The present invention relates to a degrader, preferably used in particle radiation therapy facilities.
Slowing down (also known as degrading”) protons or ions (particles) from a particle accelerator, such as for example a cyclotron, to an energy compatible with the range of the particles in tissue, is done by sending the beam through a degrader. A degrader is an amount of material (a block or several layers) with a low atomic number (Z), such as graphite. The degrader, or a part of it, is inserted in the trajectory of the particle beam in order to reduce the energy of the particles. The total thickness of the amount of that material determines the energy of the out-going particles. However, the particles crossing such material are also scattered by the nuclei in the material. Due to that so-called multiple scattering process the out-going beam obtains an increase in diameter as well as an increase in angular spread (divergence). In order to prevent unwanted beam losses in the beam transport system following the degrader, usually a set of collimators is mounted immediately behind the degrader. When degrading to low energies the scattering increases. At those collimators the losses thus increase, which yields a decrease of the transmission, which is the fraction of the beam intensity from the accelerator, that reaches the patient. In proton therapy that causes an undesirable increase of treatment time.
At the Paul Scherrer Institute (PSI), located at CH-5232 Villigen PSI, Switzerland, the proton therapy facility PROSCAN is formed of a 250 MeV superconducting cyclotron COMET and a beam transport system that guide the beam to four treatment areas, Gantry 1, Gantry 2, Gantry 3 and OPTIS. The energy required for the patient treatment is in the range between 70 MeV and 230 MeV. Energy modulation is performed via a graphite degrader (see
Low energies are used in the treatments in the gantries for dose deposition in the (parts of the) tumors that are located at a shallow position below the skin. The lowest beam energy of 70 MeV requires the maximum energy degradation in PROSCAN and beam transmission is most challenging (see the transmission curve in
Unfortunately, the multiple scattering process in an energy degrader increases the beam size, divergence and energy spread beyond the beamline and gantry acceptance. It is unavoidable to use one or more collimator systems and an energy selection system to cut those quantities to those that fit in the acceptance of the following beam transport system, to prevent unwanted beam losses in the beam transport system. That limited acceptance depends on the energy, the geometrical layout of the beam transport system and on the setting of the magnets in the system. At PSI a typical beam transmission is on the order of 0.2% for beams with an energy of 70 MeV.
Typically, for lower energies the intensity loss can be compensated by an increase of beam current at the cyclotron exit. The maximum beam current provided by the cyclotron is, however, limited. Hence, it is useful to examine the possibilities of reducing the emittance increase in the degrader that leads to the beam loss in the subsequent beam transport system.
The usual choice for degrader material is graphite, which is a material with a low atomic number Z, in order to limit the multiple scattering amplitude. Beryllium is also a degrader material with a low Z, thus also causing a low divergence increase and it has been used as degrader material at some places. However, it has a disadvantage of creating a strong neutron flux during the degrading of high energy protons. Furthermore, Beryllium is known to have some toxic characteristics. Plastics are also used, but they suffer from geometry changes due to the radiation damage and heat deposition in the material.
It is accordingly an object of the invention to provide a degrader including boron carbide, which overcomes the hereinafore-mentioned disadvantages of the heretofore-known degraders of this general type and which has excellent degrading capabilities without generating neutron flux and not having toxic properties.
With the foregoing and other objects in view there is provided, in accordance with the invention, a degrader comprising degrading active material, wherein the degrading active material includes Boron Carbide B4C.
This degrader evokes an amount of multiple scattering that is lower than in graphite for the same energy loss. At PSI it has been found that the use of B4C increases the transmission by at least 35% for the beam degradation to low energies, which is a significant and useful amount of beam intensity increase in particle therapy. The B4C-material does not become more radioactive than graphite, so that there will be no additional problems at service activities. Further, B4C when used as a degrading active material does not have any toxic properties.
Suitably, the degrading active material can be assembled as plates or as wedges.
Also suitably, the degrading active material can be mounted on one or more actuators that can bring the degrader or a part of the degrader in a position where it is crossed by the particle beam.
Other features which are considered as characteristic for the invention are set forth in the appended claims.
Although the invention is illustrated and described herein as embodied in a degrader including boron carbide, it is nevertheless not intended to be limited to the details shown, since various modifications and structural changes may be made therein without departing from the spirit of the invention and within the scope and range of equivalents of the claims.
The construction and method of operation of the invention, however, together with additional objects and advantages thereof will be best understood from the following description of specific embodiments when read in connection with the accompanying drawings.
Test measurements using boron carbide (B4C) as degrader material are discussed in the following in comparison with the conventional graphite, which is currently used in many proton therapy degraders. B4C is a material of lower average atomic weight and higher density than graphite. Calculations predict that, compared with graphite, the use of B4C results in a lower emittance behind the degrader. Downstream of the acceptance defining collimation system at the entrance of the following beam lines, a higher beam transmission occurs, especially at low beam energies. This is of great interest in particle therapy applications as it allows either a reduction of the beam intensity extracted from the cyclotron or a reduction of the treatment time.
The results of experiments carried out at the PROSCAN facility at the Paul Scherrer Institute are discussed hereinafter. The simulations of a B4C-degrader have predicted an increase in the beam transmission of approximately 31% compared to graphite, for beam degradation from 250 to 84 MeV. The experiment carried out with a B4C block reducing the energy to 84 MeV yielded a transmission improvement of 37% compared with the carbon degrader set to that energy.
In the experiment, a B4C block 1 shown in
The beam current has been measured after the cyclotron and after the energy selection system for both the B4C block and the graphite wedge degrader for a beam energy of 84 MeV. The transmission, defined as the ratio of these beam currents, was 0.59% for the B4C block and 0.43% for the graphite degrader. Hence, it was found that the measured transmission is 37.2% higher for the B4C block compared to the graphite degrader for the same energy.
Measurements have successfully been performed to compare the experimental behavior with the calculated performance of B4C as potential degrader material. Both the simulation and the measurement yield an improvement of the transmission at 84 MeV in the order of 35% using a B4C degrader compared to the currently used graphite degrader. Further a 30% lower dose rate at the B4C degrader is expected compared with the graphite degrader. Thus, the level of B4C activity after the irradiation is considered as acceptable and uncritical with respect to machine maintenance.
Number | Date | Country | Kind |
---|---|---|---|
15196840.1 | Nov 2015 | EP | regional |
This application is a continuation, under 35 U.S.C. § 120, of copending International Application PCT/EP2016/077563, filed Nov. 14, 2016, which designated the United States; this application also claims the priority, under 35 U.S.C. § 119, of European Patent Application EP 1 519 6840.1, filed Nov. 27, 2015; the prior applications are herewith incorporated by reference in their entirety.
Number | Date | Country | |
---|---|---|---|
Parent | PCT/EP2016/077563 | Nov 2016 | US |
Child | 15991161 | US |