The present invention relates to field of Intraocular Lenses (IOLs). In particular, the present invention relates to Intraocular Lenses wherein an electro-active element provides at least a portion of the IOL's refractive power, or prismatic power, or at least a portion of the tinting.
Intraocular lenses (IOLs) are typically permanent, plastic lenses that are surgically implanted inside of the eyeball to replace or supplement the eye's natural crystalline lens. They have been used in the United States since the late 1960s to restore vision to cataract patients, and more recently are being used in several types of refractive eye surgery.
The natural crystalline lens is critical component of the complex optical system of the eye. The crystalline lens provides about 17 diopters of the total 60 diopters of the refractive power of a healthy eye. Further, a healthy crystalline lens provides adjustable focusing when deformed by the muscular ciliary body that circumferentially surrounds the crystalline lens. As the eye ages, the flexibility of the crystalline lens decreases and this adjustable focusing is diminished. Thus, this critical crystalline lens almost invariably loses flexibility with age, and often loses transparency with age due to cataracts or other diseases.
Most intraocular lenses used in cataract surgery may be folded and inserted through the same tiny opening that was used to remove the natural crystalline lens. Once in the eye, the lens may unfold to its full size. The opening in the eye is so small that it heals itself quickly without stitches. The intraocular lenses may be made of inert materials that do not trigger rejection responses by the body.
In most cases, IOLs are permanent. They rarely need replacement, except in the instances where the measurements of the eye prior to surgery have not accurately determined the required focusing power of the IOL. Also, the surgery itself may change the optical characteristics of the eye. In most cases, the intraocular lenses implanted during cataract surgery are monofocal lenses, and the optical power of the IOL is selected such that the power of the eye is set for distance vision. Therefore, in most cases the patient will still require reading glasses after surgery. Intraocular lens implants may be static multifocal lenses, which attempt to function more like the eye's natural lens by providing clear vision at a distance and reasonable focus for a range of near distances, for patients with presbyopia. Not all patients are good candidates for the multifocal lens; however, those who can use the lens are some what pleased with the results.
More recently, accommodative IOLs have been introduced. These accommodative IOLs actually change focus by movement (physically deforming and/or translating within the orbit of the eye) as the muscular ciliary body reacts to an accommodative stimulus from the brain, similar to the way the natural crystalline lens focuses. While these offer promise, accommodative IOLs still have not been perfected. In spite of these limited successes, the multi-focal IOL and present accommodative IOLs still have a substantial decrease in performance when compared to a healthy natural crystalline lens.
Another ocular lens that holds promise for correcting presbyopia is the Small Diameter Corneal Inlay (SDCI). The Small Diameter Corneal Inlay (SDCI) is a prescription lens that is inserted into the corneal tissue to create an effect similar to a bifocal contact lens. Corneal Inlays (SDCI) are early in their development and it is still too early to understand how well they will function and also how effective they will become.
While all these emerging surgical procedures have their merits, they all have a substantial decrease in performance when compared to a young healthy natural crystalline lens. The present invention addresses these shortcomings by providing an intraocular lens that behaves in a manner similar to the natural crystalline lens.
Over the past decade, the miniaturization of semiconductor chips, sophisticated earphones, non-volatile solid-state memory, and wireless communication (including blue tooth, and other short-range wireless technologies) have ushered in a revolution in personal electronic components and audio listening devices that allows wearers to listen to music in a portable, hands-free manner. In addition, recent research and development has resulted in the development of accessories and features for eyeglasses such as, by way of example only: electro-active spectacle lenses which provide the wearer with variable focus capability, electro-active spectacle lenses that allow for a varying index matrix needed to correct higher order aberrations to create a supervision effect, electronic heads up displays that are associated with eye glasses, electrochromic lenses that change color and tint by way of electrical activation, and also the addition of audio and communication systems that are associated with eyeglasses. These new electronic eyeglass applications have created a significant need for a convenient, comfortable and aesthetically pleasing way to provide power to the eyeglass frame and lenses. More and more, the eyeglass frame is becoming a platform for associating and housing various electronic accessories.
Currently, there is no known way to electrify the eyeglass frame in a manner that provides a combination of pleasing aesthetics, comfort, convenience, and also allows for the proper ergonomics. While comfort, convenience and ergonomics are important, the proper fashion look of the eyeglass frame is what takes priority when the consumer makes a purchase decision. If the eyeglass frame is thicker or more bulky looking than normal, then the purchase decision can be impacted in a negative manner. In addition, if the eyeglass frame is heavier than normal, red inflamed sore spots will occur on either side of the bridge of one's nose or the top of the ears. In the case of active work or sports, such as, by way of example only, construction work, running, biking, walking, rowing, and horseback riding, the heavier eyeglass frames are, the more prone they are to slide down ones nose, and thus the alignment of the lens optics will not be optimal.
An illustrative aspect of the invention provides an intraocular lens system comprising an electro-active lens comprising multiple independently controllable zones or pixels, and a controller capable of being remotely programmed.
The present subject matter provides an inventive solution, which addresses and corrects this pressing need. The invention does this in a manner that is allows for the eyeglass frames to continue to appear like conventional fashionable eye glass frames whether they be dress glasses, sport glasses or goggles, security glasses or goggles, sunglasses or goggles. It also takes the added weight of the power source off of the eyeglass frame and places this weight were it is barely noticed if at all. Finally, it provides for doing this in a most ergonomic and convenient manner.
According to first aspects of the invention, eyewear comprising an electronic docking station may be provided, whereby the docking station provides power to a docked electrical component.
According to further aspects of the invention, eyewear comprising a camera may be provided, whereby the camera is controlled by a remote controller.
According to further aspects of the invention, eyewear comprising a heads up display may be provided, wherein the heads up display is housed in a visor affixed to the eyewear.
Other aspects of the invention will become apparent from the following descriptions taken in conjunction with the following drawings, although variations and modifications may be effected without departing from the spirit and scope of the novel concepts of the disclosure.
The present invention can be more fully understood by reading the following detailed description together with the accompanying drawings, in which like reference indicators are used to designate like elements.
Hereinafter, various embodiments of the invention will be described. As used herein, any term in the singular may be interpreted in the plural, and alternately, any term in the plural may be interpreted to be in the singular.
Electro-active materials comprise optical properties that may be varied by electrical control. For example, transmission of light may be controlled to produce tinting or a sunglass effect. Further, the index of refraction may be electrically controlled to produce focusing and or prismatic effects. One class of electro-active material is liquid crystals. Liquid crystals comprise a state of aggregation that is intermediate between the crystalline solid and the amorphous liquid. The properties of liquid crystals may be controlled electrically, thermally, or chemically. Many liquid crystals are composed of rod-like molecules, and classified broadly as: nematic, cholesteric, and smectic.
There are several characteristics of electro-active materials which are useful in IOLs. First, the optical characteristics may be generated by thin layers (rather than by the curvature of conventional lenses which may require thick lenses). These thin layers may be placed in locations where it may be difficult to place conventional lenses, for example in the anterior chamber of the eye (between the iris and the crystalline lens). In addition, it is possible to stack (place in series optically) the electro-active layers in such a manner as to get an additive effect for the overall optical power created, including prism, conventional refractive error, or higher order aberration correction, in a thin structure that may be placed in either the anterior or the posterior chamber of the eye.
Second, the optical characteristics may be actively controlled. For example, an electro-active lens may designed to become darker (more tinted, and transmit less light) under bright light conditions. This tinting may be generated automatically by measuring the brightness using, for example, a photodiode or solar cell. Alternately, the tinting may be controlled by the decisions of the user by way of a remote control.
Similarly, the focus of an electro-active lens may be controlled electrically. The focus may be controlled automatically using, for example, a range finder, or a tilt meter, or triangulation based on the direction of both eyes, the forces exerted on the lens by the muscles of the eye. Alternately, the focus may be controlled by the decisions of the user by way of a remote control.
Third, electrical control creates the potential for correcting complex and high order visual defects. Conventional intraocular lenses are limited to addressing certain visual defects for various manufacturing reasons. However, an electro-active lens with a large number of individually addressable controlled small elements (for example, an array of very small pixels) may address very complex and high order visual defects. Further, the control may be simplified by creating individually addressable elements in arbitrary configurations, such as a series of concentric circles, or a series of approximately concentric ellipsis, or whatever customized configuration efficiently corrects the visual defect. The design, manufacture, and control of an array of small pixels has similarities with the manufacture of Liquid Crystal Displays (LCDs). Correction of complex visual defects such as higher order aberrations of the eye creates the possibility of “superhuman” visual acuity, wherein the vision is not limited by the lenses (either biological or corrective), but rather is limited by the inherent anatomy and physics of the photoreceptor cells in the retina. 20/10 vision or better is possible even before additional magnification is considered. Further, it is possible for an electro-active lens to act as a telescope or as a microscope.
Fourth, electrical control creates the potential for changing the optical characteristics of the electro-active IOL as desired. For example, the desired optical characteristics may be determined after the IOL is surgically implanted in order to compensate for any changes that occur during surgery, or for that matter an error in calculating or estimating the post surgery refractive error. Similarly, the optical characteristics of the IOL may be varied over time to compensate for changes in the user's eye. For example, if the user has a degenerative disease that affects a portion of the retina, then it is possible to remotely cause the implanted electro-active IOL to create prismatic power or even change its prismatic power in order to shift the image to a portion of the retina that is undamaged. By way of example only, each month (or as needed) the image may be shifted to the remaining undamaged portion of the retina with the highest concentration of receptor cells. This change can be accomplished post-surgically and remotely (meaning without additional surgery).
Fifth, electrical control creates the potential for the user to automatically or instinctively control the focus. For example, contractions of the muscular ciliary body can be measured by an piezoelectric element (as a strain gauge), and these contractions can then be used as a control input to electrically adjust the focus of the IOL, similar to the way the ciliary body would focus the natural crystalline lens by physical deformation. Additionally, in theory, the focus could be controlled by electrical signals directly from the brain. Recent development with artificial limbs use this technique.
Sixth, electrical control creates the potential to shift the field of view, and thus compensate for diseases that prevent the eyeball from moving. Nervous signals to diseased muscles (that can no longer move the eye) may be intercepted, translated, and used to electrically shift the field of view.
Seventh, there are many types of electro-active element configurations. These configurations include: pixelated (typically a two dimensional array of pixels similar to a liquid crystal monitor on a computer), rotationally symmetric pixelated (for example, a set of concentric circles), and diffractive. Electro-active individually addressable pixelated diffractive lenses may use concentric ring shaped electrodes to product the diffractive lens power with varying index of refraction without physically machining, molding or etching diffractive elements into the surface of the lens.
The electro-active element may be used in combination with a conventional lens, wherein the conventional lens may provide a base refractive power. The electro-active element may be used in combination with a diffractive lens having a machined, molded, or etched surface or geometry. The electro-active element may be used in combination with a second electro-active element, wherein each may perform a different function. For example, the first electro-active element may provide focus, and the second may provide tinting or may serve as an electrically controlled aperture, or the second could cause a prismatic shift of the image to the healthy area of a retina of a deceased eye.
Eighth, as discussed above, it is possible to electrically replace many of the optical functions of a natural eye: tinting may replace or augment the light reducing effect of the contraction of the iris, focusing may replace the natural deformation of the crystalline lens, focusing and prismatic shifting may replace movement of the eyeball, and so forth. Among other factors, the present invention addresses: positioning the IOL, energy storage, energy recharging, power generation, control, steering of the line of site to a targeted region of the retina altering the refractive power of the eye, augmenting or replacing the accommodative power of the crystalline lens, remote tuning post surgery of the electro-active IOL. Tuning comprises altering the power of the IOL and/or altering the location of the focus on the retina of the IOL.
As background, the optical components of the eye will be described in detail. Light entering the eye first enters the cornea 124. The cornea 124 is transparent and provides about 40 diopters of the approximately 60 diopters total refractive power of the eye. Light then passes through the pupil 118. The pupil 118 is an aperture, and is variable in diameter from 1 mm to at least 8 mm. This gives an aperture range in excess of f20-f2.5, and a ratio of 32:1 for the amount of light permitted to enter the eye. The iris 114 serves as an adjustable diaphragm creating a pupil 118. The light then passes through the crystalline lens 122. The crystalline lens 122 is a transparent, encapsulated, biconvex body which is attached circumferentially to the ciliary body 112. The crystalline lens 122 contributes about 17 diopters to the total refractive power of a relaxed eye. The refractive power of the crystalline lens 122 may be altered by contractions of the ciliary muscles in the ciliary body 112, which deform the crystalline lens 122 and alter its refractive power. The light then passes through the vitreous humor 138 and finally contacts the retina 136. The retina 136 is the sensory neural layer of the eyeball and may be considered as an outgrowth of the brain, and is connected to the brain through the optic nerve 134. Near the center of the retina 136, the macula lutea 132 contains a central region of highest visual sensitivity called the fovea centralis or foveola (see
Thus, the human eye has an adjustable diaphragm (iris 114) and an adjustable refractive power (due to the ciliary body 112 deforming the crystalline lens 124).
An IOL can be placed in one of three locations: in the anterior chamber 120, which is between the cornea 124 and the iris 114; or in the posterior chamber (not shown) which is between the iris 114 and the crystalline lens 122; or as a replacement for the crystalline lens 122.
Generally, if the crystalline lens is diseased or damaged, then an IOL may be used to replace the crystalline lens. This IOL replacement for the crystalline lens may be accommodative, or non-accommodative. Replacing the crystalline lens allows the IOL to be conveniently positioned inside of a clear bag-like capsule that previously held the natural crystalline lens, and also allows the possibility of retaining some variable focus capability through interaction with the muscular ciliary body which circumferentially surrounds the clear bag-like capsule. In other cases, the IOL is placed extra capsulary (without the bag-like capsule).
However, if the crystalline lens is still functional, then it may be preferable to leave the crystalline lens undisturbed and to place the electro-active IOL into either the posterior chamber or the anterior chamber 120 of the eye, or into the corneal tissue similar to the Small Diameter Corneal Inlay (SDCI) discussed above. In these embodiments, the electro-active IOL could, by way of example only, provide optical power to correct for conventional refractive errors, correct for non-conventional refractive errors, create a prismatic image shifting effect that moves the location of focus to a healthier area of the retina, and add a tint, as opposed to replacing the optical power of the otherwise healthy crystalline lens.
Conventional refractive error is defined as one or more of: myopia, hyperopia, presbyopia, and regular astigmatism. Non-conventional (or higher order) refractive errors are defined as all other refractive errors or aberrations which are not conventional refractive error.
In many cases, the electro-active IOL may be used during cataract surgery when the existing crystalline lens is defective. In this case, the electro-active IOL will actually replace the removed defective existing crystalline lens, and may provide a range of electro-active optical correction including conventional and/or non-conventional refractive errors, as well as provide refractive power to make up for the lost optical power resulting from the removal of the crystalline lens. In addition, the electro-active IOL can provide for the ability to accommodate without any movement, translation or change in its surface geometry. This is accomplished by localized programmed changes in the index of refraction of the electro-active IOL.
The most common and advanced cataract surgery technique is phacoemulsification or “phaco.” The surgeon first makes a small incision at the edge of the cornea and then creates an opening in the membrane that surrounds the cataract-damaged lens. This thin membrane is called the capsule. Next, a small ultrasonic probe is inserted through the opening in the cornea and capsule. The probe's vibrating tip breaks up or “emulsifies” the cloudy lens into tiny fragments that are suctioned out of the capsule by an attachment on the probe tip. After the lens is completely removed, the probe is withdrawn leaving only the clear (now empty) bag-like capsule, which may act as support for the intraocular lens (IOL).
Phacoemulsification allows cataract surgery to be performed through a very small incision in the cornea. Stitches are seldom needed to close this tiny entry, which means that there is less discomfort and quicker recovery of vision than with other surgical techniques. Small incisions generally do not change the curvature of the cornea (unlike larger incisions that were required with older surgical techniques). Small incisions for more rapid rehabilitation of vision and possibly less dependence on glasses for good distance vision.
After removal of the cataract-damaged lens, an artificial intraocular lens (IOL) may be implanted. The IOL may be produced from soft acrylic or solid medical-grade silicone. IOLs may be folded so they can be implanted with a small injector, which uses the same incision through which the phaco probe was inserted at the beginning of the procedure. As the IOL is implanted, it may be allowed to unfold and anchor itself behind the eye's pupil over the remaining clear capsule. The IOL(s) to be implanted may be selected based on power calculations made before surgery. In the case of the present invention, the electro-active IOL may also be selected based on the range of electro-active correction required, the type of any other ocular disease being treated, and any special needs of the patient.
In most cases, the electro-active element would contribute typically +2.5 Diopters, +2.75 Diopters, +3.0 Diopters, or +3.25 Diopters of optical power. The base lens portion (which the electro-active element is in optical communication) which would contribute most, if not all, of the approximately 17 Diopters normally provided by the crystalline lens, would be measured and selected prior to surgery. However, unlike a conventional IOL, an electro-active IOL allows for remote tuning of its optical power (for example, in case the calculations made prior to surgery are not optimum after surgery).
The ciliary body expands and contracts as the eye attempts to focus from near to far and from far to near. The ciliary body movement may produce tension and/or compression of the piezoelectric film 212 which produces electricity. The electricity may be transferred through charging leads 220 and used to charge the charge storage capacitor 216 (or a rechargeable battery). The charge storage capacitor 216 may power the electro-active lens element 218 and any related control circuitry (not shown). Typically the electro-active lens element 218 requires approximately 1.0 to 5.0 volts, with a preferred range of 1.5 to 2.5 volts. These relatively low voltages decrease the risk involved with surgical placement of electrical devices.
The electrical characteristics of the piezoelectric film 212 under tension or compression may be used as a gauge to determine the desired viewing distance, and may be used to focus the electro-active lens. Thus, it is possible for the user to instinctively and automatically control the focus of the electro-active IOL 200 using the muscular ciliary body. The contractions of the muscular ciliary body previously focused the subject's crystalline lens by physically deforming it. Using the electro-active IOL 200 the instinctive and automatic contractions of the muscular ciliary body will change the electrical characteristics of the piezoelectric film 212, and these electrical changes may be monitored by a processor disposed, for example, on a chip (not shown) and used to electrically, variably focus the electro-active IOL 200. Alternatively, the piezoelectric film 212 may be used solely as a gauge for focusing, in which case, the electro-active IOL 200 would be provided with a different source of power.
In some embodiments, the piezoelectric film may be attached circumferentially to the ciliary body by multiple attachment tabs (more than two) in order to take advantage of the natural circumferential contraction and expansion of the surrounding ciliary body.
One or more lens anchors 214 may be used to stabilize the electro-active lens in the desired location. For example, a lens anchor 214 may be used to center the electro-active lens inside of the capsule or “bag” or membrane which formerly contained the natural crystalline lens (creating an intracapsular IOL). Alternately, the lens anchor 214 may be attached to the ciliary muscle directly, and thus be outside of the capsule (creating an extracapsular IOL).
Multiple lens anchors 214 may be used. For example, 3 or 4 lens anchors 214 may be used. The lens anchors 214 may have different shapes, customized to the specific application.
An optional base lens 252 may provide a base refractive power using a conventional lens configuration, and may be equivalent in refractive power to the crystalline lens when no accommodation is needed. The base lens 252 may also serve as a means of encapsulating the electro-active element in a hermetically sealed enclosure that consists of a biocompatible material similar to those materials currently used to make IOLs, by way of example only, soft acrylic or solid medical-grade silicone.
Other sources of power may include: solar cells, inductive charging, conductive charging, laser, thermoelectric, and harnessing the mechanical energy from blinking. The capacitor 216 (or optionally, a battery) may be recharged inductively with a pair of special glasses (spectacles) that may also remotely turn off the electro-active lens while the battery is being recharged. The special glasses may also be configured to provide vision correction while the battery is recharging.
In some embodiments, the capacitor 216 in the electro-active IOL 200 may be charged with a special pillow that has very light gauge wires through which current runs. The pillow may thus be used to charge the batteries inside the electro-active IOL 200 at night while the patient sleeps. An exemplary arrangement of this type is illustrated in
Alternately, the electro-active IOL may not have a capacitor 216 or battery, but may be constantly powered conductively by an externally located battery, or may be constantly powered inductively by an externally located inductively coupled power supply, or solar cell, or solar cell coupled to a properly tuned laser, or a thermal-electric power supply that generates electricity by dumping body heat (typically 98 degrees F.) into the relatively cool ambient air (typically 70 degrees F.).
The electro-active lens element 618 may be connected by power connections 622 to a rechargeable battery ring 624 or a capacitor (not shown), and may be charged by induction coils or by piezoelectric elements as in previously described embodiments.
In some embodiments, the correction provided by the electro-active IOL may vary depending upon the needs of the patient and the desired results. In some embodiments the electro-active element may only provide correction for presbyopia. In some embodiments, the electo-active IOL may provide remote fine tuned conventional correction. In some embodiments, the electo-active IOL may provide higher order (non-conventional) aberration corrections, by way of example only, coma, spherical aberration, trefoil, and other higher order aberrations. In some embodiments the electro-active element may also adjust the position of the image on the retina, by way of creating a prismatic shift of the image electronically. When correcting for higher orders aberrations and or correcting a prismatic shift of where the image is located on the retina, the electro-active IOL may utilize a plurality of pixels. A prismatic shift of the image is very useful in patients having conditions, by way of example only, macula degeneration of the retina (which may include alterations in color due to disease or specific degeneration of the macula lutea), macula holes, retinal tears, and neurological abnormalities that cause scotomas or a loss of vision in particular segments of the visual pathway (such as blind or dark spots in the field of vision, and blurred vision). It should be pointed out that in each of the use embodiments above the inventive electro-active IOL can be tuned remotely post surgery to effect the optimized effect desired.
While much consideration has been given to powering an electro-active lens, some electro-active materials retain their optical power in the absence of applied electricity (such as by way of example only, a bi-stable liquid crystal). Using these type of electro-active materials, the prismatic power, an additive or subtractive power that is additive or subtractive to the base optical power of the IOL, and/or the higher order corrections could be set while the device is being powered, and then would remain set after the power is removed. This may negate the need for recharging the power source in the IOL. If the patient's vision changes and requires new correction, he could return to the eye-care professional and have the IOL adjusted to a new combination of prismatic and/or higher order correction. The changes could be externally powered remotely. For example, the external power may be RF energy similar to the way RFID tags work today, where the reading device provides the power to the RFID tag inductively so that the RFID can transmit it's information to the RFID reader.
In same manner as the RFID tags, a tuning instrument for changing the IOL power could provide power to the controller on the electro-active IOL, so that the controller could change the voltages on the electrodes of the IOL thus setting the localized index of refraction that determines the optical properties of the electro-active IOL.
Alternately, the power may also be supplied optically by shining a bright light or eye-safe laser into the eye and onto a photocell built into the electro-active IOL that would then provide the temporary electrical power needed to adjust the optical power of the electro-active IOL. This system may also be used for communication, in addition to supplying power.
Bi-stable twisted nematic, cholesteric and ferroelectric liquid crystals have been used in flexible low cost LCD displays, and similar materials may be used in the electro-active elements of an IOL. This type of electrically adjusted (but otherwise non-powered) prismatic adjustment, additive or subtractive, for retinal disease tuning or higher order aberration correction may be added to (i.e., placed in optical series with) any accommodative non electro-active IOL that corrects for presbyopia. For example, electro-active elements could be placed in optical series with non-electrical or non-powered IOLs, such as non electro-active IOLs that mechanically change their optical power by changing one or more surface curvatures and/or the position of the IOL in the eye.
The addition of the electro-active lens or electro-active elements may be accomplished in at least three ways: first, a separate electro-active IOL may be placed in non-touching optical communication (optical series) with the non-electro-active accommodating IOL; second, an electro-active element can be built into one of the IOL's surfaces that does not change contour during accommodation; and third, an electro-active element may be placed inside of a layered non-electro-active.
For example, an electro-active element could be added in the anterior chamber and used in optical series with an individual's functioning crystalline lens. In this case, the crystalline lens will provide natural accommodation, and the electro-active IOL may steer the image to a healthier part of the retina, or may tune the non-electroactive IOL, or may correct for higher order aberration.
As noted above, in some embodiments, it may be a major advantage to tune or adjust the electro-active IOL remotely. After inserting the electro-active IOL in the eye, the optical power and the prismatic power can be fine-tuned remotely to accomplish the optimal vision correction to correct for conventional refractive error, or higher order aberrations, or the precise location of the image on the retina. Further, the IOL could be tuned again at a later date to compensate for changes in the eye over time, due to disease or aging. In cases of correcting solely for conventional refractive error, the electro-active IOL could either utilize diffraction or pixelation or both. The electro-active element may also perform any number of these functions in combination, as required by the patient's conditions and at the discretion of the eye care professional.
Shown in
The invention improves upon the conventional eye glass chord, chain or Croakie by modifying it to allow for not only being uses as a means of securing the eye glass frames to ones head, but in addition to provide for a means away from the eye glass frame to house or support the power source, and of course electrical connections. The invention further provides for off loading certain electrical accessories and features from the eyeglass frame, as well as the electrical connections to be detachable and re-attachable to the eyeglass frame in a very convenient and user-friendly manner. In one application of the invention, electrical connections are provided within the temple pieces of the glasses that allow the electrical signal (digital or analog) to travel to the lens by way of electrical conductors located internally in the frame. In another inventive embodiment, the electrical connectors are located on the outer surface of the temple and applied, by way of example only, with an adhesive film. In this case, the connectors are built into the film and then the film is affixed to the temple or temples. In still other cases, the connectors are applied directly to the frame and then covered by the adhesive film, which then connects to the lens.
The invention shown in the figure provides an electronic enabling tether that contains a power source such that it can be securely hung from the rear of the frame temples and be allowed to extend down to the wearer's upper back, just below the neck. The power source, in some embodiments, can be further secured to the wearer's back by: locating it under the shirt, using, by way of example only, an adhesive patch, Velcro applicator, snap, or clamp to adhere the unit to the wearer's back or shirt. Securing in this way prevents the unit from flopping around while the wearer is walking, jogging or engaged in some other athletic exercise or active work. When the invention is affixed to either one's body or shirt it should have enough length to allow the wearer to bend their head down at the neck without unduly tightening or pulling tautly on the audio unit. In most cases the power source is small and lightweight enough to be confined solely within the inventive tether. Therefore, it is not necessary to affix the enabling tether to one's body or shirt, etc.
In certain embodiments, elastic or rubber fittings are used to secure the inventive electronic enabling tether to the temple or temples. These embodiments may allow for a notch or grove to be placed or built into the temple. In certain other embodiments, the end of the temple or temples provides for a circular fastener, which may or may not be conductive, to which the invention is secured using, by way of example only, a clip.
The inventive electronic enabling tether is connected mechanically and electrically to the frames in a removable fashion. The inventive electronic enabling tether in certain embodiments utilizes a magnet connecting means. In other embodiments, no magnet is used. One such embodiment where a magnetic connector is used allows for the tether to be separated at some point near the mid-line of the tether for easy removal. In other embodiments, the tether is magnetically connected to the temple by way of a magnet attraction/receiving member that is built into the temple connection device, such as by way of example only, an elastic, plastic, or metal fastener that connects the tether to the temple or eyewear frame. In certain cases where power is being supplied to the eyewear, the magnetic connection device also serves as an electrical conductor to provide the electrical connection from the inventive tether to the eyewear (lenses and/or frame). The power source contained within the electronic tether can be either rechargeable or non-rechargeable, in which case it will need to be readily accessible or removable within the tether to be changed from time to time.
The spectacle lenses can be constructed to contain a micro-optical display that is visible to the wearer, located in a fixed space in such a manner as to not obstruct the central vision area of the leases. In this version, an audio unit is replaced or enhanced by additional electronic capability to supply video or informational data. For example, if the unit contained a cell phone or PDA, emails can be transmitted to the micro-optical display or telephone calls can be transmitted to earphones. In this second function, a microphone would have to be added into the spectacle frames near the nose bridge to allow for two-way communication. The inventive electronic enabling tether provides the needed power and the potential offloading capability from the eyewear of items that need to be electronically connected but do not need to reside on the eyeglass frame or lenses.
Thus, the invention contained herein solves a pressing and growing need of enabling electronic frames in a manner that allows for the proliferation of various electronic applications that are now being applied to eye wear. It does this while preserving the fashion aesthetics, comfort and ergonomics of the electronic eyeglasses as compared to the current popular conventional non-electronic eyeglasses.
When reading about the inventive embodiments disclosed herein, it should be pointed out that the words “stem or temple” have the same meaning in what is disclosed herein as do the words clip-on and snap-on. A clip-on can be either monocular (attaching to one eyewire or one half of the frame front) or binocular (attaching to both eyewires or the complete frame front). Further, the electronic tether can be affixed to hinged temples, hinge-less temples, the frame front, or for that matter anywhere on the eyewear. The term eyewear is meant to be interpreted broadly, and may include one or more of a frame, lens, tether, and/or clip-on. The tether is considered an electronic tether when an electrical connection is affixed to it or travels within it. A temple is considered to be an electronic temple if an electrical connection is affixed to it or travels through it. A frame is considered to be an electronic frame if an electrical connection is affixed to it or travels through it. A lens is referred to as an electronic lens when electricity affects the lens' optical power or tint. A lens can be that of a fixed/static lens or a dynamic focusing electronic lens. The word tether includes that of a Croakie, chord, chain, and connecting attachment from one temple to another. Clip-ons can be that of electronic when an electrical connection is associated with the clip-on or non-electronic when no electrical connection is associated with the clip-on. Tints can be that of an electro-chromic tint, a photochromic tint, or a fixed imbedded tint.
In
By decoupling the fixed lens from the electro-active lens, an electro-active focusing lens product could be offered with far fewer SKUs. In fact, the invention anticipates having a limited line of electro-active focusing electronic clip-ons that have preset decentrations. By way of example only, the electronic clip-ons could be available with near vision inter-pupillary measurements of 63 mm, 60 mm and 57 mm, as shown in
While the above discussion was directed to electro-active focusing near and intermediate lenses, it should be pointed out that the invention contemplates electro-active lenses that are full or partial pixilated lens(es), full or partial diffractive lens(es) or a combination of both. In addition, the invention contemplates the electronic clip-ons or electronic snap-ons that house an electroactive lens or lenses that corrects for only higher order aberrations. The electronic clip-on or electronic-snap on would be used to allow the patient to see better than 20/20, perhaps better than 20/10 by correcting his or her higher order aberrations. In this case, the inter-pupillary measurement would be set for one's distance vision needs. This proper inter-pupillary measurement would be properly established by way of the location of the higher order aberration correction location within each clip-on lens. It should be pointed out that in this inventive embodiment the clip-on correcting the patient's higher order aberration(s) can be that of either a fixed static non-electronic lens or that of an electronic pixelated lens.
The power source and/or controller 2450 is attached to the electrifiable frame temple 2410 in any of the manners described herein. The electronic snap-on or electronic clip-on device 2460 containing the electro-active elements 2470 is slightly over-sized to that of frame 2420 so that the side of the conventional lens is covered from view by a person looking at the side of the frame.
Connection points 2483, 2484 that are either mechanical of magnetic are located on the electronic frame 2481. The electronic clip-on lenses 2485 also include connection points 2486 and 2487 similar to the ones on the electronic frame. The electronic clip-on lenses may include electro-active lenses 2488, 2489 for electronic focusing to supplement the focusing power of the fixed lenses 2495, 2496 located in the electronic frame 2481. In other inventive embodiments, the electronic lenses may be electrochromic lenses that create a variable, electronically-controlled tint or a combination of an electro-chromic tint and electro-active focusing lenses to either correction higher order aberrations, provide presbyopia correction, or focus for conventional needs, for that matter.
The details in
It should be pointed out that nearly all the inventive embodiments described herein can be made to work with rimmed frames, rimless frames, hinged temples, and hingeless temples. Also, the present invention described herein could also be used with Clic Goggles™ that utilize non-electronic eyewear that joins together at the frame bridge to form a frame from two separate eyewear pieces that are connected by way of a tether in the back. This tether secures the Clic Goggle™ eyewear to ones head after the two eyepieces are attached at the bridge. Additionally, the present invention includes electronic and non-electronic connections made by magnetic means, mechanical means, utilizing pins and friction fits and other physical connection techniques, including the combination of magnetic and mechanical connections.
The electronic stem or temple on the front portion of the electronic frame 3431 is sized to fit into the stem on the back portion of the electronic frame tether 3430. In the Clic™ product, the stems or temples on the front portion of the frame are solid plastic. In the present invention, these stems or temples become electrical stems or temples and need to be either hollow to allow for the conductors 3440 and 3441 to be extended down to the lenses, or electronic connections can be applied to the external surface of the stems or temples as taught in
The conductive pairs may be as long as the fully extended length of the electronic frame stems or temples and may be flexible so that they do not break or crack when the front stems are pushed all the way into the back electronic frame stems or temples. A similar set of mechanical locks (not shown) can be placed in the electronic frame stems or temples to hold the position of the front frame stems or temple sections to that of the back frame stems or temples sections. The present invention may join together at the bridge of the nose with any number of methods described herein, including magnets 3450 and 3451.
Utilizing the inventive embodiment allows for a continuous end-to-end electrical circuit that is never disconnected when the electronic eyewear is taken off and decoupled. In this inventive embodiment, the electrical connection to either the speakers, the electronic lenses or the electronic clip-ons remains intact. When utilizing a product where the connection is in the front eyewear bridge, two monocular electronic clip-ons may be used. In this case, each monocular electronic clip-on is applied separately so that it is possible to decouple the eyewear in the bridge without having to take off the clip-on first. However in still other embodiments, a one piece binocular electronic clip-on is used and when this occurs the binocular clip-on may be removed prior to decoupling the eyewear.
While the electronic docking station in this inventive embodiment was located on the back portion of the frame tether, the docking station might also be located anywhere that makes sense on the frame, for example on the frame stem or temple. Once again it should be pointed out that any electronic audio and/or video device can be fabricated to function within the electronic docking station. These could be, by way of example only, an Apple iPod®, MP3 player, tape cassette, satellite radio, conventional radio, pager, cell phone transceiver, microDVD or video file player, video transceiver, etc.
In certain other inventive embodiments, a mirror optical splitter is included within the lens housed by the clip-on and an optical image is directed through the lens house within the clip-on where it optically communicates with the optical splitter housed within the lens. In this case the clip-on allows for a virtual image to appear as if it is floating in space in front of the wearer.
This inventive embodiment allows one to make electrical connections through a frame hinge without actually running wires through the frame hinge. It also allows one to break the frame from the temple to place the frame and frame tether over one's head. In practice the break-away magnetic frame hinge can be placed on both sides of the frame or on just one side of the frame. In the cases where the break-away magnetic frame hinge is used on just one side of the frame, the other side of the frame may include a conventional frame hinge or no frame hinge. While the break-away magnetic frame hinge has been illustrated with electrical connectivity, it is understood that the break-away magnetic frame hinge may be used for non-powered lenses and as such would only require a single magnet on either frame or temple (or both sides) of the frame hinge. Alternatively the electrical connections could be made without using the magnets as electrical contacts. In this case a single magnet on either the frame or temple side of the hinge could be used as long as the electrical contacts are properly insulated from one another.
It is should be noted that the invention contemplates the placement of the magnet on the temple and the metal hinge piece on the frame front as shown in
In another inventive embodiment of the invention, an electronic tether is used in association with a frame having two breakaway magnetic hinges, one for each side of the frame front. In this embodiment the magnets are located on the breakaway stems and the electronic tether is connected to the rear of each stem. It should be pointed out that the magnet breakaway hinge could be used for electronic eyewear or non-electronic eyewear. Also, those active individuals such as athletes and children will benefit greatly by having eyewear with breakaway hinges. Further, this inventive embodiment solves a nuisance that has been prevalent within the optical industry for decades, that being hinge screws that come loose or fall out.
The inventive embodiment solves this historical problem by doing away with the hinge screw and replacing it with a magnet. While the preferred shape of the magnet is that of cylindrical shape as shown in
Finally, it should be pointed out that the structure to which the magnet of a magnetic hinge is attracted or attached to can be of any shape to provide the proper functionality. By way of example only, it can be an open cylinder (see
While the inventors have illustrated many specific examples of how to provide power and/or drive signals to an electrically activated lens using an electronic tether or an electrified frame, it is understood that other methods may be contemplated by those ordinarily skilled in the art. Such additional methods or designs are considered within the scope and spirit of the present invention. It is also understood that the various features, while shown in separate illustrations, could be used in any number of combinations and still be within the scope of the present invention.
In some embodiments, while an electro-active lens may be used to provide vision correction as described in the present invention, the electro-active lens may also be used to provide a sunglass or tinting effect electro-actively. By using special liquid crystal layers or other electro-chromic materials, the electro-active IOL of the present invention can reduce the amount of light that hits the retina when the light levels in the environment become uncomfortably high, or reach a level that can be dangerous to the eye. The sunglass effect may be triggered automatically when a light sensor built into the IOL receives an intensity of light beyond some threshold level. Alternately, the sunglass effect may be switched remotely by the user using a wireless communication device couple to the control circuitry in the IOL. This electro-active sunglass effect may occur in milliseconds or less, in contrast to the relatively slow reaction time of seconds (or more) for commercial photosensitive chemical tints in conventional lenses. One factor in determining the reaction time of electro-active lenses is the thinness of the liquid crystal layer. For example, a 5 micron layer of liquid crystal may react in milliseconds.
Similarly, the focusing of the electro-active elements may be performed automatically by using a range finder, or a tilt meter (near distance when looking down, far distance when looking straight), or may be controlled remotely by the user using a wireless communication device.
There are a number of electro-chromic materials. One type consists of transparent outside layers of electrically conductive film that has inner layers which allow the exchange of ions. When a voltage is applied across the outer conductive layers, ions move from one inner layer to another, causing a change in tinting of the electro chromic material. Reversing the voltage causes the layer to become clear again. The electro-chromic layers can have variable light transmittance during operation, from about 5 to 80 percent. This type of electro chromic glazing has “memory” and does not need constant voltage after the change has been initiated. Further, it can be tuned to block certain wavelengths, such as infrared (heat) energy.
Another electro-chromic technology is called suspended particle display (SPD). This material contains molecular particles suspended in a solution between the plates of glass. In their natural state, the particles move randomly and collide, blocking the direct passage of light. When switched on, the particles align rapidly and the glazing becomes transparent. This type of switchable glazing can block up to about 90 percent of light. Also liquid crystal has been used to provide electro-chromic effects in sunglasses.
The systems and methods, as disclosed herein, are directed to the problems stated above, as well as other problems that are present in conventional techniques. Any description of various products, methods, or apparatus and their attendant disadvantages described in the “Background of the Invention” is in no way intended to limit the scope of the invention, or to imply that invention does not include some or all of the various elements of known products, methods and apparatus in one form or another. Indeed, various embodiments of the invention may be capable of overcoming some of the disadvantages noted in the “Background of the Invention,” while still retaining some or all of the various elements of known products, methods, and apparatus in one form or another.
This application is a continuation of U.S. application Ser. No. 15/920,634, filed Mar. 14, 2019, which is a continuation of U.S. application Ser. No. 15/613,733, filed May 6, 2017, now U.S. Pat. No. 10,159,563, which is a continuation of Ser. No. 14/816,249, filed Aug. 3, 2015, now U.S. Pat. No. 10,172,704, which is a is continuation of U.S. application Ser. No. 13/779,320, filed on Feb. 27, 2013, now U.S. Pat. No. 9,124,796, which is a continuation of U.S. application Ser. No. 11/261,035, filed Oct. 28, 2005, now U.S. Pat. No. 8,778,022, which claims the benefit of the following provisional applications: U.S. Provisional Application No. 60/692,270 filed Jun. 21, 2005; U.S. Provisional Application No. 60/687,341 filed Jun. 6, 2005; U.S. Provisional Application No. 60/687,342 filed Jun. 6, 2005; U.S. Provisional Application No. 60/685,407 filed May 31, 2005; U.S. Provisional Application No. 60/679,241 filed May 10, 2005; U.S. Provisional Application No. 60/674,702 filed Apr. 26, 2005; U.S. Provisional Application No. 60/673,758 filed Apr. 22, 2005; U.S. Provisional Application No. 60/669,403 filed Apr. 8, 2005; U.S. Provisional Application No. 60/667,094 filed Apr. 1, 2005; U.S. Provisional Application No. 60/666,167 filed Mar. 30, 2005; U.S. Provisional Application No. 60/661,925 filed Mar. 16, 2005; U.S. Provisional Application No. 60/659,431 filed Mar. 9, 2005; U.S. Provisional Application No. 60/623,947 filed Nov. 2, 2004; U.S. Provisional Application No. 60/623,946 filed Nov. 2, 2004; and U.S. Provisional Application No. 60/636,490 filed Dec. 17, 2004, all of which are hereby incorporated in their entireties by reference. The following applications, provisional applications, and patents are incorporated by reference in their entirety: U.S. application Ser. No. 11/232,551 filed Sep. 22, 2005; U.S. Pat. No. 6,918,670 issued Jul. 19, 2005; U.S. application Ser. No. 11/183,454 filed Jul. 18, 2005; U.S. Provisional Application No. 60/692,270 filed Jul. 21, 2005; U.S. Provisional Application No. 60/687,342 filed Jun. 6, 2005; U.S. Provisional Application No. 60/687,341 filed Jun. 6, 2005; U.S. Provisional Application No. 60/685,407 filed May 31, 2005; U.S. Provisional Application No. 60/679,241 filed May 10, 2005; U.S. Provisional Application No. 60/674,702 filed Apr. 26, 2005; U.S. Provisional Application No. 60/673,758 filed Apr. 22, 2005; U.S. application Ser. No. 11/109,360 filed Apr. 19, 2005; U.S. Provisional Application No. 60/669,403 filed Apr. 8, 2005; U.S. Provisional Application No. 60/667,094 filed Apr. 1, 2005; U.S. Provisional Application No. 60/666,167 filed Mar. 30, 2005; U.S. Pat. No. 6,871,951 issued Mar. 29, 2005; U.S. application Ser. No. 11/091,104 filed Mar. 28, 2005; U.S. Provisional Application No. 60/661,925 filed Mar. 16, 2005; U.S. Provisional Application No. 60/659,431 filed Mar. 9, 2005; U.S. application Ser. No. 11/063,323 filed Feb. 22, 2005; U.S. Pat. No. 6,857,741 issued Feb. 22, 2005; U.S. Pat. No. 6,851,805 issued Feb. 8, 2005; U.S. application Ser. No. 11/036,501 filed Jan. 14, 2005; U.S. application Ser. No. 11/030,690 filed Jan. 6, 2005; U.S. application Ser. No. 10/996,781 filed Nov. 24, 2004; U.S. Provisional Application No. 60/623,947 filed Nov. 2, 2004; U.S. application Ser. No. 10/924,619 filed Aug. 24, 2004; U.S. application Ser. No. 10/918,496 filed Aug. 13, 2004; U.S. application Ser. No. 10/863,949 filed Jun. 9, 2004; U.S. Pat. No. 6,733,130 issued May 11, 2004; U.S. application Ser. No. 10/772,917 filed Feb. 5, 2004; U.S. Pat. No. 6,619,799 issued Sep. 16, 2003; U.S. application Ser. No. 10/664,112 filed Aug. 20, 2003; U.S. application Ser. No. 10/627,828 filed Jul. 25, 2003; U.S. application Ser. No. 10/387,143 filed Mar. 12, 2003; U.S. Pat. No. 6,517,203 issued Feb. 11, 2003; U.S. Pat. No. 6,491,391 issue Dec. 10, 2002; U.S. Pat. No. 6,491,394 issued Dec. 10, 2002; and U.S. application Ser. No. 10/263,707 filed Oct. 4, 2002.
Number | Name | Date | Kind |
---|---|---|---|
2170287 | Kinnebrew | Aug 1939 | A |
2437642 | Henroteau | Mar 1948 | A |
2576581 | Edwards | Nov 1951 | A |
3161718 | De Luca | Dec 1964 | A |
3183523 | Harrison | May 1965 | A |
3245315 | Marks et al. | Apr 1966 | A |
3248460 | Naujokas | Apr 1966 | A |
3309162 | Kosanke et al. | Mar 1967 | A |
3614215 | Leo | Oct 1971 | A |
3738734 | Tait et al. | Jun 1973 | A |
3791719 | Kratzer et al. | Feb 1974 | A |
4050814 | Mcfadden | Sep 1977 | A |
4062629 | Winthrop | Dec 1977 | A |
4174156 | Glorieux | Nov 1979 | A |
4181408 | Senders | Jan 1980 | A |
4190330 | Berreman | Feb 1980 | A |
4190621 | Greshes | Feb 1980 | A |
4257691 | Brooks | Mar 1981 | A |
4264154 | Petersen | Apr 1981 | A |
4279474 | Belgorod | Jul 1981 | A |
4300818 | Schachar | Nov 1981 | A |
4320939 | Mueller | Mar 1982 | A |
4373218 | Schachar | Feb 1983 | A |
4395736 | Fraleux | Jul 1983 | A |
4418990 | Gerber | Dec 1983 | A |
4423929 | Gomi | Jan 1984 | A |
4457585 | Ducorday | Jul 1984 | A |
4461550 | Legendre | Jul 1984 | A |
4466703 | Nishimoto | Aug 1984 | A |
4466706 | Lamothe, II | Aug 1984 | A |
4529268 | Brown | Jul 1985 | A |
4564267 | Nishimoto | Jan 1986 | A |
4572616 | Kowel et al. | Feb 1986 | A |
4577928 | Brown | Mar 1986 | A |
4601545 | Kern | Jul 1986 | A |
4609824 | Munier et al. | Sep 1986 | A |
4712870 | Robinson et al. | Dec 1987 | A |
4753514 | Kubik | Jun 1988 | A |
4756605 | Okada et al. | Jul 1988 | A |
4772094 | Sheiman | Sep 1988 | A |
D298250 | Kildall | Oct 1988 | S |
4781440 | Toda | Nov 1988 | A |
4787733 | Silva | Nov 1988 | A |
4787903 | Grendahl | Nov 1988 | A |
4795248 | Okada et al. | Jan 1989 | A |
4807630 | Malinouskas | Feb 1989 | A |
4813777 | Rainville et al. | Mar 1989 | A |
4816031 | Pfoff | Mar 1989 | A |
4818095 | Takeuchi | Apr 1989 | A |
4836652 | Oishi et al. | Jun 1989 | A |
4842400 | Klein | Jun 1989 | A |
4869588 | Frieder et al. | Sep 1989 | A |
4873029 | Blum | Oct 1989 | A |
4880300 | Payner et al. | Nov 1989 | A |
4890903 | Treisman et al. | Jan 1990 | A |
4904063 | Okada et al. | Feb 1990 | A |
4907860 | Noble | Mar 1990 | A |
4909626 | Purvis et al. | Mar 1990 | A |
4919520 | Okada et al. | Apr 1990 | A |
4921728 | Takiguchi et al. | May 1990 | A |
4927241 | Kuijk | May 1990 | A |
4929865 | Blum | May 1990 | A |
4930884 | Tichenor et al. | Jun 1990 | A |
4944584 | Maeda et al. | Jul 1990 | A |
4945242 | Berger et al. | Jul 1990 | A |
4952048 | Frieder et al. | Aug 1990 | A |
4952788 | Berger et al. | Aug 1990 | A |
4955712 | Barth et al. | Sep 1990 | A |
4958907 | Davis | Sep 1990 | A |
4961639 | Lazarus | Oct 1990 | A |
4968127 | Russell et al. | Nov 1990 | A |
4981342 | Fiala | Jan 1991 | A |
4991951 | Mizuno et al. | Feb 1991 | A |
5015086 | Okaue et al. | May 1991 | A |
5030882 | Solero | Jul 1991 | A |
5050981 | Roffman | Sep 1991 | A |
5066301 | Wiley | Nov 1991 | A |
5067795 | Senatore | Nov 1991 | A |
5073021 | Marron | Dec 1991 | A |
5076665 | Petersen | Dec 1991 | A |
5089023 | Swanson | Feb 1992 | A |
5091801 | Ebstein | Feb 1992 | A |
5108169 | Mandell | Apr 1992 | A |
5114628 | Hoefer et al. | May 1992 | A |
5122974 | Chance | Jun 1992 | A |
5130856 | Tichenor et al. | Jul 1992 | A |
5142411 | Fiala | Aug 1992 | A |
5147585 | Blum | Sep 1992 | A |
5150234 | Takahashi et al. | Sep 1992 | A |
5171266 | Wiley et al. | Dec 1992 | A |
5173723 | Volk | Dec 1992 | A |
5178800 | Blum | Jan 1993 | A |
5182585 | Stoner | Jan 1993 | A |
5184156 | Black et al. | Feb 1993 | A |
5200859 | Payner et al. | Apr 1993 | A |
5208688 | Fergason et al. | May 1993 | A |
5219497 | Blum | Jun 1993 | A |
5229797 | Futhey et al. | Jul 1993 | A |
5229885 | Quaglia | Jul 1993 | A |
5231430 | Kohayakawa | Jul 1993 | A |
5239412 | Naka et al. | Aug 1993 | A |
D342063 | Howitt et al. | Dec 1993 | S |
5299053 | Kleinburg et al. | Mar 1994 | A |
5305028 | Okano | Apr 1994 | A |
5306926 | Yonemoto | Apr 1994 | A |
5324930 | Jech | Jun 1994 | A |
D350342 | Sack | Sep 1994 | S |
5352886 | Kane | Oct 1994 | A |
5359444 | Piosenka et al. | Oct 1994 | A |
5375006 | Haas | Dec 1994 | A |
5382986 | Black et al. | Jan 1995 | A |
5386308 | Michel et al. | Jan 1995 | A |
5401175 | Guimond et al. | Mar 1995 | A |
5411537 | Munshi et al. | May 1995 | A |
5424927 | Schaller et al. | Jun 1995 | A |
5440357 | Quaglia | Aug 1995 | A |
5443506 | Garabet | Aug 1995 | A |
5451766 | Van | Sep 1995 | A |
5455638 | Kallman et al. | Oct 1995 | A |
5488439 | Weltmann | Jan 1996 | A |
5512371 | Gupta et al. | Apr 1996 | A |
5522323 | Richard | Jun 1996 | A |
5552841 | Gallorini et al. | Sep 1996 | A |
5585871 | Linden | Dec 1996 | A |
5606743 | Vogt et al. | Feb 1997 | A |
5608567 | Grupp | Mar 1997 | A |
5608808 | Da | Mar 1997 | A |
5615588 | Gottschald | Apr 1997 | A |
5653751 | Samiy et al. | Aug 1997 | A |
5654786 | Gerald | Aug 1997 | A |
5668620 | Kurtin et al. | Sep 1997 | A |
5682223 | Menezes et al. | Oct 1997 | A |
5683457 | Gupta et al. | Nov 1997 | A |
RE35691 | Theirl et al. | Dec 1997 | E |
5702819 | Gupta et al. | Dec 1997 | A |
5712721 | Large | Jan 1998 | A |
5715337 | Spitzer et al. | Feb 1998 | A |
5728155 | Anello et al. | Mar 1998 | A |
5739959 | Quaglia | Apr 1998 | A |
5757458 | Miller et al. | May 1998 | A |
5777719 | Williams et al. | Jul 1998 | A |
5815233 | Morokawa et al. | Sep 1998 | A |
5815239 | Chapman et al. | Sep 1998 | A |
5859685 | Gupta et al. | Jan 1999 | A |
5861934 | Blum et al. | Jan 1999 | A |
5861936 | Sorensen | Jan 1999 | A |
5877876 | Birdwell | Mar 1999 | A |
5886822 | Spitzer | Mar 1999 | A |
5900720 | Kallman et al. | May 1999 | A |
5903395 | Rallison | May 1999 | A |
5905561 | Lee et al. | May 1999 | A |
5949521 | Williams et al. | Sep 1999 | A |
5953098 | Lieberman et al. | Sep 1999 | A |
5956183 | Epstein et al. | Sep 1999 | A |
5963300 | Horwitz | Oct 1999 | A |
5971540 | Ofner | Oct 1999 | A |
5980037 | Conway | Nov 1999 | A |
5988816 | Quadri | Nov 1999 | A |
5999328 | Kurtin et al. | Dec 1999 | A |
6007363 | Renk | Dec 1999 | A |
6023372 | Spitzer et al. | Feb 2000 | A |
6034653 | Robertson et al. | Mar 2000 | A |
6040947 | Kurtin et al. | Mar 2000 | A |
6050687 | Bille et al. | Apr 2000 | A |
6050717 | Kosugi et al. | Apr 2000 | A |
6069742 | Silver | May 2000 | A |
6086203 | Blum et al. | Jul 2000 | A |
6086204 | Magnante | Jul 2000 | A |
6091546 | Spitzer | Jul 2000 | A |
6091832 | Shurman et al. | Jul 2000 | A |
6095651 | Williams et al. | Aug 2000 | A |
6099117 | Gregory | Aug 2000 | A |
6115177 | Vossler | Sep 2000 | A |
6139148 | Menezes | Oct 2000 | A |
6145987 | Baude et al. | Nov 2000 | A |
6157291 | Kuenster et al. | Dec 2000 | A |
6165123 | Thompson | Dec 2000 | A |
6188525 | Silver | Feb 2001 | B1 |
6191881 | Tajima | Feb 2001 | B1 |
6199984 | Menezes | Mar 2001 | B1 |
6199986 | Williams et al. | Mar 2001 | B1 |
6204974 | Spitzer | Mar 2001 | B1 |
6213602 | Smarto | Apr 2001 | B1 |
6270220 | Keren | Aug 2001 | B1 |
6271915 | Frey et al. | Aug 2001 | B1 |
6282449 | Kamerling et al. | Aug 2001 | B1 |
6299311 | Williams et al. | Oct 2001 | B1 |
6305802 | Roffman et al. | Oct 2001 | B1 |
6324053 | Kamijo | Nov 2001 | B1 |
6325508 | Decreton et al. | Dec 2001 | B1 |
6338559 | Williams et al. | Jan 2002 | B1 |
6350031 | Lashkari et al. | Feb 2002 | B1 |
6390623 | Kokonaski et al. | May 2002 | B1 |
6396622 | Alden | May 2002 | B1 |
6437762 | Birdwell | Aug 2002 | B1 |
6437925 | Nishioka | Aug 2002 | B1 |
6464363 | Nishioka et al. | Oct 2002 | B1 |
6491391 | Blum et al. | Dec 2002 | B1 |
6491394 | Blum et al. | Dec 2002 | B1 |
6501443 | Mcmahon | Dec 2002 | B1 |
6517203 | Blum et al. | Feb 2003 | B1 |
6542081 | Torch | Apr 2003 | B2 |
6554425 | Roffman et al. | Apr 2003 | B1 |
6569199 | Dotan et al. | May 2003 | B1 |
6609794 | Levine | Aug 2003 | B2 |
6614408 | Mann | Sep 2003 | B1 |
6616275 | Dick et al. | Sep 2003 | B1 |
6616279 | Davis et al. | Sep 2003 | B1 |
6618208 | Silver | Sep 2003 | B1 |
6619799 | Blum et al. | Sep 2003 | B1 |
6626532 | Nishioka et al. | Sep 2003 | B1 |
6631001 | Kuiseko | Oct 2003 | B2 |
6638304 | Azar | Oct 2003 | B2 |
6643552 | Edell et al. | Nov 2003 | B2 |
6652096 | Morris et al. | Nov 2003 | B1 |
6667471 | Bos et al. | Dec 2003 | B2 |
6682195 | Dreher | Jan 2004 | B2 |
6705729 | Piers et al. | Mar 2004 | B2 |
6709105 | Menezes | Mar 2004 | B2 |
6709107 | Jiang et al. | Mar 2004 | B2 |
6709108 | Levine et al. | Mar 2004 | B2 |
6714133 | Hum et al. | Mar 2004 | B2 |
6733130 | Blum et al. | May 2004 | B2 |
6738199 | Nishioka | May 2004 | B2 |
6768536 | Okuwaki et al. | Jul 2004 | B2 |
6769767 | Swab et al. | Aug 2004 | B2 |
6774871 | Birdwell | Aug 2004 | B2 |
6778246 | Sun et al. | Aug 2004 | B2 |
6793340 | Morris et al. | Sep 2004 | B1 |
6830193 | Tanaka | Dec 2004 | B2 |
6833938 | Nishioka | Dec 2004 | B2 |
6840619 | Dreher | Jan 2005 | B2 |
6851805 | Blum et al. | Feb 2005 | B2 |
6857741 | Blum et al. | Feb 2005 | B2 |
6859333 | Ren et al. | Feb 2005 | B1 |
6871951 | Blum et al. | Mar 2005 | B2 |
6883916 | Menezes | Apr 2005 | B2 |
6886938 | Menezes | May 2005 | B1 |
6893124 | Kurtin | May 2005 | B1 |
6894751 | Payne et al. | May 2005 | B2 |
6902271 | Perrott et al. | Jun 2005 | B2 |
6918570 | Ahn | Jul 2005 | B2 |
6918670 | Blum et al. | Jul 2005 | B2 |
6948818 | Williams et al. | Sep 2005 | B2 |
6951391 | Michael et al. | Oct 2005 | B2 |
6955433 | Benjamin et al. | Oct 2005 | B1 |
6956682 | Benjamin | Oct 2005 | B2 |
6976982 | Santini et al. | Dec 2005 | B2 |
6986579 | Blum et al. | Jan 2006 | B2 |
7008054 | Kurtin et al. | Mar 2006 | B1 |
7009757 | Nishioka et al. | Mar 2006 | B2 |
7018040 | Blum et al. | Mar 2006 | B2 |
7019890 | Meredith et al. | Mar 2006 | B2 |
7023594 | Blum et al. | Apr 2006 | B2 |
7034619 | Lynch | Apr 2006 | B2 |
7041133 | Azar | May 2006 | B1 |
7077519 | Blum et al. | Jul 2006 | B2 |
7085065 | Silver | Aug 2006 | B2 |
7130664 | Williams | Oct 2006 | B1 |
7133172 | Nishioka | Nov 2006 | B2 |
7137702 | Piers et al. | Nov 2006 | B2 |
7137952 | Leonardi et al. | Nov 2006 | B2 |
7159981 | Kato | Jan 2007 | B2 |
7159983 | Menezes et al. | Jan 2007 | B2 |
7188948 | Blum et al. | Mar 2007 | B2 |
7192136 | Howell et al. | Mar 2007 | B2 |
7195353 | Blum et al. | Mar 2007 | B2 |
7209097 | Suyama et al. | Apr 2007 | B2 |
7229173 | Menezes | Jun 2007 | B2 |
7255437 | Howell et al. | Aug 2007 | B2 |
7290876 | Duston et al. | Nov 2007 | B2 |
7380936 | Howell et al. | Jun 2008 | B2 |
7396126 | Blum et al. | Jul 2008 | B2 |
7401918 | Howell et al. | Jul 2008 | B2 |
7404636 | Blum et al. | Jul 2008 | B2 |
7438410 | Howell et al. | Oct 2008 | B1 |
7461936 | Jannard | Dec 2008 | B2 |
7475984 | Blum et al. | Jan 2009 | B2 |
7481531 | Howell et al. | Jan 2009 | B2 |
7500746 | Howell et al. | Mar 2009 | B1 |
7500747 | Howell et al. | Mar 2009 | B2 |
7543934 | Howell et al. | Jun 2009 | B2 |
7581833 | Howell et al. | Sep 2009 | B2 |
7607775 | Hermanson et al. | Oct 2009 | B2 |
7621634 | Howell et al. | Nov 2009 | B2 |
7677723 | Howell et al. | Mar 2010 | B2 |
7760898 | Howell et al. | Jul 2010 | B2 |
7771046 | Howell et al. | Aug 2010 | B2 |
7792552 | D et al. | Sep 2010 | B2 |
7806525 | Howell et al. | Oct 2010 | B2 |
7831055 | Frerking et al. | Nov 2010 | B2 |
7922321 | Howell et al. | Apr 2011 | B2 |
7988283 | Jannard | Aug 2011 | B2 |
8025396 | Power | Sep 2011 | B1 |
8089511 | Yamamoto | Jan 2012 | B2 |
8109629 | Howell et al. | Feb 2012 | B2 |
8174569 | Tanijiri et al. | May 2012 | B2 |
8337013 | Howell et al. | Dec 2012 | B2 |
8337014 | Kokonaski et al. | Dec 2012 | B2 |
8430507 | Howell et al. | Apr 2013 | B2 |
8434863 | Howell et al. | May 2013 | B2 |
8465151 | Howell et al. | Jun 2013 | B2 |
8500271 | Howell et al. | Aug 2013 | B2 |
8708483 | Kokonaski et al. | Apr 2014 | B2 |
8770742 | Howell et al. | Jul 2014 | B2 |
8905541 | Blum et al. | Dec 2014 | B2 |
8915588 | Blum et al. | Dec 2014 | B2 |
8931896 | Blum et al. | Jan 2015 | B2 |
9122083 | Blum et al. | Sep 2015 | B2 |
9124796 | Blum et al. | Sep 2015 | B2 |
10092395 | Blum et al. | Oct 2018 | B2 |
10114235 | Blum et al. | Oct 2018 | B2 |
10159563 | Blum et al. | Dec 2018 | B2 |
10379575 | Blum et al. | Aug 2019 | B2 |
20010005230 | Ishikawa | Jun 2001 | A1 |
20010055094 | Zhang | Dec 2001 | A1 |
20020049374 | Abreu | Apr 2002 | A1 |
20020060525 | Sagano et al. | May 2002 | A1 |
20020101568 | Eberl | Aug 2002 | A1 |
20020140899 | Blum et al. | Oct 2002 | A1 |
20020149739 | Perrott et al. | Oct 2002 | A1 |
20020186346 | Stantz et al. | Dec 2002 | A1 |
20030004546 | Casey | Jan 2003 | A1 |
20030018383 | Azar | Jan 2003 | A1 |
20030067585 | Miller et al. | Apr 2003 | A1 |
20030086052 | Zelman | May 2003 | A1 |
20030103413 | Jacobi et al. | Jun 2003 | A1 |
20030112523 | Daniell | Jun 2003 | A1 |
20030151721 | Lai et al. | Aug 2003 | A1 |
20030199978 | Lindsey et al. | Oct 2003 | A1 |
20030208265 | Ho et al. | Nov 2003 | A1 |
20030210377 | Blum et al. | Nov 2003 | A1 |
20040000733 | Swab et al. | Jan 2004 | A1 |
20040008157 | Brubaker et al. | Jan 2004 | A1 |
20040008319 | Lai et al. | Jan 2004 | A1 |
20040010310 | Peyman | Jan 2004 | A1 |
20040108971 | Waldern et al. | Jun 2004 | A1 |
20040117011 | Aharoni et al. | Jun 2004 | A1 |
20040130677 | Liang et al. | Jul 2004 | A1 |
20040160573 | Jannard et al. | Aug 2004 | A1 |
20040179280 | Nishioka | Sep 2004 | A1 |
20040186533 | Greenberg et al. | Sep 2004 | A1 |
20040196435 | Dick et al. | Oct 2004 | A1 |
20040239874 | Swab et al. | Dec 2004 | A1 |
20040246440 | Andino et al. | Dec 2004 | A1 |
20050049578 | Tu et al. | Mar 2005 | A1 |
20050073739 | Meredith et al. | Apr 2005 | A1 |
20050078274 | Howell et al. | Apr 2005 | A1 |
20050099594 | Blum et al. | May 2005 | A1 |
20050113912 | Feenstra et al. | May 2005 | A1 |
20050124983 | Frey et al. | Jun 2005 | A1 |
20050237485 | Blum et al. | Oct 2005 | A1 |
20050256571 | Azar | Nov 2005 | A1 |
20060044510 | Williams et al. | Mar 2006 | A1 |
20060066808 | Blum et al. | Mar 2006 | A1 |
20060095128 | Blum et al. | May 2006 | A1 |
20060113054 | Silvestrini | Jun 2006 | A1 |
20060122531 | Goodall et al. | Jun 2006 | A1 |
20060164593 | Peyghambarian et al. | Jul 2006 | A1 |
20060173367 | Stuart et al. | Aug 2006 | A1 |
20060177086 | Rye et al. | Aug 2006 | A1 |
20060183986 | Rice et al. | Aug 2006 | A1 |
20070030442 | Howell et al. | Feb 2007 | A1 |
20080143954 | Abreu et al. | Jun 2008 | A1 |
20080218684 | Howell et al. | Sep 2008 | A1 |
20090256977 | Haddock et al. | Oct 2009 | A1 |
20090296044 | Howell et al. | Dec 2009 | A1 |
20100103076 | Yamamoto | Apr 2010 | A1 |
20100177277 | Kokonaski et al. | Jul 2010 | A1 |
20100292599 | Oleson et al. | Nov 2010 | A1 |
20110071459 | Rickard et al. | Mar 2011 | A1 |
20110249230 | Blum | Oct 2011 | A1 |
20130215374 | Blum et al. | Aug 2013 | A1 |
20130235332 | Blum et al. | Sep 2013 | A1 |
20130242253 | Blum et al. | Sep 2013 | A1 |
20130250135 | Blum et al. | Sep 2013 | A1 |
20130250233 | Blum et al. | Sep 2013 | A1 |
20150335420 | Blum et al. | Nov 2015 | A1 |
20150378180 | Blum et al. | Dec 2015 | A1 |
20170176777 | Blum et al. | Jun 2017 | A1 |
20180221137 | Blum et al. | Aug 2018 | A1 |
20180329234 | Blum et al. | Nov 2018 | A1 |
20190110887 | Blum et al. | Apr 2019 | A1 |
Number | Date | Country |
---|---|---|
89113088 | Oct 2001 | CN |
19891013088 | Oct 2001 | CN |
2911723 | Jun 2007 | CN |
201222131 | Apr 2009 | CN |
201464741 | May 2010 | CN |
4223395 | Jan 1994 | DE |
0154962 | Sep 1985 | EP |
0233104 | Aug 1987 | EP |
0237365 | Sep 1987 | EP |
0578833 | Jan 1994 | EP |
0649044 | Apr 1995 | EP |
0918248 | May 1999 | EP |
2169417 | Jul 1986 | GB |
2170613 | Aug 1986 | GB |
1122719 | May 2009 | HK |
S5576323 | Jun 1980 | JP |
S58113912 | Jul 1983 | JP |
S61156227 | Jul 1986 | JP |
S61502221 | Oct 1986 | JP |
S6357044 | Mar 1988 | JP |
H01237610 | Sep 1989 | JP |
H0423579 | Jan 1992 | JP |
H0461495 | Feb 1992 | JP |
H05100201 | Apr 1993 | JP |
H0728002 | Jan 1995 | JP |
H08508826 | Sep 1996 | JP |
H11352445 | Dec 1999 | JP |
2001252346 | Sep 2001 | JP |
2001522063 | Nov 2001 | JP |
2002533158 | Oct 2002 | JP |
2003230590 | Aug 2003 | JP |
2000138858 | Sep 2007 | JP |
2007323062 | Dec 2007 | JP |
2008545287 | Dec 2008 | JP |
2011229024 | Nov 2011 | JP |
8505466 | Dec 1985 | WO |
9423334 | Oct 1994 | WO |
9425510 | Nov 1994 | WO |
9923524 | May 1999 | WO |
0038593 | Jul 2000 | WO |
0106298 | Jan 2001 | WO |
02057836 | Jul 2002 | WO |
03007851 | Jan 2003 | WO |
03050472 | Jun 2003 | WO |
03068059 | Aug 2003 | WO |
2004008189 | Jan 2004 | WO |
2004015460 | Feb 2004 | WO |
2004015481 | Feb 2004 | WO |
2004034095 | Apr 2004 | WO |
2004072687 | Aug 2004 | WO |
2006120416 | Nov 2006 | WO |
2010080999 | Jul 2010 | WO |
Entry |
---|
Anderson, M. “Adaptive Optics: Liquid Crystals Lower the Cost of Adaptive Optics” Laser Focus World (Dec. 1999). |
Bertsch, A. et al., “The Sensing Contact Lens”, Medical Device Technology (2006); 17: 19-21. |
Bradley, Arthur “Profile: Larry N. Thibos, PhD., and Donald T. Miller, PhD.” Indiana Journal of Optometry; 2:I (Spring 1999). |
Danfield, C., Phakic Intraocular Lens Implant, WordcatTop Quality Articles, 2 pages, Retrieved from the Internet on Nov. 20, 2008: http://www.wordcat.co.uk/articles/phakic-intraocular-lens-implant/279/. |
Davis, Robert A. “Computer Vision Syndrome—The Eyestrain Epidemic” Review of Optometry (Sep. 15, 1997). |
Donald T. Miller, Xin Hong, and Larry N. Thibos, “Requirements for the segmented spatial light modulators for diffraction-limited imaging through aberrated eyes,” G.D. Love, ed. Proceedings of the 2nd International Workshop on Adaptive Optics for Industry and Medicine, World Scientific, Singapore, 63-68 (Jul. 1999). |
Eggers, T. et al., “Wireless Intra-ocular Pressure Monitoring System Integrated in an Artificial Lens”, Presented at the First Annual International IEEE-EMBS Special Topic Conference on Microtechnologies in Medicine & Biology, Lyon, France, Oct. 12-14, 2000; Paper 7: 466-469. |
Eyecare Business (Oct. 1997). |
International Preliminary Report on Patentabiity for Application No. PCT/US2005/039101 dated May 8, 2007. |
International Preliminary Report on Patentability for PCT/US2009/044168 dated Nov. 17, 2010. |
International Search Report corresponding to the PCT/US09/037544 application dated May 20, 2009. |
International Search Report for Application No. PCT/US 08/54721 dated Aug. 20, 2008. |
International Search Report for Application No. PCT/US2005/039101 dated Jul. 7, 2006. |
International Search Report for application PCT/US 08/54721 dated Aug. 20, 2008. |
International Search Report for PCT/US2009/044168 dated Oct. 7, 2009. |
International Search Report of Application No. PCT/US08/51649 dated Jul. 7, 2008. |
Kowel, Stephen T., et al “Focusing by electrical modulation of refraction in a liquid crystal cell” Applied Optics23:2 (Jan. 15, 1984). |
Lazarus, Stuart M. “The Use of Yoked Base-Up and Base-In Prism for Reducing Eye Strain at the Computer” Journal of the American Optometric Association (Apr. 1996). |
Leonardi, M. et al., “A Soft Contact Lens with a MEMS Strain Gage Embedded for Intraocular Pressure Monitoring”, Transducers '03; The 12th International Conference on Solid Slate Sensors, Actuators and Microsyslems, Boston, Jun. 8-12, 2003; 362.5: 1043-1046. |
Leonardi, M. et al., “First Steps toward Noninvasive IOP—Monitoring with a Sensing Contact Lens”, Investigative Ophthalmology & Visual Science (2004); 45(9): 3113-3117. |
Machine English Translation of JP2000-138858 (May 16, 2000). |
Miller, Donald T. et al. “Requirements for the segmented spatial light modulators for diffraction-limited imaging through aberrated eyes,” G.D. Love, ed. Proceedings of the 2nd International Workshop on Adaptive Optics for Industry and Medicine, World Scientific, Singapore, 63-68 (Jul. 1999). |
Naumov, A.F. “Control Optimization of Spherical Modal Liquid Crystal Lenses”, Optics Express 4:9; Optical Society of America (Apr. 26, 1999). |
Naumov, A.F. “Liquid Crystal Adaptive Lenses with Modal Control” Optics Letters, 23:13 Optical Society of America (Jul. 1, 1998). |
Optics, Org, Dec. 19, 2006 “Liquid Lenses Eye Commercial Breakthrough” Opto & Laser Europe (Nov. 2003). |
Pitchon, E.M. et al., “First In-Vivo Human Monitoring of Intraocular Pressure Fluctuation and Ocular Pulsation by a Wireless Soft Contact Lens Sensor.” Congress of the European Glaucoma Society, Berlin, Jun. 2008; Congres annuel de la Societe francaise d'ophtalmologie, Paris, May 2008; ARVO Meeting (The Association for Research in Vision and Ophthalmology), Apr. 27-May 1, 2008, Fort Lauderdale American Glaucoma Society, 18th Annual Meeting, Mar. 2008, Washington, 1 page. |
Supplementary European Search Report of Application No. EP 05824718 dated Nov. 19, 2007. |
Tarascon et al., “Issues and challenges facing rechargeable lithium batteries” Nature 2001, 414:359-367 (Nov. 15, 2001). |
Thibos, Larry N., et al. “Vision through a liquid-crystal spatial light modulator” Adaptive Optics Conference; Durham, UK (1999). |
Thibos, Larry N., et. al. “Use of Liquid-Crystal Adaptive-Optics to Alter the Refractive State of the Eye; Optometry and Vision Science” 74:7; American Academy of Optometry (Jul. 1997). |
Thibos, Larry, N., et al. “Electronic Spectacles for the 21 Century” Indian Journal of Optometry, 2:1 (Spring 1999). |
Walter, P. et al., “Development of a completely encapsulated intraocular pressure sensor”, Ophthalmic Research (2000); 32: 278-284. |
Number | Date | Country | |
---|---|---|---|
20190317551 A1 | Oct 2019 | US |
Number | Date | Country | |
---|---|---|---|
60692270 | Jun 2005 | US | |
60687341 | Jun 2005 | US | |
60687342 | Jun 2005 | US | |
60685407 | May 2005 | US | |
60679241 | May 2005 | US | |
60674702 | Apr 2005 | US | |
60673758 | Apr 2005 | US | |
60669403 | Apr 2005 | US | |
60667094 | Apr 2005 | US | |
60666167 | Mar 2005 | US | |
60661925 | Mar 2005 | US | |
60659431 | Mar 2005 | US | |
60636490 | Dec 2004 | US | |
60623947 | Nov 2004 | US | |
60623946 | Nov 2004 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 15920634 | Mar 2018 | US |
Child | 16454823 | US | |
Parent | 15613733 | Jun 2017 | US |
Child | 15920634 | US | |
Parent | 14816249 | Aug 2015 | US |
Child | 15613733 | US | |
Parent | 13779320 | Feb 2013 | US |
Child | 14816249 | US | |
Parent | 11261035 | Oct 2005 | US |
Child | 13779320 | US |