The present application relates to devices, systems and methods for evaluating hemostasis in a subject by analysis of a test sample from the subject to determine one or more indices of hemostasis.
Hemostasis, the physiological control of bleeding, is a complex process incorporating the vasculature, platelets, coagulation factors (FI-FXIII), fibrinolytic proteins, and coagulation inhibitors. Disruption of hemostasis plays a central role in the onset of myocardial infarction, stroke, pulmonary embolism, deep vein thrombosis and excessive bleeding. Consequently, in vitro diagnostics (IVD) are critically needed to quantify hemostatic dysfunction and direct appropriate treatment. This need is particularly acute during cardiac surgeries requiring cardiopulmonary bypass (CPB), where post-surgical bleeding is a common complication requiring transfusion of blood products.
Existing IVDs include endpoint biochemical assays, platelet aggregation assays, and clot viscoelastic measurement systems. Endpoint biochemical assays such as the prothrombin time (PT) and the partial thromboplastin time (PTT) are widely used to assess coagulation. However, these tests measure only a part of the hemostatic process and operate under non-physiological conditions incorporating only the function of plasma. As a result of these limitations, complications such as postoperative bleeding often occur despite normal perioperative PT and PTT measurements.
Activated clotting time (ACT) is an endpoint assay that is most often applied in support of CPB. This assay applies strong initiation of the surface activation (intrinsic) pathway to quantify heparinization. Limitations of the ACT include its disregard for platelet function, lysis, and coagulation kinetics along with the use of large aliquots of whole blood (WB) (generally 2 mL) and moving mechanical parts. For these reasons, the ACT is used for rapid assessment of heparinization and associated protamine reversal with limited utility for additional applications.
Platelets play a crucial role in the progression of coagulation and quelling arterial bleeding. Furthermore, the modern cell-based theory of hemostasis recognizes that platelets play a modulating role in coagulation. Platelet function is monitored clinically via both central lab assays and point of care (POC) tests, which use anticoagulated WB. Both approaches are limited in that they use platelet aggregation as a proxy for overall platelet function. Furthermore, disabling coagulation, these methods neglect the interaction between platelets and the coagulation cascade.
Techniques that monitor the viscoelastic properties of WB, such as thromboelastography (TEG) and rotational thromboelastometer (ROTEM), circumvent many of the limitations of endpoint biochemical assays and platelet aggregation assays by measuring the combined effects of all components of hemostasis. TEG has been shown to diagnose hyperfibrinolysis in bleeding patients, indicate transfusion requirements better than standard biochemical assays, and reduce transfusion requirements during CPB when used with transfusion algorithms. While these tests offer valuable clinical information, the devices are typically complex to operate and difficult to interpret. Moreover, the TEG applies relatively large shear strains, which transgress the non-linear viscoelastic regime, thereby disrupting clot formation. For these reasons, the TEG sees very limited utility as a POC test.
Provided are devices, systems and methods for evaluation of hemostasis. For example, provided are sonorheometric devices for evaluation of hemostasis in a subject by in vitro evaluation of a test sample from the subject. An example device comprises a cartridge having a plurality of test chambers each configured to receive a test sample of blood from the subject. Each test chamber comprises a reagent or combination of reagents.
A first chamber of the plurality comprises a first reagent or a combination of reagents that interact with the test sample of blood received therein. A second chamber of the plurality comprises a second reagent or combination of reagents that interact with the test sample of blood received therein. The first and second chambers are configured to be interrogated with sound to determine a hemostatic parameter of the test samples.
The example device can further comprise a third chamber having a third reagent or combination of reagents that interact with the test sample of blood received therein and a fourth chamber having a fourth reagent or combination of reagents that interact with the test sample of blood received therein. The third and fourth chambers are also configured to be interrogated with sound to determine a hemostatic parameter of the tests samples. Example reagents are selected from the group consisting of kaolin, celite, glass, abciximab, cytochalasin D, thrombin, recombinant tissue factor, reptilase, arachidonic acid (AA), adenosine diphosphate (ADP), and combinations thereof. Optionally, the reagents are lyophilized prior to interacting with the test samples.
The example devices can be used in a system comprising a transducer for transmitting ultrasound into one or more chamber and for receiving reflected sound from the chamber and the test sample therein. The system can further comprise at least one processor configured to determine a hemostasis parameter from the received sound. The parameters are optionally selected from the group consisting of TC1, TC2, clot stiffness, clot formation rate (CFR), TL1 and TL2. The processor is optionally further configured to determine an intrinsic pathway coagulation factors index, an extrinsic pathway coagulation factors index, a platelets index, a fibrinogen index, and a fibrinolysis index value. The intrinsic and extrinsic coagulation factors are optionally combined to form a coagulation factors index.
Also provided are sonorheometric methods for evaluation of hemostasis in a subject, comprising a cartridge having at least two test chambers. Each test chamber comprises a reagent or combination thereof. Blood from the subject is introduced into the test chambers to mix with the reagents and ultrasound is transmitted into each test chamber. Sound reflected from the blood reagent mixture in the test chamber is received and processed to generate a hemostasis parameter. The parameters are optionally selected from the group consisting of TC1, TC2, clot stiffness, clot formation rate (CFR), TL1 and TL2. The disclosed methods can further include determining an intrinsic pathway coagulation factors index, an extrinsic pathway coagulation factors index, a platelets index, a fibrinogen index, and a fibrinolysis index value. The intrinsic and extrinsic coagulation factors are optionally combined to form a coagulation factors index. The reagents or combinations thereof are optionally lyophilized prior to mixing with the blood.
Further provided are sound focusing assemblies. An example sound focusing assembly includes a rigid substrate permeable by sound and an elastomeric couplant permeable by sound. The elastomeric couplant is positioned relative to the rigid substrate to create an interface between the elastomeric couplant and the rigid substrate, wherein the interface focuses sound transmitted through the assembly.
These and other features and advantages of the present invention will become more readily apparent to those skilled in the art upon consideration of the following detailed description and accompanying drawings, which describe both the preferred and alternative embodiments of the present invention.
The present invention now will be described more fully hereinafter with reference to specific embodiments of the invention. Indeed, the invention can be embodied in many different forms and should not be construed as limited to the embodiments set forth herein; rather, these embodiments are provided so that this disclosure will satisfy applicable legal requirements.
As used in the specification, and in the appended claims, the singular forms “a,” “an,” “the,” include plural referents unless the context clearly dictates otherwise.
The term “comprising” and variations thereof as used herein are used synonymously with the term “including” and variations thereof and are open, non-limiting terms.
As used throughout, by a “subject” is meant an individual. The subject may be a vertebrate, more specifically a mammal (e.g., a human, horse, pig, rabbit, dog, sheep, goat, non-human primate, cow, cat, guinea pig or rodent), a fish, a bird or a reptile or an amphibian. The term does not denote a particular age or sex.
The cartridge further includes a port 106 for applying a vacuum to the cartridge. When a vacuum is applied at the port 106, the biological fluid introduced at the inlet 102 into the channel 202 the fluid is propelled along the channel 202 towards the port 106.
As shown in
Referring again to
Each test chamber is configured to accept a quantity of the biological fluid into the open space. In reference to test chamber 116, illustrated in detail in
The biological fluid can also exit each respective test chamber and continue along an exit channel 130 towards the port 106. Thus, fluid introduced at the inlet 102 flows under vacuum through the device channels and into the test chambers. From each test chamber (110, 112, 114, 116), the biological fluid continues to flow along exit channels towards the vacuum.
Proximate the port 106 each exit channel may direct the flowing biological fluid into a hydrophobic filter at location 222, 220, 218 and 216 respectively. The filters or filter prevents movement of the biological fluid out of the cartridge 100 at the port 106. Because the volume of the channels and the test chamber are fixed, the vacuum can pull the biological fluid into the cartridge until the channels and each test chamber is filled with the biological fluid.
Pressure can be controlled within the cartridge 100 to, for example, manage flow rate within the consumable 100 and to mitigate reliability issues related to possible user misuse. To measure the properties of a target biological sample, such as a blood sample, a user of the hemostasis system optionally attaches a blood filled syringe to the cartridge 100 unit. There exists the possibility that the user of the hemostasis system 300 (see
Inadequate mixing of the biological sample with the reagents described herein may result in variation of hemostasis measurements. Rapidly aspirating the blood sample is optionally used to provide increased mixing of the reagents with the biological sample, such as a blood sample. This is optionally achieved by creating a pressure differential between the cartridge and the aspirating mechanism of the hemostasis system.
In this regard,
The level of sample in each chamber can also be monitored. For example, as shown in
Whether a desired level has been reached in a given chamber can be indicated by a LED or other visual indicator. Employing a single light beam from an LED emitter 802 reflecting off the chamber at a blood detection target reservoir 224, which is then detected by a detector 800 can be optionally used to optically monitor chamber fluid level.
For example, blood entering a test chamber reduces reflection of light originating from an emitter 802 located alongside the detector 800, and pointed at the test chamber. A dual beam approach can be used whereby two sources of different wavelengths were reflected off the test chamber. Blood has a deep red color that can be differentiated by comparing the red wavelength reflection to that of another colour.
The difference in intensity of the reflected red light alone is sufficient to determine when blood has entered a chamber. The red light intensity reflected from the test chamber containing blood was about one-half that of the well containing air, and about two-thirds of that from the well containing water.
To control the temperature of the biological sample entering the test chambers the cartridge 100 can comprise a heat exchanger in communication with the channel 204. The heat exchanger can be used to maintain, elevate or lower the temperature of the biological fluid before analysis in each test chamber. Optionally, the temperature of biological fluid for analysis in each test chamber is the same such that common portion of the channel system, as shown in
For example, to heat the biological fluid, it can be passed through the channel 204 through a polystyrene labyrinth held against a copper block. The copper block can be thin (for example under 2 mm) and sized just larger than the labyrinth to minimize the thermal mass. A thermistor can be embedded in the block so that a control circuit could maintain a steady set temperature in the block. A heater is used that optionally comprises two Watlow® (St. Louis, Mo.) serpentine foil heating elements bonded to a flexible kapton plastic substrate, and the interface between the block and the heater can be a thin layer of silicone heatsink compound.
Various flow rates, for example, up to and including 5.99 ml/min or 6.0 ml/min can be used, and power input to the heater can be varied optionally between 8 and 16 Watts. Blood or other biological fluid can be heated in the cartridge from ambient temperature (approximately 20° C.) to 37° C. at a nominal flow rate of 6 ml/min, which is fast enough to fill the cartridge in 20 seconds. The surface area of the labyrinth used was less than 8 cm2.
Physiologically, the process of coagulation is highly dependent on the temperature at which it takes place. Under normal conditions, coagulation occurs at body temperature (37° C.), which is optimal for the proper enzymatic action of the clotting factors in the cascade.
Blood can be warmed from its incoming temperature, ranging between 18° C. and 37° C., to an arbitrary or desired temperature, such as body temperature, of 37° C. by passing through a serpentine channel in close proximity to a heater block. To accomplish the heating in a short time over a short path the block can be warmed to almost 60° C. when the incoming blood is at the lower end of its temperature range. The temperature of the blood can also be measured and the heater block can optionally be adjusted to a temperature, ranging from 40° C. to 58° C.
To measure the temperature a sensor can be incorporated in the system 300 (
Temperature changes can occur in the body due to fever or in hospital settings such as the emergency room (ER) or operating room (OR). Trauma patients arriving at the ER are treated with large volumes of intravenous saline, which lowers body temperature to as much as 17° C. In the OR, patients undergoing cardiac bypass surgeries (CPB) have their entire blood volume pass through a lung-heart machine, which also lowers blood temperature and can adversely affect coagulation. Also, if there is a lag of time between the time of blood draw and the measurement, the temperature of blood is given time to change.
Styron® 666 (Styron Inc. Berwyn, Pa.) polystyrene and the microfluidic heat exchanger channel 204 allows a blood sample to be warmed by a copper block outside of the cartridge that is kept at a constant 37° C. When a sample enters the cartridge at temperatures substantially lower than 37° C., it is optionally desirable to use a cartridge modified to allow for more rapid heating of the biological sample. For example, in a model that simulates the temperature changes over time of blood entering the polystyrene cartridge at 17° C., Styron® 666 was found to reduce ability to heat blood and the blood exiting the heat exchanger did not reach 37° C. These shortcomings of Styron® 666 are due to its relatively low thermal conductivity. When more rapid or efficient heating of the biological sample is desired that is possible through Styron® 666, the cartridge can include materials with higher thermal conductivity than Styron® 666. For example, a thermally conductive polymer (E1201®) from Cool Polymers Inc. (North Kingstown, R.I.) with improved thermal conductivity properties can be used. This polymer can form a portion of the cartridge between the heating block and the channel 204. By using this polymer in a portion of the cartridge between the heating block and sample, the sample can be more efficiently heated. For example,
Cartridges optionally include both materials, E1201® and Styron® 666, in order to improve the heat transfer to the sample with E1201® on the heated side while maintaining flow visibility on the other side of the consumable with the Styron® 666. Another alternative is to use E1201® as an insert that fits over the copper heater and into a chassis made out of Styron® 666. This is optionally accomplished by overmolding the separate pieces into one single piece or affixing the E1201® to the Styron® chassis by means such as laser, ultrasonic or RF welding. Changing the geometry of the E1201® insert to fit into the larger chassis as a puzzle piece can further improve assembly of the separate parts and help seal the microfluidic flow chambers.
It may also be desirable to cool the biological fluid in the cartridge. In these example, and similar to when heating is desired, the cartridge can include materials with higher thermal conductivity than Styron® 666. For example, the thermally conductive polymer (E1201®), described above, with improved thermal conductivity properties can be used. This polymer can form a portion of the cartridge between a cooling device, such as a peltier cooling device, and the channel 204. Using this polymer in a portion of the cartridge between the cooling device and sample, the sample can be efficiently cooled.
Each test chamber can comprise one or more reagents useful in the analysis of one or more indices of hemostasis. Optionally, the reagents are lyophilized. Optionally, one or more lyophilized bead type reagent is used. For example, the lyophilized bead can be a LyoSphere® produced by BioLyph (Minnetonka, Minn.). A self-contained lyophilized bead is a format that allows for immunochemical and clinical chemistry reagents requiring two or three components that are incompatible as liquids because of their pH level or reaction to one another to coexist compatibly. Because such lyophilized beads are stable and nonreactive, chemicals can be packaged together in the same test chamber.
To produce lyophilized reagents, a lyophilizer device can be used. For example, the reagent for a given test chamber can be frozen to solidify all of its water molecules. Once frozen, the product is placed in a vacuum and gradually heated without melting the product. This process, called sublimation, transforms the ice directly into water vapor, without first passing through the liquid state. The water vapor given off by the product in the sublimation phase condenses as ice on a collection trap, known as a condenser, within the lyophilizer's vacuum chamber. Optionally, the lyophilized product contains 3% or less of its original moisture content. The lyophilized product, which may be a pellet, can then be positioned in each test chamber. Once placed in a test chamber, the test chamber can be sealed to prevent unwanted rehydration of the product.
To locate the lyophilized reagents in the test chambers, the components can first be lyophilized and then the resulting lyophilized product can be placed in the test chambers. Using UV cure epoxy glue or a welding process (such as ultrasound or RF welding), the lens assembly is sealed over each of the test chambers. The assembled cartridge can be sealed in a vapor proof barrier (e.g. a bag) and the vapor barrier can be sealed to preserve the dehydrated nature of the product in the test chambers. When ready for use, the cartridge can be removed from the bag or vapor barrier and placed into an analysis system 300, which is described in further detail below.
Anti-static treatment of plastic cartridges is optionally used with the lyophilized reagents. Lyophilized reagents are inherently devoid of water, granting them significant electrical insulation.
Materials that are electrical insulators more readily build up static charge than materials that act as electrical conductors. This can create problems with process control when assembling the cartridges and loading the reagents. Since the cartridges are optionally made from an electrically insulating material (polystyrene, for example), it is not likely to dissipate a static charge build up within the lyophilized reagents. As a result, lyophilized reagents can statically adhere to the interior walls of the consumable. In order to prevent this from occurring, three techniques are optionally implemented to remove static build-up.
Air ionization is a method that passes directed, ionized air over a target material to neutralize residual static charge on the material surface. Directing ionized air at one or more cartridge test chamber and/or the reagents during the assembly process improves manufacturability by reducing the adherence of the reagent bead to the cartridge test chambers.
A second method implements cartridge construction using a plastic material that exhibits significantly more conductivity than standard injection molding materials. RTP PermaStat® (Winona, Mass.) plastics are an example of such materials. The use of this material for the cartridge reduces the adhesion of the lyophilized reagents to the cartridge test chamber walls.
Third anti-static, liquid sprays are used to temporarily create a dust-free coating on optical lenses and equipment. These sprays reduce static charge on the target surface and are useful for static reduction during the cartridge assembly process.
When the lyophilized reagents are exposed to the fluid sample, they can generate foam that floats at the surface of the sample in the test chambers. As illustrated in
As shown in
Reagents are mixed with the sample before testing is initiated. Mixing of the reagents can be accomplished using passive and/or active mechanisms. Passive methods include, for example, the use of serpentine channels and embedded barriers to create flow turbulence. Active methods include, for example, magnetic beads, pressure perturbation, and artificial cilia. The consumable cartridge contains a lens that focuses ultrasound energy within the sample that can be used to generate streaming and mixing. The lens, also referred to herein as a lens assembly, or sound focusing assembly, is designed using a soft material, such as a thermoplastic elastomer 134, in conjunction with a rigid substrate 132, such as polystyrene. This combination provides a dry ultrasound coupling that does not require the use of any fluid or gel couplant. Note that the same lens and ultrasound driver used for hemostasis measurement can be used in this matter to provide mixing. Increasing acoustic energy for mixing can be delivered by, for example, increasing pulse length, pulse amplitude or pulse repetition frequency.
Mixing can also be provided by a variable magnetic field applied by a series of coils placed outside a test chamber or each test chamber. A small magnetic bead or magnetic stirrer can be placed within a test chamber and when the fluid sample enter the chamber, the current across the coils can be modulated in order to generate a variable magnetic field. This generates motion of the magnetic bead or magnetic stirrer which in turns generates mixing of the sample with the reagent.
The exposure of blood to surface proteins, such as in the case of collagen or von Willebrand factor (vWF) on damaged blood vessel walls is an essential part of the coagulation process. These proteins not only contribute to the clotting cascade but also modulate several steps leading to clot formation and hemostasis.
Although exposure to these proteins is essential to the coagulation cascade, standard point-of-care (POC) coagulation assays and devices fail to take this interaction into account. Optionally, the test well(s) and/or channel(s) of a consumable cartridge, such as those described herein, are coated with such surface proteins for the measurement of coagulation within a POC medical device.
The use of surface protein coatings includes collagen, vWF, fibronectin and any other molecule that modulates coagulation such as fibrinogen and thrombin. A layer of protein on a substrate (glass, polystyrene, polypropylene) creates binding sites that allow the mediation of receptor-ligand interactions between the substrate and other biological materials such as blood in a manner that improves the assessment of coagulation or provides new testing information.
The interior surfaces of a consumable cartridge can be coated using for example: (1) a layer of such proteins by covalent binding using linker molecules, (2) covalent binding using photochemistries or (3) simple protein adsorption. Linker molecules such as streptavidin or avidin and biotin can be used for this purpose. With linker molecules, the surface of any interior portion of the cartage that will be exposed to the biological sample is biotinylated (coated with a layer of biotin) using commercially available biotin that is conjugated to a reactive group that non-specifically and covalently binds with the substrate. A solution with a high concentration of streptavidin or avidin, which have high affinity for biotin, is added to create a layer of streptavidin/avidin bound biotin. Addition of biotinylated protein (collagen, vWF, fibronectin, thrombin, fibrinogen) then creates a layer of protein bound to the test well surface that specifically affects coagulation through interactions with plasma proteins and platelets.
Protein adsorption can be accomplished by filling the wells with a highly concentrated protein solution. Adsorption to the plastic surface takes place almost immediately depending on temperature, ph, surface charges, surface morphology and chemical composition. The solution can then be removed and the surface air dried. Brushing a highly concentrated protein solution on the surface of the wells or dipping the wells into such a solution will accomplish the same purpose.
The concentration of molecules in the solutions used for coating, whether using linker proteins or adsorption, can be changed to modulate the amount of protein that binds the substrate and, thus, modulate the effects on the coagulation cascade in a way that is relevant to physiology and hemostasis.
Referring again to
Between each couplant 134 and the open space of each test chamber is a rigid substrate 132. The rigid substrate and the couplant form an interface that focuses ultrasound transmitted (e.g. lens assembly) by an ultrasonic transducer into the chamber's open space and onto any biological fluid and/or reagents in the chamber. The rigid substrate of the lens can comprise a material which allows sound to pass and that can act to focus ultrasound at some level within the space. Optionally, the rigid substrate comprises a styrene, such as, for example Styrene® 666.
The lens assembly may be glued or welded to the surface 101 to secure the lens in place in an orientation that allows the desired focusing of sound. Alternatively, the lens assembly is optionally manufactured together with the surface 101. In this regard, the rigid substrate 132 can be molded with the surface 101 and the couplant 134 can be overmolded on the rigid substrate. A wide variety of materials can be used to construct the device. For example, plastics can be used for single use, disposable cartridges.
Each test chamber (116, 114, 112 and 110) can have a lens assembly positioned over the large opening of each chamber's open space. In this way, each chamber can be separately interrogated by focused ultrasound.
When placed in the analysis system 300, the couplant 134 can be placed in acoustic communication with a transducer for supplying ultrasound through the lens assembly and into a test chamber. Optionally, an intermediate layer of an acoustically permeable material is positioned between an ultrasonic transducer and the couplant. For example, and intermediate layer or block of Rexolite® can be used. The intermediate layer can be forced against the couplant and can be in acoustic contact with the transducer.
Sound generated by a transducer passes through the intermediate layer, through the couplant, through the rigid substrate, and is focused within the biological sample and reagent in the test chamber. Some of the sound directed into chamber contacts the distal interior surface 111 of the test chamber, which is defined by the surface 126. Optionally, the surface is polystyrene. The distal interior surface has a know geometry and is positioned at a know distance from the ultrasound source. The distal interior surface 111 is used as a calibrated reflector, which is used to estimate the speed of sound and attenuation of sound in a test chamber at base line and during the process of clot formation and clot dissolution. These measurements can be used, for example, to estimate hematocrit of the subject along with the indexes of hemostasis. The sound generated by the transducer can be focused within the biological sample in a test chamber using a parabolic mirror that is coupled to the biological sample using an elastomer.
d=f(x,y)+√{square root over ((xf−x)2+(yf−y)2+(zf−f(x,y))2)} (1)
Where d is the total distance from the face of the acoustic source to the focus. If the distance is set from the origin to the reflector as z0, then the total path-length is:
d=z0+√{square root over (xf2+yf2+(zf−z0)2)} (2)
The shape of the reflector can be determined by solving for f(x,y) as follows:
If z0 is set, then the equation 2 above can be evaluated and substituted into equation 10 above to yield an equation for the surface of the reflector. The reflector is a parabolic section. Example parameters are optionally an 8 mm aperture with a focus at 16 mm laterally, 4 mm in range and with an offset between the mirror and aperture of 0.5 mm. A diagram of this geometry is shown in
The cartridge 100 can be positioned into pocket 302 of an analysis system 300. As shown in
At least a portion of the sound is reflected by the biological sample positioned therein the chamber, and a portion of the sound transmitted into the chamber can also be reflected from the chamber distal surface 111. The reflected ultrasound can be received by the ultrasonic transducer and transmitted to the system for processing. Thus the cartridge and the analysis system 300 may be in communication such that data and other operational or processing signals may be communicated between the cartridge and the analysis system.
A suitable analysis system 300 can therefore comprise one or more processing devices. The processing of the disclosed methods, devices and systems can be performed by software components. Thus, the disclosed systems, devices, and methods, including the analysis system 300, can be described in the general context of computer-executable instructions, such as program modules, being executed by one or more computers or other devices. Generally, program modules comprise computer code, routines, programs, objects, components, data structures, etc. that perform particular tasks or implement particular abstract data types. For example, the program modules can be used to cause the transmission of ultrasound having desired transmit parameters and to receive and process ultrasound to evaluate hemostasis indices of a sample from the subject. The software can also be used to control the heating of the biological sample using the heat exchanger and to monitor and indicate the fill level of a given chamber. The processor can also be used to perform algorithms, to determine hemostatic indices and hematocrit. In some examples, the software can be used to back-out determined hematocrit from determined hemostatic indices. The determined hemostatic indices and hematocrit can be displayed to a medical professional or medical agent for the purpose of making medical decisions for a subject.
Thus, one skilled in the art will appreciate that the systems, devices, and methods disclosed herein can be implemented via a general-purpose computing device in the form of a computer. The computer, or portions thereof, may be located in the analysis system 300. The components of the computer can comprise, but are not limited to, one or more processors or processing units, a system memory, and a system bus that couples various system components including the processor to the system memory. In the case of multiple processing units, the system can utilize parallel computing.
The computer typically comprises a variety of computer readable media. Exemplary readable media can be any available media that is accessible by the computer and comprises, for example and not meant to be limiting, both volatile and non-volatile media, removable and non-removable media. The system memory comprises computer readable media in the form of volatile memory, such as random access memory (RAM), and/or non-volatile memory, such as read only memory (ROM). The system memory typically contains data such as data and/or program modules such as operating system and software that are immediately accessible to and/or are presently operated on by the processing unit.
In another aspect, the computer can also comprise other removable/non-removable, volatile/non-volatile computer storage media. By way of example, a mass storage device, which can provide non-volatile storage of computer code, computer readable instructions, data structures, program modules, and other data for the computer. For example and not meant to be limiting, a mass storage device can be a hard disk, a removable magnetic disk, a removable optical disk, magnetic cassettes or other magnetic storage devices, flash memory cards, CD-ROM, digital versatile disks (DVD) or other optical storage, random access memories (RAM), read only memories (ROM), electrically erasable programmable read-only memory (EEPROM), and the like.
Optionally, any number of program modules can be stored on the mass storage device, including by way of example, an operating system and software. Each of the operating system and software, or some combination thereof, can comprise elements of the programming and the software. Data can also be stored on the mass storage device. Data can be stored in any of one or more databases known in the art. Examples of such databases comprise, DB2®, Microsoft® Access, Microsoft® SQL Server, Oracle®, mySQL, PostgreSQL, and the like. The databases can be centralized or distributed across multiple systems.
In another aspect, the user can enter commands and information into the computer via an input device. Examples of such input devices comprise, but are not limited to, a keyboard, pointing device (e.g., a “mouse”), a touch screen, a scanner, and the like. These and other input devices can be connected to the processing unit via a human machine interface that is coupled to the system bus, but can be connected by other interface and bus structures, such as a parallel port, game port, an IEEE 1394 Port (also known as a Firewire port), a serial port, or a universal serial bus (USB).
In yet another aspect, a display device 304, such as a touch screen, can also be connected to the system bus via an interface, such as a display adapter. It is contemplated that the computer can have more than one display adapter and the computer can have more than one display device. For example, a display device can be a monitor, an LCD (Liquid Crystal Display), or a projector.
Any of the disclosed methods can be performed by computer readable instructions embodied on computer readable media. Computer readable media can be any available media that can be accessed by a computer. By way of example and not meant to be limiting, computer readable media can comprise computer storage media and communications media. Computer storage media comprise volatile and non-volatile, removable and non-removable media implemented in any method or technology for storage of information such as computer readable instructions, data structures, program modules, or other data.
The reagents in each test chamber, also referred to as a test well, can include all the reagents needed for evaluating one or more indices of hemostasis.
Optionally the cartridge is a single use, disposable cartridge with pre-loaded lyophilized reagents. The cartridge can be used with whole blood from a subject. The cartridge or assay components include the following for fresh whole blood samples. Four separate wells containing lyophilized reagents to which 1.6 ml of fresh whole blood is added. Each test well utilizes around 300 μl of fresh whole blood along with the following reagents:
The devices systems and methods use the phenomenon of acoustic radiation force to measure changes in mechanical properties (e.g. stiffness) of a blood sample during the processes of coagulation and fibrinolysis. These changes are representative of the role of the four key components of hemostasis: (i) plasma coagulation factors, (ii) platelets, (iii) fibrinogen, and (iv) fibrinolytic factors of the plasma. The basic approach is shown in
A series of N focused ultrasound pulses are sent into a blood sample at short intervals ΔT (ΔT is on the order of microseconds), as shown schematically in panel A. Each pulse generates a small and localized force within the blood as acoustic energy is absorbed and reflected during propagation. This force, which is concentrated around the focus of the ultrasound beam, induces a small displacement within the blood sample that depends upon the local mechanical properties. These displacements are on the order of 40 microns or less at the focus of the ultrasound beam.
Each pulse also returns an echo, as a portion of its energy is reflected from within the blood sample. Because the sample moves slightly from one pulse transmission to the next, the path length between the fixed ultrasound emitter and any given region within the target increases with pulse number. This change in path length can be estimated from differences in the arrival times of echoes from the same region. The ensemble of these delays forms a time-displacement curve that holds combined information about viscoelastic properties of the sample. These time-displacement curves are shown in
When the blood sample is in a viscous fluid state, the application of the acoustic force generates large displacements. As coagulation is activated and fibrinogen is cross-linked into fibrin strands, the sample behaves as viscoelastic solid and the induced displacement reduce as the stiffness of the sample increases. The interaction of platelets and the fibrin mesh also further reduce the induced displacements as the clot's stiffness increases. As the clot progresses into the phase of fibrinolysis, the fibrin mesh is dissolved by the fibrinolytic enzymes and the sample returns to viscous fluid, exhibiting increasing displacements.
The evolution of the magnitude of the induced displacements over time is therefore directly related to the changes in mechanical properties of the blood sample during hemostasis. A curve obtained with this method is shown in
Acoustic radiation force results from the transfer of momentum that occurs when a propagating acoustic wave is either absorbed or reflected. This body force acts in the direction of the propagating wave, and can be approximated by the following expression:
where α[m−1] is the acoustic attenuation coefficient, c [m/s] is the speed of sound, I(t) [W/m2] is the instantaneous intensity of the ultrasound beam, PII is the pulse intensity integral, ΔT [s] is the time interval between successive ultrasound pulse transmissions, and indicates a time averaged quantity.
The acoustic energy used by the instrument to generate acoustic radiation force is comparable with the acoustic energy typically used for common medical ultrasound procedures such as color Doppler imaging. The estimated maximum acoustic intensity is on the order of 2.5 W/cm2 (time average), which results in a temperature increase of the blood sample of 0.01° C. for each measurement ensemble (performed roughly every 6 seconds).
As the blood sample rapidly changes from viscous fluid to viscoelastic solid during coagulation and back to viscous fluid after clot lysis, the applied acoustic radiation force is adaptively changed to induce displacements above the noise threshold, but below levels that could induce mechanical disruption (typically below 40 microns).
The magnitude of the force is adjusted to follow the changes in mechanical properties of the blood sample by varying the time interval ΔT between successive pulses, as shown in equation 1. The maximum displacement induced during the (m−1)th acquisition is used to determine whether the force should be increased or decreased for the mth acquisition, based on predetermined threshold values. This adaptive process allows characterization of five orders of magnitude in stiffness without generating high strain within the blood sample that could alter the dynamics of coagulation and fibrinolysis.
As shown in equation (1), the applied acoustic radiation force changes as a function of acoustic attenuation and speed of sound, both of which change as a function of coagulation. The system uses the echoes returning from within the cartridge to estimate changes in these parameters and normalize the acoustic radiation force.
Acoustic radiation force is generated using conventional piezoelectric materials that act as acoustic emitters and receivers. These materials deform when a voltage is applied across them, and conversely generate a voltage when they are deformed. Similar to optics, an acoustic lens can be placed in front of the piezoelectric material to focus acoustic energy on a single focal point.
In the example systems, method, and devices piezoelectric disks are used that have an active diameter of 7.5 mm. The acoustic lens is provided by the curved shape of the disposable cartridge. Four disks are placed side by side to send sound in the four test wells in a disposable. The frequency of vibration of these piezoelectric disks is centered at 10 MHz, well within the range of frequencies used in conventional ultrasound imaging.
Ultrasound echo signals returning to the transducers from the blood samples are first filtered to remove electronic noise, digitized, and further processed within an embedded processor in the system. A flow chart of the data analysis steps performed by the system is shown in
Time Delay Estimation
Once an ensemble of N pulses is sent into the blood sample and the returning echoes are obtained, time delay estimation (TDE) is performed to estimate a local time-displacement curve, similar to that shown in
A variety of “off-the-shelf” algorithms are available to perform this operation. TDE is a common signal processing step in application fields ranging from RADAR, SONAR, and medical ultrasound imaging (Doppler).
Curve Fitting
The viscoelastic properties of the blood sample during hemostasis are modeled using a modified model consisting of the well-known Voigt-Kelvin mechanical model with the addition of inertia. While the dynamic changes in viscoelasticity of blood during hemostasis are certainly complex, the modified Voigt-Kelvin model is simple and robust, and it has been well validated in the past.
Each time-displacement curve is fitted to the characteristic equation of the modified Voigt-Kelvin model to estimate a variety of parameters relating to the viscoelastic properties of the sample. These parameters include relative elasticity, relative viscosity, time constant, and maximum displacement. The mathematical expression of the equation of motion for the modified Voigt-Kelvin model is
where ξ is the damping ratio, ω is the natural frequency, and s is the static sensitivity.
Among the parameters obtained by the curve fitting, the system uses the estimated displacement magnitude at 1 second as a qualitative measure of the stiffness of the sample. When blood is in viscous fluid state, the displacement at 1 second is high. As the blood coagulates this displacement decreases proportionally to the generation of the fibrin mesh and activity of platelets. The value increases again during the process of fibrinolysis.
Estimate Indices of Hemostatic Function
The displacement values obtained at 1 second for each data acquisition are compiled to form a curve showing relative stiffness as a function of time (
Indices of hemostasis are calculated by fitting a sigmoidal curve to the stiffness-time curve (
A summary of the parameters generated for each test chamber is presented in the table 2:
In order to isolate the four main components of hemostasis, four measurements are performed in parallel within the disposable cartridge using a combination of agonists and antagonists in each of four wells. The measurements in each well are combined to form indices of hemostasis as shown in the table 3:
Many modifications and other embodiments of the invention set forth herein will come to mind to one skilled in the art to which this invention pertains having the benefit of the teachings presented in the foregoing description. Therefore, it is to be understood that the invention is not to be limited to the specific embodiments disclosed and that modifications and other embodiments are intended to be included within the scope of the appended claims. Although specific terms are employed herein, they are used in a generic and descriptive sense only and not for purposes of limitation.
This application is a continuation application of U.S. patent application Ser. No. 13/397,398, filed Feb. 15, 2012, which is currently pending and which claims the benefit of and priority to U.S. Provisional Application No. 61/443,088, filed Feb. 15, 2011. Each of these applications is incorporated by reference herein in its entirety.
Number | Name | Date | Kind |
---|---|---|---|
4112740 | Brandestini | Sep 1978 | A |
4558589 | Hemmes et al. | Dec 1985 | A |
4695956 | Leveen et al. | Sep 1987 | A |
4705756 | Spillert et al. | Nov 1987 | A |
4814247 | Spillert et al. | Mar 1989 | A |
4852577 | Smith et al. | Aug 1989 | A |
4900679 | Spillert et al. | Feb 1990 | A |
5016469 | Henderson | May 1991 | A |
5056357 | Dymling et al. | Oct 1991 | A |
5104975 | McCormick et al. | Apr 1992 | A |
5205159 | Carr, Jr. | Apr 1993 | A |
5234839 | McCormick et al. | Aug 1993 | A |
5273517 | Barone et al. | Dec 1993 | A |
5311908 | Barone et al. | May 1994 | A |
5331964 | Trahey et al. | Jul 1994 | A |
5473536 | Wimmer | Dec 1995 | A |
5487387 | Trahey et al. | Jan 1996 | A |
5605154 | Ries et al. | Feb 1997 | A |
5606971 | Sarvazyan et al. | Mar 1997 | A |
5655535 | Friemel et al. | Aug 1997 | A |
5657760 | Ying et al. | Aug 1997 | A |
5673699 | Trahey et al. | Oct 1997 | A |
5744898 | Smith et al. | Apr 1998 | A |
5800781 | Gavin | Sep 1998 | A |
5810731 | Sarvazyan et al. | Sep 1998 | A |
5854423 | Venegas | Dec 1998 | A |
5899861 | Friemel et al. | May 1999 | A |
5921928 | Greenleaf et al. | Jul 1999 | A |
6039691 | Walker et al. | Mar 2000 | A |
6083159 | Driscoll, Jr. et al. | Jul 2000 | A |
6114135 | Goldstein | Sep 2000 | A |
6117081 | Jago et al. | Sep 2000 | A |
6135957 | Cohen et al. | Oct 2000 | A |
6213950 | Cespedes et al. | Apr 2001 | B1 |
6221672 | Baugh | Apr 2001 | B1 |
RE37171 | Busche et al. | May 2001 | E |
6225126 | Cohen et al. | May 2001 | B1 |
6264609 | Herrington et al. | Jul 2001 | B1 |
6270459 | Konofagou et al. | Aug 2001 | B1 |
6277074 | Chaturvedi et al. | Aug 2001 | B1 |
6283917 | Jago et al. | Sep 2001 | B1 |
6371912 | Nightinggale et al. | Apr 2002 | B1 |
6402704 | McMorrow | Jun 2002 | B1 |
6412344 | Danicich et al. | Jul 2002 | B1 |
6454714 | Ng et al. | Sep 2002 | B1 |
6494834 | Konofagou et al. | Dec 2002 | B2 |
6508768 | Hall et al. | Jan 2003 | B1 |
6514204 | Alam et al. | Feb 2003 | B2 |
6535835 | Rubin et al. | Mar 2003 | B1 |
6537819 | Cohen et al. | Mar 2003 | B2 |
6573104 | Carr, Jr. et al. | Jun 2003 | B2 |
6613573 | Cohen | Sep 2003 | B1 |
6632678 | Aiken et al. | Oct 2003 | B2 |
6685646 | Cespedes et al. | Feb 2004 | B2 |
6687625 | Srinivasan et al. | Feb 2004 | B2 |
6692439 | Walker et al. | Feb 2004 | B1 |
6716168 | Nock et al. | Apr 2004 | B2 |
6726629 | Frinking et al. | Apr 2004 | B1 |
6764448 | Trahey et al. | Jul 2004 | B2 |
6787363 | Cohen et al. | Sep 2004 | B2 |
6797519 | Cohen et al. | Sep 2004 | B2 |
6890299 | Cohen et al. | May 2005 | B2 |
6951544 | Trahey et al. | Oct 2005 | B2 |
7179652 | Cohen et al. | Feb 2007 | B2 |
7192726 | Carr, Jr. et al. | Mar 2007 | B1 |
7202048 | Carr, Jr. | Apr 2007 | B2 |
7207939 | Husher | Apr 2007 | B2 |
7261861 | Kautzky | Aug 2007 | B2 |
7374538 | Nightingale et al. | May 2008 | B2 |
7399637 | Wright et al. | Jul 2008 | B2 |
7422905 | Clague et al. | Sep 2008 | B2 |
7439069 | Nippoldt et al. | Oct 2008 | B2 |
7524670 | Cohen et al. | Apr 2009 | B2 |
7732213 | Cohen et al. | Jun 2010 | B2 |
7912661 | Zeng | Mar 2011 | B2 |
7972271 | Johnson et al. | Jul 2011 | B2 |
8058023 | Gurbel | Nov 2011 | B2 |
8740818 | Walker et al. | Jun 2014 | B2 |
20020013530 | Cespedes et al. | Jan 2002 | A1 |
20020040187 | Alam et al. | Apr 2002 | A1 |
20020081741 | Braun, Sr. | Jun 2002 | A1 |
20030013958 | Govari et al. | Jan 2003 | A1 |
20030073244 | Cohen et al. | Apr 2003 | A1 |
20030105398 | Vitek | Jun 2003 | A1 |
20030113929 | Baugh | Jun 2003 | A1 |
20030170883 | Martin | Sep 2003 | A1 |
20030171676 | Trahey et al. | Sep 2003 | A1 |
20030204141 | Nock et al. | Oct 2003 | A1 |
20040068184 | Trahey et al. | Apr 2004 | A1 |
20040076546 | Bissett | Apr 2004 | A1 |
20040167403 | Nightingale et al. | Aug 2004 | A1 |
20040214337 | Kautzky | Oct 2004 | A1 |
20050004463 | Chen et al. | Jan 2005 | A1 |
20050015001 | Lec | Jan 2005 | A1 |
20050053305 | Li et al. | Mar 2005 | A1 |
20050148899 | Walker et al. | Jul 2005 | A1 |
20050215901 | Anderson et al. | Sep 2005 | A1 |
20050216987 | Murakami | Sep 2005 | P1 |
20070038095 | Greenleaf et al. | Feb 2007 | A1 |
20070059840 | Cohen et al. | Mar 2007 | A1 |
20070184508 | Cohen et al. | Aug 2007 | A1 |
20070276236 | Jong | Nov 2007 | A1 |
20080038828 | Cohen et al. | Feb 2008 | A1 |
20080194967 | Sliwa et al. | Aug 2008 | A1 |
20080249408 | Palmeri et al. | Oct 2008 | A1 |
20080261261 | Grimes et al. | Oct 2008 | A1 |
20080297169 | Greenquist | Dec 2008 | A1 |
20090112483 | Cohen | Apr 2009 | A1 |
20100274130 | Anand | Oct 2010 | A1 |
20110034805 | Walker et al. | Feb 2011 | A1 |
20110151491 | Dennis et al. | Jun 2011 | A1 |
20110172661 | Desinger et al. | Jul 2011 | A1 |
20110252352 | Viola et al. | Oct 2011 | A1 |
20120232803 | Viola et al. | Sep 2012 | A1 |
20120252127 | Bansil et al. | Oct 2012 | A1 |
20120294767 | Viola et al. | Nov 2012 | A1 |
20130190584 | Walker et al. | Jul 2013 | A1 |
Number | Date | Country |
---|---|---|
2011237383 | Jul 2014 | AU |
1816306 | Aug 2006 | CN |
101035479 | Sep 2007 | CN |
2555704 | Feb 2013 | EP |
2676143 | Dec 2013 | EP |
2011035162 | Mar 2011 | WO |
2011127436 | Oct 2011 | WO |
2012159021 | Feb 2012 | WO |
2013105987 | Oct 2013 | WO |
Entry |
---|
Amukele, et al., “Comparison of plasma with whole blood prothrombin time and fibrinogen on the same instrument,” American Journal of Clinical Pathology, vol. 133, No. 4, Apr. 2010, pp. 550-556. |
Anderson, “Multi-Dimensional Velocity Estimation with Ultrasound Using Spatial Quadrature,” IEEE Transactions on Ultrasonics Ferroelectrics & Frequency Control, vol. 45, No. 3, 1998, pp. 852-861. |
Anderson, “Preventing Deep Vein Thrombosis and Pulmonary Embolism,” Center for Outcomes Research, U Mass Med Center, 1998, 23 pages. |
Becker, R., “Cell-based models of coagulation: a paradigm in evolution,” Journal of Thrombosis and Thrombolysis, vol. 20, No. 1, Aug. 2005, pp. 65-68. |
Beer: Center for Reproductive Immunology & Genetics, “Thrombophilia: Inherited and Acquired,” 6 pages, http://repro-med.net/papers/thromb.php. Mar. 30, 2005. |
Bell, et al., “Thrombelastographic evaluation of coagulation in transurethral prostatectomy,” British Journal of Urology, vol. 78, No. 5, 1996, pp. 737-741. |
Bercoff et al., “In vivo breast tumor detection using transient elastography,” Ultrasound in Medicine & Biology, vol. 29, No. 10, 2003, pp. 1387-1396. |
Bercoff, et al., “Supersonic Shear Imaging: A New Technique for Soft Tissue Elasticity Mapping,” IEEE Transactions on Ultrasonics Ferroelectrics & Frequency Control, vol. 51, No. 4, 2004, pp. 396-409. |
Bilgen, et al., “Error analysis in acoustic elastography. II. Strain estimation and SNR analysis”, Journal of the Acoustical Society of America, vol. 101, 1997, pp. 1147-1154. |
Bohs, et al., “A Real Time System for Quantifying and Displaying Two-Dimensional Velocities using Ultrasound,” Ultrasound in Medicine & Biology, vol. 19, No. 9, Jul. 1993, pp. 751-761. |
Bombeli, et al., “Updates in perioperative coagulation: physiology and management of thromboembolism and haemorrhage,” British Journal of Anaesthesia; vol. 93, No. 2, Aug. 2004, pp. 275-287. |
Bonnefous, et al., “Time Domain Formulation of Pulse-Doppler Ultrasound and Blood Velocity Estimation by Cross Correlation,” Ultrasonic Imaging 8, 1986, pp. 73-85. |
Brock, et al., “Assessing Thrombin Generation at the Point of Care,” Clinical Chemistry, vol. 55, No. 3, Mar. 2009, pp. 398-399. |
Carr, M., “In vitro assessment of platelet function,” Transfusion of Medicine Reviews, vol. 11, No. 2, Apr. 1997, pp. 106-115. |
Carroll, et al., “Measurement of functional fibrinogen levels using the Thrombelastograph,” Journal of Clinical Anesthesia, vol. 20, No. 3, May 2008, pp. 186-190. |
Carter, G., “Coherence and time delay estimation,” Proc IEEE, vol. 75, No. 2, 1987, pp. 236-255. |
Chakroun et al., “The influence of fibrin polymerization and platelet-mediated contractile forces on citrated whole blood thromboelastography profile,” Thrombosis and Haemostasis, vol. 95, No. 5, May 2006, pp. 822-828. |
Chandler, et al., “Development of a rapid emergency hemorrhage panel,” Tranfusion, vol. 50, No. 12, Dec. 2010, pp. 2547-2552. |
Chandler, et al., “Estimating the rate of thrombin and fibrin generation in vivo during cardiopulmonary bypass,” Blood, vol. 101, No. 11, Jun. 2003, pp. 4355-4362. |
Chaturvedi, et al., “Testing the limitations of 2-D companding for strain imaging using phantoms,” IEEE Transactions on Ultrasonics Ferroelectrics & Frequency Control, vol. 45, 1998, pp. 1022-1031. |
Chavez, J., “A novel thrombelastograph tissue factor/kaolin assay of activated clotting times for monitoring heparin anticoagulation during cardiopulmonary bypass,” Anesthesia and Analgesia; vol. 99, No. 5 Nov. 2004, pp. 1290-1294. |
Cohn et al., “An elasticity microscope. Part I: Methods,” IEEE Transactions on Ultrasonics Ferroelectrics & Frequency Control, vol. 44, 1997, pp. 1304-1319. |
Cohn et al., “An elasticity microscope. Part II: Experimental Results,” IEEE Transactions on Ultrasonics Ferroelectrics & Frequency Control, vol. 44, 1997, pp. 1320-1331. |
Craft, et al., “A novel modification of the Thrombelastograph assay, isolating platelet function, correlates with optical platelet aggregation,” The Journal of Laboratory and Clinical Medicine, vol. 143, No. 5, May 2004, pp. 301-309. |
Curry, et al., “Convention and near-patient tests of coagulation,” British Journal of Anaesthesia, vol. 7, No. 2, Apr. 2007, pp. 45-50. |
Dahlback, B., “Blood Coagulation,” Lancet, vol. 355, No. 9215, May 2000, pp. 1627-1632. |
Despotis, et al., “Monitoring of hemostasis in cardiac surgical patients: impact of point-of-care testing on blood loss and transfusion outcomes,” Clinical Chemistry, vol. 43, No. 9, Sep. 1997, pp. 1684-1696. |
Embree, et al., “Volumetric Blood Flow via Time-Domain Correlation: Experimental Verification,” IEEE Transactions on Ultrasonics Ferroelectrics & Frequency Control, vol. 37, No. 2, May 1990, pp. 176-189. |
Emelianov et al., “Ultrasound Elasticity Imaging of Deep Venous Thrombosis,” Proc. IEEE Ultrasonics Symp., 2000, pp. 1791-1794. |
Evans, et al., “Rheometry and associated techniques for blood coagulation studies,” Medical Engineering and Physics, vol. 30, No. 6, Jul. 2008, pp. 671-679. |
Fatemi et al., “C-Scan Imaging by Radiation Force Stimulated Acoustic Emission Method,” Proc. IEEE Ultrasonics Symp., 1996, pp. 1459-1462. |
Fatemi, et al., “Application of radiation force in noncontact measurement of the elastic parameters,” Ultrasonic Imaging, vol. 21, No. 2, Apr. 1999 pp. 147-154. |
Fatemi, et al., “Ultrasound-Stimulated Vibro-Acoustic Spectography,” Science Magazine, vol. 280, No. 5360, 1998, pp. 82-85. |
Ferraris, et al., “2011 Update to the Society of Thoracic Surgeons and the Society of Cardiovascular Anesthesiologists Blood Conservation Clinical Practice Guidelines,” Annals of Thoracic Surgery, vol. 91, 2011, pp. 944-982. |
Fertner, et al., “Comparison of Various Time Delay Estimation Methods by Comptuer Simulation,” IEEE Transactions on Acoustics, Speech, and Signal Processing, vol. 34, No. 5, 1986, pp. 1329-1330. |
Flax, et al., “Phase-Aberration Correction Using Signals From Point Reflectors and Diffuse Scatterers: Basic Principles,” IEEE Transactions on Ultrasonics Ferroelectrics & Frequency Control, vol. 35, No. 6, Nov. 1988, pp. 758-767. |
Freedman, et al., “A Meta-Analysis of Thromboembolic Prophylaxis Following Elective Total Hip Arthroplasty,” Journal of Bone and Joint Surgery, vol. 82-A, 2000, pp. 929-938. |
Gaetano, G. de, et al., “Effect of Platelets on Clot Structuration, a Thrombelastographic Study,” Thrombosis Research, vol. 3, No. 4, pp. 425-435, 1973. |
Ganter, et al., “Active, personalized, and balanced coagulation management saves lives in patients with massive bleeding,” Anesthesiology, vol. 133, No. 5, Nov. 2010, pp. 1016-1018. |
Gallippi, et al., “Adaptive clutter filtering via blind source,” Ultrasonic Imaging, vol. 24, No. 4, 2002, pp. 193-214. |
Gallippi, et al., “BSS-based filtering of physiological and ARFI-induced tissue and blood motion,” Ultrasound in Medicine and Biology, vol. 29, No. 11, 2003, pp. 1583-1592. |
Gallippi, et al., “Complex blind source separation for acoustic radiation force impulse imaging in the peripheral vasculature, in vivo,” IEEE Ultrasonics Symposium, vol. 1, 2004, pp. 596-601. |
Ganter, et al., “Coagulation monitoring: current techniques and clinical use of viscoelastic point-of-care coagulation devices,” Anesthesia and Analgesia, vol. 106, No. 5, May 2008, pp. 1366-1375. |
Gauss, et al., “Adaptive Imagining in the Thyroid Using Fundamental and Harmonic Echo Data,” presented at IEEE Ultrasonics Symposium, 1999, pp. 1515-1519. |
Gauss, et al., “Wavefront Estimation in the Human Breast,” presented at SPIE Medical Imaging, vol. 4325, 2001, pp. 172-180. |
Giunta, et al., “Estimation of Global Motion Parameters by Complex Linear Regression,” IEEE Transactions on Image Processing, vol. 8, No. 11, 1999, pp. 1652-1657. |
Glidden, Paul F., et al., “Thromboelastograph Assay for Measuring the Mechanical Strength of Fibrin Sealant Clots,” Clinical and Applied Thombosis/Hemostasis, vol. 6, No. 4, Oct. 2000, pp. 226-233. |
Gottumukkala, Vijaya N., et al., “Assessing Platelet and Fibrinogen Contribution to Clot Strength Using Modified Thromboelastography in Pregnant Women,” Anesth Analg, vol. 89, 1999, pp. 1453-1455. |
Greilich, Philip E., et al., “A Modified Thromboelastographic Method for Monitoring c7E3 Fab in Heparinized Patients,” Anesth Analg, vol. 84, 1997, pp. 31-38. |
Greilich, Philip E., et al., “Near-Site Monitoring of the Antiplatelet Drug Abciximad Using the Hemodyne Analyzer and Modified Thrombelastograph,” Journal of Cardiothoracic and Vascular Anesthesis, vol. 13, No. 1, Feb. 1999, pp. 58-64. |
Gurbel, et al., “Platelet function monitoring in patients with coronary artery disease,” Journal of the American College of Cardiology, vol. 50, No. 19, Nov. 2007, pp. 1822-1834. |
Hardisty R. M. et al, “Fibrinogen as a Co-factor in the Reaction of Platelets with Kaolin,” May 7, 1966, Nature Publishing Group, Edition 210, vol. 644 (http://www.nature.com/nature/journal/v210/n5036/abs/210644a0.html). |
Harris, et al., “Evaluation of recurrent thrombosis and hypercoagulability,” American Family Physician, vol. 56, No. 6, Oct. 1997, pp. 1591-1596, pp. 1601-1602. |
Hartley, et al., “Characteristics of Acoustic Streaming Created and Measured by Pulsed Doppler Ultrasound,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 44, No. 6, Nov. 1997, pp. 1278-1285. |
Hartley, et al., “Doppler Measurement of Acoustic Streaming,” IEEE Ultrasonics Symposium Proceedings, 1995, pp. 1537-1540. |
Hett, et al., “Sonoclot Analysis,” British Journal of Anaesthesia, vol. 75, No. 6, Dec. 1995, pp. 771-776. |
Hirsh, et al., “How we diagnose and treat deep vein thrombosis,” Blood, vol. 99, 2002, pp. 3102-3110. |
Hirsh, et al., “Management of deep vein thrombosis and pulmonary embolism. A statement for healthcare professionals,” Council on Thrombosis (in consultation with the Council on Cardiovascular Radiology), American Heart Association, vol. 93, 1996, 55 pages. |
Hoffman, et al., “A cell-based model of hemostasis,” Thrombosis and Haemostasis, vol. 85, No. 6, Jun. 2001, pp. 958-965. |
Huang, et al., “Characterization of Blood Properties from Coagulating Blood of Different Hematocrits Using Ultrasonic Backscatter and Attenuation”, Japanese Journal of Applied Physics, vol. 45, No. 9A, 2006, pp. 7191-7196. |
Huang, et al., “Detection of blood coagulation and clot formation using quantitative ultrasonic parameters,” Ultrasound in Medicine and Biology, vol. 31, No. 11, Nov. 2005, pp. 1567-1573. |
Ickx, Brigitte, “Point-of-Care Monitoring of Haemostasis in the OR and the ICU,” European Society of Anaesthesiologists. Jun. 5, 2004, pp. 79-83. |
International Search Report and Written Opinion of the International Searching Authority, received in corresponding application PCT/US2010/049342, Nov. 16, 2010. |
International Search Report and Written Opinion of the International Searching Authority, received in corresponding application PCT/US2011/031832, Dec. 15, 2011. |
International Preliminary Report on Patentability and Written Opinion, dated Oct. 8, 2013, in connection with International Application No. PCT/US2012/025270. |
International Search Report, dated Sep. 30, 2013, in connection with International Application No. PCT/US2012/025270. |
International Preliminary Report on Patentability and Written Opinion, dated Aug. 27, 2013, in connection with International Application No. PCT/US2012/025278. |
International Search Report, dated Aug. 20, 2013, in connection with International Application No. PCT/US2012/025278. |
International Preliminary Report on Patentability and Written Opinion, dated Nov. 19, 2013, in connection with International Application No. PCT/US2012/038553. |
International Search Report, dated Jan. 2, 2013, in connection with International Application No. PCT/US2012/038553. |
International Preliminary Report on Patentability and Written Opinion, dated Oct. 9, 2012, in connection with International Application No. PCT/US2011/031832. |
International Preliminary Report on Patentability and Written Opinion, dated Mar. 20, 2012, in connection with International Application No. PCT/US2010/049342. |
Ivandic, et al., “Determination of clopidogrel resistance by whole blood platelet aggregometry and inhibitors of the P2Y12 receptor,” Clinical Chemistry, vol. 52, No. 3, Mar. 2006, pp. 383-388. |
Jacovitti, et al., “Discrete Time Techniques for Time Delay Estimation,” IEEE Transactions on Signal Processing, vol. 41, No. 2, Feb. 1993, pp. 525-533. |
Jensen, “A New Method for Estimation of Velocity Vectors,” IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control, vol. 45, No. 3, 1998, pp. 837-851. |
Jensen, Estimation of Blood Velocities Using Ultrasound, 1996, pp. 195-225. |
Jensen, et al., “Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducers,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 39, 1992, pp. 262-267. |
Jolliffe, IT., “Principal Component Analysis,” Springer Series in Statistics, 2nd edition, Springer, NY, 2002, pp. 1-8. |
Kadi, et al., “On the performance of regression and step-initialized IIR Clutter Filters,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 42, 1995, pp. 827-837. |
Kasai, et al., “Real-time Two-Dimensional Blood Flow Imaging Using an Autocorrelation Technique,” IEEE Ultrasonics Symposium, vol. 32, No. 3, 1985, pp. 458-464. |
Katori, et al., “The effects of platelet count on clot retraction and tissue plasminogen activator-induced fibrinolysis on thrombelastography,” Anesthesia and Analgesia, vol. 100, No. 6, Jun. 2005, pp. 1781-1785. |
Keresztes, et al., “The PFA-100: analysis and interpretation of a platelet function measurement,” The Journal of Cardiovascular Nursing, vol. 20, No. 6, 2005, pp. 405-407. |
Kettner, S.C., et al., “Use of abciximab-Modified Thrombelatography in Patients Undergoing Cardiac Surgery,” Anesth Analg, vol. 89, 1999, pp. 580-584. |
Khurana, Sandeep, et al., “Monitoring Platelet Glycoprotein IIb/IIa-fibrin Interaction with Tissue Factor-Activated Thromboelastography,” J Lab Clin Med, vol. 130, No. 4, 1997, pp. 401-411. |
Khurana, Sandeep, et al., “Thromboelastography Can Rapidly Bioassay Fibrinogen,” Anesthesiology, vol. 85, No. 3A, Sep. 1996, p. A457. |
Koepke, J., “Point-of-Care Coagulation Testing,” Laboratory Medicine, vol. 31, No. 6, Jun. 2000, pp. 343-346. |
Kruse, et al., “A new high resolution color flow system using an eigendecomposition-based adaptive filter for clutter rejection,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 49, No. 10, 2002, pp. 1384-1399. |
Ledoux, et al., “Reduction of the clutter component in Doppler ultrasound signals based on singular value decomposition: a simulation study,” vol. 19, No. 1, 1997, pp. 1-18. |
Lerner, et al., “Sono-elasticity: medical elasticity images derived from ultrasound signals in mechanically vibrated targets,” Ultrasound in Medicine & Biology, vol. 16, 1998, pp. 317-327. |
Libgot, R., et al., “High frequency ultrasound characterization of the blood clotting process: intra- and inter-individual variations,” 2005 IEEE Ultrasonics Symposium, IEEE, vol. 4, 2005, pp. 2259-2262. |
Loupas, et al., “An axial Velocity Estimator for Ultrasound Blood flow imaging, by means of a two-dimensional autocorrelation approach,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 42, No. 4, 1995, pp. 672-688. |
Lubinski, et al., “Adaptive strain estimation using retrospective processing medical US elasticity imaging,” IEEE Transactions on Ultrasonics Ferroelectrics & Frequency Control, vol. 46, 1999, pp. 97-107. |
Mahla, et al., “Thromboelastography for monitoring prolonged hypercoagulability after major abdominal surgery,” Anesthesia and Analgesia, vol. 92, No. 3, Mar. 2001, pp. 572-577. |
Malinin, et al., “Validation of a VerifyNow-P2Y12 cartridge for monitoring platelet inhibition with clopidogrel,” Methods and Findings in Experimental and Clinical Pharmacology, vol. 28, No. 5, Jun. 2006, pp. 315-322. |
Mauldin, et al., “Robust Principal Component Analysis and Clustering Methods for Automated Classification of Tissue Response to ARFI Excitation,” Ultrasound in Medicine & Biology, vol. 34, No. 2, 2008, pp. 309-325. |
Mauldin, Jr., F.W., et al., “Adaptive Force Sonorheometry for Assessment of Whole Blood Coagulation,” Clin Chim Acta, vol. 411, Issues 9-10, 2010, pp. 638-644. |
Ng, et al., “A Comparative Evaluation of Several Algorithms for Phase Aberration Correction,” IEEE Transactions on Ultrasonics Ferroelectrics & Frequency Control, vol. 41, No. 5, Sep. 1994, pp. 631-643. |
McAleavey, et al., “Estimates of echo correlation and measurement bias in acoustic radiation force impulse imaging,” IEEE Transactions on Ultrasonics Ferroelectrics & Frequency Control, vol. 50, 2003, pp. 631-641. |
Nielson, er al., “Effects of coagulation factor deficiency on plasma coagulation kinetics determined via thrombelastography: critical roles of fibrinogen and factors II, VII, X and XII,” Acta Anesthesiologica Scandanavia, vol. 49, No. 2, Feb. 2005, pp. 222-231. |
Nightingale, et al., “Shear-Wave Generation Using Acoustic Radiation Force: In Vivo and Ex Vivo Results,” Ultrasound in Medicine & Biology, vol. 29, No. 12, 2003, pp. 1715-1723. |
Nightingale, et al., “Acoustic Radiation Force Impulse Imaging: In Vivo Demonstration of Clinical Feasibility,” Ultrasound in Medicine & Biology, vol. 28, 2002, pp. 227-235. |
Nightingale, et al., “Acoustic remote palpation: initial in vivo results,” presented at IEEE Ultrasonics Symposium, 2000, pp. 1553-1558. |
Oberhardt, et al., “Dry reagent technology for rapid, convenient measurements of blood coagulation and fibrinolysis,” Clinical Chemistry, vol. 37, No. 4, Apr. 1991, pp. 520-526. |
O'Donnell, et al., “Role of the Thrombelastograph as an adjunctive test in thrombophilia screening,” Blood Coagulation and Fibrinolysis, vol. 15, No. 3, Apr. 2004, pp. 207-211. |
O'Donnell, et al., “Internal Displacement and Strain Imaging using Ultrasonic Speckle Tracking,” IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control, vol. 41, 1994, pp. 314-325. |
Ophir, et al., “Elastography: A Quantitative Method for Imaging the Elasticity of Biological Tissues,” Ultrasonic Imaging, vol. 13, No. 2, 1991, pp. 111-134. |
Packham, M., “Role of platelets in thrombosis and hemostasis,” Canadian Journal of Physiology and Pharmacology, vol. 72, No. 3, Mar. 1994, pp. 278-284. |
Palmeri, et al., “Ultrasonic tracking of acoustic radiation force-induced displacements in homogeneous media,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 53, No. 7, 2006, pp. 1300-1313. |
Parsons, et al., “Age Determiniation of Experimental Venous Thrombi by Ultrasonic Tissue Characterization,” Journal of Vascular Surgery, vol. 17, 1993, 470-478. |
Patil, et al., “3D prostate elastography: algorithm, simulations and experiments,” Physics in Medicine & Biology, vol. 52, No. 12, 2007, pp. 3643-3663. |
Perry, et al., “Point-of-care testing in haemostasis,” British Journal of Haematology, vol. 150, No. 5, Sep. 2010, pp. 501-514. |
Pivalizza, et al., “Perioperative thromboelastography and sonoclot analysis in morbidly obese patients,” Canadian Journal of Anaesthesia, vol. 44, No. 9, Sep. 1997, pp. 942-945. |
Rao, G., “Need for a point-of-care assay for monitoring antiplatelet and antithrombotic therapies,” Stroke, vol. 40, No. 6, Jun. 2009, pp. 2271-2272. |
Riou, et al., “Fast adaptive eigenvalue decomposition: a maximum likelihood approach,” IEEE International Conference on Acoustics, Speech, and Signal Processing, vol. 5, 1997, pp. 3565-3568. |
Rubin, et al., “Clinical application of sonographic elasticity imaging for aging of deep venous thrombosis: preliminary findings,” Journal of Ultrasound in Medicine, vol. 22, 2003, pp. 443-448. |
Sakharov, et al., “Acceleration of Fibrinolysis by High-Frequency Ultrasound: The Contribution of Acoustic Streaming and Temperature Rise,” Thrombosis Research, vol. 100, No. 4, 2000, pp. 333-340. |
Sarvazyan, et al., “Shear Wave Elasticity Imagining—A New Ultrasonic Technology of Medical Diagnostics,” Ultrasound in Medicine and Biology, vol. 24, 1998, pp. 1419-1436. |
Schmitt, C., et al., “Characterization of blood clot viscoelasticity by dynamic ultrasound elastography and modeling of the rheological behavior,” Journal of Biomechanics, vol. 44, No. 4, 2011, pp. 622-629. |
Shi, Quantitative Investigation of Acoustic Streaming in Blood, J. Acoust. Soc. Am. 111, Feb. 2002, pp. 1110-1121. |
Shi, et al., “Color Doppler Detection of Acoustic Streaming in a Hematoma Model,” Ultrasound in Medicine and Biology, vol. 27, No. 9, 2001, pp. 1255-1264. |
Shi, et al., “Color Doppler imaging of acoustic streaming in blood and clot,” IEEE Ultrasonics Symposium, vol. 2, 1999, pp. 1315-1318. |
Shi, et al., “Experimental Investigation and Finite Element Simulation of Streaming in Blood in Cylindrical Models,” IEEE Ultrasonics Symposium, vol. 2, 2000, pp. 1509-1512. |
Shih, C-C, et al., “In Vitro Assessments of Viscoelastic Properties of Fibrin Clot by Using Acoustic Radiation Force on a Solid Sphere,” International Ultrasonics Symposium Proceedings, IEEE, 2010, pp. 479-482. |
Shung, et al., “Ultrasonic characterization of blood during coagulation,” Journal of Clinical Ultrasound, vol. 12, No. 3, 1984, pp. 147-153. |
Skovoroda, et al., “Tissue elasticity reconstruction based on ultrasonic displacement and strain images,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 42, No. 4, 1995, pp. 747-765. |
Srinivasan, et al., “Elastographic imaging using staggered strain estimates,” Ultrasonic Imaging, vol. 24, 2002, pp. 229-245. |
Strobach, P., “Low-rank adaptive filters,” IEEE Trans Signal Process, vol. 44, No. 12, 1996, pp. 2932-2947. |
Sugimoto, et al., “Tissue Hardness Measurement Using the Radiation Force of Focused Ultrasound,” Proc. IEEE Ultrason. Symp., 1990, pp. 1377-1380. |
Sumino, et al., “Measurements of ultrasonic pulse arrival time differences produced by abdominal wall specimens,” Journal of the Acoustical Society of America, vol. 90, No. 6, 1991, pp. 2924-2930. |
Thuerlemann, et al., “Monitoring thrombin generation by electrochemistry: development of an amperometric biosensor screening test for plasma and whole blood,” Clinical Chemistry, vol. 55, No. 3, Mar. 2009, pp. 505-512. |
Toner, et al., “Blood-on-a-chip,” Annual Review of Biomedical Engineering, vol. 7, 2005, pp. 77-103. |
Torr, “The Acoustic Radiation Force,” Am. J. Phys., vol. 52, 1984, pp. 402-408. |
Trahey, et al., “Synthetic receive aperture imaging with correction for motion and for tissue inhomogeneities. II. Effects of and correction for motion,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 39, No. 4, 1992, pp. 496-501. |
Traverso C, Arcelus JI, Gomez E, Luna D, Lopez-Cantarero M, Garcia JM. “Prospective assessment of the risk of deep vein thrombosis in elective abdominal surgery. Predictive role of [Thrombelastograph® analysis].” Thromb Haemorrh Disorders. 1993:71:9-15. |
Vig, et al., “Thromboelastography: a reliable test?,” Blood Coagulation and Fibrinolysis, vol. 12, No. 7, Oct. 2001, 555-561. |
Viola, et al., “A Comparison between spline-based and phase-domain time-delay estimators,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 53, No. 3, 2006, pp. 515-517. |
Viola, Francesco, et al., “A Novel Ultrasound-Based Method to Evaluate Hemostatic Function of Whole Blood,” Clin Chim Acta, vol. 411, Nos. 1-2, 2010, pp. 106-113. |
Viola, et al., “A comparison of the performance of time-delay estimators in medical ultrasound,” IEEE Transactions on Ultrasonics Ferroelectrics & Frequency Control., vol. 50, 2003, pp. 392-401. |
Viola, et al., “A Spline Based Algorithm for Continuous Time Delay Estimation Using Sampled Data,” IEEE Transactions on Ultrasonics Ferroelectrics & Frequency Control, in press, 2005, pp. 80-93. |
Viola, et al., “Analysis of Clot Formation with Acoustic Radiation Force,” SPIE Proceedings, vol. 4689, 2002, pp. 235-242 and pp. 1-2. |
Viola, et al., “Comparison of Time Delay Estimators in Medical Ultrasound,” IEEE Ultrasonics Symposium, vol. 2, 2001, pp. 1485-1488. |
Viola, et al., “Efficient and Accurate Spline-Based Time Delay Estimation,” IEEE Ultrasonics Symposium, vol. 2, 2004, pp. 870-873. |
Viola, et al., “Imaging Viscoelastic Properties of the Vitreous,” Ultrasonics Symposium, vol. 2, 2001, pp. 1623-1626. |
Viola, et al., “Radiation Force Imaging of Viscoelastic Properties with Reduce Artifacts,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 50, No. 6, 2003, pp. 736-742. |
Viola, et al., “Sonorheometry: A new Method for Assessing coagulation potential,” IEEE Ultrasonics Symposium, vol. 1, 2007, pp. 1001-1004. |
Viola, et al., “Sonorheometry: A Noncontact Method for the Dynamic Assessment of Thrombosis,” The Annals of Biomedical Engineering, vol. 32, 2004, pp. 696-705. |
Viola, et al., “Ultrasound echo decorrelation due to acoustic radiation force,” IEEE Ultrasonics Symposium Proceedings, vol. 2, 2002, pp. 1903-1906. |
Volei{hacek over (s)}is, A., et al., “Ultrasonic method for the whole blood coagulation analysis,” Ultrasonics, vol. 40, May 2002, pp. 101-107. |
Walker, et al., “A Fundamental Limit on Delay Estimation Using Partially Correlated Speckle Signals,” IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control, vol. 42, 1995, pp. 301-308. |
Walker, et al., “A Fundamental Limit on the Accuracy of Speckle Signal Alignment,” IEEE Ultrasonics Symposium Proceedings, vol. 3, 1994, pp. 1787-1791. |
Walker, et al., “A Method of Imagining Viscoelastic Parameters with Acoustic Radiation Force,” Physics in Medicine and Biology, vol. 45, No. 6, 2000, pp. 1437-1447. |
Walker, et al., “Application of Acoustic Radiation Force in Ophthalmic Ultrasound,” Proc. IEEE Ultrason. Symp., vol. 2, 1997, pp. 1291-1295. |
Walker, et al., “Real-Time Imaging of Tissue Vibration Using a Two-Dimensional Speckle Tracking System,” IEEE Ultrason. Symp., 1993, pp. 873-877. |
Walker, et al., “The Significance of Correlation in Ultrasound Signal Processing,” SPIE Proceedings, vol. 4325, 2001, pp. 159-171. |
Westbrook, et al., “Protocol based on thromboelastograph (TEG) out-performs physician preference using laboratory coagulation tests to guide blood replacement during and after cardiac surgery: a pilot study,” Heart, Lung, and Circulation, vol. 18, No. 4, Aug. 2009, pp. 277-288. |
Whitten, et al., “Thromboelastography: past, present, and future,” Anesthesiology, vol. 92, No. 5, May 2000, pp. 1223-1225. |
Yu, et al., “Single-Ensemble-Based Eigen-Processing Methods for Color Flow Imaging,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Controls, vol. 55, No. 3, 2008, pp. 573-587. |
Number | Date | Country | |
---|---|---|---|
20160139159 A1 | May 2016 | US |
Number | Date | Country | |
---|---|---|---|
61443088 | Feb 2011 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 13397398 | Feb 2012 | US |
Child | 15003325 | US |