The invention described herein relates to flow chamber assemblies and systems and methods of using the same. More specifically, the invention relates to a flow chamber assembly that allows cells or biologic, reagents to be subjected to laminar flow conditions and to provide a predetermined level of shear stress to the cells or biological reagents. Suitably the flow chamber assembly is suitable for use with a microscope to allow viewing of cells or biological reagents under flow conditions and allows the cells or biological reagents to be collected or harvested from the flow chamber assembly after exposure to flow conditions.
Many cells in the human body are constantly subjected to fluid shear stresses. Endothelial cells, for example, line the inner surface of blood vessels and are exposed to fluid flow. in order to study the differentiation and behavior of such cells under well quantifiable fluid shear stresses and different conditions, flow chambers are needed. The flow chamber design described herein emerged from a need to subject adherent cells or other biological reagents to laminar flow conditions and evaluate their response to well-quantifiable fluid shear stresses. While some flow chambers are currently available, they all have various limitations.
Initially flow chambers were designed by compressing a gasket between two parallel flat plates (GlycoTech); however, the height of the gasket changes based on the force applied to the plates. Therefore the height varies with every experiment and the shear stress is not constant or quantifiable. Similar designs (FlexFlow™) further require the use of vacuum pumps to prevent leakage of fluid during a flow experiment. These systems are further limited in that only one experimental condition can be studied per experiment. The Flexcell Streamer® was created in an attempt to overcome the latter problem. It allows several glass slides to be exposed to the same shear stress at once but lacks a viewing window for real-time visualization of cells under flow.
More recently, microfluidic flow chambers have been designed in an effort to study cells under many different conditions per single flow experiment. Ibidi's® multi-channel slide can be used for multiple flow assays in parallel and its channel contents can be visualized in real-time.
However, thee flow channels require a microscope for read-out and are not compatible with microtiter plate readers.
CellAsic engineered MiCA™ plates, which integrate their microfluidic channels in the middle of three consecutive wells of microtiter plates. In its design, the outer two wells serve as inflow and outflow reservoirs, respectively. Since gravity is responsible for the perfusion of the culture chamber in MiCA™ plates, the shear stress cannot be varied. Therefore it is not possible to model arterial, venous and capillary fluid shear stresses.
Moreover, none of these microfluidic devices are suitable to harvest a sufficiently large number of cells for DNA or RNA purification experiments after exposure to flow. They are also not convenient to use for the purification of specific proteins expressed in cells after flow, e.g. to perform a Western blot, because the microfluidic channels do not harbor sufficiently large numbers of cells. Furthermore, these devices are not suitable as replacement for cell culture plastic ware and therefore routinely require disposable plastic flasks for culturing cells prior to flow experiments. It would also be difficult to mass-produce these devices at low cost, which limits their use as a more universal research tool in life sciences. Lastly, these devices cannot function without external components, such as external pumps, to subject cells to different shear stresses, and are therefore not practical to utilize for automated processes, e.g. robotics.
A flow chamber assembly for subjecting cells or other biological reagents to laminar flow conditions and methods of using the flow chamber assembly are provided herein. The flow chamber assembly includes a bottom plate having at least one well with a bottom surface adapted to receive the cells or biological reagents, a top plate having at least one flow protrusion positioned and shaped to fit into the well of the bottom plate and a sealing element positioned between the top plate and the bottom plate when the top plate and the bottom plate are removably attached. The flow chamber assembly further includes a flow path comprising a fluid feeding channel, an inflow bay, a laminar flow section, an outflow bay and a fluid exit channel all of which are in fluid communication with each other as described and shown herein. In an alternative embodiment the top plate and the bottom plate may form a unitary construction or be attached such that the top plate and bottom plate once attached are not removably attached.
The laminar flow section of the flow chamber assembly is suitable for subjecting the cells or biological reagents to laminar flow conditions when the top plate is attached to the bottom plate and the flow chamber assembly is in operation. The flow chamber assembly is constructed such that the width, height and hydraulic diameter of the laminar flow section are known and fixed to allow for laminar flow. The flow rate, fluid density and viscosity can be altered to provide a calculated shear stress to the cells or biological reagents in the laminar flow section of the flow chamber assembly. In addition, the bottom surface of the well and the lower surface of the flow protrusion may be made of an optically clear material such that the cells or biological reagents can be monitored optically using microscopes or microtiter plate readers. The flow chamber assembly provided herein may comprise more than one set of paired wells and flow protrusions connected in series or in parallel to ensure equivalent treatment of more than one sample.
In another aspect, a top plate for a flow chamber assembly is provided. The top plate is capable of being removably attached to a bottom plate comprising wells to form a flow chamber assembly. The top plate includes at least one flow protrusion positioned and shaped to fit into the well of the bottom plate, and a flow path comprising a fluid feeding channel, an inflow funnel, an outflow funnel, and a fluid exit channel.
In another aspect, systems for subjecting cells or biological reagents to laminar flow conditions to provide a predetermined level of shear stress to the cells or biological reagents are provided. The systems include the flow chamber assembly described herein connected to an external pump capable of pumping a perfusate through the flow chamber assembly and a reservoir connected in series to the flow chamber assembly and the pump. An alternative system includes a flow chamber assembly, which includes within the flow chamber assembly a pumping mechanism capable of pumping a perfusate through the flow chamber assembly and a reservoir, Thus, the flow chamber assembly can be used without external connections to the flow path. The pumping mechanism may he driven by an external motor, may be part of or positioned within the flow chamber assembly or may be external to the flow chamber assembly.
In a further aspect, methods of using the flow chamber assemblies and the systems described herein are provided. The methods include adding perfusate including the cells or biological reagents to be analyzed to the well of the bottom plate and attaching the top plate to the bottom plate of the flow chamber assembly. Then the perfusate is pumped from a reservoir through the fluid feeding channel into the inflow hay, through the laminar flow section and back out through the outflow bay and the fluid exit channel to achieve laminar flow in the laminar flow section of the flow path. The methods allow the cells or biological reagents to be exposed to a predetermined shear stress and biological properties of the cells or biological reagents to be assayed.
The bottom plate 14 can be utilized for cell culture with a traditional microtiter top plate and thus can replace use of tissue-culture flasks to prepare cells before exposure to flow conditions. Thus the flow chamber assembly provided herein will reduce the steps in the routine work-flow. The flow chamber assembly is designed to be mass-produced at low cost via injection molding. Other advantages of one or more aspects will be apparent from a consideration of the drawings and ensuing description.
With reference to
The sealing element, shown as a gasket 20 in
In
The port 18 (including both the inlet port and the outlet port) is designed to connect to the fluid feeding channels 32 on the outside surface of the top plate 12 (FIG. 2 and 3-5). It allows for connection of flexible tubing to allow entry of the perfusate into the flow chamber assembly.
Since the height of the flow chamber assembly may be restricted such that the assembled chamber fits under a condenser of a microscope on a microscope stage, the fluid feeding channel opposite to the port in this embodiment is <0.7 cm deep. To maintain the condition that the major resistance in the flow chamber assembly shall occur in the wells 22 in the laminar flow section, the area opposite the port may be wider to compensate for the decrease in depth, e.g. 0.2 cm wide and 0.4 cm deep. This area is denoted as basins 36 in
With reference to
The bottom plate 14 is of such geometry and dimensions that it can be inserted into standard microtiter plate readers. The wells 22 are arranged in such pattern that the light beams of a microtiter plate reader transverses the wells. For example, the wells may be spaced such that the well center to well center distance is a multiple of 9 mm consistent with use in a 96 well microtiter plate reader. Microtiter plate readers can detect biological, chemical and physical events of samples in the array of wells in the bottom plate of the high-throughput multi-well flow chamber assembly. The high-throughput flow chamber assembly may be used for assays that are based on but not limited to the detection of and quantification of time-resolved fluorescence energy transfer, time-resolved fluorescence, fluorescence resonance energy transfer, fluorescence intensity, fluorescence polarization, bioluminescence resonance energy transfer, luminescence detection, spectrometer absorbance, alpha screen, simultaneous dual emission, bottom reading, etc. Therefore, the bottom plate 14 of the high-throughput flow chamber assembly can be used analogous to a standard microtiter plate.
The bottom plate 14 may further comprise bumpers 30 and a rim 28, which define an elevation over the remaining surface of the bottom plate. In one embodiment, this remaining surface of the bottom plate is covered with a gasket 20 as shown in
The top plate 12 and the bottom plate 14 may be made such that the orientation of the top plate relative to the bottom plate is maintained. This directionality may be achieved through various means for example, one or more of the threaded holes for receiving a screw during attachment of the top plate to the bottom plate may be offset as compared to the others so that to align the threaded hole in the bottom plate with the hole in the top plate the top and bottom plate must be oriented to each other in the same configuration. In this embodiment, if the top plate was rotated by 180 degrees with respect to the bottom plate the holes would not align and the screws could not be used to affect attachment of the top plate and the bottom plate. In an alternative embodiment, the edge or corner of one side of the top plate may be rounded or have a shape distinct from the other edges or corners and the bottom plate has a corresponding rounded edge or corner such that the two plates only fit together in one orientation. This strategy for maintaining directionality of fit between a bottom plate and a top plate is commonly found in commercially available microtiter plates. The gasket may also be fitted with a rounded corner or edge similar to that of the top and bottom plates. In another alternative, the wells 22 of the bottom plate and the flow protrusions 24 of the top plate may be configured within the plates such that the top plate flow protrusions 24 can only fit into the wells 22 of the bottom plate in one orientation. Finally, the gasket 20 may be designed with an opening and either the top plate or the bottom plate may have a bumper or pin that extends through the opening in the gasket and engages with a receiver on the opposite plate to allow for directionality in the orientation of the top plate and the bottom plate.
With reference now to
Initially, it is important to understand that in order to achieve laminar flow in the laminar flow section, the flow chamber assembly must be designed such that several important conditions are met. First, the flow must be laminar, which can be verified by calculating its Reynolds number (Re), which is the ratio of inertial forces to viscous forces. If viscous forces predominate, Re is small and the flow is laminar or ‘fully developed’—usually for Re <2300. When inertial forces predominate, the flow becomes more and more random until it is turbulent, as is the case for Re>4000. We can calculate Re according to equation 1:
Where ρ is the fluid density, Q is the flow rate, μ is the viscosity, w and h are the width and height of the chamber, respectively, and Dh is the hydraulic diameter, defined according to equation 2:
Second, for the velocity field and shear stress to be independent of the distance along the laminar flow section (i.e. fully developed), the distance from the inlet bay to the laminar flow section must be longer than the entrance length 50, Le, shown in
Le=0.04 hRe
Thus, a section of the bottom surface of the well is not part of the laminar flow section of the flow chamber assembly, but is the entrance length and the length can be calculated using the above formula.
The portion of the bottom surface of the well in which the flow is not laminar can be excluded during analysis of cells or biological reagents using microscopes or microtiter plate readers or can be excluded when adding cells or other biological reagents to the wells. Several other alternative strategies may be employed to avoid a possible confounding effect of non-laminar or not fully developed flow on adherent cells at the bottom of each well. The flow chamber can he constructed in such a fashion that the total length of the well, L, is much larger than the inflow length, Le (L>>Le). In the flow chamber of
Third, in order to ensure that the velocity and shear stress in the lateral direction do not vary significantly from the value for one-dimensional channel flow (ΔPh/2L), the ratio h/w must be much less than 1. For the average wall shear stress under two-dimensional flow conditions to be 95% of the wall shear stress under one-dimensional flow, h/w must be equal to 0.10, and for the wall shear stress under two-dimensional flow conditions to be >95% of the wall shear stress under one-dimensional flow, h/w must be equal to <0.1.
As an example these conditions are met for the following dimensions. Each individual well is 0.8 cm wide (w=0.8 cm), 1 cm deep and 2.7 cm long. The protrusions are 0.79 cm wide, 2.5 cm long and 1.05 cm tall, When the flow chamber is closed the top plate rests against the rim and bumpers of the bottom plate and a flow path of height h=0.05 cm is established. In this embodiment the bottom plate of the flow chamber is 8.5 cm wide and 12.75 cm long. It has 12 wells arranged in pairs of two. Each pair of wells is spaced apart by 1 cm. Each of these spaces is endowed with a 0.4 cm wide, 2.7 cm long and 0.1 cm tall bumper.
The flow Q through a single well, that is necessary to exert a shear stress of 15 dynes/cm2 acting tangentially on cells adherent to the bottom of the well, can be calculated according to equation (4)
where Q is the desired flow rate, τ is the target shear stress acting tangentially on the cells (15 dynes/cm2), w is the width of the well (0.8 cm), h is the height of the flow path (0.05 cm), and μ is the viscosity of the perfusate (flow medium). Assuming a typical value of viscosity (μ) for medium as 0.9 cP (0.009 g·cm−1·s−1), Q in our example is =0.56 ml/s or 34 ml/min. Since the hydraulic diameter is defined according to equation (2), Dh is =0.094 cm. Therefore the Reynolds number, Re, according to equation (1) is ˜145.27, indicating fully-developed or laminar flow after the inlet length, Le. In our rectangular well Le can be calculated according to equation (3) and is =0.29 cm in this example, Since L>>Le a possible confounding effect of non-laminar or not fully developed flow on adherent cells at the bottom of each well can be avoided.
It should be understood from the above description that the specific dimensions can be varied, as long as the specified conditions described above defining laminar flow inside the wells arc met. In one embodiment, for example, the height h of the flow path may be reduced to 0.04 cm, 0.03 cm, 0.02 cm, etc. Further the viscosity μ of the medium may be altered by adding dextran to cell medium in order to decrease the flow Q necessary to achieve a desired shear stress.
A flow chamber assembly may be constructed with different numbers of wells and flow protrusions. For example, the wells may be designed with 2, 3, 4, 6, 8, 12, 16, 20, 24, 30, 36, 48 or more wells. The flow chamber assembly shown in
The wells can be perfused separately, such that each well is connected to its own flow path. Alternatively, wells can be arranged in parallel (as shown in
Other arrangements are possible. For example, three wells may be arranged in parallel as schematically depicted in
In another embodiment, wells may be arranged in series, such that a compound has to transverse the first well before it can transverse the second well, etc. In this embodiment, one drug may be studied as it interacts with cells in the first well, e,g. is metabolized by hepatocytes seeded into the first well, then reaches cancer cells, which the drug is supposed to affect, and lastly reaches kidney cells to study potential nephrotoxic effects.
In designing flow chamber assemblies with various configurations of flow paths, the geometry of the fluid feeding channels must be considered. The fluid feeding channels may be round or rectangular and of such geometry that the magnitude of flow of perfusate into each well is equal. The feeding channels should be large enough so that the wells in the closed chamber represent the major resistance to the flow. In one embodiment the fluid feeding channels in the top plate are 0.1 cm wide and 0.7 cm deep. Such dimensions represent approximately 4% of the resistance to the flow through the individual wells. Therefore the above condition is satisfied.
In an alternative embodiment such as that shown in
Q
0
=Q
1
+Q
2
+Q
3 (5)
Assuming a round fluid feeding channel, the radii of the bifurcation of the fluid feeding channel can be adjusted such that 1/3rd of the flow is directed to one well and 2/3rds of the flow is directed to the other two wells.
Let the radius of the inlet fluid feeding channel be R0, the radius of the channel directing fluid to one well be R1, and the radius directing flow to two other wells be R2,3. Since the flow through the inlet fluid feeding channel, Q0, equals the area of the inlet fluid feeding channel it traverses with velocity v0, we can write:
Q0=πR02v0
(a)
Further, we can express the flow in the different channels as:
Q
0
=Q
1
+Q
2,3 (b)
And
Q1=⅓Q0 (c)
And
Q2,3=⅔Q0 (d)
We assume the velocities to be equal in the two branch points, such that:
v1
=⅓
v0
(e)
and
v1
=
v2,3
=⅓
v0
(f)
Since Q1=⅓Q0, it follows that πR12v1
=⅓πR02
v0
, and therefore, that the radius in the inflow fluid feeding channel is equal to its first branch point (
R1=R0 (g)
Since Q2,3=⅔Q0, it follows that πR2,32v2,3
=π⅔R02
v0
and R2,32⅓
v0
2/3R02
v0
.
Therefore, the radius in the 2nd branch point is 1,4-times the radius of the inlet fluid feeding channel:
R2,3=√{square root over (2)}R0 (h)
Note that the inflow bay into the wells may be constructed such that the flow fans out from the fluid feeding channel 32 into the well in the funnel section 34 of the inlet hay in order to minimize any ‘jetting’ of the flow at that point (see 34 in
Similar computational analyses can be performed for fluid feeding channels of square geometry where w0 is the inlet channel width and h0 the inlet channel height:
Q0=w0h0v0
(i)
Assuming the height is constant throughout such that
h0=h1=h2,3 (j)
and the velocities are equal in the inlet fluid feeding channel and two branch points,
v1
=
v2,3
=
v0
(k)
then
w1=⅓w0 (l)
And
w2=⅔w0 (m)
Note that with similar computational analyses, the geometry in channels with multiple branch points can be described such that each well in a multi-well flow chamber assembly receives the same magnitude of flow.
In an alternative embodiment, shown in
If the mixing chamber 56 is designed to function as a pulse dampener, its size can be constructed such that it stores a small amount of perfusate, dV, and has a total volume, VT. Its purpose is to dampen the puke generated by a pump that drives the perfusate through the flow chamber assembly. As an example, we can designate the minimum pressure in our circuit as PA and the maximum pressure at PB. The relationship between the pressures and volumes in the pulse dampener can be expressed with Boyle-Mariotte's law as:
PAVA=PBVB=PTVT (n)
Further assuming that
VA=0.9 VT, (o)
it follows that
PT=0.9 PA (p)
And assuming that
V
B
=V
A
−dV (q)
we can define the relationship between the total volume of the mixing chamber, the maximal and minimal pressures and the amount of perfusate stored in the chamber as:
Alternatively, to ‘build-in’ pulse dampeners, the flow chamber assembly may be connected via tubing with separate external pulse dampeners, which are connected in series to the inflow tubing running into the flow chamber assembly as described more fully below.
In an alternative embodiment, a pulse resembling the heartbeat may be desirable. This can be achieved by interposing a pulse wave generator into a laminar flow circuit or by directly generating a pulse wave of defined frequency and amplitude with a microprocessor controlled pump.
In order to ensure that the flow chamber assembly does not leak during flow, a variety of leak-proof closure mechanisms may be utilized. A snap seal design can be constructed by providing grooves 54 on either the top or bottom plate of the flow chamber assembly with interlocking teeth 56 that fit into corresponding grooves 54 and are located on the opposite plate of the flow chamber assembly. For example, a groove 56 in the top plate may fit into a corresponding tooth 54 in the bottom plate (
An alternative method of sealing the flow chamber assembly is to utilize a vacuum seal. For this purpose, a vacuum line, which is commonly available in laboratories, may be connected to either the top Or bottom plate of the flow chamber assembly. This will provide a negative pressure, which will keep the components of the flow chamber assembly tightly held together.
A screw closure device as described in relation to
As noted above, the top plate 12 and the bottom plate 14 are suitably made out of transparent material. Suitably, the top plate and the bottom plate are constructed of a plastic material, In one embodiment this material is grade 1A optically-clear polystyrene. Other suitable materials may be used: The protrusions are carved out as shown in at least
It may be necessary to use multiple components, which are molded separately, e.g. the top plate differently from the bottom plate. The bottom of the flow chamber or individual wells might be assembled separately by inserting glass cover slips as viewing windows compatible with inverted microscopes. Other manufacturing processes that may be applied include hot embossing, soft lithography, casting, ultraviolet embossing, vacuum forming, laser manufacturing and micromachining techniques.
The flow chamber can be connected to a flow circuit consisting of a reservoir 100, pump 102, and flow chamber assembly 10 as depicted in
For the flow chamber assembly described in
The various parts of the flow circuit may be connected with microbore tubing. One piece of tubing 114 may be used to connect the flow chamber assembly 10 to the inlet 106 of the reservoir 100. A second piece of tubing 118 may be used to connect the outflow 108 of the reservoir 100 to the flow chamber assembly 10. The tubing may contain one or more stopcocks 116 or other means to control the flow of the perfusate or to gain access to the perfusate to allow sampling of the perfusate during a flow experiment. A peristaltic pump 102 can force the perfusate through the flow circuit. Other pumps known in the art may also be used in the flow circuit shown in
In one embodiment of the flow chamber assembly shown in
The basins 36 described above, may be modified to act as internal reservoirs for the perfusate or medium may be located between wells along the flow path. Such basins 36 may be endowed with a through-hole or valve 60 to enable filling the assembled flow chamber with perfusate. In one application a robot may pipet a solution or drug into the assembled flow chamber assembly through the valve 60. In another application of this embodiment, the through-hole or valve 60 can be utilized to obtain perfusate samples at predetermined time points, e.g. via a robotically driven arm with a needle, where said needle is inserted into the through-hole or valve 60 to aspirate a small amount of perfusate for analysis of metabolites. The through-hole or valve 60 can also enable gas exchange with the outside environment if the chamber is placed inside a CO2 incubator.
In one embodiment, a software program is utilized to control the speed of the magnetic centrifugal pumps. By modulating the rotation speed and the ‘on,’ ‘off’ frequency, different flow conditions can be recreated. These include, but are not limited to, physiological or pathological heart rhythms.
In one embodiment, the fluid feeding channels 32 of separate flow circuits are connected to create more complex circuits. One example is illustrated in
In another embodiment, the rotation of a motor contained within the flow chamber assembly is transmitted to the impeller 58 via direct mechanical coupling. A small output shaft may penetrate the bottom plate housing of the motor at a predetermined area and engage the impeller. A small rubber washer can provide a seal around the output shaft in order to prevent perfusate from leaking into the motor containing compartment. The impeller can be designed with a tapered opening in its central area such that an output shaft with a predetermined geometry slides into an orientation until it engages the impeller without slipping. Such mechanism will ensure that multiple impellers can be engaged to multiple output shafts simultaneously when pressing the impeller-containing plate onto a motor containing board.
In an alternative, but related embodiment shown in
The high throughput flow chamber assembly can be utilized for DNA or RNA or protein quantification of cells after exposure to shear stress. It also can be used for, but is not limited to, high-throughput drug development or screening studies, toxicological assays, chemical screening studies, screening compounds for impurities, dose-response studies, drug dissolution profile generation, drug target development, drug discovery research, evaluation of pathological states, antimicrobial compound screening, microbial adhesion studies, the evaluation of biofilm production and/or bacterial growth under different drug conditions or combinations of drugs or compounds or types of compounds under the influence of fluid shear stress, viral studies, viral infection studies, cell differentiation studies, cell morphology analyses, shear stress-depending cell differentiation studies, matrix invasion, transmigration studies, cell growth and proliferation studies, cell migration assays, basement membrane formation and degradation research, cell-cell interaction assays, cellular adhesion strength quantification, cell spreading studies, dynamic cell adhesion studies, cell rolling studies, cell-ligand interaction studies, protein-cell interaction assays, protein-protein interaction assays, including antibody binding assays, signal transduction studies, cell-signal pathway analysis, ion-flux studies, ELISAs, apoptosis assays, angiogenesis research, vasculogenesis research, cancer research, wound healing assays, thrombosis research (including platelet adhesion studies and platelet aggregation studies), further nucleofection studies, transfection studies, molecular profiling, immunoassays, lipophilicity assays, enzyme kinetic studies, kinetic turbidimetric analyses, NADH and NADPH quantification, membrane fluidity studies, colorimetric assays, solubility assays, as well as other specific applications in the fields of cardiovascular research, immunology research, and parasitology. Performance of these assays will be completed using cells or other biological reagents such as proteins, antibodies, metabolites, platelets, cellular membranes and the like.
Note that while we have described a specific arrangement of the wells in the above examples, the wells may be arranged in different patterns. Similarly, the wells may have different dimensions and geometry and may be grouped in different combinations. Therefore, the flow chamber may be designed with either one inflow and outflow channel only, or with a multitude of separate inflow and outflow channels that may correspond to a multitude of different flow circuits.
While the invention has been described in connection with what is presently considered to be the most practical and preferred embodiment, it is to be understood that the invention is not to be limited to the disclosed embodiment, but on the contrary, is intended to cover various modifications and equivalent arrangements. The following examples are meant only to be illustrative and are not meant as limitations on the scope of the invention or of the appended claims. All references cited herein are hereby incorporated by reference in their entireties.
The flow chamber assembly may be used for culture of cells prior to a flow experiment. The sterile-packaged bottom plate of the flow chamber assembly is placed under a laminar flow hood and the package material removed. Adherent cells, e.g. endothelial progenitor cells, are pipetted into the rectangular wells of the bottom plate using sterile technique (
The layer of cells in the bottom wells are subjected to laminar flow fluid shear stress. The sterile-packaged top plate and gasket of the flow chamber are placed under a laminar flow hood and the package material removed. The gasket is placed onto the bottom plate in the depression between rim and bumpers using sterile gloves or forceps (
The assembled flow chamber assembly is inserted into a flow circuit, which in one embodiment comprises of a pump, tubing, pulse dampeners, fluid reservoirs and optionally a heating plate. In one embodiment of the flow chamber shown in
The desired shear stress acting tangentially on the cell layer inside the wells is chosen by programming the pump at a predetermined flow rate Q (for a specific width w of the wells and height h of the flow path inside the wells and viscosity μ of the perfusate). The viscosity of the perfusate is optionally increased to a predetermined value by the addition of dextran, which reduces the flow rate necessary to achieve a predetermined shear stress (see equation (4)). The flow chamber may be maintained at 37° C. by placing it onto a heating plate, and the pH of the perfusate is held constant by using a perfusate that at least in part consists of 4-(2-hydroxyethyl)-1-piperazineethanesulfonic acid (HEPES) buffer. Alternatively, the flow chamber assembly may be placed inside an incubator at 37° C. and 5% CO2.
Operation of Flow Chamber Assembly with Plate Reader
The bottom part of the flow chamber is suitably compatible with microtiter plate readers and enables measurement of physical, chemical, or biological events inside the wells. Bioassays may be performed inside the wells after the exposure of cells to physiological fluid shear stresses and to drugs or toxic agents for toxicity studies (
Operation of Flow Chamber Assembly for Read-out with Microscope
The fully assembled flow chamber assembly may be placed onto a microscope stage while it is being perfused in a flow circuit (
The wells in the bottom plate are constructed out of transparent material of such thickness that a standard 20× or 40× microscope objective (without oil immersion) allows visualization of adherent cells inside that well. The inside surfaces of the wells can be visualized with bright field illumination, phase contrast microscopy or fluorescent microscopy.
A microscope may be outfitted with an automatic stage that, when combined with a suitable software program, may scan the individual wells in the bottom plate of the flow chamber assembly in a predetermined matter to obtain images either before, after, or at various time points during flow experiments. Following this operation, the flow chamber assembly allows for time-lapse microscopy and live-cell imaging during a flow experiment. For use with upright microscopes, the flow chamber assembly may simply be inverted.
In contrast to currently available microfluidic chambers, large numbers of cells may be harvested after flow exposure for analytical techniques, such as reverse transcription polymerase chain reaction (RT-PCR). Cells may be exposed to predetermined shear stresses and/or drugs. Total RNA is isolated for gene expression analysis (
Cells may be exposed to predetermined flow conditions. After exposure to flow, cells may be lysed directly inside the wells of the bottom plate. Protease inhibitors are then added to prevent the digestion of proteins by cells own enzymes. A Western blot or other protein analysis assay can then be performed to detect proteins of interest in the samples (
The multi-well flow chamber assembly described herein may also be used for drug development or drug toxicity studies by adding different drugs or toxins to the flow circuits and testing cells' responses to agents of interest. Bioassays may be used in conjunction with a microliter plate reader for instantaneous read-out.
The flow chamber assembly may be utilized for the evaluation of biofilm production and/or bacterial growth under different drug conditions or fluid shear stresses. The individual circuits are filled with bacteria and test drugs and the formation of biofilms in the individual wells and circuits is evaluated with phase contrast microscopy.
The adhesion of platelets (PLTs) to endothelial cells or matrix proteins inside the wells may be quantified using time-lapse microscopy. The study variables include compounds that inhibit platelet function or compounds that activate endothelial cells. Furthermore, the effect of different shear stresses on PLT adhesion may be studied.
The wells of the bottom plate may be filled with matrigel and stem cells seeded into said matrigel-filled wells. The chamber is then closed and perfused with shear stresses mimicking capillary flow conditions, At predetermined time periods, the cells are imaged to study their morphological changes during flow. After a predetermined time period, the chamber is opened and a bioassay added to the wells. Following, the bottom plate is covered with a lid, the chamber incubated for a predetermined time period and then evaluated with a microtiter plate reader.
In addition to the examples of methods of using the flow chamber assemblies described above many other methods of using the flow chamber assembly will be apparent to those skilled in the art. The flow chamber assembly may be used for antimicrobial compound screening, microbial adhesion studies, chemical screening studies, screening compounds for impurities, dose-response studies, drug dissolution profile generation, drug target development, drug discovery research, evaluation of pathological states, viral studies, viral infection studies, matrix invasion studies, transmigration assays, cell growth and proliferation studies, cell migration assays, basement membrane formation and degradation research, cell-cell interaction assays, cellular adhesion strength quantification, cell spreading studies, dynamic cell adhesion studies, cell rolling studies, cell-ligand interaction studies, protein-cell interaction assays, protein-protein interaction assays, signal transduction studies, cell-signal pathway analysis, ion-flux studies, ELISAs, apoptosis assays, angiogenesis research, vasculogenesis research, cancer research, wound healing assays, thrombosis research, further nucleofection studies, transfection studies, molecular profiling, immunoassays, lipophilicity assays, enzyme kinetic studies, kinetic turbidimetric analyses, NADH and NADPH quantification, membrane fluidity studies, solubility assays, as well as other specific applications in the fields of cardiovascular research, immunology research, and parasitology.
This patent application claims the benefit of priority of U.S. Provisional Patent Application No. 61/568,252, filed Dec. 8, 2012, which is incorporated herein by reference in its entirety.
Filing Document | Filing Date | Country | Kind | 371c Date |
---|---|---|---|---|
PCT/US12/68781 | 12/10/2012 | WO | 00 | 6/9/2014 |
Number | Date | Country | |
---|---|---|---|
61568252 | Dec 2011 | US |