The present apparatus relates, generally, to non-destructive testing of weld integrity and strength in the attachment weld of pad eyes and other lifting lugs.
Presently, the weld integrity of pad eye or other lifting lug welds are only tested by x-rays or liquid penetrant. This testing is at best random and cannot insure the safety or reliability of the pad eyes especially after many cycles. The failure of the pad eyes can cause equipment damage and destruction as well as compromise the safety of workers and by-standers. In particular, when drill string piping is off loaded, from a barge or supply boat, the failure of the pad eyes does cause the loss of human life due to the extreme weight of the pipe and its uncontrolled fall.
Pad eyes and lifting lugs are primarily used as an attachment point for any rigging employed to hoist, transport, or secure heavy equipment. These pad eyes are typically welded either to the equipment or to some device on which the equipment is transported. The strength of these welds cannot be easily tested after they have been manufactured. Usually, the only indication of weakness is discovered upon the complete failure of the attachment weld.
Currently, there are similar approaches to the present device disclosed in other patents. However, there is no prior art for the method or apparatus for testing the pad eye welds. U.S. Pat. No. 4,676,110 issued to James Hodo, et al., discloses a fatigue testing apparatus. However, this apparatus utilizes a method of destructive testing which would render the pad eye useless. Other prior art for pull testing is disclosed in U.S. Pat. Nos. 5,844,142 and 5,918,284, both issued to John Blanch, et al. However, these systems are not portable, are not for larger loads, and are only intended for testing the products during manufacturing. These systems are also used to test the strength for one-time use only products, such as surgical sutures and needles. The pad eye welds must withstand a vast number of loading cycles, with a varied amount of loads, throughout their useful life.
There are other prior art testing tools such as disclosed in U.S. Pat. No. 6,186,011 B1, issued to Pey Min Wung, et al., which tests the failure modes of spot welds on sheet metal. Another testing tool disclosed in U.S. Pat. No. 6,216,531 B1, issued to Joe Zhou, tests the shear strength of adhesive bonded materials. However, both of these inventions are based on pre-manufacture testing, do not consider cyclic loading over the useful life of the product, and cannot be adapted for portability. These testing tools also cannot be adapted to perform testing of finished products or to test the weld integrity before each use.
It is thus a desire to have a testing apparatus which is portable and can quickly and accurately check the integrity of a pad eye and its attachment weld before each field use. The desired apparatus should be portable, self-contained, easily transportable, and environmentally safe in order to test the pad eye welds at almost any location. This testing device should be capable of being hydraulically operated as well as by other available pressurized fluid sources. This device could consist of one or multiple pressurized fluid cylinders depending on the required test loads and the configuration of the apparatus. The effective area of the cylinder piston and the pressure applied to the cylinder would determine the capacity of the apparatus. The fluid pressure is preferably supplied by a hand pump for currently optimum portability; however, other types of pumps could be utilized. A flow manifold would be needed to control the flow direction as well as measure the pressure applied to the cylinder(s). One end of the cylinder(s) would be attached to the same base as is the pad eye or attached to its own base plate. The other end is attached to a cross bar, bridge plate, or similarly functioning member. The cross bar, bridge plate, or similarly functioning member could be further supported by either cylinders or support beams.
For a further understanding of the nature of the present device, reference should be had to the following detailed description, taken in conjunction with the accompanying drawings, in which like elements are given the same or analogous reference numbers and wherein:
It should be appreciated that the footings 9 and the attachment plates 17 are an illustrative method of attaching the apparatus to the pad eye 19 or lifting lug. Other ways of attachment which provide adequate support and connection can be employed without departing from the scope thereof. The test apparatus can be mounted to a base plate 37 in order to provide support for the test apparatus. As seen in
As shown in
Each of the two bridge plates 1 is preferably substantially identical flat rectangular bars made of steel or a material with similar strength properties. These said bridge plates 1 have a plurality of apertures 23 which are used for the fixed connections between the two bridge plates 1 and the cylinder shaft 13 and the two support beams 3. It should be appreciated that the hole 23 diameter and consequently the diameter of the pin 39 varies in size depending on the load which will be tested and are typically sized by calculation based on the maximum load contemplated. Preferably, there is a plurality of apertures 23 located on the right and left lengthwise ends of the bridge plate 1 and also located substantially symmetrically around the lengthwise and widthwise centers of the said bridge plate 1. Preferably, said apertures 23, located on the right and left lengthwise ends, should be substantially symmetrically located at each end of the bridge plates. Preferably, said apertures 23, at each end of said bridge plate 1, are substantially symmetrically located with respect to said bridge plate 1 and each other. The apertures 23 located approximately midway of both the lengthwise and widthwise centers of said bridge plates 1 are preferably substantially symmetrically located at said midway point. The multiple apertures 23 are used for allowing vertical and horizontal adjustment of the test apparatus. The cylinder shaft 13 is preferably fixedly attached at substantially the lengthwise midpoint of the bridge plate 1 by pin 39. The two bridge plates 1 are fixedly attached at opposing ends, left and right side, and to the outside face of the two support beams 3 by pin 39.
The two support beams 3 are preferably hollow rectangular beams made of steel or a material with similar strength properties. These said support beams 3 have a plurality of apertures 23 on all four faces of the rectangular beam which are used for the fixed connections between the two support beams 3 and the two bridge plates 1 and the two footings 9. It should be appreciated that the hole 23 diameter and consequently the diameter of the pin 39 varies in size depending on the load which will be tested and are typically sized by calculation based on the maximum load contemplated. Preferably, the apertures 23, on each face of the support beams 3, are substantially centered lengthwise on each face of the support beam 3 and are substantially symmetrically spaced from the top of said support beam 3. The multiple apertures 23 are used for allowing vertical and horizontal adjustment of the testing apparatus. The bottom end of the two support beams 3 has a hole on each of the four faces. Preferably, the said apertures 23 are substantially symmetrically located on the lengthwise centerline of each face of the said support beam 3. As described above and shown in
The two footings 9 are preferably comprised of a steel, or a material with similar strength properties, channel bar. Preferably, the two channel sides have matching apertures 23 on each face. Preferably, said hole diameters are substantially symmetrically centered relative to the said channel bar height and substantially symmetrically spaced between the respective apertures centers on each channel wall face. The two footings 9 will preferably be attached by welding to the test apparatus base plate 37 or attached by clamps to the pad eye base plate 21. Preferably, the two footings 9 are substantially horizontally and symmetrically centered on each side of the pad eye 19. This said placement, of the two footings 9, is assured through the substantially symmetric connection of the support beams 3 to bridge plate 1. This said positioning insures that the testing apparatus will provide an approximately equal upward force on the pad eye 19.
The two attachment plates 17 are comprised of two substantially identical pieces of flat steel, or a material with similar strength properties, bar. The attachment plates 17 have a plurality of apertures 23. The hole 23 diameters are substantially centered along the lengthwise centerline of the flat face of the attachment plates 17. The said apertures 23 are further substantially symmetrically located at opposing lengthwise ends of the said attachment plates 17. The attachment plate 17 is fixedly attached on the upper end, by pin 39, to the end of the bottom adaptor 15. The said bottom adaptor 15 is fixedly attached, by pin 39, pinned between the two attachment plates 17. The lower end of said attachment plates 17 is attached to the pad eye 19 by pin 39. The pad eye 19 is sandwiched between the two attachment plates.
An alternative embodiment of this structure, shown in
It should be appreciated that the footings 9 and the attachment plates 17 are an illustrative method of attaching the apparatus to the pad eye 19 or lifting lug. Other ways of attachment which provide adequate support and connection can be employed without departing from the scope thereof. The test apparatus can be mounted to a base plate 37 in order to provide support for the test apparatus, it can be used without the base plate 37 and attached via clamps 38 or other suitable method directly to the base support structure to which the pad eye 19, lifting lug, or other lifting connection is attached.
As shown in
Each of the two bridge plates 1 is preferably substantially identical flat rectangular bars made of steel or a material with similar strength properties. These said bridge plates have a plurality of apertures 23 which are used for the fixed connections between the two bridge plates 1 and the cylinder shafts 13, of both cylinders 5, and the support beam 47. It should be appreciated that the hole 23 diameter and consequently the diameter of the pin 39 varies in size depending on the load which will be tested and are typically sized by calculation based on the maximum load contemplated. Preferably, there are a plurality of apertures 23 located on the right and left lengthwise ends of the bridge plate 1 and also located substantially symmetrically around the lengthwise and widthwise centers of the said bridge plates 1. Preferably, the said apertures 23, located on the right and left lengthwise ends, should be substantially symmetrically located at each end of the bridge plates 1. Preferably, said apertures 23, at each end of said bridge plate 1, are substantially symmetrically located with respect to said bridge plate 1 and each other. The said apertures 23 located approximately midway of both the lengthwise and widthwise centers of said bridge plates 1 are preferably substantially symmetrically located at said midway point. The multiple apertures 23 are used for allowing vertical and horizontal adjustment of the test apparatus.
The support beam 47 is preferably a fabricated hollow rectangular beam, having a top side and a bottom side, in which two opposing bottom sides extend, in the lengthwise direction, beyond the other two opposing sides. The said opposing extended sides form a channel at the bottom end of the said support beam 47. The support beam 47 has a plurality of apertures 23 on all four faces of the rectangular beam which are used for the fixed connections between the support beam 47 and the two bridge plates 1 and the two footings 9. It should be appreciated that the hole 23 diameter and consequently the diameter of the pin 39 varies in size depending on the load which will be tested and are typically sized by calculation based on the maximum load contemplated. Preferably, said apertures 23, on each face of the support beam 47 are substantially centered lengthwise on each face of the support beam 47 and are preferably substantially symmetrically spaced from the top of said support beam 47. The multiple apertures 23 are used for allowing vertical and horizontal adjustment of the testing apparatus. As shown in
The two footings 9 are preferably comprised of a steel, or a material with similar strength properties, channel bar. Preferably, the two channel sides have matching apertures 23 on each face. Preferably said hole diameters are substantially symmetrically centered relative to the said channel bar height and substantially symmetrically spaced between the respective hole centers on each channel wall face. The two footings 9 will preferably be attached by welding to the test apparatus base plate 37 or attached by clamps to the pad eye base plate (not shown). Preferably, the two footings 9 are substantially horizontally and symmetrically centered on each side of the pad eye 19. This said placement, of the two footings 9, is assured through the substantially symmetric connection of the right and left cylinders 5 to the bridge plates 1. This said positioning insures that the testing apparatus will provide an approximately equal upward force on the pad eye 19.
a is substantially similar to
It should be appreciated that one skilled in the art would recognize that the pistons utilized in all embodiments of this invention could be attached to the pad eye, lug, or test piece via the use of alternate attachment means, such as but not limited to, cables or tension wires. It should also be appreciated that one skilled in the art would recognize that cables or tension wires could be utilized in attaching pistons to a variety of frameworks including, but not limited to, support columns, bridge plates, cross bars, mounted framework, arms, brackets, or footing. It should also be appreciated that one skilled in the art would recognize that cables or tension wires could be utilized in attaching the test piece to a variety of frameworks or bridges including, but not limited to, support columns, bridge plates, cross bars, mounted framework, arms, brackets, or footing.
It should be understood that the frames in these embodiments could be modified in many aspects to achieve the substantially same function and substantially same result of the present device. For instance, the bridge plate 1 could be curved, angled, have additional members attached thereto, and the like. The structure and frame could be further modified using springs or other mechanisms to exert a tensioning force on the pad eye or lifting lug attachment weld.
Operating Apparatus
Preferably, as shown in
As shown in
In use, the present device can be transported in the disassembled condition to any location where some pad eye or lifting lug will be utilized or requires testing. One possible criterion for selecting a particular embodiment may be the required test load.
In use, prior to selecting the proper embodiment of the present device, the desired test load must be known. Typically, this value will be approximately 1.5 times greater than the rated load capacity of the pad eye or lifting lug. If such a rated load is not readily accessible, the test value would be approximately 1.5 times the weight that will be supported by the pad eye or lifting lug. It should be understood that the said multiplying factor of 1.5 is only a preferred safety factor; therefore, the value can vary depending on a user's experience or preference and should not be used as a limiting factor of the scope of the claimed apparatus. After determining the said load test value, the proper testing apparatus will be comprised of the correct size and number of cylinders 5 which can generate the required test load. The required test load is generated by the cylinder 5 and is produced by the combination of the pressure, produced by the hand pump 40 and measured by the manifold pressure gauge 73, acting over the effective area of the cylinder. The effective area of the cylinder is calculated based on the diameter of the piston 53 within the cylinder 5. As is known to those in the art, the effective area of the piston 53 can be obtained from the cylinder manufacturer. As is also well known to those in the art, the said effective area is multiplied by the pressure to predetermine the produced test load. After determining the test load as described above, selecting the cylinder size, as described above, and determining the required pressure to produce the said test load, as described above, the hand pump 40 is used to produce the required pressure.
In use, the test apparatus is assembled as shown in
Turning the directional lever 79 ninety degrees from the position described in the above paragraph will cause the fluid to flow into the upper cylinder fitting 7 when the pump 61 is actuated by the moving the pump handle 71 in alternating upwardly and downwardly directions as described above. The said pump handle 71 movement causes the pump to push fluid through one of the hoses, through the flow control manifold 50, and into the upper cylinder fitting 7. This causes the said fluid to flow above the piston 53 and begin forcing the piston 53 in a downwardly direction. As the piston 53 begins to move in the said downward direction, the fluid, resting below the piston 53 begins to be forced out of the lower cylinder fitting 7A, through the pressure containing hose connected to the lower cylinder fitting 7A, through the flow control manifold 50, through the hand pump 40, and into the reservoir 59. This described flow of the fluid and the accompanied downward movement of the piston 53 causes the cylinder shaft 13 to retract.
In use, the single cylinder version of the present device produces an upward force on the pad eye or lifting lug when the cylinder shaft 13 retracts. Therefore, the directional lever 79 shall be moved into the position which causes the fluid to flow into the upper cylinder fitting 7.
In use, the double cylinder version of the present device produces an upward force on the pad eye or lifting lug when the cylinder shaft 13 extends. Therefore, the directional lever 79 shall be moved into the position which causes the fluid to flow into the lower cylinder fitting 7A.
In use, the pressure gauge 73 will indicate the system pressure. As the pump handle 71 is moved, as described above, the pressure, indicated on the pressure gauge will increase. When the indicated pressure is approximately the same as the calculated pressure corresponding to the required test load, as described above, then the proper test load has been applied. It should be understood that this paragraph describes the operation regardless of the direction of fluid flow or the position of the directional valve handle 79.
In use, after the test is completed, the directional valve 79 should be turned ninety degrees, and the pump 61 should be actuated by the pump handle 71, as previously described, until the system pressure, as indicated by the gauge or other monitoring device, is completely relieved.
It should be appreciated that the directional lever 79 does not actually control flow but rather moves the directional valve 77 into a position that changes the flow direction. The detailed workings of the directional valve 77 are well known to those skilled in the art.
Those who are skilled in the art will readily perceive how to modify the present apparatus still further. For example, most of the illustrated connections utilize pins, however, it should be recognized that other methods of connection may be utilized, such as threaded connectors or if the unit will be modified for permanent installation as opposed to portable, the connections could be welded. Further, the frames or structures, of this apparatus, do not need to be comprised of substantially vertical and horizontal members. These members could be curved, angled, or joined in a manner to provide the substantially same function and substantially same result in testing the pad eye or other lifting lug attachment welds.
In
In
In
In
It may be seen from the preceding description that a new and improved pad eye testing apparatus and method has been provided. Although very specific examples have been described and disclosed, the invention of the instant application is considered to comprise and is intended to comprise any equivalent structure and may be constructed in many different ways to function and operate in the general manner as explained hereinbefore. Accordingly, it is noted that the embodiment of the new and improved pad eye testing apparatus and method described herein in detail for exemplary purposes is of course subject to many different variations in structure, design, application and methodology. Because many varying and different embodiments may be made within the scope of the inventive concept(s) herein taught, and because many modifications may be made in the embodiment herein detailed in accordance with the descriptive requirements of the law, it is to be understood that the details herein are to be interpreted as illustrative and not in a limiting sense.
This application is a continuation-in-part of prior application Ser. No. 10/365,105, filed Feb. 12, 2003.
Number | Date | Country | |
---|---|---|---|
Parent | 10365105 | Feb 2003 | US |
Child | 11047127 | Jan 2005 | US |