This application relates to a device and method for monitoring pressure in a body cavity.
Traditionally, physicians relied on visual cues or physical examination to detect increase in intra-abdominal pressure (IAP). More recently Dr. Kirkpatrick and colleagues, in an article “Is Clinical Examination an Accurate Indicator of Raised Intra-Abdominal Pressure in Critically Injured Patients,” CJS, June 2000, 43, No. 3, 207-211, showed that IAP measured through the patient's bladder was significantly more accurate than physical examination. That is, it was demonstrated that the clinical abdominal examination was insensitive and inaccurate when compared with urinary bladder pressure measurements.
Various tools for measuring IAP have been developed over the years. Many researchers have documented IAP measurements through almost every natural or manmade orifice in the body. Earlier crude forms of measuring IAP used bladder catheters, nasogastric tubes, and rectal tubes attached to a manometer. The nasogastric or the rectal route was better suited in rare cases of bladder rupture or situations where bladder catheters were contraindicated. However, due to local interferences, the nasogastric and the rectal tube measurements were neither reproducible nor logical as were the bladder catheters.
Thus, measuring of IAP through the bladder became more suitable. In 1989 Iberti and colleagues in an article entitled, “Determination of Intra-abdominal Pressure Using a Transurethral Bladder Catheter: Clinical Validation of the Technique,” Anesthesiology, January 1989, 70(1), 47-50, validated the correlation of IAP using a catheter inserted in the bladder. Their study was key in using bladder pressure as the gold standard for measuring IAP. In 1995, Kron and colleagues published a study in “The Measurement of Intra-Abdominal Pressure as a Criterion for Abdominal Re-exploration, 1984 Ann Surg., 199, 28-30, comparing catheters in various body locations for measuring IAP. They measured IAP from the stomach using a nasogastric tube, from the rectum using a modified rectal tube, from the bladder using a modified bladder catheter, and direct abdominal pressure using a laparoscopic insufflator needle. They found that the bladder catheter had the best measurement of IAP and that the gastric and the rectal catheter measurements were less reliable due to dependence on the position of the catheter. Thus, clinicians generally agreed that the urinary bladder is the best-suited location for measurement of IAP.
The need for measuring IAP has become more important as physicians increasingly realized that organ failure and death were directly related to increase in IAP in certain high-risk patients. High abdominal pressure has been found to cause a decrease in function of the intestines, liver and blood vessels resulting in adverse consequences for the patients. Consequently, accurate measurement of IAP can help decrease patient morbidity and mortality. It has also been more recently discovered that pediatric and neonate population may also have need for IAP measurement to determine specific conditions.
Currently, there are few products available on the market to measure the IAP through the bladder. One device, the Bard IAP device, has a “valve clamp” which diverts urine from the main catheter drainage channel to measure IAP via converting hydrostatic pressure to a readable pressure gauge. This mechanism of IAP measurements is archaic and does not provide continuous pressure measurement when used with the standard 2-channel bladder drainage catheter. Two other manufacturers, Holtech and ConvaTec, also use a column of urine by connecting their kit to an existing bladder catheter. Their systems are cumbersome and the IAP readings are also not continuous. Biometrix has developed an IAP monitoring device which like other manufacturers relies on tapping into the main bladder drainage catheter, using a valve to measure the hydrostatic pressure. In 2008 Sugrue and colleagues, in an article “Prospective Study of Intra-Abdominal Hypertension and Renal Function after Laparotomy, British Journal of Surgery, 1999, 82, 235-238, suggested the use of 3-channel bladder drainage catheter so that the smaller channel, which was used for bladder irrigation, could be used to attach a pressure-monitoring device. The use of an extra channel made it possible to have continuous bladder drainage while measuring the bladder pressure. However, this bladder catheter did not provide a continuous pressure read because intermittently the operator needed to add 50 ml of water or saline to the bladder to record the IAP pressure. Thus, the pressure reading at best was intermittent since pressure readings were not performed when fluid was being added to the bladder. Consequently, although this was a step toward increasing the amount of pressure readings/recordings, it still was unable to conduct continuous pressure monitoring. Furthermore, it was still the same cumbersome IAP device set up which required a skilled person to add water before each IAP reading. Control of the amount of water added is critical since adding too much water to the bladder can falsely increase the pressure readings and also increase infection risk, thus further complicating the use.
It has also been recognized that most patients that have a need for measurement of IAP also need to have continuous drainage of the urinary bladder and thus devices need to account for this process.
Consequently, current devices placed in the bladder for measuring pressure require a continuous water column to maintain pressure readings. Thus, they fail to measure IAP continuously but only measure pressure intermittently. They also all rely on tapping into an existing bladder drainage catheter, which adds complications. Furthermore, they do not reduce the complexity of the procedure since they require constant retrograde insertion of a relatively large amount of fluid into the bladder, e.g., 50 cc, which increases the ICU workload. Still further, these devices increase the risk of complications and infections associated with fluid injection into the bladder. Fluid injection is also complicated since it needs to be closely monitored since too much fluid in the bladder can give false elevation of IAP readings, causing clinicians to take unnecessary steps in response to what is mistakenly believed is excess IAP.
It would therefore be advantageous to provide a device insertable into the bladder that accurately measures abdominal pressure without requiring adding water to the bladder to obtain such pressure readings. Such device would advantageously avoid the complications and risks associated with such fluid insertion. Furthermore, it would be advantageous if such device could continuously measure bladder pressure without interruption. This would advantageously enable a constant monitoring of IAP so critical time periods are not missed. It would further be advantageous to provide a device that improves the accuracy of the pressure reading in the bladder to more accurately determine IAP so necessary steps can be taken to address IAP only when warranted. Still further, it would be advantageous if such device could satisfy the foregoing needs and provide these enumerated advantages while being simple to use so that so that any of clinical staff with basic knowledge of bladder catheter insertion will be able to insert the device without relying on specially trained staff members. It would also be advantageous to provide such devices with these advantages for insertion into other body cavities for accurately measuring pressure within the cavity without the need for injecting fluid into the cavity.
The present invention overcomes the deficiencies and disadvantages of the prior art. The present invention advantageously provides a catheter insertable into the cavity of the patient to determine pressure without requiring insertion of water or other fluid into the body cavity. The present invention provides catheters insertable into various regions of the patient such as the bladder to measure intra-abdominal pressure or maternal uterine contraction pressure or the uterine cavity to measure intrauterine pressure, the abdominal cavity, etc. The catheters can be used for example in rectal, abdominal, esophageal, cardiac, etc. procedures. The catheters of the present invention utilize a gas-charged chamber to measure pressure across a large surface area, and thus, accurately determine pressure, and enable pressure to be measured continuously without interruptions to add water to the cavity.
In some embodiments, an outer fluid filled balloon provides a fluid transmission medium for an inner pressure sensing balloon.
Some embodiments of the catheter of the present invention utilize a stabilizing balloon to help retain the catheter in the bladder during the procedure.
In accordance with one aspect of the present invention, a catheter is provided which is insertable into a patient for monitoring pressure. The catheter includes a first lumen having a wall and at least one side opening in the wall and an expandable outer balloon at a distal portion of the catheter. The outer balloon has a first outer wall and receives fluid to move from a first condition to a more expanded condition, and expands radially outwardly with respect to the catheter. An expandable inner balloon is positioned within the first lumen of the catheter and has a second outer wall and an elongated portion extending proximally through the first lumen, and has a gas containing chamber to monitor pressure within the patient. The outer balloon has a circumferential area greater than a circumferential area of the inner balloon wherein in response to pressure exerted on the first outer wall of the expanded outer balloon fluid within the outer balloon enters the at least one opening in the wall of the lumen to exert a pressure on the second outer wall of the expanded inner balloon to deform the inner balloon and compress the gas within the inner balloon. A pressure sensor communicates with the gas containing chamber for measuring pressure based on compression of gas caused by deformation of the expanded inner balloon resulting from deformation of the expanded outer balloon.
In some embodiments, the outer balloon is inflated via a second lumen independent of the first lumen. In some embodiments, the inner balloon in the expanded position remains within the confines of the first lumen.
In some embodiments, a chamber is provided containing a plurality of openings communicating with the interior of the outer balloon.
In some embodiments, the catheter includes an additional lumen and a stabilizing balloon, the additional lumen communicating with the stabilizing balloon to inflate the stabilizing balloon to stabilize the position of the catheter. The stabilizing balloon can be positioned proximal of the outer balloon.
In some embodiments, the inner and outer balloons have a coating to increase impermeability.
In some embodiments, the pressure sensor is contained within a hub and the hub includes an elongated member extending distally therefrom, and connection of the hub to a first port of the catheter automatically inserts the elongated member into the catheter to advance air into the inner balloon to expand the inner balloon.
In some embodiments, the first lumen is not vented to atmosphere when the pressure sensor is connected to the catheter and advances gas to expand the inner balloon.
In some embodiments, the gas within the inner balloon and/lumen is air to provide an air containing chamber.
In some embodiments, an elongated member is positioned within the tubular portion of the inner balloon to decrease the volume of gas within the tubular portion.
In some embodiments, a third balloon is positioned within the outer balloon, the third balloon being less compliant than the outer balloon and forming an inner liner of the outer balloon to maintain an expanded condition of the outer balloon.
In accordance with another aspect of the present invention, a catheter insertable into a patient for monitoring pressure within a body cavity without insertion of fluid into the cavity is provided, the catheter including a wall having at least one side opening and an expandable outer balloon at a distal portion of the catheter having a first outer wall and movable from a first condition to a more expanded condition. An inner balloon is movable to a more expanded condition, the inner balloon having a second outer wall and a gas containing chamber. The second outer wall of the inner balloon is radially spaced from the first outer wall of the outer balloon, the outer balloon acting as a medium for transfer of fluid to a second outer wall of the inner balloon to deform the inner balloon for monitoring fluid pressure. The inner balloon has an elongated portion extending proximally through a lumen of the catheter. In response to pressure exerted on the first outer wall of the expanded outer balloon, fluid within the outer balloon enters the at least one side opening in the wall of the catheter to exert a pressure on the second outer wall of the expanded inner balloon to deform the inner balloon and compress the gas within the inner balloon to provide a pressure measurement. A pressure sensor communicates with the gas containing chamber of thinner balloon for measuring pressure based on compression of gas caused by deformation of the expanded inner balloon resulting from deformation of the expanded outer balloon, the pressure sensor measuring pressure at multiple times during a procedure without injecting fluid within the body cavity as the outer balloon provides the fluid transfer medium.
In some embodiments, the elongated portion of the inner balloon along with an enlarged portion of the inner balloon forms the gas chamber to monitor pressure within the patient. In some embodiments, the second outer wall of the inner balloon does not expand outside the lumen of the catheter when the inner balloon is in the expanded condition.
In accordance with another aspect of the present invention, a method for measuring pressure within a body cavity without insertion of fluid is provided including the steps of:
The method can include the step of transmitting the pressure readings to an external monitor.
In some embodiments, deformation of the outer balloon is in response to pressure exerted on an outer wall of the expanded outer balloon and upon such deformation, gas within the outer balloon enters one or more openings in the catheter to communicate with the outer wall of the expanded inner balloon to exert a pressure on and deform the inner balloon and compress the gas within the inner balloon.
The method may further comprise the step of connecting to the catheter a hub containing a pressure transducer to automatically advance gas into the inner balloon to expand the inner balloon. In some embodiments, the step of connecting the hub automatically connects a temperature sensor to a connector within the hub.
In accordance with some aspects of the present invention, catheters are insertable into the bladder and utilized for measuring intra-abdominal pressure. In some such embodiments, the gas containing chamber monitors pressure within the bladder to thereby monitor pressure within an abdomen of the patient. In some embodiments, the pressure transducer measures average pressure continuously throughout insertion of the catheter within the urethra without requiring infusion of water into the bladder.
In some embodiments, a second lumen communicates with the bladder to remove fluid from the bladder. In some embodiments, the second lumen has a side opening distal of the inner and outer balloons; in other embodiments the side opening is proximal of the inner and outer balloons. The catheter can include a third lumen communicating with the outer balloon to expand the outer balloon.
In some embodiments, the catheter has a fourth lumen and a temperature sensor positioned within the fourth lumen to measure core body temperature. A wire can extend from the temperature sensor through the fourth lumen and external of the catheter into the hub connected to the catheter. The hub can have a first opening to receive a connector of the wire to automatically connect the temperature sensor to a cable extendable from the hub and connectable to an external temperature monitor.
In some embodiments, connection of the pressure sensor to the catheter a) automatically connects the temperature sensor to a temperature monitor cable; and b) automatically advances air through the first lumen to expand the inner balloon.
In accordance with another aspect of the present invention, a method for measuring intra-abdominal pressure is provided comprising the steps of:
The method can further include the step of draining the bladder through the second lumen of the catheter. In some embodiments, the step of obtaining pressure readings obtains average pressure.
In accordance with another aspect of the present invention, a multi-lumen catheter for monitoring intra-abdominal pressure is provided. The catheter includes an elongated body configured and dimensioned for insertion into a bladder of a patient, a first lumen, a second lumen, and a third lumen, the lumens being independent. A first balloon is positioned at a distal portion and the first lumen communicates with the first balloon. The second lumen communicates with the bladder to remove fluid from the bladder. The first balloon and first lumen are filled with a gas to form a gas filled fully closed chamber to monitor pressure within the bladder to thereby monitor pressure within an abdomen of the patient. A pressure sensor measures pressure within the bladder based on deformation of the first balloon in response to pressure within the bladder exerted on an outer wall of the balloon, the pressure sensor measuring bladder pressure continuously and communicating with an external monitor to visually display pressure readings, the sensor providing continuous pressure measurements throughout its duration of insertion without requiring infusion of water into the bladder.
In accordance with another aspect of the present invention, a system for monitoring intra-abdominal pressure is provided comprising a catheter having an elongated body configured and dimensioned for insertion into the bladder of a patient, a first lumen, a second lumen, a third lumen, and a first balloon at a distal portion. The first lumen communicates with the first balloon and the second lumen communicates with the bladder to remove fluid from the bladder. The first balloon and first lumen are filled with a gas to form a gas filled fully closed chamber to monitor pressure within the bladder to thereby monitor pressure within an abdomen of the patient. A pressure sensor measures bladder pressure continuously and communicates with an external monitor to visually display pressure readings, the sensor providing continuous pressure measurements during its insertion without requiring infusion of water into the bladder. An indicator indicates if the measured pressure exceeds a threshold value.
The indicator can be a visual and/or audible indicator.
In accordance with another aspect, the present invention provides a method for measuring intra abdominal pressure comprising the steps of a) providing a catheter having first and second lumens and a balloon; b) inserting the catheter into a bladder of a patient; c) injecting gas into the first lumen of the catheter to expand the balloon from a deflated condition to a partially inflated condition; d) obtaining a first pressure reading of the bladder based on deformation of the balloon without injecting fluid into the bladder; e) transmitting the first pressure reading to an external monitor connected to the catheter; f) obtaining a second pressure reading of the bladder based on deformation of the balloon without injecting fluid into the bladder; g) transmitting the second pressure reading to the external monitor connected to the catheter; and h) obtaining consecutive continuous pressure readings of the bladder without injecting fluid into the bladder.
The method can include measuring the temperature of a body of a patient utilizing a temperature sensor within the first lumen.
In accordance with another aspect of the present invention, a multi-lumen catheter is provided that is insertable into a patient for monitoring pressure. The catheter comprises an expandable outer balloon at a distal portion of the catheter, the outer balloon having a first outer wall and receiving fluid to move from a first condition to a more expanded condition. A chamber is positioned within the outer balloon and contains a plurality of openings communicating with the interior of the outer balloon. An expandable inner balloon is positioned within the chamber and has a second outer wall. A first lumen communicates with the inner balloon, the inner balloon and first lumen forming a gas filled chamber to monitor pressure within the patient, wherein the outer balloon has a circumferential area greater than a circumferential area of the inner balloon, wherein in response to pressure exerted on the first outer wall of the expanded outer balloon fluid within the outer balloon enters one or more of the openings in the chamber to exert a pressure on the second outer wall of the expanded inner balloon to deform the inner balloon and compress the gas within the inner balloon and the first lumen to provide a finer measurement. A pressure sensor communicates with the gas filled chamber for measuring pressure based on compression of gas caused by deformation of the expanded inner balloon resulting from deformation of the expanded outer balloon.
In accordance with another aspect of the present invention, a multi-lumen catheter insertable into a patient for monitoring pressure is provided, the catheter comprising a shaft having first, second and third lumens, an expandable outer balloon at a distal portion of the catheter having an outer wall and receiving fluid via the second lumen to move from a first condition to a more expanded condition and a chamber positioned within the outer balloon containing a plurality of openings communicating with the interior of the outer balloon. An expandable inner balloon is positioned within the chamber and has an outer wall and a tubular portion extending within the first lumen. The outer balloon has a circumferential area greater than a circumferential area of the inner balloon, wherein in response to pressure exerted on the outer wall of the expanded outer balloon, fluid within the outer balloon enters one or more of the openings in the chamber to exert a pressure on the outer wall of the expanded inner balloon to deform the inner balloon and compress the gas within the inner balloon. A pressure sensor communicates with the gas within the inner balloon for measuring pressure based on compression of gas caused by deformation of the expanded inner balloon resulting from deformation of the expanded outer balloon.
In accordance with another aspect of the present invention, a multi-lumen catheter insertable into a patient for monitoring pressure is provided comprising a shaft having a first lumen, a second lumen and a third lumen. The third lumen of the catheter has an opening for drainage of a cavity. An expandable outer balloon is positioned at a distal portion of the catheter, the outer balloon having an outer wall and receiving fluid via the second lumen to move from a first condition to a more expanded condition. A plug is positioned within the third lumen of the catheter to provide a distal region and an expandable inner balloon is positioned within the distal region distal of the plug, the inner balloon having an outer wall and further having a tubular portion extending within the first lumen. The tubular portion has an angled portion so the tubular portion extends from the distal region into the first lumen. The outer balloon has a circumferential area greater than a circumferential area of the inner balloon, wherein in response to pressure exerted on the outer wall of the expanded outer balloon fluid within the outer balloon and exerts a pressure on the outer wall of the expanded inner balloon to deform the inner balloon and compress the gas within the inner balloon. A pressure sensor communicates with the gas within the inner balloon for measuring pressure based on compression of gas caused by deformation of the expanded inner balloon resulting from deformation of the expanded outer balloon.
In accordance with another aspect of the present invention, a multi-lumen catheter insertable into a patient for monitoring pressure is provided. The catheter comprises a catheter shaft having a distal end formed of a first material, an expandable outer balloon at a distal portion of the catheter having an outer wall and receiving fluid to move from a first condition to a more expanded condition, and an expandable inner balloon positioned within the outer balloon, the inner balloon having a second outer wall. A connecting pin is positioned at the distal end of the catheter and is positioned in the distal opening of the catheter extending distally therefrom. The connecting pin is composed of material different than the first material and the inner balloon is composed of material different than the first material and attached to the core pin. A first lumen communicates with the inner balloon and extends through the catheter shaft, the first lumen radially spaced from the connecting pin. The inner balloon and first lumen form a gas filled chamber to monitor pressure within the patient, wherein the outer balloon has a circumferential area greater than a circumferential area of the inner balloon, wherein in response to pressure exerted on the outer wall of the expanded outer balloon a pressure is exerted on the outer wall of the expanded inner balloon to deform the inner balloon and compress the gas within the inner balloon and the first lumen to provide a finer measurement. A pressure sensor communicates with the gas filled chamber for measuring pressure based on compression of gas caused by deformation of the expanded inner balloon resulting from deformation of the outer balloon.
In accordance with another aspect of the present invention, a multi-lumen catheter insertable into a patient for monitoring pressure is provided. The catheter includes a catheter body, a first balloon at a distal portion of the catheter balloon having a first outer wall expandable from a first condition to a more expanded condition and a second balloon having a second outer wall. The first balloon is external of the second balloon and the second balloon forms a fluid containing chamber to monitor pressure within the patient. A third balloon is positioned external of the first balloon such that the first balloon is positioned within the third balloon, the first balloon being less compliant than the third balloon and forming an inner liner of the third balloon to maintain an expanded condition of the third balloon. In response to pressure exerted on an outer wall of the expanded third balloon fluid within the first balloon enters one or more openings in the catheter to exert a pressure on the second outer wall of the expanded second balloon to deform the second balloon and compress the fluid within the second balloon to provide a pressure measurement.
The catheter can include a pressure sensor communicating with the fluid chamber for measuring pressure based on compression of fluids caused by deformation of the expanded second balloon resulting from deformation of the expanded first and third balloons.
In some embodiments the fluid chamber is a gas containing chamber which can in some embodiments be an air containing chamber.
In some embodiments, the catheter has a first lumen and the second balloon has an elongated portion extending through the first lumen and forming an elongated channel, the elongated channel along with the second balloon forming the gas filled chamber to monitor pressure within the patient.
The catheter can include one or more additional lumens to communicate with the bladder to remove fluid from the bladder and/or to inflate a retention balloon and/or inflate the first balloon. Preferably, the inner space of the first and second balloons are not in fluid communication so they are independently inflatable and deflatable.
The second balloon in some embodiments is maintained centered within the first lumen of the catheter so the second outer wall of the second balloon does not contact an inner wall of the first lumen of the catheter.
The catheter can have a sensor to measure core body temperature and a plurality of wires extending from the temperature sensor through the catheter.
In accordance with another aspect of the present invention, a method for measuring intra-abdominal pressure is provided comprising the steps of:
In some embodiments, deformation of the outer balloon is in response to pressure exerted on an outer wall of the expanded outer balloon, and upon such deformation fluid within the intermediate balloon enters one or more openings in the catheter to communicate with an outer wall of the expanded inner balloon to exert a pressure on and deform the inner balloon and compress the gas within the inner balloon to provide a finer pressure measurement. Preferably, fluid within the intermediate balloon does not enter inside the inner balloon.
Various uses of the catheter are provided including for example, the gas chamber monitoring pressure within a patient's bladder to thereby monitor pressure within an abdomen of the patient, monitoring pressure within a patient's bladder to thereby monitor uterine contraction pressure or monitoring pressure within a uterus of the patient to determine if excessive pressure is being applied to fallopian tubes of the patient.
In some embodiments, connecting a hub containing a pressure transducer to the catheter automatically advances gas into the inner balloon to expand the inner balloon and can also automatically connect a temperature sensor to a connector within the hub.
So that those having ordinary skill in the art to which the subject invention appertains will more readily understand how to make and use the surgical apparatus disclosed herein, preferred embodiments thereof will be described in detail hereinbelow with reference to the drawings, wherein:
Increased abdominal pressure can cause many adverse conditions including diminishing the function of the intestines, liver, and blood vessels. Simply viewing or feeling the abdomen does not provide sufficient information or reading of health conditions.
It is recognized that urinary bladder pressure directly correlates to the intra-abdominal pressure. Although pressure readings can be determined by access to the esophagus or rectum, the bladder has been found to be the most accurate and the least invasive. In trauma or burn patients for example, time is critical and the less complicated the method for determining bladder pressure the better the clinical results.
The catheters of the present invention measure abdominal pressure via measurement of bladder pressure without filling the bladder with water. This avoids the risks associated with retrograde filling of the bladder with water as such retrograde filling not only increases the complications and workload for the intensive care (IC) staff and can create inaccuracies by providing false elevation of IAP readings, but can adversely affect the patient by increasing the risk of infection. Furthermore, by avoiding refilling of the bladder, bladder pressure can be measured continuously. This is because in devices requiring filling the bladder with water, water needs to be periodically added to the bladder to replace the water drained from the bladder and measurement readings are interrupted during water insertion. Due to these repeated interruptions, pressure cannot be read continuously. Note in some cases, as much as 50 cc of fluid needs to be repeatedly added to the bladder.
The catheters of the present invention efficiently and effectively measure bladder pressure without requiring filling the bladder with water. Also, as will become apparent from the discussion below, the catheters of the present invention provide a more accurate reading of pressure and enable continuous monitoring of the bladder pressure. This is all achieved in an easy to insert device.
It should be noted that the catheters of the present invention can be utilized for measuring other pressure in a patient and are not limited to intra-abdominal pressure. The catheters of the present invention can also be inserted into a variety of body cavities of the patient and can be used for monitoring pressure of various body regions. These catheters can be used in various body cavities for measuring pressure without requiring insertion of water into the body cavity and thus have the numerous advantages associated with requiring water as described herein.
Furthermore, in some embodiments, the catheters of the present invention have a dual sensor to provide a backup pressure reading. In some embodiments, a dual pressure balloon arrangement is provided. These various embodiments are discussed in more detail below.
Referring now to the drawings and particular embodiments of the present invention wherein like reference numerals identify similar structural features of the devices disclosed herein, there is illustrated in
The catheter 10 of the present invention can in some embodiments include an alarm or indicator to alert the user if pressure within the bladder, which correlates to pressure within the abdomen, rises to an unacceptable level, i.e., beyond a threshold or predetermined value (pressure). The indicator or alarm can be on the catheter or alternatively on an external device such as the monitor as discussed in more detail below. The alarm can also be connected via wireless connection to a phone or remote device to alert the appropriate personnel. The indicator or alarm can alternatively or in addition be activated if a change in pressure measurement exceeds a specified rate over a specified period of time.
Turning now to details of the catheter 10, which is also referred to herein as the device 10, and with initial reference to
Note
A sensor 30 is positioned within lumen 14 adjacent balloon 16. The wire(s) 32 are shown extending through lumen 14, the sensor 30 and wire(s) 32 being of sufficiently small size so as not to interfere with air flow though lumen 14. The sensor 30 measures pressure of the bladder. The sensor 30 is part of a transducer for converting the variation in pressure to an electrical signal for transmission to an external monitor. The pressure sensor can also include a temperature sensor, or alternatively another sensor for sensing temperature could be provided, to measure core temperature of the body as seen inside the bladder. Transmission wire(s) 34 of the temperature sensor extend adjacent wire 32 through lumen 14 and terminate external of the catheter 10 for connection to an external monitor. The transducer can be wired directly to the monitor or alternatively wired to a converter external of the catheter for converting the signal received by the transducer and transmitting a signal to the monitor, e.g., a bedside monitor, to display the pressure readings. This is shown schematically in
Wires 32 and 34 can extend though lumen 14 and exit side port 15 for connection to a converter or monitor or alternatively can be inserted through the lumen 14, piercing the wall to enter the lumen 14 distal of the side port.
An alarm system can also be provided wherein the system includes a comparator for comparing the measured pressure (and/or temperature) to a threshold (predetermined) value, and if such threshold is exceeded, an indicator, e.g., an alarm, is triggered to indicate to the hospital personnel the excessive pressure and/or temperature. An alarm system can alternatively or in addition be activated if a change in pressure measurement exceeds a specified rate over a specified period of time. This would alert the staff to an imminent risk prior to intra-abdominal pressure exceeding a certain value, e.g., 20 mm hg, since due to this link, the relationship between intra-abdominal pressure and abdominal cavity volume is believed to be linear up to an intra-abdominal pressure of 12-15 mm hg and increasing exponentially thereafter.
The alarm system can be part of the catheter (as shown in
The lumen 14 and space 16a within balloon 16 together form a closed gas, e.g., air, chamber, i.e., the lumen 14 forming an air column. With the balloon 16 filled with air, pressure on the external wall of the balloon will force the balloon to deform inwardly, thereby compressing the air contained within the balloon space 16a and within the lumen 14. The pressure sensor 30 is located in a distal portion of the lumen 14 at the region of the balloon 16 and thus is positioned at the distal end of the air column. Therefore, the pressure is sensed at the distal region as the sensor 30 detects change in air pressure in lumen 14 due to balloon deformation. Placement of the sensor 30 at a distal location provides a pressure reading closer to the source which advantageously increases the accuracy because it reduces the risk of transmission issues by reducing the amount of interference which could occur due to water, air, clots, tissue, etc. if the transmission is down the air lumen (air column).
Additionally, the pressure measurement occurs about a more circumferential area of the balloon 16 providing a pressure reading of a region greater than a point pressure sensor reading. Also, average pressure over an area of the bladder wall can be computed. Thus, the area reading gleans information on pressure over more of the bladder wall. Stated another way, the balloon has a relatively large surface area with multiple reference points to contribute to average pressure readings of the surface around it by the sensor.
The air column is charged by insertion of air through the side port 15 which communicates with lumen 14. The side port 15 includes a valve to provide a seal to prevent escape of air from a proximal end. The balloon 16 can be composed of impermeable material, or in alternative embodiments, a permeable or semi-permeable material with an impermeable coating. This seals the air column at the distal end to prevent escape of air through the distal end, i.e., through the wall of the balloon 16. Thus, with the lumen sealed at the proximal and distal ends, a closed air system is provided, and without the requirement for repeated water insertion, a fully closed unit is provided.
In some embodiments, when the lumen 14 is air charged, the balloon 16 is not fully inflated. This improves the accuracy of the balloon 16 transmitting pressure from external the balloon to the interior of the balloon and into the lumen, i.e., air column, by ensuring the balloon has sufficient compliancy to prevent the balloon from introducing artifact into the pressure reading which would diminish its accuracy.
In some embodiments, the pressure balloon 16 is of a size to receive at least about 3 cc (3 ml) of fluid. However, other sizes/volumes are also contemplated such as about 2 cc or about 1 cc. Additionally, these volumes represent the maximum volume of fluid for the balloon, however, as noted above, in preferred embodiments, the pressure balloon 16 is not fully inflated so it would receive less than the maximum volume. Thus, with a balloon of X volume, the fluid would receive X-Y fluid, with Y representing the amount of desired extra space to achieve desired compliancy of the balloon while still enabling sufficient inflation of the balloon to achieve its pressure induced deformation function.
Note in this embodiment, the stabilizing balloon 26 is positioned proximal of the pressure balloon 16. Also, in this embodiment, the stabilizing balloon 26 is larger than the pressure balloon 16. By way of example, the stabilizing balloon 26 can have a fully expanded diameter of about 23 mm and the pressure balloon 16 can have a fully expanded diameter of about 15 mm, although other dimensions or diameters for these balloons are also contemplated. By way of example, the stabilizing balloon 26 can have a capacity of about 10 cc (10 ml) of air, although other sizes/volumes are also contemplated. Note these sizes/volumes for both balloons are provided by way of example and other sizes are also contemplated. Alternatively, the stabilizing balloon can be the same size or smaller than the pressure balloon. Various shapes of the balloons are also contemplated.
Additionally, although the balloon 26 is positioned proximal of the balloon 16, it is also contemplated that the balloon 26 be positioned distal of balloon 16. The axial spacing of the balloons 16, 26 enable the stabilizing balloon 26 to engage the bladder wall to provide a sufficient radial force thereon for securing/mounting the catheter within the bladder without interfering with the function of balloon 16.
It should be appreciated that although the stabilizing balloon is shown in the embodiment of
Note that although only one sensor is shown in
As shown, the sensor 30 and its transmission wires are located in the same lumen 14 also used for initial inflation gas, e.g., air, for balloon 16 and for the air charged column. This minimizes the overall transverse cross-section (e.g., diameter) of the catheter 10 by minimizing the number of lumens since additional lumens require additional wall space of the catheter. However, it is also contemplated in an alternate embodiment that the sensor is in a dedicated lumen separate from the inflation lumen 14. This can be useful if a larger sensor or additional wires are utilized which would restrict the air lumen if provided therein. This is also useful if a specific sized lumen for the sensor and wires is desired to be different than the sized lumen for the air column. Provision of a separate lumen is shown in the cross-sectional view of
Turning now to the use of the catheter 10, the catheter 10 is inserted into the bladder. Note catheter 50 would be used in the same manner. The balloon 26 is inflated to secure the catheter 10 in place during the procedure by insertion of a fluid (liquid or gas) through side port 28 which is in fluid communication with lumen 24. The system is charged by inflation of the balloon 16, i.e., preferably partial inflation for the reasons discussed above, by insertion of air via a syringe through port 15 which is in fluid communication with lumen 14. As discussed above, the catheter 10 is a closed system with the balloon 16 sealed so that air inserted through lumen 14 and into balloon 16 cannot escape through balloon 16. Thus, a closed chamber is formed comprising the internal space 16a of the balloon 16 and the internal lumen 14 communicating with the internal space 16a of balloon 16. With the balloon 16 inflated, pressure monitoring can commence. When external pressure is applied to an outer surface 16b of the balloon 16, caused by outward abdominal pressure which applies pressure to the bladder wall and thus against the wall of balloon 16, the gas e.g., air, within the chamber is compressed. The sensor 30 at the distal end of lumen 14 provides continuous pressure readings, converted to an electrical signal by the transducer within the distal end of lumen 14, and then electrically communicates through wire(s) 32 extending through lumen 14, exiting through the proximal side port 15 and connected to an external monitor. Note the wire can terminate at the proximal end in a plug in connector which can be connected directly to the monitor or alternatively plugged into a converter to convert the signals from the transducer in the embodiments wherein the converter is interposed between the wires and monitor (see e.g., the system of
In the embodiments wherein an indicator is provided, if the measured pressure exceeds a threshold value, and/or a change in pressure measurement exceeds a specific rate over a specific time period, the indicator would alert the clinician, e.g., via a visual indication or an audible indication that the threshold is exceeded. The indicator in some embodiments can include an audible or visual alarm (shown schematically in
In the embodiments of
Sensor 80 is positioned in lumen 64 for sensing pressure in response to balloon deformation in the same manner as sensor 30. Sensor 82 is positioned in lumen 64 distal of sensor 80 for measuring core temperature. Temperature sensor 82 can be a thermocouple, a thermistor or other types of temperature sensors. As shown in
In use, catheter 60 is inserted into the bladder and stabilizing balloon 76 is inflated to secure the catheter 60 in place. The system is charged by inflation of the balloon 66, i.e., preferably partially inflated for the reasons discussed above, by insertion of gas, e.g., air, through port 65 which is in fluid communication with lumen 64 in a closed system formed by the internal space 66a of the balloon 66 and the internal lumen 64 communicating with the internal space 66a of balloon 66. With the balloon 66 inflated, pressure monitoring can commence as external pressure applied to an outer surface of the balloon 66 compresses the gas within the gas containing chamber. The sensor 80 at the distal end of lumen 64 provides continuous pressure readings, converted to an electrical signal by the transducer within the distal end of lumen, and then electrically communicates through wires 82 extending through lumen 64 to an external monitor either directly or via a converter. The sensor 82 at the distal end of lumen 64 provides continuous temperature readings via wires 83 communicating directly or indirectly with the monitor, Although, the system is capable of continuous pressure and continuous temperature monitoring, as with the other systems disclosed herein, it can also be adapted if desired for periodic monitoring so the pressure and/or temperature readings can be taken at intervals or on demand by the clinician.
In the alternative embodiment of
In the alternative embodiment of
Note the wire(s) of the sensor 102 can terminate at the proximal end in a plug in connector which can be connected directly to the monitor or alternatively plugged into a converter to convert the signals from the transducer in the embodiments where the converter is interposed between the wires and monitor (see e.g. the system of
The pressure sensor 130 is carried by catheter 110 and positioned within the balloon 116 to measure pressure in response to deformation of the balloon in response to pressure exerted on an outer wall of balloon 116. The pressure transducer can include the sensor 130 or can be a separate component positioned at a proximal end of the catheter external of the catheter 110. The temperature sensor 132 can be positioned within the balloon 116, can be part of sensor 130, or alternatively positioned within lumen 114 (as shown in
In all other respects, catheter 110 is identical to catheter 60 and therefore for brevity further discussion is not provided since the structure and function of the balloons, lumens, continuous pressure monitoring, etc. as well as the aforedescribed alternative arrangements of catheter 60, are fully applicable to the catheter 110.
As discussed above, the pressure balloons disclosed herein have a large circumferential area (and large volume) to provide multiple reference points for pressure readings and to provide an average pressure to enable more accurate readings. Thus, the pressure balloon provides for gross measurement. In an alternate embodiment shown in
The inner balloon 143 and outer balloon 142 can be separately/independently inflated and closed with respect to each other so there is no communication, e.g. passage of gas or liquid, between the inner and outer balloons 143, 142.
In the embodiments disclosed herein having inner and outer balloons, the outer balloon acts a medium of transmission to the inner pressure balloon. That is, as the outer balloon is deformed, the fluid within the outer balloon acts against the outer wall of the inner balloon to deform the inner balloon and pressurize the gas, e.g., air, within the chamber of the inner balloon for pressure measurement. With the outer balloon functioning as a transmission medium, the bladder (or other body cavity in which the catheter is inserted) does not need to be filled with fluid. Thus, the catheter can be used in a voided cavity, e.g., without interstitial fluid. The radial spacing between the wall of the outer balloon and wall of the inner balloon provides space for transmission of the fluid within the outer balloon to deform the inner balloon. The spacing can be achieved in various ways which are described below and include for example, radial separation, a chamber interposed between the inner and outer balloons, a wall of the catheter interposed between the inner and outer balloons, etc. The advantages of not requiring insertion of fluid during pressure measurement are discussed below.
The pressure transducer and pressure sensor 150 can be positioned within the lumen 144 in the same manner as sensor 30 of
The catheter 140 can optionally include a stabilizing balloon 145 similar to balloon 76 of
In use, catheter 140 is inserted into the bladder and stabilizing balloon 145 is inflated to secure the catheter 140 in place. The system is charged by inflation of the inner balloon 143, i.e., preferably partially inflated for the reasons discussed above, by insertion of air through a side port which is in fluid communication with lumen 144 in a closed system formed by the internal space 143a of the inner balloon 143 and the internal lumen 144 communicating with the internal space of inner balloon 143. Outer balloon 142 is filled, i.e., preferably partially inflated for the reasons discussed above, via injection of air through a separate lumen. With the outer balloon 142 inflated, pressure monitoring can commence as external pressure applied to the larger circumferential outer surface of the outer balloon 142 compresses and deforms the outer balloon 142 which compresses the inner balloon 143 as fluid presses the outer wall of the inner balloon. As the inner balloon 143 is compressed and deformed in response to compression/deformation of the outer balloon 142 based on changes to bladder pressure, the sensor 150 at the distal end of lumen 144 provides continuous pressure readings, converted to an electrical signal by the transducer within the distal end of lumen 144, and then electrically communicates through wires 152 extending through lumen 144 to an external monitor either directly or via a converter. Although, the system is capable of continuous pressure and continuous temperature monitoring, as in the other embodiments disclosed herein it can also be adapted if desired for periodic monitoring so the pressure and/or temperature readings can be taken at intervals or on demand by the clinician. Note fluid does not need to be present in the cavity to achieve the pressure readings.
Note that although separate lumens are provided for the inflation of inner balloon 143 and outer balloon 142, in an alternate embodiment, a single lumen can be utilized to inflate both balloons 143 and 142.
Note that the larger balloon 142′ can be used with the catheters of any of the embodiments described herein. Thus, a pressure balloon of the larger size balloon 142′ can be used instead of the smaller pressure balloons illustrated in the drawings. Note the size of the balloons is provided by way of example and are not necessarily drawn to scale comparatively to the other components.
Note that the positioning of the side opening for drainage of
Note that the pear shaped balloon 142″ can be used with the catheters of any of the embodiments described herein. Thus, a pressure balloon of the pear shape of balloon 142″, and of larger or smaller size if desirable, can be used instead of the pressure balloons illustrated in the drawings.
The pressure transducer and pressure sensor are external to catheter 200 and mounted to port 218 at the proximal end 201 of catheter 200. More specifically, a transducer hub or housing, designated generally by reference numeral 240, contains the sensor and pressure transducer and is mounted to the angled side port 218. In the embodiment of
To charge the system, when the hub 240 is mounted to the side port 218, the elongated member 242 extends into lumen 214 to advance air through the air lumen 214 into inner balloon 204 to expand inner balloon 204. In some embodiments, 0.2 cc of air can be displaced/advanced by the member 242, although other volumes are also contemplated. Thus, as can be appreciated, mounting of the hub 240 to the catheter 200 automatically pressurizes the air lumen/chamber and expands the inner balloon 204. Note the inner balloon 204 can be partially or fully inflated (expanded), dependent on the amount of air advanced into the inner balloon 204. Further note that the lumen 214 is not vented to atmosphere when the transducer hub 240 is attached and air is advanced through the air lumen. The port 218 can include a closable seal through which the elongated member 242 is inserted but maintains the seal when the elongated member 242 remains in the lumen 214. This maintains a closed system.
Lumen 214 which is used to inflate the inner balloon 204 and create the air column has an opening at a distal region to communicate with the interior of inner balloon 204. Lumen 212 of catheter 200 has an opening at a distal region to communicate with the outer balloon 202 to fill the outer balloon 202. Angled port (extension) 222 at the proximal end of catheter 200 receives an inflation device to inflate, either fully or partially, the outer balloon 202.
Note as in the other embodiments disclosed herein, air is described as the preferred gas for creating the column and expanding the balloon, however, other gasses are also contemplated, for each of the embodiments herein.
The outer balloon 202 can be shaped such that a distal region 207a (
The inner and outer balloons 204, 202 can by way of example be made of urethane, although other materials are also contemplated such as silicone or EVA.
A temperature sensor 230 (
As can be appreciated, connection of the transducer hub 240 to the catheter 200 (port 218) a) automatically connects the temperature sensor 230 to a connector for communication with a temperature monitor cable; and b) automatically advances air through the first lumen 214 to expand the inner balloon 204.
The catheter 200 can optionally include a stabilizing balloon 206 similar to balloon 76 of
In the embodiment of
In use, catheter 200 is inserted into the bladder and stabilizing balloon 206 is inflated to secure the catheter 200 in place. The system is charged by inflation of the inner balloon 204, i.e., preferably partially inflated for the reasons discussed above, by advancement of air through lumen 214 upon attachment of the pressure transducer 240 to the port 218 of catheter 200. Such attachment moves elongated member 242 into lumen 214 to displace the air (or other gas) already in the lumen 214 to expand the inner balloon 204. A closed system is formed by the internal space 204a of the inner balloon 204 and the internal lumen 214 communicating with the internal space 204a of inner balloon 204. In a preferred embodiment, additional air does not need to be added to the balloon 204/lumen 214. Outer balloon 202 is filled, i.e., preferably partially inflated for the reasons discussed above, via injection of air through the separate port 222 which communicates with lumen 212 of catheter 200. With the outer balloon 202 inflated, pressure monitoring can commence as external pressure applied to the larger circumferential outer surface of the outer balloon 202 compresses and deforms the outer balloon 202 which exerts a force on the outer wall of inner balloon 204 (via fluid contact with the outer wall of the inner balloon) and compresses the inner balloon 204. As the inner balloon 204 is compressed and deformed in response to compression/deformation of the outer balloon 202 based on changes to bladder pressure, the pressure sensor within the external hub 240 attached at the proximal end of the catheter 200 provides continuous pressure readings, converted to an electrical signal by the transducer of the sensor within the hub 240, and then electrically communicates through a connector, e.g. cable 245, to an external monitor either directly or via a converter to display pressure readings. Although, the system is capable of continuous pressure and continuous temperature monitoring, it can also be adapted if desired for periodic monitoring so the pressure and/or temperature readings can be taken at intervals or on demand by the clinician. Temperature readings are also taken during the procedure as temperature sensor 230 is connected to a temperature monitor via wire 232 connected to a connector of hub 240 which is connected to the temperature monitor to display temperatures. The temperature monitor can be separate from the pressure display monitor or alternatively integrated into one monitor. Cable 245 can connect to the temperature monitor as well (directly or via a converter) or a separate cable extending from the hub 240 could be provided for connection to the temperature monitor.
Note that although separate lumens are provided for the inflation of inner balloon 202 and outer balloon 204, in an alternate embodiment, a single lumen can be utilized to inflate both balloons 202 and 204. In such embodiment, catheter 200 can have one less angled port and one less lumen since inflation of the outer balloon 202 would be through port 218 and lumen 214.
The proximal and distal end of the inner balloon 204 in the illustrated embodiment are within the confines of the outer balloon 202, i.e., the proximal end of the inner balloon 204 is distal of the proximal end of the outer balloon 202 and the distal end of the inner balloon 204 is proximal of the distal end of the outer balloon 202. Thus, in this illustrated embodiment, the inner balloon 204 is fully encapsulated within the outer balloon 202.
With the inner/outer balloon arrangement, the larger outer surface of the outer balloon 202 takes gross measurements and then the forces are concentrated on the smaller inner balloon 204 to amplify/concentrate pressure on the small area of the inner balloon so small changes can be detected and waves transmitted to the pressure transducer (via the length of the lumen to a proximal transducer, e.g., an external pressure transducer).
As noted above, preferably no additional air needs to be added after mounting of hub 240. However, it is also contemplated that in alternate embodiments a port can be provided in communication with hub 240 to enable subsequent injection of air though lumen 214 and into inner balloon 204. Additionally, outer balloon 202 can in some embodiments receive additional fluid injection via port 222 during the procedure.
Catheter 400 has a shaft 402 having a distal region (portion) 402a terminating in a distal opening 402b. Distal opening 402b receives core pin 410 therein. Core pin 410, also referred to as a bonding pin or a connecting pin, has a proximal end 414a dimensioned for insertion in a press fit through opening 402b and into the lumen in distal region 402a of shaft 402. In the illustrated embodiment, the proximal end 414a has a non-circular shape, e.g., a triple lobe or Y shape, corresponding to the shape of the opening 402b. The distal end of the pin 410 has a reduced diameter cylindrical portion 414b which receives thereover a distal tip 412 (also referred to herein as a distal plug).
Catheter 400 further has a retention (stabilizing) balloon 404, an inner balloon 408 and an outer balloon 406. The retention balloon 404 is spaced proximally of the outer balloon 406 and inner balloon 408. The outer balloon 406 encapsulates the inner balloon 408 such that the inner wall 408c of the inner balloon 408 is contained within the outer balloon 406. The outer wall 406c of outer balloon 40c is exposed to the patient, e.g., the bladder. The retention balloon 404 functions in the same way as the retention (stabilizing) balloons described above and can be of varying shapes as described herein. The inner and outer balloons 408, 406 function in the same way as the inner and outer balloons 204, 202 of catheter 200.
As shown in the cross-sectional view of
Catheter 400 also has three angled extensions/ports at its proximal end 420 (
The steps of manufacture (assembly) of the balloons to the catheter will now be described with reference to
Next, with reference to
In some embodiments, the outer balloon 406 is folded over itself as shown in
Side opening(s) 418 in catheter 400 communicate with the drainage lumen 432 for draining the bladder. As shown, the drainage opening(s) 418 in this embodiment is positioned between the a) outer balloon 406/inner balloon 408 and b) retention balloon 404. More than one drainage opening can be provided. It should be appreciated that such location of the drainage opening(s) between the retention balloon and pressure balloon(s), rather than distal of the pressure balloon(s) can be utilized with any of the catheter embodiments disclosed herein.
As noted herein, the catheters of the present invention can be utilized for measuring other pressure in a patient and are not limited to intra-abdominal pressure nor limited to measuring bladder pressure.
In the foregoing embodiments, the inner balloon is positioned within the outer balloon (with its outer wall radially spaced from the outer wall of the outer balloon) and deformation of the outer balloon based on changes in pressure within the patient, e.g., within the bladder in response to abdominal pressure, causes deformation of the inner balloon as the fluid within the outer balloon (or wall) exerts a pressure against the wall of the inner balloon. This deforms the inner balloon to provide a pressure reading. In the alternate embodiment of
With reference now to
The outer balloon 456 has a proximal end 456a attached to the shaft 451 and a distal end 456b attached to the chamber 452. The outer balloon 456 can include a cuff 457 like cuff 411 of outer balloon 406 of
The chamber 452 forms the distal end region of the catheter 450 as it extends distally from the distal edge 451a of shaft 451. With reference to
Chamber 452 also includes a plurality of openings 462 (only some of which are labeled for clarity) to provide fluid (liquid or gas) flow from the outer balloon 456 against the outer wall 458b of the inner balloon 458 retained within the chamber 452. Thus, the chamber 452 separates the outer wall 458b of the inner balloon 458 from the interior of the outer balloon, except for the communication through the openings 462 in the chamber 452. (The inner and outer balloons are sealed from each other so they are not in fluid communication with each other). One arrangement of chamber openings is shown by way of example in
The inner balloon 458 has balloon portion 459a and a tube portion 459b. The balloon portion 459a as referred to herein is the portion or region that expands within the catheter (e.g., chamber) and is the region that is within the outer balloon that receives fluid contact from the fluid in the outer balloon. The balloon portion 459a has a larger diameter than the tube portion 459b as shown in
Note in the embodiments wherein the inner balloon has an elongated tubular portion extending through the lumen of the catheter, the tubular portion can be integral with the enlarged region of the balloon that receives the fluid contact from the outer balloon. In alternative embodiments, the tubular portion can be a separate component attached to the balloon portion. In other embodiments, the tubular portion can be in the form of a metal tube extending through the lumen of the catheter and attached to the tail of the inner balloon. In any of these versions, the tubular portion along with the balloon portion form the gas chamber. In some embodiments, the tubular portion can extend through the entire or almost the entire length of the lumen of the catheter and terminate at the proximal end adjacent the distal end of the elongated member (rod) of the hub used to inflate the inner balloon. In other embodiments, it can terminate more distal.
A sealant or plug 480 can be provided, the plug 480 positioned around the tubular portion 459b of the inner balloon 458 which closes off the lumen 467 distal of the drainage hole 463. The plug 480 could also help maintain the centering of the tubular portion 459b. Plug 480 abuts inner wall 478 of chamber 452 as shown in
A thermistor can be positioned adjacent the drainage opening(s) 463 and the temperature sensor wires can be positioned in lumen 484, running parallel to the balloon tubular portion 459b. Alternatively, the temperature sensor wires can be positioned in the same lumen as the lumen for filling the outer balloon or alternatively an additional lumen can be provided for the temperature sensor wire(s).
The outer balloon 456 could include a coating such a parylene to change the modulus of the balloon. That is, such coating could stiffen the balloon so it is not to continuously expanding under pressure, which could cause a reduced pressure reading. The coating can also cover all or part of the catheter which could add lubricity.
The inner and outer balloons disclosed herein can be filled to various volumes. In one example, the inner balloon can be filled to a volume of about 0.16 cc and the outer balloon can be filed to about 10 cc.
The inner and outer balloons disclosed in the various embodiments can be coated to reduce their permeability. That is, to prevent escape of air, the balloons can be made of an impermeable material and/or the balloons can be made of a permeable material and coated with an impermeable material. As used herein, impermeability means there is no (or negligible) pressure decay in the balloon over a period of time in which the catheter remains inserted into the body. This period of time could be for example from one day to up to 30 days. This period of time of catheter insertion, due to current hospital and clinical protocols, typically does not exceed 30 days, but impermeability of the balloons of the present invention can also mean little or no leakage for a longer period of time, e.g., 45 days, depending on the protocol for duration of catheter insertion.
Various features affect pressure decay which include the density of the balloon material, the material and/or structure of the balloon, and the type and/or density of coating on the balloon. The catheters of the present invention provide pressure loss management by limiting loss of pressure resulting from escape of air (or other gas) from the system. The pressure management, affected by these parameters, also needs to be balanced with the size and patient comfort restraints, which the catheters of the present invention achieve.
Additionally, the amount of dead space in the system can affect decay because if there is more gas (e.g., air) in the system, the percentage loss will have less of an overall effect than if there is less gas in the system. However, a larger gas chamber acts as a dampening affect so there is less responsiveness. Therefore, the catheters of the present invention achieve this balance of accurate pressure reading (due to maximized responsiveness) while minimizing or eliminating pressure decay in the necessary time period.
Note that when used in the bladder, there is an osmotic effect with urine. That is, urine sucks out the air from the system as the pH changes. The impermeability of the balloon helps to limit or eliminate this osmotic effect.
As noted herein, the catheters of the present invention can be utilized for measuring other pressure in a patient and are not limited to intra-abdominal pressure nor limited to measuring bladder pressure. Thus, the double balloon/chamber structure of
Catheter 450 can be used in the same manner as the catheters described above, and such aforedescribed use(s) are fully applicable to catheter 450, the difference being the caged inner balloon. Thus in use, catheter 450 is inserted into the cavity, e.g., bladder, and stabilizing balloon 454 is inflated to secure the catheter 450 in place. The system is charged by inflation of the inner balloon 458, i.e., preferably partially inflated for the reasons discussed above, by insertion of air through a side port which is in fluid communication with tubular portion 459b, or by mounting of the transducer hub as described herein, in a closed system formed by the internal space of the inner balloon 458 and the internal catheter lumen and/or tubular portion 459b communicating with the internal space of inner balloon 458. Outer balloon 456 is filled, i.e., preferably partially inflated for the reasons discussed above, via injection of fluid such as air or saline through a separate lumen. With the outer balloon 456 inflated, pressure monitoring can commence as external pressure applied to the larger circumferential outer surface of the outer balloon 456 compresses and deforms the outer balloon 456 which forces fluid within the outer balloon 456 through openings 462 in chamber 452 against the outer wall of the inner balloon 458, compressing the inner balloon 458. As the inner balloon 458 is compressed and deformed in response to compression/deformation of the outer balloon 456 based on changes to bladder pressure, the sensor provides continuous pressure readings, communicated to an external monitor. Although, the system is capable of continuous pressure and continuous temperature monitoring, as in the other embodiments disclosed herein it can also be adapted if desired for periodic monitoring so the pressure and/or temperature readings can be taken at intervals or on demand by the clinician.
An alternate embodiment of the chamber (cage) is illustrated in
Like chamber 452, chamber 490 has a series of openings 494 to provide fluid (liquid or gas) flow from the outer balloon 514 against the outer wall of the inner balloon 510 retained within the chamber 490. Thus, the chamber 490, like chamber 452, separates the outer wall of the inner balloon 510 from the interior of the outer balloon 514, except for the communication through the openings 494 in the chamber 490. One arrangement of chamber openings is shown by way of example in
Chamber 490 has a reduced diameter proximal extension 502 for press fit and attachment within the distal end of the catheter shaft. Slot 504 in chamber 490 communicates, i.e., aligns, with lumen 528 for inflating outer balloon 514. A wire can extend through the inner balloon 510 and be embedded in a wall 493 (pierces the wall 493) to help stabilize and center the inner balloon 510.
The catheter has four lumens: 1) lumen 522 receiving the tubular portion 512 of the inner balloon 510; 2) lumen 528 communicating with outer balloon 514 for inflating outer balloon 514; 3) lumen 525 communicating with the retention balloon 516 to inflate retention balloon 516; and 4) drainage lumen 524 having one or more side openings 520 at a distal region of the catheter for drainage of the bladder. In this embodiment, the tubular portion 512 of the inner balloon 510 is positioned in a separate lumen as shown. Except for the angled portion 513, the inner balloon 510 is identical to the inner balloon 458 of
Inner balloon 538 functions in the same manner as inner balloon 458. Catheter 530 has an outer balloon 546 and a retention (stabilizing) balloon 548 which function in the same manner as outer balloon 456 and retention balloon 454 of
This embodiment of
Catheter 600 has a retention balloon 604 identical to retention (stabilizing) balloon 206 of catheter 200 (or other stabilizing balloons disclosed herein), an inner balloon 608 and an outer balloon 606. An intermediate balloon 614 is positioned within the outer balloon 606 and external of the inner balloon 608.
Turning to
Catheter 600 also has a hub portion with three angled extensions/ports (
In the alternate embodiment of
A thermistor can be placed adjacent the drainage opening 610 of catheter 600 for temperature readings, and the thermistor wire(s) can extend through a lumen of the catheter 600 such as lumen 644 for the inner balloon as in catheter 600 or alternatively through the drainage lumen or through a separate lumen, for electrical connection to a temperature monitor.
Catheter 600 has an inner drainage lumen 646 having a plug or seal 662 (
Distal of plug 662 is a distal region or area 623 which forms an enlarged area of the lumen 644 which receives the inner balloon 608. Inner balloon 608 has a balloon region 608a and a tubular region 609 extending proximally therefrom. Tubular region 609 extends within lumen 644 which is separate (independent) from the drainage lumen 646. Thermistor wires discussed below can also extend through lumen 644.
The inner balloon 608 has an elongated shape (
In an alternate embodiment, the inner balloon can include an inner component 639 placed within the inside diameter. The inner component 639 (
The intermediate balloon 614 can be “pear shaped” or “liberty bell shaped” as shown in
The intermediate balloon 614 forms an inner liner of the outer balloon 606 to therefore act a like a coating to stiffen the outer balloon 606. The intermediate balloon 614 is composed of a material, e.g., a thermoplastic material, that is less compliant than the outer balloon 606. For example, the intermediate balloon 614 can be composed of EVA which would counteract expansion of the more compliant outer balloon 606. The outer balloon 606 is composed of a more compliant material such as silicone. With a compliant balloon, it would continue to expand (stretch) so the pressure would drop. The inner liner prevents this. Note that outer balloon 606 preferably has a smoother surface to shield the thermoplastic intermediate balloon 614 during insertion and use to reduce patient discomfort. The intermediate balloon 614 in some embodiments has a wall thickness less than the wall thickness of the outer balloon 606. Note that other compliant/non-compliant materials are also contemplated.
In an alternate embodiment, upon inflation of the intermediate balloon 614, the outer balloon 606 would peel back. Therefore, the outer balloon 606 would act as a sheath.
In use, the intermediate balloon 614 is expanded via injection of fluid, e.g. air saline, etc. through the port 622. Expansion of the intermediate balloon 614 causes expansion of the outer balloon 606. Deformation of the outer balloon 606 based on changes in pressure within the patient, e.g., within the bladder in response to abdominal pressure, causes deformation of the intermediate balloon 614. As the intermediate balloon 614 is deformed, fluid within the intermediate balloon 614 passes through the opening(s) 644a within the catheter to apply a pressure against the outer wall of the inner balloon 606 to deform the inner balloon 606 to provide pressure readings in the same manner as the other embodiments disclosed herein.
Turning back to the transducer hubs mountable to the catheter,
In the aforedescribed embodiments, mounting of the transducer hub a) automatically connects the temperature sensor to a connector for communication with a temperature monitor cable; and b) automatically advances air through the first lumen to expand the inner balloon. In the embodiment of
More specifically, transducer hub or housing, designated generally by reference numeral 300, contains the pressure transducer and sensor 309 and is mounted to the angled side port 282. In the embodiment of
The housing 290 attached to catheter 280 has a proximal opening 294 and a channel (lumen) 296 to receive an elongated (rod-like) member or nose 302 extending distally from transducer hub 300. As shown channel 296 has a first diameter region 296a to match with the lumen 283 of the port 282, a second larger diameter region 296b proximal of region 296a to receive the male rod 302 of the hub 300, and a still larger diameter region 296c proximal of region 296b to receive the valve 299 and valve 298 and allow expansion thereof. As shown, valve 298 is dome shaped and is distal of valve 299. Conical cap 293, proximal of valve 299, provides a lead in to the valve 299 for the rod 302. Thermistor pins 292 receive thermistor connectors 308. Note valves 288, 299 are one example of valves that can be provided as other valves to provide an airtight seal are also contemplated. A single valve is also contemplated.
Hub 300 is mounted to connector 290 and includes a housing 304 from which a pair of distally extending snap fit connector arms 306 extend. The latch arms 306 are sufficiently flexible to enable attachment and have an enlarged distal portion 307, illustratively shown as arrow shaped although other enlarged shapes could be provided. The elongated member 302 extends between the latch arms 306. When the hub 300 is mounted to the connector 290, the elongated member 302 extends into the channel 296 to advance air to inflate the inner balloon. The enlarged ends 307 of latch arms 306 enter recesses 291 and engage shoulders 291a to retain the hub 300. Note to release (disconnect) the hub 300, the ends 307 are pressed radially inwardly to disengage from shoulder 291a and the hub 300 is pulled proximally. Note that alternatively a different number of latch arms could be provided.
The housing (connector) 290 has a lumen 296 for communication with the lumen 283 in the side port 282 of catheter 280 which communicates with the air lumen and inner balloon of the catheter 280. As noted above, the lumen 296 is dimensioned to receive the elongated rod 302 of transducer hub 300. The wire for the sensor extends in housing 300. When transducer hub 300 is attached to connector 290, such attachment inserts the elongated rod 302 into lumen 296 to advance air though the air lumen in the catheter and into the balloon 204. (Note the air lumen extends into its angled side port 282). In the embodiments discussed above wherein the balloon has a tubular portion, the proximal end of the tubular portion could terminate adjacent the distal end of the rod (elongated member of this or other embodiments) or slightly distally thereof or in the lumen of the side port 283. The elongated member 302 also has a channel or lumen 305 extending therethrough to allow the pressure wave to travel through to the pressure sensor. Although in preferred embodiments no additional air needs to be injected into balloon 204 after attachment of hub 300, it is also contemplated that a port or opening can be provided in hub 300 to receive an injection device for injection of additional air. Such additional air can communicate with and flow through channel 305 of elongated member 302, into the air lumen and balloon 204 for inflation, or alternatively, a side port or opening in the angled port downstream (distal) of the elongated member 302 could be provided. Attachment of hub 300 to housing 290 also automatically connects thermistor connectors 308 to thermistor pins 292 to automatically connect the temperature sensor to the hub 300 for communication via a cable to a temperature monitor.
To charge the system, when the hub 300 is mounted to the side port 282 via attachment to connector 290, the elongated member 302 extends into lumen 296 to advance air through the air lumen into inner balloon 204 (or the pressure balloon in the embodiments with a single pressure balloon) to expand the balloon 204. That is, connection of the transducer hub 300 to the catheter 280 (port 282) automatically advances air through the connector lumen 296, the port lumen 283 and the first lumen 214 (or tubular portion of the balloon) to expand the balloon 204. (Such connection also automatically connects the temperature sensor to the hub 300). In some embodiments, 0.2 cc of air can be displaced/advanced by the member 102, although other volumes are also contemplated. Thus, as can be appreciated, mounting of the hub 300 to the catheter 280 automatically pressurizes the air lumen/chamber and expands the balloon. Note the balloon can be partially or fully inflated (expanded), dependent on the amount of air advanced into the balloon. Further note that preferably the lumen is not vented to atmosphere when the transducer hub 300 is attached and air is advanced through the air lumen. The port 282 includes a closable seal, e.g., valves 298 and 299, through which the elongated member 302 is inserted but maintains the seal when the elongated member 302 remains in the lumen 296. Note that catheter 280 is identical in all other respects to catheter 200 so that the description of catheter 200 and its components and function (and alternatives) are fully applicable to catheter 280, the difference being the connector 290 of catheter 292 to receive transducer hub 300. The transducer hub is also different, e.g., has latch arms and a different configuration.
In the alternative embodiment of
To disconnect (release) the hub 320, ends 317 of latch arms 316 are pressed radially inwardly to disengage from shoulder 326 so hub 320 can be pulled proximally out of connector 310.
In the alternate embodiment of
The housing (connector) 670 attached to catheter 600 has a proximal opening 672 and an internal channel (lumen) 680 (
Instead of valves as in the embodiment of
Hub 330 is mounted to connector 670 and includes a housing 331 from which a pair of distally extending snap fit connector arms 332 extend. The latch arms 332 are sufficiently flexible to enable attachment and have an enlarged distal portion 332a, illustratively shown as arrow shaped although other enlarged shapes could be provided. The elongated member (rod or nose) 334 extends between the latch arms 332. When the hub 330 is mounted to the connector 670, the elongated member 334 extends into the channel 680 of connector 670 to advance air though inner balloon tubular portion 609 to inflate the inner balloon 608. The enlarged ends 332a of latch arms 332 enter recesses 674 and engage shoulders 674a to retain the hub 330. The arrowhead tips of the latch arms 332 are at an acute angle to create a positive lock so that once engaged with the shoulder 674a, the hub 330 cannot be pulled proximally unless the latch arms 332 are released. To release (disconnect) the hub 330, the ends 332a are pressed radially inwardly to disengage from shoulder 674a and the hub 330 is pulled proximally. Note that alternatively a different number of latch arms could be provided. The distally extending thermistor pins 335 engage thermistor connectors within barrels 678 of connector 670 which have an interior stop to limit insertion of hub 330. This connection of the hub 330 and connector 670 limit lateral and longitudinal movement to ensure accurate pressure readings as lateral movement, for example, could change the pressure. The longitudinal movement is restricted in a proximal direction by the latch arms/shoulder engagement and in a distal direction by the full insertion of the thermistor pins 335 or alternatively by engagement of the hub 330 and housing 670 distal and proximal surfaces. The O-ring engagement of the elongated rod 334 limits lateral movement of the rod 334 and therefore restricts lateral movement of the hub 330.
Printed circuit board 336 containing the pressure sensor is mounted within hub 330. It can be bonded to a support within the hub. Note a digital pressure sensor can be used instead of an analog sensor in this embodiment as well as the other embodiments disclosed herein.
When transducer hub 330 is attached to connector 670, such attachment inserts the elongated rod 334 into lumen 680 to advance air though the tubular portion 609 of the inner balloon 608 and into the inner balloon 608 to inflate the balloon 608. The elongated member 334 also has a channel or lumen 337 (
In use, to charge the system, when the hub 330 is mounted to the catheter side port 620 via attachment to connector (housing) 670, the elongated member 334 extends into lumen 680 to advance air through the lumen into balloon 608 to expand the balloon 608. That is, connection of the transducer hub 330 to the catheter 600 (port 620) automatically advances air through the connector lumen 680 and tubular portion 609 to expand the inner balloon 608. Such connection also automatically connects the temperature sensor to the hub 330. In some embodiments, 0.2 cc of air can be displaced/advanced by the elongated member 334, although other volumes are also contemplated. Thus, as can be appreciated, mounting of the hub 330 to the catheter 600 automatically pressurizes the air lumen/chamber and expands the balloon. Note the balloon can be partially or fully inflated (expanded), dependent on the amount of air advanced into the balloon. Further note that preferably the lumen is not vented to atmosphere when the transducer hub 330 is attached and air is advanced through the air lumen.
Note the lumen which is used to inflate the pressure balloon and create the air column has an opening at a distal region to communicate with the interior of the pressure balloon. If an outer balloon is provided, an additional lumen can be provided in the catheter to communicate with the outer balloon to fill the outer balloon and an additional angled port (extension) at the proximal end of the catheter would receive an inflation device to inflate, either fully or partially, the outer balloon.
Note in each of the embodiments disclosed herein, air is described as the preferred gas for creating the column and expanding the balloon, however, other gasses are also contemplated for each of the embodiments.
The pressure balloons of the embodiments herein can be symmetrically shaped as shown or shaped such that a distal region has an outer transverse cross-sectional dimension, e.g., diameter, greater than an outer transverse cross-sectional dimension, e.g., diameter, of the proximal region. A smooth transition (taper) can be provided between the distal region and proximal region, although other configurations are also contemplated. The inner (and outer) balloon can by way of example be made of urethane, although other materials are also contemplated.
The wire connector of the foregoing embodiments can plug into the openings of a connector positioned on or in the hub. The wire connector can be internal of the hub with an opening in the wall of the hub to enable access for the wire connector. Also note that alternatively the wire can include a female connector and the hub can have a male connector. Other types of connectors/connections are also contemplated.
In alternate embodiments, any of the catheters disclosed here can include a pulse oximetry sensor to measure oxygen saturation in the urethral or bladder tissue. The sensor can be located either proximal or distal to the pressure balloon and/or either proximal or distal to the stabilizing balloon. It could also alternatively be mounted within one of the balloons.
It is also contemplated that in some embodiments a backup system be provided to determine pressure. The backup system can provide a double check of pressure readings to enhance accuracy. Such backup system can be used with any of the embodiments disclosed herein to provide a second pressure reading system. One example of such backup system is disclosed in
In use, catheter 160 is inserted into the bladder and stabilizing balloon 168 is inflated to secure the catheter 160 in place. The system is charged by inflation of the balloon 167, i.e., preferably partially inflated for the reasons discussed above, by insertion of air through side port 172 which is in fluid communication with the air lumen in a closed system formed by the internal space of the balloon 167 and the internal lumen 164 communicating with the internal space of balloon 167. With the balloon 167 inflated, pressure monitoring can commence as external pressure applied to an outer surface of the balloon 167 compresses the air (or other gas) within the chamber. The sensor 162 at the distal end of lumen 64 provides continuous pressure readings, converted to an electrical signal by the transducer within the distal end of lumen, and then electrically communicates through its transmission wires extending through the air lumen to an external monitor either directly or via a converter. Additionally, pressure within the air charged column is measured at a proximal region by sensor 169 within side port 172 of catheter 160. The sensor 162 at the distal end of lumen 164 provides continuous pressure readings, and such pressure readings can be confirmed by the proximal sensor. Such pressure readings can be performed continuously (along with continuous temperature monitoring) or alternatively can also be adapted if desired for periodic monitoring so the pressure and/or temperature readings can be taken at intervals or on demand by the clinician. Thus, air pressure readings at a proximal end plus microtip pressure readings at the distal end are provided. The sensors 162 and 169 can electrically communicate with an external monitor to display both pressure readings from sensors 162, 169, or alternatively, if the pressure readings are different, they can be averaged to display a single measurement. Clearly, other displays of information can be provided to display the information from the two sensors 162, 169.
The sensors disclosed herein can be microtip sensors within the air (or other gas) lumen or balloon. In alternative embodiments, fiber optic sensors within the air (or other gas) lumen or balloon can by utilized to transmit circumferential/area pressure. The pressure transducers can be housed within the catheter or alternatively external to the catheter. Additionally, core temperature sensors can be part of the pressure sensor or a separate axially spaced component.
The multi-lumen or single lumen catheters disclosed herein provide an air (or other gas) charged balloon (air containing chamber) giving precise readings of intra-abdominal pressure (or for other pressure measurements) and the systems are charged via insertion of air through a side port. The multi-lumen catheters are easily inserted into the bladder in the same manner as standard bladder drainage catheters and enable continuous drainage of urine while continuously recording IAP without interrupting urine flow and without requiring retrograde filling of the bladder with water. Thus, these catheters provide a closed system. The catheters also have a balloon providing a large reservoir (large capacity) and large circumferential area/interface for obtaining more information from the bladder over multiple reference points (rather than a single point sensor) that provides an average pressure to provide a more accurate assessment of the surrounding environment as pressure measurement is not limited to one side of the bladder but can determine measurements on the opposing side as well. The balloon can have a sufficiently large circumferential area so that it is in contact with the bladder wall, and in some embodiments, could distend the bladder wall, thus enabling pressure measurement without insertion of fluid into the bladder. When used in other body cavities for other pressure measurements, the pressure balloon of the multi-lumen or single lumen catheters disclosed herein can be of sufficiently large to contact or in some embodiments, distend the cavity wall, thus enabling pressure measurement without insertion of fluid into the cavity. The balloon, as noted above, of the multi-lumen or single lumen catheters disclosed herein can be impermeable or have an impermeable membrane (as defined herein) to prevent escape of gas to prevent loss of accurate pressure readings.
As noted above the catheters in some embodiments can be connected to a bedside monitor through either a wire or blue-tooth wireless connection. The system can also in some embodiments include an indicator or alarm system to alert the staff at the site as well as remote staff through wired or wireless connections to external apparatus, e.g., hand held phones or remote monitors.
As noted above, an alarm or indicator can be provided in some embodiments to alert the staff. The indicator can be a visual indicator such as a light, LED, color change, etc. Alternatively, or additionally, the indicator can be an audible indicator which emits some type of sound or alarm to alert the staff. The indicator can be at the proximal region of the catheter or at other portions of the catheter, e.g., at a distal end portion, where known imaging techniques would enable the user to discern when the indicator is turned on. It is also contemplated that in addition to providing an alert to the user in some embodiments, the pressure monitoring system can be tied into a system to directly reduce abdominal pressure so that if the pressure exceeds a threshold level (value), the abdominal pressure can automatically be reduced. In such systems, an indicator can be provided on the proximal portion of the catheter, e.g., at a proximal end outside the patient's body, or separate from the catheter. The sensor can be in communication with the indicator, either via connecting wires extending through a lumen of the catheter or a wireless connection. The sensor can be part of a system that includes a comparator so that a comparison of the measured pressure to a predetermined threshold pressure value is performed and a signal is sent to the indicator to activate (actuate) the indicator if the measured pressure exceeds the threshold pressure to alert the clinician or staff that pressure within the abdomen is too high and a signal is also sent to a device or system to automatically actuate the device or system to reduce the abdominal pressure. If the measured temperature is below the threshold, the indicator is not activated. A similar system can be used for temperature measurement and indication.
It is also contemplated that a micro-air charged sensor could be provided in the retention (stabilizing) balloon.
It is also contemplated that microtip sensors and/or fiber optic sensors can be utilized to measure pressure, and these sensors can be utilized instead of or in addition to the air pressure readings utilizing the aforedescribed balloon(s) for measuring pressure.
Pulse oximeters for measuring oxygen levels (oxygen saturation) in the urethral and/or bladder tissue could also be provided. In some embodiments, the pulse oximetry sensors can be positioned on the catheter proximal to the retention balloon. Alternatively, the sensors can be positioned within the retention balloon, on the catheter distal to the pressure balloon or on other regions of the catheter. Another channel in the catheter can be provided for the sensor and its connector to external devices, e.g. readers.
The catheters disclosed herein are designed for insertion into the bladder. However, it is also contemplated that they can be adapted for insertion into the rectum, colostomy pouch, stomach, supra-pubic bladder drain, or other orifice directly connected with the abdominal cavity. They can also be inserted into other areas connected with other cavities. Uses include by way of example, cardiac use, labor and delivery use, rectal placement for abdominal cavity, use for gastric pressure, esophageal motility, endocranial pressures ERCP, gall bladder, etc.
Although the apparatus and methods of the subject invention have been described with respect to preferred embodiments, those skilled in the art will readily appreciate that changes and modifications may be made thereto without departing from the spirit and scope of the present invention as defined by the appended claims.
This application claims priority from provisional application Ser. No. 62/865,360, filed Jul. 1, 2019 and from provisional application Ser. No. 62/771,040, filed Nov. 24, 2018. The entire contents of each of these applications are incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
62771040 | Nov 2018 | US | |
62865360 | Jun 2019 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 16675358 | Nov 2019 | US |
Child | 18127804 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 18127804 | Mar 2023 | US |
Child | 18640067 | US |