The U.S. Food and Drug Administration (FDA) has developed a set of industry guidelines for applicants (e.g., pharmaceutical companies) who are planning product quality studies to measure bioavailability (BA) and establish bioequivalence (BE) in support of new drug applications (NDA) or abbreviated new drug applications (ANDA) for locally acting nasal sprays using metered-dose spray pumps. These guidelines include specific recommendations for BA and BE studies of prescription corticosteroids, antihistamines, anticholinergic drug products, and the over-the-counter (OTC) mast-cell stabilizer cromolyn sodium. The recommendations include seven tests and associated metrics and lifestages shown in Table 1, all of which should be conducted using validated analytical methods to characterize the in vitro performance of the products.
The Beginning of Life (BOL) lifestage is defined as the first actuation (s) following the labeled number of priming actuations. The End of Life (EOL) lifestage is defined as the actuation (s) corresponding to the label claimed number of actuations.
The FDA recommends using automated actuation systems when conducting the tests listed in Table 1, to decrease variability in drug delivery due to operator factors and, thus, increase the sensitivity for detecting potential differences between products. The FDA also recommends that the automated actuation system includes settings for force, velocity, acceleration, stroke length, and other relevant parameters. The FDA further recommends that the selection of appropriate settings used with the automated actuation system be relevant to proper usage of the product by the trained patient. The settings may be available from pump suppliers or by conducting an exploratory study in which the relevant parameters are varied to simulate in vitro performance upon hand actuation.
With regard to Tests 1 and 7 in Table 1, the FDA specifically recommends determining the delivered (e.g., emitted or ex-actuator) drug mass from the units. In conjunction with these FDA recommendations, the United States Pharmacopoeia (USP) provides specific test methods that should be followed for testing the delivered dose uniformity of nasal spray products, including the use of an automated actuation system. Both the FDA and USP recommendations state that the nasal spray product to be tested should be prepared as directed on the label and instructions for use, which invariably implies shaking and priming the product. Though not stated directly in either the FDA or the USP recommendations, measuring the delivered shot weight (e.g., the weight of the delivered spray) is often used as the primary indicator of pump delivery performance and as a supplemental measurement for delivered dose uniformity. Overall, these recommendations pose many challenges to organizations involved in testing nasal sprays, including:
The currently accepted method for performing spray content uniformity testing relies heavily on manual operations for actuation, sample weighing, shaking, data collection, and data analysis. Some companies employ a completely manual method with paper records, while other companies use a combination of automated actuation and manual sample collection with some electronic record keeping.
Regardless of the method employed, manual operations related to actuation, sample collection, and/or weighing are known to be fraught with problems such as human error, production inefficiencies, operator repetitive stress, and low data integrity. Additionally, paper records create a significant 21 CFR Part 11 (requirements related to electronic records and signatures) compliance challenge in today's pharmaceutical environment. These problems lead to production bottlenecks and are likely to cause additional testing due to current Good Manufacturing Practices (cGMP) requirements when data discrepancies appear, both of which may seriously affect the manufacturer's profitability. Some processes, however, are best handled manually. For instance, manually moving a sample or samples of collected doses to an automated sample preparation system may be perfectly acceptable because the sample preparation system may be a shared resource in a separate laboratory.
According to one embodiment, an automated system for testing a spray pump assembly includes a robotic handler, a tray for holding multiple spray pump assemblies and collectors, a spray pump assembly actuator, and a weighing device such as a balance. The robotic handler transports the spray pump assemblies and collectors between the tray, the spray pump assembly actuator, and the balance to facilitate automated testing of spray pump assemblies. The testing may include performing shot weight and spray content uniformity tests. The tray may include sensors associated with each spray pump assembly and collector to sense the presence of each spray pump assembly and collector. A system computer may use the sensor information to assist a user in loading the tray and to ensure proper operation of the system.
In one embodiment, the robotic handler includes an electromechanical gripper. The electromechanical gripper may include a rotary motor, a left- and right-handed linear screw, and first and second gripper elements. The first and second gripper elements are movably coupled to respective left- and right-handed portions of the linear screw. The linear screw, in turn, is coupled to the rotary motor. When a system controller sends control signals to the rotary motor, the rotary motor drives the linear screw-rail assembly to move the first and second gripper elements in opposite linear directions.
The first and second gripper elements include jaws to grasp objects (e.g., a spray device assembly) when the gripper elements are driven together or to release objects when the gripper elements are driven apart. The gripper elements may include movable jaws and sensors to sense movement of the jaws, for example, when the robotic handler moves an object held by the jaws towards a stationary object the robotic handler continues move in the same direction after the object makes contact with the stationary object.
The spray pump assembly may include a spray pump clamp which the robotic handler or actuator may more easily handle. The clamp may include a threaded aperture centered about a central axis of the clamp with a first diameter at a bottom side of the clamp and a second diameter at a top side of the clamp. The clamp may be secured to a nozzle tip of the spray pump by simultaneously inserting the nozzle tip into the aperture of the clamp and rotating the clamp until the clamp is secured to the nozzle tip.
The system may further include a nozzle tip dabber which the robotic handler may use to keep the nozzle tip of the spray pump clean. In one embodiment, the nozzle tip dabber includes a base, an absorbent pad, and a flexible pad. The flexible pad is attached to the base and the absorbent pad is attached to the flexible pad. The flexible pad improves the cleaning capabilities of the absorbent pad.
In another embodiment of an automated system for testing a spray pump assembly, the system includes a first testing region with a first testing device, a second testing region with a second testing device, an elevator assembly connecting the first and second testing regions, and a spray pump assembly actuator attached to the elevator assembly. The elevator assembly may move the actuator between the first and second testing regions to automate performing multiple tests on spray pump assemblies. The first testing region of the system may employ a robotic handler for handling and transporting spray pump assemblies. In one example embodiment, the first testing device is an analytical balance and the second testing device includes either (i) a camera and a first laser configured to measure spray pattern, or (ii) a second laser and a receiver configured to measure droplet size distribution through laser diffraction, or both.
The foregoing will be apparent from the following more particular description of example embodiments of the invention, as illustrated in the accompanying drawings in which like reference characters refer to the same parts throughout the different views. The drawings are not necessarily to scale, emphasis instead being placed upon illustrating embodiments of the present invention.
A description of example embodiments of the invention follows.
As described in more detail below, embodiments of the automated spray pump testing system 100 may incorporate the example features and benefits set forth in Table 2.
As indicated in Table 2, embodiments of the automated spray pump testing system 100 are stand-alone systems with electrical and computer network interfaces. Standalone systems provide various benefits including superior vibration isolation for faster weighing performance; simpler installation and supervision of operation; independence from laboratory variables such as bench space and availability, air currents, and fume exhaust; and integration of proven technologies.
The system 100 may include a unified steel frame 110 to provide simplified construction and vibration isolation for the system's testing and measurement devices. The elements of the system 100 include a testing area 112, a system computer 114, system controller 117, and Input/Output (I/O) devices 116 through which a user may interact with the system 100. As illustrated in
The testing area 112 of the system 100 includes a robotic device handler 120 that safely and reliably handles and transports devices and collectors within the testing area 112. In this embodiment, the robotic device handler 120 is an intelligent four-axis design that is able to handle spray devices of varying shapes and sizes. According to this embodiment, the robotic device handler 120 may be programmed to perform shaking and intra-actuation nozzle tip dabbing or cleaning of the spray devices. The robotic device handler 120 provides highly automated operation. The system 100 may operate without user intervention except for loading and unloading devices and samples, and for handling error conditions (e.g., network unavailability, mishandling of device or collection vessel, lack of airflow). Also, the system 100 may run uninterrupted over a period of time consistent with pharmaceutical production equipment available today, without routine maintenance other than cleaning.
The testing area 112 is enclosed within the steel frame 110 and sliding glass access doors 118. The sliding glass access doors 118 provide an operator an interface to the testing area 112 for loading or unloading spray device samples. The sliding glass access doors 118 may include tempered glass panels to reduce the effects of static charge buildup on the system's testing and measurement devices and to provide adequate operator safety.
The robotic handler 120 features an electromechanical gripper 220 for grasping and releasing spray devices or collectors. The electromechanical gripper is designed to minimize the amount of moving parts. The electromechanical gripper includes two stiff gripper elements 224a-b or gripper arms with movable jaws 226a-b and stationary jaws 228a-b. The gripper elements 224a-b are movably coupled to a low mass, high performance, linear screw-rail assembly 222. The linear screw-rail assembly 222 includes a half left-handed and half right-handed screw-rail spindle. The linear screw-rail assembly 222, in turn, is coupled to a rotary motor 221 through a drive coupler 223, which includes two pulleys and a drive belt. In this configuration, the rotary motor may drive the linear screw-rail assembly 222 to cause the gripper elements 224a-b to move in opposite directions (either towards or away from each other depending on the rotational direction that the rotary motor 221 drives the linear screw-rail assembly 222).
The electromechanical gripper 220 connects to a vertically-oriented linear screw-rail assembly 242 and a z-axis motor assembly 240. The z-axis of the robotic handler 120 is designed to be large and not to induce vibrations. The vertically-oriented (z-axis) linear screw-rail assembly 242 is movably coupled to an x-axis screw-rail assembly 232. When a rotary motor 241 of the z-axis motor assembly 240 drives the vertically-oriented linear screw-rail assembly 242, the vertically-oriented linear screw-rail assembly 242 together with the electromechanical gripper 220 and z-axis motor assembly 240 move with respect to the x-axis linear screw-rail assembly 232 along a z-axis (relative to the system 100). Another rotary motor 231 is coupled to the x-axis linear screw assembly 232 (see also
The x-axis linear screw assembly 232, in turn, movably couples to a first y-axis linear screw assembly 236 and a second y-axis linear screw assembly 238 (see
As illustrated in
The system controller 117 generates control signals to control the motion of the robotic handler 120 and the actuator 250 based on the commands from the system computer 114 and sensor signals from sensors integrated into the robotic handler 120 and the actuator 250. As described in more detail below, the holder tray 210 may also include sensors for sensing the presence of each spray device 311, waste collector 213, collection vessel 215, and nozzle tip dabber 218. The system computer 117 is configured to receive, process, and display, via an Input/Output (I/O) device 116, measurement data from the analytical balance 255 and sensor signals from the holder tray 210, actuator 250, or robotic handler 120. For example, the system computer 114 may process sensors signals from the holder tray 210 and visually or otherwise present to the user via I/O devices 116 which elements (e.g., spray device, waste collector, collection vessel) need to be inserted and where they should be inserted.
As indicated in Table 2 above, the system 100 may further incorporate an air handling subsystem to minimize the adverse effects of prolonged user exposure to the spray formulations. For example, the air handling subsystem may remove and/or control odors. The air handling subsystem may also use carbon-based filtration and provide sufficient airflow throughout the testing area 112. The air handling subsystem, however, may be designed so that air currents do not detract from the performance of the analytical balance 255.
The system computer 114 may include a database 115 to store measurement data or other system information. In addition, the system computer 114 may run on a secure database software platform in order to comply with 21 CFR Part 11 and other company standards (e.g., actuation system and method) and for consistent data storage and retrieval formats. The software platform may be centered on processing the results from the numerous device actuations or “actuation events”. Actuation events may be linked directly to the devices from which originated (e.g. a unique identifying number such as lot number or manufacturing batch ID), in addition to other information such as who actuated the device, when the actuation occurred, etc. Additionally, unifying the actuation events by device type, lot number, and device identifier may be implemented to simplify data management and analysis and to allow a complete testing actuation history to be formed.
The system software may use a relational database management system as its core records management entity subsystem. With the system software, users can manage the machine- and human-readable data associated with spray drug product testing (e.g. acquire, process, analyze, etc.) thus allowing companies to:
The system computer 114 and corresponding system software can properly handling the following example failure events (along with any other critical system events):
The balance 255 or weighing device, is disposed on a granite table 355 which, in turn, is disposed on the metal frame 110 (
The system 100 may automatically measure the delivered and/or metered dose/shot weight, depending on user programmable inputs, with high resolution (e.g., 0.1 mg). The delivered shot/dose weight is computed as the weight difference of the appropriate collection apparatus (e.g. waste collector or collection vessels) before and after actuation. The metered dose weight is computed as the weight difference of the spray device before and after actuation. Table 4 indicates test measurements, measurement modes and elements for weighing.
The analytical balance 255 may be automatically calibrated using two internal calibration weights (providing a minimum of a three point calibration). The automatic calibration procedure may be programmable (e.g. daily, weekly, monthly) and be executed as part of normal operation of the system 100.
The nozzle tip dabber 218 may be stored in a nozzle tip dabber holder 318 that includes a sensor 319 to sense the presence of the nozzle tip dabber 218. As described further below, the nozzle tip dabber 218 is used to maintain a nozzle tip of a spray device free of residue which may affect performance of the spray bottle 311.
Many nasal spray drug products are formulated as thixotropic suspensions and are delivered via a mechanical pump device. A thixotropic material exhibits a decrease in viscosity with increasing applied shear stress or shear rate, followed by a time dependent recovery when the shear load is removed. (Tomato ketchup is a good example of a thixotropic material because it does not start to flow from its container unless sufficient shear is imposed on it, e.g., short stroke, jerk-action shaking motion). The thixotropic nature of nasal spray drug formulations can seriously affect the performance of the emitted sprays. Therefore, to prepare the spray devices properly for actuation and subsequent shot weight measurements, embodiments of the system computer 114 may be programmed to perform various shaking routines with the robotic handler 120.
For certain spray devices, the shaking routine must not tilt the spray devices from a vertical axis, nor impart any foaming into the formulation. Based on a survey of current laboratory practices, and the characteristics of many nasal spray formulations, the shaking routine may operate in either of two modes:
Both modes may be programmed with various parameters for the shaking routine including amplitude, frequency, and duration. In various embodiments, the shaking routine may be executed prior to or during a test run.
After moving the device to the actuator 408, the system computer gives commands to the robotic handler to secure the device in the actuator 410 and to release the device 412. In steps 414-418, the system computer issues commands to the robotic handler to move the electromechanical gripper to the desired collection vessel at a collection vessel holding area 414, to pick up the collection vessel 416, and to move the collection vessel 418 to a balance pan. After the robotic handler releases the collection vessel 420, the balance weighs the collection vessel 422.
After weighing the collection vessel 422, the robotic handler picks up the collection vessel 424 and moves it to the actuator 426. The robotic handler positions the collection vessel over the nozzle tip of the device in the actuator and maintains the collection vessel at that position 428. In step 430, the system computer issues commands to the actuator to actuate the device for a number of repetitions corresponding to one dose (e.g., in many cases, two actuations are needed per dose, corresponding to one actuation per nostril). Before ending 437, the robotic handler moves the collection vessel to the balance pan 432 and releases the collection vessel 434 so that it can be weighed by the balance 436.
In a process similar to process 400, a waste collector may be used instead of the collection vessel to perform pump delivery testing on a spray device using the delivered weighing mode (Table 4). Table 5 highlights basic example operations and functions of the robotic handler 120. It should be understood that the terms “pick up” and “release” are meant to convey the spirit of what the robotic handler may do, not how the robotic handler actually does it.
In a process similar to process 400,
Embodiments of the system 100 may provide various actuation measurement modes. In a priming mode, the system computer 114 may record in the database 115 the force and position versus time profiles acquired from the actuator. In a delivered or metered shot weight with individual tare mode, the system 100 weighs each shot based on taring the balance 255 after each actuation. In both modes, waste collectors 213 collect each shot according to a given test method. For delivered shot weight measurements the waste collector is weighed and for metered shot weight measurements the device is weighed. In a dose content uniformity mode, the system 100 collects individual doses in an appropriately sized collection vessel (e.g., a standard laboratory collection tube), and records the shot weight by measuring the weight differential of the collection vessel before and after actuation (i.e., the delivered shot weight method).
The system 100 described above is able to match at least the performance of a trained laboratory technician on a per-shot basis. Based on test runs of the system 100, including nozzle tip dabbing and balance taring after each actuation, a trained laboratory technician can collect, on average, approximately 50 delivered shot weight measurements in 25 minutes (or one measurement every 30 seconds on average) using a Mettler-Toledo AX-204 4-place analytical balance. Embodiments of the system 100 are designed not to allow any misreads of shot weight to occur due to machine malfunctions under normal operating conditions.
The force coupler 251 is coupled to the linear screw-rail assembly 535 such that the motor 531 may drive the screw rail spindle 532, which, in turn, drives the force coupler 251 to actuate the spray bottle 311. The actuator electronics 540 communicate with the system computer 114 and system controller 117 via connectors 551-555. For example, the actuator electronics 540 may receive commands from the system computer 114 to apply a specified force to actuate the spray bottle 311. The actuator 250 also includes a power connector 557 through which to provide power to the actuator electronics 540, the motor 531, and other components of the actuator 250.
The actuator 250 includes a receiver 511 for receiving and securing the collet 312 to the actuator 250. The actuator 250 may further include a sensor 515 for sensing the presence of the collet 312. The sensor may include a photoelectric sensor, magnetic sensor, or any other sensor for detecting the presence of the collet 312. The sensor 515 may communicate sensor signals through the sensor electronics 540 to the system computer 114. The receiver 511 may be spring-loaded to firmly secure the spray device in place during actuation.
As illustrated in
The electromechanical gripper 220 includes a motor and linear screw-rail assembly 721 that drives first and second gripper elements 224a-b to grasp or release a spray device, collection vessel, or a waste collector. Each gripper element 224a-b includes movable jaws 226a-b, stationary jaws 228a-b, springs 729a-b, magnetic sensors 724a-b, and magnets 726a-b. The upper jaws 226a-b and springs 729a-b together are movably coupled to the gripper elements 224a-b to provide vertical compliance when, for example, a nozzle tip dabber held by the upper jaws 226a-b is being used to dab a spray device having a nozzle tip with an unknown height.
The magnetic sensors 724a-b are fixably mounted in respective gripper elements 224a-b and magnets 726a-b are mounted in respective upper jaws 226a-b. When the sensors 724a-b and corresponding magnets 726a-b misalign, the magnetic sensors 724a-b indicate to the system computer (e.g., system computer 114) that the object, such as a waste collector, has made contact with another object, such as a spray device. The magnetic sensors 724a-b may also provide information about the position of the upper jaws 226a-b relative to the gripper elements 224a-b.
As illustrated in
The upper jaws 226a-b include V-grooves to allow the electromechanical gripper to firmly and securely grasp various objects. In this embodiment, the jaws 226a-b and 228a-b are designed with a V-shape with a predefined angle to further improve the electromechanical gripper's ability to grasp spray devices and other objects.
An advantage of the nozzle tip dabber 318 is that it does not affect the performance of the spray device or otherwise alter the shot weight of the sprays. The nozzle tip dabber may be used to clean the nozzle tip area after each shot in the worst case scenario or after an actuation group in a minimal scenario.
After process 1100 starts 1101, the actuator actuates a spray device into a collector 1102, such as a collection vessel. In steps 1104-1110 the robotic handler moves to the actuator 1104, picks up the collector 1106, places the collector at the appropriate location in the device holder tray 1108, and releases the collector 1110. In step 1112, the robotic handler moves to the tip dabber holding area and, in step 1114, picks up the tip dabber with the movable jaws of the robotic handler.
In step 1116, the robotic handler moves the tip dabber to the actuator and positions the tip dabber such that an area of the tip dabber (e.g., an unused area) is aligned with the nozzle tip of the spray device. In step 1118, the robotic handler moves the tip dabber towards the nozzle tip of the spray device. If the robotic handler detects movement of the movable jaws 1120-1121, indicating that the tip dabber has made contact with the nozzle tip of the spray device, the robotic handler moves the tip dabber away from the nozzle tip of the device 1122. Alternatively, the robotic handler may not move the tip dabber away from the nozzle tip of the device, but stop the robotic handler so that the tip dabber maintains contact with the nozzle tip of the spray device for a given period of time. Before ending 1127, the robotic handler moves the tip dabber to the tip dabber holding area 1124 and releases the tip dabber 1126.
According to one estimate, the maximum fluid capacity of currently marketed nasal spray products is approximately 40 mL. Therefore, this embodiment of the system is capable of handling a maximum of 400 mL (40 mL×10 spray devices 1211) of fluid in one run in a dose content uniformity test and 800 mL (40 mL×20 spray devices 1211) in a pump delivery test. The system is also capable of handling identical spray devices from the same or different batches.
If the holder tray sensors detect the absence or presence of elements at the holder tray 1312, the system computer indicates to the user the absence or presence of elements at all locations of the holder tray 1314. Until the testing is complete 1316, 1317, the system computer continues to monitor for the absence and presence of elements at the holder tray 1310. In step 1319, the process 1300 ends.
The collection vessel base 1413b includes a first inner wall 1411 to provide a first collection area 1412 and a second inner wall 1415 to provide a second collection area 1416 between the first and second inner walls 1411, 1415. The second inner wall 1415 and the second collection area 1416 may compose a “plug” that is inserted into the collection vessel base 1413b. The first collection area 1412 collects the majority of formulations that are ejected from the spray bottle 311. The remaining formulation that is not collected in the second collection area 1412 is collected in the second collection area 1416.
The spray pump collection assemblies 1400a-b include collets 312. The collet 312 has an aperture with threads 1411. The aperture of the collet 312 is tapered such that the threads 1411 of the aperture grip the spray bottle's nozzle tip as the collet 312 is screwed onto the spray bottle's nozzle tip. The collet 312 does not effect the spraying function of the spray bottle 311 because the threads 1411 of the aperture grasp the base of the spray bottle nozzle tip. The collet 312 facilitates easy handling of the spray bottle 311, for example, by a robotic handler or an actuator.
After performing optical measurements 1714, the actuator, along with the spray device, return to the DCU and pump delivery testing region 1716. Before ending 1725, the robotic handler moves to the actuator 1718, picks up the spray device 1720, moves the spray device to the holder tray 1722, and releases the spray device 1724.
While this invention has been particularly shown and described with references to example embodiments thereof, it will be understood by those skilled in the art that various changes in form and details may be made therein without departing from the scope of the invention encompassed by the appended claims.
It should be understood that any of the above-described flow diagrams of
This application claims the benefit of U.S. Provisional Application No. 60/858,257, filed on Nov. 10, 2006. The entire teachings of the above application are incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
60858257 | Nov 2006 | US |