The present invention relates generally to devices and methods for a dosimeter. More specifically, the present invention relates to devices and methods for detecting radiation-induced cell damage resulting from delivered radiation exposure.
Without limiting the scope of the disclosed devices and methods, the background is described in connection with devices and methods for detecting radiation-induced cell damage by directly measuring DNA damage resulting from delivered radiation exposure.
A dosimeter is a device that measures exposure to ionizing radiation. These devices are often utilized to protect humans from radiation exposure, as well as the measurement of radiation dose in applications such as medical and industrial. There are several types of dosimeters such as the electronic personal dosimeter (EPD), the film badge dosimeter, the quartz fiber dosimeter, and the thermo luminescent dosimeter (TLD). Each of these dosimeters relies on different mechanisms of action to obtain their measurements.
Known approaches in the art utilize an exposure-based approach. That is, the measurement of dose is based on delivered radiation exposure. This approach often requires calibration of machines/devices to correctly measure or estimate exposure to dose. For small, field doses, a set of correction factors are used on top of extrapolations from exposure to dose and then dose to biological damage.
As an example, in radiation therapy, ionizing radiation is used to shrink tumors and kill cancer cells by damaging the DNA within the actively replicating cancer cells. These procedures have traditionally involved treatment areas of 4 cm×4 cm and larger. The process for determining the resulting cellular damage involves using mathematical models specific to the radiation equipment and equipment calibration dosimeter to convert the machine output to an estimated biological cellular effect. Stereotactic radiation therapies, Stereotactic Radiosurgery (SRS) for treatment of brain tumors, and Stereotactic Body Radiotherapy (SBRT) for treatment of body tumors, target much smaller areas. These target areas are often at or below 1 cm×1 cm and some of the assumptions used in the models for larger fields do not accurately reflect the delivered dose. This is particularly critical in SRS, which uses very small field sizes and requires accuracy of 1-2 mm. What is desired therefore, is a dosimeter that circumvents the calculated connections between exposure, dose, and biological damage with one that directly measures DNA damage resulting from delivered radiation. The measured DNA damage would now directly correlates with the biological cellular effect allowing for more accurate delivery of the intended dose.
In view of the foregoing, it is apparent that there exists a need in the art for a device and method to directly measure biological damage resulting from delivered radiation exposure, which overcomes, mitigates, or solves the above problems in the art. It is the purpose of this invention to fulfill this and other needs in the art, which will become apparent to the skilled artisan once given the following disclosure.
The present invention, therefore, provides for devices, systems, and methods directed to a dosimeter.
In one embodiment, the device is comprised of magnetic streptavidin beads and strands of DNA configured with a biotin on one end and fluorescein on the other end. The biotin causes the DNA strand to adhere to the magnetic streptavidin bead. The configured DNA strand adhered to the magnetic streptavidin bead will be referred to as the marker element. In embodiments, the device is further comprised of a solution element and a body element. The marker elements are held in the solution element that is held by the body element. In embodiments, the solution element is a physiological solution.
The DNA dosimeter or device in embodiments is utilized using the following steps: the DNA dosimeter or device is measured with a fluorescence reader prior to being radiated. The measured fluorescence will serve as the baseline. The DNA dosimeter or device produces a signal that is proportional to the number of configured DNA strands held in the body element. The DNA dosimeter or device is then radiated. A magnet is then placed up against the body element and the contents within the body element are then washed. The magnetic streptavidin beads stick to the magnet and remain in the body element. If a double strand break has occurred for a DNA strand, the attached fluorescein is washed away. Thus, when the fluorescence of the DNA dosimeter or device is read after the delivered radiation exposure, the amount fluorescence amount removed from the original signal is proportional to the probability of a double strand break.
In summary, the present invention discloses devices, systems, and methods directed to a dosimeter. More specifically, the present invention relates to devices, systems, and methods for detecting radiation-induced cell damage by directly measuring DNA damage resulting from delivered radiation exposure.
The accompanying drawings, which are incorporated in and form a part of the specification, illustrate a preferred embodiment of the present invention, and together with the description, serve to explain the principles of the invention. It is to be expressly understood that the drawings are for the purpose of illustration and description only and are not intended as a definition of the limits of the invention. In the drawings:
Disclosed herein are devices, systems, and methods directed to detecting radiation-induced cell damage by directly measuring DNA damage resulting from delivered radiation exposure. The numerous innovative teachings of the present invention will be described with particular reference to several embodiments (by way of example, and not of limitation).
Reference is first made to
The DNA dosimeter or device in embodiments is utilized using the following steps: the DNA dosimeter or device is measured with a fluorescence reader prior to being radiated. The measured fluorescence will serve as the baseline. The DNA dosimeter or device produces a signal that is proportional to the number of configured DNA strands held in the body element. The DNA dosimeter or device is then radiated. This is shown in
Reference is next made to
The following is illustrative of the method and steps for the preparation of a 400 μL sample of the DNA dosimeter at regular concentration.
PCR Setup (400 μL)
During all steps we should keep all materials in ice.
Running the Gel
Attaching the Beads (Immobilization Procedure—100 μL of PCR+100 μL of Beads)
Regular concentration is the concentration of the DNA double-strand-break dosimeter. It has been shown that using a more concentrated dosimeter might increase the signal to noise ratio and the response at low doses.
Checking the Fluorescence
Dilute both beads and supernatant with a factor of 50 (1 μL B/S+49 μL dH2O)
Place 25 μL of the supernatant+6 μL of Binding buffer+19 μL of PBS in the 1st well.
Place 25 μL of the beads+25 μL of binding buffer in the 2nd well.
Place 50 μL of PBS in the 3rd well.
In embodiments, the streptavidin beads are opaque. Opacity affects the efficiency of reading the fluorescence signal of the DNA strands, which attach to the beads. In embodiments, the binding efficiency may also be calculated in the following manner. During the immobilization procedure, the PCR product is attached to the streptavidin beads. The PCR represents the total number of DNA double strands attempting to attach, while the supernatant in the immobilization procedure represents the number of DNA double strands which did not attach to the beads. Theoretically, the subtraction of PCT and supernatant fluorescence signals will represent the number of DNA double strands attached to the streptavidin beads. It is recommended to read at least 50 μL in each plate well for more precise readings, but reading high volumes of PCR will oversaturate the response. In embodiments smaller volumes of both PCR and supernatant are utilized to calculate binding efficiency. These small volumes represent equivalent volumes and ratios to what was used during the immobilization procedure. The PBS buffer was used for dilution purposes and to achieve a volume of 50 μL in each plate well.
To calculate the binding efficiency, in an embodiment, the following steps are taken:
In a first well, place is 1 μL of PCR product diluted with 49 μL of PBS (PCR).
In a second well, placed is 4 μL of supernatant diluted with 46 μL of PBS (S).
In a third well. placed is 50 μL of PBS for the background signal (BG).
The disclosed devices, systems, and methods are generally described, with examples incorporated as particular embodiments of the invention and to demonstrate the practice and advantages thereof. It is understood that the examples are given by way of illustration and are not intended to limit the specification or the claims in any manner.
To facilitate the understanding of this invention, a number of terms may be defined below. Terms defined herein have meanings as commonly understood by a person of ordinary skill in the areas relevant to the present invention. Terms such as “a”, “an”, and “the” are not intended to refer to only a singular entity, but include the general class of which a specific example may be used for illustration. The terminology herein is used to describe specific embodiments of the invention, but their usage does not delimit the disclosed device or method, except as may be outlined in the claims.
Alternative applications for this invention include using the disclosed devices, systems, and methods for measuring shallow skin doses as well as any organization concerned with the impact a new drug could have in regards to DNA damage. Consequently, any embodiments comprising a one piece or multi piece system having the structures as herein disclosed with similar function shall fall into the coverage of claims of the present invention and shall lack the novelty and inventive step criteria.
It will be understood that particular embodiments described herein are shown by way of illustration and not as limitations of the invention. The principal features of this invention can be employed in various embodiments without departing from the scope of the invention. Those skilled in the art will recognize, or be able to ascertain using no more than routine experimentation, numerous equivalents to the specific systems and methods described herein. Such equivalents are considered to be within the scope of this invention and are covered by the claims.
All publications, references, patents, and patent applications mentioned in the specification are indicative of the level of those skilled in the art to which this invention pertains. All publications, references, patents, and patent application are herein incorporated by reference to the same extent as if each individual publication, reference, patent, or patent application was specifically and individually indicated to be incorporated by reference.
In the claims, all transitional phrases such as “comprising,” “including,” “carrying,” “having,” “containing,” “involving,” and the like are to be understood to be open-ended, i.e., to mean including but not limited to. Only the transitional phrases “consisting of” “consisting essentially of,” respectively, shall be closed or semi-closed transitional phrases.
The devices, systems, and/or methods disclosed and claimed herein can be made and executed without undue experimentation in light of the present disclosure. While the devices, systems, and methods of this invention have been described in terms of preferred embodiments, it will be apparent to those skilled in the art that variations may be applied to the devices, systems, and/or methods and in the steps or in the sequence of steps of the methods described herein without departing from the concept, spirit, and scope of the invention.
More specifically, it will be apparent that certain components, which are both shape and material related, may be substituted for the components described herein while the same or similar results would be achieved. All such similar substitutes and modifications apparent to those skilled in the art are deemed to be within the spirit, scope, and concept of the invention as defined by the appended claims.
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US17/16619 | 2/4/2017 | WO | 00 |
Number | Date | Country | |
---|---|---|---|
62290985 | Feb 2016 | US |