The invention relates generally to apparatus and methods for conducting analysis of three-dimensional cell samples (e.g., tissue samples).
It has long been established that distinct differences in the metabolic phenotype of cancer cells are linked to underlying mechanisms that provide a selective advantage for survival and proliferation. However, the precise mechanisms that trigger tumorgenesis are poorly understood. It has been postulated that glycolytic adaptation is a survival mechanism that allow tumors to proliferate in a microenvironment characterized by low pH and oxygen tension. These adaptations of “the Warburg shift” provide a selective advantage to the tumor through increased glucose uptake and ATP synthesis in order to meet the demands for biosynthesis, energy and reducing equivalents.
Recent advances in the development of instruments that measure the flux of key analytes indicating aerobic metabolism (O2), glycolysis (H+), and intermediary metabolism (CO2) within the microenvironment may provide insight to the underlying mechanism of malignant transformation. However, these systems are designed and optimized for use in cell-based assays, may lack environmental control, and generally do not facilitate the measurement of multicellular tissue samples because of constraints on chamber size, difficulty in immobilization and perfusion of the sample.
Seahorse Bioscience, Billerica, Mass., launched the XF96 “Extracellular Flux Analyzer” in 2007. Since that time the product has been adopted as a technology platform for making quantitative measurements of mitochondrial function and cellular bioenergetics. XF measurements are performed in a fully integrated instrument that measures the concentrations of various analytes (O2, H+, CO2) in the extracellular media of a cell based assay. Analyte concentrations are measured non-invasively, within a small volume about the cells, providing quantitative measurements for changes in analyte concentrations as a function of time from which bioenergetic flux (example: dO2/dt=oxygen consumption rate, dpH/dt=extracellular acidification rate) can be determined. XF measurements are based on a method in which a small, temporary, measurement volume is created around the cells, or a sensor is placed in close proximity to the cells. Measurements under these conditions amplifies changes in concentrations allowing highly sensitive, time resolved measurements to be collected from a set of optical sensors. Once the measurement is made, the plunger (probe) is lifted and the medium around the cells is restored to its original condition. This nondestructive method allows multiple measurements to be serially collected for a biological sample under various conditions of stimulation (basal, environmental change, compound stimulus).
By measuring key metabolic parameters such as oxygen consumption rate (OCR) and extracellular acidification rate (ECAR), a profile of the bioenergetic phenotype may be developed based on the substrate and pathway (glycolysis or oxidative phosphorylation) for generating energy and biosynthesis. The product allows quantitative measurements of mitochondrial function and cellular bio-energetics of cells.
A need exists for a system that allows for measurement of key analytes of, for example, aerobic metabolism, glycolysis and intermediary metabolism in multicellular tissue.
The instrument and methods described herein allow the measurement of bioenergetic parameters in tissue samples or cell populations, e.g., tumor tissue samples, to generate its metabolic profile, thereby enabling, for example, a better understanding of the mechanisms of malignant progression. Accordingly, embodiments of the invention enable testing of physiologically relevant hypotheses that to date could not be tested otherwise, for example, by enabling the measurement and quantification of phenotypic shifts in tumors.
Embodiments of the invention comprise an apparatus for conducting analysis of three-dimensional cell samples. The apparatus may include a plate of wells that are loaded with tissue samples which are tested by placing the plate in a machine specifically adapted as disclosed herein to conduct assays in wells of the plate. The plate typically defines a plurality of wells for holding respective samples and sample media, where at least one of the wells, typically the entire array of wells, includes a sample nesting site disposed therein. A bore is disposed above the nesting site, the bore being dimensioned to interfit with a plunger that moves vertically down within the bore and within the sample media disposed in the well. In practice it is preferred to maintain the well(s) or well plate in position on a stage or platform and robotically move the plunger(s), but it is possible but not preferred to keep the plungers stationary and move the plate. The structure also defines a media channel that is in fluid communication with the sample nesting site. The media channel permits media displaced by the plunger to flow through and expose the sample to fresh media as the plunger moves.
One or more of the following features may be included. The plunger preferably moves relative to the nesting site to induce perfusion of media about the sample, preferably on both the down and up stroke. Alternatively, the plunger and sample nesting site may move together to expose the sample to different regions of media. The media channel may include a fluid path that returns media perfused about the sample back to media disposed in the well. The bore, sample nesting site, and/or media channel may be defined by a removable well insert.
The sample nesting site may include a media permeable platform defined by or as a part of the removable well insert. The well may include a sump in fluid communication with the media channel. The sample nesting site may be disposed within the sump where media collects and flows through the media channel.
The media channel may include a fluid path defining a closed loop beneath the surface of media in the well to permit media perfusive flow about the sample on both upward and downward movement of a plunger within the bore. A check valve may be included in the media channel to inhibit backflow of spent media from the channel to the sample during upward movement of the plunger.
The apparatus may include sensors for detecting the concentration of solutes secreted from or absorbed by the sample in the media disposed about the sample. The sensors may be disposed at the bottom of at least one of the wells. The sensors may comprise fluorophores sensitive to the concentration of one or more solutes in the media mounted on the plungers to measure in regions adjacent the tissue sample, for example, oxygen, CO2 or H+ concentration. The sensors may comprise beads coated with specific binders of specific cytokines, chemokines, hormones, or other biomolecules absorbed by or secreted from the tissue which become immobilized on the beads. The beads may be probed in situ in a sump in a well to enable detection of the concentrations of molecules of interest secreted or taken up by the tissue. Alternatively, the beads may be separated from the media and analyzed. In some embodiments, in place of beads, spotted antibodies may be disposed in a bottom portion of the well in fluid connection with the media perfusing through the well. The spotted antibodies may be used to detect other molecules of interest.
The apparatus may include plungers adapted for reciprocating movement within the bores of respective wells. In a preferred embodiment, fluorophore sensors may be disposed on the plungers for detecting the concentration of solutes in media disposed about the sample. The sensors may measure the concentration of oxygen, carbon dioxide, and/or hydrogen ions dissolved in media about the sample.
The plurality of wells may define a multi-well plate including, for example, 24 or 96 wells.
A source of oxygen, carbon dioxide, and/or a biologically inert gas may be in fluid communication with media in a well or a headspace above the surface of media in the wells for controlling the composition of gas in the headspace or in the media. A source of a solution of a biologically active substance may be in fluid communication with media in wells for exposing a sample to the substance.
A three-dimensional cell growth scaffold may be disposed on the sample nesting site.
In another aspect, embodiments of the invention may include an apparatus for conducting analysis of three-dimensional cell samples. The apparatus may include a well for holding a sample and sample media, the well including a sample nesting site. A bore may be disposed above the nesting site, the bore being dimensioned to interfit with a plunger that moves vertically down within the bore and within media disposed in the well. The apparatus may also include a media channel in fluid communication with the sample nesting site that permits media perfusion about the sample, and a plunger adapted for reciprocal movement in the bore to impel media perfusion about the sample.
The apparatus may optionally include a sensor for detecting the concentration of a dissolved media component in media about the sample nesting site.
In yet another aspect, embodiments of the invention feature an insert for a well of a culture plate for adapting the well to implement perfusion of a three-dimensional cell culture sample disposed therein. The insert includes a structure defining (i) a sample nesting site comprising a media-permeable platform, and (ii) disposed thereabove a bore dimensioned to interfit with a plunger which moves vertically down within the bore and within media disposed in a well. The insert may also define a media channel in fluid communication with the sample nesting site, which permits media perfusion about the sample impelled by a plunger.
In another aspect, the embodiments of the invention feature a method of experimenting with a three-dimensional cell culture sample, e.g., a tissue sample, biopsied sample, or cell scaffold holding cells, so as to maintain viability of the sample and exercise control over its microenvironment. The method includes providing a structure defining a well including a sample nesting site, a bore dimensioned to interfit with a plunger disposed above the sample nesting site, a media channel in fluid communication with the sample nesting site, and a plunger adapted for reciprocal movement in the bore. A sample is placed on the sample nesting site in medium within the well. The plunger is moved within the bore to impel media flow about the sample and through the channel to perfuse the sample with media.
One or more of the following features may be included. A gas may be added to the media or to headspace in the well above the media to modify the microenvironment about the sample by altering dissolved gas composition. A solution of a biologically active substance such as a drug, drug candidate, or toxin may be added to the media to modify the microenvironment about the sample by exposing the sample to the biologically active substance.
The concentration of one or more solutes in media about the sample may be measured. A plurality of measurements separated in time may be made of the concentration of one or more solutes in media about the sample.
A metered amount of one or more gases and/or one or more solutes may be added to media in the well thereby setting the microenvironment in the medium about the sample to a predetermined point. The microenvironment may be set to a hypoxic condition. An oxygen scavenger may be added to the medium.
A human biopsied tissue sample may be placed on the nesting site, potential therapeutic drugs may be added to the media, and the effect of the drugs on the sample may be assessed.
The sample may be a tumor sample, and a metered amount of one or more gases and/or one or more solutes may be added to media in the well thereby to set the microenvironment in the medium about the sample to a predetermined point simulating the microenvironment of the tumor sample in vivo.
The method may be multiplexed by providing a plurality of the wells including sample nesting sites, bores dimensioned to interfit with plungers disposed above the respective sample nesting sites, media channels in fluid communication with each of the sample nesting site, and plungers adapted for reciprocal movement in each of the bores. Sample may be placed on the sample nesting sites in media within each of a plurality of wells. The plurality of plungers may be moved within the bores to impel media flow about the samples and through the channels thereby to perfuse the samples with media. The concentration of one or more solutes in media about the samples in the plurality of the wells may be measured one or more times.
Embodiments of the invention enable the measurement of one or more properties of one or more three-dimensional cell samples (e.g., tissue samples, biopsied samples, or cell scaffolds holding cells) that are disposed in, for example, one or more wells of a multiwell plate. Plates of various specific forms embodying the invention and well inserts that adapt more conventional plates to structures embodying the invention may be designed by the skilled artisan in view of this disclosure, and may be manufactured using conventional molding techniques from polymeric materials such as polyethylene terephthalate or polystyrene.
These elements of embodiments of the invention will now be discussed in detail. The bottom surface of the well 110 may define a depression that acts as a sump during use. The depression may be used to orient and control the positioning of samples in the well and its bottom surface to serve as a sample nesting site 150. The well may be one of many wells of a multiwell plate, that may be designed to a standard “Society for Biological Screening” SBS footprint having a 6 mm well diameter with the depression at the bottom of the well having a depth of, e.g., 0.5 mm and a diameter of, e.g., 3 mm. The dimensions of the depression may be selected in view of the analysis to be performed in the well. For example, the volume should be sufficient to hold a sample. The exemplary dimensions indicated above are suitable for holding a sample that is approximately 300 μm thick. Other dimensions may be provided, depending on analytical needs.
The removable insert 100 may be generally cylindrical, and sized and configured to slide into the well with a slight interference fit to its interior wall. See
One currently preferred set of dimensions would define a 3 mm plunger stroke within a 3 mm bore so that the displaced volume of the pump (a bit over 20 mm3 or 20 μl) is approximately 4 times the volume of the channels. Accordingly, as currently contemplated, the plunger and insert are sized such that the plunger moves at least 20 μl of media when inserted into the bore. The smaller the clearance between bore and plunger the better the pump efficiency, so a minimum clearance of 0.001 inch and a maximum of 0.01 inch is a reasonable range. Larger clearances are contemplated in some embodiments to permit fresh media to flow downwardly around the piston on its upward stroke, particularly in embodiments including a check valve function as described below.
Referring to
As discussed hereinafter in more detail, including with reference to
Media channels that may be defined by the insert permits media displaced by the plunger to leave the proximity of the sample, e.g., by flowing up the perimeter walls through the annulus defined by an outer surface of the insert and an inner sidewall of the well. In some embodiments (not shown) the media channel may extend from a bottom portion of the insert to a higher portion of the insert. Referring to
Referring to
Each of the media channels may include a fluid path defining a closed loop beneath the surface of media in the well, permitting media perfusive flow about the sample on both upward and downward movement of the plunger within the bore.
The insert may be fabricated by injection molding. Surface wettability may be increased by treatments known per se, e.g., performing a plasma pretreatment of the inserts to eliminate entrapment of gas bubbles in the perfusion channels by creating a more hydrophilic surface.
The bore preferably is situated above the nesting site, guides the plunger vertically down into the insert within the sample media disposed in the well, and creates hydrostatic pressure and media movement about the sample. Thus, the plunger may be adapted for reciprocating movement within the bore disposed in the well, e.g., the bore defined by the insert.
In some embodiments, the sample nesting site may include a scaffold 200 attached to the insert or to the bottom of the sump. A scaffold is a three-dimensional porous solid such as a collagen membrane that mimics the parenchyma of tissue and its surrounding structure in vivo. Such scaffolds are available commercially and may be fabricated from gels or fibrous/porous media, e.g., Alvetex® Scaffold or 3D BioTek scaffold material. Alvetex® Scaffold is a highly porous, cross-linked polystyrene scaffold that has been section into 200 μm thick membrane. The resulting material is inert and does not degrade during normal use. It has been adapted to fit a variety of conventional cell culture plastic-ware formats. Alvetex® Scaffold provides a suitable 3D structure in which cells can proliferate, migrate, differentiate, and function in an appropriate niche environment. Cells maintain a 3D shape and form close interactions with adjacent cells.
Referring to
The apparatus preferably includes sensors 180, in addition to the optional beads described above, e.g., disposed on the plungers, for detecting the concentration of solutes in media disposed about the sample. The sensors may measure the concentration of dissolved oxygen, carbon dioxide, or hydrogen ions in media about the sample. Measurements spaced in time permit assessment of the health or metabolic efficiency of the sample under various condition of its microenvironment.
Referring to
Various types of sensors may be utilized with the apparatus depending on the analysis to be performed and its selected configuration, including oxygen sensors, such as oxygen-quenched fluorescent sensors, pH sensors, including fluorescent sensors, ISFET and impedance sensors, CO2 sensors, including bicarbonate buffer coupled and ammonium dye coupled fluorescent sensors as well as other CO2 sensors; various ion and small molecule sensors; large molecule sensors including surface plasmon resonance sensors and sensors exploiting the principle of Wood's anomaly; acoustic sensors; and microwave sensors. In certain embodiments, a conventional plate reader may be used.
Preferred sensors are fluorophores. Many fluorescent sensing compounds and preparations are described in the art and many are available commercially from, for example, Molecular Probes Inc. and Frontier Scientific, Inc. The currently preferred oxygen sensor is a fluorophore with the signal inversely proportional to oxygen concentration such as a porphyrin or rhodamine compounds immobilized as a particle or homogenously distributed in an oxygen permeable polymer, e.g., silicone rubber. The currently preferred compound is porphyrin. The currently preferred pH sensor is a fluorescent indicator dye, fluorescein, whose signal decreases upon protonation of the dye, and which is either entrapped in a particle that is suspended in a carrier polymer, or covalently attached to a hydrophilic polymer. Useful fluorescent CO2 indicator sensor typically are based on a pH sensitive transducer, with the fluorescence being indirectly modulated by the production of carbonic acid due to reaction of carbon dioxide with water. See, e.g., O. S. Wolfbeis, Anal. Chem. 2002, 74, 2663-2678. A fluorophore that detects glucose also can be used, such as one based on a non-enzymatic transduction using a boronic probe that complexes with glucose, resulting in a charge transfer that modulates the fluorescence of the probe, or an enzymatic glucose transducer that couples a glucose oxidase to a fluorescent oxygen sensor, with the binding and oxidation of glucose resulting in a quantitative modulation of the oxygen sensor. It also is within the scope of embodiments of the invention to employ a fluorophore or other type of sensor sensitive to biological molecules such as, for example, lactate, ammonia, or urea. A lactate sensor can be based on an enzymatic sensor configuration, with lactate oxidase coupled to a fluorescent oxygen sensor, and with the binding and oxidation of lactate resulting in a quantitative modulation of the oxygen sensor. An ammonia or ammonium ion sensor can be configured with immobilization of a protonated pH indicator in a hydrophobic, gas permeable polymer, with the fluorescence output quantitatively modulated by reaction with transient ammonia. A urea sensor can be based on an enzymatic sensor configuration, with urease coupled to a fluorescent ammonia transducer, and with the binding and reduction of urea to ammonia, resulting in modulation of the ammonia sensor fluorescence. The nature of the sensor generally does not form an aspect of embodiments of this invention.
In use, the insert guides the plunger to provide perfusion by creating hydrostatic pressure in the column of medium above the tissue sample in the insert and/or in the depression at the bottom of the well. As the plunger reciprocates vertically through the column, medium is forced to flow across and sometimes through the tissue and exits the chamber through a series of channels around the perimeter of the insert and upwardly between the outer surface of the insert and the inner wall of the well. By moving the plunger up and down, medium is moved across the tissue, replenishing nutrients, providing oxygen, and sweeping away wastes. Accordingly, the microenvironment around the sample may be continuously perfused between measurements. As the plunger moves into the bottom position, resting on or just above the insert, its motion is stopped, the small transient volume is created, and measurements are made. Efficiency of perfusion through the insert may be increased by altering the stroke height, speed and clearances between the plunger and the insert.
Referring to
Each of the ports 460 may have a cylindrical, conic or cubic shape, open through planar element 430 at the top, and closed at the bottom except for a small hole, i.e., a capillary aperture, typically centered within the bottom surface. The capillary aperture is adapted to retain test fluid in the port, e.g., by surface tension, absent an external force, such as a positive pressure differential force, a negative pressure differential force, or possibly a centrifugal force. Each port may be fabricated from a polymer material that is impervious to gasses, test compounds, or from any other solid material. When configured for use with a multiwell microplate 400, the liquid volume contained by each port may range from 500 μl to as little as 2 μl, although volumes outside this range are contemplated.
Referring to
The cartridge 410 may be attached to the plunger, or may be located proximal to the plunger without attachment, to allow independent movement. The cartridge 410 may include an array of compound storage and delivery ports assembled into a single unit and associated with a similar array of plungers.
The apparatus may also feature a removable cover 420 for the cartridge 110 or for multiwell plate 400. The configuration of cartridge 110 as a cover for multiwell plate 400 may help prevent evaporation or contamination of a sample or media disposed in wells 110. The cover 420 may also be configured to fit over the cartridge 110 thereby to reduce possible contamination, to maintain the content of the gas in the wells, or reduce evaporation of fluids disposed in the ports 460 of the cartridge 410.
In particular, one way for achieving this is with the configuration described with respect to
In some embodiments, a source of a solution of a biologically active substance may be in fluid communication with media in wells for exposing a sample to the substance
To control the operation and timing of the solenoid valve, the instrument software may be modified to facilitate control of the valve/timing and to expose some of the calculation variables used during calibration. For example, to calculate molar concentration of O2 in the medium, the concentration at calibration is preferably known and input into the calculation table. Under some conditions the initial calibration value (F or current ambient concentration) may not be known. In this case, calibration and solution of equation (1) (see example 2 below) may be achieved by injecting sodium sulfite into a set of control wells and calibrating the system based on a known F0 value. To calculate these results, certain coefficients may be made accessible in the software. A separate window may be created in the software to facilitate access to these variables, valve control and calculation of calibration coefficients.
The instrument may be tested using a well characterized cell line (mouse C2C12) to verify proper operation and control of the gas system. A series of tests may be conducted to demonstrate the ability to purge O2 from medium and create a hypoxic microenvironment around the sample. These tests may include:
1. Calibration of the instrument under known and unknown ambient O2 concentrations
2. Verify performance of the gas delivery system and the ability to drive environmental O2 levels to desired value (<5% PPO). This may be verified within the instrument by looking at the O2 level data. The readout from the instrument may provide a view that presents this data.
An alternative to controlling O2 and pH within the sample environment may be to enclose the entire instrument in an environmental chamber and pump down the chamber to the desired levels. This alternative approach may be less desirable, as it may be very costly, take up a lot of lab space, and require long periods of time to achieve the desired levels around the tissue. By the time these O2 levels are achieved the tissue may be dead.
The compound storage and delivery apparatus 610 is controlled by a controller 680, that may be integrated with a computer 690, that may control the elevator mechanism, the multiplexer, and the pressure source. The controller 680 may, thereby, permit delivery of a test fluid from a port to a corresponding well when an associated sensor is disposed in the well.
The apparatus described herein is a modification of the apparatus disclosed in US 20080014571, referenced above, and enables experimentation with and analysis of three-dimensional cell culture samples, such as a tissue sample, a biopsied sample, or a cell scaffold holding cells. Viability of the sample may be maintained and control exercised over its microenvironment. In certain embodiments, a gas may be added to the media or to a headspace in the well above the media to modify the microenvironment about the sample by altering dissolved gas composition. In certain other embodiments, a solution of a biologically active substance may be added to the media to modify the microenvironment about the sample by exposing the sample to a biologically active substance. A metered amount of one or more gases and/or one or more drugs or other solutes may be added to media in the well to set the microenvironment in the medium about the sample to a predetermined point. The microenvironment in the well may be set to a hypoxic condition. The concentration of one or more solutes in media about the sample may be measured. A plurality of measurements, separated in time, of the concentration of one or more solutes in media about the sample may be taken.
In certain other embodiments, the method includes adding an oxygen scavenger such as sodium sulfite to the medium.
A human biopsied tissue sample may be placed on the nesting site, potential therapeutic drugs may be added to the media, and the effect of the drugs on the sample may be assessed.
The following examples illustrate certain exemplary and preferred embodiments and applications of the instant invention, but are not intended to be illustrative of all embodiments and applications.
The Seahorse XF96 flux analyzer was developed and optimized to measure bio-energetic activity in cell based assays. XF measurements are based on a patented method in which a small, temporary measurement volume is created around a monolayer of cells which are adherent to the bottom of the wellplate. A small volume is created when the plunger (probe) is lowered to the bottom of the well (3.8 mm diameter) and engages a set of standoffs (0.20 mm height) to create a volume of approximately 2.25 microliters. The present design of the wellplate is not optimal for use with tissue samples because: (1) samples need to be immobilized and orientated to prevent them from shifting between measurements, (2) lack of a consistent and homogeneous supply of nutrients (perfusion) of the three-dimensional samples, and (3) a larger measurement chamber is required that will accommodate samples up to 200 micrometers thick. In order to adapt the geometry of the wells for use with tissue samples, disclosed herein is a disposable 96-well plate and plunger (probe) system to work within the Seahorse XF96 instrument that immobilizes, orientates, and provides perfusion of tissue samples.
The custom wellplate and perfusion inserts described above are suitable for use with tissue samples. As discussed above, the wellplate may be designed to a standard “Society for Biological Screening” SBS footprint having a 6 mm well diameter with a depression at the bottom of the well 0.5 mm deep, 3 mm diameter, used to orient and control the positioning of samples in the depression at the bottom of the well. The perfusion inserts are designed to slide into the well, with a slight interference to the wall so that the insert is held in place over the sample, and to immobilized the sample in the depression at the bottom of the well.
A wellplate, inserts, and plungers (probes) may be assembled and installed in a specialized heatsink that provides alignment and thermal control of the samples. Appropriate position offsets and a calibration protocol for the sensors may be developed. The calibration protocol consists of determining the appropriate volumes, diffusion constants, and sensor gains that are unique to the plate, insert, and probe geometry. These constants may be calculated using calibration reagents titrated to known concentrations to develop a set of coefficients that describe the signal outputs as a function of analyte concentration (O2 or H−). For example; when calibrating the pH sensor the optical signals (based on a 16 bit readout) are normalized to a starting H− concentration (pH 7.4) by determining the excitation intensity for each probe that provides a desired starting signal at pH 7.4. The signal at different H+ concentration is recorded to develop a standard curve. The coefficients for the standard curve are then loaded into the instrument so that each sensor is calibrated over a range of concentrations to be measured. A similar calibration for the Oxygen sensor may be performed by normalizing the outputs under ambient O2 concentrations (PPO=155 mm Hg), and determining a second calibration point by injecting sodium sulfite (an oxygen scavenger) to generate a calibration point at zero (PPO=0 mm Hg). The Stern Volmer quenching constant may be calculated based on equation 1 below and this relationship used to calculate the O2 level at each time point during the experiment.
K=1/O2(F0/F−1) (equation 1)
Where
K=Stern Volmer constant
F0=signal output at zero O2
F=signal output at ambient
O2=concentration of O2 at ambient
Once the calibration coefficients are established and uploaded to the instrument a series of tests may be conducted to determine the best geometry for the perfusion inserts. These tests may involve taking repeated OCR/ECAR measurements of a standard, well characterized cell line such as C2C12 fibroblasts. After each measurement the plungers (probes) may be oscillated to perfuse the measurement chamber and optimize the timing, Z travel, and speed to get the best perfusion. The best perfusion may be determined based on the ability of the system to restore the measurement chamber to the starting concentration of O2 and pH.
A prototype device was fabricated to evaluate various geometric properties and applications of embodiments of the invention, as discussed below in examples 4-6. The device includes a plurality of cylindrical vessels fitted with a polycarbonate perfusion insert. The insert is fitted to the bottom of the vessel to form a pair of chambers within the vessel that are connected through a 1.5 mm port. The upper chamber consists of a reservoir of media which is approximately 30 times the volume of the lower chamber. The inside diameter of the insert engages the Seahorse sensor cartridge to form a piston-like pump having a stroke length of approximately 5 mm and a diameter of 3 mm. The plunger of the Seahorse sensor cartridge, when moved through the inner diameter of the insert, forces media from the upper section of the chamber, through the port, into the lower chamber from which it exits through a set of peripheral vents between the inside diameter of the vessel and the outside diameter of the insert. With each stroke of the cartridge, a volume of fluid is perfused through the lower chamber such that the fluid turnover is a minimum of 2 times the static volume. At the bottom of the lower chamber is disposed a depression in the vessel (1 mm×0.25 mm) where, during use, a spherical micro-tissue is positioned for measurement. During a measurement, the cartridge comes to rest on a surface where the sensors are positioned in the center of the port which isolates the top chamber from the bottom chamber. The sensors are thus positioned over the micro-tissue, sealing the lower chamber from the upper chamber. Within this reduced volume of fluid, measurements are collected from the sensors to record the oxygen and pH concentrations at various timed intervals. From this data the oxygen and pH flux (dO/dT), (dpH/dT) are determined. This method has been used to evaluate the geometric properties of the insert to optimize fluid turnover while maintaining tissue position and to facilitate measurements of the metabolic properties of the tissues.
Feasibility for measuring single 3D spheroids using prototype inserts and plates has been demonstrated. For example, referring to
The ability to perfuse, maintain and measure metabolic profiles of single islets using a prototype chamber has been demonstrated. For example, referring to
3D micro-tissues ranging in size from 50-600 microns have been generated. For example, using the hanging drop method, cells are seeded with 40 μl of media into a commercially available seeding lid provided by InSphero Bioscience (Zurich, Switzerland). Referring to
Referring to
The invention may be embodied in other specific forms without departing from the spirit or essential characteristics thereof. The foregoing embodiments are therefore to be considered in all respects illustrative of the invention described herein. Various features and elements of the different embodiments can be used in different combinations and permutations, as will be apparent to those skilled in the art. Scope of the invention is thus indicated by the appended claims rather than by the foregoing description, and all changes which come within the meaning and range of equivalency of the claims are therefore intended to be embraced herein.
This application claims the benefit of priority under 35 U.S.C. §119(e) to U.S. Provisional Application Ser. No. 61/725,781 filed Nov. 13, 2012, the entirety of which is incorporated herein by reference.
Number | Name | Date | Kind |
---|---|---|---|
4016617 | Cardus et al. | Apr 1977 | A |
4065357 | Groves | Dec 1977 | A |
4217868 | Grather et al. | Aug 1980 | A |
D258145 | Potts | Feb 1981 | S |
4256832 | Findl et al. | Mar 1981 | A |
D260428 | Fekete | Aug 1981 | S |
D268130 | Easton | Mar 1983 | S |
4405375 | Gibson et al. | Sep 1983 | A |
4461328 | Kenney | Jul 1984 | A |
4498510 | Minshew, Jr. et al. | Feb 1985 | A |
D280131 | Takasugi | Aug 1985 | S |
D280663 | Albon et al. | Sep 1985 | S |
4599315 | Terasaki et al. | Jul 1986 | A |
D285118 | Huang | Aug 1986 | S |
4711851 | McNamara et al. | Dec 1987 | A |
D300245 | Navarro et al. | Mar 1989 | S |
D301167 | Raybould et al. | May 1989 | S |
4879097 | Whitehead et al. | Nov 1989 | A |
D324426 | Fan et al. | Mar 1992 | S |
5104804 | Humphries et al. | Apr 1992 | A |
5120421 | Glass et al. | Jun 1992 | A |
D328135 | Fan et al. | Jul 1992 | S |
D332145 | Wada et al. | Dec 1992 | S |
5200051 | Cozzette et al. | Apr 1993 | A |
D335348 | Frenkel et al. | May 1993 | S |
D339869 | Schea, III et al. | Sep 1993 | S |
5250419 | Bernard et al. | Oct 1993 | A |
5264103 | Yoshioka et al. | Nov 1993 | A |
5278048 | Parce et al. | Jan 1994 | A |
5286364 | Yacynych et al. | Feb 1994 | A |
5309085 | Sohn | May 1994 | A |
5345213 | Semancik et al. | Sep 1994 | A |
D351661 | Fischer | Oct 1994 | S |
D359125 | Livingston | Jun 1995 | S |
5459300 | Kasman | Oct 1995 | A |
5466348 | Holm-Kennedy | Nov 1995 | A |
5468605 | Harris et al. | Nov 1995 | A |
5495850 | Zuckerman | Mar 1996 | A |
5496697 | Parce et al. | Mar 1996 | A |
5512492 | Herron et al. | Apr 1996 | A |
5536662 | Humphries et al. | Jul 1996 | A |
5567598 | Stitt et al. | Oct 1996 | A |
5622872 | Ribi | Apr 1997 | A |
5728541 | Kornblith | Mar 1998 | A |
5766875 | Hafeman et al. | Jun 1998 | A |
5774214 | Prettyjohns | Jun 1998 | A |
5792426 | Portmann et al. | Aug 1998 | A |
5830138 | Wilson | Nov 1998 | A |
D404497 | Lahm et al. | Jan 1999 | S |
D404831 | Yamazaki et al. | Jan 1999 | S |
5959297 | Weinberg et al. | Sep 1999 | A |
5998517 | Gentle, Jr. et al. | Dec 1999 | A |
6030917 | Weinberg et al. | Feb 2000 | A |
D421498 | Livingston | Mar 2000 | S |
D423679 | Jenkins et al. | Apr 2000 | S |
6078698 | Lorton et al. | Jun 2000 | A |
6080574 | Berndt | Jun 2000 | A |
D428657 | Ward | Jul 2000 | S |
6083761 | Kedar et al. | Jul 2000 | A |
6146967 | Thakur et al. | Nov 2000 | A |
D438631 | Miller | Mar 2001 | S |
D438632 | Miller | Mar 2001 | S |
D438633 | Miller | Mar 2001 | S |
6280586 | Wolf et al. | Aug 2001 | B1 |
6306658 | Turner et al. | Oct 2001 | B1 |
6376233 | Wolf et al. | Apr 2002 | B1 |
6380605 | Verhaegen | Apr 2002 | B1 |
6395506 | Pitner et al. | May 2002 | B1 |
6395555 | Wilson et al. | May 2002 | B1 |
6416967 | Kornblith | Jul 2002 | B2 |
D461554 | Lafond et al. | Aug 2002 | S |
6468736 | Brooker | Oct 2002 | B2 |
D466219 | Wynschenk et al. | Nov 2002 | S |
6486947 | Modlin et al. | Nov 2002 | B2 |
D467080 | Zimmerman | Dec 2002 | S |
6627158 | Peltier | Sep 2003 | B1 |
6653124 | Freeman | Nov 2003 | B1 |
6673532 | Rao | Jan 2004 | B2 |
D486580 | Abdel-Model | Feb 2004 | S |
D492419 | Farina | Jun 2004 | S |
6766817 | da Silva | Jul 2004 | B2 |
6821787 | Neilson et al. | Nov 2004 | B2 |
6835574 | Neilson et al. | Dec 2004 | B2 |
6880158 | Basso et al. | Apr 2005 | B1 |
6881584 | Lenhard et al. | Apr 2005 | B1 |
6887680 | Kornblith | May 2005 | B2 |
6900027 | Kornblith | May 2005 | B1 |
6918404 | Dias da Silva | Jul 2005 | B2 |
6933129 | Kornblith | Aug 2005 | B1 |
7066586 | da Silva | Jun 2006 | B2 |
D529184 | Bargh et al. | Sep 2006 | S |
7276351 | Teich et al. | Oct 2007 | B2 |
D565742 | Parunak et al. | Apr 2008 | S |
D566291 | Parunak et al. | Apr 2008 | S |
7638321 | Teich et al. | Dec 2009 | B2 |
D613418 | Ryan et al. | Apr 2010 | S |
7704475 | Bull et al. | Apr 2010 | B2 |
D617468 | Marquordt et al. | Jun 2010 | S |
D618821 | Larsen | Jun 2010 | S |
D619257 | Meschenmoser et al. | Jul 2010 | S |
D624661 | Himmelsbach et al. | Sep 2010 | S |
7795012 | Lehmann et al. | Sep 2010 | B2 |
D628305 | Gorrec et al. | Nov 2010 | S |
D628306 | Blanc et al. | Nov 2010 | S |
7851201 | Teich et al. | Dec 2010 | B2 |
D631557 | Tajima et al. | Jan 2011 | S |
D632402 | Sattler et al. | Feb 2011 | S |
D651802 | Riedesel et al. | Jan 2012 | S |
D657473 | Miyashita et al. | Apr 2012 | S |
8202702 | Neilson et al. | Jun 2012 | B2 |
D669594 | Cao et al. | Oct 2012 | S |
D672053 | Chen et al. | Dec 2012 | S |
D673293 | Demas et al. | Dec 2012 | S |
D674112 | Demas et al. | Jan 2013 | S |
D686311 | Mori | Jul 2013 | S |
D686749 | Trump | Jul 2013 | S |
D694904 | Banes et al. | Dec 2013 | S |
D694906 | Priebe et al. | Dec 2013 | S |
D694908 | Okihara | Dec 2013 | S |
D696419 | Fusellier et al. | Dec 2013 | S |
8658349 | Teich et al. | Feb 2014 | B2 |
D701972 | Ohmae | Apr 2014 | S |
8697431 | Teich et al. | Apr 2014 | B2 |
D707847 | Motadel et al. | Jun 2014 | S |
D714957 | Smith | Oct 2014 | S |
D717470 | Demas et al. | Nov 2014 | S |
D720468 | Calderwood et al. | Dec 2014 | S |
20010039045 | Chan et al. | Nov 2001 | A1 |
20010051353 | Kornblith | Dec 2001 | A1 |
20020025547 | Rao | Feb 2002 | A1 |
20020059945 | Maiefski | May 2002 | A1 |
20020098592 | Neilson et al. | Jul 2002 | A1 |
20020098593 | Nelson et al. | Jul 2002 | A1 |
20020132360 | Neilson et al. | Sep 2002 | A1 |
20020146345 | Neilson et al. | Oct 2002 | A1 |
20020146836 | Neilson et al. | Oct 2002 | A1 |
20020168679 | Naus et al. | Nov 2002 | A1 |
20020182720 | Gevaert et al. | Dec 2002 | A1 |
20020192638 | Kornblith | Dec 2002 | A1 |
20030059807 | Roach et al. | Mar 2003 | A1 |
20030124029 | Webb | Jul 2003 | A1 |
20030162285 | Tajima | Aug 2003 | A1 |
20040023375 | Kornblith et al. | Feb 2004 | A1 |
20040072722 | Kornblith et al. | Apr 2004 | A1 |
20040077075 | Jensen et al. | Apr 2004 | A1 |
20040086888 | Kornblith et al. | May 2004 | A1 |
20040107986 | Neilson et al. | Jun 2004 | A1 |
20040110301 | Neilson et al. | Jun 2004 | A1 |
20040121454 | Jury et al. | Jun 2004 | A1 |
20040197905 | Hafeman | Oct 2004 | A1 |
20050054028 | Teich et al. | Mar 2005 | A1 |
20070037285 | Ehret et al. | Feb 2007 | A1 |
20070087401 | Neilson et al. | Apr 2007 | A1 |
20080014571 | Teich et al. | Jan 2008 | A1 |
20080031774 | Magnant et al. | Feb 2008 | A1 |
20100227385 | Teich et al. | Sep 2010 | A1 |
20130040855 | Takayama et al. | Feb 2013 | A1 |
20140186876 | Teich et al. | Jul 2014 | A1 |
Number | Date | Country |
---|---|---|
40 16 617 | Nov 1991 | DE |
42 17 868 | Dec 1993 | DE |
44 05 375 | Aug 1995 | DE |
101 36 005 | Nov 2002 | DE |
103 29 983 | Mar 2005 | DE |
0128438 | Dec 1984 | EP |
0 363 262 | Apr 1990 | EP |
0402917 | Dec 1990 | EP |
0545284 | Jun 1993 | EP |
0 722 136 | Jul 1996 | EP |
1416041 | Aug 2004 | EP |
2636452 | Sep 2013 | EP |
279233 | Oct 2000 | FR |
WO-8809808 | Dec 1988 | WO |
WO-9308464 | Apr 1993 | WO |
WO-9322678 | Nov 1993 | WO |
WO-9403583 | Feb 1994 | WO |
WO-9429708 | Dec 1994 | WO |
WO-9522406 | Aug 1995 | WO |
WO-9815645 | Apr 1998 | WO |
WO-9955827 | Nov 1999 | WO |
WO-9960630 | Nov 1999 | WO |
WO-0032308 | Jun 2000 | WO |
WO-0036410 | Jun 2000 | WO |
WO-0071669 | Nov 2000 | WO |
WO-0185901 | Nov 2001 | WO |
WO-0200336 | Jan 2002 | WO |
WO-0202736 | Jan 2002 | WO |
WO-0208385 | Jan 2002 | WO |
WO-0211881 | Feb 2002 | WO |
WO-02061858 | Aug 2002 | WO |
WO-02072423 | Sep 2002 | WO |
WO-02083852 | Oct 2002 | WO |
WO-02099386 | Dec 2002 | WO |
WO-03000557 | Jan 2003 | WO |
WO-03004596 | Jan 2003 | WO |
WO-03059518 | Jul 2003 | WO |
WO-2004065618 | Aug 2004 | WO |
WO-2004094060 | Nov 2004 | WO |
Entry |
---|
Haber et al., “Flow Sensor Driven Nanodispensing: The Path to More Reliable Liquid Handling Operations”, American Laboratory, Oct. 2004, pp. 32-36. |
International Preliminary Report on Patentability in PCT/US2013/069839 dated May 28, 2015 8 pages. |
Isao Karube, “Microbial Sensor”, Journal of Biotechnology, 15, (1990), pp. 255-266. |
Klaus Riedel et al., “Microbial Sensors: Fundamentals and Application for Process Control”, J. Chem. Tech. Biotechnol. 44, (1989), pp. 85-106. |
Kraus et al. “Biosensing with Cellular Systems”, Bioscope, 1, pp. 24-33, 1993. |
Y.I. Korpan et al., “A Cell Biosensor Specific for Formaldehyde Based on pH-Sensitive Transistors Coupled to Methylotrophic Yeast Cells with Genetically Adjusted Metabolism”, Analytical Biochemistry, 215, (1993), pp. 216-222. |
Yicong et al., “Drug evaluations using a novel microphysiometer based on cell-based biosensors”, Sensors & Actuators B 80:215-221 (2001). |
“Footprint Dimensions”, Society for Biomolecular Sciences SBS, ANSI American National Standards Institute, ANSI/SBS Jan. 2004, Jan. 25, 2006. |
“How to Adjust pH Levels,” Office Action mailed Sep. 28, 2010 in U.S Appl. No. 11/486,440. |
The Nature of ATP, ATP and Biological Energy, (as printed from Internet on Oct. 4, 2005, http://www.emc.maricopa.edu/faculty/farabee/BIOBK/BioBookATP.html). |
“Well Positions”, Society for Laboratory Automation and Screening SLAS, ANSI American National Standards Institute, ANSI/SLAS Apr. 2004 (formerly recognized as ANSI/SBS Apr. 2004), Oct. 13, 2011. |
Ainscow et al., “Top-down control analysis of ATP turnover, glycolysis and oxidative phosphorylation in rat hepatocytes,” Eur. J. Biochem., 263(3):671-85 (Aug. 1999). |
Amano et al., “Measuring respiration of cultured cell with oxygen electrode as a metabolic indicator for drug screening,” Human Cell 12(1):3-10 (1999). |
Andreescu et al., “Autonomous Multielectrode System for Monitoring the Interations of Isoflavonoids with Lung Cancer Cells,” 76 Anal. Chem. 8, pp. 2321-2330 (2004). |
Andreescu, S. et al., “Advanced electrochemical sensors for cell cancer monitoring,” Methods, vol. 37 pp. 84-93 (2005). |
B. Cunningham, P. Li, B. Lin, J. Pepper, “Colorimetric resonant reflection as a direct biochemical assay technique,” Sensors and Actuators B, vol. 81, p. 316-328, Jan. 5, 2002. |
Beebe D.J., Mensing G.A., Walker G.M. (2002) “Physics and applications of microfluidics in biology.” Annu. Rev. Biomed. Eng., 4, 261-86. |
Beebe D.J., Moore J.S., Bauer J.M., Yu Q., Liu R.H., Devadoss, C., Jo B.H. (2000) Functional hydrogel structures for autonomous flow control inside microfluidic channels. Nature, 404, 588-90. |
Bousse, L., Cohen, C., Nikiforov, T., Chow, A., Kopf-Sill, A.R., Dubrow, R. and Parce, J.W. (2000) “Electrokinetically Controlled Microfluidic Analysis Systems.” Annu. Rev. Biophys. Biomol. Struct. 29, 155-181. |
Brecht & Gauglitz, “Optical probes and transducers,” Biosensors and Bioelectronics, 10, p. 923-936, 1995. |
Burd et al., “Tumor oxygenation and acidification are increased in melanoma xenografts after exposure to hyperglycemia and meta-iodo-benzylguanidine,” Rediation Research 159:328-335 (2003). |
Buttgereit et al., “A hierarchy of ATP-consuming processes in mammalian cells,” Biochem. J., 1995 Nov. 15;312 (Pt 1):163-7. |
Clark, L.C. Jnr. Ann. NY Acad. Sci. 1962; 102:29-45. |
Criddle et al. “Simultaneous Measurement of Metabolic Heat Rate, CO2 Production, and O2 Consumption by Microcalorimetry” Analytical Biochem. 1991, 194:413-417. |
De Beer, Dirk, “Micro-Electrodes.” Immolilized Cells, Chapter 10 2001, 85-100. (mailed Feb. 2, 2012). |
Deshpande et al., “Microplates with integrated oxygen sensing for medium optimization in animal cell culture,” Cytotechnology 46:1-8 (2004). |
Ekelund et al., “Microphysiometry: new technology for evaluation of anticancer drug activity in human tumor cells in vitro,” Anti-Cancer Drugs 9:531-538 (1998). |
Examination Report mailed Aug. 31, 2012 for European Patent Application No. 04788615.5 filed Sep. 8, 2004, 4 pages. |
Extended European Search Report from EP Application No. 10184182.3. |
Ferguson et al. “Simultaneous monitoring of pH, CO2, and O2 using an optical imaging fiber” Analytica Chemica Acta, 1997, 340: 123-131. |
Flora K and J Brennan, “Comparison of Formats for the Development of Fiber-Optic Biosensors Utilizing Sol-Gel Derived Materials Entrapping Fluorescently-Labeled Proteins.” Analyst, 1999, 124, 1455-1462. |
Gatti et al., “Oxygen microoptodes: a new tool for oxygen measurments in aquatic animal ecology,” Marine Biology, 2002, 140:1075-1085. |
Ge X, Kostov Y, and G Rao. High Stability non-invasive autoclavable naked optical CO2 sensor. 2003. Biosensor and Bioelectronics 18:pp. 857-865. |
Gesinski RM, Morrison JH, Toepfer JR. “Measurement of oxygen consumption of rat bone marrow cells by a polarographic method.” J Appl Physiol. 1968; 24(6):751-754. |
Gump et al., “TAT transduction: the molecular mechanism and therapeutic prospects,” Trends Mol. Med., 13(10):443-48 (2007). |
Guppy, J. Cell Phys. 170:1-7 (1997). |
Handbook of Fluorescent Probes and Research Products published by Molecular Probes, Inc., Eugene, Oregon, USA, http://www.probes.com/handbook/ (accessed Mar. 12, 2004), Table of Contents, 2 pages. |
Hasselbrink E.F. Jr., Shepodd T.J., Rehm J. (2002) “High-pressure microfluidic control in lab-on-a-chip devices using mobile polymer monoliths.” Anal. Chem. 74, 4913-18. |
Hua S.Z., Sachs F., Yang D.X., Chopra H.D. (2002) “Microfluidic actuation using electrochemically generated bubbles.” Anal. Chem. 74, 6392-96. |
Huber et al., “Direct optical immunosensing (sensitivity and selectivity),” Sensors and Actuators B, 6, p. 122-126, 1992. |
International Preliminary Report on Patentability for PCT/US2004/029163, Dec. 15, 2005, 15 pages. |
International Preliminary Report on Patentability for PCT/US2007/013998, Jan. 22, 2009, 12 pages. |
International Search Report and Written Opinion for PCT/US2004/029163, Mar. 2, 2005, 12 pages. |
International Search Report and Written Opinion for PCT/US2007/ 013998, Apr. 8, 2008, 19 pages. |
International Search Report and Written Opinion for PCT/US2013/069839, Jun. 3, 2014, 9 pages. |
International Search Report for International Application No. PCT/US03/38294, Apr. 2004. |
Invitation to Pay Additional Fees & Partial Internation Search for International Application No. PCT/US2007/013998, mailed Feb. 1, 2008. |
Jekabsons et al., “Bioenergetic analysis of cerebellar granule neurons undergoing apoptosis by potassium/serum deprivation,” Cell Death Differ. 13(9):1595-610 (Sep. 2006) (Epub Jan. 20, 2006). |
Jin et al., “A biosensor concept based on imaging ellipsometry for visualization of biomolecular interactions,” Analytical Biochemistry, 232, p. 69-72, 1995. |
Jordan & Corn, “Surface Plasmon Resonance Imaging Measurements of Electrostatic Biopolymer Adsorption onto Chemically Modified Gold Surfaces,” Anal. Chem., 69:1449-1456 (1997). |
Karasinski, J. et al., “Detection and identification of bacteria using antibiotic susceptibility and a multi-array electrochemical sensor with pattern recognition,” Biosensors and Bioelectronics, vol. 22, pp. 2643-2649 (2007). |
Lehmann, M, Baumann W, Brischwein M, Gahle H-J, Freund I, Ehret R, Dreschler S, Palzer H, Kleintges M, Sieben U and Wolf B. “Simultaneous measurement of cellular respiration and acidification with a single CMOS ISFET. 2001.” Biosensors & Bioelectronics. 2001;16:195-203. |
Linder, V., Sia, S., and Whitesides, G. “Reagent-Loaded Cartridges for Valveless and Automated Fluid Delivery in Microfluidic Devices.” American Chemical Society 2005; 77(1):64-71. |
Lou et al., “Mitochondrial uncouplers with an extraordinary dynamic range,” Biochem J., 407(1):129-40 (Oct. 2007). |
Ländesmäki I, Scampavia LD, Beeson C, and Ruzicka J. “Detection of Oxygen Consumption of Cultured Adherent Cells by Bead Injection Spectroscopy.” Anal. Chem. 1999; 71: 5248-5252. |
Maharbiz et al., “Silicon microbial bioreactor arrays,” Poster 83, 1st Annual International IEEE-EMBS Special Topic Conference on Microtechnologies in Medicine & Biology, Oct. 12-14, Lyon, France. |
McConnell, H.M., Owicki, J.C., Parce, J.W., Miller, D.L., Baxter, G.T., Wada, H.G. and Pitchford, S. (1992) “The Cytosensor Microphysiometer: Biological Applications of Silicon Technology” Science 257: 1906. |
Metzger, R., Deglmann, C.J., Hoerrlein, S., Zapf, S. and Hilfrich, J. (2001) Toxicology 166, 97-108. |
Morhard et al., “Immobilization of antibodies in micropatterns for cell detection by optical diffraction,” Sensors and Actuators B, 70, p. 232-242, 2000. |
Motterlini et. al., “Depression of Endothelial and Smooth Muscle Cell Oxygen Consumption by Endotoxin,” American Journ. of Physio. vol. 275, No. 168 p. 776-782, Sep. 1998. |
O'Riordan TC, Buckley D., Ogurtsov V, O'Connor R., Papkovsky DB “A cell viability assay based on monitoring respiration by optical oxygen sensor.” Anal. Biochem. 2000; 278(2):221-227. |
Office Action in Chinese Patent Application No. 200480029825, mailed Jul. 18, 2008. |
Office Action in Chinese Patent Application No. 200480029825, mailed Nov. 28, 2008 (translation). |
Office Action in Chinese Patent Application No. 200780031522.6, dated Feb. 1, 2011. |
Office Action in Chinese Patent Application No. 200780031522.6, dated Jul. 15, 2010 (translation). |
Office Action in Chinese Patent Application No. 200780031522.6, dated Jul. 29, 2011 (translation). |
Office Action in Indian Patent Application No. 1170/DELNP/2006, mailed Oct. 6, 2008 (translation). |
Official Action in European Patent Application No. 04788615.5-1234, dated Mar. 12, 2008, 4 pages. |
Official Action in European Patent Application No. 04788615.5-1234, dated Sep. 8, 2010, 6 pages. |
Owicki, J.C., Bousse, L.J., Hafeman, D.G., Kirk, G.L., Olson, J.D., Wada, H.G. and Parce, J.W. (1994) “The Light-Addressable Potentiometric Sensor: Principles and Biological Applications.” Ann. Rev. Biophys. Biomol. Struct. 23: 87-113. |
Paitan et al., “Monitoring Aromatics Hydrocarbons by Whole Cell Electrochemical Biosensors,” Analytical Biochemistry, 335:175-183 (2004). |
Panten U and Klein H. “O2 consumption by isolated pancreatic islets, as measured in a Microincubation system with a Clark-type electrode.” Endocrinology 1982; 111:1595-1600. |
Parce W, Owicki J, Kercso K, Sigal G, Wada H, Muir V, Bousse L, Ross K, Sikic B, and McConnell H. 1989. “Detection of Cell-Affecting Agents with a Silicon Biosensor.” Science. 1989; 246(4927):243-247. |
Pattison R, Swamy J, Mendenhall B, Hwang C, and Frohlich B. “Measurement and Control of Dissolved Carbon Dioxide in Mammalian Cell Culture Processes Using an in Situ Fiber Optic Chemical Sensor.” 2000. Biotechnology Prog. 16:769-774. |
Pouli, A.E., Karagenc, N., Arden, S., Bright, N., Schofield, G.S., Hutton, J.C. & Rutter, G.A. (1998) “A phogrin-aequorin chimaera to image Ca2+ in the vicinity of secretory granules.” Biochem. J., 330, 1399-1404. |
Prokop et al., “NanoLiterBioReactor: long-term mammalian cell culture at nanofabricated scale,” Biomedical Microdevices 6(4):325-339 (2004). |
Robiolio et al., “Oxygen diffusion and mitochondrial respiration in neuroblastoma cells,” Am. J. Physiol. 256 (6 Pt 1):C1207-1213 (Jun. 1989). |
Rumsey et al., “Cellular Energetics and the Oxygen Dependence of Respiration in Cardiac Myocytes Isolated from Adult Rat” Journal of Biological Chemistry. 265(5):15392-15399. 1990. |
Scott et al., “Energy transduction in intact synaptosomes. Influence of plasma-membrane depolarization on the respiration and membrane potential of internal mitochondria determined in situ,” Biochem. J. 186(1):21-33 (Jan. 1980). |
Seaver et al. “Hydrogen Peroxide Fluxes and Compartmentalization inside Growing Eschericha coli”J. Bacteriol., 2001, 183: 7182-7189. |
Shenoy MA, Biaglow JE, Varnes ME, Hetzel FW. “Inhibition of cultured human tumor cell oxygen utilization by chlorpromazine.” Adv Exp Med Bio1.1983;159:359-68. |
Terada, “Uncouplers of oxidative phosphorylation,” Environ. Health Perspect., 87:213-18 (1990). |
Thorsen, T., Maerkl, S.J. and Quake, S.R. (2002) Microfluidic Large-Scale Integration Science 298, 580-586. |
Tolosa L, Kostov Y, Harms P, Rao G. “Noninvasive measurement of dissolved oxygen in shake flasks.” Biotechnol Bioeng Dec. 5, 2002;80(5):594-97. |
Unger, M.A., Chou, H-P, Thorsen, T., Scherer, A, and Quake, S.R. (2000) Science 288, 113-116. |
Van der Gun et al., “Serum insensitive, intranuclear protein delivery by the multipurpose cationic lipid SAINT-2,” J. Control Release, 123:228-238 (2007). |
Vanderkooi et. al., “An Optical Method for Measurement of Dioxygen Concentration Based upon Quenching of Phosphorescence, ” J. Biol. Chem., 262 (12):5476-5482 (Apr. 1987). |
Wada, H.G. Indelicato, S.R., Meyer, L. Kitamura, T., Miyajima, A., Kirk, G., Muir, V.C. and Parce, J.W. (1993) “GM-CSF Triggers a Rapid Glucose Dependent Extracellular Mediated Activation of Acid Production.” J. Cell Physiol. 154: 129-138. |
Wiley, C and Beeson, C. (2002) “Continuous measurement of glucose utilization in heart myoblasts.” Analytical Biochemistry 304, 139-146. |
Wilson et al., “The Oxygen Dependence of Mitochondrial Oxidative Phosphorylation Measured by a New Optical Method for Measuring Oxygen Concentration,” J. Biol. Chem., 263:2712-2718 (1988). |
Wodnicka M, Guarino RD, Hemperly JJ, Timmins MR, Stitt D, Pitner JB. “Novel fluorescent technology platform for high throughput cytotoxicity and proliferation assays.” Journal of Biomolecular Screening. 2000; 5:141-152. |
Wolfbeis OS, 2002. “Fiber-Optic Chemical Sensors and Biosensors.” Annal of Chem. 2002; 74:2663-2678. |
Yang et al., “Reversible and repeatable linear local cell force response under large stretches,” Experimental Cell Research, 2005, Apr., 305:42-50. |
International Search Report and Written Opinion in PCT/US2015/033815, dated Sep. 9, 2015, 10 pages. |
Number | Date | Country | |
---|---|---|---|
20140170671 A1 | Jun 2014 | US |
Number | Date | Country | |
---|---|---|---|
61725781 | Nov 2012 | US |