The present disclosure is directed to single-use medical devices and implements (hereafter “devices”). More specifically, the present disclosure is directed to single-use medical devices that, when subjected to autoclave processing for potential reuse, deform, destruct and/or provide a visual indication that the devices are not reusable.
Historically, medical devices, such as those utilized in surgery (e.g., scalpels, forceps, spreaders, scrapers, etc.) were reusable. That is, the devices were intended to be re-sterilized/decontaminated and reused after a medical procedure. Such re-sterilization has advanced over time from simply immersing the devices or tools in alcohol to the current practice of autoclave sterilization. In the latter regard, medical devices are inserted into an autoclave chamber where they are subjected to elevated temperatures, pressures and, commonly, steam. By way of example, many autoclaves used for sterilizing medical devices subject those devices to pressurized saturated steam at approximately 121° C. (250° F.) for around 15-20 minutes depending on the size of the load and the contents.
While autoclave sterilization is an industry accepted means for cleaning reusable medical devices, the process is not always one-hundred percent effective. That is, in some cases bacteria, pathogens or other contaminants survive the autoclave process. This is more common for devices that have internal passageways where such contaminants may be partially protected during the autoclave process. The inability to fully sterilize reusable devices can result in infection risks, when utilized with subsequent patients. Accordingly, there has been an ongoing shift in the medical field to single-use medical devices. Such devices are designed to be utilized once in a hospital or clinic and then disposed. In this regard, single-use devices are often delivered in sterilized packages, which are opened shortly before or during a medical procedure. In such an arrangement, no sterilization is required before use. Such single-use devices are intended to eliminate re-sterilization problems sometimes encountered by reusable medical devices. To ensure the single-use devices are not reused, manufacturers often put a single-use labels and warnings on the packaging and/or ISO symbols on the medical device itself alerting medical professionals that the device should be disposed after use.
The present inventor(s) has recognized that many single use devices (SUDs) are erroneously reused. By way of example, during surgery, it is common for all devices or tools used during surgery be placed in a bucket or tray after use. In many instances, reusable devices and SUDs may be comingled. If SUDs are comingled with reusable devices, the SUDs may be erroneously re-sterilized in an autoclave and reused. Further, SUDs may be purposely re-sterilized as a cost saving measure. However, as SUDs are not designed for reuse, operating parts of re-sterilized SUDS can be dull, loose, bent or otherwise out of calibration after autoclaving. The presented systems and methods prevent such unintentional or other off-label reuse of SUDS on patients.
In an arrangement, a destruct element (e.g., thermally deformable element) is incorporated into a SUD that deforms during autoclaving. Most commonly, at least a portion of the destruct element has a melting point below an autoclaving process temperature. That is, the destruct element has a low melting point of less than about 250° F., less than about 230° F. or even less than about 180° F. When subjected to autoclave processing temperatures, the destruct element deforms. Such deformation may provide a visual and/or tactile indication that the device is damaged and not intended for reuse. Size shape and color of the destruct element may be chosen to improve visual discernability. In some arrangements, deformation of the destruct element may prevent operation of a functioning or moving component(s) of the SUD. In any arrangement, a subsequent user of the device has in indication that the SUD has been unintentionally autoclaved in an off-label attempt to sterilize the SUD and that the SUD should not be utilized with a patient.
In an embodiment, a destruct element is disposed on a handle of a SUD. In such an embodiment, the destruct element may provide a visual or tactile indication that the SUD has improperly been exposed to an autoclave process.
In an embodiment, one or more destruct elements form or are disposed against a moving element of a SUD. Such SUDS may include, without limitation, retractable surgical knives and swiveling awls. In such embodiments, the destruct element(s) may prevent ordinary movement of the SUD once the SUD has been exposed to an autoclave process.
The foregoing and other aspects, features, details, utilities, and advantages of the present disclosure will be apparent from reading the following description and claims, and from reviewing the accompanying drawings.
Reference will now be made to the accompanying drawings, which at least assist in illustrating the various pertinent features of the presented inventions. The following description is presented for purposes of illustration and description and is not intended to limit the inventions to the forms disclosed herein. Consequently, variations and modifications commensurate with the following teachings, and skill and knowledge of the relevant art, are within the scope of the presented inventions. The embodiments described herein are further intended to explain the best modes known of practicing the inventions and to enable others skilled in the art to utilize the inventions in such, or other embodiments and with various modifications required by the particular application(s) or use(s) of the presented inventions.
Broadly, the present disclosure is directed to single-use medical devices (SUDS) that are configured to deform (e.g., destruct) if re-sterilized in an autoclave process. Generally, each SUD includes a destruct element (e.g., thermally deformable element) that deforms in response to elevated temperatures present during autoclaving. Typically, at least a portion of the destruct element is formed of a material having a melting point below an autoclaving process temperature. That is, at least a portion the destruct element is made of a material having a melting point of less than about 250° F., less than about 230° F. or even less than about 180° F. When subjected to elevated temperature, during an autoclave processing, the destruct element deforms. Preferably, the destruct element grossly deforms. In this regard, materials having a melting point that is significantly less than the autoclaving temperature (e.g., at least 20° F.) may be preferred. However, this is not a requirement. Such deformation may provide a visual indication that the device is damaged and should not be used with a patient. In some arrangements, such deformation may prevent operation of a functioning or moving component(s) of the SUD. In any arrangement, a subsequent user of the device has an indication that the SUD has been unintentionally autoclaved in an off-label attempt to sterilize the SUD and that the SUD should not be utilized with a patient.
In the illustrated embodiment of the retractable surgical knife 30, various components may be thermally deformable elements (e.g., destruct elements). For instance, the slider 36, thumb actuator 42 and/or the biasing pin 48 may be formed of a material having a low melting point (e.g., below an autoclaving temperature). Accordingly, when subjected to elevated temperatures during autoclaving that exceed the melting point of these component, the components may deform. In the case of the slider 36, thumb actuator 42 and biasing pin 48, such deformation may prevent operation of the retractable knife 30. That is, such deformation may prevent the advancement and retraction of the blade 32 rendering the retractable knife 30 inoperable. For instance, the slider 36 may be a thermally deformable element that may melt to and, for example, adhere to the interior of the handle 38 preventing movement of the blade 32 once the slider 36 has re-solidified (e.g., after cooling). Likewise, additional thermally deformable components may melt and adhere to adjacent components. For instance, the actuator 42 may melt and adhere to the handle 38 and the biasing pin 48 may melt into and adhere with the spring 50. Stated otherwise, one or more components may at least partially melt together and/or adhere to adjacent components preventing subsequent movement. In addition, the knife 30 may include one or more destruct elements that provide a visual indication that the knife has been re-sterilized. For instance, the distal ferrule 52 and/or the proximal ferrule/cap 54 may be formed of a low melting point material that deform during an autoclaving process like the destruct element discussed in
Though discussed above as utilizing a destruct element that deforms in response to elevated temperatures, it will be appreciated that the destruct element may be any component that provides an indication that the SUD should not be reused after exposure to elevated temperatures.
The aperture 123 extending through the ball 124 is elongated, which allows the housing 130 and attached awl 132 to pivot about the end of the shaft 122 in at least first and second axes. A friction bearing 126 (hereafter “bearing”) and spring 128 provide friction between the housing 130 and the ball 124. More specifically, when assembled, the spring 128 and a generally cylindrical bearing 126 are disposed within the housing 130 such that a proximal end of the bearing 126 is compressed against an outside surface of the ball 124 by the spring 128, which is itself compressed between a distal end of the bearing 126 and an inside surface (not shown) of the housing 130. Once assembled, the friction of the bearing 126 on the outside surface of the ball 124 provides resistance to movement, which allows positioning the awl 132 at a desired angle.
The bearing 126 is a thermally deformable destruct element in accordance with the present disclosure. That is, the bearing 126 has a low melting point of less than about 250° F., less than about 230° F. or even less than about 180° F. When subjected to autoclave processing temperatures, the bearing (e.g., destruct element) deforms or partially melts. Once the bearing 126 has sufficiently softened during an autoclave sanitation process, the spring 128 helps deform the bearing 126 by pushing it into the ball 124. In response to the pressure exerted by the spring 128, the bearing deforms into the functional swivel area (e.g., ball 124 and aperture 123, housing 130 and pin 129). Further, the spring 128 may become disposed within the distal end of the bearing 126. After an autoclaving process and the bearing material has re-solidifies (e.g., once cooled), the awl SUD 120 is inoperable. This is, due to the deformation of the bearing 126 during autoclaving into the functional swivel area, the housing 130 and awl 132 retain a fixed position relative to the shaft.
As noted above, the destruct elements are made of a material having a melting point that is less than a standard medical autoclaving temperature of approximately 121° C. or 250° F. A variety of materials may be utilized so long as those materials can be sterilized prior to the first use of a SUD (e.g., where such sterilization is by means other than elevated temperatures). Most commonly, the destruct material is a polymer. However, this is not a requirement. In one embodiment, the destruct material is Polycaprolactone (PCL), a biodegradable polyester with a low melting point of around 140° F. (60° C.). In another embodiment, paraffin wax may be utilized as the destruct material. Paraffin wax has a low melting point of approximately 154° F. In a further embodiment, the destruct material is a low-density polyethylene having a melting point of less that about 200° F. (90° C.). In a yet further embodiment, the destruct material is Polyethylene Glycol (PEG), which has a melting point of less than about 144° F. (62° C.). In various embodiments, the destruct material may be a material that acts similar to an adhesive when softened or melted. For example, various polymers may become sticky when softened/melted. Such materials tend to adhere to adjacent components when softened/melted. Once cooled (e.g., after autoclaving) such materials effectively bond to adjacent components thereby preventing movement of these components. Other materials are possible and considered within the scope of the present disclosure.
All directional references (e.g., distal, proximal, upper, lower, upward, downward, left, right, leftward, rightward, top, bottom, above, below, vertical, horizontal, clockwise, and counterclockwise) are only used for identification purposes to aid the reader's understanding of the present disclosure, and do not create limitations, particularly as to the position, orientation, or use of the any aspect of the disclosure. Joinder references (e.g., attached, coupled, connected, and the like) are to be construed broadly and may include intermediate members between a connection of elements and relative movement between elements. As such, joinder references do not necessarily infer that two elements are directly connected and in fixed relation to each other. It is intended that all matter contained in the above description or shown in the accompanying drawings shall be interpreted as illustrative only and not limiting. Changes in detail or structure may be made without departing from the spirit of the invention as defined in the appended claims.
The present application claims the benefit of the filing date of U.S. Provisional Patent Application No. 63/217,546 having a filing date of Jul. 1, 2021, the entire contents of which is incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
63217546 | Jul 2021 | US |