Embodiments of the subject matter described herein relate generally to drug delivery systems and more specifically to systems for controlling the infusion rate of insulin based on state variable feedback.
The pancreas of a normal healthy person produces and releases insulin into the blood stream in response to elevated blood plasma glucose levels. Beta cells (β-cells), which reside in the pancreas, produce and secrete the insulin into the blood stream, as it is needed. If β-cells become incapacitated or die, a condition known as Type I diabetes mellitus (or in some cases if β-cells produce insufficient quantities of insulin, Type II diabetes), then insulin must be provided to the body from another source.
Traditionally, since insulin cannot be taken orally, insulin has been injected with a syringe. More recently, use of infusion pump therapy has been increasing, especially for delivering insulin for diabetics. For example, external infusion pumps are worn on a belt, in a pocket, or the like, and deliver insulin into the body via an infusion tube with a percutaneous needle or a cannula placed in the subcutaneous tissue. As of 1995, less than 5% of Type I diabetics in the United States were using infusion pump therapy. Presently, over 7% of the more than 900,000 Type I diabetics in the United States are using infusion pump therapy, and the percentage of Type I diabetics that use an infusion pump is growing at an absolute rate of over 2% each year. Moreover, the number of Type I diabetics is growing at 3% or more per year. In addition, growing numbers of insulin-using Type II diabetics are also using infusion pumps. Physicians have recognized that continuous infusion provides greater control of a diabetic's condition, and are also increasingly prescribing it for patients. Although offering control, pump therapy can suffer from several complications that make use of traditional external infusion pumps less desirable for the user.
In insulin pumps, it is common to use fast acting insulin as opposed to the slower acting insulin that is used for injections, because pumps allow changing of insulin profiles. As insulin companies develop faster acting insulin, the faster acting insulin is often adopted quickly. However, current pumps are still limited by the speed of the insulin they are using.
A processor-implemented method is presented here. The method can be used to control an insulin infusion device for a user. Certain embodiments of the method involve the operation of a processor architecture having at least one processor device to obtain a current insulin on board (IOB) value that represents an estimate of active insulin in the body of the user. The method continues by calculating, by the processor architecture, an IOB rate based at least in part on the obtained current IOB value. The method continues by determining, by the processor architecture, an adjusted insulin infusion rate based at least in part on the calculated IOB rate and an uncompensated insulin infusion rate. The processor architecture selects a final insulin infusion rate for the insulin infusion device, wherein either the determined adjusted insulin infusion rate, the uncompensated insulin infusion rate, or a current basal rate is selected as the final insulin infusion rate.
Also presented here is a processor-implemented method of controlling an insulin infusion device for a user. Certain embodiments of the method begin by generating a current IOB value that represents an estimate of active insulin in the body of the user. The method continues by calculating an IOB rate based at least in part on the generated current IOB value, obtaining an uncompensated insulin infusion rate, and determining an adjusted insulin infusion rate in accordance with the expression AdjustedRate(n)=max(0; PIDRate(n)−IOBRate(n)). The method continues by selecting a final insulin infusion rate in accordance with the expression
In this expression: AdjustedRate(n) is the determined adjusted insulin infusion rate; PIDRate(n) is the obtained uncompensated insulin infusion rate; IOBRate(n) is the calculated IOB rate; FinalRate(n) is the selected final insulin infusion rate; and Basal is a current basal rate maintained by the insulin infusion device for the user.
Also presented here is a tangible and non-transitory electronic storage medium having processor-executable instructions that, when executed by a processor architecture comprising at least one processor device, perform a method of controlling an insulin infusion device for a user. In certain embodiments, the method begins by estimating a current IOB value that indicates an amount of active insulin in the body of the user. The method continues by calculating an IOB rate based at least in part on the estimated current IOB value, determining an adjusted insulin infusion rate based at least in part on the calculated IOB rate and an uncompensated insulin infusion rate, and selecting a final insulin infusion rate for the insulin infusion device, wherein either the determined adjusted insulin infusion rate, the uncompensated insulin infusion rate, or a current basal rate is selected as the final insulin infusion rate. The method then provides the selected final insulin infusion rate to regulate delivery of insulin by the insulin infusion device.
An electronic device is also presented here. Certain embodiments of the electronic device include a processor architecture and at least one memory element associated with the processor architecture. The at least one memory element stores processor-executable instructions that, when executed by the processor architecture, perform a method of controlling an insulin infusion device for a user. The method involves: computing a current IOB value that indicates an amount of active insulin in the body of the user; calculating an IOB rate based at least in part on the computed IOB value; determining an adjusted insulin infusion rate based at least in part on the calculated IOB rate and an uncompensated insulin infusion rate; and selecting a final insulin infusion rate for the insulin infusion device. The selecting step selects either the determined adjusted insulin infusion rate, the uncompensated insulin infusion rate, or a current basal rate as the final insulin infusion rate.
An electronic controller for an insulin infusion device is also presented here. The electronic controller includes a processor architecture comprising at least one processor device, and at least one memory element associated with the processor architecture. The at least one memory element stores processor-executable instructions that, when executed by the processor architecture, provide an IOB compensation module that estimates a current IOB value that indicates an amount of active insulin in the body of the user, calculates an IOB rate based at least in part on the estimated current IOB value, and determines an adjusted insulin infusion rate based at least in part on the calculated IOB rate and an uncompensated insulin infusion rate. The IOB compensation module selects a final insulin infusion rate for the insulin infusion device, wherein the final insulin infusion rate is selected as either the determined adjusted insulin infusion rate, the uncompensated insulin infusion rate, or a current basal rate. The IOB compensation module then provides the selected final insulin infusion rate to regulate delivery of insulin by the insulin infusion device.
An exemplary embodiment of an electronic device is also provided here. The electronic device includes a processor architecture having at least one processor device, and at least one memory element associated with the processor architecture. The at least one memory element stores processor-executable instructions that, when executed by the processor architecture, perform a method of controlling an insulin infusion device for a user. The method operates the insulin infusion device in a closed-loop mode to deliver insulin to the body of the user, obtains current insulin-delivered data that indicates an amount of insulin delivered by the insulin infusion device during a most recent sampling period, obtains current sensor data that indicates a current sensor glucose value for the user corresponding to the most recent sampling period, and processes historical insulin-delivered data and historical sensor data, for a plurality of historical sampling periods prior to the most recent sampling period, to obtain predicted sensor glucose values for a historical time period. The method continues by calculating a difference between the current sensor glucose value and a predicted current sensor glucose value for the most recent sampling period, wherein the predicted sensor glucose values for the historical time period include the predicted current sensor glucose value. The method continues by generating an alert when the difference exceeds a threshold error amount.
The following detailed description also relates to a tangible and non-transitory electronic storage medium having processor executable instructions that, when executed by a processor architecture comprising at least one processor device, perform a method of controlling an insulin infusion device for a user. The method involves operation of the insulin infusion device in a closed-loop mode to deliver insulin to the body of the user. The method continues by identifying, from historical sensor glucose values for the user, a baseline historical sensor glucose value obtained during a begin-training sampling period. The method calculates a plurality of candidate solutions to a sensor glucose prediction model, wherein each of the plurality of candidate solutions is calculated as a function of a bounded initial condition and historical insulin delivered data for the user, and wherein the bounded initial condition is influenced by the baseline sensor glucose value. The method continues by selecting a best-matched solution from the calculated plurality of candidate solutions, based on a comparison of predicted sensor glucose values from the calculated plurality of candidate solutions to a first portion of the historical sensor glucose values. The predicted sensor glucose values from the best-matched solution are compared to a second portion of the historical sensor glucose values, wherein the first portion of the historical sensor glucose values corresponds to a distant history period, the second portion of the historical sensor glucose values corresponds to a recent history period, and the distant history period occurred before the recent history period that data samples. The method continues by generating an alert, in response to the comparing, when the second portion of the historical sensor glucose values deviates from the best-matched solution by at least a threshold error amount.
Also presented here is an embodiment of an electronic controller for an insulin infusion device. The electronic controller includes a processor architecture comprising at least one processor device, and at least one memory element associated with the processor architecture. The at least one memory element stores processor-executable instructions that, when executed by the processor architecture, provide a model supervisor module to obtain, during closed-loop operation of the insulin infusion device, insulin-delivered data that indicates an amount of insulin delivered by the insulin infusion device during a most recent sampling period, and current sensor data that indicates a current sensor glucose value for the user corresponding to the most recent sampling period. The model supervisor module defines a model training period and a model prediction period for a historical period of time, and finds a best-matched solution to a sensor glucose prediction model, relative to historical sensor glucose values obtained during the model training period, wherein the best-matched solution is a function of a baseline sensor glucose value obtained during the model training period, and is a function of historical insulin-delivered data for the user obtained during the historical period of time. The model supervisor module compares at least one predicted sensor glucose value from the best-matched solution to at least one historical sensor glucose value corresponding only to the model prediction period, and generates an alert, in response to the comparing, when the at least one historical sensor glucose value deviates from the at least one predicted sensor glucose value by at least a threshold error amount.
Also included below is a detailed description of a processor-implemented method of controlling an insulin infusion device for a user. The method may begin by operating the insulin infusion device in a closed-loop mode to deliver insulin to the body of the user. The method continues by obtaining current insulin-delivered data that indicates an amount of insulin delivered by the insulin infusion device during a most recent sampling period, obtaining current sensor data that indicates a current sensor glucose value for the user corresponding to the most recent sampling period, and processing historical insulin-delivered data and historical sensor data, for a plurality of historical sampling periods prior to the most recent sampling period, to obtain predicted sensor glucose values for a historical time period. The method then calculates a difference between the current sensor glucose value and a predicted current sensor glucose value for the most recent sampling period, wherein the predicted sensor glucose values for the historical time period include the predicted current sensor glucose value. An alert is generated when the difference exceeds a threshold error amount.
Also included below is a detailed description of a processor-implemented method of controlling an insulin infusion device for a user. The method may begin by operating the insulin infusion device in a closed-loop mode to deliver insulin to the body of the user. The method continues by identifying, from historical sensor glucose values for the user, a baseline historical sensor glucose value obtained during a begin-training sampling period. Next, a plurality of candidate solutions to a sensor glucose prediction model is calculated, wherein each of the plurality of candidate solutions is calculated as a function of a bounded initial condition and historical insulin delivered data for the user, and wherein the bounded initial condition is influenced by the baseline sensor glucose value. The method continues by selecting a best-matched solution from the calculated plurality of candidate solutions, based on a comparison of predicted sensor glucose values from the calculated plurality of candidate solutions to a first portion of the historical sensor glucose values. At least one predicted sensor glucose value from the best-matched solution is compared to a second portion of the historical sensor glucose values, wherein the first portion of the historical sensor glucose values corresponds to a distant history period, the second portion of the historical sensor glucose values corresponds to a recent history period, and the distant history period occurred before the recent history period that data samples. An alert is generated, in response to the comparing, when the second portion of the historical sensor glucose values deviates from the best-matched solution by at least a threshold error amount.
Another embodiment of a processor-implemented method of controlling an insulin infusion device for a user is also presented below. The method involves operating the insulin infusion device in a closed-loop mode to deliver insulin to the body of the user, defining a model training period and a model prediction period for a historical period of time, and finding a best-matched solution to a sensor glucose prediction model, relative to historical sensor glucose values obtained during the model training period, wherein the best-matched solution is a function of a baseline sensor glucose value obtained during the model training period, and is a function of historical insulin-delivered data for the user obtained during the historical period of time. The method continues by comparing at least one predicted sensor glucose value from the best-matched solution to at least one historical sensor glucose value corresponding only to the model prediction period. An alert is generated, in response to the comparing, when the at least one historical sensor glucose value deviates from the at least one predicted sensor glucose value by at least a threshold error amount.
Additional processor-implemented methods of controlling an insulin infusion device are also presented below. For example, one method involves operating a processor architecture having at least one processor device to obtain sensor calibration data for a continuous glucose sensor that generates a sensor variable indicative of blood glucose of the user. This method continues by identifying a most recent calibration factor from the sensor calibration data, the most recent calibration factor representing a first conversion value applicable to convert a first value of the sensor variable to a first blood glucose value. This method also identifies a prior calibration factor from the sensor calibration data, the prior calibration factor representing a second conversion value applicable to convert a second value of the sensor variable to a second blood glucose value, and the prior calibration factor corresponding to an earlier time relative to the most recent calibration factor. The method regulates entry into a closed-loop operating mode of the insulin infusion device, based on the most recent calibration factor and the prior calibration factor.
Also provided is a tangible and non-transitory electronic storage medium having processor-executable instructions that, when executed by a processor architecture, perform an exemplary embodiment of a method of controlling an insulin infusion device for a user, the method including: obtaining sensor calibration data for a continuous glucose sensor that generates a sensor variable indicative of blood glucose of the user; identifying a most recent calibration factor from the sensor calibration data, the most recent calibration factor representing a first conversion value applicable to convert a first value of the sensor variable to a first blood glucose value; identifying a prior calibration factor from the sensor calibration data, the prior calibration factor representing a second conversion value applicable to convert a second value of the sensor variable to a second blood glucose value, and the prior calibration factor corresponding to an earlier time relative to the most recent calibration factor; and regulating entry into a closed-loop operating mode of the insulin infusion device, based on the most recent calibration factor and the prior calibration factor.
An embodiment of an electronic device is also presented here. The electronic device includes: a processor architecture having at least one processor device; and at least one memory element associated with the processor architecture. The memory element stores processor-executable instructions that, when executed by the processor architecture, perform a method of controlling an insulin infusion device for a user. The method involves: obtaining sensor calibration data for a continuous glucose sensor that generates a sensor variable indicative of blood glucose of the user; identifying a most recent calibration factor from the sensor calibration data, the most recent calibration factor representing a first conversion value applicable to convert a first value of the sensor variable to a first blood glucose value; identifying a prior calibration factor from the sensor calibration data, the prior calibration factor representing a second conversion value applicable to convert a second value of the sensor variable to a second blood glucose value, and the prior calibration factor corresponding to an earlier time relative to the most recent calibration factor; and regulating entry into a closed-loop operating mode of the insulin infusion device, based on the most recent calibration factor and the prior calibration factor.
Also provided is an embodiment of an electronic controller for an insulin infusion device. The electronic controller includes: a processor architecture having at least one processor device; and at least one memory element associated with the processor architecture, the at least one memory element storing processor-executable instructions that, when executed by the processor architecture, provide a closed-loop initiation module. The initiation module is configured to: obtain a most recent calibration factor for a continuous glucose sensor that generates a sensor variable indicative of blood glucose of a user, the most recent calibration factor representing a first conversion value applicable to convert a first value of the sensor variable to a first blood glucose value; obtain a prior calibration factor for the continuous glucose sensor, the prior calibration factor representing a second conversion value applicable to convert a second value of the sensor variable to a second blood glucose value; obtain calibration timestamp data for the most recent calibration factor and the prior calibration factor; and regulate entry into a closed-loop operating mode of the insulin infusion device, based on the most recent calibration factor, the prior calibration factor, and the calibration timestamp data.
Another embodiment of a processor-implemented method of controlling an insulin infusion device for a user is also provided. This method involves: operating a processor architecture to initiate a closed-loop operating mode of the insulin infusion device; in response to initiating the closed-loop operating mode, obtaining a most recent sensor glucose value for the user; and calculating a difference between the most recent sensor glucose value and a target glucose setpoint value. When the calculated difference is less than or equal to a minimum threshold value, the method adjusts a closed-loop insulin infusion rate over time, based on a fixed final target glucose value that is derived from the target glucose setpoint value. In contrast, when the calculated difference is greater than the minimum threshold value, the method adjusts the closed-loop insulin infusion rate over time, based on a dynamic final target glucose value that decreases over time toward the target glucose setpoint value.
Also presented here is a tangible and non-transitory electronic storage medium having processor-executable instructions that, when executed by a processor architecture, perform a method of controlling an insulin infusion device for a user. The method involves: initiating a closed-loop operating mode of the insulin infusion device; in response to initiating the closed-loop operating mode, obtaining a most recent sensor glucose value for the user; and calculating a difference between the most recent sensor glucose value and a target glucose setpoint value. When the calculated difference is less than or equal to a minimum threshold value, a closed-loop insulin infusion rate is adjusted over time, based on a fixed final target glucose value that is derived from the target glucose setpoint value. When the calculated difference is greater than the minimum threshold value, the closed-loop insulin infusion rate is adjusted over time, based on a dynamic final target glucose value that decreases over time toward the target glucose setpoint value.
Another exemplary embodiment of an electronic device is also presented here. The electronic device includes a processor architecture and at least one memory element associated with the processor architecture. The at least one memory element stores processor-executable instructions that, when executed by the processor architecture, perform a method of controlling an insulin infusion device for a user. The method initiates a closed-loop operating mode of the insulin infusion device and, in response to initiating the closed-loop operating mode, obtains a most recent sensor glucose value for the user. The method continues by calculating a difference between the most recent sensor glucose value and a target glucose setpoint value. When the calculated difference is less than or equal to a minimum threshold value, the method adjusts a closed-loop insulin infusion rate over time, based on a fixed final target glucose value that is derived from the target glucose setpoint value. When the calculated difference is greater than the minimum threshold value, the method adjusts the closed-loop insulin infusion rate over time, based on a dynamic final target glucose value that decreases over time toward the target glucose setpoint value.
Another exemplary embodiment of an electronic controller for an insulin infusion device is also provided here. The electronic controller includes: a processor architecture; and at least one memory element associated with the processor architecture, the at least one memory element storing processor-executable instructions that, when executed by the processor architecture, provide a closed-loop start-up module. The start-up module is configured to: initiate a closed-loop operating mode of the insulin infusion device; in response to initiating the closed-loop operating mode, obtain a most recent sensor glucose value for the user; and calculate a difference between the most recent sensor glucose value and a target glucose setpoint value. When the calculated difference is less than or equal to a minimum threshold value, the start-up module adjusts a closed-loop insulin infusion rate over time, based on a fixed final target glucose value that is derived from the target glucose setpoint value. When the calculated difference is greater than the minimum threshold value, the start-up module adjusts the closed-loop insulin infusion rate over time, based on a dynamic final target glucose value that decreases over time toward the target glucose setpoint value.
Yet another processor-implemented method of controlling an insulin infusion device for a user is presented here. The method calculates a maximum insulin infusion rate for the user based on a fasting blood glucose value associated with the user, a total daily insulin value associated with the user, and fasting insulin delivery data that is indicative of insulin delivered to the user during a fasting period. The maximum insulin infusion rate is applicable during a period of closed-loop operation of the insulin infusion device. The method continues by obtaining a first closed-loop insulin infusion rate for the user, wherein the first closed-loop insulin infusion rate is obtained for a current sampling point during the period of closed-loop operation. The method provides a second closed-loop insulin infusion rate for the user when the obtained first closed-loop insulin infusion rate is greater than the calculated maximum insulin infusion rate, wherein the second closed-loop insulin infusion rate is less than the first closed-loop insulin infusion rate.
Also provided is a tangible and non-transitory electronic storage medium having processor-executable instructions that, when executed by a processor architecture, perform a method of controlling an insulin infusion device for a user. The method involves: calculating a maximum insulin infusion rate for the user based on a fasting blood glucose value associated with the user, a total daily insulin value associated with the user, and fasting insulin delivery data that is indicative of insulin delivered to the user during a fasting period, wherein the maximum insulin infusion rate is applicable during a period of closed-loop operation of the insulin infusion device. The method obtains a first closed-loop insulin infusion rate for the user, wherein the first closed-loop insulin infusion rate is obtained for a current sampling point during the period of closed-loop operation. The method continues by providing a second closed-loop insulin infusion rate for the user when the obtained first closed-loop insulin infusion rate is greater than the calculated maximum insulin infusion rate, wherein the second closed-loop insulin infusion rate is less than the first closed-loop insulin infusion rate.
An additional embodiment of an electronic device is also presented here. The electronic device includes: a processor architecture having at least one processor device; and at least one memory element associated with the processor architecture, the at least one memory element storing processor-executable instructions that, when executed by the processor architecture, perform a method of controlling an insulin infusion device for a user. The method involves: calculating a maximum insulin infusion rate for the user based on a fasting blood glucose value associated with the user, a total daily insulin value associated with the user, and fasting insulin delivery data that is indicative of insulin delivered to the user during a fasting period, wherein the maximum insulin infusion rate is applicable during a period of closed-loop operation of the insulin infusion device; obtaining a first closed-loop insulin infusion rate for the user, wherein the first closed-loop insulin infusion rate is obtained for a current sampling point during the period of closed-loop operation; and providing a second closed-loop insulin infusion rate for the user when the obtained first closed-loop insulin infusion rate is greater than the calculated maximum insulin infusion rate, wherein the second closed-loop insulin infusion rate is less than the first closed-loop insulin infusion rate.
Another embodiment of an electronic controller for an insulin infusion device is also provided. The electronic controller includes: a processor architecture having at least one processor device; and at least one memory element associated with the processor architecture, the at least one memory element storing processor-executable instructions that, when executed by the processor architecture, provide a closed-loop insulin limit module. The closed-loop insulin limit module is configured to: obtain a fasting blood glucose value associated with the user, the fasting blood glucose value corresponding to a fasting period for the user; obtain a total daily insulin value associated with the user; obtain fasting insulin delivery data that is indicative of insulin delivered to the user during the fasting period; and calculate a maximum insulin infusion rate for the user based on the obtained fasting blood glucose value, the obtained total daily insulin value, and the obtained fasting insulin delivery data, wherein the maximum insulin infusion rate is applicable during a period of closed-loop operation of the insulin infusion device.
This summary is provided to introduce a selection of concepts in a simplified form that are further described below in the detailed description. This summary is not intended to identify key features or essential features of the claimed subject matter, nor is it intended to be used as an aid in determining the scope of the claimed subject matter.
A more complete understanding of the subject matter may be derived by referring to the detailed description and claims when considered in conjunction with the following figures, wherein like reference numbers refer to similar elements throughout the figures.
The following detailed description is merely illustrative in nature and is not intended to limit the embodiments of the subject matter or the application and uses of such embodiments. As used herein, the word “exemplary” means “serving as an example, instance, or illustration.” Any implementation described herein as exemplary is not necessarily to be construed as preferred or advantageous over other implementations. Furthermore, there is no intention to be bound by any expressed or implied theory presented in the preceding technical field, background, brief summary or the following detailed description.
Techniques and technologies may be described herein in terms of functional and/or logical block components, and with reference to symbolic representations of operations, processing tasks, and functions that may be performed by various computing components or devices. Such operations, tasks, and functions are sometimes referred to as being computer-executed, computerized, software-implemented, or computer-implemented. It should be appreciated that the various block components shown in the figures may be realized by any number of hardware, software, and/or firmware components configured to perform the specified functions. For example, an embodiment of a system or a component may employ various integrated circuit components, e.g., memory elements, digital signal processing elements, logic elements, look-up tables, or the like, which may carry out a variety of functions under the control of one or more microprocessors or other control devices.
When implemented in software or firmware, various elements of the systems described herein are essentially the code segments or instructions that perform the various tasks. The program or code segments can be stored in any tangible and non-transitory processor-readable medium. The “processor-readable medium” or “machine-readable medium” may include any medium that can store or transfer information. Examples of the processor-readable medium include an electronic circuit, a semiconductor memory device, a ROM, a flash memory, an erasable ROM (EROM), a floppy diskette, a CD-ROM, an optical disk, a hard disk, or the like.
The various tasks performed in connection with a process described herein may be performed by software, hardware, firmware, or any combination thereof. It should be appreciated that a described process may include any number of additional or alternative tasks, the tasks shown in a particular figure need not be performed in the illustrated order, and a described process may be incorporated into a more comprehensive procedure or process having additional functionality not described in detail herein. Moreover, one or more of the tasks shown in the figures could be omitted from an embodiment of a described process as long as the intended overall functionality remains intact.
As shown in the drawings for purposes of illustration, the invention is embodied in a closed loop infusion system for regulating the rate of fluid infusion into a body of a user based on feedback from an analyte concentration measurement taken from the body. In particular embodiments, the invention is embodied in a control system for regulating the rate of insulin infusion into the body of a user based on a glucose concentration measurement taken from the body. In preferred embodiments, the system is designed to model a pancreatic beta cell (β-cell). In other words, the system controls an infusion device to release insulin into a body of a user in a similar concentration profile as would be created by fully functioning human β-cells when responding to changes in blood glucose concentrations in the body.
Thus, the system simulates the body's natural insulin response to blood glucose levels and not only makes efficient use of insulin, but also accounts for other bodily functions as well since insulin has both metabolic and mitogenic effects. However, the algorithms must model the β-cells closely, since algorithms that are designed to minimize glucose excursions in the body, without regard for how much insulin is delivered, may cause excessive weight gain, hypertension, and atherosclerosis. In preferred embodiments of the present invention, the system is intended to emulate the in vivo insulin secretion pattern and to adjust this pattern consistent with the in vivo β-cell adaptation experienced by normal healthy individuals. The in vivo β-cell response in subjects with normal glucose tolerance (NGT), with widely varying insulin sensitivity (SI), is the optimal insulin response for the maintenance of glucose homeostasis.
Preferred embodiments include a glucose sensor system 10, a controller 12 and an insulin delivery system 14, as shown in
Generally, the glucose sensor system 10 includes a glucose sensor, sensor electrical components to provide power to the sensor and generate the sensor signal 16, a sensor communication system to carry the sensor signal 16 to the controller 12, and a sensor system housing for the electrical components and the sensor communication system.
Typically, the controller 12 includes controller electrical components and software to generate commands for the insulin delivery system 14 based on the sensor signal 16, and a controller communication system to receive the sensor signal 16 and carry commands to the insulin delivery system 14.
Generally, the insulin delivery system 14 includes an infusion device and an infusion tube to infuse insulin 24 into the body 20. In particular embodiments, the infusion device includes infusion electrical components to activate an infusion motor according to the commands 22, an infusion communication system to receive the commands 22 from the controller 12, and an infusion device housing to hold the infusion device.
In preferred embodiments, the controller 12 is housed in the infusion device housing and the infusion communication system is an electrical trace or a wire that carries the commands 22 from the controller 12 to the infusion device. In alternative embodiments, the controller 12 is housed in the sensor system housing and the sensor communication system is an electrical trace or a wire that carries the sensor signal 16 from the sensor electrical components to the controller electrical components. In other alternative embodiments, the controller 12 has its own housing or is included in a supplemental device. In another alternative embodiment, the controller is located with the infusion device and the sensor system all within one housing. In further alternative embodiments, the sensor, controller, and/or infusion communication systems may utilize a cable, a wire, fiber optic lines, RF, IR, or ultrasonic transmitters and receivers, or the like instead of the electrical traces.
System Overview
Preferred embodiments of the invention include a sensor 26, a sensor set 28, a telemetered characteristic monitor 30, a sensor cable 32, an infusion device 34, an infusion tube 36, and an infusion set 38, all worn on the body 20 of a user, as shown in
The controller 12 processes the digital sensor values Dsig and generates commands 22 for the infusion device 34. Preferably, the infusion device 34 responds to the commands 22 and actuates a plunger 48 that forces insulin 24 out of a reservoir 50 located inside the infusion device 34, as shown in
In alternative embodiments, the closed-loop system can be a part of a hospital-based glucose management system. Given that insulin therapy during intensive care has been shown to dramatically improve wound healing, reduce blood stream infections, renal failure, and polyneuropathy mortality, irrespective of whether subjects previously had diabetes (See Van den Berghe G. et al., NEJM 345: 1359-67, 2001, which is incorporated by reference herein), the present invention can be used in this hospital setting to control the blood glucose level of a patient in intensive care. In these alternative embodiments, since an intravenous (IV) hookup is typically implanted into a patient's arm while the patient is in an intensive care setting (e.g., ICU), a closed loop glucose control can be established which piggy-backs off the existing IV connection. Thus, in a hospital based system, IV catheters which are directly connected to a patient vascular system for purposes of quickly delivering IV fluids, can also be used to facilitate blood sampling and direct infusion of substances (e.g., insulin, anticoagulants) into the intra-vascular space. Moreover, glucose sensors may be inserted through the IV line to give real-time glucose levels from the blood stream. Therefore, depending on the type of hospital based system, the alternative embodiments would not necessarily need the described system components such as the sensor 26, the sensor set 28, the telemetered characteristic monitor 30, the sensor cable 32, the infusion tube 36, and the infusion set 38 as described in the preferred embodiments. Instead, standard blood glucose meters or vascular glucose sensors as described in provisional application entitled “Multi-lumen Catheter,” filed Sep. 27, 2002, Ser. No. 60/414,248, which is incorporated herein in its entirety by reference, can be used to provide the blood glucose values to the infusion pump control and the existing IV connection can be used to administer the insulin to the patient.
It is important to appreciate that numerous combinations of devices in the hospital-based system can be used with the closed loop controller of the present invention. For example, as described in
In other modifications to the system, a manual blood glucose/intravenous insulin infusion system can be used where frequent manual entry of blood glucose values from a standard blood glucose meter (e.g., YSI, Beckman, etc.) and extrapolate the values at more frequent intervals (preferably 1 min) to create a surrogate signal for calculating IV-insulin infusion. Alternatively, a sensor blood glucose/intravenous insulin infusion system can use a continuous glucose sensor (e.g., vascular, subcutaneous, etc.) for frequent blood glucose determination. Moreover, the insulin infusion can be administered subcutaneously rather than intravenously in any one of the previous examples according to the controller described below.
In still further alternative embodiments, the system components may be combined in a smaller or greater number of devices and/or the functions of each device may be allocated differently to suit the needs of the user.
Controller
Once the hardware for a closed loop system is configured, such as in the preferred embodiments described above, the effects of the hardware on a human body are determined by the controller. In preferred embodiments, the controller 12 is designed to model a pancreatic beta cell (β-cell). In other words, the controller 12 commands the infusion device 34 to release insulin 24 into the body 20 at a rate that causes the insulin concentration in the blood to follow a similar concentration profile as would be caused by fully functioning human β-cells responding to blood glucose concentrations in the body 20. In further embodiments, a “semi-closed-loop” system may be used, in which the user is prompted to confirm insulin delivery before any insulin is actually delivered.
A controller that simulates the body's natural insulin response to blood glucose levels not only makes efficient use of insulin but also accounts for other bodily functions as well since insulin has both metabolic and mitogenic effects. Controller algorithms that are designed to minimize glucose excursions in the body without regard for how much insulin is delivered may cause excessive weight gain, hypertension, and atherosclerosis. In preferred embodiments, of the present invention, the controller 12 is intended to emulate the in vivo insulin secretion pattern and to adjust this pattern to be consistent with in vivo β-cell adaptation. The in vivo β-cell response in subjects with normal glucose tolerance (NGT), with widely varying insulin sensitivity (SI), is the optimal insulin response for the maintenance of glucose homeostasis.
The β-Cell and PID Control
Generally, the in vivo β-cell response to changes in glucose is characterized by “first” and “second” phase insulin responses. This biphasic insulin response is clearly seen during hyperglycemic clamps applied to NGT subjects, as shown in
The biphasic insulin response of a β-cell can be modeled using components of a proportional, plus integral, plus derivative (PID) controller. A PID controller is selected since PID algorithms are stable for a wide variety of non-medical dynamic systems, and PID algorithms have been found to be stable over widely varying disturbances and changes in system dynamics.
The insulin response of β-cells during a hyperglycemic clamp is diagrammed in
Proportional Component Response: UP=KP(G−GB)
Integral Component Response: UI=KI∫t
Derivative Component Response:
Where
UP is the proportional component of the command sent to the insulin delivery system,
UI is the integral component of the command sent to the insulin delivery system,
UD is the derivative component of the command sent to the insulin delivery system,
KP is a proportional gain coefficient,
KI is an integral gain coefficient,
KD is a derivative gain coefficient,
G is a present blood glucose level,
GB is a desired basal glucose level,
t is the time that has passed since the last sensor calibration,
t0 is the time of the last sensor calibration, and
IB is a basal insulin concentration at t0, or can also be described as UI(t0).
The combination of the PID components that model the two phases of insulin response by a β-cell is shown in
The components of the PID controller can also be expressed in its discrete form:
Proportional Component Response: Pconn=KP(SGfn−Gsp)
Integral Component Response: Iconn=Iconn−1+KI(SGfn−Gsp); Icon0=Ib
Derivative Component Response: Dconn=KDdGdtfn
Where KP, KI, and KD are the proportional, integral, and derivative gain coefficients, SGf and dGdtf are the filtered sensor glucose and derivative respectively, and the superscript n refers to discrete time.
An acute insulin response is essential for preventing wide postprandial glycemic excursions. Generally, an early insulin response to a sudden increase in glucose level results in less total insulin being needed to bring the glucose level back to a desired basal glucose level. This is because the infusion of insulin increases the percentage of glucose that is taken up by the body. Infusing a large amount of insulin to increase the percentage of glucose uptake while the glucose concentration is high results in an efficient use of insulin. Conversely, infusing a large amount of insulin while the glucose concentration is low results in using a large amount of insulin to remove a relatively small amount of glucose. In other words, a larger percentage of a big number is more than a larger percentage of a small number. The infusion of less total insulin helps to avoid development of insulin resistance in the user. As well, first-phase insulin is thought to result in an early suppression of hepatic glucose output.
Insulin sensitivity is not fixed and can change dramatically in a body depending on the amount of exercise by the body. In one study, for example, insulin responses in highly exercise-trained individuals (individuals who trained more than 5 days a week) were compared to the insulin responses in subjects with normal glucose tolerance (NGT) during a hyperglycemic clamp. The insulin response in exercise-trained individuals 444 was about one-half of the insulin response of the NGT subjects 446, as shown in
In preferred embodiments, a closed loop control system may be used for delivering insulin to a body to compensate for β-cells that perform inadequately. There is a desired basal blood glucose level GB for each body. The difference between the desired basal blood glucose level GB and an estimate of the present blood glucose level G is the glucose level error GE that must be corrected. The glucose level error GE is provided as an input to the controller 12, as shown in
If the glucose level error GE is positive (meaning that the present estimate of the blood glucose level G is higher than the desired basal blood glucose level GB) then the controller 12 generates an insulin delivery command 22 to drive the infusion device 34 to provide insulin 24 to the body 20. In terms of the control loop, glucose is considered to be positive, and therefore insulin is negative. The sensor 26 senses the ISF glucose level and generates a sensor signal 16. The sensor signal 16 is filtered and calibrated to create an estimate of the present blood glucose level 452. In particular embodiments, the estimate of the present blood glucose level G is adjusted with correction algorithms 454 before it is compared to the desired basal blood glucose level GB to calculate a new glucose level error GE to start the loop again.
If the glucose level error GE is negative (meaning that the present estimate of the blood glucose level is lower than the desired basal blood glucose level GB) then the controller 12 reduces or stops the insulin delivery depending on whether the integral component response of the glucose error GE is still positive.
If the glucose level error GE is zero, (meaning that the present estimate of the blood glucose level is equal to the desired basal blood glucose level GB) then the controller 12 may or may not issue commands to infuse insulin depending on the derivative component (whether the glucose level is raising or falling) and the integral component (how long and by how much glucose level has been above or below the basal blood glucose level GB). In “semi-closed loop” embodiments, the user is prompted before the controller 12 issues the commands to infuse insulin. The prompts may be displayed to the user on a display, sounded to the user, or otherwise provide an indication to the user that the system is ready to deliver insulin, for example a vibration or other tactile indication. In addition, the amount of insulin to be delivered may be displayed, with or without other information, such as the total amount infused for the day or the potential effect on the user's blood glucose level by the insulin delivery. In response, the user may indicate that the insulin should or should not be delivered, for example by selecting a button, key, or other input. In further embodiments, there must be at least two keystrokes so that insulin is not delivered by accident.
To more clearly understand the effects that the body has on the control loop, a more detailed description of the physiological effects that insulin has on the glucose concentration in the interstitial fluid (ISF) is needed. In preferred embodiments, the infusion device 34 delivers insulin through the cannula 56 of the infusion set 38 into the ISF of the subcutaneous tissue 44 of the body 20. And the insulin 24 diffuses from the local ISF surrounding the cannula into the blood plasma and then spreads throughout the body 20 in the main circulatory system, as described in the block diagram of
In addition, insulin has direct and indirect effects on liver glucose production. Increased insulin concentration decreases liver glucose production. Therefore, acute and immediate insulin response not only helps the body to efficiently take up glucose but also substantially stops the liver from adding to the glucose in the blood stream. In alternative embodiments, insulin is delivered more directly into the blood stream instead of into the interstitial fluid, such as delivery into veins, arteries, the peritoneal cavity, or the like. And therefore, any time delay associated with moving the insulin from the interstitial fluid into the blood plasma is diminished. In other alternative embodiments, the glucose sensor is in contact with blood or body fluids other than interstitial fluid, or the glucose sensor is outside of the body and measures glucose through a non-invasive means. The embodiments that use alternative glucose sensors may have shorter or longer delays between the blood glucose level and the measured blood glucose level.
Selecting Controller Gains
In preferred embodiments, the controller gains KP, KI, and KD, are selected so that the commands from the controller 12 cause the infusion device 34 to release insulin 24 into the body 20 at a rate, that causes the insulin concentration in the blood to follow a similar concentration profile, as would be caused by fully functioning human β-cells responding to blood glucose concentrations in the body. In preferred embodiments, the gains may be selected by observing the insulin response of several normal glucose tolerant (NGT) individuals, with healthy normally functioning β-cells. The first step in determining a set of controller gains is to take periodic measurements of blood glucose and blood insulin concentrations from the group of NGT individuals. Second, each individual in the group is subjected to a hyperglycemic clamp, while continuing to periodically measure and record the blood glucose and blood insulin concentrations. Third, a least squares curve fit is applied to the recorded blood insulin concentrations measured over time for each individual. The result is a set of curves representing the insulin responses to the hyperglycemic clamp for each individual of the group. Fourth, the curves are used to calculate the controller gains KP, KI, and KD, for each individual. And finally, the proportional gains from each of the individuals are averaged together to obtain an average proportional gain, KP, to be used in a controller 12. Similarly, the integral gains, KI, and the derivative gains, KD, are averaged to obtain an average integral gain, KI, and an average derivative gain, KD, for the controller 12. Alternatively, other statistical values may be used instead of averages such as, maximums, minimums, the high or low one, two or three sigma standard deviation values, or the like. The gains calculated for various individuals in a group may be filtered to remove anomalous data points before statistically calculating the gains to be used in a controller.
In an example, a least squares curve-fitting method is used to generate representative insulin response curves from two fasted individuals in a group, as shown in
The controller gains may be expressed in various units and/or may be modified by conversion factors depending on preferences for British or S. I. Units, floating-point or integer software implementation, the software memory available, or the like. The set of units for the controller gains in Table 1 is:
KP: (mU of insulin)/min/(Kg of body weight) per (mg of glucose)/(dl of plasma);
KI: (mU of insulin)/min/(Kg of body weight) per (mg of glucose)/(dl of plasma) min.; and
KD: (mU of insulin)/min/(Kg of body weight) per (mg of glucose)/(dl of plasma)/min.
In alternative embodiments, other curve fitting methods are used to generate the insulin response curves from the measurements of blood insulin concentrations.
An estimate of an insulin clearance rate (k), the individual's body weight (W), and the insulin sensitivity SI are needed to calculate the controller gains from the insulin response curves for each NGT individual. The insulin clearance rate (k) is generally proportional to body weight and is well documented in literature. The individual's insulin sensitivity SI may be measured using an intravenous glucose tolerance test, a hyperinsulinemic clamp, or in the case of a diabetic, comparing the individual's daily insulin requirement to their daily carbohydrate intake.
In particular embodiments, two parameters, the insulin sensitivity SI and the insulin clearance rate k, are measured for each individual. In other embodiments, the insulin clearance rate k is estimated from literature given the individual's body weight. In other particular embodiments, longer or shorter insulin clearance times are used. In still other embodiments, all of the parameters are estimated. In additional embodiments, one or more parameters are measured, while at least one parameter is estimated from literature.
In other alternative embodiments, the controller gains are calculated using a group of individuals with similar body types. For example, the insulin response to a hyperglycemic clamp may be measured for several tall, thin, NGT, males in order to calculate the controller insulin response gains for each individual in the group. Then the gains are statistically combined to generate a set of representative controller gains for tall, thin, NGT, males. The same could be done for other groups such as, but not limited to, short, heavy, NGT, females; medium height, medium weight, highly exercised trained, females; average height and weight 10 year olds; or the like. Then the controller gains are selected for each individual user based on the group that best represents them. In further alternative embodiments, controller gains are uniquely selected for each individual user. In particular embodiments, the controller gains for a user are selected based on measurements of insulin sensitivity, insulin clearing time, insulin appearance time, insulin concentration, body weight, body fat percentage, body metabolism, or other body characteristics such as pregnancy, age, heart conditions, or the like.
In other alternative embodiments, the controller gains are estimated as a function of a user's body weight W and insulin sensitivity SI. A series of observations are used to justify this method. The first observation is that the controller gains are proportional to each other. In other words, small changes in glucose concentration cause a small derivative response UD, a small proportional response UP and a small integral response UI. And larger changes in glucose concentration cause a proportionally larger derivative response UD, a proportionally larger proportional UP response and a proportionally larger integral response UI, as shown in
While there are multiple sources for the information needed to calculate the insulin clearance rate k, one source is the article “Insulin clearance during hypoglycemia in patients with insulin-dependent diabetes mellitus”, written by Kollind M et al., published in Horm Metab Res, 1991 July; 23(7):333-5. The insulin clearance rate k is obtained from the insulin infused divided by the steady state plasma insulin concentration. An insulin clearance constant Ak, which is independent of an individual's body weight, may be obtained by dividing the insulin clearance rate k (measured from a particular individual) by the individual's body weight. The insulin clearance constant Ak is generally the same for all humans, except under extenuating circumstances such as after an individual has contracted HIV, other metabolic affecting diseases, or the like.
The disposition index (DI) for a human given a change in glucose concentration is available from information presented in the article “Quantification of the relationship between insulin sensitivity and beta-cell function in human subjects. Evidence for a hyperbolic function”, written by Khan S E et al., published in Diabetes, 1993 November; 42(11):1663-72.
Both the disposition index DI and the insulin clearance rate k may be measured directly from tests. The disposition index DI may be calculated given the first phase insulin response measured form a glucose clamp test and the individual's insulin sensitivity measured from an insulin sensitivity test. The insulin clearance rate k may be measured from an insulin clearance test. The glucose clamp test and the insulin clearance test are described in the above-mentioned articles and are well known in the art. The insulin sensitivity SI may be measured using an intravenous glucose tolerance test or a hyperinsulinemic clamp test.
Given these observations, then the following parameters may be measured from an NGT individual's insulin response to a glucose clamp: a desired first phase insulin response φ1, the ratio of KD to KP, and the ratio of KD to KI. Then the derivative gain KD may be calculated from the first phase insulin response φ1 using the constants k and DI. And finally KP and KI may be calculated using the ratios of KD to KP and KD to KI.
The first phase insulin response φ1 may be observed in a NGT individual as the area under the insulin response curve during approximately the first 10 minutes of a glucose clamp. The increase in the glucose concentration during the glucose clamp is ΔG=(G−GB), where G is equal to Gc, the glucose concentration during the clamp, and GB is the basal glucose concentration before the clamp.
The importance of the first phase insulin response φ1 has been emphasized by studies indicating that, in subjects with normal glucose tolerance (NGT), the product of first phase insulin response φ1 and insulin sensitivity (SI) is a constant known as the disposition index, DI=φ1SI. Therefore,
For a different ΔG there is a different φ1 and therefore a different DI. But, the ratio DI/ΔG is substantially constant even for different individuals with different insulin sensitivities.
The insulin sensitivity SI is defined as the percentage of the glucose concentration that the body tissues will take up for a given amount of insulin. The β-cell naturally adapts to changes in insulin sensitivity by adjusting the amount of insulin it secretes during the first phase insulin response φ1. This suggests that the body naturally seeks an optimal level of glucose tolerance. A controller that mimics this characteristic of the β-cell more accurately simulates the body's natural insulin response.
The instantaneous insulin response (RI) may be calculated given the insulin clearance rate (k) and the first phase insulin response φ1, RI=kφ1.
The insulin clearance rate k is proportional to body weight (W), therefore substituting a proportional constant Ak and the user's body weight W for k and replacing φ1 with the ratio of DI over SI yields the following equation:
The instantaneous insulin response RI may also be expressed as the product of the derivative gain KD and the change in glucose concentration ΔG, RI=KDΔG.
Setting the two equations for RI equal to each other and solving for KD yields,
As mentioned above, DI/ΔG and Ak are constants available or calculated from data in published literature. Combining the constants into a single constant, Q,
yields an equation for the derivative gain KD that is a function of the user's body weight W and the user's insulin sensitivity SI,
Once the derivative gain KD is calculated, the proportional and integral gains are calculated using ratios. The ratio of KD/KP can be set to the dominant time constant for insulin action, ranging from 10-60 minutes, but more typically 20-40 minutes and preferably 30 minutes. For example, calculating KP given KD using a time constant of 30 minutes, yields the following relationship:
In a similar fashion, the ratio of KD/KI can be set to the average ratio measured from a population of NGT individuals. And KI can be calculated from KD.
In particular embodiments, the user enters their body weight W and insulin sensitivity SI into the device that contains the controller. Then the controller gains are automatically calculated and used by the controller. In alternative embodiments, an individual enters the user's body weight W and insulin sensitivity SI into a device and the device provides the information to the controller to calculate the gains.
A study was conducted to confirm that the insulin response for an individual could be reproduced using the glucose sensor as an input. In the study, glucose and insulin measurements were taken while a hyperglycemic clamp was applied to a NGT individual. The glucose level measurements, shown in
Three different devices were used to measure the individual's blood glucose during the study. Blood glucose meter readings 460 from periodic blood samples taken from the individual are represented by the dots in
Fuzzy Logic to Select Between Multiple Sets of Controller Gains
In preferred embodiments, one set of controller gains is used for a particular individual. In alternative embodiments, more than one set of controller gains is used, and fuzzy logic is used to select between sets of controller gains and to determine when to change from one set of controller gains to another. In particular alternative embodiments, the controller gains are different if the glucose level is above or below the desired glucose basal level. In other alternative embodiments, the controller gains are different if the glucose level is increasing or decreasing. A justification for different sets of gains comes from physiological studies that indicate that β-cells turn off faster than they turn on. In still other alternative embodiments, the controller gains are different depending on whether the glucose level is above or below the desired glucose basal level and whether the glucose level is increasing or decreasing, which results in four sets of controller gains. In additional alternative embodiments, the controller gains change depending on the magnitude of the hypoglycemic excursion. In other words, the controller gains for small changes in glucose are different than those for large changes in glucose.
Self-Tuning Controller Gains
Further embodiments may include a controller that self-tunes one or more the gains, KP, KI, KD to accommodate changes in insulin sensitivity. In particular embodiments, previous measurements of glucose levels are compared to the desired basal glucose level GB. For example, the desired basal glucose level GB is subtracted from the previous glucose level measurements. Then any negative values, within a predefined time window, are summed (in essence integrating the glucose level measurements that were below the basal glucose level GB). If the resulting sum is greater than a pre-selected hypoglycemic integral threshold, then the controller gains are increased by a factor (1+α). Conversely, if the integral of the glucose level measurements that were measured above the basal glucose level GB within the predefined time window is greater than a pre-selected hyperglycemic integral threshold, then the controller gains are decreased by a factor (1−α).
In particular embodiments, the predefined time window over which the glucose concentration integrals are evaluated is generally 24 hours, and the controller gains are adjusted if needed at the end of each predefined time window. In alternative embodiments, the integrals of the glucose level measurements are continuously calculated over a moving window of time, and if either integral exceeds a threshold, the gains are immediately adjusted. In particular embodiments, the moving time window is one hour, and the time window may be restarted whenever the gains are adjusted. In other alternative embodiments, the time window is longer or shorter depending on the sensor accuracy, the rate at which an individual's insulin sensitivity changes, the computational capabilities of the hardware, or the like.
In particular embodiments, the adjustment amount (α) is 0.01. In alternative embodiments, the adjustment amount α is greater or smaller depending on the sensor accuracy, the rate at which an individual's insulin sensitivity changes, the rate at which the sensor sensitivity SI changes, or the like. In still other alternative embodiments, the adjustment amount α is made larger or smaller depending on the amount that the integral of the measured glucose levels exceeds a threshold. In this way, the gains are adjusted by greater amounts if the measured glucose level G is significantly deviating from the desired blood glucose level GB and less if the measured glucose level G is closer to the desired blood glucose level GB. In additional alternative embodiments, the controller employs a Kalman filter.
State Variable Feedback
While the primary signal determining the β-cell's insulin response is glucose, there also exists a putative effect of insulin per se to inhibit insulin secretion. This effect may be directly related to the concentration of insulin in plasma (IP(t)), or mediated through some signal proportional to insulin effect (IEFF(t)). The β-cell can likely directly sense these signals (i.e., directly sense insulin concentration and secondary signals proportional to insulin effect such as free fatty acid). Feedback from these intermediary signals is analogous to what is known as state variable feedback; that is feedback, whereby the variable being controlled (glucose in this case) is used together with feedback of each intermediary signal that affects the variable (insulin concentration in plasma and interstitial fluid). With this type of feedback, undesirable slow kinetic process can be made to appear much faster than they are. For example, if β-cell insulin secretion were inhibited by a signal proportional to insulin concentration in the interstitial fluid where it acts, the delay between plasma and interstitial insulin could be made to appear to be shorter. For the artificial closed-loop algorithm, or for “semi-closed-loop” algorithms, this beneficial effect can be achieved by using “state observers” (mathematical equations that predict the insulin concentration in various parts of the body knowing the history of past insulin delivery). In “semi-closed loop” algorithms, the algorithms are the same as for closed loop algorithms but there is a user confirmation step before any insulin is actually administered. By using state variable feedback, it is possible to make the insulin in an insulin pump act faster than the insulin actually is.
To estimate subcutaneous insulin concentration, plasma insulin concentration, and insulin effect, the following equations may be used:
Wherein ISC is the estimate of normalized insulin concentration in the subcutaneous space, IP is the estimate of normalized insulin concentration in the plasma, IEF is the estimate of insulin effect on glucose, α1 is the rate constant between insulin delivery and the subcutaneous insulin compartment, α2 is the rate constant between subcutaneous insulin and plasma compartments, α3 is the rate constant between the plasma compartment and the insulin effect. ID is the delivered insulin, which can be a function of the three state variables (ISC, IP, and IEF).
In particular embodiments, an open loop fixed base rate plus user requested bolus would result in the bolus being increased a certain amount and the basal rate subsequently decreased the same amount in accordance to the following formula:
ID′=(1+γ1+γ2+γ3)/ID−γ1ISC−γ2IP−γ3IEF
Wherein ID is the user requested basal (U/h) plus bolus (U) profile and ID′ is the state feedback adjusted profiles. Note that for a given kinetic excursion the total amount of insulin requested (area under curve of ID) and delivered (area under curve of ID′) is identical. Here, γ1, γ2, and γ3 are state-feedback gains (scalars). Careful choice of these gains the pump to correct its delivery rate to compensate for delays associated with the dispersion of insulin from the bolus injection into the subcutaneous layer of the patient, to the plasma, and to its actual insulin effect/action on the body. Thus, by estimating how much insulin from a bolus is in the subcutaneous layer, the plasma, or is actually acting on the patient's glucose level (state variables ISC, IP and IEF), it is possible to optimize delivery of insulin over time to the patient. Using state feedback the bolus is increased by an amount (1+γ1+γ2+γ3) that is gradually taken away from future insulin delivery (−γ1ISC−γ2IP−γ3IEF). As a result, the apparent insulin pharmokinetic curve appears faster. This is akin to developing a faster acting insulin, but it is achieved algorithmically by rearranging the distribution of the insulin delivery per unit bolus by delivering more upfront and removing the extra amount at a later time. The three gains can be chosen to move the time delays (1/α1, 1/α2, and 1/α3) to any arbitrary locations. In control theory, this is known as pole placement.
State feedback can be used in open loop and closed loop insulin delivery algorithms and with “semi-closed-loop” delivery algorithms. State feedback can be used in conjunction with a Proportional-Integral-Derivative (PID) or any other type of closed loop controller. γ1 is the feedback gain multiplied to ISG, γ2 is the feedback gain multiplied to IP, and γ3 is the feedback gain multiplied to IEF.
The physical state space form directly taken from the equations above is:
The finite difference form is calculated as follows (wherein eX indicates an exponential function):
Define: k1=e−α
ISC=(1−k1)(ID(i−1))+k1ISC(i−1) (eq 1b)
IP(i)=(1−k2)(ISC(i))+k2IP(i−1) (eq 2b)
IEF(i)=(1−k3)(IP(i))+k3IEF(i−1) (eq 3b)
The Laplace Form is as follows, wherein s represents the Stäckel determinant used in Laplace equations:
To obtain the transfer function of insulin delivery with state feedback, the control equation is as follows, wherein E represents the error between the actual glucose concentration and the desired glucose concentration (G−GD):
ID=PID·E−γ1ISC−γ2IP−γ3IEFF (eq 6)
Substituting equations (eq 1c), (eq 4) and (eq 5) into (eq 6) and rearranging, the following transfer functions are obtained, wherein GM is a gain multiplier:
The computation of the gain multiplier is also obtained in the state variable feedback method. When state variable feedback is used, the gain multiplier (GM) is a scalar that forces a step response to reach the same steady value whether state feedback is used or not. In other words, GM ensures that the total amount given per unit of bolus will be the same in both cases. In the case of state feedback, more insulin is given up front, but this extra insulin is taken away later. To calculate GM in particular embodiments, the “final value theorem” from control systems is used. The final value theorem states that to evaluate the steady state of any transfer function T(s) given any input X(s), the steady state output response to the input is given by:
γSS(t→∞)=lims→0(sT(s)X(s))
The Laplace form of a step input is given by
and the steady state solution of the final value theorem simplifies to:
γSS(t→∞)=lims→0(T(s)).
In the case when there is no state feedback, (γ1, γ2 and γ3=0), the steady state solution may be obtained from equation (eq 7) to be as follows:
ID(t→∞)=1 (eq 11)
With state feedback without the gain correction factor, the steady state solution is:
GM is then evaluated as the ratio of equation (eq 12) to equation (eq 11) to obtain the following: GM=1+γ1+γ2+γ3.
Using state variable feedback, a closed loop control equation and state feedback gain are determined for pole placement. Specifically, the gains are calculated for the insulin delivery equation shown above. In particular embodiments, they are determined as follows: First, with state feedback, the denominator of equations (eq 7), (eq 8), (eq 9), and (eq 10) is:
D=s3+(α1+α2+α3+γ1α1)s2+(α1α2+(α1+α2)α3+γ2α1α2+(α2+α3)γ1α1)s+(α1α2α3+γ3α1α2α3+γ2α1α2α3+γ1α1α2α3) (eq 14)
To get the poles of the system in the equations (eq 7), (eq 8), (eq 9), or (eq 10), D may be set equal to zero yield the characteristic equation:
s3+(α1+α2+α3+γ1α1)s2+(α1α2+(α1+α2)α3+γ2α1α2+(α2+α3)γ1α1)s+(α1α2α3+γ3α1α2α3+γ2α1α2α3+γ1α1α2α3)=0 (eq 16)
If the desired system poles or roots of (eq 16) are defined by eigenvalues λ1, λ2 and λ3, then the characteristic equation can be written as:
(s−λ1)(s−λ2)(s−λ3)=0
Expanding and collecting like powers of s, (eq 16) can be written as:
s3−(λ1+λ2+λ3)s2+(λ1λ2+λ1λ3+λ2λ3)s−λ1λ2λ3=0 (eq 17)
Setting the coefficients of like powers of s equal to each other we have the system of equations:
α1+α2+α3+γ1α1=−(λ1+λ2+λ3) (eq 18)
α1α2+α1α3+α2α3+γ2α1α2+γ1α1(α2+α3)=λ1λ2+λ1λ3+λ2λ3 (eq 19)
α1α2α3+γ3α1α2α3+γ2α1α2α3+γ1α1α2α3=λ1λ2λ3 (eq 20)
This results in three equations and three unknowns, γ1, γ2 and γ3. Therefore, the unknown gains can be solved for in terms of the desired poles λ1, λ2, λ3, and system time constants, α1, α2 and α3. These formulas enable us to control the desired pharmacokinetics of insulin as it appears in the different compartments:
Thus, through the above calculations, the gains can be calculated and used in the control equation for insulin delivery of:
ID=PID·E−γ1ISC−γ2IP−γ3IEF
PID is the output of a PID controller of any other closed loop (or “semi-closed-loop”) controller. The gains are generally calculated just once, but could be calculated more often if desired. The control equation may be calculated on a repeated basis, after predetermined periods of time or continuously. For example, and without limitation, it may be calculated every five, ten, thirty or sixty minutes. Just the state variable portion (γ1ISC−γ2IP−γ3IEF) may be updated or the entire equation may be updated. By updating the control equation, it is possible to continually improve the delivery of insulin to the patient.
A control feedback block diagram of an embodiment of a pump using state variable feedback is shown in
Modifying the PID Controller to Incorporate an Integrator Leak
In preferred embodiments, the PID control response was described with constant gain components, KP, KI, KD. Although the preferred control response guarantees zero steady-state error (i.e., steady state glucose minus a desired basal glucose (GB)=0), inherently, the integral component UI/=KI∫t
with initial condition UI(t0).
The parameter KLEAK is the reciprocal time constant of the rate of leaking (τLEAK in min=1/KLEAK), where τLEAK is a tuning parameter that can be set based on empirical data, and be tied with the other gain components KP, KI, KD. However, the current realization of the artificial β-cell has τLEAK as a user input. UI can also be expressed in discrete form by standard methods.
Post-Controller (Lead/Lag) Compensator
In preferred embodiments, commands are issued from the controller without regard to where in the body the insulin delivery system will infuse the insulin. In essence, the assumption is that the insulin is either delivered directly into the blood stream for immediate use by the body, or that any time delays caused by delivering the insulin somewhere in the body other than the blood stream can be compensated for by adjusting KP, KI, and KD. In this case, the commands generally model a β-cell insulin secretion profile, an example of which is shown in
where:
The time constant for insulin clearance P1 may be calculated using the following equation:
where:
Or the time constant for insulin clearance P1 may be obtained by providing insulin to an individual that does not generate his own insulin, and then periodically testing blood samples from the individual for insulin concentration. Then, using an exponential curve fitting routine, generate a mathematical expression for a best-fit curve for the insulin concentration measurements, and observe the time constant in the mathematical expression.
Given the same insulin dose (delivered at time zero all at once) into the subcutaneous tissue, instead of directly into the blood plasma, the concentration of insulin in the blood plasma would begin to rise slowly as insulin diffuses from the interstitial fluid ISF into the blood plasma, as shown in
A bi-exponential equation may be used to model the insulin concentration in blood plasma given an insulin dose delivered to the subcutaneous tissue:
where:
The time constants may be calculated using the quadratic formula:
In alternative embodiments, a post-controller lead-lag compensator 522 is used to modify the commands (UPID) to compensate for the insulin delivery delay and/or the insulin clearance rate k, as shown in
where 1/α and 1/γ are the lead and lag constants respectively, s is the Laplace variable, and UCOMP is the compensated commands calculated by the lead-lag compensator 522.
The PID controller generates commands (UPID) for a desired insulin delivery rate into the blood plasma. The commands UPID are calculated and issued periodically depending on the update rate for the control loop, which is selected based on a maximum anticipated rate of change of the blood glucose level, an insulin delivery system minimum insulin dosage, insulin sensitivity, a maximum and a minimum acceptable glucose concentration, or the like. The commands UPID are used as inputs to the post-controller lead-lag compensator 522.
In particular embodiments, the compensated commands (UCOMP) issued from the post-controller lead-lag compensator 522 uses more than one value from the controller. In particular embodiments, post-controller lead-lag compensator 522 uses the present command (UPIDn) and the previous command (UPIDn−1) to calculate a compensated command UCOMP per a compensation equation:
UCOMPn=(1−γ)UCOMPn−1UPIDn+(1−α)UPIDn−1, where:
This is a first forward difference equation. However, other forms can be used alternatively (e.g., first backward or bilinear), but all result in a compensated control output (UCOMP) that is comprised of a weighted history of both past PID outputs (UPID)), and past compensated outputs (UCOMP).
An alternative method of modifying the commands (UPID) to compensate for the insulin delivery delay and/or the insulin clearance can be performed based on a weighted history of past insulin delivery. By giving the most recent delivery history more weight, the weighted history of the previous insulin delivered can then be subtracted from the present PID control output to yield a compensated control output. Expressed in Laplace domain this results in:
where E is the Laplace transformed error signal (G−GB), λ determines how much the PID output is reduce in proportion to the weighted history of past control outputs, and α is the reciprocal time constant determining how long a history is weighted (the preferred value of α would be equal to the reciprocal dominant time constant or subcutaneous insulin appearance, P2). Solving the compensated signals as a function of the error results in:
which is identical to the previously described lead-lag compensation.
In other alternative embodiments, additional previous command values may be used. In still other alternative embodiments, the compensation equation compensates for both time constants P2 and P3.
In still more alternative embodiments, the controller gains are modified to include the effects of the post-controller lead/lag compensator so that the post-controller lead/lag compensator is not needed to modify the commands to account for the insulin delivery delay.
In particular embodiments, the insulin delivery system provides finite insulin doses into the body in response to commands from the controller. The smallest amount of insulin that the insulin delivery system can deliver is the minimum finite insulin dose. The controller may generate commands for a dose of insulin to be delivered that is not a whole number multiple of the minimum finite insulin dose. Therefore, either too much or too little insulin is delivered by the insulin delivery system in response to the commands. In particular alternative embodiments, the post-controller lead-lag compensator truncates the command to the nearest whole number multiple of the minimum finite insulin dose and adds the remaining commanded volume of insulin to the next command. In other alternative embodiments, a compensator rounds the command to the nearest whole number multiple of the minimum finite insulin dose. In still other alternative embodiments, other methods are used to compensate for the difference between the commands and the nearest whole number multiple of the minimum finite insulin dose. In other embodiments, no compensation is needed.
Eliminating the Lead-Lag Compensator with Feedback of Predicted Plasma Insulin
Yet in another alternative embodiment, the PID control commands may be modified to emulate the effect of plasma insulin on a β-cell to determine optimal insulin administration by feeding back a predicted plasma insulin based on the subcutaneous insulin infusion. The net effect of such feedback is to replace an undesired dynamic with a more desirable one and achieve a plasma insulin profile that a β-cell would achieve. This can be seen as follows (using Laplace transformed variables). Assume the relation between glucose above basal (G−GB) and insulin delivery (ID) is described by a linear transfer function ID(s)=C(s)(G(s)−GB), where C(s) may be, but is not necessarily, described by the PID controller transfer function. If the β-cell is using peripheral insulin (IP(s)) levels to suppress insulin secretion the predicted rate of insulin delivery would be modified as: ID(s)=C(s)(G(s)−GB)−kIp(s).
For portal insulin delivery the relation between ID(s) and plasma insulin IP(s) is known to be approximated by a single time delay:
Substituting IP(s) value into the previous formula and making k large results in:
which would completely cancel the undesirable time constant 1/α. In practice a lower value of k would be used resulting in:
where γ=α+kk1 (i.e., something greater than α). Thus, the effect for the β-cell, of adding a plasma insulin feedback is to replace the portal insulin delivery time constant (α) with a faster time constant (γ=α+kk1; γ>α). In block diagram form:
which is equivalent to:
To apply this mechanism to subcutaneous insulin delivery all that is needed is the transfer function between sc insulin delivery and plasma insulin. This transfer function is well approximated by a bi-exponential time course (bolus response) or:
in the limiting case as kk2/(s+α1)(s+α2)>>1 this is approximately equal to
where again, the undesirable time constants associated with subcutaneous insulin delivery have been eliminated. In practice they would just be replaced with more desirable rate constants (i.e., faster time constants).
Correction of Hypoglycemic Excursion Around ˜200 Minutes (Wind-Down)
Previous modeling of β-cells using a PID controller gave excellent predictability of the “first” and “second” phase insulin responses during prolonged periods of increased glucose appearance. However, if the periods of increased glucose appearance is followed by a rapid decrease in glucose appearance, the PID controller would not be able to correctly predict the wind down of the insulin response to lower glucose levels.
In preferred embodiments, the hypoglycemic excursion is corrected by modifying the PID controller to a PD control with Adaptive Proportional Gain (or Bilinear PID controller), which is modified form of the original PID equations. As described previously, the discrete PID algorithm is as follows:
Proportional Component Response:
Pconn=KP(SGfn−Gsp),
Integral Component Response:
Iconn=Iconn−1+KI(SGfn−Gsp);Icon0=Ib, and
Derivative Component Response:
Dconn=KDdGdtfn,
where KP, KI, and KD are the proportional, integral, and derivative gain coefficients, SGf and dGdtf are the filtered sensor glucose and derivative respectively, and the superscript n refers to discrete time.
In the Bilinear PID controller, the proportional gain KP is based on the integrated error term. The magnitude of each component's contribution to the insulin response is described by the following equations:
Pconn=KPn(SGfn−INT)
Dconn=KDdGdtfn
KPn=KPn−1+KI(SGfn−Gsp), where KP0=KP0
Where the proportional gain now integrates at rate KI (initial value KP0) and the proportional component is related to an intercept value (INT) where (INT<Gsp). The modified formulation can be seen to fit the hypoglycemic glucose excursion without systematic error as the adaptive PD line shown as a dashed line in
In additional embodiments, the Bilinear PID controller can also incorporate an integrator leak by modifying the formula to multiply the previous KP with a value such as α as follows:
KPn=αKPn−1+KI(SGfn−Gsp), where α≈0.99
An alternative method of correcting the hypoglycemic glucose excursion can be performed by integrator clip into the PID control. PID controllers generally have integrator-reset rules that prevent excessive “winding” and such a rule can be used to correct the hypoglycemic glucose excursion. For example, the integrator can be clipped as follows:
If (SG≦60 mg/dl AND Iconn−1>KP(SP−60)) then Iconn−1=KP(SP−60)
This equation resets the integrator such that if the sensor glucose falls below 60 mg/dl the insulin delivery is zero for all stable or falling sensor glucose signals. The clipping limit represents an absolute threshold, similar to the human counter regulatory response.
However, other approaches that may emulate the β-cell more accurately include the use of piecewise continuous functions. For example, the following function allows for progressive clipping to be tuned:
This equation introduces two additional tuning parameters (γ0 and T1) and starts to check the integrator output at a higher threshold. For example, if γ0=5 and T1=100 mg/dl, the integrator output would be clipped to 4 KP60 if glucose fell to 90 mg/dl, 3 KP60 if glucose fell to 80 mg/dl and so forth until glucose reached 60 where it would be clipped at KP60. Other functions than that proposed in the above equation (e.g. functions based on the rate of fall of glucose, or percent decrease in Icon) may alternatively be used.
System Configurations
The following sections provide exemplary, but not limiting, illustrations of components that can be utilized with the controller described above. Various changes in components, layout of various components, combinations of elements, or the like may be made without departing from the scope of the embodiments of the invention.
Before it is provided as an input to the controller 12, the sensor signal 16 is generally subjected to signal conditioning such as pre-filtering, filtering, calibrating, or the like. Components such as a pre-filter, one or more filters, a calibrator and the controller 12 may be split up or physically located together, and may be included with a telemetered characteristic monitor transmitter 30, the infusion device 34, or a supplemental device. In preferred embodiments, the pre-filter, filters and the calibrator are included as part of the telemetered characteristic monitor transmitter 30, and the controller 12 is included with the infusion device 34, as shown in
In preferred embodiments, the sensor system generates a message that includes information based on the sensor signal such as digital sensor values, pre-filtered digital sensor values, filtered digital sensor values, calibrated digital sensor values, commands, or the like. The message may include other types of information as well such as a serial number, an ID code, a check value, values for other sensed parameters, diagnostic signals, other signals, or the like. In particular embodiments, the digital sensor values Dsig may be filtered in the telemetered characteristic monitor transmitter 30, and then the filtered digital sensor values may be included in the message sent to the infusion device 34 where the filtered digital sensor values are calibrated and used in the controller. In other embodiments, the digital sensor values Dsig may be filtered and calibrated before being sent to the controller 12 in the infusion device 34. Alternatively, the digital sensor values Dsig may be filtered, and calibrated and used in the controller to generate commands 22 that are then sent from the telemetered characteristic monitor transmitter 30 to the infusion device 34.
In further embodiments, additional optional components, such as a post-calibration filter, a display, a recorder, and a blood glucose meter may be included in the devices with any of the other components or they may stand-alone. Generally, if a blood glucose meter is built into one of the devices, it will be co-located in the device that contains the calibrator. In alternative embodiments, one or more of the components are not used.
In preferred embodiments, RF telemetry is used to communicate between devices, such as the telemetered characteristic monitor transmitter 30 and the infusion device 34, which contain groups of components. In alternative embodiments, other communication mediums may be employed between devices such as wires, cables, IR signals, laser signals, fiber optics, ultrasonic signals, or the like.
Filtering
In preferred embodiments, the digital sensor values Dsig and/or the derivative of the digital sensor values are processed, filtered, modified, analyzed, smoothed, combined, averaged, clipped, scaled, calibrated, or the like, to minimize the effects of anomalous data points before they are provided as an input to the controller. In particular embodiments, the digital sensor values Dsig are passed through a pre-filter 400 and then a filter 402 before they are passed to the transmitter 70, as shown in
Pre-Filter
In particular embodiments, the pre-filter 400 uses fuzzy logic to determine if individual digital sensor values Dsig need to be adjusted. The pre-filter 400 uses a subset of a group of digital sensor values Dsig to calculate a parameter and then uses the parameter to determine if individual digital sensor values Dsig need to be adjusted in comparison to the group as a whole. For example, the average of a subset of a group of digital sensor values Dsig may be calculated, and then noise thresholds may be placed above and below the average. Then individual digital sensor values Dsig within the group are compared to noise thresholds and eliminated or modified if they are outside of the noise thresholds.
A more detailed example is provided below to more clearly illustrate, but not limit, an embodiment of a pre-filter. A group of eight digital sensor values Dsig are shown in
In preferred embodiments, when any of a group's values are outside of the noise thresholds 406 or 408 then a warning flag is set. If one to three values are outside of the noise thresholds 406 or 408, a “noise” flag is set. If more than three values are outside of the noise thresholds 406 or 408, a “discard” flag is set which indicates that the whole group of values should be ignored and not used. In alternative embodiments, more or less values need be outside of the thresholds 406 or 408 to trigger the “noise” flag or the “discard” flag.
In preferred embodiments, each digital sensor value Dsig is checked for saturation and disconnection. To continue with the example of
Similarly, in preferred embodiments, if a digital signal value Dsig is below a disconnect threshold 412, then a “disconnect” flag is set indicating to a user that the sensor is not properly connected to the power supply and that the power supply or sensor may need replacement or recalibration. In further particular embodiments, if a digital sensor value Dsig is below the disconnect threshold 412, the individual value may be ignored, changed to a value equal to the average line 404, or the entire group of values associated with the individual digital sensor value Dsig may be ignored. In preferred embodiments, the disconnect threshold 410 is set at about 20% of the full range of values. Higher or lower disconnect threshold levels may be used depending on an expected signal range of the sensor, sensor system noise, sensor gains, or the like.
In alternative embodiments, other methods are used to pre-filter the digital sensor values Dsig such as rate-of-change thresholds, rate-of-change squared thresholds, noise thresholds about a least squares fit line rather than about the average of a subset of a group's values, higher or lower noise threshold lines, or the like.
Noise Filter
After the digital sensor values Dsig are evaluated, and if necessary, modified by the pre-filter 400, the digital sensor values Dsig are passed to the filter 402. The filter 402 may be used to reduce noise in particular frequency bands. Generally the body's blood glucose level 18 changes relatively slowly compared to a rate at which digital sensor values Dsig are collected. Therefore, high frequency signal components are typically noise, and a low pass filter may be used to improve the signal to noise ratio.
In preferred embodiments, the filter 402 is a finite impulse response (FIR) filter used to reduce noise. In particular embodiments, the FIR filter is a 7th order filter tuned with a pass band for frequencies from zero to 3 cycles per hour (c/hr) and a stop band for frequencies greater than about 6 c/hr, as shown in an example frequency response curve 414 in
In alternative embodiments, other methods are used to pre-filter the digital sensor values Dsig such as rate-of-change thresholds, rate-of-change squared thresholds, noise thresholds about a least squares fit line rather than about the average of a subset of a group's values, higher or lower noise threshold lines, or the like.
Delay Compensation Filter
Aside from noise reduction, a filter may be used to compensate for time delays. Ideally, a sensor would provide a real time, noise-free measurement of a parameter that a control system is intended to control, such as a blood glucose measurement. However, realistically there are physiological, chemical, electrical, and algorithmic sources of time delays that cause the sensor measurement to lag behind the present value of blood glucose.
A physiological delay 422 is due to the time required for glucose to move between blood plasma 420 and interstitial fluid (ISF). The delay is represented by the circled double headed arrow 422 in
A chemical reaction delay 424 is introduced by the sensor response time, represented by the circle 424 surrounding the tip of the sensor 26 in
There is also a processing delay as the analog sensor signal Isig is converted to digital sensor values Dsig. In preferred embodiments, the analog sensor signal Isig is integrated over one-minute intervals and then converted to a number of counts. In essence an A/D conversion time results in an average delay of 30 seconds. In particular embodiments, the one-minute values are averaged into 5-minute values before they are sent to the controller. The resulting average delay is two and one half minutes. In alternative embodiments, longer or shorter integration times are used resulting in longer or shorter delay times. In other embodiments the analog sensor signal current Isig is continuously converted to an analog voltage Vsig and a A/D converter samples the voltage Vsig every 10 seconds. Then six 10-second values are pre-filtered and averaged to create a one-minute value. Finally, five 1-minute values are filtered and then averaged creating a five-minute value resulting in an average delay of two and one half minutes. Other embodiments use other electrical components or other sampling rates and result in other delay periods.
Filters also introduce a delay due to the time required to acquire a sufficient number of digital sensor values Dsig to operate the filter. Higher order filters, by definition, require more digital sensor values Dsig. Aside from the most recent digital sensor value Dsig, FIR filters use a number of previous values equal to the order of the filter. For example, a 7th order filter uses 8 digital sensor values Dsig. There is a time interval between each digital sensor value Dsig. To continue with the example, if the time interval between digital sensor values Dsig is one minute, then the oldest digital sensor value Dsig used in a 7th order FIR filter would be seven minutes old. Therefore, the average time delay for all of the values used in the filter is three and a half minutes. However, if the weighting factors associated with each of the values are not equal then the time delay may be longer or shorter than three and one half minutes depending on the effects of the coefficients.
Preferred embodiments of the invention include a FIR filter that compensates for both the various time delays, of up to about 30 minutes as discussed above, and high frequency noise, greater than about 10 c/hr also discussed above. Particular embodiments employ a 7th order Weiner type FIR filter. The coefficients for the filter are selected to correct for time lags while simultaneously reducing high frequency noise. An example of a frequency response curve 426 is shown in
In alternative embodiments, other types of filters may be used as long as they remove a sufficient portion of the noise from the sensor signal. In other alternative embodiments, no time compensation is used if the rate of change in the blood glucose level is slow compared to the time delay. For example, a five-minute delay between blood plasma glucose and a sensor measurement does not have to be corrected for a closed loop glucose control system to function.
Derivative Filter
Further embodiments may include a filter to remove noise from the derivative of the sensor signal before the controller uses it. A derivative is taken from the digital sensor values Dsig, which results in digital derivative sensor values (dDsig/dt). The digital derivative sensor values dDsig/dt are passed through a FIR filter. In particular embodiments, the derivative filter is at least a 7th order FIR filter tuned to remove high frequency noise. In alternative embodiments, higher or lower order filters may be used and the filters may be tuned to remove various frequencies of noise. In other alternative embodiments, a derivative is taken from the glucose level error GE values and then passed through a derivative filter 526, as shown in
Calibration
In preferred embodiments, after filtering, the digital sensor values Dsig are calibrated with respect to one or more glucose reference values. The glucose reference values are entered into the calibrator and compared to the digital sensor values Dsig. The calibrator applies a calibration algorithm to convert the digital sensor values Dsig, which are typically in counts into blood glucose values. In particular embodiments, the calibration method is of the type described in U.S. patent application Ser. No. 09/511,580, filed on Feb. 23, 2000, entitled “GLUCOSE MONITOR CALIBRATION METHODS”, which is incorporated by reference herein. In particular embodiments, the calibrator is included as part of the infusion device 34 and the glucose reference values are entered by the user into the infusion device 34. In other embodiments, the glucose reference values are entered into the telemetered characteristic monitor transmitter 30 and the calibrator calibrates the digital sensor values Dsig and transmits calibrated digital sensor values to the infusion device 34. In further embodiments, the glucose reference values are entered into a supplemental device where the calibration is executed. In alternative embodiments, a blood glucose meter is in communication with the infusion device 34, telemetered characteristic monitor transmitter 30 or supplemental device so that glucose reference values may be transmitted directly into the device that the blood glucose meter is in communication with. In additional alternative embodiments, the blood glucose meter is part of the infusion device 34, telemetered characteristic monitor transmitter 30 or supplemental device such as that shown in U.S. patent application Ser. No. 09/334,996, filed on Jun. 17, 1999, entitled “CHARACTERISTIC MONITOR WITH A CHARACTERISTIC METER AND METHOD OF USING THE SAME”, which is incorporated by reference herein.
In preferred embodiments, to obtain blood glucose reference values, one or more blood samples are extracted from the body 20, and a common, over-the-counter, blood glucose meter is used to measure the blood plasma glucose concentration of the samples. Then a digital sensor value Dsig is compared to the blood glucose measurement from the meter and a mathematical correction is applied to convert the digital sensor values Dsig to blood glucose values. In alternative embodiments, a solution of a known glucose concentration is introduced into the subcutaneous tissue surrounding the sensor 26 by using methods and apparatus such as described in U.S. patent application Ser. No. 09/395,530, filed on Sep. 14, 1999, entitled “METHOD AND KIT FOR SUPPLYING A FLUID TO A SUBCUTANEOUS PLACEMENT SITE”, which is incorporated by reference herein, or by using injection, infusion, jet pressure, introduction through a lumen, or the like. A digital sensor value Dsig is collected while the sensor 26 is bathed in the solution of known glucose concentration. A mathematical formula such as a factor, an offset, an equation, or the like, is derived to convert the digital sensor value Dsig to the known glucose concentration. The mathematical formula is then applied to subsequent digital sensors values Dsig to obtain blood glucose values. In alternative embodiments, the digital sensor values Dsig are calibrated before filtering. In additional alternative embodiments, the digital sensor values Dsig are calibrated after pre-filtering and before filtering. In other alternative embodiments, the sensors are calibrated before they are used in the body or do not require calibration at all.
Sensor Signal Processing Systems
Before filtering and calibrating, generally the sensor signal is processed to convert the sensor signal from a raw form into a form acceptable for use in the filters and/or calibrator. In preferred embodiments, as shown in
A/D Converters
Various A/D converter designs may be used in embodiments of the present invention. The following examples are illustrative, and not limiting, since other A/D converters may be used.
I to F (Current to Frequency (Counts)), Single Capacitor, Quick Discharge
In preferred embodiments, the integrator 72 consists of a first Op-Amp 92 and a capacitor 82, shown in
When the capacitor voltage (A′) is discharged to VrefL, the comparator output (B′) returns to low, thus forming a pulse. The low comparator output (B′) opens the reset switch 84 allowing the capacitor 82 to begin charging again.
Virtually simultaneously, the low comparator output (B′) also triggers the reference voltage switch 88 to open and the inverter output (C′) triggers reference voltage switch 90 to close resulting in changing the comparator reference voltage from VrefL back to VrefH.
I to F, Single Reversible Capacitor
In alternative embodiments, two or more integrator switches are used to control the polarity of one or more capacitors. A particular embodiment is shown in
The high comparator output (B″) pulse causes the capacitor polarity to reverse using the following method. The high comparator output (B″) triggers the second integrator switch 112 to close while virtually simultaneously the inverter 118 inverts the comparator output (B″). And the low inverter output (C″) pulse triggers the first integrator switch 110 to open. Once the capacitor's polarity is reversed, the capacitor 116 discharges at a rate proportional to the analog sensor signal Isig. The high comparator output (B″) pulse also triggers the reference voltage of the comparator to change form VrefH the low reference voltage (VrefL). When the capacitor voltage (A″) is discharged to VrefL, the comparator output (B″) returns to low. The low comparator output (B″) opens the second integrator switch 112 and virtually simultaneously the high inverter output (C″) closes the first integrator switch 110 allowing the capacitor 116 to begin charging again. The low comparator output (B″) also triggers the comparator reference voltage to change from VrefL back to VrefH.
An advantage of this embodiment is that sensor signal errors, which may be created due to capacitor discharge time, are reduced since the magnitude of the analog sensor signal Isig drives both the charging and the discharging rates of the capacitor 116.
I to F, Dual Capacitor
In further alternative embodiments, more than one capacitor is used such that as one capacitor is charging, at a rate proportional to the magnitude of the analog sensor signal Isig, another capacitor is discharging. An example of this embodiment is shown in
In summary, as the blood glucose level 18 increases, the analog sensor signal Isig increases, which causes the voltage coming out of the integrator 72 to ramp up faster to the high reference voltage VrefH, which causes the comparator 74 to generate pulses more often, which adds counts to the counter 76 faster. Therefore, higher blood glucose levels generate more counts per minute.
The charge storage capacity for the capacitors used in the integrator 72, and the reference voltages VrefH, and VrefL are selected such that the count resolution for counts collected in a one-minute period at a glucose level of 200 mg/dl represents a blood glucose measurement error of less than 1 mg/dl. In particular embodiments, VrefH is 1.1 volts and VrefL is 0.1 volts. Higher or lower reference voltages may be selected based on the magnitude of the analog sensor signal Isig, the capacity of the capacitors, and the desired measurement resolution. The source voltage V+ is set to a voltage sufficiently high to discharge one or more capacitors quickly enough that the discharge times do not significantly reduce the number of counts per minute at a blood glucose level of 200 mg/dl.
Pulse Duration Output Feature
In preferred embodiments, the transmitter 70 transmits the digital sensor values Dsig from the buffer 78 whenever triggered by the clock 80. However, in particular embodiments, the user or another individual may use a selector 96 to choose other outputs to be transmitted from the transmitter 70, as shown in
In particular embodiments, a pulse duration counter 98 counts clock pulses from a pulse duration clock 100 until the pulse duration counter 98 is reset by a rising or falling edge of a pulse from the comparator 74, as shown in
I to V (Current to Voltage), Voltage A/D
Alternative methods may be used to convert the analog sensor signal Isig from an analog current signal to a digital voltage signal. The analog sensor signal Isig is converted to an analog voltage Vsig using an Op Amp 302 and a resistor 304, as shown in
In alternative embodiments, a buffer 310 is not needed and the digital sensor values Dsig are sent from the A/D converter directly to the transmitter 70. In other alternative embodiments, the digital sensor values Dsig are processed, filtered, modified, analyzed, smoothed, combined, averaged, clipped, scaled, calibrated, or the like, before being sent to the transmitter 70. In preferred embodiments, the clock 308 triggers a measurement every 10 seconds. In alternative embodiments, the clock 308 runs faster or slower triggering measurements more or less frequently depending on how quickly the blood glucose level can change, the sensor sensitivity, how often new measurements are needed to control the delivery system 14, or the like.
Finally, in other alternative embodiments, other sensor signals from other types of sensors, as discussed in the section “Sensor and Sensor Set” below, are converted to digital sensor values Dsig if necessary before transmitting the digital sensor values Dsig to another device.
Additional Controller Inputs
Generally, the proportional plus, integral plus, derivative (PID) insulin response controller uses only glucose (digital sensor values Dsig) as an input. Conversely, in a normally glucose tolerant human body, healthy β-cells benefit from additional inputs such as neural stimulation, gut hormone stimulation, changes in free fatty acid (FFA) and protein stimulation etc. Thus in other alternative embodiments, the PID controller, as discussed above, can be augmented with one or more additional inputs. In particular alternative embodiments, the user may manually input supplemental information such as a start of a meal, an anticipated carbohydrate content of the meal, a start of a sleep cycle, an anticipated sleep duration, a start of an exercise period, an anticipated exercise duration, an exercise intensity estimation, or the like. Then, a model predictive control feature assists the controller to use the supplemental information to anticipate changes in glucose concentration and modify the output commands accordingly. For example, in a NGT individual, neural stimulation triggers the β-cells to begin to secrete insulin into the blood stream before a meal begins, which is well before the blood glucose concentration begins to rise. So, in alternative embodiments, the user can tell the controller that a meal is beginning and the controller will begin to secrete insulin in anticipation of the meal.
In other alternative embodiments, the user or another individual may manually override the control system or select a different controller algorithm. For instance, in particular alternative embodiments, an individual may select to normalize to a basal glucose level immediately, and instead of using the β-cell emulating PID controller another controller would take over such as a PID controller with different gains, a PD controller for rapid glucose adjustment, or the like. Additional alternative embodiments allow an individual to turn off the integral component of the PID controller once the glucose level is normalized and no meals are anticipated. In other particular alternative embodiments, the user may select to turn off the controller entirely, therefore disengaging the closed loop system. Once the closed loop system is not controlling insulin dosing, the user may program the infusion device with a basal rate, variable basal rates, boluses, or the like, or the user may manually enter each individual dosage when it is needed.
In still other alternative embodiments, more than one body characteristic is measured, and the measurements are provided as inputs to a controller. Measured body characteristics that may be used by the controller include, but are not limited to, the blood glucose level, blood and/or ISF pH, body temperature, the concentration of amino acids in blood (including arginine and/or lysine, and the like), the concentration of gastrointestinal hormones in blood or ISF (including gastrin, secretin, cholecystokinin, and/or gastro inhibitory peptide, and the like), the concentration of other hormones in blood or ISF (including glucagons, growth hormone, cortisol, progesterone and/or estrogen, and the like), blood pressure, body motion, respiratory rate, heart rate, and other parameters.
In NGT individuals, the glucose-induced secretion of insulin by healthy β-cells may be as much as doubled in the presence of excess amino acids. Yet, the presence of excess amino acids alone, without elevated blood glucose, only mildly increases insulin secretions according to the Textbook of Medical Physiology, Eighth Edition, written by Arthur C. Guyton, published by W. B. Saunders Company, 1991, Ch. 78, pg. 861, section “Other Factors That Stimulate Insulin Secretion”. In particular alternative embodiments, amino acid concentrations are estimated or measured, and the controller's insulin response increases when amino acid concentrations are sufficiently high.
In NGT individuals, the presence of sufficient quantities of gastrointestinal hormones in the blood causes an anticipatory increase in blood insulin, which suggests that β-cells release insulin before increases in blood glucose due to an individual's anticipation of a meal. In particular alternative embodiments, the concentration of gastrointestinal hormones is measured or estimated, and when concentrations are high enough to indicate that a meal is anticipated, the controller commands are adjusted to cause insulin introduction into the body even before the blood glucose level changes. In other alternative embodiments, the controller uses measurements or estimates of other hormones to modify the rate of insulin secretion.
In NGT individuals, the body's cells take up glucose during periods of heavy exercise with significantly lower levels of insulin. In alternative embodiments, physiologic parameters such as body motion, blood pressure, pulse rate, respiratory rate, or the like, are used to detect periods of heavy exercise by the body and therefore provide inputs to the controller that decreases (or eliminates) the amount of insulin infused into the body to compensate for glucose concentrations.
Sensor Compensation and End-of-Life Detection
In particular embodiments, the sensor sensitivity 510 may degrade over time, as shown in
As the sensor resistance Rs increases, the analog sensor signal Isig response to a given glucose concentration decreases. In preferred embodiments, the decrease in the analog sensor signal Isig may be compensated for by identifying the amount that the sensor resistance Rs has changed since the last calibration and then using the change in resistance in a correction algorithm 454 to adjust the analog sensor signal value. A compensation value calculated by the correction algorithm 454 is used to increase the sensor analog signal value. The compensation value increases over time as the sensor resistance Rs increases. The correction algorithm 454 includes at least one value that varies with changes in sensor resistance Rs. In particular embodiments, a low pass filter is applied to the sensor resistance Rs measurement to decrease high frequency noise before evaluating how much the sensor resistance Rs has changed since the last calibration.
In alternative embodiments, the sensor resistance Rs may be calculated using different equations. For instance, a sensor resistance Rs2 may be calculated as:
Rs2=(V0−Vcnt/Isig)
In particular embodiments, V0 is the same voltage as Vset. An advantage of this approach is that it accounts for the voltage level Vset, which can vary from sensor to sensor and/or monitor to monitor, and/or as the analog sensor signal changes. This removes the noise and/or offset associated with variations in Vset, and can provide a more accurate indication of sensor resistance. In other particular embodiments, V0 is set at −0.535 volts, which is a commonly used voltage for Vset. In further embodiments, V0 is calculated from paired measurements of Vcnt and Isig. Using least squares or another curve fitting method, a mathematical equation representing the curve (typically a straight line equation) is derived from the relationship between Vcnt and Isig. Then, V0 is obtained by extrapolating the curve to find the value for Vcnt when Isig is zero.
In preferred embodiments, the sensor is recalibrated or replaced when the change in the sensor resistance Rs since the last calibration exceeds a threshold, or the rate of change of the sensor resistance dRs/dt exceeds another threshold. In particular embodiments, the rate of change of the sensor resistance dRs/dt may be compared to two thresholds as shown in
In an example shown in
In particular embodiments, the amount of time dt over which the derivative of the sensor resistance Rs is taken is the entire time since the last calibration. In other embodiments, the amount of time dt over which the derivative is taken is fixed, for example over the last hour, 90 minutes, 2 hours, or the like.
In alternative embodiments, the sensor is recalibrated or replaced when the integral of the sensor resistance Rs over a predetermined time window (∫ Rs d/dt) exceeds a predetermined resistance integral threshold. An advantage to this approach is that it tends to filter out potential noise that could be encountered from a signal that includes occasional spikes, sudden variations in voltage levels, or the like. Preferably, the integral of the sensor resistance Rs is calculated over a time window (such as 15 minutes, or the like) based on Rs measurements obtained at set rates (such as 1 minute, 5 minutes, or the like) during the time window. In alternative embodiments, the time windows may be longer or shorter and different sampling rates may be used, with the selection being dependent on noise, response of the system, sampling rate used in the controller, or the like. In further embodiments, the time windows and sampling rates may change over time, such as when approaching the end of the expected sensor life, or as the equations indicate that the sensor is degrading, or the like.
Like above, multiple thresholds may be used. For instance, if ∫Rs d/dt exceeds a “replacement” threshold then a warning is provided to the user to replace the sensor. And if ∫Rs d/dt exceeds a “recalibrate” threshold then a warning is provided to the user to recalibrate the sensor. In further alternative embodiments, the counter electrode voltage Vcnt is used to evaluate other characteristics such as, sensor accuracy, sensor bio-fouling, sensor function, sensor voltage operating range, sensor attachment, or the like.
pH Controller Input
In alternative embodiments, the controller uses measurements of both the interstitial fluid (ISF) glucose level and a local pH in the ISF surrounding the sensor to generate commands for the infusion device. In particular alternative embodiments, a single multi-sensor 508 located in the subcutaneous tissue is used to measure both the glucose level and the pH. The tip of the multi-sensor 508 that is placed into the subcutaneous tissue with three electrodes is shown in
In additional alternative embodiments, the pH level near the sensor may be used to detect infection. By tracking trends in the pH over time, a dramatic change in pH may be used to identify that an infection has developed in proximity to the sensor. A warning is used to notify the user to replace the sensor.
The pH sensor may be used in other embodiments. When insulin is not available to assist the body to use glucose, the body shifts to consuming fat for energy. As the body shifts from using glucose to using almost exclusively fat for energy, concentrations of keto acids (acetoacetic acid and β-hydroxybutyric acid) increase from about 1 mEq/liter to as high as 10 mEq/liter. In particular alternative embodiments, the pH level is measured to detect increases in keto acids in the body. In embodiments of the present invention, a warning is provided to the user when the ISF pH level is too low.
A side effect of the increased of keto acid concentrations is that sodium is drawn from the body's extra cellular fluid to combine with the acids so that the body can excrete the acids. This leads to increased quantities of hydrogen ions, which greatly increases the acidosis. Severe cases lead to rapid deep breathing, acidotic coma and even death. In other alternative embodiments, an ion-selective electrode (ISE) is used to detect changes in sodium concentration. A special membrane is used to coat the ISE so that it only senses changes in sodium concentration. In particular alternative embodiments, the ISE is a fourth electrode added to the glucose sensor. In another alternative embodiment, a three-electrode system is used with a silver-silver chloride reference electrode REF, an Ir Ox counter electrode CNT, and a sodium ion-selective (Na ISE) working electrode WRK.
While pH measurements, end-of-life measurements, hormone measurements, or the like, add inputs to the controller that can significantly affect the accuracy of insulin delivery, the basic input to the controller is generally a glucose measurement. The glucose measurement is provided by the sensor system. And once the controller uses the glucose measurement to generate commands, the delivery system executes the commands. The following is a detailed description of several apparatus embodiments for the sensor system and the delivery system.
Sensor System
The sensor system provides the glucose measurements used by the controller. The sensor system includes a sensor, a sensor set to hold the sensor if needed, a telemetered characteristic monitor transmitter, and a cable if needed to carry power and/or the sensor signal between the sensor and the telemetered characteristic monitor transmitter.
Sensor and Sensor Set
In preferred embodiments, the glucose sensor system 10 includes a thin film electrochemical sensor such as the type disclosed in U.S. Pat. No. 5,391,250, entitled “METHOD OF FABRICATING THIN FILM SENSORS”; U.S. patent application Ser. No. 09/502,204, filed on Feb. 10, 2000, entitled “IMPROVED ANALYTE SENSOR AND METHOD OF MAKING THE SAME”; or other typical thin film sensors such as described in commonly assigned U.S. Pat. Nos. 5,390,671; 5,482,473; and 5,586,553 which are incorporated by reference herein. See also U.S. Pat. No. 5,299,571.
The glucose sensor system 10 also includes a sensor set 28 to support the sensor 26 such as described in U.S. Pat. No. 5,586,553, entitled “TRANSCUTANEOUS SENSOR INSERTION SET” (published as PCT Application WO 96/25088); and U.S. Pat. No. 5,954,643, entitled “INSERTION SET FOR A TRANSCUTANEOUS SENSOR” (published as PCT Application WO 98/56293); and U.S. Pat. No. 5,951,521, entitled “A SUBCUTANEOUS IMPLANTABLE SENSOR SET HAVING. THE CAPABILITY TO REMOVE OR DELIVER FLUIDS TO AN INSERTION SITE”, which are incorporated by reference herein.
In preferred embodiments, the sensor 26 is inserted through the user's skin 46 using an insertion needle 58, which is removed and disposed of once the sensor is positioned in the subcutaneous tissue 44. The insertion needle 58 has a sharpened tip 59 and an open slot 60 to hold the sensor during insertion into the skin 46, as shown in
In preferred embodiments, the sensor 26 has three electrodes 42 that are exposed to the interstitial fluid (ISF) in the subcutaneous tissue 44 as shown in
The most prominent reaction stimulated by the voltage difference between the electrodes is the reduction of glucose as it first reacts with GOX to generate gluconic acid and hydrogen peroxide (H2O2). Then the H2O2 is reduced to water (H2O) and (O−) at the surface of the working electrode WRK. The O− draws a positive charge from the sensor electrical components, thus repelling an electron and causing an electrical current flow. This results in the analog current signal Isig being proportional to the concentration of glucose in the ISF that is in contact with the sensor electrodes 42. The analog current signal Isig flows from the working electrode WRK, to the counter electrode CNT, typically through a filter and back to the low rail of an op-amp 66. An input to the op-amp 66 is the set voltage Vset. The output of the op-amp 66 adjusts the counter voltage Vcnt at the counter electrode CNT as Isig changes with glucose concentration. The voltage at the working electrode WRK is generally held to ground, the voltage at the reference electrode REF is generally equal to Vset, and the voltage Vcnt at the counter electrode CNT varies as needed.
In alternative embodiments, more than one sensor is used to measure blood glucose. In particular embodiments, redundant sensors are used. The user is notified when a sensor fails by the telemetered characteristic monitor transmitter electronics. An indicator may also inform the user of which sensors are still functioning and/or the number of sensors still functioning. In other particular embodiments, sensor signals are combined through averaging or other means. If the difference between the sensor signals exceeds a threshold then the user is warned to recalibrate or replace at least one sensor. In other alternative embodiments, more than one glucose sensor is used, and the glucose sensors are not of the same design. For example, an internal glucose sensor and an external glucose sensor may be used to measure blood glucose at the same time.
In alternative embodiments, other continuous blood glucose sensors and sensor sets may be used. In particular alternative embodiments, the sensor system is a micro needle analyte sampling device such as described in U.S. patent application Ser. No. 09/460,121, filed on Dec. 13, 1999, entitled “INSERTION SET WITH MICROPIERCING MEMBERS AND METHODS OF USING THE SAME”, incorporated by reference herein, or an internal glucose sensor as described in U.S. Pat. Nos. 5,497,772; 5,660,163; 5,791,344; and 5,569,186, and/or a glucose sensor that uses florescence such as described in U.S. Pat. No. 6,011,984 all of which are incorporated by reference herein. In other alternative embodiments, the sensor system uses other sensing technologies such as described in Patent Cooperation Treaty publication No. WO 99/29230, light beams, conductivity, jet sampling, micro dialysis, micro-poration, ultra sonic sampling, reverse iontophoresis, or the like. In still other alternative embodiments, only the working electrode WRK is located in the subcutaneous tissue and in contact with the ISF, and the counter CNT and reference REF electrodes are located external to the body and in contact with the skin. In particular embodiments, the counter electrode CNT and the reference electrode REF are located on the surface of a monitor housing 518 and are held to the skin as part of the telemetered characteristic monitor, as shown in
Sensor Cable
In preferred embodiments, the sensor cable 32 is of the type described in U.S. patent application Ser. No. 60/121,656, filed on Feb. 25, 1999, entitled “TEST PLUG AND CABLE FOR A GLUCOSE MONITOR”, which is incorporated by reference herein. In other embodiments, other cables may be used such as shielded, low noise cables for carrying nA currents, fiber optic cables, or the like. In alternative embodiments, a short cable may be used or the sensor may be directly connected to a device without the need of a cable.
Telemetered Characteristic Monitor Transmitter
In preferred embodiments, the telemetered characteristic monitor transmitter 30 is of the type described in U.S. patent application Ser. No. 09/465,715, filed on Dec. 17, 1999, entitled “TELEMETERED CHARACTERISTIC MONITOR SYSTEM AND METHOD OF USING THE SAME” (published as PCT Application WO 00/19887 and entitled, “TELEMETERED CHARACTERISTIC MONITOR SYSTEM”), which is incorporated by reference herein, and is connected to the sensor set 28 as shown in
In alternative embodiments, the sensor cable 32 is connected directly to the infusion device housing, as shown in
In other alternative embodiments, the telemetered characteristic monitor transmitter includes a receiver to receive updates or requests for additional sensor data or to receive a confirmation (a hand-shake signal) indicating that information has been received correctly. Specifically, if the telemetered characteristic monitor transmitter does not receive a confirmation signal from the infusion device, then it re-sends the information. In particular alternative embodiments, the infusion device anticipates receiving blood glucose values or other information on a periodic basis. If the expected information is not supplied when required, the infusion device sends a “wake-up” signal to the telemetered characteristic monitor transmitter to cause it to re-send the information.
Insulin Delivery System
Infusion device
Once a sensor signal 16 is received and processed through the controller 12, commands 22 are generated to operate the infusion device 34. In preferred embodiments, semi-automated medication infusion devices of the external type are used, as generally described in U.S. Pat. Nos. 4,562,751; 4,678,408; 4,685,903; and U.S. patent application Ser. No. 09/334,858, filed on Jun. 17, 1999, entitled “EXTERNAL INFUSION DEVICE WITH REMOTE PROGRAMMING, BOLUS ESTIMATOR AND/OR VIBRATION CAPABILITIES” (published as PCT application WO 00/10628), which are herein incorporated by reference. In alternative embodiments, automated implantable medication infusion devices, as generally described in U.S. Pat. Nos. 4,373,527 and 4,573,994, are used, which are incorporated by reference herein.
Insulin
In preferred embodiments, the infusion device reservoir 50 contains HUMALOG® lispro insulin to be infused into the body 20. Alternatively, other forms of insulin may be used such as HUMALIN®, human insulin, bovine insulin, porcine insulin, analogs, or other insulins such as insulin types described in U.S. Pat. No. 5,807,315, entitled “METHOD AND COMPOSITIONS FOR THE DELIVERY OF MONOMERIC PROTEINS”, and U.S. Patent Application Ser. No. 60/177,897, filed on Jan. 24, 2000, entitled “MIXED BUFFER SYSTEM FOR STABILIZING POLYPEPTIDE FORMULATIONS”, which are incorporated by reference herein, or the like. In further alternative embodiments, other components are added to the insulin such as polypeptides described in U.S. patent application Ser. No. 09/334,676, filed on Jun. 25, 1999, entitled “MULTIPLE AGENT DIABETES THERAPY”, small molecule insulin mimetic materials such as described in U.S. patent application Ser. No. 09/566,877, filed on May 8, 2000, entitled “DEVICE AND METHOD FOR INFUSION OF SMALL MOLECULE INSULIN MIMETIC MATERIALS”, both of which are incorporated by reference herein, or the like.
Infusion Tube
In preferred embodiments, an infusion tube 36 is used to carry the insulin 24 from the infusion device 34 to the infusion set 38. In alternative embodiments, the infusion tube carries the insulin 24 from infusion device 34 directly into the body 20. In further alternative embodiments, no infusion tube is needed, for example if the infusion device is attached directly to the skin and the insulin 24 flows from the infusion device, through a cannula or needle directly into the body. In other alternative embodiments, the infusion device is internal to the body and an infusion tube may or may not be used to carry insulin away from the infusion device location.
Infusion Set
In preferred embodiments, the infusion set 38 is of the type described in U.S. Pat. No. 4,755,173, entitled “SOFT CANNULA SUBCUTANEOUS INJECTION SET”, which is incorporated by reference herein. In alternative embodiments, other infusion sets, such described in U.S. Pat. Nos. 4,373,527 and 4,573,994, are used, which are incorporated by reference herein. In alternative embodiments, other infusion sets, such as the Rapid set from Disetronic, the Silhouette from MiniMed, or the like, may be used. In further alternative embodiments, no infusion set is required, for example if the infusion device is an internal infusion device or if the infusion device is attached directly to the skin.
Configurations with Supplemental Devices
In further alternative embodiments, the pre-filter, filters, calibrator and/or controller 12 are located in a supplemental device that is in communication with both the telemetered characteristic monitor transmitter 30 and the infusion device 34. Examples of supplemental devices include, a hand held personal digital assistant such as described in U.S. patent application Ser. No. 09/487,423, filed on Jan. 20, 2000, entitled “HANDHELD PERSONAL DATA ASSISTANT (PDA) WITH A MEDICAL DEVICE AND METHOD OF USING THE SAME”, which is incorporated by reference herein, a computer, a module that may be attached to the telemetered characteristic monitor transmitter 30, a module that may be attached to the infusion device 34, a RF programmer such as described in U.S. patent application Ser. No. 09/334,858, filed on Jun. 17, 1999, entitled EXTERNAL INFUSION DEVICE WITH REMOTE PROGRAMMING, BOLUS ESTIMATOR AND/OR VIBRATION CAPABILITIES (published as PCT application WO 00/10628), which is incorporated by reference herein, or the like. In particular embodiments, the supplemental device includes a post-calibration filter, a display, a recorder, and/or a blood glucose meter. In further alternative embodiments, the supplemental device includes a method for a user to add or modify information to be communicated to the infusion device 34 and/or the telemetered characteristic monitor transmitter 30 such as buttons, a keyboard, a touch screen, and the like.
In particular alternative embodiments, the supplemental device is a computer in combination with an analyte monitor and a RF programmer. The analyte monitor receives RF signals from the telemetered characteristic monitor transmitter 30, stores the signals and down loads them to a computer when needed. The RF programmer sends control signals to the infusion device 34 to reprogram the rate of insulin infusion. Both the analyte monitor and the RF programmer are placed into separate communication stations. The communication stations include IR transmitters and IR receivers to communicate with the analyte monitor and the RF programmer. The sensor signal values are transmitted via the telemetered characteristic monitor transmitter 30 to the analyte monitor located in one of the communication stations. Then the sensor signal values are communicated through the IR receiver in a first communication station and to the computer. The computer processes the sensor signal values through one or more filters, calibrators, and controllers to generate commands 22. The commands are sent to a second communication station and sent to an RF programmer by the IR transmitter in the communication station. Finally the RF programmer transmits the commands 22 to the infusion device 34. The communication station, analyte monitor and infusion device 34 may be of the type described in U.S. patent application Ser. No. 09/409,014, filed on Sep. 29, 1999 entitled COMMUNICATION STATION FOR INTERFACING WITH AN INFUSION PUMP, ANALYTE MONITOR, ANALYTE METER OR THE LIKE (published as a PCT application WO 00/18449), which is incorporated by reference herein. Alternatively, the RF programmer may be omitted and the infusion device may be placed in a communication station, or the infusion device may receive the commands without the use of an RF programmer and/or a communication station.
Overnight Closed-Loop System
A closed-loop insulin delivery system of the type described herein may utilize a variety of control algorithms to regulate the delivery of insulin to the body of the patient in a safe and predictable manner. Overnight operation of a closed-loop insulin infusion system should be carefully controlled in an automated way that need not depend on patient, user, or caregiver interaction. In this regard, a number of safeguards can be implemented with the system. These safeguards are intended to provide actionable sensor glucose readings, assess the accuracy of sensor readings, and constrain insulin delivery based upon possible sensor over-read conditions. These safeguards will alert the user and allow the patient to take appropriate actions. Therefore, these safeguards will mitigate the potential risks of overnight closed-loop control.
The control algorithm utilized by the system may be considered to be one type of safeguard in that it emulates the effect of insulin inhibiting insulin secretion. The system may also implement sensor performance safeguards. For example, a closed-loop initiation algorithm determines if the system can enter the closed-loop mode by calculating a recent calibration factor. The initiation algorithm checks the time between recent and prior calibration factors and determines the relative sensor error between the readings. As another example of a sensor safeguard, the system may employ a model supervisor during the closed-loop mode. The model supervisor checks that the sensor glucose readings are adequate for use during overnight closed-loop mode by comparing model-predicted sensor glucose values in real-time against actual sensor glucose values. If the model-predicted glucose values and the actual values differ significantly, the system triggers a fail-safe alert indicating a faulty sensor. This fail-safe alert can be generated in response to a number of sensor issues, such as sensor drift, sensor dislodgement, sensor compression artifact, etc.
The system may also implement a target glucose level safeguard. In this regard, a start-up algorithm can be deployed to provide a smooth transition between the open-loop mode and the closed-loop mode by gradually adjusting the target glucose level while in the closed-loop mode. The adjusted target glucose is used by the closed-loop control algorithm until the adjusted target glucose converges to a particular setpoint. At that time, the setpoint can be used for future dosing calculations during the closed-loop mode.
The system may also utilize at least one insulin limit as an insulin delivery and sensor performance safeguard. In this context, the insulin limit constrains the maximum amount of insulin delivered to the patient at any time in order to avoid over-delivery of insulin by the closed-loop control system due to potential sensor faults. In practice, the insulin limit is a value that is specific to each patient and is calculated based on the patient's delivered insulin during a fasting period, fasting blood glucose, and insulin sensitivity.
The system may also employ one or more insulin delivery safeguards. For example, an insulin delivery timeout continuously monitors (during closed-loop operation) if the patient is receiving insulin at the insulin limit for a prolonged period of time and, if so, triggers a fail-safe alert. This safeguard also monitors if the system is not delivering insulin for a prolonged period of time and, if so, triggers a fail-safe alert. A correction bolus is another insulin delivery safeguard. The system calculates an insulin bolus dosage for mitigating hyperglycemia at the commencement of closed-loop mode if the patient is above a designated blood glucose threshold. The determination can be achieved by acquiring a blood glucose meter reading at the initiation of closed-loop mode. The correction bolus is calculated based on the patient's insulin sensitivity, the amount of insulin on board, and a glucose target. Insulin on board (IOB) compensation is yet another insulin delivery safeguard. IOB compensation estimates the amount of insulin on board based on manual boluses administered, such that the system can effectively account for the IOB. In this regard, the manual boluses may be subtracted from the insulin dose that is calculated by the PID-IFB control algorithm.
The system may also implement one or more communication safeguards. For example, a “missed sensor transmission” feature continuously monitors data being received by the controller. For missed data packets totaling less than 15 minutes of operating time, the system remains in closed-loop mode. During this time, however, the system continues to calculate the insulin dose using the closed loop control algorithm based on the last valid sensor glucose value. For missed data packets totaling 15-60 minutes, the safeguard will switch to a pre-programmed safe basal rate, defined as half the patient's night time basal rate. If the controller starts receiving data packets during the safe basal rate timeframe, the system will again switch to the closed-loop mode. For missed data packets totaling more than 60 minutes, the system will switch to the open-loop mode where it will deliver a pre-programmed basal rate (which may be set by a caregiver).
The exemplary closed-loop control algorithms, methodologies, and techniques described in more detail below may be based around a PID control algorithm of the type presented in the preceding sections of this disclosure. In certain embodiments, the closed-loop control algorithms utilize a PID insulin feedback (PID-IFB) control algorithm. More specifically, the PID-IFB control algorithm cooperates with other algorithms, processes, and controls that represent additional safeguards that may apply during overnight use (and/or during other periods of use). These additional safeguards may include, without limitation: the use of an “Insulin Limits” parameter; a closed-loop initiation circuit that is based on glucose sensor calibration; an insulin on board (IOB) compensation algorithm; monitoring missed transmissions; and monitoring sensor glucose against predicted sensor glucose.
In practice, optimal or desired settings for the Insulin Limits parameter should be determined In this regard, the Insulin Limits parameter serves as an input to the controller logic for each patient, and it imposes an upper limit to the insulin delivery rate as an additional safety feature to avoid over-delivery of insulin by the controller due to potential sensor error. In certain embodiments, the Insulin Limits parameter is calculated from an amount of insulin delivered to the patient during a designated fasting period, a fasting blood glucose value of the patient, and the patient's insulin sensitivity.
Referring again to
In other embodiments, the functionality of the glucose sensor system 10 could be integrated into the insulin delivery system 14, perhaps as an interchangeable disposable module that attaches to the housing of the insulin delivery system 14. In yet other embodiments, the functionality of the controller 12 could be incorporated into the insulin delivery system 14 such that a separate and distinct controller device need not be carried by the patient. Indeed, the control software utilized by the controller 12 can be ported for installation in an insulin infusion pump, a pump monitor device, or the like to implement the functionality of the controller 12 in those devices if so desired. In further embodiments, a single hardware device platform could be suitably designed to accommodate the functionality of the insulin delivery system 14, the glucose sensor system 10, and the controller 12. These and other possible implementations are contemplated by this disclosure, and the particular manner in which the closed-loop system is configured and deployed is not intended to limit or otherwise restrict the scope or application of the closed-loop control techniques described herein.
Although not shown in
After entering the closed-loop mode, the insulin delivery system 14 sends sensor glucose (SG) values, sensor Isig values, calibration factors, “insulin delivered” values, and other data as needed to the controller 12 in accordance with a predetermined schedule, e.g., at five minute intervals. The controller 12 determines the desired insulin dose based on the closed-loop algorithm to maintain the patient at a target glucose setpoint, and communicates suitable control data and instructions to the insulin delivery system 14. The insulin delivery system 14 responds to deliver the insulin dose specified by the controller 12 to the user.
The host electronic device that implements the controller 900 may be realized as a monitor device for an insulin infusion device, where the monitor device and the insulin infusion device are two physically distinct hardware devices. In another embodiment of the system, the host electronic device that implements the controller 900 may be realized as a portable wireless device, where the portable wireless device and the insulin infusion device are two physically distinct hardware devices. The portable wireless device in this context may be, without limitation: a mobile telephone device; a tablet computer device; a laptop computer device; a portable video game device; a digital media player device; a portable medical device; or the like. In yet other system embodiments, the host electronic device and the insulin infusion device are physically and functionally integrated into a single hardware device. In such embodiments, the insulin infusion device will include the functionality of the controller 900 as presented here.
Certain embodiments of the controller 900 include a plurality of cooperating functional modules that are designed and configured to determine the insulin dose to be delivered to keep the patient at the target glucose setpoint during an overnight closed-loop operating mode. In this regard, the illustrated embodiment of the controller 900 may include the following functional modules, without limitation: a closed-loop initiation module 902; a start-up module 904; a proportional integral derivative insulin feedback (PID-IFB) control module 906; an insulin limit module 908; an insulin on board (IOB) compensation module 910; an insulin delivery timeout module 912; a model supervisor module 914; and a missed transmission module 916.
Referring to
In some embodiments, the closed-loop initiation module 902 may consider certain sensor performance criteria that prevents closed-loop initiation. Such criteria may include, without limitation: (1) during start-up when the calibration is not stable; (2) when the sensor sensitivity changes significantly; (3) when sensors may be calibrated with a potentially invalid meter reading thereby changing the sensor sensitivity significantly; (4) any other situation that could cause a mismatch between the sensor and meter for a number of most recent calibrations spaced over a designated period of time (e.g., the two most recent calibrations).
The illustrated embodiment of the closed-loop initiation module 902 receives at least the following items as inputs: a meter (measured) BG value 920; at least one sensor calibration factor 922 (i.e., calibration measurements, calibration data, etc.); the sensor Isig value 924; and timestamp data 926 that indicates the calibration time associated with the BG value 920 and the sensor calibration factor 922. Some or all of this input data may be provided directly or indirectly by the insulin delivery system 14 (see
The closed-loop initiation module 902 analyzes the input data (both current values and historical values) to determine whether or not the system is allowed to enter into the closed-loop mode. For example, the closed-loop initiation module 902 may: check the period between two consecutive calibration timestamp values; compare recent and prior calibration factor values; and the like. The “outputs” of the closed-loop initiation module 902 correspond to two operating modes of the system. More specifically, the closed-loop initiation module 902 controls whether the system remains operating in the open-loop mode 928 or whether the system starts the closed-loop mode 930.
Referring to
Referring to
The start-up module 904 receives sensor glucose (SG) values 940 as an input, and the functionality of the start-up module 904 may be initiated in response to the start of the closed-loop mode 930 (this trigger mechanism is represented by the dashed arrow 942 in
In certain embodiments, the start-up module 904 calculates a final target glucose value 946, which serves as an input to the PID-IFB control module 906. The final target glucose value 946 enables the system to make a smoother transition between open-loop and closed-loop modes (by gradually adjusting the final target glucose value 946). The start-up module 904 may utilize the target glucose setpoint value 944 to calculate the final target glucose value 946. In this regard, the start-up module 904 elevates the final target glucose value 946 to the same level as the sensor glucose value at the start of the closed-loop mode, provided the sensor glucose is above a certain threshold. As time progresses, the final target glucose value 946 gradually decreases back to the target glucose setpoint value 944 (usually in approximately two hours). Referring to
As an additional safeguard, the insulin limit module 908 cooperates with the PID-IFB control module 906 to provide an upper insulin limit that is calculated based on the patient's insulin intake during a designated fasting period, the patient's fasting blood glucose, and the patient's insulin sensitivity. This insulin limit imposes an upper limit to the insulin delivery rate to avoid over-delivery of insulin by the system due to potential sensor error.
The PID-IFB control module 906 may be configured to carry out the control processes described in more detail above with reference to
The insulin delivered 954 is a parameter or value that indicates the amount of insulin that has been delivered to the patient by the insulin delivery system. Thus, the insulin delivered 954 may indicate recent boluses (typically by Units) delivered over a period of time. In certain implementations, the insulin delivered 954 corresponds to the amount of insulin delivered in the last sampling time, which may be, without limitation: one minute; five minutes; thirty seconds; or any designated sampling time. The insulin delivered 954 may also indicate the amount of insulin delivered by the delivery system as basal or boluses in any defined period of time in the past (e.g., the last N hours) or the amount of insulin delivered by the system in the last sampling cycle. In practice, the PID-IFB control module 906 (and the IOB compensation module 910) may be “initialized” to collect and save historical values for the insulin delivered 954 as needed. Thereafter, the insulin delivered 954 can simply indicate an amount of insulin administered by the system during the last sampling time period if by a bolus or basal channels.
As mentioned above, the PID-IFB control module 906 may utilize the upper insulin limit 959, which is a patient-specific parameter. In certain embodiments, the upper insulin limit 959 may be entered by the user, a caregiver, or the like. Alternatively, the insulin limit module 908 may be responsible for calculating or otherwise managing the upper insulin limit 959 if so desired. The upper insulin limit 959 imposes an upper limit to the insulin delivery rate as an additional safety feature to avoid over-delivery of insulin by the controller 900 due to potential sensor error. Thus, if the PID-IFB control module 906 recommends a dose higher than the insulin limit 959, the insulin limit 959 will be utilized to constrain the insulin delivered to the insulin limit value. In addition, implementation of the insulin limit 959 will “freeze” the integral component of the PID to its previous value to prevent integral windup, which can cause continuous integrating of the glucose error until it reaches maximum values. In certain embodiments, the upper insulin limit 959 has a default value set at five times the patient's basal rate. Hence, if the maximum value is reached, the PID-IFB control algorithm will be fairly aggressive in calculating an insulin dose. Accordingly, to minimize integral windup, the insulin limit 959 is fed back to the PID-IFB control module 906 (as depicted in
The PID-IFB control module 906 operates as described previously to calculate a current insulin dose 958 as an output value (the current insulin dose 958 is also referred to herein as the uncompensated insulin infusion rate, PIDRate(n)). In practice, the current insulin dose 958 is typically expressed as an infusion rate (Units/Hour). In the context of this description, the current insulin dose 958 may represent a closed-loop infusion rate that has already been subjected to limiting by the insulin limit module 908, and which may be subjected to further adjustment or compensation by the IOB compensation module 910. Thus, the output of the insulin limit module 908 (the upper insulin limit 959) represents a potentially limited insulin dose to be provided by the PID-IFB control module 906—if no limit is imposed, then the insulin limit 959 has no effect on the output of the PID-IFB control module 906; otherwise, the current insulin dose 958 will be the same as the upper insulin limit 959. Referring again to
The IOB compensation module 910 receives at least the following items as inputs: the current insulin dose 958; and information regarding manual boluses delivered 960. The manual boluses delivered 960 may be provided directly or indirectly by the insulin delivery system 14, a translator device, a monitor device, and/or any device in the closed-loop system (see
Referring to
If query task 1018 determines that the closed-loop mode should continue (the “No” branch of query task 1018), then the control process 1000 may check whether it is time to perform another iteration of the control routine. In other words, the control process 1000 may check for the next sampling time (query task 1022). If it is time for the next iteration, then the control process 1000 may return to task 1010 and repeat the computations with the next set of data values. For example, the next iteration of the control routine may obtain and process the current values of some or all of the following parameters, without limitation: the SG value 940; the SG rate of change; the sensor Isig value 924; the amount of insulin delivered 954; and the manual boluses delivered 960. This allows the control process 1000 to adjust the final insulin infusion rate in an ongoing manner in accordance with a predetermined schedule, a designated sampling rate, or the like.
The insulin delivery timeout module 912 monitors if the patient is receiving continuous delivery of insulin at the maximum insulin limit or the minimum allowable infusion of zero Units/Hour for a time specified by the controller. Accordingly, the insulin delivery timeout module 912 may receive the insulin delivered 954 as an input. If the specified time is exceeded, the system will trigger a fail-safe alert 966. Otherwise, the system remains in the closed-loop operating mode 968.
Referring back to
The missed transmission module 916 is suitably configured to monitor the following, without limitation: the sensor Isig values 950; the SG values 940; and the sensor calibration factors 952. More particularly, the missed transmission module 916 continuously monitors to check whether the system is receiving data packets that convey the necessary information and input values. For missed data packets totaling less than a lower threshold of time (e.g., 15 minutes), the system remains in the closed-loop mode, as indicated by block 980 in
To summarize, the controller 900 determines whether to enter into the closed-loop mode in response to at least the recent meter BG values 920, the sensor calibration factors 922, and the calibration timestamp data 926. The controller 900 utilizes the closed-loop initiation module 902 to check if the sensor calibration time between the last two calibration values is within an acceptable range, and whether any change between the two calibration values (recent and prior value) is acceptable. If so, the controller 900 will switch the system into the closed-loop mode. Once the system is in the closed-loop mode, the controller 900 will periodically receive data packets (e.g., every five minutes) that include the current SG value 940, the current sensor Isig values 950, the insulin delivered 954, the sensor calibration factors 952, and manual boluses delivered 960. In certain embodiments, each of the data packets received by the controller 900 includes data collected during the previous 24-hour period.
The start-up module 904 utilizes the SG values 940 and the target glucose setpoint value 944 to calculate the final target glucose value 946. In some embodiments, the target glucose setpoint value 944 is set to 120 mg/dL, although other settings could be used if so desired (a typical range of settings may be, for example 70-300 mg/dL). This results in a smoother transition between open-loop and closed-loop modes by gradually adjusting the final target glucose value 946. The final target glucose value 946 is sent to the PID-IFB control module 906 for use as one input that influences the calculation of the final insulin dose 962.
The PID-IFB control module 906 utilizes the final target glucose value 946, the current and past SG values 940, the SG rate of change values, and the insulin delivered 954 to determine the insulin infusion rate (the current insulin dose 958) in order to achieve euglycemia. As an additional safeguard, the upper insulin limit 959 (calculated based on the patient's insulin intake during a fasting period, fasting blood glucose, and insulin sensitivity) from the insulin limit module 908 is input into the controller 900 for each patient to impose an upper limit to the insulin delivery rate to avoid over-delivery of insulin by the controller 900. The PID-IFB control module 906 considers the upper insulin limit 959 before sending the current insulin dose 958 to the IOB compensation module 910, which estimates insulin on board from manual boluses, before or during closed-loop operation, in order to calculate the final insulin dose 962. The final insulin dose 962 may be communicated from the controller 900 directly or indirectly to the insulin delivery system 14 such that the final insulin dose 962 can be delivered to the patient during closed-loop operation.
Additional safeguards could be implemented to monitor the system during closed-loop operation, such that the system exits the closed-loop mode when certain criteria are not met. For example, the controller 900 may cause the system to exit the closed-loop mode if more than a designated number of consecutive data packets are missed. This assumes that the controller 900 usually receives data packets (from the insulin delivery system 14, from a monitor, from a translation device, or the like) in a continuous manner during closed-loop operation. Thus, if the controller 900 detects that more than a threshold number of consecutive data packets are not received as expected, the system will be commanded to exit the closed-loop mode. This functionality is associated with the missed transmission module 916, as described previously.
Moreover, the model supervisor module 914 estimates the user's glucose concentration in an ongoing manner, based on the insulin delivered 954, the sensor Isig values 950, and the sensor calibration factors 952. If the difference between the model-predicted glucose and the sensor glucose value is greater than a stated threshold, the controller 900 may cause the system to exit the closed-loop mode.
As summarized above, the controller 900 employs a number of modules or functions that cooperate to regulate the delivery of insulin during closed-loop operation: the closed-loop initiation module 902; the start-up module 904; the PID-IFB control module 906; the insulin limit module 908; and the IOB compensation module 910. Moreover, the controller 900 may employ a number of modules that perform various safeguarding functions during closed-loop operation. These safeguarding modules may include: the insulin delivery timeout module 912; the model supervisor module 914; and the missed transmission module 916.
Closed-Loop Initiation Module
The closed-loop initiation module 902 checks for changes in sensor sensitivity and determines whether or not the system is allowed to enter the closed-loop mode. To this end, the closed-loop initiation module 902 receives, obtains, or accesses sensor calibration data that is associated with a continuous glucose sensor that generates a sensor variable (e.g., an electrical current such as Isig), which in turn is indicative of the current blood glucose level of the user/patient. Referring again to
In the context of the closed-loop initiation module and its related functionality, a sensor calibration factor represents a translation or conversion value that is applicable to convert a given value of a sensor variable (e.g., an Isig value) to a corresponding blood glucose value. For the exemplary embodiment presented here, Isig is an electrical current that is expressed in nanoamperes (nA), blood glucose is expressed in mg/dL, and, therefore, sensor calibration factors are expressed in
The closed-loop initiation module 902 analyzes the sensor calibration factors and checks for a series of conditions pertaining to the calibration and/or sensitivity of the continuous glucose sensor used by the system, based on the sensor calibration data that is processed at that time. If all the conditions are met, the controller 900 initiates the closed-loop operating mode. If this criteria is not met, the system remains in the open-loop operating mode.
As explained in more detail below, the closed-loop initiation process 5000 utilizes at least two sensor calibration factors to determine whether or not to allow the insulin infusion device to transition into the closed-loop operating mode. One of the two calibration factors is the most recent calibration factor, CFR. To this end, the process 5000 identifies or calculates the most recent calibration factor from the received sensor calibration data (task 5004). In accordance with one methodology, task 5004 identifies the most recent calibration factor that was received at task 5002, based on the received calibration timestamp data. In accordance with an alternative methodology, task 5004 calculates the most recent calibration factor, using the current Isig value and the current meter BG value (which may also be received with the sensor calibration data, as mentioned above). When the most recent calibration factor is calculated in this manner, it will represent the current or instantaneous calibration factor rather than a historical or previously-computed calibration factor.
Although the most recent calibration factor can be generated using any suitable methodology that leads to an accurate representation, the exemplary embodiment presented here calculates the most recent calibration factor (CFR) in accordance with the expression CFR=meterBG/(Isig−2). In this expression, meterBG is the most recent meter blood glucose value and Isig is the sensor current obtained at the time of the meterBG value received from the continuous glucose sensor. For this particular implementation, the most recent calibration factor is calculated based on only one sampling point, i.e., only one meter BG reading and only one Isig reading. This simplified derivation is preferred because it enables the process 5000 to quickly and effectively obtain the recent calibration factor with little to no delay.
Referring again to
In accordance with the exemplary timing requirements described in the preceding paragraph, by definition there will be no intervening calibration factors obtained between the most recent calibration factor (CFR) and the prior calibration factor (CFP). In contrast, and as explained in more detail below with reference to
The timeline diagram 5024 also indicates a window of time 5025 that is defined between a first (minimum) time 5026 and a second (maximum) time 5028. For this particular example, the first time 5026 and the second time are defined relative to tR. More specifically, the first time 5026 is defined to be tR−tDiffmin, where tDiffmin is the minimum allowable time difference relative to tR. Although other time values could be utilized, tDiffmin equals two hours for this example. Accordingly, the first time 5026 is labeled tR−2 in
The timeline diagram 5024 also indicates a window of time 5029 that is defined between the time t and a recent time 5030. For this particular example, the recent time 5030 is defined relative to the time t. More particularly, the recent time 5030 is defined to be t−tRecent, where tRecent is a predetermined time period that is utilized by the closed-loop initiation module 902 to determine the “freshness” of the most recent calibration factor. Although other time values could be utilized, tRecent equals two hours for this example. Accordingly, the recent time 5030 is labeled t−2 in
The timeline diagram 5024 also indicates a window of time 5031 that is defined between the first time 5026 (i.e., the time defined by tR−tDiffmin) and tR (which is obtained from the timestamp data for the most recent calibration factor). The closed-loop initiation process 5000 is performed when tP (which is obtained from the timestamp data for the prior calibration factor) is not within the window of time 5031.
Referring back to
In accordance with certain embodiments, one of the conditions relates to the recency of the most recent calibration factor (query task 5008). In this regard, the process 5000 permits entry into the closed-loop operating mode only when the recent calibration factor is relatively new or “fresh”. More specifically, closed-loop operation is allowed only when t−tRecent≦tR≦t. For the example presented here, tRecent is two hours and, therefore, query task 5008 checks whether the most recent calibration factor has a timestamp that is less than two hours old. If the timing of the most recent calibration factor does not satisfy this condition (the “No” branch of query task 5008), then the process 5000 remains in the open-loop operating mode (task 5020). This example assumes that the most recent calibration factor is less than two hours old (see
As mentioned previously, the recent calibration factor may be calculated on the fly. For such embodiments, the recency check of query task 5008 need not be performed because the system can safely assume that the calculated recent calibration factor is less than two hours old. Alternatively, query task 5008 could still be performed even though the recency condition will always be satisfied.
In accordance with certain embodiments, one of the conditions relates to the timing of the prior calibration factor, CFP. In this regard, the process 5000 permits entry into the closed-loop operating mode only when the prior calibration factor falls within a designated window of time (query task 5010). More specifically, closed-loop operation is allowed only when tR−tDiffmax≦tP≦tR−tDiffmin. For the example presented here, tDiffmax is eight hours and tDiffmin is two hours. Accordingly, query task 5010 checks whether the prior calibration factor has a timestamp that falls between tR−8 and tR−2. If the timing of the prior calibration factor does not satisfy this condition (the “No” branch of query task 5010), then the process 5000 remains in the open-loop operating mode (task 5020). This example assumes that the prior calibration factor falls within the window of time of interest (see
In accordance with certain embodiments, one of the conditions relates to the value of the most recent calibration factor, CFR. In this regard, the process 5000 permits entry into the closed-loop operating mode only when the most recent calibration factor falls within a specified range of values (query task 5012). More specifically, closed-loop operation is allowed only when CFmin≦CFR≦CFmax, where CFmin is a minimum acceptable calibration factor value and CFmax is a maximum acceptable calibration factor value. For the example presented here, CFmin is
and CFmax is
Accordingly, query task 5012 checks whether the most recent calibration factor is within this range of values. If not, (the “No” branch of query task 5012), then the process 5000 remains in the open-loop operating mode (task 5020). This example assumes that the most recent calibration factor falls within the designated range.
In accordance with certain embodiments, one of the conditions relates to the value of the prior calibration factor, CFP. In this regard, the process 5000 permits entry into the closed-loop operating mode only when the prior calibration factor falls within a specified range of values (query task 5014). More specifically, closed-loop operation is allowed only when CFmin≦CFP≦CFmax. Accordingly, query task 5014 checks whether the prior calibration factor is also within this range of values. If not, (the “No” branch of query task 5014), then the process 5000 remains in the open-loop operating mode (task 5020). This example assumes that the prior calibration factor falls within the designated range.
In accordance with certain embodiments, one of the conditions relates to the change between the prior and most recent calibration factors. In this regard, the process 5000 may regulate entry into the closed-loop operating mode by permitting entry only when the change between the most recent calibration factor and the prior calibration factor is less than or equal to a threshold value. The example described here calculates a percentage change value, CFchange, using the values of the prior and most recent calibration factors (task 5016). In some embodiments, CFchange is calculated in accordance with the expression:
If CFchange is less than or equal to a designated threshold value (the “Yes” branch of query task 5018), then the closed-loop initiation process 5000 determines that the continuous glucose sensor is accurate for purposes of the closed-loop operating mode and the process 5000 continues by initiating the closed-loop operating mode (task 5022). If, however, CFchange is greater than the designated threshold value (e.g., 35% for this example), then the “No” branch of query task 5018 is followed and the insulin infusion device remains in the open-loop mode (task 5020).
Referring again to
Recall that the closed-loop initiation process 5000 and the timeline diagram 5024 relate to a situation where the prior calibration factor, CFP, falls outside of the window of time 5031, which is defined between the first time 5026 and tR. Indeed, CFP in
The process 5050 is similar to the closed-loop initiation process 5000 in many ways. Indeed the process 5050 may perform tasks 5002, 5004, and 5006 in the manner described above. Moreover, certain aspects of the process 5050 are similar or identical to counterpart aspects described previously for the process 5000. For the sake of brevity and simplicity, such common aspects will not be redundantly described here in the context of the process 5050.
The closed-loop initiation process 5050 regulates entry into the closed-loop operating mode of the insulin infusion device, based on the most recent calibration factor, the prior calibration factor, and the associated calibration timestamp data. In this regard, the process 5050 permits entry into the closed-loop mode only when certain conditions are met. If any of the conditions are not satisfied, then the process 5050 prevents entry into the closed-loop mode.
In accordance with certain embodiments, one of the conditions relates to the recency of the most recent calibration factor (query task 5052). Query task 5052 is equivalent to query task 5008 of the process 5000, which was described previously. For the example presented here, query task 5052 checks whether the most recent calibration factor has a timestamp that is less than two hours old. If the timing of the most recent calibration factor does not satisfy this condition (the “No” branch of query task 5052), then the process 5050 remains in the open-loop operating mode (task 5066). This example assumes that the most recent calibration factor is less than two hours old (see
In accordance with certain embodiments, one of the conditions relates to the value of the most recent calibration factor, CFR. In this regard, the process 5050 permits entry into the closed-loop operating mode only when the most recent calibration factor falls within a specified range of values (query task 5054). Query task 5054 is equivalent to query task 5012 of the process 5000, which was described previously. Thus, query task 5054 checks whether the most recent calibration factor is within the specified range of values. If not (the “No” branch of query task 5054), then the process 5050 remains in the open-loop operating mode (task 5066). This example assumes that the most recent calibration factor falls within the designated range.
As an additional safeguard, the process 5050 also checks for the presence of at least one eligible calibration factor that was determined or obtained prior to the most recent calibration factor, and prior to the second most recent calibration factor (query task 5056). In this context, a prior calibration factor is “eligible” if it was determined or obtained during the specified window of time 5025 (see
It should be appreciated that the closed-loop initiation module may be suitably configured to select any eligible calibration factor for use as CFP2. For example, the process 5050 may select the newest eligible calibration factor (i.e., the most recent calibration factor that falls within the designated window of time), the oldest eligible calibration factor, the median eligible calibration factor, a randomly selected eligible calibration factor, the first-discovered eligible calibration factor, or the like. In certain embodiments, and for purposes of this example, it is assumed that the process 5050 has designated the most recent eligible older calibration factor as CFP2. Accordingly, there will be no intervening calibration factors obtained between tP2 and tR−2 (as depicted in
The process 5050 may perform task 5058 to identify or designate some or all of the intervening calibration factors that were obtained within the window of time between tR−tDiffmin (e.g., tR−2 as shown in
It should be appreciated that the process 5050 could be suitably designed to designated and consider other calibration factors if so desired. For example, if the process 5050 is able to select any eligible calibration factor as CFP2 (rather than always selecting the most recent eligible calibration factor), then the process 5050 may also identify and handle additional intervening calibration factors that were obtained after CFP2 was obtained. In other words, one or more intervening calibration factors could fall within the window of time 5025. As another example, if the process 5050 is able to select any calibration factor that falls within the window of time 5031 as CFP (rather than always selecting the second most recent calibration factor, relative to CFR), then the process 5050 may also identify and handle additional intervening calibration factors that were obtained during the window of time 5031. In such a scenario, one or more intervening calibration factors could be obtained during the window of time defined between tP and tR. Moreover, in certain embodiments, it may be desirable to disregard any intervening calibration factors (other than CFP) that fall within the window of time 5031, based on the assumption that accurate results can be obtained using only the second most recent calibration factor, CFP.
In accordance with certain embodiments, one of the conditions for entering the closed-loop operating mode relates to the values of the intervening calibration factors, CFINT. In this regard, the process 5050 permits entry into the closed-loop operating mode only when each of the intervening calibration factors falls within a specified range of values (query task 5060). More specifically, closed-loop operation is allowed only when CFmin≦CFINT≦CFmax for all intervening calibration factors under consideration. Accordingly, query task 5060 checks whether all of the intervening calibration factors are within this range of values. If not, (the “No” branch of query task 5060), then the process 5050 remains in the open-loop operating mode (task 5066). This example assumes that all of the identified intervening calibration factors fall within the designated range.
Thereafter, the process 5050 continues in the manner described above for tasks 5016, 5018, and 5022 of the process 5000. Thus, the process 5050 calculates the percentage change value, CFchange, using the values of the most recent calibration factor and the eligible calibration factor that was selected for use as CFP2 (task 5062), and compares CFchange to the designated threshold value (query task 5064). If CFchange is less than or equal to the threshold value (the “Yes” branch of query task 5064), then the closed-loop initiation process 5050 determines that the continuous glucose sensor is accurate for purposes of the closed-loop operating mode and the process 5050 continues by initiating the closed-loop operating mode (task 5068). If, however, CFchange is greater than the designated threshold value (e.g., 35% for this example), then the “No” branch of query task 5064 is followed and the insulin infusion device remains in the open-loop mode (task 5066).
It should be understood that the various conditions checked by the process 5000 and the process 5050 need not be performed in the illustrated order, or in any particular order. Indeed, some or all of the conditions could be checked in parallel if so desired. Of course, if any one of the conditions is not satisfied, then the process 5000, 5050 may immediately exit rather than perform any remaining checks.
As mentioned previously, the functionality of the closed-loop initiation module can be implemented using a suitably written software program, application, or executable logic. In this regard, the following sections contain descriptions of the desired functionality in a format that is helpful for purposes of software development, application programming, and the like.
Closed-Loop Initiation Module: First Representation
Certain embodiments of the closed-loop initiation module 902 (see
In some embodiments, the closed-loop initiation module 902 is implemented in the form of a series of processing steps. Using the logic described below, the closed-loop initiation module 902 decides whether to let the system enter the closed-loop mode.
Case A
Case A refers to the scenario depicted in
If (tP is not in the time window (tR−tDiffmin:tR)), the following logic is checked.
If any of the above conditions is not met, the system remains in the open-loop mode. Thus, in order to enter the closed-loop mode, new calibration(s) that satisfy all of the conditions in Case A (or Case B as described below) will be required.
Case B
Case B refers to the scenario depicted in
If any of the above conditions is not met, the system remains in the open-loop mode. Thus, in order to enter the closed-loop mode, new calibration(s) that satisfy all of the conditions in Case A or Case B will be required.
In accordance with certain variations of the closed-loop initiation module 902, the system requests a meter BG and related calibration when entering the closed-loop mode. In such alternative embodiments, therefore, the closed-loop initiation module 902 uses the meter BG and Isig to calculate CFR. Thus, in such an implementation, the sensor current would also be an input to the closed-loop initiation module 902. Accordingly, because CFR is now being calculated by the closed-loop initiation module 902 itself, the conditions in Case A and Case B (i.e., checking whether t-tRecent≦tR≦t) will always be met.
In particular implementations, some of the parameters mentioned above can be fixed. In this regard, the following values may be utilized in an exemplary embodiment. It should be appreciated that these values are provided here for illustrative purposes only, and that an implementation of the closed-loop initiation module 902 may utilize different values if so desired.
tRecent=120 minutes
tDiffmin=120 minutes
tDiffmax=480 minutes
CFmin=2.5 mg/dL per nA
CFmax=6 mg/dL per nA
Closed-Loop Initiation Module: Second Representation
In accordance with some embodiments, the functionality of the closed-loop initiation module 902 can be represented as follows. The closed-loop initiation module 902 may be implemented in the form of a series of case steps. In this regard, the closed-loop initiation module 902 first calculates the recent calibration factor value (CFR) using the most recent meter BG and Isig values as shown in the following Equation A1:
CFR=meterBG/(Isig−2) (eq A1)
Here, CFR is the recent calibration factor value, meterBG is the meter BG value, and Isig is the sensor Isig value. The “−2” in Equation A1 represents a constant offset that is used by the calibration algorithm when calculating calibration factors and sensor glucose.
Using the logic described below for Case C or Case D, the closed-loop initiation module 902 decides whether to let the system enter the closed-loop mode. Each case condition is dependent upon the time at which the most recent prior calibration factor (CFP) was obtained.
Case C
Case C corresponds to a scenario where the time of prior calibration is greater than 120 minutes before the most recent calibration. In addition, the recent calibration factor (CFR) and the prior calibration factor (CFP) are within limits as shown in the following logical expressions.
CFmin≦CFR≦CFmax (eq A2)
CFmin≦CFP≦CFmax (eq A3)
Here, CFR is the recent calibration factor value, CFP is the prior calibration value, CFmin is the minimum value for the calibration factor that is set as 2.5 mg/dL per nA, and CFmax is the maximum value for the calibration factor that is set as 6 mg/dL per nA.
For Case C, the time of prior calibration (tP) occurs between two and eight hours before time of recent calibration factor as shown in the following logical expression.
tR−tDiffmax≦tP≦tR−tDiffmin (eq A4)
Here, tP is the time of prior calibration, tR is the time when CFR was obtained, tDiffmax is the maximum time difference between the recent calibration and the prior calibration (which is set as 480 minutes (8 hours)), and tDiffmin is the minimum time difference between the recent calibration and the calibration prior to the recent calibration (which is set as 120 minutes (2 hours)).
For Case C, the calibration change (CFchange) is less than 35% as shown in the following logical expression, where CFchange is calculated according to Equation A6.
CFchange=(abs(CFR−CFP)/CFP)×100 (eq A5)
CFchange≦CFchangeTh (eq A6)
Here, CFchange is the percentage change in calibration factor from the previous calibration factor to a current calibration factor for any pair of calibration factors, CFchangeTh is the threshold for acceptable CFchange, (which is set to 35% for this example), CFR is the most recent calibration factor value, and CFP is the last calibration factor value before CFR.
If all of the foregoing conditions are met for Case C (Equations A2-A6), the closed-loop initiation module 902 can initiate the methodology for calculating the correction bolus (if needed). If, however, any condition is not met, the controller 900 remains in the open-loop mode. Thus, in order to enter the closed-loop mode, new calibration(s) that satisfy all of the conditions in Case C or Case D will be required.
Case D
Case D corresponds to a scenario where the time of prior calibration is less than 120 minutes before the most recent calibration. If the prior calibration is less than two hours before the recent calibration, an additional prior calibration factor (CFP2) is included in the analysis. This allows the closed-loop initiation module 902 to evaluate sensor sensitivity that has at least a two hour time span.
For Case D, the closed-loop initiation module 902 finds an earlier second prior calibration factor (CFP2) that is the most recent within a time window of two to eight hours before the time of the recent calibration factor (CFR) as shown in the following logical expression.
tR−tDiffmax≦tP2≦tR−tDiffmin (eq A7)
Here, tP2 is the time when the second prior calibration factor (CFP2) is obtained, tR is the time when CFR was obtained, tDiffmax is the maximum time difference between tP2 and tR (which is set as 480 minutes (8 hours) for this example), and tDiffmin is the minimum time difference between tP2 and tR (which is set as 120 minutes (2 hours) for this example).
For Case D, the closed-loop initiation module 902 also determines if more than one calibration factor (CF1. . . CFN) is available between the time of the second prior calibration factor (CFP2) and the time of the recent calibration factor (CFR), as shown in the following logical expression.
tP2≦t1 . . . tn≦tR (eq A8)
Here, t1. . . tn is the time when more calibration factors (CF1. . . CFN) are observed, tR is the time when CFR was obtained, and tP2 is the time when CFP2 is obtained.
For Case D, all calibration factors including recent calibration factor (CFR), prior calibration factor (CFP), second prior calibration factor (CFP2) and CF1. . . CFN are within limits as shown in the following logical expressions.
CFmin≦CFR≦CFmax (eq A9)
CFmin≦CFP≦CFmax (eq A10)
CFmin≦CFP2≦CFmax (eq A11)
CFmin≦CF1 . . . CFN≦CFmax (eq A12)
Here, CFR is the recent calibration factor, CFP is the prior calibration factor, CFP2 is the second prior calibration factor, CF1. . . CFN are calibration factors obtained between tP2 and tR, CFmin is the minimum value for the calibration factor (which is set as 2.5 mg/dL per nA for this example), and CFmax is the maximum value for the calibration factor (which is set as 6 mg/dL per nA for this example).
For Case D, the calibration change (CFchange) between CFR and CFP2 is less than 35%, as shown in the following logical expression where CFchange is calculated according to Equation A15.
CFchange=(abs(CFR−CFP2)/CFP2)×100 (eq A13)
CFchange≦CFchangeTh (eq A14)
Here, CFchange is the percentage change in calibration factor from the previous calibration factor to a current calibration factor for any pair of calibration factors, CFchangeTh is the threshold for acceptable CFchange (which is set to 35% for this example), CFR is the most recent calibration factor value, and CFP2 is the most recent calibration factor value in the time range described in Equation A7.
If all of the foregoing conditions are met for Case D (Equations A7-A14), the closed-loop initiation module 902 can initiate the methodology for calculating the correction bolus (if needed). If, however, any condition is not met, the controller 900 remains in the open-loop mode. Thus, in order to enter the closed-loop mode, new calibration(s) that satisfy all of the conditions in Case C or Case D will be required.
Correction Bolus with IOB Compensation
As described above, a correction bolus 932 may be commanded at the beginning of the closed-loop mode. The purpose of the correction bolus is to provide an insulin dose for mitigating hyperglycemia at the initiation of the closed-loop mode. This can be achieved by first acquiring a blood glucose meter reading value immediately prior to closed-loop initiation. If that BG meter reading value is above a certain correction threshold (CTH, which is 180 mg/dL for this example), the controller 900 will deliver an insulin dose based on the patient's insulin sensitivity (ISF, mg/dL/Units), insulin on board, and the desired target glucose level (TG, mg/dL) in order to bring the subject's glucose level to the target glucose level.
In accordance with certain implementations, the correction bolus (CB) is delivered based on the meter BG value (in mg/dL) acquired at the start of the closed-loop mode, as shown below:
Here, CB is the correction bolus, BG is the blood glucose meter value (mg/dL), TG is the target glucose (mg/dL), ISF (see Equation A18) is the patient's adjusted insulin sensitivity factor (mg/dL/Units), CTH is the correction threshold for blood glucose beyond which a correction bolus will be delivered (mg/dL), and IOB(n) is the active insulin on board from manual boluses (Units) where n is the current sampling point as described previously.
ISF=ISFfactor×ISF0 (eq A18)
Here, ISF is the patient's adjusted insulin sensitivity factor (mg/dL/Units), ISF0 is the patient's established insulin sensitivity factor (mg/dL/Units) and ISFfactor is an ISF adjusting factor (unit less). The default value for ISFfactor is set to one, which makes ISF=ISF0. However, for this particular example, ISFfactor has been assigned as an adjustable parameter with a range of 0.5 to 2 in order to provide greater flexibility for optimizing the patient's insulin sensitivity factor.
Here, CB is the correction bolus expressed in Units. It should be appreciated that Equation A19 is utilized because the controller 900 can only deliver positive boluses.
Start-Up Module
Referring again to
The process 5100 may begin by starting or initiating the closed-loop operating mode of the insulin infusion device (task 5102). In this regard, task 5102 may be triggered by the closed-loop initiation module 902 in the manner described above. The process 5100 responds to the initiation of the closed-loop operating mode by receiving or accessing the most recent sensor glucose value for the user, e.g., the sensor glucose value obtained for the current sampling point (task 5104). Referring to
The process 5100 continues by calculating a difference between the most recent sensor glucose value (obtained at task 5104) and the target glucose setpoint value (task 5106). This difference, DeltaGlu, may be calculated in accordance with the following expression: DeltaGlu=SG−Setpoint, where SG is the most recent sensor glucose value and Setpoint is the target glucose setpoint value (it should be appreciated that DeltaGlu could be derived using a more complex relationship or formula if so desired). At the beginning of the closed-loop operating mode, the start-up module 904 calculates the difference between the current SG value 940 and the target glucose setpoint value 944, as indicated by Equation 51 below.
In Equation 51, SG is the most recent sensor glucose value, n is the current sampling point, Setpoint is the target glucose setpoint value defined by user, and m is the sampling time during closed-loop operation (m=1 indicates the start of the closed-loop mode, and m increases with each sample received during the closed-loop mode). DeltaGlu(n) is forced to zero when m>1 as well as in the following situation described in Equation 52. In other words, DeltaGlu(n) is equal to the calculated difference only for the initial calculation of the final target glucose value 946. In contrast, DeltaGlu(n) is equal to zero for all subsequent calculations of the final target glucose value 946.
Referring again to
In accordance with the embodiments described herein, task 5110 adjusts the closed-loop insulin infusion rate in accordance with a suitable PID-IFB control algorithm. This closed-loop control is performed in an ongoing manner until it is time to exit the closed-loop operating mode (the “Yes” branch of query task 5112). Thus, task 5110 is repeated for each sampling point to update the final insulin dose 962 as needed while the closed-loop mode remains active.
Referring to the “Yes” branch of query task 5108, if the initial value of DeltaGlu is greater than MinDeltaGlu, then the process 5100 calculates and adjusts the closed-loop insulin infusion rate over time, based on a dynamic final target glucose value that gradually decreases over time toward the target glucose setpoint value (task 5114). This action ensures that the patient does not experience a “bump” associated with a significant change in insulin infusion rate at the outset of the closed-loop operating mode. Rather, the process 5100 transitions the insulin infusion rate in a manner that contemplates the difference between the patient's actual sensor glucose reading and the desired target glucose setpoint.
In accordance with the embodiments described herein, task 5114 adjusts the closed-loop insulin infusion rate in accordance with a suitable PID-IFB control algorithm. This closed-loop control is performed in an ongoing manner until it is time to exit the closed-loop operating mode (the “Yes” branch of query task 5116). Thus, task 5114 is repeated for each sampling point to update the final insulin dose 962 as needed while the closed-loop mode remains active. Notably, the dynamic final target glucose value will eventually settle to a value that is equal to (or substantially equal to) the target glucose setpoint value, as described in more detail below. This settling time will typically occur about two hours after the start of the closed-loop mode.
The process 5150 is iterative in that a final target glucose value is generated for each sampling point. For purposes of this description, the variable n represents the current sampling point. Accordingly, for the first iteration of the process 5150, n is assigned an initial value, such as one (task 5152). As mentioned above with reference to Equation 51, the process 5154 calculates the initial value of DeltaGlu as follows: DeltaGlu(n)=SG(n)−Setpoint (task 5154). In other words, DeltaGlu(n) is equal to the calculated difference between the initial sensor glucose value and the target glucose setpoint value only for the initial calculation of the dynamic glucose setpoint value. Otherwise, DeltaGlu(n) is equal to zero for all subsequent calculations of the dynamic glucose setpoint value (following the initial calculation of the dynamic glucose setpoint value).
In practice, therefore, the start-up module 904 forces DeltaGlu to zero if it is less than a certain threshold set in the controller 900 (referred to as MinDeltaGlu). As will become apparent from the following description, forcing DeltaGlu to zero in this manner results in the calculation of the final insulin dose 962 based on the fixed final target glucose value (as explained above with reference to task 5110 of the process 5100). Otherwise, DeltaGlu(n) retains its value if it is more than the threshold set in the controller 900, as described in Equation 52:
Here DeltaGlu is the difference between the current SG value 940 and the defined target glucose setpoint value 944, calculated from Equation 51 above, and MinDeltaGlu is the minimum difference (set in the controller 900) allowable between the current SG value 940 and the target glucose setpoint value 944. For purposes of the process 5150, it is assumed that the initial value of DeltaGlu(n) exceeds MinDeltaGlu.
The process 5150 may continue by calculating the dynamic glucose setpoint value (DynSP) for the current sampling point (task 5156). For this particular implementation, DynSP(n) is calculated in accordance with a discretized second order transfer function model. More specifically, DynSP(n) is calculated in accordance with the expression set forth in Equation 53:
DynSP(n)=cd1·DynSP(n−1)+cd2·DynSP(n−2)+cn0·DeltaGlu(n)+cn1·DeltaGlu(n−1) (eq 53)
Here, DynSP is the dynamic setpoint value, n is the current sampling point, n−1 is the last sampling point, and n−2 is the second to last sampling point. For this example, DynSP(n−1) is equal to zero for the initial calculation of the dynamic glucose setpoint value (i.e., for the first iteration of the process 5150). Similarly, DynSP(n−2) is equal to zero for the initial calculation of the dynamic glucose setpoint value, and DeltaGlu(n−1) is equal to zero for the initial calculation of the dynamic glucose setpoint value.
The parameters cd1, cd2, cn0, and cn1 are the coefficients of the discretized second order transfer function model. Each of the parameters cd1, cd2, and cn1 is calculated based on the two time constants (τsp1 and τsp2) of the discretized second order transfer function model, as indicated below:
In the above equations Ts indicates the sampling interval in minutes, and τsp1 and τsp2 are the time constants of the setpoint model. Moreover, axx1 is the reciprocal of the time constant τsp1, axx2 is the reciprocal of the time constant τsp2, eaxx1 is the exponential decay factor for τsp1, eaxx2 is the exponential decay factor for τsp2, and daxx21 is the difference between the reciprocal of τsp1 and τsp2.
It should be appreciated that the coefficient values could be adjusted if so desired to define the trajectory characteristics of DynSP over time. In this regard, the coefficient values could be changed as needed to alter the decay timeline, to alter the shape of the decay trajectory curve, or the like.
Referring again to
FinalTarget(n)=Setpoint+DynSP(n) (eq 54)
Notably, FinalTarget(n) corresponds to the final target glucose value 946 shown in
At the end of the closed-loop operating mode (the “Yes” branch of query task 5160), the process 5150 exits. If the closed-loop operating mode remains active (the “No” branch of query task 5160), then the variable n is incremented by an appropriate amount, such as one (task 5162). As mentioned above, DeltaGlu(n) is forced to zero for all iterations of the process 5150 other than the first iteration. Accordingly, the process 5150 sets DeltaGlu(n) to zero (task 5164) before returning to task 5156 to calculate the next value of the dynamic glucose setpoint value. Referring again to Equation 53, the current value of DynSP, i.e., DynSP(n), is calculated from two previous values of DynSP. Moreover, Equation 53 results in a second order decay of DynSP over time, such that DynSP effectively approaches zero and, consequently, the dynamic final target glucose value eventually approximates the target glucose setpoint value (see Equation 54).
Referring again to Equation 53, DynSP will be zero for the initial iteration if DeltaGlu(n) is forced to zero (which occurs when DeltaGlu(n) is less than or equal to the minimum threshold value). Moreover, DynSP will be zero for all subsequent iterations because DeltaGlu(n) is forced to zero for subsequent iterations, and because the historical values of DynSP (namely, DynSP(n−1) and DynSP(n−2)) will also be zero. Accordingly, FinalTarget(n) will be equal to Setpoint for all sampling points when the “No” branch of query task 5108 is followed (see
In particular implementations, some of the parameters mentioned above for the start-up module 904 can be fixed. In this regard, the following values may be utilized in an exemplary embodiment. It should be appreciated that these values are provided here for illustrative purposes only, and that an implementation of the start-up module 904 may utilize different values if so desired.
PID-IFB Control Module
The PID-IFB control module 906 calculates the current insulin dose 958 based on the current and past sensor glucose values 940, the sensor Isig values 950, the rate of change of the sensor glucose values, the sensor calibration factor 952, the final target glucose value 946, the target glucose setpoint value 944, insulin limits such as the upper insulin limit 959, and the insulin delivered 954 in order to achieve euglycemia. In certain embodiments, the PID-IFB control module 906 receives its inputs every five minutes, therefore the calculation of the insulin feedback components considers how often the input data is being received by the controller 900.
The PID-IFB control module 906 calculates the current insulin dose 958 in accordance with Equation 55:
u(n)=P(n)+I(n)+D(n)−γ1ISC−γ2IP−γ3IEFF (eq 55)
Note that PIDRate(n)≡u(n). In Equation 55, the insulin infusion rate u(n) represents the current insulin dose 958 shown in
The exemplary implementation of the control algorithm employed by the PID-IFB control module 906 is implemented in discrete (sampled) time. In the notation used, n is the current time step, where t=nTs, and t is the continuous time (in minutes) when using a sampling period of Ts.
The proportional component P(n) is calculated as follows:
P(n)=Kp[SG(n)−Final Target(n)] (eq 56)
Here, Kp is the overall controller gain (expressed in Units/hour per mg/dL), SG(n) is the current sensor glucose, n indicates the current sampling point, and Final Target(n) is the calculated final target glucose setpoint from Equation 54. It should be appreciated that Kp is a patient-specific value and, therefore, the actual value of Kp will vary from one person to another. Although the range can vary depending upon the patient, in most typical scenarios, the value of Kp may be within the range of 0.008 to 0.200.
The integral component I(n) can be calculated as follows:
Here, I(n−1) is the integral component from the previous sampling point, n indicates the current sampling point, n−1 indicates the previous sampling point, Kp is the overall controller gain, Ts is the sampling period, T1 is the integral time constant, SG(n) is the current sensor glucose, and Final Target(n) is the calculated final target glucose setpoint from Equation 54.
The derivative component D (n) can be calculated as follows:
D(n)=Kp×TD×dSGdt(n) (eq 58)
Here, Kp is the overall controller gain, TD is the derivative time constant, dSGdt(n) is derivative of the sensor glucose value (pre-filtered to remove noise), and n indicates the current sampling point.
The parameters that need to be set (tuned) for the controller 900 are: KP, τI, and τD (see Equations 56, 57, and 58). For the first three PID parameters, they will be calculated in accordance with the previous studies mentioned below (these previous approaches have resulted in good controller performance). The nominal (i.e., for the case with no insulin feedback) controller gain KP0 is calculated based on the subject's insulin total daily dose ITDD, (in Units/day), as indicated in Equation 59:
Here, KP0 is the default controller gain, and ITDD is the subject's insulin total daily dose in Units/Day.
The purpose of insulin feedback is to allow the controller 900 to deliver more insulin up-front (e.g., at the onset of a meal disturbance), but to prevent over-infusion of insulin, in a similar fashion to the bolus estimator calculations used in existing insulin pumps to prevent the stacking of boluses. Therefore, when insulin feedback is used, the controller gain KP can be adjusted so that at steady state (i.e., basal delivery conditions) the insulin delivery system will deliver the same amount of insulin as in the nominal case. This is accomplished by multiplying the nominal controller gain KP0 (calculated for no insulin feedback) by (1+γ1+γ2+γ3) as indicated below:
KP=KPfactor·KP0·(1+γ1+γ2+γ3) (eq 60)
Here, KP is the overall controller gain, KPfactor is the gain factor for KP, KP0 is the default controller gain, γ1 (0.64935) is a tuning coefficient for the subcutaneous insulin concentration, γ2 (0.34128) is a tuning coefficient for plasma insulin concentration, and γ3 (0.0093667) is a tuning coefficient for effective insulin concentration.
The integral component I(n) is also equipped with anti-windup and saturation constraints to address integral windup issues. This is achieved by calculating an integral clip value (upper limit for the integral component, IClip) as indicated by the following equations:
Here, Imaxfactor is the gain factor for Imax, and Basal is the subject's night time basal rate. In accordance with these expressions, IClip becomes equal to the Imax (which is a constant value) when the sensor glucose value is greater than an upper threshold (UnwindHigh). In certain embodiments, the value of (expressed in Units/hour) may reach about 15. In a typical case, Imax may have a default value of 5.0. When the sensor glucose value is between the upper and lower threshold (UnwindLow), then IClip becomes equal to Iramp(n), which is calculated as indicated in Equation 62.
Here, Kp is the overall controller gain, SG(n) is the current sensor glucose, UnwindLow is the sensor glucose lower threshold, UnwindHigh is the sensor glucose upper threshold, Imax is a constant value, and Setpoint(n) is a user-defined target glucose setpoint.
Finally, if sensor glucose is below the UnwindLow threshold, IClip assumes the value of Ilow(n), which can be calculated by Equation 61:
Ilow(n)=Kp[Setpoint(n)−Unwindlow] (eq 63)
Here, Kp is the overall controller gain, Setpoint is the target glucose defined by the user, and UnwindLow is the sensor glucose lower threshold.
The integral component I(n) as calculated in Equation 57 must be less than or equal to the IClip value as shown in Equation 64:
The insulin feedback components correspond to the remaining terms. As mentioned above, for this particular example γ1=0.64935, γ2=0.34128, and γ3=0.0093667 (the tuning coefficients), while the parameters ISC(n), IP(n), and IEFF(n) correspond to the states of the insulin pharmacokinetic model corresponding to, respectively, the subcutaneous, plasma, and effective site compartments. Therefore, the amount of insulin delivered is reduced in proportion to the predicted insulin concentration in the different compartments.
The model describing the insulin pharmacokinetics (insulin PK) is given by the following expressions:
Here, Ts is the sampling time (which is five minutes for this example), and τs is the time constant for estimated subcutaneous insulin level in minutes (which is set to 50 minutes for this example).
Here, Ts is the sampling time, which is five minutes for this example, τs is the time constant for estimated subcutaneous insulin level in minutes, which is set to 50 for this example, and τp is the time constant for estimated plasma insulin level in minutes, which is set to 70 for this example.
IEFF(n)=α31×ISC(n−1)+α32×IP(n−1)+α33×IEFF(n−1)+β3×ID(n) (eq 66)
For Equation 66:
Here, Ts is the sampling time, which is five minutes for this example, τs is the time constant for estimated subcutaneous insulin level in minutes, which is set to 50 for this example, τp is the time constant for estimated plasma insulin level in minutes, which is set to 70 for this example, and τe is the time constant for estimated interstitial insulin level in minutes, which is set to 55 for this example.
ID(n) is the calculated and administered insulin infusion.
The notation (n−1) indicates values at the previous time step.
ISC is the subcutaneous insulin model estimate/prediction.
IP is the plasma insulin model estimate/prediction.
IEFF is the effective site insulin model estimate/prediction.
For this particular example, the insulin PK model parameters α11, α21, α22, α31, α32, α33, β1, β2, and β3 are set to 0.9802, 0.014043, 0.98582, 0.000127, 0.017889, 0.98198, 1.1881, 0.0084741, and 0.00005, respectively. These values were calculated from PK-PD data for insulin as part of an empirical study and investigation. It should be appreciated that the specific values presented here merely reflect one possible set of suitable values, and that any one or more of these values may be adjusted as appropriate for the particular embodiment.
As mentioned above, previous studies may be leveraged for purposes of configuring and/or operating the PID-IFB control module 906. A first study (Study 1; Panteleon et al., 2006) looked at the effect of changing the controller gain in a study on eight diabetic dogs. The nominal gain was calculated based on the total daily dose of insulin for one experiment; in duplicate experiments this nominal gain was increased and decreased by fifty percent. Insulin delivery in the six-hour postprandial period tended to increase with increasing gain, but this did not achieve statistical significance. This was due to feedback, with lower insulin delivered as glucose levels fell. In essence, the higher gains tended to give more insulin earlier (achieving much better glucose regulation), and reducing the infusion later, whereas the lower gains tended to keep insulin infusion above basal for a longer period of time due to the glucose levels remaining significantly above target. Note that the controller gain affects all of the components of the PID algorithm, including the integral and derivative terms.
A second study (Study 2; Steil et al., 2006) applied a PID controller to ten human subjects. In this study the nominal controller gain was about 42 percent higher than that proposed here, with the integral time constants the same (therefore also a higher integral response), but slightly lower derivative time constants (defined in terms of rising or falling blood glucose, instead of day or night response). In the proposed embodiment presented here, the night-time derivative time constant is just slightly lower than that used in Study 2.
A third study (Study 3; Weinzimer et al., 2008) applied a PID controller to seventeen human subjects. For a subset of eight subjects, no pre-meal bolus was given, while in the remaining nine subjects about 50 percent of the standard meal bolus was given about fifteen minutes before the meal. The controller tuning was the same as that proposed herein. In both cases, performance was acceptable, with the pre-meal bolus helping to lower the postprandial peak glucose excursion. When compared with home treatment using standard pump therapy, both closed-loop algorithms were superior, reducing both glucose excursions above 180 mg/dl and below 70 mg/dl.
One of the observations from Study 3 was that the meal-related insulin infusion persisted above pre-meal levels for at least four hours after the meal. This result led to the introduction of insulin feedback to the algorithm, which serves to compensate for insulin already delivered, while at the same time allowing for more aggressive action at the start of meals, where it is needed.
In particular implementations, some of the parameters mentioned above for the PID-IFB control module 906 can be fixed. In this regard, the following values may be utilized in an exemplary embodiment. It should be appreciated that these values are provided here for illustrative purposes only, and that an implementation of the PID-IFB control module 906 may utilize different values if so desired.
The values of the PID parameters mentioned above (KP, τI, and τD) are not expected to change, but it may be desirable to adjust them if doing so would improve the response of the glucose excursion. The nominal values for these PID parameters, together with exemplary allowable ranges, are shown in Table 2.
The lower bound for the controller gain KP0 would lower it by 95 percent from the nominal, resulting in an overall less aggressive response of the controller 900, and resulting in less insulin delivered for the same glucose level and trend. The upper bound allows this gain to be increased up to 20 times from the nominal. Although this would result in more insulin being given, this is always with respect to the actual glucose level and its derivative, rapidly falling blood glucose, even with an elevated blood glucose level, results in a decreased delivery of insulin due to the derivative component.
The integral time constant τI determines how fast a deviation in the blood glucose from the desired glucose target accumulates. A higher value results in a slower response to persistent deviations in glucose from target. Lastly, the derivative time constant τD can safely be decreased all the way to zero, as the lower it is the less insulin will be delivered. Anything much higher than the upper bound of 500 minutes would make the controller too sensitive to small changes in the rate of change of the sensor glucose, which even though it is not necessarily dangerous, is still undesirable in terms of overall performance.
Insulin Limit
As mentioned previously with reference to
The illustrated embodiment of the insulin limit module 908 receives at least the following items as inputs: a fasting blood glucose value 5202 that is associated with the user/patient; a total daily insulin value 5204 that is associated with the user/patient; and fasting insulin delivery data 5206 that is associated with the user/patient. Some or all of this input data may be provided directly or indirectly by the insulin delivery system 14 (see
The operation of the insulin limit module 908 will now be described with reference to
The fasting blood glucose value 5202 indicates a patient-specific measure of blood glucose taken after a period of fasting. For example, the fasting blood glucose value 5202 could be based on one or more blood glucose measurements obtained the morning after the patient wakes up (assuming that no food was ingested overnight, etc.). In practice, the fasting blood glucose value 5202 may be obtained from a continuous glucose sensor, a blood sample device—such as a finger stick measurement device, or a combination thereof. Of course, the actual fasting blood glucose of the patient may vary from one day to another, on a weekly basis, or the like. Accordingly, the current fasting blood glucose value 5202 obtained at task 5254 may be an average value, a median value, or any statistical value that is calculated based on at least some historical patient data. The fasting blood glucose value 5202 could be calculated instantaneously (based on historical data) whenever the process 5250 is performed, or the fasting blood glucose value 5202 could be pre-calculated and stored in accordance with any desired schedule, such that pre-calculated values can be accessed at task 5254. Regardless of the manner in which the fasting blood glucose value 5202 is obtained or derived, the specific value preferably remains constant during the period of closed-loop operation.
The TDI value 5204 indicates an amount of insulin delivered to the patient over a specified 24-hour period. The amount of insulin delivered to a person will usually vary from day to day, depending on a variety of factors. Accordingly, the TDI value 5204 obtained at task 5254 may be an average value, a median value, or any statistical value that is calculated based on at least some historical patient data. In certain embodiments, for example, the TDI value 5204 is calculated as a running average based on the last five days, wherein the insulin delivery information is recorded and provided by the insulin infusion device. In practice, the TDI value 5204 could be calculated instantaneously (based on historical data) whenever the process 5250 is performed, or the TDI value 5204 could be pre-calculated and stored in accordance with any desired schedule, such that pre-calculated values can be accessed at task 5254. Regardless of the manner in which the TDI value 5204 is obtained or derived, the specific value preferably remains constant during the period of closed-loop operation.
For this implementation, the fasting insulin delivery data 5206 indicates an amount of insulin delivered to the user during a specified fasting period, such as the fasting period that precedes the time corresponding to the fasting blood glucose value 5202. In this regard, if the fasting blood glucose value 5202 is obtained for a given ten-hour window of time, then the fasting insulin delivery data 5206 will also correspond to that same ten-hour window of time. Consequently, the fasting insulin delivery data 5206 is correlated to the fasting blood glucose value in this manner. In certain embodiments, the fasting insulin delivery data 5206 includes information about each insulin delivery operation (that occurred during the designated fasting period). In this context, the fasting insulin delivery data 5206 may include, without limitation: a quantity of insulin (in Units); an insulin delivery rate (in Units/hour); timestamp data; and other data for each insulin delivery operation of interest. Accordingly, the fasting insulin delivery data 5206 includes timing information associated with insulin delivery during the fasting period, rather than a simple accumulated total volume or quantity of insulin.
Referring again to
In certain embodiments, the maximum insulin infusion rate, Umax, is calculated in accordance with the following expression (Equation 67):
Here, Umax is the maximum insulin infusion rate (expressed in Units/hour) bounded at BGLBL (see Equation 67a below), and Ibasal,0 is an estimated basal rate (expressed in Units/hour) that will bring the patient's fasting blood glucose to a value equal to FBG0. Thus, Ibasal,0 is a basal rate that in theory results in the fasting blood glucose value 5202 obtained for the patient (see
BGLBL=Setpoint−ILB (eq 67a)
Here, BGLBL (mg/dL) is a lower blood glucose limit that corresponds to operation of the insulin infusion device at Umax, Setpoint is the target glucose setpoint value 944 (
Here, KI is insulin gain, and IS is an insulin sensitivity value for the user, expressed in mg/dL/U. For this particular embodiment, KI is estimated by the “1800 rule” as IS=1800/TDI (where TDI is the TDI value 5204). Accordingly, IS and KI are derived from the TDI value 5204 obtained for the user. In a typical implementation, the value of Umax is calculated on a per-patient basis, and the typical range for Umax is between 0.5 to 3.0 Units/hour.
Referring back to
As explained previously with reference to the PID-IFB control module 906, the insulin dose, PIDRate(n), is calculated in accordance with a PID-IFB control algorithm that considers a proportional component, an integral component, and a derivative component. The process 5250 is suitably designed to inhibit or prevent windup of the integral component when the insulin infusion rate is limited. In this regard, the process 5250 may initiate an integral windup prevention scheme (task 5264) when the closed-loop insulin infusion rate calculated by the PID-IFB control module 906 is greater than the calculated insulin limit 959. For this exemplary embodiment, the PID-IFB control module 906 responds to the application of the insulin limit 959 by effectively freezing the integral component while permitting continued updating of the proportional and derivative components of the PID-IFB control algorithm.
Thus, if the insulin limit 959 is active, then the integral component of the PID algorithm is frozen to its previous value. This feature can be employed to assist in the prevention of integral windup. This information is used by the PID-IFB control module 906 for use during the next PID calculation, as indicated here:
In Equation 70, I(n) represents the current value of the integral component of the PID-IFB control algorithm. The integral windup scheme enables the proportional and derivative components of the PID-IFB control algorithm to remain active for purposes of dynamically adjusting PIDRate(n) in a manner that does not accumulate error that would otherwise be associated with the integral component.
Referring again to
In certain embodiments, Umax is calculated once for each period of closed-loop operation. Accordingly, if the next sampling point for the current period of closed-loop operation has been reached (the “Yes” branch of query task 5270), the process 5250 returns to task 5258 to calculate the next closed-loop insulin infusion rate (which may or may not need to be limited by the previously calculated value of Umax), and to proceed as described above. If the closed-loop operating mode has ended, then the process 5250 exits such that it can be restarted at the beginning of the next period of closed-loop operation.
IOB Compensation Module
The manual boluses delivered 960 may be collected and saved in association with the IOB compensation module 910 for use as bolus history 992. In this regard, the bolus history 992 may include any number of past bolus amounts administered during a given period of time. The IOB compensation module 910 may also use a number of constants, parameters, coefficients, configuration settings, gain values, or the like (for simplicity, the constants 994 shown in
The IOB compensation module 910 provides an additional safeguard that estimates the insulin on board the body of the patient from manual boluses that were delivered prior to activation of the closed-loop mode, to compensate the final insulin dose 962 and help avoid over-delivery of insulin. When the system initially enters the closed-loop mode, the IOB compensation module 910 considers the manual boluses delivered 960 that have been administered over a defined period of time (e.g., the last eight hours), and subtracts the manual boluses. Thereafter, during the closed-loop mode, the IOB compensation module adjusts for manual boluses that were delivered during each sampling period (e.g., every five minutes).
In the context of the process 1100, the IOB compensation module 910 calculates insulin on board from manual boluses at each cycle and compensates the controller output rate (insulin infusion rate) if active IOB is greater than a certain threshold. Accordingly, the process 1100 calculates, generates, or otherwise obtains a current IOB value (task 1104) that represents an estimate of active insulin in the body of the user. In certain embodiments, the active IOB is estimated in accordance with a discretized three-compartment insulin pharmacokinetic (PK) model, as indicated below:
IOB(n)=ci1·IOB(n−1)+ci2·IOB(n−2)+ci3·IOB(n−3)+cb0·Ubolus(n)+cb1·Ubolus(n−1)+cb2·Ubolus(n−2) (eq 71)
Here, IOB is the active insulin on board, Ubolus is the amount of manual bolus delivered in units per sample, n is the current sampling point, n−1 is the last sampling point, n−2 is the second to last sampling point, and n−3 is the third to last sampling point. Accordingly, the process 1100 obtains the current IOB value, IOB(n), based at least in part on historical bolus delivery data for the user (see the manual boluses delivered 960 and the bolus history 992 in
ci1=eaxx3+eaxx4+eaxx5 (eq 71a)
ci2=−(eaxx3×eaxx4+(eaxx3+eaxx4)×eaxx5) (eq 71b)
ci3=eaxx3×eaxx4×eaxx5 (eq 71c)
cb0=1
cb1=dprod×(−(daxx22×eaxx3+daxx22×eaxx4)×axx3×axx4+(daxx31×eaxx3+daxx31×eaxx5)×axx3×axx5−(daxx32×eaxx4+daxx32×eaxx5)×axx4×axx5) (eq 71d)
cb2=dprod×(daxx22×eaxx3×eaxx4×axx3×axx4+daxx32×eaxx4×eaxx5×axx4×axx5−daxx31×eaxx3×eaxx5×axx3×axx5) (eq 71e)
Where:
axx3=1/τSC (eq 71f)
axx4=1/τP (eq 71g)
axx5=1/τeff (eq 71h)
eaxx3=e−axx3·TsC (eq 71i)
eaxx4=e−axx4·TsC (eq 71j)
eaxx5=e−axx5·TsC (eq 71k)
daxx22=axx4−axx3 (eq 711)
daxx31=axx5−axx3 (eq 71m)
daxx32=axx5−axx4 (eq 71n)
dprod=−1/(daxx22×daxx31×daxx32) (eq 71o)
In the above equations TsC indicates a modified sampling interval in minutes, which can be calculated as TsC=Ts*6/CurveSpeed, where Ts is the sampling interval and CurveSpeed is the insulin on board decay speed rate in hours. τSC, τP, and τeff are the respective time constants of the subcutaneous, plasma, and effective compartments of the insulin PK model.
IOB (in Units) as calculated by Equation 71 represents the residual active insulin in the body from manual boluses (that may have been administered before the start of the closed-loop mode or during closed-loop operation) which must be taken into account for calculation of the final insulin delivery rate. This is achieved by first calculating the IOB rate (task 1106) and then deducting the IOB rate from the PID-IFB calculated infusion rate, as indicated below. Accordingly, in certain situations the process 1100 determines an adjusted insulin infusion rate (task 1108) based at least in part on the calculated IOB rate and the uncompensated insulin infusion rate, PIDRate(n):
Note that the process 1100 calculates the IOB rate, IOBRate(n), based at least in part on the current IOB value, IOB(n). The IOB rate, which is expressed in Units per hour (U/h), represents the amount of active insulin accumulated from manual boluses in the body per unit of time. Hence, this additional insulin which is already present in the body is deducted from the controller delivery rate (PIDRate). This accounts for all the manual boluses that have been administered by the user, and minimizes the possibility of over-delivery by the controller. Here, GainIOB is the IOB decay rate in h−1, and MinIOB is the minimum IOB required to compensate the PIDRate (where MinIOB is expressed in Units). Thus, the IOB rate is calculated to be equal to the current IOB value multiplied by an IOB decay rate when the current IOB value is greater than a minimum IOB value, and is calculated to be equal to zero when the current IOB value is less than or equal to the minimum IOB value. In this regard, MinIOB is a minimum threshold for IOB, below which the effect of IOB on glucose is considered to be negligible; hence not needed to be compensated.
As reflected in Equation 73, the process 1100 selects the adjusted insulin infusion rate to be the maximum of zero or the difference between the uncompensated insulin infusion rate and the calculated IOB rate (from task 1106). Note that the difference between PIDRate and IOBRate could be negative, as these values are calculated from different sources. PIDRate is the controller calculated infusion rate and IOBRate is the accumulated active insulin in the body obtained from manual boluses. Accordingly, Equation 73 ensures that the AdjustedRate does not fall below zero.
Next, the process 1100 may calculate, select, or otherwise determine a final insulin infusion rate (task 1110). In certain embodiments, the final insulin infusion rate (the final insulin dose 962 in
As indicated by this expression, the process 1100 selects either the adjusted insulin infusion rate (AdjustedRate(n)), the uncompensated insulin infusion rate (PIDRate(n)), or the current basal rate (Basal) to serve as the final insulin infusion rate (FinalRate(n)) for the insulin infusion device. Here, PIDRate is the insulin infusion rate as calculated by the PID-IFB control module 906 and Basal is the current pre-programmed pump basal rate. Accordingly, task 1110 selects the final insulin infusion rate to be equal to the uncompensated insulin infusion rate when the current basal rate is greater than or equal to the uncompensated insulin infusion rate. In contrast, when the current basal rate is less than the uncompensated insulin infusion rate, task 1110 selects the final insulin infusion rate to be either the current basal rate or the adjusted insulin infusion rate, whichever is higher.
In the context of task 1110, the PIDRate is used as the FinalRate (when PIDRate is less than or equal to Basal) to allow the controller to “apply brakes” (in other words, suppress insulin delivery rate) in order to prevent any potential hypoglycemia. On the other hand, when PIDRate is greater than Basal, the FinalRate will be the maximum of Basal or AdjustedRate, which ensures that the insulin adjustment only accounts for the insulin coming from boluses, and not the basal. A lower bound (i.e., the value of Basal) is applied to the FinalRate when PIDRate is greater than Basal; this lower bound is utilized to prevent over-compensation of insulin on board under these circumstances.
The process 1100 may continue by communicating or otherwise providing the final insulin infusion rate, FinalRate(n), to the insulin infusion device (task 1112). For embodiments where the process 1100 is executed natively by the insulin infusion device itself, then the process 1100 may simply provide the final insulin infusion rate to the processing logic or fluid delivery control module of the infusion device. In turn, the insulin infusion device responds by regulating delivery of insulin in accordance with the final insulin infusion rate.
This description assumes that the process 1100 is repeated for each sampling time. For the next sampling time, therefore, the value of n may be incremented by one (or by any desired amount) to establish an index for the next iteration of the process 1100 (task 1114). Thereafter, the process 1100 may return to task 1102 to obtain the newest input data values and to repeat the various tasks described above. Accordingly, the process 1100 facilitates regulation of insulin delivery to the body of the user in a controlled and ongoing manner by continuously adjusting the final insulin infusion rate while the system is operating in the closed-loop mode.
In certain embodiments, it may be desirable to adjust some of the parameters utilized by the IOB compensation module 910 if doing so would improve performance. The nominal values for these parameters, together with exemplary allowable ranges, are shown in Table 3.
Insulin Delivery Timeout Module
The insulin delivery timeout module 912 is suitably designed and configured to continuously monitor (during the closed-loop mode) if the patient is receiving insulin at the insulin limit (Umax) or no insulin (Umin, which may be defined as insulin delivery at little to no Units/Hour) for a prolonged period of time. If one of these insulin delivery conditions is detected, the system will issue a warning and continue operating under the closed-loop mode. As mentioned previously, the insulin delivery timeout module 912 may process the insulin delivered 960 as an input.
Accordingly, the insulin delivery timeout module 912 introduces an additional safeguard that checks for a series of steps as described below for the delivery of insulin at the insulin limit (Umax Timeout) or no insulin (Umin Timeout) for a prolonged period of time. This is achieved by calculating the total amount of insulin delivered by the system during the closed-loop mode in a pre-specified moving window that is identified as the insulin time window.
Regarding the Umin Timeout condition, once the insulin time window for Umin (which may be defined as delivering insulin at zero Units/Hour) has been reached from the start of the closed-loop mode, the system will monitor the amount of insulin being delivered at the user specified insulin time window and compare it with the amount that could have been delivered if operating at the patient's basal rate for the same time span, as shown in the following logical expression.
Here, Pump Delivery Rate (Units/Hour) is the infusion rate, which is either equal to the FinalRate from Equation 74 (i.e., the infusion rate calculated by the controller during the closed-loop mode), or a pre-programmed overnight basal rate that is used during open-loop operation. The quantity UTotWinMin is the total amount of insulin delivered (in Units) by the closed-loop control algorithm in a user-specified insulin time window for Umin, and the quantity UBasalWinMin is the total amount of insulin that could be delivered if operating at a pre-programmed overnight basal rate in the same insulin time window for Umin. The parameter MinDeliveryTol is a user-specified tolerance, in percentage of UBasalWinMin, that the system has to deliver in order to stay in the closed-loop mode.
In accordance with this particular example, closed-loop control continues as long as the total amount of insulin delivered during the insulin time window (which is set to 120 minutes for this example) by the system is greater than the total amount of insulin that might have been delivered if operating at five percent (which is the default minimum tolerance for this example) of basal. Moreover, a fail-safe alert is triggered once the total amount of insulin delivered during the insulin time window (120 minutes) by the system is less than five percent of basal.
Regarding the Umax Timeout condition, once the insulin time window for Umax has been reached from the start of the closed-loop mode, the system will monitor the amount of insulin being delivered at the user specified insulin time window and compare it with the amount that might be delivered if operating at the Umax rate for the same time span, as shown in the following logical expression.
Here, Pump Delivery Rate is the infusion rate, which is equal to either the FinalRate or a pre-programmed overnight basal rate that is used during operation in the open-loop mode. The quantity UTotWinMax is the total amount of insulin delivered (in Units) by the closed-loop control algorithm in a user-specified insulin time window for Umax, and the quantity UmaxWinMax is the total amount of insulin that could have been delivered in the user-specified moving window for Umax if operating at a calculated Umax rate. The parameter MaxDeliveryTol is a user-specified tolerance, in percentage of UmaxWinMax, that the system must adhere to in order to stay in the closed-loop mode.
In accordance with this particular example, closed-loop control continues as long as the total amount of insulin delivered during the insulin time window (which is set to 600 minutes for this example) by the system is less than the total amount of insulin that might have been delivered if operating at 95% (which is the default maximum tolerance for this example) of Umax. Moreover, a fail-safe alert is triggered once the total amount of insulin delivered during the insulin time window (600 minutes) by the system is greater than the total amount that might have been delivered if operating at 95% of Umax.
In certain embodiments, it may be desirable to adjust some of the parameters utilized by the insulin delivery timeout module 912 if doing so would improve performance. The nominal values for these parameters, together with exemplary allowable ranges, are shown in Table 4.
Model Supervisor Module
The model supervisor module 914 is similar to the insulin delivery timeout module 912 in that it monitors and polices the system during closed-loop operation. In practice, closed-loop systems are only aware of signals (inputs) that are being provided by the measurement device(s). If measurements deviate from true values, the control system may react to the deviation. When using continuous glucose sensors for diabetes, the sensors provide the measurements to the closed-loop control system and, based on these measurements, insulin is delivered to the subject. Accordingly, sensor performance and integrity should be closely monitored. Fortunately, there is a relationship between the insulin and meal intake to the glycemic response. This relationship can be translated into a mathematical model that is able to predict the sensor glucose response based upon the insulin delivered. The sensitivity of sensor glucose to insulin delivered is patient-specific (the sensitivity can usually be learned over a period of three to six days for each patient).
The model supervisor module 914 uses a mathematical model that enables individualization of a patient's unique blood glucose time-dependent response. The model describes the sensor glucose time-dependent response as a function of insulin and meal intake. The exemplary mathematical model described herein has a number of benefits and advantages: it is linear, physiologically-based, and includes only parameters that have direct connection to measurable data (sensor glucose and insulin delivered). These features are important because a linear model is easy to analyze and predict. Moreover, a physiological-based model facilitates an understanding of the origin of the predictions (e.g., insulin sensitivity, meal intake, etc.), and use of measurable data reduces the need to estimate unobserved variables (e.g., metabolism, cell activity, etc.).
As described in more detail below, the model supervisor module 914 considers a “moving window” that includes a historical time period 1122 that is defined from the present (most recent sampling period 1120) back to the beginning of the LTH. The moving window considered by the model supervisor module 914 may also include insulin history that precedes the LTH, as depicted in
The process 1150 represents one iteration that is performed for a current sampling point or time, which corresponds to the most recent sampling period. This example assumes that the insulin infusion device is already operating in the closed-loop mode (task 1152) to deliver insulin to the body of the user, and that the process 1150 receives relevant data in accordance with a predetermined schedule (e.g., a sampling period of five minutes). Accordingly, the process 1150 receives, obtains, or accesses a variety of inputs that may have an influence on the operation of the model supervisor module 914 (task 1154). For example, the process 1150 may receive at least the following data for the current sampling period: current insulin-delivered data that indicates an amount of insulin delivered by the insulin infusion device during the most recent sampling period; current sensor data that indicates a current sensor glucose value for the user, which corresponds to the most recent sampling period; and a current sensor calibration factor, which may be needed to compensate for recent meter-based calibrations. Any amount of historical data could also be received during task 1154 if so desired. Thus, some amount of redundancy may be built into the system (which may be desirable to account for missed transmissions, lost packets, or the like). The sensor data may be received and processed in any suitable form. For example, a continuous glucose sensor may generate Isig (electrical current) values that can be mapped to sensor glucose values. The model supervisor module 914 may be suitably configured to process Isig values directly, or it could translate or map the raw Isig values into any desired representation.
The process 1150 may also access or retrieve historical data that was received for past sampling periods (task 1156). Task 1156 may represent an initialization routine that populates a grid, matrix, or other type of database structure as needed to prepare the model supervisor module 914 for the various calculations, analyses, and functions described in more detail below. It should be appreciated that subsequent iterations of the process 1150 (which are performed in an ongoing manner during the closed-loop mode) need not repeat the initialization of the historical data. Rather, task 1156 may simply adjust the data history to reflect newly received data. For the embodiments described here, the following historical data may be processed by the model supervisor module 914, without limitation: historical insulin-delivered data for the user; and historical sensor glucose values for the user. The historical insulin-delivered data may correspond to amounts of insulin delivered by the insulin infusion device during each historical sampling period of interest, and the historical sensor glucose values may correspond to respective sensor glucose measurements obtained during each historical sampling period of interest. In certain implementations, each historical sensor glucose value may be associated with or derived from a historical Isig value and a sensor calibration factor.
The process 1150 is iterative in nature, and each iteration considers data associated with the defined historical period of time (see
Next, the process 1150 processes at least some of the historical data to determine a best-matched solution to a sensor glucose prediction model (task 1160). Task 1160 may be considered to be a training procedure that finds the best-fitting sensor glucose prediction function, which in turn can be used to check (predict) the integrity and quality of the glucose sensor. In certain embodiments, the sensor glucose prediction model is expressed as a fourth order ordinary differential equation that, when solved given the initial conditions, provides model-predicted sensor glucose values. Notably, task 1160 uses the actual sensor glucose values obtained during the model training period (and does not use any of the actual sensor glucose values obtained during the model prediction period) to determine which candidate solution will be selected as the best-matched solution. Conceptually, task 1160 generates a plurality of curves (or discrete values that may be used to visualize curves for purposes of this explanation) and compares the portion of the curves within the model training period to the actual sensor glucose values obtained during the model training period. In an ideal scenario with a perfect match, one of the generated curves will precisely track the actual sensor glucose values within the model training period. In practice, however, the generated curves will deviate from the actual sensor glucose values. Accordingly, task 1160 identifies the calculated curve that best matches the actual sensor values. It should be appreciated that this best-matching curve also includes model-predicted sensor glucose values that extend beyond the model training period and into the model prediction period.
The process 1150 may continue by comparing at least one historical sensor glucose value obtained during the model prediction period to at least one corresponding predicted sensor glucose value of the best-matched solution (task 1162). In certain embodiments, task 1162 checks only one actual sensor glucose value: the current sensor glucose value obtained for the most recent sampling period. In other embodiments, any or all of the sensor glucose values obtained during the model prediction period could be analyzed during task 1162. The straightforward example described here only considers the current sensor glucose value such that the comparison in task 1162 is simple and straightforward. In this regard, task 1162 may calculate a difference between the current sensor glucose value (i.e., the most recent historical value) and the predicted current glucose value for the most recent sampling period (the difference may be expressed as an absolute value), and the process 1150 may continue by comparing the calculated difference to a threshold error amount (query task 1164). In other embodiments, the comparison performed during task 1162 may involve a more advanced methodology, e.g., curve-fitting that considers more than one sampling point in the model prediction period, statistical analyses, or the like. For example, rather than calculating error on a point-by-point basis, the process 1150 could utilize any appropriate methodology to determine whether or not the historical sensor glucose values in the model prediction period deviate (by at least a threshold amount) from the corresponding model-predicted values in the best-matched solution.
If the calculated error between the model-predicted glucose value(s) and the corresponding actual historical sensor glucose value(s) is less than or equal to the error threshold, or otherwise satisfies the predetermined criteria monitored by the model supervisor module 914, then the “No” branch of query task 1164 is followed and the process 1150 continues to the next sampling period (task 1166). At this point, the process 1150 returns to task 1152 such that the core of the process 1150 can be repeated to consider the data received for the next sampling period. Thus, the oldest data considered by the previous iteration of the process 1150 is disregarded, the newly received data is designated as the “most recent” data, and the historical time period or “analysis window” for the current iteration of the process 1150 shifts by one sampling period (see
If the calculated error exceeds the threshold error amount (the “Yes” branch of query task 1164), then the process 1150 may generate an alert, an alarm, and/or a message (task 1168). In practice, an alert, alarm, or message can be initiated by the model supervisor module 914 for rendering, annunciation, delivery, playback, etc. For example, an alert could be presented at the insulin infusion device, at a remote monitoring station, at a handheld controller device, or the like. In certain embodiments, the process 1150 switches from the closed-loop mode to the open-loop mode (or to some type of safe operating mode with reduced insulin delivery) when the threshold error amount is exceeded (task 1170).
One important aspect of the process 1150 relates to the manner in which the best-matched sensor glucose prediction model is chosen (see task 1160). In this regard,
As mentioned previously, the exemplary sensor glucose prediction model utilized here is expressed as a fourth order ordinary differential equation. In accordance with conventional mathematics, the model-predicted sensor glucose values (G) in time are calculated as a function of the two model prediction initial conditions G0 and dG0. Here, G0 is the estimated sensor glucose value for the begin-training sampling period 1124 (the start of LTH in
For the exemplary embodiment presented here, the initial condition dG0 is bounded in a simple manner that is based on a predetermined parameter (which may be adjustable): dG0=±grad_bound. In contrast, the boundary for the initial condition G0 is based on (or is otherwise influenced by) a baseline historical sensor glucose value obtained during the model training period, such as the sensor glucose value that was obtained during the begin-training sampling period 1124. Thus, the process 1180 may identify, from the historical sensor glucose values, the baseline sensor glucose value to be used for purposes of calculating the boundary for the initial condition G0: G0=SGk-LTH±0.14·SGk-LTH where SGk-LTH is the baseline sensor glucose value obtained for the earliest sampling period in the historical time period under analysis (see
The process 1180 may then continue by determining, calculating, or otherwise obtaining the next set of initial conditions (task 1184). The manner in which the process 1180 selects and steps through the different initial conditions is unimportant in this context. The current set of initial conditions is used to calculate a candidate solution to the sensor glucose prediction model (task 1186). As mentioned above, each candidate solution is calculated as a function of the two bounded initial conditions. Moreover, each candidate solution is calculated as a function of estimated plasma insulin for the user, which in turn is calculated as a function of an amount of insulin delivered to the user. Accordingly, task 1186 may estimate plasma insulin for the user based on: the current insulin-delivered data (obtained for the most recent sampling period); the historical insulin-delivered data; and an insulin basal rate for the user. In practice, task 1186 considers the total insulin (basal, bolus, and any other insulin delivered) for all sampling periods. This allows the process 1180 to obtain the candidate solution to the sensor glucose prediction model based at least in part on the estimated plasma insulin and based at least in part on the baseline sensor glucose value obtained at the earliest sampling period under analysis.
The process 1180 may continue by generating a training error value, quantity, or function for the candidate solution (task 1188). The training error may be calculated by comparing predicted sensor glucose values from the candidate solution to the corresponding historical sensor glucose values, to obtain a metric that indicates how closely the predicted values match the actual values. In certain embodiments, the training error is based only on predicted values and actual values for the model training period (LTH in
If the process 1180 has considered all of the initial condition combinations (the “Yes” branch of query task 1190), then the process 1180 may proceed to a task 1192. If more sets of initial conditions remain (the “No” branch of query task 1190), then the process 1180 may return to task 1184, retrieve the next set of initial conditions, and continue as described above. Task 1192 is performed after a plurality of different candidate solutions have been calculated, using the different sets of initial conditions. Task 1192 may be performed to select the best-matching candidate solution from the plurality of calculated solutions. For this particular embodiment, the selection is based on the training errors generated during task 1188. For example, the candidate solution having the lowest training error may be selected as the best-matching solution.
It should be appreciated that the process 1180 need not be performed in the illustrated sequence, and that some tasks could be performed in parallel. For example, the calculation of the training errors (task 1188) may instead be performed after all of the candidate solutions have been obtained and saved. Moreover, the process 1180 could be designed to immediately eliminate candidate solutions (following completion of task 1188) that have a training error that exceeds a predetermined error threshold. As another option, the process 1180 could be designed to immediately designate a candidate solution as the best-matched solution if the associated training error satisfies certain criteria.
The foregoing concepts and methodologies may be implemented in a practical embodiment of the model supervisor module 914. The following description relates to two possible embodiments that implement the general concepts presented above. It should be appreciated that the particular embodiments described below are not exhaustive, and that the description of the embodiments is not intended to limit or restrict the scope or application of the subject matter presented here.
Model Supervisor Module: First Representation
The model supervisor module 914 is suitably designed and configured to detect potentially faulty sensor measurements. The model supervisor module 914 may utilize a mathematical model that is trained offline. For example, parameters that could be estimated offline estimated include, without limitation: KI (insulin gain in mg/dL per U/h; τ1 (first insulin time constant, in minutes); τ2 (second insulin time constant, in minutes); Ibasal (basal insulin, in U/h); and SGbase (blood glucose (BG) at fasting, in mg/dL, when Ibasal insulin is delivered.
The model supervisor module 914 trains the model prediction initial conditions, G0 and dG0 every sampling time. G0 and dG0 represent the BG (mg/dL) and BG derivative (mg/dL/min) estimate values at k-LTH (see
G0=CGMk-LTH±0.14·CGMk-LTH
dG0=±grad_bound (eq 77)
For Equation 77, CGMk-LTH is the CGM measurement at sampling time k-LTH, and grad_bound is a predefined absolute maximum BG derivative in time (mg/dL/min)
The model supervisor module 914 estimates plasma insulin, Ip, at k-LTH using the insulin history records from k-LTH-insulin history and k-LTH (see
In Equation 80, Fault 1 indicates a faulty sensor, Fault 0 indicates a non-faulty sensor, and Fault −1 indicates that there is not enough information to decide. Referring again to
In certain embodiments, some of the parameters used by the model supervisor module 914 may be adjustable. Table 5 identifies the adjustable parameters, along with some exemplary values for the parameters.
The following equations describe the mathematical model equations in Laplace transform form:
In this expression, α=3500, β=120, and ÎP is the IP in deviation form. ÎP0 and dIP0 are the ÎP and derivative initial conditions, respectively.
All the insulin states are formulated in deviation form from the given insulin value Ibasal as it is expressed by the following Equation 82:
Îx=Ix−Ibasal (eq 82)
In Equation 82, x represents D, in, or P.
The following Equation 83 expresses BG in deviation form from SGbase. Note that Equation 83 represents one suitable expression for the sensor glucose prediction model, which is a fourth order ordinary differential equation.
In Equation 83, the following relationships hold:
α=120+τ1+τ2
β=3500+120τ1+120τ2+τ1τ2
χ=3500τ1+3500τ2+120τ1τ2
δ=3500τ1τ2
Moreover, in Equation 78, Ĝ, KI, Ĝ0, dG0, τ1, and τ2 are the BG in deviation form from SGbase, the insulin gain, the BG initial conditions in deviation form, the BG derivative initial conditions, and two time constants, respectively.
Model Supervisor Module: Second Representation
In accordance with some embodiments, the functionality of the model supervisor module 914 can be represented as follows. As mentioned above, the model supervisor module 914 estimates the user's glucose concentration in real-time based on the insulin delivered, the sensor Isig values, and sensor calibration factors. If the model-predicted sensor glucose value (SG) and the actual SG value differ significantly, the system will trigger a fail-safe alert that indicates that the collected data contains unexplained behavior, which may in turn be associated with a faulty sensor and/or insulin delivery, or an unannounced meal intake.
The time frames and reference time periods for the model supervisor module 914 are defined as shown in
The following equations describe the mathematical model in Laplace transform form. Equation 84 provides an estimate of the plasma insulin, and Equation 85 provides the model-predicted SG values. Accordingly, the model supervisor module 914 in accordance with this particular embodiment estimates the plasma insulin as follows:
For this example, ε=3500, γ=120, ÎP is the estimated plasma insulin in deviation form, (s) refers to Laplace transform form, and ÎD is the insulin delivered from the system in deviation form. Moreover, ÎP0 is the estimated plasma insulin in deviation form for the sampling time identified as k-LTH (see
The insulin states described above are formulated in deviation form from a given insulin value Ibasal as it is expressed by the following equation:
Îx=Îx−Ibasal (eq 85)
In Equation 85, x represents D or P (where D refers to insulin delivered and P refers to plasma insulin), and Ibasal,0 is the estimated basal rate defined for each user to bring the patient to a fasting blood glucose (FBG) of the value FBG0 (in mg/dL).
For this second embodiment, the model-predicted sensor glucose value, Ĝ, in time is calculated in accordance with Equation 83 and the associated relationships, as described for the first embodiment of the model supervisor module 914. In this regard, Ĝ is the model-predicted SG value in deviation form from FBG0 (estimated blood glucose using meter blood glucose readings at the end of the night period), (s) refers to Laplace transform form, τ1, and τ are the two insulin time constants identified for each patient, which are related to how fast a patient reacts to insulin, KI is the insulin gain, and ÎP is the estimated plasma insulin in deviation form. Moreover, Ĝ0 is the estimated SG value (in mg/dL) in deviation form for the sampling time of k-LTH (see
The estimated blood glucose values are calculated as a function of the model prediction initial conditions, G0 and dG0. For this particular embodiment, the estimations for G0 and dG0 are bounded as formulated by the following equations. Note that these initial conditions and their boundaries were also described above with reference to task 1182 of the sensor model training process 1180.
G0=SGk-LTH±0.14·SGk-LTH (eq 86)
dG0=±grad_bound (eq 87)
Here, G0 is the estimated SG value (in mg/dL) value for the sampling time of k-LTH, SGk-LTH is the SG measurement for the sampling time of k-LTH, dG0 is the derivative of the estimated SG value (in mg/dL/min) for the sampling time of k-LTH, and grad_bound is a predefined absolute maximum SG derivative in time (mg/dL/min). For certain embodiments, grad_bound is a fixed parameter. For the example presented here, grad_bound has a value of 5 mg/dL/min.
The model prediction facilitates the calculation of two values: Terr and Perr. Terr is defined as the mean absolute error between the model predicted SG values and the actual SG records for the sampling time identified as k-LTH and k as calculated by Equation 88 below. Perr is defined as the mean absolute error between the model predicted SG values and the actual SG records for sampling times identified as k to present (see
Here, Terr is defined as the mean absolute error between the model predicted SG values (Modeli) and the SG records (SGi) for the sampling time identified as k-LTH and k.
Here, Perr is defined as the percentage of error between the model prediction and the SG measurement at the present (most recent) sampling time.
In accordance with this particular implementation, the model supervisor module 914 estimates the fault scenario based on Equation 90, where Fault 1 signifies a faulty sensor, Fault 0 indicates a non-faulty sensor, Fault 3 indicates a training error, and Fault −1 indicates that there is not enough data available to make a determination.
In Equation 90, err1 is the upper threshold for the mean absolute error (see Equation 88). Thus, if the training error is above this threshold, a fault cannot be triggered because the credibility of the training is suspect. The err2 is the lower threshold for Equation 89. If the prediction value of the model and the sensor measurement at present are above this threshold and the training error is less than err1, then a fault will be triggered. The err3 defines a lower threshold for the training period. If Equation 88 indicates a value that is above this threshold, then a warning associated with poor training can be triggered.
The top plot 1206 in
Referring to the top plot 1206, there is good agreement between the model predicted SG values (represented by the solid line 1212) and the actual SG measurements (represented by the dots). In other words, the actual measurements do not deviate significantly from the predicted values. In certain embodiments, the model supervisor module 914 only compares actual measurement values that are within the LPH time frame. In accordance with one exemplary embodiment, the model supervisor module 914 determines the fault status based solely on the most recently obtained data, i.e., the information received for the last sampling time. For this example depicted in
Referring to the middle plot 1208, of
In certain embodiments, some of the parameters used by the model supervisor module 914 may be adjustable. Table 6 identifies some adjustable parameters for this embodiment, along with some exemplary values for the parameters.
Missed Transmission Module
The missed transmission module 916 continuously checks whether the controller receiving data packets (including SG values) for processing. The missed transmission module 916 keeps the system operating in the closed-loop mode for when less than a stated number of data packets are missed (e.g., less than four data packets in a row, a total number of data packets that represent a timespan of less than 15 minutes, or the like). During this time, the system will continue to calculate the insulin dose using the closed-loop control algorithm based on the last valid sensor glucose value or sensor Isig value. For missed data packets that represent a time longer than a lower time threshold and longer than an upper time threshold (e.g., between 15 and 60 minutes), the missed transmission module 916 switches the system to a pre-programmed safe basal rate, which may be defined as half the patient's nighttime basal rate. If the controller starts receiving data packets during the safe basal rate timeframe, the system will switch back to the closed-loop mode. For missed data packets that represent a time longer than the upper time threshold, however, the missed transmission module 916 switches the system to the open-loop mode to deliver a pre-programmed nighttime basal rate, which may be set by a healthcare provider or a caregiver.
The missed transmission module 916 checks for different scenarios pertaining to when and what kind of packet is lost during transmission. Different steps are executed depending on the type of lost transmission. The details of four different scenarios are described below.
Case 1
If the sensor Isig value and the SG value are both received by the controller, then:
Case 2
If the sensor Isig value is not received, but the SG value is received by the controller, then:
Case 3
If the sensor Isig value is received, but the SG value is not received by the controller, then:
Case 4a
If neither the sensor Isig value nor the SG value are received by the controller (i.e., both values are not received), and if:
ZOH Count≦ZOH Count Max
then:
Case 4b
If neither the sensor Isig value nor the SG value are received by the controller (i.e., both values are not received), and if:
ZOH Count>ZOH Count Max
then:
In accordance with certain embodiments, ZOH Count Max has a fixed value of two, and Timeout Count Max has a fixed value of 45, although different values may be used as appropriate to the particular implementation. Moreover, the safe basal rate used by the missed transmission module 916 may be adjustable. In this regard, the safe basal rate may be adjustable within a range of about zero to five Units/Hour.
While at least one exemplary embodiment has been presented in the foregoing detailed description, it should be appreciated that a vast number of variations exist. It should also be appreciated that the exemplary embodiment or embodiments described herein are not intended to limit the scope, applicability, or configuration of the claimed subject matter in any way. Rather, the foregoing detailed description will provide those skilled in the art with a convenient road map for implementing the described embodiment or embodiments. It should be understood that various changes can be made in the function and arrangement of elements without departing from the scope defined by the claims, which includes known equivalents and foreseeable equivalents at the time of filing this patent application.
This application is a continuation-in-part of U.S. patent application Ser. No. 13/870,902, filed on Apr. 25, 2013 (titled “Insulin On Board Compensation For A Closed-Loop Insulin Infusion System”), which claims the benefit of: U.S. provisional patent application No. 61/694,950, filed Aug. 30, 2012 (titled “Closed Loop System”); U.S. provisional patent application No. 61/694,961, filed Aug. 30, 2012 (titled “Closed Loop Mobile System”); and U.S. provisional patent application No. 61/812,874, filed Apr. 17, 2013 (titled “Closed Loop System”). This application is also a continuation-in-part of U.S. patent application Ser. No. 13/870,907, filed on Apr. 25, 2013 (titled “Sensor Model Supervisor For A Closed-Loop Insulin Infusion System”), which claims the benefit of: U.S. provisional patent application No. 61/694,950, filed Aug. 30, 2012 (titled “Closed Loop System”); U.S. provisional patent application No. 61/694,961, filed Aug. 30, 2012 (titled “Closed Loop Mobile System”); and U.S. provisional patent application No. 61/812,874, filed Apr. 17, 2013 (titled “Closed Loop System”). This application is also a continuation-in-part of U.S. patent application Ser. No. 13/870,910, filed on Apr. 25, 2013 (titled “Safeguarding Measures For A Closed-Loop Insulin Infusion System”), which claims the benefit of: U.S. provisional patent application No. 61/694,950, filed Aug. 30, 2012 (titled “Closed Loop System”); U.S. provisional patent application No. 61/694,961, filed Aug. 30, 2012 (titled “Closed Loop Mobile System”); and U.S. provisional patent application No. 61/812,874, filed Apr. 17, 2013 (titled “Closed Loop System”). This application also claims the benefit of: U.S. provisional patent application No. 61/694,950, filed Aug. 30, 2012 (titled “Closed Loop System”); U.S. provisional patent application No. 61/694,961, filed Aug. 30, 2012 (titled “Closed Loop Mobile System”); and U.S. provisional patent application No. 61/812,874, filed Apr. 17, 2013 (titled “Closed Loop System”).
Number | Name | Date | Kind |
---|---|---|---|
3631847 | Hobbs, II | Jan 1972 | A |
4212738 | Henne | Jul 1980 | A |
4270532 | Franetzki et al. | Jun 1981 | A |
4282872 | Franetzki et al. | Aug 1981 | A |
4373527 | Fischell | Feb 1983 | A |
4395259 | Prestele et al. | Jul 1983 | A |
4433072 | Pusineri et al. | Feb 1984 | A |
4443218 | Decant, Jr. et al. | Apr 1984 | A |
4494950 | Fischell | Jan 1985 | A |
4533346 | Cosgrove, Jr. et al. | Aug 1985 | A |
4542532 | McQuilkin | Sep 1985 | A |
4550731 | Batina et al. | Nov 1985 | A |
4559037 | Franetzki et al. | Dec 1985 | A |
4562751 | Nason et al. | Jan 1986 | A |
4573994 | Fischell et al. | Mar 1986 | A |
4671288 | Gough | Jun 1987 | A |
4678408 | Nason et al. | Jul 1987 | A |
4685903 | Cable et al. | Aug 1987 | A |
4731051 | Fischell | Mar 1988 | A |
4731726 | Allen, III | Mar 1988 | A |
4755173 | Konopka et al. | Jul 1988 | A |
4781798 | Gough | Nov 1988 | A |
4803625 | Fu et al. | Feb 1989 | A |
4809697 | Causey, III et al. | Mar 1989 | A |
4826810 | Aoki | May 1989 | A |
4871351 | Feingold | Oct 1989 | A |
4898578 | Rubalcaba, Jr. | Feb 1990 | A |
5003298 | Havel | Mar 1991 | A |
5011468 | Lundquist et al. | Apr 1991 | A |
5019974 | Beckers | May 1991 | A |
5050612 | Matsumura | Sep 1991 | A |
5078683 | Sancoff et al. | Jan 1992 | A |
5080653 | Voss et al. | Jan 1992 | A |
5097122 | Colman et al. | Mar 1992 | A |
5100380 | Epstein et al. | Mar 1992 | A |
5101814 | Palti | Apr 1992 | A |
5108819 | Heller et al. | Apr 1992 | A |
5153827 | Coutre et al. | Oct 1992 | A |
5165407 | Wilson et al. | Nov 1992 | A |
5247434 | Peterson et al. | Sep 1993 | A |
5262035 | Gregg et al. | Nov 1993 | A |
5262305 | Heller et al. | Nov 1993 | A |
5264104 | Gregg et al. | Nov 1993 | A |
5264105 | Gregg et al. | Nov 1993 | A |
5284140 | Allen | Feb 1994 | A |
5299571 | Mastrototaro | Apr 1994 | A |
5307263 | Brown | Apr 1994 | A |
5317506 | Coutre et al. | May 1994 | A |
5320725 | Gregg et al. | Jun 1994 | A |
5322063 | Allen et al. | Jun 1994 | A |
5338157 | Blomquist | Aug 1994 | A |
5339821 | Fujimoto | Aug 1994 | A |
5341291 | Roizen et al. | Aug 1994 | A |
5350411 | Ryan et al. | Sep 1994 | A |
5356786 | Heller et al. | Oct 1994 | A |
5357427 | Langen et al. | Oct 1994 | A |
5368562 | Blomquist et al. | Nov 1994 | A |
5370622 | Livingston et al. | Dec 1994 | A |
5371687 | Holmes, II et al. | Dec 1994 | A |
5376070 | Purvis et al. | Dec 1994 | A |
5390671 | Lord et al. | Feb 1995 | A |
5391250 | Cheney, II et al. | Feb 1995 | A |
5403700 | Heller et al. | Apr 1995 | A |
5411647 | Johnson et al. | May 1995 | A |
5482473 | Lord et al. | Jan 1996 | A |
5485408 | Blomquist | Jan 1996 | A |
5497772 | Schulman et al. | Mar 1996 | A |
5505709 | Funderburk et al. | Apr 1996 | A |
5543326 | Heller et al. | Aug 1996 | A |
5569186 | Lord et al. | Oct 1996 | A |
5569187 | Kaiser | Oct 1996 | A |
5573506 | Vasko | Nov 1996 | A |
5582593 | Hultman | Dec 1996 | A |
5586553 | Halili et al. | Dec 1996 | A |
5593390 | Castellano et al. | Jan 1997 | A |
5593852 | Heller et al. | Jan 1997 | A |
5594638 | Iliff | Jan 1997 | A |
5609060 | Dent | Mar 1997 | A |
5626144 | Tacklind et al. | May 1997 | A |
5630710 | Tune et al. | May 1997 | A |
5643212 | Coutre et al. | Jul 1997 | A |
5660163 | Schulman et al. | Aug 1997 | A |
5660176 | Iliff | Aug 1997 | A |
5665065 | Colman et al. | Sep 1997 | A |
5665222 | Heller et al. | Sep 1997 | A |
5685844 | Marttila | Nov 1997 | A |
5687734 | Dempsey et al. | Nov 1997 | A |
5704366 | Tacklind et al. | Jan 1998 | A |
5750926 | Schulman et al. | May 1998 | A |
5754111 | Garcia | May 1998 | A |
5764159 | Neftel | Jun 1998 | A |
5772635 | Dastur et al. | Jun 1998 | A |
5779665 | Mastrototaro et al. | Jul 1998 | A |
5788669 | Peterson | Aug 1998 | A |
5791344 | Schulman et al. | Aug 1998 | A |
5800420 | Gross et al. | Sep 1998 | A |
5807315 | Van Antwerp et al. | Sep 1998 | A |
5807336 | Russo et al. | Sep 1998 | A |
5814015 | Gargano et al. | Sep 1998 | A |
5822715 | Worthington et al. | Oct 1998 | A |
5832448 | Brown | Nov 1998 | A |
5840020 | Heinonen et al. | Nov 1998 | A |
5861018 | Feierbach et al. | Jan 1999 | A |
5868669 | Iliff | Feb 1999 | A |
5871465 | Vasko | Feb 1999 | A |
5879163 | Brown et al. | Mar 1999 | A |
5885245 | Lynch et al. | Mar 1999 | A |
5897493 | Brown | Apr 1999 | A |
5899855 | Brown | May 1999 | A |
5904708 | Goedeke | May 1999 | A |
5913310 | Brown | Jun 1999 | A |
5917346 | Gord | Jun 1999 | A |
5918603 | Brown | Jul 1999 | A |
5925021 | Castellano et al. | Jul 1999 | A |
5933136 | Brown | Aug 1999 | A |
5935099 | Peterson et al. | Aug 1999 | A |
5940801 | Brown | Aug 1999 | A |
5951521 | Mastrototaro et al. | Sep 1999 | A |
5954643 | VanAntwerp et al. | Sep 1999 | A |
5956501 | Brown | Sep 1999 | A |
5960403 | Brown | Sep 1999 | A |
5965380 | Heller et al. | Oct 1999 | A |
5972199 | Heller et al. | Oct 1999 | A |
5978236 | Faberman et al. | Nov 1999 | A |
5997476 | Brown | Dec 1999 | A |
5999848 | Gord et al. | Dec 1999 | A |
5999849 | Gord et al. | Dec 1999 | A |
6009339 | Bentsen et al. | Dec 1999 | A |
6011984 | Van Antwerp et al. | Jan 2000 | A |
6032119 | Brown et al. | Feb 2000 | A |
6043437 | Schulman et al. | Mar 2000 | A |
6081736 | Colvin et al. | Jun 2000 | A |
6083710 | Heller et al. | Jul 2000 | A |
6088608 | Schulman et al. | Jul 2000 | A |
6101478 | Brown | Aug 2000 | A |
6103033 | Say et al. | Aug 2000 | A |
6119028 | Schulman et al. | Sep 2000 | A |
6120676 | Heller et al. | Sep 2000 | A |
6121009 | Heller et al. | Sep 2000 | A |
6134461 | Say et al. | Oct 2000 | A |
6143164 | Heller et al. | Nov 2000 | A |
6162611 | Heller et al. | Dec 2000 | A |
6175752 | Say et al. | Jan 2001 | B1 |
6183412 | Benkowski et al. | Feb 2001 | B1 |
6246992 | Brown | Jun 2001 | B1 |
6254586 | Mann et al. | Jul 2001 | B1 |
6259937 | Schulman et al. | Jul 2001 | B1 |
6329161 | Heller et al. | Dec 2001 | B1 |
6408330 | DeLaHuerga | Jun 2002 | B1 |
6418332 | Mastrototaro et al. | Jul 2002 | B1 |
6424847 | Mastrototaro et al. | Jul 2002 | B1 |
6461331 | Van Antwerp | Oct 2002 | B1 |
6472122 | Schulman et al. | Oct 2002 | B1 |
6484045 | Holker et al. | Nov 2002 | B1 |
6484046 | Say et al. | Nov 2002 | B1 |
6503381 | Gotoh et al. | Jan 2003 | B1 |
6514718 | Heller et al. | Feb 2003 | B2 |
6544173 | West et al. | Apr 2003 | B2 |
6553263 | Meadows et al. | Apr 2003 | B1 |
6554798 | Mann et al. | Apr 2003 | B1 |
6558320 | Causey, III et al. | May 2003 | B1 |
6558351 | Steil et al. | May 2003 | B1 |
6560741 | Gerety et al. | May 2003 | B1 |
6565509 | Say et al. | May 2003 | B1 |
6579690 | Bonnecaze et al. | Jun 2003 | B1 |
6591125 | Buse et al. | Jul 2003 | B1 |
6592745 | Feldman et al. | Jul 2003 | B1 |
6605200 | Mao et al. | Aug 2003 | B1 |
6605201 | Mao et al. | Aug 2003 | B1 |
6607658 | Heller et al. | Aug 2003 | B1 |
6616819 | Liamos et al. | Sep 2003 | B1 |
6618934 | Feldman et al. | Sep 2003 | B1 |
6623501 | Heller et al. | Sep 2003 | B2 |
6641533 | Causey, III et al. | Nov 2003 | B2 |
6654625 | Say et al. | Nov 2003 | B1 |
6659980 | Moberg et al. | Dec 2003 | B2 |
6671554 | Gibson et al. | Dec 2003 | B2 |
6676816 | Mao et al. | Jan 2004 | B2 |
6689265 | Heller et al. | Feb 2004 | B2 |
6728576 | Thompson et al. | Apr 2004 | B2 |
6733471 | Ericson et al. | May 2004 | B1 |
6734162 | Van Antwerp et al. | May 2004 | B2 |
6746582 | Heller et al. | Jun 2004 | B2 |
6747556 | Medema et al. | Jun 2004 | B2 |
6749740 | Llamos et al. | Jun 2004 | B2 |
6752787 | Causey, III et al. | Jun 2004 | B1 |
6809653 | Mann et al. | Oct 2004 | B1 |
6881551 | Heller et al. | Apr 2005 | B2 |
6892085 | McIvor et al. | May 2005 | B2 |
6893545 | Gotoh et al. | May 2005 | B2 |
6895263 | Shin et al. | May 2005 | B2 |
6916159 | Rush et al. | Jul 2005 | B2 |
6932584 | Gray et al. | Aug 2005 | B2 |
6932894 | Mao et al. | Aug 2005 | B2 |
6942518 | Liamos et al. | Sep 2005 | B2 |
7153263 | Carter et al. | Dec 2006 | B2 |
7153289 | Vasko | Dec 2006 | B2 |
7396330 | Banet et al. | Jul 2008 | B2 |
7500949 | Gottlieb et al. | Mar 2009 | B2 |
20010044731 | Coffman et al. | Nov 2001 | A1 |
20020013518 | West et al. | Jan 2002 | A1 |
20020055857 | Mault et al. | May 2002 | A1 |
20020082665 | Haller et al. | Jun 2002 | A1 |
20020137997 | Mastrototaro et al. | Sep 2002 | A1 |
20020161288 | Shin et al. | Oct 2002 | A1 |
20030060765 | Campbell et al. | Mar 2003 | A1 |
20030078560 | Miller et al. | Apr 2003 | A1 |
20030088166 | Say et al. | May 2003 | A1 |
20030144581 | Conn et al. | Jul 2003 | A1 |
20030152823 | Heller | Aug 2003 | A1 |
20030176183 | Drucker et al. | Sep 2003 | A1 |
20030188427 | Say et al. | Oct 2003 | A1 |
20030199744 | Buse et al. | Oct 2003 | A1 |
20030208113 | Mault et al. | Nov 2003 | A1 |
20030220552 | Reghabi et al. | Nov 2003 | A1 |
20040061232 | Shah et al. | Apr 2004 | A1 |
20040061234 | Shah et al. | Apr 2004 | A1 |
20040064133 | Miller et al. | Apr 2004 | A1 |
20040064156 | Shah et al. | Apr 2004 | A1 |
20040073095 | Causey, III et al. | Apr 2004 | A1 |
20040074785 | Holker et al. | Apr 2004 | A1 |
20040093167 | Braig et al. | May 2004 | A1 |
20040097796 | Berman et al. | May 2004 | A1 |
20040102683 | Khanuja et al. | May 2004 | A1 |
20040111017 | Say et al. | Jun 2004 | A1 |
20040122353 | Shahmirian et al. | Jun 2004 | A1 |
20040152622 | Keith et al. | Aug 2004 | A1 |
20040167465 | Mihai et al. | Aug 2004 | A1 |
20040263354 | Mann et al. | Dec 2004 | A1 |
20050038331 | Silaski et al. | Feb 2005 | A1 |
20050038680 | McMahon et al. | Feb 2005 | A1 |
20050154271 | Rasdal et al. | Jul 2005 | A1 |
20050192557 | Brauker et al. | Sep 2005 | A1 |
20060229694 | Schulman et al. | Oct 2006 | A1 |
20060238333 | Welch et al. | Oct 2006 | A1 |
20060293571 | Bao et al. | Dec 2006 | A1 |
20070088521 | Shmueli et al. | Apr 2007 | A1 |
20070135866 | Baker et al. | Jun 2007 | A1 |
20070173761 | Kanderian, Jr. et al. | Jul 2007 | A1 |
20070276279 | Echauz et al. | Nov 2007 | A1 |
20080154503 | Wittenber et al. | Jun 2008 | A1 |
20080228056 | Blomquist et al. | Sep 2008 | A1 |
20080269723 | Mastrototaro et al. | Oct 2008 | A1 |
20090081951 | Erdmann et al. | Mar 2009 | A1 |
20090082635 | Baldus et al. | Mar 2009 | A1 |
20100017141 | Campbell et al. | Jan 2010 | A1 |
20100057041 | Hayter | Mar 2010 | A1 |
20100121314 | Iobbi | May 2010 | A1 |
20100185175 | Kamen et al. | Jul 2010 | A1 |
20110028279 | Chen | Feb 2011 | A1 |
20110106011 | Cinar et al. | May 2011 | A1 |
20110208155 | Palerm et al. | Aug 2011 | A1 |
20110208156 | Doyle, III et al. | Aug 2011 | A1 |
Number | Date | Country |
---|---|---|
1973768 | Jun 2007 | CN |
101675438 | Mar 2010 | CN |
102576375 | Jul 2012 | CN |
4329229 | Mar 1995 | DE |
0319268 | Nov 1988 | EP |
0806738 | Nov 1997 | EP |
0880936 | Dec 1998 | EP |
1338295 | Aug 2003 | EP |
1631036 | Mar 2006 | EP |
2218831 | Nov 1989 | GB |
2006510467 | Mar 2006 | JP |
2007516814 | Jun 2007 | JP |
2010512945 | Apr 2010 | JP |
2010519623 | Jun 2010 | JP |
2010521222 | Jun 2010 | JP |
2012502693 | Feb 2012 | JP |
WO 9620745 | Jul 1996 | WO |
WO 9636389 | Nov 1996 | WO |
WO 9637246 | Nov 1996 | WO |
WO 9721456 | Jun 1997 | WO |
WO 9820439 | May 1998 | WO |
WO 9824358 | Jun 1998 | WO |
WO 9842407 | Oct 1998 | WO |
WO 9849659 | Nov 1998 | WO |
WO 9859487 | Dec 1998 | WO |
WO 9908183 | Feb 1999 | WO |
WO 9910801 | Mar 1999 | WO |
WO 9918532 | Apr 1999 | WO |
WO 9922236 | May 1999 | WO |
WO9929230 | Jun 1999 | WO |
WO 0010628 | Mar 2000 | WO |
WO0010628 | Mar 2000 | WO |
WO 0019887 | Apr 2000 | WO |
WO0018449 | Apr 2000 | WO |
WO0019887 | Apr 2000 | WO |
WO 0048112 | Aug 2000 | WO |
WO0078210 | Dec 2000 | WO |
WO 02058537 | Aug 2002 | WO |
WO 03001329 | Jan 2003 | WO |
WO 03094090 | Nov 2003 | WO |
WO 2005065538 | Jul 2005 | WO |
WO2011162970 | Dec 2011 | WO |
WO 2012058694 | May 2012 | WO |
Entry |
---|
Garg et al. (Diabetes Technology & Therapeutics, 2017, 19(3), 1-7). |
Steil et al. (J. Clin. Endocrinol. Metab. 2011, 96(5), 1402-1408). |
Garry M. Steil, et al., The Effect of Insulin Feedback on Closed Loop Glucose Control, Journal of Clinical Endocrinology & Metabolism, May 1, 2011, pp. 1402-1408, vol. 96, Issue No. 5. |
Mikhail Loutseiko, et al., Closed-Loop insulin Delivery Utilizing Pole Placement to Compensate for Delays in Subcutaneous Insulin Delivery, Journal of Diabetes Science and Technology, Nov. 1, 2011, pp. 1342-1351, vol. 5, Issue No. 6. |
Roman Hovorka, et al., Nonlinear Model Predictive Control of Glucose Concentration in Subjects With Type 1 Diabetes, Physiological Measurement, 2004, pp. 905-920, vol. 25. |
C. Choleua, et al., Calibration of a Subcutaneous Amperometric Glucose Sensor Part 1 Effect of Measurement Uncertainties on the Determination of Sensor Sensitivity and Background Current, Aug. 1, 2002, pp. 641-646, vol. 17, Issue No. 8. |
Gianni Marchetti, et al., An Improved PID Switching Control Strategy for Type 1 Diabetes, Mar. 1, 2008, pp. 857-865, vol. 35, Isue No. 3. |
Marchetti et al: “A feedforward-feedback glucose control strategy for type 1 diabetes mellitus”, Journal of Process Control, Oxford, GB, vol. 18, No. 2, Oct. 24, 2007 (Oct. 24, 2007), pp. 149-162, XP022455375, ISSN: 0959-1524, DOI: 10.1016/J.JPRONCONT.2007.07.008 *chapter 3: equations (1)-(4); chapter 7: “Feedforward-feedback control”. |
Patek S D et al: “Modular Closed-Loop Control of Diabetes”, IEEE Transactions on Biomedical Engineering, IEEE Service Center, Piscataway, NJ, USA, vol. 59, No. 11, Apr. 3, 2012 (Apr. 3, 2012), pp. 2986-2999, XP011490242, ISSN: 0018-9294, DOI: 10.1109/TBME.2012.2192930 * section III.A and III.B.1-III.B.2 (pp. 2988-2990). |
Greet Van Den Berghe G. et al., Intensive Insulin Therapy in Critically III Patients, The New England Journal of Medicine, Nov. 8, 2001, pp. 1359-1367, vol. 345, No. 19. |
M. Kollind, et al., Insulin clearance during hypoglycemia in patients with insulin-dependent diabetes mellitus, Horm Metab Res, Jul. 1991, pp. 333-335. |
Steven E. Khan, et al., Quantification of the Relationship Between Insulin Sensitivity and Beta-Cell Function in Human Subjects. Evidence for a Hyperbolic Function, Diabetes, Nov. 1993, pp. 1663-1672, vol. 42. |
Arthur C. Guyton, Insulin, Glucagon, and Diabetes Mellitus, Textbook of Medical Physiology, Eighth Edition, 1991, Chapter 78, pp. 855-867, W. B. Saunders Company. |
Antonios E. Panteleon, et al., Evaluation of the Effect of Gain on the Meal Response of an Automated Closed-Loop Insulin Delivery System, Diabetes, Jul. 2006, pp. 1995-2000, vol. 55. |
Garry M. Steil, et al., Feasibility of Automating Insulin Delivery for the Treatment of Type 1 Diabetes, Diabetes, Dec. 2006, pp. 3344-3350, vol. 55. |
Stuart A. Weinzimer, et al., Fully Automated Closed-Loop Insulin Delivery Versus Semiautomated Hybrid Control in Pediatric Patients With Type 1 Diabetes Using an Artificial Pancreas, Diabetes Care, May 2008, pp. 934-939, vol. 31. |
Gary M. Steil, et al., Metabolic Modelling and the Closed-Loop Insulin Delivery Problem, Diabetes Research and Clinical Practice, 2006, pp. S183-S186, vol. 74. |
PCT Search Report (PCT/US02/03299), dated Oct. 31, 2002, Medtronic Minimed, Inc. |
(Animas Corporation, 1999). Animas . . . bringing new life to insulin therapy. |
Bode B W, et al. (1996). Reduction in Severe Hypoglycemia with Long-Term Continuous Subcutaneous Insulin Infusion in Type I Diabetes. Diabetes Care, vol. 19, No. 4, 324-327. |
Boland E (1998). Teens Pumping it Up! Insulin Pump Therapy Guide for Adolescents. 2nd Edition. |
Brackenridge B P (1992). Carbohydrate Gram Counting a Key to Accurate Mealtime Boluses in Intensive Diabetes Therapy. Practical Diabetology, vol. 11, No. 2, pp. 22-28. |
Brackenridge, B P et al. (1995). Counting Carbohydrates How to Zero in on Good Control. MiniMed Technologies Inc. |
Farkas-Hirsch R et al. (1994). Continuous Subcutaneous Insulin Infusion: A Review of the Past and Its Implementation for the Future. Diabetes Spectrum From Research to Practice, vol. 7, No. 2, pp. 80-84, 136-138. |
Hirsch I B et al. (1990). Intensive Insulin Therapy for Treatment of Type I Diabetes. Diabetes Care, vol. 13, No. 12, pp. 1265-1283. |
Kulkarni K et al. (1999). Carbohydrate Counting a Primer for Insulin Pump Users to Zero in on Good Control. MiniMed Inc. |
Marcus A O et al. (1996). Insulin Pump Therapy Acceptable Alternative to Injection Therapy. Postgraduate Medicine, vol. 99, No. 3, pp. 125-142. |
Reed Jet al. (1996). Voice of the Diabetic, vol. 11, No. 3, pp. 1-38. |
Skyler J S (1989). Continuous Subcutaneous Insulin Infusion [CSII]With External Devices: Current Status. Update in Drug Delivery Systems, Chapter 13, pp. 163-183. Futura Publishing Company. |
Skyler J S et al. (1995). The Insulin Pump Therapy Book Insights from the Experts. MiniMed•Technologies. |
Strowig S M (1993). Initiation and Management of Insulin Pump Therapy. The Diabetes Educator, vol. 19, No. 1, pp. 50-60. |
Walsh J, et al. (1989). Pumping Insulin: The Art of Using an Insulin Pump. Published by MiniMed•Technologies. |
(Intensive Diabetes Management, 1995). Insulin Infusion Pump Therapy. pp. 66-78. |
Disetronic My Choice™ D-TRON™ Insulin Pump Reference Manual. (no date). |
Disetronic H-TRON® plus Quick Start Manual. (no date). |
Disetronic My Choice H-TRONplus Insulin Pump Reference Manual. (no date). |
Disetronic H-Tron®plus Reference Manual. (no date). |
(MiniMed, 1996). The MiniMed 506. 7 pages. Retrieved on Sep. 16, 2003 from the World Wide Web: http://web.archive.org/web/19961111054527/www.minimed.com/files/506—pic.htm. |
(MiniMed, 1997). MiniMed 507 Specifications. 2 pages. Retrieved on Sep. 16, 2003 from the World Wide Web: http://web.archive.org/web/19970124234841/www.minimed.com/files/mmn075.htm. |
(MiniMed, 1996). FAQ: The Practical Things . . . pp. 1-4. Retrieved on Sep. 16, 2003 from the World Wide Web: http://web.archive.org/web/19961111054546/www.minimed.com/files/faq—pract.htm. |
(MiniMed, 1997). Wanted: a Few Good Belt Clips! 1 page. Retrieved on Sep. 16, 2003 from the World Wide Web: http://web.archive.org/web/19970124234559/www.minimed.com/files/mmn002.htm. |
(MiniMed Technologies, 1994). MiniMed 506 Insulin Pump User's Guide. |
(MiniMed Technologies, 1994). MiniMed™ Dosage Calculator Initial Meal Bolus Guidelines / MiniMed™ Dosage Calculator Initial Basal Rate Guidelines Percentage Method. 4 pages. |
(MiniMed, 1996). MiniMed™ 507 Insulin Pump User's Guide. |
(MiniMed, 1997). MiniMed™ 507 Insulin Pump User's Guide. |
(MiniMed, 1998). MiniMed 507C Insulin Pump User's Guide. |
(MiniMed International, 1998). MiniMed 507C Insulin Pump for those who appreciate the difference. |
(MiniMed Inc., 1999). MiniMed 508 Flipchart Guide to Insulin Pump Therapy. |
(MiniMed Inc., 1999). Insulin Pump Comparison I Pump Therapy Will Change Your Life. |
(MiniMed, 2000). MiniMed® 508 User's Guide. |
(MiniMed Inc., 2000). MiniMed® Now [I] Can Meal Bolus Calculator / MiniMed® Now [I] Can Correction Bolus Calculator. |
(MiniMed Inc., 2000). Now [I] Can MiniMed Pump Therapy. |
(MiniMed Inc., 2000). Now [I] Can MiniMed Diabetes Management. |
(Medtronic MiniMed, 2002). The 508 Insulin Pump A Tradition of Excellence. |
(Medtronic MiniMed, 2002). Medtronic MiniMed Meal Bolus Calculator and Correction Bolus Calculator. International Version. |
Abel, P., et al., “Experience with an implantable glucose sensor as a prerequiste of an artificial beta cell,” Biomed. Biochim. Acta 43 (1984) 5, pp. 577-584. |
Bindra, Dilbir S., et al., “Design and in Vitro Studies of a Needle-Type Glucose Sensor for a Subcutaneous Monitoring,” American Chemistry Society, 1991, 63, pp. 1692-1696. |
Boguslavsky, Leonid, et al., “Applications of redox polymers in biosensors,” Sold State Ionics 60, 1993, pp. 189-197. |
Geise, Robert J., et al., “Electropolymerized 1,3-diaminobenzene for the construction of a 1,1′-dimethylferrocene mediated glucose biosensor,” Analytica Chimica Acta, 281, 1993, pp. 467-473. |
Gernet, S., et al., “A Planar Glucose Enzyme Electrode,” Sensors and Actuators, 17, 1989, pp. 537-540. |
Gernet, S., et al., “Fabrication and Characterization of a Planar Electromechanical Cell and its Application as a Glucose Sensor,” Sensors and Actuators, 18, 1989, pp. 59-70. |
Gorton, L., et al., “Amperometric Biosensors Based on an Apparent Direct Electron Transfer Between Electrodes and Immobilized Peroxiases,” Analyst, Aug. 1991, vol. 117, pp. 1235-1241. |
Gorton, L., et al., “Amperometric Glucose Sensors Based on Immobilized Glucose-Oxidizing Enymes and Chemically Modified Electrodes,” Analytica Chimica Acta, 249, 1991, pp. 43-54. |
Gough, D. A., et al., “Two-Dimensional Enzyme Electrode Sensor for Glucose,” Analytical Chemistry, vol. 57, No. 5, 1985, pp. 2351-2357. |
Gregg, Brian A., et al., “Cross-Linked Redox Gels Containing Glucose Oxidase for Amperometric Biosensor Applications,” Analytical Chemistry, 62, pp. 258-263. |
Gregg, Brian A., et al., “Redox Polymer Films Containing Enzymes. 1. A Redox-Conducting Epoxy Cement: Synthesis, Characterization, and Electrocatalytic Oxidation of Hydroquinone,” The Journal of Physical Chemistry, vol. 95, No. 15, 1991, pp. 5970-5975. |
Hashiguchi, Yasuhiro, MD, et al., “Development of a Miniaturized Glucose Monitoring System by Combining a Needle-Type Glucose Sensor With Microdialysis Sampling Method,” Diabetes Care, vol. 17, No. 5, May 1994, pp. 387-389. |
Heller, Adam, “Electrical Wiring of Redox Enzymes,” Acc. Chem. Res., vol. 23, No. 5, May 1990, pp. 128-134. |
Jobst, Gerhard, et al., “Thin-Film Microbiosensors for Glucose-Lactate Monitoring,” Analytical Chemistry, vol. 68, No. 18, Sep. 15, 1996, pp. 3173-3179. |
Johnson, K.W., et al., “In vivo evaluation of an electroenzymatic glucose sensor implanted in subcutaneous tissue,” Biosensors & Bioelectronics, 7, 1992, pp. 709-714. |
Jöonsson, G., et al., “An Electromechanical Sensor for Hydrogen Peroxide Based on Peroxidase Adsorbed on a Spectrographic Graphite Electrode,” Electroanalysis, 1989, pp. 465-468. |
Kanapieniene, J. J., et al., “Miniature Glucose Biosensor with Extended Linearity,” Sensors and Actuators, B. 10, 1992, pp. 37-40. |
Kawamori, Ryuzo, et al., “Perfect Normalization of Excessive Glucagon Responses to Intraveneous Arginine in Human Diabetes Mellitus With the Artificial Beta-Cell,” Diabetes vol. 29, Sep. 1980, pp. 762-765. |
Kimura, J., et al., “An Immobilized Enzyme Membrane Fabrication Method,” Biosensors 4, 1988, pp. 41-52. |
Koudelka, M., et al., “In-vivo Behaviour of Hypodermically Implanted Microfabricated Glucose Sensors,” Biosensors & Bioelectronics 6, 1991, pp. 31-36. |
Koudelka, M., et al., “Planar Amperometric Enzyme-Based Glucose Microelectrode,” Sensors & Actuators, 18, 1989, pp. 157-165. |
Mastrototaro, John J., et al., “An electroenzymatic glucose sensor fabricated on a flexible substrate,” Sensors & Actuators, B. 5, 1991, pp. 139-144. |
Mastrototaro, John J., et al., “An Electroenzymatic Sensor Capable of 72 Hour Continuous Monitoring of Subcutaneous Glucose,” 14th Annual International Diabetes Federation Congress, Washington D.C., Jun. 23-28, 1991. |
McKean, Brian D., et al., “A Telemetry-Instrumentation System for Chronically Implanted Glucose and Oxygen Sensors,” IEEE Transactions on Biomedical Engineering, Vo. 35, No. 7, Jul. 1988, pp. 526-532. |
Monroe, D., “Novel Implantable Glucose Sensors,” ACL, Dec. 1989, pp. 8-16. |
Morff, Robert J., et al., “Microfabrication of Reproducible, Economical, Electroenzymatic Glucose Sensors,” Annuaal International Conference of teh IEEE Engineering in Medicine and Biology Society, Vo. 12, No. 2, 1990, pp. 483-484. |
Moussy, Francis, et al., “Performance of Subcutaneously Implanted Needle-Type Glucose Sensors Employing a Novel Trilayer Coating,” Analytical Chemistry, vol. 65, No. 15, Aug. 1, 1993, pp. 2072-2077. |
Nakamoto, S., et al., “A Lift-Off Method for Patterning Enzyme-Immobilized Membranes in Multi-Biosensors,” Sensors and Actuators 13, 1988, pp. 165-172. |
Nishida, Kenro, et al., “Clinical applications of teh wearable artificel endocrine pancreas with the newly designed needle-type glucose sensor,” Elsevier Sciences B.V., 1994, pp. 353-358. |
Nishida, Kenro, et al., “Development of a ferrocene-mediated needle-type glucose sensor covereed with newly designd biocompatible membrane, 2-methacryloyloxyethylphosphorylcholine -co-n-butyl nethacrylate,” Medical Progress Through Technology, vol. 21, 1995, pp. 91-103. |
Poitout, V., et al., “A glucose monitoring system for on line estimation oin man of blood glucose concentration using a miniaturized glucose sensor implanted in the subcutaneous tissue adn a wearable control unit,” Diabetologia, vol. 36, 1991, pp. 658-663. |
Reach, G., “A Method for Evaluating in vivo the Functional Characteristics of Glucose Sensors,” Biosensors 2, 1986, pp. 211-220. |
Shaw, G. W., et al., “In vitro testing of a simply constructed, highly stable glucose sensor suitable for implantation in diabetic patients,” Biosensors & Bioelectronics 6, 1991, pp. 401-406. |
Shichiri, M., “A Needle-Type Glucose Sensor—A Valuable Tool Not Only for a Self-Blood Glucose Monitoring but for a Wearable Artifiical Pancreas,” Life Support Systems Proceedings, XI Annual Meeting ESAO, Alpbach-Innsbruck, Austria, Sep. 1984, pp. 7-9. |
Shichiri, Motoaki, et al., “An artificial endocrine pancreas—problems awaiting solution for long-term clinical applications of a glucose sensor,” Frontiers Med. Biol. Engng., 1991, vol. 3, No. 4, pp. 283-292. |
Shichiri, Motoaki, et al., “Closed-Loop Glycemic Control with a Wearable Artificial Endocrine Pancreas—Variations in Daily Insulin Requirements to Glycemic Response,” Diabetes, vol. 33, Dec. 1984, pp. 1200-1202. |
Shichiri, Motoaki, et al., “Glycaemic Control in a Pacreatectomized Dogs with a Wearable Artificial Endocrine Pancreas,” Diabetologia, vol. 24, 1983, pp. 179-184. |
Shichiri, M., et al., “In Vivo Characteristics of Needle-Type Glucose Sensor—Measurements of Subcutaneous Glucose Concentrations in Human Volunteers,” Hormone and Metabolic Research, Supplement Series vol. No. 20, 1988, pp. 17-20. |
Shichiri, M., et al., “Membrane design for extending the long-life of an implantable glucose sensor,” Diab. Nutr. Metab., vol. 2, No. 4, 1989, pp. 309-313. |
Shichiri, Motoaki, et al., “Normalization of the Paradoxic Secretion of Glucagon in Diabetes Who Were Controlled by the Artificial Beta Cell,” Diabetes, vol. 28, Apr. 1979, pp. 272-275. |
Shichiri, Motoaki, et al., “Telemetry Glucose Monitoring Device with Needle-Type Glucose Sensor: A useful Tool for Blood Glucose Monitoring in Diabetic Individuals,” Diabetes Care, vol. 9, No. 3, May-Jun. 1986, pp. 298-301. |
Shichiri, Motoaki, et al., “Wearable Artificial Endocrine Pancreas with Needle-Type Glucose Sensor,” The Lancet, Nov. 20, 1982, pp. 1129-1131. |
Shichiri, Motoaki, et al., “The Wearable Artificial Endocrine Pancreas with a Needle-Type Glucose Sensor: Perfect Glycemic Control in Ambulatory Diabetes,” Acta Paediatr Jpn 1984, vol. 26, pp. 359-370. |
Shinkai, Seiji, “Molecular Recognitiion of Mono- and Di-saccharides by Phenylboronic Acids in Solvent Extraction and as a Monolayer,” J. Chem. Soc., Chem. Commun., 1991, pp. 1039-1041. |
Shults, Mark C., “A Telemetry-Instrumentation System for Monitoring Multiple Subcutaneously Implanted Glucose Sensors,” IEEE Transactions on Biomedical Engineering, vol. 41, No. 10, Oct. 1994, pp. 937-942. |
Sternberg, Robert, et al., “Study and Development of Multilayer Needle-type Enzyme-based Glucose Microsensors,” Biosensors, vol. 4, 1988, pp. 27-40. |
Tamiya, E., et al., “Micro Glucose Sensors using Electron Mediators Immobilized on a Polypyrrole-Modified Electrode,” Sensors and Actuators, vol. 18, 1989, pp. 297-307. |
Tsukagoshi, Kazuhiko, et al., “Specific Complexation with Mono- and Disaccharides that can be Detected by Circular Dichroism,” J. Org. Chem., vol. 56, 1991, pp. 4089-4091. |
Urban, G., et al., “Miniaturized multi-enzyme biosensors integrated with pH sensors on flexible polymer carriers for in vivo applciations,” Biosensors & Bioelectronics, vol. 7, 1992, pp. 733-739. |
Ubran, G., et al., “Miniaturized thin-film biosensors using covalently immobilized glucose oxidase,” Biosensors & Bioelectronics, vol. 6, 1991, pp. 555-562. |
Velho, G., et al., “In vivo calibration of a subcutaneous glucose sensor for determination of subcutaneous glucose kinetics,” Diab. Nutr. Metab., vol. 3, 1988, pp. 227-233. |
Wang, Joseph, et al., “Needle-Type Dual Microsensor for the Simultaneous Monitoring of Glucose and Insulin,” Analytical Chemistry, vol. 73, 2001, pp. 844-847. |
Yamasaki, Yoshimitsu, et al., “Direct Measurement of Whole Blood Glucose by a Needle-Type Sensor,” Clinics Chimica Acta, vol. 93, 1989, pp. 93-98. |
Yokoyama, K., “Integrated Biosensor for Glucose and Galactose,” Analytica Chimica Acta, vol. 218, 1989, pp. 137-142. |
Number | Date | Country | |
---|---|---|---|
20140066889 A1 | Mar 2014 | US |
Number | Date | Country | |
---|---|---|---|
61694950 | Aug 2012 | US | |
61694961 | Aug 2012 | US | |
61812874 | Apr 2013 | US |
Number | Date | Country | |
---|---|---|---|
Parent | 13870902 | Apr 2013 | US |
Child | 13966120 | US | |
Parent | 13870907 | Apr 2013 | US |
Child | 13870902 | US | |
Parent | 13870910 | Apr 2013 | US |
Child | 13870907 | US |