This invention relates to electro-spun cardiovascular implants.
Current cardiovascular substitutes and implants encounter risks due to coagulation, infections, degeneration, and no growth possibilities. Tissue engineering is a fairly new approach which uses patient's own cells and a biodegradable polymer scaffold to make autologous tissue that is able to grow, adapt and repair. Polymeric scaffolds can be constructed from biocompatible, non-toxic polymers. The choice of polymer and the technique used to make the scaffold effects the mechanical properties exhibited by the scaffold.
For cardiac tissue engineering, the most commonly used biodegradable synthetic scaffold materials are polyglycolic acid (PGA), polylactic acid (PLA), polyhydroxybutyrates (PHB), ε-polycaprolactone (PCL) or their copolymers. US20150173921 teaches the use of supramolecular compounds as the basis of the biodegradable synthetic scaffold for cardiovascular implants.
Despite the recent advances, there is still an unmet medical need, in particular, since cardiovascular substitutes and implants such as heart valves, and hereby preferably for aortic or pulmonary heart valves, have to pass high standards as provided by the FDA and the relevant guidelines such as ISO 5840-1:2015, ISO 5840-2:2015 and ISO 5840-3:2015. An important emphasis for said high standards is hereby placed on hydrodynamic, durability, and, in particular, on fatigue testing to ensure reasonable assurance of safety, effectiveness and FDA approval.
The present invention addresses this unmet need by providing tissue engineered cardiovascular implant with enhanced durability and fatigue resistance, while maintaining the effectiveness as a cardiovascular implant.
A biodegradable cardiovascular implant is provided for growing cardiovascular tissue in a patient, comprising an electro-spun network having {in alternate embodiments described as ‘consisting essentially of’ or ‘consisting of’} supramolecular compounds having hard-blocks covalently bonded with soft-blocks, wherein the soft-blocks are a polycarbonate soft-blocks each with a molecular weight range of 500-2000, and wherein the hard-blocks comprise 2-ureido-4[1H]-pyrimidinone (UPy) compounds and chain extenders at a range of 1.5 to 3 for the chain extenders over the UPy compounds.
Embodiments of the invention show much enhanced durability and fatigue resistance, while maintaining the effectiveness as a cardiovascular implant.
The present invention provides tissue engineered cardiovascular implant with enhanced durability and fatigue resistance, while maintaining the effectiveness as a cardiovascular implant.
Supramolecular compounds are defined as hard-blocks covalently bonded with soft-blocks. The hard blocks are based on UPy moieties. The soft block is the backbone of the supramolecular compounds. Polycarbonate (PC) was used as it showed surprisingly benefit for the purposes and objectives of this invention, especially compared to polycaprolactone.
The ratio between the soft block and the hard block has an influence on the material properties. Herein, we describe that ratios of components within the hard block section has a tremendous impact on properties such as durability. We disclose here a specific combination of ratios within the hard block and length of the polymer used to form soft block that lead to enhanced mechanical properties (durability). Specifically, polycarbonate with a molecular weight range of 500-2000 provide enhanced durability and reduced fatigue compared to e.g. polycaprolactone. The hard block is composed of the Upy component, a diisocyanate and a chain extender. The ratio (R) within the hard-blocks for 2-ureido-4[1H]-pyrimidinone (UPy) compounds and chain extenders at a range of 1.5 to 3 for the chain extenders over the UPy compounds.
PCL Polymer—XP1, XP2
To synthetize XP1, telechelic hydroxy terminated polycaprolactone with a molecular weight of 800 g/mol (30.0 g, 37.5 mmol, dried under vacuum), 1,6-hexanediol (4.4 g, 37 mmol), and UPy-monomer (6.3 g, 37 mmol) were dissolved in dry DMSO (105 mL) at 80° C. To this reaction mixture was added hexamethylene diisocyanate (18.8 g, 111.5 mmol) while stirring, followed by the addition of one drop of tin dioctoate. This reaction mixture was stirred overnight at 80° C. The next day, the reaction mixture was cooled to 25° C. and its viscosity was lowered by the addition of additional DMSO in order to precipitate the mixture in water. The polymer was collected as white elastic solid, redissolved in chloroform/methanol (7/3 v/v) and reprecipitated in an excess methanol. This resulted in a clear elastic solid after drying under vacuum at 50° C. SEC (THF, PS− standards): Mn=13 kg/mol, D=1.6. See also WO2014185779A1. XP2 is synthesized in a similar manner, with the exception of the quantity of 1,6-hexanediol which is increased to 74 mmol. The composition of XP1 and XP2 polymers are summarized in TABLE 1.
PC Polymer—XP3
Polymers made with polycarbonates with molecular weight varying from 500 to 3000 g/mol were synthetized in a similar manner as for XP1. The changes were made depending on the length of the polycarbonate and the desired ratio between the components. Molar ratio can be expressed as followed. A (polycarbonate) is fixed at 1. B (chain extender) varies between 0 and 3, D (Upy) from 0.3 to 2 and and C is always equal to 0.8 to 1.2 times the total molar amount of A plus B plus D. Molar ratio B/D is noted R. For the purpose of this invention, XP3 was synthesized by using a polycarbonate molecular weight of 2000 g/mol and selecting a molar ratio R of 2. The composition of XP3 is summarized in TABLE 1.
If not otherwise mentioned, the thickness in the examples is 500 μm for all polymers.
Description of the Test
The accelerated wear test is a test aimed at evaluating durability of devices. The device is subjected to pressure conditions simulating in vivo conditions and the number of cycles before failure is recorded. Details about durability assessment are described in ISO 5840-3:2013.
First the PVs are tested for 20 min at 30/10 mmHg (20 mmHg) to check opening of the leaflets: maximum/mean systolic pressure gradient, effective orifice area and regurgitation. This pressure condition is classified as a normotensive pulmonary condition (ISO 5840-3:2013). At the start, and after 20 min data and high-speed movies are acquired. The maximum systolic pressure gradient should be below 25 mmHg (Drossner et al. Pediatr Cardiol. 2008 May; 29(3):649-52 doi: 10.1007/s00246-007-9191-y.), and the effective orifice area should be larger than 30% of the geometric orifice area.
Subsequently the valve is tested for 20 hours at 90/35 mmHg, which is classified as very severe hypertensive pulmonary conditions (ISO 5840-3:2013). At the start, after 20 minutes, and after 20 hours (or after failure) data and high-speed movies are acquired.
The materials tested are summarized in TABLE 1 and results are shown in TABLE 2. We observe that the PC-based polymer provides better results after 20 hours compared to PCL based polymers. XP3 shows the best fatigue resistance in the series tested, with no observed tears.
Pulmonary valve electrospun with PCL polyol (described in PCL/synthesis) as the leaflet material were tested in a valve tester. The pulmonary-valved conduit are evaluated after 20 hours at 90/35 mmHg (very severe hypertensive pulmonary conditions according to ISO 5840-3:2013). Leaflet made with PCL polyol showed tears and failure. On the contrary, leaflets made with PC polyol showed good results (TABLE 2).
The enhanced fatigue resistance of PC based polymers was further tested in aortic conditions (120/80 mmHg). The polymers were dissolved and further electro-spun and assembled to a stent to form an aortic valve. The valves were further tested in aortic conditions at 10 Hz. Within the PC based polymer family, it was then possible to discriminate which polymers gave the best results.
An additional feature that can also influence durability is the alignment of the fibers within the scaffold. The preferred fiber alignment is circumferential around an imaginary axis of the implant wherein the axis points in the direction of blood flow in case of a tubular implant. We can clearly see that alignment enables the increase in fatigue resistance from
Complementary Information
1. Ranges (Durability Focus)
2. Scaffold Structure
This application claims priority from U.S. Provisional Patent Application 62/611,431 filed Dec. 28, 2017, which is incorporated herein by reference.
Number | Name | Date | Kind |
---|---|---|---|
10766995 | Coury | Sep 2020 | B2 |
10876222 | Edouard Naz | Dec 2020 | B2 |
20090130172 | Dankers | May 2009 | A1 |
20100280594 | Sahatjian | Nov 2010 | A1 |
20150173921 | Lavrijsen | Jun 2015 | A1 |
20160115272 | Mes | Apr 2016 | A1 |
20180015202 | Dankers | Jan 2018 | A1 |
20180274131 | Naz | Sep 2018 | A1 |
20180325646 | Burke | Nov 2018 | A1 |
Number | Date | Country |
---|---|---|
WO2005042641 | May 2005 | WO |
WO2014185779 | Nov 2014 | WO |
WO-2016120456 | Aug 2016 | WO |
WO2016120456 | Aug 2016 | WO |
WO-2019129640 | Jul 2019 | WO |
Entry |
---|
Drossner et al. A management strategy for mild valvar pulmonary stenosis. Pediatr Cardiol. May 29, 2008(3):649-52 doi: 10.1007/s00246-007-9191-y. |
Number | Date | Country | |
---|---|---|---|
20190201588 A1 | Jul 2019 | US |
Number | Date | Country | |
---|---|---|---|
62611431 | Dec 2017 | US |