The present invention relates to implantable prosthetic devices for hip-joint replacement in a human body. More particularly, the present disclosure relates to a prosthetic acetabulum cup and liner assembly as well as a prosthetic ball attached to a prosthetic femur, the acetabulum cup and liner assembly controllably constraining the ball in the acetabulum cup and liner assembly.
During recent years the number of people requiring a joint replacement has been increasing. A paper entitled “Developing PEEK Polymer as a Bearing Material for Implants,” John Devine, ©2006 Medical Devices & Diagnostic Industry, reports that each year approximately 1.4 million joint replacement procedures are performed worldwide. Of joint replacement procedures, a paper entitled “Failure Analysis of Composite Femoral Components for Hip Arthroplasty,” Chaodi Li, PhD; et al., Journal of Rehabilitation Research & Development, Vol. 40, No. 2, March/April 2003, pp. 131-146, estimated that at the time of its publication 800,000 total hip replacements were be performing annually worldwide. However, as described below it appears that there still exist obstacles to providing a prosthetic hip replacement that matches the original healthy hip joint. A significant concern in hip replacement surgery is the ease with which the prosthetic hip replacement can dislocate immediately after surgery. The range of motion (“ROM”) provided by a prosthetic hip replacement is another concern especially for younger, more active recipients.
The natural human hip is considered a relatively frictionless ball and socket joint that is enclosed by a soft tissue capsule. A ball-like head of the femur rotates within a socket or acetabulum situated in the pelvis. The soft tissue capsule is comprised of ligaments; the ilio-femoral, ischio-femoral and pubo-femoral ligaments being external to the joint while the ligamentum teres is an internal ligament. A primary function of these ligaments is to retain the femur lightly in the acetabulum, prevent extension of the femur much beyond the straight position, and limit the extent of abduction/adduction and movements of rotation.
The ball-like head of the femur is connected to the thigh bone by a neck which is angularly disposed relative to the femoral axis and relative to the vertical axis of the body. Thus any load applied by the body through the hip and femoral neck to the thigh bone and leg and any impact, such as caused by walking, jumping and the like applied by the leg and thigh bone through the femoral neck and hip to the body, is transmitted angularly through the femoral neck. This angular transmission of the load and forces through the femoral neck results in high stresses and high sheer-loads applied to the femoral neck. These high stresses, when normally applied can cause dislocation of the femoral head from the acetabulum or hip socket, and fracture or breaking of the femoral neck. For ninety-five percent (95%) of normal US adults, dislocating the femoral head from the acetabulum requires an estimated force between one hundred and twenty five (125) to two hundred (200) lbs. In older people the femoral neck often becomes brittle, and in both older and younger people is subject to fracture.
Presently, if the natural hip-joint displays an appropriate anomaly, sufficient damage or diseased state, the natural hip-joint is usually replaced by an implantable prosthetic hip-joint replacement. The prosthetic hip-joint replacement includes a substantially spherically-shaped head that is attached to the femur by a neck and stem which fits into the medullary canal. The prosthetic hip-joint replacement also includes a corresponding artificial socket implanted into the acetabulum, which may be suitably enlarged for the purpose. Conventional prosthetic hip-joint sockets normally embody an acetabulum-type cup and liner assembly having a spherically-shaped cavity which receives and rotatably supports the substantially spherically-shaped head. The acetabulum-type cup is suitably secured in various ways to the acetabulum pocket of the pelvis. In this way, the implantable prosthetic hip-joint replacement establishes a ball and socket-type joint which permits ordinary-type of articulated motion provided by the natural hip-joint.
During a hip-joint reconstructive procedure, the ligamentous capsule around the natural hip-joint is usually resected. When the ligaments of the natural joint are resected during the reconstructive procedure, the artificial joint is inherently less stable and subject to dislocation. Consequently, there exists an increased potential for artificial joint dislocation when a total hip-joint prothesis implantation into a patient causes ligamentous laxity. For example, after undergoing a hip replacement procedure a patient is strongly advised to avoid any pressure while in a crossed leg position. Until the ligamentous capsule around the natural hip-joint heals sufficiently, pressure applied in a crossed leg position can easily dislocate the substantially spherically-shaped head from the prosthetic acetabulum cup liner.
In an attempt to prevent dislocation, some total joint implant devices have been designed to constrain the substantially spherically-shaped head within the artificial acetabulum. However, these constrained devices transfer greater forces to the acetabulum cup due to a lever effect that would normally cause dislocation. Therefore, higher stresses occur at the bone/acetabular component interface that results in loosening of the acetabulum cup. U.S. Pat. No. 4,676,798 exemplifies one type of a constrained implantable prosthetic hip-joint.
To replace the natural socket, some prosthetic acetabulum cup assemblies include a metallic shell for attachment to a suitably enlarged acetabulum. Such prosthetic acetabulum cup assemblies may include a polymer bearing which is inserted into the metallic shell that provides a hemispherical bearing surface for receiving the prosthesis’ substantially spherically-shaped head. Frequently, the polymer bearing component is non-symmetrical and includes a built-up lip around a portion of the hemispherical bearing surface to reduce the likelihood that an implanted spherically-shaped head may become dislocated from the hemispherical bearing surface.
Examples of known implantable prosthetic hip-joint replacements which address some of the aforementioned problems using an acetabulum cup assembly having locking mechanism appear in U.S. Pat. Nos. 5,049,158 and 4,380,090. In particular, the latter patent discloses that a retaining-ring for an acetabulum cup assembly is “preferably made of a resistant metal such as vitallium or stainless steel”.
Retaining-rings made of other materials such as silicone and Ultra-High Molecular Weight Polyethylene (“UHMWPE”) are also known. However, many of these alternative materials exhibit problems. For example, acetabulum cup assemblies which include either a silicone or UHMWPE retaining-ring have exhibited a wide range of push-in and push-out force. Such acetabulum cup assemblies having a retaining-ring have also exhibited problems related to the shell/insert interface being too loose, and also restrict the ROM.
U.S. Pat. No. 4,936,855 discloses an implantable prosthetic hip-joint replacement having an acetabulum cup assembly adapted for receiving a femur ball. The acetabulum cup assembly includes an insert having a stepped entry that provides a cavity which receives the femur ball. The insert's stepped entry receives a split retaining-ring. A ball shaped portion of the prosthesis introduced into the entry displaces the locking ring inwardly into a larger stepped portion of the entry so the locking ring can expand to allow passage of the ball. After the ball passes the split retaining-ring, the ring contracts and slides over the ball to a locking position in a smaller portion of the insert's stepped entry.
U.S. Pat. No. 5,782,930 also discloses an a implantable prosthetic hip-joint replacement having an acetabulum cup assembly adapted for receiving a femur ball. The acetabulum cup assembly includes:
U.S. Pat. No. 6,916,342 discloses an a constrained implantable prosthetic hip-joint replacement that includes:
Accordingly, it is desirable to provide an improved retaining-ring for an acetabulum cup assembly. It is also desirable that an improved retaining-ring for an acetabulum cup assembly be mass-producible particularly to assure consistency in the locking mechanism and to assure the quality of the acetabulum cup assembly's components. Moreover, it is desirable to provide an acetabulum cup assembly having a locking mechanism that exhibits consistent push-in and pull-out forces.
An object of the present disclosure is to provide an improved prosthetic hip-joint having a femoral head which is constrained against dislocation.
Another object of the present disclosure is to provide an improved prosthetic hip-joint having a femoral head which is constrained against dislocation that is able to rotate through a planar angle which exceeds at least one-hundred fifty-three degrees)(153°).
Yet another object of the present disclosure is to provide an improved prosthetic hip-joint having a femoral head which is constrained against dislocation by a preestablished amount of force which is adjustable during implantation of the prosthetic hip-joint.
Yet another object of the present disclosure is to provide an improved prosthetic hip-joint which improves rehabilitation from an implantation procedure.
Yet another object of the present disclosure is to provide an improved prosthetic hip-joint which reduces the time interval required to heal from an implantation procedure.
Yet another object of the present disclosure is to provide an improved prosthetic hip-joint which by permitting a greater ROM reduces the time interval required to heal from an implantation procedure.
Briefly, an improved prosthetic hip-joint in accordance with the present disclosure includes:
These and other features, objects and advantages of the present disclosure will be understood or apparent to those of ordinary skill in the art from the following detailed description of various embodiments illustrated in the drawing figures.
The perspective diagram of
The prosthetic hip-joint 20 further includes a prosthetic femoral assembly 28. The femoral assembly 28 includes a prosthetic, ball-shaped femoral head 32 and a prosthetic femoral stem 34. A first end 36 of the femoral stem 34 is fixed to the femoral head 32 while a second end 38 of the femoral stem 34 distal from the first end 36 is adapted for implantation into a medullary canal of a femur. Except for possible material selections described in greater detail below, the femoral head 32 and the femoral stem 34 are conventional. Securing the liner assembly 26 into the acetabulum cup 22 permits the femoral head 32 to be received into the liner assembly 26 within the acetabulum cup 22 as described below.
The exploded perspective illustration of the prosthetic hip-joint 20 appearing on the right hand side of
As also depicted in
As also illustrated in
The liner assembly 26 also includes a split constraining ring 92, best illustrated in
The liner assembly 26 depicted in
The slots 106 of the constraining ring 92 are respectively adapted to mate with tangs 112 of a pliers-like constraining ring installation tool 114 depicted in
After the constraining ring 92 is properly installed about the inner surface 64 of the rotating liner 42, fabrication the liner assembly 26 is completed by installing the circlip 102 into the groove 98 of the constraining ring 92. When the circlip 102 is initially installed into the groove 98, the splits in the constraining ring 92 and the circlip 102 are aligned. Functionally, the circlip 102 is basically a standard seger ring, also called a snap ring. Accordingly, two apertures 104 pierce the circlip 102 to permit its expansion using a conventional snap ring pliers during its installation into the groove 98.
Implantation of the prosthetic hip-joint 20 begins conventionally with securing the acetabulum cup 22 to the pelvis bone 24 with cancellous bone screws that pass through the screw apertures 68. It is imperative that the screw heads lay entirely within the acetabulum cup 22 thereby avoiding contact the liner assembly 26. Then the liner assembly 26 is initially secured to acetabulum cup 22 by manually threading the stem 54 of the securing bolt 56 into the projecting boss 58.
Completely fixing the liner assembly 26 to the acetabulum cup 22 is preferably performed using a cage assembly 122 depicted in
The cage assembly 122 also includes an electric motor 142 that is secured to the cage 124 for rotating a square drive shaft 144 very slowly with very high torque. A release clutch, not shown in any of the drawings that couples the electric motor 142 to the drive shaft 144, disconnects the electric motor 142 from the drive shaft 144 preferably when the torque on the drive shaft 144 reaches a preestablished value of 400 inch-lbs. Alternatively, the torque at which the clutch disconnects the electric motor 142 from the drive shaft 144 may be adjustable. Before mating the projecting prongs 126 with the notches 132, a pin 146, which is threaded at a first end 148 and which includes a square cavity not illustrated in any of the FIGs. within the first end 148, is fitted onto the drive shaft 144 and screwed into mating threads 152 included in the cage 124. A hexagonally-shaped Allen key 154 projects from a second end of the pin 146 furthest from the first end 148 thereof. As illustrated in
Using the motorized cage assembly 122 for tightening the securing bolt 56 with the prongs 126 of the cage 124 engaging and fully mated with the notches 132 of the acetabulum cup 22 confines all tightening torque to the mated cage assembly 122 and acetabulum cup 22. In this way the acetabulum cup 22 and the cage assembly 122 avoid transmitting any torque to the cancellous bone screws securing the acetabulum cup 22 to the pelvis bone 24 while fixing the liner assembly 26 to the acetabulum cup 22.
After the liner assembly 26 has been fastened to the acetabulum cup 22, the femoral head 32 must be installed into the femoral ball liner 72. The constraining ring 92 includes a slightly tapered inner surface 164 that is substantially coplanar with the groove 96, and is located beyond an equator of the femoral head 32 when the femoral head 32 is mated with the inner surface 84 of the femoral ball liner 72. Arranged in this way, more than one-half of the ball-shaped femoral head 32 lies between the inner surface 164 and the surface 76 of the securing bolt 56 that abuts the outer surface 74 of the femoral ball liner 72. For a femoral head 32 having a diameter of approximately one inch and four-hundred and nineteen thousandths of an inch (1.419 inch diameter), when the constraining ring 92 is properly fitted to the rotating liner 42 the inner surface 164 preferably has a diameter that is approximately seven-thousandths (0.007) of an inch smaller than the diameter of the femoral head 32. During installation of the femoral head 32 into the femoral ball liner 72, the installation tool 114 is used for expanding the constraining ring 92 approximately twenty thousandths (0.020) of an inch to allow the femoral head 32 to pass easily through the inner surface 164 and mate with the inner surface 84 of the femoral ball liner 72. After the femoral head 32 is installed within the femoral ball liner 72, wiggling the constraining ring 92 from side-to-side with respect to the rotating liner 42 with the installation tool 114 may be required to effect disengagement of the prongs 126 of the installation tool 114 from the slots 106 of the constraining ring 92.
When the ball-shaped femoral head 32 is positioned in the femoral ball liner 72 and the constraining ring 92 is fully and properly installed on the rotating liner 42, during normal hip movement with the femoral head 32 fully seated in the femoral ball liner 72 there exists a gap of approximately two and one-half thousandths (0.0025) of an inch between the inner surface 164 and the femoral head 32. If a force were applied to the femoral head 32 tending to dislocate it from the femoral ball liner 72, due to the narrow space between the inner surface 164 and the femoral head 32 initially a piston-like effect resists outward movement of the femoral head 32. The femoral head 32 would have to move outward away from the securing bolt 56 approximately fifty thousandths (0.050) of an inch against this resistance before the femoral head 32 contacts the inner surface 164 of the constraining ring 92. Dislocating the femoral head 32 from the liner assembly 26 requires applying a force to the constraining ring 92 which expands the diameter of the inner surface 164 approximately seven-thousandths (0.007) of an inch. If the femoral head 32 were to contact the inner surface 164, it is readily apparent that the combined precise shape of the ring 92 and the properties of material from which it is made constrain the femoral head 32 to remain within the liner assembly 26 by the inherent resistance of the constraining ring 92 to expanding the diameter of the inner surface 164. Furthermore, the circlip 102 located in the groove 96 of the constraining ring 92 further increases resistance to expanding the diameter of the inner surface 164 by some amount of force regardless of the orientation of the split in the circlip 102 with respect to the split in the constraining ring 92.
As described above, the constraining ring 92 in combination with the circlip 102 establish an amount of force required to dislocate the femoral head 32 from the liner assembly 26. During or even before implantation of the prosthetic hip-joint 20, differing orientations for the split in the constraining ring 92 with respect to the split in the circlip 102 permit. adjusting the preestablished amount of force required for dislocating the femoral head 32.
The cross-sectional views of the prosthetic hip-joint 20 appearing in
Because the prosthetic hip-joint 20 permits the constrained femoral head 32 to rotate eighty-five degrees)(85°) both clockwise and counterclockwise in the same plane with respect to the acetabulum cup 22 from the unrotated orientation depicted in
As is readily apparent to those skilled in the art, everything included in the prosthetic hip-joint 20 either must be made entirely from a biocompatible material, or may have an internally located non-biocompatible material that is entirely encased within an impermeable layer of a biocompatible material.
It is also readily apparent that different size prosthetic hip-joints 20 are required to provide a proper fit to a particular individual. In the following descriptions, various different component dimensions of the prosthetic hip-joint 20 are intended for use with a femoral head 32 having a diameter of approximately one inch and four-hundred and nineteen thousandths of an inch (1.419 inch diameter).
Acetabulum Cup 22
While the acetabulum cup 22 may be made of a biocompatible cobalt-chrome material, for various reasons alternative biocompatible materials may be preferable. For example, the acetabulum cup 22 could be made from titanium such as titanium 6AL4V, or from carbon-carbon material. However, making the acetabulum cup 22 from titanium requires interposing an intermediate liner 172 depicted in
Whichever material is chosen for the acetabulum cup 22, the outer surface juxtaposed with the pelvis bone 24 should have a coating of porous titanium. This porous titanium coating can be applied by metal spraying, plasma spraying or vapor deposition. Whichever material is selected for the acetabulum cup 22 and possibly for the intermediate liner 172, the inner surface thereof juxtaposed with the outer surface 44 of the rotating liner 42 must be highly polished to permit smooth movement of the rotating liner 42.
For a femoral head 32 having a diameter of approximately one inch and four-hundred and nineteen thousandths of an inch (1.419 inch diameter), the outer surface of the acetabulum cup 22 is preferably formed with a radius of approximately one inch and two-hundred and twenty thousandths of an inch (1.220 inches). If due to the particular material selected for the acetabulum cup 22 the prosthetic hip-joint 20 lacks a intermediate liner 172, such a acetabulum cup 22 has a wall thickness of approximately one-hundred and fifty thousandths of an inch (0.150 inch). If the prosthetic hip-joint 20 includes the intermediate liner 172, the acetabulum cup 22 has a wall thickness of approximately one-tenth of an inch (0.100 inch), and the intermediate liner 172 has a wall thickness of fifty thousandths of an inch (0.050 inch).
Rotating Liner 42
The rotating liner 42 may be made from any one of four (4) different materials or combination of materials listed below.
Materials which may be used for the rotating liner 42 include:
Making the rotating liner 42 or other part of the prosthetic hip-joint 20 from a polymeric material such as PEEK reinforced with carbon-carbon, the carbon-carbon reinforcement must be completely embedded within the biocompatible polymeric material. For such reinforcements the carbon-carbon must be woven and processed to yield a very porous mesh before being mated with the polymeric material. Such an extremely strong, porous carbon-carbon reinforcement is then placed in an injection mold and the polymeric material injected concurrently on both sides of the mold thus keeping the carbon-carbon reinforcement centered in the finished product such as the rotating liner 42.
For a femoral head 32 having a diameter of approximately one inch and four-hundred and nineteen thousandths of an inch (1.419 inch diameter), the outer surface of the rotating liner 42 is preferably formed with a radius of approximately one inch and sixty-seven thousandths of an inch (1.067 inch). Such a acetabulum cup 22 has a wall thickness of approximately one-hundred twenty thousandths of an inch (0.120 inch). The aperture 52 has a diameter at the outer surface of the rotating liner 42 of approximately one inch and three-hundred twenty-five thousandths of an inch (1.325 inch) and at the inner surface of approximately one inch and two-hundred twenty thousandths of an inch (1.220 inch).
Securing Bolt 56
The securing bolt 56 must be made of cobalt-chrome or the highest quality stainless steel material. The surface 76 of the securing bolt 56 that is juxtaposed with the outer surface 74 of the femoral ball liner 72 must be accurate and highly polished. The external thread on the stem 54 of the securing bolt 56 is preferably 0.375 inches in diameter, 0.375 inches long and has a pitch of 24 threads per inch (“TPI”). The thread on the stem 54 must be class 3A, which has a tolerance of 0.0000 inch. The mating internal thread within the boss 58 of the acetabulum cup 22 must be class B that has a tolerance of 0.0000 inch. The mating threads of the stem 54 and the boss 58 respectively being class 3A and B and being tightened to a torque of 400 inch-lbs as described above ensures that the securing bolt 56 will not loosen.
For a femoral head 32 having a diameter of approximately one inch and four-hundred and nineteen thousandths of an inch (1.419 inch diameter), the head 62 of the securing bolt 56 is preferably formed with an outer surface having a radius of approximately nine-hundred forty-four thousandths of an inch (0.944 inch). The head 62 preferably has a wall thickness of approximately one-hundred ten thousandths of an inch (0.110 inch), and a radius of curvature for the inner surface of the securing bolt 56 of approximately eight-hundred thirty-five thousandths of an inch (0.835 inch). The head 62 preferably subtends a half-angle of fifty degrees)(50°) measured at its center of curvature.
Femoral Ball Liner 72
The femoral ball liner 72 may be made from:
For a femoral head 32 having a diameter of approximately one inch and four-hundred and nineteen thousandths of an inch (1.419 inch diameter), the outer surface of the femoral ball liner 72 is preferably formed with a radius of approximately eight-hundred thirty-three thousandths of an inch (0.833 inch), and a wall thickness of approximately one-hundred and twenty-two thousandths of an inch (0.122 inch).
Constraining Ring 92
Because the constraining ring 92 is the major component for resisting unintended dislocation of the femoral head 32 from the liner assembly 26, it must be made of a high quality metal and heat-treated to have characteristics equivalent to spring steel. The constraining ring 92 can be cobalt-chrome, spring steel grade stainless steel or titanium which must also be spring steel grade. This part has to be made to the highest standards of accuracy and the surface finish has to he exceptional. The split in the constraining ring 92 must be sufficiently large to ensure that the constraining ring 92 attaches firmly to the rotating liner 42. From a mechanical engineering perspective, firm attachment between the constraining ring 92 and the rotating liner 42 indicates that the split between confronting ends of the constraining ring 92 should not be smaller than fifteen thousandths (0.015) of an inch. However, determining the size of the split in the constraining ring 92 requires considering Wolff's law which concerns the ingress of human tissue towards foreign objects present in the human body.
For a femoral head 32 having a diameter of approximately one inch and four-hundred and nineteen thousandths of an inch (1.419 inch diameter), the constraining ring 92 is preferably formed with a radius for the outer circular surface of approximately one inch and forty-six thousandths of an inch (1.046 inch). The inner surface 164 of the constraining ring 92 is preferably formed with a radius of approximately seven-hundred ten thousandths of an inch (0.710 inch), and subtends an angle of approximately ten degrees)(10°) measured at its center of curvature which lies in the plane of the constraining ring 92 that abuts the open end of the rotating liner 42.
Circlip 102
Similar to the constraining ring 92, the circlip 102 can be cobalt-chrome, spring steel grade stainless steel or titanium which must also be spring steel grade. The finish of the circlip 102 is preferably left rough on the inner circular surface while the outer circular surface must be smooth and polished to ensure that human tissue does not easily migrate towards it and bond. To increase the circlip's resistance to expansion by the constraining ring 92, it may be advantageous to transversely knurl the inner circular surface of the circlip 102 which contacts the groove 96 of the constraining ring 92. From a mechanical engineering perspective, the split in the constraining ring 92 is preferably one-eighth (0.125) of an inch. However, this gap could be smaller, possibly as small as fifteen thousandths (0.015) of an inch. Similar to the constraining ring 92, determining the size of the split in the circlip 102 requires considering Wolff's law which concerns the ingress of human tissue towards foreign objects present in the human body.
For a femoral head 32 having a diameter of approximately one inch and four-hundred and nineteen thousandths of an inch (1.419 inch diameter), the circlip 102 is preferably formed with an outer surface having a radius of approximately one inch and twenty-eight thousandths of an inch (1.028 inch), and an inner surface having a radius of approximately nine-hundred and forty-eight thousandths of an inch (0.948 inch). The circlip 102 is preferably formed with a thickness of approximately seventy-seven thousandths of an inch (0.077 inch).
Femoral Head 32
As stated previously, the femoral head 32 included in the prosthetic hip-joint 20 can be entirely conventional. Consequently, giving due consideration to the service life of the prosthetic hip-joint 20 the femoral head 32 can be made of a ceramic material or of a metal such as cobalt-chrome, stainless steel, vitallium or other metal, perhaps with a ceramic coating applied thereto.
Femoral Stem 34
As stated previously, the femoral stem 34 included in the prosthetic hip-joint 20 can be entirely conventional. However, it appears that making the femoral stem 34 from fiberglass encased in carbon fiber reinforced PEEK material or pure PEEK-Optima may be advantageous. Using high tensile strength “S” fiberglass allows forming the femoral stem 34 with a directional lay up that provides a modulus of elasticity for the femoral stem 34 similar to that of human bone. Using “S” fiberglass for the femoral stem 34 is particularly advantageous because the material resists fatigue. The auto industry has found properly fabricated fiberglass springs to be advantageous because they do not break due to fatigue. A slight spring effect, which can be designed into a femoral stem 34 fabricated using “S” fiberglass, should reduce transmission of shock to the acetabular area while avoiding fatigue failure of the femoral stem 34. Furthermore, by exhibiting mechanical properties that more closely resemble that of human bone the femoral stem 34 would conform with Wolff's law more consistently than presently used metal femoral stems 34. Encasing the “S” fiberglass in PEEK or PEEK-Optima appears to offer a synergy between properties of the two materials.
Although the present invention has been described in terms of the presently preferred embodiment, it is to be understood that such disclosure is purely illustrative and is not to be interpreted as limiting. For example, while the acetabulum cup 22 is preferably secured to the pelvis bone 24 by cancellous bone screws, the acetabulum cup 22 may also be glued thereto or anchored there by spikes extending from its outer surface. Consequently, without departing from the spirit and scope of the disclosure, various alterations, modifications, and/or alternative applications of the disclosure will, no doubt, be suggested to those skilled in the art after having read the preceding disclosure. Accordingly, it is intended that the following claims be interpreted as encompassing all alterations, modifications, or alternative applications as fall within the true spirit and scope of the disclosure.
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/US07/23838 | 11/13/2007 | WO | 00 | 5/8/2009 |
Number | Date | Country | |
---|---|---|---|
60858779 | Nov 2006 | US |