The present disclosure generally relates to the field of sound/ultrasound technology and, more particularly, relates to a method for calculation of detecting depth and speed of moving objects based on speed changes of sound/ultrasound.
Transmission of ultrasound pulses is actually energy traveling of acoustic pulses in transmitting medium. If there is acoustic impedance during the transmission of the ultrasound pulses, the acoustic impedance will resist the movement of the ultrasound pulses. Currently, it is supposed that speed of the ultrasound pulses is identical in the same medium during the transmission. But, in the invention, the speed of the ultrasound pulses is considered as gradually reduced during the transmission due to the acoustic impedance of the transmitting medium, which gradually depletes the energy of the ultrasound pulses. According direct relationship between the acoustic impedance and the speed of the ultrasound pulses in the transmitting medium, higher speed of the ultrasound pulses will meet higher acoustic impedance and consume more its energy during the transmission. So, the question is if the speed of the ultrasound pulses can still keep the same as currently supposed when its energy is gradually reduced until exhausted? If the speed of the ultrasound pulses is gradually reduced during transmission, the detecting depth may be wrong based on calculating the detecting depth with fixed ultrasound speed for the ultrasound pulses with different frequencies.
Current ultrasound theory connects frequencies of the ultrasound pulses with their detecting depth, with lower frequency of the ultrasound pulses having deeper detecting depth. But, a thin piezoelectric (PZT) element can make the ultrasound pulses with a high frequency as well as a low frequency, which means the ultrasound pulses with both frequencies sending from the same PZT element have the same level of energy. So, the question is what are main factors that affect the detecting depth of the ultrasound pulses?
Ultrasound pulses can be reflected by motionless or moving objects, and according Doppler theory, it is currently considered that forward moving objects can compress the frequency of the ultrasound pulses and reversely moving objects decompress the frequency of the ultrasound pulses. So, Doppler mechanism has been widely used to measure the velocity of the moving objects based on frequency shift, such as medical ultrasound machine and Doppler radar. For the pulsed wave ultrasound, aliasing is explained with insufficient Doppler sampling rate of the frequency domain analysis. But, the theory of the frequency domain can not completely solve the aliasing problem of the pulsed wave ultrasound and the color ultrasound.
Thus, there is a need to overcome above problems to provide methods for more accurately calculating the detecting depth of ultrasound pulses, increasing the detecting depth of high frequency ultrasound, correctly calculating the speed of the moving objects and correcting the aliasing for the pulsed wave and the color ultrasound.
One aspect of the invention, correcting the transmitting distance of ultrasound pulses can rectify the registration of the detecting depth, which improves the quality of ultrasound images. Current sound/ultrasound theories and applications are based on the identical average speed in the same transmitting medium with various frequencies of the ultrasound pulses. But, the invention is based the speed reduction of sound/ultrasound during the transmission in the medium due to the loss of their energy caused by acoustic impedance. For ultrasound, calculating of the detecting depth of ultrasound pulses based on the identical average speed of the ultrasound pulses will cause miscalculation of the detecting depth due to different average speeds for the ultrasound pulses with different frequencies. Because a length and a density of the ultrasound pulses can affect the average speed of the ultrasound pulses, they can be used to calculate the ultrasound speed reducing coefficient and correct the registration of detecting depth of ultrasound pulses, which improve the quality of images.
In another aspect of the invention, changing thickness and density of piezoelectric (PZT) elements and sound speed in the PZT elements can regulate intensity of the ultrasound pulses, which affect their detecting depth. The detecting depth of the ultrasound pulses is not directly related to their frequencies, but related to the intensity of the ultrasound pulses. The thickness and the density of PZT elements decide the length and density of the ultrasound pulses, and the sound speed in the PZT elements decides the maximal speed of the ultrasound pulses in the transmitting medium. So, selecting the PZT with greater density and higher speed of ultrasound pulses in the PZT elements will increase the detecting depth for high frequency ultrasound.
Another aspect of the invention is detecting the speed of moving objects based on time of flight (TOF) shift of time domain analysis for a continuous wave, a pulsed wave and a color ultrasound. It is based on the speed changes of reflected ultrasound pulses by the moving objects, which change the TOF and the TOF shift of the ultrasound pulses. No matter in the continuous wave or the pulsed wave or the color ultrasound, when checking the speed of blood flow, the ultrasound system always detects the reflected ultrasound pulses from certain locations where ultrasound beam cross with blood vessels to calculate the TOF shift. So, the speed of the moving objects can be calculated based on the TOF shift. A angle between ultrasound beams and a direction of moving objects decides the value of the TOF shift. Calculating the speed of moving objects by simultaneously detecting TOF from two separated PZT elements from same gate avoids the effect of tortuous blood vessels and variant performances of sonographers, which improves the testing results with accuracy, simplification and reproducibility.
In the invention, the theory of above TOF and TOF shift can be used to completely correct an aliasing for the pulsed wave and the color ultrasound no matter how fast the speed of the moving objects will be. A calculated TOF is based on the average speed of ultrasound pulses in the transmitting medium and distance between transducer and the gate. A detected TOF is the time that the ultrasound system interprets from emitted ultrasound pulses and reflected ultrasound pulses. An actual TOF, which is an actual traveling time of the ultrasound pulses between transducer and the detecting objects. If the speed of the moving objects is too fast, which makes the actual TOF excesses its aliasing limit, the ultrasound system will misinterpret the reflected ultrasound pulses and generate the aliasing TOF. For the forward moving objects, the aliasing limit for the actual TOF is less than the value of half calculated TOF. If the actual TOF is smaller than the aliasing limit, the ultrasound system will misinterpret the reflected ultrasound pulse and add a value of calculated TOF into the actual TOF, which generates the aliasing TOF. Then the aliasing TOF is greater than the calculated TOF. So, the aliasing TOF shift is below the baseline, which represents the moving objects toward opposite direction. For reversely moving objects, their TOF aliasing limit is that the actual TOF is greater than the value of one and half calculated TOF. If the actual TOF is greater than its aliasing limit, the ultrasound system will misinterpret the detected TOF and subtract a value of calculated TOF from the actual TOF. Then the aliasing TOF is smaller than the calculated TOF. So, the aliasing TOF shift is above the baseline, which represents the moving objects as forward direction. So, in the invention, a computer program is designed to identify and correct the aliasing TOF shift no matter how fast the speed of the moving objects will be. Identifying and correcting the aliasing TOF shift can also be used to differentiate the colors of aliasing from the colors of the turbulent flow, which benefits clinical judgment and diagnosis.
The continuous ultrasound doesn't contain information of depth and the pulsed wave ultrasound may generate a problem of aliasing. In the invention, a method of coding ultrasound pulses is used to take advantages of the continuous wave ultrasound and the pulsed wave ultrasound, which can obtain information of depth and speed of the detecting objects at the same time, and avoid the aliasing. Within each pulse duration, different numbers of ultrasound pulses are sent, which is assigned a pulse character. Coded pulse characters are emitted with a same rest period, which is between each pulse duration. Coding pulse characters endow each pulse character with information of a sending time. The TOF of the pulse characters can be obtained between the time of sending the pulse characters and receiving the same pulse characters, and TOF shift can be obtained from the difference of sending rest period and receiving period of the pulse characters. Therefore, the depth and the speed of the detecting objects can be obtained from the TOF and the TOF shift.
Based on the speed reduction of the ultrasound/sound during the transmission, the intensity of the ultrasound pulses, the TOF, and the TOF shift can more accurately present the relationship between the ultrasound pulses with the depth and the speed of the detecting objects than the results from the identical ultrasound speed and the Doppler shift. Other aspects or embodiments of the present disclosure can be understood by those skilled in the art in light of the description, the claims, and the drawings of the present disclosure.
The following drawings are merely examples for illustrative purposes according to various disclosed embodiments and are not intended to limit the scope of the present disclosure.
a is a schematic illustration of spectrum for TOF shift of continuous wave ultrasound;
b is a schematic illustration of TOF shift and the profile of TOF shift for forward moving objects of pulsed wave ultrasound;
c is a schematic illustration of TOF shift and profile of TOF shift for reversely moving objects of pulsed wave ultrasound;
a is a schematic illustration of profile of aliasing TOF shift for forward flow of pulsed wave ultrasound;
b is a schematic illustration of profile of corrected TOF shift for forward flow of pulsed wave ultrasound;
a is a schematic illustration of profile of aliasing TOF shift for reversed flow of pulsed wave ultrasound;
b is a schematic illustration of profile of corrected TOF shift for reversed flow of pulsed wave ultrasound;
Reference will now be made in detail to exemplary embodiments of the disclosure, which are illustrated in the accompanying drawings. Wherever possible, the same reference numbers will be used throughout the drawings to refer to the same or like parts.
Transmitting in a medium, a sound pulse contains its energy, which decides its sound intensity. A pulse of the sound intensity includes its length, density and speed. The length and density of the sound pulse is related to thickness and density of materials that create the sound. The thicker material gives a longer sound pulse, which is like different sounds from different chords of a violin or a piano. The density of the sound is related to density of materials that create the sound. The higher density of the materials is, the greater density of the sound will be, such as the sound difference launched from wood or metal. The sound speed is also related to the speed sent from the material, such as different sound strength when hitting a key of a piano with different forces. Hitting the key more strongly will bring louder sound, which is related to faster sound speed. So, the intensity of the sound pulse is the multiplication value of its length, density and speed. A sound pulse with greater intensity will travel further, and the speed of the sound pulse will gradually reduce due to the acoustic impedance of transmitting medium, which gradually depletes the energy of the sound.
Sound intensity(kg/MS)=sound length(M)×sound density(Kg/M3)×sound speed(M/S)
M=meter; Kg=kilogram; S=second
Piezoelectric elements (PZT) in a transducer of ultrasound machine emit ultrasound pulses with their intensity, which is related to the length, density, and speed of the ultrasound pulses. Nowadays, the speed of the ultrasound pulses with different frequencies is considered as identical in the same medium. But, actually the speed of the ultrasound pulses is not fixed at the same speed as supposed during the transmission, and it will gradually reduce due to an acoustic impedance of the transmitting medium. As bullets shooting from a machine gun, their speed is gradually reduced due to loss of their energy caused by resistance of air. But, frequency of the bullets at any sites of trajectory may be kept the same. The transmission of the ultrasound pulses has the similar mechanism. The acoustic impedance is decided by density of the transmitting medium and the average speed of the ultrasound pulses in the transmitting medium. During the transmission, the speed of the ultrasound pulses gradually reduces due to the loss of their energy caused by the acoustic impedance, which will finally exhausts the energy of the ultrasound pulses. But, Ultrasound pulses keep the same frequency during the transmission, including their reflected frequency. A rate of the speed change of the ultrasound pulses is related to the density of the transmitting medium, sound speed in the transmitting medium, and the length and density of the ultrasound pulses. So, a speed reducing coefficient can be used to express their relationship with the speed changes.
One aspect of the invention is calculation of detecting depth of the ultrasound pulses based on the speed reduction of the ultrasound pulses during the transmission. Nowadays, calculating the detecting depth is based on an average speed of the ultrasound pulses in the transmitting medium, and the average speed of the ultrasound pulses is considered as identical for the ultrasound pulses with different frequencies in a transmitting medium, which may miscalculate the detecting depth due to a variation of the average speed of the ultrasound pulses with the different frequencies. As the ultrasound pulses leave the PZT elements and enter the transmitting medium, the speed of the ultrasound pulses is at their maximal speed. Then, under effect of the acoustic impedance, the speed of the ultrasound pulses will gradually reduce during the transmitting process. The longer the ultrasound pulses travel, the slower the speed of the ultrasound pulses will be. So, the speed reducing coefficient of the ultrasound pulses can be used to calculate a distance shift. The speed reducing coefficient is directly proportional to the density of the transmitting medium and the sound speed in the PZT elements, and inversely proportional to the length and density of the ultrasound pulses. The sound speed in the PZT elements is directly correlated to the maximal speed of the ultrasound pulses in the transmitting medium as they enter the transmitting medium. The depth shift is caused by speed reduction of the ultrasound pulses during the transmission. The depth shift of the ultrasound pulses is a half value of multiplication result of the speed reducing coefficient and maximal speed in the transmitting medium and traveling time of the ultrasound pulses. Vm is a maximal speed of the ultrasound pulses as they just enter a transmitting medium. t is the traveling time from emitting to receiving the ultrasound pulses. Vavg is a average speed of the ultrasound pulses in the transmitting medium. The detecting depth is a half value of multiplication result of the average speed and the traveling time.
Depth shift=speed reducing coefficient×Vm×t/2
V
avg
=V
m×(1−speed reducing coefficient)
Detecting depth=Vavg×t/2
Currently there is just one average speed of the ultrasound pulses in each transmitting medium. For instance, the speed of the ultrasound pulses in the soft tissue is about 1540 meter/second. But a change of the length or density of the ultrasound pulses affects their average speed in the transmitting medium. Such as there may be a difference of the speed reducing coefficient between the ultrasound pulses with 4 MHz and the ultrasound pulses with 8 MHz because of the difference of the length of the ultrasound pulses. So, their average speed in the soft tissue may be different, and their detecting depth may be also different at the same traveling time. Calculating the detecting depth with a fixed average speed may cause an error in their actual traveling depth. Because multiple factors affect the energy and the speed of the ultrasound pulses, such as reflection, absorption and scattering, the mathematic equation of the average speed of the ultrasound pulses just mainly reflects a relationship of the average speed with the length of the ultrasound pulses, which represents different length of the ultrasound pulses with different average speed. Therefore, it is necessary to more accurately calculate the detecting depth based on the different average speed of the ultrasound pulses in the transmitting medium according to the change of the length of the ultrasound pulses.
In another aspect of the invention, increasing the intensity of the ultrasound pulses increases their detecting depth. As the power transferred per unit area, the intensity of the ultrasound pulses equals the multiplication value of the length, density and speed of the ultrasound pulses. Increasing one or more of the thickness and the density of PZT elements and sound speed in the PZT elements increases the intensity of the ultrasound pulses, which increase their detecting depth.
In current ultrasound theory, the attenuation coefficient is directly proportional to the frequency of the ultrasound pulses. The lower the frequency of the ultrasound pulses is, the smaller the attenuation coefficient will be. As the frequency of the ultrasound pulses is inversely proportional to the thickness of the PZT elements, the more thickness of PZT has lower frequency of the ultrasound pulses.
Frequency=sound speed in PZT/2×PZT thickness
attenuation coefficient(dB/cm)=frequency (MHz)/2
Actually, the thickness and the density of the PZT elements are directly related to the length and density of the ultrasound pulses, which affect the intensity of the ultrasound pulses. The more thickness and the density of the PZT elements are, the greater length and density of the ultrasound pulses will be, which increase the intensity of the ultrasound pulses. As a heavier ball has ability of further traveling distance, the ultrasound pulses with greater intensity will have greater penetrating depth. The speed of the ultrasound pulses in the PZT elements is directly correlated to a characteristics of the PZT elements and a of electric output on the PZT element. So, changing the thickness and the density of the PZT elements and the sound speed in the PZT elements can regulate the intensity of the ultrasound pulses and their detecting depth.
Sound intensity=sound length×sound density×sound speed
Sound intensity=acoustic impedance×traveling distance of ultrasound pulses
Currently, in order to increase the frequency of the ultrasound pulses, the thickness of the PZT elements is reduced, which decreases the length of the ultrasound pulses and their detecting depth. But, in the invention, selecting the PZT materials with higher sound transmitting speed and increasing electric output on the PZT elements will increase a frequency of the ultrasound pulses. At the same time, increasing the density of the PZT elements increases the density of the ultrasound pulses, but not just decreasing the thickness of the PZT elements, it will increase the frequency as well as the intensity of the ultrasound pulses. As the result, it increases the detecting depth for high frequency ultrasound.
Frequency=sound speed in PZT/2×PZT thickness
Currently increasing frequency of the ultrasound pulses is used to increase axial resolution. Actually, the axial resolution is decided by length of the ultrasound pulses, which is related to the thickness of the PZT elements. But, the frequency of the ultrasound pulses is not directly related to the axial resolution of the ultrasound pulses, because thin PZT elements can generate high frequency as well as low frequency of the ultrasound pulses. Less thickness of the PZT elements generates shorter ultrasound pulses and smaller numerical values of spatial pulse length, which improve the axial resolution.
At the same time, the thickness of the PZT elements also limits the highest frequency a ultrasound system can reach, which is less than a value of sound speed in PZT elements divided by a length of the ultrasound pulses.
Highest frequency<sound speed in PZT/PZT thickness
Nowadays, Doppler mechanism is widely used to detect the speed of moving objects. According Doppler mechanism, a distance of sound resources, such as a coming or leaving motorcycle, is changing. Therefore, a frequency of sound pulses emitted from an engine of the motorcycle is relatively compress or decompressed due to the movement, which can be used to calculate to its speed.
As containing the energy, the ultrasound pulses can be reflected by motionless or moving objects. No matter in the continuous wave or the pulsed wave or the color ultrasound, when checking speed of blood flow, the ultrasound system always detects the reflected ultrasound pulses from fixed locations where the ultrasound beam cross with blood vessels to calculate TOF shift of the reflected ultrasound pulses. So, distances from the reflecting sites are fixed. It is more like playing table tennis, a racket hits a ball at a fixed location and changes a speed of the reflected ball, which changes its TOF. Comparing to motionless objects, moving objects will change the rebounding force to the ultrasound pulses. As in the
As the length and density of the ultrasound pulses is directly correlated to the thickness and density of the PZT elements, changing the length and density of the ultrasound pulses also affects their TOF and TOF shift. For the forward moving objects, increasing the length and density of the ultrasound pulses will have smaller rate of increased speed of the reflected pulses. It elongates their TOF and reduces their TOF shift. Decreasing the length and density of the ultrasound pulses have greater rate of increased speed of the reflected ultrasound pulses, which will shorten their TOF and increase their TOF shift. For reversely moving objects, increasing the length and density of the ultrasound pulses will have smaller rate of decreased speed of the reflected ultrasound pulses. It shortens its TOF and decreases their TOF shift. Decreasing the length and density of the ultrasound pulses have greater rate of decreased speed of the reflected ultrasound pulses, which elongates their TOF and increase their TOF shift.
So, one aspect of the invention is calculating the speed of the moving objects based on the TOF shift for the continuous wave or the pulsed wave or the color ultrasound. As mentioned above, the speed of the ultrasound pulses will gradually reduce, and the moving objects will generate the rebounding force shift, which changes the speed of the reflected ultrasound pulses, their TOF and TOF shift. Therefore, the TOF shift can more accurately present relationship between the speed of the moving objects and characters of the ultrasound pulses.
Currently, it is considered that speed of the ultrasound pulses is fixed in the same medium during the transmission. The moving objects will change the frequency of the reflected ultrasound pulses. The forward moving objects will compress the reflected frequency, which is higher than the emitted frequency. Its Doppler shift is above the baseline. The reversely moving objects will decompress the reflected frequency, which is lower than the emitted frequency. Its Doppler shift is below the baseline. So, calculating Doppler shift of the continuous wave (CW) ultrasound is based on difference between the reflected frequency and the emitted frequency. Vobjects is a speed of the moving objects, f is a frequency of a transducer, and V is a speed of the ultrasound pulses in a transmitting medium.
Doppler shift=reflected frequency−emitted frequency
The invention discloses that a TOF shift of the CW ultrasound is used to calculate the speed of the moving objects. The TOF shift of the CW ultrasound is difference between a time of emitting period and a time of reflected period. There are two PZT parts in CW ultrasound transducer. As in the
emitted period=the time between previous and following emitted pulses
reflected period=the time between previously and following reflected pulses
TOF shift=emitted period−reflected period
As 101 in
As 100 in
There is only one part of PZT elements in a transducer of the pulsed wave ultrasound, which sends and receives ultrasound pulses. So, the transducer has to receive previously reflected pulses before sending next emitted pulses. In order to detect speed of the moving objects, a gate is set with a certain distance. So, based on the average speed of the ultrasound pulses and the distance between the transducer and the gate, a calculated TOF can be obtained as A in
TOF shift=calculated TOF−detected TOF
For the pulsed wave ultrasound, there is the aliasing, which is caused by the ultrasound system misinterpreting the detected TOF from the reflected ultrasound pulses. If the speed of the moving objects is too fast, and makes the actual TOF excesses its aliasing limit, the ultrasound system will misinterpret it and the detected TOF becomes an aliasing TOF. Then the aliasing TOF shift is located on opposite side of the baseline, which presents the moving objects as toward opposite direction. The aliasing TOF shift also disrupts continuation of the profile of the TOF shift.
For forward moving objects, the aliasing limit of the actual TOF is less than the value of half calculated TOF. if the actual TOF is smaller than its aliasing limit, the ultrasound system will misinterpret the reflected pulses, and the aliasing TOF is a value of a actual TOF adding a calculated TOF, which is larger than the calculated TOF (104 in
Aliasing TOF=actual TOF+calculated TOF
Aliasing TOF shift=calculated TOF−aliasing TOF
Aliasing TOF shift=−actual TOF
So, in the invention, a computer program is designed to identify and correct the aliasing TOF shift. For the forward moving objects, the actual TOF is smaller than calculated TOF, and its TOF shift is above the baseline. As the speed of moving objects is increased, its actual TOF keeps decrease and smaller than the calculated TOF, and the TOF shift keeps increase and above baseline. But, after the actual TOF excesses its aliasing limit, the aliasing TOF becomes greater than the calculated TOF, and the aliasing TOF shift becomes below the baseline. The computer program will trace and compare the value of the following TOF and TOF shift with the value of the previous TOF and TOF shift. If the value of the TOF and the TOF shift approaches the value of half calculated TOF, and the value of following TOF shift is below the baseline, which discontinues the profile of the TOF shift. It is the aliasing TOF shift. After identifying the aliasing TOF shift, the ultrasound system will register the corrected TOF shift by subtracting the value of the aliasing TOF shift from one calculated TOF (116 in
TOF shift=calculated TOF−actual TOF
Aliasing TOF shift=−actual TOF
corrected TOF shift=calculated TOF−|aliasing TOF shift|
After rectifying the registering errors of TOF shift, the value of the corrected TOF shift will keep increase as increase of the speed of the moving objects, and the tip of the profile of the TOF shift is away from the baseline (84 in
For the reversely moving objects, the rebounding force is reduced, which decreases the reflected speed of the ultrasound pulses and increases their TOF, which is greater than the value of the calculated TOF. So, the value of the TOF shift is below the baseline. For the reversely moving objects, the aliasing limit of the actual TOF is larger than the value of one and half calculated TOF. If the value of the actual TOF excesses its aliasing limit, the ultrasound system will misinterpret the reflected ultrasound pulses and the aliasing TOF is the value of the actual TOF subtracting a calculated TOF, which is smaller than the calculated TOF. So, the aliasing TOF shift will be above the baseline; as the speed of the reversely moving objects keeps increase, the aliasing TOF is increased but the aliasing TOF shift is decreased, which make the tip of the profile of TOF shift is toward baseline (83 in
Aliasing TOF=actual TOF−calculated TOF
Aliasing TOF shift=calculated TOF−aliasing TOF
Aliasing TOF shift=2×calculated TOF−actual TOF
After identifying the aliasing TOF shift, the computer program will rectify the aliasing TOF shift by subtract the value of a calculated TOF from the value of the aliasing TOF shift, which is based on following equations:
TOF shift=calculated TOF−actual TOF
aliasing TOF shift=2×calculated TOF−actual TOF
correct TOF shift=aliasing TOF shift−calculated TOF
After correcting the aliasing TOF shift, the corrected TOF shift will increase as the speed of the moving objects keeps increase, which makes the tip of the profile of the corrected TOF shift away from the baseline. The corrected TOF shift will reestablish the continuation of the profile of the TOF shift (
Another method of avoiding the happening of aliasing is modifying the computer program in the ultrasound system to prevent adding or subtracting the value of a calculated TOF into the detected TOF after the actual TOF excesses its aliasing limit.
Differentiating Color of Aliasing from Color of Turbulent Flows for Color Ultrasound
For the color ultrasound, ultrasound system automatically sets different baselines at regular distance along the ultrasound beam. The TOF from reflected ultrasound pulses is compared with their respective baseline and get their TOF shift. Then colors are assigned according to a value of the TOF shift to represent a velocity of the moving objects. But, there are similar color patterns between color of the aliasing and color of turbulent flows. For the aliasing pattern, the color of the aliasing mistakenly presents as the moving objects toward opposite side after the actual TOF excesses its aliasing limit. For the turbulent flows, the color of the turbulent flows truly presents their moving direction. So, this will make the difficulties for clinical judgment and diagnosis for pathological situations. In the invention, differentiating the color of the aliasing from the color of the turbulent flows is based on the characters of the TOF shift of different colors.
For the color of the aliasing in
But, for the color of the turbulent flows in
So, differentiating and correcting the aliasing TOF shift for the color of aliasing from the TOF shift for the color of turbulent flows will benefit the clinical judgment and diagnosis for truly pathological conditions.
Calculation of Speed of Moving Objects with Two Separated PZT Elements Without Need to Adjust the Angle of the Ultrasound Beam
The speed value of the moving objects is important in judging some pathological conditions, such as stenosis of blood vessels. The angle of the ultrasound beam with the direction of the moving objects decides the value of TOF shift, which affects the calculation of the speed of the moving objects. Currently in order to get accurate speed of the moving objects, it is important to adjust the angle of the ultrasound beam with the direction of the moving objects within 45 to 60 degree. But, the tortuous blood vessels and variant performances of sonographers often derive different speed values of the moving objects from a same testing site, which increases the difficulties in the clinical diagnosis. In the invention, by simultaneously checking the TOF shifts at one detecting site from two separated PZT elements, the speed value of the moving objects can be accurately calculated without the need to adjust the angle of the ultrasound beams with the direction of the moving objects, which simplifies the operating procedures and avoids the variation of detection.
As in the
Calculation of Detecting Depth and Moving Speed of Objects with Coded Ultrasound Pulses
Because the pulsed wave ultrasound causes the aliasing and the continuous wave ultrasound loses information of distance, a method of coding ultrasound pulses can combine advantages of the pulsed wave ultrasound and the continuous wave ultrasound as well as avoid their disadvantages. This method is more like coding genomic sequence of deoxyribonucleic acid (DNA). A pulse duration is a time that ultrasound pulses are sent. During each pulse duration, different numbers of ultrasound pulses are sent, which is assigned a pulse character. For instance, just one pulse within the pulse duration is assigned as a pulse character A, two pulses as a pulse character C, three pulses as a pulse character G, and four pulses as a pulse character T. A resting period is a time between each adjacent pulse duration, and the resting period for the emitted ultrasound pulses keeps identical. A transducer of the ultrasound system contains one pair or more of sending PZT elements and receiving PZT elements. The sending PZT elements send ultrasound pulses with specific coded pulse characters, such as ATC GCG . . . , which is like codes of a DNA sequence. By this way, it actually endows information of emitting time for each pulse character. The receiving PZT elements receive reflected ultrasound pulses, which contain the same codes of pulse characters, such as A′T′C′ G′C′G′ . . . , which contains information of receiving time for each pulse character. Therefore, TOF can be obtained from the time between the emitting time and the receiving time of the related pulse characters. At the same time, TOF shift can also obtained from the time difference of emitting resting time and receiving resting time of the related pulse characters. For instance, TOF of pulse character A can be obtained from a time between the emitted pulse character A and the reflected pulse character A′, and TOF shift can be obtained from a time difference between the emitted resting period of the pulse character A and the receiving resting period of the pulse character A′. If reflections of the ultrasound pulses from one point keep identical TOF and their TOF shift is zero, it means the reflections coming from motionless objects. So, the TOF can be used to calculate the depth or distance of the motionless objects. If the TOF and the TOF shift keep variable, it means the reflections coming from moving objects. So, the TOF shift can be used to calculate the speed of the moving object. Because the moving objects change the TOF, which can not be used to calculate its actual location. By adding the TOF shift into the TOF, the TOF shift will compensate the changed part of TOF. For the forward moving objects, the TOF shift is positive, which will compensate shortened TOF. For the reversely moving object, the TOF shift is negative, which will offset extended TOF. So, a sum of the TOF shift with the TOF can be used to calculate the depth or distance of the moving objects. The depth or distance of the moving objects is a half multiplication value of the sum of TOF shift and TOF with the average speed of ultrasound in the transmitting medium. Both the speed and location of the moving objects can be used in imaging the color ultrasound. So, the method of the coded ultrasound pulses combines the advantages of the pulsed wave ultrasound and the continuous wave ultrasound in the ultrasound system, which can obtain the information of distance and speed of the detecting objects at the same time (
Improving Imaging Quality with Coded Ultrasound Pulses
Because each PZT element may not only receive the reflected ultrasound pulses emitted by itself but also receive the reflected ultrasound pulses emitted from other PZT elements, which cause noise and artifacts, such as mirror image or refraction. The noise and artifacts will affect quality of ultrasound images. With each PZT element emits its specific codes of the pulse characters, after receiving reflected ultrasound pulses, the ultrasound system will compare the received codes of the pulse characters with the emitted ones, and register locations of the reflections that have the same received codes of the pulse characters with the emitted codes of the pulse characters to an area belonging to the PZT element that emits the codes of the pulse characters. By this way, it may improve the noise and the artifacts.
Ultrasound just occupies sound wave with frequencies of more than 20 kilohertz. Actually sound pulses with any frequencies have the same mechanisms mentioned as above. So, the applications in detecting depth and calculating a speed of moving objects as mentioned above can be used in the sound pulses with any other sound frequencies, such as radar and sonar.
Other applications, advantages, alternations, modifications, or equivalents to the disclosed embodiments are obvious to those skilled in the art and are intended to be encompassed within the scope of the present disclosure.
This application claims priority to U.S. patent application Ser. No. 13/341,928 filed on Dec. 31, 2011, and U.S. patent application Ser. No. 14/305,074 filed on Jun. 16, 2014, and U.S. patent application Ser. No. 14/532,125 filed on Nov. 4, 2014, and U.S. patent application Ser. No. 14/645,475 filed on Mar. 12, 2015, the entire contents of all of which are incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
Parent | 13341928 | Dec 2011 | US |
Child | 14692777 | US | |
Parent | 14305074 | Jun 2014 | US |
Child | 13341928 | US | |
Parent | 14532125 | Nov 2014 | US |
Child | 14305074 | US | |
Parent | 14645475 | Mar 2015 | US |
Child | 14532125 | US |