The present invention pertains to back support systems, in particular to a dynamic back support system that provides back support during forward flexion through a three-dimensional range of motions.
Standing in a forward flexed position is an awkward posture due to the flexion extension moment created by the trunk as it bends forward away from the centerline. To hold this position the low back and hip extensor muscles must work continuously to counterbalance the moment composed of the weight of the upper body and external load, when applicable, times the length of the lever arm. Since the back extensors (erector spinae) work on a small lever arm an amplified tensile force is needed and as a result applies large compressive and shear forces through the lumbar spine. Stationary loading has been shown to increase the incidence of low back pain (Silverstein, Silverstein and Franklin, 1996), therefore lightening the biomechanical load on the lumbar spine should reduce the incidence of low back injuries. Once an injury has occurred, workers may be delayed in their return to work as they wait for sufficient healing and strengthening to take place so as their physical capacity matches the work demands.
Standing at work is a prevalent work posture (Tissot et al. 2005) but to date there are no effective assistive devices that lower the workload of the trunk muscles and reduce the compressive forces through the spinal joints (Swie and Sakamoto, 2004). In sitting, chair backrests have been shown to lower the muscular electromyography activity of the back extensors and abdominals (Makhsous, 2003) but this is not helpful when an individual leans forward to carryout a work task in front. An on-body personal lifting assist device (PLAD) has been shown to offset the spinal loads and reduces the electromyography of the erector spinae by 14%-21% and the compression and shear forces at L4/L5 by 13% to 15% (Abdoli et al. 2007, Lotz et al. 2007, Graham et al. 2008, Frost et al. 2008). PLAD is modeled on the concept of human muscle through the use of an elastic element that acts as an external muscle force generator but it has certain limitations associated with being an on-body device and appears best suited for dynamic work such as manual handling tasks with large vertical lifting. Alternate work postures include sitting or standing upright in a neutral position. Standing compared to sitting produces lower compressive forces through the lumbar spine (Callaghan & McGill, 2001), Psychophysically, however, the preference and perceived effort is mixed. Yates and Karwowski, (1992) found subjects perceived sitting to be harder and the authors attributed the difference to change in lumbar curvature. Kim et al., (2004) also found a higher perceived load in sitting but this was only for smaller subjects. Johnson and Nussbaum (2003), found that perceived effort was higher in a standing waist bend posture possibly due to more stability gained from leaning onto the fixture. An early study by Aaras et al. (1988) also showed that a seated posture was usually preferred despite increased load on the shoulder muscles and attributed to improved precision, stability, increased mobility and less load on legs and feet, less energy expenditure. Standing upright with the flexion extension moment neutral is ideal, however, this is not always possible and a forward flexed posture is needed when a work surface is below the elbow height or extreme reach beyond the length of the extended arm is needed. These two factors are unique to individuals and cannot always be accommodated by traditional ergonomic strategies.
Low back pain is the most common musculoskeletal complaint of workers with annual costs estimated at $12 Billion in 2002 in Canada (WorksafeBC, 2003) and $90 Billion in 1998 in the United States (Luo et al. 2004). Standing work has been identified as a risk factor for lower back pain as a result of the cumulative spinal loading and physiological work demands that result from the flexion moment that is created when the trunk leans forward from the upright neutral position. With as little as 10 degrees of forward lean the compressive loads through the lower lumber discs doubles (Takahashi et al., 2006). A recent population survey in Quebec showed that 58% of workers reported standing at work of which less than less 20% report that they can alternate position by sitting or walking (Tissot et al., 2005). Keyserling (1992), reported 89% of 335 surveyed manufacturing and warehouse jobs involved mild trunk flexion of less than 20 degrees. The degree of forward inclination depends upon individual anthropometry, design of workstation and nature of the work. The amount of forward leaning can be minimized by mechanical strategies that adjust workstation to fit the individual but this is not always practical, safe and can be costly.
An ergonomic device that safely lowers compressive loads may be effective in addressing the above-noted problems.
According to an aspect of an embodiment, provided is a device for supporting a portion of the upper body during forward lean, the device comprising:
a hinge joint provided on a base plate;
a central pillar extending upwards from the hinge joint, the hinge joint enabling movement of the central pillar through a range of motions;
a support plate situated on the central pillar at an opposite end from the hinge joint, the support plate providing support to a user in the region of the user's breast plate;
at least one resistor element operably associated with the central pillar to receive stresses delivered via at least one linkage unit during periods of forward lean, thereby transferring at least a portion of the upper body weight to the at least one resistor element.
According to some embodiments, the hinge joint is a ball and socket joint.
According to some embodiments, said range of motions of said hinge joint includes up to 80° flexion-extension, up to ±70° lateral bend, up to ±70° rotation, and combinations thereof.
According to some embodiments, each at least one resistor element is associated with a corresponding linkage unit.
According to some embodiments, a plurality of resistor elements is provided, each resistor element being operably associated with a respective linkage unit.
According to some embodiments, the at least one resistor element is located below the base plate, opposite the side comprising the hinge joint, and wherein the linkage unit operably associates the resistor element to the central pillar.
According to some embodiments, the at least one resistor element undergoes compression during periods of forward lean to offset a portion of the upper body weight away from the user.
According to some embodiments, the at least one resistor element is adjustable to suit anthropometric differences and user preferences.
According to some embodiments, the at least one resistor element is a spring.
According to some embodiments, the spring is a coil spring.
According to some embodiments, the at least one resistor element is a hydraulic actuator.
According to some embodiments, the device further comprises height adjustability in the central pillar to enable height adjustments of the support plate.
According to some embodiments, the height adjustability of the central pillar is provided by way of a telescoping arrangement between a first support shaft, and a second support shaft.
According to some embodiments, the device further comprises a springed telescoping element to provide length variability in the central pillar to allow for a degree of length variability during usage.
According to some embodiments, the springed telescoping element provides up to approximately 10 cm of length variability.
According to some embodiments, the device further comprises a pelvic support pad located on the base plate to enabling a user's hips to lean up against the device.
According to some embodiments, the pelvic support pad is pivotable through a range of motions of up to ±20°.
According to some embodiments, the device further comprises a clamp mechanism to allow for attachment of the device to a table top.
According to some embodiments, the central pillar extends below the base plate, terminating at a support platform upon which a user may stand, enabling the device to be used as a stand-alone unit.
According to some embodiments, the central pillar below the base plate is height adjustable.
According to some embodiments, the device is configured for attachment to a chair allowing for support during a forward lean sitting position.
Embodiments will now be described, by way of example only, with reference to the attached Figures, wherein:
The dynamic trunk support (DTS) device is a mechanical device that reduces biomechanical loading and physiological work of the lower back by continuously supporting the trunk during forward leaning. It has been determined that supporting a portion of the upper body weight reduces the flexion moment thus reducing the muscular work necessary to counterbalance the moment and resulting spinal load. The device is intended to reduce the incidence of lower back injury, promote an early return to work following a lower back injury and increase stability in standing. The DTS device described herein is designed to support a portion of the body weight through the bony ribcage while allowing full axial movement.
Turning now to
Shown in
Referring now to
The resistor elements 38 can be adjustable to suit anthropometric differences as well as user preference. The resistor elements can be independently adjustable, or alternatively adjustable as a complete unit. In the case of springs, the springs themselves can be interchanged to suit a particular user. In the case of hydrautic actuators, built in adjustments can be used to alter their compression characteristics.
The extent of compression of each of the resistor elements 38 is dependent upon the extent of lean, and the direction of lean relative to the neutral position. As one will appreciate, in a forward lean directed towards the right-hand side, compressive stresses are delivered via the linkage units to the central and side resistor elements 38 on the left-hand side of the device 10, and visa versa. Motion of the central pillar 22 through the allowable range of forward lean has the effect of dynamically transferring to the resistor elements 38 a portion of the upper body weight. Support of the upper body is accomplished by way of redirecting the load through the DTS into the linkage units, in particular the resistor elements 38 in the form of compressive stresses, thereby supporting the forward lean throughout.
In addition to the plurality of linkage units provided, at least one further restraint linkage 46 can be provided on the side opposite the linkage units to prevent the central pillar 22 from falling towards the user when the user backs away from the DTS device 10. The restraint linkage 46 is not intended to support any loads. As such, it is not necessary to provide the restraint linkage 46 with a resistor element 38 as detailed above for the other linkage units.
To accommodate different users, the height of the support plate 24 can be adjustable. Optimal placement upon a user is generally 2 cm below the sternal notch. For this adjustability, as shown in
In some embodiments, during forward lean and movement through the range of motions allowable with the DTS device 10, the height of the support plate 24, in particular the distance between the hinge joint 20 and the support plate 24 allows for a degree of length variability. In the absence of this variability, that is with a central pillar 22 of fixed length, the support plate 24 is apt to shift upon the user's chest, causing discomfort over extended periods of use. As such, the DTS device 10 can be provided with a mechanism that can accommodate a range of length variability, while not substantially departing from the predefined height as previously established. To accomplish this, the central pillar 22 can be provided with a springed telescoping element that can compress or extend in the range of 1 to 10 cm to accommodate the aforementioned variations in length during usage. The springed telescoping element may be integrated into the aforementioned locking mechanism, or may be provided as a separate element on the central pillar 22. For example, as shown in
The base plate 34 is generally located at the height of the pelvic bone. To support the user in a forward lean configuration, and to provide added comfort during use, a pelvic support pad 62 is can be provided, as shown in
The DTS device 10 is suitable for use as a stand-alone device, or in combination with a table top. An exemplary stand-alone device is shown in
For a DTS device 10 suitable for use in combination with a table top, the device 10 is provided with a clamp mechanism 72 such as the one shown in
A combination table top/stand alone configuration is also possible. For example, the base plate 34 can be configured for interchangeable attachment to either the clamp mechanism as shown in
The support plate 24 and optional pelvic support pad 62 is constructed using materials generally known in the art of chair manufacture. In general, as shown in
The concept of supporting the trunk in forward lean can also be applied to sitting work where the leaning forward position still creates a flexion moment that the erector spinae must work to balance. Sitting has been shown to have higher compressive disc loads. Shown in
It may also be advantageous to configure the stand-alone or table-top versions of the DTS device with a fold-away or retractability feature. In particular, in the event the DTS device presents an obstruction or impediment to work when forward leaning is not necessary, the ability to retract or fold-away the device would be desirable.
In use, the DTS device can provide (1) continuous partial support through a range of motion, (2) three dimensional trunk movement, (3) load transfer over the most stable part of the anterior rib cage, (4) no interference with arm mobility, (5) no compression in the thoracic outlet area.
In one study that analyzed single plane forward flexion through a trunk angle of 0 to 50°, the DTS device was able to reduce the compression forces at the L4-L5 joint between 10 to 70% at 10 and 50 degrees respectively (see
The impact of the DTS device upon asymmetrical postures, spinal loading, trunk muscle activity, reach distance, comfort and perceived effort with dynamic trunk movement was assessed.
The DTS device may find application in a range of areas, including (1) industry, healthcare and the service sector, (2) orthopaedic rehabilitation, and (3) home support for the disabled and elderly.
The DTS device may also be used as a prescriptive clinical work brace for orthopaedic rehabilitation for injured workers with lower back injuries. It could be used in workplaces to facilitate a safe return to work following a significant low back injury such as disc protrusion, spinal fracture or instability. Further, the DTS device could be a prescriptive leaning device for elderly individuals at home who experience problems with balance and/or generalized weakness. It could be used as a postural support while carrying out activities of daily living, for example while brushing teeth or combing hair.
The DTS device may be marketed as a preventative ergonomic device for use in the workplace, particularly in the manufacturing, healthcare and the service sector where a significant portion of work is carried out in forward leaning trunk postures. The device reduces a worker's exposure to compressive and shear forces through the lumbar spine that eventually translate into lower back discomfort, productivity loss and injury claims. Other benefits from the use of the DTS device include increased core stability with an offbalancing reach, reduction in the physiological cost of work resulting from lessened postural muscular workload, promotion of a suitable spinal posture rather than a forward slump, and limitation of rotational movements during forward flex.
The DTS may be used as a clinical assistive device used in the early postoperative stage following a lower back injury and/or surgery or at a later recovery stage to protect bones and joints from compressive and shear joint forces and contractile soft tissue from excessive muscular tension following repair of a spinal fracture or disc protrusion. The DTS device could be adapted to provide an adjustable level of support that can be changed according to the stage of healing and tissue tolerance and adjustable limits for range of motion in three dimensions. For example, a surgeon could prescribe maximum of 30 degrees forward flexed trunk with no rotation and 80% support. In the later stage of recovery, typical functional restoration programs are directed to progressively increase tissue loading until there is adequate tolerance and reduced fear of pain or re-injury. The DTS device may serve as an adjunct to existing rehabilitation interventions. The DTS device may promote an earlier return to pre injury or modified work by simply lowering the spinal loads and muscular workload of the job and reducing the likelihood of an early onset of local back muscle fatigue.
The DTS device may also be used to assist seniors with diminished balance, generalized weakness, or shortened reach distance. Walking aids used for balance require the use of both hands and therefore are not available to carry out simple tasks. Given the DTS device does not limit arm use, it could be used as a postural support while carrying out activities of daily living, such as brushing teeth or combing hair.
Postural supports are generally rigid in nature and can impede normal work movement patterns and be uncomfortable to wear or use. The DTS device is based on a biomechanical principle using leverage and support to lower physiological demands. It is based on dynamic splinting providing partial support, thereby limiting the risk that the spinal joints and discs are kept in a static position. Physiological movement is active assisted thereby preventing muscle atrophy from disuse and work specific deconditioning.
It will be appreciated that, although embodiments have been described and illustrated in detail, various modifications and changes may be made. While several embodiments are described above, some of the features described above can be modified, replaced or even omitted. For example, while the hinge joint is described as a ball and socket joint, other joint mechanisms providing the range of angular motions may be possible, for example a universal joint. The DTS device can also be modified to include a lubrication mechanism to reduce friction in the joint. Still further alternatives and modifications may occur to those skilled in the art. All such alternatives and modifications are believed to be within the scope of the invention and are covered by the claims appended hereto.