Generally speaking, the invention relates to medical equipment, and more particularly, to carts and power supplies suitable for medical equipment.
Many pieces of medical equipment are designed to be used with multiple patients. This is particularly true of expensive electronic diagnostic equipment, like ultrasound probes. While disposable covers and other kinds of disposable barriers are often used to minimize some types of equipment contamination where and when possible, equipment still gets contaminated, especially if it must enter a body cavity in normal use.
Procedures for decontaminating equipment vary widely according to the type of equipment and the types of harsh exposure that the equipment can sustain without damage. For example, handheld metal medical and surgical tools are often autoclaved for high temperature sterilization. Medical devices that include plastic components, or other components that cannot withstand high temperatures, are often chemically disinfected or, in some cases, disinfected by exposure to UV light.
The TROPHON® 2 disinfection apparatus (Nanosonics Limited, Sydney, Australia) is one example of a disinfection apparatus that is particularly adapted for surface, transvaginal, and transrectal ultrasound probes. Within a generally rectilinear cabinet, the device uses a chemical disinfectant mist driven by ultrasonic vibration to achieve disinfection.
Many pieces of medical equipment are portable. Many pieces of decontamination equipment are not. This means that in most cases, contaminated equipment must be brought to the decontamination equipment. This can cause serious inconvenience for medical staff and an impediment to workflow. If the decontamination equipment can be moved, the means for doing so are often imperfect, and do not provide for all of the equipment's needs.
One of those needs is power. When working with many pieces of medical equipment and disinfection equipment, it is desirable to provide a continuous source of power. If a piece of equipment is turned off, it may be necessary to subject it to time-consuming start-up or recalibration procedures. Most pieces of medical equipment are supplied with alternating-current (AC) power from traditional wall outlets. Uninterruptable power supplies (UPS), which use batteries to store energy and circuitry to deliver that energy as high-voltage AC power, are also becoming more common.
Most UPS units are ill-suited for sensitive medical equipment. For example, most lower-end UPS units on the market provide a stepped or square-wave AC voltage waveform, instead of the pure sinusoidal waveform provided by the typical power grid. While some equipment can function using square-wave AC, square-wave UPS systems can also cause unreliability and equipment failure. However, there are very few UPS systems that are particularly adapted for medical equipment and can also facilitate equipment portability.
One aspect of the invention relates to an equipment cart for medical and disinfection equipment. The equipment cart is designed to mount disinfection equipment in a position near the base of the cart and may include structure to fix the disinfection equipment in place. Casters are provided on the underside of the base to allow for movement of the equipment cart. A telescoping support post supports a work surface at a position above the base and disinfection equipment. The work surface may provide a fully-equipped disinfection workstation with a storage drawer and holders for disinfecting wipes, gloves, and pieces of equipment. Various locking mechanisms may be present, e.g., to lock the storage drawer, and to lock the piece of disinfection equipment to the base of the equipment cart. The equipment cart may have an electrical system to supply power to the disinfection equipment and other peripherals.
Another aspect of the invention relates to an electrical system for a medical equipment cart. The electrical system routes power from a high-voltage AC power source to the equipment when a high-voltage AC power source is available. The electrical system also includes a battery and battery charging circuit that are charged using the AC power source when it is available. When the AC power source is not available, the electrical system uses power from the battery to generate a pure AC sine wave, which is then stepped up to high voltage. The pure AC sine wave is generated by generating a first voltage signal and a second voltage signal and sending both voltage signals through a comparator. The first and second voltage signals may be, e.g., triangular waves, with the first signal being a higher-frequency “fast” wave and the second signal being a lower-frequency “slow” wave. The output of the comparator is further processed to generate the pure AC sine wave. For example, a half-wave rectifier and a pair of flip-flops may be used to generate the pure AC sine wave from the output of the comparator.
Other aspects, features, and advantages of the invention will be set forth in the description that follows.
The invention will be described with respect to the following drawing figures, in which like numerals represent like features throughout the description, and in which:
The equipment cart 10 is intended to serve as a mobile disinfection workstation, with a base 14 that includes casters 16 for movement. In the illustrated embodiment, there are four casters 16, positioned in a rectangular layout supporting the base 14, although any number of casters 16 may be used, so long as that number of casters 16 provides for a stable, moveable base 14. The front casters 16 each have a standard frictional braking mechanism, the actuation levers 18 for which can be seen in
The base 14 includes a lower shelf 20, on which the disinfection equipment 12 rests. A single telescoping support post 22 rises from the base 14 to provide support for the other elements of the equipment cart. At its upper extent, a work surface 24 is cantilevered from the support post 22. A drawer 26 lies under the work surface 24.
One particular advantage of the equipment cart 10 is that by mounting the disinfection equipment on a low shelf, like the lower shelf 20, the weight of the disinfection equipment 12 is less likely to tip the equipment cart 10. Low mounting also means that the equipment cart 10 can include a workspace, like the work surface 24, at the appropriate height for a user.
As shown in the view of
The orientation of the disinfection equipment 12 on the cart, and the manner in which it opens and is used, may affect the arrangement of the equipment cart 10 and its base 14 and lower shelf 20. In this embodiment, the disinfection equipment 12 has the form of a cabinet with a hinged door 13 that swings open outwardly. Thus, the height of the berm 32 at the front of the base 14 is sufficient to help retain the disinfection equipment 12, but not so high as to obstruct the movement of the door 32.
With the disinfection equipment 12 exploded away, other details of the equipment cart 10 can be seen in the view of
Thus, fully equipped, the equipment cart 10 may serve as a workstation that has all necessary equipment to remove gross soil and contamination from a piece of medical equipment before it is placed in the disinfecting equipment. A laptop and printer may be installed on the work surface 24 and the shelf-holder 44 in order to keep disinfection or other related compliance records.
This particular configuration of the equipment cart 10 is not the only possible configuration.
Both equipment carts 10, 50 include locks and security measures. Chemical disinfection may involve chemicals that are toxic, corrosive, strong oxidizers, or are otherwise hazardous. For example, chemicals like 35% hydrogen peroxide are common. Because of this, it is helpful if the equipment cart 10, 50 has some locks. For example, the drawer 26 may be locked, either with a key lock mechanism, or with a proximity locking mechanism, like a radio-frequency identification (RFID) locking mechanism. An RFID locking mechanism uses a low-power radio-frequency transmitter to interrogate a nearby RF element, which may be either powered or unpowered. If the nearby element transmits the correct identifier, the locking mechanism unlocks.
In addition to securing peripherals and chemicals within the drawer 26, locking mechanisms may be present elsewhere as well.
The plate 28 and its openings 30 take advantage of an existing cant to the feet 46 of the disinfection equipment 12. As those of skill in the art will realize, many kinds of cooperating engaging features may be used to lock a structure such as the plate 28 to a piece of disinfection equipment.
The equipment carts 10, 50 have their own onboard electrical systems. As can be seen in
The main circuit 100 provides AC power to the electrical outlets 54 with a pure AC sine wave, even when it is not connected to internal power. To do this, it includes both high-voltage portions and low-voltage portions. (While the definition of “high voltage” varies according to the authority one consults, for purposes of this description, the term will refer to voltages over 50V.) Much of the high-voltage portion 102 of the circuit 100 is shown in
The main circuit 100 illustrated in
In the main circuit 100 of
The switch S1 is connected to a relay U8.1 that connects either to the AC power, indicated as U2 in
Power from the AC power U2 in
As those of skill in the art will realize, a full diagram of the battery charge portion 106 is included in
As was noted briefly above, when the circuit 100 is not drawing power from AC mains, it draws from the battery V1 and modulates that power into a pure AC sine wave within the low-voltage signal generation portion 104 of the circuit 100. The low-voltage pure AC sine wave is then stepped up from low voltage to high voltage by the transformer T2.
Much of the low-voltage signal generation portion 104 is shown in
More specifically, the two waveforms of different characteristics are triangular or sawtooth-type waveforms in this embodiment. The first of the two sawtooth-type waveforms is generated by an NE555P timer IC U3. The connection of the 555 timer IC U3 with the resistors R13, R14 and capacitor C12 places the 555 timer IC in an astable configuration, allowing it to act as an oscillator. The voltage across the capacitor C12 is a triangular or sawtooth waveform in this configuration, and that waveform is sent to the noninverting input of an LM741CN op amp U.
A broader, “slower” triangular waveform is generated by a CD4047 multivibrator IC U1 in astable free-running operating mode. This output is connected to the inverting input of the LM741CN op amp U.
As was described briefly above, the output from the op amp U is first sent to two diodes D5, D8 in a half-wave rectifier configuration. The output of those diodes is sent to two IRF3205 flip-flops M1, M2, shown in
Unless otherwise noted, all electronic components in the circuit 100 are manufactured by, or can be obtained from, Texas Instruments, Inc. (Dallas, Tex., United States). As those of skill in the art will understand, the topology and components shown in
While the invention has been described with respect to certain embodiments, the description is intended to be exemplary, rather than limiting. Modifications and changes may be made within the scope of the invention, which is defined by the appended claims.
This application claims priority to, and the benefit of, U.S. Provisional Patent Application No. 63/139,603, filed Jan. 20, 2021. The contents of that application are incorporated by reference herein in their entirety.
Number | Date | Country | |
---|---|---|---|
63139603 | Jan 2021 | US |