1. Field of the Invention
This invention relates to an electrode for an electrochemical device and a method for detecting hydrogen peroxide using the electrode.
2. Description of the Related Art
Hydrogen peroxide (H2O2) is a reactive oxygen species and a byproduct of several types of oxidative metabolism. Because accurate determination of H2O2 is of practical importance in the clinical, environmental and industrial fields, increasing interest has focused on fabrication of reliable H2O2 biosensors. Due to their high selectivity and sensitivity, electrochemical devices have been used extensively to detect H2O2.
In electrochemical devices, electrodes used to detect analytes are either enzyme-based or enzyme-free electrodes. Enzyme-free electrodes can be made by forming an active layer on a conductor. The materials for the active layer can be inorganic materials (i.e., metallic nano-particles, metallic oxides, transition metals, carbon nanotubes, etc.), organic materials (i.e., polyamic acid, polyaniline, poly(aniline-co-p-aminophenol)) or organic-inorganic materials (i.e., polyaniline-carbon nanotubes).
As to an electrochemical device using the enzyme-free electrode, an oxidation potential (approximately 0.5 V to 0.7 V) is usually measured in this device to determine the amount of hydrogen peroxide. Since the oxidation potential is susceptible to interference with other undesired substances, such as uric acid (UA) and ascorbic acid (AA), in a test sample, the specificity for hydrogen peroxide is reduced and the accuracy of the test result would be adversely affected. Therefore, improvements for the electrodes of the electrochemical devices are aimed at detecting H2O2 at reduction potential to eliminate the interference of interfering molecules.
W. Zhao et al. disclosed a multi-wall carbon nanotube/silver nanoparticle nanohybrids modified gold electrode for H2O2 sensors (Talanta 2009, 80, 1029-1033). This electrode can be operated under −0.15 V to −0.6 V, and has a sensitivity of 1.42 μA/mM, linear detection range of 0.05 mM to 17 mM, and a response time as low as 5 seconds.
The inventors of the present invention have published using poly (N-butyl benzimidazole)-modified gold electrode for the detection of hydrogen peroxide [Analytica Chimica Acta 2011, 693, 114-120]. The modified electrode detects hydrogen peroxide in the presence of carboxylic acid. The modified electrode has a detection range of 12.5 μM˜5.0 mM, with a sensitivity of 419.4 μA/mM·cm2, and a response time of 6.3 seconds.
The inventors of the present invention have also published electrode sensors to detect H2O2 by modifying Au (gold) electrodes with poly(amic acid-benzothiazole) (PAA-BT), poly(amic acid-benzoxazole) (PAA-BO), poly(amide-benzoxazole) (PA-BT) or poly(amide-benzothiazole) (PA-BO) (Biomaterial, 2011, 32, 4885-4895). These modified Au electrodes can detect H2O2 in the presence of acetic acid. PAA-BT-modified Au electrode has a sensitivity of 280.6 μA/mM·cm2, 0.025 mM to 5.0 mM detection range, and a response time of 5.2 seconds. PAA-BO-modified Au electrode has a sensitivity of 311.2 μA/mM·cm2, 0.025 mM to 2.5 mM detection range, and a response time of 3.9 seconds.
Accordingly, the detection of hydrogen peroxide and other analytes of interest using an electrochemical device would be ideal when the detection thereof occurs at reduction potential to prevent detection of undesired analytes. In addition, the electrochemical device should have a short response time and high sensitivity.
Therefore, an object of the present invention is to provide an electrode for an electrochemical device that can be used to detect H2O2 under reduction potential and that has short response time and high sensitivity.
According to a first aspect of the present invention, an electrode for an electrochemical device comprises:
a conductor; and
an active layer that is formed on the conductor and that includes a polymer having a functional group represented by the following formula (A) or (B) and a carboxylated material containing a carboxylic acid group;
wherein in formula (A), X is O or S; R1, R2, R3 and R4 are independently hydrogen, a C1 to C12 alkyl group, a C1 to C12 alkoxy group, an ether group, a cycloalkoxy group, a halogen group, a halogenalkyl group, a hydroxyl group, a trifluoromethoxy group, a trimethylflouro group, or a phenyl group;
wherein in formula (B), R5, R6, R7 and R8 are independently hydrogen, a C1 to C12 alkyl group, a C1 to C12 alkoxy group, an ether group, a cycloalkoxy group, a halogen group, a halogenalkyl group, a hydroxyl group a trifluoromethoxy group, a trimethylflouro group, or a phenyl group; and
wherein the carboxylated material is selected from the group consisting of a carboxylic acid-containing water-soluble polymer, a carboxylated carbon material, and the combination thereof.
According to a second aspect of the present invention, a method for detecting hydrogen peroxide includes the steps of:
contacting a test sample with an electrode of claim 1 such that nitrogen on the functional group of the polymer of the active layer on the electrode is oxidized;
applying a constant voltage to the electrode to reduce the oxidized nitrogen of the polymer of the active layer such that an electrical current is generated; and
measuring the electrical current.
According to a third aspect of the present invention, a method for detecting an analyte includes the steps of:
contacting a test sample with an electrode of claim 1, in the presence of an oxidase, such that nitrogen on the functional group of the polymer of the active layer on the electrode is oxidized;
applying a constant voltage to the electrode to reduce the oxidized nitrogen of the polymer of the active layer such that an electrical current is generated; and measuring the electrical current.
Other features and advantages of the present invention will become apparent in the following detailed description of the preferred embodiments of the invention, with reference to the accompanying drawings, in which:
This invention provides an electrode for an electrochemical device, which comprises:
a conductor; and
an active layer that is formed on the conductor and that includes a polymer with a functional group represented by the following formula (A) or (B) and a carboxylated material containing a carboxylic acid group.
In formula (A), X is O or S; R1, R2, R3 and R4 are independently hydrogen, a C1 to C12 alkyl group, a C1 to C12 alkoxy group, an ether group, a cycloalkoxy group, a halogen group, a halogenalkyl group, a hydroxyl group, a trifluoromethoxy group, a trimethylflouro group, or a phenyl group.
In formula (B), R5, R6, R7 and R8 are independently hydrogen, a C1 to C12 alkyl group, a C1 to C12alkoxy group, an ether group, a cycloalkoxy group, a halogen group, a halogenalkyl group, a hydroxyl group, a trifluoromethoxy group, a trimethylfluoro group, or a phenyl group.
The conductor for the electrode can be any materials that have conductivity. In a preferred embodiment of this invention, the conductor is gold (Au) electrode.
The carboxylated material is selected from the group consisting of a carboxylic acid-containing water-soluble polymer, a carboxylated carbon material, and the combination thereof.
The carboxyl acid group of the carboxylated material would react with hydrogen peroxide (H2O2).
Materials that have a carboxyl acid group, that consist of a plurality of pores and that do not affect the conductivity of the conductor can be used as the carboxylated carbon material of this invention. The pores can increase the surface area of the active layer.
The carboxylated carbon material is selected from the group consisting of carboxylated carbon tube, carboxylated graphene, carboxylated carbon spheres, and combinations thereof.
Examples of the carboxylated graphene include 1-keto-2-vinyl-butyric acid graphene and 1-keto-2-butyric acid graphene.
Since the carboxylic acid-containing water-soluble polymer is water soluble, certain carboxylic acid-containing water-soluble polymer would be dissolved in water, thus rendering the formation of pores in the active layer. Examples of the carboxylic acid-containing water-soluble polymer include, but are not limited to, polyacrylic acid, poly (2-ethylacrylic acid), poly(2,6-dihydroxymethyl-4-methylphenol-co-4-hydroxy benzoic acid), poly(acrylic acid-co-maleic acid), poly(styrene-co-methacrylic acid), and combinations thereof.
Preferably, the molecular weight of the carboxylic acid-containing water-soluble polymer is in the range of 2,000 to 3,000,000.
Preferably, the molecular weight of the polymer having the functional group (A) or (B) is in the range of 3,000 to 400,000.
The polymer with the functional group of formula (A) is selected from the group consisting of polyamic acid and polyamide derivatives, and has a repeating unit represented by the following formula (I), II), or (III).
In formula (I), (II), and (III), X is independently O or S.
In formula (I), when X is O, the polymer is poly(amic acid-benzoxazole) (PAA-BO). When X is S, the polymer is poly(amic acid-benzothiazole) (PAA-BT).
In formula (II), when X is O, the polymer is poly(amide-benzoxazole) (PA1-BO). When X is S, the polymer is poly(amide-benzothiazole) (PA1-BT).
In formula (III), when X is O, the polymer is poly(amide-benzoxazole) (PA2-BO). When X is S, the polymer is poly(amide-benzothiazole) (PA2-BT).
In examples of this invention, the active layer includes PAA-BO, PA1-BO, PA1-BT or PA2-BT.
The polymer with the functional group of formula (B) is a polyanline derivative. The polyaniline derivative has a first repeating unit represented by the following formula (PAn-1):
The polyaniline further includes a second repeating unit represented by the following formula (PAn-2)
The polyaniline derivative containing the first and second repeating units, (PAn-1) and (PAn-2), is used in an example of this invention.
Preferably, the weight ratio of the polymer to the carboxylated material ranges from 1:0.1 to 1:130.
The electrode of this invention can be made by well known methods. For example, the electrode can be made by applying a solution prepared by mixing a diamine polymer and a carboxylated material on a conductor, followed by drying in an oven. Application of the solution on the conductor can be conducted by coating or dripping the solution on the conductor or dipping the electrode into the solution.
The electrode of this invention can be assembled into an electrochemical device with other components, e.g., a counter electrode, a reference electrode, a buffer, an ammeter, and any elements used for an electrochemical device known to those skilled in the art.
This invention also provides a method for detecting hydrogen peroxide, which includes the steps of: contacting a test sample with the electrode such that nitrogen on the functional group of the polymer of the active layer on the electrode is oxidized; applying a constant voltage to the electrode to reduce the oxidized nitrogen of the polymer of the active layer such that an electrical current is generated; and measuring the electrical current.
Specifically, a three-step mechanism by which the electrochemical device detects H2O2 is proposed:
Preferably, the constant voltage is −0.4 V or −0.5 V.
Preferably, the aforementioned electrode can detect H2O2 in the presence of an organic acid. Examples of the organic acids include acrylic acid, acetic acid, formic acid, maleic acid, succinic acid, oxalic acid, citric acid, tartaric acid, lactic acid, and malic acid.
The current invention also provides a method for detecting an analyte, which comprises: contacting a test sample with an electrode in the presence of an oxidase, thereby oxidizing the nitrogen on the functional group of the polymer of the active layer on the electrode; applying a constant voltage to the electrode to reduce the oxidized nitrogen of the polymer of the active layer such that an electrical current is generated; and measuring the electrical current.
Specifically, the proposed mechanism is similar to that of detecting H2O2, except that, the detection of the analytes of interest requires a suitable oxidase to oxidize the analytes thereby forming H2O2. The proposed mechanism is as follows:
Preferably, the analyte is selected from the group consisting of glucose, cysteine, hypoxanthine, lactic acid, sterigmatocystin, glutamate, choline and cholesterol.
The oxidase is chosen according to the analyte of interest. Preferably, the oxidase is selected from the group consisting of glucose oxidase, copper/zinc superoxide dismutase, hypoxanthine oxidase, lactate oxidase, aflatoxin-oxidase, glutamate oxidase, choline oxidase and cholesterol oxidase.
Preferably, the oxidase is fixed on the electrode.
Preferably, the electrode of the present invention can detect the aforementioned analytes in the presence of an organic acid. Examples of the organic acid include acrylic acid, acetic acid, formic acid, maleic acid, succinic acid, oxalic acid, citric acid, tartaric acid, lactic acid, and malic acid.
An electrochemical device used in the present invention includes the electrode of this invention, a counter electrode, a reference electrode, a buffer, and an ammeter. All electrochemical measurements were performed in 40 mL of 0.2 M phosphate buffered saline (PBS) (pH 7.0) at 25° C.
All measurements were conducted by applying a constant voltage of either −0.4 V or −0.5 V, and stabilized for 150 seconds. Thereafter, current was recorded after various concentrations of H2O2 were sequentially added (from low to high) into the electrochemical device.
All measurements were taken when the electrode is responsive to each concentration change with a signal to noise ratio of at least 3. The detection limit and linear range can be extrapolated from plots of current vs. time plot or response current vs. H2O2 concentration.
The response time is time interval between the current changes of two adjacent stages.
The sensitivity (μA/mM·cm2) for H2O2 was also studied and is a ratio of the slope of the curve of the current vs. H2O2 concentration plot to the surface area of the Au electrode.
3.62 g (10 mmol) of PAn was dissolved in 50 mL of DMSO to obtain a solution A; and 0.72 g (10 mmol) of polyacrylic acid was dissolved in 50 ml of DMSO to obtain a solution B. 2 ml of solution A and 8 mL of solution B were evenly mixed to form a mixture. 5 μL of the mixture was dripped on an Au electrode having 0.1 cm2 of surface area and dried for 24 hours at 30° C. in a vacuum oven, followed by washing with deionized water to remove excess polyacrylic acid. The electrode was dried, thus obtaining an electrode for Example 1-1.
The preparation methods for Examples 1-2 and 1-3 were similar to that of Example 1-1, except that the mixing ratios of solutions A and B are different (see Table 1).
1. Observation of Electrode Surface
2. Determination of the Effect of Various Concentrations of Hydrogen Peroxide on Response Current
As shown in Table 1, the response time exhibits an inverse correlation with the amount of polyacrylic acid. The sensitivity also increases with the increase of the amount of polyacrylic acid. This indicates that the polyacrylic acid promotes the oxidation of polyaniline and increases the responsiveness to the reductive current.
3. Interference
−0.5 V was applied to the electrochemical device having the electrode of Example 1-2 and the current was stabilized for 150 seconds. 1 mM of H2O2 solution was added into the electrochemical device. After the stabilization of the current, different interfering molecules were added at intervals of 40 seconds. Specifically, 1.0 mM of ascorbic acid (AA) and 1.0 mM of uric acid (AA), both in PBS, were added sequentially into the electrochemical device. The results are shown in
As shown in
4. Determination of Stability of the Electrode
The stability of the electrode is defined as the percentage change of the electrical output (current) between a given time point during the exposure to an environment of interest to that before the exposure to the environment of interest. The electrochemical device containing the electrode of Example 1-2 was preserved under 30° C. As shown in
1.5 mg (4×10−3 mmol) of polyaniline was dissolved in 1 mL of DMSO to obtain a solution A. 10 mg of carboxylated graphene (Ga-COOH) was dissolved in 1 mL of DMSO to obtain a solution B. 0.05 mL of the solution A and 0.95 mL of the solution B were evenly mixed, followed by addition of 13.5 mL of DMSO to obtain a mixed solution. 5 μL of the mixed solution was evenly coated on a Au electrode (with 0.0415 cm2 surface area) followed by drying in a vacuum oven at 30° C. for 12 hours, thus obtaining an electrode of Example 2-1. The electrode contains a porous active layer composed of polyaniline and Ga-COOH.
The preparation methods for Examples 2-2 to 2-5 were similar to that of Example 1-1, except that the mixing ratios of solutions A and B, and the volume of DMSO were different (see Table 2).
The electrochemical testings in Examples 2-1 to 2-5 were similar to those in Examples 1-1 to 1-5, except that a constant voltage of −0.4 V was applied to the electrochemical device.
As shown in Table 2, the addition of Ga-COOH promotes the oxidation of polyaniline and increases the responsiveness to reductive current.
1.5 mg (4×10−3 mmol) of polyaniline was dissolved in 1 mL of DMSO to obtain a solution A. 10 mg of Ga-COOH was dissolved in 1 mL of DMSO to obtain a solution B. 0.4 mL of solution A and 0.6 mL of solution B were evenly mixed, followed by adding 10.3 mL of deionized water and even mixing using an ultrasonic vibrator to obtain a mixture containing a precipitate of polyaniline-encapsulated Ga-COOH. 10 mg of glucose oxidase (purchased from Sigma-Aldrich) was added into the mixture, and adsorption of glucose oxidase onto the precipitate was allowed for three hours at 4° C. Thereafter, centrifugation was performed to remove excess un-adsorbed glucose oxidase to obtain the precipitate with glucose oxidase. 11.3 mL of deionized water was added to the precipitate with glucose oxidase to form a solution C. 5 μL of solution C was dripped onto an Au electrode having a surface area of 0.0415 cm2. The Au electrode was dried in a vacuum oven at room temperature for 10 minutes, thus obtaining an electrode of Example 2-6. The electrode comprises an active layer having polyaniline, Ga-COOH, and glucose oxidase.
The preparation method for Example 2-7 was similar to that of Example 2-6, except that in Example 2-7, 0.3 mL of solution A and 0.7 mL of solution B were mixed with 9 mL of deionized water. Thereafter, 10 mL of deionized water was used to dissolve the precipitate to obtain the solution C. The parameters for preparing the electrodes of Examples 2-6 and 2-7 are listed in Table 3.
Various concentrations of glucose solution ranging from 1 mM to 5 M were prepared. A constant voltage of −0.4 V was applied to the electrode and stabilized for 150 seconds. Thereafter, 0.1 mL of various concentrations of solution C was added to the electrochemical device at intervals of 30 seconds. The results are shown in
As shown in Table 3, the response time and sensitivity are better in the electrode obtained from Example 2-6.
2 mg of PA1-BT was dissolved in 5 mL of NMP to obtain a solution A. 2 mg of Ga-COOH was dissolved in 5 mL of NMP to obtain a solution B. 0.5 mL of solution A and 0.5 mL of solution B were evenly mixed to obtain a mixture. 5 μL of the mixture was dripped onto an Au electrode with a surface area of 0.07 cm2, followed by drying in a vacuum oven at 50° C. for 24 hours. An electrode with an active layer composed of PA1-BT/Ga-COOH was thus obtained.
The preparation method for an electrode composed of PA1-BT and Ga-COOH was similar to that of Example 3, except that PA1-BT was substituted with PA1-BO.
Various concentrations of H2O2 ranging from 1 mM to 8 M were prepared. A constant voltage of −0.5 V was applied to the electrodes from Examples 3 and 4 and stabilized for 150 seconds. Thereafter, 0.1 mL of various concentrations of H2O2 were added to the electrochemical device at intervals of 30 seconds, and the response current was recorded.
The inset of
As shown in Table 4, the response time and sensitivity are better in the electrode obtained from Example 4.
0.618 g (1 mmol) of PA2-BT was dissolved in 10 ml of DMSO to obtain a solution A. 0.072 g (1 mmol) of polyacrylic acid was dissolved in 10 mL of DMSO to obtain a solution B. 0.1 mL of the solution A and 0.9 mL of the solution B were evenly mixed to obtain a mixture. 5 μL of the mixture was dripped onto an Au electrode having a surface area of 0.057 cm2, followed by drying in a vacuum oven at 30° C. for 24 hours. The electrode was washed with deionized water to remove excess polyacrylic acid and dried, thus obtaining an electrode of Example 5-1.
Preparation methods for Examples 5-2 and 5-3 were similar to that of Example 5-1, except that the mixing ratios of the solutions A and B were different, and are listed in Table 5.
Similar to the testing conducted in Examples 3 and 4, various concentrations of H2O2 were added to the electrochemical device.
As shown in Table 5, the electrode from Example 5-2 has a shorter response time and higher sensitivity.
6 mg of PAA-BO was dissolved in 1 mL of NMP to obtain a solution A. 4 mg of Ga-COOH was dissolved in 1 mL of NMP to obtain a solution B. 40 μl of the solution A, 20 μl of the solution B and 20 μl of NMP were evenly mixed to obtain a mixture. 5 μL of the mixture was dripped onto an Au electrode having a surface area of 0.0616 cm2, followed by drying in a vacuum oven at 40° C. for 8 hours, thus obtaining an electrode of Example 6-1.
Preparation methods for Examples 6-2 to 6-4 were similar to that of Example 6-1, except that the mixing ratios of the solutions A and B were different, and are listed in Table 6.
Various concentrations of H2O2 ranging from 1 mM to 5M were prepared. A constant voltage of −0.5 V was applied to the electrochemical device and stabilized for 150 seconds. Thereafter, 0.1 mL of various concentrations of H2O2 was added thereto at intervals of 25 seconds, and the current was recorded.
As shown in Table 6, the electrode from Example 6-1 has the shortest response time (1.3 seconds). The electrode from Example 6-2 has the highest sensitivity (1037.6 μA/mM·cm2) and broadest detection range (0.0025-12.5 mM).
The preparation method for the electrode of Example 7 was similar to that of Example 6-1, except that, 40 μL of solution A, 80 μL of solution B and 20 μL of NMP were evenly mixed to obtain a mixture.
The electrode from Example 7 was placed in 40 mL of PBS (pH=7) saturated with oxygen, and 170 U of glucose oxidase (purchased from Sigma-Aldrich) was added thereto. An electrochemical device was further setup by connecting the electrode and an ammeter by a wire, and connecting a reference electrode to an ammeter. An electrochemical device for Example 7 was thus obtained.
Various concentrations of glucose solution ranging from 1 mM to 5 M were prepared. The glucose solutions were added to the electrochemical device that was saturated with oxygen, and each addition was allowed to react for 90 seconds. Thereafter, a constant voltage −0.5 V was applied and current was allowed to stabilize for 100 seconds before the recording of the current.
The inset of
To sum up, the electrode of the present invention which comprises an active layer including a polymer having a reactive functional group of formula (A) or (B) and a carboxylated material can be used to detect H2O2 and other analytes at reduction potential (−0.4 V or −0.5 V), and has a shorter response time (1.3 seconds), improved sensitivity (up to 1037.6 μA/mM·cm2), broader detection range (0.0025˜12.5 mM), and a lower detection limit (0.331 mM).
While the present invention has been described in connection with what are considered the most practical and preferred embodiments, it is understood that this invention is not limited to the disclosed embodiments but is intended to cover various arrangements included within the spirit and scope of the broadest interpretations and equivalent arrangements.
Number | Date | Country | Kind |
---|---|---|---|
099125888 | Aug 2010 | TW | national |
099125889 | Aug 2010 | TW | national |
099127939 | Aug 2010 | TW | national |
This application is a continuation-in-part (CIP) of U.S. patent application Ser. No. 13/105,304, filed on May 11, 2011, the entire disclosure of which is incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
Parent | 13105304 | May 2011 | US |
Child | 13482264 | US |