The present disclosure relates to indirect determination of parameters used by water treatment facilities such as biological oxygen demand (BOD), chemical oxygen demand (COD), total organic carbon (TOC), trihalomethane formation potential (THMFP), etc. based on components identified using absorbance and fluorescence spectral analysis.
Water treatment plants, including those that treat surface water sources must generally meet various government requirements with respect to both the effluent distributed from the plant as well as intermediate by-products that may be created during the treatment process. In the United States, water treatment plants are required by the Environmental Protection Agency (EPA) to reduce the total organic carbon (TOC) concentration using a process of coagulation prior to disinfection of the finished water with halogenated/chlorinated disinfectants.
While various tools or instruments for monitoring TOC have been developed, online monitoring of TOC alone does not provide necessary information on the aromatic properties of the sample that are needed to determine the effective coagulation and disinfection dosages needed to prevent formation of toxic disinfection by products (DBPs). The aromaticity is the primary characteristic of the TOC that determines the chemical reactivity with halogenated disinfectants that results in toxic, carcinogenic DBPs.
Current convention uses separate measurements of the absorbance at 254 nm (A254 nm) and TOC concentration with separate instruments/detectors for the purposes of evaluating the effectiveness of the coagulation using what is known as the specific UV absorbance calculation, SUVA=A254 (m−1)/TOC (mg/l). The TOC and SUVA techniques do not provide a reproducible evaluation for different water sources because the aromatic properties of the organic composition of the source water often varies for a particular source over time, as well as among different sources. Further, the SUVA parameter is often imprecise due to the lack of kinetic simultaneity of the parameters of the TOC meter and absorbance detectors, as well as inherent propagated noise/interferences of the separate detection methods as conventionally implemented. Both TOC and A254 are prone to interferences of several types. Use of independent fluorescence data provides a means of ameliorating the influence of primary interferences.
While previous regulations for water treatment plants in the United States required averaged readings of disinfection byproduct formation levels of an entire distribution system, more recent EPA regulations require monitoring of local averages of disinfection byproduct formation levels in different regions of the distribution system. Monitoring only system wide averages may not be sufficient to detect local regions with higher propensity to form DBPs, which may be in violation of the more recent regulations outlined in the EPA Disinfection Byproduct Rule 2 (DBPR2). This of course heightens the need for more precise and accurate TOC and aromaticity evaluations for local regions of the treatment processes.
The imprecision of A254, TOC and SUVA as calculated using EPA specified methods (such as EPA Method 415.3) may also be attributed to the fact that both detector readings are aggregate, single point readings and therefore lack qualitative information on the effects of the coagulation treatment with respect to reactive organic species. As previously described, a number of interferences or confounding factors must be considered with respect to their effect on the readings, including inorganic carbon, metals like iron, and unknown contaminants that may or may not fluoresce. In addition, online TOC meters are highly prone to falling out of calibration, as are online DBP meters (gas chromatographs). Most water treatment plants cannot afford to install and maintain these pieces of equipment. However, as a result of the recent regulatory requirements, many water treatment plants in the United States are considering major infrastructure changes (tens of millions of dollars) including addition of ozone-destruction and ion exchange processes, such as the MIEX resin treatment process, for example.
In one embodiment, a method implemented by a computer for determining a quality monitoring parameter of a liquid sample includes controlling an instrument to measure a first fluorescence emission spectrum and a second fluorescence emission spectrum of the sample over a plurality of wavelengths emitted in response to an excitation wavelength, controlling the instrument to measure an absorbance spectrum of the sample at the plurality of wavelengths, and correcting the first and second fluorescence emission spectra for inner filter effects based on the absorbance spectrum to generate first and second corrected fluorescence emission spectra, respectively. The method may also include determining a first peak emission wavelength of the first corrected fluorescence emission spectrum and at least a second peak emission wavelength of the second corrected fluorescence emission spectrum, the second peak emission wavelength different from the first peak emission wavelength, determining, using the computer, a value for a first component in the sample relating to the quality monitoring parameter and associated with the first peak emission wavelength and a value for a second component in the sample relating to the quality monitoring parameter and associated with the second peak emission wavelength, and calculating, using the computer, a value for the quality monitoring parameter based on at least one coefficient representing a relationship between fluorescence emission spectra of the first and second components and the quality monitoring parameter.
A computer-implemented method for determining a water treatment parameter includes receiving, by a computer, measurements of a fluorescence emission spectrum of a water sample including a first peak emission wavelength and at least a second peak emission wavelength, emitted in response to an excitation wavelength, receiving, by the computer, an absorbance measurement obtained at the excitation wavelength of the water sample, determining, using the computer, a ratio of the measurements at either the second peak emission wavelength, or a sum of measurements at a plurality of peak emission wavelengths including at least the first peak emission wavelength and the second peak emission wavelength, to the first peak emission wavelength, and calculating, using the computer, a value for the water treatment parameter based on a linear combination of at least the ratio and the absorbance measurement. The method may provide indirect determination of parameters used by water treatment facilities such as biochemical oxygen demand (BOD), chemical oxygen demand (COD), total organic carbon (TOC), Trihalomethane and Haloacetic Acid formation potentials (HAAFP and THMFP, respectively), etc. based on a ratio of components identified using absorbance and fluorescence spectral analysis rather than relying on instruments, such as a TOC meter, that are subject to more interferences and related variance in readings.
In one embodiment, a computer-implemented method for controlling a water treatment process includes measuring a first fluorescence emission spectrum of a pre-process water sample over a predetermined wavelength range produced in response to an excitation wavelength, normalizing the first fluorescence emission spectrum to a predetermined peak value, measuring a second fluorescence emission spectrum of a post-process water sample over the predetermined wavelength range produced in response to the excitation wavelength, normalizing the second fluorescence emission spectrum to the predetermined peak value, comparing the first and second peak normalized fluorescence emission spectra to determine a change in dissolved organic carbon (DOC), and controlling the water treatment process based on the change in DOC.
Embodiments may also include a system for monitoring a water treatment process that incorporates a coagulation-settling or flocculation-only process. The system may include a first instrument positioned for online sampling of an inlet to the coagulation-settling process, the first instrument measuring a first fluorescence emission spectrum of an inlet sample in response to a first excitation wavelength, a second instrument positioned for online sampling of an outlet from the coagulation-settling process, the second instrument measuring a second fluorescence emission spectrum of an outlet sample in response to the first excitation wavelength, and a computer in communication with the first and second instruments and configured to compare the first fluorescence emission spectrum and the second fluorescence emission spectrum for controlling the coagulation-settling (or flocculation) process.
Systems and methods according to various embodiments of the present disclosure may provide a number of advantages. For example, embodiments may be used to provide a more rapid, precise and accurate indirect determination of one or more water treatment parameters used for monitoring or control of water treatment facilities such as biological oxygen demand (BOD), chemical oxygen demand (COD), total organic carbon (TOC), and trihalomethane formation potential (THMFP) using online absorbance corrected fluorescence excitation-emission spectroscopy.
Various embodiments provide simultaneous determination of the TOC, aromaticity and SUVA parameters using one or more online instruments positioned at one or more key points of a water treatment system. A synchronized calibrated network of online instruments provides full ultraviolet-visible (UV-VIS) absorbance and corrected fluorescence emission spectra on the order of milliseconds to seconds for real-time or near real-time monitoring and control of one or more treatment processes. Embodiments of the system and method of this disclosure facilitate cost-effective improvement of current water treatment infrastructure methodology (such as using enhanced coagulation or additional granular activated carbon) to ameliorate predicted disinfectant byproduct (DPB) spikes as well as to avoid overdosing of coagulant when DBP potential is low.
The online absorbance and fluorescence spectral analyses according to embodiments of this disclosure may be used to identify or flag unknown contaminations as new components. The system or method for indirect determination of a water treatment parameter according to embodiments of this disclosure include a model that uses the spectral analyses to determine various application or process specific treatment parameters to monitor and/or control the treatment process. The system and method of various embodiments may be used to detect membrane fouling agents for a variety of membrane systems including: reverse osmosis, microfiltration, ultrafiltration and membrane bioreactors and forward osmosis (ceramic membranes). Fouling agents may have associated spectral peaks. As such, the system and method may be used for a variety of treatment plants including desalination, wastewater recycling and industrial treatment as well as for shipboard ballast purification systems using membrane technology. The model may be tuned for various applications including surface water treatment, wastewater treatment, or industrial treatment processes. In various embodiments, the model is tuned to protein like peaks for sewage treatment to provide an indirect determination of biologic oxygen demand (BOD) or chemical oxygen demand (COD). Similarly, the model may be tuned to oil peaks for an oil recycling application, etc. Oxygen concentration is another factor that can be quantified in the model as an influencer of the quantum yield of the component species for ozone treatment monitoring by evaluating the absorbance and fluorescence data. Oxygen in aqueous solution, associated with ozone treatment, lowers the fluorescent quantum yield (by collisional quenching) of a given chemical species but normally does not lower its absorbance extinction coefficient. Hence by evaluating the change in fluorescence intensity and absorbance one can evaluate the oxygen concentration and changes in the concentration as a function of the treatment process.
Other applications and advantages of a system or method according to embodiments of the present disclosure include use to effectively monitor and accurately determine the replacement period for bio-active carbon (BAC) filters by analyzing the levels or ratios of C1-C4 components. The BAC activity primarily influences the protein-like peaks following ozone treatment. This can potentially save millions of pounds of activated carbon and millions of dollars a year. One water treatment facility has estimated that it can extend the useful life of their BAC filter mats for up to several years at huge cost savings with an accurate measurement of effectiveness rather than following the recommendation for annual replacements by the carbon mat suppliers.
Using instruments with one or more multichannel detector(s), collection and processing of the absorbance and fluorescence data is effectively instantaneous (within seconds) relative to many offline analysis strategies. Based on the representative embodiments disclosed, algorithms can be easily calibrated and validated to precisely and accurately quantify the compounds removed specifically by coagulation, ozone, or other processes.
The online absorbance and fluorescence measurements may be combined with other online monitor quality metrics including nephelometric turbidity units (NTU), chlorine dose and residual, pH, alkalinity (hardness), and temperature, for example, which all can easily be incorporated into a continuously updated, calibrated predictive model for highly effective determinations of disinfection byproduct formation potential and TOC.
The above advantages and other advantages and features of the present disclosure will be readily apparent from the following detailed description of the preferred embodiments when taken in connection with the accompanying drawings.
Various representative embodiments of systems and methods according to the present disclosure are described in detail. However, it is to be understood that the representative embodiments are merely exemplary and systems and methods according to the present disclosure may be embodied in various and alternative forms. The figures are not necessarily to scale and some features may be exaggerated or minimized to show details of particular components. Therefore, specific structural and functional details disclosed herein are not to be interpreted as limiting, but merely as a representative basis for teaching one of ordinary skill in the art to variously employ the present invention. Simplified flowcharts or block diagrams may be provided to illustrate operation of a representative embodiment of a system or method. Those of ordinary skill in the art will appreciate that the order of steps or processes may not be required for particular applications, that some steps may have been omitted for ease of illustration and description, and that steps or processes may be repeated individually and/or as a subset of the illustrated steps or processes. Likewise, all illustrated or described steps may not be needed to provide one or more the advantages described herein.
As those of ordinary skill in the art will also understand, various features of the present disclosure as illustrated and described with reference to any one of the Figures may be combined with features illustrated in one or more other Figures to produce embodiments of the present disclosure that are not explicitly illustrated or described. The combinations of features illustrated provide representative embodiments for typical applications. However, various combinations and modifications of the features consistent with the teachings of the present disclosure may be desired for particular applications or implementations.
A simplified block diagram illustrating a representative water treatment facility having various water treatment processes is shown in
While a number of variations exist, most surface water treatment facilities have a number of key processes that include dilution, coagulation/flocculation, settling, filtration, and disinfection. Various other chemical treatments or processes may also be used depending on the type of source water and expected contaminants. The complexity and type of treatments or processes may depend on the quality of the source water and variation over time based on the effectiveness of source-water protection and management programs.
As illustrated in
Depending on the particular application and implementation, a single (or common) instrument may be used to measure or monitor a water treatment parameter before and after a particular treatment process rather than having separate instruments, as represented by optical instruments 110, and 126. Plumbing 106, 108 may be provided to route online samples from upstream and downstream of one or more treatment processes to a single/common instrument. While this may provide cost savings by reducing the number of instruments needed for monitoring of the desired treatment processes, the cost savings may be reduced by additional water sample routing and control to a common location, as well the additional manual or automatic hardware required to change the sample flow through the instrument. Instruments or monitors may be positioned at key locations relating to essential steps in the treatment plant as previously described. These usually include at least the inlet, after coagulation-settling, and at the plant effluent. Online monitors positioned at these locations, or one or more central or common monitors that receive online samples from these locations, may be used to sample the water flow at regular time intervals. Water treatment parameters and associated measurement or determination protocols may be developed for particular types of treatment facilities. Representative water treatment parameters and associated representative measurement methods or protocols according to embodiments of the present disclosure are described below.
Water treatment facility 100 may include a flocculation and/or coagulation process as generally represented at 140a and 140b, respectively. As known, coagulation/flocculation may include addition and mixing of chemical coagulants or filter aids that cause small suspended particles to stick together to form larger particles that more readily settle out or can be more easily filtered. The type and quantity of coagulants may depend on the microbial and chemical components of the water. This process in combination with a settling process may be used to remove organic carbon materials from the water to reduce formation of undesirable disinfection byproducts during the subsequent disinfection treatment/process 160. The coagulation process may be monitored and/or controlled by comparing measurements from instrument 124 positioned downstream or subsequent to coagulation process 140 with one or more upstream instruments, such as instruments 110, 122 as described in greater detail herein.
A settling or sedimentation process (not shown) may follow coagulation/flocculation process 140a-b, particularly in applications treating water that contains significant solids. The settling or sedimentation process slows the flow of the water in a basin or pond to allow heavier items to settle to the bottom rather than being carried in the flow to the next process. Online instrument 124 may be positioned downstream of such a sedimentation process with measurements used to determine one or more water treatment parameters used to monitor or control the process by comparing measurements with an upstream instrument, such as instrument 122 or 110, for example. A settling or pre-sedimentation process (not shown) may be used prior to the dilution process 104 in some applications.
After settling or sedimentation, a single or multi-step filtration process 150 may employ a variety of methods to filter particles from the water. The type of filtration may vary based on the raw water quality. As filtration implies, water flows through a material that captures and removes particles, organisms, and/or contaminants. Granular media such as sand, crushed anthracite coal, or granular activated carbon (GAC) are often used as filter media. Different types and sizes of media may be layered and may operate at different flow rates. Membrane filtration may also be used in some applications, but may not be well-suited for highly contaminated source waters because of membrane clogging. Membrane filtration is gaining use in the United States for special applications and in combination with other types of filtration. Online instrument 126 may be positioned downstream of filtration process 150 with associated measurements used to determine water treatment parameters for monitoring or controlling the process. In the representative embodiment illustrated, online optical instrument 126 may receive water samples after filtration process 150, as well as samples after disinfection process 160.
While filtration and the steps prior to filtration focus on the physical removal of contaminants in the water, disinfection is used to kill or inactivate bacteria and viruses that may pass through the physical removal steps. Viruses and organisms like giardia are effectively killed by chlorine. Some organisms, such as Cryptosporidium may be resistant to chlorine, but are susceptible to treatment by ozone and ultraviolet light. In some countries, ozone and UV light may be used without chlorination to kill bacteria and other organisms. In the representative application for an optical instrument and associated monitoring and control of a water treatment facility, disinfection process 160 provides chemical disinfection, but may also generate undesirable disinfection byproducts when reacting with organic carbon components. Disinfectants may include chlorine, chloramines (chlorine plus ammonia), ozone, ultraviolet light, and chlorine dioxide, for example. The advantage of chlorination is that it continues to kill bacteria as water moves through the distribution system 170. Its disadvantage is the possibility of potentially harmful disinfection byproducts. Excess chlorine in water can combine with organic material in the water to form substances such as trihalomethanes, which have been linked to various adverse human health effects over a lifetime exposure. Use of one or more online instruments to monitor and/or control one or more water treatment processes or parameters according to embodiments of the present disclosure may reduce disinfection byproducts by more accurate monitoring and control of one or more upstream processes, such as coagulation process 140b, in addition to more accurate monitoring and control of the disinfection process 160.
Various other treatment processes, may be used to treat or condition the water at various points of the overall treatment process. For example, chemicals may be added to drinking water to adjust its hardness (softness), pH, and alkalinity to reduce corrosion in the distribution system 170, which may include pipelines, storage tanks, and building plumbing systems, for example. Similarly, fluoride may be added to the water to enhance dental health.
As generally illustrated in
Light passing through cuvette 214 is coupled to fiber link 220 and routed to multi-channel detector 250, implemented by a CCD device in various embodiments. Fiber link 220 includes a linear array of fibers to route the light transmitted through cuvette 214 to grating 232 and a separate portion of detector 250 so that a single CCD or other multichannel detector 250 may be used for both absorbance and fluorescence measurements to reduce instrument cost. This is particularly beneficial in reducing system costs associated with incorporating a network of instruments/monitors to monitor or control multiple steps or processes within a water treatment facility according to embodiments of the present disclosure. In addition, use of a fiber coupling or link 220 facilitates insertion of light injection from a solid state reference material and/or a calibrated light source 114 for wavelength and spectral accuracy in calibration of multichannel detector 250. This facilitates in-situ calibration of the instrument as described in greater detail herein.
Light generated by fluorescence of the sample within flow-through cuvette 214 passes through an emission filter 230 to grating 232, which separates the light into wavelength bands before passing through order sorting emission filter 246 to multi-channel detector 250. Similarly, as previously described, light passing through cuvette 214 travels through fiber coupling link 220 and is routed to grating 232 and to detector 250 for obtaining absorbance measurements. Instrument 110 includes a computer 252 to receive the absorbance and fluorescence measurements and correct the fluorescence spectrum using the absorbance measurements. The measurements may be communicated over a local or wide area network 116 to a separate computer/server 120, 150 for use in determining various water treatment parameters and for monitoring and controlling the water treatment process. Alternatively, the computer associated with any particular instrument may be used to determine the water treatment parameters and communicate them to a central monitoring/control computer, server, or system. The computer associated with each instrument may also perform various instrument control functions such as control of sample water routed to cuvette 214, control of one or more shutters (not specifically illustrated) positioned to block light from light source 202, or in the absorbance or fluorescence optical paths for use during calibration etc. The computer may also correct measurements using the reference detector 212.
Other embodiments of the instrument could include use of separate multichannel detector-spectrographs to measure fluorescence and absorbance. However, use of an instrument having a single multichannel detector-spectrograph to simultaneously measure fluorescence and absorbance may provide various advantages. The main advantage of such an instrument is the full spectral correction and coordinated wavelength calibration afforded by the multichannel detector for both absorbance and fluorescence along with streamlined rapid data collection and processing.
Optical instruments/detectors 110, 122, 124, 126 are positioned at select treatment steps to monitor complete absorbance spectrum and inner-filter effect corrected fluorescence emission spectra at selected excitation wavelengths including but not limited to 254 nm (A254) for total and aromatic carbon concentration, key chlorophyll wavelengths, key petroleum or contaminant wavelengths, protein peaks for BOD/COD etc. Regular recording of DOC can be used to calibrate optical online meters as a function of treatment steps and account for quality differences within and between treatment facilities or plants.
In the representative embodiments of
The absorbance at 254 nm (represented by A254 nm) corresponds to:
Various process parameters such as C1, C2, A254, pH, temperature, etc. may be selected to indirectly determine the desired monitoring parameter, such as DOC, TOC, THMFP, etc. Selection of suitable parameters may include identifying the parameters such as C1, C2 pH temperature, etc. having some relationship or correlation with the monitoring parameter(s). Once these parameters are identified, a linear combination of the parameters may be determined. For example, a linear combination such as xA254+yC1/C2+zpH+ . . . . The relationship between measured values and the linear product is used to determine the slope and x, y, and z scaling factors using any of a number of curve fitting strategies, such as multiple linear regression (MLR), for example. This process is repeated over time with measurements from representative samples for the particular application to fine tune the slope, scaling factor(s), and/or base parameters used in the linear combination to achieve a desired correlation between the base parameter and the monitoring parameter.
In one embodiment for monitoring a water treatment process, TOC was selected as the monitoring parameter. To select the base parameters and determine the slope and scaling factors for the linear combination used to indirectly determine TOC, the component concentrations or values for C1 and C2 were selected that correlate with the SUVA value because the SUVA correlates with the aromatic DOC component. The A254/DOC values correlate with the aromatic component concentration of C1 and C2. The C1 component is the humic component with the highest aromaticity and molecular weight (MW) and the strongest correlation with high SUVA aromaticity index. The C2 component is the fulvic/humic component with lower MW and aromaticity and hence correlates with a low SUVA aromaticity index. C3 and other components with deeper UV absorbance and emission than C1 and C2 are generally less influenced by, and therefore not well correlated with coagulation. They are much lower in aromaticity and tend to vary independently of coagulation as compared to C1 and C2 so they are not effective predictors of aromaticity. As such, C3 and C4 were not selected because including these components in an MLR prediction of TOC or DOC would result in poor correlation and predictive capability of the selected process monitoring parameter.
The integral of the DOC fluorescence spectrum measured from 290-600 nm in response to excitation at 254 nm is represented by:
The quantum contribution of each component (C1-C4) excited at 254 nm is defined in general as:
ΦCx=A254Cx·T·pH
Where temperature (T) and (pH) may influence quantum contributions but generally the contributions statistically are negligibly different among the four components and pH also plays a minor role across treatment process for the four components, meaning temperature and pH are normally not large factors in typical surface water treatment applications, but are accounted for in the model to provide broader applicability.
Water treatment directly alters the apparent molar extinction coefficient (ε254 nm) primarily by changing the relative concentrations of the components (Cx), as opposed to changing their quantum yields or generating new fluorescent components. As such, evaluating the (normalized) spectral profile of the fluorescence spectrum excited at 254 nm provides a direct measurement of the relative concentrations of the components C1-C4. The normalized spectral profiles captured from samples taken before and after a particular treatment process or series of treatment processes may be used to monitor and/or control the process(es) as described in this disclosure. Evaluation of the spectral profiles may be performed using various techniques as generally represented and described with reference to
The largest influence of coagulation (as a representative treatment example) is to reduce the apparent molar extinction coefficient (ε254 nm) by reducing the concentration of C1 relative to the total or sum of components of Ctot=(C1+C2+C3), which provides a direct measurement of the aromatic concentration. The ratio of C1:Ctot is linearly related to, and the primary cause of the change in slope of A254/DOC, which is also referred to as the specific UV absorbance (SUVA). In many applications, various treatment processes may not affect all of the components C1-C4 equally. As such, the particular ratio used to monitor or control a particular process may vary. For example, the concentrations of C3 and C4 are not significantly changed by a coagulation treatment process. As such, the process can be monitored, evaluated, and/or controlled by determining corresponding treatment parameters, such as DOC, using a ratio of C1:C2 rather than C1:Ctot.
As one can see from the above description, C1:Ctot=xSUVA where x is a simple linear scaling factor such that the relationship remains linear for:
DOC=A254 nm/ε254 nm·1
because changes in the molar extinction coefficient (ε254 nm) are linearly related to the sum of the components represented by Ctot in this example.
As previously described, C4 is not normally detected as a natural component in surface water treatment. However, it may be monitored for detection of particular types of pollutants and is highly prevalent in sewage sourced wastewater and water sources receiving treated sewage. The changes in components represented by C1-C4 can be monitored continuously and evaluated by comparing spectra from the different detectors 110, 122, 124, 126, etc. in the network within a particular treatment facility 100, or among treatment facilities. Representative comparisons are illustrated and described with respect to
The optical signal processing of the present disclosure is designed to monitor two main sources of relative and absolute component variation of the DOC. For natural and manmade raw source variation, the relative DOC component contribution varies depending on several factors of the source water including rain and storm events, fire events, leaf fall events, algal blooms, sewage discharge, etc. For treatment process variation, the primary effect of the coagulation and filtration processes is to influence the total DOC as well as the relative aromatic carbon component ratios. The main function of the treatments is to reduce the probability of the formation of regulated and undesirable halogenated disinfection byproducts by reducing the precursor components contained within the DOC.
The flowcharts of
With continuing reference to
Alternatively, or in combination, an absolute concentration ratio of components may be determined as represented by block 318. In one embodiment, an absolute concentration ratio of C1:C2 is determined along with the corresponding A254 measurement for use as linear coefficients (x and y) with the intercept fixed at 0 mg/l to predict DOC. According to embodiments of the present disclosure, predicted DOC may be represented by a generally linear function (x[C1]:[C2]+yA254) as described in greater detail with reference to
A multiple linear regression (MLR) is used to determine the x and y coefficients and corresponding slope as represented by block 320 and as described in greater detail with reference to
The calibration may be periodically checked as generally represented by block 324 although this is not part of the calibration process per se, but is provided for purposes of illustration of a representative water treatment process. In particular, the predictive MLR model may be applied in continuous sampling mode with periodic re-evaluation compared to daily batch analyses, for example. The C1:C2 ratio may also be determined as a surrogate of the Aromatic Index and compared to SUVA.
As illustrated in the flowchart of
The absorbance spectral data (250 nm to 600 nm=−log(l/lo)) is used to calculate a corrected fluorescence spectrum to adjust for primary and secondary inner filter effects (IFE) as represented at 348. The fluorescence emission spectrum excited at 254 nm provides quantitative information on the TOC fractions (C3, C4) associated with proteins (aromatic amino acids), but more importantly (for the representative embodiment illustrated) the lower molecular weight and aromaticity (C2) and the higher molecular weight and aromaticity humic and fulvic acid components (C1) associated with the SUVA parameter. The selected water treatment parameters, SUVA and TOC in this example, are then calculated or determined based on the fluorescence intensity C3, fluorescence intensity ratio C2:C1 and the absorbance A254 as represented at 350. The TOC can be determined based on the absorbance and fluorescence data using the predetermined relationship between the measured TOC using a TOC meter and the predicted TOC, which may be represented as the linear combination of the absorbance and fluorescence as illustrated and described in greater detail herein, for example.
Other water treatment parameters may also be determined or calculated using the absorbance and fluorescence measurements. In the representative embodiments illustrated in
The fluorescence spectra data 430 can be analyzed using various quantitative analysis strategies as generally represented by blocks 440, 442, 444, and 446 and illustrated and described in greater detail with reference to
As demonstrated by the strong positive correlation illustrated in data 418, the absorbance and fluorescence data can be used to indirectly determine a water treatment parameter, such as TOC, using a corresponding model without the need for a corresponding TOC meter. This provides various advantages with respect to the ability to accurately monitor the treatment process with an online instrument as previously described. In particular, use of the absorbance and fluorescence data provide a reproducible evaluation for different water sources that may have variable aromatic properties over time or between sources. The kinetic simultaneity of the absorbance and fluorescence measurements provide a more precise determination of SUVA that eliminates propagated noise and interferences of the separate detection methods that rely on a TOC meter as conventionally implemented. The more precise and accurate TOC and aromaticity evaluations afforded according to embodiments of the present disclosure facilitate compliance with more stringent regulations for water treatment plants.
While the representative embodiment illustrated in
As previously described, the changes in C1-C4 or other components being monitored in particular applications can be monitored continuously and evaluated by comparing absorbance and corrected fluorescence spectra from different detectors positioned to analyze samples from desired points within a particular treatment process or series of processes. Various numerical operations may be employed for quantitatively evaluating the spectral shape change associated with the change in concentration ratios from a particular source or as a result of a particular treatment process or series of processes.
As illustrated and described with reference to
As illustrated in
Because the system and method rely primarily on the relative changes in the normalized fluorescence spectral shape, area, and maxima, the component concentration changes for the fluorescence spectra can also be calculated numerically based on ratiometric, subtractive, and/or derivative analyses to speed calculation rather than using a calibrated classical least squares (CLS) approach to fit the fluorescence data to the known component spectral shapes as previously described. In addition, absorbance and fluorescence ranges can be extended to include chlorophylls and accessory pigments from algae with appropriate excitation filters and gratings in the associated online instrument(s). In addition, the relationship between protein components (C3) can also be used to monitor biochemical and chemical oxygen demand parameters for waste-water treatment as previously described.
During development and model validation, samples were collected to compare the fluorescence and absorbance spectral data of raw source water from a canal and after coagulation/filtration treatment. The model used to treat the data was first validated using complete fluorescence excitation-emission matrices (EEMs) corrected for fluorescence inner filter effects and with parallel sample measurements of the total organic carbon measured independently with a TOC meter. The EEM data set originally comprised 204 samples and was fit to a validated three component parallel factor analysis model with a high degree of correlation (greater than 98%) and split half validated correlation (greater than 97.8%). The three components were clearly identified as a low molecular weight humic/fulvic component (C1), a high-molecular weight humic/fulvic component (C2), and protein components (C3). Analysis of samples taken after a coagulation/filtration treatment demonstrated that the treatment lowers the relative concentration of C2 compared to C1 with little if any effect on C3.
The SUVA data calculated from the A254 nm and TOC parameters was compared to the ratio of the C2:C1 components as a function of analysis date for samples collected over the course of several days to monitor the effect of sample variation. Comparison of the SUVA parameter to the C2:C1 ratio was used to demonstrate that changes in SUVA (measured using separate absorbance and TOC meter readings) may be more precisely measured using the C2:C1 ratio measured simultaneously with absorbance and corrected fluorescence instrument according to embodiments of the present disclosure. The SUVA calculations showed raw water was greater than 4 and settled and effluent samples were generally around 2-3 with significant (>19 to 13%) scatter in the treated samples.
The C2:C1 ratio, which is a linear factor in relation to TOC, mirrored the SUVA trend with a lower coefficient of variance in the treated samples (<5 to 3%) for better resolution of aromaticity changes among the three treatments. Monitoring of the A254 and TOC meter data as a function of analysis date was performed to illustrate the consistency of the changes in the source water composition. The measured TOC and A254 values used for the SUVA calculations were used to demonstrate that treatment resulted in two linear relationships between the raw and treated water consistent with the SUVA and C2:C1 ratios. Based on this data, it is clear the changes in the linear relationships between the raw and treated data sets as a function of TOC is consistent with the systematic change in the molar extinction of the samples that is determined by the relative reduction in the C2:C1 component ratio according to Beer's Law.
As such, the present disclosure recognizes that knowing the C2:C1 ratio, concentrations, or other correlated values, and A254 value enables the calculation (by the process of multiple linear regression or similar methods) of the TOC for the data sets using one predictive equation. The multiple linear regression equation results indicate a high degree of precision and accuracy of the TOC (R2>0.988). Therefore, knowing the fluorescence emission spectrum and absorbance spectrum allows simultaneously determination of the SUVA and TOC concentrations with one instrument that can operate online for TOC ranges from <<0.5 to >10 mg/l with high confidence. Online instruments according to the present disclosure can determine these and similar water treatment parameters or indicators within a matter of seconds for each sample using instruments having few moving parts.
The sample flow can be temporarily halted with manually or automatically controlled valves to route a particular sample to the flow through cuvette as represented at 912 for subsequent absorbance and fluorescence determination to provide analysis synchronization as described below. The complete UV-NIR absorbance spectrum is measured from 230-800 nm using an order sorting filter equipped (OSF) CCD-spectrograph until a desired signal to noise is obtained for each detector location as indicated at 914. The fluorescence path is shuttered to prevent scattered light from entering this path to the CCD during absorbance measurement at 916. The absorbance path is corrected by the reference diode signal to account for light source fluctuations as represented at 918.
The sample's fluorescence emission spectrum is excited at 254 nm and the signal is integrated until the desired signal to noise is reached for each detector location as represented at 930. The emission path is filtered to eliminate any excitation light and the absorbance/transmittance path is shuttered to prevent illumination of the CCD through this path during fluorescence measurement at 932. The fluorescence spectrum is corrected for emission spectral response and the reference diode at 934. The A254 nm value is evaluated from the absorbance spectrum and if it exceeds the threshold saturation value an associated message is generated at 936. If no saturation warning is recorded, then the absorbance spectral data is used to apply the primary and secondary inner filter effect corrections to the fluorescence spectral data to achieve a correct fluorescence response at 938.
The fluorescence spectra from each detector are processed by calculating a cumulative integral curve where the total integral is normalized to unity at 940. The treated water signals from each step are subtracted from the source water signal to evaluate the effect of the associated treatment process at 942. The difference curve is evaluated in terms of area and wavelength peak and width to determine when the treatment process, which is a coagulation process in one representative embodiment, has had the desired effect 944. The cumulative integrals and difference curve data are further processed with the A254 nm value using the previously described wavelength coordinate evaluations and equations at 946 to provide the raw signal data parameters used to calculate/calibrate the total dissolved organic carbon (DOC) equivalent.
As also shown in
The DOC is calculated using the multiple linear regression derived parameter coefficients x and y based on the independent A254 nm and C2:C1 (or Ctot:C1) intensity ratios, respectively as represented at 960 according to: DOC=x(A254 nm)*y (C2:C1), where x and y are linear coefficients, typically the regression is constrained to an intercept of 0 and slope of unity to increase the predictive correlation capacity and significance. The A254 nm and C2:C1 (or Ctot:C1) signals would be calibrated using TOC meter data sampled in a batch mode comparison from the corresponding processing steps (locations) at corresponding times using the simple multiple linear regression (MLR) formula above to determine x and y for predictive capacity. The term y*(C2:C1) is linearly correlated by definition to the parameter more widely known as the specific UV absorbance coefficient (SUVA) since C2:C1 (or Ctot:C1) represents the change in the relative component concentrations for the water samples which determines the linear relationship between A254 nm and DOC.
The linear analysis is based on recognition that the molar extinction coefficients for each component remain constant as a function of treatment such that the A254 nm changes only represent changes in the absolute and relative component concentration(s). In addition, the fluorescent quantum yields for each component also remain constant (both absolutely and in relation to each other) during treatment such that changes in the normalized fluorescence spectrum and integral shapes also only indicate relative changes in the component concentrations. Minor changes in quantum yield can be accounted for in the fluorescence spectral modeling when measured/detected as a function of the relationship between absorbance and fluorescence. Such changes may be associated with pH, temperature, and dissolved oxygen concentration.
As previously described, the A254 nm and the C2:C1 or (Ctot:C1) ratio can also be used to accurately predict disinfection byproduct formation potential in a multiple linear regression scheme potentially including several variables measured with other corresponding meters in the plant operation. Correlated/determinant parameters can include, but would not be limited to, chlorine dose and residual, contact times, temperature, pH, alkalinity and turbidity. A typical trihalomethane formation potential prediction (THMFP) equations is defined as:
THMFP (ppb)=a(A254 nm)+b(C2:C1)+c(pH)+d([C1−])+e(T)+f[Alk])
where a, b, c, d, and e are linear coefficients. Typically, as with the TOC prediction, an intercept of 0 and slope of unity can be constrained in the linear equation fit parameters to increase the predictive correlation.
As those of ordinary skill in the art will appreciate, representative embodiments previously described may use a ratio of components, such as C2:C1 or C1:C2. Alternatively, or in combination, component concentrations, component values, or other parameters that correlate with the component concentrations or values may be used depending on the particular application and implementation. For example, the peak height associated with a particular component can be used as previously described. Furthermore, various embodiments do not require measurement of the emission spectra or dispersing of the emitted light. Rather, a diode may be used as the detector with acceptable results for some applications. Similarly, IFE correction may not be required in some applications.
As such, various embodiments of a water treatment or similar process monitor according to the present disclosure may provide savings or reduced use of associated treatment chemicals, such as coagulation chemicals and granulated activated carbon (on the order of tens of thousands of dollars per month for a typical application). Similarly, various embodiments may provide savings or reduced use of Biologically Activated Carbon (bGAC) in terms of activity needed to remove DOC. Likewise, embodiments may reduce membrane fouling and associated pumping energy as well as reducing membrane damage and increasing membrane useful life for Reverse Osmosis (RO), Forward Osmosis (FO), Membrane Bioreactors (MBR), Microfiltration (MF) and Ultrafiltration (UF) applications. Online monitoring of water treatment parameters according to various embodiments may be used for optimization of ozone treatment and reduction in taste and odor objections. Furthermore, various embodiments may be used to improve detection of organic contaminants including oil.
While exemplary embodiments are described above, it is not intended that these embodiments describe all possible forms of a system or method for analyzing a sample to determine a water treatment parameter or indicator according to the present disclosure. Rather, the words used in the specification are words of description rather than limitation, and it is understood that various changes may be made without departing from the spirit and scope of the disclosure. As previously described, the features of various representative embodiments may be combined in ways that are not explicitly illustrated or described to form further embodiments. While various embodiments may have been described as providing advantages or being preferred over other embodiments with respect to one or more desired characteristics, as one of ordinary skill in the art is aware, one or more characteristics may be compromised to achieve desired system attributes, which depend on the specific application and implementation. These attributes include, but are not limited to: cost, strength, durability, life cycle cost, marketability, appearance, packaging, size, serviceability, weight, manufacturability, ease of assembly, operation, etc. Any embodiments described herein as less desirable than other embodiments or prior art implementations with respect to one or more characteristics are not outside the scope of the disclosure and may be desirable for particular applications.
This application is a divisional of U.S. application Ser. No. 15/586,542 filed May 4, 2017, which is a continuation of U.S. application Ser. No. 14/527,752 filed Oct. 29, 2014, the disclosures of which are hereby incorporated by reference in their entirety.
Number | Name | Date | Kind |
---|---|---|---|
3666362 | Chance | May 1972 | A |
4279511 | Maute et al. | Jul 1981 | A |
4305660 | Kallet | Dec 1981 | A |
4475813 | Munk | Oct 1984 | A |
4927265 | Brownlee | May 1990 | A |
5285254 | De Sa | Feb 1994 | A |
5414508 | Takahashi et al. | May 1995 | A |
5645799 | Shah et al. | Jul 1997 | A |
6236456 | Giebeler et al. | May 2001 | B1 |
6970241 | DeSa | Nov 2005 | B1 |
7209223 | Hull et al. | Apr 2007 | B1 |
7265827 | Slutter et al. | Sep 2007 | B2 |
7324202 | Leonard et al. | Jan 2008 | B2 |
7569839 | Gilmore et al. | Aug 2009 | B2 |
7595881 | Leonard et al. | Sep 2009 | B2 |
8901513 | Gilmore et al. | Dec 2014 | B2 |
20030034292 | Rela | Feb 2003 | A1 |
20050057753 | Mosley et al. | Mar 2005 | A1 |
20050264803 | Jones | Dec 2005 | A1 |
20060219630 | Abe et al. | Oct 2006 | A1 |
20070037135 | Barnes et al. | Feb 2007 | A1 |
20080123095 | Hubner et al. | May 2008 | A1 |
20080174767 | Leonard et al. | Jul 2008 | A1 |
20080192249 | Babichenko et al. | Aug 2008 | A1 |
20080272312 | Tuschel | Nov 2008 | A1 |
20100308234 | Harju et al. | Dec 2010 | A1 |
20110186753 | Dixon | Aug 2011 | A1 |
20110311987 | Fukui | Dec 2011 | A1 |
20120228519 | Gilmore et al. | Sep 2012 | A1 |
20130064776 | El Harrak | Mar 2013 | A1 |
20160123882 | Gilmore et al. | May 2016 | A1 |
Number | Date | Country |
---|---|---|
1840483 | Oct 2006 | CN |
103765211 | Apr 2014 | CN |
0614085 | Sep 1994 | EP |
0616211 | Jan 1999 | EP |
2002328091 | Nov 2002 | JP |
2003088882 | Mar 2003 | JP |
2003090787 | Mar 2003 | JP |
2003090797 | Mar 2003 | JP |
2007014955 | Jan 2007 | JP |
2007326047 | Dec 2007 | JP |
0063680 | Oct 2000 | WO |
2011069067 | Jun 2011 | WO |
Entry |
---|
Zsolnay et al. Differentiating with fluorescence spectroscopy . . . , Chemosphere, vol. 38, No. 1, pp. 45-50, 1999 (Year: 1999). |
Cohen et al. (Fluorescent components of organic matter in wastewater: Efficacy and selectivity of the water treatment, Water Research, vol. 55, pp. 323-334, Mar. 2, 2014 (Year: 2014). |
Baghoth et al. (Tracking natural organic matter . . . , Water Research, vol. 45, pp. 797-809, Sep. 15, 2010 (Year: 2010). |
Li et al. (THMs and HAAs formation by tryptophan during chlorination disinfection, 2009 International Conference of Energy and Environment Technology, pp. 888-891 (Year: 2009). |
Kling. “Fluorometers Dissected, Fluorometry offers lower detection limits for various applications”, Analytical Chemistry Mar. 1, 2000, p. 219A-224A. |
Holland et al. “A Unique Computer Centered Instrument for Simultaneous Absorbance and Fluorescence Measurements”, Analytical Chemistry Jan. 1973, vol. 45, No. 1, p. 145-153. |
Monte et al. “Minimizing Uncertainty for Traceable Fluorescence Measurements—The BAM Reference Fluorometer”, Federal Institute for Materials Research and Testing (BAM) Working Group Optical Spectroscopy, I.3, Richard-Willstatter-Strabe 11, D-12489 Berlin, Germany. 2005, 1 Page. |
“PHAR 7633 Chapter 24, Pharmaceutical Analysis, Drug Quantitation”, http:www.boomer.org/c/p4/c24-c2403.html, Retrieved Dec. 9, 2010, 8 Pages. |
“Fluorescence spectroscopy—Wikipedia, the free encyclopedia”, http://en.wikipedia.org/wiki/Fluorescence_spectroscopy, Retrieved Dec. 9, 2010, 5 Pages. |
“STAR Grant R832738: Rapid Detection of Trace Endocrine Disrupting Chemicals in Complex Mixtures: A Full-Spectrum Deconvolution Technique with a UV-Transparent Passive Concentrator”, http://www.epa.gov/ppcp/project/star-grant1.html, Retrieved Aug. 6, 2010, 2 Pages. |
J.R. Lakowicz “Principles of Fluorescence Spectroscopy”, ISBN No. 978-0-387-31278-1, p. 27-61. 2006. |
Chen et al. “Light Source Compensation in Absorption and Transmission Spectral Measurements”, Acton Research Corporation, p. 1-3, 1998. |
McPherson “Additive or Subtractive Mode Double Monochromator”, http://www.mcphersoninc.com/spectrometers/uvvisir/model275d.htm, 2006, p. 1-2. |
International Preliminary Report on Patentability dated Sep. 10, 2013, for corresponding PCT Application PCT/US2012/027833 filed Mar. 6, 2012. |
International Search Report for PCT/US2012/027833, dated May 30, 2012. |
Extended European Search Report for European Patent Application No. 12754324.7 dated Aug. 21, 2014. |
A simple correction method of inner filter effects affecting FEEM and its application to the PARAFAC decomposition, by Xavier Luciani, Stéphane Mounier, Roland Redon, André Bois; 2009; 28 Pages <https://hal.archives-ouvertes.fr/hal-00807769>. |
Fluorescence of Dyes in Solutions with High Absorbance. Inner Filter Effect Correction, by Alexander V. Fonin, Anna I. Sulatskaya, Irina M. Kuznetsova, Konstantin K. Turoverov; Jul. 2014 | vol. 9 | Issue 7 | e103878; 8 pages <www.plosone.org>. |
Inner filter correction of dissolved organic matter fluorescence, by Dolly N. Kothawala, Kathleen R. Murphy, Colin A. Stedmon, Gesa A. Weyhenmeyer, and Lars J. Tranvik; Limnology. and Oceanography.: Methods 11, 2013, 616-630; 2013, by the American Society of Limnology and Oceanography, Inc.; 15 pages. |
Singapore Search Report dated Oct. 19, 2020, for related Singapore Appln. No. 10201909423P; 3 Pages. |
Singapore Written Opinion dated Oct. 30, 2020 for related Singapore AppIn. No. 10201909423P; 7 Pages. |
Baghoth, S. A. et al., Tracking Natural Organic Matter (NOM) in a Drinking Water Treatment Plant Using Fluorescence Excitation Emission Matrices and PARAFAC. Water Research, Sep. 15, 2010, vol. 45, pp. 797-809. |
Chinese Office Action for Chinese Patent Application No. 2015800592238 dated May 8, 2019. |
RK. Henderson et al.; Fluorescence as a Potential Monitoring Tool for Recycled Water Systems: A Review; Water Research 43; ScienceDirect; 2009; pp. 863-881. |
Partial European Search Report for European Patent Application No. 15853829.8 dated Mar. 14, 2018. |
Supplementary European Search Report for European Application No. 15853829.8 dated May 29, 2018. |
Search Report and Written Opinion for Singapore Patent Application No. 11201703257R dated Mar. 21, 2018. |
Yang, L. et al.; Estimating the Concentration and Biodegradability of Organic Matter in 22 Wastewater Treatment Plants Using Fluorescence Excitation Emission Matrices and Parallel Factor Analysis; Sensors; Jan. 20, 2014; vol. 14; pp. 1771-1786. |
Hur, J. et al.; Prediction of BOD, COD, and Total Nitrogen Concentrations in a Typical Urban River Using a Fluorescence Excitation-Emission Matrix with PARAFAC and UV Absorption Indices; Sensors; Jan. 16, 2012; vol. 12; pp. 972-986. |
Hao, R. et al.; Use of Three-dimensional Excitation and Emission Matrix Fluorescence Spectroscopy for Predicting the Disinfection By-product Formation Potential of Reclaimed Water; Water Research, Nov. 1, 2012; vol. 46; pp. 5765-5776. |
Guo, W. et al.; Characterization of Dissolved Organic Matter in Urban Sewage Using Excitation Emission Matrix Fluorescence Spectroscopy and Parallel Factor Analysis; Journal of Environment Sciences; Nov. 2010; vol. 22, No. 11; pp. 1728-1734. |
International Preliminary Report on Patentability for International Application No. PCT/US2015/056138 dated May 2, 2017. |
Second Singapore Written Opinion dated Octobe Aug. 8, 2022 for related Singapore Application No. 10201909423P; 6 Pages. |
Number | Date | Country | |
---|---|---|---|
20210255105 A1 | Aug 2021 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 15586542 | May 2017 | US |
Child | 17307341 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 14527752 | Oct 2014 | US |
Child | 15586542 | US |