The present invention is related to a dual-tube mass flow sensor combining the technologies of the Coriolis flow measurement and the thermal flow measurement, specially targeting gas applications.
Coriolis flow sensors are based on the Coriolis principle, that is when a mass moving in a rotating system, Coriolis force will be produced. Coriolis flow sensor have many advantages:
Coriolis flow sensor also has its limitation, a major one is the difficulty to use them in gas applications stemmed from the low density of gases. For the Coriolis flow sensor, the signal strength is directly proportion to the mass flow rate. As gases have much low densities than liquids, for the same pressure drop, the Coriolis tube will flow much low mass flow rate, this will make the signal much weaker when flowing gases, especially at low end of the flow, the background noise will make the measurement impossible. The consequences are: first, the ranges of gas applications are much narrower; another one is the minimum detectable mass flow rate or resolution is not low enough. For liquid applications, the turn-down ratio of the Coriolis sensor can easily reach to 200:1, but for gas applications, it will be difficult to reach to a 50:1 turn-down ratio. In many cases, especially for lighter gases, it can only reach 20:1 turn-down ratio or worse.
To solve the issues of using the Coriolis technology in gas applications, the thermal flow measurement technology is included in this invention.
Thermal mass flow measurement is based on the thermal cooling effect of the flowing fluid. They usually use one or more heated sensing element(s), placed in the vicinity or inside of the flow path, by measuring the temperature change of the element(s) caused by cooler fluid to decide the flow rate. The major advantages of thermal flow sensors are:
In this invention, a thermal measurement tube is arranged in parallel with the Coriolis tube. In the low flow end, such as below 10%, the thermal measurement will take over. In this way, the flow sensor will keep all the benefits of Coriolis measurement, but compensated the shortcoming of it at the low flow end.
In the thermal tube, two coils are wound on it and they are heated by the currents flow through, by measuring the temperature changes brought by the flow, the flow rate can be measured. The calibration of the thermal measurement will be based on the Coriolis measurement, so all the benefits of Coriolis measurement will be kept. The results from the thermal measurement will be combined with the Coriolis measurement results to cover the whole flow range.
Thermal tube 2 has a cover 11 in the middle of its top horizontal beam. Inside cover 11, there are two coils 12 and 13 wound on thermal tube 2 (
To create a stable thermal environment, cover 11 made of copper sheet is installed.
They are laser-welded to sensor base 3 at 21 and 22. For Coriolis tube 1, to strengthen the connection and provide a stable support, it will be brazed at 23.
In this embodiment, Coriolis tube 1 is made of 316L with an ID 2.286 [mm], OD 2.413 [mm] and 194 [mm] long. With a pressure drop 14.7 [psi], the full flow rate for water is close to 110 [kg/h].
When flowing Nitrogen, the flow rate under 14.7 [psi] pressure drop is about 1 [kg/h], or 57 [SLM], which is shown on
Depend on the gas and other factors, when the flow rate below certain percentage of the full flow rate (here is 50 [SLM]), the error will be unacceptable. We assume that for Nitrogen this percentage is 5%. For Hydrogen, Helium and other light gases, this percentage may be 10% or higher. As a demonstration, we will use 10% as a divider, below 10%, that is 5 [SLM], the thermal measurement will be used to measure the flow rate and above 10%, Coriolis measurement will be used. From the pressure drop calculation (
We now need to find out at 0 to 7 [sccm], what kind of output thermal sensor can provide. In the thermal measurement of the mass flow rate, coils 12 and 13 will be heated up. There are different schemes to do the heating, such as constant current, constant temperature or constant temperature drop. In this demonstration, we will use the constant current scheme as shown in
The coil temperature change will result in its resistance change:
R=R
0[(1+α(T˜T0)], (1)
where: R and R0 are the current and the initial coil resistances, respectively;
α is the temperature coefficient (1/K), for the resistant wire used, this value is around 0.0045;
T and T0 are the current and the initial coil temperatures, respectively.
If we assume that a constant 12-mA current i is applied to both coils and we also assume that the initial resistances for both coils are 308 ohms. Based on these values and Equation (1), the coil resistance change is showing in
The voltage drops V across each coil can be calculated by
V=i·R, (2)
They are plotted in
With an addition of thermal tube, the calibration is a little more complicated the one with only Coriolis tube. As shown in
Q
T
=Q
c
+Q
t, (3)
where QT, Qc, and Qt are the flow rates of total, Coriolis tube and thermal tube.
For the Coriolis flow, the flow rate Qe is a linear function of phase angle difference, that is
Q
C
=M·φ, (4)
where φ is the phase angle difference between upstream leg and downstream leg of the sensor; and M is a constant.
For the thermal tube flow, we can use a two-order polynomial equation to fitting the curve:
Q
t
=a+b·φ+c·φ
2, (5)
where a, b and c are fitting coefficients.
We can combine Equations (3), (4) and (5) together as
Q
T
=a+(M+b)·φ+c˜φ2=A+B·φ+C·φ2, (6)
where constants A, B and C can be decided by the calibration and saved in the PCB RAM for the later use in operation.
It can be seen from
Below 10% of the flow rate, we will totally rely on the thermal output. Depend on the curve linearity, different scheme can be used to interpolate the data. For the near-straight-line V-Q curve as shown in
The calibration will also decide the value φ10%, that is the phase difference angle when the flow rate is 10% of the full flow. These values will be saved and retrieved during measurement. The procedure will be: first check whether the phase angle is above φ10, if yes, use Equation (6) to get flow rate; if not, use Equation (7). If the thermal output is not very linear, then more sophisticated linearization and interpolation scheme should be used.
It is known that the thermal sensor is not very age-stable, that is one of the reasons that people are trying to switch to other measuring technologies or trying to recalibrate the thermal sensor in-line in recent years. With this invention, the thermal sensor output can be recalibrated easily. If it is a controller, the recalibration can be implemented per schedule, such as every 6 months, or even each power-up. For example, at each power-up or scheduled recalibration instant, the controller will control the flow rate flowing from zero up to pass the thermal-Coriolis division flow rate (10% in this demonstration), while passing the 10% flow rate Q10%, the V10% will be recorded down and saved in the RAM. If it is a meter, it can also be recalibrated in-line with a little help. For example, at each power up, by using manual control valve or system-controlled valve upstream of the unit to make the flow rate going up from zero to pass the Q10% and record down the V10%.
Thermal sensor usually has better than 1% sensitivity. For the case showing here, it means that the sensitivity is 50 [sccm], 1% of Q10%, which is 5 [SLM]. The total turn-down ratio will be: 50/50,000=1:1000, an astonishing number.
As the thermal measurement is under the control of Coriolis measurement, the hybrid sensor will keep the benefit of Coriolis sensor, such as fluid insensitivity, etc.
In other embodiments, the full flow rates of the Coriolis sensor tubes can be in 10 [kg/h], 1000 [kg/h] or high flow rate levels. The diameters of the Coriolis tubes can be different with the same size of the thermal tube. From the accuracy point of view, higher flow rate units benefit more, because the flow rate of the thermal tube flow will take smaller part of the total flow. For flow rate 1000 [g/h] or less, it may lose too much accuracy due to the error caused by the thermal tube. In such case, one tube doing both Coriolis measurement and thermal measurement functions may be more suitable (in another patent). For some lighter gases, such as Hydrogen and Helium, the thinner thermal tube, such as 0.008″ ID, may be needed. For the thermal tube, instead of heating coils, MEM film sensor may be used.