The field of the present invention pertains generally to fluorescence techniques used in the measurement of chemical interactions and chemical properties, including, more specifically, the measurement of fluorescence lifetime for determining the degree of binding of chemical substances or for determining the properties of the chemical environment of a substance.
Fluorescence techniques are known for studying chemical properties and processes. Fluorescence emission is a process in which a fluorophore is excited to a higher energy state by absorption of a photon at some excitation wavelength. The molecule decays via emission of a fluorescence photon on time-scales on the order of 10 μs to 1 μs. The fluorescence lifetime is defined as the average time the fluorophore spends in the excited state. Various scientific applications involve the measurement of fluorescence lifetime because it represents an intrinsic molecular property of the fluorophore and can be affected by small changes in the fluorophore's direct environment. For example, fluorescence lifetime measurements are used in high-throughput screening for drug discovery. If one substance is labeled with a fluorophore and combined in solution with a second substance, the lifetime of the fluorophore typically will change if the two substances interact. Affinity analysis can be performed by measuring the variation in lifetime with the relative concentration of the two substances. Alternatively, kinetic studies can be carried out by monitoring the lifetime as a function of time after the two substances are mixed. In some cases, the substances being studied may exhibit intrinsic fluorescence, thereby eliminating the need for labeling. One example is the study of protein-protein interactions using the intrinsic fluorescence of tryptophan, tyrosine, or phenylalanine, three aromatic amino acid residues contained in most proteins. Turconi, et al. give an overview of fluorescence lifetime techniques for drug discovery in “Developments in fluorescence lifetime-based analysis for ultra-HTS”, Drug Discovery Today, Vol. 6, No. 12 (Suppl.) 2001 and in the references therein.
Measurements of fluorescence lifetime have been carried out using either a time-domain or a frequency-domain technique. In the time-domain technique, the sample is typically excited with a pulse of light from a pulsed laser and the fluorescence light is measured using a detector with single-photon sensitivity. The detector measures the time delay between the excitation pulse and the first detected photon. The fluorescence lifetime distribution is usually determined by using many repeated pulses and building up a histogram of the measured time delays. Unfortunately, the pulsed laser sources and single-photon detectors are relatively expensive. Because detection is typically done at the single-photon level, it can require a significant amount of time to build-up enough data to approximate the fluorescence lifetime distribution. One disadvantage of the frequency-domain approach is that it is not a direct measurement of the fluorescence lifetime distribution. Rather, it provides an estimate of the mean lifetime based on the phase shift between a detected signal and the excitation signal. When the fluorophore exhibits multi-exponential time decay, extrapolation of the lifetime from the phase shift data is more difficult. Usually this requires measurements at more than one modulation frequency. In some cases, a complete measurement of the lifetime distribution yields evidence of particular chemical interactions that is not evident in a measurement of the mean lifetime alone. This data is not readily obtained with frequency-domain instrumentation. A further disadvantage of the frequency-domain approach is the need for accurate high-frequency analog electronics. An overview of both the time-domain and frequency-domain techniques can be found in the above-referenced article by Turconi, et al.
U.S. Pat. No. 5,565,982 discloses a time-resolved spectroscopy system using digital processing techniques and two low power, continuous wave light sources. The disclosed system requires two light transmitters of different wavelengths modulated with separate codes for interrogating a sample of interest. Properties of the sample are inferred by differential comparison of the return signals from each of the two light sources. It is undesirable to have two distinct light sources due to the cost and complexity involved. Furthermore, the noise level associated with a measurement made with two separate light sources will be higher than with a single source even if the codes used to drive the two sources are orthogonal.
A system and method capable of addressing these disadvantages while providing acceptable fluorescence lifetime measurements for whatever application the measurement is being used is needed.
The inventions presented herein provide for direct measurements of fluorescence lifetime using any light source modulated with a known digital pattern. A preferred system uses a low-power continuous-wave light source and low-cost detector. Preferably the measurement system is implemented with digital electronics. One embodiment of the system and methods disclosed comprises a continuous-wave light source modulated with a digital waveform for interrogating a sample, a photo-sensitive detector for measuring the fluorescence light from the sample, and electronics for sampling the detector output and performing a correlation of the output signal with the modulation waveform. Other embodiments include electronics and software for calculating the parameters of the fluorescence lifetime distribution from the measured correlation.
A functional block diagram of a preferred fluorescence measurement system 100 is depicted in
In the preferred fluorescence measurement system 100, the detection optics 6 preferably include a second 3 mm diameter fiber bundle located between the optical filter and the optical detector 7. The optical detector 7 converts the fluorescence optical waves 21 to an electronic signal. In the preferred fluorescence measurement system 100, the optical detector 7 is preferably a 0.5 mm-diameter silicon avalanche photodiode (APD) manufactured by Pacific Silicon Sensor. In other embodiments, the optical detector 7 may be a PIN photodiode, a photomultiplier tube, a charge-couple device, or other suitable photosensitive element. As previously stated, the optical detector 7 preferably converts detected fluorescence optical waves 21 into an electronic signal which is communicated to the detected signal conditioner 8. The detected signal conditioner 8 preferably formats the signal so it may be converted to discrete samples by an Analog to Digital (A/D) converter 9. The A/D converter 9 outputs a detected response signal 19. The detected response signal 19 is communicated to a signal detector 10, where it is preferably correlated with the electronic reference signal 17 to extract a sample transfer characteristic.
Information about the temporal properties of the photons is preferably calculated from the sample transfer characteristic. This information preferably includes such properties as the fluorescence lifetime. The estimate of fluorescence lifetime can preferably be used to estimate characteristics such as the degree of chemical binding or to infer properties of the chemical environment surrounding the fluorescing material.
Another embodiment of the fluorescence measurement system 100 includes an optical reference generator 22. The optical reference generator 22 preferably includes an optical splitter 12A or 12B that routes a portion of the modulated optical wave 20 to a secondary optical detector 13. The position of the optical splitter 12A or 12B can be either before or after the light delivery optics. The output of the secondary optical detector 13 is preferably routed to a secondary signal conditioner 14 whose output is communicated to a secondary A/D converter 15. The secondary A/D converter 15 preferably outputs a source reference signal 18 which can be correlated with the detected response 19 to extract the sample transfer characteristic. Using the source reference signal 18 as opposed to the electronic reference signal 17 allows the filtering of the temporal properties of the signal conditioner 2 and the modulated optical source 3 from the measured transfer characteristic.
The preferred hardware implementation of the A/D converter module and its interfaces to the signal detector 10 are shown in
The acquisition synchronizer 92 is preferably synchronized with an externally provided synchronization clock (SClk) 40 which is also preferably used to synchronize the signal generator 1. The signals CClk[1..N] are preferably generated within the acquisition synchronizer 92 and preferably have the same frequency as SClk 40 but are offset in phase from SClk 40 in N fixed increments of (360÷N)°, with the phase of CClk[1] set to the fixed offset of Z°. In the preferred system the internal clock generation capabilities of the Xilinx FPGA are used to implement the acquisition synchronizer 92 directly. The A/D converters 90 preferably perform their conversions in sync with the conversion clocks 96 such that they generate samples at N discrete sample times spread evenly throughout the fundamental sample interval defined by the period of SClk 40. The effective sample rate for the array of converters is preferably N times the rate defined by SClk 40. This process of using multiple A/D converters sampling out of phase to increase the effective sample rate is what we call parallel over-sampling. In the preferred fluorescence measuring system, parallel over-sampling results in an effective sample rate of 2 Gsamples/sec. The offset value Z allows the entire sample set to be offset by some phase from the synchronization clock 40. The acquisition synchronizer 92 preferably is configured such that the value of Z can be varied synchronously with the modulation frame, or with a block of frames called a frame block. This allows Z to follow a sequence of K values smaller than (360÷N)° such that on successive modulation frames/frame blocks the effective sampling phases (relative to the synchronization clock) take on K values intermediate to those created by the N conversion clocks in any given frame. In this case preferably the input signal at any given A/D converter 90 will be sampled at K discrete phases over K blocks. The detected response 19 is preferably assumed to be stationary with respect to the start of the code pattern block over that time interval. The preferred K discrete sampling phases correspond to K discrete sample times and the effective temporal resolution of the sampling process is preferably increased by a factor of K. This process is referred to as temporal over-sampling.
In the preferred fluorescence measuring system the value of Z is always zero and temporal over-sampling is achieved by adjusting the phase of the modulation as described below rather than by adjusting the phase of the A/D converter sampling. Preferably the FIFOs latch input data to the A/D converters 90 synchronously with the corresponding conversion clock 96. The FIFO 91 output data is preferably provided to the internal components of the signal detector 10 synchronously with the synchronization clock 40 such that all further processing is synchronized with the synchronization clock 40.
The preferred implementation of the Temporal Response Analysis Engine 11 are shown in
The functional blocks of the preferred signal generator 1 are shown in
The modulation signal 16 for both the LFSR 30 or pattern memory implementation is preferably buffered by an output buffer 35 to make the signals 16 more robust when driving external components. Timing for presentation of the code pattern bits is preferably controlled by a generation synchronizer 34 which preferably generates the master clock (MClk) 38 for the LFSR 30 and the address sequencer 33. The master clock 38 is preferably synchronized to a system synchronization clock (SClk) 40 which preferably controls both code pattern generation and response signal acquisition. MClk 38 preferably operates at the same frequency as SClk 40 but is preferably offset in phase by an amount specified by the phase input 39, which is preferably an externally programmable parameter. This phase offset allows the relative phase between the modulation signal 16 and the detected response 19 to be adjusted. If the phase is adjusted by some increment, (360÷K)°, at the end of each code pattern block or set of blocks the detected response resulting from the modulation signal will preferably be sampled at K discrete phases over K blocks. In this embodiment of the fluorescence measuring system as with the preferred embodiment, the detected response 19 is assumed to be stationary with respect to the start of the code pattern block over that time interval so that the K discrete sampling phases correspond to K discrete sample times and the effective temporal resolution of the sampling process is increased by a factor of K.
This temporal over-sampling is functionally equivalent to the technique described for temporal over-sampling in the A/D converter embodiment. In other embodiments the external phase specification may represent the phase increment rather than the absolute phase, and the generation synchronizer 34 may increment the phase internally.
The preferred implementation of the LFSR 30 is shown in
The preferred functional blocks for the signal detector 10 are shown in
The details of the preferred frame accumulator 50 or 51 are shown in
The details of the preferred frame correlator 55 is shown in
The geometric relationship shown in
A method for using the present invention for examining chemical binding is as follows. A substance that exhibits fluorescence is placed in a sample holder. The substance may naturally exhibit fluorescence or it may be a material that has been modified by the addition of a fluorescent label. The fluorescence from the sample is then measured as described above to obtain a temporal transfer characteristic. One or more additional substances is then added to the sample holder and allowed to interact with the first substance. The fluorescence is measured again. By comparing the temporal transfer characteristic obtained before the second material was added to that obtained after the material is added, one can estimate the change in the system caused by adding the second material. If the two materials interact, the width and/or shape of the measured temporal transfer characteristic typically will change. A binding curve is generated by measuring the change as a function of the relative concentrations of the two materials.
The present invention can also be used for monitoring the kinetics of chemical interactions. In this case, the output of the fluorescence measurement system is monitored continuously or at multiple discrete time intervals after the second substance is added to the sample. The kinetics of the interaction is determined by measuring the change in the temporal transfer characteristic as a function of time.
In one embodiment, the present invention was utilized to investigate the binding of biotin, a colorless crystalline vitamin of the vitamin B complex, to streptavidin, a protein that has a high affinity to biotin. The streptavidin was labeled with Cy5, a fluorescing dye that can be excited with wavelengths around 635 nm. The starting solution consisted of 1 μM concentration of streptavidin in a buffer solution. The light source was a Sanyo DL5038-21 635 nm diode laser. The photodetector was a 0.5 mm-diameter APD, part number AD500-1.3G-TO5 from Pacific Silicon Sensor. The Temporal Response Analysis Engine was implemented with a 2.5 Gsample/sec data acquisition card from Z-Tec for signal detection and a Tektronix DG2040 digital pattern generator for signal generation. Correlation calculations were carried out in software on a personal computer. The result of the correlation is the sample transfer characteristic. A change in the width of the sample transfer characteristic is a direct measure of the change in fluorescence lifetime. The laser was modulated at a bit rate of 125 Mb/sec with a 31-bit PRBS code. Measurements of the sample transfer characteristic width were made for different concentrations of biotin added to the solution containing the labeled streptavidin. As the concentration of biotin increased, the fluorescence lifetime of the Cy5 dye changed due to the binding of biotin molecules to the streptavidin molecules. This change in lifetime was reflected as a change in the width of the sample transfer characteristic. For, each measurement, the code sequence was repeated 20 times, with the data averaged over the 20 cycles. Correlation was performed on the averaged data. A plot of change in transfer characteristic width as a function of biotin concentration is shown in