The present disclosure generally relates to medicinal inhalers and, more particularly, to medicinal inhalers that include carbon dioxide as a propellant.
Delivery of aerosolized medicament to the respiratory tract for the treatment of respiratory and other diseases can be done using, by way of example, pressurized metered dose inhalers (pMDI), dry powder inhalers (DPI), or nebulizers. pMDIs are familiar to many patients who suffer from asthma or from chronic obstructive pulmonary disease (COPD). pMDI devices can include an aluminum canister, sealed with a metering valve, that contains medicament formulation. Generally, the medicament formulation is a solution and/or suspension of one or more medicinal compounds in a liquefied hydrofluoroalkane (HFA) propellant.
In pulmonary pMDIs, the sealed canister can be provided to the patient in an actuator—a generally L-shaped plastic part that includes a generally vertical tube that surrounds the canister plus a generally horizontal tube that forms a patient portion (e.g., a mouthpiece or nosepiece) that can define an inspiration (or inhalation) orifice.
The canister typically includes a metering valve that is crimped onto an appropriately-sized metal can. The metal can is typically made of aluminium, having a wall thickness of approximately 0.5 mm. The canister contains a formulation typically including liquid propellant(s), drug(s), co-solvent(s) and excipient(s). To prevent loss of the formulation, (primarily the liquid propellant), the metering valve contains rubber components that form seals.
Historically, the propellants in most pMDIs had been chlorofluorocarbons (CFCs). However, environmental concerns during the 1990s led to the replacement of CFCs with hydrofluoroalkanes (HFAs) as the most commonly used propellant in pMDIs. Although HFAs do not cause ozone depletion they do have a stated high global warming potential (GWP), which is a measurement of the future radiative effect of an emission of a substance relative to that of the same amount of carbon dioxide (CO2). The two HFA propellants most commonly used in pMDIs are HFA134a (CF3CH2F) and HFA 227 (CF3CHFCHF3) having stated 100-year GWP values of 1300 to 1430 and 3220 to 3350, respectively.
Various other propellants have been proposed over the years. Among them, carbon dioxide (CO2) has been mentioned as a potential propellant for pMDIs, but no pMDI product has been successfully developed and commercialized using carbon dioxide as a propellant.
It has now been found that despite CO2's major differences from other MDI propellants (such as much higher vapor pressure and different density, polarity, solubility, and component interaction characteristics) a practical pMDI can be made using CO2. This can be very useful due to CO2's stated lower GWP (GWP value of 1).
In particular, it has been found that conventional pMDI metering valves can have serious problems with CO2-containing pMDIs, such as, by way of example, leakage and valve performance. The valve stem seal, which is a dynamic seal through which the valve stem slides during actuation, can be particularly problematic. However, it has also now been found that using significantly harder stem seal materials, having a Shore D hardness of 45 to 80, and a valve stem seal opening sized sufficiently smaller than the valve stem passing through it so that the valve stem stretches the hard seal material, can address these problems, providing significant performance and leak prevention benefits. Moreover, contrary to the teachings of U.S. Pat. No. 6,032,836, this can be achieved without the need to use a lip seal design with an integral spring.
It has further been found that, unlike conventional pMDI valves, metered dose valves of the various embodiments disclosed herein can be made with an axial (relative to the axis of the MDI canister and motion of actuation) valve stem movement, but without using a spring to return the valve stem after actuation, instead relying on the increased pressure of the CO2 to do so.
Accordingly, some embodiments of the present disclosure provide a pMDI including a reservoir containing a pressurized formulation of medicament and CO2, and equipped with a metering valve having a metering chamber, a metering valve stem, and a metering valve stem seal having an opening through which the metering valve stem passes to form a dynamic seal between the metering valve stem and outside atmosphere. The metering valve stem seal has a Shore D hardness of 45 to 80 and its opening is adapted to be stretched wider by the metering valve stem passing through it than it would be absent the valve stem.
Some embodiments provide a pMDI having a reservoir containing a pressurized formulation of medicament and CO2. The pMDI is equipped with a valve including a valve housing having a first end, a second end exposed to pressurized formulation in the reservoir, and walls defining a metering chamber for receiving the formulation from the reservoir. The valve further includes a metering valve stem located within the metering chamber, the metering valve stem having a first end exposed to atmosphere and a second end exposed to pressurized formulation in the reservoir. At least a portion of the metering valve stem proximate the first end has a diameter larger than the diameter of at least a portion of the metering valve stem proximate the second end. The valve further includes a metering valve stem seal member in contact with the first end of the valve housing and the metering valve stem, where the metering valve stem seal member has a Shore D hardness of about 45 to 80. The valve has a primed position where the formulation can freely flow between the reservoir and the metering chamber and where the metering valve stem seal member seals the metering chamber from the outside atmosphere. The valve has an actuated position where the second end of the metering valve stem seals against the second end of the valve housing to seal the metering chamber from the reservoir, thereby defining a metered volume of formulation within the metering chamber, and where the metering valve stem includes a flow path allowing the formulation to freely flow between the metering chamber and the outside atmosphere.
Some embodiments of the present disclosure provide a pMDI having a reservoir containing a pressurized formulation of medicament and CO2. The pMDI is equipped with a valve including a valve housing having a first end, a second end with an opening exposed to pressurized formulation in the reservoir, and walls defining a metering chamber for receiving the formulation from the reservoir. The valve further includes a metering valve stem located within the metering chamber, the metering valve stem having a first end exposed to atmosphere and a second end exposed to pressurized formulation in the reservoir. At least a portion of the metering valve stem proximate the first end has a diameter larger than the diameter of at least a portion of the metering valve stem proximate the second end. The valve further includes a metering valve stem seal member in contact with the first end of the valve housing and the metering valve stem, where the metering valve stem seal member has a Shore D hardness of about 45 to 80 and does not include a spring. The valve has a primed position where the formulation can freely flow between the reservoir and the metering chamber, and where the metering valve stem seal member seals the metering chamber from the outside atmosphere. The valve has an actuated position where the opening at the second end of the valve housing is sealed from the reservoir, thereby defining a metered volume of formulation within the metering chamber, and where the metering valve stem includes a flow path allowing the formulation to freely flow between the metering chamber and the outside atmosphere.
The present disclosure also provides inhalers including valves of the embodiments described above where the inhaler further includes a canister having a reservoir and an actuator including an actuator housing configured to enclose at least a portion of the valve and canister.
Some embodiments of the present disclosure provide an inhaler including a canister having a reservoir containing a pressurized formulation of medicament and CO2. The inhaler further includes a metering valve having a valve housing having a first end, a second end exposed to pressurized formulation in the reservoir, and walls defining a metering chamber for receiving the formulation from the reservoir. The metering valve further includes a metering valve stem located within the metering chamber, the metering valve stem having a first end exposed to atmosphere and a second end exposed to pressurized formulation in the reservoir. The metering valve further includes a metering valve stem seal member in contact with the first end of the valve housing, where the metering stem seal has a Shore D hardness of about 45 to 80. The valve has a primed position where the formulation can freely flow between the reservoir and the metering chamber and where the metering valve stem seal member seals the metering chamber from the outside atmosphere. The valve has an actuated position where the opening at the second end of the valve housing is sealed from the reservoir, thereby defining a metered volume of formulation within the metering chamber, and where the metering valve stem includes a flow path allowing the formulation to freely flow between the metering chamber and the outside atmosphere. The canister has a lower canister body, a portion of the lower canister body forming the first end of the valve housing and having an opening configured to accept the metering valve stem. The metering valve stem has a retention feature that is sized so that it cannot pass through the lower canister body opening. The inhaler further includes an actuator including an actuator housing configured to enclose at least a portion of the valve and canister.
Other features and aspects of the present disclosure will become apparent by consideration of the detailed description and accompanying drawings.
Throughout this disclosure, singular forms such as “a,” an,” and “the:” are often used for convenience; singular forms are meant to include the plural unless the singular alone is explicitly specified or is clearly indicated by the context. Numerical ranges, for example “between x and y” or “from x to y”, include the endpoint values of x and y.
Some terms used in this application have special meanings, as defined herein. All other terms will be known to the skilled artisan and are to be afforded the meaning that a person of skill in the art at the time of the invention would have given them.
Elements in this specification that are referred to as “common,” “commonly used,” “conventional,” “typical,” “typically,” and the like, should be understood to be common within the context of the compositions, articles, such as inhalers and pMDIs, and methods of this disclosure; this terminology is not used to mean that these features are present, much less common, in the prior art. Unless otherwise specified, only the Background section of this Application refers to the prior art.
The present disclosure will be described with respect to particular embodiments and with reference to certain drawings, but the invention is not limited thereto. The drawings described are only schematic and are non-limiting. In the drawings, the size of some of the elements can be exaggerated and not drawn to scale for illustrative purposes.
The medicament formulation includes one or more active pharmaceutical ingredients and liquid CO2. The CO2 acts a propellant to propel the formulation from canister 14 into mouthpiece 17 and then to the patient. The formulation can further include one or more excipients. In some embodiments one or more of the active pharmaceutical ingredients and excipients can be dissolved in the liquid CO2. In some embodiments one or more of the active pharmaceutical ingredients and excipients can be dispersed or suspended in the liquid CO2.
The medicament formulation can include any suitable pharmaceutical ingredients or medicinal compositions, e.g., one or more of the medicinal compositions described in U.S. Provisional Application No. 62/962,018, filed Jan. 16, 2020, and entitled MEDICINAL COMPOSITIONS FOR CARBON DIOXIDE BASED METERED DOSE INHALERS.
Turning now to
The canister 14 has a generally cylindrical form with upper housing 34 and lower housing 32 connected by screw thread 36 and sealed by canister seal member 35 to form reservoir 37 for holding the medicament formulation. In some embodiments the screw thread attachment can be replaced by another attachment, such as welding and/or gluing together of the upper housing 34 and lower housing 32. The upper housing 34 includes pressure fill valve 38 to enable filling of the canister 14 with medicament formulation. The lower housing 32 of canister 14 houses a portion of the metering valve 26. The reservoir 37 and metering valve stem 24 share a central longitudinal axis A.
In
In any embodiment, one or more of the sealing members, including canister seal member 35, metering valve stem seal member 59, pressure fill valve stem seal member 50, canister seal member 35, fill valve seal member 43, and a narrowed internal portion 68 (
In
As can be seen in Table 1, a surface area of 8 mm2 for the surface 83 of the abutment 58 for an inhaler having a formulation pressure of 50 bar would provide a Force to Fire of about 40N.
The mid-section 57 and pin section 60 of the metering valve stem 24 are situated within the valve housing 64, which has a metering chamber 69 open at both a first end 53 and a second end 55 of the valve housing. The valve housing 64 has an upper section 67 proximate the second end 55 with internal dimensions that allow for clearance between the pin section 60 of the metering valve stem 24 and the valve housing 64 when the valve 26 is in the primed position, while forming an interference fit between the mid-section 57 and the valve housing 64 when the valve is in the actuated position. In particular, the valve stem 24 transitions from the mid-section 57 to the pin section 60 via a beveled surface 61, and the valve housing upper section 67 includes narrowed internal portion 68 that is dimensioned to form an interference fit with the mid-section 57 of the valve stem. The narrowed internal portion 68 acts as a seal member against the mid-section 57 of the valve stem 24. The metering chamber 69 is disposed between the metering valve stem 24, the valve housing 64, and the metering valve stem seal 59 when in the actuated position (or more particularly, just prior to reaching the actuated position as the valve transitions from primed to actuated position). The size of the metering chamber 69 defines a metered volume that can be about 25 to 100 microliters, 25 to 80 microliters, 40 to 65 microliters, 25 microliters, 50 microliters, or 62 microliters. A lower section 85 of the valve housing 64 has a circumferential step 66, which is formed in an inner face 87 (
An upper or second end 55 of valve housing 64 is urged towards the lower end of the canister 14 by a valve retainer in the form of the valve clamp 74, which has an external screw thread (not shown), and which engages an internal screw thread (also not shown) on the inner surface of the lower housing 32. Any suitable technique or techniques can be utilized to connect the valve clamp 74 and the lower housing 32, e.g., threads, interference fit, adhesives, etc. The valve clamp 74 has a central opening, which permits flow of the medicament formulation within the reservoir 37 and from the reservoir into the upper or second end 55 of the metering chamber 69 of the valve. The valve clamp 74 has an integral bolt 80 to facilitate the tightening of the upper section 67 of the valve housing 64 against the lower housing 32. Alternative techniques of attaching the upper section 67 of the valve housing 64 against the lower end of the canister 14 can be employed for example, an interference fit, welding, and/or gluing.
Alternative flow paths to allow for medicament formulation to flow from the reservoir 37 into the second end 55 of the metering chamber 69 of the valve 26 can be used. For example, a bottle emptier could surround the exterior of the valve housing 64 so that formulation would flow from the base of the canister 14, up the side of the valve housing, and then down into the second end 55 of the valve housing and the metering chamber 69. Alternatively, one or more openings into the metering chamber 69 could be in the sidewall of the upper section 67 of the valve housing 64 at a point above the narrowed internal portion 68.
The opening of the metering valve 26 will now be considered in further detail with reference to
In
Exemplary materials used to form the sealing members, in particular, the metering valve stem seal member 59, include, but are not limited to, polytetrafluoroethylene (PTFE), ethylene tetrafluoroethylene (ETFE), fluorinated ethylene propylene (FEP), polyethylene, particularly high-density polyethylene or linear low-density polyethylene, polyamide, such as nylon, and polypropylene. The metering valve stem seal member 59 can have a hardness of about 45 to 80, 50 to 70, 50 to 60, or 55 to 60 Shore D as measured by ASTM D2240-15 in a standard atmosphere of 23° C. and 50% relative humidity. In embodiments all of the seal members, including for example the canister seal member 35, metering valve stem seal member 59, pressure fill valve stem seal member 50, canister seal member 35, and fill valve seal member 43 are all formed from the same material. In embodiments the metering valve stem seal member 59 is formed from PTFE. In embodiments the narrowed internal portion 68 of the valve housing 64 is high-density polyethylene.
The valve stem 24 preferably has a hardness equal to or greater than the hardness of the metering valve stem seal member 59 and/or the narrowed internal portion 68 of the valve housing. Suitable valve stem hardness values are about greater than 45, greater than 50, greater than 60, greater than 70, greater than 80, 50 to 120, 60 to 110, or 70 to 100 Shore D as measured by ASTM D2240-15 in a standard atmosphere of 23° C. and 50% relative humidity. In some embodiments the hardness of the valve stem 24 is only equal to or greater than the hardness of the metering valve stem seal member 59 and/or the narrowed internal portion 68 of the valve housing 64 in the region of the valve stem that contacts each respective portion. Exemplary materials used to form the valve stem 24 include, but are not limited to, nylon, polyacetal or polyoxymethylene, and stainless steel.
In embodiments the metering valve stem seal member 59 has an internal diameter that is smaller than the external diameter of the portion of the valve stem 24 that interacts with the metering valve stem seal member 59. The internal diameter of the metering valve stem seal member 59 can be about 2.0 to 4.0, 2.0 to 3.0, 2.5 to 2.7, or 2.57 mm. The cross section can be about 1.0 to 3.0, 1.5 to 2.0, 1.7 to 1.9, or 1.78 mm. The cross-section is the width of the metering valve stem seal member 59 in the direction perpendicular to the valve stem 24, for example, the diameter of an O-ring. The external diameter of the portion of the valve stem 24 that interacts with the metering valve stem seal member 59 can be about 2.2 to 4.4, 2.2 to 3.3, 2.7 to 3.0, or 2.8 mm. In some embodiments the difference between the external diameter of the valve stem portion that interacts with the metering valve stem seal member 59 and the internal diameter of the metering valve stem seal member 59 is about 0.2 to 0.4, 0.2 to 0.3, or 0.23 mm. In some embodiments the ratio of the external diameter of the valve stem portion that interacts with the metering valve stem seal member 59 and the internal diameter of the metering valve stem seal member can be about 1.02 to 1.20, 1.05 to 1.15, 1.08 to 1.12, or 1.10
In some embodiments, one or more of the sealing members, with the exception of the metering valve stem seal member 59, in the pMDI 10, including canister seal member 35, pressure fill valve stem seal member 50, canister seal member 35, and fill valve seal member 43, can be formed of a rubber material, such as nitrile rubber, ethylene propylene diene (EPDM) terpolymer, or butyl rubber.
The canister 14 is generally pressurized to between 45 and 80 bar, often to between 50 and 70 bar, and sometimes to between 50 and 60 bar. In general, the canister 14 is configured to withstand a pressure of at least 80 Bar. The pressure is selected so that at least a portion of the CO2 in the canister 14 is present as a liquid. The total amount of formulation is desirably selected so that at least a portion of the CO2 in the canister 14 is present as a liquid after a predetermined number of medicinal doses have been delivered. The predetermined number of doses can be 30 to 200, 60 to 200, 60 to 120, 60, 120, 200, or any other suitable number of doses. The total amount or volume of liquid formulation in the canister 14 can be about 1 to 30 mL, 2 to 25 mL, 5 to 20 mL, or 10 to 20 mL. The total amount of formulation is typically selected to be greater than the product of the predetermined number of doses times the metering volume of the metered valve. This helps to ensure that the amount of each dose remains relatively constant through the life of the inhaler.
The process of filling the canister 14 will now be considered. The liquid medicament formulation is pumped through the pressure fill valve 38 at a pressure of 45 to 80 bar, 45 to 70 bar, 45 to 60 bar, 45 bar, 50 bar, 55 bar, or 60 bar. It flows through the pressure fill valve 38 and into the reservoir 37. It then passes into the second end 55 of the valve housing 64 and into the metering chamber 69. By virtue of the clearance between the pin section 60 of the metering valve stem 24 and the inner wall 90 of the metering chamber 69 of the valve housing 64 the medicament formulation is permitted to pass into the metering chamber. The medicament formulation is prevented from leaving the canister 14 by the seal formed by the metering valve stem seal member 59 when the valve is in its primed, or rest, position.
In use, the patient actuates the inhaler 10 by pressing downwardly on the end cap 40 of the canister 14. This moves the canister 14 into the body 16 of the actuator 12 and presses the canister metering valve stem 24 against the stem socket 22, resulting in the canister metering valve 24 opening and releasing a metered dose of medicament into the expansion chamber 28 within the stem post 20. The expansion chamber 28 delivers the dose into the mouthpiece 17 via the jet orifice 30 and from there the dose passes into the patient's mouth. It should be understood that other modes of actuation, such as breath-actuation, can be used as well and would operate as described with the exception that the force to depress the canister would be provided by the device, for instance by a spring or a motor-driven screw, in response to a triggering event, such as patient inhalation.
In embodiments, the canister 14 is urged from the actuated position to the primed position by the pressurized propellant alone and without the aid of a return spring. In contrast, other pMDIs use a spring in the valve to bias the valve stem towards the primed position. Other inhalers are also known where springs are employed to assist with the sealing elements of the valve. In embodiments, the sealing elements of the valve 26 are not assisted by springs. Thus, the valve 26 can act as a springless valve (i.e., neither stem bias nor seal assist springs used). This can be advantageous as it reduces part count and eliminates components that can be susceptible to drug deposition. Furthermore, inclusion of a spring will increase the force to actuate the valve 26, as the user will need to overcome the force exerted on the valve stem 24 by both formulation and the spring.
It will be appreciated that in the above exemplary embodiment, the metering valve 26 is situated within the canister internal storage volume. It is entirely feasible within the scope of the disclosure that some or all of the metering valve 26 be positioned externally to the storage volume, for example by use of a manifold through which medicament formulation could be delivered to the metering valve.
The metering valve 26 can have any suitable leak rate. In one or more embodiments, the metering valve 26 can have a leak rate of no greater than 725 mg/yr. Further, in one or more embodiments, the metering valve 26 can have a leak rate of no greater than 525 mg/yr.
Some valves include a tank or reservoir seal disposed within a valve housing of the valve. In such valves, the reservoir seal can become displaced from its desired location adjacent to a second end of the valve housing, thereby diminishing the effectiveness of the seal in preventing pressurized formulation from entering the valve between an opening in the housing and a valve stem. This leakage of formulation can affect shot weights of the valve. To overcome this potential displacement of the seal, a spring can be disposed within the housing that engages the seal and holds the seal in place; however, such spring can make operation of the valve more challenging for users as it increases the Force to Fire of the inhaler.
In one or more embodiments of valves described herein, one or more techniques can be utilized to prevent displacement of the reservoir seal during use without the need for a spring. For example,
The metering valve 126 includes a valve housing 164 that has a first end 153, a second end 155 exposed to pressurized formulation in a reservoir (e.g., reservoir 37 of
The metering valve 126 also includes metering valve stem 124 located within the metering chamber 169. The metering valve stem 124 includes a first end 192 exposed to atmosphere and a second end 194 exposed to pressurized formulation in the reservoir. A metering valve stem seal member 159 is in contact with the first end 153 of the valve housing 164 and the metering valve stem 124. A fill channel 177 is disposed in the metering valve stem 124 adjacent to the second end 194 of the stem. The fill channel 177 can be adapted to allow pressurized formulation from the reservoir to pass through the fill channel and into the metering chamber 169 when the fill channel extends between the reservoir and the metering chamber. The fill channel 177 can take any suitable shape or shapes and have any suitable dimensions. The metering valve stem 124 also includes a radial port 162 that extends from an outer surface 179 of the valve stem 124 to a blind hole 156 disposed in the valve stem. The blind hole 156 extends to an opening at the first end 192 of the valve stem 124. The radial port 162 is adapted to allow pressurized formulation disposed within the metering chamber 169 to pass through the radial port and into the blind hole 156 where it can be directed out of the valve 126 and into a jet orifice (i.e., jet orifice 30 of
The metering valve stem 124 can take any suitable shape or shapes and have any suitable dimensions. In one or more embodiments, the metering valve stem 124 has a constant cross-sectional area along a stem axis that extends between the first end 192 and the second end 194 of the valve stem.
The metering valve 126 also includes a reservoir seal 198 disposed adjacent to the second end 155 of the valve housing 164. As used herein, the phrase “adjacent to the second end” means that the reservoir seal 198 is disposed closer to the second end 155 of the valve housing 164 than to the first end 153 of such housing. The reservoir seal 198 can be disposed in any suitable location relative to the second end 155 of the valve housing 164. In the embodiment illustrated in
The reservoir seal 198 can be retained within the housing 164 using any suitable technique or techniques. For example, the reservoir seal 198 can be friction-fit within the housing, adhered to the walls 190 of the housing, mechanically fastened to the housing, etc. In the embodiment illustrated in
In use, the valve 126 has a primed position (
In one or more embodiments, the valve 126 can be urged from the actuated position to the primed position by the pressurized propellant alone and without the aid of a return spring. In contrast, other pMDI use a spring in the valve to bias the valve stem towards the primed position. Other inhalers are also known where springs are employed to assist with the sealing elements of the valve. In one or more embodiments, the sealing elements of the valve 126 are not assisted by springs. Thus, the valve 126 can act as a springless valve (i.e., neither stem bias nor seal assist springs used).
As mentioned herein, the reservoir seal 198 can be disposed in any suitable location relative to the second end 155 of the valve housing 164. For example,
The reservoir seal 298 can be retained on the outer surface 275 of the valve housing 264 using any suitable technique or techniques, e.g., the seal can be friction-fit to the second end 255 of the valve housing. In one or more embodiments, the reservoir seal 298 can be attached to the second end 255 of the valve housing 264 using any suitable attachment technique, e.g., adhesives, mechanical fasteners, etc. In one or more embodiments, the reservoir seal 298 is retained on the outer surface 275 of the valve housing 264 by a valve clamp (e.g., valve clamp 74 of valve 26 of
Canister Filling/Inhaler Preparation—Examples 1 to 5
Formulations were prepared using a refillable two-part 12 g canister (Modern Combat Sports UK) equipped with a filling valve and an outlet valve. The canister was opened and a quantity of liquid containing the non-propellant components of the composition was dispensed into the open canister. The canister halves were attached to each other and the canister was filled through the filling valve with a quantity of carbon dioxide from a cylinder containing carbon dioxide using a 12 g carbon dioxide cylinder charger (Modern Combat Sports, UK). The canister was shaken and then allowed to rest for approximately 15 minutes.
The outlet valve of the filled canister was attached to a sealed manifold equipped with a metering valve as shown in
Through Life Shot Weight—Examples 1 to 5
A filled canister was prepared as described above and connected to an inhaler as described above. Units were tested with an aluminum actuator with a plastic insert having a 0.319 mm spray orifice. The inhaler was actuated with the plume directed towards waste collection. The weight of the inhaler was determined before and after each shot and the shot weight was determined from the difference. Results are reported as the number of shots for the shot weight to reach approximate steady state, the number of shots produced at steady state, and the average and standard deviation of the shot weights at steady state.
Approximately 8.3 g of a 10% (w/w) ethanol in carbon dioxide composition was prepared in a 12 g canister according to the canister filling procedure. The canister was allowed to rest for approximately 15 minutes. The canister was then connected to an inhaler device as described above. The through life shot weights were determined as described above. After eleven shots the inhaler produced 78 shots with an average of 58.3+/−2.5 mg.
Approximately 8.2 g of a 20% (w/w) ethanol in carbon dioxide composition was prepared in a 12 g canister according to the canister filling procedure. The canister was allowed to rest for approximately 15 minutes. The canister was then connected to an inhaler device as described above. The through life shot weights were determined as described above. After three shots the inhaler produced 85 shots with an average of 62.0+/−0.5 mg.
A solution of beclomethasone dipropionate (Teva) in ethanol (99.5% Acros) was prepared by adding 0.23 g beclomethasone to 10 ml ethanol. A 0.87 g aliquot of the beclomethasone dipropionate in ethanol solution and 6.09 g of carbon dioxide were added to a refillable 12 g canister. The canister was shaken and then allowed to rest for approximately 15 minutes. The canister was connected to an inhaler device. The through life shot weights were determined. After three shots the inhaler produced 64 shots with an average of 64.5+/−0.5 mg.
A mixture of micronized fluticasone propionate (Hovione) and ethanol (100% BP/EP Hayman, Essex, UK) was prepared by adding 109 mg of fluticasone propionate to 3.95 g of ethanol. The mixture was stirred to form a homogeneous suspension. A 0.84 g aliquot of the fluticasone propionate and 6.4 g of carbon dioxide were added to a refillable 12 g canister. The canister was shaken and then allowed to rest for approximately 15 minutes. The canister was connected to the inhaler device. The through life shot weights were determined. After two shots the inhaler produced 62 shots with an average of 63.0+/−8.2 mg.
A mixture of micronized salbutamol sulphate (Teva API, Israel) and ethanol (100% BP/EP Hayman, Essex, UK) was prepared by adding 0.133 g of salbutamol sulphate and mixing with 7.89 g of ethanol. The mixture was stirred to form a homogeneous suspension. A 12 g canister was filled with a 0.846 g aliquot of the salbutamol sulphate (SS) in ethanol suspension, 0.862 g ethanol, and 7.142 g of carbon dioxide. The canister was shaken and then allowed to rest for approximately 15 minutes. The canister was connected to an inhaler device. The through life shot weights were determined. After three shots the inhaler produced 70 shots with an average of 64.3+/−2.7 mg.
Canister Filling/Inhaler Preparation—Example 6
Units were prepared using a refillable three-part canister (Kindeva Drug Delivery L.P., St. Paul, Minn.) equipped with a filling valve and a metering valve. The metering valve was a modified Kindeva™ Spraymiser™ valve core assembly consisting of a valve stem, spring, outer and inner seal, fitted into the canister ferrule with a metering tank and held in place with a valve retainer. A diaphragm (ID 0.094″ OD 0.347″) was punched from a 1 mm thick sheet (Hytrel 4556, Du Pont) and used to provide the external seal around the metering valve stem and a seal (ID 0.069″ OD 0.152″) was punched from the same sheet and used to provide an internal seal. Carbon dioxide was dispensed into the canister by attaching the canister halves to each other and filling through the filling valve with a quantity of carbon dioxide from a cylinder containing carbon dioxide using a 12 g carbon dioxide cylinder charger (Modern Combat Sports, UK). The canister was then fired to function using a firing block (4 shots) manufactured by Kindeva Drug Delivery L.P.
Through Life Shot Weight—Example 6
A filled canister was prepared as above. Shot weight was determined using a firing block manufactured by Kindeva Drug Delivery L.P. The inhaler was actuated with the plume directed towards waste collection. The weight of the inhaler was determined before and after each shot, and the shot weight was determined from the difference. The result is reported as the number of shots for the shot weight to reach approximate steady state, the number of shots produced at steady state, and the average and standard deviation of the shot weights at steady state.
A unit was prepared as described above was filled with carbon dioxide. Shot weight testing was completed on the seventh day after filling. After 1 shot, the inhaler produced 73 shots with an average 38.9+/−6.5 mg.
The embodiments described above and illustrated in the figures are presented by way of example only and are not intended as a limitation upon the concepts and principles of the present disclosure. As such, it will be appreciated by one having ordinary skill in the art that various changes in the elements and their configuration and arrangement are possible without departing from the spirit and scope of the present disclosure.
All references and publications cited herein are expressly incorporated herein by reference in their entirety into this disclosure.
Various features and aspects of the present disclosure are set forth in the following claims.
This application claims the benefit of U.S. Provisional Application No. 62/939,093, filed Nov. 22, 2019, and U.S. Provisional Application No. 62/962,018, filed Jan. 16, 2020, the disclosures of which are incorporated by reference herein in their entireties.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US2020/061056 | 11/18/2020 | WO |
Number | Date | Country | |
---|---|---|---|
62939093 | Nov 2019 | US |