The human circulatory system includes arteries that direct oxygen-rich blood throughout the body. The veins are the blood vessels that return the oxygen-poor blood and waste products from the body back to the heart to be recycled through the lungs and liver. Veins include tiny valves that keep the blood moving back toward the heart, rather than collecting at an extremity.
Deep vein thrombosis (DVT) occurs when a blood clot forms in one or more of the deep veins of the body or when one or more of the valves in a vein has been compromised by a clot. DVT can develop from certain medical conditions that affect how the blood clots or that affect blood flow, typically in extremities such as the legs. DVT can be very serious because the blood clots can break loose, travel through the blood stream and lodge in another location, blocking blood flow to the body in that location.
DVT can occur when a person's legs remain still for long periods because the leg muscles are not contracting to help blood circulate. DVT can often occur during and as a result of surgery. It has been found that DVT conditions arise after a patient has been on an operating table for as little as 20 minutes. The DVT risk increases for prolonged recovery times after surgery during which the patient may spend the great majority of each day in bed. A treatment of choice to reduce the risk of blood clots and DVT is to get the patient up and walking as soon as possible after the surgery.
Another preferred treatment, usually in addition to walking, is the use of a compression device that is wrapped around the extremity, usually the lower leg. The compression device applies intermittent compression to the limb to promote blood flow through the veins back to the heart. The cyclic compression can also promote the natural release of substances in the body that help prevent clots. The typical DVT compression device is a pneumatic device that pumps air into a hollow cuff encircling the affected limb to apply pressure to the limb. This pressure squeezes the veins, forcing blood out of the veins toward the heart. The pressure is released by venting the cuff, allowing it to deflate. This cycle of inflation and deflations continues for as long as the cuff is worn by the patient.
For DVT prevention, patient compliance is a necessity, meaning that the patient wears an active DVT cuff for the prescribed time and the patient leaves the hospital bed to walk for a prescribed duration. However, patient compliance is often very problematic. One problem is that a DVT cuff is uncomfortable to wear for extended lengths of time, yet the recommendations to prevent DVT can exceed in upwards of 18 hours a day. Some DVT cuffs include means for monitoring the amount of time the cuff has been activated and run through its pressure cycle. However, some patients—particularly patients for whom the DVT cuff is prescribed for home care—find ways to “trick” the DVT cuff by mounting the cuff on a rigid object and allowing the cuff to inflate and deflate on the inanimate object.
Another problem is that the DVT cuff is not conducive to patient mobility. The typical DVT cuff requires a source of pressurized air to inflate the cuff during the pressure cycle. Early systems utilized a large pump unit that sat on the floor next to the patient's bed. Smaller pumps were later developed that could be carried by the patient. However, many patients, particularly elderly patients, lack the strength and/or stamina to carry around a pneumatic pump connected to a DVT cuff worn on the patient's leg. Moreover, the pneumatic hose between the pump and the cuff can be an entanglement nuisance.
There is a need for a compression device that is particularly suited for DVT prevention and that is mobile. There is also a need for a compression device that can ensure patient compliance, or at least ensure that the non-compliant patient cannot “trick” the DVT cuff into appearing to have been properly used.
A compression device comprises a disposable wrap that is configured to be wrapped around the limb of a patient, and a reusable controller that is removably mounted to the disposable wrap. The controller is a non-pneumatic device that is operable to contract the wrap around the patient's limb in a controlled fashion and according to a predetermined compression protocol. In one aspect, the compression protocol is adapted as a prophylaxis for deep vein thrombosis, although other compression protocols are possible.
In one aspect, the controller includes a DC motor and transmission to gear down the rotational output speed of the motor to a speed suitable for use in contracting the wrap. The wrap is connected at a looped end to a D-ring connected to a pull strap that is in turn mounted to a pulley that rotates with the motor to wind the pull strap at least partially around the pulley. The opposite end of the wrap includes a controller mount that allows for removable mounting or attachment of the controller to the wrap. In one embodiment, the controller mounting arrangement includes a load cell at the interface between the wrap and the controller that is configured to measure a tension force generated as the wrap is tightened on the patient's limb. In one specific embodiment, the controller mounting arrangement utilizes a load cell axle engaged within a pair of clips affixed to the wrap. In another specific embodiment, a keyed hinge arrangement is provided between the wrap and a housing of the controller. The controller mounting arrangement is configured to allow the controller to be removed from the wrap and replaced with another controller as desired.
The controller can include an accelerometer or position sensor to sense the physical position and movement of the patient. Data from the accelerometer or position sensor are provided to an on-board microprocessor that generates compliance data that can be uploaded or displayed on a display screen of the compression device.
In another feature, an RF chip or tag is provided on the wrap that can be specifically associated with a patient. The controller includes an RF sensing circuit that detects the RF chip and reads information from the chip, including a unique identifier. Concordance between the unique identifier on the chip and a data base of known valid identifiers maintained in the controller is required before the controller is operable. The unique identifier associated with the wrap, and thus with the patient, follows the wrap regardless of which controller is mounted to the wrap. This feature allows the same wrap to be recognized as the patient moves from one unit of a hospital to another.
The compression device of the present disclosure is a non-pneumatic wearable device that permits patient mobility. Thus, the patient is not restricted to a hospital bed or chair during a compression protocol. Moreover, the sensors and microprocessor of the controller is configured to monitor the amount of time that the patient spends laying down/reclined, seated/standing or moving while wearing the device. The controller displays information indicative of the manner of activity while wearing the device.
In another feature of the present disclosure, the non-pneumatic mobile compression device disclosed herein is configured to apply a compression profile that reduces the risk of DVT. In particular, the controller of the device is configured to apply compression to the patient's limb/leg that achieves a blood flow velocity that has been found to reduce or eliminate the risk of DVT. The device is operable to generate a blood flow velocity that is about three times greater than the baseline velocity of the patient.
In a further feature of the compression device, the device is configured so that the compression applied to the patient's limb is produced by contact between the patient and the disposable wrap and not by contact between the housing of the controller and the patient. In particular, the controller is provided with a housing that is curved in the surface facing the patient's body that avoids contact between that surface and the patient. This attribute not only prevents the application of pressure to the patient by a rigid body, it also provided an air flow passageway to dissipate heat generated by the controller and improve comfort of the patient.
For the purposes of promoting an understanding of the principles of the disclosure, reference will now be made to the embodiments illustrated in the drawings and described in the following written specification. It is understood that no limitation to the scope of the disclosure is thereby intended. It is further understood that the present disclosure includes any alterations and modifications to the illustrated embodiments and includes further applications of the principles disclosed herein as would normally occur to one skilled in the art to which this disclosure pertains
A compression device 10, shown in
The wrap includes a flap 17 fastened at one end to the wrap 12. The flap is arranged beneath the controller 14 and can operate to protect the patient's skin from any heat generated by the controller 14 or by patient's skin. The flap 17 may be formed of the same material as the wrap 12, or may be formed of a different material adapted to cushion the skin from pressure induced the controller and/or heat from the controller or the patient's skin. When the wrap 12 encircles a limb the flap 17 is not applying any pressure to the limb since it has a free end beneath the controller 14.
The controller 14 includes a base plate 42 and a cover 44 that contains the drive components and electronics of the cuff. The cover 44 can be fastened to the base plate 42 at a plurality of latches 47 preferably located at the corners of the plate, as shown in
The wrap 12 includes an end loop 24 that is configured to be removably wrapped around a D-ring 22 connected to the controller 14. The end loop 24 can include releasable facing surfaces, such as a hook-and-loop or VELCRO®-type fastener, so that the wrap can pass through the D-ring and overlap itself to form the end loop. It can be appreciated that the releasable facing surfaces can have a length sufficient to allow varying amounts of overlap. This allows the DVT cuff to be snugly wrapped around the patient's limb, regardless of the size of the patient.
The opposite end of the wrap includes a mounting arrangement 40 that includes a pair of clips 60 affixed to a mounting plate 61, as best seen in
The clips 60 are configured to removably receive an axle 58, and in one embodiment can be in the form of spring clips or the like that can be elastically pushed to allow entry of the axle into the clip. The clips are sufficiently flexible to allow the axle to be pushed into the clip, but also sufficiently strong to prevent the axle from being dislodged during a compression cycle of the cuff 10. The axle provides a connection to a load cell 57, as best seen in the enlarged view of
The other end of the wrap that includes the end loop 24 is connected to the D-ring 22, which is itself connected to a pull strap 20 that passes through a slot 46 in the housing 44, as shown in
As noted above, the load cell 57 provides one connection interface between the controller 14 and the wrap 12 that is encircling the patient's limb. Since the axle 58 is retained on the wrap by the clips 60, the axle, and thus the pull bar 30 is pulled by a circumferential force as the wrap is tightened around the circumference of the patient's limb. This force thus tends to bend the load cell plate 57 since one end of that plate is fastened to the pull bar and the other end is essentially cantilever mounted to the base plate 42 of the controller 14. As the plate bends, the strain gage 57c mounted to the surface of the plate elongates. The strain gage 57c is connected by wires 57d to the electronics of the controller that is configured to interpret the measured strain, and convert this measured strain to a force value.
In an alternative embodiment, the load cell 57 is eliminated in favor of a direct mount between the pull bar 30 and the controller 14, or more particularly the base plate 42 of the controller. In this embodiment, the circumferential force generated in the wrap as it is tightened about the patient's limb can be determined by a motor-related sensor. One such sensor can be a current sensor for the motor 32 that measures the current through the DC motor. The current required to maintain the motor rotational speed (at a given voltage) is a measure of the resistive force from the wrap as it is tightened. The current sensor can be connected to the electronics of the controller that is configured to interpret the measured current and convert this current to a force value.
In one feature of the DVT cuff, the controller 14, and particularly the base plate 42, defines a curved surface 45 facing the patient's limb when the cuff is wrapped around the limb, as best seen in
Returning to the drive train for the controller 14, the pulley 34 can be coupled to the motor 32 by way of a transmission 33 that is configured to reduce the rotary speed and increase the torque of the output driving the pulley. In one specific embodiment, the transmission can be configured for a 488:1 speed reduction. For a DVT device, a certain compression protocol requires a no-load output speed of at least 30 rpm and a torque of at least 310 in-oz. The motor specifications and the reducer drive train of the transmission can be selected to achieve these output characteristics.
The motor 32 is driven by control circuitry 50 that controls the activation of the motor to wind and unwind the pull strap 20 about the pulley 34. The control circuitry thus includes a microprocessor 52 and a motor controller 53. The microprocessor includes one or more stored programs that control the motor controller according to a compression profile and that control the transfer of data to and from the controller 14. The control circuitry 50 can include a pulley sensor 53, electrically connected to the microprocessor, which is configured to determine the position of the pulley as it rotates to wind and unwind the pull strap 20. The load cell 57 (or current sensor in the alternative embodiment) is also electrically connected to the microprocessor and is configured to provide a measure of the tension in the wrap 12, which is directly related to the amount of compression applied to the patient's limb. For certain features of the DVT cuff 10, the control circuitry can also include an accelerometer 55 electrically connected to the microprocessor and operable to provide motion data indicative of the position, attitude and movement of the patient.
The cuff 10 is provided with a visual display 15 in the cover 44 that is also connected to the microprocessor. The display 15 can display information regarding the operation of the cuff and/or indicative of the compliance of the patient wearing the cuff. In one aspect, the display can be a touch screen device that allows medical personnel to scroll through different screens displaying different information. The display 15 can be an electronic paper or E-ink display that reduces the power requirements for maintaining the display. A battery (not shown) is contained within the controller 14, such as in the space between the microprocessor 52 and the base plate 42, to provide electrical power to all of the electrical components of the control circuitry 50. The battery is preferably rechargeable. The controller can include a jack for receiving a cable for connecting to a charging station, or can include circuitry permitting proximity charging of the battery.
In a further feature of the disclosed DVT cuff, the control circuitry 50 includes an RF (radio frequency) sensor 56 in communication with the microprocessor 52. The RF sensor 56 is configured to detect an RF chip 65 integrated into the wrap 12. In one embodiment shown in
The RF chip 65 is also configured to store data regarding the operation of the DVT cuff 10 and the patient's compliance. In one aspect, the chip is provided with sufficient memory to store data continuously for 30 days. The microprocessor of the controller 14 is configured to upload the stored data from the RF chip, via the RF circuit 56, into an on-board memory within the microprocessor 52. It is noted that the controller can be configured to limit the cumulative data displayed to the preceding 48 hour period, rather for the entire 30 day period stored in the RF chip memory.
An alternative embodiment of the DVT cuff is shown in
The hinge beam 75 is mounted between a pair of mounts 76 projecting from the base plate 42′ of the controller 14′. The hinge beam 75 is configured as a rectangular beam, as described above, for introduction into and rotation within the keyed slot and channel 73a, 73b. The controller 14′ can be otherwise configured like the controller 14, including the curved base plate 42′ and the cover 44′ defining a pull strap slot 46′ through which the pull strap (not shown) extends. The drive mechanism and control circuitry 50 can be the same for the controller 14′ as in the controller 14. However, in this embodiment, since the controller 14′ is mounted to the wrap by way of the keyed hinge interface, the cuff 10′ does not include the load cell feature of the cuff 10 that is configured to determine the load or force applied to the patient through the cuff. Instead, in this embodiment, the motor can include the current sensor discussed above that is used to determine the motor current during compression, to thereby determine the tension force in the wrap, which correlates to the compressive force applied to the patient's limb.
The wrap 12′ includes an RF chip 65′ similar to the RF chip 65 of the wrap 12. However, in this embodiment, the chip 65′ can be mounted on or embedded in the mounting pad 70. The chip 65′ is thus positioned, like the chip 65, to be detected by the RF circuitry 56 of the control circuitry.
The mounting pad 70 can incorporate ventilation openings 71. Similarly, the flap 17′ may also incorporate ventilation openings or perforations, such as the openings 71. In this specific embodiment, the flap 17′ is not formed of the same breathable material as the wrap 12′, but is instead formed of a semi-rigid but pliable material, such as a low-density foam, in particular a PORON® foam. The flap formed of the low-density foam can have a basic shape that follows the curvature of the patient's limb, but is pliable enough to flex as needed to avoid exerting pressure on the skin. In this instance, the ventilation perforations 71 in the flap 17′ are beneficial to provide air flow to the patient's skin in contact with the flap. Although the openings or perforations 71 are shown as circular, they could have other configurations, such as elongated slots through the pad 70 and flap 17′.
In both embodiments of the DVT cuff shown in
The controller is not intended to be disposable, but is instead reusable with every authenticated and authorized cuff. Since the controller is not specific to any particular cuff it is capable of being used with a number of cuffs, which is particularly useful in a hospital setting. Since the DVT cuff is not continuously worn and used by a patient, a single controller can be used to control the compression protocol for a number of patients, with each patient being uniquely identified by the cuff 12, 12′ issued to that patient. The cuff remains with the patient at all times, but the controller can be maintained in a separate storage unit. In a hospital, each ward or unit of the hospital can have its own collection of controllers, all capable of being used interchangeably with all patient-specific cuffs in every ward or unit of the hospital. Thus, a patient undergoing surgery may wear a DVT cuff that is operating during the surgery to prevent the onset of DVT condition. When the surgery is complete, the controller is removed and kept with the surgical unit, and the patient is transferred to a recovery ward or ICU where a controller maintained by that ward or unit can be engaged to the patient's cuff to continue DVT preventative treatment during recovery. If the patient is moved to a longer-term care room, the recovery ward controller is removed and the controller maintained by the care ward is engaged to the patient's cuff. When the patient is released but DVT treatment is still prescribed, the patent can take his/her assigned cuff 12, 12′ home together with a separately prescribed controller for home use. Once the treatment is complete or the risk of DVT has passed, the patient can dispose of the cuff and return the controller 14, 14′ to the medical facility.
In one feature of the present disclosure, a kiosk 80 can be provided that includes a number of bays 82 for storing several controllers 14, 14′, as shown in
The kiosk can also include a module 88 for use in charging the replaceable batteries. Another module 87 can incorporate disinfection equipment, such as a UV-C lamp, that can aid in the disinfection of a controller after each use. The kiosk may also include a number of bays 85 for storing new wraps 12, 12′ for initial distribution to a patient.
Returning to the controller 14, 14′ associated with a wrap 12, 12′, the microprocessor 52 can execute software or firmware that monitors various attributes of the DVT cuff and the patient and then displays pertinent information on the display 15. An exemplary data display is shown in
As reflected in
All of this information gives the medical personnel or care-giver a complete picture of the patient's compliance with the compression protocol and mobility regimen. “Early Mobility” or “Progressive Mobility” programs have been found to lower the incidence of hospital-acquired or recovery-acquired events, including not only DVT but also pressure ulcers and infections. Mobility protocols have also been linked to reductions in length of stay at the hospital, re-admission rates and overall costs of stay. The ready availability of patient compliance and activity information can allow the medical personnel to address deviations from the recommended prophylactic protocol.
The DVT cuff can be placed on the patient's limb, such as his/her leg, as described above. The end loop 24 can be used to slightly tighten the wrap 12, 12′ around the leg, with sufficient tightness to hold the cuff in place. A power switch (not shown) on the controller 14, 14′ is actuated to activate the microprocessor 52 and initiate a start-up screen on the display 15. The microprocessor first checks the pulley sensor 54 to determine whether the pulley 34 is in its proper initial or “home” position. If not then the microprocessor will direct the motor controller 53 to operate the motor 32 in an “unwind” direction, such as counter-clock wise, The motor remains energized until the pulley sensor 54 detects the pulley at its home position.
Once the pulley is homed, the microprocessor prompts the operator with a display of a “Pretension” button on the touch screen display. When the operator presses the “Pretension” the microprocessor sends a command to the motor controller to set the motor rotational direction to the “wind” direction, namely clockwise in the present example. The microprocessor then sends a second command to the motor controller to energize the motor and set the motor speed to a pre-determined speed, preferably a mid-range rotational speed for the motor. As the motor operates the transmission 33 reduces the motor rotational speed to a suitable mid-range speed for the pulley, such as 10-15 rpm. As the pulley retracts the wrap, the microprocessor monitors the force applied to the wrap via the load cell 57. Alternatively, or in addition, the microprocessor can monitor motor current, as discussed above, which varies as a function of the load applied to the wrap (or more precisely the reaction load experienced by the controller). When a minimum pre-tension force is achieved, approximately 1 pound in a specific example, the microprocessor directs the motor controller to stop the motor and hold the pulley at its current location. The wrap is thus pre-tensioned at a known amount of compression on the patient's leg. In one embodiment, a new home position of the pulley can be set corresponding to the position of the pulley in the pre-tensioned state of the wrap.
With the wrap and controller properly installed and the desired pre-tensioning achieved, the microprocessor issues a notification on the display 15 that the compression protocol will begin. In one exemplary embodiment of the compression cuff 10, the compression protocol can be for DVT prevention. However, it is understood that other compression protocols are contemplated and can be readily executed with the present cuff. The series of instructions from the microprocessor 52 to the motor controller 53 are generated by software/firmware executed by the microprocessor. This software can be configured as a generic series of commands that read compression variables from a stored database, such variables including on-off times, dwell times, power levels and the like. This database can be contained within the memory of the microprocessor or downloaded from a remotely stored database. As a further alternative, the software itself can be application specific with all of the protocol-specific variables hard-wired into the software commands. It is thus contemplated that the database of variable and/or protocol specific software can be patient-specific and incorporated into each controller 14 being used by the particular patient. In this respect, the variables database can be stored in the RF chip 65, 65′ associated with the patient's cuff, and then uploaded into each controller 14 connected to the patient's cuff.
Returning to the operation of the drive system for the cuff 10, when the compression protocol begins, the microprocessor sends a command to the DC motor controller circuit to set the motor direction to clockwise and to set the power value for the motor to full power. In one embodiment, the motor controller is a pulse-width modulated controller, in which case the full power mode corresponds to a PWM input of 254 for a 100% duty. In one specific embodiment, in the full power mode of the motor, the pulley rotates at approximately 30 revolutions per minute (rpm) with a torque of 310-inch ounces. During compression, the microprocessor continuously monitors the force of the compression wrap via the load cell 57 (the DC motor current). When the force being applied to the wrap equals the pretension force plus a pre-determined offset force, such as about 7 lb. in one example, the microprocessor sends a “stop” command to the DC motor controller which de-energizes the DC motor. The microprocessor holds the position of the pulley for 500 milliseconds. After the “hold”, the microprocessor sends a set counter-clockwise motor direction command to the DC motor controller and sets the motor power to “low” speed, which can correspond to a PWM input of 60 for a 25% duty cycle. As the motor turns counter-clockwise, to release the pull strap 20 and relieve the wrap compression, towards the home position, the microprocessor monitors the force until the pretension force is met, after which the microprocessor sends a stop command to the DC motor controller. In an alternative embodiment in which a new home position of the pulley is reset corresponding to the pre-tensioning position of the pulley, the microprocessor can monitor the pulley sensor and send a stop command when the pulley reaches the updated home position. After the stop command is executed, the microprocessor updates the compression duration time and resets the cycle timer to zero. When the cycle timer reaches a pre-determined dwell time, such as 60 seconds, the compression process is re-played.
As described above, the compression achieved by the DVT cuff is effectuated by a small DC motor 32 within the controller 14. The cuff 12, 12′ is fastened at one end to the housing of the controller, either directly or via a load cell 57 as described above. The opposite end of the cuff, the end loop 24, is connected to the D-ring 22 at the end of the pull strap 20. The pull strap is fastened to the rotating pulley 34 so that rotation in one direction, such as clockwise, causes the pull strap to wind around the pulley. As the strap winds it pulls the D-ring, which pulls the wrap 12, essentially shortening the effective length of the wrap and tightening it around the patient's limb/leg.
As mentioned above, the microprocessor 52 of the controller 14, 14′ can be programmed to many different compression protocols. In the exemplary embodiment, the cuff 10 serves as a DVT cuff for the prevention of deep vein thrombosis in a patient's limb, particularly the leg. In order to prevent DVT the goal is to push the blood up the femoral vein toward the heart. However, simply exerting pressure on the lower leg and pushing blood toward the heart has not been found to eliminate the risk of DVT. Instead, achieving a particular flow velocity through the femoral vein is essential to good DVT prophylaxis. In particular, a flow velocity that is about three times the baseline flow velocity through the femoral vein for the patient has been found to be effective in preventing DVT.
In one aspect of the present disclosure, an optimum compression protocol for DVT prevention has been developed for implementation in the non-hydraulic compression cuff disclosed herein. The graph shown in
In the second stage, or the first stage of the repeated compression protocol, the motor is driven at its maximum speed for less than one second until a predetermined maximum tension force in the wrap is reached. In one embodiment, this maximum force can range from 5.5-6.5 lbf greater the pre-tension force, corresponding to a maximum tension force in the wrap of between about 6.5 lbf to about 7.5 lbf (for a 1.0 lbf pre-tension). It has been found that the requisite upward blood flow of three times the normal flow femoral vein flow rate or velocity is achieved not only by the amount of compressive force applied to the limb by the tensioning of the wrap, but also by the rapidity of the application of that compressive force. Thus, in the exemplary embodiment, the DVT cuff achieves the maximum applied force in less than about one second. This pressure is maintained for the hold segment shown in
The fourth segment, or third segment of the compression cycle, relieves the tension force in the wrap, and thus the compression force on the patient's limb, but does so gradually to allow the blood flow velocity to return to the normal baseline velocity for the patient. Thus, the motor is reversed and driven at about one-fourth of the motor speed during the third segment of the repeated compression cycle. In the illustrated embodiment, the motor is driven at about a 25% duty cycle over a period of about three seconds. At the end of the release segment, the DVT cuff is returned to its pretensioning state (1.0 lbf in the embodiment) and the motor is de-activated for a predetermined dwell time before another compression, hold, and release cycle is commenced. As described above, this dwell time can be about 60 seconds. The controller 14, 14′ repeats these three segments for a prescribed treatment period, which can range from 15 minutes to an hour, or from 15-60 compression cycles, depending on the patient needs. With each compression cycle (compression, hold and release) the blood velocity follows the profile shown in
It is noted that the graph on
As described above, the DVT cuff 10, 10′ includes a removable and replaceable controller 14, 14′ that includes control circuitry 50 for controlling the operation of the cuff, namely the pre-tensioning and compression stages as well as data collection and retrieval. The control circuitry 50 includes a microprocessor 52 and associated digital memory that includes software and/or firmware that controls the operation of the cuff.
In the next step 208 a display is provided that allows the operator to select from the two operational modes of the DVT cuff—mobility and DVT prophylaxis, or DVT prophylaxis only. In both modes the DVT compression protocol is enabled, but in the first mode the patient is expected to move apart from the hospital bed. The selection of the mode depends on the patient treatment protocol. If “mobility & DVT” is selected the controller sends the display to the screen in step 209 that allows the operator to enter an elapsed time for use of the DVT cuff in the mobility mode. Once the mode has been selected the controller displays that the controller is ready for use in step 210 after which the controller powers down in step 211.
The flowchart 300 is provided in
It is understood that conventional Bluetooth pairing technology can be implemented between the controller and the kiosk. It should also be understood that the pairing step requires activation of the kiosk according to the flowchart 302. Thus, when the kiosk data processor 84 is activated an initial set-up screen is displayed in step 313 that allows the operator to set the date and time and then activate the pairing sequence in step 314. A pairing screen is displayed on the kiosk processor 84 as shown in step 315 in which a table of uniquely identified cuff controllers within the vicinity of the kiosk are detected. The user can select the appropriate controller for pairing, after which a successful pairing is displayed in step 316.
The flowchart 400 in
On the other hand, if the RFID is authenticated the controller writes a start date and time to the RF chip 65 of the wrap 12, 12′ and stores the identifier of the wrap in the memory of the controller 14, 14′. The controller checks in step 407 whether the two writes were successful, and if not generates an error message in step 408 and returns controller to the initial step 402. If the writes were found to be successful in step 407 then program flow proceeds to step 409 in which the pre-tensioning of the strap is conducted. In this first step, the patient, or preferably the medical personnel, adjusts the end loop 24 of the strap on the D-ring 22 of the controller 14, 14′ to initially tension the wrap on the patient's limb, typically the leg. In the first step 410 the controller measures the force in the wrap and determines whether the proper amount of pre-tensioning, or tightness, of the wrap has been achieved. In one specific embodiment, that force value is 1.0 lbf, which has been found to be an optimum starting tension for the DVT prophylaxis protocols. If the amount of pretension is not at the desired value, the controller seeks to determine whether the wrap is too loose or too tight in step 411. If it is too loose a message is displayed in step 412 and if too tight a commensurate message is provided in step 413. In step 414 the patient or medical personnel adjusts the end loop 24 on the D-ring 22 to adjust the pre-tension of the cuff. This process continues until the wrap is properly tensioned. In an alternative embodiment, if the wrap is less than the desired pre-tension force by a pre-determined amount, the controller can activate the motor 32 to pull the D-ring 22 until the requisite pre-tensioning force is reached. Of course, if the current wrap force is greater than the desired pre-tension force the motor cannot relieve the tension in the wrap—only adjusting the loop on the D-ring can reduce the initial tension in the wrap.
Once the amount of pre-tension or initial force has been achieved the controller initiates the DVT protocol in step 415 and displays a “DVT Prophylaxis Running” message on the controller screen. In step 417 it is determined whether the DVT cuff is to be operated in the DVT-only mode or in the DVT+mobility mode. This determines whether the “DVT Prophylaxis Running” screen continues in step 418 or whether additional displays for the mobility function are displayed in step 419 (see
As explained above, the DVT cuff of the present disclosure contemplates the removal and replacement of a controller from the wrap of a particular patient. The present disclosure also contemplates removing a current wrap for a patient and replacing it with a new wrap. After an extended period of use a wrap may become soiled with sweat or other fluids so that a new wrap is required. The wraps disclosed herein are intended to be disposable so there is no particular benefit to removing, cleaning and replacing a particular wrap, especially in a hospital setting. The method for changing a given wrap for a new wrap is illustrated in the flowchart 500 of
The mobility displays are provided in the flowchart 600 of
With respect to patient compliance, as discussed above compliance to a DVT protocol is often problematic. Likewise, determining the level of patient compliance has always be difficult, often requiring first-hand knowledge of the medical personnel as to whether the patient has engaged in the requisite physical activity and activated the DVT cuff according to the prescribed protocol. The DVT cuff 10 of the present disclosure provides the medical personnel with significant information to assess the level of compliance for a particular patient. In addition to the various displays described above, the pre-tensioning steps also assure compliance. If the cuff is not properly wrapped on the patient's limb with the proper amount of pre-tension force, the controller will not allow the DVT prophylaxis sequence to commence. The RF chip of the patient's wrap can store time and date information regarding the starting and completion of a DVT prophylaxis sequence, information that can be accessed by the medical personnel to verify patient compliance. Moreover, the controller can display information indicative of patient compliance, such as the “DVT Prophylaxis Running” message (see steps 416, 418 in
Once a particular controller is no longer in use the controller can be stored, such as in the kiosk 80 described above. In this instance, the controller and kiosk follow a flowchart 700 for the storage of the controller. The flowchart 701 for the controller includes a first step 703 for storing the controller under two scenarios. In the first scenario, the controller is set-up and ready for use, while in the second scenario the controller has just been used by a patient. In both cases the controller is turned off and the replaceable battery removed for placement in a charging station in one of the bays 82 of the kiosk 80. In step 704 it is determined whether the controller contains patient data uploaded from the RF chip of the patient's wrap. If no, then control passes to step 705 in which a “Ready for Use” message is displayed in step 706. It is noted that in one embodiment of the present disclosure the display 15 is an e-ink display so that the “Ready for Use” display remains on the screen even when the controller is powered down.
If the controller includes uploaded patient data the controller displays a message in step 707 and activates the wireless or Bluetooth communications between the controller and the kiosk in step 708. The controller times out after a predetermined “connection” time and determines whether the data was successfully downloaded to the kiosk in step 709. If not then the controller returns to steps 707, 708 to attempt the download again. If the download was successful, the controller clears its memory of the patient data, resets any control variables that may have been modified, activates the “Ready for Use” display in step 706 and powers down the controller.
On the kiosk side 702 of the flowchart 700, the kiosk processor displays the selection screen in step 711 in which the user can select “Service” to move to the display in step 712. This screen allows the user to select from the service functions of uploading patient data, updating the controller software/firmware or updating the kiosk software/firmware. For the controller storage, the user selects uploading the patient data and the kiosk processor automatically connects with the previously paired controllers in step 713. The automatic download process occurs in step 714 followed by a message on the kiosk processor that the download was complete, along with the identifier for the particular controller. It is understood that multiple controllers can be stored in a given kiosk so the downloads may be from multiple controllers. The patient-related data is maintained in a memory of the kiosk processor for subsequent review and/or processing by medical personnel. The kiosk may be paired with another device, other than a DVT cuff controller, which allows downloading of patient data to the device, such as a smart phone or smart pad, which can be reviewed by the medical personnel.
At the end of a DVT session by a patient, it is desired to remove the controller from the wrap associated with the patient. Flowchart 800 in
The present disclosure should be considered as illustrative and not restrictive in character. It is understood that only certain embodiments have been presented and that all changes, modifications and further applications that come within the spirit of the disclosure are desired to be protected.
Number | Name | Date | Kind |
---|---|---|---|
7354407 | Quintana | Apr 2008 | B2 |
20170252252 | Wyatt | Sep 2017 | A1 |
20170312161 | Johnson | Nov 2017 | A1 |
20200253814 | Freidenrich | Aug 2020 | A1 |
20200261303 | Bennett | Aug 2020 | A1 |
Number | Date | Country |
---|---|---|
WO-2015089383 | Jun 2015 | WO |
Number | Date | Country | |
---|---|---|---|
20200315905 A1 | Oct 2020 | US |