BRIEF DESCRIPTION OF THE DRAWINGS
The drawings are only for purposes of illustrating embodiments and are not to be construed as limiting the claims.
FIG. 1 illustrates a physiological parameter analysis system that detects ECG electrode misplacement from acquired ECG signals and optionally corrects the misplaced electrodes.
FIG. 2 illustrates a method that detects ECG electrode misplacement from acquired ECG signals and optionally corrects the misplaced electrodes.
FIG. 3 graphically illustrates electrode placement for a standard twelve lead ECG system.
FIG. 4 graphically illustrates electrode placement for the EASI ECG system.