The National Institutes of Health (NIH) has noted that computed tomography (CT) use is up over 2000% in the last 20 years. In the 1990's, CT machines were primarily used for post-diagnostic tests. They are now often used for pre-diagnostic tests especially in the case of trauma such as car accidents, etc. Accordingly, CT use has become so prevalent that the cumulative radiation dosage which patients are exposed to during their lifetime has increased dramatically. This long term exposure risks cancerous outcomes as a result of the CT scans. Thus, the cost-benefit analysis arises: (1) one or more CT exam vs. (2) the possibility of long term health complications. X-ray radiation is generally considered carcinogenic, or cancer creating. The goal of this project is to allow the positive use of CT, while minimizing the amount of radiation used.
Many aspects of the present disclosure can be better understood with reference to the following drawings. The components in the drawings are not necessarily to scale, emphasis instead being placed upon clearly illustrating the principles of the present disclosure. Moreover, in the drawings, like reference numerals designate corresponding parts throughout the several views.
The present disclosure describes systems and methods for improving the focusing of computed tomography (CT) machines, such as all of those currently in use. This will allow one to image portions of the human anatomy, such as the spine, shoulders, hips, and elbows, without exposing an entire slice of the human body to a full radiation (x-ray) dose.
Computerized or computed tomography (CT) has become a standard diagnostic tool in modern medicine. In the last 30 years, however, CT use has become so prevalent that the cumulative radiation dosage which patients are exposed to during their lifetime has increased dramatically. The risk/reward tradeoffs for these diagnostic tools are hard to quantify. Often times a physician is only interested in a limited region of interest (ROI) which has been identified through clinical means or previous imaging. We believe that the risk can be reduced when a limited ROI is of interest by significantly reducing the total radiation dosage. The reward of this imaging will be preserved, since the images of the ROI will be nearly identical to those achieved with full radiation dosage. Large radiation dosages are well known to be potential causes for cancer.
The original investigations into region of interest (ROI) tomography, called local tomography, did not return the actual density of the ROI. Rather these returned an altered or transformed image which did preserve edges between varying tissues. The goal of local tomography is returning the actual image of the ROI. This is possible when one realizes that the low frequency components of the image are the only components which need non-local x-rays for their estimation. The high frequency components can be measured with the local line integral measurements which pass through the ROI. Thus, one does not need to send high dosage radiation on paths which do not intersect the ROI to find the high frequency components of the image. In
In the figure of
Thus, to provide a complete and accurate reconstruction of the ROI, two earlier methods feature (a) regular sampling of the x-rays through the ROI in order to reconstruct the high frequency components of the image via a computer processor of the CT scanner, and (b) sparse sampling of the x-rays which do not intersect the ROI, allowing for the recovery of low frequency components of the image. The combination of these methods results in an accurate reconstruction of the ROI with greatly reduced radiation dosages. These earlier methods work very well but only allow 0-1 sampling, i.e. either an x-ray is measured along a line integral or it is not. This is not feasible with current computerized tomography mechanisms.
The present disclosure concentrates on new improved methods and systems, which allow for variable sampling of the x-rays by featuring regular dosage radiation through the ROI & dramatically reduced dosage radiation through the portions of the body outside of the ROI. The lower dosage measurements outside of the ROI will be sufficient because they will only be utilized to determine the low-frequency components of the ROI image.
Thus, the lower signal to noise ratio (SNR) measurements (at lower radiation dosages) can be averaged producing adequate SNR measurements for these low frequency components. The detailed high-frequency components of the image will have standard SNR measurements (at standard radiation dosages). Thus, the image in the ROI can be made arbitrarily close to that using complete radiation dosage, with dramatically lower dosages.
Various systems and methods of the present disclosure will allow physicians to view and monitor regions of interest using radiation dosages outside the ROI which are 10% or less of the standard dosage. This work will augment, rather than compete with, efforts which primarily focus on improving receiver sensitivity.
Current clinical practice is to choose a body segment to be imaged, and expose the entire portion to the same radiation levels. Thus, if the right shoulder is the region of interest, the entire chest from the neck to below the shoulder will be exposed to the same amount of radiation. In accordance with teachings of the present disclosure, physicians will be enabled to ask for only an image of the right shoulder or ROI. Accordingly, systems and methods of the present disclosure may then concentrate radiation on that ROI and use a greatly reduced amount of radiation on the rest of the upper chest cavity and neck. Similarly, when doing spinal imaging, the radiation will be concentrated on that portion of the spine which is of interest, minimizing the radiation elsewhere, as demonstrated by
This focused tomography may be especially meaningful concerning diagnostic endeavors with children and pregnant women. The present disclosure discusses the imaging of tissue structures in cylindrical volumes (while also contemplating the use of non-cylindrical volumes). The mathematical methods of the present disclosure promise the possibility of imaging within formed structures that can directly follow the outline of the anatomy under concern. There are many organs in the human body which are well known to be susceptible to radiation exposure. Minimizing risk to these organs is one focus of the present disclosure.
This focused tomography can be accomplished through a thorough understanding of the CT reconstruction process. CT data is generally collected in a helical fan-beam geometry. For the discussion of the reconstruction process, it is easier if we assume that this data has been reformatted into a parallel beam data set. Thus, the present disclosure assumes that the data can be viewed as a function of two variables, angle {right arrow over (θ)} and distance r from the origin, or that the data, which is referred to as the projections or sinogram, has the form
Pƒ(r,{right arrow over (θ)})=∫ƒ(r{right arrow over (θ)}+t{right arrow over (θ)}⊥)dt, (1)
where t is an arbitrary variable chosen for the purposes of integration. The basics of the CT projection is illustrated in
To this end, let us recall that the natural coordinates for tomography are x=r{right arrow over (θ)}+t{right arrow over (θ)}⊥=r. Moreover, {right arrow over (x)}·{right arrow over (θ)}⊥=r. We want to see what the Fourier coefficients propagating at a fixed direction {right arrow over (θ)} have in common with the projections, so the present disclosure considers
where s is radial polar component of the polar representation of the Fourier transform. Thus, a central slice of the two dimensional Fourier transform of ƒ(x,y), i.e. ƒ(s{right arrow over (θ)}) can be obtained from the one dimensional projections of the function or Pƒ(r, {right arrow over (θ)}). Formally stated, we have Theorem 2.1 (Radon Transform, or Central Slice Theorem), as described below.
Theorem 2.1.
The one dimensional Fourier transform of Pƒ({right arrow over (θ)},r) is given by the central slice of the two dimensional Fourier transform, or ƒ(s{right arrow over (θ)}). Mathematically,
From this formula one can quickly derive the Filtered Backprojection formula, which was the basis of the 1979 Nobel Prize.
where |w(s)| is the frequency cut-off window.
By choosing w(s) appropriately, we can make the approximation above arbitrarily small. If we denote −1(|s|w(s))=k(r) then, by the convolution theorem, we have
ƒ({right arrow over (x)})≈∫(Pƒ(r,{right arrow over (θ)})*k(r)))({right arrow over (x)}·{right arrow over (θ)})dθ, (5)
where * denotes convolution.
One problem with Equation (5) is that the kernel k(r) is very broad as a function of r, and as a result, radiation measurements must be taken far from the region of interest. The reason for this kernel being broad is the jump discontinuity of the derivative of the function |s| at the origin from −1 to 1. Recall that |s| is the necessary term due to the polar coordinates used in the Fourier inversion of the filtered backprojection formula (4). The basic theorems of Fourier analysis dictate that this kernel cannot decay quickly.
This problem is solved by separating the discontinuity at the origin of |s| into separate portions: |s|w2(s) at the origin and |s|(1−w2(s)) away from the origin, as illustrated in
Referring back to
Thus, we have k(r)=kl(r)+kh(r). The filtered backprojection algorithm now looks like
and, we will reconstruct, via a computer processor, the low and high frequency terms of ƒ({right arrow over (x)}) separately. The kernels are illustrated in
The energy of kh(t) is contained within the interior 9 pixels of the current digitization or 9/512 to an accuracy of 1/10000. The energy concentration of kl(r), similarly measured, takes 175 terms. The low frequency terms take a great deal of non-local information, and the high frequency terms can be measured locally.
Initially, there seems to be no advantage to the change to two kernels kl(r) and kh(r) from a radiation reduction standpoint. The low frequency kernel will require the gathering of large quantities of data from outside the region of interest. Thus, there is no apparent win in the fact that the high frequency component ƒh({right arrow over (x)}) can be calculated from completely local measurements. One must understand the structure of the projections, and corresponding structure of the filtered backprojection algorithms to see how to solve the problems with the low frequency reconstruction ƒl({right arrow over (x)}). The structure theorem for the projections or Radon transform states that
P
ƒ({right arrow over (θ)},r)=(1−r2)−1/2 i=0∞Tl(r)hl(θ), (8)
where Tl(r) are the Chebyshev polynomials. Taking the Fourier transform of this yields
where Jl(s) are the Bessel functions, and hl(θ) is a trigonometric polynomial of order l. The key to understanding Equation (9) is that the low frequency terms in s, which are the Bessel functions, are only multiplied in frequency by the low order terms hl(θ). Thus, the low frequency terms do not have to be measured for many angles θ in order to accurately determine the complete low frequency components of the image.
The sampling of the projections, or Radon transform, is illustrated in
The present disclosure will now outline the refinements and improvements for a fan-beam geometry, as well as a parallel beam geometry. Accordingly,
As described previously, earlier methods concentrated on 0-1 sampling. Namely, a linear x-ray beam would be either sampled or not sampled. The present disclosure will relax this condition and attempt to find an optimal solution to minimize radiation. Thus, in various embodiments, techniques of the present disclosure will either sample at the necessary high-dosage rate, which is required for appropriate SNR and resolution in the ROI, or a variable lower-dosage rate, which is all that is necessary outside the ROI.
This optimal solution to this problem is necessarily better than the optimal solution to the 0-1 sampling problem. Any time more variables are added to an optimization problem, the solution necessarily gets better. Moreover, this will allow us to design the system in a way that is easily implementable in hardware. In one embodiment, static or non-adaptive/dynamic attenuating filters are used in front of the x-ray transmitter, to alter the beam for the appropriate reduced dosage.
This approach is now illustrated, both mathematically and visually. Assume for now that the region of interest is circular (with radius r1) and that the x-ray scanner is centered on the center of the circle when θ=0. Then, there is a distance d from the isocenter of the scanner to the center of the ROI, in which the ROI is assumed to be a circle with radius r1. As the x-ray transmitter moves with θ, the center of the circle will then be a distance d(θ)=d sin(θ) off of the center of the gantry. Thus, we want to gather a full data set of the x-rays which pass through the ROI, which represent a first data set Pƒh(θ,r)=Pƒ(θ,r) where 0≤0≤π and r ∈ [d sin(θ)−r1, d sin(θ)+r1]. This first data set is gathered with full radiation dosage, just as if you were going to image the whole slice. Therefore, it will have a relatively high SNR. The notation pƒh(θ,r) recognizes that this data will be used to reconstruct the high frequency details of the image.
A second data set is then gathered from all of the lines or projections which did not intersect the region of interest. The second data set is a low frequency data set pƒl(θ,r)=Pƒ(θ, r), where r ∉[d sin(θ)−r1, d sin(θ)+r1]. This low frequency data set is gathered with minimal radiation, will have very low SNR, and will only be needed to reconstruct the low frequency portion of the image and will not affect the final image in the ROI.
The data sets are combined to get an approximate, noisy sinogram or Radon transform Pƒ(θ,r)≈Pƒh(θ,r)+Pƒl(θ,r), noting that all of the Radon transform has been sampled, some of it at high SNR and some at low SNR. The final reconstruction will be
The first term Pƒh(r, θ)*kh(r) is the term which will yield most of the high resolution image, and is highly sampled through the ROI. This is the foundation of the reconstruction. The second term Pƒl(r, θ)*kh(r) will be essentially zero, since it is the convolution of high frequency data versus a low frequency kernel. The third term Pƒh(r, θ)*kl(r) is essentially zero for the same reason—it is the convolution of low frequency data and a high frequency kernel. The last term, Pƒl(r, θ)*kl(r) is the low frequency term which is essential. At first one might think that this would end our ability to accomplish the task of lowering the radiation levels, since these low frequency terms are global and cannot be measured locally. We must remember the structure of the Radon transform to minimize this non-local information.
Thus, there is a very high SNR estimate for the high frequency components inside the ROI. We must recall the structure of the Radon transform to realize why the low frequency component is not affected by the low SNR estimates. Recall from Equation (11) that the low frequency components are only affected by low frequency sines and cosines with respect to θ or the Fourier transform
with N being very small. Restating this, if you look at the Fourier transform from a polar viewpoint, the small circular components are controlled by very low-order sines and cosines. Therefore, since we are estimating very few parameters in the low frequency component and have a great number of data samples, the law of large numbers will yield a very solid estimate for the low frequency component. This can be accomplished even if this data is gathered at very low SNR levels, i.e. with very little radiation.
This process is illustrated in
To address this inquiry,
Noise analysis and modification can be done on the unprocessed raw data which comes from the photon detectors. Let's recall Equation (1):
P
ƒ(r, {right arrow over (θ)})=∫ƒ(r{right arrow over (θ)}+t{right arrow over (θ)}⊥)dt,
which is central to CT.
The data Pƒ(r, {right arrow over (θ)}) is actually pre-processed in the following way. The actual raw data from the CT detectors, Pƒ(r, {right arrow over (θ)}) is given by
R
ƒ(r, {right arrow over (θ)})=exp (−Pƒ(r, {right arrow over (θ)})).
Thus, the utilized data is gained by the expression
P
ƒ(r, {right arrow over (θ)})=−ln(Rƒ(r, {right arrow over (θ)})).
The reason that we want to step back to Rƒ(r, {right arrow over (θ)}) is that the statistics of measuring Rƒ are very well known. A CT machine sends photons through the medium (i.e. the patient) along a straight line, and the resulting arrivals are classic Poisson arrivals with their mean being (the natural log of) the actual desired result Pƒ(r, {right arrow over (θ)}) and a variance which decreases according to how many photons are sent. Thus, lower radiation measurements of Rƒ are just noisier in a classical Poisson measurement process.
For the following trials, data is acquired by imaging a human cadaver at 4 different radiation dose levels: 2 milligray (mGy), 8 mGy, 18 mGy, and 60 mGy. To put perspective on this, most clinical abdominal CT scans reconstructed with filtered back-projection will use around 18-20 mGy for adults, with 60 mGy far exceeding the norm. Scanners using iterative reconstruction algorithms may use closer to 8 mGy for the same scan. For children or infants, lower doses such as 8 mGy or 2 mGy are routinely used. This is partially due to two reasons: 1) The much smaller body size, meaning its far easier to penetrate through the body and 2) Increased attention to dose reduction, due to the higher radiosensitivity of children and the longer lifespan they have in which to develop a cancer with a long latent period.
For these trials, the CT scanner that was used did not permit further dose reduction below 2 mGy. Future clinical implementation of this technique will require the development of an adjustable collimator mechanism, which will allow us to reduce the ionizing radiation exposure to areas outside the ROI to about 10% or less of the normal clinical dose. For the present disclosure, using 18 mGy for the region of interest (ROI) and 2mGy for the exterior, non-ROI projections adequately simulates the ROI/non-ROI dose ratio that is anticipated to perform well with the focused tomography method.
For the clinical trials, measurements are made at 18 mGy for the region of interest (ROI) and at 2mGy for the exterior, non-ROI projections. For simplicity two different modes are specified for the acquired images. The first is denoted by Focused Tomography 1 (FCT1) by which the present disclosure uses the lower dosage measurement outside the ROI and the higher dosage measurements inside the ROI with no transition region between them. The second is denoted by Focused Tomography 2 (FCT2) by which we have a smooth windowed transition or ramp between the lower dosage and the higher dosage measurements.
The first clinical study is of the sacrum which is illustrated in
As used in acquisition of the images of
A second clinical study is of the lower thoracic spine which is illustrated in
As used in acquisition of the images of
Accordingly,
Next,
Referring to
Referring next to
The same techniques are illustrated in
For
To assess the comparative image quality of the two methods (i.e. full dosage ROI scanning and reduced dosage ROI scanning), various metrics can be utilized. The obvious first metric can be simple mathematical mean square error measurement. These measurements do not, however, generally yield a true measure of the image quality. Therefore, visual quality metrics can be utilized, which have been developed in computer vision over the past 15 years to assess the relative value of both methods. Finally, a significant number of qualified physicians, surgeons, and radiologists can be enlisted to view this work. The study can be made to be double blind by showing images within the ROI from both the full exposure and reduced exposure methods.
In various embodiments, static attenuation filters can be developed and used in conjunction with scanners, in accordance with various embodiments of the present disclosure. The static filters will focus the radiation on the ROI and eliminate unnecessary non-local radiation. These static filters will only be able to gather data on a centralized ROI. In various embodiments, adaptive filters may be used as alternatives to static filters.
Additionally, in various embodiments, designed filters will be active, gathering data through a non-centered ROI at near real-time gantry speeds. Cramer-Rao statistical bounds will be attempted to prove that minimal data sets are being used. Also, in various embodiments, designed filters will be fully automated, so that a physician can choose his/her ROI, and have the machine optimally reduce (e.g., via a controller device) the radiation dose in accordance with the present disclosure.
Accordingly, in one embodiment, a designed filter will be a static filter, in which we will look to build a filter, or filters, which will alter the output of the CT machine producing a desirable localized radiation profile. This is possible with a number of materials, such as aluminum, tungsten, or lead, by merely adjusting the depth of the filter. A stable material which will not degrade under the radiation exposure and which can be made thin enough to not interfere with the gantry is used in certain embodiments. In one embodiment, a designed filter will be a mechanically active, materially static filter. Accordingly, after having a static filter design, the filter can be mechanically moved, keeping the focus of the radiation on the region of interest.
In the present disclosure, cylindrical ROIs are considered. This is consistent with the spine and shoulders, but perhaps not optimal for hip imaging. However, it is contemplated that the mechanically-active filter would be capable of essentially imaging any ROI, including non-cylindrical ROIs. The radiation reduction of an exemplary cylindrical system is very substantial. We do not imagine decreasing this by more than 2 or 3 times with more advanced methods.
The following describes the sequence of steps required to perform the novel method of focused computed tomography (CT) in one embodiment. In general, a computed tomography scanner features a ring or cylinder for a gantry, in which a subject is positioned. An x-ray tube and an x-ray detector are positioned opposite of each other and rotate around the gantry as x-ray images are acquired. A body scanning filter, such as bowtie filter, is generally positioned in front of the x-ray tube to shape the x-ray beam and reduce the range of x-ray energies that reach the subject, such as reducing the beam intensity at the periphery of the x-ray beam that is transmitted. Additionally, a static pre-patient collimator may be positioned between the filter and the patient. In accordance with the present disclosure, an additional adaptive collimator device can also be positioned between the pre-patient collimator and the patient.
Accordingly, in one embodiment, an exemplary CT scanner of the present disclosure provides two sliding collimators at the location indicated in
Additionally, the position of the opening may be adjustable laterally to create a focused field-of-view anywhere within the scan field-of-view, as shown in
In one embodiment, the sliding collimators are made of a material with sufficient thickness and density to reduce the measured air kerma to approximately one-tenth of the air kerma that would be measured in the beam exiting the bowtie filter and any other pre-patient filter. The exact thickness of the collimator depends on the beam quality of the CT scanner, which is dependent on manufacturer and model.
For structural integrity and the ability to adjust position quickly, the material can be a hard metal that can achieve this attenuation with a thickness of no more than a few millimeters. Appropriate materials that meet the requirements of sufficient thickness/density and structural integrity include but are not limited to copper, lead, or tungsten, in various embodiments. Approximate thicknesses would be about 0.3 mm of copper, lead, or tungsten, in various embodiments.
In one embodiment of an exemplary scanning procedure, the patient is placed on the table and positioned by the CT technologist for the CT scan. No special positioning of the patient specific to the application of a focused CT is required. The technologist acquires anteroposterior (AP) and lateral topograms. The anatomy to be included in the focused CT is marked on both topograms (in
The acquisition parameters such as kV, mA, rotation time, and pitch should be the same as those that would be used in clinical exams without the sliding collimators. No adjustment to these techniques is required. Systems using mA modulation calculated from the topograms should continue to function correctly with the sliding collimators in place. Systems calculating mA modulation on the fly will likely need to be switched to a manual mA, in various embodiments.
In various embodiments, the sliding collimators adjust continuously or repeatedly during the CT acquisition to restrict the SFOV to the area of interest (as shown in
In a further discussion of exemplary reconstruction techniques of the present disclosure, it is noted that the data in any CT Scanner or x-ray device is classically a Poisson process. Some of the fundamentals of a Poisson process X is that if mean or expectation of the process is E(X)=A, then the standard deviation is σ(X)=√{square root over (λ)}. A standard way to define the quality of the signal that a CT is receiving is the signal to noise ratio (SNR), which is often defined as:
Thus the signal to noise ratio is improved if we can increase E(X). This is certainly true in CT. If we increase the intensity of the beam, I0, then we will get a better x-ray measurement. Unfortunately this exposes the patient to higher radiation levels, which are well documented to be harmful to the health of the human body. Thus, a goal of the present disclosure is to reduce the radiation dosage level while preserving the quality of the image in the region of interest (ROI), where the quality of the image is dominated by the SNR. As such, exemplary methods of the present disclosure involve maintaining a high level of intensity or dosage through the region of interest and reducing the level of intensity elsewhere.
One approach (“Approach A”) to CT reconstruction utilizes raw projection data from the CT x-ray detector. Accordingly,
Thus, the adaptive collimator 2710 is configured to allow a standard CT radiation dosage through the ROI, and a lower dosage by a factor of 0<a<1 outside of the ROI. The received signal is given by:
P
ƒ
r(θ,r)=Pƒ(θ,ri)+Pƒα(θ,r0),
where Pƒr is the total received signal, ri are the x-rays which pass through the ROI, and r0 are the x-rays which do not intersect the ROI, and pƒα are the projections or x-rays which are taken at a lower dosage α (that represents the radiation dosage factor that is applied to the areas outside the region of interest during CT scanning).
The final result for the total received signal is obtained by the equation:
Note that
will have the same expected values as Pƒ(θ, r), but will
have the lower signal to noise ratio associated with the lower dosage, where 1/∝ may be considered to be a multiplicative correction factor. The present disclosure has shown that this multiplicative reconstruction approach 2760 produces diagnostic CT images 2770 which have the quality of the higher dosage within the ROI and produces images with the quality of the lower dosage outside of the ROI.
The above approach requires that raw data projection data 2740 collected during CT scanning and typically stored as a sinogram file is available for further processing using techniques of the present disclosure to produce the desired high-quality CT images. However, manufacturers of CT scanners typically do not provide such raw data to be output for processing to external equipment or to anyone outside of their organization. Thus, reconstruction techniques developed by third parties that require raw projection data and may be an improvement over the manufacturer's reconstruction techniques 2750 (that are built into the manufacturer's scanner equipment) are unable to produce their own CT images using their own CT reconstruction techniques and must use the CT images 2780 output by a manufacturer's CT scanner using the manufacturer's CT reconstruction techniques.
However, in accordance with the present disclosure, an exemplary alternative reconstruction technique of the present disclosure utilizes the CT scan images 2780 output or made available by a manufacturer's CT scanner in order to produce improved CT scan images 2795 using third-party CT reconstruction techniques. Referring back to
This exemplary technique can be utilized if CT scan images 2780 are available from a CT scanner machine, which is often times the case. So Ir and I0 represent the received and initial intensities, then the basic principle of exponential attenuation/tomography says that:
Therefore, it follows that:
Now, if we artificially manipulate Ir by 0<α<1 in accordance with embodiments of the present disclosure, we have:
for the received intensity measurement.
Recalling that ln(ab)=ln(a)+ln(b), we have:
Thus, we get:
Thus, the correction factor applied in the alternate reconstruction technique 2790 is additive rather than multiplicative, which is advantageous from a signal to noise ratio perspective. The results should therefore be identical with the exception of minor numerical rounding errors. Accordingly, the flow chart of
In accordance with various embodiments of the present disclosure, a CT machine first generates a CT reconstructed image using the CT machine's default reconstruction algorithm and then the reconstructed image is corrected to account for a different radiation dosage level that was applied to areas outside a region of interest during the scan by adjusting pixel values in the reconstructed image using an additive correction factor (e.g., In(a)) where a represents the radiation dosage factor that is applied to the areas outside the region of interest during CT scanning. Such a technique can be employed when manufacturers of CT scanners do not make raw projection data available to anyone outside their organizations, even to customers who use their CT scanners.
Thus, referring back to
Functionality of an exemplary CT scanner and disclosed reconstruction techniques in certain embodiments of the present disclosure or portions thereof can be implemented in hardware, software, firmware, or a combination thereof. Such software or firmware can be stored in a computer readable medium, such as memory and be executed by a suitable instruction execution system. If implemented in hardware, the hardware can be implemented with any or a combination of the following technologies, which are all well known in the art: a discrete logic circuit(s) having logic gates for implementing logic functions upon data signals, an application specific integrated circuit (ASIC) having appropriate combinational logic gates, a programmable gate array(s) (PGA), a field programmable gate array (FPGA), etc.
Stored in the memory 2804 are both data and several components that are executable by the processor 2802. In particular, stored in the memory 2804 and executable by the processor 2802 are various disclosed methods of the present disclosure, such as, but not limited to, exemplary CT reconstruction techniques 2812. Also stored in the memory 2804 may be a data store 2814 and other data. The data store 2814 can include raw projection data or CT scan images or potentially other data. In addition, an operating system may be stored in the memory 2804 and executable by the processor 2802. The I/O devices 2808 may include input devices, for example but not limited to, a keyboard, touchscreen, mouse, recording devices, and/or sensors, etc. Furthermore, the I/O devices 2808 may also include output devices, for example but not limited to, a display, a printer, communication adapters, etc.
In the context of this document, a “computer-readable medium” can be any means that can contain, store, communicate, or transport a program for use by or in connection with an instruction execution system, apparatus, or device. The computer readable medium can be, for example but not limited to, an electronic, magnetic, optical, electromagnetic, infrared, or semiconductor system, apparatus, or device. More specific examples (a nonexhaustive list) of the computer-readable medium would include the following: an electrical connection (electronic) having one or more wires, a portable computer diskette (magnetic), a random access memory (RAM) (electronic), a read-only memory (ROM) (electronic), an erasable programmable read-only memory (EPROM or Flash memory) (electronic), an optical fiber (optical), and a portable compact disc read-only memory (CDROM) (optical).
It should be emphasized that the above-described embodiments are merely possible examples of implementations, merely set forth for a clear understanding of the principles of the present disclosure. Many variations and modifications may be made to the above-described embodiment(s) without departing substantially from the principles of the present disclosure. All such modifications and variations are intended to be included herein within the scope of this disclosure.
This application is a continuation-in-part of co-pending U.S. non-provisional patent application entitled “Focused Tomography,” having application Ser. No. 16/986,014, filed Aug. 5, 2020, which claims priority to U.S. provisional application entitled, “Focused Tomography,” having application No. 62/884,230, filed Aug. 8, 2019, all of which are entirely incorporated herein by reference in their entireties.
Number | Date | Country | |
---|---|---|---|
62884230 | Aug 2019 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 16986014 | Aug 2020 | US |
Child | 18372819 | US |