oct 2010 | US |