The present invention relates to an apparatus and a method for fixation of ankle syndesmosis.
Ankle syndesmosis disruptions are usually caused by severe external rotation ankle injuries. Surgery is recommended to reduce and internally fix the diastasis to prevent lateral talar shift, which could otherwise lead to post-traumatic arthrosis. Such surgical treatment usually involves tibio-fibular transfixation using a syndesmosis screw as recommended by the A.O. group (Arbeitsgemeinschaft für Osteosynthesefrage (Association for the Study of Internal Fixation)). Disadvantages of syndesmosis screw fixation include the need for a second operation for implant removal; implant fatigue and breakage; and loss of diastasis reduction following implant removal. Furthermore, prolonged non-weight bearing to avoid implant breakage prior to removal may cause further morbidity. In addition, studies have shown ligament healing to be inhibited by full immobilisation.
Movement of the distal fibula relative to the tibia is seen in normal ankle motion. Rigid fixation of the ankle syndesmosis, therefore, prevents normal physiological movement, until the rigid fixation device is removed, loosens or breaks.
Various methods of syndesmosis fixation have been studied before, including bioabsorbable implants (Thordarson D B, Hedman T P, Gross D, Magre G. “Biomechanical evaluation of polylactide absorbable screws used for syndesmosis injury repair” Foot Ankle Int 1997; 18: 622-7) and flexible implants (Miller R S, Weinhold P S, Dahners L E. “Comparison of tricortical screw fixation versus a modified suture construct for fixation of ankle syndesmosis injury: a biomechanical study” J Orthop Trauma 1999; 13: 39-42; Seitz W H Jr, Bachner E J, Abram L J, Postak P, Polando G, Brooks D B, Greenwald A S. “Repair of the tibiofibular syndesmosis with a flexible implant” J Orthop Trauma 1991; 5: 78-82). Seitz used a suture-button fixation using a large polyethylene button, as is commonly used for tendon repair pull-out sutures and a No. 5 braided polyester suture. Seitz's operative technique involved opening both the medial and lateral sides of the ankle. On biomechanical testing, failure occurred through the polyethylene button at an average of 20 kg of tension, and through the suture at 28 kg. Clinical testing in 12 patients showed good results, one patient having a symptomatic medial button. Buttons were routinely removed at 8 to 12 months, and were all found to be intact. Miller compared a modified suture construct against tricortical screw fixation at 2 cm and 5 cm above the ankle mortise. This method also required opening both the medial and lateral sides of the ankle. No. 5 braided polyester suture was looped through two holes drilled across the distal tibia and fibula. Similar results were seen for the suture and screw fixations, with a better holding strength for both groups at 5 cm.
It is an object of the present invention to overcome the problems associated with the prior art, whilst permitting normal physiological movement of the fibula relative to the tibia.
An apparatus for performing ankle syndesmosis repairs according to an exemplary aspect of the present disclosure includes, inter alia, a first button, a second button, and a suture connecting the first button and the second button. The first button and the second button are stainless steel buttons. At least one of the first button and the second button is oblong. The suture includes multiple strands that extend between the first button and the second button. A first free end of the suture is tensionable to shorten a length of the suture between the first button and the second button and thereby move the first button and the second button closer together.
In a further non-limiting embodiment of the foregoing apparatus, the suture is a braided polyethylene suture.
In a further non-limiting embodiment of either of the foregoing apparatuses, the suture is non-absorbable.
In a further non-limiting embodiment of any of the foregoing apparatuses, the suture is double looped through the first button and the second button.
In a further non-limiting embodiment of any of the foregoing apparatuses, the suture is passed through at least one opening in both the first button and the second button.
In a further non-limiting embodiment of any of the foregoing apparatuses, the suture is arranged to include at least four strands extending between the first button and the second button.
In a further non-limiting embodiment of any of the foregoing apparatuses, the suture includes a second free end, and the first free end and the second free end extend through the second button.
In a further non-limiting embodiment of any of the foregoing apparatuses, the first free end and the second free end are tied together in a knot over the second button.
In a further non-limiting embodiment of any of the foregoing apparatuses, in use, the first button is adapted to rest against a medial cortex of a tibia and the second button is adapted to rest against a lateral cortex of a fibula.
In a further non-limiting embodiment of any of the foregoing apparatuses, a pull-through device is connected to the first button by a second suture.
A method of ankle syndesmosis repair according to another exemplary aspect of the present disclosure includes, inter alia, drilling a hole through a fibula and a tibia, passing a first button through the hole until the first button exits on a medial side of the tibia, flipping the first button so it rests against a medial cortex of the tibia, approximating a second button to a lateral side of the fibula by applying traction to a suture that extends between the first button and a second button, and tying a knot in free ends of the suture to secure the second button against a lateral cortex of the fibula.
In a further non-limiting embodiment of the foregoing method, passing the first button includes connecting the first button to a pull-through device with a pull-through suture, and inserting the suture passing device through the hole to advance the first button device horizontally through the hole.
In a further non-limiting embodiment of either of the foregoing methods, flipping the first button includes applying traction to the pull-through suture while applying counter-traction to the suture until the first button device pivots from a position generally parallel to the hole to a position generally transverse to the hole.
In a further non-limiting embodiment of any of the foregoing methods, the method includes removing the pull-through suture after flipping the first button.
In a further non-limiting embodiment of any of the foregoing methods, the suture is double looped through the first button and the second button.
In a further non-limiting embodiment of any of the foregoing methods, applying the traction to the suture includes applying traction to the free ends of the suture, the free ends extending through the second button.
In a further non-limiting embodiment of any of the foregoing methods, the method includes visualizing movement of the first button using an image intensifier as the first button is passed through the hole.
In a further non-limiting embodiment of any of the foregoing methods, the method includes visualizing flipping of the first button using the image intensifier.
In a further non-limiting embodiment of any of the foregoing methods, drilling the hole includes drilling through the fibula with a drill bit, and drilling through the tibia using the same drill bit.
In a further non-limiting embodiment of any of the foregoing methods, the method includes visualizing the drill bit during the drilling using an image intensifier.
These and other features and advantages of the present invention will become apparent from the following description of the invention that is provided in connection with the accompanying drawings and illustrated embodiments of the invention.
The apparatuses, methods and buttons of the present invention are illustrated with respect to the following drawings:
a show an anterior view and a schematic view, respectively, of a normal acromioclavicular joint;
a show an anterior view and a schematic view, respectively, of a Rockwood Type III acromioclavicular joint dislocation, with superior migration of the clavicle with respect to the acromium;
a show a plan and an undersurface view, respectively, of a first or second suture anchor in the form of a washer of the present invention;
b illustrates the mobile positioning of the washer against an arcuate undersurface of the screw-head of a bone anchor;
a-16f illustrate, in sequence, the steps of a method according to the present invention.
The present invention provides minimally invasive, flexible fixation of the ankle syndesmosis whilst resisting tibio-fibular diastasis. It allows physiological micromotion at the ankle syndesmosis. There is no need for routine removal of the implant and its use should enable patients to weight-bear at an earlier stage.
The present invention is indicated for use in the fixation of ankle syndesmosis tibio-fibular diastasis (splaying apart). These are typically seen in Weber C-type ankle injuries, caused by severe pronation-external rotation forces. The fibula is fractured above the level of the syndesmosis. A medial ankle injury (malleolar fracture or deltoid ligament rupture) is also usually present. Reduction and fixation of the ankle syndesmosis is necessary to prevent lateral talar shift, which can lead to premature ankle osteo-arthritis.
Thus, referring to the accompanying drawings, the apparatus of the present invention comprises a pair of buttons 10, which, in the preferred embodiment illustrated are 9 mm by 3.5 mm in dimension, more particularly in length and width respectively. The buttons 10 are preferably formed from titanium or stainless steel, although it will of course be appreciated that any other suitable material could be used, in particular any suitable bioabsorbable material. The pair of buttons 10 each have a first aperture 12 and a second aperture 14 which, in the preferred embodiment illustrated, are triangular in shape, each of the first and second apertures 12, 14 having an apex 16, the respective apices 16 preferably being directed away from one another and being located substantially about a longitudinal mid-line of the button 10. Referring in particular to
In the present embodiment, leading and trailing edges of the button 10 of the present invention are substantially symmetrical, although it will be appreciated that this is not a requirement of the present invention. Specifically, the leading edge 24 of the button 10 of the present invention should be blunt and should have a width sufficient to reduce the possibility that the leading edge 24 of the first button 10 follows the second or pull through suture 20 through the intact medial skin or to catch or skewer any soft tissue structures between the bone and the medial skin, as will be described in detail hereinafter.
The button 10 of the third aspect of the present invention may be provided with apertures 12, 14 which are countersunk (not illustrated) so as to allow easier threading passage of the first and second sutures 18, 20. Care needs to be taken in such countersinking, to avoid compromising the mechanical strength of the first and second apertures 12, 14 of the button 10 of the present invention.
The first suture 18 used in the apparatus of the present invention can be of any material, which is suitable for this purpose, whether absorbable or non-absorbable, provided it is sufficiently strong. A number 5-strength braided polyester (ETHIBOND—Trade Mark) suture is preferred. This is a non-absorbable suture which knots easily without slipping.
The second suture 20 used in the present invention can be of any material which is suitable for this purpose, provided it is of at least 0-strength.
The pull through needle 22 can be of any dimensions, provided it is long enough to span the ankle. Its tip can be either “taper cut” or “cutting”.
Set-Up
The patient is positioned supine on a radiolucent operating table (not shown). Intra-operative fluoroscopy is necessary during the procedure. The patient and all theatre personnel should be adequately protected for x-ray radiation. A sandbag (not shown) is placed under the ipsilateral buttock to facilitate internal rotation of the leg. Antibiotic prophylaxis and the use of a tourniquet are recommended.
Instrumentation
An A.O. small fragment set (or equivalent) should be used for fracture osteosynthesis. The 3.5 mm drill bit is required for drilling the hole 30 through both the fibula 26 and tibia 28, for the first button 10 and first and second sutures 18, 20 to pass through, as illustrated in
Fracture Fixation
Osteosynthesis should be undertaken according to A.O. principles of internal fixation. It is recommended that fractures (not shown) in the lower half of the fibula 26 should be fixed. High fibular fractures (Maisonneuve injury) can be managed by addressing the syndesmosis diastasis only. Care should be taken not to injure the superficial peroneal nerve during the lateral approach to the fibula 26; the nerve passes posteriorly to anteriorly as it pierces the deep fascia. A one-third tubular plate usually provides sufficient stability and can be contoured easily to sit on the bone. The use of a lag screw for fracture compression is rarely required, once fibular length and rotation have been corrected.
Syndesmosis Reduction
The syndesmosis is reduced by internal rotation of the ankle, at around 30° of plantar flexion. This does not result in an over-tightening of the syndesmosis. Reduction should be confirmed using the image intensifier.
Drilling
All four cortices are drilled from the open lateral side using the 3.5 mm drill bit. The drill (not shown) should be angled at 30° upwards from the horizontal, at a distance of 2-3 cm above the ankle joint. Placing a finger on the medial aspect of the leg can help with aiming and feel when the drill has passed through. The drill hole 30 may go through one of the holes of a one-third tubular plate (not shown), if needed. To ensure accurate placement, drilling should be performed under image intensifier control.
Button Placement
The long straight needle 22 with pull-through, second suture 20 is passed through the drill-hole 30 and out the intact medial skin (see
Post-Operative Management
Following wound closure, the ankle should be placed in either a well-padded below-knee cast or backslab, ensuring the ankle is kept in a neutral position. The patient should be kept non-weight bearing for the first two weeks, and then allowed to partial weight-bear (50%) from two to six weeks in cast, depending on fracture stability. Full weight bearing can be allowed out of cast at six weeks.
Implant Removal
Routine removal of the suture-button construct is not required. If, for any reason, it needs to be removed, this can be performed simply by small incisions over the medial and lateral buttons 10, cutting the first suture 18 as it loops through the button 10 and removing the pair of buttons 10 and the first suture 18.
Phase One aims to reproduce a cadaver model of a syndesmosis injury, with a medial deltoid ligament rupture. An intact fibula simulates an anatomically fixed fracture. Phase two compares the suture-button versus conventional A.O. screw fixation following total intraosseous membrane (IOM) division, in a model resembling a Maisonneuve injury.
Material and Methods
Sixteen embalmed cadaver legs (eight pairs) were used. For each leg (not shown), the tibia and foot were fixed to a customised jig using Steinman pins. The foot was fixed to a mobile footplate so that the centre of rotation was directly under the centre of the ankle joint. External rotation moment was applied tangential to the centre of rotation at a radius of 25 cm. 1 kg of weight used therefore corresponds to approximately 2.5 Newton-meters of torque. The syndesmosis was exposed via an antero-lateral approach. Marker pins were placed in the tibia and fibula at the level of the syndesmosis to aid clinical and radiographic measurements. Clinical measurements were made using vernier calipers. In order to reduce bias, x-rays received a coded label to help blind subsequent review. The distance between the tips of the marker pins was measured on the mortise view x-ray. The stress lateral view was found to be less reliable, due to lack of reproducibility.
A 5 kg (12.5 Nm) load was used for all phase one measurements. Following baseline readings, the medial deltoid and syndesmotic ligaments were divided. Measurements of diastasis were taken following 5 cm, 10 cm and total intraosseous membrane division.
In phase two, left and right ankles were randomised to receive a suture-button 10 (4 mm×11 mm; the button being a conventional button marketed by Smith & Nephew Inc. under Endo-Button®) or A.O. standard (4.5 mm) screw fixation (not shown). In both groups, the syndesmosis was first reduced by internal rotation of the footplate. A hole was then drilled from lateral to medial, at 30° anterior to the horizontal, 2 cm superior to the ankle joint.
In the suture-button group of the present invention, a 4 mm drill hole 30 was drilled through all four cortices. The no. 5 braided polyester first suture 18 was looped twice through first and second apertures 12, 14 of the first and second buttons 10. The second suture 20 was threaded through the first aperture 12 of the first button 10 and also through the needle 22. This needle 22 was passed into the drill hole 30 from the lateral side and out through the intact medial skin. Using the leading pull-through suture 20, the first button 10 was advanced horizontally along the drill hole 30 until it has exited the medial tibial cortex. Using the leading pull-through second suture 20, whilst maintaining traction on the braided polyester first suture 18, the first button 10 was flipped to engage and anchor against the medial tibial cortex. The second suture 20 was then pulled out. The second button 10 was tightened against the lateral fibular cortex by further manual traction on the braided polyester first suture 18. The first suture 18 was securely tied over the second button 10 when flush with the lateral fibular cortex. The progress of the first button 10 may be followed intra-operatively using an x-ray image intensifier (not shown), if available.
In the comparative group (A.O. screw), a 3.2 mm drill hole was drilled through all four cortices. The hole was measured, tapped and an A.O. 4.5 mm cortical screw inserted to engage all four cortices, maintaining the reduction of the syndesmosis, without compression.
Measurements of syndesmosis diastasis were taken both under direct vision and radiographically at increasing external rotation torques. Torque loads were increased in increments of 1 kg, to a maximum of 8 kg or until fracture or implant failure. In four ankles (two per group), fixations were also tested at 5 cm above the ankle joint, having removed the fixations at 2 cm, in order to determine the optimum level of fixation placement.
Results
In phase one, the mean values of the measured diastasis above the baseline value at 5 cm, 10 cm and total intraosseous (IOM) division under 5 kg (12.5 Nm) load were 3.7 mm, 5.5 mm and 7.2 mm, respectively (see
In phase two, there was a gradual diastasis with increasing torque load in both groups, which was probably due to the quality of the bone. The mean diastasis from baseline for the suture-Endo-Button® and the A.O. screw groups for torque loads increasing at 1 kg intervals, up to 8 kg, are shown in Table 2. These differences were not statistically significant (p=0.7, unpaired t-test,
The apparatus and method of the present invention did give a more consistent performance, though. The distribution of standard deviations for A.O. screw fixation was 0.64 mm higher than that for the apparatus and method of the present invention (95% C.I. 0.46 to 0.84, Hodges-Lehmann estimation of shift).
There were no implant failures in either group. There were two fibular fractures in the A.O. screw group, prior to reaching the 8 kg load (5 kg, 8 kg). Only measurements prior to fracture were used for analysis. By comparison, there was one fibular fracture in the group of the present invention (8 kg). Comparing fixation placement at 2 cm versus 5 cm showed no significant difference (Table 2).
Discussion
The cadaver model in this study was tested using a jig (not shown) generating external rotation torque, which reproduces the mechanism of syndesmosis injury and, therefore, reflects the clinical situation.
Syndesmosis diastasis is seen with increasing intraosseous membrane division, under an external rotation torque load. This corroborates the findings of previous studies, showing a significantly larger diastasis with greater intraosseous membrane division.
Regarding the level of placement of the fixation, there was a trend towards better fixation at 2 cm, although only a small sample size was tested (Table 2).
Flexible fixation gives a more physiological end-result, allowing for micromotion at the distal tibio-fibular joint. Implant fatigue or breakage is less likely and routine removal is not essential. This avoids the complication of loss of reduction following removal of fixation. Earlier weight-bearing may be allowed, depending on the overall fracture configuration.
The advantages of the suture-button technique are that it is simple, flexible, minimally invasive as the medial side does not need to be opened, and has given a consistent performance on biomechanical testing. Clinical testing of the suture-button in ankle injuries that require reduction and fixation of a syndesmosis diastasis is recommended.
Patients with Weber C ankle fractures who had suture-button fixation, were compound with a cohort of patients who had syndesmosis screw fixation.
Methods
8 patients had suture-button fixation. The buttons used in Example 2 were conventional buttons supplied by Smith & Nephew Inc. and marketed under Endo-Button®. A retrospective cohort of 8 patients with similar Weber C fractures, treated using syndesmosis screw fixation, were recalled for clinical and radiological evaluation. Outcome was assessed using the American Orthopaedic Foot and Ankle Surgeons (AOFAS) score on a 100-point scale.
Results
Patients with screw fixation had a mean AOFAS score of 79 (range: 61-100) at an average follow-up of four months (range: 3-6 months). The suture-button group had a mean score of 92 (range: 76-100) at three-month review (p=0.02, unpaired t-test). Six of the screw group required further surgery for implant removal, compared to none of the suture-button group (p=0.007, Fisher's exact test).
Conclusion
Patients treated using the suture-button 10 regained a better functional outcome, within a shorter time frame. The technique is minimally invasive, as the medial side is not opened, and allows tibio-fibular micromotion whilst resisting diastasis. The need for secondary surgery for implant removal is significantly lessened. The suture-button technique may become the gold standard for syndesmosis diastasis injuries.
The present invention also provides minimally invasive, flexible fixation of the AC joint dislocation by resisting superior migration of the clavicle with respect to the coracoid process. It allows physiological micromotion at the AC joint. There should be no need for routine removal of the implant.
The present invention is indicated for use in the fixation of AC joint dislocation. These are typically seen in Rockwood type III AC joint dislocations, usually caused by severe downward blunt trauma to the point of the shoulder, or acromium. Typically, the clavicle is upwardly displaced as a result of the injury because of disruption to the AC and coracoclavicular ligaments. Reduction and fixation of displaced AC joint dislocations are necessary to prevent painful deformity and loss of function.
a show anterior and schematic views of a normal shoulder 10.
Referring to
A Rockwood type III AC joint dislocation is characterized by the disruption of the AC and the coracoclavicular ligaments 18, 24, respectively. As shown in
Repair of the type III shoulder dislocation according to the present invention is an out-patient procedure performed with a general anesthetic. The procedure is done with the patient lying supine on the operating table, preferably in the “deck-chair” position to allow the surgeon full access to the affected shoulder.
Referring to
Referring to
Reference is now made to
The washer 60 also has a substantially centrally located bone screw-retaining aperture 62. In the illustrated embodiment, the aperture 62 has a diameter of about 4.6 mm and the washer 60 is adapted to allow mobile positioning against an arcuate undersurface 69 of the head of the bone screw 68 (illustrated in
Referring to
The first suture 70 used in the apparatus can be made from any material which is suitable for this purpose, whether absorbable or non-absorbable, provided it is sufficiently strong. A number 5-strength braided polyester (FIBERWIRE®) suture is preferred. This is a non-absorbable suture which knots easily without slipping. The second suture 74 can be made from any material which is suitable for this purpose, and preferably should be at least 0-strength.
The pull through needle 72 can be of any dimensions, provided it is long enough to span the clavicle 12 or the coracoid process 14 of the shoulder 10. The needle 72 is preferably about 100 mm in length. The needle's body can either be straight or curved. The needle's tip can be either “taper cut” or “cutting.”
In the present embodiment, leading and trailing edges of the button 50 are substantially symmetrical, although it will be appreciated that this is not a requirement of the present invention. Specifically, the leading edge 56 (illustrated in
The following sets out the procedure, as shown in
Set-Up
The patient is positioned in a “deck-chair” position on the operating table (not shown). A sandbag (not shown) can be placed under the scapula to ease access to the shoulder region. A longitudinal or horizontal incision of about 5 cm is made on the skin, at the front of the shoulder, overlying the coracoid process 14 and the clavicle 12. The clavicle 12 and the superior surface of the coracoid process 14 are exposed by blunt dissection. As explained in detail below, if the clavicle hole 80 is to be drilled (
Instrumentation
A 3.5 mm drill bit is required for drilling a hole 80 through the clavicle 12. A 2.5 mm drill bit is required for drilling a hole 82 into the base of the coracoid process 14 of the scapula (
Button Placement
As illustrated in
In
The two trailing ends of the first suture 70 (
The volume between the arcuate undersurface 69 of the bone screw 68 and the coracoid process 14 defines the maximum flexibility of the washer 60 therebetween. The designed flexibility is helpful in increasing the tolerance for non-aligned drill holes and the like.
Post-Operative Management
Following wound closure, the shoulder should be placed in a shoulder immobilizer for three weeks. Gentle range of motion exercises can begin after three weeks. Full range exercises can be allowed after six weeks.
Implant Removal
Routine removal of the first suture anchor-suture-second suture anchor construct is not required. If, for any reason, it needs to be removed, this can be performed simply by re-opening the surgical incision, cutting the first suture 70 as it loops through the button 50 and removing the button 50. The screw 68 and washer 60 can be removed easily using the screwdriver.
It is noted that the above description and drawings are exemplary and illustrate preferred embodiments that achieve the objects, features and advantages of the present invention. It is not intended that the present invention be limited to the illustrated embodiments. Any modification of the present invention which comes within the spirit and scope of the following claims should be considered part of the present invention.
Number | Date | Country | Kind |
---|---|---|---|
S2002/0504 | Jun 2002 | IE | national |
This is a continuation of U.S. application Ser. No. 14/933,269, filed on Nov. 5, 2015, which is a continuation of U.S. application Ser. No. 13/970,269, filed Aug. 19, 2013, which is a divisional of U.S. application Ser. No. 11/482,038, filed Jul. 7, 2006, now U.S. Pat. No. 8,512,376, which claims the benefit of U.S. Provisional Application No. 60/697,125 filed on Jul. 7, 2005, and which is a continuation-in-part of U.S. application Ser. No. 10/233,122, filed Aug. 30, 2002, now U.S. Pat. No. 7,235,091, which claims priority under 35 U.S.C. § 119 to IE S2002/0504, filed Jun. 20, 2002. The entire disclosures of all of the above priority applications are incorporated herein by reference.
Number | Name | Date | Kind |
---|---|---|---|
2765787 | Pellet | Oct 1956 | A |
3176316 | Bodell | Apr 1965 | A |
3762418 | Wasson | Oct 1973 | A |
4187558 | Dahlen et al. | Feb 1980 | A |
4301551 | Dore et al. | Nov 1981 | A |
4400833 | Kurland | Aug 1983 | A |
4776851 | Bruchman et al. | Oct 1988 | A |
4790850 | Dunn et al. | Dec 1988 | A |
4792336 | Hlavacek et al. | Dec 1988 | A |
4851005 | Hunt et al. | Jul 1989 | A |
4863471 | Mansat | Sep 1989 | A |
4917700 | Aikins | Apr 1990 | A |
4932972 | Dunn et al. | Jun 1990 | A |
4988351 | Paulos et al. | Jan 1991 | A |
5024669 | Peterson et al. | Jun 1991 | A |
5026398 | May et al. | Jun 1991 | A |
5129902 | Goble et al. | Jul 1992 | A |
5171274 | Fluckiger et al. | Dec 1992 | A |
5211647 | Schmieding | May 1993 | A |
5217495 | Kaplan et al. | Jun 1993 | A |
5219359 | McQuilkin | Jun 1993 | A |
5263984 | Li et al. | Nov 1993 | A |
5266075 | Clark et al. | Nov 1993 | A |
5306290 | Martins et al. | Apr 1994 | A |
5306301 | Graf et al. | Apr 1994 | A |
5320626 | Schmieding | Jun 1994 | A |
5397357 | Schmieding et al. | Mar 1995 | A |
5409490 | Ethridge et al. | Apr 1995 | A |
5562669 | McGuire | Oct 1996 | A |
5575819 | Amis et al. | Nov 1996 | A |
5628756 | Barker et al. | May 1997 | A |
5643266 | Li et al. | Jul 1997 | A |
5645588 | Graf et al. | Jul 1997 | A |
5830234 | Wojciechowicz | Nov 1998 | A |
5921986 | Bonutti | Jul 1999 | A |
5931869 | Boucher et al. | Aug 1999 | A |
5961520 | Beck, Jr. et al. | Oct 1999 | A |
5964764 | West et al. | Oct 1999 | A |
6056752 | Roger | May 2000 | A |
6099530 | Simonian et al. | Aug 2000 | A |
6099568 | Simonian et al. | Aug 2000 | A |
6110207 | Eichhorn et al. | Aug 2000 | A |
6117160 | Bonutti | Sep 2000 | A |
6159234 | Bonutti et al. | Dec 2000 | A |
6193754 | Seedhom | Feb 2001 | B1 |
6203572 | Johnson et al. | Mar 2001 | B1 |
6238395 | Bonutti | May 2001 | B1 |
6283996 | Chervitz et al. | Sep 2001 | B1 |
6296659 | Foerster | Oct 2001 | B1 |
6325804 | Wenstrom, Jr. et al. | Dec 2001 | B1 |
6517578 | Hein | Feb 2003 | B2 |
6533802 | Bojarski et al. | Mar 2003 | B2 |
6635073 | Bonutti | Oct 2003 | B2 |
6641596 | Lizardi | Nov 2003 | B1 |
6716234 | Grafton | Apr 2004 | B2 |
7097654 | Freedland | Aug 2006 | B1 |
7235091 | Thornes | Jun 2007 | B2 |
7494506 | Brulez et al. | Feb 2009 | B2 |
7686838 | Wolf et al. | Mar 2010 | B2 |
7749250 | Stone et al. | Jul 2010 | B2 |
7776039 | Bernstein et al. | Aug 2010 | B2 |
7819898 | Stone et al. | Oct 2010 | B2 |
7828855 | Ellis et al. | Nov 2010 | B2 |
7875057 | Cook et al. | Jan 2011 | B2 |
7875058 | Holmes, Jr. | Jan 2011 | B2 |
7905903 | Stone et al. | Mar 2011 | B2 |
7914539 | Stone et al. | Mar 2011 | B2 |
8109965 | Stone et al. | Feb 2012 | B2 |
8118836 | Denham et al. | Feb 2012 | B2 |
8162997 | Struhl | Apr 2012 | B2 |
8206446 | Montgomery | Jun 2012 | B1 |
8231654 | Kaiser et al. | Jul 2012 | B2 |
8388655 | Fallin et al. | Mar 2013 | B2 |
8512376 | Thornes | Aug 2013 | B2 |
8821551 | Zeetser et al. | Sep 2014 | B2 |
9259217 | Fritzinger et al. | Feb 2016 | B2 |
20010041938 | Hein | Nov 2001 | A1 |
20020019634 | Bonutti | Feb 2002 | A1 |
20020161439 | Strobel et al. | Oct 2002 | A1 |
20020198527 | Muckter | Dec 2002 | A1 |
20030114929 | Knudsen et al. | Jun 2003 | A1 |
20030130694 | Bojarski | Jul 2003 | A1 |
20030236555 | Thornes | Dec 2003 | A1 |
20040015171 | Bojarski et al. | Jan 2004 | A1 |
20040059415 | Schmieding | Mar 2004 | A1 |
20040073306 | Eichhorn et al. | Apr 2004 | A1 |
20040116963 | Lattouf | Jun 2004 | A1 |
20040236373 | Anspach, III | Nov 2004 | A1 |
20040243235 | Goh et al. | Dec 2004 | A1 |
20040267360 | Huber | Dec 2004 | A1 |
20050004670 | Gebhardt et al. | Jan 2005 | A1 |
20050033363 | Bojarski et al. | Feb 2005 | A1 |
20050065533 | Magen et al. | Mar 2005 | A1 |
20050070906 | Clark et al. | Mar 2005 | A1 |
20050137704 | Steenlage | Jun 2005 | A1 |
20050149187 | Clark et al. | Jul 2005 | A1 |
20050171603 | Justin et al. | Aug 2005 | A1 |
20050203623 | Steiner et al. | Sep 2005 | A1 |
20050261766 | Chervitz et al. | Nov 2005 | A1 |
20060067971 | Story et al. | Mar 2006 | A1 |
20060095130 | Caborn et al. | May 2006 | A1 |
20060142769 | Collette | Jun 2006 | A1 |
20060190041 | Fallin et al. | Aug 2006 | A1 |
20060264944 | Cole | Nov 2006 | A1 |
20060265064 | Re et al. | Nov 2006 | A1 |
20070021839 | Lowe | Jan 2007 | A1 |
20070083236 | Sikora et al. | Apr 2007 | A1 |
20070118217 | Brulez | May 2007 | A1 |
20070162123 | Whittaker et al. | Jul 2007 | A1 |
20070162125 | LeBeau et al. | Jul 2007 | A1 |
20070179531 | Thornes | Aug 2007 | A1 |
20070225805 | Schmieding | Sep 2007 | A1 |
20070239209 | Fallman | Oct 2007 | A1 |
20070239275 | Willobee | Oct 2007 | A1 |
20070250163 | Cassani | Oct 2007 | A1 |
20070270857 | Lombardo et al. | Nov 2007 | A1 |
20080046009 | Albertorio et al. | Feb 2008 | A1 |
20080082128 | Stone | Apr 2008 | A1 |
20080177302 | Shurnas | Jul 2008 | A1 |
20080188935 | Saylor et al. | Aug 2008 | A1 |
20080188936 | Ball et al. | Aug 2008 | A1 |
20080195148 | Cook et al. | Aug 2008 | A1 |
20080208252 | Holmes | Aug 2008 | A1 |
20080215150 | Koob et al. | Sep 2008 | A1 |
20080228271 | Stone et al. | Sep 2008 | A1 |
20080234819 | Schmieding et al. | Sep 2008 | A1 |
20080243248 | Stone et al. | Oct 2008 | A1 |
20080275553 | Wolf et al. | Nov 2008 | A1 |
20080275554 | Iannarone et al. | Nov 2008 | A1 |
20080300683 | Altman et al. | Dec 2008 | A1 |
20080312689 | Denham et al. | Dec 2008 | A1 |
20090018654 | Schmieding et al. | Jan 2009 | A1 |
20090030516 | Imbert | Jan 2009 | A1 |
20090036893 | Kartalian et al. | Feb 2009 | A1 |
20090054982 | Cimino | Feb 2009 | A1 |
20090062854 | Kaiser et al. | Mar 2009 | A1 |
20090082805 | Kaiser et al. | Mar 2009 | A1 |
20090187244 | Dross | Jul 2009 | A1 |
20090216326 | Hirpara et al. | Aug 2009 | A1 |
20090228017 | Collins | Sep 2009 | A1 |
20090234451 | Manderson | Sep 2009 | A1 |
20090265003 | Re et al. | Oct 2009 | A1 |
20090275950 | Sterrett et al. | Nov 2009 | A1 |
20090306776 | Murray | Dec 2009 | A1 |
20090306784 | Blum | Dec 2009 | A1 |
20090312776 | Kaiser et al. | Dec 2009 | A1 |
20100049258 | Dougherty | Feb 2010 | A1 |
20100049319 | Dougherty | Feb 2010 | A1 |
20100100182 | Barnes et al. | Apr 2010 | A1 |
20100145384 | Stone et al. | Jun 2010 | A1 |
20100145448 | Monies De Oca Balderas et al. | Jun 2010 | A1 |
20100211075 | Stone | Aug 2010 | A1 |
20100211173 | Bárdos et al. | Aug 2010 | A1 |
20100249930 | Myers | Sep 2010 | A1 |
20100268273 | Albertorio et al. | Oct 2010 | A1 |
20100268275 | Stone et al. | Oct 2010 | A1 |
20100274355 | McGuire et al. | Oct 2010 | A1 |
20100274356 | Fening et al. | Oct 2010 | A1 |
20100292792 | Stone et al. | Nov 2010 | A1 |
20100305709 | Metzger et al. | Dec 2010 | A1 |
20100312341 | Kaiser et al. | Dec 2010 | A1 |
20100318188 | Linares | Dec 2010 | A1 |
20100324676 | Albertorio et al. | Dec 2010 | A1 |
20100331975 | Nissan et al. | Dec 2010 | A1 |
20110040380 | Schmieding et al. | Feb 2011 | A1 |
20110046734 | Tobis et al. | Feb 2011 | A1 |
20110054609 | Cook et al. | Mar 2011 | A1 |
20110087284 | Stone et al. | Apr 2011 | A1 |
20110098727 | Kaiser et al. | Apr 2011 | A1 |
20110112640 | Amis et al. | May 2011 | A1 |
20110112641 | Justin et al. | May 2011 | A1 |
20110118838 | Delli-Santi et al. | May 2011 | A1 |
20110137416 | Myers | Jun 2011 | A1 |
20110184227 | Altman et al. | Jul 2011 | A1 |
20110196432 | Griffis, III | Aug 2011 | A1 |
20110196490 | Gadikota et al. | Aug 2011 | A1 |
20110218625 | Berelsman et al. | Sep 2011 | A1 |
20110224729 | Baker et al. | Sep 2011 | A1 |
20110238179 | Laurencin et al. | Sep 2011 | A1 |
20110270278 | Overes et al. | Nov 2011 | A1 |
20110276137 | Seedhom et al. | Nov 2011 | A1 |
20110282350 | Kowarsch et al. | Nov 2011 | A1 |
20110288635 | Miller et al. | Nov 2011 | A1 |
20110301707 | Buskirk et al. | Dec 2011 | A1 |
20110301708 | Stone et al. | Dec 2011 | A1 |
20120046746 | Konicek | Feb 2012 | A1 |
20120046747 | Justin et al. | Feb 2012 | A1 |
20120053630 | Denham et al. | Mar 2012 | A1 |
20120065732 | Roller et al. | Mar 2012 | A1 |
20120089143 | Martin et al. | Apr 2012 | A1 |
20120109299 | Li et al. | May 2012 | A1 |
20120123474 | Zajac et al. | May 2012 | A1 |
20120123541 | Albertorio et al. | May 2012 | A1 |
20120150297 | Denham et al. | Jun 2012 | A1 |
20120165938 | Denham et al. | Jun 2012 | A1 |
20120197271 | Astorino et al. | Aug 2012 | A1 |
20120296345 | Wack et al. | Nov 2012 | A1 |
20130023928 | Dreyfuss | Jan 2013 | A1 |
20130023929 | Sullivan et al. | Jan 2013 | A1 |
20130331886 | Thornes | Dec 2013 | A1 |
20170209196 | Zajac et al. | Jul 2017 | A1 |
Number | Date | Country |
---|---|---|
29910202 | Sep 1999 | DE |
20101791 | Jun 2001 | DE |
0440991 | Aug 1991 | EP |
1108401 | Jun 2001 | EP |
1707127 | Oct 2006 | EP |
2238944 | Oct 2010 | EP |
2007002561 | Jan 2007 | WO |
2008091690 | Jul 2008 | WO |
Entry |
---|
E.P. Su, et al., “Using Suture Anchors for Coracoclavicular Fixation in Treatment of Complete Acromioclavicular Separation,” The American Journal of Orthopedics, May 2004, pp. 256-257. |
Number | Date | Country | |
---|---|---|---|
20170209140 A1 | Jul 2017 | US |
Number | Date | Country | |
---|---|---|---|
60697125 | Jul 2005 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 11482038 | Jul 2006 | US |
Child | 13970269 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14933269 | Nov 2015 | US |
Child | 15483338 | US | |
Parent | 13970269 | Aug 2013 | US |
Child | 14933269 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 10233122 | Aug 2002 | US |
Child | 11482038 | US |