There is a strong clinical need to improve the resolution of magnetic resonance imaging (MRI) for the detection of small pathological lesions. A salient example is Cushing's disease (CD): a potentially fatal disorder caused by an adrenocorticotropin hormone (ACTH)-producing pituitary tumor. Clinical magnetic resonance imaging (MRI) of the pituitary gland sometimes fails to detect small pituitary tumors due to limited signal-to-noise ratio (SNR) and spatial resolution. While the median size of pituitary tumors (microadenomas) causing CD is 5 mm, a significant percentage are less than 3 mm in size. Currently, MRI is unable to detect up to 50% of microadenomas in CD. This failure of diagnostic imaging thwarts the primary and optimal treatment of CD: surgical excision of the offending tumor. In such cases without an imaging-identifiable tumor, neurosurgeons may need to resort to surgical exploration and systemic slicing of the pituitary gland to identify the small pituitary tumors. For example, neurosurgeons must consider surgically “exploring” the anterior pituitary gland by making multiple parallel incisions typically spaced 2-3 mm apart with the hope of fortuitously encountering the tumor. In addition to the real possibility of not finding a tumor, this technique adds the risk of permanently damaging the normal gland.
Standard pituitary MRI protocols generate multi-slice 2-dimensional (2D) images with a typical in-plane resolution of 0.7×0.7 mm2 and through-plane slice thickness of 3 mm. When considering various shapes of the pituitary gland, partial volume averaging, and motion-related degradation, it is not surprising that MR images with an in-plane pixel size of 0.7 mm commonly fail to detect lesions smaller than 3 mm.
One of the common factors limiting MRI spatial resolution is signal-to-noise ratio (SNR). Two approaches for increased SNR are to use higher strengths (e.g., 7T MRI scanner) and to design application-specific radiofrequency (RF) coil arrays. These two are generally additive when combined with each other. Advancements have been made with flexible RF coil arrays. Conforming the coil elements to the patient's surface anatomy achieves higher SNR in directly adjacent regions, which unfortunately is of limited value for certain regions of the anatomy located inside a subject, for example, imaging of the pituitary given that the pituitary gland is located centrally within the cranium. Another approach to improve SNR is to place a separate receive-only RF coil in close proximity to the imaging target. One example in clinical use, the endorectal coil designed for prostate imaging, has had limited use due to patient discomfort related to the relatively large diameter of the endorectal component. One prior study adopted the endorectal prostate coil for pituitary imaging. The study demonstrated a potential 10-fold increase in SNR by positioning the coil apparatus withing the sphenoid sinus via a sublabial approach in a cadaver. However, the design included a potential concern that the coil is needed to be positioned blindly, given the complete obstruction of the surgical corridor by the probe.
In accordance with an embodiment, an RF coil apparatus for magnetic resonance imaging (MRI) includes a flexible RF coil configured to be positioned in a cavity of a subject proximate to a region of interest located internally in the subject and a circuit assembly coupled to the flexible RF coil and configured to be positioned external to the subject. The circuitry assembly can include adjustable tune and match components. The RF coil apparatus can further include a first connector connected between the flexible RF coil and the circuit assembly and a second connector coupled to the circuit assembly and configured to connect to an MRI system.
In accordance with another embodiment, a magnetic resonance imaging (MRI) system includes a magnet system configured to generate a polarizing magnetic field about at least a portion of a subject, a magnetic gradient system including a plurality of magnetic gradient coils configured to apply at least one magnetic gradient field to the polarizing magnetic field; and a radio frequency (RF) system configured to apply an RF field to the subject and to receive magnetic resonance signals from the subject. The RF system can include an RF coil apparatus. The RF coil apparatus includes a flexible RF coil configured to be positioned in a cavity of a subject proximate to a region of interest located internally in the subject and a circuit assembly coupled to the flexible RF coil and configured to be positioned external to the subject. The circuit assembly can include adjustable tune and match components. The RF coil apparatus can further include a first connector connected between the flexible RF coil and the circuit assembly and a second connector coupled to the circuit assembly and the RF system.
The present invention will hereafter be described with reference to the accompanying drawings, wherein like reference numerals denote like elements.
The present disclosure describes a radiofrequency (RF) coil apparatus that includes a flexible RF coil having dimensions configured to provide a small field-of-view and to allow positioning of the flexible RF coil inside a cavity of a subject (e.g., a patient) and in close proximity to a region of interest internal to the subject, enabling high signal-to-noise-ratio (SNR) MRI of the region of interest. The increased SNR provided from the RF coil apparatus can enable markedly higher resolution imaging compared to conventional external RF coils. The RF coil apparatus can advantageously enable a much increased spatial resolution than that currently provided with standard MR imaging. In some embodiments, the increased SNR can enable control over the measurement time (e.g., faster acquisitions of images or snapshots). The RF coil apparatus may also include a circuit assembly (or box) coupled to the flexible RF coil and configured to be located external to the subject, and a connector configured to couple the circuit assembly to an MRI system. In some embodiments, the flexible RF coil can be a miniature single-loop flexible coil. In some embodiments, the miniature flexible RF coil can be, for example, a butterfly coil, a figure-of-eight coil, or an array of coils. In some embodiments where the miniature flexible RF coil is an array of coils, the array of coils can be configured to look outward from the cavity (e.g., a sphenoid sinus cavity) in which the RF coil is positioned. The flexible RF coil may be coupled to the circuit assembly using a connector such as, for example, a coaxial cable. The flexible RF coil and a portion of the RF coil connector can form an intracavity portion of the RF apparatus that can be inserted into the cavity of the subject.
The circuit assembly can include adjustable tune and match components configured to allow fine tuning of the RF coil. Advantageously, the circuit assembly can be located outside of the subject when the RF coil is inserted into a cavity of the subject. Locating the circuit assembly external to the subject can enable the dimensions of the RF coil to be minimized to fit inside the body of the subject. In some embodiments, the circuit assembly can also include a pre-amplifier. In some embodiments, the region of interest can be the pituitary gland and the flexible RF coil of the RF coil apparatus is configured with dimensions to allow insertion of the flexible RF coil through a nostril of the subject (i.e., endonasally) and into the sphenoid sinus cavity where the flexible RF coil can be positioned near the pituitary gland. For example, the single-coil flexible RF coil can be configured to be positioned millimeters from or placed against the pituitary gland. In some embodiments, the flexible RF coil of the RF coil apparatus can be designed with different diameters, geometries, resonance frequencies, and placement configurations. In some embodiments, different flexible RF coils with different sizes and shapes can be provided for an RF coil apparatus such that an RF coil with an optimal shape and dimensions may be selected based on the specific anatomy of each patient.
where dmax is the maximum distance of interest from the RF coil 102. Placement of the RF coil 102 inside a cavity of a subject and in close proximity to a region of interest internal to the subject can require that the RF coil 102 be flexible to be able to bend slightly as needed. In some embodiments, the flexible RF coil 102 may be configured to be placed in close proximity to the region of interest or placed against the region of interest. In some embodiments, the orientation of the RF coil 102 when placed in the cavity and in close proximity to the region of interest can be parallel to the orientation of the main magnetic field, B0, generated by an MRI system (e.g., MRI system 200 shown in
The flexible RF coil 102 may be coupled to a circuit assembly 104. In some embodiments, the flexible RF coil 102 may be coupled and connected to the circuit assembly 104 using a first connector 108 (e.g., a cable). In some embodiments, the first connector 108 may be a coaxial cable. The circuit assembly 104 can be advantageously configured to be located remotely and external to the subject when the RF coil 102 is inserted into a cavity of the subject. Locating the circuit assembly 104 external to the subject can enable the dimensions of the RF coil 102 to be minimized to fit inside the body of the subject. The circuit assembly 104 can include a housing and circuit elements or components. In some embodiments, the circuit assembly 104 can include adjustable tune and match components 106 (shown in
As mentioned above, the RF coil apparatus 100 can be configured to allow positioning of the flexible RF coil 102 inside a cavity of a subject (e.g., a patient) and in close proximity to a region of interest internal to the subject.
In
Endonasal placement of the RF coil 102 can require that the RF coil 102 be flexible and configured to be able to bend slightly beyond a U-shape in order to pass by the nostril. In some embodiments, an RF coil diameter up to 2.5-cm could be inserted in a nostril without hyperangulation (kinking). Once past the nostril, further advancement of the flexible RF coil 102 into the sphenoid sinus cavity can be easy and safe.
A circuit assembly 104 of the RF coil apparatus can be configured to be located outside the body of the subject 112. A first connector 108 (e.g., a cable) of the RF coil apparatus 100 can be used to connect and couple the flexible RF coil 102 to the circuit assembly 104. For example, a coaxial cable (e.g., 50 Ω, 1 mm diameter, 20 cm in length) can be used to connect (or couples) the RF coil 102 to the circuitry assembly 104. In some embodiments, the circuit assembly 104 can include a housing and adjustable tune and match components 106. In some embodiments, the circuit assembly can also include a preamplifier (not shown). In some embodiments, the circuit assembly 104 can be a 3D printed circuit box. The first connector 108 can have a length that allows the circuit assembly 104 to be outside of and external to the subject 112 when the RF coil 102 is inserted into the sphenoid sinus cavity near the pituitary gland. In some embodiments, the first connector 108 is configured to enable the circuit assembly 104 (and tune and match components 106) to be positioned at multiples of a half wavelength away from the RF coil 102 so that the tune and match components 106 can have the same tune and match effects as if they were placed on the RF coil 102. For the example of pituitary MRI imaging as shown in
In some embodiments, the RF coil apparatus 100 configured for MR imaging of the pituitary gland can be used in a clinical implementation where the RF coil apparatus 100 may be used as part of the initial portion of an endoscopic endonasal operation, including the sphenoidotomy, drilling (removal) of intrasphenoidal septa, and other sphenoid bone if necessary (so called “conche” sella). In this example application, the anterior sellar bone can remain to protect the sellar contents per se. A saline soaked collagen sponge can be placed over the clival recess and reduce air-bone artifacts. The sterile RF coil (e.g., flexible RF coil 102) can then be placed through a single nostril of the subject and then positioned as much as possible parallel to the ground floor or scanner bed (which are both parallel to the B0 field). An intrasphnoid balloon can then be inflated to secure the RF coil 102 against the sellar face. The patient (or subject) can then undergo an intra-operative MRI scan with, for example, the immediate interpretation of the images to be able to identify a previously unidentifiable lesion.
The following example sets forth, in detail, ways in which the RF coil apparatus of the present disclosure was evaluated and ways in which the RF coil apparatus of the present disclosure may be used or implemented, and will enable one of ordinary skill in the art to more readily understand the principles thereof. The following example is presented by way of illustration and are not meant to be limiting in any way.
In this example study, the spatial distribution of the image SNR of the flexible RF coil (e.g., the RF coil 102 shown in
In this example study, the disclosed RF coil design was evaluated using a custom-built phantom which allowed the precise measurement of SNR. The ideal orientation of the RF coil 102 inside the subject may be parallel to the orientation of the main magnetic field, B0. As the surgical positioning for endoscopic surgery is supine, however, with the sphenoid sinus anatomy this RF coil orientation may not be anatomically possible in some cases. Therefore, in this example study, the phantom was designed to allow the study of the effect of coil angulation relative to the B0 field. Two aims of this example study were to 1) investigate spatial distribution of the image SNR for various coil rotation angles (θ) using a numerical simulation model and phantom experiments, and 2) test the feasibility of increased SNR within the pituitary gland based on simulated surgical placement results.
Modeling of MRI RF coils can be an important step in coil design and development. In this example study, a 3D coil model was developed in simulation software to study the magnetic field distribution of the RF coil 102. In this example study for pituitary imaging, a circular loop RF coil with a 20 mm inner diameter and a trace width of 3 mm was set up in the frequency domain. The RF coil was assigned as Perfect Electric Conductor surface and the current flowing in the coil was set to 1 A. The sample properties in the simulation were set up according to the material properties of the agar-carrageenan gel used in the phantom. For this finite element simulation, a maximum element size of 0.5 mm was used on the ROI, and the simulated fields from the RF coil, Bi fields, at each vertex were for post-processed using known post-processing software.
In this example study, the simulated amplitude of the effective transverse field B1xyeffective at the resonance frequency within the ROI was evaluated and then compared with the MRI scan results. The magnetic field components can be simulated with the RF coil plane parallel to the B0 field (θ=0°) as shown by diagram 130 in
where B1x, B1y, and B1z are the magnetic field components for the RF coil 102 (e.g., an Rf receiving coil) in x, y, z directions at θ=0°. In this example, B1x, B1y, and B1z remain constant during the rotation.
An experimental setup was designed for evaluation of the RF coil apparatus 100. In this example study, a 3D-printed phantom was designed and manufactured to include a cavity to roughly mimic the sphenoid sinus dimensions. A surrounding cylindrical jar allowed for easy rotation of the assembly effectively tilting the RF coil rotation angle relative to the B0 field. In this example study, the cavity of the phantom holds the RF coil and a resolution plate was placed directly on the outside of the cavity of the phantom to measure the SNR at the location where the pituitary gland for a subject would be. To assess imaging resolution, five holes, ranging from 1 mm to 2.8 mm in diameter (e.g., 1 mm, 1.6 mm, 2 mm, 2.4 mm, and 2.8 mm), were drilled into the resolution plate, for example, an acrylic plate (2.5 cm thick and 7.5 cm wide), which was attached under the cavity of the phantom. These five smaller holes can be used for visual demonstration. In addition, a center hole of 12.7 mm in diameter was drilled to provide sufficient volume for SNR measurements. The cavity of the phantom and the resolution plate were then fixed inside a transparent cylindrical plastic jar, parallel to the jar wall. In this example study, the plastic jar was chosen to have a similar size as a human head, 13 cm in diameter and 12 cm in height. The plastic jar was rested on a pair of 3D-printed supporters, so the jar was able to be rotated and set at the desired RF coil (and scan) angle. In this example study, outside of the cavity can surrounded by agar gel. The plastic jar, including the holes in the resolution plate, was filled with agar gel, which, for example, consists of distilled water, 1% agar powder, 2% Kappa carrageenan, and 22 μmol/kg of gadolinium contrast. The RF coil was placed inside the cavity of the phantom. In this example study, a portable vector network analyzer (VNA) was used to tune and match the RF coil after placing the RF coil inside the cavity of the phantom. The tune and match circuitry was connected to a pre-amplifier, and MRI scans were performed on the phantom.
In this example study, the tune and match of the flexible RF coil 102 (shown in
In this example study, the T1/T2 value of the agar phantom was measured to be 1250/64 ms, with T1/T2 map sequences. For example, standard resolution proton density Turbo Spin Echo (SD PD-TSE) sequences can be used to compare the flexible RF coil with a commercial 20-channel head coil, both quantitatively and qualitatively. In this example study, proton density was used because it is a direct measure of the maximum signal. For the example experiments to evaluate the RF coil apparatus, the two-dimensional (2D) SD PD-TSE sequence (e.g., echo time=9.1 ms; repetition time=3000 ms; refocusing angle=160°; bandwidth=250 Hz/pixel; acquisition matrix size=320×320×15; field of view=220×220×45 mm3; resolution=0.7×0.7×3 mm3; phase over sampling 0%; scan time 03:09 mm:ss; no parallel imaging) was scanned at 10 different coil rotation angles, ranging from 0° to 90°.
Given the expected higher SNR, in this example study the miniature flexible RF coil was also scanned with a 2D high-resolution proton density Turbo Spin Echo (HD PD-TSE) sequence at 0° and 60° coil angles. Images can be reconstructed from the frequency data directly via inverse Fast Fourier Transform (iFFT). In this example study, the HD PD-TSE scan (e.g., echo time=14 ms; repetition time=3000 ms; refocusing angle=160°; bandwidth=250 Hz/pixel; acquisition matrix size=320×320×35; field of view=64×64×25 mm3; resolution=0.2×0.2×0.7 mm3; phase over sampling 100%; scan time 06:21 mm:ss; no parallel imaging) was also performed on the commercial head coil using the same scan sequence. The commercial head coil images can be sum-of-square combined after coil reduction.
In this example study, for each angle, SNR measurements for the single-channel flexible RF coil were calculated from two repeated standard-resolution 2D PD-TSE scans. For this experiment, the region of interest (ROI) was divided into five cylindrical slices—each with 1 cm diameter and 3 mm thickness (e.g., as shown in
In this example study, SNR measurements were calculated with a known method for magnitude images of a single-coil array. SNR can be calculated as the ratio of signal and noise (SNR=S/σ). The signals can be measured as the mean intensity within the ROI, for Example, as given by:
where N is the number of samples and A is the pixel intensity. The noise can be measured as the background standard deviation on a signal-free region, for example, as given by:
In this example experiment, the signal-free region was selected within the acrylic plastic part of the resolution block.
For the 20-channel commercial head coil, the SNR was calculated based on Kellman's method for root-sum-of-squares magnitude combining images, which is the gold standard for multi-channel phased array coils. The scaled noise covariance matrix was calculated from averaging pixel SNR within ROI from two repeated standard-resolution 2D PD-TSE scans. In this example study, standard resolution proton density Turbo Spin Echo (PD-TSE) MRI scans were performed on the phantom for SNR measurements for both the miniature flexible RF coil and the commercial head coil, and as mentioned above, a coil simulation model was developed to characterize the performance of the miniature flexible RF coil. For evaluation of the flexible RF coil and comparison with the conventional head coil, the SNR maps and the amplitudes of the simulated effective transverse Bi field distributions were plotted for θ from 0° to 90°, at defined ROIs from 4.5 mm to 16.5 mm distance to the coil (as shown in
In this example study, S11 (the reflection coefficient) was recorded and then compared with the simulated S11 for the loaded and unloaded cases.
As the coil angle increases, the overall SNR and the amplitude of the B1xyeffective within the ROI can decrease. Because of the circular shape of the small flexible RF coil used in this example study, the magnetic field from the RF coil may not be uniform, and dead spots, where B1xyeffective drops to zero, were observed in the in-plane results. When the rotation angles increased from 0° to 90°, the dead spot gradually moved from the edge of the ROI to the center of the ROI in both experiment and simulation. In this example study, the simulated field distribution qualitatively matched with the scan experiment SNR maps.
In this example study, the effective transverse fields predicted by simulation fields were normalized at a single point, the mean B1xyeffective at θ=0° at 4.5 mm below the coil. By setting this one point equal to the experimentally measured SNR in this example study, it can be seen that the simulations of magnetic field amplitude tracks with the experimentally measured SNR, with an error of 1.1% 0.8%. As shown in
In this example study, the SNR improvement using the miniature flexible RF coil compared to a commercial head coil was estimated using the simulated effective field at θ=30°. The SNR improvement factors of the disclosed flexible RF coil compared to the commercial head coil can be estimated based on the mean SNR from the scan of the miniature RF coil and the commercial head coil. In this example study, the MR imaging of the pituitary gland had a 12 to 19 times SNR improvement compared to the commercial head coil at the region close to the flexible RF coil, and at least 3 times of SNR improvement at the region further away from the RF coil. In an example “worst-case-scenario with an RF coil rotation angle of 60 degrees and a ROI depth of 16.5 mm, the flexible RF coil still produced a 2-fold relative increase in SNR.
The increased SNR from the miniature flexible RF coil enabled a markedly higher resolution MR imaging compared to the commercial head coil. In this example study, The voxel size of the high-resolution sequence was approximately 1/50th of the standard-resolution. Because the SNR is proportional to the voxel size, the flexible RF coil can enable a much-increased spatial resolution of that currently used with standard 3T imaging. At this reduced voxel size, the inadequate SNR associated with the commercial coil was demonstrable. The example phantom study suggests that pituitary adenomas of 1 mm and smaller may be detectable using the disclosed miniature intrasphenoidal flexible RF coil.
In this example, study, multiple aspects of the electromagnetic behavior and performance of the flexible RF coil were accurately simulated using a simulation model. In some embodiments, the simulation model may be implemented using known simulation software. The simulation of the effective magnetic field aligned with the experimentally measured SNR across a clinically relevant range of coil angles and distance, both in-plane pixel-wise and through-plane. The consistency of these two groups of simulation data and experiment data validates both the numerical simulation model and SNR experiments. The SNR from repeated scans also had little difference demonstrating precise SNR measurements.
In the example study, the coil simulation model was validated with the phantom scan experiment, and the coil simulation model can be important in studying the interaction between the RF fields from the surface coil and the ROI or the phantom. As demonstrated in the example study, the simulated coil field can potentially be used to predict the SNR improvement of using the miniature flexible RF coil compared to a commercial RF coil (e.g., a commercial head coil). It can also allow the simulation of the performance of other RF coil designs with different diameters, geometries, resonance frequencies, and placement configurations, and therefore accelerate the development of improved RF coil designs. Furthermore, the simulation model can enable selection of an optimal flexible RF coil size and shape from a set of existing flexible RF coil designs based on the specific anatomy (e.g., the specific physical spatial constraints of the anatomy) of each subject.
The pulse sequence server 210 functions in response to instructions provided by the operator workstation 202 to operate a gradient system 218 and a radiofrequency (“RF”) system 220. Gradient waveforms for performing a prescribed scan are produced and applied to the gradient system 218, which then excites gradient coils in an assembly 222 to produce the magnetic field gradients Gx, Gy, and Gz that are used for spatially encoding magnetic resonance signals. The gradient coil assembly 222 forms part of a magnet assembly 224 that includes a polarizing magnet 226 and a whole-body RF coil 228. A subject 250 (e.g., a patient) may be positioned in the magnet assembly 224 on a patient table (or scanner table) 252.
RF waveforms are applied by the RF system 220 to the RF coil 228, or a separate local coil to perform the prescribed magnetic resonance pulse sequence. Responsive magnetic resonance signals detected by the RF coil 228, or a separate local coil (e.g., the flexible RF coil 202 shown in
The RF system 220 also includes one or more RF receiver channels. An RF receiver channel includes an RF preamplifier that amplifies the magnetic resonance signal received by the coil 228 to which it is connected, and a detector that detects and digitizes the I and Q quadrature components of the received magnetic resonance signal. The magnitude of the received magnetic resonance signal may, therefore, be determined at a sampled point by the square root of the sum of the squares of the I and Q components:
and the phase of the received magnetic resonance signal may also be determined according to the following relationship:
The pulse sequence server 210 may receive patient data from a physiological acquisition controller 230. By way of example, the physiological acquisition controller 230 may receive signals from a number of different sensors connected to the patient, including electrocardiograph (“ECG”) signals from electrodes, or respiratory signals from a respiratory bellows or other respiratory monitoring devices. These signals may be used by the pulse sequence server 210 to synchronize, or “gate,” the performance of the scan with the subject's heart beat or respiration.
The pulse sequence server 210 may also connect to a scan room interface circuit 232 that receives signals from various sensors associated with the condition of the patient and the magnet system. Through the scan room interface circuit 232, a patient positioning system 234 can receive commands to move the patient to desired positions during the scan.
The digitized magnetic resonance signal samples produced by the RF system 220 are received by the data acquisition server 212. The data acquisition server 212 operates in response to instructions downloaded from the operator workstation 202 to receive the real-time magnetic resonance data and provide buffer storage, so that data is not lost by data overrun. In some scans, the data acquisition server 212 passes the acquired magnetic resonance data to the data processor server 214. In scans that require information derived from acquired magnetic resonance data to control the further performance of the scan, the data acquisition server 212 may be programmed to produce such information and convey it to the pulse sequence server 210. For example, during pre-scans, magnetic resonance data may be acquired and used to calibrate the pulse sequence performed by the pulse sequence server 210. As another example, navigator signals may be acquired and used to adjust the operating parameters of the RF system 220 or the gradient system 218, or to control the view order in which k-space is sampled. In still another example, the data acquisition server 212 may also process magnetic resonance signals used to detect the arrival of a contrast agent in a magnetic resonance angiography (“MRA”) scan. For example, the data acquisition server 212 may acquire magnetic resonance data and processes it in real-time to produce information that is used to control the scan.
The digitized magnetic resonance signal samples produced by the RF system 220 are received by the data acquisition server 212. The data acquisition server 212 operates in response to instructions downloaded from the operator workstation 202 to receive the real-time magnetic resonance data and provide buffer storage, so that data is not lost by data overrun. In some scans, the data acquisition server 212 passes the acquired magnetic resonance data to the data processor server 214. In scans that require information derived from acquired magnetic resonance data to control the further performance of the scan, the data acquisition server 212 may be programmed to produce such information and convey it to the pulse sequence server 210. For example, during pre-scans, magnetic resonance data may be acquired and used to calibrate the pulse sequence performed by the pulse sequence server 210. As another example, navigator signals may be acquired and used to adjust the operating parameters of the RF system 220 or the gradient system 218, or to control the view order in which k-space is sampled. In still another example, the data acquisition server 212 may also process magnetic resonance signals used to detect the arrival of a contrast agent in a magnetic resonance angiography (“MRA”) scan. For example, the data acquisition server 212 may acquire magnetic resonance data and processes it in real-time to produce information that is used to control the scan.
Images reconstructed by the data processing server 214 are conveyed back to the operator workstation 202 for storage. Real-time images may be stored in a data base memory cache, from which they may be output to operator display 204 or a display 236. Batch mode images or selected real time images may be stored in a host database on disc storage 238. When such images have been reconstructed and transferred to storage, the data processing server 214 may notify the data store server 216 on the operator workstation 202. The operator workstation 202 may be used by an operator to archive the images, produce films, or send the images via a network to other facilities.
The MRI system 200 may also include one or more networked workstations 242. For example, a networked workstation 242 may include a display 244, one or more input devices 246 (e.g., a keyboard, a mouse), and a processor 248. The networked workstation 242 may be located within the same facility as the operator workstation 202, or in a different facility, such as a different healthcare institution or clinic.
The networked workstation 242 may gain remote access to the data processing server 214 or data store server 216 via the communication system 240. Accordingly, multiple networked workstations 242 may have access to the data processing server 214 and the data store server 216. In this manner, magnetic resonance data, reconstructed images, or other data may be exchanged between the data processing server 214 or the data store server 216 and the networked workstations 242, such that the data or images may be remotely processed by a networked workstation 242.
The present invention has been described in terms of one or more preferred embodiments, and it should be appreciated that many equivalents, alternatives, variations, and modifications, aside from those expressly stated, are possible and within the scope of the invention.
This application is based on, claims priority to, and incorporates herein by reference in its entirety U.S. Ser. No. 63/296,958 filed Jan. 6, 2022 and entitled “Flexible Radiofrequency (RF) Coil For Small Field-Of-View Magnetic Resonance Imaging (MRI).”
Filing Document | Filing Date | Country | Kind |
---|---|---|---|
PCT/US2023/060246 | 1/6/2023 | WO |
Number | Date | Country | |
---|---|---|---|
63296958 | Jan 2022 | US |