This development relates generally to systems, devices and methods for excluding the left atrial appendage (LAA). In particular, systems, devices and methods for excluding the LAA using an expandable foam implant with a deployable and compliant frame are described herein.
Atrial fibrillation (Afib) is a condition in which the normal beating of the left atrium (LA) is chaotic and ineffective. The left atrial appendage (LAA) is a blind pouch off the LA. In patients with Afib blood stagnates in the LAA facilitating clot formation. These clots (or clot fragments) have a tendency to embolize or leave the LAA and enter the systemic circulation. A stroke occurs when a clot/clot fragment embolizes and occludes one of the arteries perfusing the brain. Anticoagulants, e.g. Coumadin, have been shown to significantly reduce the stroke risk in Afib patients. These drugs reduce clot formation but also increase bleeding complications including hemorrhagic strokes, subdural hematoma, and bleeding in the gastrointestinal tract.
There are about eight million people in the US and EU with Afib. About 4.6 million of these patients are at a high risk for stroke and would benefit from anticoagulation. A large portion of these patients cannot take anticoagulants due to an increased bleeding risk, leaving their stroke risk unaddressed. The prevalence of Afib increases with age.
Existing devices for occluding the LAA have drawbacks. Existing devices are offered in many sizes and must be closely matched to the highly variable LAA anatomy. This is difficult to do using fluoroscopy and often requires adjunctive imaging in the form of transesophageal echocardiography (TEE), cardiac CT and MRI, all with three dimensional reconstructions. If the device is significantly oversized, the LAA ostium may become overstretched leading to tearing, resulting in bleeding into the pericardial space. If the device is too small, it will not adequately seal the ostium and may be prone to embolization. Even if sized correctly, the device forces the oval LAA ostium to take the round shape of the device, often resulting in residual leakage at the edges due to poor sealing.
Existing devices require sufficient spring force or stiffness to seal and anchor to surrounding tissue. If too stiff, these devices may lead to leaking of blood through the tissue into the pericardial space which may lead to cardiac tamponade. Furthermore, the geometry of these devices limits repositioning once the implant is fully expanded. Existing devices also complicate delivery by requiring positioning in the LAA coaxial to the axis of the LAA.
There is therefore a need for an improved LAA occlusion device.
The embodiments disclosed herein each have several aspects no single one of which is solely responsible for the disclosure's desirable attributes. Without limiting the scope of this disclosure, its more prominent features will now be briefly discussed. After considering this discussion, and particularly after reading the section entitled “Detailed Description,” one will understand how the features of the embodiments described herein provide advantages over existing systems, devices and methods for left atrial appendage (LAA) occlusion.
The following disclosure describes non-limiting examples of some embodiments. For instance, other embodiments of the disclosed systems and methods may or may not include the features described herein. Moreover, disclosed advantages and benefits can apply only to certain embodiments and should not be used to limit the disclosure.
Devices and methods are described for occluding the LAA (LAA) to exclude the LAA from blood flow to prevent blood from clotting within the LAA and subsequently embolizing, particularly in patients with atrial fibrillation. An LAA occlusion device is delivered via transcatheter delivery into the LAA and anchored using a compliant frame and foam body. The device conforms to the oval shape of the LAA with superior sealing effect, does not require an excessive number of sizes and thus negates the need for extensive pre-procedure imaging, and can be delivered off-axis thereby allowing for simpler delivery procedure, among other advantages.
A foam body, which can be tubular in shape, and a compliant frame inside or within the foam body, are described that are collapsed for delivery and then expand in place within the LAA. The foam body may have a coating at least partially on the outer surface(s) of the foam body. The coating may be a layer of Polytetrafluoroethylene (PTFE). The device is anchored by structural anchors of the frame and/or by tissue ingrowth from the left atrium (LA) and LAA into the foam. Some embodiments are additionally or alternatively anchored by independent or integrated repositionable anchors, by barbs, and/or by distal anchoring elements. For example, anchors extending from a compliant frame are described which deploy through the compressible foam plug. In some embodiments, repositionable atraumatic anchor system embodiments are also disclosed which can be independent structures or integral to the foam plug and/or skin.
The foam body may be at least partially covered by a proximal end cover. The cover may be an expanded Polytetrafluoroethylene (ePTFE) cover. The cover provides several advantages, such as the following: sufficiently strong to enable handling of the plugs without tearing; allow for repositioning and retrieval of the plugs; provides a thromboresistant surface within the LA which will encourage formation of a neointima; assist in the creation of occlusion zones designed to encourage thromboresistance and endothelialization from the blood and adjacent tissue and anchoring zones designed to promote fast and tenacious tissue ingrowth into the compressible implant from the adjacent non-blood tissue; and can assist in closure at the ostium. The cover, e.g. a layer, jacket or skin, etc., can be independent or can be attached to the foam body, for example with sutures, adhesives, etc. In some embodiments, retrieval finials can be attached at one or more points to aid in retrieval of an embolized device and to increase radiopacity.
Some embodiments are tracked over a guidewire and have a guidewire lumen within the foam that is expandable, to allow for placement of the guidewire, and then is self-closing upon removal of the guidewire. Some embodiments do not require a guidewire lumen. Further, some embodiments may be multi-functional and include features for ablation, pressure-sensing, drug-elution, pacing, electrical isolation, etc.
In one aspect, a left atrial appendage occlusion device comprises a tubular foam body and a cover. The tubular foam body extends axially from a proximal end to a distal end. The cover includes at least a portion that covers the proximal end. The portion that covers the proximal end includes a series of openings therethrough. The foam and cover are configured to allow for a flow rate of water axially through the device of at least four liters per minute, with the water at about sixty-eight degrees Fahrenheit (F) and an upstream pressure of about twenty-five millimeters of Mercury (mmHg).
Various embodiments of the various aspects may be implemented. In some embodiments, the body may include a compressible side wall extending between the proximal end and the distal end and defines a central cavity. The left atrial appendage occlusion device may further comprise an expandable support coupled with the body and configured to compress the side wall against a wall of a left atrial appendage. The foam body may include a proximal face having an area, and the series of openings in the cover may collectively provide an open area that is at least five percent of the area of the proximal face. The open area may be at least ten percent of the area of the proximal face. The open area may be at least fifteen percent of the area of the proximal face. The flow of water may be along a flow axis, and the tubular foam body may extend axially along a device axis. The device axis may be angled with respect to the flow axis by at least thirty degrees. The device may be configured to allow for an off-axis flow rate of water through the device of at least four liters per minute, with the water at about sixty-eight degrees Fahrenheit (F) and an upstream pressure of about twenty-five millimeters of Mercury (mmHg). The off-axis flow may be at least thirty degrees from the axial flow.
In another aspect, a left atrial appendage occlusion device comprises a tubular foam body and a cover. The tubular foam body extends axially from a proximal end to a distal end. The cover includes at least a portion that covers the proximal end. The portion that covers the proximal end includes a series of openings therethrough. The foam body includes a proximal face at the proximal end having an area. The series of openings in the cover collectively provide an open area that is at least five percent of the area of the proximal face.
Various embodiments of the various aspects may be implemented. In some embodiments, the open area may be at least ten percent of the area of the proximal face. The open area may be at least fifteen percent of the area of the proximal face. The foam and cover may be configured to allow for a flow rate of water axially through the device of at least four liters per minute, with the water at about sixty-eight degrees Fahrenheit (F) and an upstream pressure of about twenty-five millimeters of Mercury (mmHg). The body may include a compressible side wall extending between the proximal end and the distal end and define a central cavity. The left atrial appendage occlusion device may further comprise an expandable support coupled with the body and configured to compress the side wall against a wall of a left atrial appendage.
In another aspect, a method of loading a left atrial appendage occlusion device into a delivery catheter is described. The method comprises positioning a proximal end of a loading body adjacent a distal end of the delivery catheter, the loading body having a sidewall defining a channel therethrough with a distal opening at a distal end that is larger than a proximal opening at the proximal end. The method further comprises retracting the left atrial appendage occlusion device proximally through the loading body to thereby radially compress the device, the device comprising a foam body, and receiving the device into the distal end of the delivery catheter.
Various embodiments of the various aspects may be implemented. In some embodiments, the loading body may comprise a frustoconical portion. The loading body may define a central longitudinal axis, and the sidewall may extend at an angle of at least five degrees with respect to the longitudinal axis. The sidewall may extend at an angle of at least ten degrees with respect to the longitudinal axis. The sidewall may extend at an angle at least fifteen degrees with respect to the longitudinal axis The sidewall may define a total angle of at least ten degrees, at least twenty degrees, or at least thirty degrees. The advancing step may comprise pulling a tether proximally through the delivery catheter. The device may be radially compressed within a delivery catheter having an outer diameter of no more than fifteen French. An inner surface of the loading body may be substantially smooth. The method may comprise radially compressing the device to a radial compressed width within the delivery catheter that is no more than twenty percent of a radially uncompressed width of the device. The radial compressed width within the delivery catheter may be no more than fifteen percent of the radially uncompressed width of the device.
In another aspect, a left atrial appendage occlusion device comprises a foam body, an expandable support and at least one anchor. The foam body has a tubular sidewall with a radial uncompressed thickness. The expandable support is coupled with the body. The at least one anchor is coupled with the support and extends through the sidewall when the foam of the sidewall is compressed, and where the at least one anchor has a radial height no greater than the radial uncompressed thickness of the sidewall.
Various embodiments of the various aspects may be implemented. In some embodiments, the device may define a central axis and the at least one anchor may be angled relative to the central axis. The at least one anchor may extend radially outward in a proximal direction at an angle of at least twenty degrees relative to a portion of the central axis that extends proximally to the device. The angle may be at least thirty degrees. The at least one anchor may extend through a radially-compressed portion of the sidewall having a radial thickness less than the radial uncompressed thickness. The left atrial appendage occlusion device may further comprise an attachment that connects the support to the sidewall and radially compresses the sidewall at the radially-compressed portion. The device may further comprise a proximal cover that covers at least a portion of a proximal face of the foam body.
In another aspect, a left atrial appendage occlusion device comprises a foam body, an expandable support and at least one anchor. The foam body has a tubular sidewall comprising at least one first portion having a first radial thickness and at least one second portion having a second radial thickness that is less than the first radial thickness. The expandable support is coupled with the body. The least one anchor is coupled with the support and extends at least partially through the at least one second portion of the sidewall.
Various embodiments of the various aspects may be implemented. In some embodiments, the device may define a central axis and the at least one anchor may be angled relative to the central axis. The at least one anchor may extend radially outward in a proximal direction at an angle of at least twenty degrees relative to the central axis. The angle may be at least thirty degrees. The at least one anchor may extend through the at least one second portion of the sidewall such that a portion of the at least one anchor extends outwardly beyond an outer surface of the at least one second portion of the sidewall. The at least one anchor may have a length equal to the radial uncompressed thickness of the tubular sidewall. The support may comprise a tubular frame portion configured to expand radially outward to compress the sidewall against a wall of a left atrial appendage after implantation of the device. The device may further comprise a proximal cover that covers at least a portion of a proximal face of the foam body.
In another aspect, a left atrial appendage occlusion device comprises a tubular foam body and an expandable support coupled with the body. The device is configured to insert into a non-cylindrical opening of a test body having a non-cylindrical profile, expand radially within the non-cylindrical opening, and conform to the non-cylindrical profile at least at the opening of the test body.
Various embodiments of the various aspects may be implemented. In some embodiments, the device may be configured to conform to the non-cylindrical profile at least at the opening of the test body and leave no radial gaps between the device and the test body opening that are greater than five millimeters. The device may leave no radial gaps that are more than four, three, two and/or one millimeter. The device may be configured to insert into a non-cylindrical opening of a test body having a non-cylindrical profile with a size and shape substantially similar to that of a native left atrial appendage. The device may be configured to insert into a non-cylindrical opening of a test body having a radial stiffness substantially similar to that of a native left atrial appendage and to assume a non-cylindrical profile at least at the opening of the test body after a period of at least thirty days, of at least sixty days, and/or of at least one hundred twenty days. The device may further comprise at least one anchor coupled with the frame and extending at least partially into the tubular foam body.
In some embodiments, the foam body may include a compressible side wall extending between a proximal end and a distal end and define a central cavity. The left atrial appendage occlusion device may further comprise an expandable support coupled with the foam body and configured to compress the side wall against an inner surface of the test body.
In another aspect, a left atrial appendage occlusion device comprises a tubular foam body and an expandable support coupled with the body. The device has a radial uncompressed width. The device is configured to compress radially to a radial compressed width that is no more than fifty percent of the radial uncompressed width.
Various embodiments of the various aspects may be implemented. In some embodiments, the radial compressed width may be no more than forty percent of the radial uncompressed width. The tubular foam body may extend along a longitudinal axis, and the radial uncompressed width may extend along a diameter of the foam body that is perpendicular to the longitudinal axis.
In another aspect, a left atrial appendage occlusion device comprises a tubular foam body and an expandable support coupled with the body. The device extends axially from a proximal end to a distal end, and the proximal end has a radial uncompressed width. The distal end is configured to compress radially to a radial compressed width that is no more than fifty percent of the radial uncompressed width of the proximal end.
Various embodiments of the various aspects may be implemented. In some embodiments, the distal end may be configured to compress radially to a radial compressed width that is no more than forty percent of the radial uncompressed width of the proximal end. The radial compressed width may be no more than thirty percent, no more than twenty percent, no more than ten percent, and/or no more than five percent, of the radial uncompressed width of the proximal end.
In another aspect, a left atrial appendage occlusion device comprises a tubular foam body and an expandable support coupled with the body. The device has an axial uncompressed length. The device is configured to compress axially to an axial compressed length that is no more than fifty percent of the axial uncompressed length.
Various embodiments of the various aspects may be implemented. In some embodiments, the axial compressed length may be no more than forty percent of the axial uncompressed length. The axial uncompressed length may extend from a proximal end of the foam body to a distal end of the foam body.
In another aspect, a left atrial appendage occlusion device is described. The device comprises a conformable, tubular foam body, a compressible side wall and an expandable support. The conformable, tubular foam body has a closed proximal end and a distal end. The compressible side wall extends between the proximal end and the distal end, and defines a central cavity. The expandable support is within the body and configured to compress the side wall against a wall of a left atrial appendage.
In some embodiments, the side wall may have an uncompressed thickness of at least about 0.5 mm. The compressible side wall may have an uncompressed thickness of at least about 1.5 mm. The compressible side wall may have an uncompressed thickness of about 2.5 mm. The compressible side wall may extend in a distal direction beyond a distal end of the support by at least about 2 mm in an unconstrained, expanded state. The compressible side wall may extend in a distal direction beyond a distal end of the support by about 5 mm in an unconstrained, expanded state. The compressible side wall may comprise a foam having a plurality of interconnected reticulations and voids, and further comprising a PTFE coating on at least some of the interconnected reticulations. The closed proximal end may comprise a foam end wall. The foam end wall may further comprise a cover. The cover may comprise ePTFE. The expandable support may be self-expandable. The expandable support may be in the central cavity. The tubular foam body may be substantially cylindrical in an unconstrained, expanded state.
In another aspect, a self-expandable, atraumatic occlusion device is described. The device is configured to conform to the side wall of a left atrial appendage. The device comprises a compressible open cell foam body, a self-expandable support and a proximal end wall. The compressible open cell foam body has a tubular foam side wall and a central cavity. The expandable support is within the cavity. The proximal end wall is on the foam body. The proximal end wall is positioned proximally of the proximal end of the support, and the foam side wall extends distally beyond the distal end of the support to form a distal, atraumatic bumper for preventing contact between the support and a wall of the left atrial appendage in an implantation in which a central longitudinal axis of the occlusion device is non-parallel to a primary longitudinal axis of the left atrial appendage.
In another aspect, a left atrial appendage occlusion device is described. The device comprises an expandable tubular foam cup and an expandable frame. The expandable tubular foam cup has a proximal end, a distal end, a tubular side wall and a proximal end wall. The side wall has a thickness of at least about 1.0 mm and a porosity of at least about 85% open void content. The expandable frame is configured to press the side wall into conforming contact with a wall of the left atrial appendage.
In some embodiments, the tubular side wall may have a thickness of at least about 2 mm. The tubular side wall may have a void content of at least about 90%. The tubular side wall may have an average pore size of at least about 100 microns. The tubular side wall may have an average pore size of at least about 200 microns. The tubular side wall may be provided with a thromboresistant coating. The thromboresistant coating may comprise PTFE. The proximal end wall may be provided with a thromboresistant cover. The frame may further comprise at least three recapture struts inclining radially inwardly in the proximal direction to a hub. The frame may comprise a plurality of axially extending side wall struts, with adjacent pairs of side wall struts joined at an apex. The frame may comprise at least six proximally facing apexes and at least six distally facing apexes. Each recapture strut may be joined to a unique proximally facing apex on the frame. The recapture struts may be integrally formed with the frame. The device may further comprise a lumen through the hub. The device may further comprise anchors to secure the device to tissue. The anchors may be flexible anchors configured to extend through the foam side wall at an inclined angle.
In another aspect, a conformable LAA occlusion device is described. The device comprises a compressible tubular foam wall. The wall comprises a reticulated, cross linked matrix having at least about 90% void content, an average cell size within the range of from about 250-500 microns, a wall thickness of at least about 2 mm and a compressive strength of at least about 1 psi. In some embodiments, the compressive strength is within a range of from about 1 psi to about 2 psi. In some embodiments, the device may have an expandable support configured to compress the side wall against a wall of a left atrial appendage.
In another aspect, a LAA occlusion device is described. The device includes an open cell foam body and an internal locking system. The body has a proximal end, a distal end and an outer skin. The proximal end is configured to face a left atrium and the distal end is configured to face the LAA following implantation in the LAA. The body can be compressed for delivery within a delivery catheter and can self-expand when removed from the delivery catheter. The internal locking system is coupled with the body and comprises at least one deployable tissue anchor. The deployable anchor is configured to deploy from a constrained configuration within the body to a deployed configuration where a tissue engaging segment of the anchor extends outside the body to secure the body within the LAA. The deployable anchor is configured to deploy to the deployed configuration after the body expands within the LAA. The deployable anchor may be retractable from the deployed configuration to a retracted configuration within the body.
In some embodiments, the internal locking system further comprises a plurality of the deployable anchors rotatably coupled with the body, wherein the plurality of anchors are configured to rotate to the deployed and retracted configurations. The internal locking system may comprise four of the deployable anchors. In some embodiments, the body further comprises a plurality of axially extending slots corresponding to the plurality of anchors, wherein each of the plurality of anchors is configured to deploy and retract through the corresponding axial slot.
In some embodiments, the internal locking system further comprises a restraint that restrains the anchor in the constrained configuration, and the anchor is deployed from the constrained configuration to the deployed configuration by removing the restraint from the anchor. The restraint may be a sheath that restrains the anchor in the constrained configuration by covering the anchor, wherein the anchor is deployed from the constrained configuration to the deployed configuration by removing the sheath from covering the anchor. The restraint may be a lasso that restrains the anchor in the constrained configuration by surrounding the anchor, and the anchor is deployed from the constrained configuration to the deployed configuration by removing the lasso from surrounding the anchor.
In some embodiments, the internal locking system further comprises a moveable mount coupled with an end of the anchor, and the anchor is deployed from the constrained configuration to the deployed configuration by axially moving the mount.
In some embodiments, the internal locking system further comprises a constraint configured to move over the anchor to cause the anchor to retract. The constraint may be a ring configured to slide over the anchor to cause the anchor to retract.
In some embodiments, the skin comprises ePTFE.
In some embodiments, the device further comprises at least one tissue ingrowth surface on a sidewall of the body.
In some embodiments, the device further comprises a plurality of openings in the skin to permit tissue ingrowth into the open cell foam body. The plurality of openings of the skin may be located in an anchoring region of the device located at least between the proximal and distal ends of the device, and the device may further comprise an occlusion region located at the proximal end of the device and configured to encourage thromboresistance and endothelialization from the blood and adjacent tissue.
In another aspect, a LAA closure system is described. The system comprises a delivery catheter and a LAA occlusion device. The delivery catheter comprises an elongate flexible tubular body, having a proximal end and a distal end and at least one lumen extending therethrough. The LAA occlusion device is configured to be compressed within the delivery catheter and to self-expand upon deployment from the delivery catheter. The device comprises a self-expandable open cell foam body coupled with an internal locking system. The internal locking system comprises a deployable anchor configured to deploy from a constrained configuration to a deployed configuration after the body expands within the LAA and is configured to retract from the deployed configuration to a retracted position within the body.
In some embodiments, the system further comprises an axially movable deployment control extending through a lumen of the body, for deploying the deployable anchor. The system may further comprise an axially movable deployment control extending through a lumen of the body, for deploying the foam body from the distal end of the closure system. The internal locking system may further comprise a restraint that restrains the anchor in the constrained configuration, and the anchor is actively deployed from the constrained configuration to the deployed configuration by removing the restraint from the anchor using an axially movable deployment control extending through a lumen of the body. The internal locking system may further comprise a moveable mount coupled with an end of the anchor, and the anchor is actively deployed from the constrained configuration to the deployed configuration by axially moving the mount using an axially movable deployment control extending through a lumen of the body.
In another aspect, a method of excluding a LAA is described. The method comprises advancing a guidewire into the LAA, advancing a distal end of a delivery catheter over the guidewire and into the LAA, and deploying a LAA occlusion device from the distal end of the delivery catheter. The device comprises an expandable foam body coupled with an internal locking system having a deployable anchor, and the body expands within the LAA upon deploying from the distal end of the delivery catheter. The method further comprises actively deploying the deployable anchor after the body expands within the LAA. The deployable anchor is configured to retract from the deployed configuration to a retracted position within the body. In some embodiments, the method further comprises retracting the deployable anchor from the deployed configuration to the retracted position.
In another aspect, a LAA occlusion device is described. The device comprises an expandable foam body and an internal locking system. The body can be compressed for delivery within a delivery catheter and can self-expand when removed from the delivery catheter. The internal locking system is coupled with the body and comprises a deployable anchor configured to deploy from a constrained configuration within the body to a deployed configuration where the anchor extends outside the body to secure the body within the LAA. The body is configured to expand upon removal from the delivery catheter, and the deployable anchor is configured to deploy to the deployed configuration after the body expands.
In another aspect, a LAA occlusion device is described. The device comprises an expandable foam body and an internal locking system. The body can be compressed for delivery within a delivery catheter and can self-expand when removed from the delivery catheter. The internal locking system is coupled with the body and comprises a deployable anchor configured to deploy from a constrained configuration within the body to a deployed configuration where the anchor extends outside the body to secure the body within the LAA. The deployable anchor is configured to retract from the deployed configuration to a retracted configuration within the body such that the body can be repositioned within the LAA.
In another aspect, a LAA occlusion device is described. The device comprises an expandable tubular frame, an expandable tubular foam layer and a tissue scaffold. The expandable tubular frame has a proximal end, a distal end and a central lumen. The expandable tubular foam layer is carried by the frame and has a thickness of at least about 0.5 mm. The emboli retention layer is carried by the frame and encloses the lumen at the proximal end.
In some embodiments, the foam layer may have a thickness of at least about 1 mm. The foam layer may have a thickness of at least about 2.5 mm. The foam layer may have a void content of at least about 80%. The foam layer may have a void content of at least about 90%. The foam layer may have an average pore size of at least about 100 microns. The foam layer may have an average pore size of at least about 200 microns. The foam layer may extend across the proximal end of the frame to form the tissue scaffold. The tissue scaffold may be provided with a thromboresistant coating. The tissue scaffold may be provided with a thromboresistant layer. The thromboresistant coating or layer may comprise PTFE. The thromboresistant coating or layer may comprise ePTFE. The frame may further comprise at least three recapture struts inclining radially inwardly in the proximal direction to a hub.
In some embodiments, the foam layer extends across the proximal end of the frame to form the tissue scaffold, and the frame may comprise a plurality of axially extending side wall struts, with adjacent pairs of side wall struts joined at an apex. The device may comprise at least six proximally facing apexes and at least six distally facing apexes. The device may comprise at least three recapture struts joined at a proximal hub, where each recapture strut has a distal end joined to the frame. Each recapture strut may be joined to a unique proximally facing apex on the frame. The recapture struts may be integrally formed with the frame. The device may further comprise a lumen through the hub.
In some embodiments, the device may comprise anchors to secure the device to tissue. The anchors may be static anchors that are configured to deploy upon deployment of the device from a delivery catheter. The anchors may be constrained anchors that are configured to be controllably released into a deployed configuration after expansion of the foam. The anchors may be dynamic anchors that are configured to be deployed from a contracted configuration to a deployed configuration and are further configured to be retracted from the deployed configuration back to a retracted configuration. The anchors may be further configured to be retracted from the deployed configuration back to the contracted configuration.
The foregoing and other features of the present disclosure will become more fully apparent from the following description and appended claims, taken in conjunction with the accompanying drawings. Understanding that these drawings depict only several embodiments in accordance with the disclosure and are not to be considered limiting of its scope, the disclosure will be described with additional specificity and detail through use of the accompanying drawings. In the following detailed description, reference is made to the accompanying drawings, which form a part hereof. In the drawings, similar symbols typically identify similar components, unless context dictates otherwise. The illustrative embodiments described in the detailed description, drawings, and claims are not meant to be limiting. Other embodiments may be utilized, and other changes may be made, without departing from the spirit or scope of the subject matter presented here. It will be readily understood that the aspects of the present disclosure, as generally described herein, and illustrated in the drawing, can be arranged, substituted, combined, and designed in a wide variety of different configurations, all of which are explicitly contemplated and make part of this disclosure.
While the above-identified drawings set forth presently disclosed embodiments, other embodiments are also contemplated, as noted in the discussion. This disclosure presents illustrative embodiments by way of representation and not limitation. Numerous other modifications and embodiments can be devised by those skilled in the art which fall within the scope and spirit of the principles of the presently disclosed embodiments.
The following detailed description is directed to certain specific embodiments of the development. In this description, reference is made to the drawings wherein like parts or steps may be designated with like numerals throughout for clarity. Reference in this specification to “one embodiment,” “an embodiment,” or “in some embodiments” means that a particular feature, structure, or characteristic described in connection with the embodiment is included in at least one embodiment of the invention. The appearances of the phrases “one embodiment,” “an embodiment,” or “in some embodiments” in various places in the specification are not necessarily all referring to the same embodiment, nor are separate or alternative embodiments necessarily mutually exclusive of other embodiments. Moreover, various features are described which may be exhibited by some embodiments and not by others. Similarly, various requirements are described which may be requirements for some embodiments but may not be requirements for other embodiments. Reference will now be made in detail to embodiments of the invention, examples of which are illustrated in the accompanying drawings. Wherever possible, the same reference numbers will be used throughout the drawings to refer to the same or like parts.
The devices and related methods are described herein in connection use in occluding, i.e. excluding, a LAA (LAA). The various figures show various embodiments of LAA occlusion devices, systems and methods for delivery of the LAA occlusion devices, and/or methods of using the device to occlude a LAA. The various systems, devices and methods described herein may include the same or similar features and/or functionalities as other LAA occlusion systems, devices and methods as described, for example, in U.S. application Ser. No. 14/203,187 entitled “DEVICES AND METHODS FOR EXCLUDING THE LAA” and filed on Mar. 10, 2014, and/or as described in U.S. Provisional Application No. 62/240,124 entitled “DEVICES AND METHODS FOR EXCLUDING THE LAA” and filed on Oct. 12, 2015, the entire disclosure of each of which is incorporated herein by reference for all purposes and forms a part of this specification.
Some embodiments of an LAA occlusion device 3000 include a foam body 3002, a deployable and compliant frame 3040, and a proximal cover 3100, as primarily shown and described for example with respect to
The heart 100 is shown in
One embodiment of an LAA occlusion device is shown in
The plug 204 may be made of polyurethane, polyolefin, PVA, collagen foams or blends thereof. One suitable material is a polycarbonate-polyurethane urea foam with a pore size of 100 μm-250 μm or 250 μm-500 μm and 90-95% void content. The foam could be non-degradable or use a degradable material such as PLA, PGA, PCL, PHA, and/or collagen. If degradable, the tissue from the LAA will grow into the foam plug and replace the foam over time. The plug 204 may be cylindrical in shape in an unconstrained expansion, but it may also be conical for example with its distal end smaller than the proximal end or reversed. It could also be oval in cross section to better match the opening of the LAA.
The foam plug 204 is oversized radially in an unconstrained expansion to fit snuggly into the LAA and may be 5-50 mm in diameter depending on the diameter of the target LAA. In a free, unconstrained state, the axial length “L” of the plug is less than its outer diameter “D” such that the L/D ratio is less than 1.0. In some embodiments, this ratio may be greater than 1.0. The compliance of the foam material is designed such that it pushes on the walls of the LAA with sufficient force to maintain the plug 204 in place but without overly stretching the LAA wall. The foam and/or skin also conforms to the irregular surfaces of the LAA as it expands, to provide a complementary surface structure to the native LAA wall to further enhance anchoring and promote sealing. Thus, the expandable foam implant described herein conforms to the native configuration of the LAA. In one embodiment, the structure of the foam may be fabricated such that squeezing axially on the opposing ends of the foam causes the foam to increase in diameter.
The ePTFE or foam material may be provided with one or two or more radiopaque markers such as radiopaque threads 210 or be filled with or impregnated with a radiopaque filler such as barium sulfate, bismuth subcarbonate, or tungsten which permit the operator to see under x-ray the plug for proper positioning in the anatomy. An x-ray image is shown in
An outer ePTFE layer may be formed from a tube with a diameter about the same diameter of the foam plug and a wall thickness between about 0.0001″ and about 0.001″ thick and serves to allow one to collapse and pull on the plug 204 without tearing the foam material. The ePTFE material also serves as the blood contacting surface facing the LA 206 and has pores or nodes such that blood components coagulate on the surface and an intimal or neointimal covering of tissue grows across it and anchors tightly to the material. Pore sizes within the range of from about 4μ to about 110μ, ideally 5-35μ are useful for formation and adherence of a neointima.
The outer covering 206 may be constructed of materials other than ePTFE such as woven fabrics, meshes or perforated films made of FEP, polypropylene, polyethylene, polyester or nylon. The covering 206 should have a low compliance (non-elastic), at least longitudinally, be sufficiently strong as to permit removal of the plug, a low coefficient of friction, and be thromboresistant. The outer covering 206 serves as a matrix to permit plug removal as most foams are not sufficiently strong to resist tearing when pulled. The plug 204 can also be coated with or contain materials, such as PTFE. Such materials may enhance the plug's 204 ultrasonic echogenic profile, thromboresistance, and/or lubricity. The plug 204 can also be coated with or contain materials to facilitate echocardiographic visualization, promote cellular ingrowth and coverage.
The outer covering 206 has holes in it to permit contact of the LAA tissue with the foam plug 204 to encourage ingrowth of tissue into the foam plug pores and/or allow blood flow therethrough. These holes may be 1 to 5 mm in diameter or may also be oval with their long axis aligned with the axis of the foam plug, the length of which may be 80% of the length of the foam plug and the width may be 1-5 mm. The holes may be as large as possible such that the outer covering maintains sufficient strength to transmit the tensile forces required for removal. The holes may be preferentially placed along the device. In one embodiment, holes are placed distally to enhance tissue ingrowth from the LAA wall.
In one implementation, the implant is provided with proximal and/or distal end caps of ePTFE, joined together by two or three or four or more axially extending strips of ePTFE. The axially extending strips are spaced apart from each other circumferentially, to provide at least two or three or four or more laterally facing windows through which the open cell foam body will be in direct contact with the tissue wall of the LAA. This outer covering could be a mesh or netting as well. As shown in
The implantable plug 204 or devices 10, 1020, 3000 (as described below) may be anchored and secured in place in the LAA by tissue ingrowth and/or with additional anchoring features. In some embodiments, the plug 204 or devices 10, 1020, 3000 may be anchored by tissue ingrowth alone.
In some embodiments, other anchoring means may be implemented. One means of adhering the foam plug in place within the LAA is to use an adhesive, such as a low viscosity cyanoacrylate (1-200 cps). The adhesive is injected into place along the sidewall near the distal end of the foam plug 208. Holes in the ePTFE covering permit the adhesive to interact between the foam plug 204 and the LAA wall 200. Injection of the adhesive may be accomplished with several means, one of which is to inject through the catheter into the center lumen 212. Passages 214 serve to guide the adhesive to the correct location. The distal end of the foam plug may be restricted at that time to prevent the adhesive from exiting the distal crimp 216. Alternatively,
Other one part adhesives including aqueous cross linking adhesives, polyurethane, PEG, PGA, PLA, polycaprolactone or a lycine-derived urethane may be used. In addition, these adhesives may be made in two components such that one component is adherent to the foam and the second injected in vivo. Also, these two component adhesives may be injected simultaneously to mix in vivo to prevent fouling of injection tubes.
An alternative anchoring means for plug 400 or device 3000, etc. is one or two or more distal anchors as shown in
Additional means of anchoring include the use of a plurality of hooks or barbs or graspers to grab the distal wall and baskets, malecots, distal foam plugs and Nitinol wire birds nests that open within the LAA and push outward on the wall or engage the protrusions of the LAA. It may be desirable to place the plug then engage the anchor as a secondary step. One such embodiment could include a multitude of nitinol wires with a ball or catch welded proximal to the anchor tip. These could be gathered with the delivery catheter then released when the ideal plug position has been confirmed.
A cross section of one embodiment is shown in
Referring to
An embodiment of a technique for placement of an LAA occlusion device is shown in
Once the Mullins sheath and dilator are in the SVC, the guidewire is removed and a trans-septal needle is placed through the dilator. The needle contains a stylette to prevent skiving off of polymeric material from the dilator lumen as it traverses to the tip. Once the needle is near the dilator tip, the stylette is removed and the needle is connected to a manifold and flushed. The Mullins sheath/dilator set and the needle (positioned within the dilator tip) are retracted into the SVC toward the RA as a unit. As the system is withdrawn down the wall of the SVC into the RA and positioned in the fossa ovale, the preferred puncture location.
Once proper position in the fossa ovale is observed, the needle is advanced across the fossa ovale into the LA. Successful trans-septal puncture can be confirmed by echo, pressure measurement, O2 saturation and contrast injection. Once the needle position is confirmed to be positioned in the LA, the sheath and dilator can be advanced over it into the LA. In some cases, the user will first pass a guidewire through the needle into the LA and into an upper pulmonary vein (typically the left) prior to crossing. Alternative options include the use of radiofrequency trans-septal needles, which are useful for crossing very thick or hypertrophic septa, or the use of a safety wire placed through the needle and utilized for the initial puncture.
Referring to
The foam plug 1204 is then pushed through the guide catheter 1200 with pusher 1202 and is shown exiting the guide catheter 1300 slowly in
An alternative to pushing the plug through the entire length of the guide catheter is that the plug 1204 may be initially located at the distal end of the guide catheter 1200 as shown in
For alternative anchors, they may be deployed, the shafts disconnected and removed. Disconnection mechanisms may be any of several types, such as threaded, electrolytic detachment, or others known in the art. In some embodiments, a suture attachment may be implemented, for example as described with respect to
As shown in
The stent 1602 may be made of Nitinol to enable it to pack into a 10, 12, 14, 16, 18 or 20F delivery catheter and expand to its desired diameter. The stent 1602 may be braided, laser cut or wire formed. Any of a variety of stent wall patterns may be utilized, depending upon the desired performance. The stent 1602 may be a balloon expandable stent, or self-expandable stent. In the illustrated embodiment, the self-expandable stent 1602 comprises a plurality of proximal apexes 1608 and distal apexes 1610 connected by a plurality of zig zag struts 1612. A hole 1606 allows passage of the guidewire for delivery. This design may be advantageous in that the expansion force exerted by the plug on the LAA can be controlled separately from the foam characteristics. Also, it may be easier to pack this concept into a smaller geometry. For example, the plug can be packed into a smaller geometry by reducing the amount of foam that must be compressed into the delivery catheter while maintaining sufficient dilation force.
Alternatively, the foam plug may be constructed of two foams. One denser core to provide force, for example radial force, and an outer softer foam to engage the tissue irregularities. The softer foam could also be located on the proximal and/or distal ends to facilitate retrieval.
Another means of adding stiffness to the foam plug is shown in
Instead of wires as shown in
Another LAA plug is shown in
Rather than using a foam, a low porosity outer bag without perforations could be placed in the LAA and then filled with a substance to provide the radial expansion. This substance may be a hydrogel, cellulose or polyvinylacetate.
Rather than requiring the use of a separate dilation device to cross the septum, the distal crimp element 1902 may be formed in a tapered manner such that it extends from the distal end of the catheter 1200 and serves as a dilating tip to dilate the opening in the septum as the catheter is advanced. See
An alternative plug design uses a foam such as cellulose sponge material that is compacted and dehydrated such that it can be packed into the guide catheter. This foam material 2202 may be packed into the guide catheter as shown in
It may be advantageous to use small barbs 2302 in
One means of removing a device that is not functioning properly is to releasably attach a retrieval suture 2400 to the implant, such as to the proximal cap 2402 which also passes proximally throughout the entire length of the guide catheter 2404 in
Deployment of the occlusion device has been discussed primarily in the context of a transvascular access. However, the implants may alternatively be deployed via direct surgical access, or various minimally invasive access pathways (e.g. jugular vein). For example, the area overlying the xiphoid and adjacent costal cartilage may be prepared and draped using standard techniques. A local anesthetic may be administered and skin incision may be made, typically about 2 cm in length. The percutaneous penetration passes beneath the costal cartilage, and a sheath may be introduced into the pericardial space. The pericardial space may be irrigated with saline, preferably with a saline-lidocaine solution to provide additional anesthesia and reduce the risk of irritating the heart. The occlusion device may thereafter be introduced through the sheath, and through an access pathway created through the wall of the LAA. Closure of the wall and access pathway may thereafter be accomplished using techniques understood in the art.
Depending upon the desired clinical performance, any of the LAA occlusion devices described herein may be provided with a drug or other bioactive agent, which may be injected via the deployment catheter, or impregnated within the open cell foam or coated on the implant. The bioactive agent may be eluted or otherwise released from the implant into the adjacent tissue over a delivery time period appropriate for the particular agent as is understood in the art. Useful bioactive agents can include those that modulate thrombosis, those that encourage cellular ingrowth, throughgrowth, and endothelialization, and potentially those that resist infection. For example, agents that may promote endothelial, smooth muscle, fibroblast, and/or other cellular growth into the implant including collagen (Type I or II), heparin, a combination of collagen and heparin, extracellular matrix (ECM), fibronectin, laminin, vitronectin, peptides or other biological molecules that serve as chemoattractants, molecules MCP-1, VEGF, FGF-2 and TGF-beta, recombinant human growth factors, and/or plasma treatment with various gases.
Anti-thrombotics can typically be separated into anti-coagulants and antiplatelet agents. Anti-Coagulants include inhibitors of factor(s) within the coagulation cascade an include heparin, heparin fragments and fractions as well as inhibitors of thrombin including hirudin, hirudin derivatives, dabigatran, argatroban and bivalrudin and Factor X inhibitors such as low molecular weight heparin, rivaroxaban, apixaban.
Antiplatelet agents include GP 2b/3a inhibitors such as epifibitide, and abciximab, ADP Receptor agonists (P2/Y12) including thienopyridines such as ticlopidine, clopidogrel, prasugrel and tacagrelor and aspirin. Other agents include lytic agents, including urokinase and streptokinase, their homologs, analogs, fragments, derivatives and pharmaceutical salts thereof and prostaglandin inhibitors.
Antibiotic agents can include, but are not limited to penicillins, cephalosportins, vancomycins, aminoglycosides, quinolonges, polymyxins, erythromycins, tetracyclines, chloraphenicols, clindamycins, lincomycins, sulfonamides, their homologs, analogs, derivatives, pharmaceutical salts and combinations thereof.
Biologic agents as outlined above maybe be added to the implant 204 and may be injected through the delivery catheter into the space between the proximal cap 206 and the foam plug 204. This may serve as a reservoir to minimize thrombus formation during the initial implantation and reduce the need for systemic anticoagulation following device implantation.
An electronic pressure sensor may be embedded into the proximal end of the foam plug which may be used to transmit LA pressure to a remote receiver outside the body for the monitoring of LA pressure which is useful to monitor cardiac function. In addition, a cardiac pacer or defibrillator may be embedded into the foam plug and attached electrically to the distal anchor. A drug delivery reservoir may be embedded with connection to the LA for controlled delivery of biologic agents as outlined above.
Another means of anchoring is shown in
Another means of anchoring the distal anchor element to the foam is shown in
The body 15 of the device 10 shown in
The internal locking system 101 may be controllably deployed a period of time after the body 15 expands. For instance, the location, orientation, etc. of the device 10 may be verified with various imaging techniques such as by fluoroscopy with injection of contrast media via the central lumen before the internal locking system 101 is deployed and the anchors secure the device 10 within the LAA. In some embodiments, even after deployment of the internal locking system 101 and anchors thereof, the anchors may be retracted to a position within the body 15 for repositioning, and/or retrieval of the device 10 from, within the LAA.
The device 10 may have any or all of the same or similar features and/or functionalities as the other plugs described herein, for example the plug 204, etc. For example, the device 10 is at least partially encapsulated within the skin 20. In some embodiments, the skin 20 may cover the proximal end of the body 15. The skin 20 may be a thin, strong outer layer. The skin 20 may be a thin, encapsulating layer. The skin 20 may be fabricated from ePTFE (expanded polytetrafluoroethylene), polyolefin, polyester, other suitable materials, or combinations thereof. In some embodiments, the skin 20 may be fabricated from bioabsorbable materials, for example polylactic acid (PLA), Polyglycolic acid (PGA), ploycaprolactone (PCL), PHA, collagen, other suitable bioabsorbable materials, or combinations thereof. The skin 20 can be oriented or otherwise modified to be elastomeric in at least one direction, such as radially.
The body 15 may be made of polyurethane, polyolefin, PVA, collagen foams or blends thereof. One suitable material is a polycarbonate-polyurethane urea foam with a pore size of 100-250 um and 90-95% void content. The body 15 may be non-degradable or use a degradable material such as PLA, PGA, PCL, PHA, and/or collagen. If degradable, the tissue from the LAA will grow into the foam body 15 and replace the foam over time. The body 15 may be cylindrical in shape in an unconstrained expansion, but it may also be conical with its distal end smaller than the proximal end, or vice versa. The body 15 may also be oval in cross section to better match the opening of the LAA.
The device 10 is oversized radially in an unconstrained expansion to fit snuggly into the LAA. The device 10 may be 5-50 millimeters (mm) and generally at least about 10 mm or 15 mm in diameter in its unconstrained configuration, for example depending on the diameter of the target LAA. The length “L” of the device 10 may be less than, similar to or greater than its diameter “D” such that the L/D ratio is less than 1.0, about or greater than about 1.0, greater than about 1.5, or greater than about 2.0. The L/D ratio may be greater than 1.0 to maximize its stability. However, in some embodiments, the L/D ratio may be less than 1.0, for example, from about 0.2 to about 0.9, or from about 0.3 to about 0.8, or from about 0.4 to about 0.6. The compliance of the material of the device 10 is designed such that it pushes on the walls of the LAA with sufficient force to maintain the plug in place but without overly stretching the LAA wall. The foam body 15 and/or skin 20 also conforms to the irregular surfaces of the LAA as it expands, to provide a complementary surface structure to the native LAA wall to further enhance anchoring and promote sealing. Thus, the expandable foam body 15 conforms to the native irregular configuration of the LAA. In some embodiments, the structure of the foam body 15 may be fabricated such that axial compression on the opposing ends of the body 15 such as by proximal retraction of a pull wire or inner concentric tube causes the foam to increase in diameter.
The body 15 and/or skin 20, for example the foam material and/or ePTFE, may be provided with one, two or more radiopaque markers, such as radiopaque threads 210 (see
The skin 20, such as an outer ePTFE layer, may have a thickness between about 0.0001 inches and about 0.0030 inches. In some embodiments, the thickness of the skin 20 may be between about 0.0003 inches and about 0.0020 inches. In some embodiments, the thickness of the skin 20 may be between about 0.0005 inches and about 0.0015 inches. The thickness of the skin 20 may be uniform, for example the same or approximately the same no matter where the thickness is measured. In some embodiments, the thickness of the skin 20 may be non-uniform, for example the thickness may be different in different portions of the skin 20.
The skin 20, such as an outer ePTFE layer, may also serve as the blood contacting surface on the proximal end of the device 10 facing the LA. The skin 20 may have pores or nodes such that blood components coagulate on the surface and an intimal or neointimal covering of tissue grows across it and anchors tightly to the skin material. Pore sizes may be within the range of from about 4μ to about 110μ. In some embodiments, the pore sizes are within the range of from about 30μ to about 90μ. In some embodiments, the pore sizes are within the range of from about 30μ to about 60μ. Such ranges of pore sizes are useful for formation and adherence of a neointima. In some embodiments, the skin 20, such as an outer ePTFE layer, may be formed from a tube with a diameter about the same diameter of the foam body 15. and allows one to collapse and pull on the body 15 without tearing the foam material.
The skin 20 may be constructed of materials other than ePTFE such as woven fabrics, meshes or perforated films made of FEP, polypropylene, polyethylene, polyester or nylon. The skin 20 may have a low compliance (e.g. non-elastic), for instance a low compliance longitudinally, may be sufficiently strong as to permit removal of the plug, may have a low coefficient of friction, and/or may be thromboresistant. The skin 20 serves as a matrix to permit plug removal as most foams are not sufficiently strong to resist tearing when pulled. The body 15 can also be coated with or contain materials to enhance its ultrasonic echogenic profile, thromboresistance, lubricity, and/or to facilitate echocardiographic visualization, promote cellular ingrowth and coverage.
The skin 20 may include holes to permit contact of the LAA tissue with the foam body 15. Exposure of the foam body 15 to the LAA or other tissue has benefits for example encouraging ingrowth of tissue into the foam plug pores and/or increasing friction to hold the body 15 in place. These holes may be 1 to 5 mm in diameter or may also be oval with their long axis aligned with the axis of the foam plug, the length of which may be 80% of the length of the foam plug and the width may be 1-5 mm. The holes may be as large as possible such that the outer covering maintains sufficient strength to transmit the tensile forces required for removal. The holes may be preferentially placed along the device 10. In some embodiments, the holes are placed distally to enhance tissue ingrowth from the distal LAA wall.
In some embodiments, the device 10 includes an occlusion region and anchoring region. The proximal portion of the device 10 facing the LA after the device is implanted in the LAA may include the occlusion region. The occlusion region may be a blood contacting surface on the proximal end of the device 10 that is thromboresistant while promoting formation of a neointima at the occlusion region. The occlusion zone encourages thromboresistance and endothelialization from the blood and adjacent tissue. The anchoring zone promotes fast and tenacious tissue ingrowth into the device 10 from the adjacent non-blood tissue. The anchoring zone may be lateral surfaces of the device 10 that interface with tissue adjacent and/or within the LAA. The anchoring zone can also include the distal end of the device 10 that faces the distal wall of the LAA after implantation.
Any of a variety of structures may be utilized as the dynamic internal locking system 101 with the device 10. In general, at least about two or four or six or more tissue anchors 120 may be actively or passively advanced from the implantable device 10 into adjacent tissue surrounding the implantation site. Following deployment of the device 10 and expansion of the body 15, a tissue engaging segment 121 of the tissue anchor 120 will extend beyond the skin by at least about one, and in some implementations at least about two or four 4 mm or more. The tissue engaging segment 121 is carried by a support segment 122 of the tissue anchor 120 which extends through the foam body 15, and may be attached to a deployment control such as a pull wire, push wire, tubular support or other control structure depending upon the desired configuration.
The locking system 101 discussed primarily herein is a passive deployment construction. Removal of a constraint allows the tissue anchors 120 to laterally self-expand to deploy into adjacent tissue. Self-expansion may be accomplished by constructing the tissue anchor 120 using nitinol, Elgiloy, stainless steel or other shape memory or spring bias materials. The constraint may be removed by proximal retraction or distal advance until the tissue anchors 120 are no longer engaged by the constraint, depending upon the construction of the locking system 101.
Alternatively, tissue anchors 120 may be deployed actively such as by distal advance, proximal retraction or rotation of a control, or inflation of a balloon positioned within the device 10 to actively drive the anchors 120 through the skin 20 or corresponding apertures on the skin 20 and into tissue. For example, a plurality of support segment 122, such as struts, may be joined at a distal end to a central hub 111, and incline radially outwardly in the proximal direction. Proximal retraction of the hub 111 will cause the tissue engaging segment 121 to advance along its axis beyond the skin 20 and into the adjacent tissue. The inclination angle of the support segment 122, for example the struts, may be reversed, in another construction, such that distal advance of the hub 111 will deploy the tissue engaging segments 121 beyond the skin 20. Proximal or distal advance of the hub 111 may be accomplished by proximal or distal advance of a control such as a control wire or inner tube releasably engaged with the hub 111.
Depending upon the desired clinical performance, the tissue anchors 120 may be retractable, such as by axial distal or proximal movement of the control depending upon the inclination angle of the anchors 120. In the embodiment primarily illustrated herein, re-sheathing the anchors 120 may be accomplished by advancing the tubular constraint along the ramped surface of the tissue anchor 120 to move the anchor 120 radially inwardly towards the central longitudinal axis of the device 10. In the case of an anchor 120 which deploys by advance along its own longitudinal axis, the anchor 120 may be retracted by advancing the control in the opposite direction from the direction advanced to deploy the anchors 120.
Referring to
The illustrated anchor 120 may have a distal region 130, a hinge region 135, and/or a proximal region 125. The distal region 130 interacts with the hub element 111. The hinge region 135 and the curvilinear geometry as shown allow the end of the proximal region 125 to extend beyond the body 15, for example beyond a sidewall of the body 15. The proximal region 125 includes a tissue engaging segment 121 configured to engage adjacent tissue. The tissue engaging segment 121 may be the entire proximal region 125 or a portion thereof, for example the tip, etc. The proximal region 125 may thus include a sharpened tissue engaging segment 121, a shaped tissue engaging segment 121, an angled tissue engaging segment 121, a thickness configured for tissue engagement, and/or other suitable features. In some embodiments, the proximal region 125 may retract back within the body 15, as further described herein. In the embodiment shown, the anchor 120 and central tube 111 are distinct elements which are affixed to one another as shown. In other embodiments, the anchor 120 and tube 111 are a single, integral unit.
The internal locking system 101 is made from biocompatible metallic wire such as Nitinol, implant grade stainless steel such as 304 or 316, or cobalt-chromium based alloys such as MP35N or Elgiloy. In some embodiments, the internal locking system 101 may be cut from a single tubular piece of metal fabricated via machining or laser cutting followed by a secondary forming or annealing step using similar materials.
The internal locking system 101 may be in a constrained configuration as the device 10 is placed in position in the LAA and the body 15 expands therein. Then, in a secondary step, the internal locking system 101 locks or otherwise secures the device 10 in the LAA by engaging the anchors 120. If the position is not considered optimal, or if the device 10 otherwise needs to be repositioned within and/or removed from the LAA, the internal locking system 101 and anchors 120 thereof can be unlocked and the device 10 repositioned and/or removed.
In the illustrated construction, deployment of the tissue anchors by distal advance of the restraint enables reversible deployment, so that subsequent proximal retraction of the restraint will retract the tissue anchors. Alternatively, proximal retraction of the restraint to release the tissue anchors will irreversibly release the tissue anchors.
In some embodiments, the anchors 401 may be fixed such that they do not move axially. For example, the anchors 401 may have a portion, such as a tissue engaging segment 121, of a fixed length extending outside the body 15. The portion of the anchors 401 extending outside the body 15 may be bent when compressed within a delivery catheter and/or sheath, and these portions of the anchors 401 may then straighten out to the configuration shown in
In some embodiments, the anchors 401 may be moveable axially. For example, the anchors 401 may not deploy or otherwise extend outside the body 15 immediately upon expansion of the body 15. Following acceptable positioning of the device 10 within the LAA, the flexible anchors 401 may then be advanced through a corresponding tube 500. The anchors 401 may move axially in any suitable manner, including those described elsewhere herein. The anchor 401 may be moved through the tube 500 either before or after the tube 500 has been moved and deployed outside the body 15, as described below.
In some embodiments, the tubes 500 are moveable and deploy outside the body 15. The tubes 500 may be moveable in embodiments having either fixed or moveable anchors 401. The tubes 500 may be pre-loaded over corresponding wire anchors 401 as shown in
In some embodiments, the tubes 500 extend from the delivery catheter to or near the outer surface of the body 15 but do not extend outside the body 15. Instead, the tubes 500 just guide the anchor 401, for example around the curve, and support the wire 401 right up to tissue penetration. The tubes 500 may set the launch angle so the anchor 401 does not buckle and hits the tissue at the right angle. In this embodiment, the anchor 401 may have relatively more stiffness than in the embodiments where the anchors 401 are relatively flexible, in order to provide a more secure anchoring of the device 10 to the tissue. It is understood the tube 500 may provide this guiding function to the corresponding anchor in any of the embodiments described herein having moveable anchors, such as the moveable anchors 401, the anchors 120, etc.
The flexible anchors 401 and/or the external stiffening tube 500 may be made from biocompatible metallic materials such as Nitinol, implant grade stainless steel such as 304V or 316LVM, cobalt-chromium based alloys such as MP35N or Elgiloy, other suitable materials, or combinations thereof. The anchor 401 length can vary from 0.1 mm to 5 mm in length with an external stiffening tube 500 that covers from 10% to 90% of the exposed length of the anchor 401.
The skin 20 at least partially surrounds the body 15 and portions of the skin 20 may or may not be attached to the body 15. The various devices 10 described herein may have the body 15 at least partially encased within the skin 20, which may be fabricated from a material such as ePTFE (expanded polytetrafluoroethylene), polyolefin, or polyester which assists with healing, anchoring, and retrieval.
The selective location of the points of attachment 700 may facilitate with formation of a circumferential rim 800 of the skin 20. The rim 800 is shown schematically in
In some embodiments, the internal locking system 101, for example anchors thereof, may be preloaded surface elements releasably constrained or otherwise locked down in a collapsed or constrained position or configuration. The internal locking system 101, for example anchors thereof, may be constrained using a restraint. The restraint may be a dissolvable polymer, a lasso, or wires that can be retracted to release the anchors. The restraint may be similar to a deadbolt. Other anchoring concepts include Velcro integral to the ePTFE, electrically orientable/ratcheting anchoring elements, unidirectional Gecko tape, or wires pre-attached to the finial 30. In some embodiments, the body 15 with skin 20 may be secured within the LAA by texturing the body 15 and exposing the body 15 to the tissue through holes in the skin 20 to increase the friction with the cardiac surface to a high enough level to prevent implant migration.
One or more of the anchors 1506 may be placed along the skin 20 or otherwise along an external surface of the body 15. One or more corresponding guides 1500 and locking loops 1504 may be located along the skin 20 or the body 15. The guides 1500 may be located on a first side of the anchor 1506 and the locking loops 1504 may be located on a second, opposite side of the anchor 1506, as shown. The anchors 1506 are held in the constrained or restrained configuration or position by a sheath cover 1502. The sheath cover 1502 may be tubular or rectangular in shape. The sheath cover 1502 constrains the anchors 1506. The sheath cover 1502 may constrain the anchors 1506 in a flat position as shown in
Various features for LAA (LAA) occlusion may be included, such as those described, for example, in U.S. patent application Ser. No. 15/290,692, filed Oct. 11, 2016 and titled DEVICES AND METHODS FOR EXCLUDING THE LAA, in U.S. patent application Ser. No. 14/203,187, filed Mar. 10, 2014 and titled DEVICES AND METHODS FOR EXCLUDING THE LAA, in European Patent Application no. EP 14779640.3, filed Aug. 24, 2015 and titled DEVICES AND METHODS FOR EXCLUDING THE LAA, and in PCT Patent Application no. PCT/US2014/022865, filed Mar. 10, 2014 and titled DEVICES AND METHODS FOR EXCLUDING THE LAA, the entire disclosure of each of which is hereby expressly incorporated by reference for all purposes and forms a part of this specification. Further additions and improvements to these and other concepts are described below. The embodiments described in the sections below may include the same or similar features and/or functionalities as the embodiments described above, and vice versa, except as otherwise noted or indicated by context.
A. Basic Plug Design Components and Improvements
Various occlusion devices and associated features are described with respect to
As shown in
The tissue scaffold 1030 may be integrally formed with the tubular wall 1028 or may be bonded thereto. The tissue scaffold 1030 and the tubular wall 1028 may have approximately the same thickness and pore characteristics, discussed below. Alternatively the tissue scaffold may comprise a different material such as ePTFE, PTFE, Dacron, or others known in the art, configured to support tissue ingrowth and isolate the LAA, but thinner than the tubular wall 1028.
An expandable, internal support structure 1032 such as a wave stent 1034 or other frame may be provided. The illustrated wave stent 1034 comprises a plurality of struts 1038, adjacent pairs of struts joining to form a plurality of proximal apexes 1041 and distal apexes 1042. The stent 1034 may be laser cut from tube stock as is known in the art. Each of at least three, and preferably at least 4 or 6 or 8 or more of the proximally facing apexes 1040 is provided with a reentry or recapture strut 1044, which incline radially inwardly in the proximal direction to a central hub 1046. Recapture struts may be cut from the same tube stock as the stent. Hub 1046 may be provided with a central lumen, such as for delivery over a guidewire, or for releasable engagement with a deployment device (not illustrated). Alternatively, the hub 1046 may be provided with an attachment such as an eyelet 1048 for receiving a suture loop. A suture or other retention element may extend distally from the deployment catheter, through the tissue scaffold 1028, through the eyelet 1048 and proximally back through the tissue scaffold 1028 and into the deployment catheter. Following satisfactory positioning of the implant 1020, the suture may be removed, releasing the implant 1020 from the deployment catheter leaving a homogeneous tissue scaffold 1030 as resilience of the material closes the suture tracks. In one preferred implementation the implant 1020 is deployed from the delivery catheter without advancing over the wire and the hub lacks a central lumen. In one embodiment, the implant is secured to the delivery system using any of a variety of means known in the art including screw mechanisms or ball in socket attachment mechanism, which can also pivot.
The tubular wall 1028 may be attached to the stent 1034 by adhesives, sutures or other bonding techniques known in the art. In the illustrated embodiment, the tubular wall 1028 is sutured to the stent 1034 and the tissue scaffold 1030 is sutured to the recapture struts 1044, with the support structure 1032 carried within the cavity 1026. Alternatively, at least a portion of the support structure 1032 may be carried on the outside surface of the tubular wall 1028 or tissue scaffold 1030. The wave stent may be embedded within the tubular wall 1028 such as by sandwiching the stent between inner and outer layers of foam which are then bonded together. Similarly the recapture struts may be enclosed between inner and outer polymer layers. The polymeric material can also be foamed around the stent so that a secondary attachment process is not required.
The frame may be expandable from a contracted delivery configuration to an expanded deployed configuration. The frame may be retractable from an expanded deployed configuration to a contracted delivery configuration. The frame may be generally tubular, e.g. circular, rounded, segmented, polygonal, other shapes, or combinations thereof in the unconstrained, expanded configuration, and preferably presses the foam into conformance with the shape of the inside surface of the LAA. This permits the deployed implant to minimize leaks, the largest of which is no more than about 4 mm or 3 mm or 2 mm or less and in some deployments essentially no leaks as viewed on color Doppler.
Some advantages of the cored-out foam “cup”, as compared to a solid foam plug, are as follows: it still behaves like a full foam plug with respect to conformability and sealing; it allows incorporation of an internal metallic frame which can be optimized to provide the desired amount of expansion for sealing and anchoring by providing an optimal radial force and an attachment point for the anchors, and a front face inside the proximal face to aid in collapse of the foam for retrieval; by sizing the metallic frame so that it is shorter in length than the foam cup, an atraumatic distal bumper is formed which is entirely foam and can be extruded from the sheath tip as the tip is advanced within the LAA; and the reduced overall volume of material aids with the following: it significantly reduces the delivery profile, it is easier to flush to remove air prior to delivery into the vascular system, and it makes the plug more porous to blood so that if embolized within the cardiovascular system, it allows more blood to flow through it.
In an embodiment, the proximally facing edges of the foam plug 1040 are chamfered to aid in loading and retrieval. Also, while there can still be a central location to allow tracking of the implant 1020 over a guidewire-type device, in some embodiments there is no lumen or there is just a slit in the foam plug 1040, as it is not desirable to have a significant residual central hole which may cause thrombus formation or allow leakage. The slit can be a single slit, double cross-shaped slit, or multiple slits. The objective is to still allow tracking over a guidewire but to be sure the hole closes completely once the guidewire is removed. In the case of a solid face, the implant 1020 may not be tracked over a guidewire and may instead, for example, be delivered through a long transseptal sheath.
Please note that the term “guidewire”, as utilized above, can mean an actual medical device sold as a guidewire or it can be a catheter, such as a pigtail catheter, which is initially placed in the LAA over which the LAAC (LAA Closure) implant 1020 is tracked.
Diameters: The LAA may vary in diameter from about 15 mm to about 33 mm and as such, the implant 1020 diameter must be able to accommodate this variation in sizes. The more the implant 1020 may accommodate various large ranges of diameters, the fewer predetermined sizes are needed, thereby simplifying the procedure for implantation. The construct of this implant 1020 is such that it may accommodate diameters less than 50% of the fully expanded diameter. The preferred plug 1040 diameters may be about 27 mm, 33 mm, and 35 mm. Ideally, only 1-2 sizes would be needed to close a large range of LAA diameters. This is a key advantage of the foam plug 1040 concept as compared to metal cage type devices with fabric tissue scaffolds.
Depth: The preferred plug 1040 length (the depth of the occluder within the LAA generally along a proximal-distal direction) is 20 mm and is independent of the diameter of the implant 1020. This allows for good implant 1020 stability while still accommodating the majority of anatomies. The distal tip of the foam plug 1040 is very soft, providing an atraumatic tip as it enters the LAA when the distal tip of the implant 1020 is allowed to protrude past the distal dip of the delivery catheter or sheath. The short depth makes placement of the plug 1040 more robust, as there is less need to align delivery catheters with the LAA, as is needed with longer devices.
Foam and Porosity: The average pore size of the foam is 250-500 microns. The foam has a very high void content (90-95%) to promote quick and thorough tissue ingrowth. The open cell foam permits blood to flow through it. If the plug 1040 should embolize, it will be open enough to allow enough blood flow to be safe until it can be retrieved. Additionally, the large void content should be beneficial for proper flushing of the implant 1020 to prevent air introduction into the vascular system. The porosity and cell size may be as described with respect to the foam body 3002 of the device 3000, shown and described for example with respect to
The compliance and thickness of the foam are designed to provide a good seal against the tissue with minimal compression. While other devices require significant oversizing to obtain a seal, this implant 1020 may require only ≤1 mm of oversizing.
LA Facing Surface: ePTFE (expanded Polytetrafluoroethylene) or PTFE as the skin/layer for the implant 1020 such as over or partially over the plug 1040, as described above, may be ideal for supporting neointimal formation without thrombus formation. While embodiments may be described with respect to ePTFE, it is understood that PTFE may also be used. It may, however, reduce safety due to the inability or reduced ability to allow blood flow through the surface of an embolized implant 1020 due to the low porosity of the ePTFE, even though blood flow may be enabled around the outer surfaces of the cup. ePTFE porosity is much lower than that of the open cell foam, so blood flow across the membrane may be negligible. It is, however, hydrophobic which is beneficial for thromboresistance. One option, to maintain the desired open porosity of the foam structure while adding the thromboresistance of ePTFE and/or PTFE, is to add a PTFE coating via vapor deposition to the foam. The thromboresistant coating may contain ePTFE or PTFE. This creates a highly porous surface that simulates the ePTFE morphology. Methods of attachment could include vapor deposition, as mentioned above or elastomeric glue (although this may eliminate porosity at the attachment points). If ePTFE is preferred, that could be attached by encasing the metallic frame in ePTFE then wrapping through the center, around the OD, and attaching via sutures.
It may be desirable to reduce the pore size of the foam to between about 30-200 um, as further described herein.
Barbs/Anchors: There are several options or types of barb designs that may be implemented for anchoring the implant 1020 within the LAA. The following are some examples: 1) Static: always engages tissue when the plug is deployed. This makes re-sheathing and repositioning of the implant 1020 more difficult. 2) Constrained: the implant 1020 can be deployed with barbs constrained. The implant 1020 can then be repositioned, as needed. The barbs are then released when the plug 1040 is in its final position. 3) Dynamic: barbs can be deployed or retracted as desired without dislodging plug. Dynamic barbs may allow for deployment and retraction, for example to reposition and/or remove the implant 1020.
In some embodiments, the implant 1020 may have different features form other implants described herein. In some embodiments, the implant 1020 may include any or all of the following features: does not have a central lumen; has a spoked element against the proximal face inside the cup in addition to the wave stent; has anchors/barbs that can preferably be activated as a secondary step following placement of the plug itself; has a layer on the proximal face, which may be similar to the proximal face 1604′ shown in
B. Endoskeleton System with Proximal Spokes
The implant 1020 may include a foam plug 1040 with a central endoskeleton that includes a proximal spoke face 1080 with several radial struts. This configuration improves the ability to retrieve (re-sheath) the foam implant 1020. The stent version depicted in
C. Proximal (Blood-Contacting) Surface of the Foam
The foam implant 1020 may be comprised of porous open cell foam. The foam can be any of a variety of currently available materials including polyurethane-based biomaterials such as polyurethane or polycarbonate-polyurethane, or polyvinyl acetal (PVA) (Ivalon®). The foam can also be reticulated, such as a net. An embodiment utilizes a non-resorbable reticulated polyurethane-based biomaterial. Additionally, resorbable foams could also be utilized including a polyhydroxyalkanoate (PHA) such as poly-4-hydroxybutyrate (P4HB) or cross-linked resorbable polyester urethane-urea scaffolds.
Pore sizes in the material may be from about 50 micron to about 800 micron, preferably from about 250 micron to about 500 micron. Such a high void content (e.g. from about 90% to about 95%) material promotes quick and tenacious tissue ingrowth and effectively mimics the extracellular matrix. While such a high void content material is desirable for tissue ingrowth, it may not be ideal for thromboresistance, which is required for the left atrial (LA) surface. A thromboresistent surface is desired on the face that faces the LA. If modifications are made to the LA facing surface to promote blood compatibility, those modifications may be extended from about 1 mm to about 20 mm, preferably from about 1 mm to about 5 mm, onto the sides of the implant 1020 to assure thromboresistance in the event the implant 1020 is deployed with a portion of the implant's 1020 side surface projecting outside the LAA and into the blood environment. If there is a hole within the proximal face, such as a guidewire lumen, the thromboresistent surface may extend at least partially within that lumen. Additionally, this thromboresistent layer should promote tissue ingrowth and endothelialization.
Various methods may be used to create a thromboresistent proximal face of the implant 1020, including but not limited to the following. For example, an expanded PTFE (ePTFE) skin or layer may be applied to the outside surface of the foam implant as described elsewhere herein. It could be attached by encasing the metallic frame in ePTFE then wrapping through the center, around the OD, and attaching to the frame. This can be done using numerous methods including suturing or adhesive bonding, including the use of elastomeric glue (although this may eliminate porosity at the attachment points). This ePTFE layer can extend into the guidewire lumen, if there is one. In addition to ePTFE, electrospun, melt blown, non-woven, knitted or woven fibers of PTFE, polyester, PGA, PLA, poly-4-hydroxybutyrate (P4HB) or other biocompatible fiber materials can be utilized to create a porous biocompatible surface.
In some embodiments, a coating of a hydrophobic material such as PTFE is applied over the proximal face using any of numerous processes known to one skilled in the art including vapor deposition coating. Ideally this coated layer would also extend partially onto the sides of the implant. While some embodiments may not include a guidewire lumen, if a central lumen is present, the coating would preferably extend at least partially (e.g. about 1 mm) into it. To promote thromboresistance, this coated layer would decrease the porosity of the blood-contacting face to a porosity from about 30 microns to about 200 microns, preferably from about 100 microns to about 150 microns. Materials which could be utilized for this include conformal coatings such as PTFE applied in a thickness of about 50-100 microns, polyurethane spray or dip coatings, albumin, polyethylene glycol (PEG), or poly(ethylene oxide) (PEO), all with or without the incorporation of heparin or nitric oxide. The PEG or PEO would ideally be attached via a grafting process. In the preferred embodiment, the outer layer would also be lubricious to assist in re-sheathing of the implant. This can be achieved with both hydrophobic materials, such as ePTFE and PTFE, and hydrophilic ones, such as PEO and PEG. In order to produce a desirable combination of porosity and blood compatibility, a two-step process may be utilized where the foam is first coated with a base layer, such as a polyurethane-based biomaterial, then in a second step a more thromboresistent and lubricious surface is created utilizing PTFE, PEG, or PEO. Heparin or other anticoagulants may be added to the final blood-contacting surface.
Another option to create smaller pore sizes with an ePTFE-like material would be to attach an electrospun layer of PTFE to the face of the foam using suturing. A very thin layer (<1 mm) can be made and attached via suturing or adhesive bonding.
Another desired property of the foam of the plug 1040 is to provide echogenicity of the implant 1020, which allows for visualization by echocardiography. To promote echogenicity, a porous surface may be adequate; however, in certain cases a hydrophilic surface may be beneficial. To promote blood compatibility and a hydrophilic surface, the preferred embodiment would be a foam implant which has a surface grafted with PEO or PEG.
D. Static Barb (Anchor) Designs
Static barbs engage tissue when the plug 1040 is deployed, e.g. expanded. While this simplifies fabrication, it makes re-sheathing and repositioning of the implant more difficult.
In some embodiments, as shown in
Referring to
Referring to
Referring to
E. Constrained Barb (Anchor) Designs
With constrained barbs 2010, the implant 1020 can be deployed in the LAA with the barbs 2010 constrained and can therefore be repositioned, as needed, prior to barb release. The barbs 2010 are released when the implant 1020 is in its final position.
In one embodiment, as shown in
As shown in
A flipping-barb option is shown in
F. Dynamic Barb (Anchor) Designs
Dynamic barbs 2020 can be deployed or retracted as desired without dislodging the implant. Dynamic barbs 2020 may be the preferred option for some procedures, for example for recapture of the implant 102 into the delivery catheter and/or sheath.
One embodiment is a tube within a tube. In this embodiment, as shown in
As shown in
In another embodiment, as shown in
Recapture or re-sheathing of the implant 1020 may initiate the barb retraction by collapsing the front spoke face, as shown in
In another embodiment, as shown in
G. Embodiments with Foam Cup, Wave Stent, and ePTFE Layer
As shown in
H. Dynamic and Static Anchor Concepts
The embodiment of an implant 2300 shown in side cross-section view and end view in
The embodiment of the implant 2350 shown in
I. Constrained Anchor Deployed in Secondary Step
The embodiment of an implant 2370 shown in
J. Internal Stent with Increased Embolization Resistance
A metallic stent-like frame 2380 with distal static barbs 2381 and a proximal “speed bump” 2382 is disclosed as shown in
K. Constrained Atraumatic Anchor
As shown in
During delivery of the implant through the vascular system and into the heart and LAA, the loops 2391 are folded toward the center of the constrained stent and are lined up next to each other, as shown in
L. Perfusion Element & Barb Options
In the event of the implant being dislodged after deployment in the LAA, it is anticipated that the device would travel into the LA, across the mitral valve and into the left ventricle, then out the left ventricular outflow track, across the aortic valve and into the aorta and distal circulation. In this journey, it is possible that the device may get lodged in any of the above structures, interrupting flow and causing distal ischemia and possible hemodynamic collapse. It therefore would be desirable for an implant 2392 to have design features which would allow for distal perfusion, for example as shown in
As shown in
Regarding barb designs, such as those described herein, in addition to having an array of barbs deployed at different distances from the LA surface, barbs can be incorporated which penetrate into the tissue to different depths. In one embodiment, the barbs placed most proximal on the implant can be longer so they penetrate deeper into the thicker proximal LAA tissue whereas those placed most distal on the implant are shorter, so they penetrate less deep into the delicate distal tissue, to maximize embolization resistance while minimizing the risk of perforation. In another embodiment the proximal and distal barbs can be the same length but can be fabricated from different diameters of wire or can be cut at different thicknesses from tubing material, so that the barbs that engage the more delicate distal LAA tissue are more flexible, or they are designed to primarily engage the internal trabeculations within the LAA, whereas the proximally located barbs penetrate the tissue. Alternatively, two barbs can be placed at each crown point of the stent, as opposed to one.
M. Multi-Functional Occlusion Devices
In some embodiments, the LAA may be electrically isolated. The LAA may be electrically isolated with an occlusion device that incorporates one or more ablative elements, such as those shown and described with respect to
N. Embodiments with Compressible Foam Body, Proximal Cover, and Compliant Frame Having Proximal Recapture Struts and Distal Tubular Body
1. Compressible Foam Body
The body 3002 is formed from a compressible material, such as foam. The body 3002 may be a foam formed from reticulated (e.g. net-like) polycarbonate polyurethane-urea. The body 3002 may be cut, formed or assembled into a cup shape, as further described. The body 3002 may have a thickness and compressibility sufficient to engage the surrounding tissue and conform to the anatomic irregularities under radial force applied by the inner frame, as further described. The use of a compressible material such as foam for the body 3002 provides a complete seal of the LAA and superior performance for LAA occlusion over existing devices, as further described. The structure of the foam of the body 3002 comprises a three dimensional network of interconnected reticulations, spaced apart to form a network of interconnected open pores, as further described. The reticulations can carry a coating, such as PTFE, while preserving the open pores, as further described.
The foam material of the body 3002 has a high porosity. “Porosity” as used herein has its usual and customary meaning and refers to open void content between the interconnected reticulations of the foam. The porosity of the body 3002 may be at least about 65%, at least about 70% at least about 75%, at least about 80%, at least about 85%, at least about 90%, at least about 95% or more. The porosity may be within the range of approximately 90-95%. The porosity may be approximately 90%. The porosity may be approximately 95%. The porosity may be 90%, 91%, 92%, 93%, 94%, or 95%. The high porosity promotes quick and tenacious tissue ingrowth, allows it to be compressed into a small catheter, and/or allows blood to pass if the implant embolizes, among other advantages.
The foam body 3002 has pores or cells formed between the interconnected reticulations of the foam material. The foam body 3002 has cells with sizes in the range of from about 250 μm to about 500 μm. The foam may have a cell size from about 125 μm to about 750 μm, from about 175 μm to about 650 μm, from about 200 μm to about 600 μm, from about 225 μm to about 550 μm, from about 275 μm to about 450 μm, less than 125 μm, or greater than 750 μm. These sizes may refer to the size of the cell prior to application of any coating, such as PTFE. The cell size may thus change, e.g. decrease, after application of the coating. The desired porosity and/or cell size may be determined based on allowing the passage of blood while blocking debris of a size capable of potentially causing ischemic stroke. The allowable size of such debris may drive the selection of the particular porosity and/or cell size. For example, the cell size from about 250 μm to about 500 μm may be based on prevention of debris of a particular size from passing through the body 3002.
In an embodiment, the foam body 3002 is made from a non-resorbable, reticulated, cross-linked, polycarbonate polyurethane-urea matrix, structurally designed to support fibrovascular tissue ingrowth, with a fully interconnected, macroporous morphology with over 90-95% void content and cell sizes ranging from 250 to 500 μm.
The body 3002 has a proximal end 3004 and a distal end 3006. In some embodiments, the axial length of the device 3000 from the proximal end to the distal end in a free, unconstrained state is 20 mm. As used herein, the “free, unconstrained” state, and the like, refers to a state of the device 3000 without any external forces applied to the device 3000 other than a normal or reactive force from a surface (e.g. table top) on which the device 3000 is placed. In some embodiments, this axial length may be from about 10 mm to about 30 mm, from about 12 mm to about 28 mm, from about 14 mm to about 26 mm, from about 16 mm to about 24 mm, from about 18 mm to about 22 mm, or about 20 mm. The body 3002 may have any of these lengths regardless of outer diameter of the body 3002.
The proximal end 3004 of the body 3002 has a proximal end wall or face 3008. The proximal face 3008 faces generally toward the LA when the device 3000 is implanted into the LAA. The device 3000 may be implanted off-axis, as further described, in which case the proximal face 3008 may not reside at a perpendicular to a longitudinal axis of the LA. The proximal face 3008 thus provides a closed proximal end 3004 of the body 3002. The closed proximal end 3004 is configured to span the ostium but the porosity, as further described, is sufficient to permit the passage of blood while blocking debris of a size capable of potentially causing ischemic stroke. This membrane may be formed by the body 3002 and/or the cover 3100. In some embodiments, the proximal face 3008 or portions thereof may be open. For example, there may not be a proximal face 3008, there may be a partial proximal face 3008, there may be a proximal face 3008 with portions removed, etc. In some embodiments, the proximal face 3008 or portions thereof is/are not included and any opening or openings is/are covered by the cover 3100. The size of any such openings in the proximal face 3008 may be driven by the desired size of embolic debris to be prevented from escaping the LAA, as further described.
The proximal face 3008 is flat or generally flat and generally perpendicular to the longitudinal axis of the device 3000. The proximal face 3008 has a circular or generally circular shape as viewed from the proximal end 3004 in an unconstrained expansion. In some embodiments, the proximal face 3008 may be flat, rounded, segmented, angled with respect to the longitudinal axis, other shapes, or combinations thereof. The proximal face 3008 may have a non-circular, polygonal, other rounded shape, other shapes, or combinations thereof, as viewed from the proximal end 3004.
The proximal face 3008 has an outer surface 3010 and an opposite inner surface 3012. The outer surface 3010 faces proximally away from the device 3000 and the inner surface 3012 faces distally toward the frame 3040. The surfaces 3010, 3012 may define outer and inner sides of the proximal face 3008. The thickness of the proximal face 3008 may be measured axially between the outer surface 3010 to the inner surface 3012. This thickness in a free, unconstrained state (e.g. uncompressed and expanded) may be from about 0.5 mm to about 5 mm, from about 1 mm to about 4 mm, from about 2 mm to about 3 mm, about 2.5 mm, or 2.5 mm. In some embodiments, the thickness may be less than 0.5 mm or greater than 5 mm. The thickness of the proximal face 3008 may be uniform or non-uniform. Thus the thickness may be greater or smaller in different regions of the proximal face 3008.
The body 3002 includes a sidewall 3014 extending distally from the proximal face 3008. The sidewall 3014 extends circumferentially about a perimeter of the proximal face 3008 to form a closed cross-section (i.e. extends circumferentially 360 degrees about the axis). The sidewall 3014 extends axially to define a tubular body concentric about the longitudinal axis of the device 3000. The longitudinal axis extends through a geometric center of the tubular body defined by sidewall 3014. The sidewall 3014 is tubular or generally tubular, e.g. cylindrical, along the axis. In some embodiments, the sidewall 3014 may be conical or frustoconical, for example where the proximal end is wider than the distal end or vice versa. The sidewall 3014 may have an outer profile at the proximal end thereof, and as viewed from the proximal or distal end, to match that of the outer perimeter of the proximal face 3008.
In some embodiments, the cross-section of the sidewall 3014 may not be closed, for example where there are openings in the sidewall 3014. Thus cross-sections taken at various locations along the longitudinal axis may or may not show a closed section. In some embodiments, the sidewall 3014 may be non-tubular, non-cylindrical, non-circular, polygonal, other rounded shapes, other shapes, or combinations thereof. In some embodiments, as shown, the sidewall 3014 may extend continuously for the entire length from the proximal end 3004 to the distal end 3006. In some embodiments, the sidewall 3014 may not extend continuously for the entire length from the proximal end 3004 to the distal end 3006. For example, the sidewall 3014 may include a plurality of disconnected sections, such as annular portions of the sidewall, located and spaced along the longitudinal axis and connected to the frame 3040.
The sidewall 3014 has an outer surface 3016 and an opposite inner surface 3018. The outer surface 3016 faces radially outward from the axis. The inner surface 3018 faces radially inward toward the axis. The thickness of the sidewall 3014 may be measured radially between the outer surface 3016 to the inner surface 3018. This thickness in a free, unconstrained state (e.g. uncompressed) may be from about 0.5 mm to about 5 mm, from about 1 mm to about 4 mm, from about 2 mm to about 3 mm, about 2.5 mm, or 2.5 mm. In some embodiments, the thickness may be less than 0.5 mm or greater than 5 mm. The thickness of the sidewall 3014 may be uniform or non-uniform. Thus the thickness may be greater or smaller in different regions of the sidewall 3014. The thickness of the sidewall 3014 may be the same or different as the thickness of the proximal face 3008. In some embodiments, the thickness of the proximal face 3008 is 2.5 mm and the thickness of the sidewall 3014 is 2.5 mm. In some embodiments, the thickness of the proximal face 3008 is about 2.5 mm and the thickness of the sidewall 3014 is about 2.5 mm.
The sidewall 3014 has a distal free end 3020 having a distal surface 3022. The distal surface 3022 is flat or generally flat and perpendicular to the longitudinal axis of the device 3000. In some embodiments, the distal surface 3022 is non-flat, angled with respect to the axis of the device 3000, curved, rounded, segmented, other shapes, or combinations thereof.
The body 3002 may have a distal opening 3024. The opening 3024 is formed by the distal free end 3020 of the sidewall 3014. The opening 3024 is at a distal end of an internal central volume or cavity 3028 of the body 3002 that is formed at least partially by the sidewall 3014, the proximal face 3008 and/or the shoulder 3030. The frame 3040 may reside within the cavity 3028, as further described. The distal opening 3024 may be completely open. In some embodiments, the distal opening 3024 may be mostly open, partially open, or closed, for example where the body 3002 has a distal face similar to the proximal face 3008 to enclose or partially enclose the cavity 3028.
The body 3002 has a shoulder 3030, shown as a bevel, that extends between the proximal face 3008 to the sidewall 3014. The shoulder 3030 may be an intersection of a proximal end of the sidewall 3014 and the proximal face 3008. The shoulder 3030 extends circumferentially about the entire perimeter of the intersection. The shoulder 3030 has an outer surface 3032. The outer surface 3032 may be a beveled surface. The outer surface 3032 is flat or generally flat in an axial direction. The outer surface 3032 extends circumferentially about the entire perimeter of the shoulder 3030. In some embodiments the shoulder 3030 and/or outer surface 3032 may be non-flat, rounded, other shapes in an axial direction, or combinations thereof. The shoulder 3030 and/or outer surface 3032 may extend circumferentially less than the entire perimeter of the shoulder 3030. The thickness of the shoulder 3030 may be measured inward perpendicularly to the outer surface 3032. The thickness of the shoulder 3030 may be the same as the thicknesses of the proximal face 3008 and/or the sidewall 3014, as described herein. In some embodiments, the thickness of the shoulder 3030 may be different from the thicknesses of the proximal face 3008 and/or the sidewall 3014. The shoulder 3030 may function as a recapture ramp, to facilitate drawing the implant proximally into the deployment catheter.
The compressibility of the body 3002 contributes to the superior sealing capability of the device 3000. The foam may be compressible to provide a larger radial “footprint” and spread out the radial forces from struts on the frame 3040, as further described. The foam body 3002 may have a compressive strength of at least 1 pound per square inch (psi) or within a range of about 1 psi to about 2 psi, or no more than about 2 psi. The “compressive strength” here refers to the pressure to compress the foam to 50% strain. With some foam materials for the body 3002, the pressure may not change from 50% strain through at least 80% strain, and the relation of pressure versus strain may be flat or generally flat. Thus, even with thicker foams for the body 3002, the body 3002 will not exert much more outward force on the tissue due to the increased thickness by itself. In an embodiment, the foam body 3002 is a reticulated, cross linked matrix having at least about 90% void content, an average cell size within the range of from about 250-500 microns, a wall thickness of at least about 2 mm and a compressive strength of at least about 1 psi. In an embodiment, the body 3002 is formed from a foam material having or substantially having the material properties indicated in Table 1. In some embodiments, the body 3002 is formed from materials described in, for example, U.S. Pat. No. 7,803,395, issued Sep. 28, 2010, and titled “Reticulated elastomeric matrices, their manufacture and use in implantable devices,” or U.S. Pat. No. 8,337,487, issued Dec. 25, 2012, and titled “Reticulated elastomeric matrices, their manufacture and use in implantable devices,” the entire disclosures of which are incorporated herein by reference.
The device 3000 may include markers 3023 (see
In some embodiments, four platinum iridium (PtIr) radiopaque (RO) tubular markers 3023 are sewn onto the distal end 3006 of the foam body 3002 to enable visualization of the distal edge of the device 3000 under fluoroscopy. In some embodiments, a PtIr marker 3023 is attached to the foam body 3002 at the location of the proximal shoulder 3030 to use as a marker during recapture of the device 3000. Visualization of the proximal and/or distal markers 3023 may facilitate with identifying the amount of recapture. If the device 3000 is recaptured up to but not including the anchors proximal 3090 inside the access sheath, the device 3000 can be redeployed and reused. If the proximal anchors 3090 are recaptured into the access sheath, the device 3000 may be removed and discarded due to permanent deformation of the anchors 3090. In some embodiments, other materials may be used for the markers 3023, such as gold or other suitable materials.
As shown in
As further shown in
As shown, two markers 3023B are visible on either side of the interior of the foam body 3002. The markers 3023B are attached through the foam and around the frame 3040. The marker 3023B may be attached, for example sutured, around a proximal face 3060 member of the frame 3040, such as one of the struts 3061. The marker 3023B may be attached to the frame 3040 just proximally of one of the proximal apexes 3084 of the frame 3040, for example at an outer curved portion 3066 of the strut 3061. There may be only one marker 3023B, or two, three, four or more markers 3023B. There may be one of the markers 3023B for each strut 3061. The markers 3023B may be used additionally to connect the frame 3040 with the foam body 3002. The markers 3023B may be sutures as described herein.
The one or more markers 3023A and/or 3023B at or near the proximal end of the device 3000 provide various desirable features. For instance, the marker 3023A at the shoulder 3030 facilitates visualization of the device 3000 during and after implantation. The typically non-circular shape of the opening of the LAA (ostium) may compress the proximal end 3004 of the device and cause the proximal end 3004 to protrude slightly in the proximal direction. However, the shoulder 3030 may provide a location for the marker 3023A where linear bulging of the foam body 3002 in the proximal direction is reduced or prevented. Thus, the marker 3023A in that location can provide a more useful visualization of the positioning of the device 3000 and reduce complexity. For example, in some embodiments, the marker 3023A at the shoulder 3030 (e.g. on an inner surface as shown) may be particularly useful during delivery, allowing for delivery using fluoroscopy imaging only without the need for echo or other ultrasound imaging. The one or more markers 3023B may provide similar benefits.
As further shown in
The inner cover 3101 may be located between the foam body 3002 and the frame 3040. As shown, the inner cover 3101 is located between the inner surface 3012 of the foam body 3002 and a proximal end of the hub 3050 of the frame 3040. The inner cover 3101 may be circular or other shapes. The inner cover 3101 may have an area sufficient to provide a barrier in between the hub 3050 and the proximal end 3004 of the foam body 3002. In some embodiments, the inner cover 3101 may extend radially to an outer circumference of the hub 3050, or it may extend radially to the sidewall 3014 such as to an inner surface 3018 of the foam body 3002, or to any radial locations in between. The inner cover 3101 may have a diameter from about 4 mm to about 22 mm, from about 5 mm to about 15 mm, from about 6 mm to about 10 mm, about 8 mm, or 8 mm. The inner cover 3101 may be flat or generally flat. The inner cover 3101 may have a thickness from about 0.0001″-0.0020″, from about 0.0002″-0.0010″, about 0.0005″, or 0.0005″ thick. The inner cover 3101 may include one or more openings 3103 such as holes therethrough. The inner cover 3101 may include two holes 3103 to receive therethrough a tether 3240 (see, e.g.,
The inner cover 3101 may prevent the hub 3050 and/or other features of the frame 3040 from directly contacting the foam material. The cover 3101 may protect the integrity of the foam body 3002 from stresses that may be imparted by the hub 3050 on the foam material. This protection may be desirable for example during loading, deployment, retrieval, re-deployment, etc. of the device 3000. The inner cover 3101 may prevent or reduce damage to the foam body 3002 from the hub 3050.
The foam body 3002 may be attached to various features of the device 3000. The body 3002 may be attached to the frame 3040 at numerous points, including for example the center of the proximal end of the frame 3040, as further described herein. Attachment can be done using suture, such as polypropylene monofilament suture, although other methods known in the art such as adhesive bonding could be utilized. The proximal row of proximal anchors 3090 may be individually attached to (e.g. inserted through) the foam body 3002 to prevent relative movement between the foam body 3002 and the frame 3040. In other embodiments, the foam body 3002 could be formed around the endoskeleton so that the metallic frame is within the foam body 3002, eliminating the need for a secondary attachment step. Attachment of the body 3002 to the frame 3040 promotes retrieval without damage to the foam body 3002, among other advantages. The attachment also ensures that a bumper 3026, further described herein, extends beyond the frame 3040 at all times, including during initial exposure of the device 3000 upon proximal retraction of the delivery sheath.
As shown in
As shown, four attachments 3001 are visible in
The foam body 3002 may include a coating. In some embodiments, there may not be a coating. In embodiments with a coating, the coating is applied to the interconnected reticulations of the foam material. The body 3002 may be coated with pure polytetrafluoroethylene (PTFE). The PTFE coating minimizes the thrombogenicity of the LA surface, while also reducing the friction of the foam body 3002 against the delivery system to facilitate ease of deployment and retrieval. The body 3002 may be coated with conformable, vacuum deposited, pure PTFE. In addition or alternatively, the body 3002 may be coated with a coating other than PTFE. The coating, whether PTFE or otherwise, may be about 0.5 μm thick, and covers at least a portion of the surface of the interconnected reticulations of the foam without occluding the pores. The coating may be applied to some or all of the foam body 3002. The coating may be applied to some or all of the outer surfaces of the foam body 3002.
In some embodiments, the thickness of the coating is from about 0.1 μm to about 1 μm, from about 0.2 μm to about 0.9 μm, from about 0.3 μm to about 0.8 μm, from about 0.4 μm to about 0.7 μm, about 0.4 μm to about 0.6 μm, or about 0.5 μm thick. In some embodiments, greater or smaller thicknesses of the coating may be applied. The coating has a uniform or substantially uniform thickness. In some embodiments, the coating may have a non-uniform thickness. For example, the portion of the body 3002 facing the LA when implanted, such as the proximal face 3008 and/or shoulder 3030, may have a thicker coating relative to a coating along the sidewall 3014 of the body 3002. In some embodiments, the outer surface 3010 of the proximal face 3008 has a PTFE coating and the proximal face 3008 also has a ePTFE cover 3100.
The coating is applied using a vapor deposition process. In some embodiments, the coating is applied through coating, vapor deposition, plasma deposition, grafting, other suitable processes, or combinations thereof. The coating is applied to the outer surfaces 3010, 3032 and 3016 of, respectively, the proximal face 3008, the shoulder 3030 and the sidewall 3014. In some embodiments the coating is applied to the outer surfaces 3010, 3032 and only partially on the outer surface 3016. In some embodiments the coating is applied to outer and inner surfaces of the body 3002.
In some embodiments, other biocompatible, thromboresistant and/or lubricious materials could be applied to the surface(s) of the foam body 3002 and/or the cover 3100. These materials may encourage tissue ingrowth. Such materials may include, for example, heparin, albumin, collage, polyethylene oxide (PEO), hydrogels, hyaluronic acid, materials that release nitric oxide, oxygen, nitrogen, amines, bioabsorbable polymers, and other biomaterials, pharmacologic agents, and surface modification materials. Additionally, the surface(s) of the body 3002 could be roughened, textured, or otherwise modified or coated to promote healing or to make it more echogenic.
2. Proximal Cover
The device 3000 may include a cover 3100, which may be an ePTFE cover as further described. Other embodiments for this outer cover 3100 are described herein, for example the cover 3101, 3300, 3150, 3151, etc. The various embodiments of the cover may have the same or similar features and/or functionalities as each other, except as otherwise noted. The cover 3100 may have a series of openings. In some embodiments, the cover 3100 may be solid and not have any openings. In some embodiments, the cover 3100 may only have openings to receive anchors and/or a tether therethrough, as further described herein. In some embodiments, the device 3000 may include an inner cover such as an inner cover 3101, as shown and described with respect to
The outer cover 3100 is a generally flat material applied over and covering at least a portion of the body 3002. The cover 3100 is on the proximal end 3004 of the device 3000. The cover 3100 covers the proximal face 3008 of the body 3002 and at least part of the sidewall 3014. The cover 3100 covers a proximal portion of the sidewall 3014. The cover 3100 has a proximal surface 3102 that at least partially faces the LA when implanted. The cover 3100 has an outer edge 3104 forming outer vertices 3106 (for clarity, only some of the outer edges 3104 and outer vertices 3106 are labelled in the figures). In some embodiments, the cover 3100 may cover only the proximal face 3008 or portions thereof. In some embodiments, the cover 3100 may extend over more of the sidewall 3014, such as the middle or distal portion thereof, or the entire sidewall 3014.
The cover 3100 may have a thickness measured perpendicularly from the proximal surface 3102 to an opposite distal surface of the cover 3100 that faces the body 3002. The cover 3100 may have a thickness of 0.001″ (inches). In some embodiments, the cover 3100 may have a thickness from about 0.00025″ to about 0.005″, from about 0.0003″ to about 0.004″, from about 0.0004″ to about 0.003″, from about 0.0006″ to about 0.002″, from about 0.0008″ to about 0.0015″, or about 0.001″. In some embodiments, the cover 3100 may have a thickness of 0.0005″. In some embodiments, the cover 3100 may have a thickness from about 0.0002″ to about 0.0008″, from about 0.0003″ to about 0.0007″, from about 0.0004″ to about 0.0006″, or about 0.0005″.
The cover 3100 may be attached to the frame 3040 through the foam body 3002. The cover 3100 may in addition or alternatively be attached to the body 3002. The cover 3100 may be attached at least two or four or six or more of the outer vertices 3106. The cover 3100 may be attached to the frame 3040 and/or body 3002 at various locations, including at the outer vertices 3106, through the proximal surface 3100, at the proximal face 3008 of the body 3002, other locations, or combinations thereof. The cover 3100 is attached using mechanical attachments, such as sutures. In some embodiments, polypropylene 6-0 sutures are used throughout the device to attach the foam body 3002, proximal cover 3100, and RO markers 3023 to the foam body 3002 and/or frame 3040. In some embodiments, the cover 3100 is attached to the frame 3040 via standard braided or monofilament suture material, such as polypropylene, ePTFE, or polyester. In some embodiments, a polypropylene monofilament is utilized. Proximal anchors 3090 of the frame 3040 (further described herein) may extend through the outer vertices 3106 of the cover 3100. Such penetrating anchors 3090 may further secure the cover 3100 in place relative to the body 3002. In some embodiments, the cover 3100 may be attached to the various parts of the device 3000 with mechanical attachments, fasteners, adhesives, chemical bonds, other suitable techniques, or combinations thereof.
As shown, the cover 3100 is formed from expanded Polytetrafluoroethylene (“ePTFE”). An ePTFE cover 3100 provides many advantages. For example, the ePTFE cover 3100 may enhance the ability to recapture the device 3000 in vivo by distributing the proximal retraction forces applied by the catheter. The cover 3100 may be an ePTFE material approximately 0.001″ thick, with the appropriate porosity to encourage healing and minimize thrombus formation, similar to the underlying PTFE coated foam.
An ePTFE cover 3100 may assist in recapture of the implant into the access sheath while providing a smooth, thromboresistant surface which encourages tissue coverage and integration. The ePTFE may cover the entire proximal face and partially covers the sides, as shown in
In some embodiments, the cover 3100 is fabricated from other thromboresistant, high strength, biocompatible materials, such as knitted or woven polyester fabrics, polypropylene, polyethylene, non-woven vascular scaffolds, porous films, or bioabsorbable scaffolds such as polylactic acid, polyglycolic acid, and co-polymers. The shape of the cover prior to attachment with the device 3000, such as shown in
The cover 3100 may be perforated with a series of openings 3120 (for clarity, only some of the openings 3120 are labelled in the figures). The openings 3120 are perforations or holes formed in the cover 3100 via laser or mechanical cutting. The openings 3120 include proximal openings 3122 and side openings 3124 (for clarity, only some of the proximal openings 3122 and side openings 3124 are labelled in the figures). When the cover 3100 is assembled with the body 3002, the proximal openings 3122 are located over the proximal face 3008 and/or shoulder 3030, and the side openings 3124 are located over the sidewall 3014. In some embodiments, the cover 3100 includes forty proximal openings 3122. In some embodiments, the cover 3100 includes forty side openings 3124. The number of openings 3120 located over the proximal face 3008 and/or shoulder 3030 when assembled with the body 3002 may range from ten to eighty, from twenty to seventy, from thirty to sixty, from thirty five to fifty, or forty openings 3120. The number of openings 3120 located over the sidewall 3014 may range from ten to eighty, from twenty to seventy, from thirty to sixty, from thirty five to fifty, or forty openings 3120.
The openings 3120 may have a variety of sizes. The openings 3120 are 0.070″ in width, e.g. minor axis, or diameter for circular openings. The openings 3120 may have a width from about 0.010″ to about 0.200″, from about 0.020″ to about 0.150″, from about 0.030″ to about 0.110″, from about 0.040″ to about 0.100″, from about 0.050″ to about 0.090″, from about 0.060″ to about 0.080″, or about 0.070″. In some embodiments, the width may be less than 0.010″ or greater than 0.200″, such as 0.25″, 0.5″ or greater. These widths may apply to circular as well as non-circular shaped openings 3120.
In some embodiments, the openings 3120 may be various shapes. The openings 3120 may be elongated slots. The openings 3120 may extend radially along the cover 3100 from or near a center portion of the proximal surface 3102 toward and/or to the outer edge 3104. The openings 3120 may be annular openings extending circumferentially along the cover 3100 and having varying radial positions. The openings 3120 may be of uniform size and shape. Some of the openings 3120 may have varied sizes and/or shapes with respect to other of the openings 3120. The openings 3120 may have various distributions or concentrations about the cover 3100. For example, the openings 3120 may be more densely located in various areas, such as along the proximal surface 3102 that faces the LA, along the shoulder 3030, etc.
The openings 3120 enable blood to flow through the device 3000. The openings 3120 may allow blood to adequately flow through the device 3000 and thereby mitigate the risk of occlusion in the bloodstream should the device 3000 embolize within the vasculature system. In some embodiments, should the device 3000 embolize, it may act as a stationary filter at low pressures but may pass through the bloodstream at higher pressures. In some embodiments, the device 3000 allows for about two to about fourteen liters, from about four to about twelve liters, from about six to about ten liters, or from about eight liters per minute of blood to pass at <30 mmHg pressure drop to prevent shock in the event of a device embolization. In some embodiments, there are forty circular openings 3120 each having a diameter of 0.070″, and allowing for approximately eight liters per minute of blood to pass at <30 mmHg pressure drop. In some embodiments, the proximal end of the device 3000 may be a foam layer such as the foam proximal face 3008 or a membrane such as the cover 3100 or both, enclosing the cavity 3028 defined within the tubular side wall 3014 of the body 3002.
In one implementation, having both the foam proximal face 3008 and the cover 3100, the foam body 3002 has the open cell structure further discussed herein that can permit the passage of blood but block escape of embolic debris. The cover 3100 may be occlusive to blood flow, and is present to provide structural integrity and reduced friction for retracting the expanded body 3002 back into the deployment catheter. In one implementation, the cover 3100 is ePTFE in a form that is substantially occlusive to blood flow, as described. In this embodiment, the cover 3100 is therefore provided with a plurality of perfusion windows or openings 3120, so that blood can pass through the open cell foam and cover 3100 but the device 3000 still benefits from the other properties of the cover 3100.
In some embodiments, the device 3000 may allow for a particular flow rate of water at specified conditions, to test the perfusion performance of the device 3000. The device 3000 may have the foam body 3002 and cover 3100 configured to allow for a flow rate of water axially through the device 3000 of at least four liters per minute. The water may be at sixty-eight degrees Fahrenheit (F) or about sixty-eight degrees F. and an upstream pressure of twenty-five millimeters of Mercury (mmHg) or about twenty-five mmHg. In some embodiments, the device 3000 may be configured to allow for flow rates under such conditions from about one liter to about seven liters, from about two liters to about six liters, from about three liters to about five liters, more than two liters, more than three liters, or more than four liters of water per minute. The particular flow rate may depend on the porosity of the foam body 3002 and the open area of the cover 3100. The particular flow rate may depend on the inner cover 3101 features as well. The cover 3100 may have particular percentages of the cover area open with the series of openings, as further described herein, to attain a particular desired flow rate. The flow rate of water at the specified conditions may be used to extrapolate or otherwise calculate the corresponding expected flow rate of blood in the body through the device 3000 should it embolize, as described herein. The device 3000 may allow for a cardiac index from about 1.6-2.4, from about 1.7-2.3, from about 1.8-2.2, from about 1.9-2.1, about 2.0, or 2.0, liters per minute per square meter. The device 3000 may have these and other flow rate capabilities either aligned or approximately aligned with the direction of flow of the fluid, or off-axis where the device 3000 is angled with respect to the direction of flow of the fluid (a flow axis), as further discussed herein for example in the section “Off-Axis Delivery and Deployment.”
The cover 3300 includes a series of openings 3320. The openings 3320 include proximal openings 3322, shoulder openings 3323, and the side openings 3324. The proximal openings 3322 are located over the proximal end 3004 of the body 3002. The shoulder openings 3323 are located over the shoulder 3030, e.g. a bevel, of the body 3002. The side openings 3324 are located over a proximal portion of the sidewall 3014 of the body 3002. The proximal anchors 3090 may extend through the side openings 3324 that are located in the outer vertices 3106. The openings 3320 may have the same or similar features and/or functionalities as the openings 3120, and vice versa. In some embodiments, the proximal anchors 3090 may extend through the cover 3300 material at or near the outer vertices 3106.
The cover 3151 further includes another embodiment of a series of openings 3171. The openings 3171 include smaller openings 3175 and larger openings 3173. The openings 3175, 3173 may have the same or similar features and/or functionalities as other cover openings described herein, such as the openings 3120, 3122, 3124, 3320, 3322, 3324, 3170, 3172 and/or 3174, and vice versa. The smaller openings 3175 may be relatively smaller, in width and/or area, than the larger openings 3173. There may be openings with widths or areas smaller than that of the smaller openings 3175, larger than that of the larger openings 3173, or anywhere in between. As shown, the openings 3173, 3175 may be generally uniformly distributed about the proximal surface 3152 of the cover 3151. The openings 3173, 3175 may be circumferentially evenly spaced or approximately evenly spaced about the cover 3151.
There may be a variety of different quantities of each of the openings 3173, 3175. There may be a total of ten, twenty, thirty, forty, fifty, sixty, seventy, eighty, ninety, one hundred, one hundred fifty, two hundred, three hundred, four hundred, or more openings of the series of openings 3171, or any lesser, greater or in between number of openings. The series of openings 3171 may be holes as shown. They may have circular shapes. They may have other shapes, including non-circular, segmented, other shapes, or combinations thereof. The openings 3171 may all have the same general shape or different shapes. In some embodiments, there may not be any holes in the cover 3151.
When the cover 3151 is assembled with the foam body 3002, the large and small openings 3173, 3175 may be located over the proximal end 3004 and/or the sidewall 3014 of the foam body 3002. When assembled with the foam body 3002, on the proximal-facing portion of the cover 3151, there may be a collective total of one hundred forty or about one hundred forty openings 3173, 3175. On this proximal-facing portion of the cover 3151, there may be a collective total from about ten to about three-hundred, from about fifty to about two hundred fifteen, from about one hundred ten to about one hundred seventy, from about one hundred twenty to about one hundred sixty, from about one hundred thirty to about one hundred fifty, or from about one hundred thirty-five to about one hundred forty-five openings 3173, 3175. On this proximal-facing portion of the cover 3151, there may be from about thirty to about fifty, from about thirty-five to about forty-five, about forty, or forty of the larger openings 3173. On this proximal-facing portion of the cover 3151, there may be from about sixty to about one hundred forty, from about eighty to about one hundred twenty, from about ninety to about one hundred ten, about one hundred, or one hundred of the smaller openings 3175.
When assembled with the foam body 3002, on the portion of the cover 3151 located over and/or near the shoulder 3030, such as over the outer surface 3032 of the foam body 3002 (see, e.g.,
When assembled with the foam body 3002, on the portion of the cover 3151 located over and/or near the sidewall 3014, such as over the outer surface 3016 of the foam body 3002 (see, e.g.,
The larger and smaller openings 3173, 3175 may have a variety of different sizes, for example as described herein with respect to the openings 3122. In some embodiments, the openings 3173, 3175 may have diameters ranging from about 0.025 inches to about 0.040 inches. In some embodiments, the larger openings 3173 may be 0.040 inches or about 0.040 inches in diameter. The larger openings 3173 may be from about 0.030 inches to about 0.050 inches, or from about 0.035 inches to about 0.045 inches, in diameter. These values may also refer to the widths, for example maximum widths, of non-circular larger openings 3173. In some embodiments, the smaller openings 3175 may be 0.025 inches or about 0.025 inches, in diameter. The smaller openings 3175 be from about 0.015 inches to about 0.035 inches, or from about 0.020 inches to about 0.030 inches, in diameter. These values may also refer to the widths, for example maximum widths, of non-circular smaller openings 3175.
The series of openings 3171 may be configured to provide a desired amount of open area through the cover 3151. This open area refers to the total area of certain openings in the cover 3151. The cover 3151 may be covering a proximal face 3008 at the proximal end 3004 of the foam body 3002. The open area may refer to openings through the portion of the cover that is over the proximal face 3008 of the foam body 3002 when assembled with the foam body 3002. The series of openings in the various covers described herein may collectively provide the open area. For example, the series of openings 3171 in the cover 3151 over the proximal face of the foam may collectively provide an open area. This is the sum of the area of the openings in the cover 3151 over the proximal face. As further example, the open area may be the sum of the proximal openings 3122 of the cover 3100. As further example, the open area may be the sum of the proximal openings 3322 of the cover 3300.
The open area may be at least five percent of the area of the proximal face 3008 of the foam body 3002. The open area may be at least six, at least seven, at least eight, at least nine, at least ten, at least eleven, at least twelve, at least thirteen, at least fourteen percent, at least fifteen, at least sixteen, at least seventeen, at least eighteen, at least nineteen, at least twenty, at least twenty-five, at least thirty, at least forty, or at least fifty percent, of the area of the proximal face 3008. The open area may be from about one to about fifty percent, from about five to about twenty percent, from about eight to about fifteen percent, from about ten to about twelve percent, or about eleven percent, of the area of the proximal face 3008. The “area” of the proximal face 3008 is understood here to refer to an area equal to Pi×R2, where R is the radius of the proximal face 3008 and extends perpendicularly from the longitudinal axis of the device 3000. Further, “R” may be measured to the inner boundary of the shoulder 3030, to the outer boundary of the shoulder 3030, or to the outer surface 3016 of the sidewall 3014. Further, as mentioned, some embodiments may not include a cover at all.
The cover 3151 may include one or more windows 3177. As shown, there may be ten windows 3177. There may be one window 3177 for each proximal anchor 3090. There may be four, six, eight, twelve, fourteen or more windows 3177, or any lesser or in between number. The windows 3177 may be openings in the cover 3151. The windows 3177 may be located at or near the outer edge 3154 of the cover 3151. The windows 3177 may be located along portions of the outer edge 3154, for example at or near the outer vertices 3156. The windows 3177 may have a shape conforming to the shape of the cover 3151 at the respective portions of the outer edge 3154. As shown, the window 3177 may be diamond or generally diamond shaped. The window 3177 may be square, rectangular, triangular, rounded, circular, segmented, flattened diamond, other polygonal shapes, other shapes, or combinations thereof. The cover 3150 may be attached to the body 3002 at the outer vertices windows 3177. The windows 3177 may have the same or similar feature and/or functionalities as the side openings 3324, described and shown in
The device 3000 with cover 3151 may have proximal anchors 3090 extending through the windows 3177. The proximal anchor 3090 may extend through the opening of the respective window 3177. The proximal anchor 3090 may extend through a distal portion of the window 3177, for example to contribute to securing the cover 3153 on the device 3000. The proximal anchors 3090 may extend through the window 3177 at a distal edge or distal vertex of the window 3177. In some embodiments, the proximal anchor 3090 may extend through the cover 3151 material, for example through material adjacent (such as distal) to the window 3177. In some embodiments, the proximal anchor 3090 may extend through various other locations within, adjacent or near the window 3177. Some of the proximal anchors 3090 may extend through first locations and other of the proximal anchors 3090 may extend through second locations of the cover 3153 different from the first locations. For instance, one or more anchors 3090 may extend through a first region of the window 3177, one or more other anchors 3090 may extend through a second region of the window 3177, still one or more other anchors 3090 may extend through other regions, such as through the cover 3153 material, etc.
The cover 3153 may include proximal vertices 3155. The proximal vertices 3155 may be formed by the outer edge 3154. The proximal vertices 3155 may be indentations along the outer edge 3154 of the cover 3153, for example angled as shown or other shapes, configurations, etc. The proximal vertices 3155 may define a region 3016A of the outer surface 3016 of the sidewall 3014. The region 3016A may be partially enveloped by the outer edge 3154 of the cover 3153. The region 3016A may receive one or more of the distal anchors 3094 therethrough. The distal anchor 3094 may extend through a distal portion of the region 3016A, or in other locations within, adjacent, or near the region 3016A. In some embodiments, the distal anchor 3094 may not extend through or near the region 3016. There may be multiple such regions 3016A of the foam body 3002 defined circumferentially about the device 3000 by the cover 3153.
The cover 3153 may include the series of openings 3320, for example as described with respect to
3. Compliant Frame
The expandable and compliant support or frame 3040 is shown, for example, in
The frame 3040 has a proximal end 3042 and an opposite distal end 3004. The frame 3040 may be tubular, e.g. cylindrical, in a free, unconstrained state. Thus the width of the proximal end 3042 may be the same or similar to the width of the distal end 3004 in the free, unconstrained state. In some embodiments, the frame 3040 or portions thereof may be conical or frustoconical, e.g. where in the free, unconstrained state the width of the proximal end 3042 is greater than the width of the distal end 3004 or vice versa.
At the proximal end 3042, the frame 3040 has a proximal hub 3050, shown as a cylindrical nipple. The hub 3050 is a rounded, structural end piece. The hub 3050 may be tubular, e.g. circular and having the cylindrical shape as shown, or may be rounded, non-circular, segmented, other shapes, or combinations thereof. The hub 3050 extends axially and may have a central lumen. The hub 3050 may be wider than it is long, or vice versa. The hub 3050 is hollow and has a sidewall defining a space therethrough, such as a longitudinal opening. In some embodiments, the hub 3050 may be partially hollow, solid, or other configurations. The hub 3050 facilitates delivery and retrieval of the device 3000, as further described. The hub 3050 may provide a central structural attachment, as further described herein. The hub 3050 may be located within the cavity 3028 at a proximal end thereof. In some embodiments, the hub 3050 may be located partially or entirely within the foam body 3002, e.g. within the proximal face 3008.
A pin 3051 is located within the hub 3050 (shown in
The frame 3040 at the proximal end 3042 includes a proximal face 3060. The proximal face 3060 may be located within the cavity 3028 at a proximal end thereof. In some embodiments, the proximal face 3060 may be located partially or entirely within the foam body 3002, e.g. within the proximal face 3008 and/or sidewall 3014. The proximal face 3060 includes a series of recapture or reentry struts 3061. The struts 3061 are located at a proximal end of the cavity 3028. In some embodiments, the struts 3061 or portions thereof may be located partially or entirely within the foam body 3002, e.g. within the proximal face 3008 and/or sidewall 3014.
The struts 3061 are elongated structural members. The struts 3061 may have rectangular, circular or other shaped cross-sections. In some embodiments, the struts 3061 have a cross-section, e.g. rectangular, with a width that is greater than a thickness such that the struts 3061 are stiffer in one direction compared to another direction. This width may be in the lateral direction or a direction generally perpendicular to the longitudinal axis of the device 3000 when the device 3000 is in the expanded configuration, with the thickness perpendicular to the width. The struts 3061 may be less stiff in the direction of flexing or bending, for example to facilitate contraction and expansion of the device 3000 in the delivery and expanded configurations. The struts 3061 may be elongated pins. The struts 3061 may extend from the hub 3050, for example, and incline radially outwardly in the distal direction from the hub 3050. The struts 3061 may be attached inside, outside, and/or at the end of the sidewall of the hub 3050. The struts 3061 may be separate parts that are then attached to the hub 3050, for example welding, bonding, fastening, other suitable means, or combinations thereof. In some embodiments, some or all of the struts 3061 and the hub 3050 may be a single, continuous structure formed from the same raw material such as a laser cut hypotube. Some or all of the struts 3061 may be attached, e.g. with sutures as described herein, to the body 3002 and/or the cover 3100 at one or more attachment locations.
Each recapture strut 3061 may include an inner curved portion 3062 connected to a distal end of the hub 3050, a middle straight portion 3064, and/or an outer curved portion 3066 (for clarity, only some of the portions 3062, 3064, 3066 are labelled in the figures). In the deployed configuration, the inner curved portion 3062 extends from the hub 3050 primarily in a distal direction and then curves to face more outwardly radially. The middle straight portion 3064 extends from the inner curved portion 3062 primarily radially but also slightly distally. The outer curved portion 3066 extends from the middle straight portion 3064 primarily in the radial direction and then curves toward the distal direction. The portions may have different shapes in the delivery configuration inside a delivery catheter. In the delivery configuration, the portions may extend primarily distally. The portions may then take the deployed configuration as described upon deployment from the delivery catheter. In some embodiments, the struts 3061 may include fewer or more than the portions 3062, 3064, 3066.
The device 3000 may include ten of the proximal recapture struts 3061. Such configuration may accompany a device 3000 having a foam body 3002 with an outer diameter of 27 mm in the free, unconstrained state. Such configuration may accompany a device 3000 having a foam body 3002 with an outer diameter of 35 mm in the free, unconstrained state. In some embodiments, the device 3000 may have from about two to about thirty, from about four to about twenty, from about six to about eighteen, from about eight to about sixteen, from about ten to about fourteen, or other numbers of struts 3061.
In the deployed configuration, each strut 3061 may extend radially outward and distally at an angle to the axis. This angle, measured relative to a portion of the axis that extends distally from the device 3000, may be from about 60° to about 89.9°, from about 65° to about 88.5°, from about 70° to about 85°, from about 72.5° to about 82.5°, from about 75° to about 80°, or other angular amounts. This angle may be much smaller when the device 3000 is in the delivery catheter. The struts 3061 may bend or flex when transitioning between, or when positioned in, the delivery and expanded configurations. The struts 3061 may bend or flex at the inner curved portion 3062, the middle straight portion 3064, and/or the outer curved portion 3066.
The proximal end 3042 of the frame 3040, such as the proximal face 3060, may therefore have a conical shape in the expanded configuration. The conical proximal face 3060 may facilitate with recapture of the device 3000 back into the delivery catheter. For example, the orientation of the struts 3061 inclining distally and radially outward from the hub 3050 in the expanded configuration provides an advantageous conical shape to the proximal face 3008 such that distal advance of the delivery sheath over the device 3000 will bias the struts 3061 inward and cause the device 3000 to stow back toward the delivery configuration and size for retrieval within the catheter.
The proximal face 3060 foreshortens considerably upon expansion of the device 3000 relative to the delivery configuration. “Foreshortening” here refers to the difference in axial length of the proximal face 3060 between the reduced delivery configuration and the expanded configuration (expanded either freely or as implanted). This length may be measured axially from the distal or proximal end of the hub 3050 to the distal ends of the outer curved portions 3066 of the recapture struts 3061. The proximal face 3060 may foreshorten by 50%, 60%, 70%, 80%, 90% or more. The proximal face 3060 has significantly more foreshortening upon expansion than the tubular body 3080, the latter of which may be referred to as the “working length” or “landing zone.” The landing zone is further described with respect to the tubular body 3080 herein.
As shown, the struts 3061 are angularly spaced about the axis in even angular increments. That is, looking at the frame 3040 from the distal or proximal end, the angles between the struts may be equal. In some embodiment, the struts 3061 may not be evenly angularly spaced about the axis as described. The struts 3061 may or may not be symmetrically disposed about the axis or about a plane that includes the axis.
In some embodiments, portions of the frame 3040 may be at various distances from the proximal end of the foam body 3002, such as the proximal end wall having the proximal face 3008. As shown in
The frame 3040 includes a tubular body 3080. The body 3080 provides a mechanical base structure for the device 3000, as further described. The tubular body 3080 is attached to a distal end of the proximal face 3060 of the frame 3040. The tubular body 3080 extends to the distal end 3044 of the frame 3040. The tubular body 3080 is attached at a proximal end to the outer curved portions 3066 of the recapture struts 3061, as further described. The tubular body 3080 may be attached to other portions of the recapture struts 3061. The tubular body 3080 of the frame 3040 may be attached to the body 3002 and/or the cover 3100, e.g. with sutures as described herein, at one or more attachment locations, as further described. The tubular body 3080 may be located within the cavity 3028. In some embodiments, the tubular body 3080 may be located partially or entirely within the foam body 3002, e.g. within the sidewall 3014.
The tubular body 3080 includes a series of proximal struts 3082 and distal struts 3086 (for clarity, only some of the struts 3082, 3086 are labelled in the figures). The proximal struts 3082 and/or distal struts 3086 may have rectangular, circular or other shaped cross-sections. In some embodiments, the proximal struts 3082 and/or distal struts 3086 have a cross-section, e.g. rectangular, with a width that is greater than a thickness, or vice versa, such that the struts 3061 are stiffer in one direction compared to another direction. The struts 3061 may be less stiff in the direction of flexing or bending, for example to facilitate contraction and expansion of the device 3000 in the delivery and expanded configurations. Proximal ends of pairs of adjacent proximal struts 3082 join at proximal apexes 3084. Each proximal strut 3082 is connected at a respective proximal apex 3084 to a respective outer curved portion 3066 of one of the recapture struts 3061. Each distal end of the proximal struts 3082 connects to a distal end of an adjacent proximal strut 3082 and to proximal ends of two distal struts 3086 at an intermediate vertex 3087. Pairs of adjacent distal struts 3086 extend distally to join at a respective distal apex 3088. A repeating pattern 3089, shown as a diamond shape, may be formed by adjacent pairs of proximal struts 3082 and adjacent pairs of distal struts 3086. Some or all of the proximal struts 3082 and/or distal struts 3086 may be attached, e.g. with sutures as described herein, to the body 3002 and/or the cover 3100 at one or more attachment locations. Some or all of the proximal struts 3082 and/or distal struts 3086 may be located within the cavity 3028. In some embodiments, some or all of the proximal struts 3082 and/or distal struts 3086 may be located partially or entirely within the foam body 3002, e.g. within the sidewall 3014.
There are the same number of proximal apexes 3084 as distal apexes 3088. As shown, there are eleven proximal apexes 3084 and eleven distal apexes 3088. The number of proximal and distal apexes 3084, 3088 may each be at least seven, at least eight, at least nine, at least ten, at least eleven, at least twelve, at least thirteen, at least fourteen, at least fifteen, at least sixteen, or fewer or more apexes. In some embodiments, there may not be the same number of proximal apexes 3084 as distal apexes 3088. In some embodiments, there may be more than one row of the pattern, e.g. diamond pattern, formed by the proximal struts 3082 and distal struts 3086. There may be two, three, four or more rows of the pattern. Some or all of the proximal apexes 3084 and/or distal apexes 3088 may be attached, e.g. with sutures as described herein, to the body 3002 and/or the cover 3100 at one or more attachment locations.
The body 3080 may be tubular, e.g. cylindrical or generally cylindrical, in the expanded configuration. The tubular body 3080 may be cylindrical, rounded, segmented, polygonal, tube-like, other shapes, or combinations thereof, all of which are subsumed non-exhaustively under the category “tubular.” The tubular shape is formed by the proximal struts 3082 and distal struts 3086 in the expanded configuration. The tubular shape may also be formed by the outer curve portions 3066 of the recapture struts 3061 in the expanded configuration. The tubular shape may also be formed by the foam body 3002 exerting an outward radial force on the frame 3040. The frame 3040 may therefore have a proximal conical section and a cylindrical working length. In some embodiments, the body 3080 may be conical or frustoconical, for example where the distal end is wider than the proximal end or vice versa.
The tubular body 3080 may be referred to as a “landing zone,” as described. This landing zone may refer to the axial length of the body 3080, from a distal-most end to a proximal-most end at the transition to recapture struts 3061, in the expanded configuration. The landing zone may have an axial length as measured from the proximal apex 3084 to the distal apex 3088. The length of the landing zone may be 10 mm or about 10 mm. The landing zone may have a length from about 5 mm to about 15 mm, from about 6 mm to about 14 mm, from about 7 mm to about 13 mm, from about 8 mm to about 12 mm, from about 9 mm to about 11 mm, or other lengths. The tubular body 3080 may foreshorten slightly upon expansion of the device 3000 relative to the delivery configuration. The tubular body 3080 has significantly less foreshortening upon expansion than the length of the proximal face 3060. The tubular body 3080 may foreshorten by no more than about 5%, 10%, 15%, 20% or 30%.
The frame 3040 self-expands upon delivery from the sheath. The proximal face 3060 and the tubular body 3080 will self-expand. Upon expansion, the radially outward portions of the tubular body 3080 will contact and compress the foam body 3002 against tissue of the LAA wall. The tubular body 3080, for example the proximal struts 3082 and distal struts 3086, will contact the inner surface 3018 of the sidewall 3014 and press against the sidewall 3014 so that the outer surface 3016 of the sidewall 3014 contacts and compresses against the LAA wall.
When compressed against the LAA wall, the foam body 3002 provides a larger “footprint” than the skeletal frame 3040 components and forms a complete seal. Thus, the sidewall 3014 acts as a force dissipation layer, spreading radial force out from the struts 3082, 3086 of the frame 3040 over a larger area than just the area of the individual struts 3082, 3086 (e.g. a larger area than just the area of the radially outer surfaces of the struts 3082, 3086). The use of the foam material in the body 3002 and the thickness of that foam, such as 2.5 mm, provide advantages in this regard over devices with thinner and less resilient materials than foam. For example, thin fabrics or similar materials that are pressed against the LAA wall with a skeletal frame will not spread the radial force out, and may even sag or otherwise bend, creating gaps and an unsealed portion of the LAA wall. The foam body 3002 as described herein will take the shape of the LAA wall to create a complete circumferential seal and will also spread out the radial forces from the frame 3040 to create a stronger seal and retention with the foam body 3002.
Further, the device 3000 described herein with the compressible body 3002 allows for a structural frame 3040 that is compliant due to the smaller required radial force from the frame 3040. For example, existing devices with a non-compressible fabric material will have a less effective seal, and so the structural elements of those devices must provide larger radial forces to compensate and ensure an effective seal, resulting in a less compliant device. In contrast, the current device 3000 provides advantages in this regard by having the compressible foam body 3002, allowing for among other things smaller radial forces from, and thus better compliance of, the frame 3040, while still providing an effective seal. This structural configuration has a cascading effect in terms of performance advantages. For instance, the compliance of the device 3000 allows for delivery off-axis while still providing an effective seal, among other advantages as further described herein.
The frame 3040 includes a series of proximal anchors 3090. Each proximal anchor 3090 extends from a respective intermediate vertex 3087. The proximal anchors 3090 may extend from other portions of the tubular body 3080. As shown, in the deployed configuration, the proximal anchors 3090 extend from the tubular body 3080 radially and proximally. The proximal anchors 3090 may extend into an adjacent region of the sidewall 3014. The proximal anchors 3090 may extend through the outer surface 3016 of the sidewall 3014 to penetrate tissue adjacent the device 3000.
The frame 3040 includes a series of distal anchors 3094. Each distal anchor 3094 extends from a respective distal apex 3088. The distal anchors 3094 may extend from other portions of the tubular body 3080. As shown, in the deployed configuration, the distal anchors 3094 extend from the tubular body 3080 radially and proximally. The distal anchors 3094 may extend into an adjacent region of the sidewall 3014. The distal anchors 3094 may extend through the outer surface 3016 of the sidewall 3014 to penetrate tissue adjacent the device 3000. The anchors 3090, 3094 may incline radially outward in a proximal direction to engage the tissue to resist proximal movement of the device 3000.
The anchors 3090, 3094 are elongated structural members. The tips of the anchors 3090, 3094 may be sharpened to facilitate tissue engagement and penetration. The anchors 3090, 3094 may be straight, extending generally along a local axis thereof. The anchors 3090, 3094 may have a curved or other non-straight proximal portion where they attach to the tubular body 3080. In some embodiments, the anchors 3090, 3094 or portions thereof may be non-straight, curved, rounded, segmented, other trajectories, or combinations thereof. In some embodiments, the tissue engaging tips may be curved. In some embodiments, the anchors 3090, 3094 may have engagement features extending radially away from the anchor 3090, 3094, such as barbs, hooks, or other features.
The cross-section of the anchors 3090, 3094 may be rectangular. In some embodiments, the cross-section may be circular, rounded, non-rounded, square, rectangular, polygonal, other shapes, or combinations thereof. The cross-sections may or may not be uniform along the length of the anchor 3090, 3094. The anchors 3090, 3094 may be about 0.006″ thick and about 0.008″ wide. The anchors 3090, 3094 may range from about 0.003″ to about 0.009″ in thickness and from about 0.003″ to about 0.015″ in width. The cross-section of the anchors 3090, 3094 may reduce in size, for example taper, toward the distal tip.
In some embodiments, the anchors 3090, 3094 in the deployed configuration are inclined at an incline angle of about 30° relative to a portion of the central axis that extends proximally from the device 3000. This incline angle may be from about 10 degrees to about 50°, from about 15° to about 45°, from about 20° to about 40°, from about 25° to about 35°, or about 30°. This incline angle of the anchors 3090, 3094 in the delivery configuration may be smaller than in the deployed configuration. The anchors 3090, 3094 may have the angle B, as shown in and described with respect to
The anchors 3090, 3094 may have various lengths. The length of the anchor 3090, 3094 is measured from a proximal end that connects to the tubular body 3080 to a distal tissue engaging tip of the anchor. In some embodiments, the length of the anchors 3090, 3094 may be from about 0.5 mm to about 10 mm, from about 1 mm to about 9 mm, from about 2 mm to about 8 mm, from about 3 mm to about 7 mm, from about 4 mm to about 6 mm, about 5 mm, or other greater or lesser lengths. In some embodiments, the anchors 3090, 3094 are 5 mm long. In some embodiments, the anchors 3090, 3094 are about 5 mm long. In some embodiments, the anchors 3090, 3094 have a length of at least 2.5 mm, at least 3 mm, at least 3.5 mm, at least 4 mm, at least 4.5 mm, at least 5 mm or more. The anchors 3090, 3094 may each be the same or similar length. In some embodiments, the anchors 3090, 3094 may not be the same length. In some embodiments, some or all of the proximal anchors 3090 may have lengths that are less than or greater than some or all of the lengths of the distal anchors 3094. The anchors 3090, 3094 may have the length L, as shown in and described with respect to
In the expanded configuration, the anchors 3090, 3094 extend for a length outside of the uncompressed sidewall 3014. This length of the anchor 3090, 3094 is measured along a local longitudinal axis of the anchor from the outer surface 3016 of the body 3002 to the distal tip of the anchor. The anchors 3090, 3094 may extend through the sidewall 3014 and/or the cover 3100, and then be trimmed so that the anchors 3090, 3094 extend beyond the sidewall 3014 and/or cover 3100 by the desired length. In a free, unconstrained state, the anchors 3090, 3094 extend about 0.5 mm beyond the outer surface 3016 of the sidewall 3014. In some embodiments, in the free, unconstrained state, the anchors 3090, 3094 extend beyond the outer surface 3016 of the sidewall 3014 for a length of from about 0.1 mm to about 1.5 mm, from about 0.2 mm to about 1.25 mm, from about 0.3 mm to about 1.0 mm, from about 0.4 mm to about 0.8 mm, from about 5 mm to about 0.6 mm, or other greater or lesser lengths. In a compressed state, such as in the delivery configuration or after implantation, the anchors 3090, 3094 extend about 1.0 mm beyond the outer surface 3016 of the sidewall 3014. In some embodiments, in the compressed state, the anchors 3090, 3094 extend beyond the outer surface 3016 of the sidewall 3014 for a length of from about 0.25 mm to about 2.5 mm, from about 0.5 mm to about 2 mm, from about 0.75 mm to about 1.5 mm, from about 0.875 mm to 1.125 mm, or other greater or lesser lengths.
The geometry of the anchors 3090, 3094 provides several advantages. For example, the relatively long length allows for flexibility of the anchors 3090, 3094. This provides for potentially less trauma to the LAA tissue should the device 3000 need to be unanchored and/or retrieved. The anchors 3090, 3094 are less susceptible to loss of strength with off-axis orientation within the LAA. Further, the anchors 3090, 3094 provide high resistance to pull out. For instance, the device 3000 may provide at least about 0.5 lb-force of dislodgment resistance from the LAA. Such pullout tests may be simulated with in vitro or benchtop models, as further described below.
The anchors 3090, 3094 in the illustrated embodiment are located in two circumferential rows. One row is located proximal to the other distal row. Each row has ten anchors each. This configuration may be incorporated, for example, in the device 3000 having a foam body 3002 with a free, unconstrained outer diameter of 27 mm. Each row may have fourteen anchors each. This configuration may be incorporated, for example, in the device 3000 having a foam body 3002 with a free, unconstrained outer diameter of 35 mm. In some embodiments, a single row of anchors 3090, 3094 may have from two to twenty-four, from four to twenty-two, from five to twenty, from six to eighteen, from seven to sixteen, from eight to fifteen, from nine to fourteen, from ten to thirteen anchors, or greater or fewer amounts of anchors 3090 or 3094. In some embodiments, there may only be one row or greater than two rows of anchors. The anchors 3090, 3094 may be spaced circumferentially in a single row.
In embodiments with multiple rows of anchors 3090, 3094, the rows may be circumferentially offset, as shown. That is, as viewed from the proximal or distal end of the device 3000, the anchors 3090, 3094 are angularly spaced apart from each other about the axis. The anchors 3090, 3094 may not be circumferentially offset, e.g. they may be evenly angularly spaced when viewed as described. The anchors 3090, 3094 are located axially at or near a middle portion of the sidewall 3014. The anchors 3090, 3094 may be located such that the tips of the anchors 3090, 3094 extend to adjacent tissue at a middle portion of the sidewall 3014. The offset and middle locations of the anchors 3090, 3094 may ensure engagement with the LAA tissue distal to the ostium. Having the anchors 3090, 3094 located at the largest width, increases the stability of the device 3000. With a cylindrical or generally cylindrical shaped device 3000, the anchors 3090, 3094 effectively sit on the largest diameter of the device 3000. The cylindrical shape provides advantages over typical LAA occluders which taper distally thus decreasing implant stability and locating the anchors on a smaller diameter than the ostial diameter of the occluding surface. In addition to adding stability, the cylindrical shape of the device 3000 along the axial length helps with dislodgement resistance by allowing the anchors 3090, 3094 to be placed on the largest diameter section of the device 3000. In some embodiments, the anchors 3090, 3094 may be located proximal, distal, or centrally along the length of the frame body 3080. In some embodiments, the anchors 3090, 3094 may not be offset and/or may not be angularly evenly spaced.
The anchors 3090, 3094 may provide advantageous flexibility, as demonstrated by pullout tests and in comparison to existing devices. For example, the device 3000 was tested to determine the force required to dislodge the device 3000 from a simulated tissue model by pulling the device 3000 proximally outward from the model. A low durometer silicone tube with a circular inner diameter (ID) was used as the model. For the device 3000 having a foam body 3002 with a 27 mm outer diameter in a free unconstrained state, tubes with ID's of 16.5 mm, 21 mm and 25 mm were tested. The pullout forces for existing devices drop off significantly going up to a 21 mm model, whereas the forces for the device 3000 drop only slightly.
In the largest diameter (25 mm) model, where there is not a lot of interference in the fit, the forces for the existing devices approach zero as the device does not engage the model wall because the anchors are sitting at a smaller diameter on a trailing edge of the device. The device 3000 consistently resists dislodgment with about 0.7 lbs of force. Since there is very little friction resisting pullout, that force is almost entirely resisted by the anchors 3090, 3094. When examining failure modes, all devices eventually begin to slide out of the model. Upon failure, the anchors 3090, 3094 fold backward or sideways before slipping starts. Assuming 0.7 lbs force is required to cause all twenty anchors 3090, 3094 to fold backward, then the force per anchors is estimated to be about 0.035 lbs.
The frame 3040 may be laser cut. The tubular body 3080 may be laser cut from a single tube. The body 3080 may be cut from a tube having a thickness from about 0.002″ to about 0.014″, or about 0.008″. The tube may have an outer diameter (OD) from about 0.05″ to about 0.30″. The tube may have an outer diameter (OD) of 0.124″ for the 27 mm device 3000 (i.e. the embodiment of the device 3000 having a foam body 3002 with an OD of 27 mm in the unconstrained, free state). The tube may have an OD of 0.163″ for the 35 mm device 3000 (i.e. the embodiment of the device 3000 having a foam body 3002 with an OD of 35 mm in the unconstrained, free state).
In some embodiments, the body 3080 is laser cut from a superelastic nitinol tube, however, numerous other biocompatible metallic materials can be utilized such as shape memory Nitinol, stainless steel, MP35N, or Elgiloy. The frame 3040 is self-expandable. In some embodiments, a balloon-expandable frame 3040 could be utilized. Additionally, the body 3080 could be fabricated from drawn wire as opposed to being laser cut from a tube.
As shown, an embodiment of the device 3000 includes the frame 3040 having ten proximal recapture struts 3061 and twenty total anchors 3090, 3094, with the foam body 3002 having an outer diameter of 27 mm. In some embodiments, the device 3000 may include the frame 3040 having fourteen proximal recapture struts 3061 and twenty-eight total anchors 3090, 3094, with the foam body 3002 having an outer diameter of 35 mm.
In one embodiment, the frame 3040 includes a proximal hub 3050, tether pin 3051, front face with ten or fourteen recapture struts 3061, a diamond pattern cylindrical body 3080, and twenty or twenty-eight anchors 3090, 3094. The frame proximal face 3060 supports recapture, the frame body 3080 supports the foam cylinder body 3002, and the anchors 3090, 3094 located on the cylinder provide resistance to embolization.
The design of the device 3000 provides numerous advantages, some of which have been described. As further example, the frame 3040 provides many advantages, including but not limited to: 1) implant radial stiffness/compliance—the frame 3040 provides enhanced radial stiffness while still being sufficiently compliant to allow for off-axis implantation, recapture, etc.; 2) dislodgement resistance—the frame 3040 provides for high pullout strength, as described; 3) transcatheter delivery—the frame 3040 can be compressed into a delivery catheter and then fully expand when delivered; 4) recapture—the frame 3040 allows for recapture/retrieval into the delivery catheter after deployment or even after implantation in the LAA; and 5) mechanical integrity—the frame 3040 has acute and long term structural integrity, for example the ability to withstand loading into the delivery catheter, deployment from the catheter, and cyclic loading/fatigue. The frame 3040 also provides a conformable structure to enable the foam body 3002 to compress against the LAA tissue to facilitate sealing and anchoring with minimal compression (oversizing). The resulting compliance of the frame 3040 provides better anchoring than existing solutions, as described.
As further example, the device 3000 seals against irregularly shaped LAA ostia and necks. For instance, a combination of a Nitinol frame 3040 with a foam body 3002 having a coating of PTFE and cover 3100 of ePTFE contribute to ability of the device 3000 to conform to the anatomy and seal against irregular projections and shapes, while providing a smooth thromboresistent LA surface.
As further example, the device 3000 provides for controlled & safe delivery. The design of the combined frame 3040 and foam body 3002 facilitates delivery in a controlled fashion by slowing the speed of expansion. The bumper 3026 acts as an atraumatic leading edge portion when delivering the implant into the LAA mitigating the risk of injury. The user has the ability to recapture and redeploy the device 3000, if necessary. A flexible tether 3240 attachment, as further described, from the delivery catheter to the device 3000 permits the device 3000 to sit tension free immediately following implantation so the user can ensure final appropriate positioning prior to release of the device 3000.
As further example, the device 3000 provides for simplified placement. The foam-covered cylindrical design makes alignment of the device 3000 with the central axis of the LAA during delivery non-critical (by allowing deployment up to, for example, 45 degrees off-axis), which is designed to simplify the implantation procedure, as further described.
As further example, the device 3000 provides for simple sizing. The foam and frame design contributes to the ability to need only two diameters (e.g., 27 mm and 35 mm) to seal the range of expected LAA configurations and diameters (e.g. targeting LAA diameters of 16 to 33 mm). The conformability of the foam and frame allow the 20 mm long implant to fit into LAA's as short as 10 mm deep. The short landing zone requirement (LAA depth) of the device 3000, combined with the need for only two implant diameters, enables treatment of a wide range of LAA anatomies with minimal need for burdensome echo and CT sizing. The conforming nature of the implant is key to facilitating a simple to use product platform that is adaptable to a variety of anatomic structures.
As further example, the device 3000 provides thromboresistant materials and design. The removable tether leaves a smooth, metal-free surface in the LA. Thromboresistant materials (PTFE-coated foam and an ePTFE cover) create a smooth LA face (no metal attachment connection) to reduce anticoagulation needs, enhance thromboresistance, and encourage endothelialization.
As further example, the device 3000 provides thin, low profile anchors 3090, 3094 around the midpoint of the device 3000 to provide secure yet atraumatic anchoring.
4. Distal Bumper
The foam body 3002 has a distal bumper 3026. The bumper 3026 may be a foam distal region of the body 3002, such as a distal portion of the sidewall 3014. The bumper 3026 may be a portion of the foam body 3002 that extends beyond the distal end 3044 of the frame 3040. The bumper 3026 may extend beyond the distal end 3044 of the frame 3040 in the delivery configuration and in the deployed configuration. The body 3002 may be attached to the frame 3040 in various locations such that the body 3002 may stretch in some embodiments, for example in the delivery configuration, to ensure the bumper 3026 extends beyond the frame 3040 upon initially retracting the sheath during delivery.
The device 3000 can conform both in length and diameter due to conformability of both the foam body 3002 and the frame 3040. This allows for the device 3000 to accommodate most patient LAA anatomies with only a couple or few different sizes of the device 3000, such as 27 mm and 35 mm outer diameter body 3002 as described herein, and one length, such as 20 mm. The frame 3040 may thus be shorter than the foam body 3002, resulting in some embodiments in about 5 mm of foam bumper 3026 distal to the distal-most end of the frame 3040. The distal bumper 3026 acts as an atraumatic tip during delivery of the device 3000 and can be compressed following implantation to allow the device 3000 to conform to appendages with a depth (landing zone) as short as 10 mm. This ability to conform both in length and diameter is due to the conformability of both the foam body 3002 and the frame 3040.
The length of the bumper 3026 may be measured axially from the distal-most end of the frame 3040 to the distal surface 3022 of the body 3002. For example, the bumper 3026 may extend from the distal apexes 3088 to the distal surface 3022. The bumper 3026 may have a length of 5 mm or about 5 mm. The bumper 3026 may have a length of about 1 mm, 2 mm, 3 mm, 4 mm, 5 mm, 6 mm, 7 mm, 8 mm, 9 mm or more. The bumper 3026 may have a length from about 2.5 mm to about 7.5 mm, from about 3 mm to about 7 mm, from about 3.5 mm to about 6.5 mm, from about 4 mm to about 6 mm, from about 4.5 mm to about 5.5 mm.
In some embodiments, the bumper 3026 may fold in response to axial and/or radial compression of the device 3000. The bumper 3026 may fold inward, for example radially inward. The folds may be in the axial or approximately in the axial direction. The folds may be circumferential or approximately in the circumferential direction. The folds may be combinations of the radial and circumferential directions, or angled with respect thereto. The folding of the bumper 3026 is further discussed herein, for example in the section “Device Compliance.”
5. Cap & Pin
As shown in
The cap 3180 is formed from titanium and the pin 3051 is formed from Nitinol or superelastic Nitinol. In some embodiments, the cap 3180 and/or pin 3051 may be formed from other materials, for example numerous biocompatible metallic or polymeric materials such as shape memory Nitinol, stainless steel, MP35N, Elgiloy, polycarbonate, polysulfone, polyether ether keytone (PEEK), or polymethyl methylacrylate (PMMA) or other materials.
The cap 3180 and pin 3051 facilitate attachment to the tether 3240. The cap 3180 and pin 3051 also mitigate damage to the foam body 3002 during recapture of the device 3000. The cap 3180 also creates an atraumatic surface for the hub 3050 of the frame 3040. For example, the cap 3180 may prevent the hub 3050 from cutting through the foam body 3002 as the device 3000 is collapsed into an access sheath. Without the cap 3180, the sharp edges of the hub 3050 may shear through the foam body 3002 during recapture of the device 3000 into the access sheath.
6. Loading System
In some embodiments, the conical portion 3212 of the loading tool 3210 has a chamfered distal edge of approximately 45°-75° (degrees), preferably 60°. In some embodiments, the conical portion 3212 has a distal inner diameter (ID) greater than the outer diameter (OD) of the device 3000 and an angle A of ideally between 15° and 25°, and in one implementation about 20°, to appropriately collapse the anchors 3090, 3094 which may protrude off the foam body 3002 surface at an angle of 30° or about 30°. The distal opening of the conical portion 3210, for example the diameter or greatest width, may be greater than the proximal opening of the conical portion 3210, for example the diameter or greatest width, that couples with the cylindrical portion 3214. The cylindrical portion 3214 may have an opening, for example diameter or greatest width, that is smaller than the distal opening of the conical portion 3210 and/or the same or similar size as the opening at the proximal end of the conical portion 3210.
The decreasing width, for example gradual taper, of the loading tool 3210 ensures, for example, that the frame 3040 folds evenly without crossovers or extra strain. The angled conical portion 3212 may ensure that the anchors 3090, 3094 fold or rotate proximally and not distally. The sidewall of the conical portion 3212 may extend at a “total” angle A as measured between two opposite portions of the sidewall, as shown in
The loading tool 3210 may be smooth or generally smooth on its inner surface or surfaces. Inner surfaces 3211, 3215 of the conical portion 3212 and/or cylindrical portion 3214 may be smooth or generally smooth. In some embodiments, these inner surfaces 3211, 3215 or portions thereof may not be smooth. In some embodiments, these inner surfaces 3211, 3215 or portions thereof may be smooth, non-smooth, rough, etched, scored, grooved, have varying degrees of roughness or smoothness, other features, or combinations thereof.
In one example, the tool 3210 may be used by positioning a proximal end of a loading body, such as the tool 3210, adjacent the distal end 3222 of the delivery catheter 3220. The loading body may have a sidewall defining a channel therethrough with the distal opening 3213 at a distal end that is larger than a proximal opening at a proximal end. The left atrial appendage occlusion device 3000 may be advanced proximally through the loading body to thereby radially compress the device 3000. The retracting step may comprise pulling the tether 3240 proximally through the delivery catheter 3220. The device may then be received into the distal end 3222 of the delivery catheter 3220. The device 3000 may be radially compressed within the delivery catheter 3220 having an outer diameter of no more than fifteen French. In some embodiments, the device 3000 may be radially compressed within the delivery catheter 3220 having an outer diameter of no more than ten, eleven, twelve, thirteen, fourteen, fifteen, sixteen, seventeen, eighteen, nineteen, or twenty French. The proximal end of the loading tool 3210 may have an inner diameter configured to provide an interference fit with the distal end 3222 of the delivery catheter 3220. The proximal end of the loading tool, 3210, such as the cylindrical portion 3214, may have an inner diameter slightly larger than the outer diameter of the delivery catheter 3220, for instance slightly larger than 5 mm for a delivery catheter 3220 having an outer diameter of fifteen French. The device 3000 may compress radially to a compressed width in a constrained state that is less than fifty, forty, thirty, twenty, ten, and/or five percent of a radial uncompressed width of the device in an unconstrained state. The radial widths here may be measured perpendicularly to a longitudinal axis of the device 3000, such as defined by the tubular foam body 3002.
The loading tool 3210 may be formed of a material that is biocompatible, strong, transparent and can be molded smooth to minimize friction, such as polycarbonate. In some embodiments, the loading tool 3210 could be formed from hard plastics like Delrin, UHMWPE, Ultem®, polyetherimide, acrylic, metals like stainless steel, aluminum, other materials, or combinations thereof. In some embodiments, the loading tool 3210 may have one or more coatings. Such coating may be applied to reduce friction and therefore loading forces. The coating may be silicone, hydrophilics, various oils, other suitable coatings, or combinations thereof.
7. Delivery System
To deliver the device 3000 to the LAA, an access sheath is placed across the interatrial septum into the LAA through which the delivery catheter 3220, containing the device 3000, is placed. The device 3000 is loaded into the distal end 3222 of the delivery catheter 3220 using the loading tool 3210, either at the point of manufacture or at the treatment site. To load the implant device 3000, the pusher 3230 and tether 3240 are pulled proximally, collapsing the implant device 3000 as it enters the distal tip of the delivery catheter 3220. Once the loaded delivery catheter 3220 is placed through the sheath into the LAA, the pusher 3230, such as a catheter or rod, is held axially steady as the delivery catheter 3220 and access sheath are simultaneously retracted proximally, deploying the implant device 3000.
The tether 3220 passes from the proximal end of the delivery catheter 3220, through an opening 3221 of the catheter pusher 3230, around the implant tether pin 3051, and back through the delivery catheter 3220. When both ends of the tether 3240 (held together by the restraint 3246 at the proximal end of the catheter) are pulled, the device 3000 is pulled into the delivery catheter 3220. Once the device 3000 is properly placed in the anatomy, one end of the ends 3242, 3244 of the tether 3240 is cut and the entire tether 3240 can be removed from the system by pulling proximally on the uncut end and sliding the cut end distally into the system and around the pin 3051, disengaging from the pin 3051. The distal end 3232 of the pusher 3230 and/or the distal end 3222 of the delivery catheter 3220 may contact, for example push against, the proximal end of the device 3000, such as in the relative locations shown in
In some embodiments, the delivery system 3201 may include other features. For example, the delivery catheter 3220 may include an injection lumen. The injection lumen may allow for injecting a radiopaque dye distal to the device 3000 following implantation to check for leaks using fluoroscopy.
8. Tether Release System
The system 3201 as shown in
The tether 3240 may extend from a proximal end of the pusher 3230, through the opening 3221 of the pusher 3230, wrap around the pin 3051, and extend proximally back through the opening 3221 of the pusher 3230 and out the proximal end of the pusher 3230, as described with respect to
The system 3201 may facilitate removal of the tether 3240 while the pusher catheter 3230 is in contact with the device 3000. Such contact may assist, for example, with avoiding or reducing inadvertent dislodgement of the device 3000 from the LAA after implantation and anchoring. For instance, during release of the tether 3240, the pusher 3230 may have the positioning relative to the device 3000 as shown in
The lock 3402 includes a proximal end 3404 and a distal end 3406. An opening 3408 defined by a sidewall 3409 extends through the lock 3402 from the proximal end 3404 to the distal end 3406. The tube 3420 extends through the opening 3408 at the proximal end 3404 of the lock 3402 and to the distal end 3406 of the opening 3408. The sidewall 3409 of the lock 3402 has a first groove 3410 extending longitudinally from the proximal end 3404 to the distal end 3406 and extending radially partially through the thickness of the sidewall 3409. The sidewall 3409 of the lock 3402 has a second groove 3412 extending longitudinally from the proximal end 3404 partially along the sidewall 3409 toward the distal end 3406 and radially partially through the thickness of the sidewall 3409.
The tether 3240 includes a first end 3243 and the second end 3245. The tether 3240 extends distally from the first end 3243 within the opening 3426 of the tube 3240 and out through the distal end 3424 of the tube 3420 to the cap 3180. The tether 3240 extends distally into the opening 3188 of the cap 3180 and around the pin 3051 and back in the proximal direction. The tether 3240 then extends proximally into the first groove 3410 of the lock 3402, around the proximal end 3404 of the lock 3402, and then distally into and through the second groove 3412. The tether 3240 terminates at the second end 3245 in a knot 3247.
In use, the knot 3247 may be secured due to the relative location of the lock 3402 and the pusher catheter 3230 inside the delivery catheter 3220. The knot 3247 may be prevented from advancing distally due to the inner diameter of the distal end of the pusher 3230 fitting tightly about the outer diameter of the lock 3402. The grooves 3410 and/or 3412 in the lock 3402 may hold the tether 3240 in an orientation that prevents the tether 3240 from slipping (e.g., if pulled hard enough) when the lock 3402 is engaged in the pusher 3230. The pusher 3230 may be advanced distally to expose the lock 3402, for example the full length f the lock 3402, or portions thereof. When the proximal end of the tether 3240 is pulled proximally, the knot 3247 falls away from the second groove 3412, advances around the proximal end 3404 of the lock 3402, advances distally by falling away from the first groove 3410, into the cap 3180 and around the pin 3051, and then distally through the opening 3408 of the lock 3402 and can be retrieved with the pusher 3230. In some embodiments, the distal end of the lock 3402 may be located axially proximal to the distal end of the tube 3420, for example to contact the device 3000 with the tube 3420 to prevent proximal movement of the device 3000 after implantation, as described above.
9. Off-Axis Delivery and Deployment
The device 3000 may be deployed off-axis within an LAA while still providing a complete, stable, and atraumatic seal. In some embodiments, the device 3000 may be deployed at an angle of at least about 15° or 25° and in some embodiments as much as 35° or 45°, for example, relative to a central longitudinal LAA axis and still provide an effective seal. The LAA axis here is defined as the geometric center of the ostium to the LAA, and tracks the best fit geometric center of the LAA cavity.
This ability of the device 3000 to be deployed off-axis is due in part to the relatively thick, compressible foam body 3002 material, the compliant frame 3040 and the cylindrical shape of the device 3000 with the foam bumper 3026. The device 3000 is stable within the LAA despite having a length that is less than the diameter, or having L/D<1. As described, the length may be 20 mm for the device 3000 having an OD of both 27 mm and 35 mm. Thus not only is flexibility and simplicity allowed with manufacturing processes by having one length, but also stability and effectiveness of the device in use. Further, the axial compressibility of the bumper 3026, combined with the axially compliant frame 3040, allows a 20 mm long device 3000 to be placed within a 10 mm deep LAA, whereas existing LAA closure devices require longer landing zones, or at least landing zones equal to the size of the length of the metallic frame.
In some embodiments, the device 3000 may be configured to allow for sufficient flow of blood in case of accidental embolization, as described herein. Further, the device 3000 may be configured to allow for sufficient flow of blood even if the device 3000 embolizes and is misaligned with the direction of flow of blood. For example, the device 3000 may define a longitudinal axis, and the direction of flow of blood may define a flow axis. The device longitudinal axis may be at an angle with respect to the flow axis and still provide for a sufficient blood flow through the device 3000 should it embolize and lodge within the circulatory system of a patient. Thus, the capabilities of the device 3000 with regard to flow of blood through the device in case of embolization, or tests thereof with water under controlled conditions, as described herein for example with respect to the section “Proximal Cover,” may also apply to the device 3000 in such off-axis configurations or orientations with the circulatory system of the body. The device axis may be at an angle of five, ten, twenty, thirty or more degrees with respect to the flow axis and still provide sufficient flow of blood through the device 3000.
10. Anchor/Foam Interface
As described, the frame 3040 with anchors 3090, 3094 and foam body 3002 may have a variety of geometries, such as lengths, thicknesses, etc. This section discusses some particular embodiments of the frame 3040, in particular the anchors 3090, 3094, and the foam body 3002.
The interface 3500 includes a portion of the tubular body 3080 of the frame 3040, having the proximal strut 3082 and the distal strut 3086, as described in further detail herein, for example with respect to
As shown in
The anchor 3090 extends at an angle B relative to the proximal strut 3082. In some embodiments, the proximal strut 3082 as shown may be considered a projection of the proximal strut 3082 onto a vertical plane that intersects the longitudinal axis of the device 3000 and the anchor 3090. Thus the angle B may be relative to such plane and/or to the strut 3082. For simplicity, the angle B will be described relative to the strut 3082. The angle B may be 30° or about 30°. The angle B may be a variety of other angles or within ranges of angles, for example as described in further detail herein with respect to the anchors 3090, 3094 in the section “Compliant Frame.” The anchor further has a radial height H. The radial height H may be the radially outermost extent of the anchor 3090, such as the proximal tip 3091 of the anchor 3090. The length L and angle B may define the radial height H of the anchor 3090. The height H may be in a direction perpendicular to the longitudinal axis of the device 3000 (see, e.g.,
Further shown is the sidewall 3014 of the foam body 3002. The sidewall 3014 has a thickness T. The thickness T extends radially outward from the inner surface 3018 to the outer surface 3016 of the sidewall 3014. The thickness T may extend radially outward perpendicularly to the longitudinal axis of the device 3000. The thickness T may be equal to a distance from a radially outer portion of the frame 3040 to the outer surface 3016 of the sidewall 3014, for example where the inner surface 3018 of the sidewall 3014 contacts the outside of the frame struts 3082, 3086. The thickness T could be the thickness of the sidewall 3014 in an unconstrained configuration, a compressed configuration while inside the delivery catheter, or a compressed configuration after implantation within the LAA, as further described. The measurement of the thickness T of the sidewall 3014 may be in the same direction as the measurement of the height H of the anchor 3090. The thickness T of the sidewall 3014 may be 2.5 mm or about 2.5 mm. The thickness T of the sidewall 3014 may be other values or ranges of values, for example as described in further detail herein in the section “Compressible Foam Body.”
As shown in
As shown in
As shown in
The local compression of the sidewall 3014 may allow for the anchor 3090 to extend proximally and outwardly beyond the outer surface 3016 of the foam body 3002. As shown, the attachment 3001 may locally compress the thickness of the sidewall 3014 such that the proximal tip 3091 of the anchor 3090 extends at the angle B for the length L beyond the outer surface 3016 of the foam body 3002. The attachment 3001 may be located proximally to the anchor 3090 as shown, or in other location, such as distally to the anchor 3090, adjacent the base of the anchor 3090, farther proximally/distally from the base of the anchor 3090, etc. The attachment 3090 may be located and configured to allow for local compression of the sidewall 3014 to allow the tip 3091 of the anchor 3090 to extend beyond the outer surface 3016 of the foam that is located directly radially inwardly of the tip 3091 of the anchor 3090. In some embodiments the attachment 3001 may be located directly radially inwardly of the tip 3091 of the anchor 3090 (e.g., directly “below” the tip 3091 of the anchor 3090 as oriented in the figure). In some embodiments, there may be multiple attachments 3001 distributed axially along the frame 3040 and all contributing to a single local compression of the foam body 3002 about a particular one of the anchors 3090.
The foam sidewall 3014 may be compressed into the configuration shown in
Design of the anchor length may be based on a balance between longer length to provide flexibility to assist with removal, and shorter length for not penetrating through the LAA wall. The anchors 3090, 3094 may be flexible and capable of bending in the distal direction due to their length. The anchors 3090, 3094 are thus less likely to tear tissue during repositioning and therefore less traumatic. The anchors 3090, 3094 may be longer than other tissue engaging features of existing solutions for LAA occlusion. In some embodiments of the device 3000, the anchors 3090, 3094 are designed to be long enough to effectively anchor into the LAA wall. The foam body 3002 and corresponding thickness of the sidewall 3014 allows the anchors 3090, 3094 to have longer length. An advantage of making the anchors 3090, 3094 longer is to increase their flexibility, making them less damaging to tissue during removal and repositioning. However, anchors 3090, 3094 beyond a certain length may penetrate through the LAA wall, which is not desirable. The foam body 3002 and thickness thereof assists with preserving the advantageous longer length of the anchors 3090, 3094 while mitigating the risk of the anchors 3090, 3094 penetrating through the LAA wall. For example, the foam sidewall 3014 between the struts of the frame 3040 and the tips 3091 of the anchors 3090, 3094 limits how far the anchors 3090, 3094 will penetrate allowing for longer and therefore more flexible anchors 3090, 3094.
For example, with a 2.5 mm foam sidewall 3014 thickness, the anchors 3090, 3094 may be 2.5 mm in axial length and formed at an angle between 30-40 degrees, or 25-45 degrees, off the struts. In some embodiments, as discussed, when the frame 3040 is first placed into the foam body 3002 and anchors 3090, 3094 pierce into the foam sidewall 3014, the tips 3091 of the anchors 3090, 3094 may not extend all the way through the foam as the anchors 3090, 3094 may be radially too short. In some embodiments, the frame 3040 OD is about 24 mm and the foam body 3002, such as foam cup shape, has a sidewall 3014 with an ID of about 22 mm. So there may be an interference fit where the frame 3040 bulges into the foam sidewall 3014. With the anchor 3090, 3094 length and angle, the tips 3091 of the anchors 3090, 3094 are at about 27 mm diameter, which corresponds to just getting to the outer surface 3016 of the sidewall 3014. As discussed, the assembly may be attached by suturing the foam body 3002, and the frame 3040 (and in some locations the cover 3100) together at every anchor-frame interface location. This may cause a dimpling of the foam body 3002 local to the corresponding anchor 3090, 3094, thus exposing a length of the anchor 3090, 3094 at the outer surface 3016. The exposed length of the anchor 3090, 3094 may be a fraction of the total anchor 3090, 3094 length. Further, the anchor 3090, 3094 length and radial height of the foam sidewall 3014 surrounding the anchor 3090, 3094 can be adjusted to expose the desired amount of the anchor 3090, 3094.
In some embodiments, the tip 3091 may be exposed beyond the foam body 3002 when the foam is compressed, but the tip 3091 may be positioned within the foam, below the outer surface 3016, when the foam is uncompressed. Thus, with the foam uncompressed the tip 3091 may not be positioned radially outwardly relative to the outer surface 3016, but with the foam compressed the tip 3091 may be positioned radially outwardly relative to the adjacent portion of the outer surface 3016. Therefore, the tip 3091 may not be exposed with “H” less than “T” in the uncompressed configuration, and the tip 3091 may be exposed with “H” greater than “T” in the compressed configuration.
11. Device Compliance
The device 3000 is capable of conforming to the geometry of the LAA. The device 3000 is designed for compliance such that it can conform to the LAA and reduce or minimize remodeling of the LAA. For example, the device 3000 may be implanted into the LAA and after a period of time the ostium or opening of the LAA may have the same or similar profile as before implantation of the device 3000. Further, the device 3000 may exhibit such properties while conforming to extreme non-circular shapes, both at the opening of the LAA and within the LAA. A single size of the device 3000 may be used for all or a wide range of patients with varying geometries, due to the compliance and other advantages.
The lengths of the axes A1, A2 may have a variety of values or ranges of values. The minor axis A1 may be from about 5 mm to about 30 mm, from about 7.5 mm to about 20 mm, from about 10 mm to about 17.5 mm, from about 12 mm to about 15 mm, about 14 mm, or 14 mm. The major axis A2 may be from about 10 mm to about 40 mm, from about 15 mm to about 37 mm, from about 20 mm to about 35 mm, from about 22 mm to about 32 mm, from about 25 mm to about 30 mm, about 27 mm, or 27 mm.
The LAA may retain the same or similar original size and shape of the LAA immediately after implantation of the device 3000 and for a period of time thereafter. In some embodiments, the anatomical geometry, for example size and shape, of the LAA will still be the same or approximately the same after implantation of the device 3000 after a period of twenty four hours or more, seven days or more, thirty days or more, six months or more, one year or more, five years or more, or longer periods. A test construct having approximately the same geometry, stiffness, etc. may be constructed to confirm the minimal long-term changes in the construct due to the device 3000. A construct having an opening with a minor axis of about 14 mm and a major axis of about 27 mm and with a stiffness generally present in normal LAA ostiums of patients may have the same or similar size and shape after implanting the device 3000 for the aforementioned periods. The device 3000 may allow for the same or similar geometry along the length of the LAA, e.g. distal to the ostium 110, for these periods of time as well, as further described.
In one example use, the device 3000 may be configured to insert into a non-cylindrical opening, having a non-cylindrical profile, of a test body. The test body may be rigid such that the test body does not deform in response to the device 3000 being implanted therein. The test body may be formed of rigid plastics, metals, etc. The opening and profile may have a size and shape substantially similar to that of a native left atrial appendage. The device 3000 may expand radially within the non-cylindrical opening, and conform to the non-cylindrical profile, which may be at least at the opening of the test body. The device 3000 may conform to the opening and have no visible gaps between the device 3000 and the opening. There may be one or more radial gaps that are each no more than five, four, three, two and/or one millimeter across at their widest portion. Such gaps may be measured radially, or perpendicularly to a longitudinal axis extending through the geometric center of the test body opening. The gap may be measured between the outer surface of the device 3000 and the inner surface of the opening of the test body. The gap may be measured at the location of maximum space between the device 3000 and the test body. The device may conform to this shape after a period of at least thirty days, at least sixty days, and/or at least one hundred twenty days after implantation. In another example use, the device 3000 may be configured to insert into a non-cylindrical opening of a test body having a size and radial stiffness substantially similar to that of a native left atrial appendage, expand radially within the non-cylindrical opening, and assume a non-cylindrical profile at least at the opening of the test body after a period of at least thirty days, at least sixty days, and/or at least one hundred twenty days.
In some embodiments, the bumper 3026 may allow for extreme shortening of the distal end 3006 of the device. In some embodiments, the bumper 3026 may fold inward to accommodate radial and/or axial constraining of the device 3000. The bumper 3026 may fold radially inward and/or proximally inward. Further, the compliant frame 3040 within the foam body 3002 may allow for further axial shortening beyond the length of the bumper 3026. The frame 3040 may fold radially and/or axially inward.
Further, the cylindrical shape of the device 3000 facilitates with sealing the LAA, even with atypical geometries of the LAA anatomy. The cylindrical shape ensures that the anchors are located at the locations of maximum width of the device 3000. The tubular body 3080 may provide a cylindrical foundation for the anchors 3090, 3094, as described herein, such that the anchors are located at the radially outer most portion of the device 3000. Such cylindrical shape of the device along its longitudinal axis assists with the device 3000 performing the necessary sealing, even in the constrained configurations shown in
The device 3000 provides many advantages over existing solutions to LAA occlusion, as described herein. A key advantage is that the device is highly compliant while still providing superior resistance to embolization. This unique feature of being more compliant yet better anchoring is counterintuitive. As compared to existing solutions, the device 3000 is much more conformable and thus able to take the oval shape of the LAA ostium, as described, while also providing superior dislodgement resistance, in some embodiments with a pull out force in bench testing of greater than 0.8 pounds (lbs). [0452]
Various modifications to the implementations described in this disclosure will be readily apparent to those skilled in the art, and the generic principles defined herein can be applied to other implementations without departing from the spirit or scope of this disclosure. Thus, the disclosure is not intended to be limited to the implementations shown herein, but is to be accorded the widest scope consistent with the claims, the principles and the novel features disclosed herein. The word “example” is used exclusively herein to mean “serving as an example, instance, or illustration.” Any implementation described herein as “example” is not necessarily to be construed as preferred or advantageous over other implementations, unless otherwise stated.
Certain features that are described in this specification in the context of separate implementations also can be implemented in combination in a single implementation. Conversely, various features that are described in the context of a single implementation also can be implemented in multiple implementations separately or in any suitable sub-combination. Moreover, although features can be described above as acting in certain combinations and even initially claimed as such, one or more features from a claimed combination can in some cases be excised from the combination, and the claimed combination can be directed to a sub-combination or variation of a sub-combination.
Similarly, while operations are depicted in the drawings in a particular order, this should not be understood as requiring that such operations be performed in the particular order shown or in sequential order, or that all illustrated operations be performed, to achieve desirable results. Additionally, other implementations are within the scope of the following claims. In some cases, the actions recited in the claims can be performed in a different order and still achieve desirable results.
It will be understood by those within the art that, in general, terms used herein are generally intended as “open” terms (e.g., the term “including” should be interpreted as “including but not limited to,” the term “having” should be interpreted as “having at least,” the term “includes” should be interpreted as “includes but is not limited to,” etc.). It will be further understood by those within the art that if a specific number of an introduced claim recitation is intended, such an intent will be explicitly recited in the claim, and in the absence of such recitation no such intent is present. For example, as an aid to understanding, the following appended claims may contain usage of the introductory phrases “at least one” and “one or more” to introduce claim recitations. However, the use of such phrases should not be construed to imply that the introduction of a claim recitation by the indefinite articles “a” or “an” limits any particular claim containing such introduced claim recitation to embodiments containing only one such recitation, even when the same claim includes the introductory phrases “one or more” or “at least one” and indefinite articles such as “a” or “an” (e.g., “a” and/or “an” should typically be interpreted to mean “at least one” or “one or more”); the same holds true for the use of definite articles used to introduce claim recitations. In addition, even if a specific number of an introduced claim recitation is explicitly recited, those skilled in the art will recognize that such recitation should typically be interpreted to mean at least the recited number (e.g., the bare recitation of “two recitations,” without other modifiers, typically means at least two recitations, or two or more recitations). Furthermore, in those instances where a convention analogous to “at least one of A, B, and C, etc.” is used, in general such a construction is intended in the sense one having skill in the art would understand the convention (e.g., “a system having at least one of A, B, and C” would include but not be limited to systems that have A alone, B alone, C alone, A and B together, A and C together, B and C together, and/or A, B, and C together, etc.). In those instances where a convention analogous to “at least one of A, B, or C, etc.” is used, in general such a construction is intended in the sense one having skill in the art would understand the convention (e.g., “a system having at least one of A, B, or C” would include but not be limited to systems that have A alone, B alone, C alone, A and B together, A and C together, B and C together, and/or A, B, and C together, etc.). It will be further understood by those within the art that virtually any disjunctive word and/or phrase presenting two or more alternative terms, whether in the description, claims, or drawings, should be understood to contemplate the possibilities of including one of the terms, either of the terms, or both terms. For example, the phrase “A or B” will be understood to include the possibilities of “A” or “B” or “A and B.”
Any and all applications for which a foreign or domestic priority claim is identified in the Application Data Sheet as filed with the present application are hereby incorporated by reference under 37 CFR 1.57. This application is a continuation in part of U.S. application Ser. No. 15/795,083, entitled DEVICES AND METHODS FOR EXCLUDING THE LEFT ATRIAL APPENDAGE and filed Oct. 26, 2017, which claims the benefit of priority to U.S. provisional patent application No. 62/459,503, entitled DEVICES AND METHODS FOR EXCLUDING THE LEFT ATRIAL APPENDAGE and filed Feb. 15, 2017, and to U.S. provisional patent application No. 62/413,632, entitled MULTIFUNCTIONAL LEFT ATRIAL APPENDAGE (LAA) OCCLUDER and filed Oct. 27, 2016, the disclosure of each of which is hereby incorporated by reference herein in its entirety for all purposes and forms a part of this specification.
Number | Name | Date | Kind |
---|---|---|---|
3063453 | Brecht | Nov 1962 | A |
3712305 | Wennerblom et al. | Jan 1973 | A |
3812856 | Duncan et al. | May 1974 | A |
3978855 | McRae et al. | Sep 1976 | A |
4061145 | DesMarais | Dec 1977 | A |
4475911 | Gellert | Oct 1984 | A |
5192301 | Kamiya et al. | Mar 1993 | A |
5456693 | Conston et al. | Oct 1995 | A |
5634936 | Linden et al. | Jun 1997 | A |
5670572 | Ott et al. | Sep 1997 | A |
5725568 | Hastings | Mar 1998 | A |
5792179 | Sideris | Aug 1998 | A |
5823198 | Jones et al. | Oct 1998 | A |
5847012 | Shalaby et al. | Dec 1998 | A |
5848040 | Tanaka | Dec 1998 | A |
5861003 | Latson et al. | Jan 1999 | A |
5865791 | Whayne et al. | Feb 1999 | A |
5969000 | Yang et al. | Oct 1999 | A |
5968091 | Pinchuk | Nov 1999 | A |
6152144 | Lesh et al. | Nov 2000 | A |
6162168 | Schweich et al. | Dec 2000 | A |
6231561 | Frazier et al. | May 2001 | B1 |
6290674 | Roue et al. | Sep 2001 | B1 |
6398758 | Jacobsen et al. | Jun 2002 | B1 |
6408981 | Smith et al. | Jun 2002 | B1 |
6419669 | Frazier et al. | Jul 2002 | B1 |
6423252 | Chun et al. | Jul 2002 | B1 |
6436088 | Frazier et al. | Aug 2002 | B2 |
6447539 | Nelson et al. | Sep 2002 | B1 |
6458100 | Roue et al. | Oct 2002 | B2 |
6551303 | VanTassel et al. | Apr 2003 | B1 |
6623450 | Dutta | Sep 2003 | B1 |
6641557 | Frazier et al. | Nov 2003 | B1 |
6651303 | Toivanen et al. | Nov 2003 | B1 |
6652555 | VanTassel et al. | Nov 2003 | B1 |
6652556 | VanTassel et al. | Nov 2003 | B1 |
6666861 | Grabek | Dec 2003 | B1 |
6689150 | VanTassel et al. | Feb 2004 | B1 |
6712804 | Roue et al. | Mar 2004 | B2 |
6712810 | Harrington et al. | Mar 2004 | B2 |
6723108 | Jones et al. | Apr 2004 | B1 |
6730108 | Van Tassel et al. | May 2004 | B2 |
6881875 | Swenson | Apr 2005 | B2 |
6941169 | Pappu | Sep 2005 | B2 |
6949113 | Van Tassel et al. | Sep 2005 | B2 |
6977323 | Swenson | Dec 2005 | B1 |
6979344 | Jones et al. | Dec 2005 | B2 |
6994092 | Van Der et al. | Feb 2006 | B2 |
7011671 | Welch | Mar 2006 | B2 |
7044134 | Khairkhahan et al. | May 2006 | B2 |
7056294 | Khairkhahan et al. | Jun 2006 | B2 |
7115110 | Frazier et al. | Oct 2006 | B2 |
7128073 | Van der Burg et al. | Oct 2006 | B1 |
7152605 | Khairkhahan et al. | Dec 2006 | B2 |
7169164 | Borillo et al. | Jan 2007 | B2 |
7192439 | Khairkhahan et al. | Mar 2007 | B2 |
7226458 | Kaplan et al. | Jun 2007 | B2 |
7291382 | Krueger et al. | Nov 2007 | B2 |
7293562 | Malecki | Nov 2007 | B2 |
7318829 | Kaplan et al. | Jan 2008 | B2 |
7344543 | Sra | Mar 2008 | B2 |
7358282 | Krueger et al. | Apr 2008 | B2 |
7427279 | Frazier et al. | Sep 2008 | B2 |
7549983 | Roue et al. | Jun 2009 | B2 |
7566336 | Corcoran et al. | Jul 2009 | B2 |
7597704 | Frazier et al. | Oct 2009 | B2 |
7695425 | Schweich et al. | Apr 2010 | B2 |
7713282 | Frazier et al. | May 2010 | B2 |
7722641 | Van Der et al. | May 2010 | B2 |
7727189 | VanTassel et al. | Jun 2010 | B2 |
7735493 | van der Burg et al. | Jun 2010 | B2 |
7747047 | Okerlund et al. | Jun 2010 | B2 |
7780683 | Roue et al. | Aug 2010 | B2 |
7803395 | Datta et al. | Sep 2010 | B2 |
7824397 | Mcauley | Nov 2010 | B2 |
7922716 | Malecki et al. | Apr 2011 | B2 |
7972359 | Kreidler | Jul 2011 | B2 |
7998138 | Mcauley | Aug 2011 | B2 |
8043329 | Khairkhahan et al. | Oct 2011 | B2 |
8052715 | Quinn et al. | Nov 2011 | B2 |
8057530 | Kusleika et al. | Nov 2011 | B2 |
8080032 | van der Burg et al. | Dec 2011 | B2 |
8097015 | Devellian | Jan 2012 | B2 |
8142470 | Quinn et al. | Mar 2012 | B2 |
8157818 | Gartner et al. | Apr 2012 | B2 |
8197496 | Roue et al. | Jun 2012 | B2 |
8197527 | Borillo et al. | Jun 2012 | B2 |
8221445 | Van Tassel et al. | Jul 2012 | B2 |
8262694 | Widomski et al. | Sep 2012 | B2 |
8287563 | Khairkhahan et al. | Oct 2012 | B2 |
8313504 | Do et al. | Nov 2012 | B2 |
8323309 | Khairkhahan et al. | Dec 2012 | B2 |
8337487 | Datta et al. | Dec 2012 | B2 |
8361111 | Widomski et al. | Jan 2013 | B2 |
8460282 | Mcauley | Jun 2013 | B2 |
8480708 | Kassab et al. | Jul 2013 | B2 |
8523897 | Van Der et al. | Sep 2013 | B2 |
8535343 | Van Der et al. | Sep 2013 | B2 |
8540760 | Paul, Jr. et al. | Sep 2013 | B2 |
8603108 | Roue et al. | Dec 2013 | B2 |
8636764 | Miles et al. | Jan 2014 | B2 |
8690911 | Miles et al. | Jan 2014 | B2 |
8647361 | Borillo et al. | Feb 2014 | B2 |
8647367 | Kassab et al. | Feb 2014 | B2 |
8663268 | Quinn et al. | Mar 2014 | B2 |
8663273 | Khairkhahan et al. | Mar 2014 | B2 |
8685055 | VanTassel et al. | Apr 2014 | B2 |
8715302 | Ibrahim et al. | May 2014 | B2 |
8715318 | Miles et al. | May 2014 | B2 |
8740934 | McGuckin, Jr. | Jul 2014 | B2 |
8764793 | Lee | Jul 2014 | B2 |
8784469 | Kassab | Jul 2014 | B2 |
8795328 | Miles et al. | Aug 2014 | B2 |
8801746 | Kreidler et al. | Aug 2014 | B1 |
8828051 | Javois et al. | Sep 2014 | B2 |
8834519 | van der Burg et al. | Sep 2014 | B2 |
8840641 | Miles et al. | Sep 2014 | B2 |
8845711 | Miles et al. | Sep 2014 | B2 |
9011551 | Oral et al. | Apr 2015 | B2 |
9034006 | Quinn et al. | May 2015 | B2 |
9089313 | Roue et al. | Jul 2015 | B2 |
9131849 | Khairkhahan et al. | Sep 2015 | B2 |
9132000 | VanTassel et al. | Sep 2015 | B2 |
9161830 | Borillo et al. | Oct 2015 | B2 |
9168043 | Van Der et al. | Oct 2015 | B2 |
9186152 | Campbell et al. | Nov 2015 | B2 |
9351716 | Miles et al. | May 2016 | B2 |
9421004 | Roue et al. | Aug 2016 | B2 |
9445895 | Kreidler | Sep 2016 | B2 |
9474516 | Clark et al. | Oct 2016 | B2 |
9554804 | Erzberger et al. | Jan 2017 | B2 |
9592058 | Erzberger et al. | Mar 2017 | B2 |
9592110 | Dan et al. | Mar 2017 | B1 |
9649115 | Edmiston et al. | May 2017 | B2 |
9693780 | Miles et al. | Jul 2017 | B2 |
9693781 | Miles et al. | Jul 2017 | B2 |
9700323 | Clark | Jul 2017 | B2 |
9730701 | Tischler et al. | Aug 2017 | B2 |
9743932 | Amplatz et al. | Aug 2017 | B2 |
9763666 | Wu et al. | Sep 2017 | B2 |
9808253 | Li et al. | Nov 2017 | B2 |
9839431 | Meyer et al. | Dec 2017 | B2 |
9849011 | Zimmerman et al. | Dec 2017 | B2 |
9861370 | Clark et al. | Jan 2018 | B2 |
9883864 | Miles et al. | Feb 2018 | B2 |
9883936 | Sutton et al. | Feb 2018 | B2 |
9913652 | Bridgeman et al. | Mar 2018 | B2 |
9943299 | Khairkhahan et al. | Apr 2018 | B2 |
11026695 | Melanson | Jun 2021 | B2 |
20020022860 | Borillo et al. | Feb 2002 | A1 |
20020049457 | Kaplan et al. | Apr 2002 | A1 |
20020095205 | Edwin et al. | Jul 2002 | A1 |
20020099390 | Kaplan et al. | Jul 2002 | A1 |
20020103492 | Kaplan et al. | Aug 2002 | A1 |
20020111637 | Kaplan et al. | Aug 2002 | A1 |
20020111647 | Khairkhahan et al. | Aug 2002 | A1 |
20020169377 | Khairkhahan et al. | Nov 2002 | A1 |
20030023266 | Borillo et al. | Jan 2003 | A1 |
20030051735 | Pavcnik | Mar 2003 | A1 |
20030120337 | Van Tassel et al. | Jun 2003 | A1 |
20030191526 | Van Tassel et al. | Oct 2003 | A1 |
20030199923 | Khairkhahan et al. | Oct 2003 | A1 |
20030204203 | Khairkhahan et al. | Oct 2003 | A1 |
20030220667 | van der Burg et al. | Nov 2003 | A1 |
20040049210 | VanTassel et al. | Mar 2004 | A1 |
20040098031 | van der Burg et al. | May 2004 | A1 |
20040122467 | VanTassel et al. | Jun 2004 | A1 |
20040127935 | VanTassel et al. | Jul 2004 | A1 |
20040215230 | Frazier et al. | Oct 2004 | A1 |
20040220610 | Kreidler et al. | Nov 2004 | A1 |
20040225212 | Okerlund et al. | Nov 2004 | A1 |
20040230222 | van der Burg et al. | Nov 2004 | A1 |
20050004652 | van der Burg et al. | Jan 2005 | A1 |
20050033287 | Sra | Feb 2005 | A1 |
20050038470 | van der Burg et al. | Feb 2005 | A1 |
20050049573 | VanTassel et al. | Mar 2005 | A1 |
20050070952 | Devellian | Mar 2005 | A1 |
20050113861 | Corcoran et al. | May 2005 | A1 |
20050149068 | Williams et al. | Jul 2005 | A1 |
20050149069 | Bertolero et al. | Jul 2005 | A1 |
20050177182 | van der Burg et al. | Aug 2005 | A1 |
20050234540 | Peavey et al. | Oct 2005 | A1 |
20050234543 | Glaser et al. | Oct 2005 | A1 |
20050267528 | Ginn | Dec 2005 | A1 |
20060009715 | Khairkhahan et al. | Jan 2006 | A1 |
20060116709 | Sepetka | Jun 2006 | A1 |
20070005147 | Levine | Jan 2007 | A1 |
20070083230 | Javois | Apr 2007 | A1 |
20070129753 | Quinn et al. | Jun 2007 | A1 |
20070135826 | Zaver et al. | Jun 2007 | A1 |
20070149988 | Michler et al. | Jun 2007 | A1 |
20070149995 | Quinn et al. | Jun 2007 | A1 |
20070179345 | Santilli | Aug 2007 | A1 |
20070270891 | McGuckin, Jr. | Nov 2007 | A1 |
20080033241 | Peh et al. | Feb 2008 | A1 |
20080125795 | Kaplan et al. | May 2008 | A1 |
20080243183 | Miller et al. | Oct 2008 | A1 |
20080294175 | Bardsley et al. | Nov 2008 | A1 |
20080312664 | Bardsley et al. | Dec 2008 | A1 |
20090005760 | Cartledge | Jan 2009 | A1 |
20090099596 | McGunkin, Jr. et al. | Apr 2009 | A1 |
20090112249 | Miles et al. | Apr 2009 | A1 |
20090143791 | Miller et al. | Jun 2009 | A1 |
20090157118 | Miller et al. | Jun 2009 | A1 |
20090264920 | Berenstein | Oct 2009 | A1 |
20090287145 | Cragg et al. | Nov 2009 | A1 |
20090306685 | Fill | Dec 2009 | A1 |
20090326577 | Johnson et al. | Dec 2009 | A1 |
20100076463 | Mavani et al. | Mar 2010 | A1 |
20100191279 | Kassab et al. | Jul 2010 | A1 |
20100228184 | Mavani et al. | Sep 2010 | A1 |
20100228279 | Miles et al. | Sep 2010 | A1 |
20100228285 | Miles et al. | Sep 2010 | A1 |
20100286718 | Kassab et al. | Nov 2010 | A1 |
20100324585 | Miles et al. | Dec 2010 | A1 |
20100324586 | Miles et al. | Dec 2010 | A1 |
20100324587 | Miles et al. | Dec 2010 | A1 |
20100324588 | Miles et al. | Dec 2010 | A1 |
20110009853 | Bertolero et al. | Jan 2011 | A1 |
20110022079 | Miles et al. | Jan 2011 | A1 |
20110022168 | Cartledge | Jan 2011 | A1 |
20110054515 | Bridgemann et al. | Mar 2011 | A1 |
20110082495 | Ruiz | Apr 2011 | A1 |
20110087271 | Sargent et al. | Apr 2011 | A1 |
20110178539 | Holmes, Jr. et al. | Jul 2011 | A1 |
20110208233 | McGunkin, Jr. et al. | Aug 2011 | A1 |
20110218389 | Gobel | Sep 2011 | A1 |
20110218566 | van der Burg et al. | Sep 2011 | A1 |
20110220120 | Frigstad et al. | Sep 2011 | A1 |
20110257674 | Evert et al. | Oct 2011 | A1 |
20110264119 | Bayon et al. | Oct 2011 | A1 |
20110307003 | Chambers | Dec 2011 | A1 |
20120010644 | Sideris et al. | Jan 2012 | A1 |
20120029553 | Quinn et al. | Feb 2012 | A1 |
20120065662 | van der Burg et al. | Mar 2012 | A1 |
20120065667 | Javois et al. | Mar 2012 | A1 |
20120157916 | Quinn et al. | Jun 2012 | A1 |
20120158022 | Kaplan et al. | Jun 2012 | A1 |
20120172927 | Campbell et al. | Jul 2012 | A1 |
20120221042 | Schwartz et al. | Aug 2012 | A1 |
20120239077 | Zaver et al. | Sep 2012 | A1 |
20120283585 | Werneth et al. | Nov 2012 | A1 |
20120283773 | VanTassel et al. | Nov 2012 | A1 |
20120316584 | Miles et al. | Dec 2012 | A1 |
20120323262 | Ibrahim et al. | Dec 2012 | A1 |
20120323267 | Ren | Dec 2012 | A1 |
20120323270 | Lee | Dec 2012 | A1 |
20120330342 | Jones | Dec 2012 | A1 |
20130006343 | Kassab | Jan 2013 | A1 |
20130012982 | Khairkhahan et al. | Jan 2013 | A1 |
20130018413 | Oral et al. | Jan 2013 | A1 |
20130018414 | Widomski et al. | Jan 2013 | A1 |
20130083983 | Zhong et al. | Apr 2013 | A1 |
20130110154 | van der Burg et al. | May 2013 | A1 |
20130116724 | Clark et al. | May 2013 | A1 |
20130165965 | Carlson et al. | Jun 2013 | A1 |
20130178889 | Miles et al. | Jul 2013 | A1 |
20130237908 | Clark | Sep 2013 | A1 |
20140005714 | Quick | Jan 2014 | A1 |
20140046360 | van der Burg et al. | Feb 2014 | A1 |
20140074151 | Tischler et al. | Mar 2014 | A1 |
20140128903 | Alferness | May 2014 | A1 |
20140257320 | Fitz | Sep 2014 | A1 |
20140277074 | Kaplan et al. | Sep 2014 | A1 |
20140330368 | Gloss | Nov 2014 | A1 |
20140336699 | van der Burg et al. | Nov 2014 | A1 |
20140371789 | Hariton et al. | Dec 2014 | A1 |
20150005810 | Center et al. | Jan 2015 | A1 |
20150039021 | Khairkhahan et al. | Feb 2015 | A1 |
20150133989 | Lubeck et al. | May 2015 | A1 |
20150196305 | Meyer et al. | Jul 2015 | A1 |
20160058539 | Vantassel et al. | Jan 2016 | A1 |
20160089151 | Siegel et al. | Mar 2016 | A1 |
20160106437 | van der Burg et al. | Apr 2016 | A1 |
20170042549 | Kaplan et al. | Feb 2017 | A1 |
20170042550 | Chakraborty et al. | Feb 2017 | A1 |
20170095238 | Rudman et al. | Apr 2017 | A1 |
20170100112 | Van Der et al. | Apr 2017 | A1 |
20170135801 | Delaney, Jr. et al. | May 2017 | A1 |
20170224354 | Tischler et al. | Aug 2017 | A1 |
20170281192 | Tieu et al. | Oct 2017 | A1 |
20170290594 | Chakraborty et al. | Oct 2017 | A1 |
20170340336 | Osypka | Nov 2017 | A1 |
20180185130 | Janardhan et al. | Jul 2018 | A1 |
20180206830 | Khairkhahan et al. | Jul 2018 | A1 |
20180250014 | Melanson et al. | Sep 2018 | A1 |
20180338824 | VanTassel et al. | Nov 2018 | A1 |
20190083075 | Onushko et al. | Mar 2019 | A1 |
20190125362 | Tischler | May 2019 | A1 |
20190336137 | Chakraborty et al. | Nov 2019 | A1 |
20200269059 | Kaiser | Aug 2020 | A1 |
Number | Date | Country |
---|---|---|
1341519 | Feb 2007 | CA |
102088927 | Jun 2011 | CN |
102006056283 | Jun 2008 | DE |
1223890 | Apr 2004 | EP |
1227770 | Sep 2004 | EP |
1225843 | Feb 2005 | EP |
308531 | Oct 2016 | EP |
1469790 | Oct 2016 | EP |
2872051 | Mar 2017 | EP |
2003-512128 | Apr 2003 | JP |
2003-529384 | Oct 2003 | JP |
2012-530551 | Dec 2012 | JP |
WO 0027292 | May 2000 | WO |
WO2009009466 | Jan 2009 | WO |
WO 2010148246 | Dec 2010 | WO |
WO2014011865 | Jan 2014 | WO |
WO 2014078531 | May 2014 | WO |
WO 2014164572 | Oct 2014 | WO |
WO2016033170 | Mar 2016 | WO |
WO 2017161283 | Sep 2017 | WO |
WO2018185255 | Oct 2018 | WO |
WO2018185256 | Oct 2018 | WO |
WO 2019033121 | Feb 2019 | WO |
Entry |
---|
Möbius-Winkler, S., Sandri, M., Mangner, N., Lurz, P., Dähnert, I., Schuler, G. The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation. J. Vis. Exp. (60), e3671, DOI : 10.3791/3671 (Feb. 28, 2012). |
Extended European Search Report in European Patent Case No. EP 14 77 9640 dated Sep. 30, 2016. |
International Search Report and Written Opinion dated Jan. 19, 2017, in International Application No. PCT/US2016/056450. |
International Search Report and Written Opinion dated Jul. 3, 2014, in International Application No. PCT/US2014/022865. |
International Search Report dated Jul. 10, 2019, in International Application No. PCT/US2019/29364. |
International Search Report issued in International Patent Application No. PCT/US2020/016854, dated Jun. 8, 2020. |
Number | Date | Country | |
---|---|---|---|
20180250014 A1 | Sep 2018 | US |
Number | Date | Country | |
---|---|---|---|
62459503 | Feb 2017 | US | |
62413632 | Oct 2016 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 15795083 | Oct 2017 | US |
Child | 15969654 | US |