The improvements generally relate to the field of biosensors, and more specifically to optical biosensors.
Point-of-need testing has tremendous value in diagnostics (medical and non-medical) and more so in circumstances when a point-of-need device is low cost, portable and provides a ‘sample-in-answer-out’ solution. However, diagnostic technologies require quantitative readings which provide sufficient information to extra diagnostic level information. This necessitates specialized instrumentation which are unsatisfactory for miniaturization or are too costly for use in point-of-need applications. There always remains room for improvement.
Microfluidic technologies have been shown to provide certain advantages by integrating microchannels, chambers, pumps, and other components to manipulate small volume samples. These types of fluidic systems minimize consumption of reagents and provide a platform for permitting the integration of many analytical procedures into a single device. While microfluidic systems provide a way of compacting the fluidic pathways, the interrogation and measurement of the analyte of interest require for complex and costly external instrumentation, which can often become even more complex when retrofitted to a smaller fluidic apparatus.
Optical sensors are known to be versatile methods of interrogating and extracting key information from an analyte. There exists various ways of interacting with an analyte or a by-product which itself interacts with the analyte. This further depends on the application, the analyte itself, and the information desired. Some products release a definite amount of energy in the form of photons when they interact with analytes which permits the extraction of relevant information from the analyte. For instance, chemiluminescence (CL) reactions are complex multi-step reactions in aqueous conditions which emit light. Briefly, the reaction of a CL agent, such as luminol for instance, paired with an oxidant, like hydrogen peroxide (H2O2), generates an intermediate that exhibits CL emission. During this oxidation reaction, the luminol (e.g., 3-aminophthalhydrazide or 5-amino-2,3-dihydro-1,4-phthalazinedione) is oxidized into a dicarboxylate ion and loses molecular nitrogen while gaining oxygen atoms, producing 3-aminophthalate. The resulting 3-aminophthalate is in an excited state because the electrons in the oxygen atoms are elevated to higher orbitals. The electrons quickly return to a lower energy level, releasing a fraction of the excess energy as photons. This reaction occurs best with a catalyst, which may be an electric potential difference for instance, which oxidizes luminol and produces the excited state. The use of an electric potential difference as a catalyst for these types of reactions are known as electrochemiluminescence (ECL) and the oxidation luminol can be used to design biosensors exhibiting high sensitivity and a wide linear range of operation.
Providing the electrical potential (electromagnetic field) and capturing the optical information is no easy feat and more so in the context of a miniaturized fluidic system where a limited quantity of analyte and of CL agent is available. Recording and analyzing the ECL emission requires specialized photodetectors. Signal collection can be achieved using single-point photodetectors (such as photomultiplier tubes or avalanche photodiodes) due to their high gain, sensitivity, and linearity. However, these types of detectors cannot be used in portable or miniaturized devices due to their size, high-voltage supply, and power consumption, and cannot be easily integrated into multiplexed detection designs.
It was found that providing a biosensor having a photodiode array, such as a CMOS sensor for instance, with a substrate coated with a conductor forming an electrode which is capable of being integrated with a microfluidic system permits to overcome at least some of the issues in interrogating analytes and capturing optical signal related therefrom. The biosensor can provide a miniaturized, low cost, low-power system which can permit the interrogation of the analyte while providing an increased photon collection due to the close proximity of ECL reactions to the detector(s).
It is understood that the use of an electric potential difference for the purposes of oxidation of luminol, as described herein is provided as an example only and should not be construed as limitative in any way. Other uses of electrical potential (methods of providing an electromagnetic field) may be provided, with or without other analytes, without departing from the present disclosure.
It is understood that the term “interrogating” refers to any action, process, potential, interaction which can lead to the emission of an optical signal from a sample. The use of an electrical potential (e.g., electromagnetic field) for interrogation of a ECL solution is provided as an example only and should not be construed as limitative in any way.
In accordance with a first aspect of the present disclosure, there is provided a biosensor apparatus comprising: a photodiode array having a first photon receiving face, a transparent substrate covering the first photon receiving face of the photodiode array, the transparent substrate having a second photon receiving face opposite the photodiode array, an electrically conductive coating covering the second photon receiving face of the transparent substrate, the electrically conductive coating being transparent and having two electrical contacts spaced apart from one another, and an analyte receiving area extending between the two electrical contacts of the electrically conductive coating.
Further in accordance with the first aspect of the present disclosure, the biosensor apparatus can for example further comprise a body of material deposited onto the transparent substrate, the body having a recessed channel forming a fluidic channel when the recessed channel is received onto the analyte receiving area. In some embodiments, the body is made of polydimethylsiloxane (PDMS). In some embodiments, the fluidic channel has a height of approximately 143 μm, a width of approximately 650 μm, and a length of about 3.5 mm. In some embodiments, the fluidic channel is a microfluidic channel. In some embodiments, the fluidic channel can receive between about 1 and 10 μL of fluid, and most preferably between about 2 and 5 μL of fluid.
Still further in accordance with the first aspect of the present disclosure, the body can for example have an inlet and an outlet fluidly connected at opposite ends of the fluidic channel.
Still further in accordance with the first aspect of the present disclosure, the photodiode array, the transparent substrate and the electrically conductive coating can for example form a first detector assembly, the biosensor apparatus can for example comprise a second detector assembly identical to the first detector assembly, the first detector assembly and the second detector assembly both facing the analyte receiving area to receive photons therefrom.
Still further in accordance with the first aspect of the present disclosure, the first detector assembly and the second detector assembly can for example form a gap therebetween, the gap having a dimension ranging between 50 μm and 200 μm.
Still further in accordance with the first aspect of the present disclosure, wherein the photodiode array can for example be a complementary metal-oxide-semiconductor (CMOS) sensor. In some embodiments, the CMOS sensor can have an active area of at least 2 mm by 2 mm. In some embodiments, the CMOS sensor can have an active area of 3.67 mm by 2.73 mm.
Still further in accordance with the first aspect of the present disclosure, the electrically conductive coating can for example have a thickness between 3 nm and 500 nm, and most preferably between 10 nm and 300 nm.
Still further in accordance with the first aspect of the present disclosure, the electrically conductive coating can for example be an Indium Tin Oxide (ITO) coating.
Still further in accordance with the first aspect of the present disclosure, the transparent substrate can for example have a thickness between 50 nm and 200 μm, and most preferably between 100 μm and 150 μm. In some embodiments, the transparent substrate can have a thickness of approximately 127 μm.
Still further in accordance with the first aspect of the present disclosure, the transparent substrate can for example be a Poly Ethylene Terephthalate (PET) film.
Still further in accordance with the first aspect of the present disclosure, the PET film can for example form a 3-dimensional pattern having a plurality of protrusions protruding from the photodiode array. In some embodiments, the protrusions form a pyramidal shape, a triangular shape, a square shape, an hexagonal shape, a cuboid shape, a dome shape, or a combination thereof.
Still further in accordance with the first aspect of the present disclosure, the biosensor apparatus can for example further comprise an external voltage source applying an electrical voltage across the electrically conductive coating via the two electrical contacts.
Still further in accordance with the first aspect of the present disclosure, the biosensor apparatus can for example further comprise a pair of conductive connectors electrically connected between the two electrical contacts and the external voltage source.
Still further in accordance with the first aspect of the present disclosure, the photodiode array can for example be a 2-dimensional photodiode array.
Still further in accordance with the first aspect of the present disclosure, the transparent substrate can for example be an electrical insulator.
In accordance with a second aspect of the present disclosure, there is provided a method of optically detecting an analyte using a biosensor apparatus, the biosensor apparatus having a photodiode array, a transparent substrate atop the photodiode array, an electrically conductive coating atop the substrate, the electrically conductive coating being transparent and having two electrical contacts spaced apart from one another, and an analyte receiving area atop the electrically conductive coating between the two electrical contacts, the method comprising: the analyte receiving area receiving an analyte containing sample; generating an electromagnetic field across the analyte receiving area via the two electrical contacts, said electromagnetic field stimulating electrochemiluminescence (ECL) light emission in the analyte containing sample; and the photodiode array generating an electrical signal based on a detection of said ECL light emission.
Further in accordance with the second aspect of the present disclosure, the analyte containing sample can for example be one of liquid and gaseous.
Still further in accordance with the second aspect of the present disclosure, said generating the electromagnetic field can for example include applying an electrical voltage along the electrically conductive coating between the two electrical points.
Still further in accordance with the second aspect of the present disclosure, the electrical voltage can for example be of at least 5 V.
Still further in accordance with the second aspect of the present disclosure, said receiving the analyte containing sample can for example include flowing the analyte containing sample within a fluidic channel extending across the analyte receiving area.
In accordance with a third aspect of the present disclosure, there is provided a biosensor apparatus comprising: a photodiode array, a substrate atop the photodiode array, an electrically conductive coating atop the substrate, and an analyte receiving area atop the electrically conductive coating, the photodiode array having a field of view encompassing the analyte receiving area and extending across the substrate and the electrically conductive coating.
Still further in accordance with the third aspect of the present disclosure, when an electrical voltage is applied across the electrically conductive coating, analytes present at the analyte receiving area can for example emit electrochemiluminescence (ECL) light emission across the electrically conductive coating, across the substrate and towards the photodiode array.
In accordance with a fourth aspect of the present disclosure, there is provided a method of optically detecting an analyte, the method comprising: positioning an analyte in a gaseous or liquid solution, in an analyte receiving area; generating an electromagnetic field in the analyte receiving area, said electromagnetic field stimulating light emission by electrochemiluminescence (ECL); and generating an electrical signal based on the detection of said light emission.
In accordance with a fifth aspect of the present disclosure, there is provided a biosensor for multiplexed optical detection of biological and chemical analytes with high sensitivity and specificity, the biosensor comprising of: a substrate for sample holding; a microfluidic system for sample handling; and an optical detector.
In accordance with a sixth aspect of the present disclosure, there is provided a method of manufacturing the biosensor apparatus, the method comprising of: manufacturing a 3D pattern onto a 2D array of photodiodes, the 3D pattern having an electrically conductive coating deposited thereon; and implementing a microfluidic system atop the electrically conductive coating.
Many further features and combinations thereof concerning the present improvements will appear to those skilled in the art following a reading of the instant disclosure.
In the figures,
FIGS. 12A1-12E show comparisons of the signal collection provided by biosensor apparatus in comparison to a commercial microscope and a smartphone, in accordance with one or more embodiments.
The biosensor apparatus 100 in the embodiment shown in
In this example, and as will be discussed below, the photodiode array 102 is a 5-megapixel backside-illuminated complementary metal-oxide-semiconductor (CMOS) sensor 116 from a commercially available Raspberry Pi™ module (OV5647, OmniVision technologies). The field of view (FOV) of the CMOS sensor 116 is 3.67 mm by 2.73 mm, which is the active area of the CMOS sensor 116, and the spatial resolution of the system is 1.4 μm, limited by the pixel size having a field of view (FOV) of 3.67 mm by 2.73 mm. However, it is understood that other CMOS sensors or chips from other suppliers and of any desirable size with different spatial resolutions or field of views may be used without departing from the present application.
The substrate 104 is a transparent Poly Ethylene Terephthalate (PET) film having been coated with Indium Tin Oxide (ITO). In other words, the electrically conductive coating of the transparent substrate 104 is made of ITO. As is perhaps best seen in
In order to choose the optimal thickness of the ITO coating 106, the resistance and optical transparency of the coating 106 are to be considered. The difference in the coating resistance leads to a different current intensity that affects the gradient (see
As will be further discussed below, other materials may be contemplated without departing from the present application, and the ranges provided for ITO can be varied for other materials and coatings without departing from the present application. These ranges can be varied to provide a balance between the resistance of the coating, heat generation, possible failure of the material due to its burning, while not unduly inhibiting its optical transparency.
Still referring to
The body 122 is made of polydimethylsiloxane (PDMS) and sealed onto the ITO-coated PET substrate 104 via an oxygen plasma treatment. The PDMS body 122 further has openings 126 forming inlets and outlets at opposite extremities of each one of the microfluidic channels 108 permitting the reception of corresponding fluidic tubings 128 and the flowing of the sample 130 with the analyte 132 therein. The sealed microfluidic channels 108 form electrochemical cells whereas the conductive ITO-coated PET substrate 104 acts as a single electrode 134.
Still referring to
The voltage applied can vary depending on the resistance of the ITO coating. In this particular example, with a coating providing a generally uniform resistance of approximately 60 ohm/square, it was determined that voltages varying from 2.5 to 5 V may be applied. A voltage of 2.5 V was sufficient to drive the desired ECL reaction, and thus the faradaic reaction occurs simultaneously at both ends of the fluidic channels 108. Since this reaction is driven by electric potential difference, increasing the voltage results in increasing the emission until 5 V. The blue ECL emission of luminol was observed at the positive side of the fluidic channel 108 starting at around 2.5 V. By increasing the voltage difference, the emission region of the channel gradually expands toward the center of the fluidic channel, allowing the ECL reaction to occur over a greater surface area. It is understood that different voltage values may be used with different concentration ranges for the fluidic solution without departing from the present application. The voltages used may be higher than 5 V in certain applications, such as to shift the concentration range to the lower limits of detection. However, higher voltages can result in higher electrical currents and material temperature, which alter the enzyme activity and can damage the ITO-coated PET substrate 104.
As is perhaps best seen in
Referring to
Attention is now brought to
Attention is now brought to
The second photodiode array 202 of the second detector assembly of the biosensor apparatus 200 is covered by a substrate 204 which, in this example, is the same PET substrate 104 provided in the first detector assembly. However, the substrate 204 is not provided with an electrically conductive coating and is not connected to an external voltage V, as is the case with the electrically conductive coating 106. It is understood that in alternate embodiments, the electrically conductive coating 106 may be provided in the substrate found in the second detector assembly instead of the first detector assembly of the biosensor apparatus 200 or on both the first and second detector assemblies without departing from the present application.
Still referring to
In this embodiment, the photodiode arrays 102 and 202 are identical CMOS sensors 116 and 216 and the substrate are the same PET material. It is understood, however, that in alternate embodiments, the CMOS sensors 116 and 216 may be different sizes, manufacturers and have different spatial resolutions without departing from the present application. Similarly the uncoated substrate 204 may differ in material, thickness, size and transmittance with respect to the substrate 104, as suitable for the application without departing from the present application.
Attention is brought to
Still referring to
Attention is now brought to
It is understood that in alternate embodiments, a second photodiode array may be placed on the opposite side of the fluidic channel 408, as was discussed in
Attention is brought to
For the purposes of completeness, an experimental methods and results for example analytes will be discussed.
The performance of the single electrode system discussed with reference to
The electrolyte used for ECL measurements was a mixture of 0.1 M sodium bicarbonate and 1 mM luminol. Specifically, to prepare the electrolyte, 0.1 M sodium bicarbonate was prepared, and the pH was adjusted accordingly by adding 1 M NaOH. It was then mixed with 10% v/v 10 mM luminol that was prepared in NaOH (0.1 M) to obtain a final concentration of 1 mM for luminol. Further, the analytes (H2O2 or UA) were added to this electrolyte for detection. The pH used for measurements was 10.7 unless otherwise stated.
For H2O2 detection, 10 mM H2O2 was prepared by diluting the concentrated H2O2 solution (9.8 M) in water and then mixed with the prepared electrolyte containing carbonate buffer and luminol. To create the calibration curve for H2O2 detection, different concentrations of H2O2 were prepared in the electrolyte.
Similarly, for UA detection, 1 mM UA was prepared in the electrolyte containing carbonate buffer and luminol for further dilutions. Further, different concentrations of UA from 25 μM to 1000 μM (25, 50, 100, 200, 300, 500, 800, and 1000 μM) were prepared by diluting the stock solution of 1 mM UA in the same electrolyte with carbonate buffer and luminol. The uricase solution of 1 mg/ml was prepared using NaOH (0.001 M, pH 11) and kept at −20° C. for further use. To evaluate the selectivity for UA detection, interfering molecules, i.e., ascorbic acid (5 μM), creatinine (10 μM), and glucose (50 μM) were added to the prepared UA solution for ECL measurement. 1% v/v of triton X-100 was added to the final solutions before each experiment to improve mixing of components.
To prepare the solutions for simulated sample analysis, artificial body fluids (saliva or Surine™) were mixed with the electrolyte containing carbonate buffer (0.1 M) and luminol (1 mM) at two different ratios (1:9 and 1:1). Further, 50 μM or 300 μM of UA was prepared in the mixture of the electrolyte and artificial body fluids using 1 mM and 6 mM stock solutions of UA, respectively, and the ECL intensities were captured. By using the calibration equation of UA detection, the concentration of UA corresponding to the obtained intensity (average of 3 experiments) was calculated, and the recovery rate was reported by comparing the measured concentration with the known (added) concentration.
In this specific example, the PDMS body of the microfluidic device was built by soft lithography. Briefly, the SU-8 photoresist on silicon wafer was exposed to 365-nm UV light using a mask aligner (the OAI Model 200) to create the positive master mold. Then, PDMS prepared in a 10:1 (monomer: crosslinker) ratio was used to replicate the pattern. This device was then sealed with the ITO-coated PET using oxygen plasma treatment (Diener Electronic, PICO) at 60 W for 2 minutes, followed by 30 minutes of incubation at 70° C. for irreversible bonding.
The ITO-coated PET substrate comprises of two distinct layers: a PET substrate with a thickness of 127 μm and an ITO coating with a thickness ranging from 10 to 300 nm. The ITO coating was selected to provide a resistance of 60 ohm/square. The resistivity of the ITO coated PET is shown in
where Rs is the sheet resistance, k is the geometrical correction factor which is 4.532 for the aforementioned experimental setup, ΔV is the measured voltage, and I is the applied current. As a result (
For UA detection, a final concentration of 200 μg/ml of uricase was added to the sample solution and incubated for 10 minutes before applying voltage. For the ECL experiments, the sample was transferred to the channels through the inlets using a syringe and a 30 G needle. The sample filled the fluidic channels as well as both inlets and outlets.
The measurements were performed in complete darkroom. A voltage of 5 V was applied to the device and the ECL emission was observed and recorded using a Raspberry Pi™ 4 computer and then analyzed via Fiji ImageJ to extract grey values corresponding to the detected photons. The measured grey values of the control channel were subtracted from those recorded for the test channel and the results were presented. Other computer or computing devices can be used in other embodiments.
In order to study the reusability (repeatability, using the same device), and reproducibility (using different devices), and selectivity (in the presence of interfering molecules), 50 and 300 μM of UA were used for ECL measurements.
For controlling the imaging parameters, a Python script was written, and the following conditions were applied: resolution was set to 2592×1944 pixels (full field of view), framerate was set to ⅙ s, shutter speed was set to a 6 s exposure time, ISO was set to 800 (gain 8×). Thirty seconds of sleep time was applied before each experiment to give the camera an appropriate time to adjust the parameters.
During the experiment, the channels are filled with the reagents at maximum capacity, i.e., 3.3 μl of mixed luminol, H2O2, and carbonate buffer. The 3.3 μl of electrolyte inside each fluidic channel contributes to the ECL emission. This amount of electrolyte corresponds to 3.3 nmol of luminol that emits light in the oxidation reaction. The height profile of the microfluidic channel is shown in
H2O2 was used to measure the ECL intensity at different voltages ranging from 2.5 to 5 V. As shown in
where ΔEch is the potential difference inside the electrochemical cell, Etot is the voltage applied by a power supply, Ich is the length of the channel and d is the distance between wires, as illustrated in
ECL intensities were recorded with H2O2 in electrolytes at different pH values. The results revealed that raising the pH decreases the intensity of the ECL, shown in
The ECL intensity corresponding to H2O2 at different concentrations was reported. At least three experiments were conducted for each concentration, and based on the average intensity of each concentration, the relationship between the H2O2 concentration and the ECL intensity, a calibration plot was presented (
To explore the capability of the biosensor apparatus to be employed in the detection and screening of an analyte that is a biomarker of several diseases, UA was investigated. UA is a product resulting from purine metabolism in the human body. Elevated UA levels in urine or serum can impair renal function and may be used as indicator of gout, cardiovascular and renal disease, hypertension, and other conditions. Low UA levels have been linked to molybdenum insufficiency, copper toxicity, and the progression of multiple sclerosis. As a result, the detection of UA in human physiological fluids is critical for diagnosing individuals with diseases linked with abnormal purine production and catabolism.
The oxidation of UA in the presence of the enzyme uricase that produces H2O2 is described in the following equation:
H2O2 is the product of the enzymatic oxidation reaction of UA, as such the luminol-H2O2 system is being used to detect UA.
Following the same procedure detailed for H2O2 detection, the calibration plot of UA was obtained by recording ECL intensities of UA at different concentrations (See
To further explore the efficiency of the reaction of the equation above, the calibration plots (
Efficiency of the reaction (percent yield)=(actual yield/theoretical yield)×100.
As shown in
The reproducibility of different devices and the reusability of one device in multiple experiments are characteristics of a biosensor apparatus permits estimating its potential for practical applications. Hence, these parameters were studied, and the results are shown in FIGS. 11A1 to 11E.
Specifically, to evaluate the reproducibility of the fabrication process, five different devices were examined to detect 50 μM and 300 μM of UA each, as the low and the high concentration within the reported linear range of this device, in the electrolyte solution. The ECL intensities are collected, and the RSD of the intensity for these five experiments was calculated to be 14.79% for 50 μM of UA and 7.52% for 300 μM of UA. These numbers indicate good reproducibility for the detection of UA at different concentrations in the linear detection range. The observed variation is due to several factors. On one hand, the resistivity of the ITO PET substrate differs slightly between devices with a tolerance of approximately 1.75% (
The manufactured device could technically be cleaned and reused. In order to test this practicality of this approach, the reusability (repeatability) of the same device for multiple experiments was then studied by measuring the SE-ECL intensities of 50 and 300 μM of UA. The ECL signal was recorded using the same device with fresh electrolyte solution each time for 5 times. The relative standard deviation (RSD) was calculated to be 7.00% and 2.06% for 50 and 300 μM of UA, respectively. After 5 experiments, the ECL intensity has gradually decreased, and after the 9th experiment, a dramatic fall in the ECL intensity was observed (results not shown). This is probably due to the damage of the ITO coating as a result of the passing current and induced heat, which consequently leads to decreased conductivity. However, since the device is inexpensive (the cost of the CMOS sensor is 1$ and the approximate cost of the microfluidic device is 3$), it can be used as a disposable device, and its reusability might not be an issue of significant concern.
There are various chemicals in body fluids that might interfere with the ECL detection of UA. For example, glucose, ascorbic acid (AA), and creatinine, which are concurrently present in body fluids, may interfere with the detection of UA. Therefore, the selectivity of the fabricated device in the presence of glucose, AA, and creatinine was studied by detecting 50 and 300 μM of UA in the electrolyte solution with and without interference. Concentrations of these interfering molecules were chosen based on their typical physiological concentrations in human saliva.
ECL UA intensities were recorded in the electrolyte with (I on the y-axis in
Furthermore, a negative control experiment that does not include UA reveals very low ECL intensities (see
The biosensor apparatus can be used for the detection of UA in body fluids including saliva and urine. Normal physiological ranges of UA are 70-320 μM in saliva and 1.49-4.46 mM in urine. In order to measure the ECL intensity of UA in body fluids, these samples need to be mixed with the electrolyte. The ratio of saliva to electrolyte was chosen to be 1:1 as it is within the linear range of the biosensor apparatus proposed herein. The calculated recovery rates of UA at final concentrations were of 50 μM and 300 μM, which match the low and high ends of the normal concentrations in saliva. To account for higher concentrations of UA present normally in urine, a high concentration of UA (similar to the condition in urine) was diluted in a mixture of electrolyte and artificial urine for a final concentration of 300 μM, such that the measurement can be performed within the linear range of the device. 50 and 300 μM of UA in the electrolyte containing artificial body fluids were detected by the device (n=3) and the ECL intensities were obtained. Based on the calibration plot, the recovery rate was calculated and shown in the table below:
The recovery rates for 50 μM UA in saliva and urine are 89.76% and 83.63%, with RSDs of 3.64% and 4.09%, respectively. The recovery rates for 300 μM UA in saliva and urine are 105.92% and 122.10%, with RSDs of 2.02% and 4.87%, respectively.
Since the physiological concentration of UA in human urine is higher than the linear range of UA calibration plot in this study, the effect of dilution on the recovery rate was also studied. After mixing the artificial urine with the electrolyte in 1:9 ratio, the recovery rates for 300 μM of UA is 111.58% with a RSD of 7.75%.
To make a fair comparison between the ECL-on-CMOS platform and other platforms, including microscopes and smartphones, a Nikon eclipse Ti2 equipped with a Prime 95B 25 mm BSI CMOS camera (cooled to −25° C.), and Samsung Galaxy S21 ultra were employed. A 4× objective lens with a 30 mm working distance was chosen for the microscope to have the same field of view and capture the full length of both microfluidic channels. The imaging parameters (ISO, gain, exposure time) were chosen to be the same as the experiment with the CMOS detector. The pro imaging mode was used in the smartphone to adjust the imaging parameters to be similar to those of the other systems. The test channel in this experiment contained 300 μM H2O2 as the analyte. The distance between the ECL reaction and the detector was 30 mm for the microscope with a 4× objective, 300 μm for the CMOS system, and approximately 10 cm for the smartphone.
The advantage of the presently disclosed system compared to other detection systems comes from the fact that the sample is located very close to the detector, leading to a large solid angle for photon collection, as illustrated in
The biosensor apparatus was adapted to different systems, including a microscope and a smartphone, and compared the detected ECL intensities under the same experimental conditions (
In this example, it was assumed that the ECL emission is characteristic to Lambertian emitters. Therefore, the collection efficiency (η) of each platform is the fraction of the cone of light subtended by the detector divided by the whole solid angle emitted from the ECL reaction.
The angle of the detected cone of light cone and the corresponding solid angle were calculated as:
where 2θ is the apex angle of the cone, D is the distance between the ITO electrode (where the ECL reaction happens) and the detector (or lens), L is the diameter of the sensor (or diameter of the entrance pupil), and Ω is the solid angle subtended by the detector. Based on equation (5), when θ=π/2, the spherical cap becomes a hemisphere having a solid angle of 2π steradians. As a result, the whole ECL emission has a solid angle of 4π steradians. Therefore, the collection efficiency of the detector was calculated as:
where η is the collection efficiency. For the microscope, D is the working distance or the focal length of the 4× lens, which is 30 mm, and θ can be calculated using the equation:
NA=n sin(θ),
where the numerical aperture (NA) of the 4× lens is 0.10, and accordingly, θ is almost 6 degrees.
Considering the whole solid angle as 360 degrees, the collection efficiency in the microscope is almost 1.7%. Using the same calculation for CMOS (this work) and measuring D=300 μm and L=3200 μm (the average length and width of the CMOS sensor), the collection efficiency is almost 40%. This calculation is not applicable for the smartphone as the focus was being changed automatically during the experiment. The main difference between these platforms is the distance between the location of the ECL reaction (from where the photons are emitted) and the detector, i.e., the distance between the ITO electrode and the detector. In the present biosensor apparatus, the reaction happens closer to the photodetector, leading to significantly higher collection efficiencies in comparison with a microscope capable of imaging a similar field of view.
Another difference between these platforms is the quantum efficiency (QE) of the imaging sensor. The sensor of the microscope (Prime 95B 25 mm BSI CMOS) has a relatively large QE (close to 95%). The CMOS sensor employed in this work (OmniVision OV5647 BSI) has a relatively lower QE of 70% (according to the manufacturer website). Considering both quantum efficiency and collection efficiency (QE and η), the microscope detector converts ˜1.6% of the ECL photons into electrons, while this number for the CMOS platform is ˜28% or almost 17.5 times higher. However, this difference is based on the calculation at room temperature. In reality, the measured improvement in the recorded ECL intensities is only a factor of approximately 10 (shown in
As can be understood, the examples described above and illustrated are intended to be exemplary only. For instance, although the transparent substrate and the electrically conductive coating are described as two separate materials joined to one another, the transparent substrate and the electrically conductive coating may be made integral to one another. Moreover, it is intended that the transparency of the substrate or of the substrate is meant to encompass any type of transparency including, but not limited to, semi, partial or full transparency at some or all of the relevant wavelengths. The scope is indicated by the appended claims.
Number | Date | Country | |
---|---|---|---|
Parent | 63307492 | Feb 2022 | US |
Child | 18164654 | US |