The present invention relates generally to the field of medical devices, and in particular, to a diabetes management system that includes a pump, a glucose system, and a remote device for communicating with the pump and glucose monitor.
Patients with diabetes, otherwise known as diabetics, have a chronic disease that is characterized by a disordered metabolism resulting in high blood sugar. There are two types of diabetes. Type 1 diabetes involves a permanent loss or destruction of the beta cells of the pancreas which produces insulin. This often results in low levels or the complete absence of naturally produced insulin in the body. Type 2 diabetes involves a combination of an unusual resistance to naturally produced insulin and a relative insulin deficiency. As a result of both types of diabetes, a patient's blood sugar must be specially regulated. In the case of Type 2 diabetes, blood sugar regulation can sometimes be accomplished through a carefully maintained diet and exercise regime. In the case of Type 1 diabetes, blood sugar regulation almost always requires insulin supplementation.
Supplementation of insulin traditionally comes in two forms, basal injections and bolus injections. Basal injections are low level insulin injections meant to cover the patient's general insulin deficiency at low insulin need times such as between meals and at night, although even during higher insulin need times the basal injections may be administered. Bolus injections are higher level insulin injections usually administered before a meal or to correct an unusually high blood glucose level.
Both Type 1 and Type 2 diabetics who require insulin supplementation have a cycle that they repeat several times a day: checking their blood sugar level (otherwise known as blood glucose level), interpreting the results of that test, and then acting to adjust their blood sugar if necessary. A patient's blood glucose level is often checked using a finger prick mechanism, capturing the blood on a testing strip or feeding a certain amount of blood into a glucose level reader. After checking blood glucose levels, diabetics need to interpret the results of their blood glucose test. Traditionally, the blood glucose level reading is simply a raw number. Normally, blood glucose levels are between 70 to 150 mg/dL. If the patient's blood glucose level falls outside of this range, the patient will need to adjust it using an insulin injection. Generally, insulin dosages are calculated based on a formula that is unique to each patient, and then adjustments are made based on several other changing factors. These factors include taking into account what and how much food the patient had or will be having shortly, if the patient has just completed or plans to complete a specified amount of physical activity, and whether the patient has switched or plans to switch into a different mode of operation such as switching from awake mode to sleep mode. Once all of these factors have taken into account and an insulin dosage has been calculated, then the insulin is injected. Insulin is injected either with a syringe or a pump. This cycle must be done as often as necessary and is complex, time consuming, uncomfortable, and sometimes even embarrassing.
In order to perform these steps there are currently products on the market that help Diabetics take blood glucose readings, interpret results, and adjust blood glucose levels. However, the current products require diabetics carry-around with them and manage a tremendous amount of equipment. Diabetics often have to carry a glucose meter, a lancet for taking blood samples, and testing strips, an insulin syringe, one or more vials of insulin or an insulin pump. Taking a blood glucose reading manually requires a finger prick blood sample, catching the blood sample on a testing strip, and feeding the testing strip into the blood glucose meter. This is painful, potentially embarrassing, prone to error, and cannot be done while the patient is involved in other activities such as exercise or sleeping. Additionally, the equipment is often aesthetically unpleasing and “medical looking”; there are too many parts to easily manage; equipment is unsightly when seen poking out from under the patient's clothing; the equipment is bulky to carry in a shirt or pants pocket; the equipment is typically not waterproof; the long cannula stretching from the pump to the injection site can be uncomfortable and irritating for patient and can get caught on other objects; and the equipment comes in a limited choice of colors. Furthermore, current products are not context aware and they do not record and learn from the testing and dosing operations performed by the patient. Currently, there are no integrated products that help a patient easily test, interpret, and dose discretely.
Diabetes can be a very fatiguing disease to live with. Diabetes is a disease that must be constantly managed. Diabetes forces people to be more regimented in their lifestyles, eating patterns, and awareness of time. Furthermore, diabetes is typically managed alone, often without much community support. Remaining motivated to provide proper self care, especially over decades of living with the disease, is a significant challenge. Despite these serious problems, current products on the market assist very little in helping diabetics comfortably live with and manage their disease, set goals, or keep motivated.
Therefore, it would be highly desirable to provide a system and method for addressing the above mentioned problems associated with wearing and using diabetic devices, checking blood glucose, interpreting test results, adjusting blood glucose, and remaining motivated to deal with the disease. Specifically, it would be desirable to provide an integrated system that has fewer and smaller components; is easier to wear; can make better use of data; keeps patients in control; is easy to learn and teach; involves less interpretation by the patient (e.g., fewer numbers); and gives the patient a platform to view and share data, view long term trends, interact with a health care professional, and provide other methods for remaining motivated.
In some embodiments, a patient wears a glucose monitor and a pump system covered by a soft, flesh colored, rubber-like shell which is comfortable and unobtrusive. A patient also carries a small, touch-screen, remote device that communicates wirelessly with the glucose monitor and the pump system. The remote device has a form factor that can be worn or carried in various ways such as on a keychain, necklace, or armband.
In some embodiments, a method for managing diabetes proceeds as follows. A glucose monitoring system attached to a patient measures a patient's blood glucose level. The glucose monitoring system wirelessly transmits the patient's blood glucose level to a remote device. The remote device wirelessly receives the blood glucose level of the patient. Then, in some embodiments, the wireless device compares the blood glucose level to a previous blood glucose level and displays the blood glucose level trending data on its display screen.
The trending data can be displayed in the form of a trending arrow, a trending graph, a change in an ambient display, or any combination of the aforementioned options. The ambient display is customizable. The patient can choose ambient display themes such as weather, digital pet, lava lamp, or patterned colors. If a change in ambient display is used to display trending information, the display image is a positive display when the trending is in a normal blood glucose level range, and is a negative image when the trending is outside of, or nearly outside of the normal range. For example, if the ambient display is weather themed, a darker sky is displayed when trending outside of the normal range and a lighter sky is displayed when trending in the normal range. If the ambient display is digital pet themed, an unhappy digital pet is displayed when trending outside of the normal range and a happy digital pet is displayed when trending in the normal range. If the ambient display is a lava lamp theme or a colored background theme, a warm color is displayed when trending outside of the normal range and a cool color is displayed when trending in the normal range.
In some embodiments, the remote device also displays the latest blood glucose level along with the trending data. In other embodiments, the remote device displays the latest blood glucose level alone. In some embodiments, the remote device also compares the measured blood glucose level to a pre-determined dosage cycle trigger level. In some embodiments, the pre-determine dosage cycle trigger level is selected from the group consisting of: a blood glucose level, a time of day, a period of time since a previous insulin dosage, and an alert message. In some embodiments, the remote device determines whether to initiate a suggested dosage cycle based on the result of its comparison to the predetermined dosage cycle trigger level. In other embodiments, the patient separately initiates a dosage cycle. If the suggested dosage cycle is initiated, then the remote device determines a suggested insulin dosage based on the blood glucose level and a patient profile, which includes historical blood glucose levels, stored on the remote device. In some embodiments, the suggested insulin dosage suggesting is at least partially based on a patient specified mode such as a sleep mode, a rest mode, an exercise mode, a work mode, a school mode, an eating mode, and a default mode. In some embodiments, the patient then accepts, rejects, or modifies the suggested insulin dosage. In some embodiments, the patient can modify the suggested insulin dosage by inputting units of carbohydrates he anticipates eating in the near future.
The remote device then calculates an insulin dosage to be administered. In some embodiments, the insulin dosage to be administered is the same as the suggested insulin dosage; in other embodiments, they are different. In most embodiments, the patient then verifies the insulin dosage to be administered. In some basal dosages, a low level insulin drip can be automatically administered without patient verification. The remote device wirelessly transmits the dosage to be administered to a pump system attached to a patient. The pump wirelessly receives the insulin dosage to be administered and then administers the corresponding amount of insulin to the patient.
In some embodiments, the blood glucose monitoring system, the pump system, and the remote device are distinct from one another. In other embodiments, the blood glucose monitoring system and the pump system can be covered under one soft shell cover. In this embodiment the pump system and blood glucose system is called a single pump-monitor system regardless of whether or not they are embodied in a single housing.
In some embodiments, if a blood glucose level reaches below a predetermined blood glucose alert level an alert is triggered. The alert can be visual, haptic, audio, or a combination of more than one of these types. In some embodiments, similar alerts will be activated if a technical problem occurs. For example, if the insulin level in the reservoir is low, there is a malfunction in the blood glucose monitoring system, there is a malfunction in the pump system, the pump is disconnected from a cannula, a cannula is disconnected from a needle, the remote device is out of range with the blood glucose monitoring system, or the remote device is out of range with the pump system an alert will also be activated.
In some embodiments, the remote device sends, either wirelessly or by means of a USB port, a patient profile to an external computing device for tracking long term trends, tracking trends in various modes, or assisting patients in goal setting.
In some embodiments, the remote device has a touch sensitive display screen, a wireless transmitter for wirelessly transmitting an insulin dosage to be administered to a patient, a wireless receiver for wirelessly receiving a blood glucose level of a patient, a processor, a power source, and a memory comprising: an operating system, a patient profile, and a dosage calculator for calculating the insulin dosage to be administered. In some embodiments, the memory further comprises a machine learning algorithm, a patient predictive model, and a monitoring algorithm.
In some embodiments, the remote device has a USB port. In some embodiments, the remote device also has an attachment means. In embodiments with an attachment means, the remote device can be attached to a necklace, arm band, keychain, or other object. In some embodiments, the remote device's display screen rotates depending on the orientation of the remote device. In some embodiments, the remote device has a hand-held form factor. In some embodiments, the remote device is no larger than 1 in by 3 inches by ½ an inch.
In some embodiments, the pump system has a pump for injecting a patient with insulin, an insulin reservoir fluidly connected to said pump for providing said pump with insulin, a cannula fluidly connected to said pump, a needle fluidly connected to said cannula, and a wireless receiver for receiving an insulin dosage to be administered to a patient from said remote device.
In some embodiments, the glucose monitoring system has a glucose sensor for sensing a blood glucose level of a patient and a wireless transmitter for transmitting said blood glucose level of a patient to said remote device.
In some embodiments, the pump system and said glucose monitoring system are covered by a soft flesh-colored shell attached to the patient for smooth and comfortable wear. In some embodiments, the shell is substantially waterproof. In some embodiments, the pump system, glucose monitoring system, and remote device are also substantially waterproof.
In some embodiments, the wireless transmitter and the wireless receiver of the remote device, the wireless receiver of the pump system, and the wireless transmitter of the glucose monitoring system communicate by a wireless means such as infrared technology, WiFi, cellular telephone technology, radio frequency technology, or Bluetooth technology.
For a better understanding of the nature and embodiments of the invention, reference should be made to the Description of Embodiments below, in conjunction with the following drawings in which like reference numerals refer to corresponding parts throughout the figures.
In some embodiments, a diabetes management system is provided that includes a unit containing both an insulin pump system and a glucose monitor system covered by a soft shell. Systems can be incorporated in a single housing or separate shells. The pump and glucose monitor systems are monitored and/or controlled by a remote device.
The remote device 104 embodies a display screen 202. In some embodiments, the display screen 202 is touch sensitive, i.e. it includes a tactile input 214 to allow the patient to interact with the device by tapping, scrolling, or sliding his finger on the display screen 202. For example, in the embodiment shown in
In some embodiments, the remote device 104 contains a port 206, such as a USB port, for communicating with an external computing device, such as a desktop computer. The information contained in the remote device's 104 memory can be uploaded via the port 206. Data, additional programs, features, and firmware, can be downloaded from the external computing device to the remote device 104 via the port 206. In other embodiments, the information can be communicated wirelessly and not via the port 206. In some embodiments, the port 104 is covered by a protective cap 208 that can be removed from covering the port 206. In other embodiments, the cap 208 is attached to the remote device 104 via a cord, swivel device, or any suitable means to keep the cap from becoming permanently separated from the port 206, while still allowing the port to plug into an external computing device. In some embodiments, the cap 208 can contain a loop 212 such that a necklace, cord, keychain or other mechanism can be strung through the loop 212 as shown in
Downloading the information from the remote device 104 to an external computing device, allows the patient to visualize long term trends such as month long glucose level trends or the patient's reactions to various levels of insulin doses during various modes of operation. Larger and more complex graphs and trend tracking may be available on the external computing device. The patient can compare his results with other patient's results on a virtual community of patients using the diabetic monitoring system, share best practice tips, share diabetic recipes, challenge other patient's to specific goals, etc. The patient can also send his information to a health care professional and get individualized advice without a need to visit a doctor's office or clinic. Interaction with a larger community may also help a patient remain motivated, get advice, feel supported by others in a similar situation, and set goals. Even when not communicating with a virtual community or viewing long term trends on an external computing device, the patient may find it easier to remain motivated, and set short and long term goals using the diabetic management system because of its ease of use and ability to unobtrusively integrate with the patient's lifestyle.
In some embodiments, the remote device 104 contains a speaker system 310 so that audio alerts or messages can be communicated the patient. Audio alerts can be any number of tones, rings, or tunes which are customizable by the patient. The volume of the audio alert can be variable. Furthermore, additional audio alerts can be added to the memory 318 of the remote device 104. In some embodiments songs can be downloaded such that the patient can play relaxing music while administering insulin or motivating music while working out. In some embodiments, the remote device 104 contains a vibration mechanism 312 such that it can communicate haptic alerts to the patient. The haptic alerts can be of various types and intensities of vibration, can be patient customizable and/or additional haptic alerts can be downloaded into the remote device 104.
In some embodiments, the remote device 104 contains a tilt-sensor 314 or other mechanism for determining the orientation of the remote device 104. The display screen 202 may re-orient its display to match the orientation of the remote device 104. In some embodiments, the remote device 104 can also contain a digital camera, media player, etc.
The remote device 104 contains a wireless transceiver 316. This wireless transceiver 316 automatically communicates with the transceiver 414 of the pump system 106 and/or the wireless transceiver 418 (
The memory 318 contains a number of elements. In some embodiments, the memory 318 contains one or more of the following: an operating system 319 that stores instructions for communicating, processing data, accessing data, storing data, searching data, etc.; a machine learning algorithm 320; communication procedures 321; preferences 322; a dosage calculator 323; modes 324; a patient predictive model 325; a monitoring algorithm 326; ambient displays 600; and a patient profile 328. The communication procedures 321 facilitate communication with the glucose monitor system 102 and the pump system 106. The dosage calculator 323 calculates patient inputted carbohydrates and produces suggested insulin dosages.
The memory 318 also contains a patient profile 328 shown in
The information stored in the patient profile 328 is used to create a patient specific predictive model 325. The machine learning algorithm 320 builds the predictive model 325. The patient predictive model 325 is generated from the historical data stored in the patient profile 328 to model how a patient reacts to a certain dosage, at a certain time, and mode. For example, it may fit a curve to the raw data, or it may use any other suitable statistical or neural technique to generate the patient predictive model 325.
The monitoring algorithm 326 receives blood glucose readings and checks these readings against pre-determined dosage cycle trigger levels for the current mode 324 stored in the memory 318. The monitoring algorithm 326 initiates a dosing cycle if a predetermined dosage cycle trigger level is reached. In some embodiments, the monitoring algorithm 326 is triggered by a low blood glucose level. In other embodiments, the monitoring algorithm 326 is triggered by a particular time of day. For example, in some embodiments, the monitoring algorithm 326 will initiate a dosing cycle at noon because the patient usually eats at that time.
Once the dosing cycle is initiated, the dosage calculator 323 uses the patient predictive model 325 to generate a suggested insulin dosage. The dosage calculator 323 accomplishes this by using the patient predictive model 325 to determine the typical dosage that has been administered in the past for a blood glucose level at a given time and mode.
The memory 318 may contain preferences 322 such as the patient's preferred audio alerts, visual alerts, haptic alerts, ambient displays, the text font and size, basal dosage, glucose monitoring schedule, and any number of other elements that the patient has optionally selected and stored. Many of these preferences can be linked to particular modes 324. Modes 324 are the pre-set options regarding glucose monitoring, glucose displaying, insulin dosing, alerts levels, the alert type, alert volume or intensity, ambient displays, and other options that are pre-set for a standard repeatable situation. Setting-up modes 324 saves the patient time customizing particular preferences. A patient can utilize as many or as few modes 324 as desired. Some modes 324 are pre-set but customizable, such as sleep mode, exercise mode, school mode, work mode, neutral mode, default mode, emergency mode, etc. The patent can also create new modes 324. The modes can be pre-set for a 24 hour period, a weekly period, a monthly period, or a combination thereof. For example, a patient may wish to pre-assign a particular mode during the times the patient is normally sleeping, eating, and involved in work or school activities. Alternatively, a patient may wish to program a slightly different mode pattern on the weekends when he stays up later and goes to a weekly exercise class. Ambient displays 600 are also stored in the memory 318. Some examples of ambient displays 600 are a color pattern, a digital pet, weather, and a lava lamp as discussed in relation to
The remote device 104 may also contain additional software programs. For example, it may contain a calendar, an address book, or the option to send and receive information such as e-mail wirelessly for sending immediate readings to a health care professional, for requesting advice, or for setting up future appointments. The remote may also contain a picture folder for customizable ambient displays 326, a calculator for manual insulin dosage calculation, or any other useful program to assist the patient in operating the diabetes management system 100. In some embodiments, the memory 318 may also contain a telephone means, which utilizes the speaker system 310 and tactile input 214 to call for help. In some embodiments, in the case of an emergency, such as the patient passing out, the telephone will automatically connect to a cellular telephone network and transmit an emergency message to a trusted friend, relative, or health care professional.
In some embodiments, the remote device 104 can act as not only a communication device with the glucose monitor system 102 and the pump system 106, but also as a complete personal digital assistant for a patient's daily life. This reduces the number of items that a patient needs to carry on a daily basis. Alternatively, in some embodiments, the functions of a remote device 104 may be transferred to an alternative handheld device such as a cellular telephone or personal digital assistant. In these embodiments, the alternative handheld device may perform the functions of the remote device 104 described above. The screen, speaker, clock, vibration mechanism, and memory of the alternative handheld device may be utilized. The pump system 106 and the glucose monitoring system 102 wirelessly communicate with the alternative device by infrared, Bluetooth, WiFi, or cellular telephone networks.
In some embodiments, the glucose monitor system 102 is positioned a certain distance away from the pump system 106 to facilitate obtaining a more accurate blood glucose level reading. Therefore, in some embodiments, the glucose monitor system 102 is positioned on another part of the patient's body, and not under the same shell 1100 as the pump system 106. In those embodiments, the glucose monitor system 102 can be positioned under a separate soft shell 1100 or may not be covered by a shell at all.
The glucose monitor system 102 includes a glucose sensor 416, a needle 420 and a transmitter or transceiver 418, as shown in
The pump system 106 includes a pump 404, an insulin reservoir 406, one or more cannula 408, one or more needles 410, and a transmitter or transceiver 414. Unlike current products, in some embodiments, the reservoir 406 is soft such that it appears less bulky under the shell 1100. In other embodiments, the reservoir 406 is hard, but due to its snug fit inside the shell 1100, the reservoir 406 appears soft. In some embodiments, the reservoir 406 is refilled by injection through the shell 1100. In some embodiments, the pump 404 and/or insulin reservoir 406 can be permanently or semi-permanently embedded in the shell 1100. In some embodiments, the pump 404 can easily slide off of the cannula 408. The pump system 106 can be worn continuously for a period of time without having to be repositioned. In some embodiments, that period of time can be one day, one week, one month, or even longer.
During a typical dosing operation, the pump's transceiver 414 receives an actual dosage transmission from the remote device 104 (
The shell 1100 allows the pump system's transceiver 414 and the glucose monitor system's 102 transceiver 418 to wirelessly communicate with the remote controller 104 (
The keychain 502 of
The armband 504 of
The necklace 506 of
These attachment means are provided as examples only. Any other attachment devices that provided a mechanism for a patient to carry the remote device can also be used. For example, in other embodiments, the functionality of the remote device may be embodied inside a watch, such that the patient may wear it on his wrist, and the ambient display is a watch face when not actively communicating diabetes related information to the patient.
As shown in
In some embodiments, the orientation of the ambient display is landscape view as is shown in
In some embodiments, the ambient display includes icons conveying general information along with a background image. For example,
The remote device 104 receives a reading of the current glucose level from a glucose monitoring system 102 at 700. The remote device 104 stores at 702 this glucose level in its memory 318 (
In some embodiments, at any time after receiving the blood glucose level at 700, the remote device 104 displays the current blood glucose level reading on a display screen at 701, either as an individual glucose level or an average glucose level over a period of time. An average glucose level display protects against statistical variations resulting from an individual glucose level measurement malfunction. The blood glucose level can be displayed continuously or periodically. For example, the blood glucose level can be displayed every minute, every five minutes, every ten minutes, once an hour, several times a day, once a day, or the like. The patient can specify how often the readings are taken and also how often the readings are displayed. In some embodiments, trending information of the current glucose measurement as compared to the previously displayed glucose measurement is displayed in the form of a graph, arrow, or change in ambient display 600. In some embodiments, the trending information can be displayed only as a change in ambient display 600 (
The dosage cycles described below normally apply to bolus dosages, although they can apply to basal dosages as well. A basal dosage is an insulin dosage lower than a bolus dosage. A bolus dosage is usually used to cover high insulin needs such as during mealtimes. A basal dosage is often set up to automatically deliver insulin between meals or at times of rest. It is not necessary that both a bolus and a basal dosage be given to a particular patient. For example, one patient may only require basal level doses while another patient may need only bolus doses. However, some patients require both basal and bolus doses.
In some embodiments, patient input at 707 begins a dosage cycle at 706. In other embodiments, the monitoring algorithm 326 (
In the embodiment where the decision to begin a dosage cycle at 706 is initiated by the monitoring algorithm 326, the remote device 104 then calculates, at 708, a suggested insulin dosage. The suggested insulin dosage is calculated at 708 based on the received blood glucose level 700 and the patient predictive model 325 (
In some embodiments, the calculation of the insulin dosage to be administered at 710 is determined by the patient input approving or modifying the suggested insulin dosage at 711. In other embodiments, if no patient input 711 is received after a pre-determined amount of time has passed the remote device 104 determines that the insulin dosage to be administered 710 is calculated as the suggested insulin dosage 708. In yet other embodiments, the patient can select the mode 324 (
In the embodiment where the decision to begin a dosage cycle at 706 is initiated by the patient at 707, the following steps vary depending on the original patient input at 707. In some embodiments, the patient initiates the dosage cycle at 706 by simply entering instructions to begin a suggested insulin dosage calculation at 707. In this embodiment, the steps are the same as described above for the dosage cycle initiated by the monitoring algorithm 326.
In some embodiments, the patient enters the approximate amount of carbohydrates that the patient plans to eat at 707. The remote device 104 calculates the suggested insulin dosage, at 708, using the dosage calculator 323 (
In other embodiments, the patient enters the amount of insulin units the patient would like to receive at 707. In this case no suggested dosage need be calculated at 708 or displayed at 709. In some embodiments, the remote device 104 proceeds immediately to the step of displaying the insulin dosage to be administered at 712. In other embodiments, the remote device calculates at 710 the insulin dosage to be administered based on the received blood glucose level 700 and the patient predictive model 325 (
No matter how dosage cycle is initialed, for the safety of the patient, prior to transmitting the actual insulin dosage for a bolus level injection, the patient confirms that the dosage should be administered at 713. This is a safety feature to protect the patient from a potentially lethal overdose of insulin. However, certain low level basal injections can be set to administer by drip. In some embodiments, basal injections by drip do not require patient input at 713 before transmission. This can be especially helpful if the patient is sleeping or otherwise engaged and does not wish to be bothered for low level basal injections. Finally, the actual dosage is wirelessly transmitted, at 714, to the pump.
The pump system 106 (
If the patient chooses the “change SETTINGS” 1008 soft button 204 (
If the patient chooses the “change MODE” button 818, the patient is taken to the Mode screen 822 where, in some embodiments, the patient can chose modes 324 (
The patient can also edit the modes by selecting the “EDIT” button 828. In some embodiments, the patient can pre-set a variety of options for each mode 324, such as the dosage for a basal injection, how often the basal injection should be administered, and how often to check the patient's blood glucose level. In some embodiments, the patient can pre-set a schedule for daily mode 324 changes. For example, the patient can set a 24 hour period of time with a sleep mode 324 from 11 PM to 6 AM, an eating mode 324 from 6 AM to 7 AM, a work mode 324 from 7 AM until NOON, a eating mode 324 from NOON to 1 PM, work mode 324 from 1 PM to 6 PM, an exercise mode 324 from 6 PM to 7 PM, eating mode 324 from 7 PM to 8 PM, and a leisure mode 324 from 8 PM to 11 PM. In this embodiment, the patient will not need to manually change modes 324 before eating, going to work, bed, etc.
Some alerts warn of a device malfunction or technical problems 902. These can relate to any technical problem occurring with the diabetes management system 100 (
Another type of visual alert is a patient specific warning 906. For example, the alert shown in
The patient can choose to perform a function to correct the problem. For example in
In some embodiments, patient warnings 906 are customized to the patient. For example, the patient can set the high glucose alert to only appear if two or three glucose readings in a row appear outside of a pre-determined range used for patient warnings 406. In other embodiments, the patient customizes the pre-determined range. In yet other embodiments, the warnings become patient specific by means of the machine learning algorithm 320 (
In the landing screen 1002, the patient will get more information about his or her blood glucose levels over a period of time in the format of a graph 1004. In some embodiments, this graph 1004 is customized to be a part of the ambient display 600. The period of time that the graph covers can be altered by the patient, such as the last hour, the last day, or the last week. In some embodiments, the graph also shows guide lines 1006. The guide lines 1006 mark the range of blood glucose levels that the patient wishes to stay between. In some embodiments, the high blood glucose level and the low blood glucose level values are displayed. In other embodiments, as is shown in
Soft buttons 204 (
If a patient chooses to “dose by CARBS” 1010, the carbohydrate adjustment screen 1013 appears. In some embodiments, the remote device 104 makes a guess as to how many carbohydrates the patient is likely to eat and displays a suggested carbohydrates amount 1014. This suggested carbohydrate amount 1014 is based on information stored in a patient's profile 328 (
The insulin dosage screen 1022 appears after the patient has finished with the carbohydrate adjustment screen 1013. However, the insulin dosage screen 1022 will appear immediately if the patient chooses the “dose by UNITS” 1012 soft button 204 in the landing screen 1002. At the insulin dosage screen 1022, an intelligent guess as to how much insulin is needed is displayed as a suggested insulin dosage 1024. In some embodiments, the amount of insulin left in the patient's body 1026 will also be displayed. In some embodiments, the total amount of insulin that will be in the patient's body after the injection 1028 is also displayed. This additional information may help the patient better understand his or her insulin needs. The patient chooses the up arrow 1016 or the down arrow 1018 to modify the bolus dosage. In some embodiments, when the patient is finished, the dosing verification screen 1030 appears automatically. In other embodiments, the patient chooses to proceed to the dosing verification screen by pressing the “NEXT” key 1020.
The dosing verification screen 1030 displays the amount of insulin to be administered 1032. The patient is then asked to verify that the insulin should be injected now by selecting the “YES” button 1034, or that the insulin should not be injected by selecting the “NO” button 1036. After the patient chooses the “YES” button 1034, the active dosing screen appears 1038 indicating that dosing is currently occurring. The remote controller 104 (
In some modes 324 (
In should also be apparent to one skilled in the art, that the above described functions could be accomplished by any similar means. The screens to do not need to look precisely like the ones shown in
In some embodiments, the shell 1100 is attached to the patient by glue around the perimeter of the shell 1100. Alternatively, it can attach through suction or any other suitable attachment means. In some embodiments, the shell 1100 includes a built in insulin reservoir in the internal cavity 1101. In some embodiments, the internal cavity is large enough to cover all of the glucose monitoring system 100 (
In some embodiments, the shell 1100 is customizable to match the patient's skin tone. In other embodiments, the shell 1100 is clear, brightly colored, patterned, or made to mimic a skin embellishment such as a tattoo, jewel, or cartoon character. In yet other embodiments, the shell 1100 is customizable such that a patient can change its color by swapping out a faceplate or layer. In some embodiments, the shell 1100 is re-usable, while in other embodiments, the shell 1100 is designed for a one time use.
In some embodiments, the shell 1100 is waterproof, and can be worn during swimming or showering. It can be worn for numerous days in a row, and may only need to be removed for maintenance such as to re-fill the insulin reservoir. In some embodiments, the shell 1100 is configured such that the insulin reservoir is refilled by an injection through the shell 1100. In this embodiment, the shell 1100 does not need to be removed for insulin reservoir refilling. A patient can wear the shell 1100 during all normal daily activities such as to work, to school, while exercising, and while sleeping.
The term “patient” has been used throughout this description. However, one skilled in the art would realize that at many times a person other than the patient could be performing the interactions with the remote device. For example, a physician or pharmacist may set the original profile of a patient before the patient uses the device. Also, if the patient was unconscious, another user can interact with the remote device to dose insulin. Alternatively, a patient may be too young or otherwise incapable of understanding how to dose themselves and a parent or guardian may be the one interacting with the remote controller instead.
The foregoing descriptions of specific embodiments of the present invention are presented for purposes of illustration and description. They are not intended to be exhaustive or to limit the invention to the precise forms disclosed. Many modifications and variations are possible in view of the above teachings. The embodiments were chosen and described in order to best explain the principles of the invention and its practical applications, to thereby enable others skilled in the art to best utilize the invention and various embodiments with various modifications as are suited to the particular use contemplated. Furthermore, the order of steps in the described methods is not necessarily intended to occur in the sequence laid out. In addition, use of singular terms also includes their plural. It is intended that the scope of the invention be defined by the following claims and their equivalents.
The present application relates to design application number ______, to Daniel Saffer et al., filed Feb. 28, 2008, entitled “Remote Control Device for a Diabetes Management System”, which also relates to design application number ______, to Daniel Saffer et al., also filed Feb. 28, 2008, entitled “Soft Shell for a Diabetes Management System”, the disclosures of which are hereby incorporated by reference in their entirety.