a is a schematic diagram of the testing system.
b is a schematic diagram of the gas absorption reactor.
a shows a schematic diagram of the gas absorption testing system of this invention which contains a gas flow section and an absorption reactor, 2. The gas flow section includes a source of N2, a gas supply, and a scrubber.
The chemical to be tested is placed in the absorption reactor either in the pure form or in a solution. Aqueous solution is preferred, but non-aqueous, e.g. sulfolane, and mixed aqueous-non-aqueous solutions may be used. The appropriate test gas (pure gas or a mixture of gases is released from the gas supply cylinder and flows through a well designed delivery system to the absorption reactor with all regulators and safety valves to make sure there is complete control of flow and pressure. The gas enters the inner tube of the reactor and bubbles through the outer solution. During the test, the absorption reactor is in a thermostated circulated solution bath to control the absorption reaction temperature. A magnetic stirring bar (not shown in Figures) is placed in the bottom of the absorption reactor to make sure the testing solution is well mixed during the test. The test parameters such as the length of flow, the rate of flow, and the concentration of the solution depend on the design of the test which in turn is related to the purpose of the testing and type of application.
b shows a diagram of a gas absorption reactor. The design consideration is based on (1) minimization of the reaction solution required, (2) maximization of the contact surface and contact time between gas and solution, (3) easy control of the reaction environment such as the absorption reaction temperature and (4) easy sampling both of incoming gas as well as the solution during the test. The absorption reactor in
An example using the gas absorption testing system of this invention is given to demonstrate the efficiency and effectiveness.
1) EETB (242 mg) is used to prepare a 0.15 M, 10 mL aqueous solution.
2) A 3 mL aliquot of the solution is placed in the absorption reactor. The reactor is placed into a circulated bath at 35° C. for absorption test.
3) A test gas mixture (89% N2, 10% CO2, and 1% H2S) is flowed at 4.2 mL per minute through the test solution for 2 hours.
4) During the test, a sample of test gas mixture, 1 μl to 5 μl (1 mg to 5 mg), is collect by a gas-sampling syringe, and a gas chromatograph with a thermal conductivity detector (GC/TCD) is used to analyze the composition of this gas mixture.
5) Table 1 lists the test specifications of the GC/TCD.
6) After 2 hours of the absorption test, a 3 μL sample of liquid solution is collected. The same GC/TCD is used to analyze the content of CO2 and H2S.
7)
8)
9) Table 2 presents the definition of selectivity and the details of the analytical instrumentation.
10) The solution after the absorption test will be used for the desorption test.
11) The gas mixture is replaced by a nitrogen gas flow with the same flow rate through the absorption reactor.
12) The circulated bath temperature is increased from 35° C. to 85° C.
13) The desorption test is continued for 2 hours. After 2 hours of desorption, a 3 μL sample of liquid solution is collected. The same GC/TCD is used to analyze the content of CO2 and H2S.
14)
15) The solution after the desorption test will be used for the re-absorption test.
16) The nitrogen gas flow is replaced with the previous N2, H2S, CO2 gas mixture flow with the same flow rate through the absorption reactor.
17) The circulating bath temperature is lowered from 85° C. to 35° C.
18) The test is continued for 2 hours. After 2 hours of the re-absorption test, a 3 μL sample of liquid solution is collected. The same GC/TCD is used to analyze the content of CO2 and H2S.
19)
20) Table 3 presents the definition of Capacity and the details of the analytical instrumentation.
21) A 0.15 M propyl mercaptan solution is prepared as a reference concentration standard.
22) A 1 mL solution, after the re-absorption test, is mixed with an equal volume of the reference standard solution.
23) A 1 μL sample of mixed solution is collected and analyzed with a gas chromatograph with a sulfur chemiluminescence detector (GC/SCD) to determine the composition of this gas mixture.
24) Table 4 presents the definition of Loading and the details of the analytical instrumentation.
25) Table 5 shows the test results for the ethoxyethanol-t-butylamine (EETB).
In this invention, a gas absorption testing system is described. This gas absorption testing system is used for testing chemicals to explore their absorption, desorption and re-absorption properties in the solution. The design and the set-up of the gas absorption reactor are disclosed. The testing procedures as well as the detailed analytical instrumentation used to accomplish the necessary tests are also described. The test data using ethoxyethanol-t-butylamine (EETB) as an example of a tested chemical is demonstrated.
The advantages of this gas absorption testing system compared with prior art or present commercial practices are:
1) much less sample (chemicals) is required to do the test (less then 0.5 gram). This makes the synthesis of new chemicals much easier.
2) micro-liter sample size is required to do the analytical measurement, which volume for the measurement is less than 0.1% of the solution and therefore, the system is not disturbed during the analytical measurement.
3) Analytical measurements are based on gas chromatography (GC) with a thermal conductivity detector (TCD) and a sulfur chemiluminescence detector (SCD). These technologies have proven precision and accuracy for all the absorption properties.
This application claims the benefit of U.S. Provisional application 60/808,010 filed May 23, 2006.
Number | Date | Country | |
---|---|---|---|
60808010 | May 2006 | US |