Current sample preparation methods for processing whole blood involve large, laboratory-based pieces of equipment, high cost, and trained users. For example, a standard laboratory centrifuge, which is required for plasma extraction from whole blood, can cost upwards of $4,000 and take up to several square feet of space and must be operated by a trained user. In addition, filtering large volumes of blood (ml and above) clogs filtering devices, making filtering an impractical alternative to use of centrifugation.
Similarly, nucleic acid purification typically occurs via several pieces of laboratory equipment, including but not limited to, vacuums, specialized tubing, specialized buffers, centrifuges and vortexes.
Currently available filtering devices, that do not have the advantages of aspects described herein include, PlasmaDrop PD-50 (MDI Membrane), Blood Separator (Nupore), Vivid Plasma Separation Membrane (Pall). See, also, Wang, S. et al. (2012). Simple filter microchip for rapid separation of plasma and viruses from whole blood, Int J Nanomedicine. 7: 5019-5028. DOI: 10.2147/IJN.532579.
Objects and advantages of embodiments of the disclosed subject matter will become apparent from the following description when considered in conjunction with the accompanying drawings.
Aspects described herein combine portable sample preparation (e.g. from whole blood) and downstream processing (e.g., nucleic acid amplification) from the purified sample. Further aspects provide an optionally low-cost, portable, plasma extraction device that is capable of utilizing minimal power consumption. Aspects described herein integrate, for example, sample preparation and nucleic amplification into a plastic microfluidic chip. Use of these aspects can be automated, is low cost, uses minimal volumes of fluid (e.g., whole blood) and does not require a trained user to carry out each step. Combining these two components (plasma extraction and nucleic acid purification) is further beneficial, as these two steps are typically performed sequentially in a lab and must be performed prior to almost all laboratory nucleic acid assays.
As described herein, the exemplary 2-membrane design, selectively filters out cells and enables processing of large volumes of blood (e.g., about 100 milliliters and above) without clogging the whole filter. Conventionally, microfluidic analysis with large volumes of sample is extremely difficult because simple filters will get clogged. Conventionally, filters are available on market are recommended to be used individually, not together, because it was thought that conventional use would increase chances of clogging as well as increased sample loss. In addition, it was thought that stacking filters on top of each other would result in folding and sticking of the filters to each other, resulting in decreased or entire loss of filtration function and putting a large space between two filters would increase the dead volume and efficiency of the process.
Aspects described herein utilize two or more filters (e.g., a double filter design) that are able to separate components of whole blood (e.g. plasma, cells, cell-free nucleic acids in circulation). The filters can be “tuned” and can be swapped out (e.g. different size pores) depending on the application needed (e.g. different targets being analyzed, such as circulating nucleic acids, or nucleic acids inside cells which require additional lysis). Conventional filters can be used. Although such filters are available on market, and they are typically not combined together. As described herein, filters can be coupled to a device for use without the need for custom fabrication. Such devices can be made out of polymer (e.g., 3D printed) or plastic (PMMA) and coupled to a standard pump (see below) without the need for special equipment.
Fluid (e.g., whole blood) can be provided to the two-filter system using any type of pump (e.g. syringe or peristaltic), and parameters can be tuned based on downstream applications (e.g. flow rate can be optimized to avoid lysis of cells if necessary). This system can process milliliter volumes of whole blood which can be coupled to downstream processes on the microliter scale, despite the differences in volume scales. The filters system can extract, for example, about 500 microliters of plasma from un-diluted whole blood without clogging. In contrast, many portable sample preparation devices are on the microfluidic scale and can only extract microliters of plasma.
In one aspect, the filter system can be coupled to a microfluidic chip for nucleic acid extraction utilizing, for example, silica magnetic beads and chaotropic salts. In this aspect, filter system can be connected to a microfluidic amplification module for nucleic acid amplification, which utilizes a small-scale thermoelectric cooler for both active heating and cooling. All microfluidic components can be fabricated in plastic (e.g., PMMA) for disposability.
The exemplary filter system can extract large volumes of components from unprocessed (undiluted) whole blood without clogging or leakage and without significant force, such as centrifugation, making sample preparation compatible with downstream assays and without requiring dilution of whole blood. The exemplary system is tunable and adaptable to different applications (e.g., tune filters and settings for extraction, and tune reagents in a nucleic acid amplification chip (i.e., lysis buffers, PCR primers and polymerases, etc.). The systems and components described herein can be used in any suitable application, including, but not limited to, separation of blood components for other medical applications (e.g., infectious disease, hematology, etc.), separation of other fluids, and integration sample prep for lab-on-chip devices. Aspects described herein can be used for any medical application requiring sample pre-processing (e.g., point-of-care or lab-based).
The exemplary filter system can utilize a 3D printed holder with grooves and an O-ring to separate the two filters. Despite the large sample volumes being processed, the whole filtration procedure can be done without clogging, and at such low pressure that a peristaltic pump can be used, allowing for point-of-care analysis. Circulating DNA is able to move through the 2-stage filter for downstream microfluidic analysis, demonstrating the low sample loss.
The exemplary two-membrane design system can be used with a downstream microfluidic chip with magnetic beads for isolating and concentrating DNA, followed by elution into a smaller volume for downstream analysis. In this aspect, the two-membrane design can be first integrated with a microfluidic chip capable of concentrating the sample into an even smaller (5-50 μL) volume than the original starting volumes of sample (0.5-5 mL), and then used for further analysis. In this aspect, the input volume for the two membrane design system can be, for example, about 500 μL-5 mL and the output from the two-membrane design system can be about 200 μL-1 mL. The input volume for the DNA purification component can be about 200 μL-1 mL and the output volume can be about 5-50 μL. An example how the microfluidic chip can concentrate the sample is shown in
Microfluidic analysis for use in aspects described herein can be carried out using a microfluidic design which enables valving to control fluid movement for temperature cycling and real-time detection of nucleic acids.
Both sides of an exemplary microfluidic chip can be made entirely out of plastic (poly(methyl) methacrylate (PMMA) or other plastics). Sealing can be done on one side with a cover slip of PMMA using isopropanol and heat, while sealing on the other side can done with a standard adhesive (PCR adhesive or other). The side can also be sealed with elastomer (i.e. PDMS) which is typically used in such valves. Thus, different materials can be used to fabricate devices (e.g., PDMS polymer, plastic, or adhesive).
In one aspect, the microfluidic chip contains vertical valving components with concentric circles that can be actuated by a solenoid, such that sealing with adhesive and actuating with a solenoid can withstand significant pressure build-up (10-15 bar) due to thermal build-up (this is non-obvious as it is very difficult to withstand high pressure build-ups in hard plastics without leakage). The valving components that can be pulsed to actuate different flow patterns and control overflow. The solenoids can also be pulsed with modulation to tune to different applications.
Microfluidic chips, as described herein can be of a mutually compatible design to accommodate placement of components for valving (i.e. solenoids), temperature control (heating/cooling), and real-time fluorescence detection.
In one example, DNA amplification and detection, temperature control, and real-time fluorescence detection are done in one chamber. A thermoelectric cooler and a K-type thermocouple can be used to achieve fast and successful thermal cycling. The thermocouple is placed under the chip and on top of the thermoelectric cooler, such that the temperature can be measured sufficiently accurately enough to perform PCR (which is a highly sensitive and temperature-specific laboratory assay). This design does not require any use of thermocouple placed directly inside the chamber. The size and thickness of the plastic is suitable for appropriate heat transfer.
On the other side of the chip is a one-sided fluorescence detection system for DNA amplification and real-time DNA detection. In contrast, conventional designs for thermocycling require elements on both sides of the chip, and same for fluorescence detection. In this aspect, the amplification/detection elements are spatially compatible with the components for the valving. The system can utilize laboratory techniques such as PCR, isothermal amplification, and other traditional laboratory assays.
Aspects described herein provide a blood sample preparation system having a filter component comprising at least a first filter and a second filter for preparing a blood sample where each filter has a different pore size, and a microfluidic amplification component for amplification of nucleic acid from the blood sample.
The term “amplification component” refers to a device or apparatus capable of amplifying (i.e., increasing in quantity) or modifying nucleic acid (i.e., changing one or more nucleotides) using any suitable mechanism (e.g., PCR, CRISPR). In one aspect, the amplification component can be coupled with or connected to the filter component, as described herein, such that a filtered blood sample can be directed through channels, for example, to the amplification component following filtering.
The term “filter” as used herein refers to device capable of separating components in a medium such as a liquid. For example, filters can be used to separate components of complex biological fluids such as blood. Such filters can be made of any suitable material, as described herein, or shape so as to adapt to use in conventional or custom filtration components.
In one aspect, the volume of the blood sample is greater than about 100 pl. In another aspect, the microfluidic amplification component is in fluid connection with the filter component. In yet another aspect, the DNA purification component is in fluid connection with the filter component and the microfluidic amplification component.
Optionally, the blood sample preparation system can have a DNA purification component (e.g., magnetic beads) for further nucleic acid purification following, for example, use of the first and second filters to separate components of blood for further analysis and processing. The term “DNA purification component” refers to a device or apparatus capable of purifying nucleic acid (i.e., removing components other than nucleic acid) through suitable methods (e.g., column chromatography, magnetic beads, PCR). For example, DNA or another nucleic acid can selectively bind to magnetic beads and be separated from a fluid.
In one aspect, the blood sample volume into the filter component is about 500 μL-5 mL and the output volume from the filter component is about 200 μL-1 mL. The fluid input volume into the DNA purification component can be about 200 μL-1 mL. This input volume can be reduced during DNA purification by binding DNA to a purification component (e.g., magnetic beads) and eluting the DNA in a smaller volume (e.g., 5-50 See, e.g.,
The filter component can include an inlet component having an inlet port for receiving a blood sample, an outlet component having an outlet port for releasing a filtered blood sample, and a filter support for retaining the first filter and second filter.
As described herein, the filter membranes can have a pore size for excluding blood components (e.g., cells). In one aspect, the pore size of the first filter is larger than the pore size of the second filter. In this aspect, the first filter can remove larger components of blood while the second filter can remove smaller blood components. In addition to the advantages described herein, use of two filters can more efficiently filter blood components, requires smaller volumes, and reduces clogging of filtering devices.
As described herein, the filter component can have a first side and a second side. The first filter can be retained on the first side of the filter component and the second filter can be retained on the second side of the filter component. Optionally, a microchannel is disposed between the first filter and the second filter.
The pore size of the filters can range from about 1 μm to about 25 μm. In one aspect, the pore size of the first filter and the second filter range from about 2 μm to about 6 μm. In another aspect, the pore size of the first filter is about 5 μm and the pore size of the second filter is about 3 μm. In yet another aspect, the pore size of the first filter is about 5 μm and the pore size of the second filter is about 2 μm. In a further aspect, the first filter is about 3 μm and the pore size of the second filter is about 2 μm.
Optionally, the blood sample preparation system has a pump for providing a blood sample. The pump can be any suitable pump (e.g., a syringe pump and a peristaltic pump). The flow rate of the pump can be from about 100 μL/min to about 600 μL/min.
In another aspect, the microfluidic amplification component includes a polymerase chain reaction chamber for amplifying nucleic acid.
Optionally, the microfluidic amplification component has at least two inlets for the polymerase chain reaction chamber for providing reagents to the polymerase chain reaction chamber. The microfluidic amplification component can also include a channel connecting the at least two inlets for mixing reagents. In another aspect, the microfluidic amplification component includes an outlet from the polymerase chain reaction chamber for disposal of waste products.
In one aspect, the microfluidic amplification component includes at least one valve for controlling flow of reagents to and from the polymerase chain reaction chamber. In yet another aspect, the at least one valve is a solenoid valve.
In one aspect, the solenoid valve comprises a sensing tip. In another aspect, the microfluidic amplification component further comprises an actuator (e.g., a membrane). The membrane can be made of any suitable material (e.g., PDMS and PMMA).
Optionally, microfluidic amplification component includes an extraction chamber disposed between the at least two inlets and the polymerase chain reaction chamber for extracting nucleic acid from the blood sample.
Aspects described herein also provide methods of filtering and amplifying a nucleic acid target from a blood sample by providing a blood sample to a filtering component; filtering the blood sample through a first filter and a second filter to form a filtered blood sample; and amplifying the nucleic acid target from the filtered blood sample.
The blood sample can be provided to the filtering component by a pump (e.g., a syringe pump and a peristaltic pump). The flow rate of the pump can be from about 100 μL/min to about 600 μL/min.
In another aspect, the first filter and second filter have a pore size, and the pore size of the first filter is larger than the pore size of the second filter. The pore size of the first filter and the second filter can range from about 1 μm to about 25 μm or about 2 μm to about 6 μm.
In one aspect, the pore size of the first filter is about 5 μm and the pore size of the second filter is about 3 μm. In another aspect, the pore size of the first filter is about 5 μm and the pore size of the second filter is about 2 μm. In yet another aspect, the pore size of the first filter is about 3 μm and the pore size of the second filter is about 2 μm.
Further aspects provide methods of filtering a blood sample and amplifying a nucleic acid target from the blood sample by providing a blood sample to a filtering component; filtering the blood sample through a first filter and a second filter to form a filtered blood sample; purifying nucleic acid from the filtered blood sample to form a purified nucleic acid sample; and amplifying the nucleic acid target from the purified nucleic acid sample. The nucleic acid can be purified from the blood sample with, for example, magnetic beads.
In this aspect, a blood sample volume into the filter component is about 500 μL-5 mL and a fluid output volume from the filter component is about 200 μL-1 mL. In another aspect, an output volume after purifying nucleic acid from the filtered blood sample is about 5-50 μL.
Exemplary devices and methods for blood plasma extraction with the specific goal of subsequent cfDNA (cell-free DNA) isolation and analysis utilizing the following features are presented below:
For this reason, an exemplary multi-stage filter device that utilizes parallel membranes of subsequently smaller pore sizes was selected. Together, these membranes eliminate the cellular components of blood, step by step. In this way, a continuous filtration can be achieved, and pure plasma can be collected in a continuous manner while avoiding fast clogging that often results in typical single membrane devices. At the same time, the multi-stage filter achieves faster separation time than sedimentation, and creates a highly purified plasma product.
An exemplary schematic illustration of the two-stage filter is presented in
Membrane materials and pores sizes can be selected based on the desired application. In one aspect, the material can be selected to facilitate analysis of the cell free DNA contained in blood plasma. Thus, a material, such as polycarbonate, with low nucleic acid binding can be chosen to avoid loss of the cfDNA analyte during the filtration process. Polycarbonate has minimal protein and nucleic acid binding.
The pore size can be selected to specifically accommodate the diameters of both red blood cells (RBCs) and WBCs. RBCs range from 6-7 μm and WBCs range from 7-25 μm. Furthermore, in literature it has been demonstrated that RBCs, due to their deformability, can pass through a pore size of 5 μm. For this reason, membranes with a pore size starting from 5 μm can be used (e.g., Whatman® Nuclepore Track-Etched, purchased Sigma-Aldrich®, with pore sizes ranging from 5 μm to 2 μm).
In another aspect, two different configurations of filter-based devices can be selected: cross-flow filtration or dead-end filtration. Dead-end filtration can be used to avoid loss of analytes, and thus selecting a flow direction that is perpendicular to the membranes and in the same direction of the collection channel could help cfDNA recovery. In addition, there are three different orientations that can be paired with dead-end filtration. Depending on the position of the membrane with respect to the direction of fluid flow, one can choose between a “membrane-on-bottom” configuration, “membrane-on-side” configuration and “membrane-on-top”configuration (e.g.,
The membrane-on-bottom configuration allows the sedimentation of cells to occur simultaneously during the filtration process, since fluid flow occurs in the same direction as gravitational sedimentation. This could lead to faster clogging of the membrane since there are two forces (filtration and sedimentation) working in concert to pull the fluid toward the membrane. Membrane-on-side and membrane-on-top configurations allow for maximize filtration capacity by redirecting the sedimentation toward the bottom of the filter chamber, maximizing the time before clogging can happen. In particular, the membrane-on-top configuration has been shown to achieve the best results in term of membrane area available without clogging for the longest amount of time.
Filtration Component Design:
In reference to
The structure from 3 different parts that can be assembled together. The symmetrical geometry of the device allows for flow from both the top and from the bottom.
The top side of the middle layer of the MOM device 112 is shown in
Inlet/outlet 124 are designed as a cone shape structure on one side, to match with a 1.5 mm in diameter tube, and an empty cone-shape on the other side. This last feature allows a symmetrical filling from the inlet as well as a symmetrical liquid collection toward the outlet. The exemplary structure was assembled with two polycarbonate membranes with 5 μm and 3 μm pore size respectively.
The four holes 114 at the corners, 0.66 cm in diameter, can permit correct alignment during the assembly of the device and integrate a support structure for the device.
The structure was tested with a flow from the bottom to the top with tinted water. A support structure made by 4 plastic poles was created to support the device. 4 ml of water was loaded into the syringe, mounted in the syringe pump, and connected to the inlet tube of the device. The outlet was connected with a 1.5 mm inner diameter tube to collect fluid. The chosen flow rate to test the device was 200 μL/min.
It took the fluid shown in
In order to prevent any possible leakage, an optional groove 136 accommodating an optional O-ring 138 was created in both the top and bottom parts as shown in
This exemplary was tested with clinical blood samples. A schematic of the complete structure of the exemplary device including groove 136 is presented in
As described herein, blood samples can be diluted in accordance with a desired dilution factor, a choice of membranes and pore sizes, and a flow rate in order to obtain a large enough quantity and a pure enough sample of plasma, from which it will be possible to conduct studies relating to analyzing circulating fragments of, for example, circulating free DNA (cfDNA).
For example, PBS (phosphate buffered saline) can be used to dilute the blood and undiluted human whole blood and samples diluted in a 1:8 ratio (i.e., 1 part of whole blood and 7 parts of PBS) can be tested.
In another example, a flow rate of 100 μl/min and two different circular polycarbonate membranes, each 47 mm in diameter, one with 5 μm pores and the other with 3 μm pores, can be used to optimize the dilution factor. In addition, at least 500 μl of plasma was collected, to have enough sample volume to perform cfDNA detection studies even with the highest dilution.
As shown in
The exemplary device was tested with 4 human blood samples with different dilution conditions: 1) whole blood (no dilution), 2) 1:4 dilution with PBS, 3) 1:6 dilution with PBS, and 4) 1:8 dilution with PBS. Once the sample entered the device, due to the chosen design, approximately 18 minutes was needed to start collecting the output at a flow rate of 100 μl/min. Output from the four conditions was collected for at least 5 minutes after appearing at the outlet. The collected outputs are shown in
A hemocytometer was used to calculate the separation efficiency, as shown in
In addition, the flow rate was analyzed. As discussed above, a flow rate of 100 μl/min was the fixed value chosen to start the experiments, since it was likely low enough to avoid hemolysis and most useful to characterize the optimal dilution factor. Thus, in order to characterize different flow rates, a combination of two membranes with 5 μm and 3 μm pores size with a 1:6 dilution factor was used.
In one aspect, a higher flow rate can be used to reduce the time needed for plasma collection which could be important in a point-of-care (POC) device. Specifically, the device was tested with three different flow rates: 1) 100 μl/min 2) 150 μl/min and 3) 200 μl/min. The experimental setup was the same as the previous set of experiments. The quality of the plasma outputs from the different flow rates were assessed again through a hemocytometer, to study the separation efficiency, and with a spectrophotometer, to study the absorbance at Hemoglobin related peaks. The output collection of the samples exiting the device related to the 150 μl/min and 200 μl/min flow rates started after around 12 minutes and 9 minutes, respectively. The collected outputs are shown in
As shown in
Membranes were tested to determine which membrane provides the best performance for this exemplary. Three polycarbonate membranes with different pores size (5 μm, 3 μm and 2 μm) were chosen to be tested in the plasma separation process.
The exemplary fixed configuration described above was used to optimize the dilution factor and flow rate used a two-membrane filtration configuration with polycarbonate membranes of 5 μm for the first filter membrane and 3 μm for the second filter membrane. Two addition configurations, 5 μm+2 μm, and 3 μm+2 μm, were also compared.
The experimental setup remained the same and the and the previously described flow rate of 100 μL/min and dilution factor of 1:6 were used. The collected samples are presented in
As shown in
As described above, the quality of the extracted plasma was assessed in terms of separation efficiency and hemoglobin level to “tune” the filter component and optimize the dilution factor, flow rate and membrane pore size combinations. The resulting purified sample was then analyzed for the presence of cfDNA.
To simulate the presence of cfDNA, a short fragment (180 bp) of the human MSTN1 gene was designed using Macvector, and synthesized by Integrated DNA Technologies (IDT®). The gene was amplified using the polymerase chain reaction (PCR) and purified with QIAquick PCR Purification Kit, purchased from Qiagen®, before being spiked in diluted blood samples. A set of primers (forward primer 5′-TTG GCT CAA ACA ACC TGA ATC C-3′ with a Tm=55.9° C. and reverse primer 5′-TTG GCT CAA ACA ACC TGA ATC C-3′ with a Tm=54.4° C.) were designed using Macvector and ordered from IDT to amplify a region of the MSTN1 synthetic DNA fragment by PCR (PCR size of 120 bp).
The gene was amplified and spiked in 5 whole blood samples at the following concentrations: 1) 25 ng/ml, 2) 250 ng/ml, 3) 500 ng/ml, 4) 750 ng/ml and 5) 1000 ng/ml. These concentrations were selected to cover a reasonable range of concentrations in which one would find cfDNA fragments in cancer patients.
The spiked human blood samples were each diluted 1:6 with PBS to a final volume of 3 ml each. The specific samples composition is described in the following list:
The filtration device described above was used to collect 500 μl of plasma from each of the above samples. Five control samples (C.1-C.5) of the same spiked concentrations and dilution factors were subject to traditional plasma extraction by centrifugation using a standard bench-top centrifuge, Sorvall® Legend® RT, at 2350 RPM at 4° C. for 10 minutes.
In a first experiment, once plasma was extracted for each case, a direct PCR of the collected samples was performed by using a Phusion Blood Direct PCR kit, purchased from ThermoFisher Scientific®, using the set of primers described previously. Moreover, an additional set of universal control primers was added to each reaction mixture. This set of control primers came as part of the Phusion Blood Direct PCR kit and is able to amplify a 237 bp fragment of the SOX21 gene housekeeping gene (derived from the Jurkat cell line), which is indicative of the presence of contamination mammalian genomic DNA and thus WBC lysis. A positive control for WBCs lysis (containing Jurkat genomic DNA) was prepared and used as a DNA control sample for lysis. The thermocycler protocol started with an initial denaturation (1 min at 95° C.), followed by 35 cycles including denaturation (1 second at 98° C.), Annealing (5 seconds at 60° C.) and extension (15 seconds at 72° C.) to end with a final extension step (1 minutes at 72° C.). After PCR the samples were loaded in a 2% agarose gel to perform gel electrophoresis, run at 140 V for ˜45 minutes, in order to check if it is possible to qualitatively detect the spiked fragments. The results of the PCR products after running in the gel are shown in
The filtration of a whole blood samples using the exemplary device can be used as a replacement for standard purification via centrifugation resulting in a lower cost and less time-consuming methodology.
In another aspect, non-specific bands can be reduced or eliminated as described. To eliminate non-specific bands, a DNA purification step following plasma extraction was performed, prior to PCR. Plasma extraction of the five samples using the micro device and the five controls using standard centrifugation was repeated, as described previously. This time, however, instead of doing direct PCR from plasma using the Phusion Blood Direct kit, cfDNA was first extracted from the plasma using the QIAamp Circulating Nucleic Acid Kit, purchased from Qiagen®. After cfDNA extraction, PCR was performed from the purified samples and from a prepared positive control for WBCs lysis (the same used in the previous experiment, containing Jurkat genomic DNA), using the Promega™ PCR Master Mix. The thermocycler protocol this time started with an initial denaturation (2 min at 95° C.), followed by 30 cycles including denaturation (30 seconds at 94° C.), Annealing (30 seconds at 60° C.) and extension (30 seconds at 72° C.) to end with a final extension step (5 minutes at 72° C.), as indicated in the kit instructions.
After PCR, all the samples were loaded in a 2% agarose gel run at 140 V for about 45 minutes. The results are shown in
From the figure, non-specific bands are no longer present. Additionally, there is very minimal WBC lysis, if any, and thus that there is no genomic background DNA contaminating the sample. Thus, a purification step, using, for example, magnetic beads-based techniques, can be integrated with the exemplary two-membrane filtration component described.
In this aspect, a device with flow rate of 100 μl/min was used to design an alternative structure with a smaller void volume, with the intent of achieving plasma separation in around 6 minutes, 3× faster than that of the previously described device. A new smaller structure was therefore designed with a geometry almost identical to the first one, but with smaller dimensions. The circular opening was reduced from 20 mm to 15 mm in diameter and the final structure, shown in
The alternative device was similarly characterized in terms of separation efficiency and plasma purity (data not shown), and was tested to see if it was possible to decrease extraction time while still detecting with the same capacity the presence of circulating MSTN1 fragments (
Plasma was extracted from each sample using the smaller device prototype. Output collection started after 6 minutes and 30 seconds, as expected based on the specific dimensions chosen, and plasma was collected for around 4 minutes. The separation process is visible in
Though the plasma exiting the device was clear at the start of the output, it quickly turned to a deep red color soon after. The final collected samples showed absorbance values of 1,183 and 1,207 at 540 nm and 576 nm, respectively, and a separation efficiency of around 75% as determined by the hemocytometer. While these values were not as optimal as the larger prototype, but it should be noted that the used parameters for flow rate and membranes were those optimized for the larger structure and thus they don't consider the smaller available area of the membranes in this smaller prototype. Despite this, PCR studies were subsequently conducted to check the detection of the spiked MSTN1 gene.
The 5 experimental samples and the 5 control samples were again prepared with the same conditions and concentrations as described previously. Direct PCR were performed by using Phusion Blood Direct PCR kit, ThermoFisher Scientific®, and both sets of primers described before (for amplification of MSTN1 120bp segments and WBCs control lysis). The positive control for WBCs lysis containing the Jurkat gene was prepared as well. After PCR the samples were loaded in 2% agarose gel to perform gel electrophoresis for 45 minutes at 140V. The results of the PCR products running in the gel are shown in
From the figure it is possible to see how, even if the quality of the plasma was lower respect the one collected from the optimized bigger device, the detection of the spiked fragments in the samples was still possible for all concentrations. Again, there were no visible differences as compared to the plasma samples extracted using standard centrifugation.
In the previous design and experiments, all fluids were controlled using a syringe pump which may not be optimal for POC. Thus, in another aspect the device is compatible with a portable peristaltic pump as described below. Additional testing was performed with more samples and to compare the device performance to a benchtop centrifuge.
A miniature peristaltic pump was purchased from Dolomite (Item #: 3200243). The pump is lightweight (11 g) and compact in size (30 mm×12 mm×14 mm), and thus was optimal for integration with our device. The pump was fitted with silicone tubing for attachment with the exemplary device.
Characterization of voltage versus flow rate with the pump was first carried out to match voltage to previously designated/optimized flow rates (
A flow rate of ˜0.6 V was selected to mimic flow rate of 100 μL/min optimized previously using the exemplary syringe pump as described herein. Next, comparisons between extraction efficiency using the peristaltic pump, syringe pump, and centrifuge were made (
Separation efficiency values were calculated using a hemocytometer and counting the total number of cells in a diluted sample prior to and after plasma extraction. As shown in
One-way ANOVA followed by Tukey's multiple comparison tests were performed for the three groups using Prism. There was no statistical difference found between the three groups (p=0.2255) by one-way ANOVA. Using Tukey's multiple comparison test, the syringe pump and peristaltic pump showed no statistical difference (p=0.5783), the syringe pump and centrifuge showed no statistical difference (p=0.2007), and the peristaltic pump and centrifuge showed no statistical difference (p=0.7601). Thus, it can be concluded that, using separation efficiency as a measure, the syringe and peristaltic pump set-ups both performed comparably to that of the centrifuge control, although the peristaltic pump had better performance compared to the syringe pump for this application.
Next, the degree of hemolysis that occurred during plasma extraction was tested for each of the three different set-ups. A plate reader (BioTek) was used to measure absorbance at 540 nm (left panel) and 576 nm (right panel) (
1:16 dilutions of sample were made to comply within the absorbance levels of the plate reader. At 540 nm, the mean normalized values were found to be 0.9753+/−0.01296, 0.01386+/−0.007945, 0.3887+/−0.3203, 0.007533+/−00518 (A.U) for whole blood, peristaltic pump, syringe pump, and centrifuge control, respectively. Using One-way ANOVA followed by Tukey's multiple comparisons test, a statistically significant difference was found amongst all four groups (p<0.0001).
A significant difference was found between whole blood vs. peristaltic pump (p<0.0001), whole blood vs. syringe pump (p=0.0002), and whole blood vs. centrifuge (p<0.0001). No significant difference was found between the peristaltic pump and centrifuge control (p=0.9856), syringe pump and centrifuge control (p=0.4691), and peristaltic pump and syringe pump (p=0.6795). At 576 nm, similar results were found. The mean normalized absorbance values at this wavelength were found to be 0.9761+/−0.01246, 0.01698+/−0.01271, 0.3643 +/−0.3263, 0.007135+/−0.003941 (A.U.) for whole blood, peristaltic pump, syringe pump, and centrifuge control, respectively.
Using One-way ANOVA followed by Tukey's multiple comparison test, a statistically significant difference was found amongst all four groups (p<0.0001). A significant difference was found between whole blood vs. peristaltic pump (p=0.0001), whole blood vs. syringe pump (p=0.0008), and whole blood vs. centrifuge control (p<0.0001). No significant difference was found between the peristaltic pump and centrifuge (p=0.8966), syringe pump and centrifuge (p=0.4679), and peristaltic pump and syringe pump (p=0.8666). Once again, it can be concluded that, while both the peristaltic pump and syringe pump set-ups are comparable to the centrifuge control for plasma extraction, the peristaltic pump performs slightly better.
Additional testing was performed on spiked whole blood samples of cfDNA fragments to test percent recovery after passage through the exemplary plasma extraction device for a quantitative assessment (
All reactions can be performed in triplicate on a Quant Studio 3 qPCR machine (Applied Biosystems).
Quantitatively circulating fragments spiked at 100% allelic frequency directly from plasma were detected (
To ensure pure nucleic acid (NA) samples free of contaminants, many POC devices employ separate NA purification systems using the method of solid phase extraction based on silica membranes or particles. In this example, separate NA purification steps can be used in addition to the two-membrane filtering devices described herein.
Solid-phase extraction (SPE) is a sample preparation process by which compounds that are dissolved or suspended in a liquid mixture are separated from other compounds in the mixture according to their physical and chemical properties. In this case the liquid mixture is the sample and the compound to be isolated is the DNA. There are several SPE methods where the difference is embedded in the silica form that is used—microparticles, gel, or membrane. All of these techniques can be exploited both outside and inside the device. Here the general workflow of silica-based extraction method is presented, considering that all the methods are composed by the same main steps (binding, washing and elution).
The binding step can consist of mixing a binding buffer, usually a chaotropic agent, with the sample and the Silica. Chaotropic buffer is used since it's able to enhance the binding affinity between DNA and silica surface via hydrogen bonding and hydrophobic inter-actions. The chaotropic agent allows to make a phosphate group free from the DNA, and in the meanwhile the water present in the buffer protonates the Silica surface. These reactions make possible the binding between the silica and the DNA since the phosphate group becomes “exposed” and a hydrophobic interaction is possible.
The washing step allows for removal of other biological components bound to the silica surface and gets rid of the salts usually present in the binding buffer. Since the wash buffer is alcohol-based, the silica is left air dried for a while in order to make all the alcohol evaporate and not interfere with the next step.
The elution step re-hydrates nucleic acids so that they can be released from the membrane and are again free inside the solution. The elution buffer can often be just deionized water.
In this design, magnetic silica beads fixed with an external magnetic field to overcome issues with trapping small particles in very small (tens of microns) scale channels that often could overcome limitations due to deformation were used.
Fe3O4 magnetic beads coated with silicon dioxide (2.5-4.5 μm) were purchased from G-Biosciences, and supplied in phosphate buffered saline, pH 7.4 with 0.09% Sodium Azide and 0.02% Tween-20. An exemplary NA purification device was designed as an exemplary hexagonal chamber 146 (11×5×1 mm) (also referred to herein as bed extraction chamber), connected to inlet 148 and outlet 150 through first valve 152 and second valve 154 (280×250 μm (width x depth) (
The exemplary device was designed using Solidworks® (see
To exploit the valves function, a holder 156 for the solenoids (
In one example shown in
In this example, the results are generated from the PCR product so to provide a comparison with the other methods. The mean value achieved was calculated and shown in Table 2. In this example, the volume of the fluid following nucleic acid purification can be adjusted to a desired volume. In addition, the volume of the fluid following nucleic acid purification can be adjusted to a desired volume (e.g., from an input volume of about 200 μL-1 mL to an output volume of about 5-50 μL).
A typical extraction curve is also reported in
The summaries in
Without being bound by theory, it is believed that using just one inlet and one outlet, all the buffers and both the waste and eluted samples flowed in the same channels, which may have led to an unwanted mixing and the presence of impurities.
In another aspect, an exemplary microfluidic lab-on-a-chip device can consider the effective cfDNA concentration in a real sample is on the order of 100-0.1 ng/ml. However, this concentration was not detectable in this example by the ThermoScientific® Nanodrop 2000, which has a detection limit of 2 ng/μL for dsDNA.
The following experiments were performed with a starting concentration on the order of 10 ng/μL, to identify a possible trend regarding performances change caused by decreasing the DNA concentration (
The results (Table 4) are quite unexpected since they furnish higher efficiency than the one achieved with a higher starting sample concentration, but still with a very high standard deviation.
In
In one aspect, a trade off could be found between the efficiency and the amount of sample used.
After the extraction procedure was optimized, PCR was performed on a purified sample, in order to understand if the two steps could actually be implemented one after the other in a final lab-on-chip (LOC) platform. If the extraction was successful, the same band found with the DNA template is found with the PCR product run over the agarose gel.
Two different samples are shown in
In another aspect, additional steps related to the extraction procedure and coupling the extraction procedure with the PCR over a unique device are provided.
In this aspect, a new chip feature has been designed as an alternative to the above described devices. An exemplary device is shown in
Optionally, PCR chamber 164 can be added to operate directly on the microfluidic platform after sample purification. Waste channel 166 can be added to discard waste. In another aspect, one of the inlets can be replaced with a channel (e.g., a long, serpentine channel) connecting two inlets to permit the mixing of the sample and the binding buffer directly over the chip (remember that in micro-scale the mixing of solutions is achieved just by diffusion, so a long channel is necessary, and a serpentine geometry enables to occupy less space as possible).
In another aspect, devices described herein can use valves to control flow of liquid through the system. Therefore, the correct functioning of the valves is very important, otherwise the solutions will be able to follow randomly the different channels.
In one aspect, the sample is loaded and binding buffer (BB) (previously mixed) through one of the inlet leaving the related valve and the one of the waste channel open. In this way, the DNA will bind to the beads while the unwanted supernatant keep flowing and is discarded. The washing step works in the same way, but the inlet will be a different one. At the elution step the elution buffer (EB) is permitted to flow but the waste channel valve is kept closed, and the one of the PCR chamber open, so that the purified sample is collected.
The valve is one important part of microfluidic devices, especially for μTAS (micro total analysis system) “sample to answer” systems, and it is often compared to the transistor in the semiconductor industry. In one aspect, a solenoid valve compatible with a microfluidic PCR chip (sample to answer) that can withstand high pressures due to PCR is provided. This valve is compatible with a variety of materials (e.g., chips made entirely of plastic, a plastic-elastomer combination, or a plastic-adhesive combination).
In this aspect, a vertical valve actuated with a solenoid is provided. This valve design meets the requirements of a long-term stable, reliable, portable and low-cost valve for a fully-integrated point-of-care (POC) PCR device. The main drawback of solenoid actuated valves is the relatively large footprint required for each valve (roughly 1 cm2) and the lack of multiplexing capabilities (i.e. the ability to control multiple fluidic channels with a single valve). However, in this example, multiplexed valves were not used for two reasons: 1) multiplexing requires more complex designs with at least three layers, and 2) for this exemplary device, only 6-10 valves are required. An advantage of multiplexed valves is the ability to control 2n valves with 2n+2 actuators. Thus, for a chip containing 10 valves or less, it is unnecessary to increase design constraints to account for more complex multiplexed valves.
The valve van be manufactured by CNC micromilling, injection molding, and hot embossing. An exemplary working principle of the valve is explained in
PMMA-PDMS VALVE: This valve is made of two different polymers—an elastomer and a thermoplastic. One advantage of PDMS is its low Young's Modulus, which reduces the force needed to deform it and thus close the valve. In one aspect, the potential increased costs associated with PDMS production may be offset by the benefits of a thin membrane without features (e.g., channels, chambers, valves, etc.).
PMMA-PMMA VALVE: This valve is made of only one material, which is an advantage for the manufacturing and leads to fewer biocompatibility issues. However, the increased stiffness of the PMMA may require more force for deflection. An appropriate valve can be selected, for example, based on the desired biocompatibility, stiffness, and cost.
Based on models and given engineering restrictions in manufacturing, the following setups were tested:
PMMA-PDMS (
PMMA-PMMA (
PMMA was chosen as material for the main body of the chip (as depicted in
The setup is divided into an actuating and a sensing circuit. All devices were connected to an Arduino Uno and ultimately to a personal computer, in order to automate, synchronize and control the experiment and collect the data.
The push solenoid is from Adafruit Industries, LLC (Product ID: 2776). The pull-back spring of the solenoid was modified to have a spring constant of 0.024 N mm-1 in order for the plug to return to the off position when no voltage is applied.
As shown in
The thickness of the PDMS membrane was experimentally optimized evaluated by first applying a pressure in the horizontal directions at the inlet of the chip. For each PDMS membrane three identical experiments were conducted. The solenoid was actuated with 7.8 V DC during the entire experiment, and the leakage across the valve was measured with the flow sensor. Pressure in the range from 0-100 kPa in increasing in steps of 6.8 kPa was applied for 10 s per step, and PDMS membranes with thicknesses of 70, 120, 170 and 240 μm were tested. For each pressure step 120 data points were collected but only 60 were taken into consideration, since the first and the last 2.5 s of each pressure step were neglected to consider the time slot that was taken for precise manual pressure adjustment. From a total of 180 (60 times three experiments) data points the average and the standard deviation were calculated and plotted in
Furthermore, the PMMA-PDMS valve was tested with a pressure applied at the outlet of the chip, in vertical direction, as indicated in the upper left corner in
This can be a result of variations in the membrane thickness, fluctuations in the battery power and pressure adjustment errors.
Reliability of Valves: To demonstrate the working principle, valve on/off cycles of 5 s each were run. The data in
Duty Cycles and Response Time: Typically, solenoids are used as digital (on/off) actuators. If a DC voltage is applied the plunger of the solenoid moves out at maxi-mum speed until it come in contact with the membrane. This leads to two undesired effects. 1) the valve wears out and 2) The impulse is transferred to the fluid in the channel. Thus, a certain control over the force of the plunger is desirable. This is achieved by applying a PWM signals with different duty cycles. For the experiment depicted in
For the PMMA-PMMA valve, the model showed that an 80 μm PMMA membrane with a radius of 1.1 mm and a 120 μm PMMA membrane with a radius of 2 mm can sufficiently be deflected in order for the valve to shut the liquid flow and with-stand the pressure. To experimentally examine the pressure resistance, three identical experiments for each of the two PMMA valves were performed. The solenoid was actuated with 7.8 V DC. The applied pressure ranged from 0-11 kPa with steps of 0.2 and 1 kPa, respectively. For every pressure step, 120 data points were collected and 60 were taken into consideration. The first and the last 2.5 s of each pressure step were neglected considering the time slot required for precise manual pressure adjustment. From the total of 180 data points the average and the standard deviation were calculated and plotted in
Power: The voltage (V), current (I) and force at full stroke (Fmax) of the solenoid were measured for duty cycles from 0-100% at 976.56 Hz with a multimeter and a scale, respectively. The operating time was then calculated with a standard value for the charge capacity C=620 mAh of a 9 V battery. The results are shown in Table 5.
Overall, the size of the valve was restricted by a hole that had to be drilled in vertical direction to connect the channels on both sides of the PMMA chip. Thus, in this example, the minimal radius of the valve was 1.1 mm. The thickness of a valve made of PMMA or PDMS can be optimized. In this aspect, the best results were achieved with a PDMS membrane of 120 μm in thickness. This valve was stable for up to 75 kPa in the horizontal direction and up to 175 kPa in vertical direction. Furthermore, it worked reliably for more than 1000 actuation cycles. The Young's modulus of PMMA is more than 1000-fold larger than the one of PDMS. Consequently, the valve was not as pressure resistant as the valve with the PDMS membrane, however, still worked for smaller pressures. The valve with the smaller radius and the 80 μm membrane withstood pressure up to 0.8 kPa and the valve with the larger radius and the 120 μm membrane up to 10 kPa. While the smaller valve did not show a stable behavior over multiple cycles, the larger valve proved to be reliable for more than 100 actuation cycles.
The exemplary valves described herein provide both portability and their reliability. The PMMA-PMMA version is inexpensive for large scale production what makes it an attractive valve for a commercial device. The PMMA-PMDS valve is still manufacturable at an attractive price and applicable to high-pressure ranges. The valves are furthermore bio- (and, specifically, PCR) compatible and have a short response time. The valve can be applied on both sides of the chip, which, theoretically, doubles the number of valves per chip.
The valves discussed above form a strong foundational basis for similar valves with different types of solenoids (off shelf). In another aspect, a continuous push solenoid with a 0.25″ stroke distance and 1 oz force (McMaster-Carr, Product #699905K172) was used. This solenoid can provide different forces at different percentages of the stroke length and can be actuated by 12 V DC. A circuit diagram for a single solenoid valve is shown below.
The use of these adjusted solenoid valves reduces overheating of the valves and is able to withstand full pressure build-up of PCR in a microfluidic chip. Further, valves with PDMS and adhesive sealings were successfully tested.
Design of chip: These valves are compatible with an exemplary plastic, microfluidic chip capable of conducting PCR as described herein. These chips are made using PMMA and have features on both sides of a single layer, making them small, easy to manufacture, and easy to use once made. As shown in
These valves have been designed to be compatible with vertical solenoid valves described above. An image of the chips, valves, and 3D printed holder is shown below:
As shown in
In terms of electronic components necessary to conduct PCR, a small Peltier heater and K-type thermocouple for heating, cooling, and temperature sensing can be used. These components were selected because they are cheap, accurate in small temperature ranges, and can be reused without touching. In one aspect, the chip is designed to be fully sealed—the thermocouple is able to provide a very accurate approximation of the temperature inside the PCR chamber even when it is not in contact with the fluid, which is possible due to the optimized design of the chip. PWM and active cooling can be integrated using this very low-cost, minimal equipment set-up. A thermocouple was sandwiched between the heater and the chamber of the chip to measure the temperature and plotted as compared to the target temperature (
A custom Arduino program was written for this example, however, it is very simple to re-write the program for other thermal conditions, such as isothermal amplification, which may be useful for additional on-chip experiments. For instance, qPCR protocol was performed on this chip and successful results were obtained that are comparable to the gold-standard device (Quant Studio 3, Applied Biosystems).
Further, proof-of-concept compatibility between valving components and PCR has been demonstrated. However, these valves and chips can be actuated in pulse patterns such that fluid flow can be controlled for use when integrating additional functionalities to the device, such as sample preparation and signal detection.
While aspects have been disclosed with reference to certain embodiments, numerous modifications, alterations, and changes to the described embodiments are possible without departing from the sphere and scope of the present disclosure, as defined in the appended claims. Accordingly, it is intended that the present disclosure not be limited to the described embodiments, but that it has the full scope defined by the language of the following claims, and equivalents thereof.
This application is a continuation of International Application No. PCT/US2019/052844 filed on Sep. 25, 2019, which claims priority to and the benefit of U.S. Provisional Patent Application No. 62/736,033 filed Sep. 25, 2018, both of which are hereby incorporated by reference in their entirety.
Number | Date | Country | |
---|---|---|---|
62736033 | Sep 2018 | US |
Number | Date | Country | |
---|---|---|---|
Parent | PCT/US2019/052844 | Sep 2019 | US |
Child | 17210083 | US |