The present disclosure relates to systems and methods for embolization of a patient. More specifically, to vials for containing microspheres as well as systems for delivering the contained microspheres to a patient for radioembolization.
Colorectal cancer (CRC) is the fourth biggest cancer killer in the world, following lung, liver and stomach cancers respectively. In Europe and North America, it is the second most common cause of death from cancer. In Asia, it is the fourth most common cause of death from cancer (GLOBOCAN 2012; WWW.globocan.iarc.fr/Default.aspx, last accessed February 2017. The liver is the most common part of the body to which colorectal cancer spreads. This is because the colon is directly connected to the liver by blood vessels (Clark M E et al. J Gastrointest Oncol 2014; 5: 374-387). Colorectal cancer can also spread to other organs such as the lungs or bones. Once the cancer starts to spread like this it is known as secondary or metastatic colorectal cancer, which is often abbreviated to mCRC.
Surgical removal (also known as resection) of the liver cancers currently provides the only realistic possibility of providing a cure for patients with tumors that have spread from the colon. However, approximately 20-30% of patients will have liver tumors that can be removed surgically. Of those patients that can be treated using resection, 15-67% (median 30%) have been reported to be alive after five years, compared to 0-6% for patients who do not undergo surgery. (Kopetz S et al. J Clin Oncol. 2009; 27(22): 3677-3683; Simmonds P C et al Br J Cancer. 2006 94(7): 982-99).
Although some patients with liver cancer can benefit from resection, many patients are not candidates given the degree of liver disease and advanced stage of their cancer at diagnosis. Although liver transplantation can be a cure, this is often not a viable option since globally, and even in the developed world, only a small fraction of all potential patients have access to transplantation. For these patients, local ablative therapies, including radiofrequency ablation (RFA), chemoembolization, and potentially novel chemotherapeutic agents, may extend life and provide palliation. (Natthida Khajornjiraphan et al. “Yttrium-90 Microspheres: A Review of Its Emerging Clinical Indications” Liver Cancer 2015; 1:6-15; Aaron K. T. Tong et al. “Yttrium-90 hepatic radioembolization: clinical review and current techniques in interventional radiology and personalized dosimetry”, Br J Radiol; 2015; 89: 2-16.)
Radioembolization, also known as Selective Internal Radiation Therapy (SIRT) is a relatively new procedure with good overall survival and is a recommended treatment for some patients with intermediate to late stage liver cancer. SIRT involves embolizing an artery feeding a target tumor with a radioactive particle so that the tumor is selectively irradiated. SIRT uses particles or microspheres containing yttrium 90 (Y90), which is a pure beta emitter Y90 having a 2.6-day half-life and a tissue penetration depth of about 2.5 mm in the liver. SIRT is performed using a minimally invasive surgical technique by an interventional radiologist using commercially available devices.
Although effective, the commercial devices for SIRT can be cumbersome to use. For example, some procedures and devices include steps where a stopcock is manually positioned to alternate between a position for delivering radioactive particles to a position for delivering water or imaging solution. A drawback to this is that in practice the radiologist must remember to change the stopcock position or the incorrect fluid will be delivered. In addition, the particles tend to be denser than the delivery solution (e.g., DI water) and tend to quickly settle at the bottom of the delivery vial. To ensure delivery of the particles the particles are disturbed by energetic bursts of incoming delivery solution so that the particles are lofted up and can be ejected from the delivery vial through an egress tube. Although useful, this method relies on the manual positioning of the tube delivering solution to the vile (e.g. using a needle), in the delivery vials by trial and error to ensure the particles are lofted up into the solution. In addition, the devices include several tubes, syringes, a catheter, connectors as well as the delivery vial to manage during preparation of the procedure, during the procedure and after the procedure. Managing these components is very important, especially since the radioactive contamination of some of the components is inherent in the procedure. Although effective, these also deliver microspheres in non-uniform pulses from the vial to the catheter.
There is therefore a need for easier to use devices for the delivery of radioactive particles in the practical application of SIRT. This disclosure addresses some of these needs.
In general, the disclosure herein relates to methods and systems for delivery of microspheres (e.g., radioembolization microspheres). The systems can include vials (for containing the microspheres) having an inclined base or a domed base, as well as inlets having controlled angles of entry which aid in dispersing the microspheres in a carrier fluid to facilitate delivery to a patient. The system can include a tubing configuration including connectors and one-way valves in order provide a microsphere delivery system that is easy to operate, such as by an interventional radiologist.
In one aspect the disclosure includes a vial having a bottom interior surface at least a portion of which has an incline angle to a horizontal plane greater than 0 degrees and a cylindrical interior wall substantially perpendicular to the horizontal plane and meeting the bottom interior surface, wherein the bottom surface and cylindrical wall form a container for containing a liquid; (e.g., the walls can have 0-10 degrees of draft, 2-5 degrees of draft due to fabrication processes, such as molding). The vial can also have an inlet channel for fluid flow in to the vial through the cylindrical wall and an outlet channel for fluid flow out of the vial through the cylindrical wall, the inlet channel defining a direction vector of fluid flow, wherein the direction vector is oriented at a first predefined angle with respect to the horizontal plane and oriented at a second predefined angle with respect to a line tangent to the cylindrical interior wall such that fluid flowing in through the inlet channel causes a rotational flow or an upwardly moving fluid vortex to be created in the vial.
Optionally, the inlet channel of the vial is disposed at a vertical position closer to the bottom interior surface than the outlet channel. Optionally, the cylindrical wall includes a ramp portion that extends inwardly from the cylindrical wall and further defines an interior surface that pushes rotational flow in the vial toward a center of the vial and wherein the ramp portion extends vertically from a position proximate to the bottom interior surface to a first predefined vertical position with respect to the outlet channel. Optionally, the ramp portion is configured to produce a higher mixing shear in a solution in a first portion of the vial in a region proximate to the inlet channel and outlet channel as compared to lower mixing shear in a solution in a second portion of the vial in regions above outlet channel or above the ramp portion. Optionally, the inlet channel comprises an inner diameter equal to or smaller than an inner diameter of the outlet channel (e.g., 1% to 80%, from about 10% to about 75%, from about 25% to about 50% smaller). Optionally, the outlet port has an inner diameter between about 1/32″ and ¼″ inches.
In some other embodiments of the vial, the bottom interior surface is approximately flat and the first predefined angle is between about 10 and 60 degrees (e.g., in a range from 20-25 degrees) with respect to the horizontal plane. Optionally, the second predefined angle of the direction vector of fluid flow through the inlet is equal to or greater than 0° (e.g., between 0 and 20°, between 0 and 10°) with respect to the line tangent to cylindrical wall.
Optionally the vial has a convex interior bottom surface. In some other options, a radius of curvature of the cylindrical wall is greater than a radius of curvature of the convex bottom surface.
In another aspect, the disclosure comprises a delivery device comprising a first syringe in fluid connection to the inlet of the vial as described herein, wherein the vial is in fluid connection to a three-way fluid connector through the outlet channel, a second syringe in fluid connection to the three-way fluid connector, and a patient interface in fluid connection to the three-way fluid connector. Optionally, the device is configured for delivery of (a) a first solution from the first syringe, through the vial, through the three-way fluid connector, and through the patient interface to a patient, and (b) a second solution from the second syringe, through the three-way connector, and through the patient interface to a patient; wherein the inlet channel is disposed for delivering the first solution to the vial at the first predefined angle such that settled particles in a particle-containing solution in the vial are dispersed in the solution and exit the vial through the outlet channel. Optionally, the patient interface is a microcatheter, for example, a microcatheter for radioembolization.
Optionally, the device further comprises a selecting-valve integrated with the three-way fluid connector for selective delivery of the first solution from the first syringe through the selecting valve to the patient or for selective delivery of the second solution from the second syringe through the selecting valve to the patient. Optionally, the selecting-valve integrated with the three-way fluid connector is configured as a three-way stopcock.
Optionally, the delivery device comprises a first one-way valve disposed between the first syringe and the inlet channel, and a second one-way valve disposed between the second syringe the three-way connector, wherein the first one-way valve allows fluid flow from the first syringe to the vial, and the second one-way valve allows fluid flow from the second syringe to the three-way fluid-connector. Optionally, the device further comprises a third one-way valve disposed between the vial outlet and the three-way fluid connector, wherein the third one-way valve allows fluid and particle flow from the vial to the patient interface when the fluid pressure on an inlet of the third one-way valve above the pressure on an outlet of the third one-way valve, for example, for some valves in the range from about 0.1 to about 1.0 psi, such as about 0.2-0.4 psi.
Optionally, the delivery device further includes a vial holder configured as a container having a top opening and a removable top cover, and wherein the container is configured to receive the vial therein through the top opening, and the removable top cover is configured for compressing an elastomeric seal against an opening at the top of the vial when the top cover is attached to the container, thereby providing a fluid seal at the top opening of the vial. Optionally, the inlet channel is provided with fluid connection to the first syringe using a first tube, the outlet channel is provided with fluid connection to the three-way valve using a second tube, and the vial holder includes an opening for allowing connection of the first tube to the inlet channel, and an opening for allowing connection of the second tube to the outlet channel. Optionally, the device further comprises a light disposed to illuminate the interior of the vial and wherein the vial and vial container are at least partially (e.g., at least 80%, e.g., at least 99%, at least about 100%). Optionally, the container and vial are at least partially opaque to ionizing radiation (e.g., at least 50% of the radiation emanating from the interior of the vial is attenuated by the vial and container). For example, the vial includes a shielding that is between about ⅛″ and ½″ thick acrylic (e.g., about ¼″ acrylic).
Optionally, the delivery device further comprises a holder for the vial and vial holder. Optionally, the holder for the vial and vial holder includes a moat channel surrounding the vial and configured for containing at least the volume of the interior of the vial, a transparent lid for enclosing the vial and three-way fluid connector, and a removable tray configured for attachment to the holder and configured for holding the three-way fluid connector and removably attaching to the vial holder. Optionally, the fluid connection from the second syringe to the three-way fluid connector is provide by removable attachment to a third tube, the fluid connection from the three-way fluid connector to the patient interface is provide by removable attachment to a fourth tube, wherein at least a portion of the first, third and fourth tubes are placed in the holder when the device is in operation.
In yet another aspect, the disclosure comprises a method for delivery of radiomicrospheres to a patient in need of radioembolization. Optionally, the method uses the device as described herein to deliver radiomicrospheres from the vial as described herein to the patient. Optionally, the method further includes delivering a contrast agent from the second syringe to the patient and visualizing the artery using the delivered contrast agent. Optionally, the radiomicrospheres include Y90.
The vial and devices described herein provide many improvements including ease of operation as well as improved delivery of microspheres during the delivery of radioactive particles, for example, in a radioembolization procedure.
This patent or application file contains at least one drawing executed in color. Copies of this patent or patent application publication with color drawings will be provided by the Office upon request and payment of the necessary fee.
The present disclosure is directed to components, systems and methods for delivering microspheres or other elements that result in embolization of a blood vessel or other biological passageway. The system can include one or more vials that serve as a reservoir for delivering the microspheres to the patient through a configuration of tubing. In addition, disclosed herein are systems for managing and operating the vials in a safe and effective manner. In accordance with some illustrative embodiments of the, the vials and systems can be used to facilitate the administration of radiomicrospheres in a radioembolization procedure administered by an interventional radiologist to a patient, such as a patient having liver cancer.
Radioembolization or SIRT is an embolization procedure and requires administration of radiomicrospheres to a major artery feeding a tumor. Since radioisotopes have a half-life, the procedure must be scheduled and the amount of spheres delivered is dependent on the activity of the radioisotope at the time of delivery. For example, Y90 undergoes β-decay having 2.28 MeV decay energy and with a half-life of 2.6 days. Therefore, for Y90 radiomicrospheres the calibrated dose must provide the appropriate amount of GBq per treatment, with the number of microspheres varying depending on the prescribed dosage and how many days after creation of the Y90. Careful dosing is critical since under-dosing can be ineffective and require additional treatments, while over dosing can lead to unwanted radiation (e.g., to other organs) and toxicity to the patient. It is therefore important to carefully dose and administer radiomicrospheres during radioembolization.
Microspheres, such as radiomicrospheres, can include ceramic materials such as glass and polymeric beads, and, in the case of radiomicrospheres, encapsulate or are functionalized with, the radioisotope. In some embodiments, microspheres used are spherical in shape and have a diameter between about 5 and 100 μm (e.g., 20-80 μm, 30-50 μm or about 35 μm). These microspheres can be delivered by an embolization microcatheter using a carrier fluid. The carrier fluid can include water and/or saline solutions. As used herein, carrier and delivery fluid or solution are equivalent terms. The microspheres are denser than the carrier solution and therefore, when the solution is quiescent, tend to settle at the bottom of a container holding the solution and microspheres. Delivery therefore requires the microspheres to be dispersed or re-dispersed into the carrier solution, so that they can be delivered through the catheter.
One embodiment of a system 10 for delivering microspheres (e.g., radiomicrospheres) is shown in
The one way valves can be of any kind and can be independently selected. For example, the valves can be one-way, check valves or Duckbill Valves (e.g., part 80065 available from Qosina Corp). The valves can have different or the same crack or opening pressures. For example, in some embodiments the valve 111 can have a crack pressure that is at least about 0.2 psi, such as between about 0.2 and 0.8 psi, or about 0.4 psi. Without being bound to a specific theory, a low crack pressure for valve 111 can reduce the shear stresses imparted on radiomicrospheres and minimize any fragmentation.
The system 10 can include an enclosure depicted as a box 124. The enclosure 124 can serve to hold some of the various components of system 10 therein and provide a shield for the operator during use of the system 10. Without limitation, the enclosure 124 can include transparent portions so as to allow viewing of components, as well including radiation opaque portions that can serve as shielding to the operator. In some embodiments, the holder is made using materials including radiation opaque material such as acrylic, wherein the thickness is sufficient for at least about 50% attenuation of ionizing radiation emanating from anywhere within the enclosure, such as from vial 108, and lines 102, 106, 112, and 110. Other materials can also be coated on the enclosure or used in construction of the holder. For example, high Z materials such as lead can be included. In some embodiments, the holder is at least about 50% transparent to visible light.
Although the one-way valve 103 can be placed anywhere along line 102, in some embodiments it is positioned so that the valve 103 is within enclosure 124. Likewise, although the one-way valve 107 can be placed anywhere along line 106, in some embodiments it is positioned so that the valve 107 is within enclosure 124. In an alternative embodiment, the three-way connector 115 can include one, two or three one-way valves that only permit flow into the ports connected to line 106 and line 110 and only permit flow out of the third port connected to line 112. In embodiments where the one-way valve 103 and 107 are configured to be within the enclosure 124, accidental exposure due to radioactive material flowing back towards syringe 100 or 104 can be minimized since the one-way valves prevent flow past regions 102′ in the direction of syringe 100, and 106′ in the direction of syringe 104.
In operation, a user (e.g., an interventional radiologist) inserts the transfemoral catheter 122 into the hepatic artery of the patient under x-ray guidance. The selective catheterization of the proper hepatic or lobar branches can be done using a large lumen microcatheter. It is useful that the radiomicrospheres are delivered selectively to the liver tumor and not delivered to other organs such as the pancreas, stomach or duodenum. Preferably, the microcatheter inserted into the hepatic artery is positioned distal to the gastroduodenal artery (GDA) to avoid the microspheres travelling to the duodenum or stomach. Other precautions can be taken to avoid the radiomicrospheres traveling to unintended locations. During the procedure, the radiologist can check periodically, e.g., using an imaging or contrast agent and x-ray, to confirm that the catheter is correctly placed and that no reflux is occurring.
Once the microcatheter 122 is correctly placed in the patient, the microsphere delivery enclosure 124 can be placed beside, and slightly higher than, the exam table with the patient. The patient can be positioned horizontally, for example in a supine position for the procedure. Syringe 104 containing DI water, and when used, contrast agent, can be connected to line 106 which is connected to three-way connection 115. Syringe 100 of DI water can be connected to the line 102 which is connected to the vial 108, and the vial 108 can be connected to the three-way connection 115 by line 110. The microcatheter line 112 can be used to connect the three-way connector 115 to the microcatheter 122 in the patient. The connections can, for example, include Luer type taper connections, although other fluid connections can be used.
Once the tubing lines are connected, the delivery of fluid and microspheres can begin. Initially, using syringe 104, 2-3 mL of DI water with contrast agent can be injected into the system 10 to check that line 112 and microcatheter 122 are securely connected and positioned. In accordance with some embodiments of the invention, using the first syringe 100, DI water can be delivered to the inlet channel 118 of the vial 108 (containing a predefined quantity of microspheres) in 0.25 to 0.5 ml pulses. The pulsation causes agitation of the microspheres in the vial 108 so that they are lofted up from the bottom of the vial 108 to the level of the outlet channel 120 and dispersed in solution so they can be delivered out through the outlet channel 120 and line 110 to the three-way connection 115 to the microcatheter line 112 which is connected to the microcatheter 122. In some embodiments, the injection is done slowly at a rate of not more than 5 mL/min. In some embodiments, pulsation is not needed and delivery is by a steady and uniform delivery of solution from the first syringe 100. In accordance with some embodiments, prior to the microspheres reaching the catheter, the second syringe 104 can be used to deliver 3-5 mL of DI water (e.g., in 0.25 to 0.5 ml pulses) to flush the microspheres through the catheter at a rate of not more than 5 mL/min. The slow and pulsed administration helps ensure even distribution and delivery of the radio microspheres to the hepatic artery and tumor and avoids reflux back up the hepatic artery. Antegrade flow to the tumor can be evaluated at any point during the procedure by administering DI water with a contrast agent through line 106 and using fluoroscopy.
The radioembolization procedure can be continued by alternatively using syringes 100 and 104 to deliver the microspheres to patient in multiple, small aliquots over about 20 minutes (or more or less time, depending on the prescribed treatment). The radiologist can periodically check the position of the microcatheter to ensure it remains correctly sited during the delivery procedure and confirms that blood is flowing forward without stasis. As the procedure progresses, the concentration of microspheres in the vial 108 decreases and the aliquot volumes of microspheres can be increased if there is good antegrade flow. After the procedure is completed, the microcatheter 122 can be removed and the patient closed.
When the reservoir is almost empty, the last of the microspheres can be ejected from the reservoir using air to clean out the lines. The final microspheres are then delivered by DI water delivered from the 102 line. The syringes can be disconnected from lines 102 and 106, and the lines can be capped. The 102 line, 106 line, 112 line, reservoir 108 as well as the three-way connector 115, and catheter 122 can be then treated as radioactive waste and disposed appropriately.
A scan of the upper abdomen, for example using a single-photon emission computerized tomography (SPECT) system, can be performed within 24 hours after implantation of the microspheres. The scan can help to confirm placement of the microspheres in the liver, for example, the SPECT scan will detect the Bremsstrahlung radiation from the Y90.
An alternative embodiment of a system 12 for delivering microspheres (e.g., radiomicrospheres) is shown in
The operation of system 12 is similar to that of system 10 except that prior to injecting DI water from syringe 100 into the system 12, the stopcock valve 114 is placed in a first position that connects line 110 to the microcatheter line 112 and prior to injecting DI water and/or contrast agent from syringe 104 into the system 12, the stopcock valve 114 is placed in a second position that connects line 106 to the microcatheter line 112.
Some embodiments can include a combination of the features as depicted and described with reference to
It is understood the a “line” as used herein includes any useful element for transferring fluid including, for example, a tube such as a flexible plastic tube, but also can including couplers, rigid tubes, or channels. The lines can also be of any dimension although lines meant for delivery of microspheres must be larger than the microspheres and in some embodiments are at least about 2, 5, 10 or more times larger than the microspheres. In some embodiments, the lines have an inside diameter in a range from about 1/32 in. to ¼ in.
In some embodiments, the syringes as described herein can be replaced by other fluid delivery devices such as a pump including a syringe pump, a displacement pump or a peristaltic pump.
In accordance with some embodiments, the vial 108 can include a cylindrical wall 414 and an inclined or domed base and the input channel 118 can be configured at a predefined angle to produce a tangential entry of fluid resulting in the formation of a vortex in the vial that causes microspheres sitting on the base to become suspended in solution above the base. In accordance with some embodiments, the cylindrical wall 414 can be substantially perpendicular to the base or bottom of the vial 108 (e.g., in a range from 92 degrees to 95 degrees in order to allow fabrication by injection molding). In accordance with some embodiments, the integrated input and output channels as well as the interior surfaces of the vial can be arranged and configured to create a rising vortex of disperse particles (e.g., microspheres) to a predefined level in the vial such that they can be delivered to the patient from the vial through the outlet channel.
As used herein, “tangential” includes vectors with angles up to about 20° from a geometric tangent. This definition is further illustrated with reference to
In some embodiments, the outlet channel 120 extends into the inner volume of the vial 108 so that the opening 821 is disposed a perpendicular distance 8210 from wall 414. In some embodiments the distance 8210 is less than or equal to the radius of curvature R, which as previously defined is half of the diameter 458. For example, the distance 8210 can be in a range of 0.01R to 0.99R, 0.05R to 0.95 R, 0.05R to 0.5 R, 0.1R to 0.3R. The positioning of the opening 821 can help in avoiding air entering the channel 120 as described herein. An air bubble can be present in vial 108 when the vial is in use. Generally, this air bubble will reside at the top of the vial. However, if the vial is tipped, a situation can occur where the air bubble is proximate to the exit channel 120. Accordingly, the placement of the opening 821 can aid in avoiding any air accidently entering the channel 120 due to the vial 108 being unintentionally or intentionally moved from its upright position. In some embodiments the opening 821 is formed on an extending piece couple to channel 120. For example, the extending piece can be a tube inserted into channel 120.
In some embodiments, the inner diameter of exit port 821 to the outlet channel 120 is about equal to the diameter of inlet port 819 to the inlet channel 118. In some embodiments, these inner diameters are about 0.062 inches (about 1.57 mm). In some embodiments, the diameter of inlet port 819 (e.g., the inner diameter) is smaller than the diameter of exit port 821 (e.g., the inner diameter). In some embodiments, the diameter of the inlet port 819 is smaller than the diameter of exit port 821 in a range from 1% to 80% (e.g., from about 10% to about 75%, from about 25% to about 50%). In general, the size of the inlet port 819 can be selected to be small enough to provide the desired flow velocity (e.g., reducing the inlet port 819 size increases the flow velocity, but can result in decreased flow volume) while enabling the predefined desired amount of flow through the exit port 821. For example, in some embodiments, the inlet port 819 is 50% smaller than exit port 821 and the exit port 821 diameter can be approximately the same diameter as the tubing 110 connected to the outlet port 821 in order to avoid restricting outward flow, reducing the potential for clogging of spheres, and to minimize internal pressure buildup in the vial 108 during delivery.
The viewing port can also be configured to accept the protrusion 1030 when the vial 108 is coupled to vial holder 1112, where the viewing port extends down the vial holder 1112 behind. This alignment can aid in orienting the vial 108, for example during a measurement of radioactivity, since the radiation of the vial when charged with radiomicrospheres is not uniform due to the inclined base 416, which settles the microspheres preferably in low end 418 when quiescent. A secondary holder, or cradle 3012, as shown in
In some embodiments, the tray 1208 is integrated as part of the base 1204 of housing 1202, such that the tray is not configured to be removable from the base 1204. In these embodiments, for example, the features such as recessed feature 1214 form a part of the base 1204. Other features such snap-fit components, cleats and fasteners for placement and securing of lines and the three-way connector 115 can also be used. Although shown as a T valve in
In addition to helping in managing the tubing and vial while in operation, in some embodiments the tray is configured to be disposed of with the vial, vial holder and associated tubing and patient interface (e.g., a microcatheter) once a procedure is completed. For example, the tray with attached tubing and vial holder can be lifted off of the base 1204 and disposed of appropriately, for example, into a radioactive waste container.
Embodiments of various aspects described herein can be defined as in any of the following numbered paragraphs:
1. A vial comprising;
a bottom interior surface at least a portion of which has an incline angle to a horizontal plane greater than 0 degrees and a cylindrical interior wall substantially perpendicular to the horizontal plane and meeting the bottom interior surface, wherein the bottom surface and cylindrical wall form a container for containing a liquid; (e.g., the walls can have 0-10 degrees' of draft, 2-5 degrees of draft due to fabrication processes such as molding)
an outlet channel for fluid flow out of the vial through the cylindrical wall;
an inlet channel for fluid flow into the vial through the cylindrical wall, the inlet channel defining a direction vector of fluid flow, wherein the direction vector is oriented at a first predefined angle with respect to the horizontal plane and oriented at a second predefined angle with respect to a line tangent to the cylindrical interior wall such that fluid flowing in through the inlet channel causes a rotational flow or an upwardly moving fluid vortex to be created in the vial.
2. The vial according to paragraph 1, wherein the inlet channel is disposed at a vertical position closer to the bottom interior surface than the outlet channel.
3. The vial according to paragraph 1 or 2, wherein the cylindrical wall includes a ramp portion that extends inwardly from the cylindrical wall and further defines an interior surface that pushes rotational flow in the vial toward a center of the vial and wherein the ramp portion extends vertically from a position proximate to the bottom interior surface to a first predefined vertical position with respect to the outlet channel.
4. The vial according to paragraph 1, wherein the ramp portion is configured to produce a higher mixing shear in a solution in a first portion of the vial in a region proximate to the inlet and outlet as compared to lower mixing shear in a solution in a second portion of the vial in regions further from the inlet and outlet.
5. The vial according to any one of the above paragraphs wherein the inlet channel comprises an inner diameter equal to or smaller than an inner diameter of the outlet channel (e.g., 1% to 80%, from about 10% to about 75%, from about 25% to about 50% smaller), optionally wherein the outlet port has an inner diameter between about 1/32″ and ¼″ inches.
6. The vial according to any of the above paragraphs, wherein
the bottom interior surface is approximately flat and the first predefined angle is between about 10 and 60 degrees (e.g., in a range from 20-25 degrees) with respect to the horizontal plane.
7. The vial according to paragraph 6, wherein the second predefined angle of the direction vector of fluid flow through the inlet is equal to or greater than 0° (e.g., between 0 and 20°, between 0 and 10°) with respect to the line tangent to cylindrical wall.
8. The vial according to any one of paragraphs 1-4, wherein
the vial has a convex interior bottom surface, and
wherein a radius of curvature of the cylindrical wall is greater than a radius of curvature of the convex bottom surface.
9. A delivery device comprising:
a first syringe in fluid connection to the inlet of the vial according to any one of paragraphs 1-8, wherein the vial is in fluid connection to a three-way fluid connector through the outlet channel to the vial;
a second syringe in fluid connection to the three-way fluid connector;
and a patient interface in fluid connection to the three-way fluid connector; wherein,
the device is configured for delivery of:
wherein the inlet channel is disposed for delivering the first solution to the vial at the first predefined angle such that settled particles in a particle-containing solution in the vial are dispersed in the solution and exit the vial through the outlet channel.
10. The delivery device according to paragraph 9, further comprising a selecting-valve integrated with the three-way fluid connector for selective delivery of the first solution from the first syringe through the selecting valve to the patient or for selective delivery of the second solution from the second syringe through the selecting valve to the patient.
11. The delivery device according to paragraph 10, wherein the selecting-valve integrated with the three-way fluid connector is configured as a three-way stopcock.
12. The delivery device according to paragraph 9, further comprising a first one-way valve disposed between the first syringe and the inlet channel, and a second one-way valve disposed between the second syringe the three-way connector, wherein
the first one-way valve allows fluid flow from the first syringe to the vial, and the second one-way valve allows fluid flow from the second syringe to the three-way fluid-connector.
13. The delivery device according to paragraph 12, further comprising a third one-way valve disposed between the vial outlet and the three-way fluid connector, wherein the third one-way valve allows fluid and particle flow from the vial to the patient interface when a fluid pressure on an inlet of the third one-way valve is at least 0.2 psi above a pressure on an outlet of the third one-way valve, for example for example between about 0.2 and 0.8 psi, or about 0.4 psi.
14. The delivery device according to any one of paragraphs 9-13, further including a vial holder configured as a container having a top opening and a removable top cover, and wherein the container is configured to receive the vial therein through the top opening, and the removeable top cover is configured for compressing an elastomeric seal against an opening at the top of the vial when the top cover is attached to the container, thereby providing a fluid seal at the top opening of the vial, and
the inlet channel is provided with fluid connection to the first syringe using a first tube, the outlet channel is provided with fluid connection to the three-way valve using a second tube, and the vial holder includes an opening for allowing connection of the first tube to the inlet channel, and an opening for allowing connection of the second tube to the outlet channel.
15. The delivery device according to paragraph 14 further comprising a light disposed to illuminate the interior of the vial and wherein the vial and vial container are at least partially (e.g., at least 80%, e.g., at least 99%, at least about 100%).
16. The delivery device according to paragraph 14 or 15, wherein the container and vial are at least partially opaque to ionizing radiation (e.g., at least 50% of the radiation emanating from the interior of the vial is attenuated by the vial and container). For example, the vial includes a shielding that is between about ⅛″ and ½″ thick acrylic (e.g., about ¼″ acrylic).
17. The delivery device according to any one of paragraph 14-16, further comprising a holder for the vial and vial holder, wherein the holder includes,
a moat channel surrounding the vial and configured for containing at least the volume of the interior of the vial,
a transparent lid for enclosing the vial and three-way fluid connector, and
a removable tray configured for attachment to the holder and configured for holding the three-way fluid connector and removably attaching to the vial holder.
18. The delivery device according to paragraph 17, wherein the fluid connection from the second syringe to the three-way fluid connector is provide by removable attachment to a third tube, the fluid connection from the three-way fluid connector to the patient interface is provide by removable attachment to a fourth tube, wherein at least a portion of the first, third and fourth tubes are placed in the holder when the device is in operation.
19. The delivery device according to any one of the above paragraphs wherein the patient interface is a microcatheter.
20. A method for delivery of radiomicrospheres to a patient in need of radioembolization, the method comprising using the device of any one of paragraphs 9-19 to deliver radiomicrospheres from the vial to the patient.
21. The method of paragraph 20, further comprising delivering a contrast agent from the second syringe to the patient and visualizing the artery using the delivered contrast agent.
22. The method of paragraph 20 or 21, wherein the radiomicrospheres include Y90.
As used herein the term “comprising” or “comprises” is used in reference to compositions, methods, and respective component(s) thereof, that are essential to the claimed invention, yet open to the inclusion of unspecified elements, whether essential or not.
As used herein the term “consisting essentially of” refers to those elements required for a given embodiment. The term permits the presence of elements that do not materially affect the basic and novel or functional characteristic(s) of that embodiment of the claimed invention.
The term “consisting of” refers to compositions, methods, and respective components thereof as described herein, which are exclusive of any element not recited in that description of the embodiment.
As used in this specification and the appended claims, the singular forms “a,” “an,” and “the” include plural references unless the context clearly dictates otherwise. Thus for example, references to “the method” includes one or more methods, and/or steps of the type described herein and/or which will become apparent to those persons skilled in the art upon reading this disclosure and so forth.
All patents, patent applications, and publications identified are expressly incorporated herein by reference for the purpose of describing and disclosing, for example, the methodologies described in such publications that might be used in connection with the present invention. These publications are provided solely for their disclosure prior to the filing date of the present application. Nothing in this regard should be construed as an admission that the inventors are not entitled to antedate such disclosure by virtue of prior invention or for any other reason. All statements as to the date or representation as to the contents of these documents is based on the information available to the applicants and does not constitute any admission as to the correctness of the dates or contents of these documents.
This application is a 35 U.S.C. § 371 National Phase Entry application of International Patent Application No. PCT/US2019/063000 filed on Nov. 25, 2019 which claims any and all benefits as provided by law including benefit under 35 U.S.C. § 119(e) of the U.S. Provisional Application No. 62/774,620, filed Dec. 3, 2018, and U.S. Provisional Application No. 62/913,461, filed Oct. 10, 2019, the contents of which are incorporated herein by reference in their entirety.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US2019/063000 | 11/25/2019 | WO | 00 |
Number | Date | Country | |
---|---|---|---|
62774620 | Dec 2018 | US | |
62913461 | Oct 2019 | US |