Fiberoptic Raman spectroscopy probes used in conjunction with autofluorescence endoscopes can improve the specificity for cancer detection compared to using autofluorescence alone. Autofluorescence in human tissue is generally blue to blue-green when excited with UV to violet wavelengths. Precancerous and cancerous tissue does not fluoresce as strongly as normal tissue and thus shows up in the visual field of the endoscope as a relatively dark patch in a brighter field. The method is very sensitive but not always specific for cancer. Fiberoptic Raman probes can be passed through the biopsy channel of the endoscope and pressed against these darker areas of tissue for an independent spectral diagnosis. The combination of the two diagnoses is more reliable than either one alone.
Raman scattering is sensitive to the concentrations of specific chemicals in the tissue which change when the tissue becomes cancerous. In the Raman process some energy is absorbed from an incident photon and converted into vibrational motion of the molecules so that the scattered photon is red-shifted in wavelength. With narrowband incident excitation, such as from a laser, the result is a “fingerprint” of narrow lines representing different molecules in the tissue with different vibrational energies. The scattering process is quite weak, however, so the amount of the Raman scattered light is low. The Raman peaks are typically superimposed on a larger background of broadband, relatively featureless tissue fluorescence. Raman scattering at near-infrared (NIR) wavelengths is particularly useful since the background noise due to tissue fluorescence is lower at NIR wavelengths than at visible wavelengths.
Most optical tissue diagnostic algorithms require comparisons of an acquired spectrum with a similar spectrum from a different tissue site, perhaps taken earlier in the same patient from a site known to be normal. Spectra are often compared to a large set of spectra which have been correlated with pathology results during the development of those algorithms. Sometimes the absolute amplitude of the overall spectrum does not matter But that is not generally the casein the lung, for instance, the broad, featureless tissue fluorescence that complicates Raman spectra is typically stronger in normal lung tissue than it is in dysplastic lung tissue. Diagnostics which depend on relative signal comparisons like this require a stable measurement system and repeatable probe placement techniques.
The design of practical Raman fiberoptic probes makes repeatable probe placement onto the tissue particularly important. Particular lens can result in a rapid falloff in collection efficiency with increasing distance from the tissue. Fractions of a millimeter in probe to tissue spacing can significantly change the level of the acquired signal. This can make diagnostics which depend on relative measurements less reliable if contact is uncertain.
The low intensity of the Raman scattered light means that the white light illumination of the endoscope's visual field must be reduced or turned off during Raman data acquisition. Longer wavelength light leaking into the probe can add noise to the Raman spectra. The low signal levels also result in typical data acquisition times of 1 second or longer. Low illumination and long acquisitions add to the clinician's difficulty in properly placing the probe, maintaining probe position and maintaining probe contact. Involuntary tissue motion and tissue folds which may hide the tip of the Raman probe further complicate the positioning of these probes.
The present invention describes a method and apparatus by which effective contact between a fiberoptic, spectroscopic probe and tissue can be verified and monitored both before and during the acquisition of spectroscopic data from a tissue area being probed. Effective contact is preferred for the relative comparison of measurements between different tissue sites and for comparisons with the previously-acquired data sets used in diagnostic programs. A monitoring system for probe contact can thus be used to indicate the likely reliability of a given diagnostic result or its information may be used to normalize integrated spectral signatures back to standard levels when probe contact is occasionally lost during a long integration. These normalized signatures can be used for a more reliable diagnosis rather than being discarded.
In a preferred embodiment the monitor system couples light into the excitation fiber at the proximal end of a fiberoptic probe along with any illumination or excitation light required by the probe for spectroscopic purposes. The fiberoptic probe delivers the combined monitor and excitation light to the tissue at the distal end of the probe. A fraction of the monitor light scattered from the tissue is collected by the probe and returned to its proximal end by means of a number of collection fibers along with the desired spectroscopic signature. This returned monitor light is separated from the light required for spectroscopic purposes and passed to a photodetector to be quantified. Since the monitor light is relatively strong relative to the Raman signal a single collection fiber can also be dedicated to the contact monitor and coupled directly to a photodetector.
The monitor light returning from the distal tip of a probe consists of two components. The first component is due to Fresnel reflections from the glass/air interface at the distal tip of the probe itself. This Fresnel component will decrease in contact with a water or a wet absorbing surface due to the reduced reflectivity of a glass/water interface. The second component is due to diffuse reflection from tissue near the distal tip and drops off rapidly with increasing distance of the probe tip from the tissue. The relative changes in the size of these two components depend upon the specific design of the optics at the distal tip of the probe, the wavelength of the light used for the contact monitor and the nature of the tissue surface. Either or both signals may be used to detect probe contact.
For a standard, forward-looking Raman probe with a ball lens at the distal tip the tissue reflectivity signal dominates the Fresnel reflection signal upon tissue contact. This is particularly true if the monitor light source is chosen with a wavelength which is not absorbed significantly by hemoglobin in the tissue (>630 nm) and if the photodetector optics are designed to reject the higher angle light reflected from the concave surface at the distal end of the probe. As a probe is slowly lowered onto tissue with a thick water layer the signal will have an initial value followed by a somewhat lower value as the tip touches the water surface followed by a significantly higher value when the tip finally touches tissue.
For forward-looking Raman probes designed with flat optical surfaces at the distal tip the falloff of the diffuse reflection signal with tissue distance is less rapid. For side-looking Raman probes designed to be inserted through biopsy needles the tissue may remain very close to the probe window with a narrow air gap that should be avoided. In these cases it may be advantageous to maximize the Fresnel reflection relative to the diffuse tissue reflection and detect a reduction in the signal upon contact with a water film on the tissue surface. The monitor wavelength chosen for this case may be one where hemoglobin absorption is stronger, such as 532 nm or 405 nm, to further minimize the tissue reflectivity.
In either case, a good estimate for a signal reference level to determine contact or no contact can be obtained by prior testing of a particular type of probe on accessible mucosal tissue such as the hand or lip. A histogram of signal levels can also be readily maintained during a particular procedure by continuously updating how often a particular monitor signal level is obtained. Initially this level represents the background signal from the distal tip of the probe in air. The first contact with tissue increases the histogram count at higher signal levels. The transition between contact and no contact is fast for forward-looking probes so that relatively few histogram counts are obtained in the transition zone. An adaptive algorithm can determine the optimal reference signal level by determining a value which is roughly equidistant between the first peak on the low side of the histogram and the first peak on the high side of the histogram, regardless of whether the algorithm is looking for a signal increase or decrease.
For a standard ball lens probe, the background signal due to the Fresnel reflection component can either be measured and subsequently subtracted during signal processing or it can be reduced by proper spatial filtering of the light exiting the collection fibers at the proximal end of the fiberoptic probe. In the ball lens Raman probe, for example, the Fresnel reflection comes from the concave final surface of the lens which strongly focuses the reflected light so that it enters the collection fibers at a very steep angle. This steep angle is essentially maintained through many internal reflections within the collection fibers so that this component exits the collection fibers at a similar steep angle. When the light exiting the collection fibers is collimated by the first lens in the spectrometer this first component of monitor light shows up as a bright ring at the outer edge of the collimated beam (the Fourier transform of high angle light). A spatial filter in this nominally collimated beam that passes low angle light nearer the center of the collimated beam can reject most of this first component. This is desirable since it reduces the sensitivity of the later monitor signal processing software processing to variations in the intensity of the monitor light source and to variations in the collection efficiency between different probes of the same configuration.
Besides the monitor light returning to the proximal photodetector there may also be additional light due to endoscope illumination and/or tissue fluorescence which happens to be at the wavelength of the monitor light source. Two methods are used to reduce this external background signal. When the monitor light source is a narrowband laser a narrow bandpass optical filter can be placed in front of the monitor signal photodetector to reject out-of-band light. Generally this first method eliminates most, but not all, of the background light. The second method is to pulse the monitor light source and record the photodetector signal both with the monitor source on and with the monitor source off. If these two measurements are taken close together in time compared to typical intensity fluctuations in the background light (say at measurement rates of 10 Hz or above) the difference between these two measurements is the desired monitor light signal due to diffuse tissue reflection (or Fresnel reflection).
Pulsing the monitor light signal has an additional advantage when the contact monitor system is used with video endoscopes. Some of the monitor light will generally be visible in the video image of the endoscope system either because the probe is not in contact or because the monitor light is transmitted through the tissue when the probe is in contact. The integrated intensity of the monitor light source during a single video frame must not be too great or the camera pixels will be saturated. By pulsing the light for a brief period of time the monitor source can be used at its maximum value for the optimum measurement signal to noise ratio while avoiding camera saturation. The differential contact measurement can be made within a few milliseconds compared to the 33 millisecond video frame period so significant attenuation of the monitor light in the video image is possible.
The video image of the monitor light is also useful for correct positioning of the fiberoptic probe during spectroscopic measurement. Given that Raman signals are very weak the endoscope lighting may be significantly attenuated or even absent during Raman measurements. In the case of the fiberoptic Raman probe the monitor light exits the distal tip of the probe in a relatively narrow beam. The spot of monitor light projected onto the tissue surface can be used to guide the probe to the proper spot on the tissue surface. Since the monitor light is preferentially from a narrowband source such as a laser it is easily filtered out of the light collected for spectroscopic purposes. Typical endoscope illumination, on the other hand, is derived from an arc lamp source with a large amount of NIR power relative to the Raman signals even after extensive optical filtering.
The choice of wavelength for the monitor light must take into consideration the detailed design of the probe it is used with. In the case of the fiberoptic Raman probe the monitor light passes through two sets of filters located at the distal tip of the Raman probe. The first filter is located at the distal end of the Raman excitation delivery fiber. This filter blocks the (longer wavelength) Raman-scattered light generated within the excitation fiber but passes the power from the Raman excitation laser source. The second filter is placed before the collection fibers to block Raman excitation light reflected from the tissue and probe tip while passing the longer wavelength Raman-scattered light from the tissue being probed. This Raman excitation light can create an additional background signal in the long glass path leading back to the spectrometer. The specific wavelength that works for a particular Raman filter set may change from one type of probe to another but long-wavelength monitor laser sources are inexpensive and several different monitor lasers can be coupled separately into the delivery fiber and switched in appropriately. Diode lasers can also be temperature tuned over several nm by varying the power to the thermoelectric coolers typically used to stabilize them.
Preferred embodiments of the present invention are described with reference to the following drawings:
A preferred embodiment of the invention is illustrated in
The relatively weak probe contact monitor system beam 110 from its laser source 112 is angularly multiplexed into the delivery fiber 106 by directing it into the coupling lens 108 at a shallow angle with scraper mirror 114. The monitor system laser is pulsed electronically with circuit 116 at the appropriate time as determined from a video synchronization pulse 118 which can be derived from the endoscope video monitor signal.
Both the Raman excitation light 102 and the probe contact monitor light 110 are carried to the distal tip of the Raman probe through the delivery fiber 106. They both pass through the Raman rod filter 120 which rejects long-wavelength Raman shifted light generated in the delivery fiber 106. Both beams then enter the ball lens (or drum lens) 122. A small quantity of each beam is reflected where they intersect the ball lens exit surface 124 and a small portion of this reflected light passes through the Raman filter 126 before entering the probe collection fibers 128. This filter 126 can be a ring filter that has characteristics to block the Raman excitation source wavelength to prevent background Raman signals from being generated in the long collection fibers 128. The contact monitor system wavelength, however, is preferably chosen so that much of it passes through this second Raman filter. Further details regarding a Raman probe system can be found in U.S. application Ser. No. 10/407,923, filed on Apr. 4, 2003, the entire contents of which is incorporated herein by reference.
The ball lens 122 focuses most of the Raman excitation light and contact monitor light onto the tissue surface 130. Some of the resulting Raman-scattered light from the tissue and some of the diffusely scattered contact monitor light 132 is refocused by the ball lens 122 and coupled back into the collection fibers 128 after passing through the Raman ring filter 126. Most of the Raman excitation light is only diffusely scattered by the tissue (and thus not wavelength-shifted) and is blocked by the Raman ring filter 126.
The Raman scattering process immediately randomizes the direction of the Raman-scattered photons with the unscattered excitation photons generally continuing deeper into the tissue. The monitor light photons, however, are redirected by diffuse scattering to exit the tissue and be collected by a light collection system. The monitor light photons are typically at shorter wavelengths and will thus scatter faster, essentially simulating the Raman-scattered photons in terms of their collection versus probe-to-tissue distance. Most of the use of the contact monitor probe is in terms of on/off collection during the data acquisition period since the transition is very fast. The intermediate stage can be measured on representative mucosal tissue for a more precise correlation of their relative signals as a function of probe-to-tissue distance.
The collection fibers at the proximal end of the fiberoptic Raman probe 134 are aligned, bonded and polished. The polished ends are imaged with lenses 136 and 138 onto the entrance slit 140 of the Raman spectrometer 142. The first lens 136 collimates the beams exiting the collection fibers and a dichroic beamsplitter 144 is used as an optical separator which reflects the visible portion of the collected light and passes the NIR portion to the spectrometer to separate the monitor and diagnostic signals. Before entering the spectrometer a high quality, narrowband rejection filter 146 reduces the intensity of the remaining Raman excitation light by five to six orders of magnitude. A red glass absorbing filter 148 rejects the remaining broadband visible light and passes the red-shifted Raman scattered light and tissue fluorescence to the spectrometer. A CCD camera 150 records the spectra of this light for later analysis and tissue diagnosis.
The monitor light reflected off of dichroic filter 144 is passed through an aperture 152 which rejects most of the angle light reflected from the ball lens at the distal tip of the Raman probe. A laser line filter 156 passes the monitor light but blocks most of the broadband light from the endoscope white light illumination or tissue fluorescence induced by the autofluorescence endoscope. The remaining monitor light and background light at the same wavelength is passed on to photodiode 158 to be measured.
The monitor light signals are pulsed but do not need to be measured at very high frequencies so the photodiode 158 can be used in the photovoltaic or zero-biased mode for the lowest noise. A buffer circuit 160 utilizes a very large feedback resistor and a low bias current operational amplifier to convert the photodiode current to a voltage followed by low-pass filtering stages before the signal is finally measured. The signal is measured before the monitor laser source 112 is turned on by a sample-and-hold circuit and analog-to-digital converter 162 and after the monitor light source has stabilized by an equivalent circuit 164. The difference between these two measurements is taken by differencing circuit 166 to eliminate the effect of more slowly-varying background light. These measurement and timing circuits may be analog and discrete or their functions may be conveniently performed within a single programmable microcontroller 168. This microcontroller can also provide the discrimination of the resulting monitor signal with the reference threshold to determine a binary contact/no contact signal or as well as implement the adaptive histogram method of determining the optimal reference threshold for a given patient.
The microcontroller can also provide the timing pulses required by the contact monitor system which are all referenced to the video synchronization square wave 170 determined externally from the video signal of the autofluorescence endoscope. This synchronization is identical to the sync pulse 118 called out elsewhere in the figure. The monitor laser pulse can be triggered in either the odd or even video field for a 29.97 Hz update rate. The trigger to perform the background measurement 172 is followed by the signal to turn on the monitor light source 174 and the signal to perform the monitor+background measurement 176.
The final result of the contact monitor is presented to the clinician with visual display 178 which may be either a visible light or a visible mark on the autofluorescence video monitor. The result is also recorded so that it can be included in the processing of the measured Raman/fluorescence signal. A Raman signal in which the probe maintained contact for 90% of the integration time can be successfully renormalized by processing with the monitor signal to what it would have been with 100% contact during the integration time.
The detailed optical model of
Hemoglobin is the primary absorbing species in tissue so the preferred choice for a monitor light source is a diode laser with a wavelength greater than 600 nm and preferably greater than 630 nm where hemoglobin absorption is low. Since tissue reflectivity will vary with the patient, with the type of tissue being probed and with the presence or absence of blood on the tissue surface an adaptive algorithm is desirable. Minimizing the background signal of the probe in air will also increase the reliability of the contact measurement. Histogram analysis can be performed on a long rolling list of the most recent contact measurements to adapt to changing tissue types or tissue states during the procedure.
Even though the contact monitor is particularly useful for Raman spectroscopic probe the system can be used for visible fluorescence probes and visible diffuse reflectance probes as well. In this case the monitor laser can be chosen from diode lasers with wavelengths between 670 and 780 nm which can be seen visually or by video endoscopes but still be outside the range of most fluorescence and diffuse fluorescence diagnostics.
The contact decision algorithm 1000 embedded in the flow chart determines whether or not the good contact counter M is incremented following any single measurement. This algorithm may look for either increased or decreased contact monitor signal depending on the design of the Raman probe in use and can use adaptive modifications to the contact/no contact threshold for the contact signal determined by the most recent histogram of contact measurements. The contact decision algorithm may also consider the stability of the contact signal over a number of recent contact measurements <=N by weighting past measurements before changing the state of the contact/no contact decision. This is effectively equivalent to limiting the frequency bandwidth of the contact measurement.
While the present invention has been described herein in conjunction with a preferred embodiment, a person with ordinary skill in the art, after reading the foregoing specification, can effect changes, substitutions of equivalents and other types of alterations to the system or method as set forth herein. Each embodiment described above can also have included or incorporated therewith such variations as disclosed in regard to any or all of the other embodiments. Thus, it is intended that protection granted by Letters Patent hereon be limited in breadth and scope only by definitions contained in the appended claims and any equivalents thereof.
The present application claims priority to U.S. Provisional Patent Application No. 61/002,723 filed Nov. 9, 2007. The entire contents of the above application is incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
61002723 | Nov 2007 | US |