Carboxylic tris-like siloxanyl monomers

Information

  • Patent Application
  • 20080000201
  • Publication Number
    20080000201
  • Date Filed
    June 30, 2006
    18 years ago
  • Date Published
    January 03, 2008
    16 years ago
Abstract
The present invention relates to polymeric compositions useful in the manufacture of biocompatible medical devices. More particularly, the present invention relates to certain carboxylic monomers capable of polymerization to form polymeric compositions having desirable physical characteristics useful in the manufacture of ophthalmic devices. The polymeric compositions comprise polymerized carboxylic hydrophilic siloxanyl monomers.
Description
EXAMPLES

All solvents and reagents are obtained from Sigma-Aldrich, Milwaukee, Wis., and used as is. The monomers 2-hydroxyethyl methacrylate and 1-vinyl-2-pyrrolidone are purified using standard techniques.


Analytical Measurements

ESI-TOF MS: The electrospray (ESI) time of flight (TOF) MS analysis is performed on an Applied Biosystems Mariner instrument. The instrument operated in positive ion mode. The instrument is mass calibrated with a standard solution containing lysine, angiotensinogen, bradykinin (fragment 1-5) and des-Pro bradykinin. This mixture provides a seven-point calibration from 147 to 921 m/z. The applied voltage parameters are optimized from signal obtained from the same standard solution. For exact mass measurements poly(ethylene glycol) (PEG), having a nominal Mn value of 400 Da, is added to the sample of interest and used as an internal mass standard. Two PEG oligomers that bracketed the sample mass of interest are used to calibrate the mass scale. Samples are prepared as 30 μM solutions in isopropanol (IPA) with the addition of 2% by volume saturated NaCl in IPA. Samples are directly infused into the ESI-TOF MS instrument at a rate of 35 μL/min. A sufficient resolving power (6000 RP m/Δm FWHM) is achieved in the analysis to obtain the monoistopic mass for each sample. In each analysis the experimental monoisotopic mass is compared to the theoretical monoisotopic mass as determined from the respective elemental compositions. In each analysis the monoisotopic mass comparison is less than 10 ppm error. It should be noted that uncharged samples have a sodium (Na) atom included in their elemental composition. This Na atom occurs as a necessary charge agent added in the sample preparation procedure. Some samples do not require an added charge agent since they contain a charge from the quaternary nitrogen inherent to their respective structure.


GC: Gas chromatography is performed using a Hewlett Packard HP 6890 Series GC System. Purities are determined by integration of the primary peak and comparison to the normalized chromatograph.


NMR: 1H-NMR characterization is carried out using a 400 MHz Varian spectrometer using standard techniques in the art. Samples are dissolved in chloroform-d (99.8 atom % D), unless otherwise noted. Chemical shifts are determined by assigning the residual chloroform peak at 7.25 ppm. Peak areas and proton ratios are determined by integration of baseline separated peaks. Splitting patterns (s=singlet, d=doublet, t=triplet, q=quartet, m=multiplet, br=broad) and coupling constants (J/Hz) are reported when present and clearly distinguishable.


Mechanical properties and Oxygen Permeability: Modulus and elongation tests are conducted according to ASTM D-1708a, employing an Instron (Model 4502) instrument where the hydrogel film sample is immersed in borate buffered saline; an appropriate size of the film sample is gauge length 22 mm and width 4.75 mm, where the sample further has ends forming a dog bone shape to accommodate gripping of the sample with clamps of the Instron instrument, and a thickness of 200+50 microns.


Oxygen permeability (also referred to as Dk) is determined by the following procedure. Other methods and/or instruments may be used as long as the oxygen permeability values obtained therefrom are equivalent to the described method. The oxygen permeability of silicon-containing hydrogels is measured by the polarographic method (ANSI Z80.20-1998) using an O2 Permeometer Model 201T instrument (Createch, Albany, Calif. USA) having a probe containing a central, circular gold cathode at its end and a silver anode insulated from the cathode. Measurements are taken only on pre-inspected pinhole-free, flat silicon-containing hydrogel film samples of three different center thicknesses ranging from 150 to 600 microns. Center thickness measurements of the film samples may be measured using a Rehder ET-1 electronic thickness gauge. Generally, the film samples have the shape of a circular disk. Measurements are taken with the film sample and probe immersed in a bath containing circulating phosphate buffered saline (PBS) equilibrated at 35° C.+/−0.2°. Prior to immersing the probe and film sample in the PBS bath, the film sample is placed and centered on the cathode premoistened with the equilibrated PBS, ensuring no air bubbles or excess PBS exists between the cathode and the film sample, and the film sample is then secured to the probe with a mounting cap, with the cathode portion of the probe contacting only the film sample. For silicon-containing hydrogel films, it is frequently useful to employ a Teflon polymer membrane, e.g., having a circular disk shape, between the probe cathode and the film sample. In such cases, the Teflon membrane is first placed on the pre-moistened cathode, and then the film sample is placed on the Teflon membrane, ensuring no air bubbles or excess PBS exists beneath the Teflon membrane or film sample. Once measurements are collected, only data with correlation coefficient value (R2) of 0.97 or higher should be entered into the calculation of Dk value. At least two Dk measurements per thickness, and meeting R2 value, are obtained. Using known regression analyses, oxygen permeability (Dk) is calculated from the film samples having at least three different thicknesses. Any film samples hydrated with solutions other than PBS are first soaked in purified water and allowed to equilibrate for at least 24 hours, and then soaked in PHB and allowed to equilibrate for at least 12 hours. The instruments are regularly cleaned and regularly calibrated using RGP standards. Upper and lower limits are established by calculating a +/−8.8% of the Repository values established by William J. Benjamin, et al., The Oxygen Permeability of Reference Materials, Optom Vis Sci 7 (12s): 95 (1997), the disclosure of which is incorporated herein in its entirety:















Material Name
Repository Values
Lower Limit
Upper Limit


















Fluoroperm 30
26.2
24
29


Menicon EX
62.4
56
66


Quantum II
92.9
85
101









Abbreviations
NVP 1-Vinyl-2-pyrrolidone
TRIS Methacryloxypropyltris(trimethylsiloxy)silane
HEMA 2-Hydroxyethyl methacrylate

v-64 2,2′-Azobis(2-methylpropionitrile)


PG 1,3-Propanediol
EGDMA Ethylene glycol dimethacrylate
SA 2-[3-(2H-Benzotriazol-2-yl)-4-hydroxyphenyl]ethyl methacrylate
IMVT 1,4-bis[4-(2-methacryloxyethyl)phenylamino]anthraquinone
Unless otherwise specifically stated or made clear by its usage, all numbers used in the examples should be considered to be modified by the term “about” and to be weight percent.
Example 1

Synthesis of carboxylic siloxanyl monomer. This method is adapted from a literature procedure: Zilkha, A and Golik, U. J. Org. Chem. 1963, 28, 2007-2009. To a stirred solution of itaconic anhydride (4.24 g, 37.8 mmol) in CHCl3 (40 mL) is slowly added 3-aminopropyltris(trimethylsiloxy)silane (15.1 mL, 38.0 mmol) at 0° C. under a nitrogen atmosphere. The cooling bath is removed and the solution allowed to warm to ambient temperature and stirred under a nitrogen atmosphere an additional 15 h. Following removal of solvent at reduced pressure, the viscous, clear oil is analyzed via mass spectrometry to confirm the presence of the proposed structure.







Example 2

Polymerization, processing and properties of films containing carboxylic siloxanyl monomers. Liquid monomer solutions containing carboxylic siloxanyl monomers from example 1 above, along with other additives common to ophthalmic materials (diluent, initiator, etc.) are clamped between silanized glass plates at various thicknesses and polymerized using thermal decomposition of the free-radical generating additive by heating 2 h at 100° C. under a nitrogen atmosphere.


Example 3

Synthesis of carboxylate siloxanyl monomer Using methods well known in the art, the product from example 1 can be converted by treatment with base and optionally ion-exchange to afford carboxylate derivatives as shown below.







Examples 4-11

Polymerization and processing of films containing of prepolymers. Liquid monomer solutions containing carboxylate siloxanyl monomer from example 3 above, along with other monomers and additives common to ophthalmic materials (diluent, initiator, etc.) can be clamped between silanized glass plates at various thicknesses and polymerized using thermal decomposition of the free radical generating additive by heating 2 h at 100° C. under a nitrogen atmosphere. Contemplated formulations are listed in table 1.

















TABLE 1





Ex-
Ex-









am-
am-


ple
ple 3
NVP
HEMA
TRIS
HFIPMA
PG
EGDMA
ν-64























4
10.0
0.0
10.0
50.0
25.0
4.0
0.5
0.5


5
10.0
20.0
0.0
45.0
20.0
4.0
0.5
0.5


6
25.0
25.0
10.0
0.0
35.0
4.0
0.5
0.5


7
10.0
50.0
10.0
25.0
0.0
4.0
0.5
0.5


8
10.0
25.0
10.0
25.0
25.0
0.0
0.5
0.5


9
10.0
25.0
10.0
25.5
25.0
4.0
0.0
0.5


10
35.0
50.0
10.0
0.0
0.0
4.0
0.5
0.5


11
45.0
50.5
0.0
0.0
0.0
4.0
0.0
0.5









Example 12

Films are removed from glass plates and hydrated/extracted in deionized H2O for a minimum of 4 h, transferred to fresh deionized H2O and autoclaved 30 min at 121° C. The cooled films are then analyzed for selected properties of interest in ophthalmic materials as described. Mechanical tests are conducted in borate buffered saline according to ASTM D-1708a, discussed above. The oxygen permeabilities, reported in Dk (or barrer) units, are measured in phosphate buffered saline at 35° C., using acceptable films with three different thicknesses, as discussed above.


Example 13
Polymerization and Processing of Ophthalmic Lenses Containing Carboxylate Siloxanyl Monomer

40 uL aliquots of a soluble, liquid monomer mix containing 13.9 parts by weight of the product from example 3, 23.3 parts TRIS, 41.8 parts NVP, 13.9 parts HEMA, 5 parts PG, 0.5 parts v-64, 1.5 parts SA, and 60 ppm IMVT are sealed between poly(propylene) anterior and posterior contact lens molds under an inert nitrogen atmosphere, transferred to an oven and heated under an inert nitrogen atmosphere 2 h at 100° C. The cooled mold pairs are separated and the dry lens released from the mold, hydrated/extracted twice in deionized H2O for a minimum of 3 min, transferred to and sealed in an autoclave vial containing a buffered saline solution and autoclaved 30 min at 121° C.

Claims
  • 1. A monomer of formula (I):
  • 2. The monomer of claim 1 having a formula selected from the following formulae:
  • 3. A monomer mix useful for making polymerized biomaterials comprising at least one monomer of claim 1 and at least one second monomer.
  • 4. The monomer mix of claim 3, further compromising in addition to the second monomer a hydrophobic monomer and a hydrophilic monomer.
  • 5. The monomer mix of claim 3 wherein the second monomer is selected from the group consisting of unsaturated carboxylic acids; methacrylic acids, acrylic acids; itaconic acid; itaconic acid esters; acrylic substituted alcohols; 2-hydroxyethyl methacrylate, 2-hydroxyethyl acrylate; vinyl lactams; N-vinylpyrrolidone (NVP) N-vinylcaprolactone; acrylamides; methacrylamide, N,N-dimethylacrylamide; methacrylates; ethylene glycol dimethacrylate, methyl methacrylate, allyl methacrylate; hydrophilic vinyl carbonates, hydrophilic vinyl carbamate monomers; hydrophilic oxazolone monomers, 3-methacryloxypropyltris(trimethylsiloxy)silane, ethylene glycol dimethacrylate (EGDMA), allyl methacrylate (AMA) and mixtures thereof.
  • 6. A biomedical device comprising a polymerized monomer mixture of claim 3.
  • 7. A method of making a biomedical device comprising: providing a monomer mixture comprising the monomer of claim 1 and at least a second monomer;subjecting the monomer mixture to polymerizing conditions to provide a polymerized device;extracting the unpolymerized monomers from the polymerized device; andpackaging and sterilizing the polymerized device.
  • 8. The method of claim 7 wherein the step of extracting is performed with non-flammable solvents.
  • 9. The method of claim 7 wherein the step of extracting is performed with water.