Diode pumped, multi axial mode intracavity doubled laser

Abstract
A diode pumped, multi axial mode, intracavity doubled laser has at least two resonator mirrors that define a resonator cavity. A Nd:YVO4 laser crystal and a LBO doubling crystal are positioned in the resonator cavity. A diode pump source supplies a pump beam to the laser crystal and produces a laser crystal beam with at least three axial modes that are incident on the doubling crystal. A frequency doubled output beam is generated with an output power of at least 1 watt.
Description




BACKGROUND OF THE INVENTION




1. Field of the Invention




This invention relates generally to diode pumped intracavity frequency doubled lasers, and more particularly to a NdYVO


4


laser with a LBO doubling crystal.




2. Description of Related Art




Continuous wave ion lasers are relatively reliable sources of continuous wave green laser light with low amplitude noise, and provide output power at the multiple watt level. These devices convert electrical power into optical power with efficiencies of only a small fraction of one percent. There are many applications that would benefit from the development of a highly efficient, low cost, diode-pumped, continuous-wave, solid-state green, blue, red, near infrared, or UV laser source, also at the multiple watt level and with comparable amplitude stability.




Certain fundamental difficulties with intracavity-frequency-doubled solid-state lasers were discovered and numerically modeled in early work by Baer. See for example T. Baer, J. Opt. Soc. Am. B., Vol. 3, No. 9, pp. 1175-1180 (1986), and U.S. Pat. Nos. 4,656,635 and 4,701,929. It was reported and disclosed that large amplitude fluctuations are observed on the green output beam and the intracavity infrared laser beam when a frequency doubling crystal such as KTP is introduced into an otherwise amplitude-stable multiaxial mode diode-pumped Nd:YAG laser. It was also reported that the large amplitude noise on the green output beam disappears when an appropriate etalon is placed in the laser cavity that forces single axial mode oscillation. In the multiaxial mode case, where 2 to 4 modes were oscillating, the green output power was seen to fluctuate with up to 100% modulation depth. Baer's experimental work and theoretical model indicated that the insertion of a frequency doubling crystal in this multiaxial mode laser resulted in nonlinear coupling of the loss of the infrared axial modes via sum frequency generation A high peak power in one axial mode induced a high nonlinear loss for the other axial modes, and caused an unexpected and undesirable pulsing effect.




As an example of the effect described by Baer, a laser with two infrared axial modes generated three green frequencies; two were doubled modes and the other a sum frequency mode. The sum frequency process couples the two infrared axial modes in a way that can cause them to switch on and off in a sequential fashion. The typical period of this mode coupling was found to be a function of the magnitude of the nonlinear conversion. For weak conversion, the period was short and the modes minimally modulated. For stronger conversion, the mode coupling period lengthened, and the modes switched on and off in pulses of high peak power, completely out of phase with each other in a semi-periodic fashion. The noise spectrum of such a laser typically showed substantial peaks in the 10 to hundreds of kHz range for either the green or infrared, and corresponded to considerable amplitude fluctuations.




A source with this type of amplitude modulation is not as generally useful as one with low amplitude noise, and therefore high amplitude stability. As an example, for applications in ophthalmology, such as retinal photocoagulation, amplitude stability is required on the time scale of the typical exposure durations for accurate control of therapeutic effects. Another example is the use of a green laser as a pump for a second laser, such as a dye or Ti:Al


2


O


3


laser. Deep amplitude modulation at certain frequencies can cause undesirable amplitude modulations on the output of the second laser.




A number of methods for stabilizing the intracavity-frequency-doubled output of a diode-pumped solid-state laser have been described and demonstrated. The most common materials; have been Nd:YAG as a laser medium and KTP as a nonlinear, doubling medium. For this reason, the most common type of phase matching is Type II. One technique that has been used in an attempt to stabilize the frequency doubled output from such systems included insertion of intracavity quarter wave plates (see M. Oka, and S. Kubota, Opt. Lett. 13, 805 (1988)). The Oka quarter-wave technique can result in two orthogonally polarized infrared eigenmodes that are not coupled by sum-frequency generation. The Oka configuration was shown to be amplitude-stable under certain conditions. However, for higher output powers this configuration requires the addition of an etalon (M. Oka et al, 1993 Advanced Solid State Laser Conference, paper AMG1). It was reported that this system could be stable for only a few hours at a time. The temperature control of the KTP is imperative with this technique. Other techniques used for stabilizing the output of intracavity-doubled solid-state lasers include optical cavity temperature control (see U.S. Pat. No. 4,884,277 issued to Anthon, et al. on Nov. 28, 1989) and forcing single frequency operation (see U.S. Pat. No. 5,164,947 issued to G. J. Lukas, et al. on Nov. 17, 1992, and W. Weichmann et al., 1995 Advanced Solid State Laser Conference, papers TuD4 and WD4).




Another method of achieving low noise operation is also based upon single frequency operation: J. Nightingale et al have developed an intracavity-doubled unidirectional ring laser with diode-pumped Nd:YAG and KTP (U.S. Pat. Nos. 5,052,815, 5,136,597, and 5,170,409, and 1994 Compact Blue Green Laser Conference, Post-deadline paper PD6).




While all of these techniques have demonstrated regimes of operation where the frequency doubled output is measured to have low amplitude noise, in all cases the techniques are difficult to implement in a reliable, low cost fashion that is resilient to changes in environmental conditions, such as temperature. The techniques employed typically must maintain an inherently amplitude-unstable system within the narrow range of parameter space for which the system is stable. The single frequency intracavity-doubled systems can suffer mode-hops that result in undesirable discontinuities in output power. To avoid this, single frequency systems must be designed to be resistant to such mode hops. Additionally, the potential for scaling currently available systems to higher power may be limited.




It would be highly desirable to provide an inherently amplitude-stable, intracavity-frequency-doubled, solid-state laser that does not require active stabilization or single axial mode operation. Additionally, there is a need for a laser of this type that remains stable over a range of environmental conditions, such as changes in ambient temperature. There is a need for an intracavity-frequency-doubled laser that does not exhibit discontinuities in output power, as with single frequency systems. There is also a need for an amplitude stable, intracavity frequency doubled laser that can be scaled to higher powers. A similar frequency-tripled system would also be useful.




SUMMARY OF THE INVENTION




It is an object of the invention to provide a continuous-wave (CW), diode-pumped, multiaxial mode, intracavity-frequency-doubled solid-state laser with high amplitude stability.




Another object of the invention is to provide a CW, diode-pumped, multiaxial mode, intracavity-frequency-doubled laser with a percent root mean square noise (% RMS) of less than about 0.5% over a wide range of operating parameters.




A further object of the invention is to provide a CW, diode-pumped multiaxial mode, intracavity-frequency-doubled laser with a percent root mean square noise (% RMS) of less than about 0.5% over a wide range of operating parameters that utilizes the combination of Nd:YVO


4


and LBO.




These and other objects of the invention are achieved in a diode pumped, multi axial mode, intracavity doubled laser. At least two resonator mirrors define a resonator cavity. A Nd:YVO


4


laser crystal and a LBO doubling crystal are positioned in the resonator cavity. A diode pump source supplies a pump beam to the laser crystal and produces a laser crystal beam with at least three axial modes that are incident on the doubling crystal. A frequency doubled output beam is generated with an output power of at least 1 watt.











BRIEF DESCRIPTION OF THE FIGURES





FIG. 1

is a schematic diagram of a multi-port, diode pumped, multiaxial mode, intracavity doubled laser resonator with high amplitude stability.





FIG. 2

is a graph illustrating frequency doubled output power as a function of incident optical input power of the laser resonator shown in FIG.


1


.





FIG. 3

is a schematic diagram of a single port, diode pumped, multiaxial mode, intracavity frequency doubled laser with high amplitude stability.





FIG. 4

is a schematic diagram of a single port, diode pumped, multiaxial mode, intracavity frequency doubled laser with high amplitude stability where a portion of the generated frequency doubled light is lost.





FIG. 5

is a schematic diagram of a two port, diode pumped, multiaxial mode, intracavity frequency doubled laser that utilizes Nd:YVO


4


and LBO.





FIG. 6

is a schematic diagram of a two port, diode-pumped, multiaxial mode, intracavity-frequency-doubled laser that utilizes Nd:YVO


4


and LBO but uses less mirrors and optical elements than does the laser in FIG.


5


.





FIG. 7

is a schematic diagram of a two port, diode pumped, multiaxial mode, intracavity frequency doubled laser that utilizes Nd:YVO


4


and LBO, further including a frequency tripling crystal.





FIG. 8

is a schematic diagram of a diode pumped multi axial mode, intracavity frequency doubled laser used as a pump source for another laser, such as a Ti:Al


2


O


3


.





FIG. 9

is a schematic diagram of a retinal photocoagulator system incorporating a diode pumped, multi axial mode, intracavity frequency doubled laser used as a laser source.











DETAILED DESCRIPTION




For the purposes of the following description of the invention, the following definitions are used:




a “high power” frequency doubled output as an output power (P


out


) equal to or greater than 1 watt.




a “high optical efficiency” as an optical-to-optical efficiency (P


out


/P


incident, diode


) that is greater than 5%. The optical pump power from the laser diode that is incident on the laser crystal is (P


incident, diode


). In the case of the fiber-coupled diode pump sources, the value of (P


incident, diode


) is the power that is emitted by the fibers.




a “high electrical to optical efficiency” for a diode pumped intracavity frequency doubled laser as an efficiency (P


out


/P


electrical, diode


) greater than 1%. The value of (P


electrical, diode


) is the value of electrical power provided to the diode, in the form of the operating current times the voltage drop across the laser diode. The value of (P


electrical, diode


) for a 20 W diode bar is typically 2.5 times that of the optical output power of the diode bar; these bars are typically 40% efficient. For this estimation of high electrical efficiency, the amount of power required to cool the diodes is not included; in some cases it may be necessary to consider this power.




a “multiaxial” mode intracavity frequency-doubled laser as a laser where on the order of 3 or more axial modes are oscillating at the fundamental infrared wavelength in the laser resonator. In some cases, the lumber is on the order of 100 to 200.




a “highly amplitude stable” intracavity frequency doubled laser, or the same laser with “high amplitude stability” as one where the output beam exhibits percent root mean square (% RMS) noise from 10 Hz to 10 MHz of less than 3%. In a preferred embodiment, this noise level can be as low as 0.20%, high amplitude stability is equated with low amplitude noise.




The diode-pumped, multiaxial mode, intracavity-doubled laser of the present invention includes at least one laser crystal and at least one doubling crystal positioned in a resonator cavity. This laser is highly amplitude stable, and the frequency-doubled output beam is of high power. A diode pump source supplies a pump beam to the laser crystal and produces an intra-cavity, infrared laser crystal beam at a plurality of axial modes that is incident upon the doubling crystal, producing a frequency-doubled output beam, also at a plurality of optical frequencies. The output beam is highly amplitude stable, meaning that the % RMS noise is less than 1%, preferably less that 0.5%, more preferably less than 0.2% and most preferably less than 0.1%. The laser is highly optically efficient, with an optical efficiency greater than 5% and preferably greater than 12%, with a preferred Nd:YVO


4


embodiment at about 23%. The laser is highly electrically-to-optically efficient, with an electrical-to optical efficiency greater than 1%, and preferably greater than 4%, with a preferred Nd:YVO


4


embodiment at about 8%. The systems are typically configured to provide good beam quality, meaning that a substantial fraction of the output power is nearly diffraction limited, or is substantially TEM


00


.




In

FIG. 1

, a bounded, standing wave resonator


10


for 1053 nm is illustrated. Resonator


10


is multi-port and diode pumped with four arms, each arm with its own optical axis. The first arm is defined by a high reflector


12


, which is highly reflecting at 1053 nm and highly transmissive at a pump wavelength of 797 nm, and a high reflector


14


, which is highly reflecting at 1053 nm and highly transmissive at 797. The distance between reflectors


12


and


14


is L1. The second arm of resonator


10


is defined by a high reflector


16


, highly reflecting at 1053 nm, and highly reflecting at 527 nm, and an output coupler


18


, highly reflecting at 1053 nm and highly transmissive at 527 nm. It has a length of L2. The third arm, with a length of L3, is defined by high reflector


14


and a high reflector


20


, which is highly reflecting at 1053 nm. High reflector


20


is, essentially, a fold mirror. The fourth arm, with a length of L4, is defined by high reflector


20


and output coupler


18


. An optional Brewster plate


34


can be placed in this arm. Although the optical elements in this embodiment are specific for 1053 nm, 97 nm and 527 nm, it will be appreciated that the optical elements can have other transmission and reflectivity characteristics, depending on pump source, laser crystal and doubling crystal.




The length of resonator


10


is L, and it equals the total lengths of the four arms. The optical path length of the resonator is very close to L. It differs from the sum of the lengths of the four arms only because of any optically dense material in the cavity, as is well known. For example, a laser crystal of length L


c


and index n


c


at a given wavelength has an optical path length of n


c


L


c


at that wavelength, as is well known to those skilled in the art.




One or more laser crystals


22


are positioned in the first arm along its optical axis. Suitable crystals include but are not limited to Nd:YLF, Nd:YAG, Nd:YVO


4


, Nd:GVO


4


, Nd:YPO


4


, Nd:BEL, Nd:YALO, and Nd:LSB and Nd:YVO


4


. A preferred crystal material is Nd:YVO


4


, particularly as illustrated in

FIGS. 3 through 7

. Positioned adjacent to reflectors


12


and


14


are a pair of lenses


24


and


26


, arranged in a telescope configuration. The output of a fiber-coupled diode pump source


28


produces a pump beam


30


that is focussed to a desired size by lenses


24


and


26


. The telescope arrangement provides for the focussing of pump beam


30


from a diode source


28


. The size of the pump beam is optimized with lenses


24


and


26


to avoid fracture of incident faces of crystals


22


while increasing useful pump power. The TEM


00


mode size diameter of the infrared beam in the laser crystals is about 1 millimeter in an embodiment that utilizes Nd:YLF, but is about 0.5 mm in an embodiment that utilizes Nd:YVO


4


. The pump beam diameter is about 0.6 to 0.7 mm in either embodiment. This pumping geometry is longitudinal pumping, or end pumping, as will be recognized by those skilled in the art. An amplitude stable, multi-axial mode, intracavity doubled laser can also be used with transversely pumped systems, or directly-pumped systems that do not use fiber-coupled diodes.




Diode pump source


28


can be a single diode, spatial emitter, diode bar or a plurality of diodes or A plurality of diode bars that have a reduced fast-axis divergence. A suitable diode source


28


is model No. OPC-A020-797-CS, available from OptoPower Corporation, City of Industry, California. Another suitable diode is denoted as BO2O. Preferred wavelengths of diode pump source


28


are in the range of 795 to 815 nm. Peak absorption wavelengths of specific laser crystals


22


are approximately as follows: Tm:YAG—785 nm; Nd:YLF—


797


; and Nd:YAG, Nd:YVO


4


—809 nm. The “797” designator in the OptoPower product number becomes “809” when an 809 nm wavelength is desired. As is now well known in the art, the wavelength of the output of such an GaAlAs based diode can be tuned by adjusting the temperature of the diode. The tuning rate is well-known to be about 0.3 nm/° C.




In the embodiment of

FIG. 1

, two laser crystals


22


are longitudinally-pumped, or end pumped. It is possible to include only one crystal and pump both ends, as in

FIGS. 5 through 7

, or alternatively, pump only one side of a single crystal


22


, as in

FIGS. 3 and 4

. As shown in

FIG. 1

, a single diode pump source


28


is used. Separate pump sources can be used for each end of the first arm of resonator


10


, or for the resonators in

FIGS. 5 through 7

. It will be appreciated that the principles of stable, multiaxial mode intracavity doubling also extend to transversely-pumped, or side-pumped lasers.




Diode pump source


28


is coupled to one or more optical fibers


32


. Preferably, a bundle of optical fibers


32


are utilized. Suitable fibers include but are not limited to those that have silica cores with silica cladding.




In one embodiment, a fiber-bundle-coupled diode bar is used to longitudinally pump Nd:YLF crystals with up to 9 W per crystal. The fast axis divergence of the diode bar is reduced by a cylindrical microlens, U.S. Pat. No. 4,785,459, issued to Baer, and each emitting array of the bar is coupled into a multimode optical fiber. This highly bright pump source is described in U.S. Pat. No. 5,127,069, issued to Baer et al. on Jun. 30, 1992. Output power is extracted from the laser resonator by intracavity frequency doubling. The nonlinear crystal can be lithium triborate, or LBO, and Type I noncritical phase matching (NCPM) can be used. The output power is typically on the order of 2 W or greater for 16 W of incident diode pump light from a single 20 W, CW, ˜797 nm diode laser bar. This corresponds to an optical efficiency (P


out


/P


incident


) of approximately 12.5%. Including a fiber bundle transfer efficiency of 85%, and a diode electrical power to optical efficiency of 40%, the electrical diode power to optical efficiency of the intracavity frequency doubled laser is on the order of 4.25%. This is a large number in comparison to a small fraction of one percent that is typical for ion laser sources.




Optionally included in resonator


10


is a Brewster plate


34


or other polarizing device, which can be used to insure operation at a particular polarization. This is particularly useful when Nd:YAG is used for the gain medium. Brewster plate


34


can be positioned along the optical axis of either the third or fourth arms. A doubling crystal


36


is positioned in the second arm. In one embodiment, doubling crystal


36


is LBO. Other suitable doubling crystals include KTP, KDP, BBO, LBO, LiNbO


3


, ad KNbO


3


. The preferred doubling crystal is LBO. When LBO is used a heating element


38


may be included. A suitable heating element


38


is a resistive heater or a thermoelectric device as available from Melior, Trenton, N.J. 08648.




LBO doubling crystal


36


is used in a Type I, non-critical-phase-matched (NCPM) geometry, where the phase-matching is controlled with temperature, typically about 145 to 175° C. when doubling from 1.053-1.064 m. to 527-532 nm. A typical temperature is 155° C. The high acceptance angle of non-critical-phase-matching (NCPM) in LBO doubling crystal


36


allows resonator


10


to be adjusted in a manner that yields high beam quality and nearly TEM


00


operation. Other types of phase matching may not preserve beam quality with tight focussing, which can cause multi-spatial mode behavior, or elliptical doubled beams.




In these embodiments of this invention, the green or doubled output beam is substantially round since Type I, non critical phase matching (NCPM) is employed, which provides a large acceptance angle and minimizes walk-off, as will be recognized by those of ordinary skill in the art. This is very helpful when a “double pass geometry” is employed. If a nonlinear crystal with nonzero walk off is used, like KTP, the double pass geometry can result in an imperfect overlap between the two doubled beams; one is generated upon the first pass, the second through the second pass. The nonlinear crystal lithium triborate, LBO, is used. The phase matching and doubling efficiency are optimized by optimizing the crystal temperature in the vicinity of 145 to 175° C., with ˜155° C. being typical for conversion of 1064 nm to a 532 nm harmonic.




When Nd:YLF is used as laser crystal


22


, and the size of pump beam


30


in the crystal


22


is optimized, a substantially TEMP


00


output beam can be generated even without an intracavity aperture. However, when other gain media are used, an aperture must sometimes be used if TEMP


00


operation is desired.




In another embodiment, power scaling is accomplished with the use of two such fiber-bundle coupled diode bars used to pump an Nd:YVO


4


laser crystal. Further, an intracavity-doubled Nd:YVO


4


laser with output greater than 1 W and nearly diffraction-limited output is produced. In the high power Nd:YVO


4


embodiments, optimum performance is typically achieved when the pump beam size in the laser crystal is slightly larger than the size of the TEM


00


mode in the laser crystal. This is in opposition to the teachings of classic mode-matching of diode-pumped solid-state lasers, in which the pump beam size is typically somewhat smaller than that of the TEMP


00


mode. The mode size can be as small as 0.8 of the pump beam size. This configuration is optimal because of strong thermal aberrations in the end-pumped laser crystal. This is also true for other materials with strong thermal focussing characteristics, including but not limited to Nd:YAG.




The output power is typically on the order of 6 W of green in a TEM


00


mode, which corresponds to an optical efficiency (P


out


/P


incident


) of approximately 23%. Including a fiber bundle transfer efficiency of 85%, and a diode electrical power to optical efficiency of 40%, the electrical diode power to optical efficiency of the intracavity frequency doubled laser is on the order of 8%. If TEMP


00


is not required, or if very low noise is not required, up to 8 W of 532 nm output can be obtained for the same 26 W of pump power, representing even higher efficiency. When optimized for low noise, i.e., high amplitude stability, the RMS noise can be lower than 0.5% for a 6 W frequency-doubled output. Typically in the embodiment, the output beam is substantially TEM


00


. If care is taken to eliminate acoustic noise associated with vibration or water cooling, RMS noise on the order of 0.2%, or less, can be achieved. The highly amplitude stable output of these lasers is relatively insensitive to environmental parameters, such as temperature.




In certain embodiments, the green beam and infrared beam are nearly diffraction limited, and the RF noise spectrum shows no heterodyne peaks other than the primary c/2L peak at about 140 MHz. This indicates that the laser oscillates on its lowest order spatial mode. The RMS noise from 10 Hz to 10 MHz can be lower than 0.5%, and even 0.1% or less, although even a noise lower than 3% RMS can be useful. An aperture is typically used to insure TEMP


00


operation with high amplitude stability in the Nd:YVO


4


embodiment.




High amplitude stability is maintained for tilting of the LBO crystal to either side of normal, for translation of the Z position of the LBO doubling crystal


36


and/or high reflector


16


, and temperature tuning the LBO by +/−3 C of the optimum temperature. These adjustments sacrifice a small amount of useful output power of resonator


10


. In a preferred embodiment, the ends of the LBO crystal


36


are cut and polished to be non-parallel to one another and non-normal to the infrared intracavity beam propagation direction through the LBO crystal


36


. Any spurious reflection from the surfaces of LBO crystal


36


that strikes the end mirror


16


can degrade the amplitude stability of the doubled output beam. These spurious beams can strike the end mirror and then couple back into the main beam of the cavity, creating a undesirable degradation in performance. It is important to prevent these spurious beams from coupling back into the main intracavity beam. It is possible to use an aperture between crystal


36


and end mirror


16


to operate in this function.




Pump beam


30


passes through laser crystals


22


and an intra-cavity infrared laser crystal beam is created. The laser crystal beam is then incident upon a face of LBO crystal


36


. In the laser of

FIG. 1

, a small waist, on the order of about 50 m. diameter, is preferably generated inside LBO doubling crystal


36


in order to produce a very high intensity infrared beam within LBO crystal


36


. High intensity is needed because the conversion of infrared to green increases nonlinearly (as the square) with the infrared intensity. LBO crystal


36


is antireflection coated at both the infrared and doubled wavelength, as is typical for an intracavity doubled laser. These coated crystals are available from Fujian-Castech in China. The coating on LBO crystal


36


must provide very low reflectance at ˜1.064 m., such as <1%, and preferably <0.1%. The coating should also provide relatively low reflectance at 532 nm, such as <1% or better. Further, the coating must handle the high average power densities within the laser. The infrared laser crystal beam is focussed to the much smaller waist diameter by the inclusion of a lens


40


between output coupler


18


and LBO doubling crystal


36


and by the use of a high reflector


16


that has focussing powers. A suitable lens


40


is antireflection coated at the infrared and doubled wavelength, and is of focal length 30 to 50 mm. High reflector


16


has a radius of curvature, R1 of about 100 mm.




The infrared laser crystal beam (for Nd:YLF its wavelength is 1.053, for NdYVO


4


it is 1064 nm), travels in two directions in the fourth arm of resonator


10


. It travels from output coupler


18


through lens


40


and LBO doubling crystal


36


to high reflector


16


, and is reflected from high reflector


16


back through the doubling crystal. As a result of this double-pass through LBO crystal


36


, an output beam at 527-532 nm is generated. Because output coupler


18


is highly transmissive at 527-532 nm, an output beam


42


at 527-532 nm is generated by resonator


10


. 527-532 nm light is generated in the fourth arm in two directions, relative to LBO doubling crystal


36


, 527-532 nm light in the right hand direction, and 527-532 nm light in the left hand direction. The positioning of LBO doubling crystal


36


in the fourth arm between output coupler


18


and high reflector


16


creates a double pass geometry because 527-532 in light is generated in both directions, and output beam


42


is the sum of these beams. The relative phases of the 1.053-1.064 m. beam and the 527-532 nm beam are important; the two beams are phase-shifted with respect to one another by double-passage through the air in between crystal


36


and mirror


16


. The mirror coating on mirror


16


adds a further phase shift. This effect is known in the art; see for example Koechner,


Solid State Laser Engineering


, Vol. 3, p. 534. The net phase shift is ideally a multiple of 2π after this roundtrip, in order that the 1.053 m. (1064) beam and the 527-532 nm beam are in phase for the second trip through the doubler crystal


36


. Some compensation for an imperfect multiple of 2π can be accomplished by adjusting the temperature of the doubling crystal


36


. This effectively optimizes the overall phase-matching between the 1.053 nm beam and the 527 nm beam. This technique clearly extends to other wavelength pair as well (i.e. 1.064 m. to 532 nm, 1.34 m. to 670 nm, 1.047 m. to 523 nm, etc.)




In one embodiment using resonator


10


of

FIG. 1

, fiber-bundle-coupled diode bars are used for diode source


28


to longitudinally pump Nd:YLF laser crystals


22


with up to 8 W per Nd:VLF crystal. Useful output power is extracted via intracavity frequency doubling and can result in more than 2 W output power at 527 nm for 16 W of incident pump light from a single 20 W CW diode laser bar. Output beam


42


is substantially round and is of high quality. Doubling efficiency can be fine tuned by varying the temperature of LBO doubling crystal


36


from a preferred temperature of about 140 to 175° C. Both the 527 nm output beam


42


and the infrared laser crystal beam are nearly diffraction-limited. The RF spectrum indicates that under certain conditions substantially no heterodyne peaks other than the c/2L peak at nearly 140 MHz are present, indicating that resonator


10


is oscillating on its lowest order spatial mode. This can typically be accomplished without an aperture when Nd:YLF is used as laser crystal


22


, but a standard aperture is sometimes needed to insure TEM


00


operation, as is usually the cage when Nd:YVO


4


is used.




In this embodiment, resonator


10


has a length L of about 1 m. L is the optical path length of the entire resonator. L is nominally equal to the total of L1, L2, L3 and L4. Additionally, resonator


10


exhibits the lowest amplitude noise when resonator alignment, LBO doubling crystal


36


angle, and LBO doubling crystal


36


temperature are adjusted in a way that results in primarily a single peak at the c/2L frequency in the RF spectrum. LBO doubling crystal


36


can be slightly off of normal incidence with respect to the cavity mode for this effect. In this state, the % RMS noise is as low as less than 3%, preferably less than 2%, and most preferably less than about 1%, as measured by a standard RMS meter over a range of 10 Hz to 10 MHz.




Additionally, in this embodiment, the optical spectrum of resonator at both the 527 nm output beam and the 1053 nm laser crystal beam indicates that typically at least 3 axial modes are oscillating at any one time. In various embodiments, at least 4, 5, 6, 7, 8, 9, 10 or more axial modes may be oscillating. The bandwidth of the intracavity infrared laser beam is about 35 GHz while 527 nm is produced, indicating that 3, 4, 5, 6, 7, 8, 9, 10 or more axial modes can oscillate, and for certain alignment configurations the optical spectrum can have a structure indicating that the LBO doubling crystal


36


may be acting like an etalon. With the LBO doubling crystal


36


removed, the infrared axial mode spectrum stabilizes, with about 3 or more axial modes oscillating, and a bandwidth of about 30 GHz. The c/2L mode spacing is about 150 MHz for a 1 meter long resonator cavity


10


. The 527 nm output beam


42


has a bandwidth of about >70 GHz.




As more infrared power from laser crystals


22


is generated, more frequency doubled output power can be obtained. Additional pumping ports and pump sources or other laser crystals, such as Nd:YVO


4


can facilitate scaling. Nd:YAG can also be used, typically in conjunction with an intracavity polarizing device. To achieve high amplitude stability, we find that a large number of infrared axial modes must oscillate. With Nd:YLF as the active medium, this was achieve with the ˜1 meter long cavity. The cavity length (or resonator length) of diode pumped resonators for intracavity doubling of the prior art were typically very short, at typically less than 10 cm. Short lengths increase the c/2L axial mode spacing, and therefore typically decrease the number of axial modes that can oscillate within the available bandwidth of the laser crystals. Another technique for broadening bandwidth would utilize a broad bandwidth material, like Nd:LMA. Another resonator structure that would provide a long resonator length and many axial modes would be a resonator structure that includes an optical fiber. A very small c/2L spacing could be accomplished with a long fiber, placed intracavity with appropriate coupling optics.




Resonator


44


, illustrated in

FIG. 3

, has a simpler geometry than resonator


10


of FIG.


1


. Resonator


44


includes a first arm that is defined by a high reflector


44


and an output coupler


46


. Positioned along an optical axis of the first an is laser crystal


22


. A diode pump source


28


delivers pump beam


30


through optical fiber


32


, or a bundle of fibers. Lenses


24


and


26


focus pump beam


30


so it is incident on laser crystal


22


, and a laser crystal beam is produced. The second arm of resonator


44


is defined by output coupler


46


and a high reflector


48


. Resonator


44


has a length L that is equal to the total lengths of the first and second arms. Positioned along an optical axis of the second arm is a doubling crystal


36


. An optional Brewster plate


34


, or polarizing device, can be positioned in resonator


44


such as in the first arm. The inclusion of Brewster plate


34


or another polarizing device is particularly desirable when Nd:YAG is used.




The laser crystal beam is reflected from output coupler


46


and is incident on doubling crystal


36


. The second arm provides a double pass geometry for the generation of the frequency doubled output beam


42


. If LBO is used as doubling crystal


36


, then a heating element, not shown, is necessary. Additionally not illustrated in

FIG. 3

is a lens disposed between doubling crystal


36


and output coupler


46


. Inclusion of the lens is dependent on the type of doubling crystal


36


used, and on the radius of curvature and focussing power of high reflector


48


and output coupler


46


.




Resonator


52


, shown in

FIG. 4

, does not include fold arms. Resonator


52


does not provide for a double pass geometry of the infrared beam through doubling crystal


36


, and a portion of frequency doubled output beam


42


is lost in resonator


52


. Again, through doubling crystal


36


frequency doubled output beam


42


is generated in both directions. However, with resonator


52


, the portion of frequency doubled output beam


42


traveling to the left of doubling crystal


36


is lost.




As shown in

FIG. 5

, a power supply is included and is associated with diode modules


28


. Each diode module


28


can be a 20 watt diode bar, commercially available from OptoPower. Diode modules


28


are coupled to a fiber bundle


32


. Each fiber bundle


32


is connected to the resonator with a quick disconnect. The quick disconnect is in the vicinity of lenses


24


and


26


, which image the output of each fiber bundle


32


into the Nd:YVO


4


crystal


56


. This imaging is achieved through pump windows


58


and


60


which are coated with standard optical coatings and are highly transmissive at the pumping wavelength of approximately 809 nm, but highly reflective at the intra cavity wavelength of 1.064 microns. A standard multilayer dielectric coating is used, and is commercially from the Components and Accessories Group of Spectra-Physics Lasers, Mountain View, Calif. Pump sources


28


pump nominally along the longitudinal axis of a Nd:YVO


4


crystal. The region around the Nd:YVO


4


crystal of this resonator is in a “Z” configuration, and a mirror


28


is positioned at the end of the infrared section of the resonator. An apparatus


68


is utilized to ensure operation in the resonators lowest order spatial mode. Mirror


62


is highly reflective at


106


and may include a curvature of about 60 cm.




Mirrors


58


and


60


are flat, and the Z configuration defined by


58


,


60


and


62


can be in the form of an infrared module


67


, or an infrared laser, in which the output coupler has been removed. Infrared module


67


is then inserted into a larger laser. It will be appreciated, however, that a separate infrared module


67


need not be included. Instead mirrors


58


,


60


and


62


may be incorporated into the larger laser. Following the beam path it strikes mirror


64


, which can have curvature or may be flat. Mirror


64


is highly reflective at


106


.




The intracavity infrared beam propagates to mirror


68


, which is a coated motor that is highly transmissive in the green, and highly reflective at 1.064 microns. The beam then passes through a lens


72


which focuses the intracavity infrared beam tightly into LBO crystal


36


. Again, LBO


36


is positioned on a heater


38


, creating a high power density inside the LBO crystal. The LBO or other doubling crystal is typically wedged and/or titled to prevent undesired feedback of any beams reflected from its surfaces. This is preferred for amplitude stable, infra-cavity doubling. The infrared beam, with a very small waist, passes through LBO crystal


36


and then strikes mirror


70


, which is a dual reflector highly reflective at 532 nm and 1.064 microns. The infrared beam is then passed in a right direction back where it came from, thus creating a laser cavity. It also reflects back the green that was generated upon the first pass through LBO crystal


36


. The green is then passed through lens


72


and out of the cavity through an output coupler


68


into output beam


42


. Output beam


42


can have output powers of at least 1, 2, 3, 4, 5, 10, 15 or 20 watts.

FIG. 5

illustrates a double pass configurations with infrared beam passing left through LBO crystal


36


and right, resulting in the generation of green. Mirror


70


reflects the left going green and then the right going green, with both of them exiting from resonator


44


as beam


42


.




Referring now to

FIG. 6

, a schematic diagram of a two port, diode-pumped, multi axial-mode, intracavity frequency doubled laser utilizing Nd:YVO


4


and LBO is illustrated. The laser at

FIG. 6

can be, (i) shorter than the resonator in

FIG. 5

, for example, approximately 0.6 meters, and (ii) is simpler. It has one less mirror and one less lens. Lens


72


and mirror


64


of

FIG. 5

are not included in FIG.


6


. Output beam


42


is directed in a different direction trough output coupler


74


. Output coupler


74


has a fairly strong curvature, is highly transmissive of green, and highly reflective of infrared. Again as in

FIG. 5

, mirrors


58


,


60


and


62


can be incorporated into a separated infrared module


67


.




Nonlinear crystal KTP


36


can be also be used effectively with the present invention if care is taken to insure both phase matching and appropriate polarization rotation effects. Polarization rotation of the fundamental infrared beam occurs because of the birefringence of KTP; with the KTP acting as a multiple order waveplate, as is well known in the art. These effects must be controlled for efficient, stable, intracavity doubling when KTP is used: see for example Nightingale et al. or Weichmann et al. This is accomplished by rotating the KTP crystal to an orientation for optimal Type II doubling, angle-tuning for maximum frequency conversion, and temperature-controlling KTP crystal


36


for a single or double pass polarization rotation that matches an eigenmode of the laser cavity with a defined polarization rotation that matches an eigenmode of the laser cavity with a defined polarization state. For example, if an eigenmode has a defined polarization state within the laser crystal medium, the single or double pass of the KTP crystal and other intracavity elements should be arranged to provide the same polarization state upon one round trip of the laser cavity. One case is where the polarization is linear (i.e., vertical or horizontal) in the gain medium; other polarization states can also be used, as described by Oka et al.




These same techniques apply to the multiaxial mode case, but the use of the highly multiaxial mode doubling technique to generate amplitude stable output has not been presented by others. In an embodiment of the present invention, using Nd:YVO


4


pumped by 26 W of diode pump power from two fiber-coupled diode bars and a resonator configuration similar to that of FIGS.


5


and


6


, approximately 5 W of TEMP


00


output is generated with approximately 0.5% RMS noise.





FIG. 7

is a schematic diagram of a two-port, diode-pumped, multi axial-mode, intracavity frequency doubled laser utilizing Nd:YVO


4


and LBO, and further includes a frequency tripling crystal


80


. Tripling crystal


80


is mounted on a heater


82


, and is positioned closely adjacent to LBO crystal


36


. Mirror


76


is trichroic and highly reflective at 1.064 microns, highly transmissive at 532 nm and 355 nm. Opposing mirror


78


is a three-wavelength high reflector, e.g., it is a high reflector at 1.064 microns, highly reflective at 532 nm and at 55 nm. With the embodiment of

FIG. 7

, output beam


42


is a UV beam. In other configurations, two or more UV beams can be generated. It will be appreciated that tripling crystal


80


can be LBO but may also be a different material. Crystal


80


is cut at a different angle from that of LBO crystal


36


in order to achieve phase matching of a different type. That is, it does not phase match the doubling of 1064 to 532 nm. Instead, it phase matches the summation of 1064 nm and 532 nm in order to optimally generate a UV beam at 355 nm. LBO crystal


36


and crystal


80


are positioned in a close proximate relationship because this is where the mode is the smallest. It is desirable to generate high intensity both in the green and in the infrared in order to enhance the nonlinear tripling process.




Referring now to

FIG. 8

, a multi-axial mode, intracavity doubled laser is used as a pump for a second laser, including but not limited to a Ti:Al


2


O


3


laser. The laser illustrated in

FIG. 8

includes a power supply


84


, diode pump sources


88


, a multi-axial mode intracavity doubled laser


90


, a doubled output beam


92


, optional optics


94


, a second laser


96


that produces a desired output beam


98


.




Referring now to

FIG. 9

, a retinal photocoagulator system includes a power supply and system controller


100


, a power supply


102


, diode pump sources


104


, a multi axial mode intracavity doubled laser


106


, optics


108


, a beam delivery device


110


, including but not limited to a fiber optic, additional optics


112


, a magnifier/optical system


114


for the physician, a slit lamp


116


, as well as other optical components such as mirrors.




The present invention is a diode pumped, multi axial mode, intracavity doubled laser with low amplitude noise. This is created by oscillating a plurality of axial modes, such as 3 or more, and in some instances 10 or more. In one embodiment, a long resonator structure creates the multi axial modes. The length of the resonator can be in the range of 0.3 m to 2 m. Other techniques can also be used to insure multiaxial mode operation. For example the laser crystal


22


can be positioned as close as possible to one end of the resonator, taking advantage of the effects of spatial hole burning which is maximized as laser crystal


22


is moved close to an end. Additionally, a high reflecting coating can be place on laser crystal


22


. Laser crystal


22


materials with broad band widths, on the order of a few hundreds of GHz, can also be used to generate the many axial modes.




Although a green output beam has been described, blue, red, near infrared and beams of other wavelengths are possible, depending on the choice of laser and doubling crystals.




Changes and modifications in the specifically described embodiments can be carried out without departing from the scope of the invention which is intended to be limited only by the scope of the appended claims.



Claims
  • 1. A diode pumped, intracavity doubled laser, comprising:at least two resonator mirrors defining a resonator cavity; an Nd:YVO4 laser crystal positioned in the resonator cavity; an LBO doubling crystal positioned in the resonator cavity; a diode pump source supplying a pump beam to the laser crystal and producing a laser crystal beam incident on the doubling crystal to produce a frequency doubled output beam with an output power of at least 1 watt, wherein a TEM00 mode size in the laser crystal is smaller than a pump beam diameter in the laser crystal and the diode pump source is configured to be coupled to a power supply.
  • 2. The laser of claim 1, wherein the output power is at least 2 watts.
  • 3. The laser of claim 1, wherein the output power is at least 3 watts.
  • 4. The laser of claim 1, wherein the output power is at least 4 watts.
  • 5. The laser of claim 1, wherein the output power is at least 5 watts.
  • 6. The laser of claim 1, wherein the output power is at least 10 watts.
  • 7. The laser of claim 1, wherein the output power is at least 15 watts.
  • 8. The laser of claim 1, wherein the output power is at least 20 watts.
  • 9. The laser of claim 1, wherein the doubled output beam has a % RMS noise of less than 0.5%.
  • 10. The laser of claim 1, wherein the doubled output beam has a % RMS noise of less than 0.3%.
  • 11. The laser of claim 1, wherein the doubled output beam has a % RMS noise of less than 0.2%.
  • 12. The laser of claim 1, wherein the doubled output beam has a % RMS noise of less than 0.1%.
  • 13. The laser of claim 1, wherein a TEM00 mode size in the laser crystal is smaller than a pump beam diameter in the laser crystal.
  • 14. The laser of claim 1, wherein the diode pump source is a diode bar.
  • 15. The laser of claim 1, wherein the diode pump source is a plurality of diode bars.
  • 16. The laser of claim 1, wherein the diode pump source is fiber-coupled.
  • 17. The laser of claim 1, wherein the output beam is substantially TEM00.
  • 18. A diode pumped, multi axial mode, intracavity doubled laser, comprising:at least two resonator mirrors defining a resonator cavity; an Nd:YVO4 laser crystal positioned in the resonator cavity; an LBO doubling crystal positioned in the resonator cavity; a diode pump source supplying a pump beam to the laser crystal and producing a laser crystal beam with at least three axial modes that are incident on the doubling crystal to produce a frequency doubled output beam with an output power of at least 1 watt, wherein a TEM00 mode size in the laser crystal is smaller than a pump beam diameter in the laser crystal and the diode pump source is configured to be coupled to a power supply.
  • 19. The laser of claim 18, wherein the output power is at least 2 watts.
  • 20. The laser of claim 18, wherein the output power is at least 3 watts.
  • 21. The laser of claim 18, wherein the output power is at least 4 watts.
  • 22. The laser of claim 18, wherein the output power is at least 5 watts.
  • 23. The laser of claim 18, wherein the output power is at least 10 watts.
  • 24. The laser of claim 18, wherein the output power is at least 15 watts.
  • 25. The laser of claim 18, wherein the output power is at least 20 watts.
  • 26. The laser of claim 18, wherein the doubled output beam has a % RMS noise of less than 0.5%.
  • 27. The laser of claim 18, wherein the doubled output beam has a % RMS noise of less than 0.3%.
  • 28. The laser of claim 18, wherein the doubled output beam has a RMS noise of less than 0.2%.
  • 29. The laser of claim 18, wherein the doubled output beam has a % RMS noise of less than 0.1%.
  • 30. The laser of claim 18, wherein the diode pump source is a diode bar.
  • 31. The laser of claim 18, wherein the diode pump source is a plurality of diode bars.
  • 32. The laser of claim 18, wherein the diode pump source is fiber-coupled.
  • 33. The laser of claim 18, wherein at least at least four axial modes are incident on the doubling crystal.
  • 34. The laser of claim 18, wherein at least at least five axial modes are incident on the doubling crystal.
  • 35. The laser of claim 18, wherein at least at least 10 axial modes are incident on the doubling crystal.
  • 36. The laser of claim 18, wherein the output beam is substantially TEMP00.
CROSS-REFERENCE TO RELATED APPLICATIONS

This application is a continuation-in-part of U.S. patent application Ser. No. 08/446,203, filed May 19, 1995, which is a continuation-in-part of U.S. patent application Ser. No. 08/446,195, filed May 19, 1995 (U.S. Pat. No. 5,638,388, issued Jun. 10, 1997), which is a continuation-in-part of U.S. patent application Ser. No. 08/191,656, filed Feb. 4, 1994 (U.S. Pat. No. 5,446,749, issued Aug. 29, 1995), all of which are incorporated herein by reference.

US Referenced Citations (18)
Number Name Date Kind
4656635 Baer et al. Apr 1987
4701929 Baer Oct 1987
4884277 Anthon et al. Nov 1989
4951294 Basu et al. Aug 1990
5052815 Nightingale et al. Oct 1991
5127068 Baer et al. Jun 1992
5136597 Nightingale Aug 1992
5151909 Davenport et al. Sep 1992
5164947 Lukas et al. Nov 1992
5170409 Nightingale et al. Dec 1992
5180378 Kung et al. Jan 1993
5253102 Okazaki Oct 1993
5331652 Rapoport et al. Jul 1994
5446749 Nighan et al. Aug 1995
5511085 Manhall Apr 1996
5638388 Nighan et al. Jun 1997
5696780 Pieterse et al. Dec 1997
5991317 Nighan et al. Nov 1999
Foreign Referenced Citations (3)
Number Date Country
0 398 570 A2 Nov 1990 EP
0 556 582 A1 Aug 1993 EP
WO 9521480 Oct 1995 WO
Non-Patent Literature Citations (5)
Entry
Baer, T., “Large-amplitude fluctuations due to longitudinal mode coupling in diode-pumped intracavity-doubled Nd:YAG lasers”, J. Opt. Soc. Am. B, vol. 3, No. 9, pp. 1175-1180 (Sep. 1986).
Oka, M. et al., “Stable intracavity doubling of orthogonal linearly polarized modes in diode-pumped Nd:YAG lasers”, Optics Letters, vol. 13, No. 10, p. 805-807 (Oct. 1988).
James, G., et al., “Elimination of chaos in an intracavity-doubled Nd:YAG laser”, Optics Letters, vol. 15, No. 20, Oct. 15, 1990, pp. 1141-1143.
Magni, V. et al., “Intracavity frequency doubling of a cw high-power TEM00 Nd:YLF laser”, Optics Letters, vol. 16, No. 24, Dec. 15, 1993, pp. 2111-2113.
Marshall, L. et al., “Intracavity Doubled Mode-Locked and CW Diode-Pumped Lasers”, IEEE Journal of Quantum Electronics, vol. 28, No. 4, Apr. 28, 1992, pp. 1158-1163.
Continuation in Parts (3)
Number Date Country
Parent 08/446203 May 1995 US
Child 09/070478 US
Parent 08/446195 May 1995 US
Child 08/446203 US
Parent 08/191656 Feb 1994 US
Child 08/446195 US