In the following, the invention is described by using a specific embodiment in the light of the enclosed figures.
The figures show:
Within the tank 2 the coolant 3 is provided. The substances, coming into question as a coolant for the cryo temperature cooling device 1, stand out by a low phase transition temperature, for example a low melting temperature, being roughly the temperature to which an object is to be cooled, a preferably high heat of fusion, preferably high boiling temperature and a preferably low vapor pressure at the maximum operation/storage temperature. Some of the substances to be considered and their relevant properties are listed in the following, not completed Table 1:
It is to be understood that Table 1 does not list all coolants to be used according to the invention, but only a selection. Especially, this selection is mainly limited to substances with a boiling temperature TSI>20° C. and a melting temperature TS<−120° C. Only butane, butene-2 cis, ethyl acetylene, butene-1 and isobutene have been mentioned in addition, even if their boiling temperature TSI is lower than 20° C. Suitable as coolants are propane and propene also, but they do have a vapor pressure of 8.7 and 10.3 bar at 21° C., therefore requiring higher standards with regard to the pressure resistance of the tank 2.
As a comparison, liquid nitrogen has a boiling temperature of −196° C. and a heat of fusion of about 161 kJ/ltr. Suitable as coolants are also substances like propane and propene, which nevertheless do have a vapor pressure of already 8.7 respectively 10.3 bar at 21° C. and therefore having to meet higher demands when it comes to the pressure stability of the tank.
Furthermore, a heat conducting element 4 in the form of a cooling finger, protruding from tank 2 through an opening 14 through the inner wall 5 and the outer wall 6, is seen in the cryo temperature cooling device 1. The cooling finger 4 is surrounded nearly completely from the coolant 3 within the tank 5. Mounted to the part of the cooling finger 4, protruding from the cryo temperature cooling device 1, is heat insulating protective cap 8. So, the complete cryo temperature cooling device 1 is isolated from the surrounding temperature as long as it is not in operation respectively not used for the cooling of an object, therefore, being able to be stored. The cooling finger 4 preferably also comprises material, providing for a good heat transfer and at the same time for the thermal connection between coolant and the object to be cooled. It may for example be made from copper. In order to further improve the heat transfer of the coolant 3 to the cooling finger, further substances could be added to the coolant 3 in order to promote this property.
The cryo temperature cooling device is used as follows: the cryo temperature cooling device 1 is charged by bringing the cooling finger 4 to a temperature below the melting point of the coolant. When using vinyl-bromide, this would, for example, be a temperature below −138° C. (135.2 K). This cooling process continues until the coolant is mainly frozen. The cooling, respectively the charging of the cryo temperature cooling device 1, is conducted with a charging device, for example a compression refrigerating machine or with liquid nitrogen.
Afterwards, in order to store the cryo temperature cooling device 1 until it is used for operational purposes, the protecting cap 8 is mounted to the cooling finger 4. Therefore, a good heat insulation of the cryo temperature cooling device 1 with regard to the outside is achieved and the frozen state of the coolant 3 can be kept fairly long.
When the cryo temperature cooling device 1 should be used for cooling purposes, for example of a detector, the protective cap 8 is removed from the cooling finger 4 and the cooling finger 4 is connected to the object to be cooled or included therein.
The cooling finger 4 of the detector 10 is thermally mounted to the detector support 15, supporting also the detector crystal 16, for example a germanium crystal. There are, nevertheless, other detector crystals suitable in order to be cooled by the inventive cryo temperature cooling device 1, for example silicon crystal detectors, CdTe-detectors and others. Preferably, the preamplifier is thermally connected to the detector support 15 also so that this is cooled, too. When the measurement respectively the operation time of the detector 10 has ended, the cryo temperature cooling device 1 can be removed from the detector 10 again in order to be recharged for a further measurement.
As the size of the cryo temperature cooling device 1 is small, it is especially suitable for the operation in combination with a measurement device in the form of a handheld device 11, which is shown schematically in a top view in
Because of its low weight and volume this handheld device 11 is especially suitable for mobile missions, for example as gamma ray detector for luggage control at airports, borders or at big events. The handheld device 11 is touched at its handle 17 and hold thereon during the measurement. Input keys 12 allow the selection for example of various functions or are designed to start and end the measurement operation. Measurement results may be metered either acoustically, for example when excessing the detected limits, or via reading the display 13.
Via a connection 18, which may be protected and isolated by a removable heat insulating cap 8, it is possible—after removal of cap 8—to “load” the cryo temperature cooling device as described above, that is it could be cooled, by keeping the heat conducting contact of the connection 18 at a temperature T until the detector is cooled and the coolant is mainly frozen. For the temperature T it is true that T<TP, whereby TP is the phase transition temperature of the coolant used according to the invention. After removing the charging device from the connection 18 and attaching the protection cap 8, the detector is kept cooled for several hours ready for operation.
The use of cooling coil 19 for the charging of the cryo temperature cooling device 1 is especially an advantage when cooled (liquefied) gases, for example nitrogen or helium or others, liquid coolants are available for the cooling process. Those cooled (liquefied) gases or coolants could then be piped directly through the inner part of the cooling coil, which allows for a higher heat transport so that the cooling can take place faster. Furthermore, the connections 18 of the cooling coil, being directed to the outside, may consist of materials, being bad heat conductors, so that the heat insulation of the cryo temperature cooling device with regard to the surrounding is improved, so that the operation time can even be improved with regard to the device shown in
In addition to the use for the cooling of detectors, the cryo temperature cooling device according to the invention can be used in any devices where cryo temperatures are necessary or wanted. This is especially true for, but not limited to further measurement applications, not the least for such applications, using superconductors, especially high Tc superconductors. One has to think firsthand to electronics, here especially to noise sensitive preamplifiers, to IR sensors, night vision devices, to SQUIDs or to applications for radar.
1 cryo temperature cooling device
2 tank
3 coolant
4 heat conducting element
5 inner wall (inner tank)
6 outer wall (outer tank)
7 intermediate space
8 protecting cap
9 measurement device
10 detector
11 handheld device
12 input key
13 display
14 opening
15 detector support
16 detector crystal, eventually with preamplifier
17 handle
18 connection
19 cooling coil
20 outer wall of the detector
21 radiation entry window
| Number | Date | Country | Kind |
|---|---|---|---|
| 10 2004 036 483.4 | Jul 2004 | DE | national |
| 10 2004 043 900.1 | Sep 2004 | DE | national |
| Filing Document | Filing Date | Country | Kind | 371c Date |
|---|---|---|---|---|
| PCT/EP05/53707 | 7/28/2005 | WO | 00 | 11/5/2007 |