1. Field
This application relates generally to sterilization systems and more particularly to sterilization systems for use in decontamination of isolators.
2. Description of Related Art
Isolators are structures designed to maintain a sterile environment for manufacturing or laboratory activities where contamination risk must be mitigated. As an example, isolators are used in the pharmaceutical industry to provide sterile environments for drug processing and/or sterility assurance testing with minimal risk of contamination by viable microorganisms. They are typically operated at a slight positive pressure to prevent introduction of outside contaminants via leakage pathways into the enclosure. As a result, isolators are not amenable to use of vacuum cycles during decontamination operations.
In an open loop sterilization or decontamination system, sterilant is added to a chamber and then withdrawn from the chamber after a dwell period. A sterilizer unit that employs a vacuum phase, as is used for medical equipment, is an example of an open loop system. A closed loop system is one in which gas from the enclosure is recirculated for the purpose of adding or removing sterilant or humidity. Typically, a closed loop system is used when the enclosure cannot support the forces associated with creating a vacuum within the enclosure. Certain gas delivery systems, as would be used with an isolator, are an example of a closed loop system. In U.S. Pat. No. 4,909,999 Cummings, et al., describe a method of introducing a sterilant vapor to a room or chamber in which the sterilant vapor is formed from hydrogen peroxide and water, and using a recirculating gas circuit. The removal of the sterilant involves the use of heat to rapidly decompose the hydrogen peroxide.
For sterilization and decontamination of isolator enclosures, vapor hydrogen peroxide (VHP) is most widely used as the sterilant. Generally, there are two types of hydrogen peroxide-based systems described: systems that dehumidify the enclosure gas with dryers, and systems that humidify the enclosure gases in order to controllably form some water and sterilant vapor condensation. For an example of using a dehumidification phase, U.S. patent application Ser. No. 11/421,265 teaches use of a dryer in a dehumidification phase. After dehumidification, conditioning is performed and VHP is injected at a high flow rate. Systems that use a dryer for dehumidification are also described in U.S. Pat. No. 5,173,258 and U.S. Pat. No. 5,906,794.
It has been shown that excess moisture buildup in an enclosure can hinder the sterilization process and rapid removal of the sterilant. As hydrogen peroxide degrades into oxygen and water, water content in the enclosure tends to increase. To avoid the problem of excess water caused by using hydrogen peroxide as a sterilant, Childers et al., describe a method of drying the gas circulating in the chamber in U.S. Pat. No. 5,173,258.
Using chlorine dioxide as the sterilant generally requires high humidity, resulting in the presence of excess water. For example, chlorine dioxide decontamination and sterilization is described US Patent Application No. 2009/0246074 A1, by Nelson, et al., wherein high levels of humidity are required. Such high levels of humidity tend to require extended aeration periods.
A system and method for decontamination of isolation enclosures includes a recirculating isolator configured to allow injection of a sterilant gas into the isolator. Levels of humidity and sterilant gas are selected to avoid condensation of either within the isolator. In an embodiment, a positive pressure is maintained throughout the sterilization process.
In an embodiment in accordance with an aspect of the present invention, nitrogen dioxide (NO2) is used as the sterilant gas. Generally, NO2 has a low boiling point and high vapor pressure at room temperature, which the inventors have found makes it particularly well suited to sterilization or decontamination of enclosures. Use of a low boiling point sterilant may allow handling in either liquid or gaseous form, as well as avoiding a need to generate extreme temperatures or requiring the isolator to be made using highly heat or cold resistant materials. Furthermore, low boiling point sterilants will not tend to condense on surfaces of the enclosure, reducing the potentially dangerous deposition of residual sterilant.
In embodiments, sterilant may be introduced to the enclosure directly, by way of a gas injection system. Alternately, sterilant may be introduced into a recirculating gas stream.
In an embodiment, illustrated in
A recirculating gas flow circuit 14 may be used to flush the contents of the pre-chamber (or, gas generating chamber) into the enclosure. This approach does not require the addition of heat to generate the NO2 gas, it can be generated at room temperature.
An optional humidifier 16 may be included within the recirculating gas flow circuit 14. A sterilant gas source 18 is in communication with the pre-chamber 12.
An alternate approach to introducing the sterilant gas to the chamber or enclosure is the use of one or more injection nozzles that directly introduce the sterilant into the enclosure volume or recirculating gas stream. With a low temperature boiling point sterilant gas, like nitrogen dioxide, nozzles at room temperature, or slightly elevated temperature, may be used to dose the liquid sterilant directly into the chamber. Where a temperature of the sterilant is close to or above the boiling point, sterilant would vaporize as it exits the nozzles.
In an embodiment, liquid nitrogen dioxide may be metered by weight or volume prior to introduction into the enclosure, recirculating gas stream, or gas generating pre-chamber. In another embodiment, a chemical composition that generates NO2 may be positionable within the pre-chamber where it may be activated to generate the NO2 for sterilization. The gas delivery may be accomplished by using a DOT approved cylinder holding a quantity of liquid NO2 (which is actually the dimer N2O4).
In an embodiment, nitric oxide (NO) can be added to the recirculating gas stream or gas generating prechamber. NO can be stored as a compressed gas in gas cylinders. The gas will mix with air in the prechamber, in the reciculating gas stream, and/or in the enclosure. Upon mixing with air, the NO will react with oxygen to form NO2.
In an embodiment, concentrations of sterilant and temperatures are selected such that the sterilant does not condense. Sterilant condensation can tend to increase the time needed to aerate the chamber of residual sterilant gas, as the condensed sterilant does not rapidly evaporate. Certain corrosive sterilants (such as hydrogen peroxide) may be damaging to materials within the isolator, or can cause injury to personnel who come into contact with condensed sterilant.
Likewise, condensing levels of humidity tend to lead to sterilant condensation. If liquid water forms on surfaces, the sterilant will tend to form a mixture (solubilize) with the water, increasing the amount of condensed sterilant. This will tend to further increase the time needed to aerate the enclosure. Therefore, embodiments employ humidity levels less than a condensing level. In an embodiment, humidity within the isolator is controlled to between 30 and 90% relative humidity, and particularly, between 70 and 85% relative humidity. In a particular embodiment, the isolator is controlled to between 55 and 70% relative humidity.
Experiments were performed to simulate effectiveness of methods as described herein. A test chamber was operated in a manner that simulated an industrial isolator system, by employing cycles with minimal changes in pressure during gas introductions. In one testing protocol, sterilant concentrations necessary to achieve a six-log reduction in spore population on commercial biological indicators (BIs) at exposure times of 5 and 10 minutes were determined.
In another testing protocol, the ability of a dry air purge to clear sterilant from the chamber in a timely manner was demonstrated. A purge of the chamber (enclosure) introduces air that does not contain sterilant into the chamber as gas containing sterilant is removed from the chamber. No vacuum was applied either prior to sterilant gas introduction or for sterilant gas removal. Instead, the cycles described herein rely on an exhaust port (or vent valve), which was left open, permitting gas to escape from the chamber, thereby maintaining a constant chamber pressure. In this manner, as gas is added to the chamber, gas displaced by the added is exhausted from the chamber. This approach simulated the gas addition and removal as would be observed in the case where a recirculating gas circuit would be used to add or remove sterilant and humidity from the enclosure.
The results of these tests are discussed below and demonstrated the ability to humidify the chamber and add lethal amounts of NO2 gas with minimal pressure increases. Cycle conditions that sterilized commercial biological indicators (BIs) with 5×106 CFUs were selected, using exposure times of 5 and 10 minutes. The ability to purge the chamber to less than 1 ppm NO2 utilizing a dry air flush of approximately 30 minutes was also demonstrated.
The specific exposure cycles performed during these tests are shown in Table 1. The duration during which NO2 was added to the chamber was varied as a means of varying the resulting concentration of NO2 in the chamber during the exposure dwell phase of the cycle. Biological indicators were placed in the chamber during each cycle in order to determine the exposure conditions that yielded a six-log spore population reduction on commercial biological indicators (BIs) exposed.
Prior to starting each cycle, 13 BIs were placed in the chamber. The BIs were widely distributed on the chamber shelf. Nine of these 13 BIs were used for fraction negative tests, where, after exposure, each BI was placed in test tubes containing tryptic soy broth and incubated. The incubated test tubes that exhibited turbidity after an appropriate incubation time were judged to be ‘positive’, and to have had viable spores on the BI placed in that test tube. Test tubes that did not exhibit growth were deemed to be ‘negative’ for surviving (viable) spores on the BIs in that test tube. The number of negative and positive BIs for each cycle were recorded.
The results of the fraction negative testing are shown by the number of negative BIs in Table 2. With the 5-min exposures, one cycle (Cycle No. 1) had one positive BI and all other 5-min cycles were negative. For the 10-min exposures, Cycles 6 and 7 resulted in nine and five positive BIs, respectively. The other three cycles yielded complete sterilization of the nine BIs. In addition to the nine BIs used for fraction negative testing, four BIs were included in each cycle for direct enumeration of surviving CFUs. The results of the plate counts are shown as the average log of recovered CFUs per BI in Table 2.
The results of the fraction negative BI testing are plotted in
If one treats the overall NO2 dose of a given cycle as the product of NO2 injection time multiplied by exposure time, then the fraction negative data for all cycles (both 5 minute and 10 minute exposure times) can be plotted on one curve as the number of negative BI's versus dose, as is shown in
Four of the BIs from each cycle were used for direct enumeration of the surviving spores. These BIs were processed with a spore recovery procedure that is known to collect a large percentage of the spores from the BI carrier. The collected spores were grown on agar plates in a manner that permitted counting of the spores collected by counting the colonies that grow on the agar plates. The resulting colony forming units (CFUs) on each agar plate were counted and the average number of CFUs per BI per cycle were recorded, and plotted against the NO2 injection time.
A Fourier Transform Infrared (FTIR) spectroscopy system was used to monitor both the NO2 and H2O gas concentrations in the chamber during each cycle. A typical concentration profile for H2O and NO2 during one of the cycles is shown in
The maximum NO2 concentration for Cycle 2 was 6.6 mg/L, which was lower than the theoretical maximum of 8.2 mg/L. This apparent reduction in sterilant concentration was attributed to two factors. The first factor was the open vent valve, intended to simulate a recirculating isolator system. This would have allowed some percentage of the sterilant to be vented out the chamber during filling, as this part of the cycle was done under a slight positive pressure, as is common with industrial enclosures. The second factor that contributed to the apparent reduction in sterilant concentration was the interaction of NO2 gas with H2O. In
A combination of FTIR spectroscopy and electrochemical sensors (EC cells) was used to measure the NO2 levels in the exhaust gas from the test unit chamber on a cycle that employed the exposure condition described by Cycle 4 in Table 2. At the end of the exposure time, a 60 minute purge of dry air at a rate of 40 LPM was used to clear the test unit chamber of sterilant. This purge rate was equal to approximately one chamber volume exchange per minute. The test chamber was 44 L in volume.
The FTIR was used to measure the exhaust gas from the test unit until the concentration of NO2 in the gas fell below 100 ppm. At that point, the exhaust gas was directed to EC Cell 1, which had been calibrated for concentrations from 0 ppm to 100 ppm. When the NO2 concentration of the exhaust gas dropped below 10 ppm, the gas was shifted towards EC Cell 2, calibrated for 0 ppm to 10 ppm NO2 measurements, for the duration of the purging process.
An exponential fit of the FTIR measurements yields an NO2 removal rate of:
y=2112e−0.013x
Upon switching to EC Cell 1, the exponential fit of the NO2 removal rate fit the following equation:
y=33.67e−0.0097x
These two measurements were similar in the rate of reduction indicating that the dry air purge of gas from the test unit chamber was the primary dynamic of NO2 removal.
If one looks at the measurement data from EC Cell 2, the first three minutes of data (minutes 8 through 11 of the purge) followed an exponential decay pattern that eventually changed slope to a point where the rate of NO2 removal was slowed significantly. The first three minutes of EC Cell 2 data was fit to:
y=4.18e−0.0056x
While the significantly slower rate of NO2 removal could be fit to the following equation:
y=0.50e−0.00013x
The change in slope of the curve may be explained by a transition from the primary NO2 removal dynamic to a secondary dynamic. The data from EC Cell 2 were used to model the transition from the primary NO2 removal dynamic to the secondary dynamic. A simple addition of the primary and secondary fits from EC Cell 2 was found to provide a good match to the actual EC Cell 2 data. This model is described by the following equation, which is the summation of the primary and secondary fits.
[NO2]=4.18e−0.0056t+0.50e−0.00013t
There was no obvious evidence for a tertiary dynamic or other unaccounted for mechanism in the NO2 removal process. The primary and secondary fits, the sum of the two fits, and the actual EC Cell 2 data are shown in
The inventors propose that the most likely source of the secondary NO2 removal dynamic is related to the structure of the chamber walls. Specifically, the Teflon coating of the test unit's chamber and the Teflon shelf within the chamber are at least partially permeable to NO2 and will tend to absorb a fraction of the NO2 gas introduced to the chamber. The chamber coating is approximately 3200 in2, while the shelf contributes roughly 600 in2. It is proposed that as the purge process progressed, the NO2 desorbed from the surface as it diffused out of the Teflon matrix. This secondary dynamic proved to be slower than the primary dynamic of NO2 displacement.
The final NO2 concentration reached after 60 min of purging was approximately 0.35 ppm.
In view of the secondary mechanism described above, it may be useful to construct an isolator in accordance with an embodiment using materials selected to have low permeability to NO2. Such low permeability materials include glass and stainless steel. Furthermore, smooth surfaces may be used to discourage adherence or embedding of contaminant, as well as reducing adsorption of NO2 or water. The relatively small surface area of more permeable polymers is not expected to influence this rapid aeration rate.
In embodiments as described above, gas ports are described for injection of sterilant gas, air, and/or humidity. In this regard there may be multiple gas ports or all gases may be introduced through a common port. Likewise the gases may pass through a manifold to improve distribution within the chamber. In this approach, it may be useful to include a valving system such that individual lines are separately controllable.
Embodiments may include temperature controls including, for example, temperature sensors, heaters and/or coolers. A humidity sensor may also be included to allow a feedback control of system humidity conditions. In an embodiment, the source of humidity is controlled to provide humidity in vapor form and to avoid delivery of water particles, which may tend to interfere with aspects of the sterilization process.
As will be appreciated, the system described may find application with a variety of gaseous sterilants, though the inventors have found particular advantage in use of nitrogen dioxide gas. In use, a sterilization cycle with NO2 employs between about 5 mg/L to 20 mg/L (roughly 0.25% to 1% at ambient pressure).
A scrubber system 20 may be located in the gas recirculation circuit, and used to capture the NO2. Alternately, it may be located in an exhaust pathway 22 used in the purge cycle as shown in
A user interface, not shown, may be incorporated allowing for programming of aspects of the system. This may include, for example, timing of stages (i.e., conveyor speed), dosage of sterilant, humidity and/or temperature, and others. The user interface may also include displays for providing a user with information regarding the defined parameters and/or indications of operating conditions of the system. Controllers can be based on computers, microprocessors, programmable logic controllers (PLC), or the like.
Although the invention has been described in detail for the purpose of illustration based on what are currently considered to be the most practical and preferred embodiments, it is to be understood that such detail is solely for that purpose and that the inventions are not limited to the disclosed embodiments, but, on the contrary, are intended to cover modifications and equivalent arrangements that are within the spirit and scope of the described embodiments. For example, it is to be understood that the present invention contemplates that, to the extent possible, one or more features of any embodiment can be combined with one or more features of any other embodiment. Likewise, embodiments may be incorporated into systems including glove boxes and clean rooms.
This application claims priority to U.S. Provisional Patent Application 61/525,424 filed Aug. 19, 2011, which is incorporated by reference in its entirety herein.
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/US2012/051425 | 8/17/2012 | WO | 00 | 1/5/2015 |
Number | Date | Country | |
---|---|---|---|
61525424 | Aug 2011 | US |