Carbon Nanotube Field Emission Micro-focus X-ray Tube

Information

  • Research Project
  • 7627961
  • ApplicationId
    7627961
  • Core Project Number
    R44EB004215
  • Full Project Number
    5R44EB004215-03
  • Serial Number
    4215
  • FOA Number
    PA-07-280
  • Sub Project Id
  • Project Start Date
    7/1/2004 - 20 years ago
  • Project End Date
    5/31/2011 - 13 years ago
  • Program Officer Name
    LOPEZ, HECTOR
  • Budget Start Date
    6/1/2009 - 15 years ago
  • Budget End Date
    5/31/2011 - 13 years ago
  • Fiscal Year
    2009
  • Support Year
    3
  • Suffix
  • Award Notice Date
    6/9/2009 - 15 years ago
Organizations

Carbon Nanotube Field Emission Micro-focus X-ray Tube

DESCRIPTION (provided by applicant): This is a revised Phase II application based on 2006 application and reviewers'comments. In this version, we made significant modifications, including: change the two rings cathode design to three rings structure, adding the synchronization gating control effort, and details of the processing and manufacturing of the carbon nanotube (CNT) cathodes and x-ray tubes. During this period, we achieved dramatic progress on the CNT x-ray technique, which makes us confident the success of this development. Substantial technological growth has spurred a new interest in imaging techniques, and their usage has grown considerably. High-resolution small animal imaging systems have emerged as important new tools for laboratory animal research. X-ray imaging is widely used in diagnostic radiology, biomedical research and a wide variety of industrial applications. We propose to design and construct a portable, low cost micro-focus x-ray tube using the carbon nanotube (CNT) field emission source as the cathode to overcome disadvantages of the hot electron sources in the present x-ray tubes. After Phase II, we expect to be able to develop the CNT field emision micro-focus x-ray tube with the following features: spatial resolution (effective focal spot size): adjustable with 10 5m, 30 5m, and 100 5m three levels;temporal resolution: adjustable, from 0.1 to 100 ms;X-ray tube current: adjustable, up to 2 mA peak current at 100 5m resolution;anode voltage: 40-60 KVp;synchronization gating control;and lifetime: up to 3000 hours. The primary application of the proposed x-ray source is for small animal imaging related medical research and diagonosis. The major research design and methods to achieve the proposed development goals include: 1) Construction of the micro-focus x-ray tube with resolution of 30 5m based on Phase I study;2) Development of the compact CNT field emission 10 5m micro-focus x-ray source meeting key Phase II targets;3) Design and assembly of the compact electronic control system for the x-ray source of Aim 2;4) Construction of a sealed and portable micro-focus x-ray tube with the specifications of Aim 2;and 5) Demonstration of the high resolution dynamic micro-CT imaging.

IC Name
NATIONAL INSTITUTE OF BIOMEDICAL IMAGING AND BIOENGINEERING
  • Activity
    R44
  • Administering IC
    EB
  • Application Type
    5
  • Direct Cost Amount
  • Indirect Cost Amount
  • Total Cost
    321032
  • Sub Project Total Cost
  • ARRA Funded
    False
  • CFDA Code
    389
  • Ed Inst. Type
  • Funding ICs
    NCRR:321031\NIBIB:1\
  • Funding Mechanism
    SBIR-STTR
  • Study Section
    ZRG1
  • Study Section Name
    Special Emphasis Panel
  • Organization Name
    XINTEK, INC.
  • Organization Department
  • Organization DUNS
    603847901
  • Organization City
    RESEARCH TRIANGLE PARK
  • Organization State
    NC
  • Organization Country
    UNITED STATES
  • Organization Zip Code
    27709
  • Organization District
    UNITED STATES