The present invention relates to intravenous infusion therapy. More specifically, the invention relates to a system, components of the system and methods associated with the system for organizing the fluid flow for applications which require an accommodation of a broad flow rate range, a wide range of input and output pressures, and a wide range of delivered fluid viscosities, such as those seen with Intravenous (IV) infusion therapy.
Conventionally, healthcare providers have had three technical options for the intravenous infusions. Many intravenous infusions are controlled by manually adjusting a resistance in the flow path between a fluid source and the patient, based on the operator's observation of the rate of drips formed within a chamber in line with the fluid flow. The flow rate range that can be controlled with this method is limited by the relatively large and fixed size of the drops and the relatively low reliability of the human operator to accurately compute the flow rate. This method is critically flawed by virtue of the fact that it requires a human observer to maintain an accurate and consistent flow rate. In many circumstances, a trained human observer is not available. This manual method also lacks an important ability to electronically record and communicate the results of the infusion.
A relatively small number of infusions are controlled with the use of a fixed volume of liquid under a fixed amount of pressure and a fixed resistance, providing a fixed flow rate. Unfortunately, the fixed rate and fixed fluid volume do not provide the flexibility required for most infusions. Similar to a manual infusion, this method does not provide the opportunity to electronically record the results of the infusion.
Because of the strong requirement for more precise control of flow rate, flexibility of fluid volumes, and the desire to keep track of the flow information, many infusions are controlled using a positive displacement fluid pump. These large fluid volume positive displacement devices are generally of the peristaltic or reciprocating piston type. Both types come at a price of complexity, size, weight, limited battery life, and significant financial cost. Early versions of positive displacement pumps created a new hazard for patients in what was known as “runaway infusion,” where the highly controlled fluid flow was suddenly uncontrolled when a door or other containment mechanism on the pump was released. In response to this undesirable feature, pumps were later required to incorporate “flow stop” mechanisms, so that the flow rate would stop entirely if the fluid tubing were removed form the flow control device. Unfortunately, the cessation of flow is sometimes as hazardous to patients as a sudden increase. Another unintended consequence of positive pumping systems is the possibility of infusing lethal amounts of air into a patient. This possibility did not exist with low pressure gravity infusions. As a result, positive displacement pumps have incorporated air detection systems to prevent this hazard, yet these alarm systems are the source of very significant nuisance alarms, resulting in operator inefficiency and patient anxiety.
The present invention recognizes that most intravenous infusions require only moderate infusion and that portability and simplicity should not be compromised for any infusion.
Intravenous infusions can be broken into two broad categories: infusions where the flow rates are relatively high with relatively low requirements for absolute flow accuracy, and infusions that must be made with maximal precision and flow continuity. The later classification includes those drugs which have powerful effects upon the cardiovascular system or powerful anesthetics. This invention covers the former classification which includes fluids for hydration, electrolytes, nutrients, antibiotics. These fluids and medications do not require great precision and comprise the overwhelming majority of infusions required for patients.
The invention is directed to fluid administration apparatus and a method for using this apparatus, comprising a fluid pathway assembly and a flow control device wherein fluid flowing through the fluid flow system is controlled via closed loop quasi-static adjustment of in-line pressure-based resistance.
The above and other features of the invention, including various novel details of construction and combinations of parts, will now be more particularly described with reference to the accompanying drawings and pointed out in the claims. It will be understood that the particular device embodying the invention is shown by way of illustration only and not as a limitation of the invention. The principles and features of this invention may be employed in various and numerous embodiments without departing from the scope of the invention.
These and other features, aspects, and advantages of the apparatus and methods of the present invention will become better understood with regard to the following description, appended claims, and accompanying drawings where:
a is a schematic representation of the fluid administration system in accordance with the present invention;
b shows the two major assemblies of the invention;
a shows an exploded view of the variable resistance mechanism of
b shows the collapsed view of
a, 8b, and 8c illustrate operation of the control knob and fluid flow restriction apparatus of the cassette;
a and 12b show a complete realization of the system of
a is a flow chart of the operations that comprise the rate targeting method;
b is a flow chart of the operations that comprise an alternative rate targeting method.
The essential components of the invention are described with reference to
Primary fluid source 102 is represented schematically as a large-volume bag. The content of primary fluid source 102 flows via primary check valve 104 and into primary input 106 to an impeller housing 108. The movement of fluid causes impeller 110 to turn in a given direction based on the geometry of impeller 110. In one embodiment of the present invention, impeller 110 is connected to permanent magnet 112, causing permanent magnet 112 to turn as impeller 110 turns. Electromagnet coils 114 may be used as a sensor to detect movement of permanent magnet 112. It is contemplated that other embodiments in accordance with the present invention will utilize a non-metallic impeller without detracting from the present invention.
Fluid exits impeller housing 108 via outlet channel 130 through an outlet tube 136 and travels via output check valve 132 to variable resistance 134, indicated schematically by a variable resistor. In practice, variable resistance 134 may optionally be realized as a variably pinched tube element (illustrated in
A secondary fluid source could include secondary fluid source 118 or injectable fluid reservoir 122. The optional connections of secondary fluid source 118 or injectable fluid reservoir 122 feed into fluid administration system 100 as indicated by the directional arrows leading from secondary fluid source 118 and injectable fluid reservoir 122 and join at secondary fluid input node 126. The selected secondary fluid source, either secondary fluid source 118 or injectable fluid reservoir 122, flows from secondary fluid input node 126 to secondary input 120 into impeller housing 108. Fluid flow arising from secondary input to impeller housing 120 and exiting from outlet channel 130 of impeller housing 108 will tend to move impeller 110. Given the proper geometry of impeller 110, the movement of impeller 110 and permanent magnet 112 will be the opposite of that when fluid flows through primary input 106 to impeller housing 108. Electromagnet coils 114 can measure both the speed and direction of permanent magnet 112 as it rotates.
Due to the symmetry of electromagnetism, electromagnet coils 114 may be used as a sensor or as an electromagnetic driver, in the same way that a loudspeaker may also be used as a microphone. It is possible, then, to use electromagnet coils 114 to apply a torque to permanent magnet 112, so that it tends to align with the magnetic field created by electromagnet coils 114. The magnetic field created by electromagnet coils 114 can be static or time variant.
The components of
The major elements of fluid pathway assembly 200 are shown in
In
A reverse view of fluid pathway assembly 200, shown in
The hardware design and recommended usage favor situations involving little overlap of fluid flows from the two directions, as described here by way of example. If a fluid source were connected to secondary fluid source connection receptacle 204 and if said source had a pressure greater than that of primary inlet line 202, then the normal action of primary one-way valve 402 would be to close. Flow from secondary fluid source connection 204 would then travel through secondary channel 406 on its way to outlet channel 412 and then to outlet tube 206. The geometry of the pathways within fluid pathway assembly 200 are such that flow from connection receptacle for secondary fluid source 204 will tend to spin in-line rotary flow impeller 210 in the opposite direction than flow arising from primary inlet line 202.
In a typical IV therapy situation, an on-going large-volume infusion from primary inlet line 202 may be interrupted by the administration of a smaller volume of antibiotic from connection receptacle for secondary fluid source 204. Common clinical practice is to infuse an antibiotic over a relatively short period of time, so that an effective level is reached in the patient's blood stream, followed by cessation of antibiotic infusion for a relatively long period of several hours. Thus, when an antibiotic is connected to connection receptacle for secondary fluid source 204 and placed higher than primary fluid reservoir 304, the infusion automatically switches over to pure antibiotic administration, which is detected and regulated by the system. The system similarly permits a rapid bolus infusion, for example from connection receptacle for secondary fluid source 204. Again, the bolus flow forces primary one-way valve 402 to close, so in-line rotary flow impeller 210 reverses it direction and allows for the separate measurement of the bolus.
a provides an exploded view illustrating the details of the resistive elements, shown schematically as variable resistance 134 in
b illustrates the components of
In another embodiment, bolus administration syringe 702 can be attached to secondary fluid source connection receptacle 204 as shown in
The close-up view of resistance control knob 208 in
b provides a transparent view when the resistive component (pinch bar 414) is at its minimum. Resistance control knob 208 has been rotated to a position where advancing nut 502 is fully retracted along the length of threaded shaft 503, away from outlet tube 206. The force exerted by occlusion spring 504 is minimal and outlet tube 206 is fully open as shown at open lumen section of outlet channel 802.
When the force of pinch bar 414 is at its maximum, outlet tube 206 is completely closed as shown at closed lumen section of outlet channel 804 in
In
When fluid pathway assembly 200 is mated with flow control device 900 to form fluid administration assembly 910, resistance control knob 208 is still visible and available to the user. Fluid pathway assembly 200 may be removably mated with flow control device 900 via mechanical means known to those of ordinary skill in the art, such as, but not limited to snaps, hook and loop type fasteners (e.g., VELCRO®) and similar attachment mechanisms. The operator (e.g., a physician, nurse or medical technician) may watch in-line rotary flow impeller 210 or drop forming chamber 302 for an approximate indication of flow rate, and subsequently watch a digital readout of flow rate on graphic display 912 inferred from the rotation of in-line rotary flow impeller 210. When the operator releases resistance control knob 208, automatic control is restored, and the motor and gear system monitors and adjusts the fluid resistance to maintain the flow rate, compensating for changing conditions such as declining fluid head height (as the bag empties), changing position of the patient's body, and physical changes in the pinched tube related to temperature and plastic creep.
If a secondary infusion is started via secondary fluid source connection receptacle 204, the operator may set a different flow rate for that infusion. The servomechanism will then maintain that flow rate in the secondary direction, until the impeller reverses direction, indicating depletion of the secondary source. The servomechanism can be programmed to respond to this flow direction reversal by using adjusting gear motor 506 to re-adjust resistance control knob 208 to restore the rate originally set for the primary infusion. Buttons or keys on the controller are not needed and control for the two flow rates is intuitive. The automatic transition from a secondary infusion to a primary infusion, upon depletion of the secondary fluid source has not heretobefore been disclosed in the art, and is made even more valuable by the simple and intuitive user interface.
In an alternative embodiment, control of resistance can be achieved entirely via adjusting gear motor 506 and drive wheel 508. When the operator turns resistance control knob 208, the effect is to control a visual indicator, such as graphic display 912, causing it to show the desired rate. The servomechanism then causes the actual rate to match the dialed-in rate. If for some reason the servomechanism is unable to achieve the dialed-in rate, the indicator can show the actual, incorrect rate (for example, zero rate with a blocked line or depleted source) while simultaneously broadcasting an alarm signal (for example, using secondary status light 914 or primary status light 916).
Note that flow control device 900 provides a uniquely simple interface, as shown in
A front plan view shown in
The interior of flow control device 900 is revealed in the perspective view of
A transparent front plan view of fluid administration assembly 910 is shown as
Measurements of such a system are illustrated in
Slower flow rates employ a combination of driving and sensing the fluid flow. In
A steady state electromagnetic field may be imposed using set of electromagnetic coils 934, as shown in
The results from a brief electromagnetic pulse, followed by a period of “reading” or sensing the movement of permanent magnet 212 are shown
The angular velocity is equal to the sum of the squares of orthogonal coil pairs. This computation is shown as angular velocity 978. Following a known magnetic impulse, it would be useful to compute the speed created by the imposed vector. Maximum angular velocity 980 provides the indication of maximum angular velocity. There are several measurements of interest in the time domain as well. One measurement, shown by way of example, time of maximum velocity 982, indicates how long it took for permanent magnet 212 to reach it maximum velocity. Measurements made as shown in
A major feature of the present invention is the quasi-static nature of the control loop. Once resistance control knob 208 has been established at a position that achieves the targeted flow rate, resistance control knob 208 is maintained in one spot to maintain said rate. Adjustments to resistance control knob 208, though the activation of adjusting gear motor 506, are required only in response to external changes in fluid pressure and resistance. The infrequent use of a motor is a key ingredient to conservation of power, which has high importance for a battery operated device. The periodic observation of in-line rotary flow impeller 210 by integrated circuitry 944 requires negligible energy consumption.
In accordance with the present invention, the apparatus of the invention affords the utilization of a sensor and an actuator with greater range and far higher degrees of resolution than was previously available. Additionally, the present invention provides for the use of quasi-static settings to provide the opportunity for vast reductions in energy consumption. The feature of high resolution for control comes from three characteristics of adjusting gear motor 506. First, adjusting gear motor 506 must be capable of moving in very small steps, certainly resulting in less than one micron of tube compression per step. Second, when the movement of said adjusting gear motor 506 has stopped, said motor must stay at the spot where movement stopped so there is no measurable backlash or mechanical hysteresis. Finally, adjusting gear motor 506 should have the ability to maintain its position without the expenditure of any additional energy.
Power consumption and required battery weight in this system can be made very low, due to the mode of operation. The flow rate need only be monitored intermittently, so that parts like integrated circuitry 944 and set of electromagnetic coils 934 need not be run continuously. Flow is propelled by gravity or by a pressurized container, rather than by pumping. Once set, fluid resistance is relatively stable, so adjusting gear motor 506 is activated only with a very low duty cycle to make small adjustments. Thus, in typical operation, integrated circuits 944 are responsive to an operator turning resistance control knob 208. Integrated circuits 944 are active until the operator has set a rate and that rate has been stabilized by the servo. The entire system then shuts down excepting for a wakeup timer that re-activates the microprocessor and pulses power to the flow measurement device, as needed to check the flow rate. On a given timed check, the control motor may require activation long enough for a small rate adjustment or may not be needed.
In one embodiment, the contained volume of in-line rotary flow impeller 210 is about 50 microliters. The rotational sensitivity of set of electromagnetic coils 934 is greater than 250 parts per revolution, yielding a sensitivity of movement on the order of 0.2 microliters (50 microliters per revolution divided by 250 counts per revolution). This sensor resolution compares favorably to the highest resolution “drop size” of 1/60th of a ml or 17 microliters.
Furthermore, a “drop counting” method does not support rapid flow rates and the user's ability to count rapid drops is limited as drops begin to merge or “stream” at higher flow rates. The present invention supports flow rates that are higher than those using drop counting methods. The ability of in-line rotary flow impeller 210 and integrated circuitry 944 to process information quickly is appreciated and known to be many orders of magnitude greater than the human. A nominally very high flow rate of 3,600 ml/h can be expressed as 1,000 microliters per second or nominally 20 revolutions per second of in-line rotary flow impeller 210. At 20 revolutions per second, integrated circuitry 944 would be required to service all of its functions in a period of 50 milliseconds. Commonly used microprocessors have many orders of magnitude of excess processing capacity for said task.
In an alternative embodiment, the sensing design provides a substitute for electromagnetic interactions. If in-line rotary flow impeller 210 were optical in nature, instead of magnetic, the rotary response of in-line rotary flow impeller 210 could be observed. The nature of the rotary encoder is not critical to this invention, so long at it reflects the true rotation of in-line rotary flow impeller 210.
a illustrates the logic of setting flow rate targets for flow originating from either primary inlet line 202 or secondary fluid source connection receptacle 204. First, a determination has been made if resistance control knob 208 has been moved by a user. If movement has occurred, then the most recently computed actual flow rate becomes the new target. This is analogous to the setting of an automotive cruise control target, where the SET button is used to maintain the currently set rate of speed.
b illustrates alternative logic of setting flow rate targets for flow originating from either primary inlet line 202 or secondary fluid source connection receptacle 204. First, a determination has been made if resistance control knob 208 has been moved by a user. If movement has occurred, then the stored targeted flow rate is incremented or decremented, depending on the direction of rotation of adjusting gear motor 506. The new targeted rate is indicated by graphic display 912 on flow control device 900 (see
While there has been shown and described what is considered to be preferred embodiments of the invention, it will, of course, be understood that various modifications and changes in form or detail could readily be made without departing from the spirit of the invention. It is therefore intended that the invention be not limited to the exact forms described and illustrated, but should be constructed to cover all modifications that may fall within the scope of the appended claims.
This application claims priority to U.S. Provisional Patent Application Ser. No. 60/608,956, filed on Feb. 20, 2004, which was converted from U.S. patent application Ser. No. 10/783,042 filed on Feb. 20, 2004 under 37 CFR 1.53(c)(2).
Number | Name | Date | Kind |
---|---|---|---|
3934473 | Griffo | Jan 1976 | A |
4489616 | Priddy | Dec 1984 | A |
4533347 | Deckert | Aug 1985 | A |
4876492 | Lester et al. | Oct 1989 | A |
Number | Date | Country | |
---|---|---|---|
20060004330 A1 | Jan 2006 | US |
Number | Date | Country | |
---|---|---|---|
60608956 | Feb 2004 | US |