This invention relates to a patch for placing over the eyelids of a patient during surgery. More particularly, there is provided an eye patch which is placed over the patient's eyes during surgery so as to prevent the drying of the eyes by the anesthesia.
General anesthesia reduces the tonic contraction of the orbicularis oculi muscle, causing lagophthalmos i.e. the eyelids do not close fully in 60% of patients.
In addition, general anesthesia reduces tear production and tear-film stability, resulting in corneal epithelial drying and reduced lysosomal protection. The protection afforded by Bell's phenomenon (in which the eyeball turns upwards during sleep, protecting the cornea) is also lost during general anesthesia.
Corneal abrasions are the most common injury; they are caused by direct trauma, exposure keratopathyor chemical injury. An open eye increases the vulnerability of the cornea to direct trauma from objects such as face masks, laryngoscopes, identification badges, stethoscopes, surgical instruments, anesthetic circuits, or drapes. Exposure keratopathy refers to the drying of the cornea with subsequent epithelial breakdown. When the cornea dries out it may stick to the eyelid and cause an abrasion when the eye reopens. Chemical injury can occur if cleaning solutions such as Betadine, chlorhexidine or alcohol are inadvertently spilt into the eye, such as when the face or mouth is being prepped for surgery. Additionally, it is important to protect the eyes during surgery when the patient is placed in the trendelenburg position—tilted so that feet are higher than the head. This position is commonly used for abdominal and gynecological surgery and the eyes must be protected from the patient's own gastric juices flowing into them.
Therefore, the anesthetist must ensure that the eyes are fully closed and remain closed throughout the procedure, in order to avoid exposure keratopathy. Seemingly trivial contact can result in corneal abrasion and the risk of this occurring is markedly increased if exposure keratopathy is already present. Corneal abrasions can be excruciatingly painful in the postoperative period, may hamper postoperative rehabilitation and may require ongoing ophthalmological review and after care. In extreme cases there may be partial or complete visual loss.
Methods to prevent perioperative corneal injuries include simple manual closure of the eyelids, taping the eyelids shut, use of eye ointment, bio-occlusive dressings and suture tarsorrhaphy. However, none of the protective strategies are completely effective; vigilance is always required i.e. the eyes need to be inspected regularly throughout surgery to check they are closed.
For many years, in most western countries, the eyes of patients undergoing general anesthesia have been routinely taped or stuck down with adhesive dressings in an attempt to combat these problems.
Unfortunately, many of the adhesives used on medical products today are temperature and time sensitive i.e. their adhesive strength may increase or decrease when applied to a body temperature and the longer they are applied, the greater the variability in their adhesiveness. What may seem the perfect adhesive strength before application can change as the operation progresses; leading to failure of stick or “over stickiness”. In the former case, the eyelids may move apart and in the latter, may cause bruising, eyelid tears and eyelash removal.
Rolls of tapes are often “laying around” the operating theatre and may not be hygienically clean. Most of these tapes are translucent and so it is not possible to see if the patient's eyes are opened or closed throughout the case. It is not uncommon for the eyelids to move open as the case progresses, even with adhesive tapes stuck onto them.
In a practical sense, these medical tapes/dressings may be difficult to remove from a patient because their ends can become stuck flush with the skin.
While various eyelid closure patches have been developed, the current designs generally provide for a continuous adhesive layer on the eyelid patch or only an adhesive layer about the perimeter of the patch. When the adhesive layer is continuous, it often sticks the wearer's eyelashes and the junction between the upper and lower eyelids, any of which may be tugged when the patch is removed. Conversely, when the adhesive is provided only around the perimeter, the closed eyelid is often not retained and may open, intentionally or inadvertently.
In at least one embodiment, the present invention provides an eyelid closure patch including a flexible transparent backing material having a main body with a width and a length and a tab extending from the main body. An adhesive is applied to one surface of the backing material such that the adhesive substantially covers the main body and the tab. An overlying cover sheet covers the adhesive on the tab to define a non-adhesive tab.
In at least one embodiment, the main body has a configuration with a linear edge and an opposed arcuate edge.
In at least one embodiment, an indicia is applied to the tab.
In at least one embodiment, the main body has a length less than 2.5 inches and a width less than 1.25 inches.
In at least one embodiment, the invention provides a method of manufacturing an eye closure patch including positioning an adhesive tape on a cover sheet; cutting through the adhesive tape and the cover sheet to define a perimeter of at least one of the eye closure patches; cutting through the cover sheet to define a first portion of the cover sheet that is a aligned with a main body portion of the eye closure patch and a second portion of the cover sheet that is aligned with a tab portion of the eye closure patch, the second portion of the cover sheet configured to remain adhered to the tab portion to define a non-adhesive tab.
The accompanying drawings, which are incorporated herein and constitute part of this specification, illustrate the presently preferred embodiments of the invention, and, together with the general description given above and the detailed description given below, serve to explain the features of the invention. In the drawings:
In the drawings, like numerals indicate like elements throughout. Certain terminology is used herein for convenience only and is not to be taken as a limitation on the present invention. The following describes preferred embodiments of the present invention. However, it should be understood, based on this disclosure, that the invention is not limited by the preferred embodiments described herein.
Referring to
In the illustrated embodiment, indicia 16 is provided on the opposite surface of the backing material 12 in the area of the tab 15 to identify the location of the tab 15 to a user. The indicia 16 may be, for example, a printed pattern, word or the like or a coloring of the backing material 12. In the illustrated embodiment, the patches 10 are manufactured such that a pair of patches 10 share a common cover sheet 20, with a cut 17 between the patches 10 (see
The backing material 12 can comprise any clear medical grade backing material such as paper, woven cloth or non-woven polyesters, foams and polymeric films, such as polyurethane, a polyester, a polyolefin (polyethylene, polypropylene), polyvinyl alcohol, polyvinyl acetate, and the like. The cover sheets 20, 22 or films can be prepared from silicone or polymeric films such as polyolefin polymers, vinylidene copolymers, fluorocarbon films, polyethylene terephthalate, acrylic polymers or the like. The films or other substrates which may be employed in the invention may have a thickness of from 0.0005 to 0.05 inch.
The adhesive 14 is supplied to the backing material 12 in sufficient amount to cause adherence. The adhesive 14 is a transparent pressure sensitive water soluble adhesive which can be easily removed and can be used with comfort by the patient. A natural or synthetic hydrocolloid provides sufficient stickiness to cause the patch to adhere sufficiently during the operation. Suitable synthetic and natural water soluble hydrocolloids and gelatins include karaya gum, guara gum, collagen, polysaccharide gum, locust bean gum, powdered pectin, gelatin, carboxymethyl cellulose and the like. As further examples, the adhesive may be a synthetic rubber, acrylate, silicone or soft silicone gel adhesive or any other suitable medical grade adhesive.
If desired, the pressure sensitive adhesive can include 0 to 10% by weight of the usual modifiers, fillers, extenders, antioxidants, stabilizers and other such ingredients known in the art for inclusion in such compositions. Thus, for example, plasticizers or solvents such as mineral oil or petrolatum may be added to improve adhesive characteristics.
The extenders can include finely divided clays, bentonites, starches or other inert ingredients normally used in adhesive compositions. Antioxidants and stabilizers can be used at levels up to about 3% by weight of the total composition. Suitable oxidants and stabilizers include bitylzimate, 2,6-ditert-butyl-4 methyl phenol sold under the trademark IONOL by Shell Chemical Company, alkylated diphenyl phenols, non-allergenic substances such as zinc oxide can also be used as a stabilizer. The adhesive preferably has a Williams plasticity number of about 1 to 4 mm according to the procedure of ASTM 0926-67 (1978).
Referring to
In the embodiment illustrated in
In each of the illustrate embodiments, the main body 13, 13′, 13″, 13′″, 13iv has a length L parallel to the linear edge 21 and a maximum width W perpendicular to the linear edge 21. Preferably, the length L is less than 2.5 inches and the width W is less than 1.25 inches, however, the width W and/or length L could be larger. In one exemplary embodiment, the main body 13, 13′, 13″, 13′″, 13iv has a length L of about 2.125 inches and a width W of about 1.0 inches. With this configuration, when the patch 10 is positioned over the wearer's eye, the adhesive 14 will contact along the wearer's closed eyelid and maintain the eyelid in the closed position.
Referring to
Referring to
Referring to
The process described allows two patches to be manufactured in an efficient, economical manner. While the various cuts 17a, 17b, 18 and 19 and the application of indicia 16 are illustrated in a particular sequence, it is understood that the invention is not limited to such. The cuts and application of indicia may be performed in different sequences and/or steps may be performed simultaneously.
Referring to
The main body 43 has a rectangular configuration with a length L and width W. Preferably, the length L is less than 2.0 inches and the width W is less than 1.25 inches. In one exemplary embodiment, the main body 43 has a length L of about 1.75 inches and a width W of about 1.0 inches. With this configuration, when the patch 40 is positioned over the wearer's eye, the adhesive 44 will contact along the wearer's closed eyelid and maintain the eyelid in the closed position.
The patches 40 are preferably supplied in the form of a strip 48 with each patch 40 being separated at an area with perforations 47 to form a tear strip. The strip 48 is preferably removably positioned on a cover sheet 50 to protect the adhesive 44. The cover sheet 50 preferably has a width greater than that of the strip 48 such that at least one end 52 extends below and beyond the tabs 45. With such a configuration, the end 52 can be easily folded away from the tab 45, as indicated by arrow A in
These and other advantages of the present invention will be apparent to those skilled in the art from the foregoing specification. Accordingly, it will be recognized by those skilled in the art that changes or modifications may be made to the above-described embodiments without departing from the broad inventive concepts of the invention. It should therefore be understood that this invention is not limited to the particular embodiments described herein, but is intended to include all changes and modifications that are within the scope and spirit of the invention as defined in the claims.
This application is a divisional application of U.S. patent application Ser. No. 15/298,281 filed Oct. 20, 2016, which claims the benefit of U.S. Provisional Application No. 62/244,424, filed Oct. 21, 2015, the contents of both applications are incorporated herein by reference.
Number | Date | Country | |
---|---|---|---|
62244424 | Oct 2015 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 15298281 | Oct 2016 | US |
Child | 16246699 | US |