The field of invention is cardiopulmonary exercise testing.
Exercise capacity is the best predictor of the future health of patients who suffer coronary artery disease or who have suffered heart failure. These diseases are the leading causes of hospitalization and mortality in the United States. Thus, exercise testing is a basic tool of clinicians, and is widely used. Analysis of expired gas during exercise is commonly known as cardiopulmonary exercise testing (“CPX”) or metabolic exercise testing, and is often referred to as exercise testing with gas analysis. CPX has been considered by many clinicians to be difficult and expensive to perform, and because of this many clinicians have foregone CPX in favor of less accurate tests that merely estimate the measurements made directly by CPX. Such tests typically require the patient to exercise under steady state conditions—that is, a constant work level—for a fixed period of time, at the end of which the patient's heart rate, breathing rate, and oxygen consumption ideally plateau out to constant levels. The constant work level is then increased to a higher constant work level for a fixed time, and the patient's measurements are again expected to plateau out at the end of that time. This process may be repeated several times.
The measurement {dot over (V)}O2 is the patient's oxygen uptake; that is, the rate of oxygen consumption by a patient during an exercise test. This measurement is sometimes referred to in terms of Mets, which are multiples of resting {dot over (V)}O2, assumed to be 3.5 milliliters per kilogram per minute. Peak {dot over (V)}O2, which is the maximum rate of oxygen consumption by a patient during an exercise test, is a good objective measurement of a patient's aerobic exercise capacity, and usually reflects cardiac function. As commonly performed, exercise testing merely estimates peak {dot over (V)}O2 from exercise duration on a treadmill, workload on a stationary bicycle or distance walked. Such estimates may be substantially influenced by factors other than the patient's medical condition, however, such as the degree of patient effort and motivation, the degree of patient familiarity with the test equipment (sometimes referred to as the training effect); the disparity between expected oxygen requirements and actual oxygen uptake widens as heart disease worsens. This gap is filled by anaerobic processes, which result in the production of lactic acid when carbohydrate is metabolized in the absence of oxygen uptake. This leads to errors when {dot over (V)}O2 is estimated by assuming the whole exercise process is fueled by aerobic metabolism.
{dot over (V)}CO2 is the rate of carbon dioxide production by a patient during exercise. {dot over (V)}CO2 relative to {dot over (V)}O2 is influenced by which substrate is metabolized (fat vs. carbohydrate) and whether anaerobic processes and lactic acid production occur. Therefore, {dot over (V)}CO2 cannot be estimated. {dot over (V)}E (minute ventilation) is the volume of air breathed per minute by a patient, which varies proportionally to {dot over (V)}CO2. {dot over (V)}CO2 relative to {dot over (V)}E is influenced by the presence of heart or lung disease. The calculation of {dot over (V)}O2 and {dot over (V)}CO2 by numerical integration of the product of expiratory airflow with O2 and CO2 concentrations over the duration of a breath is taught in the prior art.
Two different sources of error are commonly found in gas analysis equipment: delay time and response time. Delay time is the time taken for the physical transport of a gas sample from the mouth to the gas analyzers. On the other hand, response time, also known as rise time, is intrinsic to a gas analyzer. Response time is the time that elapses between exposure of a gas sample to a gas analyzer and an output signal from the gas analyzer achieving 67% of the full-scale signal that would correspond to the actual concentration of the gas. For example, if a gas sample containing carbon dioxide at a 10% concentration were exposed to a gas analyzer, the response time of that gas analyzer would be the time taken for that gas analyzer to output a signal indicating a 6.7% concentration of carbon dioxide. The errors introduced by the delay time and the response time prevent the accurate time synchronization of O2 and CO2 signals with separately-measured flow signals that do not experience delay time and response time errors, and thus prevent realtime measurement of {dot over (V)}O2 and {dot over (V)}CO2 and real-time calculation of derived parameters that depend on {dot over (V)}O2 and {dot over (V)}CO2, such as {dot over (V)}E/{dot over (V)}O2 and {dot over (V)}E/{dot over (V)}CO2.
Before calibrating a CPX system, it is often desirable to purge it of remnants of test gas or previous reference gas and ensure that ambient air is present in the system. This is essential for calibration, because if the system is not filled with ambient air before calibration, it will not be at a standard baseline state for the initiation of calibration.
Masks for collecting gas during exercise testing are known in the art, and may be used instead of the traditional mouthpiece and noseclip. However, tradeoffs are made between patient comfort during use, ease with which the operator can place the mask on the patient, and security of attachment to the patient. Typically masks which securely attach to the patient during exercise are difficult to put on the patient, and are uncomfortable; such discomfort can distract the patient during CPX and result in submaximal effort, or in early test termination due to patient discomfort.
The present invention is directed toward a method and apparatus for cardiopulmonary exercise testing.
In a first, separate aspect of the invention, a simulated breath, composed of a known volume of calibration gas containing known concentrations of oxygen, carbon dioxide and nitrogen approximating those of exhaled air, is released within a cardiopulmonary exercise testing apparatus at a flow rate and pressure profile similar to an exhaled breath. The cardiopulmonary exercise testing apparatus measures the flow rate and composition of this gas. Those measurements serve as input for a software program that calculates the necessary compensation and calibration factors for gas sensor delay time, gas sensor response time, gas sensor zero offset, gas sensor span adjustment, and flow sensor calibration. The software program uses these compensation and calibration factors to co-align the gas concentration measurement signals and the flow rate signals such that integration of flow and gas concentration signals can be accomplished breath by breath during exercise testing.
In a second, separate aspect of the invention, measurements of {dot over (V)}O2 and {dot over (V)}CO2, exhaled breath flow rate ({dot over (V)}E), heart rate, and oxygen saturation, as well as derived factors of diagnostic importance, are displayed in a series of four charts that organize and present this information for ease of use and interpretation to facilitate diagnosis.
In a third, separate aspect of the invention, a single pump is used to both purge the cardiopulmonary exercise test apparatus of calibration gas before calibration or testing and to draw the sample gas through the gas analyzers during the calibration procedure or patient testing.
In a fourth, separate aspect of the invention, a face mask used to collect a patient's exhaled breath possesses a plurality of pins. Each headstrap contains a hole corresponding to a headstrap pin, and is attached to the face mask by placing the hole over the corresponding headstrap pin. Each headstrap can be adjusted and secured in a single step, and quickly and easily removed from its corresponding headstrap pin.
Detection and analysis of human respiratory gas exchange during exercise, especially when combined with pulse rate and blood oxygen saturation data, provide important diagnostic information. Exercise requires the integrated responses of the cardiovascular, pulmonary, and musculoskeletal systems, which together reflect a patient's overall health. These exercise responses provide a functional assessment of the cardiovascular, pulmonary and metabolic systems which cannot be achieved by any test performed while the patient is at rest.
The present invention is designed to sense, analyze and display cardiopulmonary data gathered breath by breath during an exercise test. One breath is preferably defined as the interval between two successive inspiratory efforts. The measurements collected breath-by-breath during the exercise test include, but are not limited to, oxygen concentration and carbon dioxide concentration in expiratory air, inspiratory and expiratory airflow, heart rate, respiratory rate, and percent blood oxygen saturation.
Before the exercise test begins, a test station 2 is preferably calibrated or verified. The test station 2 is part of an exercise test apparatus 4.
Calibration
The objective of calibration is to zero and scale an oxygen analyzer 6 and a carbon dioxide analyzer 8, determine the transit delay time for a gas sample to travel from the patient's mouth to the oxygen analyzer 6 and the carbon dioxide analyzer 8, and determine the response time (also known as rise time) of the oxygen analyzer 6 and the carbon dioxide analyzer 8. The physical length and diameter of the gas sample tubing, and the pump flow which draws the gas sample, influence the delay time and rise time.
Referring to
A computer 18 is a part of the exercise test apparatus 4, and is attached to the test station 2. The user of the test station 2 may advantageously attach the test station 2 to an existing computer 18 that it already owns, in order to reduce costs. Optionally, however, the computer 18 may be included within the test station 2. The computer 18 is preferably electronically connected to a display 20 and a printer 22. The display 20 may be, for example, a monitor or a liquid crystal display.
The test station 2 includes an analog-to-digital (“A/D”) converter 24 that converts analog signals from various sensors associated with the test station 2 into digital signals for transmission to the computer 18, and that converts digital control signals from the computer 18 into analog signals transmitted to various actuators associated with the test station 2. Preferably, the A/D converter 24 and the computer 18 are interconnected via an external communications cable 26. The computer 18 preferably stores all sensor data transmitted to it until commanded by an operator to delete it.
The operator initiates calibration via the computer 18. The computer 18 transmits a command to the test station 2 to open the inlet valve 28 and close the outlet valve 42. Optionally, the inlet valve 28 is directly connected to the compression bottle 32, and no inlet connection line 34 is used. This command is transmitted through the external communications cable 26 to the A/D converter 24, which translates the command into analog form and sends a signal through a valve command wire 30 to the inlet valve 28, commanding the inlet valve 28 to open. Calibration gas 12 from the container 10 then enters the test station 2 through the pressure regulator 14, the calibration gas inlet hose 16 and the inlet valve 28. The inlet valve 28 is preferably connected to a compression bottle 32 by an inlet connection line 34. Optionally, the inlet valve 28 may be attached directly to the compression bottle 32, thereby eliminating the inlet connection line 34. The compression bottle 32 is filled to a known pressure, which is measured by a pressure sensor 36 located within the compression bottle 32. The temperature of the calibration gas within the compression bottle 32 is measured by a temperature sensor 38 located within the compression bottle 32. Measurements from the pressure sensor 36 and the temperature sensor 38 are transmitted through sensor wires 40 to the A/D converter 24, which in turn transmits those measurements to the computer 18. Because the volume of the compression bottle 32 is known, and the pressure sensor 36 and the temperature sensor 38 measure the final pressure and temperature of the calibration gas 12 within the compression bottle 32, the compressed volume of the gas within the compression bottle 32 can be calculated by the computer 18. Preferably, the compressed volume of calibration gas 12 within the compression bottle ranges from 0.5-1.0 liters; however, any volume may be used as long as its amount is accurately known or measured.
An outlet valve 42 is attached to the compression bottle 32. An outlet valve hose 43 is attached at one end to the outlet valve 42, and at its other end to a calibration port 44. The length and volume of the outlet valve hose 43 is preferably kept as small as possible in order to minimize dead space. The calibration port 44 is attached to a shell 46 of the test station 2. The shell 46 is simply the enclosure which preferably defines the outer surface of the test station 2. While the test station 2 may be open to the environment, it is preferable to enclose it for safety, durability, and attractiveness, among other reasons.
The calibration port 44 is preferably shaped to provide for a pressure fit with an adapter 54. The adapter 54 is attached to and in flow communication with a flow sensor 56. The adapter 54 also possesses a gas analysis outlet port 60. Preferably, the flow sensor 56 is a pneumotachograph constructed as taught by U.S. Pat. No. 4,905,709, which is hereby incorporated by reference. However, any other type of accurate flow sensor may be used to measure flow rate, if desired. Preferably, the flow sensor 56 is connected to the adapter 54 in such a manner as to enable the flow sensor 56 to be easily attached to and removed from the adapter 54, but to keep it securely fastened to the adapter 54 during calibration and exercise testing, such as by a pressure fit. The flow sensor 56 possesses a flow sensor outlet port 58.
A flow sensor outlet hose 62 is attached at one end to the flow sensor outlet port 58 on the flow sensor 56, and at the other end to a first flow sensor outlet hose connector 64. The first flow sensor outlet hose connector 64 mates with a second flow sensor outlet hose connector 66, which is preferably attached to the shell 46 of the test station 2. The first flow sensor outlet hose connector 64 and the second flow sensor outlet hose connector 66 may be any connectors that enable convenient connection and disconnection from one another but have low dead space to prevent expansion and mixing. A pressure transducer hose 68 is located within the test station 2, and is connected at one end to the second flow sensor outlet hose connector 66 and at the other end to a pressure transducer 70. Preferably, the pressure transducer 70 is a differential pressure transducer possessing a port open to ambient air and compares the pressure of the ambient air to the pressure of the gas transmitted to it through the pressure transducer hose 68.
A sampling hose 72 is attached at one end to the gas analysis outlet port 60 on the adapter 54 and at the other end to a first gas analysis outlet hose connector 74. The first gas analysis outlet hose connector 74 mates with a second gas analysis outlet hose connector 16, which is preferably attached to the shell 46 of the test station 2. The first gas analysis outlet hose connector 74 and the second gas analysis outlet hose connector 76 may be any connectors that enables convenient connection and disconnection from one another but have low dead space to prevent expansion and mixing. An internal gas transfer hose 78 is located within the test station 2, and is connected at one end to the second gas analysis outlet hose connector 76 and at the other end to a purge valve 80. The purge valve 80 is located within the test station 2, and is capable of accepting gas flow from the internal gas transfer hose 78 and switching that gas flow to one of two outlets attached to it.
One outlet of the purge valve 80 is a purge outlet 82, and the other outlet of the purge valve 80 is a gas analysis outlet 84. Preferably, a purge valve outlet hose 86 is attached at one end to the purge valve 80 and at the other end to a T-connector 88. Preferably, the first T-connector 88 is further attached to a pump inlet hose 90 and to a gas analyzer outlet hose 92. The pump inlet hose 90 is connected at one end to the T-connector 88 and at the other end to a pump 94. A pump outlet hose 96 is connected at one end to the pump 94, and at the other end to the shell 46 of the test station 2 such that the pump outlet hose 96 vents outside the test station 2.
Several variations of the connections disclosed above will be apparent to those skilled in the art For example, the pump 94 may optionally be placed adjacent to the shell of the test station 2 such that it vents directly outside the test station 2, eliminating the need for the pump outlet hose 96. The first T-connector 88 can optionally be connected directly to the pump 94, eliminating the pump inlet hose 90. Alternately, the purge valve outlet hose 86, the gas analyzer outlet hose 92, and the pump inlet hose 90 may be interconnected by methods or mechanisms other than the T-connector 88, although the T-connector 88 is preferred due to its low cost, ease of use, and positive contribution to maintainability.
One end of a gas analysis inlet hose 98 is attached to the gas analysis outlet 84 of the purge valve 80, and the other end is connected to the carbon dioxide analyzer 8. The carbon dioxide analyzer 8 is connected to the oxygen analyzer 6 by an analyzer connector hose 100. One end of the gas analyzer outlet hose 92 is connected to the oxygen analyzer 6, and the other end is connected to the T-connector 88. The oxygen analyzer 6 is preferably connected to the carbon dioxide analyzer 8 in series in this order. The oxygen analyzer 6 typically offers some resistance to gas flow through it, and thereby results in downstream mixing of gases through discrete breaths that have passed through it. However, in the preferred embodiment, the carbon dioxide analyzer 8 has low resistance and substantially no mixing of gas within. Thus, if the oxygen analyzer 6 offers such resistance or causes downstream mixing of gases from discrete breaths, the carbon dioxide analyzer 8 is preferably placed first in a series arrangement Otherwise, the gases passed on from the oxygen analyzer 6 can be mixed, negating the breath-by-breath analysis desired from the exercise test apparatus 4. Optionally, the oxygen analyzer 6 and the carbon dioxide analyzer 8 may be arranged in parallel, for example, by having the gas analysis inlet hose 98 branch to both analyzers. Such a parallel arrangement, however, requires additional pneumatic hoses, adding to cost, complexity, and size.
While the compression bottle 32 is being filled, the test station 2 is purged. The computer 18 transmits a command to the purge valve 80 to close the gas analysis outlet 84 and open the purge outlet 82. This command is transmitted through the external communications cable 26 to the A/D converter 24, which translates the command into analog form and sends a signal through a purge valve command wire 102 to the purge valve 80. The pump 94 is activated automatically when power is applied to the test station 2, and remains on as long as the test station 2 is on. The pump 94 thus draws in ambient air through the flow sensor 56, pulling it through the adapter 54, the sampling hose 72, the internal gas transfer hose 78, the purge valve 58, the purge valve outlet hose 86, the T-connector 88, and the pump inlet hose 90, into the pump then expelling that ambient air from the test station 2 through the pump outlet hose 72. The pump thereby purges those components with ambient air.
Purging continues for a preset duration that is sufficient to allow for the complete filling of the compression bottle 32 and for complete purging. This purging duration is a function of the flow rate generated by the pump 94, the preset pressure regulator 14, and the volume of the components of the test station 2 that are purged. The preset purging duration is stored in the computer 18. After the preset purging duration is complete, the computer 18 issues a command to the purge valve to close the purge outlet 82 and open the gas analysis outlet 84. This command is transmitted through the external communications cable 26 to the A/D converter 24, which translates the command into analog form and sends a signal through the purge valve command wire 102 to the purge valve 80. The pump 94 remains on. An additional time period, preferably five seconds, is allowed for the carbon dioxide analyzer 8 and the oxygen analyzer 6 to measure and record the concentrations of CO2 and O2 in the ambient air drawn into the carbon dioxide analyzer 8 and the oxygen analyzer 6 during purging. During the last second of that five-second period, the gas concentration transitions caused by purging are typically substantially complete, and the gas concentrations reach a substantially constant plateau. The concentration of O2 and CO2 in ambient air is known. Thus, the output signals from the carbon dioxide analyzer 8 and the oxygen analyzer 6, corresponding to the measured amounts of CO2 and O2, respectively, serve as the baseline signals for establishing the scaling factors and offsets for each analyzer. The analog output signals from the oxygen analyzer 6 and the carbon dioxide analyzer 8 are preferably voltages, the level of which corresponds to a given gas concentration. The output signals from the oxygen analyzer 6 travel through an oxygen analyzer wire 110 to the A/D converter 24, where they are converted into digital form and transmitted through the external communications cable 26 to the computer 18. The signal output from the oxygen analyzer 6 for this is five-second time period is stored by the computer 18, and the plateau value over the last second of that time period is averaged over that one-second time to generate the constant SignalBO2. SignalBO2 is stored in the computer 18. Similarly, the output signals from the carbon dioxide analyzer 8 travel through a carbon dioxide analyzer wire 108 to the A/D converter 24, where they are converted into digital form and transmitted through the external communications cable 26 to the computer 18. The signal output from the carbon dioxide analyzer 8 for this five-second time period is stored by the computer 18, and the plateau value over the last second of that time period is averaged over that one-second time to generate the constant SignalBCO2. SignalBCO2 is stored in the computer 18. More or less time than five seconds may be allotted for these measurements, if desired; however, five seconds is generally more than enough time to allow the gas concentration to stabilize and to calculate SignalBO2 and SignalBCO2.
The computer 18 then issues a command to the outlet valve 42 to open. This command is transmitted through the external communications cable 26 to the A/D converter 24, which translates the command into analog form and sends a signal through an outlet valve command wire 106 to the outlet valve 42. The outlet valve 42 is opened far enough in a short enough time to release the calibration gas 12 from the compression bottle 32 at a flow rate and pressure, over the duration of calibration gas 12 outflow from the compression bottle 32, that are similar to that of an exhaled breath. Indeed, the outflow of calibration gas 12 from the compression bottle 32 may be accompanied by a whooshing sound approximating the sound made by a person exhaling after a deep breath.
When the outlet valve 42 opens, the pressurized calibration gas 12 within the compression bottle 32 rushes out through the outlet valve 42, passing through the outlet valve hose 43, the calibration port 44, and the adapter 54. A portion of the calibration gas 12 entering the adapter 54 is drawn off from the adapter 54 through the gas analysis outlet port 60, due to the suction of the pump 94 which is in flow communication with the gas analysis outlet port 60. A portion of the calibration gas 12 thus travels through the sampling hose 72, the first gas analysis outlet hose connector 74, the internal gas transfer hose 78, the purge valve 80, and the gas analysis inlet hose 98 to the carbon dioxide analyzer 8. The carbon dioxide analyzer 8 measures the amount of CO2 in the calibration gas 12, and transmits that analog measurement to the A/D converter 24 through the carbon dioxide analyzer wire 108. The A/D converter 24 converts that analog signal to a digital signal and transmits it to the computer 18 via the external communications cable 26.
The calibration gas 12 then flows from the carbon dioxide analyzer 8 to the oxygen analyzer 6 via the analyzer connector hose 100. The oxygen analyzer 6 measures the amount of O2 in the calibration gas 12, and transmits that analog measurement to the A/D converter 24 through the oxygen analyzer wire 110. The A/D converter 24 converts that analog signal to a digital signal and transmits it to the computer 18 via the external communications cable 26.
The calibration gas 12 within the oxygen analyzer 6 is then drawn through the gas analyzer outlet hose 92, the T-connector 88, and the pump inlet hose 90 into the pump 94, where it is then expelled from the test station 2 through the pump outlet hose 96.
The flow sensor 56 attached to the adapter 54 is in flow communication with the pressure transducer 70 via the flow sensor outlet port 58, the flow sensor outlet hose 62 and the pressure transducer hose 68. Measurements from the pressure transducer 70 are transmitted to the A/D converter 24 through the pressure transducer signal wire 91. The A/D converter 24 converts that analog signal to a digital signal and transmits it to the computer 18 via the external communications cable 26. The computer 18 then converts the pressure measurement into a flow measurement; the flow rate of the calibration gas 12 is proportional to the difference in pressure between the pressure measured by the pressure transducer 70 and the pressure of the ambient air. The computer 18 then determines the volume of the flow through the flow sensor 56 by integrating the flow rate with respect to time.
The calibration process continues for an additional period of time after the pressure within the compression bottle 32 has reached substantial equilibrium with ambient pressure. Preferably, this additional time period is five seconds. At that time, flow essentially ceases and the gas concentration measurements from the carbon dioxide analyzer 8 and the oxygen analyzer 6 reach substantial equilibrium. During the last second of this period, delay time and rise time notwithstanding, the carbon dioxide analyzer 8 and the oxygen analyzer 6 have typically achieved substantially full response and are substantially accurately sensing the known concentrations of the CO2 and O2, respectively, within the calibration gas 12. During this time period, the oxygen analyzer 6 and the carbon dioxide analyzer 8 measure and record the concentrations of O2 and CO2 within the calibration gas 12. The concentrations of O2 and CO2 within the calibration gas 12 are known. Thus, the output signals from the oxygen analyzer 6 and the carbon dioxide analyzer 8, corresponding to the measured amounts of O2 and CO2, respectively, serve as the response signals for establishing the scaling factors and offsets for each analyzer. The output signals from the oxygen analyzer 6 travel through an oxygen analyzer wire 110 to the A/D converter 24, where they are converted to digital form and transmitted through the external communications cable 26 to the computer 18. The signal output from the oxygen analyzer 6 for this five-second time period is stored by the computer 18, and the plateau value over the last second of that time period is averaged over that one-second time to generate the constant SignalAO2. SignalAO2 is stored in the computer 18. Similarly, the output signals from the carbon dioxide analyzer 8 travel through an carbon dioxide analyzer wire 108 to the A/D converter 24, where they are converted to digital form and transmitted through the external communications cable 26 to the computer 18. The signal output from the carbon dioxide analyzer 8 for this five-second time period is stored by the computer 18, and the plateau value over the last second of that time period is averaged over that one-second time to generate the constant SignalACO2. SignalACO2 is stored in the computer 18. More or less time than five seconds may be allotted for these measurements, if desired; however, five seconds is generally more than enough time to allow for stable measurement and calculation of Signal AO2 and Signal ACO2.
Preferably, the calibration factors are calculated according to a simple linear information. The oxygen calibration factors are calculated by the computer 18 according to the following equations:
SlopeO2=(RefO2−BaselineO2)/(SignalAO2−SignalBO2) (1)
OffsetO2=BaselineO2−SlopeO2*SignalBO2=RefO2−Slope*SignalAO2 (2)
where:
Thus, because RefA and Baseline are known, and SignalAO2 and SignalBO2 have been calculated by the computer 18, SlopeO2 and OffsetO2 may be easily calculated.
Similarly, the carbon dioxide calibration factors is calculated by the computer 18 according to the following equations:
SlopeCO2=(RefCO2−BaselineCO2)/(SignalACO2−SignalBCO2) (3)
OffsetCO2=BaselineCO2−SlopeCO2*SignalBCO2=RefCO2−SlopeCO2*SignalACO2 (4)
where:
The quantities SlopeO2 and SlopeCO2 correspond to the span associated with the oxygen analyzer 6 and the carbon dioxide analyzer 8, respectively. Similarly, the quantities OffsetO2 and OffsetCO2 correspond to the offset of the oxygen analyzer 6 and the carbon dioxide analyzer 8, respectively.
Optionally, other curve fitting techniques may be used to determine the slope and the offset, if desired, for both O2 and CO2, especially if the oxygen analyzer 6 or the carbon dioxide analyzer 8, or both, are nonlinear.
The quantities calculated in Equations (1), (2), (3), and (4) above-SlopeO2, OffsetO2, SlopeCO2, and OffsetCO2—are then used to compensate for gas sensor span and offset, using the following equations:
CmeasuredO2(t)=Csignal/O2(t)*SlopeO2+OffsetO2 (5)
CmeasuredCO2(t)=Csignal/CO2(t)*SlopeCO2+OffsetCO2, (6)
where
As taught by Noguchi et. al., “Breath-by-breath {dot over (V)}CO2 and {dot over (V)}O2 require compensation for transport delay and dynamic response,” J. Applied Physiology, January 1982, p. 79-84, the output signal of a gas analyzer, such as the carbon dioxide analyzer 8 or the oxygen analyzer 6, closely follows first order kinetics in responding to a step change in gas concentration.
Thus, it is necessary to determine D and R so that the inverse Laplace transform can be applied to breath by breath measurements from a patient to generate CcompensatedO2 and CcompensatedCO2 on a breath by breath basis.
As can be seen from
The rise time 206, or R, is determined by measuring the time elapsed from the beginning of the gas signal deviation from baseline (time t2) until the time t3 at which 67% of the full scale gas response has occurred. The 67% level of full concentration 208 is shown in
Measurement of the rise time R and the delay time D is necessary in order to compensate for the measurement error inherent in current gas analyzers exhibiting first-order kinetics. Additionally, calibration factors must be calculated for each gas analyzer in order to accurately convert its output signals to gas concentrations.
The result of calibration is a set of calibration factors stored in the computer 18: R, D, SlopeO2, OffsetO2, SlopeCO2, and OffsetCO2.
After calibration is complete, the computer 9 preferably generates a graph or a tabular chart, or both, on the display 20, allowing the operator to visually check that calibration was successful. If a graph is shown on the display 20, it preferably applies the calibration factors and the inverse Laplace transform disclosed above to the gas analyzer output signals measured during calibration, in a manner that will be disclosed in greater detail below with regard to the patient exercise test. Proper operation is indicated if the leading edges of the O2 concentration and CO2 concentration are lined up with the flow rate signal, and if the leading edges of the O2 concentration and CO2 concentration signals are nearly vertical and free from overshoot.
Preferably, calibration is performed once at the beginning of each testing day. Daily calibration is preferred because it balances the need for accurate testing with the time required for calibration. However, if desired, calibration may be performed more frequently. In addition, calibration may be performed less frequently if gas sensors and analyzers are used which do not drift substantially with time. In that case, the operator may choose to perform a verification step instead of calibration.
Verification
Verification is a process by which the operator of the test station 2 can verify that the rise time R and the delay time D measured during calibration, and the slope and offset calculated during calibration, are correct. Typically, calibration is performed when the test station 2 is first activated on a day in which testing is to be conducted, and verification is periodically performed afterward to ensure that the calibration factors are correct and no drift has occurred. Of course, either calibration or verification can be performed at any time to ensure that the calibration factors are accurate.
Verification proceeds identically to calibration, with the exception that delay time and rise time are not measured, and slope and offset are not calculated. Instead, the delay time and rise time previously measured during calibration and stored in the computer 18, and the slope and offset previously calculated during calibration and stored in the computer 18, are applied to the calibration gas measured during the simulated breath. The operator can then inspect the results on the display 20 to ensure that the integrated function of the exercise test apparatus 4 is operating properly. Proper operation is indicated if the leading edges of the O2 concentration, CO2 concentration, and flow rate signals are lined up, and if the rise times are free from overshoot. If the operator is not satisfied with the results of verification, calibration should be initiated or repeated.
Patient Date Entry
A set of patient data items is entered into the computer 18, which preferably stores them. This patient data set preferably includes the patient's name, date of birth, height, weight, gender and medication usage, as well as the type of work device used or to be used. This data is used to calculate normal reference values including but not limited to maximum heart rate, Peak {dot over (V)}O2, anaerobic threshold, and maximum breathing capacity. Calculation of these values from the patient data set entered in this step is well known in the medical literature. Preferably, the patient data set is stored with the computer 18 and may be used for future tests, eliminating the need to reenter the patient data set for a given patient if that patient is retested in the future. The patient data set preferably may be retroactively edited for any individual test.
Patient Testing
The operator may choose to perform exercise testing alone, or in combination with pre-exercise or post-exercise spirometry testing. Spirometry may also be performed alone if desired. The results of exercise testing, and of pre-exercise and post-exercise spirometry, are preferably stored in the computer 18. Spirometry alone may be performed with a given patient if an exercise test was previously done for that patient without spirometry. The spirometry results are preferably stored in the computer 18 in association with the exercise test results for each tested patient. Spirometry and exercise testing will now be described in detail.
Spirometry
Spirometry data is used to calculate the patient's maximum breathing capacity (MBC) and is an excellent screening test for many pulmonary disorders. By integrating the ability to first make these resting measurements prior to performing an exercise test, then using this data to predict the patient's MBC, then assessing the ventilatory and gas exchange responses of a patient during exercise and comparing those responses to the patient's MBC, CPX is better able to distinguish pulmonary from cardiac causes of exercise limitation as well as make a more comprehensive evaluation of the respiratory system.
During spirometry, the flow sensor 56 need not be attached to the adapter 54. However, as in calibration, the flow sensor outlet hose 62 is attached at one end to the flow sensor outlet port 58 on the flow sensor 56, and at the other end to the first flow sensor outlet hose connector 64. The first flow sensor outlet hose connector 64 is attached to the shell 46 of the test station 2. A pressure transducer hose 68 is located within the test station 2, and is connected at one end to the first flow sensor outlet hose connector 64 and at the other end to a pressure transducer 70.
Initially, the operator zones the pressure transducer 70 by keeping the flow sensor 56 still, for example, by placing the flow sensor 56 on the calibration port 44 or on a surface in a location where the air is still. The flow through the flow sensor 56 while it is at rest is effectively zero. The operator enters a command for zeroing into the computer 18. Analog output from the pressure transducer 70 is transmitted over the pressure transducer signal wire 91 to the A/D converter 24, where it is converted to digital form and transmitted to the computer 18 via the external communications cable 26. The output signal from the pressure transducer 70 during zeroing corresponds to a flow rate of zero. The computer averages that output signal over a short period of time, and the computer 18 then equates the averaged output signal from the pressure transducer 70 with zero flow. The value of that averaged output signal is stored in the computer 18.
The spirometry test then begins. Referring to
The patient then makes a maximum expiratory effort into the adapter 54. That is, the patient, exhales as forcefully as possible for as long as possible into the adapter 54. The two most important variables measured during spirometry are Forced Expiratory Volume in 1 Second (FEV1) and Forced Vital Capacity (FVC), which is the total exhaled breath volume during a maximum expiratory effort. The FEV1 and FVC measurements are well known in the medical literature. As stated above, the computer 18 determines the volume of the flow through the flow sensor 56 by integrating the measured flow rate with respect to time. The patient's maximum breathing capacity (MBC) is then calculated from the FEV1 measurement, based on published equations. Karlman Wasserman et. al., Principles of Exercise Testing and Interpretation at 79 (1984). The FEV1, FVC and MBC for that maximum expiratory effort are then displayed on the display 20.
Preferably, the patient then repeats a maximum expiratory effort until two consistent results are recorded. The display 20 preferably includes an incentive bar or other graphic which demarcates the patient's previous best effort, to encourage the patient to perform maximum expiratory efforts. The computer 18 compares the spirometry results between each trial to ensure that they are within the standards of the American Thoracic Society. American Thoracic Society Board of Directors, Robert O. Crapo et. al., “Standardization of Spirometry: 1994 Update,” Am. J. Respiratory and Critical Care Medicine 152:1107-1136 (1995). Consistent results between trials indicate that the patient has in fact exerted a maximum expiratory effort in each trial. After three or more trials and two consistent results within the standards of the American Thoracic Society, there is no need for the patient to repeat a maximum expiratory effort.
The data for the best trial are preferably displayed on the display 20 as a flow vs. volume curve, and the numerical values for FEV1 and FVC are shown on the graph as well as in tabular form with reference values and predicted values. The values of FEV1, FVC, and MBC are stored in the computer 18.
Spirometry may optionally be omitted if the patient has previously been tested and the patient's previous spirometry data has been retained on the computer 18. Otherwise, omission of spirometry, while allowable within the scope of the present invention, results in the loss of diagnostic information.
Patient Exercise Test
Referring to
Referring to
To secure the face mask 140 to a patient, the face mask 140 is placed over the patient's mouth and nose. The headstrap 144 is then placed such that one of the headstrap pins 142 goes through the hole 146. Optionally, the face mask 140 and headstrap 144 can be designed to interconnect via a plurality of headstrap pins 142 and holes 146. Each end of the headstrap is then brought to the rear of the patient's head. The end of the headstrap 144 possessing the attachment region 152 is pulled through the loop 150 until a snug but comfortable fit is achieved against the patient's head. The end of the headstrap 144 possessing the attachment region 152 is then folded over to come into contact with the headstrap 144, thereby becoming attached to it via the quick-release attachment means associated with the attachment region 152 and the headstrap 144. Thus, the headstrap 144 can be adjusted and secured in a single step. Another advantage of the headstrap 144 is that it can be quickly and easily removed by a patient or the operator simply by lifting the headstrap 144 over the corresponding headstrap pin 142, because the headstrap 144 is held onto the headstrap pin 144 solely due to the tension in the headstrap 144.
Preferably, one headstrap 144 is positioned around the patient's head over the ears and another headstrap 144 is positioned around the patient's head under the ears, providing for a secure fit and minimizing fit difficulties arising from varying head sizes and shapes, and from head or facial hair.
Referring to
An oximeter 112 is preferably non-invasively attached to the patient, and measures blood oxygen saturation. The non-invasive oximeter 112 is known in the art, and may be readily obtained in the market. The oximeter 112 is connected to an oximeter wire 114, through which the oximeter 112 transmits data to the A/D converter 24. The analog blood oxygen saturation data is then converted to digital form and transmitted through the external communications cable 26 to the computer 18. Oxygen saturation is the percent of hemoglobin loaded with oxygen. The actual amount of O2 carried by a volume of blood (that is, O2 content) is dependent on both oxygen saturation and hemoglobin concentration. Low oxygen saturation (SaO2) reflects poor lung function and will contribute to poor exercise capacity.
Preferably, the oximeter 112 is also capable of measuring the pulse of the patient, and transmits pulse data to the A/D converter 24 for transmission to the computer 18 in the same manner as the oximeter data. However, a heart rate monitor 116, such as an electrocardiograph (EKG) or telemetry-type pulse detector may be optionally connected to the test station 2 for more precise measurement of the patient's heart rate. If the heart rate monitor 116 is used, it is connected to the A/D converter 24 via a heart rate monitor wire 118. The analog signals from the heart rate monitor 116 are then converted to digital form and transmitted through the external communications cable 26 to the computer 18. If the analog heart rate monitor 116 outputs waveforms rather than heart rate, the computer 18 converts this information into heart rate by measuring the interval between successive beats.
The operator then begins the exercise test via the computer 18. The computer 18 issues a command to the purge valve 80 to close the purge outlet 82 and open the gas analysis outlet 84. This command is transmitted through the external communications cable 26 to the A/D converter 24, which translates the command into analog form and sends a signal through the purge valve command wire 102 to the purge valve 80. If the purge outlet 82 is already closed and the gas analysis outlet 84 is already open, this condition is maintained. The pump 94 is already on, as described earlier.
After the operator initiates the test, the computer 18 begins collecting data on a breath-by-breath basis from the pressure transducer 70, the carbon dioxide analyzer 8, the oxygen analyzer 6, and the oximeter 112, as well as from the heart rate monitor 116 if used. Preferably, a resting, baseline period of observation is recorded at the beginning of the test.
The patient then begins exercise on an exercise machine (not shown), preferably a treadmill or stationary bicycle. However, other exercises or exercise machines may be used so long as they allow the patient to work continuously at incremental work loads. Optionally, the computer 18 is electronically connected to such an exercise machine, allowing the computer 18 to control its speed and/or monitor the patient's work level.
The patient's exhaled breath passes from the patient's nose and/or mouth through the face mask 140 into the adapter 54. A portion of the patient's breath thus entering the adapter 54 is drawn off from the adapter 54 through the gas analysis outlet port 60, due to the suction of the pump 94 which is in flow communication with the gas analysis outlet port 60. That portion of the patient's exhaled breath thus travels through the sampling hose 72, the first gas analysis outlet hose connector 74, the internal gas transfer hose 78, the purge valve 80, and the gas analysis inlet hose 98 to the carbon dioxide analyzer 8. The carbon dioxide analyzer 8 measures the amount of CO2 in the patient's exhaled breath, and transmits that analog measurement to the A/D converter 24 through the carbon dioxide analyzer wire 108. The A/D converter 24 converts that analog signal to a digital signal and transmits it to the computer 18 via the external communications cable 26.
The patient's breath then flows from the carbon dioxide analyzer 8 to the oxygen analyzer 6 via the analyzer connector hose 100. Preferably, the carbon dioxide analyzer 8 and the oxygen analyzer 6 are arranged in series in that order, but they may be arranged in the opposite order or in parallel if desired, at the penalty of more-complex plumbing within the test station 2. The oxygen analyzer 6 measures the amount of O2 in the patient's exhaled breath, and transmits that analog measurement to the A/D converter 24 through an oxygen analyzer wire 110. The A/D converter 24 converts that analog signal to a digital signal and transmits it to the computer 18 via the external communications cable 26.
The patient's exhaled breath is then drawn from the oxygen analyzer 6 through the gas analyzer outlet hose 92, the T-connector 88, and the pump inlet hose 90 into the pump 94, where it is then expelled from the test station 2 through the pump outlet hose 96.
The flow sensor 56 attached to the adapter 54 is in flow communication with the pressure transducer 70 via the flow sensor outlet port 58, the flow sensor outlet hose 62 and the pressure transducer hose 68. Measurements from the pressure transducer 70 are transmitted to the A/D converter 24 through the pressure transducer signal wire 91. The A/D converter 24 converts that analog signal to a digital signal and transmits it to the computer 18 via the external communications cable 26. The computer 18 then converts the pressure measurement into a flow measurement as described above. Thus, the flow rate and overall volume exhaled during each breath are determined.
While the patient exercises, the oximeter 112 measures the patient's blood oxygen saturation level. The oximeter 112 generates an analog electrical signal corresponding to the measured blood oxygen saturation level, which travels along the pulse oximeter wire 114 to the A/D converter 24. If the oximeter 112 is used to measure the patient's heart rate as well, it also generates an analog electrical signal based on the patient's heart rate, which travels along the oximeter wire 114 to the A/D converter 24. The A/D converter 24 converts that analog signal to a digital signal and transmits it to the computer 18 via the external communications cable 26.
If the patient's pulse rate is measured by a heart rate monitor 116, an analog electrical signal travels through the heart rate monitor wire 118 to the A/D converter 24. The A/D converter 24 converts that analog signal to a digital signal and transmits it to the computer 18 via the external communications cable 26. If the analog heart rate monitor 116 outputs waveforms rather than heart rate, the computer 18 converts this information into heart rate by measuring the interval between successive beats.
The patient's respiratory rate and tidal volume (the volume of each individual breath) are calculated by the computer 18 based on the output signals of the pressure transducer 70. One breath is preferably defined as the interval between two successive inspiratory efforts. The pressure transducer 70 detects an inspiratory effort by noting the time when the gas pressure it measures reverses direction. Preferably, the computer 18 then calculates the time difference between each inspiratory effort and the succeeding one, and converts that time into terms of breaths per minute. Tidal volume is also calculated from the signal output of the pressure transducer 70. As described above, the pressure transducer 70 measures flow based on the pressure difference between ambient air and the pressure experienced by the flow sensor 56. The computer 18 determines the flow for a single breath by converting output from the pressure transducer 70 to flow rate over the time between the two successive inspiratory efforts that define that breath, then integrating that flow rate with respect to time to determine tidal volume.
In a preferred embodiment, during the exercise test, the computer 18 applies the stored calibration factors to the output signals on a breath-by-breath basis from the oxygen analyzer 6 and the carbon dioxide analyzer 8. As described above, Equation (7) is used to determine the concentration of oxygen sensed by the oxygen analyzer 6 and Equation (8) is used to determine the concentration of carbon dioxide sensed by the carbon dioxide analyzer 8. However, the computer 18 can also store the raw breath-by-breath output signals from the oxygen analyzer 6 and the carbon dioxide analyzer 8, and apply the calibration factors to that data after the exercise test has been completed. The computer 18 synchronizes the exhaled air flow signal (computed from the output signals from the pressure transducer 70) with the corresponding instantaneous compensated gas concentrations for O2 and CO2. The computer 18 then multiplies the instantaneous compensated oxygen concentration at time t by the instantaneous flow rate at time t, resulting in a product O2Flow. Similarly, the computer 18 multiplies the instantaneous compensated carbon dioxide concentration at time t by the instantaneous flow rate at time t, resulting in a product CO2Flow. By integrating O2Flow and CO2Flow with respect to time over the duration of the exhaled breath, the computer 18 determines the volume of oxygen consumed and carbon dioxide produced over that breath. By then dividing those volumes by the measured duration of the entire breath cycle, the computer 18 calculates the rate of oxygen consumption and of carbon dioxide production. Appropriate correction factors for temperature and water vapor content are applied, as are known in the literature.
During the test, the computer 18 preferably displays on the display 20 the breath-by-breath measurements and calculations of {dot over (V)}E, {dot over (V)}O2, {dot over (V)}CO2, respiratory rate, heart rate, respiratory exchange ratio (the ratio of {dot over (V)}CO2 to {dot over (V)}O2, also referred to as RER) and SaO2, in tabular form. Further, the computer 18 preferably graphs on the display 20 heart rate relative to {dot over (V)}O2, with the expected maximum {dot over (V)}O2 and heart rate displayed for reference. Other relationships may be graphed during the course of the test, if desired.
The operator terminates the test when the patient has worked to a predetermined heart rate (such as a percentage of the maximum predicted heart rate), the patient exhibits a potentially dangerous change in heart rate, O2 saturation, or other monitored or measured parameters, or the patient stops exercising due to symptoms such as exhaustion, breathlessness, distress, or other reasons. Data recording continues until the operator inputs a prompt to the computer 18 indicating that the test is over. The operator then inputs into the computer 18 the reason for test termination, which is stored in the computer 18. This input is preferably made by selecting from a predetermined list of options provided by the computer 18. This information is preferably included in the analysis of the patient's test results.
Analyzing and Displaying Test Results
In a preferred embodiment, the computer 18 displays selected test data in real time on the display 20. At the completion of the test, the computer 18 displays complete test results on the display 20 and prints them on the printer 22.
With respect to the data gathered on a breath-by-breath basis, the computer 18 preferably utilizes a three-breath rolling average to reduce measurement noise. That is, data taken during each breath is averaged with data from the two preceding breaths in calculating variables based on that data. However, more than three breaths may be averaged, or other methods of noise reduction may be used to reduce measurement noise if desired.
The computer 18 first uses the height, weight, age and gender information input during the earlier process of patient data entry to calculate predicted values for FEV1, FVC, ideal body weight, peak {dot over (V)}O2, anaerobic threshold and maximum heart rate, using well-known published regression equations. The computer 18 then adjusts the predicted peak {dot over (V)}O2 and anaerobic threshold, depending on the type of exercise performed, based on published data. The computer 18 may optionally adjust the predicted peak {dot over (V)}O2 and anaerobic threshold based on the type of exercise machine used, such as a stationary bicycle or a treadmill.
The computer 18 compares the full range of recorded values of {dot over (V)}O2 over the entire exercise test to determine the maximum value of {dot over (V)}O2, referred to as peak {dot over (V)}O2, and stores peak {dot over (V)}O2, as well as the time that peak {dot over (V)}O2 was achieved. The computer 18 then locates the values of {dot over (V)}CO2, heart rate, SaO2, {dot over (V)}E, and tidal volume (the volume of a single breath) measured at the time that peak {dot over (V)}O2 occurred, and stores them. Optionally, work rate at peak {dot over (V)}O2 is stored as well.
After the exercise test, the computer 18 preferably produces a series of four graphs on the display 20. Each graph preferably plots individual data points, one per breath, preferably averaged over a three-breath rolling average as described previously. All of these four graphs may be printed together on one sheet of paper for convenience of use and interpretation.
1. Heart rate v. {dot over (V)}O2 and Stroke Volume vs {dot over (V)}O2
Referring to
2. {dot over (V)}E vs. {dot over (V)}CO2 and SaO2 vs. {dot over (V)}CO2
Referring to
3. {dot over (V)}CO2 v. {dot over (V)}O2, {dot over (V)}E/{dot over (V)}CO2 v. {dot over (V)}O2, and {dot over (V)}E/{dot over (V)}O2 vs. {dot over (V)}O2
Referring to
Anaerobic threshold (AT) is a submaximal indication of cardiovascular fitness that correlates with peak {dot over (V)}O2. It provides complementary diagnostic and prognostic information, and predicts performance in endurance athletes. AT is also an effort-independent parameter. The computer 18 preferably calculates AT with the V-slope algorithm described in William L. Beaver et al., “A new method for detecting anaerobic threshold by gas exchange,” J. Applied Physiology 60:2020-2027 (1986). The point of greatest inflection in the curve is determined; this inflection point 314 corresponds with the work level ({dot over (V)}O2) at which excess CO2 is produced due to the buffering of lactic acid, which is produced in increased quantities as exercise continues. That point of greatest inflection corresponds to the patient's anaerobic threshold. On the ventilatory equivalent graphs, at the anaerobic threshold, the ratio of {dot over (V)}E/{dot over (V)}O2 will start to increase while the {dot over (V)}E/{dot over (V)}CO2 line remains stable or begins to decrease. This transition point should correspond to the V-slope inflection point. The operator may visually determine if both points on this graph are aligned; alignment means there is more confidence in the AT data.
AT is an indicator of the patient's level of physical conditioning. A normal, healthy patient has a ratio of the {dot over (V)}O2 associated with AT the peak {dot over (V)}O2 (the “AT ratio”) of 40-60%. The AT ratio may rise to 70-80% for well-conditioned endurance athletes. A patient with a normal AT ratio, but a low peak {dot over (V)}O2 and a low AT compared to his or her predicted AT, has a cardiovascular or metabolic problem. In contrast, a patient exhibiting low peak {dot over (V)}O2 with a normal AT and a high AT ratio may have a pulmonary problem, pain, or other limiting factors, but likely not left ventricular dysfunction.
4. Heart rate & {dot over (V)}O2 v. time
Referring to
The selection of data and the grouping of relationships between parameters are intimately linked to the logic of the analysis algorithms. Aerobic exercise capacity is objectively measured by {dot over (V)}O2, which is more accurate than estimating it from work duration or intensity. Measuring the anaerobic threshold and relating it to peak {dot over (V)}O2 helps distinguish metabolic and cardiovascular disorders from other factors such as poor motivation, musculoskeletal pain, or lung disease. Lung disease can be distinguished from heart disease by the presence of O2 desaturation with excessive ventilation. Pulmonary disease is indicated by abnormal resting spirometry measurements, abnormal ventilatory and O2 saturation responses during exercise, and limitation of breathing reserve at the end of exercise. Breathing reserve is generally not exhausted in normal individuals or cardiac-limited subjects. All of these relationships are evident on the first graph 180, the second graph 190, the third graph 200, and the fourth graph 210. The detailed numerical data is also available, so that skilled practitioners can draw their own conclusions.
5. Other analysis
The computer 18 records the SaO2 levels monitored during the exercise test by the oximeter 112. The SaO2 measured during the exercise test should not decrease by more than 4% during the exercise test; it if does, lung disease or pulmonary vascular disease may be irresponsibleresponsible. The computer 18 calculates the percentage increase or decrease in SaO2 measured during the exercise test by comparing the SaO2 measurements during the baseline period of the test with the SaO2 measurements at peak {dot over (V)}O2. If SaO2 decreases by more than 4% during the exercise test, the computer 18 indicates this on the display 20 and/or in a report printed on the printer 22 after the test. The SaO2 levels measured during the test are also displayed on the second graph 190, as described above. The operator can visually inspect the second graph 190 to determine if any unusual variations in SaO2 occurred during the test.
The computer 18 also compares the predicted maximum heart rate with the measured maximum heart rate. The difference between them is called the heart rate reserve, which may indicate relative cardiovascular stress.
Preferably, the computer 18 generates a narrative report and optionally prints it on the printer 22, identifying the key variables discussed above and their deviations, if any, from normal, and indicates the implications of these abnormalities.
While the work “patient” has been used in this application, this does not limit the scope of the present invention to a medical setting. The present invention may also be used in health clubs, athletic training programs, military or police screening and training, disability evaluation and other settings where it is desirable to measure or improve physical condition and endurance.
Reference to the A/D converter 24 does not prohibit the use of digital sensors, digital command apparatus, or a set of individual A/D converters instead of or in addition to the single A/D converter. The single A/D converter 24 working with analog sensors and actuators is preferred due to the cost savings and simplicity of using merely one, rather than a plurality.
Reference to electronic wiring in the present invention is made for clarity of description, and does not prohibit wireless connections between the electronic parts disclosed herein. Wires are preferred due to their cost savings and simplicity at the present time.
A preferred method for measuring and analyzing exhaled breath for diagnosis of cardiopulmonary disease, and many of its attendant advantages, has thus been disclosed. It will be apparent, however, that various changes may be made in the steps of this method and their arrangement without departing from the spirit and scope of the invention, the steps hereinbefore described being merely a preferred or exemplary embodiment thereof. Therefore, the invention is not to be restricted or limited except in accordance with the following claims and their legal equivalents.
This is a divisional patent application of co-pending patent application Ser. No. 09/322,320 filed May 28, 1999.
Number | Name | Date | Kind |
---|---|---|---|
4368740 | Binder | Jan 1983 | A |
4463764 | Anderson et al. | Aug 1984 | A |
4909259 | Tehrani | Mar 1990 | A |
4930519 | Anderson et al. | Jun 1990 | A |
4966141 | Bacaner et al. | Oct 1990 | A |
5435315 | McPhee et al. | Jul 1995 | A |
5462504 | Trulaske et al. | Oct 1995 | A |
5474090 | Begun et al. | Dec 1995 | A |
Number | Date | Country | |
---|---|---|---|
Parent | 09322320 | May 1999 | US |
Child | 09686315 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 09686315 | Oct 2000 | US |
Child | 10969792 | US |