The embodiments disclosed herein relate to mass transfer, and in particular to apparatus, systems and methods for controlling mass transfer of gases into liquids, and more particularly to apparatus, systems and methods for controlling carbonators and other gas transfer systems.
There are numerous industrial processes and types of equipment used to promote the mass transfer of gases into liquids. In many cases, the mass transfer of a gas into a liquid is limited by the mass-transfer resistance at the gas-liquid interface and the diffusion of the gas away from this interface. For example, the binary diffusion coefficient of carbon dioxide in air is 0.139 sq.cm/sec, while the binary diffusion coefficient for carbon dioxide in water is 0.00002 sq.cm/sec.
Since the diffusivity of a gas within a gas is typically around 1,000-10,000 times greater than the diffusivity of a gas into a liquid, dispersion of the liquid is important for effecting mass transfer of a gas into a liquid. For example, if a liquid can be dispersed as droplets having a characteristic droplet length roughly equal to the square root of the binary diffusion coefficient (e.g. for carbon dioxide into water, √0.00002=0.0044 cm, or 44 micrometers), then the diffusion will tend to be extraordinarily rapid.
Generally, to provide for optimum mass transfer rates, all of the liquid should be provided with a similar droplet size having the characteristic diffusion length. Any quantity of liquid that has a larger droplet size will not provide for rapid diffusion, and will not reach equilibrium in the surrounding gas environment within a brief period of time (as is the case with the smaller droplets).
In many prior art systems, the mass-transfer resistance may be partially overcome by increasing the gas-liquid surface (e.g. by performing mechanical work on the liquid). For example, some systems use powerful mechanical mixers to agitate the liquid. Other systems create small bubbles of gas by pressing a gas through small orifices, and then the bubbles are allowed to rise through a liquid column. However, neither of these approaches is particularly good at overcoming the mass-transfer resistance.
One technique that would be beneficial is to cause the liquid to be dispersed into the gas, rather than the gas into the liquid. In practice, however, this is very difficult to achieve. Some prior art systems attempt this using high-pressure nozzles to disperse a liquid as fine droplets. Other systems use a two-phase flow of gas and liquid through a nozzle at lower pressure. However, these types of systems are also generally undesirable, as they may require high-pressure, pressure-boosting pumps to be used, or make an undesirable use of gas to disperse the liquid (e.g. using two-phase nozzles). In particular, when attempting a precision transfer of gas into liquid, two-phase nozzles are often unacceptable as the amount of gas required to accomplish the required breakup of the liquid is normally not the quantity of gas that is desired to be transferred into the liquid.
Accordingly, such systems are not appropriate for many applications, especially where precise control of the ratio of gas to liquid is desired, such as in beverage carbonation (e.g. for soda pop and similar beverages).
One known carbonated beverage system is described in U.S. Pat. No. 3,877,358 to Karr. Karr describes a system for continuously preparing a carbonated product and filling containers. Carbon dioxide is applied to a liquid under carbonating pressure and the carbonated liquid is passed to a stabilizing tank maintained at a pressure at least equal to carbonating pressure. Thereafter, the liquid is directed to a pressure reduction tank in which the pressure is lowered to a level just sufficient to produce the desired flow rate to the filler. By reducing the pressure prior to filling, a filler of the gravity or vacuum type may be employed without excessive foaming. Or, if a counter pressure filler is employed, operating pressure is considerably lowered.
Another system is described in U.S. Pat. No. 5,156,871 to Goulet et al. Goulet et al. describe an apparatus for providing carbonating of water, that includes a carbonating tank having a carbon dioxide inlet, a water inlet, and a carbonated water outlet. The carbonating tank is pivotally mounted to a rigid structure and connected to an electric motor for providing an undulating or rocking motion of the carbonator about its pivot mounting. The motion of the carbonating tank provides for carbonating of the water held therein.
Another carbonator is described in U.S. Pat. No. 5,792,391 to Vogel et al. Vogel et al. describes a carbonator comprising a tube cylinder having a closed and an open end. A disk is removably retained in the open end for providing access into the interior volume thereof. The disk provides for mounting thereto of water and carbon dioxide gas inlets, a carbonated water outlet, a safety relief valve and a water level sensor. A rigid retaining wire is bent into a square configuration wherein radiussed corners thereof cooperate with slots in the open end of the cylinder to retain the disk therein. Manipulation of the retaining wire provides for removal of the disk from the cylinder when the carbonator is not pressurized.
Another carbonation system is described in European patent application no. EP 0 873 966 to Hassell. Hassell describes an in-line carbonation system for carbonating water or a non-carbonated pre-mix mixture. A flow restrictor meters a predetermined quantity of pressurized carbon dioxide gas to a T-fitting for carbonation with a known quantity of the liquid as delivered thereto by a pump. The carbon dioxide and liquid flow therefrom through a turbulator for enhancing of the absorption of the gas by the liquid, The liquid and gas then flow through a heat exchange cooling coil for further absorption of the gas as the liquid is cooled in the coil. The coil is connected to a dispensing valve for dispensing of the cooled carbonated liquid.
European patent application no. EP 0 677 740 to Butts describes a carbon dioxide measurement and control system which comprises a thermal conductivity carbon dioxide sensor and a humidity sensor which produce separate and independent output signals which are compared in a microprocessor, thereby generating a humidity compensated carbon dioxide gas measurement which is accurate to within plus or mins 0.1 percent.
Finally, PCT patent application no. WO/2002/059534 to Chadwell et al. describes a network topology for communications within and between food service equipment items generally having one or more food service equipment items, each having a controller and at least one functional subsystem configured as a slave to the controller, and a master interface and a bus slave interface associated with each controller. Each master interfaces is adapted for electrical communication over an intra-equipment communication channel with a slave interface associated with the functional subsystem. Each bus slave interface is adapted for communication, external to the food service equipment item, over an inter-equipment communication channel. Each bus slave interface is substantially electrically identical to each slave interface.
Another technique for dispersing liquid uses violent impaction of the liquid against a set of rotating blades. However, impaction is also undesirable, as the impacted liquid tends to be dispersed as droplets of poly-disperse sizes (e.g. some droplets are quite small while other droplets may be quite a bit larger). As discussed above, the larger droplets will tend not to reach equilibrium along with the smaller droplets, and thus do not provide for good diffusion of the liquid.
Furthermore, if the time provided for dispersion is extremely brief, then only a portion of the poly-disperse droplets may achieve a target gas content, and this proportion will be a complex function of the integrated gas transferred into the droplets of various sizes.
In the specific case of beverage carbonation (e.g. for soda pop and similar beverages), there are numerous examples of systems involving the mixing of bulk carbon dioxide and water, for example McCann et al. in U.S. Pat. No. 5,855,296; Hancock and May in U.S. Pat. No. 4,850,269; Burrows in U.S. Pat. Nos. 5,073,312 and 5,071,595; Vogal and Goulet in U.S. Pat. No. 5,792,391; Goulet in U.S. Pat. No. 5,419,461; Notar et al. in U.S. Pat. No. 5,422,045; Bellas and Derby in U.S. Pat. Nos. 6,935,624 and 6,758,462; Hoover in U.S. Pat. No. 4,745,853; and Singleterry and Larson in U.S. Pat. No. 5,842,600.
Some example systems include the use of a spinning turbine within a carbonator. For example, U.S. Pat. No. 5,085,810 (Burrows) describes using jets of liquid to drive an impeller that is affixed to an elongated shaft supporting a series of discs that are submerged in a liquid. In this case, the impeller is not driven by a motor, but instead is driven by the force of the incoming liquid, which is used to rotate the shaft (and thus cause the discs attached to the shaft to also rotate). Burrows does not focus on liquid dispersion through impaction, but is instead an effort to eliminate the drive motor normally used to rotate the submerged discs.
A nearly identical arrangement is described by Koenig and Erlanger in U.S. Pat. No. 610,062, published in 1898. Again, the incoming liquid is allowed to impinge upon an impeller so as to cause an elongated shaft to rotate, which rotates additional impellers submerged within a body of liquid, causing mixing.
U.S. Pat. No. 4,804,112 by E. L. Jeans describes a liquid entering a pressurized vessel containing carbon dioxide gas being allowed to impact a bladed rotor. The mechanism of causing the break-up of the liquid into droplets is impaction upon the blades of the rotor. As will be understood by those skilled in the art, impaction involves the turbulent breakup of the liquid, and results in the production of droplets varying widely in size (e.g. droplets with poly-disperse sizes). In addition, the size of the droplets generated by Jeans is generally large (e.g. larger than 75 micrometers) unless extraordinary impaction velocities are achieved (i.e. velocities approaching the speed of sound in a liquid).
Any large droplets formed through the use of impaction inhibits achieving gas absorption equilibrium, and hence a significant volume of the liquid in such systems will have insufficient gas saturation. Accordingly, elevated pressure must be used to achieve the target gas content within the liquid under such conditions. However, this creates a potential for exceeding the target saturation, especially if the liquid and gas are left within the carbonation chamber for an extended period of time.
Accordingly, there is a need in the art for improved apparatus, systems and methods for mass transfer of gases into liquids.
In some embodiments described herein, a fine dispersion of liquid is generated using a spinning disc apparatus or a rotating capillary apparatus to generate small liquid particles. The small liquid particles are then dispersed into gas to carry out the mass transfer of the gas into the liquid droplets. The liquid particles may then coalesce with the chamber and/or against the walls of the chamber, and be subsequently collected for extraction.
Generally, it is desirable that the liquid dispersion produces an exact droplet size, or at least a dispersion of liquid droplets that are almost entirely and reliably below a critical size, so as to closely approach equilibrium with the surrounding gas within extremely brief time scales. In some examples, it would be desirable to perform such dispersion in less than a few seconds, and in some cases within tens of milliseconds.
The embodiments described herein generally form droplets of uniform or near-uniform size through the use of elegant physics for droplet formation and by balancing forces at the edge of a generally flat spinning disc or within a rotating capillary. In addition, the power consumption for such embodiments tends to be very low. Furthermore, the edge velocities and angular velocities required to achieve essentially complete reduction of the liquid into the required droplet size tend to be quite modest.
Some embodiments as described herein provide a simple apparatus that tends to produce a uniform and precise dispersion of a liquid into a mist or spray having a specific droplet size and with minimal potential for any significant volume of the liquid being dispersed as over-sized droplets (e.g. droplets that are larger than desired).
In some examples, this dispersion may be carried out within a space or chamber that operates at elevated pressure so as to cause a gas to rapidly dissolve into the liquid droplets and approach equilibrium saturation during the flight time of the droplets (e.g. between when they are thrown or disengage from the spinning disc and before contacting the walls of the chamber).
To accomplish the required mass transfer within the brief flight time of the droplets, the droplets generally should be extremely small. Furthermore, the distance between the edge of the spinning disc and the walls of the chamber should be sufficient to allow the droplets to closely approach saturation with the surrounding gas prior to being arrested against the walls. If the droplets are sufficiently small, they will slow and even come to rest before engaging the chamber walls and thus their contact time with the gas can be extended.
According to one aspect, there is provided an apparatus for mass transfer of gas into a liquid, comprising a tank that defines a chamber for receiving the gas, and at least one surface provided within the chamber, each surface having an inner region, an outer region and an edge adjacent the outer region, wherein each surface is configured to receive the liquid at the inner region and rotate such that the liquid flows on the surface from the inner region to the outer region, and, upon reaching the edge of the surface, separates to form liquid particles that move outwardly through the gas in the chamber, and wherein the liquid particles are sized so that the gas is absorbed by the liquid particles to produce a mixed liquid saturated with the gas during a brief flight time of the liquid particles through the chamber.
According to another aspect, there is provided a carbonator for mass transfer of carbon dioxide into water, comprising a tank that defines a chamber for receiving the carbon dioxide, and at least one surface provided within the chamber, each surface having an inner region, an outer region and an edge adjacent the outer region, wherein each surface is configured to receive the water at the inner region and rotate such that the water flows on the surface from the inner region to the outer region, and, upon reaching the edge of the surface, separates to form water particles that move outwardly through the carbon dioxide in the chamber, and wherein the water particles are sized so that the carbon dioxide is absorbed by the water particles to produce a carbonated water saturated with the carbon dioxide during a brief flight time of the water particles through the chamber.
According to yet another aspect, there is provided a method for mass transfer of gas into a liquid, comprising the steps of providing a chamber having the gas therein, providing at least one surface within the chamber, each surface having an inner region, an outer region and an edge adjacent the outer region, providing a liquid to the inner region of each surface, and rotating the surface at an angular velocity selected such that the liquid will move from the inner region to the outer region, and, upon reaching the edge, separates from the at least one surface to form at least one liquid particle that moves outwardly through the gas, wherein the liquid particles are sized so that the gas is absorbed by the liquid particles to produce a mixed liquid saturated with the gas during a brief flight time of the liquid particles through the chamber.
According to another aspect, a system for controlling mass transfer of a gas into a liquid, comprising a mass transfer apparatus configured to provide the mass transfer of the gas into the liquid, a temperature sensor configured to monitor the temperature of the liquid provided to the mass transfer apparatus, a pressure sensor configured to monitor the pressure within the mass transfer apparatus, and a control unit configured to communicate with the temperature sensor and the pressure sensor and in response adjust the pressure within the mass transfer apparatus to encourage reaching an equilibrium condition within the mass transfer apparatus and provide for consistent gas absorption into the liquid.
The control unit may be configured to adjust the supply of the gas to the mass transfer apparatus to control the pressure in the mass transfer apparatus. The control unit may adjust the supply of the gas to the mass transfer apparatus using a valve.
In some embodiments, the gas may be carbon dioxide and the mass transfer apparatus may be a carbonator. In other embodiments, the gas is nitrogen.
In various embodiments, the liquid may be water, beer, fruit juice, flavoured juice, coffee, a coffee flavored beverage, tea, a pre-mix, or other suitable liquids.
The control unit may be configured to control at least one operating characteristic of the mass transfer apparatus. The control unit may also be configured to control at least one operating characteristic of a pump configured to supply the liquid to the mass transfer apparatus.
The mass transfer apparatus may comprise a tank that defines a chamber for receiving the gas therein; and at least one surface provided within the chamber, each surface having an inner region, an outer region and an edge adjacent the outer region; wherein each surface is configured to receive the liquid at the inner region and rotate such that the liquid flows on the surface from the inner region to the outer region, and, upon reaching the edge of the surface, separates to form liquid particles that move outwardly through the gas in the chamber; and wherein the liquid particles are sized so that the gas is absorbed by the liquid particles to produce a liquid saturated with the gas during a flight time of the liquid particles within the chamber.
According to another aspect, there is provided a system for controlling the mass transfer of a gas into a liquid, comprising a mass transfer apparatus configured to provide mass transfer of the gas into the liquid; a temperature sensor configured to monitor the temperature of the liquid; and a control unit configured to communicate with the temperature sensor and adjust the operation of the mass transfer apparatus based on the temperature of the liquid.
The system may further comprises a pressure sensor configured to monitor the pressure within the mass transfer apparatus, and the control unit may be configured to communicate with the pressure sensor and adjust the operation of the mass transfer apparatus based on the pressure therein.
The control unit may be configured to maintain an equilibrium condition within the mass transfer apparatus based on the temperature of the liquid and the pressure within the mass transfer apparatus.
The drawings included herewith are for illustrating various examples of apparatus, systems and methods of the present specification and are not intended to limit the scope of what is taught in any way. In the drawings:
Illustrated in
The apparatus 10 generally includes a tank 12 that defines a chamber 14 into which the gas and liquid may be generally received for effecting the mass transfer.
The apparatus 10 also generally includes a disc 20 that is provided within the chamber 14. The disc 20 has a surface configured to receive a liquid thereon and can rotate so as to cause a fine dispersion of liquid particles to be ejected from the edges thereof, as will be described in greater detail below.
The tank 12 may be a pressure vessel or any other suitable vessel, and may be capable of operating at elevated pressures according to the desired operating conditions of the apparatus 10. For instance, in some examples, the tank 12 may be configured to operate up to pressures of 3 atmospheres or greater.
As shown, the tank 12 may include a separate top tank head 16 and bottom tank head 18, each having upper and lower mounting flanges 22, 24 extending outwardly from the perimeter thereof. The mounting flanges 22, 24 may be coupled together using one or more fasteners (e.g. bolts 28, washers 30 and nuts 32) so as to secure the upper tank head 16 and lower tank head 18 together to define the chamber 14 therebetween.
In some examples, a flange gasket 26 may be provided between the flanges 22, 24 so as to help seal the tank heads 16, 18 together and to inhibit leaks.
As shown, each of the upper and lower tank heads 16, 18 have outer walls generally located around the perimeter of the chamber 14. For example, the upper tank head 16 has a peripheral upper chamber wall 34, and the lower tank head 18 has a peripheral lower chamber wall 36.
As shown, the upper tank head 16 has a bulkhead fitting 38 (or liquid inlet fitting). The bulkhead fitting 38 is configured to be coupled to a liquid supply (e.g. using a hose, not shown) so that liquid may be pumped into the chamber 14 during use of the apparatus 10.
The upper tank head 16 may include an upper puck 40 for securing the bulkhead fitting 38 to the tank head 16. The upper puck 40 may help to stabilize the upper tank head 16 so as to provide for a more secure coupling of the bulkhead fitting 38. In some examples, the bulkhead fitting 38 and upper puck 40 may be welded to the upper tank head 16.
The bulkhead fitting 38 is coupled to an inlet spout 42 that extends generally downwardly into the chamber 14. The inlet spout 42 is configured to provide liquid to an inner region 20a of the spinning disc 20 during use of the apparatus 10, as will be described in greater detail below.
The upper tank head 16 also generally includes a gas inlet 44 (shown in
The lower tank head 18 also generally includes an outlet fitting 46. The outlet fitting 46 is configured to allow extraction of the gas and liquid mixture (e.g. using a hose, not shown) that is generated by the apparatus 10 and which tends to collect in the lower tank head 18 during use.
The apparatus 10 may also include a pH sensor 48, which may be coupled to the lower tank head 18 using a sensor fitting 50. The pH sensor 48 has a sensor tip 52 that extends into the chamber 14 and is configured to measure the pH levels of the gas-liquid mixture that collects in the lower tank head 18.
Based on the pH levels observed by the pH sensor 48, the properties of the gas-liquid mixture can be monitored and decisions may be made about the operation of the apparatus 10, such as whether additional quantities of liquid and/or gas should be added to the apparatus 10, and/or whether the gas-liquid mixture is ready for extraction via the outlet fitting 46.
In some examples, the tank 12 also includes a float switch 54 mounted to the lower tank head 18 via a switch fitting 56. The float switch 54 may be configured to monitor the level of the gas-liquid mixture within the lower tank head 18. Based on the height of the mixture, the float switch 54 may be used to trigger extraction 46 of the mixture, control the rate of liquid flowing in through the inlet spout 42, and/or take other actions.
In particular, the float switch 54 can ensure that the level of mixed liquid in the chamber 14 remains below the surface of the disc 20, so that liquid from the inlet spout 42 does not immediately contact the mixture, but is first dispersed by the disc 20 (as will be described in greater detail below). This also tends to ensure that the mixture does not interfere with the rotation of the disc 20.
The apparatus also generally includes a drive mechanism 60 configured for rotating or spinning the disc 20 about an axis of rotation A. The drive mechanism 60 may generally be any suitable drive (e.g. a magnetic drive) and may include an inner rotor 62 configured to rotate and an outer rotor 64 that is mechanically coupled to an electric motor or other suitable source of powered rotation. For instance, in this example, the inner and outer rotors 62, 64 are magnetically coupled so that the inner rotor 62 rotates when the outer rotor 64 is caused to rotate.
The inner rotor 62 is generally coupled to a shaft 66 that extends upwardly into the chamber 14. The shaft 66 has an upper portion 66a that is coupled to the disc 20 so that as the inner rotor 62 rotates, the shaft 66 and disc 20 also rotate.
The shaft 66 may be received within a shaft housing 68 configured to support and stabilize the shaft 66 and disc 20 during rotation. One or more journal bearings 70 may be provided between the shaft 66 and housing 68 so as to inhibit wear during rotation. In some examples, the journal bearings 70 may be plastic, or any other suitable material.
In some examples, a cap 72 may extend downwardly from the bottom of the lower tank head 18. The cap 72 may house elements of the drive mechanism 60 (e.g. the inner rotor 62 and a lower portion of 66b of the shaft 66) generally below the tank 12, which may facilitate the operation of the drive mechanism 60 (e.g. the magnetic coupling between the inner and outer rotors 62, 64).
As shown, the cap 72 may be coupled to a lower puck 78 provided in the lower tank head 18 using one or more fasteners 74, and may have a gasket 76 provided between the lower puck 78 and the barrier 72 to assist with inhibiting leaks.
In some examples, the inner rotor 62 may be coupled to a thrust bearing 80 (which may be plastic or any other suitable material).
The drive mechanism 60 may be used to rotate the disc 20 at elevated speeds selected according to the desired operating conditions of the apparatus 10. For example, the disc 20 may be rotated at speeds up to and including 3600 RPM. Alternatively, the disc 20 may be rotated at speeds of greater than 3600 RPM.
In some examples, the tank 12 may also include a safety release valve (not shown) so as to inhibit an overpressure situation from forming within the chamber 14, and which could otherwise damage the components therein and/or cause the tank 12 to crack or burst.
As shown, the disc 20 generally has a flat upper surface (as shown in
In some examples, the disc 20 may be made of a metal (e.g. steel, aluminum, etc.). In other examples, this disc 20 may be made of another material that is suitable for rotation at elevated speeds, such as high-strength plastics or ceramics.
During use of the apparatus 10, liquid (e.g. water) may be fed to the inner region 20a of the disc 20 using the inlet spout 42, and the drive mechanism 60 may be used to rotate the disc 20 about the axis of rotation A.
As shown, a lower end portion 42a of the inlet spout 42 may be positioned adjacent or directly above the upper surface of the disc 20. Accordingly, the liquid can be directed onto the disc 20 in a generally smooth manner (e.g. without violent impaction that could cause poly-disperse sizes of droplets to be formed).
The rotation of the disc 20 generally causes the liquid to move from the inner region 20a outwardly towards an outer region 20b of the disc 20. As the liquid moves outwardly, it tends to spread upon the surface of the disc 20, generally forming a thin film.
Once the liquid reaches the outer edge 21 of the disc 20, it may collect at the edge, and then eventually separate from the edge 21 as particles or droplets.
Once separated, the particles of liquid will fly outwardly through the surrounding atmosphere in the chamber 14 towards the chamber walls 34, 36. During this flight, the particles will interact with gas fed into the chamber 14 using the gas inlet 44 (e.g. carbon dioxide). In some example, the gas may be continuously fed into the chamber 14. In other examples, the gas may be intermittently fed into the chamber 14.
Generally, the liquid particles are sufficiently small that the gas will rapidly dissolve into them and approach equilibrium saturation during the flight time of the particles (e.g. between disengaging from the spinning disc 20 and contacting the walls 34, 36 of the chamber 14). In some examples, the flight time is less than 100 milliseconds. In yet other examples, the flight time is less than 50 milliseconds.
To accomplish the required mass transfer within the brief flight times of the droplets, the droplets should be extremely small and be of exact or very similar droplet sizes, or at least be almost entirely and reliably below a critical droplet size, so as to closely approach equilibrium with the surrounding gas. For example, in some examples, the droplets should be less than 100 microns in diameter. In other examples, the droplets should be less than 60 microns in diameter.
Furthermore, the distance between the edge 21 of the spinning disc 20 and the walls 34, 26 of the chamber 14 should be selected to allow the droplets to closely approach saturation with the surrounding gas prior to being arrested against the walls 34, 36. Accordingly, the chamber 14 should have a chamber diameter C sufficiently larger than disc diameter D such that the droplets have an extended life within the atmosphere prior to their coalescence into larger droplets or against a surface of the chamber walls 34, 36.
Generally, the chamber diameter C will be selected such that the droplets will tend to come to rest within the atmosphere before contacting the chamber walls 34, 36. Thus, the particles will have an extended life within the gas prior to coalescence so as to obtain a desired equilibrium level.
However, in some cases, the chamber diameter C may be sufficiently small so that the droplets tend to reach the walls 34, 36 before being arrested by the atmosphere in the chamber 14, thus coalescing on the walls 34, 36.
Once arrested within the atmosphere (or on the walls 34, 36), the gas-liquid droplets will tend to collect and/or grow and will eventually fall into the lower tank head 18, where they can be subsequently extracted via the outlet fitting 63. In this manner, the apparatus 10 can be used to provide for mass transfer of gases into liquids.
Generally, the following equation can be used to estimate the diameter of water droplets produced by the spinning disc 20:
d=4/[Ω(Dρ/σ)1/2] (1)
where d is the droplet diameter in centimeters, Ω is the rate of rotation of the disc 20 in revolutions per minute (RPM), D is the disc diameter in centimeters, ρ is the density of the liquid medium being dispersed as droplets, and σ is the surface tension of the liquid medium.
In some cases, where the liquid does not perfectly wet the spinning disc 20, this equation should be corrected by dividing the answer by cos(φ), where φ is the wetting contact angle. For example, water often does not have a wetting reaction with metal surfaces (e.g. a metal spinning disc 20). Accordingly, in some examples such surfaces may be chemically or physically modified (e.g. using a coating) to provide hydrophilic surfaces, where cos(φ) is roughly equal to unity.
It has been found that the roughly monodisperse droplets produced by the spinning disc 20 travel a given fixed distance in the surrounding gaseous medium (based on the operating conditions of the apparatus 10) before their velocity declines to essentially the ambient drag velocity within the gas. The result is a cloud of droplets accumulating in a dense and stationary ring at a generally fixed distance from the spinning disc 20. This fixed distance generally follows the form:
X/d=P (2)
where X is the distance the primary droplet travels from the spinning disc in centimeters before the droplet loses their kinetic energy and come roughly to rest, and P is a constant that may be determined by observation. For example, for water droplets released into air at ambient pressure, P is equal to 2540.
Substituting equation (1) into (2), and adding a term to account for the viscosity of a surrounding atmosphere in the chamber 14 (e.g. carbon dioxide) under pressure as compared with ambient air, the following equation may be obtained to solve for the distance X:
X=10,100/[Ω(Dρ/σ)1/2]*(ηair/ηco2) (3)
The ratio of viscosities for air (171 micro Poise) and carbon dioxide (139 micro Poise) is approximately 1.23, and this is roughly independent of the surrounding gas pressure. The surface tension of water is approximately 72 dynes/cm, and the density of water is 1.00 grams/cm3, all at a temperature of approximately 4° C.
In some embodiments, the maximum flow rate, Qmax of liquid that can be fed onto the spinning disc 20 is limited by the volume that would “flood” the surface and inhibit the formation of small droplets. This maximum flow rate is roughly equal to:
Q
max=π2D2Ωd=(4π2D2)/(Dρ/σ)1/2 (4)
According to one example, the apparatus 10 was configured with the spinning disc 20 having a disc diameter D of 10 cm, and using an AC synchronous motor to drive the drive mechanism 60.
When operating such an apparatus 10 with the disc 20 rotating at 3600 RPM, a carbon dioxide atmosphere with an absolute pressure of 45 psi (roughly 3 atmospheres) within the chamber 14, and water as the liquid, droplets of 0.00298 cm (roughly 30 micron) can be produced. Under these conditions, droplets of this size tend to be thrown a distance of approximately 9.2 cm from the edge 21 of the disc 20 prior to being arrested by their friction within the surrounding gas.
Accordingly, the chamber diameter C should be made larger than 28.4 cm to enhance the contact time between droplets or particles and the surrounding atmosphere in the chamber 14 and provide for improved dispersion of the carbon dioxide into the water. After coalescing, the gas-liquid mixture can be collected in the bottom tank head 18, and subsequently extracted.
Alternatively, the chamber diameter C may be selected to be less than 28.4 cm if it is desired that the liquid droplets impact the walls 34, 36 of the chamber 14 rather than become entrained within the surrounding atmosphere.
The roughly 30 micron droplets produced by the spinning disc 20 in this example will tend to achieve approximately 97% equilibrium with the surrounding carbon dioxide atmosphere in approximately 0.05 seconds after leaving the edge 21 of the disc 20. However, because of time spent by the liquid spreading upon the surface of the disc 20 (prior to separation from the edge 21), the actual equilibrium results are generally better than is predicted by the diffusion into droplets alone.
If the walls 34, 36 of the chamber 14 in this example are selected to be larger than the specified 28.4 cm, then the droplets produced by the spinning disc 20 will tend to accumulate within a dense cloud at this distance, and will have much greater residence time within the gas atmosphere of the chamber 14 prior to coalescing into larger droplets.
The maximum recommended flow rate (Qmax, calculated using equation (4) above) for this particular example is approximately ten liters of liquid per minute. It can be seen by inspection of equation (4) that the maximum flow rate of the apparatus 10 can be improved by increasing the size of the disc 20, and not through an increase in the speed of rotation of the disc 20. The system can be operated above the Qmax value, but generally only in cases where mass transfer is favored, such as in carbonation.
In some examples, a rotating capillary may be used in an apparatus instead of the spinning disc 20. For example, illustrated in
The rotor assembly 90 generally includes one or more surfaces sized and shaped so as to define at least one capillary, and is configured to be rotated at an angular velocity selected such that liquid received in an inner region will adopt an unsaturated condition on each surface (as the liquid moves outwardly) such that the liquid flows as a film along the at least one surface and does not continuously span the capillary. Upon reaching the edge of the capillary, the liquid separates to form particles or droplets.
As shown, the rotor assembly 90 typically includes a set of circular plates (e.g. an upper plate 92 and a lower plate 94) spinning together on a hub or spindle 96. The upper and lower plates 92, 94 are spaced apart by a gap distance “d” and generally define the capillary therebetween.
In this embodiment, the liquid is provided into an inner region 97 of the rotor assembly 90 using a feed tube 98. The liquid is then allowed to flow into the capillary (e.g. between the two plates 92, 94, in some cases via apertures 99 in the feed tube 98). As the rotor assembly 90 rotates, the liquid moves outwardly between the plates 92, 94, reaching the edges 93, 95 of the plates and eventually separating from the edges 93, 95 as particles (e.g. fine ligaments, droplets or fibers, depending upon the properties of the liquid and the operating conditions of the rotor assembly 90).
In some examples, the liquid may transition from saturated flow (e.g. flow that spans the gap width d) to unsaturated flow (e.g. flow that does not span the gap width but which exists as thin films) within the capillary and before separating from the edges 93, 95. In an unsaturated condition, the liquid does not span the entire gap width, but rather exists as separate thin films on the surfaces of each of the upper plate 92 and lower plate 94, as urged by the increasing centripetal force as the liquid moves toward the outer edges 93, 95 of the plates 92, 94.
The use of such a spinning rotor assembly 90 tends to allow roughly double the flow rates, since two surfaces are being used for the release of the droplets.
In some examples, the rotor assembly 90 may be provided and operated within a tank 12 in a manner similar to that of the disc 20 as described above.
In some examples, the two plates 92, 94 may be coated with a hydrophilic medium or other coating to facilitate a transition from saturated flow within the capillary to unsaturated flow.
In some examples, as shown in
In other examples, the edges 93, 95 may be blunt edges. In yet other examples, each of the edges 93, 95 may be bifurcated (e.g. the edges 93, 95 may be V-shaped or U-shaped) so as to provide an upper edge and lower edge on each of the edges 93, 95.
In some examples, three or more plates may be stacked together in an array in a rotor assembly. For example, the rotor assembly 90 may be modified by providing one or more intermediate rotor plates between the upper plate 92 and lower plate 94. These intermediate rotor plates will cooperate with the upper and lower plates 92, 94 so as to define capillaries between each pair of opposing surfaces. The intermediate plates may have sharp edges, blunt edges, bifurcated edges, or any combination thereof.
Further details on the rotor assemblies that may be used are described in the PCT Patent Application No. PCT/CA2009/000324 entitled “Apparatus, Systems and Methods for Producing Particles Using Rotating Capillaries”, filed on Mar. 16, 2009 in the Canadian Intellectual Property Office, the entire contents of which are hereby incorporated by reference.
According to some of the embodiments described herein, it is possible to achieve exceptionally high performance mass transfer of gases (e.g. carbon dioxide) into liquids (e.g. water).
Generally, the apparatus, systems and methods described can be used within very small or very large-scale applications, especially when such gas transfer is accomplished at elevated pressures and where the uniformity and proportion of gas transferred to each unit of liquid must be exceptionally precise.
For example, if the liquid reaches greater than 95% of equilibrium with the surrounding gas atmosphere in less than 50 msec, then the apparatus as described herein might be considered to be a “near perfect” mass transfer device, where liquid always emerges at the desired gas saturation regardless of the actual residence time.
The apparatus and methods can be used in applications where the mass-transfer process might support chemical or biological processes, or for use in producing carbonated liquids. Some examples include oxygen transfer to support fermentation, aerobic digestion, gas-liquid chemical engineering processes or three-phase processes (e.g. where a solid is dispersed in a liquid that contains a dispersed or dissolved gas).
One typical case is carbonation, where it is desirable to transfer a precise volume of carbon dioxide gas into a precise quantity of water. Excessive carbonation at elevated pressure tends to result in undesirable foaming or “flashing” of carbonated drinks dispensed through a nozzle (as in post-mix applications). Alternatively, inadequate carbonation results in a “flat tasting” drink.
Failure to obtain optimal carbonation is said to be the single most common and pervasive source of quality control problems in carbonated beverage production in post-mix systems, even more common than problems with syrup blending (e.g. Brix control).
By using the various embodiments described herein, it may be possible to accomplish a precise transfer of gas into a liquid without complexity or recourse to complex sensors, feedback loops, or controls. Instead, it is achieved through nearly instantaneous accomplishment of the desired equilibrium using physics and mass-transfer principles.
However, in some embodiments it may be desirable to provide for additional sensing or monitoring or both to provide for desired levels of mass transfer (e.g. for carbonation systems).
For example, in beverage systems, the quality of a carbonated beverage is considered to be very dependent upon the amount of carbon dioxide absorbed into the water or other liquid in the beverage.
Carbonation normally has three variables that affect the amounts of carbon dioxide, nitrogen or other gas that will be absorbed into the water or other liquid pre-mix: the pressure of the fluid, the temperature of the fluid, and the equilibrium amount of gas (e.g. carbon dioxide) that can be absorbed into the liquid.
Prior art carbonators are normally slow to reach the desired equilibrium state. Accordingly, this tends to result in over-carbonation of the liquid (e.g. excessive amounts of carbon dioxide are absorbed into the water or pre-mix liquid) or under-carbonation of the liquid (e.g. too little carbon dioxide is absorbed into the water or pre-mix liquid). Generally, pressures in these prior art carbonation systems are manually calibrated “on site” (e.g. where the carbonation occurs) in an attempt to provide the desired equilibrium depending on the temperature of the liquid supply (e.g. the water supply temperature). However, the liquid supply temperature is prone to change, which can lead to deviations from the desired equilibrium operating conditions.
However, new developments in carbonator technology, including those based on the teachings as described herein, now provide for the ability to more quickly reach equilibrium states for gas absorption (in some cases virtually instantaneously). Accordingly, using the teachings as described herein, the desired equilibrium can be more consistently reached regardless of the various temperatures and pressures experienced in the system.
In particular, these apparatus and systems can encourage reaching an equilibrium condition within the mass transfer apparatus and provide for consistent gas absorption, and can be used to address changes in the temperature of the liquid supply.
For example, shown in
The system 400 generally includes a carbonator apparatus 402 (or other mass transfer apparatus, which could be a mass transfer apparatus 10 as described above configured to operate as a carbonator).
A pump 404 may be used to provide water or another liquid (e.g. a liquid pre-mix) to the carbonator 402 through one or more fluid conduits. The pump 404 may be used to adjust the pressure of the incoming liquid supply to ensure that adequate liquid pressure is provided.
A temperature sensor 406 monitors the temperature of the water or other liquid as it is provided to the carbonator 402. The temperature sensor 406 can send information about the water temperature to an electronic control unit 408.
The control unit 408 could include a microprocessor, a dedicated control unit, a software control unit, or other suitable control systems. The control unit 408 is generally used to control one or more operational aspects of the system 400 as will be described below.
Carbon dioxide (or another gas, e.g. nitrogen) may be provided to the system 400 from a tank 412 or another gas supply.
In some cases, the incoming gas may pass through a gas regulator 414, which may regulate the pressure of the incoming gas (e.g. reduce high pressure carbon dioxide in the tank 412 to lower pressure carbon dioxide suitable for use in the carbonator 402).
The supply of carbon dioxide or other gas to the carbonator 402 may be controlled by a valve 416 (e.g. a solenoid or another variable switch mechanism).
A pressure sensor 410 can monitor the pressure within the carbonator 402 and provide pressure information to the control unit 408.
During operation, the control unit 408 can be used to control various operational aspects of the system 400. For example, the control unit 408 may monitor the temperature of the incoming liquid (e.g. using the temperature sensor 406) and the pressure within the carbonator (e.g. using the pressure sensor 410), and then repeatedly control the operating pressure within the carbonator 402 to achieve desired operating conditions in the carbonator 402 (e.g. for equilibrium to occur based on the current water temperature) by opening and closing the valve 416.
As the water temperature changes, the control unit 408 can adjust the pressure within the carbonator 402 to desired levels (e.g. to maintain the desired equilibrium) using the valve 416 to provide for efficient operation of the carbonator 402.
In some embodiments, the control unit 408 may be provided with one or more lookup tables for determining the desired pressure for the carbonator 402 based on the measured temperature of the liquid supply.
In some embodiments, the control unit 408 may be configured so as to operate the valve 416 to control the supply of gas to the carbonator 402 only under certain conditions (e.g. when the carbonator 402 is operating and not when the carbonator is deactivated).
In some embodiments, the control unit 408 may be used to control the operating characteristics of the carbonator 402 (e.g. by adjusting the velocity of the rotating capillary apparatus, etc.).
In some embodiments, the control unit 408 may be used to control the operating characteristics of the pump 402. For example, the system 400 may include a pressure sensor configured to monitor the pressure from the water supply and activate the pump accordingly.
In some embodiments, one or more heat transfer devices (e.g. heaters, coolers, heat exchangers, etc.) may be used to adjust one or more of: the temperature of the incoming liquid, the temperature of the incoming gas, and the temperature of the carbonated output.
In various embodiments, the liquid may be a beer, a fruit juice, a flavored juice, a coffee or coffee-flavored beverage, or a tea.
Turning now to
Generally, in beverages such as beer it is desirable that the taste of the beverage be as consistent as possible (e.g. taste between various batches should not change). One important factor in taste is carbonation (e.g. the amount of carbon dioxide in the beer). However, with prior art carbonation systems it is quite common for beer to be under-carbonated (in some cases as much as 66% of beer will be under-carbonated), which is generally undesirable. However, using the system 500 as described herein it may be possible to provide for more accurate carbonation of beverages such as beer.
The system 500 generally includes a carbonator 502 or another mass transfer apparatus, which could for example be any of the carbonators or mass transfer apparatus as described herein. The carbonator 502 is coupled to and receives water (or another liquid) from a liquid supply 504 (e.g. a water supply tank).
Generally, the water in the water supply tank 504 is very pure, and may be bottled water or water treated by for example, demineralization and/or filtration (e.g. using reverse osmosis). In some embodiments the water in the water supply tank 504 may be as pure as possible so as to minimize (or at least greatly reduce) any impact that the water may have on the taste of the carbonated beer (or other beverage) that is output by the system 500, as will be explained below.
In some embodiments, the water supplied to the carbonator 502 may be cooled using a chiller 503. This may be useful to cool the carbonator 502 and to obtain desired operating characteristics therein (e.g. lowering the pressure of fluids within the carbonator, etc.).
The carbonator 502 is also coupled to a gas supply 504 (e.g. a gas supply tank) for receiving carbon dioxide or another gas therefrom.
Generally, the carbonator 502 is operated at temperatures and pressures selected to introduce large quantities of gas (e.g. carbon dioxide) into small amounts of liquid (e.g. water) so as to output a “highly gassed” liquid (e.g. carbonated water), also referred to as a “carrier fluid” (e.g. the small amounts of water act as a “carrier” for large quantities of carbon dioxide). Furthermore, when using the carbonator 502 to carbonate beer, the carbonator 502 is generally operated so that the small amount of water allows the desired levels of carbonation to be achieved without the water diluting the taste of the beer.
For example, in some embodiments the “highly gassed” carbonated water carrier fluid output by the carbonator 502 may be provided at a ratio of about ten volumes of carbon dioxide gas to one volume of water (or more), where “volume” is defined as the volume the carbon dioxide gas would occupy if it were removed from the liquid (e.g. water) at atmospheric pressure and a predetermined temperature (e.g. at STP or “standard temperature and pressure”, namely at a temperature of 25 degrees Celsius and a pressure of 1 atmosphere) as compared to the original (degassed) volume of liquid. In other embodiments, the carbonated water output may be provided at 10 to 20 volumes. In other embodiments, the carbonated water output may be provided at more than 20 volumes.
In some embodiments, the carbonator 502 operates at high pressures (such as between 100 and 200 psi, or even higher). In such embodiments, the chiller 503 may be useful to cool the carbonator 502 to reduce the operating pressures therein as described above.
The highly gassed carbonated water output by the carbonator 502 is generally fed into one or more mixers 508 for mixing the highly gassed carbonated water with beer and/or another beverage. In some embodiments, the carbonated water output by the carbonator 502 may be fed directly into the mixer 508. In other embodiments, carbonated water output by the carbonator 502 may be fed to one or more holding tanks 509 before being sent to the mixer 508.
As shown, one or more beers or other beverages may be fed into the mixer 508 so as to be mixed with the highly gassed carbonated water. For example, in the illustrated embodiment, the system 500 includes three supplies of beverages being fed into the mixer 508, namely, a first beer 510, a second beer 512, and another beverage 514.
In some cases, the beers 510, 512, and beverage 514 may be mixed with the highly gassed carbonated water in separate mixers 508, for example, when it is undesirable to mix the beers 510, 512, and beverage 514 with each other. The beers 510, 512, and beverage 514 may then exit their respective mixers 508 as a first carbonated beer 511, a second carbonated beer 513 and a carbonated beverage 515.
In other embodiments, it may be possible to mix a first one of the beers 510, 512, and beverage 514 in the mixer 508, and then mix a subsequent one of the beers 510, 512, and the beverage 514 in the mixer 508. Furthermore, the mixer 508 may be cleaned before mixing the subsequent one of the beers 510, 512, and the beverage 514 in the mixer 508.
Generally, the beers 510, 512 and other beverages entering the mixer 508 may be flat (e.g. with no carbonation therein) or may be partially carbonated (e.g. with slight amounts of carbonation). Accordingly, the system 500 can be operated so as to carbonate the beverage, for example, such that the carbonated beers 511, 513 and carbonated beverage 515 have desired levels of carbonation.
For example, it is common for some beers to have desired carbonation levels of 1.5 to 3 volumes of carbon dioxide as compared to the volume of beer.
Accordingly, the concentration of carbon dioxide gas in the highly gassed carbonated water can be set and/or adjusted (based on the quantities that will be added in the mixer 508) to achieve the desired carbonation levels in the carbonated beers 511, 513 and carbonated beverages 515 output by the system 500. In some cases, the carbonated beers 511, 513 and carbonated beverages 515 output by the system 500 may be between 1.5 to 3 volumes of carbon dioxide as compared to the volume of beer or beverage.
The system 500 can be operated to carbonate different types of beverages including concentrated beverages and non-concentrated beverages. For example, some beers and certain other beverages may be provided in a non-concentrated form, and it may be undesirable to add significant amounts of water to the non-concentrated beverage that would otherwise dilute the beverage. Accordingly, carbonating such non-concentrated beverages using the system 500 may be beneficial as relatively large quantities of carbon dioxide gas can be introduced into the beverage while only introducing relatively small amounts of water (thus tending not to significantly dilute the beverage).
Generally, the system 500 may provide a number of other benefits. In particular, the system 500 can be operated on site (e.g. at a restaurant) so that carbonated beer 511, 513 or other beverages 515 can be produced with highly accurate amounts of carbon dioxide, thus providing consistency of taste and freshness.
Furthermore, since the beers 510, 512, and other beverages 514 need not be carbonated when being delivered to the restaurant or other location, this can inhibit the need to use pressurized tanks or other containers when transporting the beer 510, 512, or beverage 514. In particular, the beer 510, 512, and beverage 514 can be transported when flat and then carbonated on site with a high degree of accuracy and without significantly diluting the taste of the beer. Furthermore, in some embodiments, the removing carbonation from the beer 510, 512, or beverage 514 during transport and storage may reduce the need to refrigerate the beer 510, 512, or beverage 514.
In some embodiments, the system 500 allows for carbonation of various different types of beer 510, 512 and other beverages 514 using only one carbonator.
In some embodiments, the system 500 may include one or more flow measurement sensors 520 (such as in-line sensors) for measuring the quantities of beer 510, 512, or other beverage 514 being fed into the mixer 508.
Turning now to
The apparatus 150 as shown generally includes a tank 152 and a rotor assembly 154 provided within the tank 152. As shown, the rotor assembly 154 has surfaces that define at least one capillary therein.
For example, the rotor assembly 154 may include one or more rotor plates that cooperate to define the capillary surfaces, including an upper rotor plate 156 near the top of the tank 152, a bottom rotor plate 158 near the bottom of the tank 152, and a plurality of intermediate rotor plates 160 between the upper and lower rotor plates 156, 158. The rotor plates 156, 158, 160 may be coupled to a common shaft or spindle 161 so that they rotate in unison as the apparatus 150 operates.
The apparatus 150 generally includes a first inlet 162 for providing a first gas into the chamber 163 of the tank 152. As shown, the inlet 162 may be located at or near the bottom of the tank 152.
The apparatus also includes a first outlet 164 for extracting fluids from the chamber 163. As shown, the first outlet 164 may be located at or near the top of the tank 152.
The apparatus 150 also includes a second inlet 166 (e.g. a feed tube) for providing liquid to the rotor assembly 154.
In some embodiments, the apparatus 150 may also include a first filter member 168. As shown, the filter member 168 may be positioned between the lower rotor plate 158 and the first inlet 162. The filter member 168 may inhibit excess liquid in the tank 152 from flowing outwardly through the first inlet 162.
In some embodiments, the tank 152 may include a fluid outlet 170 for draining excess liquid from the chamber 163 of the tank 152. As shown, the fluid outlet 170 may be provided near the filter member 168.
During use, a gas may be fed into the tank 152 through the first inlet 162. The gas may then pass upwardly through the filter member 168 into the chamber 163, where the gas can interact with the particles of liquid as they separate from the surfaces of the spinning rotor plates 156, 158, 160. The resulting mixture can then be extracted through the outlet 164.
Generally, the apparatus 150 may provide for increased rates of chemical reaction, enhanced mass transfer, or both, as compared to conventional systems (which may use for example use packed beds of materials to effect mixing between a liquid and a gas).
For example, in some embodiments the particles of liquid will react with the gas within the chamber according to some chemical reaction, optionally with or without the presence of a catalyst.
In some embodiments, at least one of the temperature and pressure within the tank 152 may be adjusted to achieve a desired chemical reaction. For example, in some embodiments the pressure in the tank 152 may be greater or lower than atmospheric pressure. In some embodiments, the temperature in the tank 152 may be raised (e.g. using heaters) or lowered (e.g. using a cooling apparatus). Raising or lowering the temperature and pressure within the tank 152 may be used to increase or reduce particular rates of chemical reaction to achieve desired results.
In one embodiment, the apparatus 150 may be used as a fluid purifier. Fluids frequently become contaminated during use and must normally be purified before they can be recycled or reused. For example, lubricants, hydraulic fluids, transformer oils, and cutting fluids often become contaminated with water, cleaning solvents, or other volatile contaminants which must be separated from the fluids before the fluids can be reused.
A variety of fluid purifiers have been previously designed based on the use of heat or vacuum or both to separate a volatile contaminant from a fluid. One problem with previous fluid purifiers is providing sufficient purification in a single pass through the purifier without harming the fluid itself. Purifiers with harsh processing conditions, such as excessive heat or excessive vacuum, may provide sufficient purification in a single pass, but they often have destructive effects on the fluids being purified. For example, the fluid can be seriously altered through the loss of low boiling point components, removal of additives, or oxidation or charring of the fluid.
On the other hand, purifiers with milder processing conditions, such as lower temperature or lower vacuum, may not harm the fluid being purified, but they often provide only partial purification in a single pass. The fluid must be pumped through the purifier many times for sufficient purification. This multi-pass approach substantially increases the amount of energy and time needed to purify the contaminated fluid.
However, using the apparatus 150, it is generally possible to provide fluid purification of large quantities of fluid and to high purity levels in a single pass without providing harsh processing conditions. In particular, a plurality of rotor plates can be stacked together to provide the desired number of capillary surfaces, thus providing for very high quantities of fluid throughput. Furthermore, this can be accomplished without elevated temperatures or pressures, which could have undesirable effects on the fluids.
For example, a contaminated liquid can be provided to the rotor apparatus 154 though the inlet 166, and the separated into particles. Since the particles tend to have very large surface area to volume ratios (as the particles are very small), contaminants within the liquid tend to be released from the particles where they can react with, and/or be absorbed by, the gas.
In some embodiments, a gas containing undesired contaminants may be purified by passing the gas through the chamber 153 and using a liquid in the rotor assembly 154 that reacts with or binds to the undesired contaminants, which will tend to remove the contaminants from the gas (and may result in the contaminants being collected as excess liquid that can be drained using the outlet 170). In this manner, the gas can be “scrubbed” or cleaned.
In some embodiments the apparatus 150 may provide for desired liquid particle sizes, including fine mists or sprays, without the need to increase the pressure within the tank 152. This is in contrast to conventional spray-type devices, which may generate sprays of liquid and gas using elevated pressures.
Generally, the apparatus 150 is viscosity independent, and can be operated with high viscosity and low viscosity liquids. Accordingly, the apparatus 150 can provide for mass transfer, and/or chemical reaction of highly viscous fluids (e.g. heavy oils, etc.), and which is normally very difficult using conventional apparatus.
Furthermore, the apparatus 150 may also be operated in conditions where the viscosity in the liquid is increasing, such as during a polymerization reaction.
In some embodiments, the apparatus 150 may operate as a heat exchanger, with hot oil or another liquid being provided and separating from the rotor plates to heat gas passing through the tank 152. In such embodiments, a filter may be provided in or before the outlet 164 so as to remove any liquid particles (e.g. oil) from the heated gas before it passes through the outlet 164.
Turning now to
This provision for two emitting surfaces (e.g. edges 177a, 177b) on each intermediate rotor plate 160 tends to double the potential for particle production and a stack of such plates increases particle production enormously as compared to a single flat disc or even a stack of discs that use a single surface.
In some embodiments, the bifurcated edges 177 may be sharpened edges so as to inhibit the formation of pools of liquids thereon.
In other embodiments, the edges of the rotor plates 156, 158, 160 may have various configurations to form particles of different sizes and shapes. For example, some intermediate rotor plates 160 could be provided with blunt edges, while other rotor plates 160 could have sharp edges or bifurcated edges 177. In this manner, it may be possible to form particles having different sizes and shapes simultaneously using the rotor assembly 154.
As shown, the rotor plates 156, 158, 160 define a plurality of capillaries 167, including a first capillary 167a between the upper rotor plate 156 and the first intermediate rotor plate 160a, a second capillary 167b between the first intermediate rotor plate 160a and the second intermediate rotor plate 160b, a third capillary 160c between the second intermediate rotor plate 160b and the third intermediate rotor plate 160c, and so on.
Each capillary 167a, 167b, 167c has a corresponding gap distance d1, d2, d3, In some embodiments, the gap distances d1, d2, d3, may each be the same or be generally similar. In other embodiments, the gap distances d1, d2, d3, may vary. For example, a first gap distance d1 may be selected so as to be greater than a second gap distance d2, and so on.
Turning now to
For example, as shown in
In some embodiments, the curved portions 202 may be made of steel or another metal.
As shown in
This wetting tends to create shear within the liquid 204, and may be beneficial in helping to reducing foaming of the liquid particles, which may occur due to air entrainment and which may be undesirable in certain embodiments.
While the above description provides examples of one or more methods and/or apparatuses, it will be appreciated that other methods and/or apparatuses may be within the scope of the present description as interpreted by one of skill in the art.
This application is a continuation-in-part of International Application No. PCT/CA2009/000323 filed on Mar. 16, 2009, and entitled “Apparatus, Systems and Methods for Mass Transfer Of Gases Into Liquids”, the entire contents of which are hereby incorporated herein by reference for all purposes, and this application is also a continuation-in-part of International Application No. PCT/CA2009/000324 filed on Mar. 16, 2009, and entitled “Apparatus, Systems and Methods for Producing Particles”, the entire contents of which are hereby incorporated herein by reference for all purposes, and this application claims the benefit of US Provisional patent application Serial No. filed May 4, 2010 and entitled “Apparatus, Systems and Methods for Gassing a Fluid By Introduction of a Highly Gassed Carrier Fluid”, the entire contents of which are hereby incorporated by reference herein for all purposes.
Number | Date | Country | |
---|---|---|---|
61331351 | May 2010 | US |
Number | Date | Country | |
---|---|---|---|
Parent | PCT/CA2009/000323 | Mar 2009 | US |
Child | 13100229 | US | |
Parent | PCT/CA2009/000324 | Mar 2009 | US |
Child | PCT/CA2009/000323 | US |