Automatic pipetting units are used in particular in applications in laboratories where liquid substances in minute quantities are to be dispensed into individual vessels of a set of vessels, or when such quantities are to be removed and dispensed into individual vessels of another set of vessels. For a high efficiency of laboratory processes, it is desirable—and for uniform test conditions in all individual vessels it is indispensable—that all individual vessels be filled with a single dispensing step. For this purpose, the pipetting head of the automatic pipetting unit is designed so that it has the same number of pipetting tips as there are individual vessels, and the pipetting tips are arranged at the same grid spacing with respect to each other as are the midpoints of the individual vessels.
Some typical vessel arrays can be, for example, microtiter plates. The individual vessels are formed by wells, which are arranged in an 8×12, 16×24 or 32×48 grid, for instance. The outer dimensions of the microtiter plates are standardized, that is, they are the same, regardless of the number of wells. Therefore, as the number of the wells increases and the grid spacing becomes smaller, the wells must be designed with increasingly smaller interior dimensions. The necessity for a precise alignment of the well centers and of the axes of the pipette tips to each other increases, so that an exact liquid dispensing into the individual wells will occur, or so that the pipette tips will enter the wells at all in a 32×48 grid. This alignment will occur when both the vessel array and also the pipetting head are aligned to a common reference basis in the automatic pipetting unit.
In the case of automatic pipetting units which are delivered from the manufacturer with a non-replaceable pipetting head, the alignment of the pipetting head in the automatic pipetting unit is effected by a one-time adjustment during assembly, and then locking in place.
In order to be able to adapt an automatic pipetting unit in a flexible manner to different vessel arrays, it is sold with several pipetting heads compatible with automatic pipetting units, and these heads must be replaced by the user himself. Special operating steps must be taken so that the different pipetting heads can be easily replaced with little effort, and the new pipetting head is aligned automatically without the need for a renewed adjustment to the reference basis.
Equipment with this kind of functionality is already known in the state of the art.
For example, the Zeiss-Sumal model AD96 produced in the year 1986 had a somewhat cumbersome method for replacement of pipetting heads. The user had to employ a special tool to do so.
Starting in 1993 the German firm of Jenoptron GmbH produced and marketed its series 96 simultaneous pipetting unit IGEL with electromotor-based and microcontroller monitored pipette head replacement unit.
In the recent document WO 02/096562 an automatic pipetting unit with a replaceable pipetting head is disclosed. The replacement of this pipetting head is described by reference to the description of the installation and locking processes with the aid of
In order to install the pipetting head 600 and to lock it into the automatic pipetting unit 500, the coupling plate 514 (
The disadvantages of this kind of design are—
The present invention is based on the problem of finding a constructive design for an automatic pipetting unit allowing the pipetting head to be easily and quickly replaced.
This problem is solved for an automatic pipetting unit according to the present invention.
The invention will be described in greater detail below based on the illustrations with reference to one exemplary embodiment. In the figures:
a is a perspective view illustrating an automatic pipetting unit ready for use;
b is a perspective view illustrating the automatic pipetting unit according to
a is a perspective view illustrating the slider crank mechanism of the automatic pipetting unit according to
b is a perspective view illustrating the slider crank mechanism of the automatic pipetting unit according to
c is a perspective view illustrating the slider crank mechanism of the automatic pipetting unit according to
a is a perspective view illustrating the automatic pipetting unit according to
b is a perspective view illustrating the automatic pipetting unit according to
a illustrates an example of automatic pipetting unit 1 according to the present invention as it appears in an operational state. Head cover 2 is located in the upper housing area. A pipetting head 3 (
b shows the head cover 2 open and the pipetting head 3 is shown as already half removed from the housing of the automatic pipetting unit 1. In order to remove the pipetting head 3, the tip magazine 4 which is locked to the pipetting head 3, is replaced by a head changing tool 22. This tool is used to protect the sealing surface during placement of the pipetting head 3 outside of the automatic pipetting unit 1, as a handle for lowering during removal and installation, and as a support surface for setting down the removed pipetting head 3.
When the head cover 2 is either fully closed or fully opened, the pipetting head 3 is located at a respective end position of its travel path. These end positions are, with the head cover 2 closed the working position (lower end position), and with the head cover 2 open the replacement position (upper end position).
In the working position, the pipetting head 3 is located in a horizontally defined position with respect to the hoisting table 21 located underneath it and integrated into the automatic pipetting unit 1. The hoisting table 21 is designed so that a vessel array positioned thereon, e.g., a microtiter plate, is aligned with respect to the pipette tips 5 so that their axes are incident on the centers of wells.
In the replacement position, the pipetting head 3 is elevated in a vertical direction with respect to the working position. During the elevation, the pipetting head 3 is automatically released from its mechanical adjustment with respect to the automatic pipetting unit 1, and its electrical connection to the automatic pipetting unit 1 is broken. It can then be removed from the automatic pipetting unit 1 at a position higher than the working position, so that the free space required for replacement of the pipetting head 3 can be reduced compared to prior known systems. That is, in particular, any apparatus located in front of the automatic pipetting unit 1—such as an automatic tip replacement unit for replacement of a tip magazine 4 with used pipette tips 5 by a tip magazine 4 with unused pipette tips 5—can remain in place during replacement of the pipetting head 3.
The automatic pipetting unit 1 has a hoisting device for hoisting and lowering of the pipetting head 3 between the replacement setting and the working position. The hoisting device is composed of two double-sided, frame-mounted, slider crank mechanisms which are connected together by means of a head cover support 12 to which the head cover 2 is attached.
As is shown in
Referring to
The crank mechanisms 6 are each connected to one end of a thrust couple 14 by means of a rotary joint 13 located at the perimeter of the crank mechanism 6. The other end of the thrust couple is connected to a common articulated thrust joint 15. The articulated thrust joint 15 is moved vertically by means of two lateral guide rails 16 in and between two guide elements 20 (see
Referring to
The mode of operation of the hoisting device will be explained in greater detail below based on the operational use of a pipetting head 3 in the automatic pipetting unit 1.
As illustrated in
In order also to be able to introduce two lugs 19 formed on the side of the pipetting head 3 and provided for self-adjustment of the pipetting head 3, unhindered into the interior of the automatic pipetting unit 1, an appropriate open space must be provided beneath the guide element 20.
In order to hold the hoisting device or the hoisting device with inserted pipetting head 3 in the replacement position against the force of gravity, stationary permanent magnets are provided on which, in this end position, the hoisting device is held after attraction of the head cover support 12.
After complete insertion of the pipetting head 3, the head cover 2 is closed by manually cancelling the attraction force of the permanent magnet—already compensated in part by the weight of the pipetting head 3—by a short downward movement of the head cover 2.
With closure of the head cover 2, which is connected rigidly to the two internal gears 10 via the head cover support 12, the movement of the head cover 2 is transferred directly to the two internal gears 10. The inner tooth segments 9 of the internal gears 10 are rolled onto the outer toothing 7 of the crank mechanisms 6 facing the cover opening, and cause them to begin a rotating motion. At the same time, the other crank mechanisms 6 are caused to rotate in the opposite direction and the articulated thrust joints 15 connected to the crank mechanisms 6 are lowered by means of the thrust couples 14. This lowering takes place automatically due to the inherent weight of the pipetting head 3 and is decelerated by means of dampers located on both sides between the articulated thrust joints 15 and the frame, so that the pipetting head 3 will be lowered gently into its working position (lower end position).
In this lower end position the pipetting head 3 automatically assumes a defined position in the automatic pipetting unit 1 and thus also with respect to the hoisting table 21 in which the pipetting head 3 is held in place due to its own weight—by means of ball elements located on the lugs 19 and a third ball element located on the pipetting head 3, with a three-point contact defined by a conical recess, a U-shaped channel, and a plane, without having to be adjusted and locked. At the same time, the electrical connection between pipetting head 3 and automatic pipetting unit 1 is established automatically.
After the insertion of the pipetting head 3 into the automatic pipetting unit 1, the head cover 2 must be closed—this step is needed anyway—in order to bring the pipetting head 3 into its working position and to keep it there.
Accordingly, for the removal process, the head cover 2 has to be opened completely and the pipetting head 3 can be removed without any other attachment elements having to be loosened.
The ordinary technician skilled in the field of this invention will see that the invention is not limited to the details of the sample embodiments presented above, but rather that the present invention can be embodied in other special formats, without departing from the scope of the invention, which is otherwise specified by the attached claims.
1 Automated pipetting unit
2 Head cover
3 Pipetting head
4 Tip magazine
5 Pipetting tips
6 Crank mechanism
7 Outer toothing
8 Pivot
9 Tooth segment
10 Internal gear
11 Front surface
12 Head cover support
12.1 Support side
12.2 Support bar
13 Rotary joint
14 Thrust couple
15 Articulated thrust joint
16 Guide rail
17 Guide channel
18 Head rail
19 Lug
20 Guide element
21 Hoisting table
22 Head changing tool
Number | Date | Country | Kind |
---|---|---|---|
20 2007 000 904 U | Jan 2007 | DE | national |
Number | Name | Date | Kind |
---|---|---|---|
6325114 | Bevirt et al. | Dec 2001 | B1 |
6846680 | Friswell et al. | Jan 2005 | B2 |
6982063 | Hamel et al. | Jan 2006 | B2 |
7541001 | Kraemer et al. | Jun 2009 | B2 |
20010039843 | Schoeppe | Nov 2001 | A1 |
Number | Date | Country |
---|---|---|
02096562 | Dec 2002 | WO |
2006123690 | Nov 2006 | WO |
Number | Date | Country | |
---|---|---|---|
20080170965 A1 | Jul 2008 | US |